Gima 35111, ABPM PULSE RATE MONITOR WITH BLUETOOTH, ABPM PULSE RATE MONITOR User Manual

Page 1
Contec Medical Systems Co., Ltd
No.112 Qinhuang West Street, Economic & Technical Development Zone, Qinhuangdao, Hebei Province, 066004 PEOPLE'S REPUBLIC OF CHINA
M35111ENRev.1.05.15
Shanghai International Trading Corp. GmbH (Europe)
Eiffestrasse 80, 20537, Hamburg, Germany
0123
PROFESSIONAL MEDICAL PRODUCTS
Gima S.p.A. - Via Marconi, 1 - 20060 Gessate (MI) Italy Italia: tel. 199 400 401 - fax 199 400 403 Export: tel. +39 02 953854209/221/225 - fax +39 02 95380056 gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
ATTENTION: The operators must carefully read and completely understand the present manual before using the product.
GIMA ABPM PULSE RATE MONITOR WITH SpO
2
User Manual
Page 2
2
Page 3
3
Chapter 1 INTRODUCTION ................................................................. 5
1.1 General Information ............................................................ 6
1.2 Button Functions ................................................................ 8
1.3 Interfaces ............................................................................ 8
1.4 Accessories ........................................................................ 9
Chapter 2 GETTING STARTED ......................................................... 10
2.1 Open the Package and Check .......................................... 10
2.2 Dry battery installation ...................................................... 10
2.3 Power on the instrument................................................... 10
2.4 Connect Sensor ................................................................ 11
Chapter 3 Function Interface ............................................................ 11
3.1 Main Interface ................................................................... 11
3.2 Measuring Interface .......................................................... 12
3.3 Measure Result Interface .................................................. 16
3.4 System Menu .................................................................... 17
3.5 ABPM Working Interface................................................... 26
Chapter 4 SpO2 MONITORING ......................................................... 28
4.1 What is the SpO2 Monitoring ............................................ 28
4.2 Precautions during SpO2/Pulse Monitoring ...................... 29
4.3 Monitoring Procedure ....................................................... 29
4.4 Maintenance and Cleaning ............................................... 30
Chapter 5 NIBP MEASURING ........................................................... 30
5.1 Introduction ....................................................................... 30
5.2 NIBP Measuring ................................................................ 31
5.3 Operation Hints ................................................................. 33
5.4 NIBP Error Message and Explanations............................. 35
5.5 Maintenance and Cleaning ............................................... 35
5.6 Transportation and Storage .............................................. 36
5.7 Key And Symbols ............................................................. 37
Chapter 6 INSTALLATION OF THE SOFTWARE .............................. 37
6.1 Demand of editor .............................................................. 37
6.2 Installation of software ...................................................... 38
Chapter 7 INTRODUCTION TO THE SOFTWARE ............................ 38
7.1 Main Interface ................................................................... 38
7.2 Wear ................................................................................. 39
INDEX
Page 4
4
7.3 Set File Path ..................................................................... 39
7.4 Set Collection Project ....................................................... 40
7.5 Download Data ................................................................. 42
7.6 Select the Case to be Edited ............................................ 42
7.7 Delete Case ...................................................................... 43
7.8 Copy Case ........................................................................ 44
7.9 Edit BP Data ..................................................................... 45
7.10 Edit Trend Graph .............................................................. 46
7.11 Edit of Diagnose Information ............................................ 48
7.12 Set BP Thresholds ............................................................ 49
7.13 Histogram ......................................................................... 50
7.14 Pie Chart ........................................................................... 50
7.15 Set Print Report ................................................................ 51
7.16 Print Report Preview ......................................................... 52
7.17 Help .................................................................................. 54
7.18 Exit .................................................................................... 54
Chapter 8 TROUBLESHOOTING GUIDE .......................................... 55
SPECIFICATION ..................................................................................... 56
APPENDIX .......................................................................................... 57
Page 5
5
Chapter 1 INTRODUCTION
For an overall introduction to the Blood Pressure Monitor, please refer to General Information. For basic operating instructions, please refer to Button Function. For allocation of interface sockets, please refer to Interfaces.
Warning Possible explosion hazard if used in the presence of ammable anesthetics or other ammable substance in combination with air, oxygen-enriched environments, or nitrous oxide.
Warning You must verify if the device and accessories can work safely and normally before using.
Warning Ensure that the environment in which the device is operated is not subject to any sources of strong electromagnetic interference, such as radio transmitters, mobile telephones, etc. Keep them far away high level electromagnetic radiation emitted from such devices may greatly affect the instrument performance.
Warning Dispose of the packaging material, observing the applicable waste control regulations and keeping it out of children’s reach.
Warning Please choose the accessories which are approved or manufactured by the manufacturer, or else it may damage the device.
Warning The monitor is only for use on one patient at a time.
Warning When the monitor is wetted, please stop using it and contact us.
Warning If Luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intravascular uid systems, allowing air to be pumped into a blood vessel.
z
z
z
z
zzz
z
Page 6
6
Warning When used with Electro-surgery equipment, you must give top priority to the patient safety.
Warning It is recommended that you check if there is any damage on the monitor or the accessories regularly, if you nd any damage, stop using it, and contact the biomedical engineer of the hospital or our Customer Service immediately. In addition, the overall check of monitor, including the safety check such as the leakage current, should be only performed by qualied personnel once every 12 months.
Note Please choose the computer which should be ensured compliance with the requirements of IEC 60950, or else it may damage the device.
Note The software was developed per IEC60601-1-4. The possibility of hazards arising from errors in the software program is minimized.
Note The monitor shall comply with the standard EN1060 series: Part 1: General requirements; Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems (in course of preparation).
Caution At the end of its service life, the product described in this manual, as well as its accessories, must be disposed of in compliance with the guidelines regulation the disposal of such products. If you have questions concerning disposal of the product, please contact us or its representatives.
1.1 General Information
Environment: Temperature Working 5~40 (˚C)
Transport and Storage -20~55 (˚C) Humidity Working 15%~80% Transport and Storage 95 % Altitude 80kPa~105kPa Power Supply 3 (V) DC P ≤2.4VA
The device has life for 10000 times of BP measure(within 5 years).
z
z
z
z
z
z
Page 7
7
Safety:
This device is debrillator protected. Note that no precautions specic to the device is required during debrillation, and debrillation discharge has no effect on the monitor. The equipment uses the gray silicone airway, in case of the effect to the equipment when debrillation device was used on the patient.
General instruction
The device is applied to Blood Pressure(BP) measure and monitor for adult, pediatric, and neonatal. It most stores 300 records of common user and 350 of ambulatory Blood Pressure data. Every record includes the detailed measure time, systolic blood pressure, diastolic blood pressure, mean blood pressure, pulse rate, error message and record number, etc. This device has friendly operation interface, and adopts 2.4inch color LCD. It integrates data review function and display function which includes large-print single record data review, data list, BP data trends chart, the current time,date,power,alarm and so on. User can power on/off the monitor,start manual measure, set system parameters and so on with ve keys in the front panel. (Please refer to “Button Functions” part for detail). There are sound and light alarm functions that the buzzer intermittently beeping and the red light ashing to prompt low power. When the measure result exceed the alarm limit, the color of the measure results style becomes red and aroses sound alarm. The user can turn on or off the alarm sound if necessary. The cuff socket is located on the top of the device and the USB socket at the bottom of the device. The stored data can be transferred to computer with the USB interface, and then various operations can be performed by using the PC software. (Please refer to “Software Functions” part for detailed contents).
Note
If there is no operation in the common user mode, the device will turn off backlight according to the “BACKLIGHT TIME” you set, and if no action for three minutes, the device will automatically turn off. When the backlight turn off in the Ambulatory Blood Pressure mode, the green indicator intermittently ashes to prompt the device in running state.
z
Page 8
8
1.2 Button Functions
All the operations of the Blood Pressure Monitor could be completed with the buttons. The names of button are on them. They are:
Press the button for a long time, then the system will start. When turning on and off the monitor, the red light and the green both ash once to prompt that the on or off operation is successful. Press it for a short time to return the boot-strap interface.
The text in the middle bottom of the screen indicate the function of this key. Whatever menus the system is in, press the button and the system immediately executes a certain function.
The text in the left bottom of the screen indicate the function of this key. Such as: The button is the alarm switch in the boot-strap interface, up key in the “System Menu”, and left key in the “trend “chart.
The text in the right bottom of the screen indicate the function of this key. Such as: the button is the data review key of current user in the boot-strap interface and down key in the “System Menu” and right key in the “trend” chart.
Start/Stop button. If measuring, press this key to cancel the current measurement
Note
During uploading data with the USB, all of the buttons are disabled. If measuring, the measurement is to be canceled..
Note
During measurement, three buttons are all disabled. The rectangular mark in the screen moving with the operation of ,
buttons is called “cursor”. Operation can be performed in any position at which the cursor can stay. When the item is not selected, the cursor is yellow; when selected, the cursor becomes red.
1.3 Interfaces
For the convenience of operation, different kinds of interfaces are in different parts of the instrument. NIBP cuff socket is at the top.
OK
z
z
OK
Page 9
9
Note Cuff is connected to the monitor through the extended gas pipe.
1 Port of extended gas pipe 2 The socket for Gas pipe
At the bottom is the socket for USB 1 the Socket for USB, connect the data line to upload data
1.4 Accessories
1 A cuff for adult 2 A USB data line 3 A disk (PC software) 4 BP extending tube 5 A pouch
Warning
Please use the special accessories supplied by the manufacturer or replace the accessories according to the requirements of the manufacturer in order to avoid making harms to patients.
1
2
Figure 1 The top external airway
Figure 2 Bottom
1
z
z
Page 10
10
Chapter 2 GETTING STARTED
• Open the package and check
• Dry battery installation
• Power on the instrument
• Connect patient sensors
2.1 Open the Package and Check
Open the package and take out the equipment and accessories carefully. Keep the package material for possible future transportation or storage. Check the components according to the packing list.
• Check for any mechanical damage.
• Check all the cables, modules and accessories.
If there is any problem, contact the distributor immediately.
2.2 Dry battery installation
The instrument will be supplied with two ‘AA’ alkaline batteries or high capacity. Before using the instrument, you shall put the battery in the battery box in the back of the monitor.
Note When you don’t use the equipment, you should take out the dry battery.
2.3 Accensione apparecchio
Press button to power on the instrument. The indicators will ash once, which shows the boot-strap is success, then end pressing, the system will enter into the main interface.
Warning
If any sign of damage is detected, or the instrument displays some error messages, do not use it on any patient. Contact biomedical engineer in the hospital or our Customer Service Center immediately.
Note
Check all the functions that possibly be used and make sure that the equipment is in good status.
z
z
z
Page 11
11
2.4 Connect Sensor
Note
For information on correct connection of NIBP cuff, refer to Figure 3
Connect the sensor between the monitor and the measure part of the patient.
Chapter 3 FUNCTION INTERFACE
• Interfaccia principale
• Menu sistema
3.1 Main Interface
Press to power on the instrument. The indicator will circularly ash once, which show the boot-strap is success, then end pressing, the system will enter into the main interface. In common user mode, if there is no key-press operation during the time which system sets, the device will turn off LCD and enter into standby mode, if there is no any operation in the standby mode, the device will automatically turn off; the “RUN” indicator ashes once every 2 seconds to prompt the device in working state. When the power is low, the battery progress bar is empty, at the same time the alarm sound occurs, and the red indicator ashes in xed time.
In the Main Interface:
Alarm-switch status is displayed in the left top of the screen, button can switch the alarm status shortly. User bar displays the current patient type (adult, pediatric, neonatal), and the amount of the common user’s data record. Current date and time is displayed in the middle top of the screen, the precision of the time is second.
z
Figure 3 Connection method
CUFF
Page 12
12
The boot-strap interface is shown as the follow
Note
All interfaces except the trend retain power icon, alarm switch, as well as a small font of the current time.
Note
After the register overow, the rst record will be overwritten, “Overow” message is shown in the boot-strap interface. The interface is shown as the follow:
3.2 Measuring Interface
Measuring interface displays real-time cuff pressure and the current measurement information. In the measurement process, except the
and the buttons, other buttons are disabled.
Figure 4 Common user boot-strap interface 1
Figure 5 Common user boot-strap interface 2
z
z
Page 13
13
The measuring interface is shown as the follow:
Note
In any interface except the measurement, press key to exit current interface and back to the boot-strap interface.
3.2.1 SpO2 SETUP
Select [PM SpO2 SETUP] in the “SYSTEM MENU”. The following menu will pop up.
z
Figure 6 Measuring interface
Figure 7 SpO
2
setup
Page 14
14
For the [PULSE SOUND] item, you can choose its state. When it is “ON”, pulse sound plays while the monitor measure the ambulatory blood pressure. When it is “OFF”, pulse sound dosen’t play while the monitor measure the ambulatory blood pressure. For the [SpO
2
ALM HI/LO] item, you can set up [SpO2 ALM H] and [SpO2 ALM
LO].Alarm occurs, when SpO
2
value overows.
For the [PR ALM HI/LO] item, you can set up [PRALM HI] and [SpO
2
ALM
LO].Alarm occurs, when PR value overows.
SpO
2
and PR ALARM analysis alarm limits:
3.2.2 Patient Information
Pick the [ABPM INF0] item in the “SYSTEM MENU” to call up the following menu.
The information in this interface can’t be changed, until reinstall the parameter in software.
PARAMETERS Max. HI Min. LO Step
SpO2 100 0 1
PR 254 0 1
Figure 8 ABPM infor
Page 15
15
3.2.3 SpO2 TABLE
Pick the [SpO2 TABLE] item in the “SYSTEM MENU” to call up the following menu.
[total]: the number is the total of all data have been measured, while accessing this interface for the rst time. The sequence number of the rst line currently will be displayed, while changing the data or turning over page. Every page can contain 10 data at most. Select the [change] item to enter the [Change The Number] menu:
When the number which you input is more than the total, it will prompt “>total”. The blood pressure is measured once per 1 s, and the most number of times is 183000.
Figure 9 SpO2 table
Figure 10 Change
Page 16
16
3.2.4 DEMO function
Select the [DEMO] item to enter DEMO status. In the DEMO status, press the “MENU KEY”, and system returns to the “MAIN MENU”.
Note
In clinical application, this function is forbidden because the DEMO will mislead the medical staff to treat the DEMO waveform and parameter as the actual data of the patient, which may result in the delay of treatment or mistreatment.
When it is ambulatory blood pressure state,[PM BP SETUP]、[PM SpO
2
SETUP] and [TABLE] can’t be changed in the “MAIN MENU”.
3.3 Measure Result Interface
The measure result includes: SYS: systolic blood pressure (mmHg/kPa) DIA: diastolic blood pressure (mmHg/kPa) PR: pulse rate (bpm)
Figure 11 Demo
Figure 12
Measure results interface 1
z
Page 17
17
If there is an error during the measurement, an error message text will ap­pear on the screen. If the ALARM SOUND is set to be on, the sound would occur. Press the SILENCE key to stop the sound and press it once more to continue.
3.4 System Menu
In the boot-strap interface, according to the text in the middle bottom of the screen, press button, then enter the system menu and execute different option operations by using keys.
3.4.1 System setup
Enter “SYSTEM SETUP” item in the [SYSTEM MENU], the following menu will pop up:
Figure 14 System menu
Figure 13 Measure result interface 2
OK
Page 18
18
“SYSTEM SETUP” menu includes: “
LANGUAGE” item has two choices: Chinese, English;
UNIT” item has two choice: mmHg, kPa;
USER PURVIEW” item cannot be operated: all;
MEASURE MODE” item has three options: adult, pediatric, neonatal,
BACKLIGHT TIME(s)”item has four choices: 15, 30, 60, 120.
Note
“Backlight Time” in the “System Settings” is used by the common user, backlight time in ambulatory blood pressure monitor (ABPM) is a xed value of 5S.
Select “ABPM Setup” item in [System Setup] menu, the following menu will pop up:
Figure 15 System setup
Figure 16 ABPM Setup
z
Page 19
19
“AWAKE INTERVAL” (min/h) could be options: 15, 30, 60, 120; “
ASLEEP INTERVAL” (min/h) could be options: 15, 30, 60, 120, 240;
AWAKE TIME” and “ASLEEP TIME” transformation unit are 30 minutes.
3.4.2 System Time
Select “System Time” item in [System Menu], the following menu will pop up:
Select “Conrm” after time setup complete, time change is successful and exit system time setup and return the previous menu. Select “Exit” to cancel the setting and return to the previous menu.
3.4.3 Data Delete
Select “Yes” in “Data Delete” menu of [SYSTEM MENU], after you press certain key, the following menu will pop up:
If press “Conrm”, the common user data will be daleted, if press “Exit”, the operation will be cancelled.
Figure 17
System time
Figure 18
Data delete
Page 20
20
3.4.4 Alarm Setup
Select “ALARM SOUND” item in [SYSTEM MENU] to switch on or off the sound of alarm. [refer to Figure 19 system menu]. Select “ALARM SETUP” item in [SYSTEM MENU], the following menu will pop up:
“SYS ALARM” and “DIA ALARM” can control closing or unsealing of the SYS and DIA alarm separately. The alarm is on or off according to the high and low limits which have been set up. When the measure result is higher than the high limit or lower than the low limit, and meanwhile, the “ALARM SOUND” is on, “SYS ALARM” or “DIA ALARM” accordingly on, the alarm will occur.
The adjustable ranges of the high and low limits of the adult mode alarm are as follows: SYS ALM: 40~270 mmHg DIA ALM: 10~215 mmHg
The adjustable ranges of the high and low limits of the pediatric mode alarm are as follows: SYS ALM: 40~200 mmHg DIA ALM: 10~150 mmHg
The adjustable ranges of the high and low limits of the neonatal mode alarm are as follows: SYS ALM: 40~135 mmHg DIA ALM: 10~100 mmHg
Figure 19 Alarm setup
Page 21
21
“Default” includes the main content: Measure mode: adult; Alarm parameter:
Alarm switch: close; Measure unit: mmHg; Ordinary user backlight time: 15s; ABPM switch: close; Asleep time: 22:00; Asleep measurement interval: 30 minutes; Awake measurement interval: 15 minutes; Awake time: 7:00; SYS Alarm switch: close; DIA Alarm switch: close.
3.4.5 ABPM Menu
Select “ABPM” menu in [System Menu], the following menu will pop up.
Use
mode
Adult Pediatric Neonatal
High pressure
high limit
140 120 100
High pressure
low limit
90 70 60
Low pressure
high limit
90 70 60
Low pressure
low limit
40 40 20
Figure 20 ABPM
Page 22
22
In “ABPM” menu, “ABPM ON-OFF” select “ON”, then prompt message of ABPM of the current user, such as:
Press button, clear the ambulatory blood pressure measure data, and enter ambulatory Blood Pressure Mode;
Press button, save ambulatory Blood Pressure measure data, and enter ambulatory Blood Pressure Mode;
Press button, give up the choice, and return the previous menu.
Figure 21 ABPM prompt menu
OK
Page 23
23
3.4.6 ABPM Data Review
3.4.6.1 ABPM “large-print” Data Review
Select “ABPM Data” item in “ABPM” menu, the following interface will pop up. Every record is an interface, and display content include: the current user, total of the current user record data, serial number of the record, stored time of the record, high pressure, low pressure, mean pressure, Pulse Rate.
3.4.6.2 ABPM Data “Table”
Press button to select “trend” in ABPM large-print data review menu, the following interface will pop up
Figure 22 ABPM “large-print”
OK
Figure 23 ABPM “Table”
Page 24
24
A interface contain 5 records, every record includes time, high pressure, low pressure, mean pressure, Pulse Rate.
3.4.6.3 ABPM Trend
Press button to select “trend”, in ABPM large-print data review menu, the following interface will pop up:
Trend interface can trace 100 record trends, if measuring data are more than 100 items, press , buttons can glide trend curve for left and right,
the scale of the vertical axis and the starting point, end point automatically adjust according to the width of the stored data. The displayed dates that are at the bottom of trends shown the trend for the rst point and last point of data recording time.
Figure 24 ABPM “trend”
OK
Page 25
25
3.4.7 Ordinary User Data Review
3.4.7.1 Ordinary User “large-print” Data Review
Press button to enter the ordinary user “large-print” data review in boot-strap interface, the following interface will pop up:
Display content is similar to ambulatory blood pressure large-print data review.
3.4.7.2 Ordinary User Data Table
Press button to pop up the ordinary user data table in the ordinary user large-print data review, the interface is as shown:
OK
Figure 26 Ordinary user data table
Figure 25 Ordinary user “large-print”
Page 26
26
Display content is similar to ambulatory blood pressure large-print data review.
3.4.7.3 Ordinary User Measure Data Trends
Press button to pop up the ordinary user data table in the ordinary user data table, the interface is as shown:
Display content is similar to ambulatory blood pressure large-print data review.
3.5 ABPM Working Interface
In ABPM working environment, backlight is only for 5 seconds, except the press any key to wake the backlight, ABPM working interface is as shown:
OK
Figure 27 Trends
Figure 28 ABPM working interface
Page 27
27
If alarm sound occurs, press SILENCE key to stop it and press it again to continue.
Long press button, the exit ABPM hint interface will pop up:
Press button to exit ABPM working environment, and enter the ordinary user working environment, and display the boot-strap interface;
Press button to exit the interface, and return the ABPM working interface.
Long press button, turn off device hint interface will pop up :
Press button to turn off the device.
Press button to exit the interface, and return the ABPM working interface.
Figure 30 Turn Off Device Prompt Interface
OK
Figure 29 ABPM exiting prompt interface
Page 28
28
Chapter 4 SpO2 MONITORING
4.1 What is the SpO2 Monitoring
SpO2 Plethysmogram measurement is employed to determine the oxygen saturation of hemoglobin in the arterial blood. If, for example, 97% hemoglobin molecules in the red blood cells of the arterial blood combine with oxygen, then the blood has a SpO
2
oxygen saturation of 97%.
The SpO
2
numeric on the monitor will read 97%. The SpO2 numeric shows the percentage of hemoglobin molecules which have combined with oxygen molecules to form oxyhemoglobin. The SpO
2
/PLETH parameter can also
provide a pulse rate signal and a plethysmogram wave.
How the SpO
2
/ PLETH Parameter Works
• Arterial oxygen saturation is measured by a method called pulse oximeter.
It is a continuous, non-invasive method based on the different absorption spectra of reduced hemoglobin and oxyhemoglobin. It measures how much light, sent from light sources on one side of the sensor, is transmitted through patient tissue (such as a nger or an ear), to a receiver on the other side. The sensor measurement wavelengths are nominally 660nm for the Red LED and 940nm for Infrared LED. Maximum optical power output for LED is 4 mW.
• The amount of light transmitted depends on many factors, most of which
are constant. However, one of these factors, the blood ow in the arteries, varies with time, because it is pulsating. By measuring the light absorption during a pulsation, it is possible to derive the oxygen saturation of the arterial blood. Detecting the pulsation gives a PLETH waveform and pulse rate signal.
• The SpO2 value and the PLETH waveform can be displayed in the main
screen.
Warning ES (Electrosurgery) equipment wire and SpO
2
cable must not be
tangled up.
Warning
Do not put the sensor on extremities with arterial catheter or venous syringe.
z
z
Page 29
29
Note
Do not perform SpO2 measuring and NIBP measuring on same arm at one time, because obstruction of blood ow during NIBP measuring may adversely affect the reading of SpO
2
value.
4.2 Precautions during SpO2/Pulse Monitoring
Note
、 Make sure the nail covers the light window; 、 The wire should be on the backside of the hand.
Note
SpO
2
waveform is not proportional to the pulse volume.
Warning
Do not use the SpO
2
sensor once the package or the sensor is found
damaged. Instead, you shall return it to the vendor.
Warning
Prolonged and continuous monitoring may increase jeopardy of unexpected change of dermal condition such as abnormal sensitivity, rubescence, vesicle, repressive putrescence, and so on. It is especially important to check the sensor placement of neonate and patient of poor perfusion or immature dermogram by light collimation and proper attaching strictly according to changes of the skin. Check per 2~3 hours the sensor placement and move it when the skin deteriorates. More frequent examinations may be required for different patients.
4.3 Monitoring Procedure
SpO2 plethysmogram measurement
1.
Switch on the monitor.
2. Attach the sensor to the appropriate site of the patient nger.
3. Plug the connector of the sensor extension cable into the SpO2 socket,
please pay attention to the direction of the sensor.
4. Please pull out the sensor When measure is over.
z
z
z
z
z
Page 30
30
4.4 Maintenance and Cleaning
Care and Cleaning
Warning
Turn off the monitor before cleaning the monitor or the sensor
Warning
Do not subject the sensor to autoclaving. Do not immerse the sensor into any liquid. Do not use any sensor or cable that may be damaged or deteriorated.
Cleaning:
• Use a cotton ball or a soft mull moistened with hospital-grade ethanol
to wipe the surface of the sensor, and then dry it with a cloth. This cleaning method can also be applied to the luminotron and receiving unit.
• The cable can be cleaned with 3% hydrogen dioxide, 70% isopropanol,
or other active reagent. However, connector of the sensor shall not be subjected to such solution.
Chapter 5 NIBP MEASURING
5.1 Introduction
• The Non-invasive Blood Pressure (NIBP) module measures
the blood pressure using the oscillometric method.
• There are two modes of measurement available: manual, automatic.
• Every mode displays the systolic, mean, diastolic blood pressure
and pulse rate.
• It is applicable for adult, pediatric, and neonatal usage.
Figure 31 Mounting of the sensor
z
z
Page 31
31
Warning
1. You must not perform NIBP measurements on patients with
sickle-cell disease or under any condition which the skin is damaged or expected to be damaged.
2. For a thrombasthemia patient, it is important to determine whether measurement of the blood pressure shall be done automatically. The determination should be based on the clinical evaluation.
3. Ensure that the correct mode setting is selected when performing measurements on children and newborn (Refer to measuring mode menu setting), and use children and newborn special cuff. It may be dangerous for using error patient mode, because the upper adult blood pressure level does not apply the children and newborn.
5.2 NIBP Measuring
Warning
• Before starting a measurement, verify that you have selected
a setting appropriate for your patient (adult, pediatric or neonate).
• Do not apply the cuff to a limb that has an intravenous infusion
or catheter in place. This could cause tissue damage around the catheter when infusion is slowed or blocked during cuff ination.
Warning
Make sure that the air conduit connecting the blood pressure cuff and the equipment is neither blocked nor tangled.
1. Plug in the air hose and switch on the system.
2.
Apply the blood pressure cuff to the patient’s upper arm following
the instructions below (Figure 5.1).
• Ensure that the cuff is completely deated.
• Apply the appropriate size cuff to the patient, and make sure that
the symbol “Φ” is over the appropriate artery. Ensure that the cuff is not wrapped too tightly around the limb. Excessive tightness may cause discoloration and eventual ischemia of the extremities.
z
z
z
Figure 32 Use cuff
Page 32
32
Nota
The width of the cuff should be either 40% of the limb circumference (50% for neonates) or 2/3 of the upper arm length. The inatable part of the cuff should be long enough to encircle 50-80% of the limb. The wrong size of cuff can cause erroneous readings. If the cuff size is in question, then use a larger cuff.
Size of reusable cuff for neonate/children/adult
Size of disposable cuff for neonate
• Make sure that the cuff edge falls within the range of mark <->.
If it does not, use a larger or smaller cuff that ts better.
3. Connect the cuff to the air hose. The limb chosen for taking the measurement should be placed at the same level as the patient’s heart. If this is not possible you should apply the following corrections to the measured values:
• If the cuff is placed higher than the heart level, add 0.75 mmHg (0.10 kPa)
for each inch of difference.
• If it is placed lower than the heart level, deduct 0.75 mmHg (0.10 kPa)
for each inch of difference.
4. Check whether the measure mode is appropriately selected. (the measure mode displays in the boot-strap interface information area).
5. Press button on the front panel, and start a measurement.
Patient type Limb perimeter Cuff width
Infant 10 ~19 cm 8 cm Child 18 ~ 26 cm 10.6 cm Adult1 25 ~ 35 cm 14 cm Adult2 33 ~ 47 cm 17 cm thigh 46 ~ 66 cm 21 cm
z
Hose
1.5 m or 3 m
Size No. Limb perimeter Cuff width
1 3.1 ~ 5.7 cm 2.5 cm 2 4.3 ~ 8.0 cm 3.2 cm
3 5.8 ~ 10.9 cm 4.3 cm 4 7.1 ~ 13.1 cm 5.1 cm
Hose
1.5 m or 3 m
Page 33
33
5.3 Operation Hints
1. To start auto measuring:
Access ABPM SETUP menu and pick the “Asleep Interval” item and “Awake Interval” item, in which the user may select the time interval value for auto measurement. After that, enter “ABPM” menu and select the entrance, enter into ABPM working environment, and the system is automatically inated measurement in accordance with the setting time interval.
Warning
Prolonged non-invasive blood pressure measurements in Auto mode may be associated with purport, ischemia and neuropathy in the limb wearing the cuff. When monitoring a patient, examine the extremities of the limb frequently for normal color, warmth and sensitivity. If any abnormality is observed, stop the blood pressure measurements.
2. To stop auto measuring:
During auto measuring, press button at any time to stop auto measurement.
3. To start a manual measuring:
• Press button to start a manual measuring in the ordinary user working
environment.
• During the idle period of auto measuring process, press button at any
time to start a manual measurement. Then press button to stop manual measurement and the system continues executing auto-measuring
program.
Note
If you are in doubt about the accuracy of any reading(s), check the patient’s vital signs by an alternative method before checking the functioning of the blood pressure monitor.
Warning
If liquid is inadvertently splashed on the equipment or its accessories, or may enter the conduit or inside the monitor, contact local Customer Service Center.
z
z
z
Page 34
34
Measurement Limitations
To different patient conditions, the oscillometric measurement has certain limitations. The measurement is in search of regular arterial pressure pulse. In those circumstances when the patient’s condition makes it difcult to detect, the measurement becomes unreliable and measuring time increases. The user should be aware that the following conditions could interfere with the measurement, making the measurement unreliable or longer to derive. In some cases, the patient’s condition will make a measurement impossible.
• PatientMovement
Measurements will be unreliable or cannot perform if the patient is moving, shivering or having convulsions. These motions may interfere with the detection of the arterial pressure pulses. In addition, the measurement time will be prolonged.
• CardiacArrhythmia’s
Measurements will be unreliable and may not be possible if the patient’s cardiac arrhythmia has caused an irregular heartbeat. The measuring time thus will be prolonged.
• Heart-lungMachine
Measurements will not be possible if the patient is connected to a heart-lung machine.
• PressureChanges
Measurements will be unreliable and may not be possible if the patient’s blood pressure is changing rapidly over the period of time during which the arterial pressure pulses are being analyzed to obtain the measurement.
• SevereShock
If the patient is in severe shock or hypothermia, measurements will be unreliable since reduced blood ow to the peripheries will cause reduced pulsation of the arteries.
• HeartRateExtremes
Measurements cannot be made at a heart rate of less than 40 bpm and greater than 240 bpm.
• RoundPatient
The thick fat layer of body will reduce the measurement accuracy, because the fat that come from the shock of arteries cannot access the cuffs due to the damping.
Page 35
35
5.4 NIBP Error Message and Explanations
5.5 Maintenance and Cleaning
Warning
• Do not squeeze the rubber tube on the cuff.
• Do not allow liquid to enter the connector socket at the front
of the equipment.
• Do not wipe the inner part of the connector socket when cleaning
the equipment.
ReusableBloodPressureCuff
The cuff can be sterilized by means of conventional autoclaving, gas, or radiation sterilization in hot air ovens or disinfected by immersion in decontamination solutions, but remember to remove the rubber bag if you use this method. The cuff should not be dry-cleaned. The cuff can also be machine-washed or hand-washed, the latter method may prolong the service life of the cuff. Before washing, remove the latex rubber bag, and for machine-washing, close the Velcro fastening. Allow the cuff to dry thoroughly after washing, then reinsert the rubber bag.
Explanation Cause
Self-test failure A/D sampling error
Loose cuff Cuff is not connected correctly Air leakage Air leakage in the valve or airway Atmospheric pressure error Valve cannot be open. Signal is too weak Object measuring the pulse is too weak or the cuff
is loose It is over the range Object measuring blood pressure is over the
measurement range Excessive movement When measuring, signal the presence of excessive
movement or pseudo-differential interference Overpressure Cuff pressure is over the scope, ADU 300 mmHg,
Newborn: 150mmHg Saturated signal Movement or other factors lead to too big signal
amplitude Air leakage There is air leakage in the airway System failure There is something wrong with NIBP module, A/D
sampling or soft of system after turning on the device It spends too much time Adult: When cuff pressure is 2KPa (15mmHg,)
it may spend 3 minute (180S) Newborn: When cuff pressure is 0.67KPa (5mmHg,)
it may spend 90S
z
Page 36
36
To replace the rubber bag in the cuff, rst place the bag on top of the cuff so that the rubber tubes line up with the large opening on the long side of the cuff. Now roll the bag lengthwise and insert it into the opening on the long side of the cuff. Hold the tubes and the cuff and shake the complete cuff until the bag is in position. Thread the rubber tubes from inside the cuff, and out through the small hole under the internal ap.
DisposableBloodPressureCuffs
Disposable cuffs are intended for one-patient use only. Do not use the same cuff on any other patient. Do not sterilize or use autoclave on disposable cuffs. Disposable cuffs can be cleaned using soap solution to prevent infection.
Note
For protecting environment, the disposable blood pressure cuffs must be recycled or disposed properly.
5.6 Transportation and Storage
A The packed device can be transported by ordinary conveyance or according to transport contract. The device cannot be transported mixed with toxic, harmful, corrosive material. B The packed device should be stored in room with no corrosive gases and good ventilation. Temperature: -20°C~55°C; Humidity: ≤95%.
Figure 33 Replace Rubber Bag in Cuff
z
Page 37
37
5.7 Key And Symbols
Chapter 6 INSTALLATION OF THE SOFTWARE
6.1 Demand of editor
Processor: Pentium IV 1.8G or more Operation System:Windows XP EMS memory: 256M and more Hard Disk: 40G or more Display: 17 inch or more CD-ROM USB: 2 or more Resolution of printer: 600 DPI
Symbols Description
Warning – See User Manual
SYS Systolic pressure MAP MAP pressure DIA Diastolic pressure PR Pulse rate (bpm) ADU Adult PED Pediatric NEO Neonatal ABPM Ambulatory Blood Pressure Monitor INFO Information
Open the alarm sound indication
Close the alarm sound indication
WEEE (2002/96/EC)
Type BF debrillator proofed applied parts
SN Serial number IPX0 Ingress of liquids rank This item is compliant with Medical Device Directive 93/42/EEC of June 14, 1993, a directive of the European Economic Community
European Representative
z
0123
Page 38
38
6.2 Installation of software
1 Place the CD-ROM in the CD-ROM compartment located on your computer. 2 If Auto Play for CDs is enabled, place CD in reader and follow instructions when they appear in the screen; otherwise follow install instructions below: a. Open Windows Explorer. b. Click on the root CD-ROM directory. c. Double click le Setup.exe. d. Follow the instructions in the screen.
Chapter 7 INTRODUCTION TO THE SOFTWARE
7.1 Main interface
Double click software by left key of the mouse to enter the main operation interface as follows:
1 Menu bar: main operation menu of the software 2 Toolbar: shortcut operation button of frequent function 3 The middle area is trend graph display area: After choosing the case
that would be edited, display the data trend graph 4 Status bar: after choosing the case that would be edited, display patient name, ID and data collection date.
Figure 34 Main operating interface
1
2
3
4
Page 39
39
7.2 Wear
Click shortcut key , and the “Lead PlaceMent” dialog box will appear..
You can wear the device accroding to the picture above. Please read the “Matters need” before use.
7.3 Set File Path
Downloaded case data will be recorded in the case path.If you want to change the case path, you can click menu item “Set le path”. After the setting dialog box appears, you can change the case path. Display as follows:
Figure 36 Set le path
Figure 35 Wear
Page 40
40
Patient le path: choose patient le path.After downloading data from t he monitor, the case le is saved in this path. If you choose “ Always use default path “ item, the default case le path is program installation le path.
7.4 Set Collection Project
Click shortcut key , or click menu bar item, and the “ Upload parameters “ dialog box will appear:
As above gure, the doctor could set parameter according to the patient status and diagnosis requirement, then the monitor could nish the collection according to the setting.Parameter explanation is as follows:
PatientName: the patient’s name Patient ID: the patient’s ID number.It is used for marking patient,
and it is exclusive in order to avoid homonymy patient arisen state
CurrentTime: Current system display time TimePeriods: AwakeTime: the patient is awake state AsleepTime: the patient is asleep state Interval: collection Interval.In order to reduce the effect to the patient sleep,
the asleep collection interval should be longer. For example as above gure:the awake time area is 7:30-22:30, and the asleep time area is 22:30-tomorrow7:30. The awake collection
Figure 37 Set Collection Parameter
Page 41
41
interval is 30 minutes, and the asleep collection interval is 60 minute. The asleep time area and awake time area will be display at the right side. When the parameter setting is nished, click “ Okay “ to upload the project to the monitor. The user could select connection way: USB or WIRELESS(the device without WIRELESS connection way).When uploading, the choice dialog box will appear to prompt patient to select connection way.If you pitch on “do not show this next time”, this dialog box will not appear when uploading or downloading, and upload according to the connection way the use has chosen.
If the user uses WIRELESS to upload, the dialog box as follows will appear. Click “Scan” to begin to search.Select device, then click “connect” to download.
Figure 38 Set Connection Way
Figure 39 Scan Vicinal WIRELESS Equipment
Page 42
42
Uploading schedule is as follows:
7.5 Download Data
Before downloading data, please insure:
1. The monitor is connected with computer accurately.
2. The monitor is on.
3. Ensure that the monitor isn’t connect with patient when the monitor is connected with computer.
Click shortcut key or menu item “Download”, then select connection way. After right connecting, begin to download data.
The schedule display is as follows:
After nishing downloading, prompt dialog box will appear, and the case data will be saved in the case le path.
7.6 Select the Case to be Edited
Before editing case, rst choose the case to be edited.Click shortcut key or menu item “open data” to enter “Select le to open” interface. Display as follows:
Figure 40 Uploading Schedule
Figure 42 Select Case
Figure 41 Downloading Schedule
Page 43
43
In this interface, you can choose the le to save case. If the le has case, the le information will be displayed with list form, including these information:Patient Name,Patient ID, Start Time and File Name.Select the case le to be edited, and click “ Okay “ or double click to open the case le and edit it.
7.7 Delete Case
If you are sure that some cases don’t be needed, you can delete then. Click menu item “Delete data”, and deleting case interface will appear which is similar to the “Select le to open” interface.Display as follows:
You can delete individual le, also can delete multile. If you want to delete multile, press “Ctrl” key, at the same time click the le to be delete. After selecting, click “Okay”, and prompt dialog box will appear. Click “Yes” to nish deleting operation, and click “ No” to cancel deleting operation.
Figure 43 Select File to Delete
Page 44
44
7.8 Copy Case
Sometimes before you edit case, you may want to save a copy of originality data.Here you need to copy the case. Our software supplies this function. Choose menu item “ Copy data “, and the interface will appear as follows:
The copying case method is similar to the deleting le method.You can select multile at the same time.After selecting the le, click “Okay” , and “Select destination directory” will appear. You can set destination director path. After setting path, click “Okay” to nish operation. Display as follows:
Figure 44 Select File to Copy
Figure 45 Set Destination Director Path
Page 45
45
7.9 Edit BP Data
You can edit every piece of data.Click shortcut key or select menu item “Edit” to display “BP Data” interface as follows:
In the interface, you can see the particular information of every data. *=0/192(0.0%):192 is total number of data, 0 is the number of deleted data,
0.0% is the percent of deleted data. Number: data collection serial number Time: collection time Date: collection data BP(mmHg): the number in front of “/” is high pressure value, and the number in back of “/” is low pressure value.The unit is mmHg. PR(BPM): pulse rate value.The unit is BPM. MAP(mmHg): Mean pressure.The unit is BPM PP(mmHg): Pressure difference between high and low blood pressure. The unit is BPM TC: error code (refer to chapter 7) Comment: comment for the data. You can also edit the data. The sign “*” stands for the deleted data (the trend graph doesn’t display this data, and statistic data doesn’t include this data). You can delete or add “*” sign by click left key of mouse in “*” area. In comment area, double click any item to edit data and add comment. The comment will be displayed in the trend graph and saved in current case le.
Figure 46 Editing BP Date Interface
Page 46
46
7.10 Edit Trend Graph
After selecting case le, the BP trend graph will be shown in the screen
automatically. In other interfaces you can click shortcut key to enter the BP trend interface .The “BP Trend” has two graph types: color ller type,
dotted line type. You can switch from one graph type to another by push the “Graph Type” button. As color ller type trend graph shown, the blue line is PR change curve, and the green area is the data area below threshold, and the red area is the data area above threshold. The display effect of color ller type trend graph is as follows:
As dotted line type trend graph shown, the blue line is PR change curve, and the yellow line is the high-low pressure threshold line, and the gray uprightness line is high-low pressure linked line of BP data for one time. The dotted line type trend graph is as follows:
Figure 47 Color Filler Type BP Trend Graph
Figure 48 Dotted Line Type BP Trend Graph
Page 47
47
You can switch from one graph type to another by the “Graph Type” button. When you move the mouse on the trend graph area, on the top of the screen the detail data information about the mouse points will be shown, including the data serial number, collection time and collection date, high/low blood pressure value, pulse rate, comment, etc. Click left button of the mouse to delete or add the data point to be shown. If the distance between the two data points is too short to move the mouse to one of the points, move the mouse to the time axis area, press left button of the mouse down, dragging the mouse rightwards to stretch the BP trend graph. Of course you can also drag the mouse leftwards to shrink the trend graph. When you stretch the trend graph, if the trend graph length is beyond the trend graph, the display area will appear a horizontal scroll bar, you can see any part of the trend graph by changing position of the scroll bar.Move the mouse to the y-axis area, press left button of the mouse down, dragging the mouse upwards to stretch the trend graph. Of course you can also drag the mouse downwards to shrink the trend graph.
Figure 49 Shrink Trend Graph(Time Axis)
Figure 50 Stretch Trend Graph (Time Axis)
Page 48
48
7.11 Edit of Diagnose Information
Diagnose information includes following items: Patient information, Current Medications, Diagnose Information and Physician information. Select “Patient information”, and display as follows:
Figure 51 Stretch Trend Graph (Y-Axis)
Figure 52 Shrink Trend Graph (Y-Axis)
Figure 53
Edit Patient’s Information
Page 49
49
Click “OK”, and the patient information content will be saved in the case le in order to expediently display next time.At the same time all the information will be displayed in the report.
7.12 Set BP Thresholds
After BP Thresholds being set up, the trend graph will renew automatically, statistic data will be recalculated. From the view select “Edit” and select “Thresholds”, then the following dialog box will appear.
The default recommended thresholds for calculating Blood Pressure Load are 140/90 for awake periods (07:00-22:00) and 120/80for asleep periods (22:00-07:00). These are the defaults used when you select the Factory Defaults button. “Day BP load value” and “Night BP load value” is important guide line which is used for knowing ponderance about heart blood vessel effect.BP load heightening is important danger factor of end-point event about heart, brain and kidney. “Circadian rhythm of BP ranges” is a ration guide line which judge circadian rhythm state, and is important factor that the doctor knows syndrome about heart, brain and kidney circadian rhythm state for the patient whose BP circadian rhythm has disappeared. Generally speaking, “≥10%” means that
the circadian rhythm is normal, and “<10%” means that the circadian rhythm
is weakened or disappeared. Setting normal value range helps to distinctly comparing of analysis data. The analytical data displays in the report.
Figure 54 Set BP Thresholds
Page 50
50
7.13 Histogram
Click shortcut key , and the following interface will appear.
All: Show all BP data statistics. Day: Show only daytime statistics. Night: Show only nighttime statistics.
7.14 Pie Chart
Click shortcut key , and the following interface will appear.
The interface is divided into four areas.The most left area is maximum, minimum, and average value. The second area is pie chart. The third area
Figure 55 Histogram
Figure 56 Pie Chart
Page 51
51
is the setting area for pie colour and corresponding value, and the BP value and be set by manual method and real-time displays.The forth area has three options including: All, Day and Night. All: Show all BP data statistics. Day: Show only daytime statistics. Night: Show only nighttime statistics.
7.15 Set Print Report
After editing previous BP data, diagnosis information and so on, the software will create a series of analysis reports. You can select all pages or some of them for printing. From the view select “Edit” and select “Congure Report”, then the following dialog box will appear.
You can click “Add Report” to add a new report. If you don’t need some report, you can click “Delete Report” to delete it. You can select an already exist report cyclostyle for printing. You may also click “Edit Report” to edit the selected report. You must select one report page at least.
Figure 57
Report Congution
Figure 58
Edit Report
Page 52
52
Demographics: include patient information, current medications, examining result summary and physician diagnosis information. Statistics Page: particular data statistic information. BP Summary Page: include BP benchmark value, BP load value, average value and diagnosis value. Colorized Graph Page: display the colorized graph of BP data. Bar Graph Page: display the bar graph of BP data Histograms Page: display the histograms graph of BP data (Selected data in the histograms interface: All, Day, Night) Pie Chart Page: display the pie chart of BP data (Selected data in the pie chart interface:All, Day, Night) Frequency Histograms Page: display the histograms graph of high/low BP data for day and night. Correlation Page: display correlation of SYS and DIA PR/MAP/PP Histograms Page: display histograms graph of PR/MAP/PP data. Measure Data Page: display all BP, PP, PR, comment and so on.
7.16 Print Report Preview
When you are sure that you want to print the report, click shortcut key or from menu select “Report”, then print preview page will appear.
Click button or from menu select “Print” to print.
In print preview interface, click button, or press “Ctrl” button glide mouse idler wheel for zoom. In Demographics page, double click the black frame of “Current Medications” and “Comments&Diagnoses” information to edit content, and the edited content could be saved in the case le. Examining result includes maximum, minimum, average value, BP threshold, BP load value and natural load value range (the range can be set in the benchmark value setting dialog box ), circadian rhythm of BP ranges (the range can be set in the benchmark value setting dialog box ), BP CV and so on. BP CV commonly adopt BP SD (standard deviation)/average value to gain. The BP CV for All, Day and Night expresses BP uctuant degree.
Page 53
53
In measure Data Page, if the time is night, the background is gray, and if the Sys or Dia is beyond threshold, the background is red.
Figure 59 Demographics Page Preview
Figure 60 Measure Data Page Preview
Page 54
54
7.17 Help
Click shortcut key to pop-up help le.It is the brief explanation for program function.
7.18 Exit
Click shortcut key to save conguration information and exit application
Figure 61 Measure Data Page Preview
Page 55
55
Chapter 8 TROUBLESHOOTING GUIDE
Cod Description on in Report Editor Solution
1 No signal Check position of cuff, tighten cuff 2 Overreach movement Remain still during BP reading 4 Measurement timeout Check air hose connections and make
certain cuff is tight 85 Airway obstructed Check air hose connections and make
certain air tubing is not crimped 86 Measurement cancelled Push START/STOP button to start reading 87 Cuff leak Check air hose and cuff 88 Safety pause Retry reading, push START/STOP button.
If problem persists return monitor for servicing
89 Cuff overpressure Check for blocked or kinked air hose 90 Battery low Replace batteries. If problem persists return
for servicing 105 Self-check failed Return for servicing 110 Out of range Retry again, if problem persists return
for servicing 115 System error Return for servicing
Page 56
56
Name Ambulatory Blood Pressure Monitor Display 2.4’’ color LCD Display NIBP Specications Measurement Method Oscillometric method Working modes Automatic Measurement Range Pressure adult 0~290mmHg(0~38.6kPa) pediatric 0~235mmHg(0~31.3kPa) neonatal 0~140mmHg(0~18.6kPa) Pulse: 40~240/min Ination adult 160mmHg pediatric 120mmHg neonatal 70mmHg Alarm Range adult mode SYS ALM: 40~270 mmHg DIA ALM: 10~215 mmHg pediatric mode SYS ALM: 40~200 mmHg DIA ALM: 10~150 mmHg neonatal mode SYS ALM: 40~135 mmHg DIA ALM: 10~100 mmHg Overpressure protect adult mode 295±5mmHg pediatric mode 240±5mmHg neonatal mode 145±5mmHg Resolution Pressure 1mmHg Pulse Rate ±2 bpm or ±2% (select larger) Measurement Accuracy Cuff Pressure Accuracy ±3mmHg Error The BP Value of the device is equivalence with the measurement value of Stethoscopy. The error meets all the conditions in the ANSI/AAMI SP-10:2002+A1:2003 +A2:2006. Operating +5ºC~+40ºC 15%RH~80%RH Temperature/ Humidity Transport and Storage -20ºC~+55ºC ≤95%RH Temperature/Humidity Atmospheric pressure 80kPa~105kPa Battery 2 “AA” alkaline batteries Accessories Standard Congure:
Adult Cuff: limb circumference 25-35cm (middle of upper arm) Software CD, User Manual, USB data line, BP extending tube, A pouch.
Separate Sale: Pediatric Cuff: limb circumference 10-19cm (middle of upperarm) Neonatal Cuff: limb circumference 6-11cm (middle of upper arm)
SPECIFICATION
Page 57
57
Guidance and manufacturer’s declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specied below. The customer of the user of the device should assure that it is used in such and environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions Group 1 CISPR 11
RF emission Class B CISPR 11
APPENDIX
Guidance and manufacturer’s declaration – electromagnetic emissionsfor all EQUIPMENT and SYSTEMS
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The device is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specied below. The customer or the user of device should assure that it is used in such an environment.
Immunity test IEC60601 Compliance Electromagnetic test level level environment - guidance
Electrostatic ±6 kV contact ±6 kV contact discharge (ESD) ±8 kV air ±8 kV air IEC61000-4-2
Power frequency 3A/m 3A/m (50/60Hz) magnetic eld IEC61000-4-8
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Floors should be wood, concrete or ceramic tile. If oor are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Page 58
58
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specied below. The customer or the user of device should assure that it is used in such an environment.
Immunity test IEC60601 Compliance Electromagnetic test level level environment - guidance
Radiated RF IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures, objects and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broa­dcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in metres (m). Field strengths from xed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Guidance and manufacturer’s declaration – electromagnetic immunity – for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
3 V/m 80 MHz to 2.5 GHz
3 V/m
d
=
3,5
P
E
1
d
=
7
P
E
1
Page 59
59
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Rated maximum Separation distance according to frequency of transmitter (m) output power of transmitter (W) 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.23
0.1 0.37 0.74 1 1.17 2.33 10 3.69 7.38 100 11.67 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures, objects and people.
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
d
=
3,5
P
E
1
d
=
7
P
E
1
Page 60
60
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment. For further information on recycling points contact the local authorities, the local recycling center or the shop where the product was purchased. If the equipment is not disposed of correctly, fines or penalties may be applied in accordance with the national legislation and regulations.
GIMA WARRANTY CONDITIONS
Congratulations for purchasing a GIMA product.This product meets high qualitative standards both as regards the material and the production. The warranty is valid for 12 months from the date of supply of GIMA. During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defected parts due to production reasons. Labor costs and personnel traveling expenses and packaging not included. All components subject to wear are not included in the warranty. The repair or replacement performed during the warranty period shall not extend the warranty. The warranty is void in the following cases: repairs performed by unauthorized personnel or with non-original spare parts, defects caused by negligence or incorrect use. GIMA cannot be held responsible for malfunctioning on electronic devicesor software due to outside agents such as: voltage changes, electro-magnetic elds, radio interferences, etc. The warranty is void if the above regulations are not observed and if the serial code (if available) has been removed, cancelled or changed. The defected products must be returned only to the dealer the product was purchased from. Products sent to GIMA will be rejected.
Loading...