Gima CHISON ECO 5 VET ULTRASOUND User guide

Page 1
Digital Color Doppler Ultrasound System
Model
ECO 5VET
V1.3
Jan 19, 2015
OPERATION MANUAL
Page 2
ECO 5VET Digital Color Doppler Ultrasound System
Regulatory Requirement
This product conforms to the essential requirements of the Medical Device Directive 93/42/EEC. Accessories without the CE mark are not guaranteed to meet the Essential Requirements of the Medical Device Directive.
This manual is a reference for the ECO 5VET. Please verify that you are using the latest revision of this document. If you need the latest revision, contact your distributor.
NOTE:
Important
1.No part of this manual may be reduced, modified, copied or reprinted, in whole or in part, without written permission from CHISON.
2.The contents of this manual are subject to change without prior notice and without our legal obligation.
3.Before operating the system, please read and understand this manual. After reading, keep this manual in an easily accessible place. If you have any question or doubt, please contact CHISON's authorized service engineer.
4.CHISON’s Warranty only cover material and parts costs for repair, but do not cover any labor cost or onsite service cost at end user's side.
1
Page 3
ECO 5VET Digital Color Doppler Ultrasound System
NOTE:
Important information
1.It is the customer’s responsibility to maintain and manage the system after delivery.
2.The warranty does not cover the following items, even during the warranty period: a)Damage or loss due to misuse or abuse with system and probes, for example, drop the probe, the liquid or the metal part fall into the system. b)Damage or loss caused by Acts of God such as fires, earthquakes, floods, lightning, etc. c)Damage or loss caused by failure to meet the specified conditions for this system, such as inadequate power supply, improper installation or environmental conditions. d)Damage or loss caused by non approved transportation by CHISON. e)Damage or loss due to use the system outside the region where the system was originally sold. f)Damage or loss involving the system purchased from a source other than CHISON or its authorized agents.
3.Do not make changes or modifications to the software or hardware of this system and probes.
4.During operate the system, if user has any doubt, difficulty or any unclear, please contact CHISON's authorized service engineer immediately. Please describe the situation clearly to solve the question in time. Before solve the question, please don’t operate the system.
5.This system shall not be used by persons other than fully qualified and certified medical personnel.
6.The System modified or repaired by people other than CHISONs qualified service engineers, CHISON shall not be liable for the system.
7.The purpose of this system is to provide physicians with data for clinical diagnosis. It is the physician’s responsibility for diagnostic procedures. CHISON shall not be liable for the results of diagnostic procedures
8.This manual contains warnings regarding foreseeable potential dangers, but user shall always be alert to dangers other than those indicated as well. CHISON shall not be liable for damage or loss that results from negligence or from ignoring the precautions and operating instructions described in this operation manual.
9.Due to negligence not following operation manual, CHISON shall not be liable for the results.
10.Each time before and after ultrasound examination, please check the probe surface, probe cable and sheath whether they are abnormal, such as cracking, peeling and deformation. Also check whether the lens is strongly fixed. Abnormal probes may cause electric shock and injure the patient. Once any abnormal, user must stop using and contact CHISON's authorized service engineer.
11.If the probe is dropped or scratched by hard part, please stop using the probe immediately. And contact CHISON's authorized service engineer to make sure the safety and effectiveness is in good condition before use.
2
Page 4
ECO 5VET Digital Color Doppler Ultrasound System
Caution: The users should read the operation manual carefully before operating the devices.
Turning on the device means the users have read the operation manual and accept the listed cautions, warnings, and notes in the manuals. If the users disagree and cannot accept the cautions, the users can ask for returning the device.
12.If there is any liquid or metal to enter to the system, please power off the system and stop using it immediately. Please first contact CHISONs authorized service engineer to make sure its safe before restart using it.
13.Please don't use solvents (such as paint thinner, benzine, or alcohol) or abrasive cleansers for cleaning the system (including monitor and probes, etc). It may corrode the system and probes.
14.While the system or probe is over life time, please refer to operation manual section 9.4
15.Important data must be backed up on external memory media. CHISON shall not be liable for loss of data stored in the memory of this system caused by operator error or accidents.
16.Please put this operation manual with the system to ensure operator and manager can reach it at any time.
17.LCD display screen may have some dark or light dots, it is normal for the LCD. It
does not mean that LCD screen is defective.
3
Page 5
ECO 5VET Digital Color Doppler Ultrasound System
Content
Chapter 1 Introduction .............................................................................................................................. 1
1.1 System Overview ...................................................................................................................... 1
1.2 Contact Information ................................................................................................................ 1
Chapter 2 System Safety........................................................................................................................... 2
2.1 Safety Overview ........................................................................................................................ 2
2.2 Electrical Safety ........................................................................................................................ 3
2.3 Labels ........................................................................................................................................ 5
2.3.1 Warning Symbols ................................................................ ................................ ............ 5
2.4 Patient Environmental Devices ............................................................................................... 6
2.5 Biological Safety ....................................................................................................................... 8
2.6 Scanning Patients and Education ........................................................................................... 9
2.6.1 Safe Scanning Guidelines ............................................................................................... 9
2.6.2 Understanding the MI/TI Display ................................................................................. 10
Chapter 3 System Introduction ............................................................................................................... 15
3.1 Consol Overview ..................................................................................................................... 15
3.2 Physical Specification ............................................................................................................. 15
3.3 System View in Different Views ............................................................................................ 15
3.4 Function Introduction ............................................................................................................ 16
3.4.1 Image Modes ................................................................................................................. 17
3.4.2 Accessories.................................................................................................................... 17
3.5 Installation Procedures .......................................................................................................... 18
3.5.1 Environment Condition ................................................................................................. 18
3.5.2 Powering the System..................................................................................................... 20
3.5.3 Probe Installment .......................................................................................................... 21
3.5.4 Accessories Installment ................................................................................................. 21
Chapter 4 Control Panel ......................................................................................................................... 23
4.1 Alphanumeric Keyboard ....................................................................................................... 23
4.2 Function Keys/Knob .............................................................................................................. 23
1
Page 6
ECO 5VET Digital Color Doppler Ultrasound System
4.2.1 SWITCH ....................................................................................................................... 23
4.2.2 PATIENT....................................................................................................................... 23
4.2.3 PROBE ......................................................................................................................... 23
4.2.4 SETUP ................................................................ .......................................................... 23
4.2.5 END .............................................................................................................................. 24
4.2.6 BODYMARK ............................................................................................................... 24
4.2.7 COMMENT .................................................................................................................. 24
4.2.8 ARROWS...................................................................................................................... 24
4.2.9 DEL............................................................................................................................... 24
4.2.10 Print ............................................................................................................................. 24
4.2.11 Archive ........................................................................................................................ 25
4.2.12 Report ......................................................................................................................... 25
4.2.13 Parameter Control Key ............................................................................................... 25
4.3 Central Control ...................................................................................................................... 25
4.3.1 ENTER ......................................................................................................................... 26
4.3.2 UPDATE ....................................................................................................................... 26
4.3.3 Exit ................................................................................................................................ 26
4.3.4 Cursor ........................................................................................................................... 26
4.3.5 Change ................................ .......................................................................................... 26
4.3.6 Dist ................................................................................................................................ 26
4.3.7 Trace ............................................................................................................................. 26
4.3.8 Calc ............................................................................................................................... 26
4.3.9 MENU ........................................................................................................................... 26
4.3.10 Trackball ..................................................................................................................... 27
4.4 The key of Image Mode ......................................................................................................... 27
4.4.1 B .................................................................................................................................... 27
4.4.2 C .................................................................................................................................... 27
4.4.3 D ................................................................................................................................... 27
4.4.4 M ................................................................................................................................... 27
4.5 Control of Image .................................................................................................................... 27
2
Page 7
ECO 5VET Digital Color Doppler Ultrasound System
4.5.1 THI ................................................................................................................................ 27
4.5.2 AIO ............................................................................................................................... 28
4.5.3 CINE ............................................................................................................................. 28
4.5.4 SAVE ............................................................................................................................ 28
4.5.5 Left and Right Invert ..................................................................................................... 28
4.5.6 Up and Down Invert...................................................................................................... 28
4.5.7 STC ............................................................................................................................... 28
4.5.8 GAIN ............................................................................................................................ 29
4.5.9 ANGLE/ZOOM Knob .................................................................................................. 29
4.5.10 DEPTH/FOCUS Knob ................................................................................................ 29
4.5.11 Information Area Indicating Machine Status .............................................................. 29
4.5.12 Indicator Light ............................................................................................................ 30
Chapter 5 Operation and Exam Mode .................................................................................................... 31
5.1 Preparing the System for Use ................................................................................................ 31
5.1.1 The Device Inspection .................................................................................................. 31
5.1.2 Power On ...................................................................................................................... 31
5.2 Choose Exam Mode ................................................................................................................ 31
5.2.1 The Probe Identification................................................................................................ 31
5.2.2 Mode Selection ............................................................................................................. 31
5.3 Patient Data Entry ................................................................................................................. 32
5.4 Image Interface Display ......................................................................................................... 33
5.5 Display Mode .......................................................................................................................... 33
5.5.1 B Mode ......................................................................................................................... 33
5.5.2 B/B Mode ...................................................................................................................... 33
5.5.3 4B Mode ....................................................................................................................... 33
5.5.4 B/M Mode ..................................................................................................................... 34
5.5.5 M Mode ........................................................................................................................ 34
5.5.6 CFM Mode .................................................................................................................... 34
5.5.7 PW Mode ...................................................................................................................... 35
5.6 B Image Adjustment .............................................................................................................. 36
3
Page 8
ECO 5VET Digital Color Doppler Ultrasound System
5.6.1 Frequency...................................................................................................................... 36
5.6.2 Dynamic ........................................................................................................................ 36
5.6.3 i-Image .......................................................................................................................... 36
5.6.4 Compound ..................................................................................................................... 36
5.6.5 SRA............................................................................................................................... 36
5.6.6 M Speed Adjustment ..................................................................................................... 36
5.6.7 Gain............................................................................................................................... 37
5.6.8 STC ............................................................................................................................... 37
5.6.9 Depth ............................................................................................................................. 37
5.6.10 Focus Pos .................................................................................................................... 37
5.6.11 Angle ........................................................................................................................... 37
5.6.12 ANGLE/ZOOM .......................................................................................................... 37
5.6.13 INVERT ...................................................................................................................... 37
5.6.14 THI .............................................................................................................................. 38
5.6.15 AIO ............................................................................................................................. 38
5.7 B Image Menu Adjustment ................................................................................................... 38
5.7.1 Scan Width .................................................................................................................... 38
5.7.2 Focus Num .................................................................................................................... 38
5.7.3 Persistence .................................................................................................................... 38
5.7.4 Line Density .................................................................................................................. 39
5.7.5 Smooth .......................................................................................................................... 39
5.7.6 Edge Enhance ............................................................................................................... 39
5.7.7 Acoustic Power ............................................................................................................. 39
5.7.8 Utility ............................................................................................................................ 39
5.7.9 Advance ........................................................................................................................ 40
5.8 CFM Image Adjustment ........................................................................................................ 40
5.8.1 Color Map ..................................................................................................................... 41
5.8.2 Color Invert ................................................................................................................... 41
5.8.3 Contrast ......................................................................................................................... 41
5.8.4 Steering Angle ............................................................................................................... 41
4
Page 9
ECO 5VET Digital Color Doppler Ultrasound System
5.8.5 Wall Thre....................................................................................................................... 41
5.8.6 Blood Effection ............................................................................................................. 41
5.8.7 Flow .............................................................................................................................. 41
5.8.8 Freq ............................................................................................................................... 42
5.8.9 Wall Filter ..................................................................................................................... 42
5.8.10 PRF ............................................................................................................................. 42
5.8.11 Persistence ................................................................................................................... 42
5.8.12 Baseline ....................................................................................................................... 42
5.9 PW Image Adjustment ........................................................................................................... 42
5.9.1 D 2D Map ..................................................................................................................... 42
5.9.2 Enhance ......................................................................................................................... 43
5.9.3 D Gamma ...................................................................................................................... 43
5.9.4 Dynamic ........................................................................................................................ 43
5.9.5 Pixel Ratio..................................................................................................................... 43
5.9.6 Color Map ..................................................................................................................... 43
5.9.7 Audio ............................................................................................................................. 43
5.9.8 Acoustic power ............................................................................................................. 43
5.9.9 Invert ............................................................................................................................. 43
5.9.10 Wall Filter ................................................................................................................... 43
5.9.11 Speed ........................................................................................................................... 44
5.9.12 PRF ............................................................................................................................. 44
5.9.13 Baseline ....................................................................................................................... 44
5.10 Full Screen Show (need to activate this function in setup) ............................................... 44
5.11 Edit Comment ....................................................................................................................... 44
5.11.1 Overview ..................................................................................................................... 44
5.11.2 Input Characters .......................................................................................................... 45
5.11.3 Input Comment Library Characters ............................................................................. 45
5.11.4 Edit Quick Comments ................................................................................................. 45
5.11.5 Input Quick Comments ............................................................................................... 45
5.11.6 Move Comments ......................................................................................................... 45
5
Page 10
ECO 5VET Digital Color Doppler Ultrasound System
5.11.7 Edit Comments ............................................................................................................ 45
5.11.8 Delete Comments ........................................................................................................ 46
5.11.9 Set the Position of Default Comment .......................................................................... 46
5.12 Set Body Mark ...................................................................................................................... 46
5.12.1 General Description .................................................................................................... 46
5.12.2 Body Mark Operation ................................................................................................. 47
5.13 Set the Direction of Arrow ................................................................................................... 47
5.14 Image and Cine Disposition ................................................................................................. 47
5.14.1 The Principle of Cine Storage ..................................................................................... 47
5.14.2 Manual Loop ................................................................ ................................ ............... 48
5.14.3 Automatic Loop .......................................................................................................... 48
5.14.4 Save and Recall Image ................................................................................................ 48
5.14.5 Save and Recall Cine .................................................................................................. 48
5.14.6 Delete images .............................................................................................................. 48
5.14.7 Send images ................................................................................................................ 48
5.15 Image Browse ....................................................................................................................... 49
5.16 Archive Management ........................................................................................................... 49
5.17 Report .................................................................................................................................... 50
5.18 DICOM ................................................................................................................................. 51
5.18.1 DICOM Worklist ......................................................................................................... 51
5.18.2 DICOM Storage .......................................................................................................... 52
5.18.3 DICOM Print .............................................................................................................. 53
5.18.4 DICOM SR ................................................................................................................. 53
Chapter 6 Measurement and Calculation ................................................................................................ 54
6.1 Keyboard for Measurement .................................................................................................. 54
6.1.1 Trackball ....................................................................................................................... 54
6.1.2 [ENTER] ....................................................................................................................... 54
6.1.3 [UPDATE] .................................................................................................................... 54
6.1.4 [DEL] ............................................................................................................................ 54
6.1.5 [Change] ....................................................................................................................... 55
6
Page 11
ECO 5VET Digital Color Doppler Ultrasound System
6.1.6 [Exit] ............................................................................................................................. 55
6.1.7 Parameters control key .................................................................................................. 55
6.2 B Mode general Measurement methods ............................................................................... 55
6.2.1 Meas. Distance .............................................................................................................. 55
6.2.2 Ellipse ........................................................................................................................... 55
6.2.3 Trace ............................................................................................................................. 56
6.2.4 Histogram...................................................................................................................... 56
6.2.5 Cross-section Diagram .................................................................................................. 57
6.3 B Fast Measurement .............................................................................................................. 57
6.4 Genearl Measurement ............................................................................................................ 58
6.5 Canine Measurement ............................................................................................................. 59
6.6 Feline Measurement ............................................................................................................... 59
6.7 Equine Measurement ............................................................................................................. 60
6.8 Bovine Measurement.............................................................................................................. 61
6.9 Ovine Measurement ............................................................................................................... 61
6.10 Genearl Measurement and Calculation in M, B/M mode ................................................. 62
6.10.1 Distance ...................................................................................................................... 62
6.10.2 Time ............................................................................................................................ 62
6.10.3 Heart rate..................................................................................................................... 62
6.10.4 Velocity ....................................................................................................................... 63
6.11 M General Measurement ..................................................................................................... 63
6.12 M Canine Measurement ...................................................................................................... 63
6.13 M Feline Measurement ........................................................................................................ 63
6.14 M Equine Measurement ...................................................................................................... 63
6.15 M Bovine Measurement ....................................................................................................... 64
6.16 M Ovine Measurement ........................................................................................................ 64
6.17 PW Fast Measurement ......................................................................................................... 64
6.18 PW Genearl Measurement .................................................................................................. 65
6.19 PW Canine Measurement .................................................................................................... 67
6.20 PW Feline Measurement ..................................................................................................... 67
7
Page 12
ECO 5VET Digital Color Doppler Ultrasound System
6.21 PW Equine Measurement .................................................................................................... 67
6.22 PW Bovine Measurement .................................................................................................... 67
6.23 PW Ovine Measurement ................................................................................................ ...... 68
Chapter 7 Preset ...................................................................................................................................... 69
7.1 General setting ........................................................................................................................ 69
7.2 Measurement .......................................................................................................................... 70
7.2.1 General measurement setting ........................................................................................ 70
7.2.2 Measurement formula setting........................................................................................ 71
7.3 Annotation .............................................................................................................................. 77
7.3.1 Annotation Library ........................................................................................................ 77
7.3.2 Edit Annotation ............................................................................................................. 78
7.3.3 Comment and Arrow Option ......................................................................................... 78
7.4 Body marks ............................................................................................................................. 79
7.4.1 Body Marks Library ...................................................................................................... 79
7.4.2 Body mark edition......................................................................................................... 79
7.4.3 Body mark Option ........................................................................................................ 80
7.5 Exam Mode ............................................................................................................................. 80
7.5.1 Exam Mode Edit ........................................................................................................... 80
7.5.2 Exam Mode Selection ................................................................................................... 80
7.5.3 Exam mode selection .................................................................................................... 81
7.6 Keyboard ................................................................................................................................. 82
7.7 DICOM ................................................................................................................................... 83
7.7.1 Add/Edit DICOM Function ........................................................................................... 84
7.8 NET Work .............................................................................................................................. 85
7.9 System ..................................................................................................................................... 85
7.9.1 System information ....................................................................................................... 85
7.9.2 Upgrade ......................................................................................................................... 85
7.9.3 Function Setting ............................................................................................................ 85
7.9.4 Installment setting ......................................................................................................... 86
7.9.5 Video VGA ................................................................................................................... 86
8
Page 13
ECO 5VET Digital Color Doppler Ultrasound System
7.9.6 Image function .............................................................................................................. 86
7.9.7 System Maintenance ..................................................................................................... 86
7.9.8 USB Video Printer Option ............................................................................................ 86
Chapter 8 System Maintenance .............................................................................................................. 87
8.1 Machine Clean ........................................................................................................................ 87
8.2 Probe Maintenance ................................................................................................................ 87
8.3 Safety Check ........................................................................................................................... 88
8.4 Malfunction Check ................................................................................................................. 89
Chapter 9 Probes..................................................................................................................................... 90
9.1 General Description ............................................................................................................... 90
9.2 Care and Maintenance ........................................................................................................... 90
9.2.1 Inspecting Probes .......................................................................................................... 90
9.2.2 Cleaning and Disinfecting ............................................................................................. 91
9.3 Probe Operation Instructions................................................................................................ 98
9.3.1 Scanning the Patient ...................................................................................................... 98
9.3.2 Operating Transvaginal probe ....................................................................................... 98
9.3.3 Cleaning and Disinfecting TV and TR Probes .............................................................. 99
9.4 Service Responsibility .......................................................................................................... 100
Appendix A: Acoustic Output Report Table ........................................................................... 102
Appendix B: Guidance and Manufacturer’s Declaration ....................................................... 132
Appendix C: Measurement Results Summary ........................................................................ 136
Appendix D: Display Accuracy and Acoustic Measurement Uncertainties .......................... 138
Appendix E: Transducer Maximum Surface Temperature ................................................... 139
9
Page 14
ECO 5VET Digital Color Doppler Ultrasound System
Chapter 1 Introduction
This manual contains necessary information for safe system operation. Read and understand all instructions in this manual before operating the system. Always keep this manual with the equipment, and periodically review the procedures for operation and safety precautions.
1.1 System Overview
Indications for Use
The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of ABD, OB, GYN, Vessel, Small Parts, MSK/PT, Nerve, Big animal, Reproduction, Small animal cardiac.
Contraindication
The system is NOT intended for ophthalmic use or any use that causes the acoustic beam to pass through the eye.
1.2 Contact Information
For additional information or assistance, please contact your local distributor or the appropriate support resource shown below:
CHISON website: www.chison.com Service Support CHISON Medical Imaging Co., Ltd.
Tel: 0086-0510-85311707 Fax: 0086-0510-85310726 E-mail: service@chison.com.cn
Placing an Order CHISON Medical Imaging Co., Ltd.
Tel: 0086-0510-8531-0593/0937 Fax: 0086-0510-85310726 Email: export@chison.com.cn
Manufacturer CHISON Medical Imaging Co., Ltd.
No.228, ChangJiang East Road, Block 51 and 53 Phase 5 Industrial Park, ShuoFang, New District, Wuxi, Jiangsu, China, 214142
1
Page 15
ECO 5VET Digital Color Doppler Ultrasound System
Chapter 2 System Safety
2.1 Safety Overview
This section discusses measures to ensure the safety of both the operator and patient. To ensure the safety of both operator and patient, please read the relevant details in this chapter carefully before operating this system.
Disregarding the warnings or violation of relevant rules may result in personal injury or even loss of life for operator or patient.
Users should observe the following precautions:
This system complies with Type BF general equipment, and the IEC standard. Do not modify this system in any way. Necessary modifications must be made only by the manufacturer or
its designated agents.
This system has been fully adjusted at the factory. Do not adjust any fixed adjustable parts. In the event of a malfunction, turn off the system immediately and inform the manufacturer or its designated
agents. The power cable of the system should only be connected to a grounded power socket. Do not remove the ground cable for any reason. Only connect this system, either electronically or mechanically, with devices that comply with the EN60601-1 standard. Recheck the leakage current and other safety performance indices of the entire system to avoid potential system damage caused by leakage from a current superposition. The system does not incorporate any specialized protective measures in the event it is configured with high-frequency operation devices. The operator should use caution in these types of applications. The system should be installed only by personnel authorized by the manufacturer. Do not attempt to install the system by yourself.
Only an authorized service engineer may perform maintenance. Only a qualified operator, or someone under qualified supervision, should use the system. Do not use this system in the presence of flammable substances, otherwise an explosion may occur. Do not continuously scan the same part of a patient or expose the patient to prolonged scanning; otherwise it
may harm the patient. When using the system for ultrasound testing, use only qualified ultrasound gel that complies with system standards. Do not unplug probe when the system is in active operation. Always go to EXAM screen when need to remove the probe. To prevent from arm or neck injury, the operator should not stay at the same position for too long during patient scanning without taking break. Do not put liquid on top of the main unit.
NOTE
*The system has built-in screen saver to avoid the tic mark on the display. It is not recommended to constantly turn on and off the unit. *To dispose of this product properly, please call your local service department.
2
Page 16
ECO 5VET Digital Color Doppler Ultrasound System
2.2 Electrical Safety
Type of protection against electric shock
Class I Equipment
CLASS I EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but includes a protective earth ground. This additional safety precaution prevents exposed metal parts from becoming LIVE in the event of an insulation failure.
NOTE: The mains supply shall be cut off after disconnecting the power line and the net power.
Degree of protection against electric shock
Type BF Applied part (for Probes marked with BF symbol) TYPE BF APPLIED PART providing a specified degree of protection against electric shock, with particular regard to allowable LEAKAGE CURRENT
Level of protection against harmful ingress of water
Parts of probe likely to come into contact with operator or patient meet the requirements of drip-proof equipment (IPX1) Parts of probe intended to be immersed in normal use meet the requirements of watertight equipment (IPX7) The IP Classification of System is Ordinary Equipment (IPX0)
Safety level when used in the presence of FLAMMABLE ANAESTHETIC MIXED WITH AIR (or WITH OXYGEN or WITH NITROUS OXIDE):
The Equipment is not suitable for use in the environment with FLAMMABLE ANAESTHETIC MIXED WITH AIR (or WITH OXYGEN or WITH NITROUS OXIDE)
Mode of operation
Continuous Operation For maximum safety, always follow these guidelines: Proper grounding of the system is critical to avoid electrical shock. For protection, ground the chassis with a three-wire cable and plug, and plug the system into a hospital-grade, three-hole outlet.
Do not remove or circumvent the grounding wire. Do not remove the protective covers on the system. These covers protect users from hazardous voltages.
Cabinet panels must remain in place while the system is in use. A qualified electronic technician must make all internal replacements.
Do not operate this system in the presence of flammable gases or anesthetics. All peripheral devices (unless certified as medical grade) that are connected to the system must be powered
through the electrical outlet through an optional isolation transformer.
3
Page 17
ECO 5VET Digital Color Doppler Ultrasound System
Notice upon Installation of Product
Separation distance and effect from fixed radio communications equipment: field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast transmitter cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ultrasound system is used exceeds the applicable RF compliance level as stated in the immunity declaration, the ultrasound system should be observed to verify normal operation. If abnormal operation is observed, additional measures may be necessary, such as re-orienting or relocating the ultrasound system or using an RF shielded examination room may be necessary. Use either power supply cords provided by or designated by CHISON. Products equipped with a power source plug should be plugged into the fixed power socket which has the protective grounding conductor. Never use any adaptor or converter to connect with a power source plug (e.g. three-prong-to-two-prong converter).
Locate the equipment as far away as possible from other electronic equipment. Be sure to use only the cables provided by or designated by CHISON. Connect these cables following the
installation procedures (e.g. wire power cables separately from signal cables). Lay out the main equipment and other peripherals following the installation procedures described in this manual.
Notice against User Modification
The user should never modify this product. User modifications may cause degradation in Electrical Safety. Modification of the product includes changes in:
Cables (length, material, wiring, etc.) System configuration/components
User modifications may cause degradation in EMC performance. Modification of the product includes changes in:
Cables (length, material, wiring, etc.) System installation/layout System configuration/components Securing system parts (cover open/close, cover screwing)
4
Page 18
2.3 Labels
Caution, consult accompanying documents. This symbol advises the reader to consult the accompanying documents for important safety related information such as warnings and pre-cautions that cannot be presented on the device itself.
Dangerous electric voltage. Unplug the main plug before opening the system!
Do not use the following devices near this equipment: cellular phone, radio receiver, and mobile radio transmitter, radio controlled toy, etc. Use of these devices near this equipment could cause this equipment to perform outside the published specifications. Keep power to these devices turned off when near this equipment.
Be careful of static.
WASTE OF ELECTRICAL AND
ELECTRONIC EQUIPMENT (WEEE): This symbol is used for Environment Protection, it indicates that the waste of electrical and electronic equipment must not be disposed as unsorted waste and must be collected separately. Please contact your local
Type-BF applied part
ECO 5VET Digital Color Doppler Ultrasound System
Real panel label
2.3.1 Warning Symbols
5
Page 19
ECO 5VET Digital Color Doppler Ultrasound System
Authority or distributor of the manufacturer for information concerning the decommissioning of your equipment.
This symbol is followed by the serial
number of the device.
MANUFACTURER: This symbol is
accompanied by the name and the address of the manufacturer.
Power On/off.
CAUTION: This Power Switch cannot isolate Mains Supply completely.
This symbol signifies that the user manual
must be read.
The “Alternating current” symbol indicates
that the equipment is suitable for alternating current only.
This symbol is followed by the manufacturing date of the device in the form YYYY-MM.
The CE mark of Conformity indicates this
equipment conforms with the Council Directive 93/42/EEC CORRECT: The correct connection of the battery connector WRONG: The wrong connection of the battery connector
2.4 Patient Environmental Devices
Left side:
1 LAN port 1 VGA port: External monitor 2 USB ports 1 Footswitch port 1 Power in port
Rear panel:
2 Probe ports 1 USB port 1 Video out port 1 Remote port
Acceptable Devices
The Patient Environmental devices shown above are specified to be suitable for use within the PATIENT ENVIRONMENT.
6
Page 20
ECO 5VET Digital Color Doppler Ultrasound System
CAUTION:
DO NOT connect any probes or accessories without approval by CHISON within the PATIENT ENVIRONMENT. DO NOT touch patient and devices without EN 60601-1 approval to avoid the leakage current risk within the PATIENT ENVIRONMENT.
Unapproved Devices
CAUTION:
DO NOT use unapproved devices. If devices are connected without the approval of CHISON, the warranty will be INVALID. The system can’t be used with HF surgical equipment; otherwise the burns to patient may occur.
Any device connected to this system must conform to one or more of the requirements listed below:
IEC standard or equivalent standards appropriate to devices. The devices shall be connected to PROTECTIVE EARTH (GROUND).
CAUTION:
Unsafe operation or malfunction may result. Use only the accessories, options and supplies approved or recommended in these instructions for use.
Peripheral used in the patient environment
The system has been verified for overall safety, compatibility and compliance with the following on-board image recording devices: B/W video printer: SONY UP-D711MD The system may also be used safely while connected to devices other than those recommended above if the devices and their specifications, installation, and interconnection with the system conform to the requirements of EN60601-1. Adapter is considered as a part of ME equipment The connection of equipment or transmission networks other than as specified in the user instructions can result in an electric shock hazard or equipment malfunction. Substitute or alternate equipment and connections require verification of compatibility and conformity to EN 60601-1 by the installer. Equipment modifications and possible resulting malfunctions and electromagnetic interference are the responsibility of the owner. General precautions for installing an alternate off-board, remote device or a network would include:
The added device(s) must have appropriate safety standard conformance and CE Marking. There must be adequate mechanical mounting of the device and stability of the combination. Risk and leakage current of the combination must comply with EN 60601-1. Electromagnetic emissions and immunity of the combination must conform to EN 60601-1-2.
Peripheral used in the non-patient environment
The system has been verified for compatibility, and compliance for connection to a local area network (LAN) via a wire LAN, provided the LAN components are EN 60601-1 compliant.
7
Page 21
ECO 5VET Digital Color Doppler Ultrasound System
General precautions for installing an alternate off-board, remote device or a network would include:
The added device(s) must have appropriate safety standard conformance and CE Marking. The added device(s) must be used for their intended purpose having a compatible interface.
2.5 Biological Safety
This product, as with all diagnostic ultrasound equipment, should be used only for valid reasons and should be used both for the shortest period of time and at the lowest power settings necessary (ALARA - As Low As Reasonably Achievable) to produce diagnostically acceptable images. The AIUM offers the following guidelines:
Clinical Safety Quoted from AIUM
Approved March 26, 1997
Diagnostic ultrasound has been in use since the late 1950s. Given its known benefits and recognized efficacy for medical diagnosis, including use during human pregnancy, the American Institute of Ultrasound in Medicine herein addresses the clinical safety of such use: There are no confirmed biological effects on patients or instrument operators caused by exposures from present diagnostic ultrasound instruments. Although the possibility exists that such biological effects may be identified in the future, current data indicate that the benefits to patients of the prudent use of diagnostic ultrasound outweigh the risks, if any that may be present.
Heating: Elevating tissue temperature during obstetrical examinations creates medical concerns. At the embryo development stage, the rise in temperature and the length of time exposed to heat combine to determine potential detrimental effects. Exercise caution particularly during Doppler/Color exams. The Thermal Index (TI) provides a statistical estimate of the potential temperature elevation (in centigrade) of tissue temperature. Three forms of TI are available: Soft Tissue Thermal Index (TIS), Bone Thermal Index (TIB) and Cranial Bone Thermal Index (TIC).
Soft Tissue Thermal Index (TIS). Used when imaging soft tissue only, it provides an estimate of potential temperature increase in soft tissue. Bone Thermal Index (TIB). Used when bone is near the focus of the image as in the third trimester OB examination, it provides an estimate of potential temperature increase in the bone or adjacent soft tissue. Cranial Bone Thermal Index (TIC). Used when bone is near the skin surface as in transcranial examination, it provides an estimate of potential temperature increase in the bone or adjacent soft tissue.
Cavitations: Cavitations may occur when sound passes through an area that contains a cavity, such as a gas bubble or air pocket (in the lung or intestine, for example). During the process of cavitations, the sound wave may cause the bubble to contract or resonate. This oscillation may cause the bubbles to explode and damage the tissue. The Mechanical Index (MI) has been created to help users accurately evaluate the likelihood of cavitations and the related adverse effects. MI recognizes the importance of non-thermal processes, cavitations in particular, and the Index is an attempt to indicate the probability that they might occur within the tissue.
8
Page 22
ECO 5VET Digital Color Doppler Ultrasound System
2.6 Scanning Patients and Education
The Track-3 or IEC60601-2-37 output display standard allows users to share the responsibility for the safe use of this ultrasound system. Follow these usage guidelines for safe operation:
In order to maintain proper cleanliness of the probes, always clean them between patients. Always use a disinfected sheath on all EV/ER probes during every exam. Continuously move the probe, rather than staying in a single spot, to avoid elevated temperatures in one part
of the patient’s body.
Move probe away from the patient when not actively scanning. Understand the meaning of the TI, TIS, TIB, TIC and MI output display, as well as the relationship between
these parameters and the thermal/cavitation bioeffect to the tissue. Expose the patient to only the very lowest practical transmit power levels for the shortest possible time to achieve a satisfactory diagnosis (ALARA - As Low As Reasonably Achievable).
2.6.1 Safe Scanning Guidelines
Ultrasound should only be used for medical diagnosis and only by trained medical personnel.
Diagnostic ultrasound procedures should be done only by personnel fully trained in the use of the equipment,
in the interpretation of the results and images, and in the safe use of ultrasound (including education as to potential hazards).
Operators should understand the likely influence of the machine controls, the operating mode (e.g. B mode) and probe frequency on thermal and cavitation hazards.
Select a low setting for each new patient. Output should only be increased during the examination if penetration is still required to achieve a satisfactory result, and after the Gain control has been moved to its maximum value.
Maintain the shortest examination time necessary to produce a useful diagnostic result.
Do not hold the probe in a fixed position for any longer than is necessary. The frozen frame and Cine loop
capabilities allow images to be reviewed and discussed without exposing the patient to continuous scanning.
Do not use endo-cavitary probes if there is noticeable self heating of the probe when operating in the air. Although applicable to any probe, take particular care during trans- vaginal exams during the first eight weeks of gestation.
Take particular care to reduce output and minimize exposure time of an embryo or fetus when the temperature of the mother is already elevated.
Take particular care to reduce the risk of thermal hazard during diagnostic ultrasound when exposing: an embryo less than eight weeks after gestation; or the head, brain or spine of any fetus or neonate.
9
Page 23
ECO 5VET Digital Color Doppler Ultrasound System
TI
Maximum exposure time (minutes)
0.7
60
1.0
30
1.5
15
2.0
4
2.5
1
Operators should continually monitor the on-screen thermal index (TI) and mechanical index (MI) values and use control settings that keep these settings as low as possible while still achieving diagnostically useful results. In obstetric examinations, TIS (soft tissue thermal index) should be monitored during scans carried out in the first eight weeks after gestation, and TIB (bone thermal index) thereafter. In applications where the probe is very close to bone (e.g. trans-cranial applications), TIC (cranial bone thermal index) should be monitored. MI> 0.3 there is a possibility of minor damage to neonatal lung or intestine. If such exposure is necessary, reduce the exposure time as much as possible.
MI> 0.7 there is a risk of cavitations if an ultrasound contrast agent containing gas micro-spheres is being used. There is a theoretical risk of cavitations without the presence of ultrasound contrast agents. The risk increases with MI values above this threshold.
TI> 0.7 the overall exposure time of an embryo or fetus should be restricted in accordance with Table 2-2 below as a reference:
Table 2-2 Maximum recommended exposure times for an embryo or fetus
Non-diagnostic use of ultrasound equipment is not generally recommended. Examples of non-diagnostic uses of ultrasound equipment include repeated scans for operator training, equipment demonstration using normal subjects, and the production of souvenir pictures or videos of a fetus. For equipment of which the safety indices are displayed over their full range of values, the TI should always be less than 0.5 and the MI should always be less than 0.3. Avoid frequent repeated exposure of any subject. Scans in the first trimester of pregnancy should not be carried out for the sole purpose of producing souvenir videos or photographs, nor should their production involve increasing the exposure levels or extending the scan times beyond those needed for clinical purposes.
Diagnostic ultrasound has the potential for both false positive and false negative results. Misdiagnosis is far more dangerous than any effect that might result from the ultrasound exposure. Therefore, diagnostic ultrasound system should be performed only by those with sufficient training and education.
2.6.2 Understanding the MI/TI Display
Track-3 follows the Output Display Standard for systems that include fetal Doppler applications. The acoustic output will not be evaluated on an application-specific basis, but the global maximum de-rated Ispta must be ≤
10
Page 24
ECO 5VET Digital Color Doppler Ultrasound System
1
0.9332
0.8128
0.7080
0.5957
2
0.8710
0.6607
0.5012
0.3548
3
0.8128
0.5370
0.3548
0.2113
4
0.7586
0.4365
0.2512
0.1259
5
0.7080
0.3548
0.1778
0.0750
6
0.6607
0.2884
0.1259
0.0447
720 mW/ cm2 and either the global maximum MI must be ≤ 1.9 or the global maximum de-rated Isppa must be
≤ 190 W/cm
1.0; Ispta.3 ≤50mW/cm
2
. An exception is for ophthalmic use, in which case the TI = max (TIS_as, TIC) is not to exceed
2
, and MI ≤ 0.23. Track-3 gives the user the freedom to increase the output acoustic power for a specific exam, and still limit output acoustic power within the global maximum de-rated Ispta 720 mW/cm2 under an Output Display Standard.
For any diagnostic ultrasonic systems, Track-3 provides an Output Indices Display Standard. The diagnostic ultrasound systems and its operation manual contain the information regarding an ALARA (As Low As Reasonably Achievable) education program for the clinical end-user and the acoustic output indices, MI and TI. The MI describes the likelihood of cavitations, and the TI offers the predicted maximum temperature rise in tissue as a result of the diagnostic examination. In general, a temperature increase of
2.5°C must be present consistently at one spot for 2 hours to cause fetal abnormalities. Avoiding a local temperature rise above 1°C should ensure that no thermally induced biologic effect occurs. When referring to the TI for potential thermal effect, a TI equal to 1 does not mean the temperature will rise 1 degree C. It only means an increased potential for thermal effects can be expected as the TI increases. A high index does not mean that bioeffects are occurring, but only that the potential exists and there is no consideration in the TI for the scan duration, so minimizing the overall scan time will reduce the potential for effects. These operator control and display features shift the safety responsibility from the manufacturer to the user. So it is very important to have the Ultrasound systems display the acoustic output indices correctly and the education of the user to interpret the value appropriately.
RF: (De-rating factor) In Situ intensity and pressure cannot currently be measured. Therefore, the acoustic power measurement is normally done in the water tank, and when soft tissue replaces water along the ultrasound path, a decrease in intensity is expected. The fractional reduction in intensity caused by attenuation is denoted by the de-rating factor (RF),
RF = 10 (-0.1 a f z)
Where a is the attenuation coefficient in dB cm-1 MHz-1, f is the transducer center frequency, and z is the distance along the beam axis between the source and the point of interest.
De-rating factor RF for the various distances and frequencies with attenuation coefficient 0.3dB cm-1 MHz-1 in homogeneous soft tissue is listed in the following table. An example is if the user uses 7.5MHz frequency, the power will be attenuated by .0750 at 5cm, or 0.3x7.5x5=-11.25dB. The De- rated Intensity is also referred to as ‘.3’ at the end (e.g. Ispta.3).
Distance Frequency (MHz) (cm) 1 3 5 7.5
11
Page 25
ECO 5VET Digital Color Doppler Ultrasound System
7
0.6166
0.2344
0.0891
0.0266
Large aperture (>1cm2)
2
TIS
Soft tissue
Unscanned
Small aperture (<1cm2)
Pencil Probe
3
TIS
Soft tissue
Scanned
Evaluated at surface
Breast color
4
TIB
Soft tissue and bone
Scanned
Soft tissue at surface
Muscle color
5
TIB
Soft tissue and bone
Unscanned
Bone at focus
Fetus head PW
6
TIC
Soft tissue and bone
Unscanned/scanned
Bone at surface
Transcranial
I’=I*RF Where I’ is the intensity in soft tissue, I is the time-averaged intensity measured in water.
Tissue Model: Tissue temperature elevation depends on power, tissue type, beam width, and scanning mode. Six models are developed to mimic possible clinical situations.
Thermal Models Composition Mode Specification Application
Soft tissue: Describes low fat content tissue that does not contain calcifications or large gas-filled spaces.
Scanned: (auto-scan) Refers to the steering of successive burst through the field of view, e.g. B and color mode.
Unscanned: Emission of ultrasonic pulses occurs along a single line of sight and is unchanged until the transducer is moved to a new position. For instance, the PW, and M mode.
TI: TI is defined as the ratio of the In Situ acoustic power (W.3) to the acoustic power required to raise tissue temperature by 1°C (Wdeg), TI = W.3/Wdeg. Three TIs corresponding to soft tissue (TIS) for abdominal; bone (TIB) for fetal and neonatal cephalic; and cranial bone (TIC) for pediatric and adult cephalic, have been developed for applications in different exams. An estimate of the acoustic power in milli-watts necessary to produce a 1°C temperature elevation in soft tissue is:
Wdeg = 210/fc,for model 1 to 4, where fc is the center frequency in MHz.
Wdeg = 40 K Dfor model 5 and 6, where K (beam shape factor) is 1.0, D is the aperture diameter in cm at the depth of interest.
MI: Cavitation is more likely to occur at high pressures and low frequencies in pulse ultrasound wave in the tissue, which contains the bubble or air pocket (for instance, the lung, intestine, or scan with gas contrast agents). The threshold under optimum conditions of pulsed ultrasound is predicted by the ration of the peak pressure to the
12
Page 26
ECO 5VET Digital Color Doppler Ultrasound System
square root of the frequency.
MI = Pr’ / sqrt(fc) Pr’ is the de-rated (0.3) peak rare-fractional pressure in Mpa at the point where PII is the
maximum, and fc is the center frequency in MHz. PII is the Pulse Intensity Integral that the total energy per unit area carried by the wave during the time duration of the pulse. The peak rare- fractional pressure is measured in hydrophone maximum negative voltage normalized by the hydrophone calibration parameter.
Display Guideline: For different operation modes, different indices must be displayed. However, only one index needs to be shown at a time. Display is not required if maximum MI is less than 1.0 for any setting of the operating mode, or if maximum TI is less than 1.0 for any setting of the operating mode. For TI, if the TIS and TIB are both greater than 1.0, the scanners need not be capable of displaying both indices simultaneously. If the index falls below 0.4, no display is needed. The display increments are no greater than 0.2 for index value less than one and no greater than 1.0 for index values greater than one (e.g. 0.4, 0.6, 0.8, 1, 2, and 3).
Display and Report
Located on the upper middle section of the system display monitor, the acoustic output display provides the operator with real-time indication of acoustic levels being generated by the system.
For Scan Only display and report MI, and start from 0.4 if maximum MI > 1.0, display in increments of 0.2.
Below is a simple guideline for the user when TI exceeds one limit exposure time to 4(6-TI) minutes based on
the ‘National Council on Radiation Protection. Exposure Criteria for Medical Diagnostic Ultrasound: I.
Criteria Based on Thermal Mechanisms. Report No.113 1992’.
Operator Control Features: The user should be aware that certain operator controls may affect the acoustic output. It is recommended to use the default (or lowest) output power setting and compensate using Gain control to acquire an image. Other than the output power setting in the soft-menu, which has the most direct impact on the power; the PRF, image sector size, frame rate, depth, and focal position also slightly affect the output power. The default setting is normally around 70% of the allowable power depending on the exam application mode.
Controls Affecting Acoustic Output The potential for producing mechanical bioeffects (MI) or thermal bioeffects (TI) can be influenced by certain controls. Direct: The Acoustic Output control has the most significant effect on Acoustic Output. Indirect: Indirect effects may occur when adjusting controls. Controls that can influence MI and TI are detailed under the bioeffect portion of each control in the Optimizing the Image chapter. Always observe the Acoustic Output display for possible effects.
Best practices while scanning HINTS: Raise the Acoustic Output only after attempting image optimization with controls that have no effect
13
Page 27
ECO 5VET Digital Color Doppler Ultrasound System
on Acoustic Output, such as Gain and TGC.
WARNING: Be sure to have read and understood control explanations for each mode used before
attempting to adjust the Acoustic Output control or any control that can affect Acoustic Output.
Use the minimum necessary acoustic output to get the best diagnostic image or measurement during an examination. Begin the exam with the probe that provides an optimum focal depth and penetration.
Acoustic Output Default Levels In order to assure that an exam does not start at a high output level, the system initiates scanning at a reduced default output level. This reduced level is preset programmable and depends upon the exam icon and probe selected. It takes effect when the system is powered on or New Patient is selected. To modify acoustic output, adjust the Power Output level on the Soft Menu.
14
Page 28
Chapter 3 System Introduction
3.1 Consol Overview
ECO 5VET Digital Color Doppler Ultrasound System
Console Overview
3.2 Physical Specification
335mm(Length)×155mm(Width)×350mm(Height)
3.3 System View in Different Views
15
Page 29
ECO 5VET Digital Color Doppler Ultrasound System
System Front View
System Side View
1. Ethernet 2. VGA 3. USB 4.Footswicth 5.Power in
1. Probe Holder 2.Probe 3.Keyboard
1. Probe 2.USB 3.Remote 4. Video Out 5.Battery Holder 6. Probe Holder
3.4 Function Introduction
Console Overview
System Rear View
1.Possess B, B/B, 4B, B/M, M,CFM,PW display mode, M, B/M mode possess 4 kinds of scan velocity;
16
Page 30
ECO 5VET Digital Color Doppler Ultrasound System
C3-A, 2.5-5.0MHz Convex Array Indications for Use: ABD, OB, GYN,URO
L7M-A, 5.3-10.0MHz Linear Array Indications for Use: Vessel, Small Parts, Equine Tendon
L7S-A, 5.3-11.0MHz Linear Array
MC6-A, 4.5-8.0MHz Micro-convex Array
2.Possess many combination of focus, total gain control 8 segments STC;
3.Possess depth scan and image left-right up-down rolling-over function;
4.Possess a lot of imaging technology for example multiple compound imaging (space frequency compound), multiple frequency, multiple zoom ratio, pan zoom, screen scroll, chroma, harmonic imaging etc;
5.Possess image processing, total gain, dynamic range, frequency, focus number, focus position, zoom, compound, scan width, line density, smooth, edge enhancement, frame, persistence, gray scale, restrain boost multi-beam, acoustic power, M velocity;
6.Possess distance proportion circumference volume rate angle histogram in B mode distance time velocity heart rate measurement function in M mode ;OB measurement software package, small parts measurement software package, ventricle function measurement software package and user defined formula;
7.Possess body mark arrow case number display, real time clock display annotation in image user-defined annotation;
8.Possess multi-language interface display User interface change, shear plate, printing ,DICOM3.0 biopsy guided functions;
9.It has permanent storage for image and cine and optional 320GB HDD. It can also be connected to removable storage via USB port. To realize mass storage ,can recall saved image for analysis;
10.Cine loop storage 256frames real time image;
11.Screen rotation function make you adjust screen angle according to users' requirement 0~30°;
12.Output standard PAL or NTSC video signal and VGA signal;
13.Print or export graphic report.
3.4.1 Image Modes
B mode B/M mode M mode 2B mode 4B mode CFM mode PW mode
3.4.2 Accessories
Transducers:
17
Page 31
ECO 5VET Digital Color Doppler Ultrasound System
Indications for Use: Vessel, Small Parts, Equine Tendon
Indications for Use: Canine Abdomen, Feline Abdomen, Small Animal Cardiac
L7V-A Linear Array, 5.3-10.0MHz Linear Array Indications for Use: Big Animal, Reproduction
MC5V-A, 4.0-7.0MHz Micro-convex Array Indications for Use: Canine Abdomen, Feline Abdomen,
Peripherals
VGA output for external monitor VIDEO output for B&W video printer LAN port output LAN for DICOM and image review station USB 2.0 for flash drive Foot switch AC/DC adapter: MDS-060AAS19 B
Input: 100-250V~1.5-0.75A, 50-60Hz Output: 19V 3.15A DELTA ELECTRONICS, INC.
Battery Pack: BT-2500, 4400mAh, DONGGUAN POWER INC
3.5 Installation Procedures
Note: Please do not turn on the power switch until finishing all the installation and necessary preparation.
3.5.1 Environment Condition
The system should be operated under the following environment.
3.5.1.1 Operation Environment Requirement
Ambient Temperature: 10 ℃~40 Relative Humidity: 30%75%RH Atmospheric Pressure: 700hPa1060hPa
3.5.1.2 Transport and Storage Environmental Requirement
The following environmental transport and storage conditions are within system tolerances: Temperature: -5º C ~ 40º C Relative Humidity: ≤ 80% non-condensing Atmosphere Pressure:700hPa ~ 1060hPa
3.5.1.3 Electrical Requirements
Power Consumption: less than 60 VA Voltage Fluctuation
18
Page 32
ECO 5VET Digital Color Doppler Ultrasound System
WARNING
Maintain a fluctuation range of less than ±10% of voltage labeling on rear panel of the system, otherwise the system may be damaged.
Grounding
Before connecting the power cable, connect the attached ground protection cable from Equipotentiality terminal on system rear panel to a specialized grounding device.
NOTE
Please follow the outlined power requirements. Only use power cables that meet the system guidelinesfailure to follow these procedures may produce system damage. Line power may vary in different geographic locations. Refer to the detailed ratings on the rear panel of the system for detailed information. Battery To avoid the battery bursting, igniting, or fumes from the battery; causing equipment damage, observe the following precautions: Do not immerse the battery in water or allow it to get wet. Do not put the battery into a microwave oven or pressurized container. If the battery leaks or emits an odor, remove it from all possible flammable sources. If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during use, recharging or storage, immediately remove it and stop using it. If you have any questions about the battery, short term (less than one month) storage of battery pack: Store the battery in a temperature range between 0 degrees C (32 degrees F) and 50 degrees C (122 degrees F). Long term (3 months or more) storage of battery pack: Store the battery in a temperature range between-20 degrees C (-4 degrees F) and 45 degrees C (113 degrees F); Upon receipt of the ECO and before first time usage, it is highly recommended that the customer performs one full discharge/charge cycle. If the battery has not been used for >2 months, the customer is recommended to perform one full discharge/charge cycle. It is
also recommended to store the battery in a shady and cool area with FCC (full current capacity).• One Full
Discharge/Charge Cycle Process:1. Full discharge of battery to let the ECO automatically shut down.2. Charge the ECO to 100% FCC (full current capacity).3. Discharge of Venue 40 for complete shutdown(takes one hour for discharge).• When storing packs for more than 6 months, charge the pack at least once during the 6 month timeframe to prevent leakage and deterioration in performance.
3.5.1.4 Operation Space
Please leave enough free space from the back of the system to ensure well ventilation.
Caution: Leave enough free space from the back of the system; otherwise, with the increasing of the
temperature inside the unit, malfunction may occur.
3.5.1.5 System Positioning & Transporting
Moving the System
When moving or transporting the system, take the precautions described below to ensure maximum safety for personnel, the system and other equipments.
Before Moving the System
19
Page 33
ECO 5VET Digital Color Doppler Ultrasound System
Press 3 s, system will forced shut down and completely switch off the system. Disconnect all cables from off-board peripheral devices (external printer, etc.) from the console.
NOTE
To prevent damage to the power cord, DO NOT pull excessively on the cord or sharply bend the cord while wrapping it.
Store all probes in their original cases or wrap them in soft cloth or foam to prevent damage. Replace gel and other essential accessories in the appropriate storage case. Ensure that no loose items are left on the console.
When Moving the System
Carry the system with handle, or put the system on the cart to move it.
NOTE
Walk slowly and carefully when moving the system. Do not let the system strike walls or doorframe.
Transporting the System
Use extra care when transporting the system in a vehicle. After preparing the system as described above, take the following additional precautions:
Only use vehicles that are suitable for transport of the system. Before transporting, place the system in its original storage carton. Load and unload the system to a vehicle parked on a level surface. Load the unit abroad the vehicle carefully and over its center of gravity. Keep the unit still and
upright.
Ensure that the transporting vehicle can bear the weight of system plus the passengers. Secure the system firmly with straps or as directed within the vehicle to prevent movement during transport.
Any movement, coupled with the weight of the system, could cause it to break loose. Drive carefully to prevent damage from vibration. Avoid unpaved roads, excessive speeds, and erratic stops or starts.
3.5.2 Powering the System
3.5.2.1 Acclimation Time
After being transported, the unit requires one hour for each 2.5 º increment if its temperature is below 10 ºC or above 40 ºC.
NOTE
Please keep at least 20 to 30 cm spare space away from the back of the system to ensure well ventilation. Otherwise, with the increasing of the temperature inside the unit, malfunction may occur.
3.5.2.2 Connecting the electric power
After making sure that the AC power supply in hospital is in normal status, and this AC voltage type matches to the power requirements indicated on the label of system, then please connect the plug of power cord to the POWER IN socket at the rear panel of the system, and connect the other end of power cord to the AC power
20
Page 34
ECO 5VET Digital Color Doppler Ultrasound System
supply socket in hospital. Please use the power cable provided by the manufacturer, other type of power cable is not allowed.
Press 1 s, system boot,
Press and pop up dialog for shut down. Click the enter key then power off.
Or press 3 s, system will forced shut down
CAUTION
Connecting the system to the wrong AC power supply may cause damage to the system and danger to the operators and animals.
3.5.3 Probe Installment
CautionPlease only use the probes provided by manufacturer for this model, other types of probes are
not allowed to use with this system! Otherwise it may cause the damage to the system and the probe.
CAUTION
Before connecting the probe, please carefully check the probe lens, probe cable and probe connector to see whether there is anything abnormal, such as cracks, falls off. Abnormal probe is not allowed to connect to the system; otherwise there is possibility of electricity shock.
Hold the probe connector lock switch, and insert the connector socket vertically. Release the probe lock switch. Check the locked probe with one hand to make sure that it’s not loose, and it’s securely connected
CAUTION
Only power supply at “turn off” state, can install / take-down the probe, otherwise it will damage the machine or the probe. When installing and disassembling probe, please put the probe head inside the probe holder, it can prevent the probe falling down to the ground.
3.5.3.1 Probe Disassembly
Turn the dead lock switch 90degree in counterclockwise direction, extract probe connector plug vertically.
3.5.4 Accessories Installment
Caution:Please only use the optional parts provided or suggested by manufacturer! Using other
types of optional devices may cause the damage to the system and the connected optional devices.
3.5.4.1 Video printer installment
1.Put video printer stably.
2.Connect cable of video printer to video port in the back of the device. And connect the other side to video signal output port in the rear side.
21
Page 35
ECO 5VET Digital Color Doppler Ultrasound System
3.Connect the printer line to print control port in the printer rear side, and connect the other side to the print control port in the unit rear side.
4.Connect power cable of video printer to power system..
5.Adjust printer parameter preset according to the type of printing paper
Caution: Do not use any other power cable to replace 3-wire power cable manufacturer provides,
otherwise there is danger of electric shock.
Video printer sign introduction
: Video signal input port
Video signal output port
Print control port
Video printer switch
3.5.4.2 Graphic printer installment
Put graphic printer stably, connect printer cable to USB port in the left side of the unit. Connect the power cable of graphic printer to power system.
CautionPlease see packing list for fundamental configuration!
22
Page 36
ECO 5VET Digital Color Doppler Ultrasound System
Chapter 4 Control Panel
4.1 Alphanumeric Keyboard
Alphanumeric Keyboard
The alphanumeric keys are used for inputting patient number, name, character and figure etc.
4.2 Function Keys/Knob
4.2.1 SWITCH
Turn on or off the device
4.2.2 PATIENT
Set up a new patient data, input name and other information.
4.2.3 PROBE
Press this key for selecting probe. It can only select the connected probe.
4.2.4 SETUP
23
Page 37
ECO 5VET Digital Color Doppler Ultrasound System
Press this key to get in or out the system setting page.
4.2.5 END
Press END key to finish the exam.
4.2.6 BODYMARK
Press this key to enter into body mark working status, select the body mark and confirm the probe scanning position on the screen. It is only available in frozen status.
4.2.7 COMMENT
Press this key to enter into comment status, and add comments in the image area on the screen.
4.2.8 ARROWS
Add arrows icon to the image area.
4.2.9 DEL
Press this key to clear the measurement lines, body mark, and comments.
4.2.10 Print
24
Page 38
ECO 5VET Digital Color Doppler Ultrasound System
1
2
3
4
5
6
7
8
9
10
PRINT1: Print the screen image by video printer connected to the system. PRINT2: Print the report by printer connected to the system (Only report page works). Or print the image in the scanning page; Or print the image in the review page.
4.2.11 Archive
File management of system, you can view and edit the patient data.
4.2.12 Report
Produce/ Save/ Recall an examination report.
4.2.13 Parameter Control Key
Increase/decrease the corresponding parameters of the screen, or open/close the function.
4.3 Central Control
25
Page 39
ECO 5VET Digital Color Doppler Ultrasound System
1. UPDATE 2.EXIT 3.Cursor 4.Change 5.MENU
6. Dist 7.Trace 8.Calc 9.ENTER 10.Trackball
4.3.1 ENTER
This multifunction key is work with trackball. The function switches with the unit status. Such as, set the cursor position, body mark position, comment position, toggle trackball function, selected the menu, and confirm the input.
4.3.2 UPDATE
This multifunction key is work with trackball. The function switches with the unit status. Such as, call the annotation and back in measuring.
4.3.3 Exit
Press this key can exit measurement, dialog, and menu.
4.3.4 Cursor
Press this key to show or hide the cursor.
4.3.5 Change
Press this key to change the menu.
4.3.6 Dist
Press this key to enter into distance measurement.
4.3.7 Trace
Press this key to enter into trace measurement, and press UPDATE to change between trace and ellipse.
4.3.8 Calc
Press this key to enter into measurement software package.
4.3.9 MENU
Press MENU-knob for second time to select the item and adjust the parameters. Press MENU-knob for third time to exit from current item. Rotate the MENU- knob to select the item
26
Page 40
ECO 5VET Digital Color Doppler Ultrasound System
4.3.10 Trackball
Trackball is the main operation tool on screen. Position calipers in measurement, the function of the trackball is different under diverse working status.
4.4 The key of Image Mode
4.4.1 B
Display B mode
4.4.2 C
Display CFM mode
4.4.3 D
Display PW mode
4.4.4 M
Press this key to change the mode between B/M and M.
4.5 Control of Image
4.5.1 THI
27
Page 41
Press this key to open or close the THI function.
4.5.2 AIO
Only press this key for automatic optimization image.
4.5.3 CINE
Only press this key to save the current cine loop.
4.5.4 SAVE
ECO 5VET Digital Color Doppler Ultrasound System
Only press this key to save the current image.
4.5.5 Left and Right Invert
Press this key to invert the image from left and right.
4.5.6 Up and Down Invert
Press this key to invert the image from up and down.
4.5.7 STC
28
Page 42
ECO 5VET Digital Color Doppler Ultrasound System
STC can be used for adjusting gain compensation in different image depth.
4.5.8 GAIN
Rotate the knob to adjust the gain of B mode and M mode.
4.5.9 ANGLE/ZOOM Knob
Rotate the knob to adjust the angle or zoom. Press this knob to change the function between angle and zoom.
4.5.10 DEPTH/FOCUS Knob
Press the knob to adjust the depth and function position. Press the knob to change the function between depth and focus position.
4.5.11 Information Area Indicating Machine Status
Left-to-Right of the up row: hard disk, cable network, USB
Hard disk: press this icon to show the capacity of disc to used save data or USB flash disk in current system Cable network: show the present situation of cable network; press this icon to show the IP address of current
system. USB: show whether this system connects USB flash disk or not, press this icon to show USB safely remove interface. Left-to-Right of the down row: input method, DICOM Task Sequence, battery gauge
Input method: press this icon to switch Chinese and English input method. DICOM Task Sequence: press this icon to show DICOM task and its situation ,To terminate the DICOM task,
delete, and so on(Need to activate DICOM) Battery gauge: show the connecting situation of the battery, just press this icon to show the present State of
29
Page 43
ECO 5VET Digital Color Doppler Ultrasound System
charge and discharge, remaining electric quantity and available time.
4.5.12 Indicator Light
From left to right: Adapter Indicator, Charge Indicator, and Sleep Indicator. Adapter Indicator: when the main unit connects to the adapter with power supply, the indicator lights, otherwise extinguishes. Charge Indicator: When the battery is charging, the indicator lights, once the battery is charged, the indicator goes out. Sleep Indicator: When the main unit is in sleep mode, the indicator lights, otherwise extinguished.
30
Page 44
ECO 5VET Digital Color Doppler Ultrasound System
Chapter 5 Operation and Exam Mode
This chapter mainly describes the process of the normal operation of the device, including the preparation before examination, how to get the image, optimize the image, add comments, body mark and so on.
5.1 Preparing the System for Use
5.1.1 The Device Inspection
(1)The device is placed stability; (2)The grid voltage AC 100-240V, 50Hz-60Hz; (3)Cable is properly connected, firm and ground, the adapter is properly connected to the device; (4)Probe is connected and fixed.
5.1.2 Power On
Long press 1 second to start the machine, wait for the system to enter the user interface, activate the probe slot into the B mode.
5.2 Choose Exam Mode
5.2.1 The Probe Identification
The system default automatically identifies the current probe type, when the probe is inserted, Press to switch the probe.
CAUTION: Please connect or disconnect the probe only after the system is freezing, in order to
ensure stability and extend the service life of the probe.
5.2.2 Mode Selection
In probe selection interface, probe and clinical application selection page is displayed, you can choose needed probe and inspection part, and press the default into the B mode, start scan detection.
NOTE: The system has been set clinical application pre-set before leaving factory, each probe has its
own pre-set. The detailed operation steps of the clinical application pre-set of the probe, please refer to the pre-set section.
31
Page 45
ECO 5VET Digital Color Doppler Ultrasound System
5.3 Patient Data Entry
Press the to display the Patient screen
Patient’s Information Screen
Function Buttons on Patient screen: [Archive]: Operation on the patient information which has already existed; [Wordlist]Recall patient information in worklist. And need to open the DICOM function; [New Patient]Create a new patient information identity; [New Study]Choose exam applications for the new patient; [End Study]Edit patient’s exam item; [OK]Save patient information; [Cancel]Cancel the operation of new patient information; Operation Methods:
(1)Move the Trackball to the position of inputting character, then input patient information by character keyboard. (2)Use the Trackball and the [ENTER] key to switch between different input options: ID, patient name, doctor's name, birthday (It can be automatically calculated when input age), age (It can be automatically calculated when input birthday), gender. (3)Select the exam items, and input the regular inspection information. (4)After inputting the required information, click on the [OK] key to save the patient information, the system will return to the B mode. (5)Recall information of the previous patient, you can use the Archive or Worklist to recall patient information to exam.
CAUTION: Creating a diagnostic record, you should check the accuracy of the patient information
before saving measurement or image; otherwise, it will be stored in the wrong patient records. After checking the patient, press the [END] key to save the patient information in the system.
32
Page 46
ECO 5VET Digital Color Doppler Ultrasound System
4
2
3
5
6
10 9 7
8
1
5.4 Image Interface Display
1, Logo 2, Control menu 3, Image status prompt 4, Image parameter area 5, System state prompt 6, Image parameter area 7, Gray-scale strip 8, Start point of scanning 9, Image Region 10, Cine loop
5.5 Display Mode
Display Mode: B, 2B, 4B, M, B/M, they can be shifted by the mode key.
5.5.1 B Mode
Press [B] Mode key, and display the single B Mode image, B Mode is the basic mode for two-dimensional scanning and diagnosis.
5.5.2 B/B Mode
Press [2B] to display double B mode images side by side. One image is in real-time status; the other is in frozen status. The real-time image has start scan marker and ruler marker .Press [2B] in B/B mode, the original active image is frozen while the original frozen image is activated.
5.5.3 4B Mode
Press [4B] to enter into 4B mode, the screen will display four B mode images side by side, but only one image is in real-time status. Pressing it again can switch the real-time status among four images.
33
Page 47
ECO 5VET Digital Color Doppler Ultrasound System
5.5.4 B/M Mode
Press [B/M] key, a real time B mode image and a real-time M-mode image will be displayed at the same time. And a sample line will appear in the B mode image area, which indicates the active sample position for M image on the B image area. Click the position on the B image area to fix the position of sampling line.
5.5.5 M Mode
Press [M] key again, B mode image will disappear; M mode image is still active on the whole screen. M mode image stands for the tissue movement status at the sampling line. The M mode image varies with time, so it is mainly used for cardiac applications.
5.5.6 CFM Mode
CFM is a Doppler mode intended to add color-coded qualitative information concerning the relative velocity and direction of fluid motion within the B mode image.
CFM is useful to see flow in a broad area. It allows visualization of flow in the CROI, whereas Doppler mode provides spectral information in a smaller area. CFM is also used a stepping stone to Doppler mode. You can use CFM to locate flow and vessels prior to activating Doppler.
In CFM mode, move the trackball to change the position of sampling box. Activate the [Steering Angle] and rotate the [MENU] knob to adjust the angle of color sampling box (if current probe is linear probe). Press [ENTER] key to fix the position of color sampling box.At this time adjusts the size of color sampling box through moving trackball. Press [ENTER] key again and move trackball to change the color sampling position again.
Press [C] key to enter into CFM mode; after [C] key light is on, rotate the [GAIN] knob for adjusting the gain of CFM.
CFM mode Exam Procedure:
Follow the same procedure as described under B mode to locate the anatomical area of interest. After optimizing the B mode image, add Color Flow. Move the color region of interest CROI as close to the center of the image as possible. Optimize the color flow parameters so that a high frame rate can be achieved and appropriate flow velocity
can be visualized.
Press [FREEZE] key to hold the image in cine memory. Record color flow image as necessary.
CFM Scanning Hints: PRF: increase/decrease the PRF on the color bar. Imaging of higher velocity flow requires increased velocity
scale values to avoid aliasing
Wall Filter: affect low flow sensitivity versus motion artifact Color Map: allow you to select a specific color map. It shows the direction of the flow and highlights the
higher velocity flows.
34
Page 48
ECO 5VET Digital Color Doppler Ultrasound System
Color Gain: amplify the overall strength of echoes processed in the CROI Persistence: affect temporal smoothing and color Doppler ‘robustness’.
5.5.7 PW Mode
Doppler is intended to provide measurement data concerning the velocity of moving tissues and fluids. PW Doppler lets you examine blood flow data selectively from a small region called the Sample Volume. The X axis represents time while the Y axis represents velocity in either a forward or reverse direction. PW Doppler is typically used for displaying the speed, direction, and spectral content of blood flow at selected anatomical sites.
PW Doppler can be combined with B mode for quick selection of the anatomical site for PW Doppler examination. The site where PW Doppler data is derived appears graphically on the B mode image (Sample Volume Gate). The Sample Volume Gate can be moved anywhere within B mode image.
PW mode Exam Procedure:
Get a good B mode image. Press [C] key to help locate the vessel you wish to examine. Press [D] key to display the sample volume cursor and gate. Position the sample volume cursor by moving the Trackball left and right. Position or re-size the sample
volume gate by moving the Trackball up and down, then press [ENTER] key. Press [UPDATE] key to display PW Doppler spectrum and the system will run in combined B+Doppler mode. The Doppler signal can be heard through the speakers.
Optimize the PW Doppler spectrum as necessary. Ensure that the sample line is parallel to the blood flow. Press [FREEZE] key to hold the trace in cine memory and stop imaging. Perform measurements and calculations, as necessary. Record results with your recording devices. Press [FREEZE] key to resume imaging. Repeat the above procedure until all relevant flow sites have been examined. Replace the probe in its respective holder.
When entering Duplex mode for the first time, the Doppler spectrum is not activated. The Doppler Sample Volume appears in the default position, and the B mode image or 2D (either B or Color) mode are active. Moving the Trackball will change the Sample Volume position. Press the [ENTER] key to toggle the Trackball function between Sample Volume Gate position and size. Press the [UPDATE] key after the Sample Volume Gate is defined to activate the Spectral Doppler mode. Press the [UPDATE] key for second time to toggle back to 2D (B or Color) update and deactivate the Spectral Doppler.
Doppler mode Scanning Hints:
The best Doppler data will be got when the scanning direction is parallel to the direction of the blood flow; when the scanning direction is perpendicular to the anatomic target, you can get the best B mode image, so you should keep the balance as you don’t usually get both an ideal B mode image and ideal Doppler data simultaneously. PRF: adjust the velocity scale to accommodate faster/slower blood flow velocity. Velocity scale determines pulse repetition frequency. Wall Filter: remove the noise caused by vessel or heart wall motion at the expense of low flow sensitivity.
35
Page 49
ECO 5VET Digital Color Doppler Ultrasound System
Baseline: adjust the baseline to accommodate faster or slower blood flows to eliminate aliasing. Angle: optimize the accuracy of the flow velocity. It estimates the flow velocity in a direction at an angle to
the Doppler vector by computing the angle between the Doppler vector and the flow to be measured. This is special useful in vascular applications where you need to measure velocity.
Doppler Gain: allow you to control the background information of spectral. Sweep Speed: control speed of spectral update.
Doppler Sample Volume Gate Position and Size (Trackball and SET)
Move the sample volume on the B mode’s Doppler cursor. The gate is positioned over a specific position within the vessel.
To move Doppler cursor position, turn the trackball left or right until positioned over the vessel. To move sample volume gate position, move the trackball up or down until positioned inside the vessel. To size sample volume gate, press [ENTER] key to toggle trackball function from sample volume gate
positioning to sizing, then move the trackball to change sample volume gate size.
5.6 B Image Adjustment
5.6.1 Frequency
In real status, press the first line key of [Frequency] to increase the frequency, and press the second line key of [Frequency] to decrease the frequency.
5.6.2 Dynamic
Dynamic range is used for adjusting the contrast resolution of B mode image and mode image, compressing or enlarging the display range of gray scale. At the real-time status, press the first line key of [Dynamic] to increase the Dynamic, and press the second line key of [Dynamic] to decrease the Dynamic. The range is 30 to 90.
5.6.3 i-Image
In real status, press the corresponding key of [i-Image] to adjust, the range is 0~3.
5.6.4 Compound
In real status, press the corresponding key of [Compound] to adjust, and the range is 0~3.
The SRA can’t be edited after opening the compound.
5.6.5 SRA
In real status, press the corresponding key of [SRA] to turn on or off.
5.6.6 M Speed Adjustment
36
Page 50
ECO 5VET Digital Color Doppler Ultrasound System
In real M status, press the corresponding key of [Speed] to adjust M speed, and the range is 1~4.
5.6.7 Gain
In real status, rotate [Gain] knob to adjust the Gain, and the range is 0~255, the step is 5.
5.6.8 STC
STC curves can be used for adjusting gain compensation in different image depth. Drag the slide of STC to adjust the value. STC curve will disappear automatically 1 second later after stopping adjustment.
5.6.9 Depth
Press [DEPTH/FOCUS] selection knob until the indicator of [DEPTH] is lit, and then rotate the knob to change the depth of image.
5.6.10 Focus Pos
Press the [DEPTH/FOCUS] selection knob until the indicator of [FOCUS POS] is lit. Then turn the knob to change Focus Position.
5.6.11 Angle
Change the angle of sampling line of sampling gate in PW mode. Press [ANGLE/ZOOM], confirm the light of ANGLE on, the sampling line rotates with the direction.
5.6.12 ANGLE/ZOOM
Press [ANGLE/ZOOM] knob until the indicator of [ANGLE] is lit, and the angle of sample gate will rotate with the rotation direction. Press [ANGLE/ZOOM] knob to confirm the ZOOM button is lit, and it will appear the zoom box, rotate the knob to select the zoom steps.
5.6.13 INVERT
B mode image and B/M mode image can be reversed horizontally and vertically.
Press the key, the displayed image is reversed in the right-left horizontal direction.
Press the key, the displayed image is reversed in the up-down direction. The horizontal flip status indicators of the upper-left corner of the image window have the following
meanings:
The meaning of the symbol “ ” indicating the probe initiative scanning position “ ” situated in the left
37
Page 51
ECO 5VET Digital Color Doppler Ultrasound System
indicates that the first scanning line in the left of the screen is corresponding to the initiative scanning position of the probe, ” situated in the right indicates that the first scanning line in the right of the screen is corresponding to the initiative scanning position of the probe.
5.6.14 THI
Turn ON/OFF THI
5.6.15 AIO
Only press this key to optimize image.
5.7 B Image Menu Adjustment
Click [Change] key or [MENU] knob to display the menu. Rotate [MENU] knob or press[Cursor] key to display the cursor ,and move the sample box to the corresponding function, press [MENU] knob to update the function, then rotate [MENU] knob to adjust the function, press [MENU] again to exit the function.
5.7.1 Scan Width
Select [Scan Width], and adjust the scan width to the corresponding size.
5.7.2 Focus Num
In B mode, 4 focus points can be selected simultaneously, and the number controlled by the depth, SRA and Compound. Move the cursor to [Focus Num] selection to adjust, and the range is 1~4.
5.7.3 Persistence
In real status, adjust the contrast and resolution. In real status, press the first line key of [Persistence] to adjust. The range is 0~7.
38
Page 52
ECO 5VET Digital Color Doppler Ultrasound System
5.7.4 Line Density
Scan Line Density function is only valid for the image in B mode, B/B mode, B/M mode or 4B mode image. The line density has two types: high density and low density. High density means better image quality while low density image has higher frame rate. To do the adjustment, please select the submenu item [LINE DENSITY] and press [MENU] to adjust the line density.
5.7.5 Smooth
Smoothness function is used for restraining the image noise and performing axial smooth processing to make the image smoother. Move the cursor to [Smooth] selection to adjust, and the range is 0~7.
5.7.6 Edge Enhance
Edge enhancement is used for enhancing the image outline. In this way the user can view the tissue structure more clearly. Move the cursor to [Edge] selection to adjust, and the range is 0~7.
5.7.7 Acoustic Power
Acoustic power means the acoustic power transmitting from the probe. At the real-time status, Move the cursor to [Acoustic power] selection to adjust, and the range is 0~15,and the least adjustable level is 1 dB/level.
5.7.8 Utility
This function includes post processing, slide show and other items. Press [MENU] knob, then choose Utility selection, it will appear Utility options.
5.7.8.1 Post Processing
5.7.8.1.1 Chroma
Adjust the type of the chroma. Update [Chroma], and rotate [MENU] knob to select the Chroma type, the range is 0~31.
5.7.8.1.2 2D Map
Select the type of the scale curve. Update [2D Map], and rotate [MENU] knob to select the scale curve type, the range is 0~4.
5.7.8.1.3 B Gamma
Adjust image gray value parameters. Update [B Gamma], and rotate [MENU] knob to select the B Gamma parameters, the range is 0~8.
5.7.8.1.4 B Rejection
Adjust image gray scale inhibition parameters Update [B Rejection], and rotate [MENU] knob to adjust B Rejection parameters, the range is 0~256.
39
Page 53
ECO 5VET Digital Color Doppler Ultrasound System
Transducer
Compatible Biopsy Guide Model
V6-A
610-1093 (10041823)
5.7.9 Advance
In B Mode, select Advance, after pressing [MENU] knob, it will appear Advance options.
5.7.9.1 Zoom Coef
Adjust the size of the ruler. Update [Zoom Coef], and rotate [MENU] knob to adjust, the range is 60%~100%.
5.7.9.2 MB
Through opening MB to improve image quality Press [MENU] to turn on or turn off the MB function.
5.7.9.3 Trapezoidal Mode
Press [MENU]to turn on or turn off the trapezoidal function.
5.7.9.4 Biopsy
Show or hide biopsy. Press the [MENU] to show or hide biopsy line. After the biopsy line shows, press the [ENTER] key to activate the adjustment function of biopsy line, horizontal rolling the trackball can translate the biopsy line, vertical rolling the trackball can adjust the line angle, press the [UPDATE] key to set the default biopsy line position.
Accessories
To order biopsy guides; and other supplies and accessories, contact CIVCO Medical Solutions: CIVCO Medical Solutions 102 First Street South, Kalona, IA 52247-9589 Telephone: 800-445-6741 (USA and Canada), +1 319-656-4447 (International) Fax: 877-329-2482 (USA and Canada), +1 319-656-4451 (International) E-mail: info@civco.com Internet: www.civco.com
NOTE
Model or part numbers in the following tables are subject to change. Biopsy Guides
5.7.9.5 Center Line
Show or hide Center Line. Press the [MENU]to show or hide Center line.
5.8 CFM Image Adjustment
Parameters in Color mode:
40
Page 54
ECO 5VET Digital Color Doppler Ultrasound System
5.8.1 Color Map
Change the type of Color Map Activate the [Color Map] and rotate [MENU] to adjust Color Map, adjustment range from 0~8
5.8.2 Color Invert
Invert the color of flow
Activate [Color Invert] and rotate [MENU] to turn on or off. Or press to realize Color invert.
5.8.3 Contrast
Adjust the contrast of color Activate [Contrast] and rotate [MENU] to adjust contrast, adjustment range from 0~8
5.8.4 Steering Angle
Adjust Steering Angle. Activate [Steering Angle] and rotate [MENU] to adjust Steering Angle, adjustment range -10/-5/0/5/10.
5.8.5 Wall Thre.
Activate [Wall Thre.] and rotate [MENU] to adjust Wall Thre., adjustment range from 0~14.
5.8.6 Blood Effection
Activate [Blood Effection] and rotate [MENU] to adjust Blood Effection, adjustment range Smooth/Resolution.
5.8.7 Flow
Adjust flow speed.
41
Page 55
ECO 5VET Digital Color Doppler Ultrasound System
Activate [Flow] and rotate [MENU] to adjust flow, adjustment range low/middle/high.
5.8.8 Freq
Adjust frequency of color mode. Press parameter control key on [Freq] to change frequency, adjustment range depends on probe type.
5.8.9 Wall Filter
Adjust Wall Filter. Press parameter control key on [Wall Filter] to change Wall Filter, adjustment range from 0~3.
5.8.10 PRF
Adjust PRF. Activate [PRF] and rotate Menu to Adjust PRF, adjustment range from 0~15.
5.8.11 Persistence
Adjust persistence. Activate persistence and rotate [MENU], adjustment range from 0~7
5.8.12 Baseline
Adjust the position of baseline. Press parameter control key on [Baseline] to change position of baseline, adjustment range from 0~6.
5.9 PW Image Adjustment
Parameters in PW mode:
5.9.1 D 2D Map
Adjust D 2D Map.
42
Page 56
ECO 5VET Digital Color Doppler Ultrasound System
Activate D 2D Map and rotate [MENU] adjustment range custom/1/2/3/4.
5.9.2 Enhance
Adjust Enhance. Activate [Enhance] and rotate [MENU] to adjust Enhance, adjustment range from 0~3.
5.9.3 D Gamma
Adjust D Gamma. Activate D gamma and rotate [MENU] to adjust D Gamma, adjustment range from 0~8.
5.9.4 Dynamic
Adjust Dynamic. Activate [Dynamic] and rotate [MENU] to adjust Dynamic, adjustment range from 0~7.
5.9.5 Pixel Ratio
Adjust Pixel Ratio. Activate [Pixel Ratio] and rotate [MENU] to adjust Pixel Ratio, adjustment range from 0~7.
5.9.6 Color Map
Change type of color map. Activate [Color map] and rotate [MENU] to adjust D color map, adjustment range from 0~8.
5.9.7 Audio
Adjust audio volume. Activate [Audio] and rotate [MENU] to adjust audio volume, adjustment range from 0~15.
5.9.8 Acoustic power
Change acoustic power. Activate [Acoustic power] and rotate [MENU] to adjust acoustic power, adjustment range from 0~10.
5.9.9 Invert
Press parameter control key on Invert, or press key to invert spectrum.
5.9.10 Wall Filter
43
Page 57
ECO 5VET Digital Color Doppler Ultrasound System
Classification
Function Description
Abdomen
Abdomen, general anatomy term
Obstetrics
Anatomy term of Obstetrics
Gynecology
Anatomy term of Gynecology
Heart
Anatomy term of Heart
Small Parts
Anatomy term of Small Parts
Lesion
Lesion term:Abdomen、Obstetrics、Gynecology、Heart、Small
Press parameter control key on [Wall Filter] to adjust wall filter, adjustment range from 0~3.
5.9.11 Speed
Press parameter control key on [Speed] to adjust speed, adjustment range from 0~2.
5.9.12 PRF
Adjust PRF. Press parameter control key on [PRF] to adjust PRF, adjustment range from 0~15.
5.9.13 Baseline
Adjust the position of baseline. Press parameter control key on [Baseline] to adjust the position of baseline, adjustment range from 0~6.
5.10 Full Screen Show (need to activate this function in setup)
Full screen the image area. Press [Gain] knob to activate the function; Press [EXIT] button to exit full screen show. When full screen shows, press [MENU] knob to show the menu of current mode (except bodymark, annotation, measurement).
5.11 Edit Comment
5.11.1 Overview
The comment is to enter text or symbols on the image, the device is Chinese and English annotation system.
ENTER COMMENT: Press enter into comments status;
Exit COMMENT: Press again or [FREEZE] key to exit;
Comment means input the words or symbols on images for making explanation. Add comments can through keyboard input directly or using the default comments. The default comments are classified by examination mode as follows:
44
Page 58
ECO 5VET Digital Color Doppler Ultrasound System
Parts
NOTE: If you need to custom the default comments, refer to the pre-set section.
5.11.2 Input Characters
Operation
1.Press key, then system will go into the comment process.
2.Move the cursor to the position where need to comments.
3.Input characters at cursor position by keyboard then press [ENTER] key to confirm.
4.Press key again to exit. The key will light off and the comments process is finished.
5.11.3 Input Comment Library Characters
1In comment status, move trackball to image area to edit; 2Press [Font size] to adjust front size of comments, the range is 10~20; 3Rotate [MENU] to select needed comments, then press [MENU] to exit ;
5.11.4 Edit Quick Comments
1Press [Edit] to pop quick comment edit box; 2Input customized comments; 3Press [Done] to finish edit, press[X]to cancel edit;
5.11.5 Input Quick Comments
1Press [Text] to select needed quick comments; 2Adjust front size of comments; 3Press [Input] to place comments in the image area;
5.11.6 Move Comments
1In comment status, move trackball to the comment, press [ENTER] key to activate it; 2Move trackball to place the comment to target area; 3Press [ENTER] again to confirm the comment;
5.11.7 Edit Comments
1In comment status, move trackball to the comment, press [ENTER] key to activate it 2Press [BACKSPACE] to delete unnecessary characters; 3Press [ENTER] to confirm;
45
Page 59
ECO 5VET Digital Color Doppler Ultrasound System
5.11.8 Delete Comments
5.11.8.1 Delete Characters
In comment status, activate the comment that need to be deleted, then press the [ENTER] key, it will display "|" on the screen, press [BACKSPACE] key to delete the character.
5.11.8.2 Delete Single Comment
Activate the comment that needs to be deleted, press [DEL] to delete comment;
5.11.8.3 Delete All Contents of the Comment
Don’t activate the single comment, press [DEL] to delete all characters that has input;
Caution: Press [DEL] key, but it will delete the measurement and body mark at the same time;
5.11.9 Set the Position of Default Comment
Operation: 1Press [Save Home Pos.] to move cursor to the initial position; 2Press [Load Home Pos.] to set the initial position ;
5.12 Set Body Mark
5.12.1 General Description
The body mark indicates patient’s examination position and the direction of probe scan on the image. Body marks are divided into: obstetric, abdomen, gynecology, heart and small parts, each has different body mark. Each type of body mark automatically is corresponding to current examination mode. Icon:
Bovine Mark
Canine Mark
Equine Mark
46
Page 60
ECO 5VET Digital Color Doppler Ultrasound System
Feline Mark
Ovine Mark
5.12.2 Body Mark Operation
Operation
1.Press to enter into the body status.
2.Select the body mark that you need.
3.Move trackball after adding the body mark image and then adjust the position of probe. Rotate [MENU] or [ANGLE] knob can adjust the probe direction. Press [ENTER] key to confirm when adjustment was finished.
4.Move the Trackball to change the position of the body mark;
5.If you want to exit from the body mark function, press again;
6.Press [EXIT]to exit body status and the body mark is fastened to the screen.
7.Press [DEL] to delete body marks.
5.13 Set the Direction of Arrow
Operation
1.Press to display arrow;
2.Press [Change] button to select cursor type: arrow or cross. And change the size of the cursor.
3.Adjust the position of probe. Rotate [MENU] or [ANGLE] knob can adjust the probe direction.
4.Press [ENTER] to confirm when adjustment was finished.
5.Press [EXIT] to exit the arrow settings.
6.Press [DEL] to clear the arrows has inputted,
5.14 Image and Cine Disposition
5.14.1 The Principle of Cine Storage
In real image status, the image can be stored in the movie memory in chronological order, maximum frames can be set. The maximum number of frames of the film storage can be set, please refer to pre-set chapter. If the movie memory is full, the recent frame saved into memory, the previous frame removed from memory.
47
Page 61
ECO 5VET Digital Color Doppler Ultrasound System
Cine loop indicate diagram
5.14.2 Manual Loop
Press [FREEZE] to freeze image, pop cine playback bar, at this time, move cursor to play by hand; Track ball rolling to the right, the loop play with ascending order side by side. Or press [Next/Pre] to play.
5.14.3 Automatic Loop
After freezing image, press [Play/Pause] to play, press it again to stop. Press [ENTER] to choose the area of automatic playback area needed.
5.14.4 Save and Recall Image
Press to save current image, the image will be displayed below the screen; If you need to recall images that have been stored, move cursor to needed image, press [ENTER] to recall it;
or you can recall archived patient’s information to recall image, please refer to archive chapter.
5.14.5 Save and Recall Cine
In freeze status, press to save cine, then it will be displayed below the screen, move cursor to needed cine, press [ENTER] to recall cine.
5.14.6 Delete images
After recalling the images , press the [Delete images] to delete the files.
5.14.7 Send images
After recalling the images, press the [Send images] to send images to USB flash disk, DICOM storage and print. Hint: Activate the DICOM before DICOM storage and print.
48
Page 62
ECO 5VET Digital Color Doppler Ultrasound System
5.15 Image Browse
Press to enter image information browsing interface. Press [ENTER] to any function in the image.
Review interface
ID: ID of Current patient. Name: Current patient’s name. Information: Enter into current patient’s information interface. Report: Enter into current patient’s report interface; Send images: Send image to USB hard disk , DICOM storage and print; Print the imageprint the image which be chosen, it will be printed as the arrangement set; Delete images: Delete selected image; Row*Column: Select image’s format; Pre page: Page up; Next page: Page back; New Exam: Exit current examination and open a new dialog box. Continue Exam: Exit image browsing interface and go on checking current patient; Archive: Open up archive management interface; Cancel: Turn off image browsing interface;
5.16 Archive Management
Archive management can search for patient’s information which has been stored in system. Press [Archive] to archive management interface, all process can be opened up by moving cursor.
49
Page 63
ECO 5VET Digital Color Doppler Ultrasound System
Archive interface
Item: Type selection, select Patient’s ID or Name; Keyword: Search for key words; Period: Time filter, select today、one week、one month、three months、 six months, recent one year and all; Multiple Choice: Multiple choice; Dada Source: Path choice, select hard disk or U disk; Patient info: Enter into patient’s information interface; Review Report: Enter into report interface; Backup Exam: Select examination information to USB hard disk; Restore Exam: Recover examination information from USB hard disk; Send Exam: Send selected examination information remotely to USB hard disk or DICOM Storage/Print
(Need to activate the DICOM);
Delete Exam: Delete selected examination information; Patient View: Change display mode of information; Expand All: Select Patient View, it will display sub-directory; Collapse All: Exit sub-directory; Select All: Select all examination information; New Exam: Exit current patient’s examination; Continue Exam: Exit archive management interface and go on checking current patient; Easy View: Exit archive management interface and open up image browsing interface; Cancel: Exit archive management interface and go on checking current patient;
5.17 Report
Press [ENTER] on the image to add the image into the report page. The report can be saved and printed. It is
convenient for the doctor to view and edit the patients’ information.
50
Page 64
ECO 5VET Digital Color Doppler Ultrasound System
Reports contain normal report, abdominal report, cardiac report, small part report etc. Move the cursor to the required report page and press [ENTER] to choose.
Click the key and the page pop up the report page of the current exam mode. Change the other exam mode report by the drop-down box.
Report Interface
ReportTitle: Report options, different kinds of report can choose, such as Normal, OB/GYN etc. Hosp: Display the hospital name. Abdomen Report: Display the kind of report. Patient Name: Display the patient name. Age: Display the patient age. Gender: Display the patient sex. Patient ID: Display the patient ID. Diagnostic: Input the diagnostic instructions. Description: Input the description of symptom. Tips: Input note information. Send DICOM SR: After activating DICOM , send DICOM structural report to server Print: Print the report with image. Export: Export the PDF report to the U disk. Save: Save the report in system. The image on the right side: Press [ENTER] on the image to add the image into the report.
5.18 DICOM
5.18.1 DICOM Worklist
51
Page 65
ECO 5VET Digital Color Doppler Ultrasound System
Press [Worklist] in Patient Interface , pop up the following dialog box
ID: input ID or some characters , fuzzy query needs server Search By: select term , ID or name; ID: display the ID of patients Name: display the names of patients BirthDate: display the birth date of patients Sex: display sex of patients Doctor: display names of doctors Acc#: display the NO. of patients Search: press this button to do search operation Apply: select the searched patient and press this button , input all patient information into the new patient
interface Clear: clear all searched content.
5.18.2 DICOM Storage
Check the save and send in setting, then DICOM storage when saving cine and images. Press send button in archive or freeze interface. DICOM Send interface as follows:
52
Page 66
ECO 5VET Digital Color Doppler Ultrasound System
DICOM Send Interface Select DICOM Storage in left , choose DICOM server and press Export button to DICOM storage. Enter DIOCM Task Sequence and watch or edit DICOM process.
5.18.3 DICOM Print
DICOM Print operation is the same as DICOM storage
5.18.4 DICOM SR
Press Send DICOM SR button in report interface, this task is added into DICOM Task Sequence.
53
Page 67
ECO 5VET Digital Color Doppler Ultrasound System
Chapter 6 Measurement and Calculation
Main content of this chapter: Normal calculation and measurement on B mode image and M mode image, OB calculation and Urology measurement etc, system can enter into corresponding measurement mode depend on current exam mode, and enter into the corresponding report depend on the measurement mode. System has built-in the default measurement according to the exam mode, the change of measurement please refer to the chapter of preset settings.
6.1 Keyboard for Measurement
6.1.1 Trackball
Trackball is used to move the cursor, main functions are as follows:
1.Before starting a measurement, use the trackball to choose the menu options;
2.After starting a measurement, move the trackball to move the cursor, during the measurement, the cursor should not be moved out image area;
3.During the Ellipse method measurement, use trackball to change the length of short axis.
4.Update the moving of the measurement result, move the trackball to change the position of the measurement result.
6.1.2 [ENTER]
During the measurement, the functions of [ENTER] key are as follows:
1.When cursor is on the menu, press the key to choose the options and start the measurement.
2.During the measurement, press the key to anchor the start point and end point.
6.1.3 [UPDATE]
1.Before the measurement, press [UPDATE] to change the measurement method, such as ellipse, trace. The
changeable measurement item has “< >”.
2.During the measurement, [UPDATE] key is used to switch the start point and end point, long axis and short axis when the measurement is not finished.
3.During the distance measurement, press the [ENTER] to fix the start point, when the end point is not fixed, press the [UPDATE] key to switch the start point and end point.
4.During the Ellipse measurement, when fix the long axis, but the short axis is not fixed, press the [UPDATE] key to switch the long axis and short one.
6.1.4 [DEL]
Main functions are as follows: In frozen status, press the [DEL], delete all the measurement results, comments and traces.
54
Page 68
ECO 5VET Digital Color Doppler Ultrasound System
6.1.5 [Change]
Press [Change] to switch other menu; Press [Exit] to exit;
6.1.6 [Exit]
Press [Exit] to exit the measurement menu.
6.1.7 Parameters control key
Press the corresponding key to update the function and use the function.
6.2 B Mode general Measurement methods
The system B mode contains Distance, Ellipse, and Trace.
6.2.1 Meas. Distance
Measurement steps:
1.Press the [Calc] key to enter into measurement. Update the [distance] item in the menu or press the quick measure key [Dist], it will display a segment “+”icon.
2.Move the “+”icon by cursor to fit the one point of the line. Press [ENTER] key to fix the start point and the cursor can be moved to the next position.
3.Press [UPDATE] key can change the activated point, and fit the other point of the line.
4.Move the cursor to the end-point, press [ENTER] again to complete the measurement.
5.After the measurement, the result will display in the measurement results area.
6.Repeat the steps from 1 to 4 to start next “distance” measurement. Press [DEL] key to delete all the measurements.
Note:
Each group of measurement is limited, if the measurement results beyond, it will begin a new group of measurement automatically.
6.2.2 Ellipse
Measurement steps:
1.Press the [Calc] key to enter into measurement. Update the [Ellipse] item in the menu or press the quick measure key [Ellipse], it will display a segment “+”icon.
2.Move the “+”icon by cursor, Press [ENTER] key to fix the point and the cursor can be moved to form a round.
3.Press [UPDATE] key can exchange the activated point and the fixed point.
4.Move the cursor to the end-point of the ellipse, press [ENTER] to fix the axis, at the same time, the next axis be updated, and can change the size of the axis by the cursor.
55
Page 69
ECO 5VET Digital Color Doppler Ultrasound System
5.Now press [UPDATE] key can exit to the step 4.
6.After fixing the next axis, can press [ENTER] key to complete the measurement.
7.After the measurement, the result will display in the measurement results area.
8.Repeat the steps from 1 to 6 to start next “ellipse” measurement. Press [DEL] key to delete all the measurements.
Note:
Each group of measurement is limited, if the measurement results beyond, it will begin a new group of measurement automatically.
6.2.3 Trace
Measurement steps:
1.Press the [Calc] key to enter into measurement. Update the [Trace] item in the menu or press the quick measure key [Trace], it will display a segment “+”icon.
2.Move the “+”icon by cursor, Press [ENTER] key to fix the point and the cursor can be moved to the next position.
3.Make the cursor tracing along the edge of required area, the traced line can be not closed..
4.Now press [UPDATE] key to cancel the tracing.
5.Press [ENTER] key again in the endpoint, the start point and end point of trace line will be closed by a straight line.
6.After the measurement, the result will display in the measurement results area.
7.Repeat the steps from 1 to 5 to start next “trace” measurement. Press [DEL] key to delete all the measurements.
Note:
Each group of measurement is limited, if the measurement results beyond, it will begin a new group of measurement automatically.
6.2.4 Histogram
Histogram is used to calculate the gray distribution of the ultrasound echo signals within a specified area. Use the rectangle, ellipse or trace method to draw along the desired measurement area. The result is shown in the form of histogram. Histogram can be measured only on the frozen image.
Measurement steps by rectangular method:
1.Click [FREEZE] key to freeze the image.
2.Press [ENTER] in [Histogram] menu to enter into measurement status.
3.Click [ENTER] to fix one apex of the rectangle.
4.Move the trace ball to change the cursor position and fix the diagonal point of the rectangle
5.Move the trace ball to change the cursor position, fix the diagonal point of the rectangle, and press [ENTER] again to confirm the measurement area. The result will display on the measurement result area.
Measure the histogram by ellipse or trace method: The method is the same as that to measure y ellipse or trace method, press [UPDATE] to change the measurement between ellipse and trace. The horizontal axis represents the gray scale of the image ranging from 0 to 255. The vertical axis represents the distribution ratio of each gray scale. The value shown on the top of vertical
56
Page 70
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Distance
Distance
cm
Refer to distance Meas.
Ratio(Distance)
Refer to distance Meas. Formula: R=D1/D2
D1: First Distance D2: Second Distance
Angle
deg
Refer to distance Meas.
Angle Range: 0°~ 180°
Area
area/circle
Area cm2 Circle cm
Refer to Ellipse and trace meas.
Ellipse and trace.
Volume
Volume (1straight line)
ml
Refer to distance Meas. Formula: V (π/6)×D3
D means: Depth
Volume (1ellipse)
ml
Refer to ellipse meas. Formula: V (π/6)×A×B2
A: Long Axis B: Short Axis
Volume (2 straight line)
ml
Refer to distance Meas. Formula: V (π/6)×D1×D22
D1: the longer distance D2: the shorter distance Volume
ml
Refer to distance
D1, D2, D3: Distance
axis represents the percentage of the maximally distributed gray in the whole gray distribution.
6.2.5 Cross-section Diagram
Cross-section Diagram is used to measure the gray distribution of the ultrasound signals in the vertical or horizontal direction on a certain profile (section). This measurement is only available in the frozen mode. Measurement steps:
1.Click [FREEZE] key to freeze the image.
2.Click [Meas.] key , and choose [B General Meas.].
3.Draw a straight line at the measuring position. The method is the same as that to measure distance.
4.The calculated result of the profile will be displayed at the centre of the screen. 1-The horizontal (or vertical) axis represents the projection of the profile line on the horizontal direction. 2-The vertical (or horizontal) axis represents the gray distribution of the corresponding points on the profile line. The range is 0 to 255.
6.3 B Fast Measurement
Press [Dist] to enter B Fast measurement in B mode. Press the corresponding parameter control key to switch the fast measurement item.
57
Page 71
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
(3 straight line)
Meas. Formula: V (π/6)×D1×D2×D3
Volume(1 straight line 1 ellipse)
ml
Refer to distance and ellipse Meas. Formula: V (π/6)×A×B×M
A: Long Axis B: Short Axis M: Distance
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Distance
cm
Refer to distance Meas.
area/circle
Area cm2 Circle cm
Refer to Ellipse and trace meas.
Ellipse and trace. Press [UPDATE] to change.
Volume (1straight line)
ml
Refer to distance Meas. Formula: V (π/6)×D3
D means: Depth
Volume (1ellipse)
ml
Refer to ellipse meas. Formula: V (π/6)×A×B2
A: Long Axis B: Short Axis
Volume (2 straight line)
ml
Refer to distance Meas. Formula: V (π/6)×D1×D22
D1: the longer distance D2: the shorter distance
Volume (3 straight line)
ml
Refer to distance Meas. Formula: V (π/6)×D1×D2×D3
D1, D2, D3: Distance
Volume(1 straight line 1 ellipse)
ml
Refer to distance and ellipse Meas. Formula: V (π/6)×A×B×M
A: Long Axis B: Short Axis M: Distance
Ratio
Ratio(distance)
Refer to distance Meas. Formula: R=D1/D2
D1: First Distance D2: Second Distance
6.4 Genearl Measurement
Click display [B], [B/B] or [4B] to enter into B, B/B or 4B mode, then click [Calc] key to enter into measurement status. Or press [Change] to choose the General measurement.
58
Page 72
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Ratio(area)
Refer to ellipse Meas. Formula: R=A1/A2
A1: First Area A2: Second Area
Angle
deg
Refer to distance Meas.
Angle Range: 0°~ 180°
Histogram
Refer to histogram
Cross-section diagram
Refer to Cross-section
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
GS cm
Refer to distance Meas.
CRL
cm
Refer to distance Meas.
HC Diameter
cm
Refer to distance Meas.
Body Length
cm
Refer to distance Meas.
Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×W×H
L: Length W: Width H: Height
Thyroid
Length Width Height
cm
Refer to distance Meas.
L: Thyroid Length W: Thyroid Width H: Thyroid Height
Bladder Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×H×W
L: Bladder Length W: Bladder Width H: Bladder Height
6.5 Canine Measurement
Choose canine exam mode. Press [Calc] to enter into canine measurement status in B, B/B, 4B mode. Or press [Change] to choose the canine measurement.
6.6 Feline Measurement
Choose feline exam mode. Press [Calc] to enter into feline measurement status in B, B/B, 4B mode. Or press [Change] to choose the feline measurement.
59
Page 73
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
GS cm
Refer to distance Meas.
CRL
cm
Refer to distance Meas.
HC Diameter
cm
Refer to distance Meas.
Body Length
cm
Refer to distance Meas.
Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×W×H
L: Length W: Width H: Height
Thyroid
Length Width Height
cm
Refer to distance Meas.
L: Thyroid Length W: Thyroid Width H: Thyroid Height
Bladder Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×H×W
L: Bladder Length W: Bladder Width H: Bladder Height
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
GS cm
Refer to distance Meas.
Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×W×H
L: Length W: Width H: Height
Thyroid
Length Width Height
cm
Refer to distance Meas.
L: Thyroid Length W: Thyroid Width H: Thyroid Height
Bladder Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×H×W
L: Bladder Length W: Bladder Width H: Bladder Height
6.7 Equine Measurement
Choose equine exam mode. Press [Calc] to enter into equine measurement status in B, B/B, 4B mode. Or press [Change] to choose the equine measurement.
60
Page 74
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
BPD
cm
Refer to distance Meas.
CRL
cm
Refer to distance Meas.
Body Length
cm
Refer to distance Meas.
Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×W×H
L: Length W: Width H: Height
Thyroid
Length Width Height
cm
Refer to distance Meas.
L: Thyroid Length W: Thyroid Width H: Thyroid Height
Bladder Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×H×W
L: Bladder Length W: Bladder Width H: Bladder Height
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
BPD
cm
Refer to distance Meas.
CRL
cm
Refer to distance Meas.
Body Length
cm
Refer to distance Meas.
GS cm
Refer to distance Meas.
Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×W×H
L: Length W: Width H: Height
Thyroid
Length
cm
Refer to distance
L: Thyroid Length
6.8 Bovine Measurement
Choose Bovine exam mode. Press [Calc] to enter into bovine measurement status in B, B/B, 4B mode. Or press [Change] to choose the bovine measurement.
6.9 Ovine Measurement
Choose ovine exam mode. Press [Calc] to enter into ovine measurement status in B, B/B, 4B mode. Or press [Change] to choose the ovine measurement.
61
Page 75
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Width Height
Meas.
W: Thyroid Width H: Thyroid Height
Bladder Volume
ml
Refer to distance Meas. Formula: V=(π /6)×L×H×W
L: Bladder Length W: Bladder Width H: Bladder Height
6.10 Genearl Measurement and Calculation in M, B/M mode
Click [M] key to enter M or B/M mode, press [Calc] key to enter into M mode measurement status. Or press [M Measurement] on the menu to enter into M mode measurement status.
6.10.1 Distance
Measurement steps:
1.Select menu item-“Distance” to enter into measurement.
2.Click on the M image area, it will display a blue dotted line with two horizontal short line. The blue dotted line represents the position need to be measured. The distance between the two short lines is the distance you
want to measure. The yellow short line represents it’s in active status. Click it and drag the short line to
anywhere you want to put. .
3.Click [UPDATE] button to active the two short lines in turns and dragon them to change the distance between them. The measurement result will be displayed on the result area.
6.10.2 Time
Measurement steps:
1.Select menu item [Time] to enter into measurement.
2.Click on the M image area, it will display two blue straight dotted line. The blue dotted line with one yellow short line on it represents it is in active status. The distance between the two straight lines stands for time you want to measure. You can drag the active straight line to anywhere you want to change the measured time.
3.Click [UPDATE] button to active the two straight lines in turns and dragon them to change the distance between them. The measurement result will be displayed on the result area .
6.10.3 Heart rate
Heart rate is used to calculate the number of heart beats per minute from cardiac image. Measurement steps:
1.Choose [Heart rate]menu item to enter into measurement.
2.The method is same as M Time measurement.
3.After the above measurement, the calculated heart rate result is displayed in the measurement result area.
4.Repeat the steps form1 to 3 to start next measurement.
62
Page 76
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Distance
cm
Refer to M distance Meas.
Time
s
Refer to M time Meas.
Velocity
cm/s
Refer to M velocity Meas.
HR One Cycle
bpm
Refer to M time Meas.
Double Cycles
bpm
Refer to M time Meas.
6.10.4 Velocity
Measurement step:
1.Choose [Velocity] menu item to enter into measurement.
2.Select the start point of the measurement and click on the screen. The yellow “+” cursor is active. Drag the cursor to the peak systolic wave.
3.Click on the Switch key to active another point. Drag the active point to the end diastolic wave.
4.Repeat 1-3 to do another measurement.
Note: The maximum number of the measurement result on the image area is one. The second
measurement result will cover the first one. The measurement result area will list all the measurement values.
6.11 M General Measurement
At real-time status, click [B/M] key twice to enter M mode, click [Calc] key to enter into M mode cardiology measurement status.
6.12 M Canine Measurement
The same with M mode general measurement.
6.13 M Feline Measurement
The same with M mode general measurement.
6.14 M Equine Measurement
The same with M mode general measurement.
63
Page 77
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Peak
Vs
cm/s
Refer to M speed meas.
Pressure(s)
mmHg
Formula Pressure4 × Vs×Vs/10000
Vd
cm/s
Refer to M speed meas.
Pressure(d)
mmHg
Formula Pressure4 × Vd×Vd/10000
SD
Formula:SD Vs/Vd
RI
Formula:SD (Vs-Vd)/Vs
Time
s
Refer to M time meas.
HR
Single wave
bpm
Refer to M HR meas.
Auto Trace/Manua l Trace
Vs
cm/s
Refer to M speed meas.
Pressure(s)
mmHg
Formula Pressure4 × Vs×Vs/10000
Vd
cm/s
Refer to M speed meas.
Pressure(d)
mmHg
Formula Pressure4 × Vd×Vd/10000
VMean
cm/s
Refer to M
6.15 M Bovine Measurement
The same with M mode general measurement.
6.16 M Ovine Measurement
The same with M mode general measurement.
6.17 PW Fast Measurement
Press [Dist] to enter PW Fast measurement in PW mode. Press the corresponding parameter control key to switch the fast measurement item.
64
Page 78
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment speed meas.
Pressure(VMean)
mmHg
Formula Pressure4 × VMean × VMean/10000
TVI
cm
SD
Formula:SD Vs/Vd
RI
Formula:SD (Vs-Vd)/Vs
PI
Formula:SD (Vs-Vd)/VMea n HR(Single wave)
bpm
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
Velocity
cm/s
Refer to M speed meas.
Distance
cm
Refer to B distance meas.
Peak
Vs
cm/s
Refer to M speed meas.
Pressure(s)
mmHg
Formula Pressure4 × Vs×Vs/10000
Vd
cm/s
Refer to M speed meas.
Pressure(d)
mmHg
Formula Pressure4 × Vd×Vd/10000
SD
Formula:SD Vs/Vd
RI
Formula:SD (Vs-Vd)/Vs
Time
s
Refer to M time meas.
Auto Trace/Manua
Vs
cm/s
Refer to M speed meas.
6.18 PW Genearl Measurement
Press [Calc] to enter PW measurement in PW mode. Press the corresponding parameter control key to switch the general measurement item.
65
Page 79
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
l Trace
Pressure(s)
mmHg
Formula Pressure4 × Vs×Vs/10000
Vd
cm/s
Refer to M speed meas.
Pressure(d)
mmHg
Formula Pressure4 × Vd×Vd/10000
VMean
cm/s
Refer to M speed meas.
Pressure(VMean)
mmHg
Formula Pressure4 × VMean × VMean/10000
TVI
cm
SD
Formula:SD Vs/Vd
RI
Formula:SD (Vs-Vd)/Vs
PI
Formula:SD (Vs-Vd)/VMea n HR(Single wave)
bpm
StD%
Distance1
cm
Refer to B distance meas.
Distance2
cm
Refer to B distance meas.
StD%
%
Formula:StD% ((D1-D2)÷D1) ×100%
D1:Distance1, D2:Distance2
StA%
Area1
cm2
Refer to B ellipse meas.
Area2
cm2
Refer to B ellipse meas.
StA%
%
Formula:StA% ((A1-A2)÷A1) ×100%
A1:Area1,A2:Area2 ICA/CCA
ICA
cm/s
Refer to M speed meas.
Pressure(ICA)
mmHg
Formula Pressure4 ×
66
Page 80
ECO 5VET Digital Color Doppler Ultrasound System
Meas. Menu
Submenu
Unit
Meas. Method/ Meas. Formula
Comment
ICA × ICA/10000
CCA
cm/s
Refer to M speed meas.
Pressure(CCA)
mmHg
Formula Pressure4 × CCA × CCA/10000
ICA/CCA
Formula ICA/CCA
Flow Volume
Diam
cm
Refer to B distance meas.
TVI
cm
Time
s
Refer to M time meas.
HR(Single wave)
bpm
SV
ml
Formula
0.785* Diam* Diam*|TVI|
CO
l/min
Formula SV*HR(Single wave)/1000
6.19 PW Canine Measurement
Refer to PW general measurement.
6.20 PW Feline Measurement
Refer to PW general measurement.
6.21 PW Equine Measurement
Refer to PW general measurement.
6.22 PW Bovine Measurement
Refer to PW general measurement.
67
Page 81
6.23 PW Ovine Measurement
Refer to PW general measurement.
ECO 5VET Digital Color Doppler Ultrasound System
68
Page 82
ECO 5VET Digital Color Doppler Ultrasound System
Function name
Setting method
Function description
Hospital Department
Input freely
Set up the hospital name which is shown at top left corner of “General Setting” dialog box, 20 characters Max. can be input
Date and Time
Input freely
Set up the system date (calendar format), select current date directly. Date format can be changed by format setting.
Date Format
Set up freely
Set up date format: Year/Month/Date, Month/Date/Year, Date/Month/Year
Time zone
Set up freely
Set up the working clock of the system.
Chapter 7 Preset
This chapter introduces the operation to make settings of the system through preset menu at preset mode. Preset function is used to set up working environment and status, parameters of each examination mode. The setting will be stored in the memory of system and not be lost even after the system is switched off. When the system is switched on, it will work automatically with the status which is required by the operator. In preset interface, all operation relies on moving trackball to required Function key position. Press [ENTER] key to start operation.
7.1 General setting
Press the [SETUP] key to enter system setting interface. User can do user-defined setting. Click [X] in the title bar or the exit key on the button can exit the system setting interface.
General Setting Interface
Form 1 General setting
69
Page 83
ECO 5VET Digital Color Doppler Ultrasound System
Language
Select language
Select the language of operation interface(Simplified Chinese English and so on)
Screen shot Type
Select the need type
Set the content which screen picture contain: only picture, image area and patient information, full screen
Frame number choice
Set up freely
Set the default frame when save the film
Options for image area
Click the button to open the setting box
Set the TGC curve, including always show, always hide and hide for 1 to 8 seconds.
Options for Transmitted Images
Click the button to open the setting box
Adjust the parameters of transmitted images: brightness, contrast, gamma.
Options for PC print image
Click the mode you need and open the setting box to choose the different parameter
Including print area and arrangement condition in the image foresee interface
default
Press button
Recover all preset to factory setting
7.2 Measurement
Measurement includes general measurement setting and measurement formula setting
7.2.1 General measurement setting
General settings can only change the display of measurement unit.
General Measurement settings interface
70
Page 84
ECO 5VET Digital Color Doppler Ultrasound System
Distance: cm, mm Area: cm2, mm2 Volume: ml, l Time: s, ms Speed: cm/s, mm/s Acceleration: cm/s2, mm/s2 Angle: deg, rad Weight: g, kg Measure Result Option: whether clear results upon unfreeze, and the automatic froze image Measure Resultthe color of the result font is alternative, including yellow/white/orange/green Follicle Method: ways to measure follicle, you can choose two distances and three distances.
7.2.2 Measurement formula setting
7.2.2.1 Interface Description-Measurement Menu
Interface for Edition of measurement Formula
Region: pull down and select needed measurement menu Manage: pop up measurement software edition interface, add modify delete change marshalling sequence in
measurement menu.
BM: display measurement of each Exam mode Up: press this button to move selected measurement term up Down: press this button to move selected measurement term down Add Folder: add a measurement item. In the left column when the term is fold there is “+” otherwise “-” Add Measurement: add a measurement item for a term in the right column there is selected item and
detailed parameter.
Add Calc Measurementadd a cal item for a measurement term Remove: remove selected measurement term or item.
71
Page 85
ECO 5VET Digital Color Doppler Ultrasound System
Caption
Display names of all items that is displayed name in measurement menu
Var Name
The name of built-in selected measurement menu, user don’t need modify while display order according to the names
Unique Name
Built-in code, user don’t need modify
Measurement rule
None: Disable rule , Repeat: Repeat this item , Sequential: measure by sequence
Default item
After choosing the Repeat and Sequential , choose one measurement or calculation to activate the measurement rule
Caption
Display the name of selected term that is display names in the measurement menu
Var Name
The name of built-in selected measurement menu, user don’t need modify while display order according to the names
Unique Name
Built-in code, user don’t need modify
Display In Menu
Check the required item and it will display on the measurement menu. The item without checking will not display on the measurement menu.
Choose measurement
Check the methods in measurement menu , press [update] to switch the method, Otherwise , the measurement is not
Default: restore all measurement term as factory setting. Save: save measurement item modification users did Check : display needed items on measurement menu, otherwise not displayed
Sheet 2 Folder Edit content description
7.2.2.2 Interface Description-Measurement Manipulation
Interface for Edition of Measurement Formula
Sheet 3 Measurement Edit content description
72
Page 86
ECO 5VET Digital Color Doppler Ultrasound System
methods
available
Display in Report
Check and display the item in measurement menu , otherwise ,the item is not displayed
Name
Measurement operation of specific measurement display in results
Unit
Data unit which measurement operation produces
Display In Report
Whether display in the report or not.
Average Method
The average rule of data
Modify
Press this button to pop up interface to modify measurement operation
Remove
Press this button to delete selected measurement operation
Caption
Display the name of selected term that is display names in the measurement menu
Var Name
The name of built-in selected measurement menu, user don’t need modify while display order according to the names
Unique Name
Built-in code, user don’t need modify
Tool Type/Tool
Select available measurement tool type B distance (B line),area/circumference(ellipse, trace),M distance(M vertical line),time(M horizontal line),M slope(M slant)
Name
Needed measurement operation of specific measurement and
7.2.2.3 Interface Description-Measurement Calculation
Interface for Edition of Measurement Formal
Sheet 4 Calc Measurement Edit content description
73
Page 87
ECO 5VET Digital Color Doppler Ultrasound System
calculation
Unit
Data unit which measurement operation produces
Display In Report
Whether display in the report or not. Up
Press this button to move measurement operation up
Down
Press this button to move measurement operation down
Average Method
The average rule of data
Add
Press this button to pop up interface to add measurement operation
Modify
Press this button to pop up interface to modify measurement operation
Remove
Press this button to delete selected measurement operation
Caption
Display the name of selected term that is display names in the measurement menu
Var Name
The name of built-in selected measurement menu, user don’t need modify while display order according to the names
Unique Name
Built-in code, user don’t need modify
Ruler Result
Needed measurement operation of specific measurement and calculation
Edit
Enter into interface to edit formula when selecting calculation item
7.2.2.4 Create Measurement Operation
Press [Add] in measurement operation interface, pop up the following dialog box
Interface for Edition of Measurement Formula
Sheet 5 Operation content description of creating new measurement
74
Page 88
ECO 5VET Digital Color Doppler Ultrasound System
Maximum
The maximum value displays in result zone and report
Minimum
The minimum value displays in result zone and report
Unit
Data unit which measurement operation produces
Display In Report
Whether display in the report or not. Method
The average rule of data
7.2.2.5 Formula Edit-Normal
It is necessary to enter into the following interface when creating measurement operation except OB
Interface for Edition of Measurement Formula
Formula: edit formula in input box via keyboard and built-in formula. Validate: press this button to check whether the formula is right or not after editing formula. Clear: clear the content in the input box Unit: select the unit of calculation consequence Measurement Items: display all available measurement operation in the measurement menu. Function: built-in formula, number input and some parameters that system needs such as BSA, SPSA.etc Cancel: cancel editing formula and close the interface OK: save edited operation and close the interface
7.2.2.6 Formula Edit-OB
It is necessary to call built-in OB formula sheet when creating OB measurement operation the following function interface is required.
Caution: the results of GA and EDD don’t require unit, the unit of this class has been built-in.
75
Page 89
ECO 5VET Digital Color Doppler Ultrasound System
Measurement Items: display created measurement term by now OB Tables: built-in OB formula table OB List
Measure Item: OB Measurement Item GA Tables: Gestational list for the current measuring project. Growth Table: Growth table for the current measuring. Info: Display the gestational age and fetal weight for the current measuring. Cancel: Cancel the operation of choosing the formula
76
Page 90
Save: Save the users’ choice of formulas.
7.3 Annotation
ECO 5VET Digital Color Doppler Ultrasound System
Annotation Setting Interface
7.3.1 Annotation Library
The annotation database of the system is classified as: abdomen、OB、GYN、Cardiac、small part、Pathological change Annotation can be made by inputting characters from the soft keyboard or recalling the terms saved in annotation database. Press [CommentLib] pull down button, pop up annotation name within system, through trackball and [ENTER] to show required annotation status.
7.3.1.1 Edit Annotation Library
Operation:
1.At the annotation status, Move the cursor to the [Edit Comment Type] button then press [ENTER] key, the annotation will be updated, and can be edited.
2.Input name into the new created annotation status box, Move the cursor to the [Create] button then press [ENTER] key, then create new annotation status and appear in selected annotation status list
3.Move the cursor to the [Delete] button ,press [ENTER] key, then delete current annotation status in the selected annotation list.
4.Alter name of current annotation status list in [Current Type Name]input box, press[ENTER] on the [Rename] button, then rename the selected annotation status name.
77
Page 91
ECO 5VET Digital Color Doppler Ultrasound System
7.3.2 Edit Annotation
Operator uses only current annotation instead of all content annotation status provides, inlay common annotation. If necessary, user can import annotation or self-compiled annotation into common annotation.
7.3.2.1 Add annotation from annotation library
Operation:
1.Select needed source annotation status via Trackball and [ENTER]
2.Select needed annotation at [CommentLib] column then press the [ENTER]key to activate this annotation.
3.Press [ENTER] on [>] button to import selected annotation into user-selected annotation status; press [ENTER] on [>] button to move selected annotation in [Comment Selected] column into source annotation.
4.Press [ENTER] on [>>]button to import all annotation in source into user-selected annotation status; Press [ENTER] on[>>] button to move all annotation in [Comment Selected] column into source annotation.
7.3.2.2 Add annotation manually
Operation:
1.Activate [Edit Comment]input box via Trackball and [ENTER] key, then input needed abbreviation and full name of annotation.
2.Press [ENTER] on [Add] button, meanwhile this handout will be added into source and user-selected annotation status.
7.3.2.3 Alter annotation
Operation:
1.Alter annotation in user-selected status, the abbreviation and full name of annotation will be displayed in [Edit Comment] box.
2.Activate needed abbreviation and full name via [ENTER] and alter via keyboard.
3.Press [ENTER] on [Modify] button, modify the annotation in both source and user-selected status.
7.3.2.4 Delete annotation in library
Operation: Select needed annotation in source status, press [ENTER] on [Delete from Lib] button then the annotation deleted.
7.3.3 Comment and Arrow Option
Optional: whether clear comments and arrows upon unfreeze. Optional: whether freeze upon comments and arrows.
78
Page 92
7.4 Body marks
ECO 5VET Digital Color Doppler Ultrasound System
Interface for Body Mark Setting
7.4.1 Body Marks Library
Built-in body marks: Bovine, Canine, Equine, Feline and Ovine. Press [BodyMarkLib] pull down button, pop up name of built-in body marks, show needed body marks via Trackball and [ENTER] key.
7.4.1.1 Edit body marks library
Operation:
1.Press [ENTER] on [Edit BodyMark Type], pop edit box.
2.Input name into the new created body marks box, Move the cursor to the [Create] button then press [ENTER] key, then the new body marks will be created and appear in selected body marks list
3.Move the cursor to the [Delete] button then press [ENTER] key, then delete current body marks in the selected list.
4.Alter current annotation status list name in [Current Type Name] input box, press [ENTER] on the [Rename] button, then rename the selected body marks.
7.4.2 Body mark edition
Operation:
1.Select needed source body marks via Trackball and [ENTER]
2.Select needed body marks at [BodyMarkLib] column then press the [ENTER] key to activate it.
3.Press [ENTER] on [>] button to import selected body marks into user-selected status; press [ENTER] on [>]
79
Page 93
ECO 5VET Digital Color Doppler Ultrasound System
button to move selected body marks in [BodyMark Selected] column into source body marks.
4.Press [ENTER] on [>>]button to import all body marks in source into user-selected status; press [ENTER] on [>>] button to move all body marks in [BodyMarks Selected] column into source .
7.4.3 Body mark Option
Optional: whether clear bodymark upon unfreeze. Optional: whether freeze upon bodymark.
7.5 Exam Mode
7.5.1 Exam Mode Edit
Press Utility in the submenu of [MENU], the parameters control area shows as following. According to the corresponding control key to turn on or off the function.
Preset: Display the current preset. Rename: Rename the current preset. Load Preset: Load the preset displayed. Save: Save the current preset. Save As: Save the current preset as others.
7.5.2 Exam Mode Selection
Choose Utility and press [MENU], the following interface pops up. Open the relate function by corresponding control area.
80
Page 94
ECO 5VET Digital Color Doppler Ultrasound System
Interface for ExamMode Setting
Probe: select needed probe and ExamMode fit for it. ExamModeLib: show all existed Exam modes. ExamMode Selected: show Exam modes in probe column. >: import selected Exam modes from ExamModeLib column to ExamMode Selected column. >>: import all Exam modes from ExamModeLib column into ExamMode Selected column. <: delete selected Exam modes in ExamMode Selected column. <<: delete all Exam modes in ExamMode Selected column. Set as default: set selected Exam modes in ExamMode Selected column as default Move up: move selected Exam modes in ExamMode Selected column up. Move down: move selected Exam modes ExamMode Selected column down. Copy to: copy the exam mode selected in ExamMode to a specified preset.
7.5.3 Exam mode selection
Operator can define needed Exam mode in detail including annotation, body marks, measurement menu import and export etc.
81
Page 95
ECO 5VET Digital Color Doppler Ultrasound System
Interface for Exam Mode Setting
ExamMode: show all existed Exam modes in system Comment: double click[ENTER]to activate widget box, can choose existed annotation status name. After
setting, the default of the Exam mode is user-selected.
BodyMark: same as Comment, select user-needed default body marks. Measurement: same as Comment, select user-needed default measurement menu. Rename: rename selected Exam mode Delete: delete selected Exam mode Export: export all built-in Exam modes into USB flash disk. Import: import all built-in Exam modes into USB flash disk. Restore: restore all Exam mode as factory setting.
7.6 Keyboard
Users can set the number buttons and the store button on the keyboard, easy to use.
82
Page 96
ECO 5VET Digital Color Doppler Ultrasound System
Interface for Keyboard Setting
Output: including store images, store films, store images to U disk, store film to U disk, etc. function options. Image mode: including image full screen, biopsy, chroma, etc. Print: including video print, pc print etc. Others: including arrow, archive.
7.7 DICOM
DICOM includes DICOM Storage , DICOM Worklist , DICOM Print and DICOM SR . If DICOM is to be applied , please make sure DICOM has been activated . In the system page of setting interface , you can check whether DICOM is open or not . If you want to activate DICOM , please contact with CHISON. There must be DICOM SCP server which has been installed with PACS or other relative DICOM server software
83
Page 97
ECO 5VET Digital Color Doppler Ultrasound System
Local AE Title: input local DICOM Title to separate the DICOM equipments in local network. Service: display the local DICOM function worklist , AE Title: display the name of local DICOM AE title , IP: display the IP of DICOM server, Port: display the port of DICOM server Timeout(s): display the delay time Available: display whether DICOM is default or not , Verify: Press verify button and display whether DICOM setting is correct or not Add: add DICOM function and pop up setting dialog, Delete: delete the existed DICOM function Set as default: set one DICOM service as default , Send while saving: check this item and enable DICOM storage while saving image or cine , send clip or
image according to activated function. Send in archive: send DICOM storage in archive or review , send clip or image according to activated function.
7.7.1 Add/Edit DICOM Function
84
Page 98
ECO 5VET Digital Color Doppler Ultrasound System
Default: check this option , set DICOM function as default; Service: choose DICOM Storage , DICOM Worklist , DICOM Print or DICOM SR; Server AE Title: input DICOM server AE name; Server IP address: input DICOM IP address; SCP port: input DICOM server SCP port; Timeout: set the delay time of DICOM; Tip: choosing the DICOM Print type must be before relative print setting.
7.8 NET Work
Set the unit’s and target unit’s IP and do the connection testing.
7.9 System
7.9.1 System information
Display the software version, hardware version, system version, etc.
7.9.2 Upgrade
Software and Hardware can be updated by USB flash drives. Software upgrade File Path: X:\update\XXX or X:\update_SN\XXX.Hardware update File Path: “X:\fpga_update\XXX”, Keyboard upgrade File Path: “X:\ keyboard_update \XXX”. X means USB flash drives.XXX means upgrade content .It should restart manually after hardware update, and after software update, machine can be restarted automatically.
7.9.3 Function Setting
85
Page 99
ECO 5VET Digital Color Doppler Ultrasound System
DICOM: Click [Open] button, it will bring up the DICOM Key Input dialog box. Input the DICOM SN, and click [OK] button to save and exit. Full screen show: refer to DICOM.
7.9.4 Installment setting
Input relevant key to open trial function and the detail please contact CHISON Company.
7.9.5 Video VGA
Choose the video data: NTSC , PAL-M, PAL-D. Video opened: Choose the item to open this function. VGA opened: Choose the item to open this function.
7.9.6 Image function
Export hardware SN and import hardware secret key, it only for engineer use.
7.9.7 System Maintenance
Only an authorized service engineer may perform maintenance.
7.9.8 USB Video Printer Option
Adjust the parameters of Video Printer OptionDark, Light, Sharpness, Gamma. Select the parameters needed to adjust, press [Confirm] button on the slider of the parameter, and move the trackball to change the parameter.
86
Page 100
ECO 5VET Digital Color Doppler Ultrasound System
Chapter 8 System Maintenance
8.1 Machine Clean
Caution: do turn off the power before cleaning and pull out the cable from socket. There is possibility
of electric shock if the device is on
Clean methods:
Use the soft dry cloth to wrap the machine. If the device is quite dirty, use wet soft cloth. After wiping the blot, use soft dry cloth to wipe dry
Caution:
1.Don’t use organic solvent such s alcohol ,otherwise surface may be ruined
2.When cleaning the machine, don’t let the liquid inflow the machine, otherwise it may malfunction and there is danger of electric shock.
3.When it is necessary to clean the probe connector and peripheral instrument, please contact Sales office contact customer service or agent of CHISON. Any self-cleaning may result in malfunction or degrading the function of device.
8.2 Probe Maintenance
The probes used by this machine can be divided into two series: body surface and intracavity During all ultrasound scan , ultrasonic radiation on the human body should be as less as possible.
Caution:
1.Only person Received professional training can use the probes.
2.Probes can’t receive pressure sterilizer, when operation in sterile area, disposable sterile probe hood should be applied.
3.Make sure not to drop the transducer on hard surface. This can damage the transducer elements and compromise the electrical safety of the transducer.
4.Be careful when operation, make sure not to scratch the probe surface.
5.Avoid kinking or pinching the transducer cable.
6.Make sure not to connect the probe to plug or put adjacent cable into any kind of liquid.
7.Keep the probe clean and dry. Power off or freeze when fixing or dismantling the probe.
8.Make sure not to use or deposit the probe in the environment above 50 degree.
9.If any abnormal phenomena of probe is found, immediately stop operation and contact with Sale Office, Customer Service department or Agents of manufacturer.
Cleaning
The cleaning procedure is fit for all probes. After operation every probe should do cleaning according to stated procedure of this passage. Inspection should be done for intracavity probe depends on condition of use
87