Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
SAFETY INSTRUCTIONS
Review and completely understand the operating and safety instructions before using this machine.
WARNING!
This machine operates on electric current. Improper operation could result in electric shock,
electrocution, or an explosion!
1. The Performer III is designed for operation on 115V/60Hz power supplies. Connection to other power
sources will damage this machine and void the warranty. Contact Gilson Technical Support if operation
on an alternative power source is required. ALWAYS make sure the available power supply matches the
device requirements. Motors are NOT explosion-proof.
2. ALWAY S check electrical wiring for loose connections and for pinched or frayed wiring.
3. ALWAY S use a properly-wired, three-pronged plug, or otherwise ground the machine. Connect the
machine to a properly-wired, three-pronged receptacle. Make sure the cord is located where no one will
trip or get tangled in it.
4. ALWAYS disconnect and lock out power supply before performing maintenance and repairs.
WARNING!
• ALWAY S unplug or disconnect machine from the power source when the unit is not in
operation.
• ALWAY S wear safety glasses when operating, maintaining, or repairing this machine.
Page 2
Page 3
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
INTRODUCTION
• Quiet, electromagnetic vibratory action.
• 0—100% amplitude control.
• Switchable tapping action.
• Precise digital timing.
The Performer III 3in (76mm) Sieve Shaker is designed for small
samples of chemicals, minerals, pharmaceuticals, powdered
metals, cosmetics, abrasives, ores, foods, and other ne powders. Eective size range is No.4—No.635 (4.75mm—20μm) using woven wire sieves. Extended size ranges are possible with
some materials.
The High frequency, 3,600vpm, electromagnetic vibratory
action with 0—100% amplitude control is ideal for ne particle separations. The solenoid actuated tapping (60 taps/min)
speeds dry separations and can be used alone for tap-settling
and bulk density tests.
The Performer lll holds up to seven full-height metal sieves plus
pan or fourteen acrylic sieves and pan. The digital timer/controller and graduated vibration control knob allow the Performer lll
to give reliable, repeatable results. The optional GAA-88 Acrylic
Spacer is useful for observing sample action during set-up or
testing.
Figure 1
UNPACKING & SETUP
Carefully inspect the Performer lll as soon as it arrives and check
all package contents. See Figure 7 for an illustration of Performer
III components and accessories. If there is damage or if parts are
missing, contact Gilson Customer Service immediately at 800444-1508 or 740-548-7298 for instructions. Save all packing materials for inspection by the freight claims adjuster if damage is
reported.
The Performer III is packaged in two pieces to minimize potential damage during shipping. Place the Performer III on a dry,
level surface.
• To assemble, position the Tapper Clamping Assembly over
the four holes in the Base with the two rubber sieve locating bumpers positioned at the rear. Insert the 1/4-20 x 3/8in
panhead screws into the mounting holes and tighten snugly
(see Figure 1).
• Connect the main power cord to the power entry module.
The 1 amp fuse and a spare are located in a drawer in this
module.
• Insert the Tapper power cord into the rear of the Tapper
Connector housing (see Figure 2) and verify the connector is
locked in place. The Performer III is now ready for use.
Tapper Power Cord & Tapper Connector Housing
Main Power Cord
Power Entry
Module
Figure 2
Page 4
Page 5
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
ASSEMBLING THE SIEVE STACK
Assemble the sieve stack by placing the desired sieves on the
collection pan. Start with the nest mesh sieves, placing progressively coarser sieves on top (see Figures 4 and 5). A maximum of seven metal-framed sieves or fourteen acrylic sieves
may be stacked on top of the pan. Use of adapters and spacers
will limit the number of sieves that can be used. The sieve cover
must be placed over the top sieve before placing the stack assembly in the Performer lll. The larger diameter side of the cover
is for acrylic sieves, and the smaller diameter for metal sieves.
Check for the best t and install.
Hold in the thumb buttons to adjust the clamping knobs on
the upright rods to the height that will allow the sieve stack to
slide into place. Pushing in on the integral buttons allows the
threaded clamp knobs to slide freely. Position the stack using
the rubber bumpers on the back of the sieve base and tapper
clamping assembly, then tighten the knobs. To remove the sieve
stack, hold the thumb buttons in and slide the top up slightly.
The clamping assembly will remain in this position, ready for the
next sieve stack. This is a useful feature when using the same size
stack repetitively.
OPERATING INSTRUCTIONS
To begin testing, set the Mode Switch to MANUAL (see Figure 6).
Adjust the Amplitude Control for optimum performance with the
material being tested. Due to varying sample characteristics, the
user should observe the minimum vibration level and time period
necessary to sift particles without breaking them down. It is possible
for excessive vibratory action to degrade the sample particles by
abrasion.
To change the time setting on the Timer, press the <UP ARROW>
key or the <DOWN ARROW> key (see Figure 5). The rst digit
on the right hand side should start to ash in half second intervals. Press the <UP ARROW> key or the <DOWN ARROW> key to
adjust the digit to desired value. To enter the current digit and
move to the next digit press the <START> key. Once the last digit
on the left is entered the timer is ready to start.
To get into the adjust mode state, press and hold both the <UP
& DOWN ARROW> keys at the same time until the current mode
is displayed. When in this state the display will show the current
mode letters, H for hours, M for minutes, and S for seconds. Once
the mode letters are being displayed, press the <UP> key, or the
<DOWN> key, to change between modes. Press the <START>
key to accept new mode.
Mode Letter:
A-MMSS B-HHMM C-SSSS D-MMMM
Sieve Cover
Metal Sieves
&
Acrylic Spacers
(7 Total)
Sieve Cover
Acrylic Sieves
(14 Sieves Max.)
Acrylic Adapter
Collection Pan
Figure 3Figure 4
Collection Pan
Page 5
Page 6
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
SPECIFICATIONS
To run press the <START> key.
Overall Size:8x11x22in (203x279x559mm)
Once running, pressing the <START> key again will pause the
timer with the current amount of time remaining on screen.
When allowed to time out the timer will display DONE, press
<ANY> key to continue.
The Setting and the Mode values are saved automatically and
restored on power up.
Vibratory Frequency: 3,600vpm
Tapping Frequency: 60 Taps/min.
Designed Particle Sizing Range:
4.75mm to 75µm (No.4 to No.200 Sieve)
Extended Particle Sizing Range:
9.5mm to 25µm (3/8in to No.500 Sieve)
Power Requirements*: 115V, 50/60Hz, 2A
*Contact Gilson Technical Support for operation of this device
on other power supplies.
Figure 5
Performer III Timer
Page 6
Page 7
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
PARTS DIAGRAMS
Mode Switch
Tapper Assembly
Timer
Sieve Base
Amplitude Control
Figure 6
Performer III Control Panel
Captive Knobs with
Release Buttons
Sieve Base Spacers
Tapper/Clamping
Power Cord
Sieve Base
1/4-20 x 3/8in
Panhead Screws
Assembly
Sieve Cover
Figure 7
Performer III Components & Accessories
Page 7
Base
Acrylic Sieve
Adapter
Page 8
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3
ACCESSORIES
Acrylic Frame Sieves with Stainless Steel Mesh
Acrylic Frame Precision Sieves with Electroformed Mesh
GILSONIC AUTOSIEVER ACRYLIC FRAME SIEVES
Sieves for the GA-6 GilSonic AutoSiever have clear acrylic frames and
are 3in (76mm) in diameter. They are available tted with conventional
ASTM E 11 stainless steel Woven-Wire cloth or ASTM E 161 Precision
Electroformed nickel mesh.
The accurac y, eciency, and size ra nge of Precision Electro formed Sieves
often make the m a better solution for p recision particle sizi ng operations.
Opening tolerances of Electroformed Sieves are ±2μm, while ASTM E 11
woven-wire tolerances are two to ten times higher in comparable sizes.
Opening size s are available to 5μm, considera bly below the 20μm smalle st
available woven-wire size, and a number of sizes are equivalent to E 11.
When calibrated with glass beads or other means, electroformed sieves
can serve as a reliable reference standard.
The GilSonic AutoSiever holds seven Woven-Wire or three Precision
Electroformed Sieves in the xed-height Stack Assembly. Each sieve
is supplied with a Certicate of Compliance to the appropriate ASTM
Standard. Clear Acrylic Spacers are available if fewer sieves are used
in the stack.
Page 8
Page 9
Gilson Company, Inc. Performer III 3in Sieve Shaker: SS-3