- 12 -
For the Multichannel Models only
The pipette must be cleaned, as described below, before it is
decontaminated. Soap solution is recommended for cleaning
PIPETMAN L.
Cleaning
The following components only can be immersed in a cleaning solution: tip-ejector,
ejector locks and ejector spacer.
Remove the tip-ejector and the ejector spacer.
Immerse the tip-ejector, ejector locks and ejector spacer in the cleaning solution
or wipe them with a soft cloth or lint-free tissue impregnated with the cleaning
solution.
Rinse the components with distilled water.
Wipe the entire pipette with a soft cloth or lint-free tissue impregnated with the
cleaning solution.
Wipe it with distilled water.
Leave the parts to dry by evaporation or wipe them with a clean soft-cloth or
lint-free tissue.
For the Single Models and Fixed Models only
The pipette must be cleaned, as described below, before it is
decontaminated. Soap solution is recommended for cleaning
PIPETMAN L.
Cleaning External
Remove the tip-ejector.
Wipe the tip-ejector with a soft-cloth or lint-free tissue
impregnated with soap solution.
Wipe the entire pipette with a soft-cloth or lint- free tis-
sue impregnated with soap solution, to remove all dirty
marks. If the pipette is very dirty, a brush with soft plastic
bristles may be used.
Wipe the entire pipette and the tip-ejector with a soft cloth
or lint-free tissue soaked with distilled water.
Refit the tip-ejector and allow the pipette to dry.
Cleaning Internal
The following components only can be immersed in a cleaning
solution: connecting nut, tip-ejector, tip-holder, piston assembly,
seal and O-ring.
Disassemble the pipette as described in the Chapter 12A
- Maintenance.
Set aside the upper part in a clean, dry place.
Clean the individual components of the lower part of the
pipette using an ultrasonic bath (20 minutes at 50°C) or
with a soft-cloth and brushes. Note that the piston assembly
and seals must be degreased with isopropanol or ethanol
before being immersed in another ultra sonic bath. Small
round brushes with soft plastic bristles may be used to
clean the interior of the tip-holder.
Rinse the individual components with distilled water.
Leave the parts to dry by evaporation or wipe them with
a clean soft-cloth or lint-free tissue.
Reassemble the pipette as described in the Chapter 12A
- Maintenance.
Autoclaving
The upper part (body) and the piston assembly of the
pipette are not autoclavable. Only the following parts may
be autoclaved: tip-ejector, tip-holder and connecting nut.
The O-ring and seal are not autoclavable; they may be cleaned
or replaced with the one specified in Chapter 15 - Spare Parts.
Clean the parts to be autoclaved, especially the tip-holder.
Put the parts in an autoclaving sack.
Autoclave for 20 minutes at 121°C, 0.1 MPa.
Check that the parts are dry before re-assembling the
pipette.
Set the pipette aside to stabilize at room temperature.
Chemical Decontamination
You may choose to decontaminate your pipette chemically, in
accordance with your own procedures. Whatever decontaminant you use, check with the
supplier of the decontaminant that
it is compatible
with stainless steel and the plastics used in the
construction of the pipette: PA (Polyamide), PBT (Polybutylene
Terephtalate), PC (Polycarbonate), PC/PBT (Polycarbonate/
Polybutylene Terephtalate), POM (Polyoxymethylene), PVDF
(Polyvinylidene Fluoride), or PP (Polypropylene).
Upper Part (handle)
Wipe the upper part (handle) of the pipette with a softcloth or lint-free tissue impregnated with the chosen
decontaminant.
Wipe the upper part of the pipette with a soft- cloth or
lint-free tissue soaked with distilled water or sterile water.
Lower Part (Volumetric module)
The following components only can be immersed in a de-
contaminant solution: connecting nut, tip-ejector, tip-holder.
Piston assembly and seals must be degreased with methyl
alcohol before being immersed in decontamination solution
in separate vessel.
Disassemble the pipette as described in the Chapter 12A.
Immerse tip-ejector, tip-holder and connecting nut in the
cleaning solution.
Degrease the piston assembly, the seals and immerse
them in another vessel.
Rinse each component with distilled water.
Leave the parts to dry by evaporation (or wipe with a soft
cloth the tip-ejector, the tip-holder and connecting nut).
Lubricate the piston assembly and the seals.
Reassemble the piston assembly, the tip-holder and the
tip-ejector.
Ret the tip-ejector as described in
“Changing the tip-ejector”.
Please note that although the lower part
of PIPETMAN L multichannel can withstand a few number of autoclaving cycles,
we do not recommend autoclaving it.
Ejector spacer
Ejector lock
Tip-ejector