Gilson PIPETMAX 268 User Manual

Page 1
PIPETMAX® 268
User’s Guide
Gilson, Inc. | PO Box 620027 | Middleton, WI 53562-0027, USA
Tel: 608-836-1551 or 800-445-7661 | Fax: 608-831-4451
Gilson S.A.S. | 19, avenue des Entrepreneurs | BP 145 | F-95400 Villiers-le-bel, FRANCE
www.gilson.com | sales@gilson.com | service@gilson.com | training@gilson.com
LT255519-07
©2018 Gilson, Inc. | All rights reserved.
Page 2
Page 3
Safety
Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-2
Safety Interlocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Pipette Heads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Tip Waste Container. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-4
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-4
Sécurité
Tension. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-2
Dispositifs de sécurité . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-3
Têtes de pipetage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-3
Conteneur pour pointes usagées . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-3
Plateau . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-4
Liquides . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-4
Pièces Détachées . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-4
1 Introduction
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Repair and Return Policies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Before Calling Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Warranty Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Non-Warranty Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Unit End-of-Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Gilson PIPETMAX® 268 User’s Guide Table of Contents-1
Page 4
2 Setup/Operation
Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
On Lab Bench or Lab Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
In Hood. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Removable Tray. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Rear Panel Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Drain Connection (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Front Panel Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Power On and Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Log In . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
User Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Change Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Edit User Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Add Users (Administrators Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Delete Users (Administrators Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Touchscreen Tablet Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Icons and Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
On-Screen Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Run Alignment Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Prepare to Run Alignment Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Run Alignment Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Run Application Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
Import Application Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Add Pipette Head(s). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Set Date and Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
Run Application Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
Pause or Stop a Running Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Pause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-23
STOP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-23
View Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Tray View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-24
Labware View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Volume View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Steps View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Run Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-27
Export and View Run Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-27
Export Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
Log Out. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-28
Table of Contents-2 Gilson PIPETMAX® 268 User’s Guide
Page 5
3 Messages
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Hood Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Insufficient Volume/Excessive Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Refill Tip Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
4 Maintenance
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Add Pipette Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Replace Pipette Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Convert Blank Head into Alignment Head. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
A Replacement Parts
PIPETMAX® 268 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
TRILUTION® micro Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Pipette Heads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Removable Trays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Tip Reload Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Tip Ejector Feet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Tip Waste Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Passive Temperature (PT) Racks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Circulating Temperature (CT) Racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Miscellaneous. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Gilson PIPETMAX® 268 User’s Guide Table of Contents-3
Page 6
Page 7

Safety

Read this section before installing and operating the PIPETMAX® 268.
This instrument is intended to be used in a laboratory environment by trained technical personnel.
For safe and proper use of this instrument, it is required that both operating and service personnel follow the instructions contained in this guide when installing, cleaning, and maintaining the instrument.
The following safety precautions must be observed during all phases of operation, service, and repair of the instrument. Failure to comply with these precautions or with specific warnings elsewhere in this user’s guide violates safety standards of design, manufacture, and intended use of the instrument. Gilson assumes no liability for the customer’s failure to comply with these requirements.
The PIPETMAX 268 has been certified to safety standards required in Canada, Europe, and the United States. Refer to the instrument rear panel label and the Declaration of Conformity document for the current standards to which the instrument has been found compliant.
Gilson PIPETMAX® 268 User’s Guide Safety-1
Page 8
Safety
The following electronic and hazard symbols may appear on the instrument:
Symbol Explanation
Safety
| Electrical power ON Sous tension Netzschalter ein
O Electrical power OFF Hors tension Netzschalter aus
The following safety notices may appear in this document:
Direct current Courant continu Gleichstrom
Caution Attention Vorsicht
Mechanical hazard Danger méchanique Mechanische Gefährdung
WARNING indicates a potentially hazardous situation which, if not avoided, may result in serious injury
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury
NOTICE indicates a potentially hazardous situation which, if not avoided, may result in equipment damage

Voltage

Access to the rear panel is necessary because the instrument must be detached from all voltage sources before service, repair, or exchange of parts.
Operate the instrument using the approved power supply provided and only at the voltage specified on the rear panel label of the instrument.
Safety-2 Gilson PIPETMAX® 268 User’s Guide
Page 9
Safety

Safety Interlocks

The instrument is configured with one of two safety interlocks. One interlock is used when the rotating cover is installed on the instrument, and the other interlock is an external sensor and is used when the instrument is located in a hood.
The safety interlocks prevent the instrument from operating when the rotating cover or hood door is open.
When the instrument is installed in the hood, the external sensor must face the door of the hood and be located within 15 cm (6 in) of the door.
Safety
PIPETMAX 268 Configured with External Safety Interlock Sensor and Placed in Hood

Pipette Heads

The instrument is capable of generating significant forces that could cause potential injury to the user. For an instrument with the rotating cover installed, the safety interlock in the cover will disable the instrument movement if the cover on the instrument is open. For an instrument installed in a hood, the external safety interlock sensor mounted on the instrument will disable the instrument movement if the hood door is open.

Tip Waste Container

The tip waste container may contain residues of flammable or infectious liquids in ejected tips.
If you use flammable liquids (e.g., ethanol 98%), treat the waste before disposing of it in accordance with your laboratory guidelines.
Dispose of infectious material, waste, or tips in accordance with national and local safety regulations.

Tray

The moving tray is capable of causing injury by pinching. For an instrument with the rotating cover installed, the safety interlock in the cover will disable the instrument movement if the cover on the instrument is open. For an instrument installed in a hood, the external safety interlock sensor mounted on the instrument will disable the instrument movement if the hood door is open.
Gilson PIPETMAX® 268 User’s Guide Safety-3
Page 10
Safety

Liquids

Observe safe laboratory practices when handling liquids. If working with biological samples or chemical substances, ensure that
Safety
th er e is pr oper ven ti lati on, a nd we ar pe rson al pro te ctiv e eq ui pmen t (PP E) , su ch as sa fety gl ass es , gl oves , and prot ec tiv e c loth ing at all times.
Refer to the Material Safety Data Sheets for solvents before use.

Replacement Parts

Be sure to use only replacement parts specified in this user’s guide. Do not repair or change parts which are not listed in this user’s guide. If it is necessary to change parts not listed, please contact your local Gilson representative.
Safety-4 Gilson PIPETMAX® 268 User’s Guide
Page 11

Sécurité

Merci de lire ces instructions avant toute installation ou utilisation du PIPETMAX® 268.
Cet instrument est destiné à être utilisé dans un environnement de laboratoire, par un personnel technique qualifié.
Pour une utilisation correcte et en toute sécurité, il est nécessaire que le personnel qui utilise et réalise la maintenance de l’instrument, suive les instructions contenues dans ce guide lors de l'installation, le nettoyage et la maintenance de l'instrument.
Les consignes de sécurité suivantes doivent être respectées durant toutes les phases de fonctionnement, d’entretien ou de réparation de l'instrument. Le non-respect de ces précautions ou des avertissements spécifiques mentionnés dans ce guide compromet les normes de sécurité de conception, de fabrication et d’utilisation prévue de l'instrument. Gilson décline toute responsabilité en cas d’'incapacité du client à se conformer à ces exigences.
Le PIPETMAX 268 a été certifié conformément aux normes de sécurité en vigueur au Canada, en Europe et aux Etats-Unis. Merci de vous reporter aux indications mentionnées sur le panneau arrière de l'instrument ainsi qu’au document de Déclaration de Conformité aux normes pour lesquelles l'instrument a été déclaré conforme.
Gilson PIPETMAX® 268 User’s Guide Sécurité-1
Page 12
Sécurité
Les symboles suivants sont susceptibles d’apparaître sur l'instrument:
Symbole Signification
Sécurité
| Electrical power ON Sous tension Netzschalter ein
O Electrical power OFF Hors tension Netzschalter aus
Les consignes de sécurité suivantes peuvent apparaître dans ce document:
Direct current Courant continu Gleichstrom
Caution Attention Vorsicht
Mechanical hazard Danger mécanique Mechanische Gefährdung
AVERTISSEMENT signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut entraîner des blessures graves
ATTENTION signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut entraîner des blessures mineures ou légères
AVIS signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut entraîner des dommages matériels

Tens ion

L'accès au panneau arrière doit être libre car l'instrument doit pouvoir être déconnecté de sa source d’alimentation avant toute
opération d’entretien, de réparation ou de remplacement de pièces.
Ne faîtes fonctionner l'appareil qu’en utilisant le bloc d'alimentation fourni et uniquement à la tension indiquée sur l'étiquette située à l'arrière de l'instrument.
Sécurité-2 Gilson PIPETMAX® 268 User’s Guide
Page 13
Sécurité

Dispositifs de sécurité

L'instrument présente deux dispositifs de sécurité. Le premier est activé lorsque le capot rotatif est installé sur l'appareil ; le second est un capteur externe, activé lorsque l'instrument est placé sous une hotte.
Les dispositifs de sécurité empêchent tout fonctionnement de l'instrument lorsque le capot ou la porte de la hotte sont ouverts.
Lorsque l'appareil est installé sous une hotte, le capteur externe doit faire face à la porte de la hotte et être positionné à 15 cm de cette porte.
Sécurité
PIPETMAX 268 Configuration sous hotte où le dispositif de sécurité est un capteur externe

Têtes de pipetage

L'instrument est capable de générer des forces importantes, susceptibles d'entraîner des blessures à l'utilisateur. Lorsque l’appareil est équipé de son capot rotatif, le dispositif de sécurité désactive tout mouvement de l'appareil lors de l’ouverture du capot. Lorsque l’instrument est installé sous une hotte, le capteur externe permet de désactiver tout mouvement de l'instrument lors de l’ouverture de la porte de la hotte.

Conteneur pour pointes usagées

Les pointes usagées peuvent contenir des résidus liquides dangereux, inflammables ou infectieux.
Si vous utilisez des liquides inflammables (ex. éthanol 98%), veillez au traitement approprié des déchets avant leur élimination, en vous conformant aux directives de votre laboratoire.
Eliminez le matériel infectieux, les déchets ou les pointes usagées en vous conformant aux réglementations de sécurité nationale et locale.
Gilson PIPETMAX® 268 User’s Guide Sécurité-3
Page 14
Sécurité

Plateau

Le plateau mobile est susceptible d'occasionner des blessures par pincement. Lorsque l’appareil est équipé de son capot rotatif, le dispositif de sécurité désactive tout mouvement de l'appareil lors de l’ouverture du capot. Lorsque l’instrument est installé
Sécurité
sous une hotte, le capteur externe permet de désactiver tout mouvement de l'appareil lors de l’ouverture de la porte de la hotte.

Liquides

Respectez les pratiques de sécurité du laboratoire lors de la manipulation de liquides. Si vous travaillez avec des échantillons biologiques ou des substances chimiques, assurez-vous que la ventilation est adéquate et portez en permanence un équipement de protection individuelle (EPI), tel que : lunettes, gants et vêtements de protection.
Reportez-vous aux Fiches de Données de Sécurité pour les solvants avant toute utilisation.

Pièces Détachées

Assurez-vous de n’utiliser exclusivement que les pièces détachées préconisées dans ce guide. Ne tentez pas de réparer ou remplacer des pièces ne figurant pas dans ce guide.
Si le remplacement de pièces ne figurant pas dans ce guide s’avérait nécessaire, merci de contacter votre représentant Gilson local.
Sécurité-4 Gilson PIPETMAX® 268 User’s Guide
Page 15

Introduction 1

The PIPETMAX® 268 is a small footprint personal pipetting station used to perform repetitive, precise, pipetting actions.
This chapter provides information on the following topics:
Unpacking
Technical Specifications
Customer Service
Repair and Return Policies
PIPETMAX® 268
with optional riser for off-bed tip disposal
Gilson PIPETMAX® 268 User’s Guide 1-1
Page 16
Introduction
1

Unpacking

PI PET MAX is delivered with m ost majo r components already assembled. Keep the original packaging in case PIPETMAX must be returned to the factory.
Unpacking
1 Remove the open-ended cardboard box.
2 Remove the box of accessories and open it.
3 Lift the printed, outer cardboard box up to remove.
4 Grip PIPETMAX at the recesses near the base. There is one recess in the front, one in the
back, and one on each side. Use these recesses when lifting PIPETMAX out of the foam-lined, cardboard tray.
5 Place PIPETMAX on a lab bench or cart or in hood.
6 Remove the bag covering PIPETMAX. You may need to lift the front and the back
slightly.
7 Remove the film protecting the rotating cover (if installed).
8 Remove shipping screws and bracket using the tool provided.
9 Store shipping screws and bracket in rear panel for future use if PIPETMAX needs to be
shipped or transported.
It is recommended that two people lift PIPETMAX out of the box, as it weighs approximately 24.9 kg (55 lbs.).
Do not attempt to lift PIPETMAX from the cover or from the X-arm (the horizontal arm). Always lift the instrument from its base.
Do not store screws in rear panel without bracket as that will damage PIPETMAX during operation.
10 Remove the shipping brace from the rear panel by loosening the thumb screws
securing the shipping brace to PIPETMAX. After removing the shipping brace, tighten the thumb screws.
11 Remove the foam block preventing the tray from moving and securing the
alignment cable cord.

Standard Equipment

After PIPETMAX and its accessories are unpacked, you should have the following:
PIPETMAX® 268 with alignment head, dual eight-channel tip ejector foot, and rotating cover or external safety interlock sensor installed
•USB Cable
•Tip Reload Block
•Tip Disposal Bin
•USB Drive
Plug for Outlet on the Rear Panel
Power Supply
Power Cords
1-2 Gilson PIPETMAX® 268 User’s Guide
Page 17
Introduction
1
Documentation
The following documents are included with the PIPETMAX® 268:
Installation Qualification/Operational Qualification (IQ/OQ) Procedures
Declaration of Conformity
Quality Control (QC) Checklist
Items Included Checklist
•Setup Guide
Gilson Quality Control Report Pipette Head (one for each pipette head)
Protocols
Alignment Protocol
Standard Liquid Handling Protocols with Documentation

Accessories

Required
Unpacking
Some accessories are required, but are ordered separately.
Tablet with TRILUTION® micro pre-loaded or TRILUTION® micro on a USB drive for installation on a PC
•Pipette Head(s)
•Removable Tray
Blister Packs of Tips
•Racks
For part numbers, refer to Appendix A,
Replacement Parts.
Optional
The following optional accessories are also available:
Tablet Stand
Riser Kit for Off-bed Tip Disposal
•Orbital Shaker
Microscan 2D Barcode Reader
For part numbers, refer to Appendix A,
Replacement Parts.
Gilson PIPETMAX® 268 User’s Guide 1-3
Page 18
Introduction
1

Technical Specifications

Please be aware of the following before operating the PIPETMAX® 268.
Changes or modifications to this device not expressly approved by Gilson could void the warranty.
The instrument complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This instrument may not cause harmful interference, and (2) this instrument must accept any interference received, including interference that may cause undesired operation.
Shielded cables must be used with the instrument to ensure compliance with the FCC Class A limits.
Technical Specifications
Technical Specification
Communication USB
Connections Three USB host ports and one USB device port
Control Touchscreen tablet or PC control via USB and TRILUTION® micro software
Dimensions (W x D x H)
Environmental Conditions
Front Panel Two USB host ports and STOP button
Liquid Contact Materials
Pipette Heads and DIAMOND® Tips
Definition
Two inputs (contact closure, TTL), two relay outputs, and one switched +12V DC 1A output
Switching voltages higher than 30V or greater than 1A of current may damage the instrument.
PIPETMAX® 268 with rotating cover
54.4 x 65.5 x 53.1 cm (21.4 x 25.8 x 20.9 in)
PIPETMAX® 268 with rotating cover installed on optional riser assembly for off-bed tip disposal
54.4 x 65.5 x 69.6 cm (21.4 x 25.8 x 27.4 in)
PIPETMAX® 268 without cover
50.8 x 64.3 x 49.5 cm (20 x 25.3 x 19.5 in)
PIPETMAX® 268 without cover installed on optional riser assembly for off-bed tip disposal
52.3 X 65 X 65.8 cm (20.6 x 25.6 x 25.9 in)
Indoor use Altitude: up to 2000 m Temperature range: 5°C–40°C Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
Description Material
Tips 100% Virgin Polypropylene
Tip Disposal Bin Pro-fax 6523 (Polypropylene with colorant)
Tip Disposal Bin (Off-bed) Polypropylene
Tip Chute
Pipette Head
MAX8x20 0.0225–3.75
MAX8x200 0.225–37.5 D200 20–200 DF200 20–200
MAX1000 1.125–187.5 D1000 100–1000 DF1000 100–1000
Sterilized option available for some tips.
Range (mL/min)
PIPETMAX® 268
Flow Rate
Pro-fax 6523 (Polypropylene with colorant) Aluminum 5052 with Hentzen, URA-ZEN, White, Matte Finish, Fine Texture Paint
DIAMOND®
Tips
D200 1–20 DF30 1–20
DL10 1–20 DFL10 1–10
Volume
Range (μL)
PIPETMAX® 268 Technical Specifications
DIAMOND®
Filter Tips
Volume
Range (μL)
(Page 1 of 2)
1-4 Gilson PIPETMAX® 268 User’s Guide
Page 19
Introduction
1
Technical Specification
Positioning Performance
Power Requirements
Definition
Machine Accuracy (without tips) in XYZ = +/-0.2 mm (0.008") System Accuracy (with current tips) in XYZ = +/-0.9 mm (0.035")
External Power Supply
Voltage Input
Frequency: 50 to 60 Hz Voltage: 100–240V AC
Voltage Output
Voltage: 24V DC Current Rating: 6.25A, 150W
Removable Tray Capacity
Safety and Compliance
9-position removable tray (microplate footprints, but not for 384-well microplates) 9-position removable tray for 384-well microplates
The PIPETMAX® 268 has been certified to safety standards specified for Canada, Europe, and the United States. Refer to the instrument rear panel label and the Declaration of Conformity document for the current standards to which the instrument has been found compliant.
Volumetric Specification for Transfers
Pipette Head
MAX8x20
MAX8x200
Weight 24.9 kg (55 lbs.)
XYZ Motion Parameters
X Min Max Default
Power 1% 100% 50%
Speed 1 mm/sec 550 mm/sec 350 mm/sec
Acceleration Speed Value
Y Min Max Default
Power 1% 100% 50%
Speed 1 mm/sec 550 mm/sec 350 mm/sec
Acceleration Speed Value
Z Min Max Default
Power 1% 100% 60%
Speed 1 mm/sec 140 mm/sec 120 mm/sec
Acceleration Speed Value
PIPETMAX® 268 (Continued)
Maximum Permissible Errors
Volume of
Distilled Water
1 μL ±0.08 ≤0.05 8 5
10 μL ±0.15 ≤0.10 1.5 1
20 μL ±0.25 ≤0.12 1.25 0.6
20 μL ±0.50 ≤0.16 2.5 0.8
100 μL ±1.00 ≤0.30 1 0.3
200 μL ±2.00 ≤0.50 1 0.25
Systematic
Error (μL)
4000 mm/sec
4000 mm/sec
4000 mm/sec
Random
Error (μL)
2
2
2
PIPETMAX® 268 Technical Specifications
Systematic
Error (%)
2200 mm/sec
900 mm/sec
650 mm/sec
Technical Specifications
Random
Error (%)
2
2
2
(Page 2 of 2)
Gilson PIPETMAX® 268 User’s Guide 1-5
Page 20
Introduction

Customer Service

Gilson, Inc. and its worldwide network of authorized representatives provide customers with the following types of assistance: sales, technical support, applications, and instrument repair.
If you need assistance, please contact your local Gilson representative. Specific contact information can be found at
www.gilson.com
Customer Service
. To help us serve you quickly and efficiently, please refer to Before Calling Us on page 1-7.
1
1-6 Gilson PIPETMAX® 268 User’s Guide
Page 21
Introduction

Repair and Return Policies

Before Calling Us

Your local Gilson representative will be able to serve you more efficiently if you have the following information:
Serial number and model number of the instruments involved.
The serial number for the PIPETMAX® 268 is located on the inside panel of the right support.
The serial number for each pipette head is located on the document supplied with it.
Installation procedure you used.
List of concise symptoms.
List of operating procedures and conditions you were using when the problem arose.
List of all instruments in the configuration and the connections to those instruments.
List of other electrical connections in the room.

Warranty Repair

1
Repair and Return Policies
Units covered under warranty will be repaired and returned to you at no charge. If you have any questions about applicability, contact your local Gilson representative.

Non-Warranty Repair

For out-of-warranty repairs, contact your local Gilson representative who will discuss service options with you and can assist in making arrangements to return the equipment, if necessary.

Return Procedure

Contact your local Gilson representative to obtain authorization before returning any Gilson equipment. To return a piece of equipment:
Carefully pack the unit to prevent damage in transit. Check with your local Gilson representative regarding proper method of shipment. No responsibility is assumed by Gilson or your local Gilson representative for damage caused by improperly packaged instruments. Indicate the authorization on the carton and on the packing slip.
Always insure for the replacement value of the unit.
Include a description of symptoms, your name, address, phone number, and purchase order to cover repair costs, return and shipping charges, if your institution requires it.

Unit End-of-Life

When a unit reaches the end of its useful life, refer to www.gilson.com for directions and information on the end-of-life policy. This is in accordance with the European Union Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE).
Gilson PIPETMAX® 268 User’s Guide 1-7
Page 22
Page 23

Setup/Operation 2

The PIPETMAX® 268 and its components should be set up in the order described in this chapter. Gilson TRILUTION® micro provides software control of the PIPETMAX® 268 during setup and operation as described in this chapter. For more information about TRILUTION® micro, view the help information in the software, accessible from the button.
This chapter includes the following information:
1 Placement
2 Removable Tray
3 Connections
4 Power On and Start Up
5 Log In
6 User Management
7 Touchscreen Tablet Overview
8 Run Alignment Protocol
9 Run Application Protocol
10 Pause or Stop a Running Protocol
11 View Results
12 Run Report
13 Export Protocol
14 Log Out
Gilson PIPETMAX® 268 User’s Guide 2-1
Page 24
Setup/Operation
PIPETMAX Configured with External Safety Interlock Sensor and Placed in Hood

Placement

Place PIPETMAX on a stable, level surface.
PIPETMAX is configured with one of two safety interlocks. One interlock is internal to the rotating cover installed on the
Placement
instrument, and the other interlock is an external sensor used when the instrument is located in a hood.
The safety interlocks prevent the instrument from operating when the rotatingcover or hood door is open.

On Lab Bench or Lab Cart

A PIPETMAX with cover can run on a lab bench or lab cart.

In Hood

If PIPETMAX was ordered without a rotating cover, it must run in a hood.
When the instrument is installed in the hood, the external sensor must face the door of the hood and be located within 15 cm (6 in) of the door.
2

Removable Tray

A removable tray for placement of labware and tip racks is required.

Options

There are two removable tray options: 9-position removable tray (microplate footprints, but not for 384-well microplates) and 9-position removable tray for 384-well microplates.
9-Position Removable Tray
When placing labware and tip racks on the 9-position removable tray, refer to the following information:
Tip racks can be placed in the back row middle and right positions, and the front row middle position ONLY. These positions have installed clips to hold the tip racks in place. It is recommended to use full tip packs because an error may be encountered if using two partial tip packs containing the same type of tip.
The tip waste chute can be placed in the back row, left position ONLY. It is recommended to place the tip disposal bin in this position. It cannot be placed in positions with installed clips.
All racks and plates can be placed in any position on the tray.
Exceptions:
384-well microplates cannot be placed on this tray.
A maximum of three thermal racks total (1, 2, or 3 PT racks, 1 or 2 CT racks, or a combination) can be placed on this tray.
Refer to the instructions provided with the racks to modify the tray to allow for rack placement.
Circulating Temperature (CT) racks can be placed in the front row, left or right positions ONLY.
Refer to the instructions provided with the rack to modify the tray to allow for rack placement.
The orbital shaker can only be placed in the front row, left or right positions ONLY.
2-2 Gilson PIPETMAX® 268 User’s Guide
Page 25
Setup/Operation
2
9-Position Removable Tray for 384-Well Microplates
When placing labware and tip racks on the 9-position removable tray for 384-well microplates, refer to the following information:
Tip racks can be placed in the back row middle and right positions, and the front row middle position ONLY. These positions have installed clips to hold the tip racks in place. It is recommended to use full tip packs because an error may be encountered if using two partial tip packs containing the same type of tip.
The tip waste chute or tip disposal bin must be placed in the back row, left position ONLY.
All racks and plates can be placed in any position on the tray.
Exceptions:
A maximum of three thermal racks total (1, 2, or 3 PT racks, 1 or 2 CT racks, or a combination) can be placed on this tray.
Refer to the instructions provided with the racks to modify the tray to allow for rack placement.
Circulating Temperature (CT) racks can be placed in the front row, left or right positions ONLY.
Refer to the instructions provided with the rack to modify the tray to allow for rack placement.
The orbital shaker cannot be placed on this tray.

Installation

While holding the handles on the sides of the tray, lower the removable tray onto the fixed tray. It is keyed and will only mount one way. Grip the removable tray and then gently rock it back and forth. If it seems that the removable tray is shifting position on the fixed tray, use a 3 mm Allen wrench and gentle pressure to adjust the spacers on either side of the fixed tray (see figures). Spacers should just barely contact the removable tray. Remove and then replace the removable tray to ensure there is not any significant interference.
Removable Tray
Gilson PIPETMAX® 268 User’s Guide 2-3
Page 26
Setup/Operation
1 2
4
5
6 7
8
9
3

Connections

Refer to the diagrams when making the connections described in this section.

Rear Panel Diagrams

Connections
PIPETMAX® 268 Rear Panel
2
1 Input/Output
2 USB host (Not Supported)
3USB host
4 Ethernet (Not Supported)
5 USB device
6 RS-232 (Not Supported)
7HDMI (Not Supported)
8Power receptacle
9Power switch (MAINS)
PIPETMAX® 268 Rear Panel Connections
2-4 Gilson PIPETMAX® 268 User’s Guide
Page 27
Setup/Operation
2
Input/Output
The input/output contacts on the rear panel are used for communication between the alignment head and TRILUTION® micro and for communication between the external safety interlock sensor and TRILUTION® micro. The contacts can also be used to control peripheral devices. Refer to the diagram on page 2-4
for the location of the input/output ports.
Contact Inputs
The bottom terminal block on PIPETMAX has two paired input contacts that are labeled 4 and 5.
Never connect voltages higher than 5V DC to an input. When using TTL signals, be sure to match GROUND connections.
Contact Outputs
The bottom terminal block on PIPETMAX has two paired, isolated-relay contact closures that are labeled 1 and 2.
DC Power Output
The top terminal block has one +12V DC output, which is labeled.
Making Connections
The following are needed to make connections:
Connections
2-conductor cable (22–30 gauge for each wire)
wire insulation stripper
A 6-foot piece of suitable cable (part number 709910206) is available for purchase from Gilson.
To make connections with the 2-conductor cable:
1 Cut the cable into pieces of appropriate length.
2 Strip about 8 mm of insulation from each end of the cable.
3 Remove the terminal block connector from PIPETMAX.
4 Press in the spring-loaded retainer for the appropriate terminal on the terminal block connector. Insert each wire into the
appropriate terminal on the terminal block connector and then release the spring-loaded retainer.
5 Reconnect the terminal block connector to PIPETMAX. Push the connector in as far as it will go. It is designed to fit snugly
into its receptacle.
6 Connect the opposite ends of the wires to the other device(s). Be sure to match ground connections.
7 Label each cable to identify the purpose of the connection.
USB Host
PIPETMAX has one functional USB host port (the middle port) on the rear panel. The top port on the rear panel is not supported.
The USB ports can be used to connect USB drives (which can also be seen by the tablet when plugged into PIPETMAX), keyboard, mouse, or any other compatible device (heater/chiller, barcode scanner, or shaker, for example).
Ethernet Port
Not supported.
RS-232
Not supported.
Gilson PIPETMAX® 268 User’s Guide 2-5
Page 28
Setup/Operation
drain
HDMI
Not supported.
USB Device
The USB device port is used for communication between PIPETMAX and the tablet or PC.
Connections
2
A USB cable is provided as a standard accessory. Refer to the diagram on page 2-4
for the location of the USB device port.
Power Cord Connection
Use the power cord on the external power supply to make the connection between the power receptacle on PIPETMAX and the external power supply.
The connection from the external power supply to PIPETMAX uses a connector with a locking collar. Check the alignment of the pins and then push it in until it clicks and locks in place. To disconnect, pull back on the locking collar and then disconnect the cable from the rear panel of PIPETMAX.
Locate the appropriate power cord for your line voltage and then make the connection between the external power supply and the AC power source.

Drain Connection (Optional)

The PIPETMAX bed includes a drain. Optionally, connect a 1/8" NPT fitting and drain tubing. Ensure that the waste container is lower than the bed of PIPETMAX.
2-6 Gilson PIPETMAX® 268 User’s Guide
Page 29
Setup/Operation
1
2
2

Front Panel Diagram

Front Panel Diagram
1USB host
2STOP
USB Host
PIPETMAX has two functional USB host ports on the front panel.
Connections
The USB ports can be used to connect USB drives (which can also be seen by the tablet when plugged into PIPETMAX), keyboard, mouse, or any other compatible device (heater/chiller, barcode scanner, or shaker, for example).
STOP
In a situation where an emergency stop is required, pressing the STOP button stops PIPETMAX immediately. The protocol stops and results are displayed (after touching OK on the notification).
Gilson PIPETMAX® 268 User’s Guide 2-7
Page 30
Setup/Operation

Power On and Start Up

To start P IPETMAX:
1 Make sure that PIPETMAX is connected to the external power supply and that the external power supply is connected to a
power source.
2 Turn power on at the external power supply.
3 Turn PIPETMAX power on using the MAINS power switch located on the rear panel. The indicator light on the front panel
illuminates.
4 Connect the tablet or PC to a power source and power it on.
Power On and Start Up
5 Start TRILUTION® micro (PC users only). TRILUTION® micro starts automatically on the tablet.
Click the Start button and then select All Programs > GilsonApplications > TRILUTION micro >
TRILUTION micro x.x.
6 Close the rotating cover or hood door. PIPETMAX will not operate with the cover or hood door open.

Log In

When prompted, enter the “User name” and “Password“, and then select Accept (or X to cancel). The default “Username” is ‘admin’ and the “Password” is ‘Gilson268’.
2
Note: Password is case-sensitive.
Note: On the tablet, the “Log in” screen also provides options for shutdown. For more
information, refer to Shutdown
on page 2-11.
2-8 Gilson PIPETMAX® 268 User’s Guide
Page 31
Setup/Operation
2

User Management

The user management feature in TRILUTION® micro provides a way for administrators to grant or restrict users access to programs and features in the software. User management is accessible from the main screen shown when the software starts.

Change Password

It is recommended to change the password for the default administrator (admin).
1 Select the “User name (Full name)” in
TRILUTION®micro to go to the “User properties” screen.
2 Enter a new “Password” for the user, and then enter
the password again in the “Password confirm” field.
3Select Save to save, or Back to return to the
previous screen without saving.

Edit User Properties

User Management
1 Select the “User name (Full name)” in TRILUTION® micro to go to the “User properties” screen.
2 Edit the “Full name” and/or “Password”. If editing the “Password”, re-enter the password in the “Password confirm” field.
For the default administrator (admin), “Full name” cannot be modified.
3Select Save to save, or Back to return to the previous screen without saving.
Administrator
1Select Manage Users in TRILUTION® micro to go to
the “User management” screen. The “Usermanagement” screen displays a list of users.
2 Select a user from the list.
3Select Edit. The “User properties” screen appears.
4 Edit any or all properties (“User name”, “Full name”,
“Password”, “Is administrator ”, “Is active”, and “Feature permissions”).
“User name” may not be modified for the
logged in user.
For the default administrator (admin), only “Password” and “Feature permissions” may be modified.
5Select Save to save, or Back to return to the previous screen without saving.
Gilson PIPETMAX® 268 User’s Guide 2-9
Page 32
Setup/Operation

Add Users (Administrators Only)

1 Select Manage users in TRILUTION® micro to go to
the “User management” screen. The “User management” screen displays a list of users.
2 Select Add to add a new user. The “New user”
screen appears.
3Enter aUser name.
User Management
It cannot be the same as any other “User name”.
It can be 1 to 50 characters in length, must
begin with an English letter or number, and may contain the following special characters: _ ()-@.and,. Spaces are valid.
4 Enter the “Full name” for the user.
It can be 1 to 100 characters in length, must begin with an English letter or number, and may contain the following
special characters: _ ()-@.and,. Spaces are valid.
5 Optionally, enter a “Password” for the user. The user can change the “Password” at any time.
I t can be 1 t o 5 0 c ha rac te rs in le ng th , m us t b egin wi th an English letter or number, and may contain the following special
characters: _ ()-@.and,. Spaces are valid.
"It is case-sensitive, but is not required. (Password can be blank.)
2
6 If a “Password” was entered, re-enter the password in the “Password confirm” field.
7 Select “Is administrator” to assign administrator access to the user.
8 By default, “Is active” is selected. Clear the check box to deny the user access to the software. The user remains in the user
list and can be changed to active by an Administrator at any time.
9 Under “Feature permissions”,
a new user has access to all software.)
10 Select Save to save, or Back to return to the previous screen without saving.
clear the check box for any software to which the user should not have access. (By default,

Delete Users (Administrators Only)

Administrators can delete users; however, the default administrator (admin) cannot be deleted.
1 Select Manage Users
a list of users.
2 Select a user from the list.
3 Select Delete. A message will appear asking if you wish to delete the user.
4 Select Ye s to delete the user. Or, select No to close the message without deleting the user.
in TRILUTION® micro to go to the “User management” screen. The “User management” screen displays
2-10 Gilson PIPETMAX® 268 User’s Guide
Page 33
Setup/Operation

Touchscreen Tablet Overview

The following information provides an explanation of unique software interactions when running TRILUTION® micro on the touchscreen tablet. It is recommended to keep the tablet plugged into a power source whenever possible.

Icons and Fields

The following icons appear on several screens in TRILUTION® micro.
Icon Description
Select to return to the “Home” screen.
Select to view the help information for the current screen.
Indicates invalid or missing data error.
On the touchscreen tablet, touch and hold the icon, and then release to display details about the error. Touch to close.
On a PC, right-click on the icon to display details about the error. Click to close.
2
Touchscreen Tablet Overview
Indicates that there is more information available to view.
On the touchscreen tablet, touch and hold the icon, and then release to display the information. Touch to close.
On a PC, right-click on the icon to display the information. Click to close.

On-Screen Keyboard

Refer to the table before for a description of some icons and buttons used in the on-screen keyboards.
Icon Description Button Description
Clear entry CE Clear entry
Cancel entry Enter Accept entry and close keyboard
Cancel Clear entry and close keyboard
Accept entry
Close Accept entry and close keyboard

Shutdown

To view the options to shut down the tablet and/or TRILUTION® micro, select .
•Touch Shut down to close TRILUTION micro and shut down the tablet.
•Touch Restart TRILUTION micro to restart only the software, and not the tablet.
•Touch Cancel to go back.
Gilson PIPETMAX® 268 User’s Guide 2-11
Page 34
Setup/Operation

Run Alignment Protocol

Run the Alignment Protocol to ensure proper alignment. It is only necessary to run this protocol when setting up the instrument for the first time, or if instructed to do so by your local Gilson representative.

Prepare to Run Alignment Protocol

1 Pass the end of the alignment cable cord with pre-wired connector through the
outlet at the back of PIPETMAX.
2 The connector is labeled BOTTOM because it connects to the bottom set of
input/output ports on the rear panel of PIPETMAX. Make the connection.
Run Alignment Protocol
3 The clips to secure the tip racks on the tray can interfere with the alignment. Ensure
that the clips ( ) are positioned as shown in the diagram below, and that the
thumbscrews have been tightened.
2
4 Close the rotating cover or the fume hood door.
5 The tablet starts TRILUTION® micro automatically on power up. If running protocols from a PC, start TRILUTION micro.
2-12 Gilson PIPETMAX® 268 User’s Guide
Page 35
Setup/Operation
2

Run Alignment Protocol

1Select Run/manage protocols.
2Select Run a protocol.
3 Select Alignment Protocol and then select Next.
4 The Scanning dialog appears while the software checks that PIPETMAX is connected.
If connected, the software goes to the next screen. If not connected, a message will appear suggesting possible solutions.
5Select Skipsetup on the “Labware setup guide” screen.
Run Alignment Protocol
6Select Run protocol.
7 Wait approximately 3 minutes while the Alignment Protocol runs.
8 A notification appears when the AlignmentProtocol is complete. Select Continue.
Gilson PIPETMAX® 268 User’s Guide 2-13
Page 36
Setup/Operation

Run Application Protocol

Import Application Protocol

1 Select to go to the “Home” screen.
2 If your protocol file is on a USB drive, connect it to one of the USB ports on the front
of PIPETMAX or the middle USB host port on the rear panel. (Thetop USB port on the rear panel is not supported.)
3 Select Manage protocols to go to the listing of protocols.
4 Select Import and then browse for your application protocol file (.sqlite).
Run Application Protocol
5 Select the file and then select Open to import the protocol file.

Add Pipette Head(s)

Each pipette head that may be used now, or in the future, must be added per the instructions that follow.
1 On the “Home “screen, select Settings.
2
2 On the “Settings” screen, select Pipette heads.
3 Do any of the following:
•If using a tablet, select Scan and then select Camera. Optionally, select the icon to switch between the front and rear camera. When prompted, scan the barcode on the pipette head.
Enter the values from the Gilson Quality Control Report supplied with each pipette head.
Using the optional barcode scanner (ordered separately), select Scan, and then scan the barcode on the pipette head.
4 Select Save.
5 Repeat steps 3 and 4 for additional pipette heads. When finished adding heads,
select Back.

Set Date and Time

Set the date and time to ensure a correct date/time stamp on protocol runs. This should only have to be done once.
On the “Settings” screen, select Date/time, and then set the date, time, and time zone.
2-14 Gilson PIPETMAX® 268 User’s Guide
Page 37
Setup/Operation
2

Run Application Protocol

The “Labware setup guide” screen displays the options for confirming the instrument setup.
Select Step-by-step wizard to do all of the following (in the order shown):
Review a list of the materials.
Review the layout of the tray, set up pipette heads, and add new pipette heads (if necessary).
Set up all of the tip racks and labware.
Review and/or set initial volumes.
For details, refer to Step-by-step Wizard
Select Browse positions manually to do any of the following:
Review the layout of the tray, review pipette heads, and set up or add new pipette heads (if necessary).
Select and set up specific tip racks and labware.
Review and/or set initial volumes.
For details, refer to Browse Positions Manually
Select Skip setup to skip setup instructions and go to the “Wizard complete” screen. Initial volumes will be set to the default values or the values used the last time the protocol was run.
on page 2-16.
on page 2-22.
Run Application Protocol
Gilson PIPETMAX® 268 User’s Guide 2-15
Page 38
Setup/Operation
Step-by-step Wizard
1 Select to go to the “Home” screen.
2 Select Run a protocol.
Select a Protocol
Run Application Protocol
1 Select your protocol, and then select Next.
2 The Scanning dialog appears while the software checks that PIPETMAX is connected.
If connected, the software goes to the next screen. If not connected, a message will appear suggesting possible solutions.
Variables
If your protocol has variables, enter a value for each, and then select Next.
2
When entering a value for a location, the following applies:
Use a colon (:) to specify a range.
Ranges can be numeric or alphanumeric.
Numeric ranges will be processed sequentially.
Alphanumeric ranges will be processed in a batch, if possible.
Use a comma (,) to separate values or ranges.
Specify the starting location first.
•A dash (-) is not valid.
Labware Setup Guide
Select Step-by-step wizard.
The Step-by step wizard includes:
Gather Materials (Materials List)
Pipette Head and Tray Setup (which includes Remove Alignment Head (If Installed),
Connect External Safety Interlock Sensor Install Pipette Head(s) Select Pipette Head Serial Number
, Finish Tip Ejector Foot Installation, and
, Install Tip Ejector Foot,
)
Review Initial Volumes
Run Protocol
Gather Materials (Materials List)
Gather the materials in the Materials list and a tipreload block for each rack, and then select Next.
2-16 Gilson PIPETMAX® 268 User’s Guide
Page 39
Setup/Operation
black plug
2
Pipette Head and Tray Setup
Wait for the software to finish “Preparing pipette head” and then open the rotating cover or the hood door.
Remove Alignment Head (If Installed)
Remove the alignment head by disconnecting the terminal block connector from the rear panel, and then removing the thumbnuts securing it to the upper pipettehead assembly. Cover the hole on the rear panel from the exterior using the blackplug supplied in the accessory package.
Connect External Safety Interlock Sensor
If your PIPETMAX does not have a rotating cover, connec t the terminal block connectors for the external safety interlock sensor to the input/output ports on the rear panel. They are labeled TOP and BOTTOM to help ensure proper connection.
Run Application Protocol
Install Tip Ejector Foot
If using a MAX8x20 or MAX8x200 pipette head, skip to Install Pipette Head(s) on page 2-18. If using the MAX1000 pipette head, you ordered a tip ejector foot kit for eight and one channel.
1 Using the supplied 3/32" hex wrench, remove the shoulder screw in the installed
tip ejector foot.
2 Remove the tip ejector foot.
3 Slide the tipejector foot for eight and one channel onto the ejectorleg.
4 Use the supplied 3/32" hex wrench to start threading the shoulder screw, but do not
fully tighten.
Gilson PIPETMAX® 268 User’s Guide 2-17
Page 40
Setup/Operation
Install Pipette Head(s)
Install the pipette head(s) on the upper pipette head assembly using the thumb nuts. Ensure proper alignment of the pipette head against the upper pipette head assembly before tightening the thumbnuts.
If using only one pipette head, install a blank head on the other side.
Run Application Protocol
Refer to the following instructions for how to convert the alignment head into a blank head. The instructions describe how to convert it and remove it; however, it can be converted while installed.
a) Push up to remove the bracket that is attached to alignment head.
b) Remove the thumb nuts securing the alignment head to the upper
pipette head assembly.
c) Pull down on the alignment pin to remove it.
d) Install the blank head on the upper pipette head assembly using the
thumb nuts.
2
If a pipette head is already installed and needs to be removed, remove the pipette head by removing the thumbnuts securing it to the upper pipette head assembly.
Finish Tip Ejector Foot Installation
If using a MAX8x20 or MAX8x200 pipette head, skip to the next step (Select Pipette Head Serial Number
Manually move the tip ejector foot down to the bottom, and while holding it down, tighten the shoulder screw.
).
2-18 Gilson PIPETMAX® 268 User’s Guide
Page 41
Setup/Operation
2
Select Pipette Head Serial Number
Select the serial number for each installed pipette head from the drop-down lists, close the rotating cover or hood door, and then select Next.
If the serial number is not displayed in the drop-down list, the pipette head must be added:
1Select Add new pipette head.
2 Do any of the following:
If using a tablet, select Scan and then select Camera. Optionally, select the icon
to switch between the front and rear camera. When prompted, scan the barcode on the pipette head.
Enter the values from the Gilson Quality Control Report supplied with each
pipette head.
Using the optional barcode scanner (ordered separately), select Scan, and then
scan the barcode on the pipette head.
3Select Save.
4 Repeat steps 2 and 3 for additional pipette heads. When finished adding heads,
select Back.
5 Select the serialnumber for each installed head, close the rotating cover or hood
door, and then select Next.
Run Application Protocol
Gilson PIPETMAX® 268 User’s Guide 2-19
Page 42
Setup/Operation
Tip Rack Assembly
Beveled Corner Back Left
Tip Setup and Labware Setup
Proceed through the “Tip setup” and “Labware setup” screens by reviewing the information, making any desired, allowable changes, and then touching Next. The screens indicate the labware to use and where to place it on the tray.
Tip Rack Assembly
Each tip rack consists of a tip pack and a tip reload block.
To assemble the tip rack:
1 Open the sealed blister pack.
2 Lift the tip pack out of the blister pack.
Run Application Protocol
3 Set the tip pack on the tip reload block. Ensure that the tip pack is fully seated on
the tip reload block.
4 Remove the cover.
Tip Rack Placement
Tip racks can only be placed in positions with installed clips.
2
To install the tip rack:
1 Loosen the thumbscrew for each clip.
2 Place the tip rack on the removable tray with the beveled corner in the
back left corner.
3 Rotate each clip so that it will effectively secure the tip rack on the tray as shown
in the image below.
4 Tighten the thumbscrew for each clip.
Changing Tip Pack During the Run
If prompted during the run to refill the tip rack, use two hands and lift carefully when removing the empty tip pack from the tip reload block. Jarring movements could cause spillage of liquids from other racks.
Tip Rack Removal
To remove the tip rack from the removable tray:
1 Loosen the thumbscrew for each clip.
2 Rotate each clip away from the tip reload block.
3 Lift the tip rack off the removable tray.
4 Remove the tip pack from the tip reload block.
2-20 Gilson PIPETMAX® 268 User’s Guide
Page 43
Setup/Operation
2
Review Initial Volumes
Review initial volumes, make any desired, allowable changes, and then select Next.
Run Protocol
1Select Run protocol.
2 When prompted, it is recommended to simulate if this is the first time running the
protocol.
Run Application Protocol
Gilson PIPETMAX® 268 User’s Guide 2-21
Page 44
Setup/Operation
Browse Positions Manually
The “Labware setup guide” screen displays the options for confirming the instrument setup.
Select Browse positions manually to do any of the following:
Review the layout of the tray, review pipette heads, and set up or add new pipette heads (if necessary).
Select and set up specific tip racks and labware.
Manual Setup
Run Application Protocol
After choosing Browse positions manually, most of the protocol setup occurs on the “Manual setup” screen.
The “Manual setup” screen displays an image of the tray and allows you to choose which tips or labware to set up and also displays an image of the pipette heads with serial numbers required for the protocol so that you can confirm that your hardware configuration matches what the software expects.
2
Pipette Head Setup
Select Pipette head setup to go to the “Pipette head and tray setup” screen for selecting or adding pipette heads. The software will remember your selection when you go back to the “Manual setup” screen.
Tip or Labware Setup
Select a tip rack or a rack in the tray to go to the setup screen for that item.
Select a tip rack to go to the “Tip setup” screen.
Select a rack to go to the “Labware setup” screen.
Review Initial Volumes
Review initial volumes, make any desired, allowable changes, and then select Next.
Wizard Complete
1 On the “Wizard complete” screen, select Run protocol.
2 When prompted, it is recommended to simulate.
2-22 Gilson PIPETMAX® 268 User’s Guide
Page 45
Setup/Operation

Pause or Stop a Running Protocol

While the run is in progress, DO NOT open the hood; however, if the hood has been opened and you choose to continue the run, observe that the run resumes as expected.

Pause

While running a protocol, select anywhere on the PIPET STATUS image to pause the run. The run will stop when it finishes its current action. The protocol run timer keeps advancing during the pause. When prompted, select Continue to resume the run, or
Stop to end the run and view the results.

STOP

To stop PIPETMAX immediately, press the red STOP button on the front of PIPETMAX.
2
Pause or Stop a Running Protocol
Gilson PIPETMAX® 268 User’s Guide 2-23
Page 46
Setup/Operation

View Results

During and after a protocol run, information is available about volumes (initial volume, volume added, volume removed, and current volume).
After a protocol run, results for that run are automatically displayed.
View Results
To view results:
1 Select to go to the “Home” screen.
2 Select Results.
3 Select a protocol, and then select Results.
2
4 Select a protocol run, and then select View. Simulated runs are identified by .

Tray View

The first screen displayed when viewing results is the “Tray view” screen.
The colors indicate the following information:
Red - Negative volume
Green - Volume in the tube or well
Blue - Empty tube or well
Pink - Tips
Select a tip rack, plate, or rack for more information about the selection. For more information, refer to Labware View page 2-25.
The top of the screen shows the status, the name of the protocol, the time it took to run or simulate the protocol and generate the results, and the time of execution for the last step.
on
2-24 Gilson PIPETMAX® 268 User’s Guide
Page 47
Setup/Operation
2

Labware View

The “Labware view” screen displays an image for the selected labware (tip rack, plate, rack, etc.) and the bed element name assigned to it.
Select a well or tube for more information about the selection. For more information, refer to Volume View
The top of the screen shows the status, the name of the protocol, the time it took to run or simulate the protocol and generate the results, and the time of execution for the last step.
Tip Rack
Source
on page 2-26.
View Results
Destination
Tip Waste
Gilson PIPETMAX® 268 User’s Guide 2-25
Page 48
Setup/Operation

Volume View

The “Volume view” screen displays information about the volume in the selected tube or well.
The top of the screen shows the status, the name of the protocol, the time it took to run or simulate the protocol and generate the results, and the time of execution for the last step.
View Results
The text above the table lists the bed element name, the well location and label, and the current volume in the well.
The table lists the actions that occurred in the well or tube in the following order:
Initial volume (if any)
Volume added and the source of the volume added (bed element name and plate index or tube number)
•Volume removed
Source
2
Destination

Steps View

The “Steps view” screen displays the steps that were run in the protocol in the order of execution.
The top of the screen shows the status, the name of the protocol, the time it took to run or simulate the protocol and generate the results, and the time of execution for the last step.
2-26 Gilson PIPETMAX® 268 User’s Guide
Page 49
Setup/Operation

Run Report

The Run Report is a subset of the results and can be exported for viewing. It cannot be viewed from within the software or on the tablet.
It includes basic information about the protocol run, the bed layout, and well tracking.
It does not include the information about volume tracking.

Export and View Run Report

1 Select to go to the “Home” screen.
2
Run Report
2Select Results.
3 Select a protocol, and then select Results.
4 Select a protocol run, and then select Export.
5 Select the file type for the exported file: .htm (default), .xml, or .csv.
6 Browse for and select the USB drive, name the file (or accept the default name, which
is the name of the protocol), and then select Save. The Run Report is saved to the location.
7Open the file on a PC.
Gilson PIPETMAX® 268 User’s Guide 2-27
Page 50
Setup/Operation

Export Protocol

To export a protocol (and its results):
1 Select to go to the “Home” screen.
2 Select Manage protocols to go to the listing of protocols.
3 Select a protocol in the list. It will highlight when selected.
Export Protocol
4 Select Export.
5 Browse for and select the destination for the protocol file, and then select Open.
The protocol is saved to the location with a .SQLITE extension.

Log Out

1 Select to go to the “Home” screen.
2 Select Return to TRILUTION micro.
3 Select Log out to allow another user to log in to the software.
2
2-28 Gilson PIPETMAX® 268 User’s Guide
Page 51

Messages 3

TRILUTION® micro will display a message if a known error is encountered. This chapter describes some of the more common errors and how to prevent or resolve them.
This chapter includes:
Communication
Simulation
Hood Open
Insufficient Volume/Excessive Volume
Refill Tip Rack
Gilson PIPETMAX® 268 User’s Guide 3-1
Page 52
Messages
On-screen pause button

Communication

If a message appears that indicates that PIPETMAX cannot be found, check your settings (including serial number) and cabling.
Communication

Simulation

TRILUTION micro includes a pre-run simulation feature that can detect errors that would be encountered during a run.

Hood Open

Close the rotating cover or hood door. PIPETMAX will not operate with the cover or hood door open.
3
While the run is in progress, DO NOT open the hood; however, if the hood has been opened, you will see the following message. If you choose to continue the run, close the rotating cover or hood door and then observe that the run resumes as expected.
The preferred way to stop a run is via the on-screen pause button.
In an emergency, press the STOP button on the front panel of PIPETMAX.
For more information, refer to Pause or Stop a Running Protocol
on page 2-23.
3-2 Gilson PIPETMAX® 268 User’s Guide
Page 53
Messages

Insufficient Volume/Excessive Volume

If there is not enough volume in a well or tube to complete an aspirate or if a dispense will overfill a well or tube, TRILUTION micro will display a message with an illustration that identifies the location of the plate on the tray and the well or tube with the incorrect volume.
To resolve this error:
1 Open the rotating cover or hood door.
2 Add or remove volume from the well or tube with the incorrect volume.
3 Close the rotating cover or hood door.
4 Enter the volume added or removed.
5 Optionally, select “Don’t continue Checking Volume” if you do not want to be
notified or other wells with insufficient or excessive volume in this protocol run.
6Touch Continue to resume the run or Stop to stop the run.
3
Insufficient Volume/Excessive Volume

Refill Tip Rack

If the during a simulation the software calculates that you do not have enough tips to run your protocol or if while running the protocol you run out of tips, the message at right appears. Open the rotating cover or hood door, replace or refill the tip rack, close the rotating cover or hood door, and then touch Continue.
Note: Use two hands and lift carefully when removing the empty tip pack from the tip reload block. Jarring movements could cause spillage of liquids from other racks.
Gilson PIPETMAX® 268 User’s Guide 3-3
Page 54
Page 55

Maintenance 4

PIPETMAX is designed to require a minimum level of maintenance.
This chapter includes information about:
Cleaning
Add Pipette Heads
Replace Pipette Heads
Convert Blank Head into Alignment Head
Gilson PIPETMAX® 268 User’s Guide 4-1
Page 56
Maintenance

Cleaning

Keep PIPETMAX clean for peak performance. Always turn the power o ff to PIPE TM AX b ef ore cl ean in g. W ipe PI PE TMA X w ith a soft cloth dampened with a mild detergent and disinfect as needed. For information about cleaning the touchscreen on the tablet,
Cleaning
refer to the documentation supplied with it.

Add Pipette Heads

When adding pipette heads it is necessary to do two things:
1 Add the pipette head to the software by scanning or entering the calibration values. For detailed instructions, refer to
Add Pipette Head(s)
on page 2-14.
4
2 Physically install the head on the instrument. For detailed instructions and a diagram, refer to Install Pipette Head(s)
page 2-18.
on

Replace Pipette Heads

When replacing or changing pipette heads, use the Step-by step wizard. Follow the steps for running an application protocol that start on page 2-14
.

Convert Blank Head into Alignment Head

To convert a blank head into an alignment head:
1 Push the alignment pin into the hole furthest back on the blank
head. The pin fits either way and there is a magnetic indentation on the pin to lock it in place in the head. You should feel a change when pushing it in and the indentation should not be visible.
2 Install the alignment head on the left side of the upper pipette
head assembly using the thumb nuts. Ensure proper alignment of the head against the upper pipette head assembly before tightening the thumbnuts.
3 Align the bracket on the blank head as shown in the illustration.
4 Pull down on the bracket to lock it in place.
4-2 Gilson PIPETMAX® 268 User’s Guide
Page 57

Replacement Parts A

PIPETMAX® 268

Part Number Description
32100000 PIPETMAX 268 WITH STANDARD COVER
32100001 PIPETMAX 268 WITH COVER CUTOUTS
32100002 PIPETMAX 268 W EXTERNAL SAFETY INTERLOCK

TRILUTION® micro Software

Part Number Description
32000302 TRILUTION® MICRO INSTALLED ON TOUCHSCREEN TABLET
32000285 TRILUTION® MICRO ON USB DRIVE FOR PC INSTALL

Pipette Heads

Part Number Description
FC10022 PIPETMAX MAX8X20 PIPETTE HEAD
FC10021 PIPETMAX MAX8X200 PIPETTE HEAD
FC10010 PIPETMAX MAX1000 PIPETTE HEAD

Removable Trays

Part Number Description
32000109 PIPETMAX 268 TRAY STANDARD 96 WELL
32000091 PIPETMAX 268 TRAY 384 WELL

Tip Reload Block

Part Number Description
32000175 PIPETMAX TIP RELOAD BLOCK
Gilson PIPETMAX® 268 User’s Guide A-1
Page 58
Appendix

Tip Ejector Feet

Part Number Description
32000146 PIPETMAX TIP EJECTOR FOOT DUAL 8 CH
32000284 PIPETMAX 268 EJECTOR FOOT KIT 8 & 1 CH

Tips

Replacement Parts
Part Number Description
F172210 DIAMOND BLISTER, DSL10, 960 AUTO TIPS
F172211 DIAMOND BLISTER, DSL10ST, 960 AUTO TIPS
F172213 DIAMOND BLISTER, DSFL10ST, 960 AUTO TIPS
F172310 DIAMOND BLISTER, DS200, 960 AUTO TIPS
F172311 DIAMOND BLISTER, DS200ST, 960 AUTO TIPS
F172313 DIAMOND BLISTER, DSF30ST, 960 AUTO TIPS
F172513 DIAMOND BLISTER, DSF200ST, 960 AUTO TIPS
F172510 DIAMOND BLISTER, DS1000, 960 AUTO TIPS
F172511 DIAMOND BLISTER DS1000ST, 960 AUTO TIPS
A
F172613 DIAMOND BLISTER DSF1000ST, 960 AUTO TIPS
F172200 DIAMOND BLISTER, DL10, 960 TIPS
F172300 DIAMOND BLISTER, D200, 960 TIPS
F172500 DIAMOND BLISTER, D1000, 960 TIPS
F172201 DIAMOND BLISTER, DL10ST, 960 TIPS
F172301 DIAMOND BLISTER, D200ST, 960 TIPS
F172501 DIAMOND BLISTER, D1000ST, 960 TIPS
F172203 DIAMOND BLISTER, DFL10ST, 960 TIPS
F172303 DIAMOND BLISTER, DF30ST, 960 TIPS
F172503 DIAMOND BLISTER, DF200ST, 960 TIPS
F172603 DIAMOND BLISTER, DF1000ST, 960 TIPS

Tip Waste Disposal

Part Number Description
32000274 TIP DISPOSAL BIN
32000177 PIPETMAX 268 RISER OFF BED TIP DISPOSAL
A-2 Gilson PIPETMAX® 268 User’s Guide
Page 59
Appendix

Racks

Part Number Description
32000196 RACK CODE 496 PCR TUBES
32000198 RACK CODE 424 1.5-2.0 ML TUBE FLIP CAP
32000197 RACK CODE 425 0.5-2.0 ML TUBE SCREW CAP
32000201 RACK CODE 440 SBS TILT RACK 10 DEG

Passive Temperature (PT) Racks

Part Number Description
32000232 RACK CODE 424 PT PSV TMP FLP CAP 1.5-2ML
32000200 RACK CODE 425 PT PSV TMP SCR CAP 0.5-2ML
32000233 RACK CODE 410 PT PSV TMP FLAT PLATE
32000238 RACK CODE 496 PT PSV TMP 96 PCR TUBE

Circulating Temperature (CT) Racks

A
Replacement Parts
A Circulating Temperature rack requires a circulating bath (ordered separately), RACK CODE 410 CT CIR TMP FLAT PLATE (ordered separately, part number 32000208), INSTALL KIT CT CIR TMP (ordered separately, part number 32000193), and can be used with the following adapters:
Part Number Description
32000262 ADAPTER 424 CT CIR TMP FLP CAP 1.5-2ML
32000261 ADAPTER 425 CT CIR TMP SCR CAP 0.5-2ML
32000260 ADAPTER 496 CT CIR TMP 96 PCR TUBE

Accessories

Part Number Description
32000199 ORBITAL SHAKER
32000244 HAND HELD MICROSCAN 2D BARCODE READER
32000241 TABLET HOLDER

Miscellaneous

Part Number Description
54210001 PIPETMAX 268 PLUG FOR BACK COVER
32000174 PIPETMAX ALIGNMENT HEAD
32000275 PIPETMAX 268 PIPETTE HEAD NUT (QTY 5) SET
32000012 USB A MALE-B MALE 2M BLK
Gilson PIPETMAX® 268 User’s Guide A-3
Loading...