Gilson PIPETMAN Neo P200N, PIPETMAN Neo P10N, PIPETMAN Neo P20N, PIPETMAN Neo P1000N, PIPETMAN Neo 8x20 User Manual

...
Page 1
e Pipetting Comfort with Improved Comfort!
IMPORTANT NOTICE!
Do not lubricate this pipette.
The use of lubricant cancels the warranty of this pipette.
Page 2
1  INTRODUCTION
PIPETMAN Neo is a fully adjustable air displacement pipette and is used with disposable tips.
PIPETMAN Neo is an evolution of the PIPETMAN Classic to address the growing susceptibility of pipette users to repetitive strain injuries. The implementation of new features to improve pipetting comfort has no impact on the legendary robustness, accuracy and precision of PIPETMAN Classic.
Six single channel models cover a volume range from
0.2 µL to 1000 µL.
Two 8 and two 12 multichannel models cover a volume
range from 2 µL to 200 µL.
Contents Page
1  INTRODUCTION 2
2  PARTS CHECK LIST 3
3  DESCRIPTION 3
4  SETTING THE VOLUME 4
5  PIPETTING 5
6  GENERAL GUIDELINES FOR GOOD PIPETTING 9
7  ACCESSORIES 10
8  GLP FEATURES 11
9  TROUBLESHOOTING 11
10  LEAK TEST 12
11  MAINTENANCE 13
12  CLEANING AND DECONTAMINATION 17
13  SPECIFICATIONS 21
14  SPARE PARTS 23
EC DECLARATION OF CONFORMITY 26
WARRANTY 27
Page 3
2  PARTS CHECK LIST
Just take a moment to verify that the following items are present:
- PIPETMAN Neo,
- User's Guide,
- Safety bag,
- Certificate of conformity (including bar-code sticker).
3  DESCRIPTION
Please refer to the following chapters for a full description of the different parts and functions of the pipette.
Color-coded push-button
Thumbwheel
Tip-ejector button
Handle
Connecting nut
PIPETMAN DIAMOND® tip
Tip-holder
Tip-ejector
Page 4
4  SETTING THE VOLUME
The volume of liquid to be aspi­rated is set using the volumeter.
Color-coded push-button
Thumbwheel
Tip-ejector button
Handle
Connecting nut
PIPETMAN DIAMOND tips
Tip-holders
Tip-ejector
Ejector spacer
Cover
Ejector support
Ejector clip
Ejector lock
Model Color of volumeter numbers
Black Red Increment
P2N µL 0.01 µL 0.002 µL P10N to P20N µL 0.1 µL 0.02 µL P100N-P200N µL - 0.2 µL P1000N 0.01 mL mL 0.002 mL
1
5
P2N
1.25 µL
2
0 7
5
P10N
7.5 µL
1 2
5
P20N
12.5 µL
P100N
0 7 5
75 µL
P200N
1 2 5
125 µL
P1000N
0
7 5
0.75 mL
Page 5
The dials are colored either black or red to indicate the position of the decimal point, depending on the model (see examples).
The volume is set by turning the thumbwheel or the push­button. The push-button makes it easier and quicker to set volumes, especially when wearing gloves. The thumbwheel may be turned using only one hand to slowly reach the required setting.
To obtain maximum accuracy when setting the volume, proceed as follows:
when decreasing the volume setting, slowly reach
the required setting, making sure not to overshoot the mark.
when increasing the volume setting, pass the required
value by 1/3 of a turn and then slowly decrease to reach the volume, making sure not to overshoot the mark.
5  PIPETTING
For optimum performance, use of PIPETMAN
DIAMOND Tips with your PIPETMAN Neo is strongly recommended. These Tips, made from pure polypropylene have the Gilson logo engraved on their collar, ensuring that you have a genuine Gilson product. Plastic tips are for a single application – they must not be cleaned for reuse.
PIPETMAN Neo can also be used with the main tip brands.
Fitting the tips
Single channel models:
To t a new PIPETMAN DIAMOND Tip, push the tip-holder into the tip using a slight twisting motion to ensure a firm, airtight seal.
For the P2N and P10N models, a dual-position adapter (plastic) is required to fit DL10 tips (long tips) or D10 tips (short tips). The metallic rod of the tip-ejector is shaped so that the adapter may be clipped to it in either of two positions.
Push-button
Thumbwheeel
Page 6
P2N and P10N models are delivered with the adapter
in place, positioned in the longer slot, ready to use DL10 tips. When D10 tips (which are shorter) are used, the adapter must be repositioned in the shorter slot as follows:
Pull the adapter down from the metallic rod.
Turn the adapter through 180°C.
Refit the adapter so that the end of the metallic rod engages the shorter slot of the adapter.
Multichannel models:
PIPETMAN DIAMOND Tip: TIPACK™ and TOWERPACK™ – ROCKY RACK™
PIPETMAN DIAMOND Tips are best fitted from the pat­ented ROCKY RACK available only in our TIPACKS and TOWER PACKS.
ROCKY RACK is the dome-shaped part of the pack that contains the tips. ROCKY RACK makes it easy to securely fit the tips to a multichannel pipette, ensuring an airtight seal on all channels without the need to use undue pres­sure or to touch the tips.
The patented ROCKYRACK system available on TOWERPACK and TIPACK makes it extremely easy to t tips on a multichannel pipette. Tips will not fall o nor will they have to be manually positioned.
Ejector spacer
PIPETMAN DIAMOND Tip compatibility
8x20 12x20
DL10, DFL10, DF30, D200
8x200 12x200
D200, D300, DF100, DF200, DF300
Short tips Long tips
Page 7
Pre-rinse the tips
Some liquids (e.g. protein-containing solutions and organic solvents) can leave a lm of liquid on the inside wall of the tip ; pre-rinse the tip to minimize any errors that may be related to this phenomenon.
Pre-rinsing consists of aspirating the rst volume of liquid and then dispensing it back into the same vessel (or to waste). Subsequent volumes that you pipette will have levels of accuracy and precision within specications.
Page 8
First stop
Second stop
Aspirate
Press the push-button to the first stop (this corresponds to the set volume of liquid).
Hold the pipette vertically and immerse the tip in the liquid (see immersion depth table, page 10). Release the push-button slowly and smoothly (to top posi- tion) to aspirate the set volume of liquid. Wait one second (time depends on model, see table) ; then withdraw the pipette-tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe, however if you do so take care to avoid touching the tip’s orifice.
Dispense
Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
Press the push-button slowly and smoothly to the rst stop.
Wait for at least a second, then press the push-button to the second stop to expel any residual liquid from the tip. Keep the push-button pressed fully down and (while removing the pipette) draw the tip along the inside surface of the vessel.
Page 9
Release the push-button, smoothly. Eject the tip by pressing rmly on the tip-ejector button.
6  GENERAL GUIDELINES FOR GOOD PIPETTING
Make sure that you op­erate the push-button slowly and smoothly.
When aspirating, keep the tip at a constant depth below the sur­face of the liquid (refer to the table).
Change the tip before aspirating a dierent liquid, sample, or reagent.
Change the tip if a droplet remains at the end of the tip from the previous pipetting operation.
Each new tip should be pre-rinsed with the liquid to be pipetted.
Liquid should never enter the tip-holder; to prevent this:
- press and release the push-button slowly and smoothly,
- never turn the pipette upside down,
- never lay the pipette on its side when there is liquid in the tip.
If you use the same tip with a higher volume, pre­rinse the tip.
For volatile solvents you should saturate the air­cushion of your pipette by aspirating and dispensing the solvent repeatedly before aspirating the sample.
When the temperature of the liquid is dierent from the ambient temperature, pre-rinse the tip several times before use.
You may remove the tip-ejector (see Chapter 11 - Maintenance) to aspirate from very narrow tubes.
After pipetting acids or other corrosive liquids that emit vapors, remove the tip-holder, rinse the
Table - Immersion Depth and Wait Time
Model Immersion Wait Time Depth (mm) (seconds)
P2N 1 1 P10N 1 1 P20N 2-3 1 P100N 2-4 1 P200N 2-4 1 P1000N 2-4 2-3 8x20, 12x20 2-3 1 8x200, 12x200 2-3 1
Page 10
10
piston, O-ring and seal with distilled water. For the model P1000N, by using a specic tip holder equipped with a lter, you can increase the life­time of the piston (see Chapter 7 - Accessories).
Do not pipette liquids having temperatures above 70°C or below 4 °C. The pipette can be used between + 4 °C and + 40 °C but the specifications may vary according to the temperature (refer to the ISO 8655-2 standard for conditions of use).
7  ACCESSORIES
To make pipetting more comfortable and more secure, Gilson has developed several accessories:
To avoid the possibility of liquid running back into the pipette, store the pipette vertically.
To identify or personalize your pipette, COLORIS™ clips are available:
With THE JIMMY™, hands free microtube opener, you can open both snap-cap and screw-cap microtubes.
To protect the piston when pipetting corrosive liq­uids, you can use a specic tip holder and lter for the model P1000N:
CARROUSEL™ Pipette stand (7 pipettes) F161401 TRIO™ stand (3 pipettes) F161405 SINGLE™ pipette holder F161406
COLORIS™ clips (mixed colors set of 10) F161301 COLORIS™ clips (red, set of 10) F161302 COLORIS™ clips (yellow, set of 10) F161303 COLORIS™ clips (green, set of 10) F161304 COLORIS™ clips (blue, set of 10) F161305 COLORIS™ clips (white, set of 10) F161306
THE JIMMY™ (set of 3) F144983
Corrosion protection kit (tip holder + a bag of 10 filters) F144570
Page 11
11
8  GLP FEATURES
The Serial Number is engraved on the body of the pipette. It provides unique identication of your pipette and the date of manufacture; Ex : AA 10369
The Bar Code on the box and the certificate of conformity provide traceability of your pipette.
9  TROUBLESHOOTING
A quick inspection of the pipette may help you to detect a problem.
You may download from the Gilson website (www.gilson.com) the “2 minute inspection”, which shows how to perform a quick diagnosis of your pipette.
Before returning any pipette to your local Gilson Service Center, ensure that it is completely free of chemical, biological, or radioactive contamination. Refer to Chapter 12 - Cleaning and Decontamination. Please use the included safety bag to return the pipette to your local Gilson Service Center.
The following table may help you to identify and correct the problem you might encounter.
Symptom Possible Cause Action
Pipette is Damaged tip-holder Replace the tip-holder* leaking sample Worn O-ring or seal Replace both parts*
Pipette won’t Worn O-ring Replace both parts* aspirate Damaged tip-holder Replace the tip-holder* Connecting nut is loose Tighten connecting nut Damaged or corroded piston Return pipette to supplier Improper repair or assembly See Chapter 11 - Maintenance
Page 12
12
* Not applicable for multichannel models
However, if you can’t solve the problem, contact your Gilson representative.
Symptom Possible Cause Action
Pipette is
Tip-holder is loose Tighten connecting nut*
not precise
Connecting nut is loose Tighten connecting nut Incorrect operator technique Operator training Damaged or corroded piston(s) Return pipette to supplier Damaged tip-holder(s) Replace the tip-holder* Worn O-ring or seal Replace both parts*
Tips fall off or Low quality tips Use PIPETMAN DIAMOND tips do not fit Damaged tip-holder(s)
Replace the tip-holder* correctly Damaged tip-ejector Replace tip-ejector Ejector spacer is missing Mount the spacer on the tip- ejector The ejector spacer is damaged Replace the ejector spacer The tip-ejector is loose Assemble the tip-ejector properly The ejector lock is misaligned Align the ejector lock
10  LEAK TEST
This test may be performed at any time to check that the pipette does not leak, especially after performing a maintenance or decontamination procedure. If a pipette fails this test, replace the O-ring and seal. After making sure that the pipette is correctly reassembled, repeat this test (not applicable to multichannel pipettes).
For the P2N to P200N, 8x20, 12x20, 8x200 and 12x200 models:
Fit a PIPETMAN DIAMOND Tip.
Set the pipette to the maximum volume given in the specifications, and pre-rinse.
Aspirate the set volume from a beaker of distilled water.
Maintain the pipette in the vertical position and wait for 20 seconds.
If a water droplet appears at the end of the tip there is a leak.
If you see no droplet, re-immerse the tip below the surface of water.
Page 13
13
The water level inside the tip should remain constant; if the level goes down there is a leak.
For the P1000N model:
Fit a PIPETMAN DIAMOND tip.
Set the pipette to the maximum volume given in the specifications.
Aspirate the set volume from a beaker of distilled water.
Maintain the pipette in the vertical position and wait for 20 seconds.
If a water droplet appears at the end of the tip, there is a leak.
11  MAINTENANCE
Routine maintenance will help keep your pipette in good condition, ensuring a continued high level of performance. Maintenance is limited to cleaning or au­toclaving the parts specified under Chapter 12 - Cleaning and Decontamination or to replacing the push-button, connecting nut, tip-ejector, tip-holder, seal and O-ring.
PIPETMAN Neo P2N and P10N should not be disassembled, so you may only replace the push-button,
tip-ejector, dual position tip-ejector and its adapter. With these pipettes if the tip-holder is damaged, the piston may also be damaged.
For multichannel models, the lower part should not be disassembled: only the push-button, the connecting nut
and the tip-ejector can be replaced.
After replacing any parts you should verify the performance of your pipette following the verification procedure available on the Gilson website (www.gilson.com). If the pipette needs to be readjusted, please contact your local Gilson authorized Service Center.
Page 14
14
Single channel models
Changing the Tip-ejector, to remove
Push the ejection button.
Push laterally the tip-ejector.
Slide and remove the tip-ejector.
To ret
Push the ejection button.
Slide the tip-ejector along the tip-holder.
Clip the tip-ejector on the body of the pipette.
Changing the Tip-holder – no tools required
Remove the tip-ejector (see above).
Unscrew the connecting nut by turning it counter­clockwise.
Carefully separate the lower and upper parts.
Remove the piston assembly, O-ring and seal.
Clean, autoclave, or replace the tip-holder.
Reassemble the pipette (refer to the figure, p 15).
Tighten the connecting nut (turn clockwise).
Refit the tip-ejector (see above).
Servicing the Piston Assembly
You may remove the piston assembly for cleaning pur­poses only. If the piston assembly is changed, the pipette must be adjusted and calibrated in a Gilson authorized Service Center.
Remove the tip-ejector (see above).
Unscrew the connecting nut by turning it counter­clockwise.
Carefully separate the lower and upper parts.
Remove the piston assembly, O-ring and seal.
Clean and decontaminate the piston assembly.
Page 15
15
Color coded push-button
Connecting nut
Seal
O-ring
Tip-holder
Tip-ejector
Piston assembly
The piston assembly must not be autoclaved.
Page 16
16
Reassemble the pipette (refer to the figure, page 15).
Tighten the connecting nut (turn clockwise).
Refit the tip-ejector (see above).
Changing the O-ring
The O-ring and seal are on the piston; they must not be autoclaved, if worn or damaged in any way (chemical or
mechanical), they must be replaced.
The dimensions of the O-ring vary depending on the model of pipette.
Remove the tip-ejector (see above).
Unscrew the connecting nut by turning it counter­clockwise.
Carefully separate the lower and upper parts.
Remove the piston assembly, O-ring and seal.
Clean or replace the seal then the O-ring.
Reassemble the pipette (refer to the figure, page 15).
Tighten the connecting nut (turn clockwise).
Refit the tip-ejector (see above).
Multichannel models
Changing the Tip-ejector
To remove the tip­ejector, keep both ejector locks de­pressed.
Pull the tip-ejector down.
To refit the tip­ejector, gently re­insert the tip- ejector vertically into the rails of the ejector support.
Pull lightly on the tip-ejector to check the position.
Ejector lock
Tip-ejector
Ejector spacer
Page 17
17
12  CLEANING AND DECONTAMINATION
PIPETMAN Neo is designed so that the parts normally in contact with liquid contaminants, can easily be cleaned and decontaminated. However, because the models P2N and P10N contain miniaturized parts, it is best not to disas­semble these pipettes yourself; please contact your local Gilson authorized Service Center.
Cleaning
The pipette must be cleaned, as described below, before it is decontaminated. Soap solution is recommended for cleaning PIPETMAN Neo.
Single channel models - External
Remove the tip-ejector.
Wipe the tip-ejector with a soft-cloth or lint-free tissue impregnated with soap solution.
Wipe the entire pipette with a soft-cloth or lint- free tissue impregnated with soap solution, to remove all dirty marks. If the pipette is very dirty, a brush with soft plastic bristles may be used.
Wipe the entire pipette and the tip-ejector with a soft cloth or lint-free tissue soaked with distilled water.
Refit the tip-ejector and allow the pipette to dry.
Single channel models - Internal
The following components only can be immersed in a cleaning solution: connecting nut, tip-ejector, tip-holder, piston assembly, seal and O-ring.
Disassemble the pipette as described in the Chapter 11 - Maintenance.
Set aside the upper part in a clean, dry place.
Clean the individual components of the lower part of the pipette using an ultrasonic bath (20 minutes at
You may refer to the decontamination procedure available on the Gilson website (www. gilson. com). Liquid must never enter the upper part (handle) of any pipette.
Page 18
18
50°C) or with a soft-cloth and brushes. Small round brushes with soft plastic bristles may be used to clean the interior of the tip-holder.
Rinse the individual components with distilled water.
Leave the parts to dry by evaporation or wipe them with a clean soft-cloth or lint-free tissue.
Reassemble the pipette as described in the Chapter 11 - Maintenance.
Multichannel models
The following components only can be immersed in a cleaning solution: tip-ejector, ejector locks and ejector spacer.
Remove the tip-ejector and the ejector spacer.
Immerse the tip-ejector, ejector locks and ejector spacer in the cleaning solution or wipe them with a soft-cloth or lint-free tissue impregnated with the cleaning solution.
Rinse the components with distilled water.
Wipe the entire pipette with a soft cloth or lint-free tissue impregnated with the cleaning solution.
Wipe it with distilled water.
Leave the parts to dry by evaporation or wipe them with a clean soft-cloth or lint-free tissue.
Refit the tip-ejector as described in page 16 (Chang­ing the tip-ejector).
Autoclaving
Single channel models
The upper part (body) and the piston assembly of the pipette are not autoclavable. Only the following parts may be autoclaved: tip-ejector, tip-holder and connecting nut. The O-ring and seal are not autoclavable; they may be cleaned or replaced with the one specified in Chapter 14 - Spare Parts.
Clean the parts to be autoclaved, especially the tip-holder.
Put the parts in an autoclaving sack.
Autoclave for 20 minutes at 121°C, 0.1 MPa.
Page 19
19
Check that the parts are dry before re-assembling the pipette.
Set the pipette aside to stabilize at room temperature.
Multichannel models
The complete lower part can be autoclaved (20 minutes at 121°C, 0.1 MPa).
The lower part has to be removed from the body.
To remove:
Keep the tip-ejector button depressed,
Turn the ejector clip counter-clockwise and separate it from the handle.
Unscrew the con­necting nut turning counter-clockwise.
Press push-button to extract the head.
To autoclave:
Clean the parts to be autoclaved, especially the tip­holders and the ejector spacer.
Put the lower part in an autoclaving sack.
Autoclave for 20 minutes at 121°C,
0.1 MPa.
Check that the auto­claved part is dry be ­fore re-assembling the pipette.
Set the pipette aside to stabilize at room temperature.
Page 20
20
To t the lower part on the body,
Press on the push-button.
Insert the lower part into the ejection rod. Release the push-button.
Screw the connecting nut turning clockwise.
Press on the tip-ejector button.
Insert the ejector clip and release the tip-ejector button.
Chemical Decontamination
You may choose to decontaminate your pipette chemi­cally, in accordance with your own procedures. Whatever decontaminant you use, check with the
supplier of the
decontaminant that it is compatible
with stainless steel and the plastics used in the construction of the pipette: PA (Polyamide), PBT (Polybutylene Terephtalate), PC (Polycarbonate), PC/PBT (Polycarbonate/ Polybutylene Terephtalate), POM (Polyoxymethylene), or PVDF (Poly­vinylidene Fluoride).
Single and Multichannel - Upper Part (handle)
Wipe the upper part (handle) of the pipette with a soft-cloth or lint-free tissue impregnated with the chosen decontaminant.
Wipe the upper part of the pipette with a soft- cloth or lint-free tissue soaked with distilled water or sterile water.
Single channel - Lower Part (Volumetric module)
The following components only can be immersed in a decontaminant solution: connecting nut, tip-ejector, tip-holder, piston assembly, seal and O-ring.
Page 21
21
Multichannel - Lower Part (Volumetric module)
The following components only can be immersed in a cleaning solution: tip-ejector, ejector locks and ejector spacer.
Remove the tip-ejector and the ejector spacer.
Immerse the tip-ejector, ejector locks and ejector spacer in the cleaning solution or wipe them with a soft-cloth or lint-free tissue impregnated with the cleaning solution.
Rinse the components with distilled water.
Wipe the entire pipette with a soft cloth or lint-free tissue impregnated with the cleaning solution.
Wipe it with distilled water.
Leave the parts to dry by evaporation or wipe them with a clean soft-cloth or lint-free tissue.
Refit the tip-ejector as described in page 16 (Chang­ing the tip-ejector).
13  SPECIFICATIONS
PIPETMAN Neo is a high quality pipette that oers ex­cellent accuracy and precision. The figures given in the “Gilson Maximum Permissible Errors” table (page 22) were obtained using PIPETMAN DIAMOND Tips. These figures are guaranteed only when genuine PIPETMAN DIAMOND Tips are used.
Each pipette is inspected and validated by qualified technicians in accordance with the Gilson Quality System. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655 standard, by type testing.
The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6).
Page 22
22
Multichannel
Maximum Permissible Errors
Model Volume Gilson ISO 8655
(Reference) (µL) Systematic Random Systematic Random error (µL) error (µL) error (µL) error (µL)
8x20 (F14401) Min. 2 ± 0.10 0.08 ± 0.40 0.20 and 10 ± 0.20 0.10 ± 0.40 0.20
12x20 (F14402) Max 20 ± 0.40 0.20 ± 0.40 0.20
8x200 (F14403) Min. 20 ± 0.50 0.25 ± 3.2 1.2
and 100 ± 1.0 0.40 ± 3.2 1.2 12x200 (F14404) Max. 200 ± 2.00 0.60 ± 3.2 1.2
Gilson Maximum Permissible Errors
Single channel Maximum Permissible Errors Model Volume Gilson ISO 8655
(Reference) (µL) Systematic Random Systematic Random error (µL) error (µL) error (µL) error (µL)
P2N Min 0.2 ± 0.024 0.012 ± 0.08 0.04 (F144561) 0.5 ± 0.025 0.012 ± 0.08 0.04 D10, DL10 Max. 2 ± 0.030 0.014 ± 0.08 0.04
P10N Min. 1 ± 0.025 0.012 ± 0.12 0.08 (F144562) 5 ± 0.075 0.030 ± 0.12 0.08 D10, DL10 Max 10 ± 0.100 0.040 ± 0.12 0.08
P20N Min. 2 ± 0.10 0.03 ± 0.20 0.10 (F144563) 10 ± 0.10 0.05 ± 0.20 0.10 D200 Max. 20 ± 0.20 0.06 ± 0.20 0.10
P100N Min. 10 ± 0.35 0.10 ± 0.80 0.30 (F144564) 50 ± 0.40 0.12 ± 0.80 0.30 D200 Max. 100 ± 0.80 0.15 ± 0.80 0.30
P200N Min. 20 ± 0.50 0.20 ± 1.60 0.60 (F144565) 100 ± 0.80 0.25 ± 1.60 0.60 D200 Max. 200 ± 1.60 0.30 ± 1.60 0.60
P1000N Min. 100 ± 3 0.6 ± 8 3 (F144566) 500 ± 4 1.0 ± 8 3 D1000 Max. 1000 ± 8 1.5 ± 8 3
The data given in the tables conform to the ISO 8655-2 Standard.
With a precise pipetting technique (see Chapter 6 - General guidelines for good pipetting) the P2N model may be used to aspirate volumes as low as
0.1 µL and the P10N model as low as 0.5 µL.
Page 23
23
14  SPARE PARTS
Single channel models
Service Kit 1st level includes:
- 3 piston seals
- 3 O-rings
- 1 tip-holder
Service Kit 2nd level includes:
- 1 push-button
- 1 connecting nut
- 1 tip-ejector for the P2N and P10N models
only:
- 1 tip-ejector and 1 adapter
A
B
C
D
E
F
A
B
C
D
E
F
F1
F2
F2
F1
Page 24
24
Tip-ejector (8 channels)
Tip-ejector (12 channels)
Ejector lock
Ejector spacer (8 channels) Ejector spacer (12 channels)
Multichannel models
F3 F4
G
H1 H2
Page 25
25
8x20 (F14401) and 12x20 (F14402)
Description 8x20 12x20
F3 - F4 Tip-ejector F507005 F507006
G Ejector lock F507008 F507008
H1 - H2 Ejector spacer F507001 F507003
8x200 (F14403) and 12x200 (F14404)
Description 8x20 12x20
F3 - F4 Tip-ejector F507005 F507006
G Ejector lock F507008 F507008
H1 - H2 Ejector spacer F507001 F507003
P20N (F144563) and P100N (F144564)
Description P20N P100N
C+D+E Service Kit 1st level F144495 F144496
A+B+F Service Kit 2nd level F161972 F161973
C+D Seal + O-ring (5 sets) F144863 F144864
P200N (F144565) and P1000N (F144566)
Description P200N P1000N
C+D+E Service Kit 1st level F144497 F144498
A+B+F Service Kit 2nd level F161974 F161978
C+D Seal + O-ring (5 sets) F144865 F144866
P2N (F144561) and P10N (F144562)
Description P2N P10N
C+D+E Service Kit 1st level F144501 F144502
A+B+F1+F2 Service Kit 2nd level F161970 F161971
C+D Seal + O-ring (5 sets) F144861 F144862
F2 Tip-ejector Adapter F144879 F144879
Page 26
26
EC DECLARATION OF CONFORMITY
PIPETMAN Neo® Single channel
P2N, P10N, P20N, P100N, P200N, P1000N
PIPETMAN Neo® Multichannel
8x20, 12x20, 8x200, 12x200
98/79/EC*
on In Vitro Diagnostic Medical Devices
comply with the requirements of the following European Directives:
Villiers-le-Bel, July 2nd, 2012
A. El Sayed H. Ledorze
General Manager Quality Manager
GILSON S.A.S.
19, avenue des Entrepreneurs BP. 145 - 95400 Villiers-le-Bel (France) Tel.: +33(0) 1 34 29 50 00 Fax: +33(0) 1 34 29 50 20 www.gilson.com
* Annex III, self-declared
The company,
Hereby certies on its sole responsibility that the products listed below:
Page 27
27
WARRANTY
Gilson warrants this pipette against defects in material under normal use and service for a period of 12 months from the date of purchase.
This warranty shall not apply to pipettes which are subject to abnormal use and/or improper or inadequate maintenance (contrary to the recommendations given in the User’s guide), including, but not limited to pipettes which have been subjected to physical damage, improper handling, or spillage or exposure to any corrosive environment. This warranty shall also be void in the event pipettes are altered or modied by any party other than Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes.
THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE, UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE, INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR IMPLIED WARRANTY.
Page 28
LT801521/D - ©2012 Gilson SAS All rights reserved July 2012English
Printed in France
Specifications subject to change without notifications - errors omitted.
Gilson, Inc. World Headquarters
3000 Parmenter Street
I
P.O. Box 620027
I
Middleton, WI 53562-0027, USA
Tel: (1) 800-445-7661 or (1) 608-836-1551
I
Fax: (1) 608-831-4451
Gilson S.A.S.
19, avenue des Entrepreneurs
I
BP 145, F-95400 Villiers-le-Bel, France
Tel: (33-1) 34 29 50 00
I
Fax: (33-1) 34 29 50 20
www.gilson.com
sales@gilson.com service@gilson.com training@gilson.com
Loading...