Gima CLASS 6 EMULATING STEAM STRIPS User guide

INSTRUCTION FOR USE
PRODUCT NAME: TYPE 5 EO INTEGRATING INDICATOR (IE3501) PRODUCT DESCRIPTION: PMS Steritest Type 5 EO Integrating Indicator Strips are designed for use in
Ethylene Oxide sterilizers with operating conditions as 60 % (±10 %) Humidity, 37 °C (±0,5 °C) and 90 minutes or 54 °C (±0,5 °C) 30 minutes at 600 ±30 mg/L EO concentration. After exposure to EO sterilization, the chemical indicator will provide an accurate visual confirmation by color
changing, indicating that sterilization conditions were met. PMS Steritest Type-5 EO Integrating indicators are classified as Type-5 according to EN ISO 11140-1 Standard. During exposure to Ethylene Oxide sterilization at specified condition, the indicator color will change from
White to Mustard Yellow.
INSTRUCTION FOR USE
Place PMS Steritest Chemical Indicator Strip in each pack, peel pouch or tray to be sterilized by
Ethylene Oxide. The strip should be placed at a location in the package that is considered to be most difficult area for Ethylene Oxide to reach.
Process the package as usual in a EO sterilizer..  At the completion of the sterilization cycle, verify that the indicator has changed color and turned to
mustard yellow.
If the color change is not visible or non-uniform, the sterilization conditions may be inadequate and
the items must be reprocessed with new/unprocessed indicator strips. .
Products stored at recommended conditions have a shelf life of three (3) years from the date of
manufacture.
Manufacture, expiry date and Lot number are printed on the product box.  Do not re-use the used Chemical Indicator Strips.
STORAGE CONDITIONS
Store at maximum 35 °C and maximum 70 %RH humidity.
Chemical Indicator Strips must be kept away from direct heat, sun light and excess moisture.
PMS Tıp Teknolojileri San. ve Tic. A.Ş
Mersin Tarsus Organize Sanayi Bölgesi12. Cad. No:2
Huzurkent/Mersin/TURKEY
Phone: +90 324 238 7042 | Fax: +90 324 238 6549
pms@pmsmedikal.com | www.pmsmedikal.com