Gima 300G VET ECG User guide

Page 1
I
Page 2
II
Preface
procedures specified in this User Manual should be followed strictly.This manual describes in
detail the operation steps which must be noted, the procedures which may result in
abnormality, and possible damage to the product or users. Refer to following chapters for
details. Failed to follow the User Manual may cause measuring abnormality, device damage or
personal injury. The manufacturer is NOT responsible for the safety, reliability and performance
issues of such results due to user’s negligence of this user manual for using, maintenance or
storage. The free service s and repairs do not cover such faults either.
The content in this user manual complies with real product. For software upgrade and
some modifications, the content in this user manual is subject to change without prior notice,
and we sincerely apologize for that.
Attentions
Before using this product, the safety and effectiveness described in the following shall be
considered:
Type of protection against electric shock: class I (AC power supply), internal powered
equipment (power supplied by battery)
Degree of protection against electric shock: type CF, defibrillation-proof applied part Working mode: continuous running equipment Enclosure protection class: IPX0 Measurement results shall be described by professional doctor combined with clinical
symptoms.
The using reliability depends on whether the operation guide and maintenance
instructions in this user manual is followed.
Service life: 5 years  Date of manufacture: see the label  Contraindications: none
Warning: To ensure the device safety and effectiveness, please use the company
recommended accessories. The maintenance and repair of the device should be done by
professional personal specified by the company. It is forbidden to refit the device.
Responsibility of the operator
The device must be operated by a professionally trained medical staff, and kept by a
special person.
The operator should read the User Manual carefully before use, and strictly follow the
operating procedure described in the User Manual.
The safety requirements have been fully considered in product designing, but the
operator can not ignore the observation of the patient and device.
The operator is responsible for providing the information of product use to the
company.
Responsibility of the company
Page 3
III
The company supplies qualified products to user in accordance with enterprise standard.
The company installs and debugs the equipment and trains the physicians by contract.
The company performs device repair in warranty period (a year) and maintenance
service after warranty period.
The company responds timely to the user's request.
The user manual is written by Contec Medical Systems Co., Ltd. All rights reserved.
Page 4
Statement
Our company owns all rights to this unpublished work and intends to maintain it as
confidential information. This user manual is used only for reference of operation, maintenance,
or repair of our device. No part of this can be disseminated to others. And our company takes
no responsibilities for any consequences and liabilities caused by using this user manual for
other purposes.
This document contains proprietary information, which is protected by copyright. All rights
reserved. Photocopy, reproduction or translation of any part in the manual without our
company's written permission is prohibited.
All information contained in this user manual is believed to be correct. Our company shall
not be liable for incidental and consequential damages in connection with the furnishing,
performance, or use of this material. This user manual may refer to information and protected
by copyrights or patents and does not convey any license under the patent rights of our
company, nor the rights of others.Our company does not assume any liability for arising out of
any infringements of patents or other rights of the third parties.
Our company owns the final explanation right to this user manual, and reserves the right
to change the content of this user manual without prior notice, and the rights to change
product technology and specification.
IV
Page 5
V
Contents
Chapter1 Overview ............................................................................................................... 1
1.1 Overview ............................................................................................................... 1
1.2 Intended use ......................................................................................................... 1
1.3 Main technical specifications ................................................................................ 1
1.4 Main Characteristics ............................................................................................. 2
1.5 Software overview ................................................................................................ 3
Chapter2 Safety Precautions ................................................................................................. 4
Chapter3 Warranty ............................................................................................................... 7
Chapter4 Working Principle and Structural Characteristics .................................................. 8
4.1 Working principle and its block diagram ............................................................... 8
4.2 Name of each part and its function ...................................................................... 9
Chapter 5 Operation Precautions ....................................................................................... 11
5.1 Precautions before use ....................................................................................... 11
5.2 Precautions during operating ............................................................................. 12
5.3 Precautions after use .......................................................................................... 12
Chapter 6 Preparations before Operation .......................................................................... 13
6.1 Installation of recording paper ........................................................................... 13
6.2 Power supply connection.................................................................................... 13
6.3 Lead cable connection ........................................................................................ 14
6.4 Electrode installation .......................................................................................... 14
Chapter 7 Operation Instructions and Parameter Setting ................................................... 16
7.1 Main Interface .................................................................................................... 16
7.2 Sample interface ................................................................................................
7.3 Case information input interface ........................................................................ 19
7.4 Case management .............................................................................................. 19
7.5 Query .................................................................................................................. 20
7.6 Review ................................................................................................................ 21
7.7 Date and time setup ........................................................................................... 22
7.8 System setup ...................................................................................................... 22
7.9 Sample setup ...................................................................................................... 23
7.10 Analysis setup ................................................................................................... 24
. 16
Page 6
VI
7.11 Print setup ........................................................................................................ 24
7.12 Lead placement ................................................................................................ 26
7.13 About ................................................................................................................ 26
Chapter 8 Troubleshooting ................................................................................................. 27
8.1 Auto shutdown ................................................................................................... 27
8.2 AC interference ................................................................................................... 27
8.3 EMG interference ............................................................................................... 27
8.4 Baseline drift ....................................................................................................... 28
8.5 Troubleshooting list ............................................................................................ 28
Chapter 9 Maintenance ...................................................................................................... 30
9.1 Battery ................................................................................................................ 30
9.2 Recording paper ................................................................................................. 31
9.3 Maintenance after use ........................................................................................ 31
9.4 Lead cables and electrodes ................................................................................. 32
9.5 Silicone rubber roller .......................................................................................... 32
9.6 Cleaning of thermal print head ........................................................................... 32
9.7 Disposal of product scrap ................................................................................... 32
9.8 Others ................................................................................................................. 32
Chapter 10 Packing List and Accessories ............................................................................. 34
10.1 Accompanying accessories ............................................................................... 34
10.2 Notes ................................................................................................................ 34
Appendix I ECG Automated Measurement&Interpretation Guide ...................................... 35
1. Preface .................................................................................................................. 35
2. Automated measurement parameters and Automated interpretation items ...... 35
3. Algorithm description ........................................................................................... 37
4. Data sources and data preprocessing ................................................................... 47
5. Process and Result of Verification ........................................................................ 51
Appendix II EMC Guidance and Manufacturer6 Declaration ............................................... 61
Page 7

Chapter1 Overview

1.1 Overview

This product is a kind of electrocardiograph, which is able to sample 12 leads ECG signals
simultaneously and print out the ECG waveform with thermal printing system. Its functions are
as follows: recording and displaying ECG waveform in auto/manual mode; measuring ECG
waveform parameters automatically, and automatic analysis; prompt for electrode-off and out
of paper; optional interface languages(Chinese/English, etc.); built-in lithium battery, powered
either by AC or DC; arbitrarily select the rhythm lead to conveniently observe abnormal heart
rhythm; case database management, etc.

1.2 Intended use

This product is suitable for hospital, scientific research, wards, ambulances and carrying
out medical consultations. It can be used by medical institutions to record human ECG signals,
collect and extract the ECG waveform.

1.3 Main technical specifications

1.3.1 Environment conditions
Operation:
a). Environment temperature: 5~40
b). Relative humidity: 25%~95%(no condensation)
c). Atmospheric pressure: 700 hPa~1060 hPa
d). Power supply: Voltage: 100-240 V
Frequency: 50 Hz, 60 Hz Input power: 150 VA
Battery: 7.4 V, 3500 mAh rechargeable lithium battery
Transportation and Storage: a). Environment temperature: -20 +55 b). Relative humidity: ≤95% c). Atmospheric pressure: 500 hPa1060 hPa
1.3.2 Input way: Floating and defibrillation protection
1.3.3 Lead: 7 leads
1.3.4 Patient leakage current: <10µA
1.3.5 Input impedance: 2.5 MΩ
1.3.6 Frequency response:
Rated input
amplitude
1.0 0.67Hz~40Hz, Sine wave ±10%a
0.5 40Hz~100Hz, Sine wave +10 %, -30 %a
Input frequency and waveform Relative output response
1
Page 8
0.25 100Hz~150Hz, Sine wave +10 %, -30 %a
0.5 150 Hz ~ 500 Hz, Sine wave
1.5 ≤1Hz,200ms, Triangle wave
a
relative to 10Hz b relative to 200 ms
+10 %, -100 %
+0 %, -10 %
1.3.7 Time constant: ≥3.2s
1.3.8 CMRR: >105 dB
1.3.9 Filter: power frequency(AC50/60 Hz), myoelectricity(25 Hz/35 Hz (-3 dB)), baseline drift
filter
1.3.10 Recording way: Thermal printing system
1.3.11 Specification of recording paper: 80 mm(W)×20 m(L) high-speed thermal paper
1.3.12 Time base selection(paper speed):
12.5 mm/s, 25 mm/s, 50 mm/s, error: ±5%
1.3.13 Gain control(sensitivity): 5,10, 20mm/mV, accuracy is ±2%; Standard sensitivity: 10
mm/mV±0.2 mm/mV
1.3.14 Auto record: record setup according to auto record format and mode, automatically
change leads, automatically measure and analyze.
1.3.15 Rhythm record: record setup according to rhythm record format and mode,
automatically measure and analyze.
1.3.16 Manual record: record according to manual record format.
1.3.17 Measurement parameters: HR, PR Interval, P Duration, QRS Duration, T Duration, QT/
QTc Interval, P/QRS/T Axis, R(V5) amplitude, S(V1) amplitude, R(V5)+S(V1) amplitude
1.3.18 Product safety type: Class I type CF defibrillation-proof applied part
1.3.19 Polarization resistance voltage: ±610 mV
1.3.20 Noise level: ≤12 µVp-p
1.3.21 ECG signal input sampling frequency: 32 kHz
1.3.22 Waveform data processing sampling frequency: 1 kHz
1.3.23 Sampling precision: 24-bit
1.3.24 The minimum detection signal: 10 Hz, 20 µV(peak-peak value) deflected sinusoidal signal
can be detected
1.3.25 Accuracy of input signal: ±5%.
1.3.26 Amplitude quantization: ≤5µV/LSB
1.3.27 Dimension: 315 mm(L)×215 mm(W)×77 mm(H)
1.3.28 Net Weight: 1.6 kg
1.3.29 Interchannel time deviation: <100 µs

1.4 Main Characteristics

1.4.1 High resolution thermal-array output system(8 dots/mm), no adjustment required.
Frequency Response is up to 150Hz.
1.4.2 Record clear and exact three channels ECG waveform and remarks in real-time and
continuously. The remark includes: lead sign, sensitivity, paper speed, filter state, etc.
a
b
2
Page 9
1.4.3 In auto mode, recording can be completed with one-button operation, which improves
work efficiency.
1.4.4 Under the best DC condition, the device can stand by for 10 hours, or print for at least 3
hours, or print 260 pieces ECG.
1.4.5 At least 1,000 pieces of medical records can be stored in the device, which facilitates
doctors to review and information statistics.
1.4.6 Beautiful and smooth appearance.
1.4.7 Degree of protection against ingress of liquid: IPX0
1.4.8 Use digital signal processing technology to conduct AC filter, baseline filter and EMG filter
on ECG signals, in order to get high-quality ECGs.
1.4.9 With auto-measurement, auto-analysis functions of regular ECG parameters, which
reduces doctor’s workload and improves working efficiency.

1.5 Software overview

The ECG analysis program shows the results after analyzing the form of the
electrocardiogram, providing auxiliary reference for doctors to make diagnosis. The analysis
result cannot be used as the only standard for diagnosis. A comprehensive evaluation should be
made by professional electrocardiogram technicians and physicians according to clinical
experience and other test results.
The device is intended for use on all patient populations, which is decided by the clinical
doctor. The analysis program only provides ECG analysis for patients above 3 years old
(including 3 years).
Name of software:embedded software
Software specification: none
Software version: V1.9.11
Version naming rules: V<major version number>.<minor version number>.<revision version
number>
The version of the software can be obtained in “About”.
Involved algorithm:
Name: ECG algorithm
Type: mature algorithm
Use: to convert ECG signals of human body into intuitive waveform images and then analyzing.
Clinical function: Electrocardiogram is an important method for clinical diagnosis of
cardiovascular disease. How to use computer to quickly, automatically and accurately analyze
ECG has been a hot topic for scholars at home and abroad. The ECG algorithm is the key to the
analysis and diagnosis of ECG signals, and its accuracy and reliability determine the
effectiveness of diagnosis and treatment of patients with heart disease.
3
Page 10

Chapter2 Safety Precautions

2.1 Ensure that the device is placed on a flat level worktable. Avoid strong vibration or impact
when moving it.
2.2 When working with AC power, the power cord must be 3-core, the frequency and voltage
value of the AC power source must match the identification on the manual and have sufficient
capacity. When the provided three-core power cord cannot be used, please use the built-in DC
power supply or replace the three-core power cord that meets the standard requirements.
2.3 A perfect power supply system and grounding are necessary in the room.
Warning: To avoid the risk of electric shock, the device must be connected a power supply
with protective grounding.
2.4 If there are any questions for the integrality of protective grounding cable or the reliability
of protective grounding cable connection can not be guaranteed, the device must be run with
built-in DC power supply.
2.5 The safety requirements have been fully considered in product designing, but the operator
can not ignore the observation of the patient and device. Cut off the power or take off the
electrode when necessary to ensure patient's safety.
2.6 Please turn off the device and unplug the power cord before replacing the fuse or cleaning
and disinfection. Don't rub the screen with sharp materials.
2.7 Keep the device from water, don't use or store it in places with high air pressure, humidity
or temperature over the standard, bad ventilation, or too much dust.
2.8 Do not use the device in the place with flammable anesthetic gases or other flammable
chemicals, otherwise there is a danger of explosion or fire.
2.9 Do not use the device in medical hyperbaric oxygen chamber, otherwise there is a danger of
explosion or fire.
2.10 This device is not intended to act directly on the human heart. If this device is used with
cardiac defibrillator or other electric stimulating devices at the same time, single-use electrodes
and ECG lead cables with defibrillation-proof function should be selected. It is better not to use
this device with other electric stimulating devices at the same time. If it is necessary, there
must be professional technician guiding on the scene, and the selected accessories should be
designated by our company.
Warning:Do not operate the instrument on parts of human body with wounds,and do not
perform measurements on parts with wounds on the surface.
2.11 When the electrocardiograph is used together with a high-frequency electrosurgical knife,
the ECG electrode should be kept away from the contact of the electrosurgical knife to prevent
burns and burning of the electrode wires caused by high-frequency sparks.
2.12 When the electrocardiograph is used together with a defibrillator, the operator should
avoid contact with the patient or the sickbed. The defibrillation electrode should not directly
4
Page 11
touch the ECG electrode to prevent sparks from burning the device and the patient.
2.13 Please do not use the electrocardiograph in the environment that is interfered by
high-power device such as high-voltage cables, X-rays, ultrasonic machines and electrizer, keep
the device away from emission sources such as mobile phones.
2.14 If other equipment is connected with this ECG device, it must be a Class I device that
complies with IEC60601-1. Because the total leakage current may hurt patient, the monitoring
of leakage current is carried out and taken charge by the connected equipment.
2.15 Notes related to EMC
The device complies with the safety standards for medical electrical equipment or system
electromagnetic compatibility in IEC60601-1-2. Electromagnetic environments exceeding the
IEC60601-1-2 standard may cause harmful interference to the device or prevent the device
from performing its intended function or degrade its performance. Therefore, if there is a
phenomenon that does not match its function during use, be sure to confirm and eliminate
adverse effects before continuing to use it. Corresponding precautions for this situation are
given in this manual. The device or system should not be used near or stacked with other devices. If it must be
used near or stacked with other devices, it should be observed and verified that the
device is working normally under the configuration it is using.
The use of ACCESSORIES other than those specified by the MANUFACTURER of the device
or system, may result in increased EMISSIONS or decreased IMMUNITY of the ME
EQUIPMENT or ME SYSTEM.
Effect from radiated electromagnetic waves:
The use of a mobile phone may affect the operation of the device. When installing medical
electrical equipment, be sure to remind people around the device to turn off mobile phones
and small radios. Effect from shock and conduction electromagnetic waves:
High frequency noise from other equipment can enter the device through the AC socket.
Please identify the source of noise, if possible, stop using the equipment. If the equipment can
not be deactivated, use noise cancellation equipment or take other measures to reduce the
impact. Effect from static electricity:
Static electricity in a dry environment(indoor) may affect the operation of the device,
especially in winter. Before using the device, humidify the indoor air or discharge the static
electricity from the cable and operator. Effect from thunder and lightning:
If there is thunder and lightning nearby, it may cause a voltage surge in the device. If you
are concerned about danger, disconnect the AC power and use the internal power supply.
2.16 Notes concerning ECG waveform measurement and analysis
2.16.1 The identification of P wave and Q wave is not always reliable with intensive EMG or AC
interference. Neither are the ST segment and T wave with baseline drift.
2.16.2 Winding and unclear end position of S wave and T wave may cause error in
5
Page 12
measurement.
2.16.3 When R wave is uninspected caused by some leads off or QRS wave low voltage, the
heart rate measurement may deviate greatly from the correct.
2.16.4 In case of QRS low voltage, ECG axis calculation and border-point identify of QRS wave
are not always reliable.
2.16.5 Occasionally, frequent ventricular premature complexes may be identified as dominant
beat.
2.16.6 Merging of versatile arrhythmia may result in unreliable measurement because of the
difficulty in distinguishing P wave in such situation.
2.16.7 The device has an automatic analysis function that automatically analyzes the obtained
ECG waveform without reflecting all the patient’s status. The results of the analysis may
sometimes not comply with the doctor’s diagnosis. Therefore, the final conclusion needs to be
comprehensively analyzed by doctors in combination with analysis results, patient clinical
characterization and other test results.
6
Page 13

Chapter3 Warranty

3.1 In normal use, under strict observance of user manual and operation notes, in case of
failure, please contact with our customer service department. Our company has the sales
record and customer archives for each device. The customer has one year free warranty service
from the date of shipping according to the following conditions. To supply all-around and quick
maintenance service for you, please mail the maintenance card to us in time.
3.2 Our company may adopt such ways as guidance, express to company or door-to-door
service, etc. to carry out warranty promise.
3.3 Even in warranty period, the following repairs are charged.
3.3.1 Faults or injuries caused by misuse that not according to user manual and operation
notes.
3.3.2 Faults or injuries caused by dropping accidentally after purchase.
3.3.3 Faults or injuries caused by repair, reconstruction, decomposition, etc. not by our
company.
3.3.4 Faults or injuries caused by improper storage or force majeure after purchase.
3.3.5 Faults or injuries caused by using improper thermal recording paper.
3.4 The warranty period for accessories and fray parts is half a year. Power cable, recording
paper, operation manual and packing material are excluded.
3.5 Our company is not responsible for the faults of other connected devices caused by the
faults of this device directly or indirectly.
3.6 The warranty will be canceled if we find the protection label has been destroyed.
3.7 For charged maintenance beyond warranty period, our company advises to continue using
"Maintenance contract regulation". Please refer to our customer service department for
details.
7
Page 14

Chapter4 Working Principle and Structural Characteristics

4.1 Working principle and its block diagram

4.1.1 The power supply unit

Principle of power supply
After the AC power supply enters the switching power supply, it is converted to DC voltage
and supplied to the DC-DC power unit, it also provides constant voltage current limiting
charging for the rechargeable lithium battery in the device through the DC-DC circuit, and
generates +5V and +8.5V voltage through the power conversion to supply power to the
corresponding modules. At the same time, the lithium battery in the device can independently
satisfy working requirements of each module in the device through the buck-boost circuit.
Note: The principle block diagram and component list are only available to service stations
or maintenance personnel designated by our company.

4.1.2 Signal acquisition unit

The signal acquisition unit uses a floating setting, which is a signal acquisition and
processing system, including analog circuit part and A/D conversion (with sampling accuracy of
24 bits) and data processing part. The analog circuit consists of signal following, amplification,
anti-aliasing low-pass filtering, lead-off detection and overload detection. CPU system is
responsible for coordinating the work of each circuit such as the A/D converter, the lead-off
detection circuit and the overload detection circuit, in order to achieve signal acquisition,
processing, and lead-off detection. Control information and A/D conversion and data
acquisition between the floating circuit and the solid circuit are transmitted through the
optoelectronic coupler.

4.1.3 Control unit

1Principle of control unit
The control system consists of printing system, button system, liquid crystal display system,
and signal acquisition system. The ECG signal sent from the signal acquisition system through
the high-speed optoelectronic coupler is received by the CPU system, after digital filtering, gain
adjustment and motor drive, it is sent to the printing system to print the ECG waveform. After
the printing is completed, the CPU system processes waveform measurement and analysis. The
CPU system also receives an interrupt signal and button code from the button system to
complete the interrupt processing. In addition, the lead-off signal, paper out detection, battery
voltage management, and automatic power-off are also managed by the CPU system. The liquid
crystal controller receives data and commands from the CPU system to complete the display of
the control state of the device.
8
Page 15
Control system
Signal acquisition
system
Display system
Button system
Printing system
Power module
4
2Principle block diagram is shown in Figure 4-1.
Figure 4-1 Block diagram of control unit

4.2 Name of each part and its function

4.2.1 Front view

1
2
3
Figure 4-2 Front view
1. Paper compartment cover
Keep the paper compartment closed, hold the printing paper
2. Display screen
Display patient’s ECG and related information
3. Button area
Control the operations of the device, and enter information.
4. Cover switch
To open or close the paper compartment cover.
Note
Do not put heavy objects on the screen or hit against it, otherwise the screen will be
damaged.
If the device is not in use, cover it to prevent liquid spills on the screen. Do not use sharp stuff to operate the buttons, otherwise it may case permanent
damage to the buttons.
9
Page 16

4.2.2 Side view

6
5
Figure 4-3 Side view
5. Lead cable interface
Connect with lead cables.
6. USB interface
Communicate with the computer. The ECG data and analysis result can be transmitted to a
computer, by using the computer, many functions can be achieved, such as archiving, managing,
and analyzing ECG data, which facilitates clinical research, organization teaching and training, as
well as program upgrade, case export and connection with external printer.
Note
Lead cables must be disconnected from patient before connecting with a computer via
the USB interface.
Operator must not touch the USB interface and patient at the same time.

4.2.3 Rear view

Figure 4-4 Rear view
7. Equipotential terminal
Connect with the potential equalization conductor.
8. Input socket
Connect with AC power cord.

4.2.4 Buttons

10
Page 17
municipal waste and must be recycled separately.
Figure 4-5 Schematic diagram of buttons
1. Startup indicator
It lights in green after turning on the device.
2. Power status indicator
Green indicates that the AC power supply is used. At this time, there is no battery in the device
or the battery is full. Red and green two colors indicate that the battery is being charged.
3. MENU
Menu button
4. MODE
When the device in sampling interface, use MODE button to select the print mode.
5. 1mV
Calibration button
6. PRINT
Print the sampled ECG waveform or end the printing.
7. Direction button
Up button
8. Direction button
Right button
9. SET
System menu and confirm.
10. Direction button
Down button
11. Direction button
Left button
12. FILTER
Set the filter mode.
13. SPEED
Change the ECG recording speed
14. SEN
Adjust the sensitivity manually.
15. ON/OFF
When the device is turned on, short press this button, it will prompt whether to shut down the
device, long press this button to turn off the device.
Waste disposal symbol. This symbol indicates that electrical and
electronic equipment waste cannot be disposed of as unsorted

Chapter 5 Operation Precautions

5.1 Precautions before use

5.1.1 For safe and effective use, please read the user manual carefully before operation.
5.1.2 Check to ensure that the device is in good condition.
11
Page 18
5.1.3 The device shall be placed on a flat surface, and moves gently to avoid strong vibration or
shock.
5.1.4 Check to ensure that the lead cables are correctly connected, and the device grounding is
correct.
5.1.5 The AC frequency and voltage should comply with the requirements, and enough current
capacity should be guaranteed.
5.1.6 When using the battery for power supply, check to ensure that the battery voltage and
battery status are in good condition, and the battery has enough power.
5.1.7 When the device is used together with other equipment, all devices and equipment
should be equipotential grounded in order to protect the user and operator.
5.1.8 Install the device where easily grounded in the room. Do not allow the patient and
patient-connected lead cables and electrodes to come into contact with other conductor parts,
including the earth or a hospital bed.
5.1.9 Clean the lead cable with neutral solvent. Do not use alcohol-based cleaners or gemicides.
5.1.10 Ensure that the device is running within the normal ambient temperature range of 5 to
40℃. If the device is stored at a higher or lower temperature, leave it in the operating
environment for approximately 10 minutes before use in order to ensure the normal work.

5.2 Precautions during operating

5.2.1 The printing can be started after the ECG waveform is stable.
5.2.2 During using, the doctor should observe the patient carefully and cannot leave the
operating site. If necessary, turn off the power or remove the electrode to ensure patient
safety.
5.2.3 The patient and the device can only be connected via lead cables through the electrodes,
in order to avoid patient touches other parts of device or conductors.
5.2.4 Patient can not move during operating.
5.2.5 Maintenance or repair to the device or accessory is not allowed during using.

5.3 Precautions after use

5.3.1 Set the states of all functions to initial states.
5.3.2 Cut off the power, gently remove the electrodes and limb clips, then remove the lead
cables, do not pull with force.
5.3.3 Clean the device and all accessories, and store them for the next use.
12
Page 19

Chapter 6 Preparations before Operation

6.1 Installation of recording paper

6.1.1 The device adopts high-speed recording paper, its specification is 80 mm(W)×20 m(L).
6.1.2 The installation method of recording paper is described as below:
1. As shown in Figure 6-1, open the paper cabinet cover, take out the paper axis, insert it into
the roll paper as shown in the figure. The paper side with grids should be faced downwards,
and then install it to proper position in the paper cabinet.
Figure 6-1 Installation of recording paper
2. Close the paper cabinet cover, it would be better to leave 2cm paper outside the exit of
paper cabinet.
Note: The recording paper should be aligned with the slot of the paper cabinet cover. It is
recommended to leave 2cm paper outside.
6.1.3 If the recording paper runs out during recording, the device will stop printing
automatically, and the screen will display a prompt of lack of paper.

6.2 Power supply connection

6.2.1 AC

Insert one end of the provided three-core power cord into the device’s input socket, and
insert the other end into a three-core power socket that meets the requirements. Ensure that
the connection is secure and reliable, and the device is automatically grounded.
When the device is used in conjunction with other medical equipment, use the supplied
potential equalization wire to connect the equipotential terminal of the device to the
equipotential terminal of the connected equipment to prevent leakage current and protect the
device.
13
Page 20

6.2.2 Battery

The device has a built-in rechargeable lithium battery, which does not need to be
re-installed by user. Check the battery's power and status before use.
Note: Connect one end of the potential equalization wire to the equipotential terminal of
the device, and connect the other end to the ground to enhance the reliability of the
grounding. Do not use other pipes as ground wire, otherwise, the patient may be in danger of
electric shock.

6.3 Lead cable connection

Connect the lead cable to the lead cable interface on the device, and fasten it to the device
with the fixing knobs at both sides of the lead cable in order to prevent bad connection and
affecting the detection.
Note: The lead cable interface can not be used for other purposes except as the input
interface of ECG signals.

6.4 Electrode installation

Proper installation of the electrodes is an important part of accurately recording the
electrocardiogram. Make sure the electrodes are in good contact. Old and new electrodes or
reusable electrodes and disposable electrodes cannot be used at the same time. If different
types of electrodes are used together, it will seriously affect the ECG recording. The electrode
or lead plug must not touch other object surfaces or conductors, such as metal beds. Please
replace them all when updating the electrodes.
Warning: Do not test on part with wounds.
Following is the configuration per American standard when using five leadwires:
Figure 6-2 Leads Placement
RA (right arm) lead: on the right foreleg.
LA (left arm) lead: on the left foreleg.
RL (right leg) lead: on the right hind leg.
LL (left leg) lead: on the left hind leg.
V (precordial) lead: exploring lead.
Note: The exploring lead is used for diagnostic purposes as needed. Otherwise, it may be
14
Page 21
left unplugged.
The chart below shows the label used to identify each leadwire. Included also is its
associated color code per American (AHA) and European (IEC) standards.
American Standard European Standard
Label Color Label Color
RA White R Red
LA Black L Yellow
LL Red F Green
RL Green N Black
V Brown C White

6.4.1 Lead-off and overload indication

The device can check the connection status of the lead at any time. If lead-off or overload
is detected, the screen will display corresponding lead code on the top left corner.
Note
In the lead-off prompt area, red font represents lead-off, yellow font represent
overload.
When the connection between lead cable and patient/the device is not reliable, and
the ECG signal can not correctly transmitted, the device displays lead-off.
In the printed report, lead-off is marked with “*”, and lead overload is marked with
“+”.
15
Page 22
Total1
New
Archive
Figure
Date And Time
Analysis Setup
System Setup
Print Setup
Sample Setup
About
title
gather
Time

Chapter 7 Operation Instructions and Parameter Setting

7.1 Main Interface

As shown in below figure:
 Status bar
Storage space usage: display the usage condition of current storage space
 Current case amount: display the total number of current case
Battery power (refer to 9.1): display current power status of the device
Time (on bottom right corner of the screen): display current time of the system
Function buttons
 New: to enter the sampling interface, finish waveform acquisition, display and report
printing, generally, the device will automatically enter this interface after powering on
Archive: to enter the case management interface, in this interface, user can query,
modify or delete case information
Figure: to view the placement of leads
Date And Time: to set time and date
System Setup: to perform system setting
Sample Setup: to perform sampling setting
Analysis Setup: to set the parameters used in automatic analysis
Print Setup: to set the print mode, print style and print content, etc.
About: to display software version No., creation time and firmware information, etc.

7.2 Sample interface

Click “New” on the main interface or press the
interface.
Note: If “Info Input” is set to “Before” in Sample Setup, the case information needs
button to enter the sampling
to be entered before acquisition (refer to 7.3).
16
Page 23
Instant Heart rate
Return
Instant Waveform
Print
Lead-off Information Status
Speed FilterGain
Lead
Print Mode
The sampling interface provides several lead display modes, including 3-lead, 6-lead and
12-lead. The following figure uses 12-lead as an example:
End sampling: After the device starts sampling, use the
button to end the
sampling, and back to the main interface.
Switch lead: When the device does not simultaneously display 12 leads, use the
and
buttons to switch displayed waveform.
Switch lead display style: use the
and buttons to switch the display style
between 3-lead, 6-lead and 12-lead.
Lead-off information: In demo mode, it displays “DEMO ECG”. In sampling mode, it
displays the detected lead status.
Print mode: use the
button to switch the print mode between Manual, Auto
4×3, Auto 3×4+1, Auto 3×4, Auto 2×6+1, Auto 2×6, Auto 3-2+1, Auto 3-2, Auto 1×12+1, Auto
1×12, Rhythm 4, Rhythm 3 and Rhythm 2.
Gain (sensitivity): use the
button to switch the gain between 5 mm/mV, 10
mm/mV and 20 mm/mV. The overall gain (sensitivity) can be checked by calibration function.
Speed: use the
button to switch the speed between 12.5 mm/s, 25 mm/s and
50 mm/s.
Filter: use the
AC+EMG, AC+DFT, EMG+DFT and AC+EMG+DFT.
button to switch the filter between no filter, AC, EMG, DFT,
In which, AC AC filter
EMG EMG filter
DFT Baseline filter
17
Page 24
Display content
Explanation
Process…
It is printing.
Waiting…
It is finishing printing.
No Paper.
Lack of paper, user should restart printing after loading paper.
Print Timeout.
Communication failure between this system and printing sub-system.
ECG Timeout
Communication failure between this system and sampling sub-system.
Low Power
Low power, it cannot start printing.
please print again after connecting a USB printer.
printing under rhythm mode
Show Speed
[12.5mm/s]/[25mm/s]/[50mm/s]
Set the speed of displayed ECG.
Display calibration signal: after pressing button,1 mV signal will occur on the
screen once.
Note: The calibration is an automatic process, user does not need to press any button.
Print/End print: use the
Auto mode: After starting to print, the system automatically prints and stores the real-time
12-lead ECG waveform. The length is determined by the relevant settings in the print setup.
Based on the settings, the automatic analysis data and conclusions are printed, and the system
automatically ends printing.
Manual mode: After starting to print, user need to switch the lead to print the waveform
of different leads, that is, the ECG printed in the manual mode is asynchronous, and the data is
not saved. User need to press the PRINT button again when the print needs to be terminated.
If lead-off appears during acquisition, the waveform printed will be marked with “*”.
If lead overload appears during acquisition, the waveform printed will be marked with “+”.
During printing, the display content of print status are including:
Print Failed
Note: You can not print until ECG waveforms are displayed on the screen.
In current interface, press
of each setting item and its description are shown in the following table:
Item Options Description
AC filter
EMG filter
DFT Filter
Rhythm Lead
When it was set to use a USB printer, no such printer was connected,
[ON]/[OFF]
[ON]/[OFF]
[ON]/[OFF]
Any lead among the 12 leads
button to start or end the printing operation.
to enter quick setup interface. The optional content
Turn on or off the AC filter.
Turn on or off the EMG filter.
Turn on or off the Baseline filter.
Set the rhythm lead that used for
Show Style [3 Leads]/[6Leads]/[12Leads] Set the display method of ECG.
Show Gain [5mm/mV]/[10mm/mV]/[20mm/m
V]
Click "OK" to apply new setup and return to sampling interface; while click"Cancel" not to
18
Set the gain of displayed ECG.
Page 25
Start
Cancel
ID
Name
Sex
Date And Time
Age
Height
Weight
/
SYS/DIA
Pace
Physician
mmHg
cm kg
Total1 Current1 / 1
Date And Time ID Name Sex TimeLen
Case list
Adv-opr
<
Review Delete
Return
<< >>
>
Sampling Time 001 Patient Name Sex TimeLen
apply and directly return to sampling interface.

7.3 Case information input interface

Due to the difference of Info Input (refer to 7.9), user could choose to input case
information (including ID, name, etc.) before or after the sampling, or not input the case
information, the dialogue box is shown as below:
After selecting a edit box, pressing
Clicking “Caps” can switch between numbers, lowercase letters, capital letters and symbols.
“Space” is the space key, press it to enter a space; “Backspace” is the backspace key, press it to
delete the last character entered. Click “OK” to confirm the entry and exit the interface.
The keyboard may have input restrictions according to the limitation of content. The
restricted keys will be grayed out and unavailable.

7.4 Case management

In the main interface, click “Archive” to enter the case management interface, as shown
below:
The above interface shows all medical records stored in the device. User can search
19
button could pop up a soft keyboard.
Page 26
Select Close
ID
Name
Cond.And
Select Conditions
Clear
Cond.or
Sex
Weight
SYS/DIA
Physician
/
kg
Age
Height
cm
necessary cases by the query function in the interface (refer to 7.5), modify any case
information and review the case waveform stored by “Review” (refer to 7.6), and delete the
case by “Delete”.
The Adv-opr includes the following contents:
List All: list all cases Query: refer to 7.5Delete All: delete all cases (caution, it is unrecoverable). Export: connect the U disk through the USB port, to export the cases in the device
into the specified folder in the U disk.
Return: return to the case management interface.
modify or delete case information by edit function, and review stored case information
(refer to 7.6).
Click
Click
Click
Click

7.5 Query

Select “Adv-opr” in the case management interface to enter its sub-menu, then select
“Query” to enter its setting interface. Input query conditions and click “Select” to obtain
expected results. After clicking “Clear”, the system will clear all query conditions that entered.
to jump to the first page of case list.
to jump to the last page of case list.
to jump to the previous page.
to jump to the next page.
“Cond.And” and “Cond.Or” indicate the matching mode of the query conditions. You can
choose one of the two. If you select “Cond.And”, the displayed query results will satisfy all the
input conditions at the same time; if you select “Cond.Or”, the displayed query results only
need to meet any of the conditions entered.
Suggestion: When there are many cases, it would be better to input accurate query
conditions and choose “Cond.And” to quickly find the case.
20
Page 27
Review Close
ID
Name
Sex
Sampling Time
Age
Height
Weight
/
SYS/DIA
Pace
Physician
mmHg
cm kg
Save
Sampling Time
Return
Review Waveform
Print
Review time / Sampling
time length
Status
Speed
<<
Gain
Print Mode
>>
Diagnose

7.6 Review

In the case management interface, select a case to be reviewed, click “Review” to enter
the following dialog box, which displays the case information. User is allowed to modify patient
information, after clicking “Save”, the information will be changed. Please note that the
modification is irreversible.
Make sure the input information is correct, click “Review” to enter the review interface,
which is similar with the sampling interface.
In current interface, user could adjust the time period of displayed waveform by
buttons, each pressing could move the waveform to corresponding direction for 1s, and
the speed and gain can be changed (refer to 7.2).
marked with “*”.
be marked with “+”.
In this interface, if lead-off appears during acquisition, the waveform reviewed will be
In this interface, if lead overload appears during acquisition, the waveform reviewed will
21
and
Page 28
Item
Options
Description
current case
mode.
rhythm mode
ECG.
ECG.
ECG.
Item
Options
Description
used..
In this interface, user can use button to change the print mode.
In this interface, user can use
If lead-off appears during acquisition, the waveform printed will be marked with “*”.
If lead overload appears during acquisition, the waveform printed will be marked with “+”.
In current interface, press
of each setting item and its description are shown in the following table:
[Auto 4×3]/[Auto 3×4]/[Auto 2×6], and
Print Mode
Rhythm Lead Any lead among the 12 leads
Show Style
Show Gain
Show Speed
Click "OK" to apply new setup and return to review interface; while click"Cancel" not to
apply and directly return to review interface.

7.7 Date and time setup

In the main interface, click “Date And Time” to set date and time.
other print modes that applicable for
[3 Leads]/[6Leads]/[12Leads]
[5mm/mV]/[10mm/mV]/[20mm/mV]
[12.5mm/s]/[25mm/s]/[50mm/s]
button to print.
to enter quick setup interface. The optional content
The system takes the
selected option as print
Set the rhythm lead that
used for printing under
Set the display method of
Set the gain of displayed
Set the speed of displayed
content of the item by

7.8 System setup

descriptions are as followings:
In current interface, user can switch the items via
and buttons.
Select “System Setup” in main menu to enter its setting interface, the options and their
[None]/[30Seconds]/[1
ScreenSaver
Minute]/[2 Minutes] /[5
Minutes]/[10 Minutes], etc.
22
and buttons, and adjust the
If there is no operation after reaching the
set time, screen saver will be active. If it is
set to “None”, this function will not be
Page 29
always keep on.
Minutes]/[60 Minutes], etc.
always keep on.
device uses in low power.
[60Hz/25Hz]/[60Hz/35Hz]
filter.
Language
[English]/[Chinese], etc.
To set the default language of system.
sound.
the system will run in
, the system will
run in sampling mode.
restore to the default options.
Item
Options
Description
[OFF]/[ON]
Turn on or off the EMG filter.
[OFF]/[ON]
Turn on or off the Baseline filter.
to input the case information
before or after sampling, or not input
0mm/mV]
mm/s]
-opr
Lead]
[30Seconds]/[1 Minute]/[2
Back-light
Auto Off
Low Power [None]/[Only once]/[Always]
Filter Freq
K-B Sound [ON]/[OFF]
Demo Mode [ON]/[OFF]
Default ————

7.9 Sample setup

Select “Sample Setup” in main menu to enter its setting interface, the options and their
descriptions are as followings:
Minutes] /[5 Minutes]/[10
Minutes]/[Always On], etc.
[None]/[1 Minute]/[3
Minutes]/[5 Minutes]/[10
Minutes]/[15 Minutes]/[30
[50Hz/35Hz]/[50Hz/25Hz]/
If there is no operation after reaching the
set time, screen backlight will turn off. If it
is set to “Always On”, the backlight will
If there is no operation after reaching the
set time, the system will automatically turn
off. If it is set to “None”, the system will
It determines the alarm method that the
To set the parameters of AC filter and EMG
If it is selected, the button makes sound
while pressing, otherwise, there will be no
If it is selected,
Demo mode; otherwise
Select “Default”, the system settings will
AC Filter
EMG Filter
DFT Filter
Info Input [Before]/[After]/[None]
Show Style
Show Gain [5mm/mV]/[10mm/mV]/[2
Show Speed [12.5mm/s]/[25mm/s]/[50
Adv
Sort Lead
[OFF]/[ON] Turn on or off the AC filter.
[3 Leads]/[6 Leads]/[12
Leads]
[Routine Lead] / [Cabrera
Set
Set the display method of ECG.
Set the gain of displayed ECG.
Set the speed of displayed ECG.
Set the arrangement mode of the lead
23
Page 30
nd Grid
Be used to set whether with
background grid.
Filter
[150Hz]], etc.
filter.
[1Hz], etc.
Hospital
————
Fill in the hospital name in the report.
“Default”, the system settings
will restore to the default options.
V5]/[V6]
heartbeat sound.
as the criterion of premature beat.
as the criterion of asystole.
as the criterion of tachycardia.
Print Mode
[Manual]/[Auto
The system takes the selected
Backgrou
Low-pass
DFT Freq
Default ————

7.10 Analysis setup

Select “Analysis Setup” in main menu to enter its setting interface, the options and their
descriptions are as followings:
The setup here will affect real-time analysis during sampling, case review and diagnosis
prompt of print report.
Item Options Description
Rhythm Lead
Heartbeat Sound [OFF]/[ON]
Premature (%) ————
[Show]/[Not Show]
[OFF]/[75Hz]/[100Hz]]/
[0.05Hz]/[0.15Hz]/[0.25Hz]/
[0.32Hz]/[0.50Hz]/[0.67Hz]/
[I]/[II]/[III]/[aVR]/[aVL]/[a
VF]/[V1]/[V2]/[V3]/[V4]/[
Set the parameters of the low-pass
Set the parameters of the DFT filter.
Select
Set the rhythm lead, be used to print
under rhythm mode.
Be used to set whether it has
The system will take the input value
Pause Time (ms) ————
Tachycardia (bpm) ————
Bradycardia (bpm) ————
Default ————

7.11 Print setup

Select “Print Setup” in main menu to enter its setting interface, the options and their
descriptions are as followings:
Item Options Description
The system will take the input value
The system will take the input value
The system will take the input value
as the criterion of bradycardia.
Select “Default”, the system settings
will restore to the default options.
24
Page 31
etc.
means the system will
adjust gain automatically to fit
print gain.
The system takes the selected
set to
, the system will take
length for printing each
[Not print], etc.
When “Print Mode” is set to
will print the average QRS
waveform in selected format.
[Not print]
by user as demand.
[Off]/[per 1 min]/[per 2
in]/[per 5
the ECG acquisition
process, the system will
operation according to the
will output according to the
4×3]/[Auto 3×4+1]/[Auto
3×4]/[Auto 2×6+1]/[Auto
2×6]/[Auto 3-2+1]/[Auto
3-2]/[Auto 1×12+1]/[Auto
1×12]/[Rhythm
4]/[Rhythm 3][/Rhythm 2],
Lead Gain [Smart]/[Current] The system will take the selected
[3 sec]/[4 sec]/[5 sec]/[6
Auto Strip
Rhythm Strip
sec]/[8 sec]/[10 sec]/[15
sec]/[20 sec]/[25 sec], etc.
[10 sec]/[15 sec]/[20
sec]/[25 sec]/[30 sec], etc.
option as the default print mode.
item as the gain mode of printing.
“Smart”
paper height. “Current” indicates
that the system will use the gain
of waveform on the screen as the
time as the time for printing each
strip.
When “Print Mode” is
“Rhythm”
the selected time as the time
waveform.
Average QRS
Auto-Diagnosis
Period
[4×3+Mark]/[4×3]/
[3×4+Mark]/[3×4]/
[2×6+Mark]/[2×6]/
[All]/[Only Data]/
[Only Conclusion]/
min]/[per 3 m
min]/[per 10 min]/[per 20
min]/[per 30 min]/[per 60
min], etc.
25
“Auto” or “Rhythm”, the system
The diagnosis contains data and
conclusion, which can be chosen
During
automatically activate the printing
selected time interval. When the
print mode is set to “Manual”, it
will output 3-lead, otherwise, it
Page 32
current setting mode.
the case after pressing “PRINT”.
Print ST
Vector
vector.
Depth
the waveform as required.
options.
Print-Save [Print and Save]/[Save no
Print]/[Print no Save]
Print
Adv-opr
Default ————
Note 1: The settings of auto strip, rhythm strip, average QRS, auto diagnosis and periodic
print are only optional in auto mode and rhythm mode.
Note 2: If printing time is less than 8s, no analysis, the sampling and printing time are the
same; if the printing time is equal to or greater than 8s, the sampling and analysis time keeps
the same with print time.

7.12 Lead placement

Select “Figure” in main menu to check the sketch map of lead placement (also refer to
6.4).
Press any button to exit this interface.

7.13 About

Select “About” in main menu to view the information about the device, it contains the
following contents:
Version: version No. of current programBuild: creation time of current program.FirmInfo: select it to check the firmware information of the device.
Device
Print
[Inside]/[Outside A4]
[Not Print]/[Print]
[1]/[2]/[3]/[4]
When sampling the case, the user
can select whether print / save
The user can select to use the
internal thermal paper or outside
USB printer to print.
Be used to set whether to print ST
The user can select the depth of
Select “Default”, the system
settings will restore to the default
26
Page 33

Chapter 8 Troubleshooting

8.1 Auto shutdown

The battery is almost running out, which causes overdischarge protection circuit action. The voltage of AC power supply is too high, which causes overvoltage protection circuit
action.

8.2 AC interference

Whether the device is grounded reliably? Whether the electrode or lead cable is connected correctly? Whether the electrodes and skin are daubed with enough conductive paste?. Whether the metal bed is grounded reliably? Whether the patient is touching the wall or metal parts of the bed? Whether the patient touches other people? Whether there is high power electric equipment working nearby? Such as X-ray machine
or ultrasonic device, etc.
Note: If the interference can not be removed after taking above measures, please use a AC
filter.

8.3 EMG interference

Whether the room is comfortable? Whether the patient is nervous? Whether the bed space is narrow? Whether patient speaks during recording? Whether the limb electrode is too tight?
Note: If the interference can not be removed after taking above measures, please use a
EMG filter. The ECG waveform recorded at this time will be slightly attenuated.
27
Page 34
1. Grounding cable is not
keep quiet.
close to each other.
Not regular waveform, large

8.4 Baseline drift

Whether the electrode installation is stable? Whether the connection of lead cables or electrodes is reliable? Whether the electrodes and patient skin are cleaned and are daubed with enough
conductive paste?
Whether it is caused by patient's movement or breathing? Whether the electrodes or leads are in bad connection?
Note: If the interference can not be removed after taking above measures, please use a
baseline filter.

8.5 Troubleshooting list

Phenomenon Cause of failure Solutions
Too large interference, disorderly waveform
Baseline burr
up-and-down, beeline figure
connected reliably.
2. Lead cables are not connected reliably.
3. There is AC interference.
4. Patient is nervous and can not
1. AC interference is large.
2. Patient nervous, and EMG interference is large.
1. Bad electrode conductivity.
2. Low battery.
3. Bad connection between electrodes and patient skin.
4. Loose connection between lead cables and the device's plug.
5. Bad connection between electrodes and lead cables.
28
1. Check the power cord and lead cables.
2. Let the patient prepare for the measurement.
1. Improve the environment.
2. If the bed is made of steel, replace it.
3. The power cable and lead cables are not parallel or too
1. Use alcohol of high quality.
2. Clean electrode slice and the skin under the electrode with alcohol.
3. Charge the battery.
Page 35
Baseline draft
Unclear waveform
1. Low power.
2. Patient movement.
1. Low battery.
2. The printer head surface is dirty.
3. The thermal paper problem.
1. Charge the battery.
2. Keep patient still.
1. Charge the battery.
2. Cut off the power, clean the printer head with alcohol, air dry.
3. Replace the thermal print paper with specified one.
29
Page 36
The battery status is unknown, generally it appears within 1 min after turning on.
Using AC power supply, and the battery is full or no battery in the device
Using battery, and the battery is low. It is recommended to charge the battery before use or adopt AC power supply.

Chapter 9 Maintenance

9.1 Battery

9.1.1 The device is designed with built-in full-sealed and maintenance-free rechargeable lithium
battery, also equipped with perfect auto-charging-discharging monitor system. When the
device is connected to AC power supply, the battery will be charged automatically. Battery
status will be displayed on right edge of LCD screen in powering on state, as shown in Table 9-1.
After absolutely discharged, the battery needs 5 hours to charge to 90%, and 5.5 hours to
charge to full capacity.
Table 9-1 Battery status display
No. Icon Description
a
b
c
d
e
f
g
Note: When charging the battery, the displayed status of battery level switches between icon
f to icon c.
9.1.2 The device can print for 3 hours or work for more than 10 hours in standby mode when
battery is completely charged. When the device is powered by battery, a battery icon will be
displayed on the LCD screen, showing the battery capacity in 5 modes. When the battery
capacity is too low for the device to operate, the device will turn off automatically to avoid
permanent damage to the battery.
Note: The above data is obtained by printing demo waveform under the test environment of
temperature 25, speed 25mm/s and gain 10mm/mV. In actual use, the operation time may
be shorten due to operation condition and environment.
9.1.3 The battery should be recharged in time after discharged completely. If not used for long
period, the battery should be recharged every 3 months, which can extend the life of the
battery.
9.1.4 When the battery can not be recharged or works no more than 10 minutes after fully
charged, please replace the battery.
Note
Do not try to dismantle the sealed battery without permission.The replacement of
battery shall be carried out by professional maintenance personal authorized by our
company, and the same model of rechargeable battery provided by our company
Using battery, and battery is full
Using battery, and battery level is 3/4 of battery full
Using battery, and battery level is 1/2 of battery full
Using battery, and battery level is 1/4 of battery full
30
Page 37
should be used.
Do not touch the positive and negative terminals of the battery directly with wire,
otherwise there is a danger of fire.
Do not use the battery near fire sources or in environments where the temperature
exceeds 60°C. Do not heat the battery or throw it into fire, water and avoid splashed by
water.
Do not puncture, hammer or strike the battery or destroy it by other ways, otherwise it
will cause battery overheat, smoke, deform or burn dangers.
Keep away from the battery when it appears leakage or emitting unpleasant smell. If
the battery electrolyte leaks onto the skin or clothes, clean with water immediately. If
the electrolyte accidentally enters your eyes, do not rub your eyes, immediately clean
with water and see a doctor.
If the battery reaches its service life, or battery smell, deform, discolor or distorted
appears, please stop using the battery and dispose it in accordance with local
regulations.

9.2 Recording paper

In order to ensure the quality of the ECG waveform, please use the high-speed thermal recording paper supplied or specified by the company. If you use unspecified recording paper, the recorded ECG waveform may be blurred, faded, and the paper feeding may not be smooth. This may even increase the wear of the device and shorten the service life of important parts such as the thermal print head. For information on how to purchase such recording paper, please contact your dealer or the company. Please be careful!
9.2.1 When using recording paper, it is absolutely not allowed to use recording paper with wax on the surface or in grayish/black color. Otherwise, the wax will stick to the heating part of the print head, resulting in abnormal work or damage of the print head.
9.2.2 High temperature, humidity and sunlight may cause the recording paper to change color. Please keep the recording paper in a dry and cool place.
9.2.3 Please do not place the recording paper under fluorescent light for a long time, otherwise it will affect the recording effect.
9.2.4 Please do not to put the recording paper together with the PVC plastic, otherwise the color of recording paper will change.
9.2.5 Please use the recording paper with specified dimension. Recording paper that does not meet the requirements may damage the thermal print head or silicone rubber roller.

9.3 Maintenance after use

9.3.1 Press
9.3.2 Unplug the power cord and lead cables. Hold the header of plug to disconnect, and do not
pull the cable with force directly.
9.3.3Clean the device and accessories, cover them up to against dust.
9.3.4 Store the device in a cool and dry place, avoid strong vibration when moving.
9.3.5 When cleaning the device, do not immerse it in the cleaner. Power supply must be cut off
button to shutdown the device.
31
Page 38
before cleaning. Use neutral detergents for cleaning. Do not use any detergent or disinfectant
containing alcohol.

9.4 Lead cables and electrodes

9.4.1 The connectivity of the lead cable can be detected by the multimeter. Check whether each wire of the lead cable is in good contact according to the following table. The resistance of each wire from the electrode plug to the corresponding pin in the lead cable plug should be less than 10Ω. The integrity of the lead cable must be checked regularly. Any lead wire damage will cause a false waveform of the corresponding lead or all leads on the ECG. The lead cable can be cleaned with neutral solvent. Do not use the detergent or germicide containing alcohol (Please do not immerse the lead cables in liquid for cleaning).
Note: The resistance of lead cable with defibrillation-proof protection function is about 10KΩ.
9.4.2 Bending or knotting will shorten the service life of the lead cable. When using it, please straighten the lead cable first.
9.4.3 The electrode should be well stored. After long time use, the surface of the electrode may oxidize and discolor due to corrosion and other factors, which may affect the signal acquisition. In this case, the electrode must be replaced.

9.5 Silicone rubber roller

The silicone rubber roller should be smooth and free of stains, otherwise it will affect the ECG recording effect. In order to remove the stains on the roller, please use a clean soft cloth damped with a small amount of alcohol to wipe it along the longitudinal direction, and scroll the roller in the paper conveying direction while wiping until it is clean.

9.6 Cleaning of thermal print head

Dirt and dust on the surface of the TPH can affect the clarity of the waveform. To clean the print head surface, open the paper compartment cover after turning off the device, use a clean and soft cloth dampened with alcohol to wipe the surface gently. For the residual stains on print head, moist it with a little alcohol first, then wipe with a soft cloth. Never use hard objects to scratch the surface, otherwise the print head will be damaged. Wait until the alcohol has evaporated, then close the paper compartment cover. The print head should be cleaned at least once a month during normal use.

9.7 Disposal of product scrap

The disposal of packaging materials, waste battery and end-of-life device should obey the local laws and regulations, and user should treat the scrapped products and materials properly according to the laws and regulations, and try to support the classification and recycling work.

9.8 Others

9.8.1 Do not open the device enclosure to avoid electric shock danger.
9.8.2 The device associated circuit schematics and critical parts list are only available to authorized service station or maintenance personnel, who is responsible for maintenance of the device.
9.8.3 The device belongs to measuring instrument. User should send the device to national designated inspection institution for inspection according to the requirements of the national
32
Page 39
metrological verification procedure. The device shall be inspected at least once per year, and all the accessories should be inspected and maintained regularly (at least once every six months).
33
Page 40
Name
Quantity
Electrocardiograph
1 pc
Limb electrodes (limb clip)
1 set
ECG lead cable
1 pc
Potential equalization wire
1 pc
Power cord
1 pc
User manual
1 pc
Recording paper
1 pc

Chapter 10 Packing List and Accessories

10.1 Accompanying accessories

When the device is shipped from the factory, the intact packaging should contain the
following contents, as shown in Table 10-1:
Table 10-1 Packing list and accessories

10.2 Notes

10.2.1 Please follow the instructions on the package when opening the package.
10.2.2 After unpacking, please check the accessories and accompanying documents in
accordance with the packing list, then start inspecting the device.
10.2.3 If the packaging content does not meet the requirement or the device does not work
properly, please contact our company immediately.
10.2.4 Please use the accessories provided by our company, otherwise the performance and
safety of the device may be affected. If accessories provided by other company need to be used,
please first consult the after-sales service of our company, or we will not responsible for any
caused damages.
10.2.5 The package shall be kept properly for future use in regular maintenance or device
repair.
34
Page 41
No.
Parameter
Unit
1
HR
bpm
2
PR-interval
ms
3
P-duration
ms
4
QRS-duration
ms
5
T-duration
ms
6
QT/QTc
ms
7
P/QRS/T electric axis
deg
8
R(V5)/S(V1)
mV
9
R(V5)+S(V1)
mV
No.
Item
2
Sinus mode Bradycardia
3
Sinus mode Tachycardia
5
Right atrium Hypertrophy
6
Dual atrium Hypertrophy
8
Cardiac electric axis normal

Appendix I ECG Automated Measurement&Interpretation Guide

1. Preface

The appendix describes the functions of ECG automated measurement and automated
interpretation. It explains the specific implementation method, algorithm and formulas related
to these two functions, as well as the content output by the automated measurement and
automated interpretation.
According to the requirement of IEC60601-2-51:2003 Medical electrical equipment - Part
2-51: Particular requirements for safety, including essential performance, of recording and
analysing single channel and multichannel electrocardiographs, Clause 50 Accuracy of operating
data, the appendix gives a description of verification process and results of the performance for
automated measurement and automated interpretation.

2. Automated measurement parameters and Automated interpretation items

The output measurement parameter, interpretation item and others that require
explanation are as follows:

2.1 Measurement parameters

2.2 Interpretation items

1 No abnormal
4 Left atrium Hypertrophy
7 QRS low voltage
35
Page 42
9
Left axis deviation
10
Right axis deviation
11
Completeness Right Bundle branch block
12
Completeness Left Bundle branch block
13
No Completeness Right Bundle branch block
14
No Completeness Left Bundle branch block
15
V1 shows RSR' type
16
Left anterior fascicular block
17
Left posterior fascicular block
18
Left ventricular hypertrophy
19
Right ventricular hypertrophy
20
I atrioventricular block
21
Early anteroseptal MI
22
Possible acute forepart anteroseptal MI
23
Old anteroseptal MI
24
Early anterior MI
25
Possible acute anterior MI
26
Old anterior MI
27
Early extensive anterior MI
28
Possible acute extensive anterior MI
29
Old extensive anterior MI
30
Early apical MI
31
Acute apical MI
32
Old apical MI
33
Early anterolateral MI
34
Possible acute anterolateral MI
35
Old anterolateral MI
36
Early high lateral MI
37
Possible acute high lateral MI
38
Old high lateral MI
39
Early inferior MI
40
Possible acute inferior MI
41
Old inferior MI
42
Early inferolateral MI
43
Possible acute inferolateral MI
36
Page 43
44
Old inferolateral MI
45
ST depression, mild anteroseptal myocardial ischemia
46
ST depression, mild anterior myocardial ischemia
47
ST depression, mild extensive anterior myocardial ischemia
48
ST depression, mild apical myocardial ischemia
49
ST depression, mild anterolateral myocardial ischemia
50
ST depression, mild high lateral myocardial ischemia
51
ST depression, mild inferior myocardial ischemia
52
ST depression, mild inferolateral myocardial ischemia
53
ST depression, anteroseptal myocardial ischemia
54
ST depression, anterior myocardial ischemia
55
ST depression, extensive anterior myocardial ischemia
56
ST depression, apical myocardial ischemia
57
ST depression, anterolateral myocardial ischemia
58
ST depression, high lateral myocardial ischemia
59
ST depression, inferior myocardial ischemia
60
ST depression, inferolateral myocardial ischemia
and diagnosis
to above description
Application site
hospitals
sensitivity and specificity.
2.3 Intended use
The intended use of the Automated Measurement&Interpretation function is shown as
below:
Application
To detect the abnormal of heart of human body, examination items refer
Population Teenagers and adults, age range: 12-87
Accuracy
Others

3. Algorithm description

This section describes the algorithm, formulas and judgment conditions for interpretation
items related to functions of ECG automated measurement and automated interpretation.
The 12-lead sync ECG waveform passes through the filter (AC, EMG, DFT (if has, and open))
into the module of automated measurement and automated interpretation.
The module of automated measurement and automated interpretation mainly includes
process of find the cardiac impulse location, find the beginning/end for each wave, amplitude
calculation, parameters calculation, and interpretations judgment based on known parameters.
The accuracy of this function is reflected by the balance performance of
This function does not generate any alarm when using, so it should be
operated by professional or trained personal.
37
Page 44
Start
End
The workflow is shown as below:
ECG waveform sampling
Recognize all R points by slope method
Waveform superposition taking R point as center
Determine the positions of each wave
Calculate the amplitudes of each wave
Get measurement parameter, interpretation item

3.1 Find the cardiac impulse location

1) Data preprocessing, obtain the absolute value trend of slope for each lead; then
superimpose each absolute value, obtain the superimposed graph of absolute value of slope.
2) Smoothing filter the superimposed graph on average of width 80ms, obtain the
analytical data source DDD.
3) Find the cardiac impulse location, give an initial threshold for searching, orderly scan the
data in the analytical data source DDD, then compare it with the threshold value:
When the value is greater than the threshold, it may be the beginning of qrs-complex.
If the distance from the previous qrs-complex to the current location is less than
150ms, then give up the location.
Otherwise, take the 1/4 of threshold value as a reference, find the beginning of
qrs-complex within 100ms before the current location.
When the value is less than the threshold value, it may be the end of qrs-complex.
Take the 1/4 of threshold value as a reference, find the end of qrs-complex.
If the found qrs-complex is wide, this qrs-complex shall be excluded. Otherwise, save
the found qrs-complex.
4) Locate: after found the qrs-complex, search the max value point between the beginning
point and end point in the ecg original data, mark the point as cardiac impulse location.
38
Page 45
5) Dynamically threshold adjustment: after found the cardiac impulse location, use the
value at the cardiac impulse location for the dynamically adaptive adjustment of the threshold
value. Define the threshold value as 1/3 of the average of the nearest three cardiac impulses.
6) After found the cardiac impulse location, compute the RR-interval and accumulate it
with the previous RR-intervals, then count the number of accumulated RR-intervals.
7) Continue searching until the end of data, and calculate the global average value for
RR-intervals at the same time.

3.2 Find the beginning/end for each wave

The beginning/end of qrs-complex has been approached in above cardiac impulse locating
process, but it is mainly in order to assist to find the cardiac impulse location; in addition, the
location is searched based on the slope threshold value, which is imprecise. Here, according to
the found cardiac impulse location, the beginning/end of qrs-complex will be sought accurately.
Name the cardiac impulse location as the peak of R-wave.
1. Read data
1) Read one data of qrs-complex: take the peak of R-wave as reference, locate directly to the
original ecg file, read a piece of data containing the qrs-complex.
2) Preprocessing: superimpose the absolute value of slope for 12-lead signals.
3) Use the preprocessed data to carry on the searching of QRS-complex, P-wave and T-wave as
the followings.
4) Read the next data of qrs-complex, repeat step 2 and step 3 until the analyzing of all
qrs-complex are finished.
2. Find QRS-complex
1) Calculate the threshold value of S-wave: search the minimal value within 200ms after the
peak of R-wave, take the value that equals to minimal value plus 0.4, as the threshold value for
finding the end of S-wave.
2) Find the beginning of Q-wave: take 0.5 as the threshold vale, search forwardly starting from
R-wave, a point that less than the threshold value, within 0ms-200ms before the peak of
R-wave, which is the beginning of Q-wave.
3) Find the end of S-wave: search backwardly starting from R-wave, a point that less than the
threshold value of the end of S-wave, within 0ms-200ms after the peak of R-wave, which is the
end of S-wave.
3. Find P-wave
1) Peak of P-wave: search the max value within 30ms-100ms before the beginning of Q-wave,
temporarily mark the point as the peak of P-wave.
2) Find the end of P-wave: search the minimal value between the peak of P-wave and the
beginning of Q-wave, the minimal value plus 0.05 is the threshold value, use the threshold
value to find the end of P-wave.
3) Find the beginning of P-wave: search the minimal value within 150ms before the peak of
P-wave, the minimal value plus 0.06 is the threshold value, use the threshold value to find the
beginning of P-wave.
4) If the found P-wave is narrow, research the P-wave according to the following steps.
39
Page 46
5) Change the searching range of 30ms-100ms to 100ms-350ms in step 1, repeat step 1-4.
6) If the found P-wave is still narrow, it means that P-wave doesn’t exist.
4. Find T-wave
1) Peak of T-wave: search the max value within 30ms-300ms after the end of QRS-complex,
save it as the peak of T-wave.
2) Threshold value of the beginning of T-wave: search the minimal value within 0ms-100ms
after the end of QRS-complex, the minimal value plus 1/10 of the peak value of T-wave is the
threshold for finding the beginning of T-wave.
3) Threshold value of the end of T-wave: search the minimal value within 200ms after the peak
of T-wave, the minimal value plus 1/10 of the peak value of T-wave is the threshold for finding
the end of T-wave.
4) Find the beginning of T-wave: in the range between the minimal value in step2 and the peak
of T-wave, find a point that less than the threshold value of the beginning of T-wave, the point
is the beginning of T-wave.
5) Find the end of T-wave: in the range between the minimal value in step3 and the peak of
T-wave, find a point that less than the threshold value of the end of T-wave, the point is the end
of T-wave.
5. Explanation of equipotential segment
In searching the QRS-complex, this algorithm adopts the analysis method of superposition of
the slopes for all leads, therefore, the equipotential segments before and after the
QRS-complex are partly included in the start and end points of the QRS-complex. It is depends
on the number of leads containing equipotential segments. If there are more leads containing
equipotential segments, the slope value will be smaller after superposition, so it is difficult to
meet the threshold condition, and only a small part of the equipotential segments is counted to
the start and end points of the QRS-complex. On the contrary, if there are less leads containing
equipotential segments, a large part of the equipotential segments will be counted to the start
and end points of the QRS-complex. Anyway, the equipotential segments before and after the
QRS-complex are partly included in the QRS-complex duration.

3.3 Amplitude measurement

After finding the position of each wave, i.e. the start and end points of P wave, QRS
complex and T wave, use the following method to measure P, Q, R, S, ST and T waves of each
lead.
1. P-wave
Calculate the average value of the data 20ms before the start point of P wave, and use this
average value as the baseline of P wave. Find the max value between the start point and end
point of P wave, the difference between the max value and the baseline would be the
amplitude of P wave.
2. Q/R/S wave
Calculate the average value of the data 10-30ms before the start point of QRS complex, and use
this average value as the baseline of QRS complex. Search boundary points that exceeding the
baseline from the start point of Q wave to the end point of S wave. Each adjacent two boundary
40
Page 47
points forms a sub-wave. Determine whether each sub-wave is a recognizable minimum wave
(see the definition below). If it is a recognizable minimum wave, first identify its direction. If it is
above the QRS baseline, it is R wave, if it is below the baseline, it is Q wave or S wave. Find the
extreme value of this wave, and the difference between the extreme value and the baseline is
the amplitude of Q/R/S wave.
Note: If there is only one downward wave, its amplitude should be respectively recorded in the
amplitude of Q wave and S wave.
3. ST segment
Take above baseline of QRS complex as the ST baseline. Calculate the differences between the
ST baseline and the points at 40ms and 60ms after the end point of QRS complex, and calculate
the average value of these two differences, the average value is the amplitude of ST segment.
4. T-wave
Calculate the average value of the data 20-50ms after the end point of T wave, and
average this value with the QRS baseline in 2, then use the result as the baseline of T wave. Find
the max value between the start point and end point of T wave, the difference between the
max value and the baseline would be the amplitude of T wave.
5. Recognition of minimum wave
The minimum wave can be recognized by the algorithm according to the requirement of
IEC60601-2-51:2003 Medical electrical equipment - Part 2-51: Particular requirements for
safety, including essential performance, of recording and analysing single channel and
multichannel electrocardiographs, Annex GG, Clause GG.5 Definition of waveforms,
measurement of minimum waves. The wave that meet the following conditions is the minimum
wave that can be recognized by the algorithm.
1)The signal part under consideration shows clearly two opposite slopes with at least
one turning point in between;
2)The signal part under consideration deviates at least 30μV from the reference level
for a duration of at least 6ms;
3)The minimum observable duration of wave under consideration is 12ms and
amplitude ≥30μV.

3.4 Calculation after intervals determination

The following parameters are determined according to the requirement of
IEC60601-2-51:2003 Medical electrical equipment - Part 2-51: Particular requirements for safety,
including essential performance, of recording and analysing single channel and multichannel
electrocardiographs, Annex GG Definitions and rules for the measurement of
ELECTROCARDIOGRAMS.
No. Parameter Calculation
1 HR 60 / RR
2 PR-interval Qs - Ps③
3 P-duration Pe④ - Ps③
41
Page 48
RR
QT
PI
180 )3 S),S(S arctan(2.0
IIIII
××+×
4 QRS-duration Se⑤ - Qs
5 T-duration Te⑦ - Ts⑥
6 QT Te⑦ - Qs
7 QTc
Electric axis formula:
P electric axis:
S
: voltage sum from the beginning point to the end
III
point of P-wave on lead III
S
: voltage sum from the beginning point to the end
I
8
P/QRS/T electric
axis
point of P-wave on lead I
QRS electric axis:
S
: voltage sum from the beginning point to the end
III
point of QRS-complex on lead III
S
: voltage sum from the beginning point to the end
I
point of QRS-complex on lead I
T electric axis:
S
: voltage sum from the beginning point to the end
III
point of T-wave on lead III
S
: voltage sum from the beginning point to the end
I
point of T-wave on lead I
9 R(V5) Height (voltage value) of R-wave on lead V5
10 S(V1) Height (voltage value) of S-wave on lead V1
Note:
RR: RR-interval Qs: beginning of the Q-wave Ps: beginning of the P-wave
42
Page 49
No.
Item
Rule of interpretation
No abnormal
Cardiac electric axis normal
Left axis deviation
Right axis deviation
branch block
block
branch block
branch block
V1 shows RSR' type
Pe: end of the P-wave Se: end of the S-wave Ts: beginning of the T-wave Te: end of the T-wave PI: 3.1415926

3.5 Interpretations judgment based on parameters

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
Sinus mode Bradycardia
Sinus mode Tachycardia
Left atrium Hypertrophy
Right atrium Hypertrophy
Dual atrium Hypertrophy
QRS low voltage
Completeness Right Bundle
Completeness Left Bundle branch
No Completeness Right Bundle
No Completeness Left Bundle
Left anterior fascicular block
No any abnormal are detected
Sinus P-wave, PR-interval between
110ms-210ms, HR≤*/min, general *=50
Sinus P-wave, PR-interval between
110ms-210ms, HR≥ */min, general *=100
P-wave of leads I, II, aVL shall meet the
conditions: width increase of P-wave≥110ms,
or P-wave displays in double-peak type,
value of peak to peak ≥40ms
For leads I, II, aVF, amplitude of P-wave
≥0.25mV, or P-wave is sharp
For leads I, II, aVF, amplitude of P-wave
≥0.25mV and P-wave duration >110ms
Voltage of I-aVF limb leads <0.5mV,
and voltage of V1-V6 chest leads <0.8mV
QRS-axis between 30 to 90 degree
QRS-axis between -90 to-30 degree
QRS-axis between 120 to 180 degree
QRS-duration>120ms, R-wave of lead
V1 or aVR is wide (width of R-wave>80ms)
QRS-duration>120ms, R-wave of lead
V5 or V6 is wide
QRS-duration<120ms, R-wave of lead
V1 or aVR is wide (width of R-wave>80ms)
QRS-duration<120ms, R-wave of lead V15 or
V6 is wide (width of R-wave>80ms)
QRS-complex of lead V1 is RSR' type
QRS-duration<110ms, QRS-axis <-30
degree, lead I and lead aVL are qR type, and
Q-wave duration<20ms, lead II, III and aVF
are rS type.
43
Page 50
I atrioventricular block
anteroseptal MI
MI
Acute apical MI
QRS-duration<110ms, QRS-axis >90
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
Left posterior fascicular block
Left ventricular hypertrophy
Right ventricular hypertrophy
Early anteroseptal MI
Possible acute forepart
Old anteroseptal MI
Early anterior MI
Possible acute anterior MI
Old anterior MI
Early extensive anterior MI
Possible acute extensive anterior
Old extensive anterior MI
Early apical MI
degree, lead I and lead aVL are rS type, lead
II, III and aVF are qR type, and Q-wave of
lead II and III <20ms.
R amplitude of lead I >1.5mV, R
amplitude of lead V5 >2.5mV, R amplitude of
lead aVL >1.2mV, R amplitude of lead
aVF >2mV, R amplitude of lead V5 minus S
amplitude of lead V1 >4mV (male) or 3.5mV
(female).
R amplitude of lead aVR >0.5mV, R
amplitude of lead V1 >1mV, R amplitude of
lead V1 minus S amplitude of lead
V5 >1.2mV, R amplitude of lead V1 is larger
than S amplitude, R amplitude of lead V5 is
smaller than S amplitude.
PQ interval >210ms
Early myocardial infarction change of leads
V1, V2, V3, no change of leads V4, V5.
Acute myocardial infarction change of
leads V1, V2, V3, no change of leads V4, V5.
Old myocardial infarction change of
leads V1, V2, V3, no change of leads V4, V5.
Early myocardial infarction change of leads
V3, V4, V5, no change of leads V1, V2, V6.
Acute myocardial infarction change of leads
V3, V4, V5, no change of leads V1, V2, V6.
Old myocardial infarction change of leads V3,
V4, V5, no change of leads V1, V2, V6.
Early myocardial infarction change of leads
V1, V2, V3, V4, V5.
Acute myocardial infarction change of leads
V1, V2, V3, V4, V5.
Old myocardial infarction change of leads V1,
V2, V3, V4, V5.
Early myocardial infarction change of leads
V4, V5, no change of leads V1, V2, V3.
Acute myocardial infarction change of leads
44
Page 51
myocardial ischemia
myocardial ischemia
anterior myocardial ischemia
myocardial ischemia
ST depression, mild anterolateral
V4, V5, no change of leads V1, V2, V3.
32
33
34
35
36
37
38
39
40
41
42
43
44
ST depression, mild anteroseptal
45
46
47
48
49
Old apical MI
Early anterolateral MI
Possible acute anterolateral MI
Old anterolateral MI
Early high lateral MI
Possible acute high lateral MI
Old high lateral MI
Early inferior MI
Possible acute inferior MI
Old inferior MI
Early inferolateral MI
Possible acute inferolateral MI
Old inferolateral MI
ST depression, mild anterior
ST depression, mild extensive
ST depression, mild apical
Old myocardial infarction change of leads V4,
V5, no change of leads V1, V2, V3.
Early myocardial infarction change of leads I,
aVL, V4, V5, V6
Acute myocardial infarction change of leads
I, aVL, V4, V5, V6.
Old myocardial infarction change of leads I,
aVL, V4, V5, V6
Early myocardial infarction change of leads I,
aVL, no change of leads II, III, aVF, V4, V5, V6.
Acute myocardial infarction change of leads
I, aVL, no change of leads II, III, aVF, V4, V5,
V6.
Old myocardial infarction change of leads I,
aVL, no change of leads II, III, aVF, V4, V5, V6.
Early myocardial infarction change of leads II,
III, aVF, no change of leads I, aVL.
Acute myocardial infarction change of leads
II, III, aVF, no change of leads I, aVL.
Old myocardial infarction change of leads II,
III, aVF, no change of leads I, aVL.
Early myocardial infarction change of leads I,
II, III, aVL, aVF.
Acute myocardial infarction change of leads
I, II, III, aVL, aVF.
Old myocardial infarction change of leads I,
II, III, aVL, aVF.
Mild ST-segment depression of leads V1, V2,
V3, and no change of leads V4, V5.
Mild ST-segment depression of leads V3, V4,
V5, and no change of leads V1, V2, V6.
Mild ST-segment depression of leads V1, V2,
V3, V4, V5.
Mild ST-segment depression of leads V4, V5,
and no change of leads V1, V2, V3.
Mild ST-segment depression of leads I, aVL,
45
Page 52
myocardial ischemia
myocardial ischemia
myocardial ischemia
myocardial ischemia
myocardial ischemia
ischemia
myocardial ischemia
ischemia
myocardial ischemia
ischemia
myocardial ischemia
V4, V5, V6.
ST depression, mild high lateral
50
51
52
53
54
55
56
57
58
59
60
Note:
Early myocardial infarction: normal Q-wave, ST elevation or ST slope elevation
Acute myocardial infarction: abnormal Q-wave, ST elevation or ST slope elevation
Old myocardial infarction: abnormal Q-wave, no ST elevation.
Abnormal Q-wave:
For leads I, II, III, avR, avL, avF, V3, V4, V5, V6, voltage of Q-wave <-0.3mV, or 4 times of
negative wave of Q-wave> voltage of R-wave and R’-wave, and/or Q-duration>40ms.
For leads V1, V2, voltage of Q-wave <-0.08mV and Q-duration>10ms.
ST elevation:
For leads I, II, III, avR, avL, avF, V4, V5, V6, the voltage of ST segment at 60ms point >0.1mV,
and for leads V1, V2, V3, the voltage at 60ms point >0.3mV.
ST slope elevation:
Voltage of ST segment at 20ms point>=voltage of J point, voltage at 40ms point >= the one
at 20ms, voltage at 60ms point >= the one at 40ms, with change of ST elevation.
ST depression, mild inferior
ST depression, mild inferolateral
ST depression, anteroseptal
ST depression, anterior myocardial
ST depression, extensive anterior
ST depression, apical myocardial
ST depression, anterolateral
ST depression, high lateral
myocardial ischemia
ST depression, inferior myocardial
ST depression, inferolateral
Mild ST-segment depression of leads I, aVL,
and no change of leads II, III, aVF, V4, V5, V6.
Mild ST-segment depression of leads II, III,
aVF, and no change of leads I, aVL.
Mild ST-segment depression of leads I, II, III,
aVL, aVF.
Severe ST-segment depression of leads V1,
V2, V3, and no change of leads V4, V5.
Severe ST-segment depression of leads V3,
V4, V5, and no change of leads V1, V2, V6.
Severe ST-segment depression of leads V1,
V2, V3, V4, V5.
Severe ST-segment depression of leads V4,
V5, and no change of leads V1, V2, V3.
Severe ST-segment depression of leads I,
aVL, V4, V5, V6.
Severe ST-segment depression of leads I,
aVL, and no change of leads II, III, aVF, V4,
V5, V6.
Severe ST-segment depression of leads II, III,
aVF, and no change of leads I, aVL.
Severe ST-segment depression of leads I, II,
III, aVL, aVF.
46
Page 53
Verification
Database
Database items
CAL30000 ANE20000 ANE20001 ANE20002
database
n
CSE diagnostic database
D_0001~D_1220

4. Data sources and data preprocessing

4.1 Data sources

According to the requirement of IEC60601-2-51:2003 Medical electrical equipment - Part
2-51: Particular requirements for safety, including essential performance, of recording and
analysing single channel and multichannel electrocardiograph, the CSE measurement database,
CSE diagnostic database, CTS calibration database and customized data shall be used to
evaluate the function of automated measurements and automated interpretations.
CAL05000 CAL10000 CAL15000 CAL20000
Automated
measuremen
t
Automated
interpretatio

4.2 CTS introduction

The CTS computerized ECG conformance testing project was launched in 1989 by the
European Union. This project laid the foundation for computerized ECG conformance testing
service. Currently, about 20 types of waveform have been designed derived from the test
signals having an infinite length, these signals are part of the CTS-ECG test database, and have
proven their effectiveness in a series of official tests. According to the requirement of
IEC60601-2-51:2003 Medical electrical equipment - Part 2-51: Particular requirements for safety,
including essential performance, of recording and analysing single channel and multichannel
electrocardiograph Clause 50.101.1, 13 data (CAL05000, CAL10000, CAL15000, CAL20000,
CAL20002, CAL20100, CAL20110, CAL20160, CAL20200, CAL20210, CAL20260, CAL20500,
CAL30000) are used in the automated parameters verification for this test.

4.3 CSE introduction

The EU CSE (Common Standards for Quantitative Electrocardiography) ECG database
contains 3-lead measurement database of collection1 and collection2, 12-lead measurement
database of collection3 and collection4, and a diagnostic database of collection5. In which, the
12-lead measurement database contains 250 groups of interference data; Diagnostic database
contains 1220 cases of short-term ECG recording. The primary development purpose of using
12-lead or 15-lead is to evaluate the performance of the automatic ECG analyzer. In addition to
the normal data, the database also includes clinically confirmed ECGs of variety cases, such as
left ventricular hypertrophy, right ventricular hypertrophy, every part of myocardial infarction
and ventricular hypertrophy accompanying myocardial infarction. The database has made a
CTS database
CSE measurement
Customized data 000001000549
CAL20002 CAL20100 CAL20110 CAL20160
CAL20200 CAL20210 CAL20260 CAL20500
MA_0001MA0125
47
Page 54
Item
Number
Normal
382
Left ventricular hypertrophy
183
Right ventricular hypertrophy
55
Biventricular hypertrophy
53
Anterior myocardial infarction
170
Inferior myocardial infarction
273
Complex myocardial infraction
104
Synthetical accuracy
1220
data
223 female, average age 59.70, standard deviation 22.63.
great contribution to the study of electrocardiology, which is, the CSE group published a report
on the recommended standard for general ECG measurements based on the investigation and
study of the database, which has been widely recognized by the world.
CSE database diagnostic items:

4.4 Customized data

4.4.1 Data description

Customized
Description
Total recording
number
Race Yellow race
Coverage of
age, gender
Sampling data 12-lead ECG data (I, , , AVR, AVL, AVF, V1, V2, V3, V4, V5, V6),
Remark The interpretation conclusion of customized data is determined by the
549
Aged from 17 to 87, average age 57.23, standard deviation 21.32;
326 male, average age 55.54, standard deviation 19.81;
sampling frequency of each channel: 1kHz, amplitude quantization:
2.4μV/LSB.
physician diagnostic results of cardiac catheterization and ultrasonic
examination, and the ECG judgment result in physical examination, the
details as blow:
1) Normal ECG
Determined by the diagnostic result that judged as normal in cardiac
catheterization and ultrasonic examination, and the result that judged as
normal in physical examination.
2) Atrium hypertrophy
Determined by the diagnostic results of ultrasonic examination.
48
Page 55
cardiac functions or shape.
3) Myocardial infarction and myocardial ischemia
Determined by the physician diagnostic results of cardiac catheterization.
4) Tachycardia, bradycardia, low voltage, axis
Determined by the diagnostic results of ultrasonic examination.
5)Conduction block
Determined by the physician diagnostic results of cardiac catheterization.
The standard of normal population in the customized database: physical
examination is normal, no heart disease or other diseases that may affect

4.5 Data coverage of verification for automated interpretation

Analyzing the content of CSE diagnostic database and customized data, the overall
condition and coverage of statistical samples are shown as below:
49
Page 56
Note:
The heart abnormalities such as posterior myocardial ischemia, early posterior MI and old posterior MI are not included in the database. These abnormalities and other heart disorders not contained in above sheet won’t be regarded as the judgment object for the verification of automated interpretation accuracy.

4.6 Data preprocessing

4.6.1 CTS preprocessing

The 16 cases (CAL05000, CAL10000, CAL15000, CAL20000, CAL20002, CAL20100, CAL20110, CAL20160, CAL20200, CAL20210, CAL20260, CAL20500, CAL30000, ANE20000, ANE20001, ANE20002) from CTS-ECG shall be processed for voltage conversion and frequency conversion for resampling as the applicable format in the system. Then cases will be imported to the device. After that, the verification of automated measurement parameters will be carried on.

4.6.2 CSE preprocessing

The cases (MA_0001~MA0125, D_0001~D_1220) from the CSE shall be processed for voltage conversion and frequency conversion for resampling as the applicable format in the system. Then cases will be imported to the device. After that, the case of MA_0001~MA0125 shall be used for the following verification of automated measurement parameters, and the case of D_0001~D_1220 shall be used for the following verification of automated interpretation.
50
Page 57
Start
CTS preprocessing
Import preprocessing data to device
ECG automated measurement parameter
Calculate the difference of measurement and reference
Calculate mean differences
Eliminate the largest two deviations from the mean
Recalculate mean difference and standard deviation
End

4.6.3 Customized data preprocessing

The customized initial case files shall be processed for voltage conversion and frequency conversion for resampling as the applicable format in the system. Then cases will be imported to the device. After that, the verification of automated interpretation will be carried on.

5. Process and Result of Verification

5.1 Verification of measurement function

5.1.1 Verification and Process for CTS measurement database

The cases (CAL05000, CAL10000, CAL15000, CAL20000, CAL20002, CAL20100, CAL20110,
CAL20160, CAL20200, CAL20210, CAL20260, CAL20500, CAL30000, ANE20000, ANE20001,
ANE20002) imported to the device shall be used to verify the automated measurement
parameters.

5.1.2 Verification and Process for CSE measurement database

Import the converted case files into the device, add appropriate database records, then
51
Page 58
waveform for all case files can be reviewed in the device, therefore the automated
measurement parameters can be obtained.
Eliminate the cases existing obvious error for the diagnostic parameters (P-wave location is
wrong) from the CSE database.
Make a comparison between the ECG analytical parameters (the beginning/end of P-wave,
QRS-complex and T-wave) and the diagnostic parameters (the beginning/end of P-wave,
QRS-complex and T-wave) provided by CSE database. Draw the two groups of waveform and
mark the location of the beginning/end of P-wave, QRS-complex and T-wave corresponding to
each case. The picture provides a visualized comparison, so the mean and standard deviation of
the differences can be calculated. According to the requirement of IEC60601-2-51:2003 Medical
electrical equipment - Part 2-51: Particular requirements for safety, including essential
performance, of recording and analysing single channel and multichannel electrocardiograph,
the four largest deviations from the mean shall be eliminated before recalculation of mean and
standard deviation of the differences.
Flow diagram of CSE measurement database verification process
52
Page 59
Eliminate obvious unfit cases
Frequency conversion
Read expert diagnostic marks
Voltage conversion
Read initial case of DCD file
Get ECG data file
Automated measurement parameters
Conclude mean value of comparison
Compare automated measurement parameter and expert diagnostic mark
Summarize the comparison results
Recalculate the mean value and variation of comparison result
Start
End
Eliminate the four largest deviations from the mean
53
Page 60
Amplitude
Mean difference (uV)
Standard deviation (uV)
P-wave
-1.70
5.72
Q-wave
7.51
18.07
R-wave
-18.05
21.70
S-wave
7.77
18.58
ST-segment
0.15
4.24
T-wave
-5.81
8.03
Interval&Duration
Mean difference (ms)
Standard deviation (ms)
P-duration
0.99
13.46
PR-interval
3.65
9.68
QRS-duration
-1.69
6.11
QT-interval
-2.32
20.69
parameters
Disclosed differences
P-duration
High frequency
-5.65
12.33

5.1.3 Verification results

5.1.3.1 Accuracy of amplitude measurements

Calibration and analytical ECGs shall be used to measure the amplitude value, the
summary as follows:
Note: In amplitude measurement, for large-amplitude ECG, such as CAL30000, it is
necessary to adjust to 0.5 times the gain before testing.

5.1.3.2 Accuracy of absolute interval and wave duration measurements

Calibration and analytical ECGs shall be used to measure the global interval and wave
duration (including Q-wave ,R-wave ,S-wave), the summary as follows:
Interval&Duration Mean difference (ms) Standard deviation (ms)
P-duration -5.70 1.88
PQ-interval -2.58 1.94
QRS-duration -0.23 3.26
QT-interval -6.70 4.37

5.1.3.3 Accuracy of interval measurements on biological ECGs

CSE database shall be used to evaluate the accuracy of interval measurements on
biological ECGs, the summary as follows:

5.1.3.4 Stability of measurements against NOISE

The test is carrying on according to MA-series data (008, 011, 013, 014, 015, 021, 026, 027,
042, 061) in CSE database.
Global
measurement
Type of added
NOISE
Mean (ms) Standard deviation (ms)
54
Page 61
P-duration
Line frequency
-0.25.
12.71
P-duration
Base-line
-4.90
33.15
QRS-duration
High frequency
-0.95
5.13
QRS-duration
Line frequency
1.35
4.71
QRS-duration
Base-line
-1.55
7.68
QT-interval
High frequency
-14.55
6.51
QT-interval
Line frequency
-8.55
20.73
QT-interval
Base-line
36.20
64.47
The biological ECGs are fed into the device in form of digital signals, then the
measurement value can be obtained by calculation.
Test condition:
a) without NOISE
b)with 25uV high frequency
c) with 50uV peak to valley 50Hz/60Hz sinusoidal line frequency NOISE
d) with 1mV peak to valley 0.3Hz sinusoidal base-line NOISE
For each NOISE level above, the differences of measurements between the NOISE-free
ECGs and the ECGs with NOISE shall be determined. The two largest deviations from the mean
shall be estimated before calculation of mean and standard deviation of differences.
55
Page 62
Frequency conversion
Voltage conversion
Read initial DCD files
Get ECG data files
Automated interpretation item
Start
Summarize comparison results, draw conclusions
End

5.2 Verification of interpretation function

5.2.1 Verification process

5.2.1.1 CSE diagnostic database

Compare automated interpretation with expert diagnosis
56
Page 63
ecg format as system required
Import to device
Start comparing
case
Auto measurement parameter
Start

5.2.1.2 Customized database

Initial case data
Draw ECG waveform
Identify QRS-complex
QRS-complex superposition
Conclude automated interpretation items
Expert diagnosis
Summarize the global statistical results of comparison for each
End
57
Page 64
r
value % 1 No abnormal
585
92.01
79.16
97.38 2 Sinus mode Bradycardia
191
96.68
99.73
98.64 3 Sinus mode Tachycardia
78
97.44
96.49
96.90 4 Left atrium Hypertrophy
51
51.09
99.89
81.82 5 Right atrium Hypertrophy
43
42.64
99.66
50.00 6 Dual atrium Hypertrophy
22
93.58
99.14
60.19 7 QRS low voltage
5
96.37
99.36
63.25 8 Cardiac electric axis normal
733
98.36
89.13
98.79 9 Left axis deviation
168
98.65
89.40
98.18
10
Right axis deviation
107
98.23
88.99
94.90
block
block
branch block
branch block
15
V1 shows RSR' type
13
90.32
91.14
65.12
16
Left anterior fascicular block
26
91.43
93.25
71.11
17
Left posterior fascicular block
18
89.29
97.37
52.63
18
Left ventricular hypertrophy
236
41.37
92.65
70.36
19
Right ventricular hypertrophy
108
39.75
93.47
65.39
20
I atrioventricular block
13
94.58
91.67
80.64
21
Early anteroseptal MI
10
83.33
99.94
90.91
anteroseptal MI
23
Old anteroseptal MI
26
92.00
98.90
86.47
24
Early anterior MI
77
93.90
88.22
71.96
25
Possible acute anterior MI
10
80.00
99.72
44.44
26
Old anterior MI
13
24.00
99.66
50.00
27
Early extensive anterior MI
24
79.67
99.43
41.18
28
Possible acute extensive anterior
16
81.82
99.66
75.00

5.2.2 Verification results

No. Item
ECGs
numbe
Sensitivit
y %
Specifici
ty %
Positive
predictive
11 Completeness Right Bundle branch
12 Completeness Left Bundle branch
13 No Completeness Right Bundle
14 No Completeness Left Bundle
22 Possible acute forepart
28 97.00 89.50 95.45
32 97.73 89.65 91.43
41 96.86 89.83 82.35
47 94.68 89.83 89.66
27 16.67 98.73 91.89
58
Page 65
MI
29
Old extensive anterior MI
30
90.91
88.05
37.04
30
Early apical MI
15
88.32
87.21
88.54
31
Acute apical MI
21
78.12
78.66
53.85
32
Old apical MI
19
79.63
89.94
80.00
33
Early anterolateral MI
36
77.51
79.94
83.33
34
Possible acute anterolateral MI
9
28.57
99.77
33.33
35
Old anterolateral MI
14
70.00
93.60
50.00
36
Early high lateral MI
16
79.65
95.78
80.42
37
Possible acute high lateral MI
8
81.60
99.94
85.71
38
Old high lateral MI
23
81.82
99.66
60.00
39
Early inferior MI
31
88.89
95.00
40.00
40
Possible acute inferior MI
11
76.00
99.60
61.11
41
Old inferior MI
101
96.07
99.24
93.44
42
Early inferolateral MI
73
98.77
96.82
75.94
43
Possible acute inferolateral MI
29
11.11
99.94
50.00
44
Old inferolateral MI
28
84.62
99.83
78.57
myocardial ischemia
myocardial ischemia
anterior myocardial ischemia
myocardial ischemia
myocardial ischemia
myocardial ischemia
myocardial ischemia
myocardial ischemia
myocardial ischemia
ischemia
45 ST depression, mild anteroseptal
46 ST depression, mild anterior
47 ST depression, mild extensive
48 ST depression, mild apical
49 ST depression, mild anterolateral
50 ST depression, mild high lateral
51 ST depression, mild inferior
52 ST depression, mild inferolateral
53 ST depression, anteroseptal
54 ST depression, anterior myocardial
7 75.36 99.55 46.67
5 81.24 99.94 33.33
13 79.83 99.13 53.59
17 76.97 99.14 43.13
25 77.54 99.08 37.64
21 80.64 99.14 47.39
12 79.73 99.60 55.16
20 80.59 99.26 50.61
4 85.41 99.72 44.44
12 87.66 98.58 34.85
59
Page 66
myocardial ischemia
ischemia
myocardial ischemia
myocardial ischemia
ischemia
myocardial ischemia
55 ST depression, extensive anterior
56 ST depression, apical myocardial
57 ST depression, anterolateral
58 ST depression, high lateral
59 ST depression, inferior myocardial
60 ST depression, inferolateral
Sensitivity: probability that a "True sample" would be determined as certain "Item" by
automated interpretation function;
Specificity: probability that a "True unfit sample" would be determined as certain "Unfit
item" by automated interpretation function;
Positive predictive value: probability that a determined "Unfit item" is a "True unfit item".
7 84.78 98.04 67.75
18 79.95 99.14 55.12
13 87.42 98.97 59.09
16 90.06 99.31 57.14
12 89.88 99.13 40.08
6 91.39 99.16 50.47
60
Page 67
Guidance and manufacturer’s declaration –electromagnetic emission
is intended for use in the electromagnetic environment specified below. The
purchaser or the user of the device should assure that it is used in such environment.
Emission test
Compliance
RF emissions CISPR 11
Group 1
IEC 61000-3-2
Guidance and manufacturer’s declaration-electromagnetic immunity
is tended for use in the electromagnetic environment specified below. The
purchaser or the user of the device should assure that it is used in such an environment.
test level
IEC 61000-4-2
± 15 kV air
±15kV air
±2kV for power supply
± 1 kV for input/output line
short interruptions
and voltage vatiations on power
<5%UT(>95%dip in UT) for
%UT(30%dip in UT) for
<5%UT(>95%dip in UT) for
5 sec
<5%UT(>95%dip in UT) for
40% UT(60%dip in UT) for 5
70%UT(30%dip in UT) for 25
<5%UT(>95%dip in UT) for 5
sec
IEC 61000-4-8
Guidance and manufacturer’s declaration – electromagnetic immunity

Appendix II EMC Guidance and Manufacturer6 Declaration

Table 1:
The device
RF emissions CISPR 11 Class A
Harmonic emissions
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2:
The device
Immunity test
Electrostatic discharge (ESD)
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
supply input lines
IEC 61000-4-11
Power frequency (50 / 60Hz)
magnetic field
Table 3:
Not applicable
IEC60601
±8kV contact
lines
±1 kV lines to lines
±2 kV lines to earth
0.5 cycle
40% UT(60%dip in UT) for 5
cycle
70
25 cycle
30 A/m 30A/m
61
Compliance level
±8kV contact
±2kV for power supply lines
Not Applicable
±1 kV lines to lines
±2 kV lines to earth
0.5 cycle
cycle
cycle
Page 68
is tended for use in the electromagnetic environment specified below. The
purchaser or the user of the device should assure that it is used in such an environment.
IEC61000-4-3
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
ones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
ss the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
red. If the measured field strength in the location in which the device or system is
device should be observed to
ay be
necessary, such as adjusting the direction or location of the device.
Guidance and manufacturer’s declaration - electromagnetic Immunity
The [Code SI] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [Code SI] should assure that it is used in such an environment
tions for
ications
(MHz)
18 Hz
± 5 kHz
1 kHz sine
LTE Band
217 Hz
The device
Immunity test IEC 60601 test level Compliance level
3 V Conducted RF IEC61000-4-6
Radiated RF
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2
teleph broadcast cannot be predicted theoretically with accuracy. To asse
conside used exceeds the applicable RF compliance level above, the verify normal operation. If abnormal performance is observed, additional measures m
Tab le 4
Radiated
RF
IEC6100
0-4-3
(Test
specifica
ENCLOS
URE
PORT
IMMUNI
TY to
RF
wireless
commun
Test
Frequenc
y
385
450
710
745
780
0,15 MHz – 80 MHz
6 V in ISM bands between
0,15 MHz and 80 MHz
3 V/m 80 MHz- 2.7 GHz 3 V/m80 MHz- 2.7 GHz
Band a)
(MHz)
380
–390
380
–390
704 –
787
Service
a)
TETRA
400
GMRS
460,
FRS 460
13,
17
Modulation
b)
Pulse
modulation
b)
FM c)
deviation
Pulse
modulation
b)
3 V 0,15 MHz – 80 MHz 6 V in ISM bands between 0,15 MHz and 80 MHz
Modulation
b) (W)
1,8 0,3 27
2 0,3 28
0,2 0,3 9
62
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Page 69
LTE Band
5
LTE Band
4, 25;
UMTS
LTE Band
7
217 Hz
5785
he
may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
equipme
nt)
810
870
930
1720
1845
1970
2450
800 –
960
1 700 –
1 990
2 400 –
2 570
GSM
800/900,
TETRA
800,
iDEN
820,
CDMA
850,
GSM
1800;
CDMA
1900;
GSM
1900;
DECT;
1, 3,
Bluetoot
h,
WLAN,
802.11
b/g/n,
RFID
2450,
Pulse
modulation
b)
18 Hz
Pulse
modulation
b)
217 Hz
Pulse
modulation
b)
217 Hz
2 0,3 28
2 0,3 28
2 0,3 28
5240
5500
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between t
transmitting antenna and the
ME EQUIPMENT or ME SYSTEM
5 100 –
5 800
WLAN
802.11
a/n
Pulse
modulation
b)
63
0,2 0,3 9
Page 70
, 50 % pulse modulation at 18 Hz may be used because
LEVELS that are appropriate for the
reduced minimum separation distance. Minimum separation distances for higher IMMUNITY
P
d
6
E =
IMMUNITY TEST LEVEL in V/m.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation
while it does not represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on
RISK MANAGEMENT, and using higher IMMUNITY TEST
TEST LEVELS shall be calculated using the following equation:
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the
Warning
Don’t near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the intensity of EM DISTURBANCES is
high.
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.”
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of
the device including cables specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
Active medical devices are subject to special EMC precautions and they must be
installed and used in accordance with these guidelines.
Note
The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
When the device is disturbed, the data measured may fluctuate, please measure
repeatedly or in another environment to ensure its accuracy.
64
Page 71
combined with the protective grounding.
Triangular band:Black

4.2.5 Symbols

AC Alternating current
Equipotential point, the equipotential point of this device is
Caution: read instructions (warnings) carefully
PATIENT
Defibrillation-proof type CF applied part
USB interface
Lead cable socket
Serial number
Manufacturer
Date of manufacture
Lot number
Latex free
Atmospheric pressure limit
Temperature limit
Humidity limit
This way up
Fragile, handle with care
Keep in a cool, dry place
Stacking limit by number
General warning label
NOTE:
Background col our:yellow
65
Page 72
Medical Device compliant with Regulation (EU) 2017/745
Follow instructions for use
Product code
Authorized representative in the European Community
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Disposal: The product must not be disposed of along with other
domestic waste. The users must dispose of this equipment by bringing it
to a specific recycling point for electric and electronic equipment
66
Loading...