Gilson P10M, P20M, P100M, P200M, P300M User Manual

...
The electronic pipette with guaranteed performance in both standard and repetitive pipetting.
USER'S GUIDE
EN
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
1 INTRODUCTION
Congratulations on the purchase of your PIPETMAN® M, the electronic pipette with guaranteed performance in both standard and repetitive pipetting modes.
PIPETMAN M is a motorized member of the PIPETMAN® family of pipettes from Gilson, requiring virtually no pipetting forces to aspirate and dispense samples. Combining PIPETMAN’s renowned accuracy, precision and robustness with user-friendly functions, PIPETMAN M will reduce pipetting fatigue and increase pipetting efficiency.
PIPETMAN M is a unique pipette with only two buttons to operate the pipette and reach all menu options:
• Push-button: set the volume by turning the button left or right and pipette with a simple push.
• Mode-button: access additional features like repetitive mode, mix or maintenance functions.
Only two buttons are needed to perform your routine pipetting tasks.
PIPETMAN M features the classic PIPETMAN look with additional functionality:
• Intuitive interface with five pipetting modes for a large number of applications:
standard pipet, repetitive, mix, reverse and custom mode.
Minimal effort: aspirate and dispense with one click of the push-button and
eject tips with ease to help reduce the risk of repetitive strain injuries (RSI).
• Maximum comfort: the lightweight and balanced design of PIPETMAN M offers an ergonomic design that rests comfortably in your hand as you pipette.
Minimized pipetting variability: the motorized piston allows for high reproducibility and accuracy while pipetting.
• Increased performance for all day pipetting comfort.
PIPETMAN M is available in 20 models covering a range from 0.5 µL
to 10 mL in single channel and 0.5 µL to 1200 µL in multichannel:
Page
1  INTRODUCTION 3
2  PARTS CHECK LIST AND ACCESSORIES 4
3  GETTING STARTED 4
4  DESCRIPTION 5
5  PIPETTING JUST LIKE WITH PIPETMAN 6
6  PIPETTING MODES 6
7  PIPETTING SPEED CONTROL 12
8  PIPETMAN TIPS 12
9  GUIDELINES FOR GOOD PIPETTING 14
10  PERSONALIZE YOUR PIPETMAN M 15
11  GOOD LABORATORY PRACTICE FEATURES 16
12  PIPETMAN M POWER MANAGEMENT 17
13  CONFIGURATION 18
14  MAINTENANCE OF YOUR PIPETMAN M 19
15  CLEANING AND DECONTAMINATION 24
16  LEAK TEST 25
17  TROUBLESHOOTING 27
18  SPECIFICATIONS 27
19  SAFETY PRECAUTIONS AND LIMITATIONS OF USE 29
20  REPLACEMENT PARTS 30
APPENDICES 31
EC DECLARATION OF CONFORMITY 52
WARRANTY 54
Mode-button
Push-button
Model Reference
Range (PIPET mode)
Color code
P10M F81022 0.5 µL to 10 µL Red
P20M F81023 2 µL to 20 µL Pale yellow
P100M F81013 5 µL to 100 µL Salmon
P200M F81024 20 µL to 200 µL Yellow
P300M F81014 20 µL to 300 µL Green
P1200M F81015 100 µL to 1200 µL Blue
P5000M F81016 500 µL to 5000 µL Purple
P10mLM F81017 1 µL to 10 mL Pale blue
Model Reference
Range (PIPET mode)
Color code
P8x10M F81025 0.5 µL to 10 µL Red
P12x10M F81026 0.5 µL to 10 µL Red
P8x20M F81027 1 µL to 20 µL Pale yellow
P12x20M F81028 1 µL to 20 µL Pale yellow
P8x100M F81018 10 µL to 100 µL Salmon
P12x100M F81019 10 µL to 100 µL Salmon
P8x200M F81029 20 µL to 200 µL Yellow
P12x200M F81030 20 µL to 200 µL Yellow
P8x300M F81031 10 µL to 300 µL Green
P12x300M F81032 10 µL to 300 µL Green
P8x1200M F81020 50 µL to 1200 µL Blue
P12x1200M F81021 50 µL to 1200 µL B lue
Contents
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
2 PARTS CHECK LIST AND ACCESSORIES
Contents:
Take a moment to verify that the following items are in the box:
• PIPETMAN M pipette
• Power Supply 5V with AC adapter and cable
• 4 Battery side tags
• 4 Ejector side tags
• Tip-ejector extension (P10M model only)
• Quick Guide
• Safety bag
• Certificate of conformity (including bar-code sticker)
• 2-year warranty
• Lubricant (except for P10M, and the multichannel models)
3 GETTING STARTED
Switch on your PIPETMAN M:
Press the push-button to activate the pipette. The start-up screen appears:
The pipette will do a self calibration test, the firmware resets the piston.
During initialization, displayed on the screen, you will see the volume range, version of firmware and the number of readjustments.
Your PIPETMAN M is preset on PIPET mode and is now ready to pipette (see Chapter 5).
PIPETMAN M goes into sleep mode after 3 minutes of inactivity. The display will shut off. Just press the push-button to reactivate the pipette, your last setting and pipetting mode will appear on the screen.
4 DESCRIPTION
PIPETMAN M Single and Multichannel models
Upper part
Push-button and volume adjustment knobMode-button: direct access to all menu
functions
Tip-ejector buttonDisplay
Elastomer plug to cover the battery charging
port
Identity-tag clipIdentity-tag window
Connecting nut – attaches handle to lower part
Lower part
Tip-ejector: removable to access tip-holderOptimized tip-holder to reduce tip tting
and ejection forces, removable for cleaning and servicing
Ejector clipEjector supportCoverEjector spacer
Screen
Pipetting ModeAspirate and dispense indicatorBattery statusPurge indicatorVolume
Reset
To reset your pipette, simultaneously press on the push- and mode- button for at least 5 seconds.
Switch off
To switch off your PIPETMAN M, press the push­button for at least 5 sec.
Accessories:
The following items are sold separately:
• POWER CARROUSEL: 5-position charging carrousel for single and multichannel models FB1001
• Battery side tags (set of 4)
F807013
• Ejector side tags (set of 4)
F807014
Please contact your Gilson Distributor for these accessories.
 
Volume range
Version of firmware
Number of readjustments
Display
The PIPETMAN M display is an OLED screen. It shows the current mode and operation step, pipetting
volume, battery indicator and piston status (aspirated and dispensed volume) in real time.
Warning messages appear in the place of volume.
PIPETMAN M is provided with a minimal battery charge. Before using your new PIPETMAN M, we strongly recommend you to fully charge the battery. PIPETMAN M charges 80 % of its full battery capacity in less than an hour, it takes three hours to fully charge the battery. Please refer to PIPETMANM Power Management (Chapter 12) for charging procedure.
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
• Press the push-button to aspirate the selected volume.
• To dispense: Click (press and release) the push-button. Three things happen: sample
dispense, automatic purge and piston reset to zero. or Keep your thumb pressing the push-button until the end of the dispense
cycle: The piston reset takes place after releasing the push-button again.
This feature enables you to remove the tip from the liquid, without
aspir
ating anything.
• Eject tip by pressing the ejection button.
Your PIPETMAN M is ready again for next pipetting cycle while maintaining the volume and the last settings!
Continue pipetting as you would with your mechanical PIPETMAN; it’s as simple as that to use your new, electronic pipette!
REPE
TITIVE Mode
The REPETITIVE Mode allows you to distribute the same volume repeatedly in pre­defined number (N) of equal aliquots. You supply the aliquot volume (AV) and the pipett
e automatically calculates the number of aliquots possible from the nominal
(maximum) volume (NV) of the pipette as follows:
N = NV/AV (e.g. 120 L x 10 for P1200M).
You also can decrease the number of repetitions (e.g. 120 L x 10-(n ≥1) for P1200M).
Application note: The REPETITIVE Mode is ideal for dispensing aliquots, e.g.; dispensing a PCR master mix into PCR
tubes or 96 plates, pipetting elution buer for DNA extraction, preparing a target for spectrometric analyses, distributing loading buer into samples, etc.
Press Mode-button to access system menu, where you can choose between
different pipette modes.
Select REPETITIVE Mode by turning the push-button. To confirm your selection click on the push-button.
• Turn the push-button half way, the screen will display: “click to change vol.”
• Press the push-button; the volume on the display will start to blink, you can now
adjust the volume:
• Set the aliquot volume: click the push-button, set the volume by turning the push-button and click
again to confirm your settings.
• Set the aliquot number: The maximal number of aliquots has been calculated by your PIPETMAN
M. The number of aliquots flashes and you can decrease this number (N – (n ≥1)) by turning the
push-button. Click again to confirm your settings.
You are ready to pipette!
1. Fit a tip suitable for the model of PIPETMAN M that you
are using (preferably use PIPETMAN TIPS for assurance of accurate and precise results; refer to chapter 8).
5 PIPETTING JUST LIKE WITH PIPETMAN
Using PIPETMAN M is as easy as using a mechanical PIPETMAN. Your PIPETMAN M is set by default on “PIPET Mode”, that means forward pipetting. You can select a volume and start to pipette.
Switch on
Press push-button.
Adjust the volume setting
• Hold your PIPETMAN M in pipette position.
• Turn the push-button half way, the screen will display: “click to change vol.”
• Now press the push-button; the volume on the display will start blinking, you can now adjust the volume:
• Turn the push-button either clockwise to decrease volume or counter clockwise to increase volume, like indicated on the button.
• Press the push-button one time when finished. The volume is locked.
You are ready to pipette!
6 PIPETTING MODES
PIPETMAN M offers more pipette modes for a large number of applications: you will find all pipetting modes in the menu (for system settings refer to the Chapters 7 and 13). Access the menu by pushing the Mode-button. To choose one of the pipetting modes, turn the push-button and confirm your
selection with a simple click!
PIPET Mode
This is the classic pipette mode (forward pipetting), your pipette is simply aspirating and dispensing.
Application note: You can use the PIPET Mode for all standard applications like DNA extraction, Plasmid Isolation,
Cloning, Dilution, PCR, qPCR and many others. You can easily pipette aqueous liquids like buer, chemical solutions (MgCl2, KCl …) and biological samples like blood, DNA and RNA.
• Fit a tip suitable for the model of PIPETMAN M that you are using (preferably use PIPETMAN TIPS for assurance of accurate and precise results; refer to Chapter 8).
START button
Optional STOP
0
V
Purge
You must always t a tip before using any pipette!
The tip can be emptied at any time during a pipetting cycle. To do so, turn the push-button quickly: the message “click to abort” will appear on the screen. Click on the push-button to validate: the liquid will be dispensed and an automatic purge followed by piston reset to zero will occur.
You must always t a tip before using any pipette!
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
2. Click push-button to aspirate total
volume. The volume aspirated will be a little bit more than required (extra volume). The extra volume is necessary to ensure equal operating conditions for each dispensed aliquot.
3. To dispense: click the push-button.
DISCARD appears on the screen. Dis­card the extra volume and continue dispensing the aliquots
. PIPETMAN M distributes the aliquot volume each time you click the push-button. The number of aliquots left to dispense is displayed on the screen.
A warning beep indicates the last aliquot!
4. To purge: Press the push-button, PURGE appears on the screen, click again to purge and reset the piston. Keep your thumb pressing the push-button until the end of the purge: the piston reset takes place after you release the push-button again. This feature enables you to remove the tip from the liquid, without aspirating anything.
5. Eject tip by pressing the ejection button.
MIX Mode
This is the classic pipette mode (forward pipetting) followed by a mixing phase, composed of repeatedly aspirating and dispensing, as well as an optional forward pipetting step.
Application note: The Mode MIX can be used to prepare a PCR master mix, enzyme restriction mix, protein solution, oligo-
nucleotide dilution and to mix samples with gel loading buer and so on…you can mix two dierent solutions or homogenize one solution. If you work with higher viscosity than water (e.g. restriction enzyme) you may change the aspiration speed (refer to chapter 7). If you work with genomic DNA you should always pipette very carefully to avoid shearing and nicking.
• Press Mode-button to access menu, where you can choose between different pipette modes. Select MIX Mode by turning the push-button, to confirm your selection click on the push-button.
• Set the first volume, named VOLUME 1: click the push-button, set the volume by turning the push­button and click to confirm your selection. This is the first volume you will aspirate and dispense. Set the second volume, named VOLUME 2: click the push-button, you can now set the second volume by turning the push-button. Click again to confirm your settings.
 If you set VOLUME 1 or VOLUME 2= 0 µL, the step will not appear. If you set VOLUME 1 and 2 = 0 µL the cycle will
start directly with the MIX step.
• Set the MIX volume: click the push-button, set the volume by turning the push-button. Confirm your settings with a simple click. This is the volume which will be aspirated and dispensed repeat­edly. “MIX” as long as you press the push-button.
• Set the additional pipetting volume, named VOLUME 3: You can choose a volume to be aspirated and dispensed after the mixing step, so you can continue routine pipetting.
START button Optional STOP
V (+ extra vol)*
Purge
Start Purge
V (vol)
0
DISCARD
ALIQUOTES
PURGE
Click to Purge
Discard
Click the push-button, set the volume by turning the push-button. Confirm your settings by a
simple click.
If you set the VOLUME 3 = 0 µL the cycle will stop after the MIX step.
You are ready to mix!
1. Fit a tip suitable for the model of PIPETMAN M that you are using (preferably use PIPETMAN Tips for assurance of accurate and precise results; refer to chapter 8).
2. To aspirate VOLUME 1 press the push-button
3. To dispense:
Click (press and release) the push-button: Three things happen: sample dispense, automatic
purge and piston reset to zero
or Keep your thumb pressing the push-button until the end of the dispense cycle: The piston reset
takes place after you release the push-button again. This feature enables you to remove the tip from the liquid, without aspirating anything.
4. To aspirate and dispense the second volume “VOLUME 2”, repeat steps 2 and 3.
5. To mix: Press the push-button; as long as you keep your thumb pressing the button, the pipette continues mixing. Release the push-button to complete the current mixing cycle.
6. To purge: press push-button, PURGE appears on the screen, click again to purge and reset the piston.
7. To pipette: If you have chosen a VOLUME 3 > 0 you can now aspirate and dispense this volume by repeating steps 2 and 3.
If you choose VOLUME1 = 0, step 2 won’t occur, if you choose VOLUME 2 = 0 µL, step 4 won’t occur, as well if you set
VOLUME 1 and 2 = 0 µL, step 1-4 will disappear. If you set VOLUME 3 = 0 µL step 7 will not occur.
You must always t a tip before using any pipette!
START button Optional STOP
0
V
Purge
Start Purge
Push-button pressed
V1
V2
V3
- 11 -- 10 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
REVERSE Mode
The REVERSE Mode is reverse pipetting with a manual pipette. During aspiration, additional liquid is added. After delivery, the excess volume remains in the tip and is discarded.
Application note: The REVERSE Mode is ideal for pipetting viscous liquids; liquids with high vapour pressure or those
which tend to foam. Your PIPETMAN M aspirates a selected volume and an excess. This excess compensates for the liquid that remains as a lm inside the tip during dispensing. For example: protein extraction, cell lysis, plasmid isolation, cell culture, buer preparation and many others…
Electrophoresis gels loading protocol: rst change the pipetting speed of your pipette (see chapter 7). Aspiration can
be done by using the standard speed (by default speed 6). Dispensing should be done very slowly and carefully, to prevent swirling and spilling of the samples, so select the lowest speed 1. Your pipette is now ready to load your gel! Please be aware that if you purge without pulling out the tip air bubbles could get into your gel!
• Press Mode-button to access system menu, where you can choose between different pipette modes. Select REVERSE Mode by turning the push-button, to confirm your selection click the push-button.
• Set the volume: click the push-button, set the volume by turning the push-button and click again to confirm your settings.
You are ready to do reverse pipetting!
1. Fit a tip suitable for the model of PIPETMAN M that you
are using (preferably use PIPETMAN TIPS for assurance of accurate and precise results; refer to Chapter 8).
2. Click push-button to aspirate the selected volume. An
amount of liquid equal to the amount of purged air is added, the volume aspirated will be a little bit more than the set volume.
3. Click push-button to dispense the volume, the addition
-
ally amount of liquid remains in the tip.
4.
To purge: press push-button, PURGE appears on the
screen, click again to purge and reset the piston. Keep your thumb pressing the push-button until the end of the purge: the piston reset takes place after you release the push-button again. This feature enables you to remove the tip from the liquid, without aspirating anything.
5. Eject tip by pressing the ejection button.
CUSTOM Mode
The CUSTOM Mode enables personalized pipetting protocol creation on a computer for transfer to PIPETMAN M through PIPETMAN M programmable Software, an easy-to-use software. T
o use this mode, the software has to be installed on the computer (file downloadable on
www.gilson.com/pipetmanm).
To install PIPETMAN M programmable software
1. Download the appropriate PIPETMAN M programmable software installation file for your operating system to the computer.
2.
Double-click the installer and follow the instructions on the screen to complete the installation.
3. When completed, the PIPETMAN M software icon will appear on desktop of the computer if Create Desktop Shortcut is checked.
4. If PIPETMAN M software is not launched automatically double click the PIPETMAN M icon on the Desktop.
5. Once installed, the PIPETMAN M software can be run by any user on the installed computer.
You can then create, edit, import or export custom protocols easily and rapidly.
To create a protocol, choose among the available tasks in PIPETMAN M programmable software:
Aspirate: aspirates specified volume into the tip
Dispense: dispenses specified volume from the tip
Dispense all: dispenses all remaining volume from the tip followed by a purge action
Repeat: enables repetition of any other tasks once or more
Mix: mixes the specific volume in one or more cycles
Purge: dispenses all remaining volume from the tip followed by purging extra volume
Wait: introduces a wait for a specified number of seconds
Wait for click: pauses the protocol until Push-button is clicked
Beep: makes pipette produce a beep sound while selected tasks are executed
In this mode, the protocols can be automatically executed, avoiding repetitive pushes on the button, helping to decrease risk or RSI.
To transfer a protocol to your PIPETMAN M pipette, click the “Transfer Custom Protocols to Pipette” button and follow the instructions on the screen. The custom mode is available in pipettes from firmware version V2.02. For more information on PIPETMAN M software, see p34.
0
Purge
START button Optional STOP
Start Purge
V (+ extra vol)*
Click to Purge
You must always t a tip before using any pipette!
- 13 -- 12 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
7 PIPETTING SPEED CONTROL
PIPETMAN M is set by default on speed 6 (maximum speed). You may need to change the speed of aspiration or dispensing depending on your application. You can change aspiration speed and dis­pensing speed independently of one other: from very slow to very fast (speed 1 – speed 6). Your speed selection is memorized for each pipetting mode until you change it again.
• Press Mode-button to access system menu, where you can find the different pipette modes, the speed menu and configuration menu.
• Select SPEED by turning the push-button, to confirm your selection click on the push-button.
• Set ASPIRATION SPEED: aspiration speed is blinking, set the speed by turning the push-button and click to confirm your selection.
• Set DISPENSING SPEED: dispensing speed is blinking, set the speed by turning the push-button and click to confirm your selection.
• After selecting the speed you go automatically back to pipetting mode.
8 PIPETMAN TIPS
PIPETMAN TIPS are made to the highest specifications; strict quality control is maintained throughout the manufacturing process. These tips are used to calibrate PIPETMAN M, therefore for optimum performance we recommend using PIPETMAN TIPS with your PIPETMAN M.
However, PIPETMAN M also offers you a high compatibility with a large number of other tips. For more information please feel free to contact your Gilson Distributor.
PIPETMAN TIPS have the Gilson logo engraved on their collar, ensuring that you have a genuine Gilson product. To ensure accuracy and precision, Gil­son’s Quality Assurance System focuses on the following critical parameters.
• PIPETMAN TIPS are made from pure polypropylene (virgin, metal-free, to avoid the possibility of contamination). They are available sterilized and with filters.
• Sterilized PIPETMAN TIPS are certified free of detectable RNases, DNases, DNA, RNA, and proteases.
• Optimized shape (revised collar for optimum sealing, thin walls, and fine point), making them easier to mount, more flexible, with no vortexing, and improved precision.
• PIPETMAN TIPS are free from even microscopic defects, especially at the orifice. All surfaces are smooth and hydrophobic, thereby avoiding the excessive retention of liquids that causes poor ac
-
curacy and a lack of precision.
Mold and cavity references are marked on the collar, ensuring the traceability. For quality assurance
purposes batch numbers appear on all packages (bags and boxes).
• They form an air-tight seal with the tip-holder, preventing the leaks that cause poor accuracy and a lack of precision.
Gilson logo ID Number
• PIPETMAN TIPS (except filter tips) may be auto­claved at 121°C for 20 minutes at 0.1 MPa.
Do not autoclave PIPETMAN Filter Tips, the lter will be damaged.
To ensure the best performance from your Gilson pipette, you shoud ALWAYS use PIPETMAN Tips (in accordance with ISO8655) system, because PIPETMAN Tips were used to establish the specications.
*
A plastic adapter is required to eject D10 and DF10 tips (short tips). The adapter is supplied with P10M. No adapter is required when using DL10 and DFL10 tips (long tips).
Model Tips model
Range
Color code
(PIPET mode)
P8x10M D10* - DL10
0.5 µL to 10 µL Red
P12x10M DF10* - DFL10
P8x20M DL10 - D200
1 µL to 20 µL Pale yellow
P12x20M DFL10 - DF30
P8x100M D200
10 µL to 100 µL Salmon
P12x100M DF100
P8x200M D200 - D300
20 µL to 200 µL Yellow
P12x200M
DF100 - DF200 - DF300
P8x300M D200 - D300
10 µL to 300 µL Green
P12x300M DF200 - DF300
P8x1200M D1200
50 µL to 1200 µL Blue
P12x1200M DF1200
Model Tips model Range Color code (PIPET mode)
P10M
D10* - DL10
0.5 µL to 10 µL Red
DF10* - DFL10
P20M
D200
2 µL to 20 µL Pale yellow
DF30
P100M
D200
5 µL to 100 µL Salmon
DF100
P200M
D200 - D300
20 µL to 200 µL Yellow
DF200 - DF300
P300M
D200 - D300
20 µL to 300 µL Green
DF200 - DF300
P1200M
D1000 - D1200
100 µL to 1200 µL Blue
DF1000 - DF1200
P5000M D5000 500 µL to 5000 µL Purple
P10mLM D10 mL 1 µL to 10 mL Pale blue
Ejector spacer D10
Ejector spacer
Short tips
Long tips
Short tips Long tips
Plastic adapter P8x10M and P12x10M
- 15 -- 14 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
9 GUIDELINES FOR GOOD PIPETTING
Aspirate and dispense
1) Fit new PIPETMAN Tip for the best results (see Chapter 18). For single channel models, push the tip-holder into the tip
using a slight twisting motion to ensure a firm and airtight seal.
For multichannel models,PIPETMAN Tips are best fitted from
the patented ROCKY RACK available only in our TIPACKs and TOWERPACKs. ROCKY RACK is the dome-shaped part of the pack that contains the tips. ROCKY RACK makes it easy to securely fit the tips to a multichannel pipette, ensuring an airtight seal on all channels without the need to use undue pressure or to touch the tips.
2) Pre-rinse the tip. Pre-rinsing consists of aspirating the first volume of liquid and then dispensing it back into the same vessel (or to waste). Subsequent volumes that you pipette will have levels of accuracy and precision within specifica­tions. Some liquids (e.g. protein-containing solutions and or
ganic solvents) can leave a film of liquid on the inside the wall of the tip; pre-rinsing the tip minimizes any errors that may be related to this phenomenon.
3) Hold the pipette vertically and immerse the tip in the liquid (see page 13). Press the push-button to aspirate the set vol­ume of liquid. Wait a few seconds (time depends on model, see page 13); then withdr
aw the pipette-tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe, however if you do so take care to avoid touching the tip’s orifice.
4) Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°). Press the push-button. Wait for at least a few seconds before releasing the push-button to expel any residual liquid from the tip. While removing the pipette draw the tip along the inside surface of the vessel.
5) Eject the tip by pressing firmly on the tip-ejector button.
General guidelines for good pipetting
a. Make sure that you fit new tips.
b. Each new tip should be pre-rinsed with the liquid to be pipetted.
c. When aspirating, keep the tip at a constant depth below the
surface of the liquid (refer to Table).
Immersion Depth and Wait Time
Model Immersion Depth Wait Time (mm) (seconds)
P10M 1 1
P20M 2-3 1
P100M 2-4 1
P200M 2-4 1
P300M 2-4 1
P1200M 2-4 2-3
P5000M 3-6 4-5
P10 mLM 5-7 4-5
P8x10M, P12x10M 1 1
P8x20M, P12x20M 2-3 1
P8x100M, P12x100M 2-4 1
P8x200M, P12x200M 2-3 1
P8x300M, P12x300M 2-3 1
P8x1200M, P12x1200M 2-4 2-3
d. Change the tip before aspirating a different liquid, sample, or reagent.
e. Change the tip if a droplet remains at the end of the tip from the previous pipetting operation.
f. Liquid should never enter the tip holder. To prevent this:
• Never turn the pipette upside down,
• Never lay the pipette on its side when there is liquid in the tip(s),
g. The Gilson Stand Adapter (see Chapter 2) is recommended for use with the CARROUSEL and SINGLE
Pipette Holder to store your PIPETMAN M pipette in the vertical position. Alternatively the POWER CARROUSEL can be used to store and charge up to 5 PIPETMAN M pipettes.
h. When pipetting liquids with temperatures different to the ambient temperature, pre-rinse the tip
several times before use in order to reach equilibrium between the temperatures of the liquid and the pipette’s dead-volume.
i. For volatile liquids you should saturate the dead-volume by aspirating and dispensing the liquid
repeatedly before aspirating the sample.
j. After pipetting acids or other corrosive liquids that emit vapors, clean the pipette, as described in
Chapter 15.
k. The pipette can be used between + 4 °C and + 40 °C, but the
specifications may vary (see Chapter 18).
l. Do not pipette liquids having temperatures above 50 °C or
below 4 °C.
10 PERSONALIZE YOUR PIPETMAN M
Tip-ejector extension for use with P10M
In order to eject D10 tips, tip-ejector extensions are supplied with P10M pipettes.
The tip-ejector extension which is made of PVDF (polyvinylidene Fluoride), is autoclavable.
To t a tip-ejector extension:
• Hold the pipette with the push-button upright.
• Hold the extension with the slot upright.
• Slide the extension over the tip-holder.
• Push the extension firmly onto the end of the tip-ejector until it clicks into place (see opposite).
To remove a tip-ejector extension (see opposite):
• Hold the pipette in one hand and grip the extension with the other.
• Gently twist the extension (either direction) and pull it away from the pipette.
Short tips Long tips
To fit To remove
Slot
Extreme temperatures can aect accuracy and precision!
- 17 -- 16 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
Personal label – Name and/or application tag:
Your PIPETMAN M offers you two options to mark your pipette. You can personalize your pipette with a name tag and/or an application tag:
Window-tag:
Remove the window by inserting a small screwdriver in the access slot.
Position the name tag on the pipette
Clip the window back in place.
Clip-tag:
Pull off the clip.
Position the name tag into the clip
Clamp the clip back in place.
11 GOOD LABORATORY PRACTICE FEATURES
Your PIPETMAN M is fully compliant to ISO8655 standard and is CE marked (for IVD and EMC directives). PIPETMAN M includes the following GLP features.
General
• Lockable volume.
• Useful volume-range is printed on the push-button.
• Volume range is displayed on the screen.
• Serial number is engraved on the body (handle) of the pipette and encoded in the firmware.
• Bar code: on the box and with the certificate (can be transferred).
• Personalization by nametags (for marking application and/or user name), see Chapter 10.
Cycle counters
• From last volume setting (to count the number of cycles in the current “run”).
• From manufacture and since the last service (see Chapter 14).
• By indicators (R1, R2, R3 … Rx) each time pipette is readjusted.
• Maintenance intervals can be specified by weeks or number of cycles elapsed.
Alarms
• Low battery warning.
• Service is overdue warning (see Chapter 14).
• Service Alert can be set “On” or “Off”.
Window-tag
Clip-tag
12 PIPETMAN M POWER MANAGEMENT
The PIPE TMAN M battery has been designed to ensure 900 pipetting cycles (750 for PIPETMAN M Multi 200 µL and 300 µL) at maximum speed (up to nine 96 well plates) between two charge cycles. For service continuity when needing more capacity, your PIPETMAN M can also be used while charging with the same performance as a fully charged pipette!
Battery charging
PIPETMAN M charges 80 % of its full battery capac­ity in less than an hour, it takes three hours to fully char
ge the battery.
Charge your pipette by using the power supply:
- Connect the AC adapter to a suitable AC power supply
- Plug the adapter into the USB port of the pipette.
- The pipette starts charging.
Charge your pipette by using stand adapter
(see Chapter 2):
- First place the stand adapter on your Gilson
SINGLE pipette holder or CARROUSEL.
- Remove the plug. Now place your pipette
on the stand adapter and plug the power transformer into the stand adapter.
- The pipette starts charging.
Battery charging of PIPETMAN M using the
POWER CARROUSEL:
- First install the POWER CARROUSEL as described above.
-
Remove the plug. Now place your pipette
on the charging position, please ensure that the pipette is properly fitted to the contacts ont the top of the carrousel (a proper fitting will trigger a beep sound).
- The pipette starts charging.
Stand adapter
Plug
USB port
- 19 -- 18 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
Working with charger connected to the pipette
PIPETMAN M can also be used for pipetting while charging.
- Plug the power supply into your pipette.
- Press the push-button to start your pipette.
- PIPETMAN M is ready to pipette.
You can also recharge PIPETMAN M by USB with your com-
puter; however it will take much longer than by power supply.
Low battery warning
PIPETMAN M has a low battery alert. As the battery runs down, the LOW BATTERY message appears.The warning is replaced by a blinking symbol of an empty battery. If it is not recharged, the pipette will switch off after a while. Settings will not be lost.
If the battery is completely discharged (= black screen, no operation) please connect the pipette to the charger for at least five minutes, then a “batt. too low ” message will appear. The pipette will display the last settings used a few minutes later and will continue charging.
Changing the battery
If the following symptoms occur, please contact your Gilson Service Center for battery change:
 PIPETMAN M battery won’t charge or will not
operate even if connected to the charger at least for 10 minutes.
 PIPETMAN M battery provides you very short
cycle time or permanently needs to be used in connected mode.
13 CONFIGURATION
Enter the menu using the Mode-button. You will find not only the pipetting programs (PIPET, REPETITIVE, MIX, REVERSE and CUSTOM) and the speed control, but also the menu “CONFIGURATION”, including “SERVICE” and “ADJUSTMENT” (refer to Chapter 14).
The navigation of the menus is always done in the same way Make your selection by rotating the push-button, enter and confirm with a simple click!
Never work with PIPETMAN M while charging by computer!
PIPETMAN M uses Lithium-Ion battery. Operation on the battery is done at user’s risk only. Opening the pipette voids the warranty!
Use only the original power transformer supplied by Gilson. Use of an incompatible power supply can damage your PIPETMAN M!
Mode-button
Push-button
The configuration menu allows you to set up the following items:
1. Volume limit: Here you can fix the maximal volume of the pipette (by default = nominal volume).
For example you have a P200M pipette, but you would like to use it with your PIPETMAN DIAMOND Filter tips DF100, you can set the maximal volume at 100 µL and you won’t risk contamination of your pipette.
2. Beeper: You can switch the tone of your pipette on or off.
3. Contrast: You can choose between values of 1 to 5 to set up the contrast of the screen.
4. Service see Chapter 14.
5. Adjustment see Chapter 14.
6. Language Available from the firmware version 1.20. Language available are English, Chinese and Japanese. To change the language, enter in the Configuration menu:
- Select your language by turning the push-button,
- To confirm your selection, click on the push-button,
- The language selected is blinking,
- To confirm your selection, click 2 seconds on the mode button.
14 MAINTENANCE OF YOUR PIPETMAN M
PIPETMAN M requires very little maintenance. However, to ensure pipette accuracy, precision and robustness for timeless use please proceed periodically with a Two-minute Inspection as recom­mended by Gilson.
Your PIPETMAN M allows you to:
• Get service information.
• Readjust the pipette to user settings.
• Return to factory settings.
• Replace spare parts (for detailed information see Chapters 14 and 19).
• Prepare the pipette for cleaning or autoclaving by “disassembly” of the par ts specified (see Chapter 15).
Service information
PIPETMAN M provides you with all required service information. It will help you establish an easy diagnosis and plan any service operation with your accredited Gilson Service Center.
You will find in the “service” menu following sub menus:
a. Disassembly b. Service Info c. Service Settings d. Pipette Info
- 21 -- 20 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
The navigation of the menus is always done in the same way Make your selection by rotating the push-button, enter and confirm with a simple click!
a. Disassembly:
The Multichannel models should not be disassembled: only the push-button and the tip-ejector can be replaced.
Only the lower part of the Single models can be disassembled, the push-button, the connecting nut and the tip-ejector can be replaced.
This option allows you to disassemble the lower part safely (protection of the piston and the actuator). If you select DISASSEMBLY, the following screen appears:
Now, you can remove the lower part of your pipette with complete security.
b. Service info provides you with maintenance information; you cannot modify any of the items:
• Number of pipetting cycles since last maintenance.
• Number of weeks since last maintenance.
c. Service settings gives you the option to organize your maintenance, you can set following service points:
• Number of cycles until next maintenance.
• Number of weeks until next service.
• Service date.
• Activate or deactivate the Service alert.
Changes of service settings will reset all the service info to zero.
d. Pipette info provides information about your pipette, you cannot modify any of the items:
• Version of firmware.
• Serial number.
• Total number of cycles (one cycle: up and down of the piston).
• Number of readjustments carried out by this pipette.
Adjustment
Press the mode-button to enter the menu. Select “configuration” by rotating the push-button and click to confirm your selection. Enter the sub-menu in the same way and select the “adjustment” menu.
The adjustment menu allows you to access the following items:
• Standard Readjustment
• Reset Settings
Standard readjustment
This menu allows you to adjust the pipette using 3 calibration points: 10 %, 50 % and 100 % of nominal volume (in according with ISO8655 recommendations).
You may want to calibrate your pipette for solutions with a density, viscosity, surface tension or vapour pressure that are different to that of water. To return to the factory settings, choose “Reset Settings”, your PIPETMAN M will be reset to the three factory calibration values which are permanently stored in the pipette’s firmware.
In accordance with ISO 8655 Gilson recommends a gravimetric procedure for pipette calibration. This gravimetric method is used to establish the mean mass of a given volume of water (taking into account evaporation losses, where necessary). After converting the mean mass to a volume (using the Z factor, see appendix II), enter the measured volumes(s) into the pipette’s memor y and the software readjusts the pipette accordingly. This method requires the strict monitoring of environmental conditions and the use of routinely controlled equipment that is adapted to the volume being measured.
Conversion to volume must take into account the density of the liquid as well as evaporation during the cycle time. For each measurement, the corresponding volume (Vi) can be calculated as follows:
Wi is the weight as read on the balance e is the mean evaporation loss during the cycle time Z expressed in L/mg, is a conversion factor incorporating density of water buoyed in air, at test temperature and barometric pressure (see Appendix II)
For volumes greater than 20 L, the evaporation factor can be disregarded.
For more information please refer to “Verification Procedure for Accuracy and Precision” which you can download from the Gilson website (www.gilson.com).
Maintenance operations
(disassembly of your PIPETMAN M t
o change and clean parts)
It is best to inspect your pipette regularly and to routinely clean and change parts as required. To help you to keep up a regular schedule and in the interests of GLP, you can configure your pipette to display an alarm before servicing is due (refer to Service Settings, page 18).
Maintenance warning
PIPETMAN M notifies you, when maintenance is due (service alert is set by default to “ON”). The following message () will appear:
after validation by a click, the following message () appears:
If you click to confirm within 10 sec. the pipette will automatically switch to “disassembly” mode so you can carry out your maintenance safely.
Wait longer than 10 seconds and the option to have a reminder in one week will appear ().
Click to confirm. Your pipette will be operational again.
Maintenance operation for single channel models
Tip-Holder and Tip-ejector
These parts must be changed if they are damaged. You may also remove these parts for cleaning or decontamination purposes.
Vi = (Wi + e) Z
The following maintenance operation should only be done when the pipette is in Mode DISASSEMBLY (see page 18)!
- 23 -- 22 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
Changing the tip-ejector
1) Keep the tip-ejector button depressed and grip the top of the tip-ejector with the other hand.
2) Gently rotate the tip-ejector counter clockwise and separate its connector from the operating rod.
3) Pull the tip-ejector away from the body of the pipette.
4) Clean or autoclave the tip-ejector and refit or replace it by reversing the procedure.
Changing the tip-holder (lower part)
After removing the tip-ejector, you may remove the lower part of the tip-holder, which is more likely to become contaminated or damaged than the upper part. Removal of the lower part is shown below; for the upper part see “Removing the Piston” (special precautions are necessary).
1) Gently rotate the lower part of the tip holder counterclock
-
wise to unscrew it from the upper part.
2)
Separate the parts and remove the O-ring (see “Changing
O-ring”)
3) Clean and if required autoclave the lower part of the tip­holder (autoclaving: 20 min. at 121°C and 0.1MPa)
4) If required, lubricate the piston (see “How to Lubricate the Piston”) and fit a new O-ring.
5) Screw the two parts together, making sure that the two parts are fully tightened, by hand.
6) Refit the tip-ejector.
After autoclaving the tip-holder and tip-ejector may change color, however this has no impact on the performance.
Changing the seal and/or O-ring (see Chapter 20)
The O-ring is positioned on the piston; it should not be autoclaved, if worn or damaged in any way, it must be replaced.
To access the O-ring, remove the tip-ejector and unscrew the lower part of the tip-holder. You should now be able to remove the O-ring from the piston. Sometimes, the O-ring may be found in the recess at the top-end of the lower part of the tip-holder. If required, lubricate the piston (see “How to Lubricate the Piston”) then fit a new O-ring by sliding it onto the piston. Reassemble the pipette. Depending on the pipette model the dimensions of the O-ring will vary (refer to Chapter 20).
Servicing the piston
You may remove the piston-assembly to clean, lubricate, or change the piston.
O-ring
Piston
Tip-holder (lower)
Tip-holder (upper)
Removing the piston
1) Remove the tip-ejector and (optionally) the lower part of the tip-holder. If you remove the lower part, take care to remove the O-ring as described in “Changing the O-ring”.
2) Unscrew the connecting nut (turn by hand, coun­terclockwise).
3)
Gently remove the connecting nut and upper tip
holder.
4) Pull off the piston assembly from the body of the pipette – separate the parts (see below).
5) Clean and autoclave (if required) the piston and holder, together with any other parts that may need to be treated in the same way (see Chapter 15).
In the case of P1200M, the piston holder, piston and
guide are a unit. Don’t try to separate them!
6) Lubricate the piston, see below.
7) Reassemble the piston and guide; then carefully insert the assembly into the body of the pipette. The guide should hold the piston assembly inside the body of the pipette. You can hear a “click” when the piston is back in place.
Take care not to touch the piston, and that it is dust-free. If needed, clean the assembly with a dust-free cloth.
8) Reassemble the upper part of the tip-holder and the connecting nut, and then refit to the body of the pipette by rotating the connecting nut clockwise until it is finger tight. Fit the O-ring (for P10M, the seal) and reassemble the lower part of the tip-holder. Refit the tip-ejector.
The seals for P10M are fragile and can be used only once. So, after unscrewing the lower part of the tip­holder you must fit a new seal.
How to lubricate the piston (except P10M)
Use only Gilson lubricant (ref: F2070902, as supplied).
Squeeze a small quantity from the tube onto a clean, lintless cloth. Use the cloth to transfer the lubricant to the piston. Ensure that the piston is evenly lubricated and that you wipe away any excess – remember only a fine film of lubricant is required (over the entire piston).
Guide Connecting nut
The specications of the pipette must be checked after changing any part!
Don’t pull on the piston to remove the assembly!
Tip-holder (lower) Tip-holder (upper) Connecting nutPiston holderPiston
Guide
- 25 -- 24 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
Maintenance operation for multichannel models
For multichannel models, the lower part should not be disassembled: only the push-button, the connecting nut and the tip-ejector can be replaced.
Changing the tip-ejector
To remove the tip-ejector, keep both ejector locks depressed. Pull the tip-ejector down.
To refit the tip-ejector, gently re-insert the tip-ejec tor vertically into the rails of the ejector support. Pull lightly on the tip-ejector to check the position.
Replace the ejector spacer
1) Gently press the tabs from the ejector spacer and remove it from the tip-ejector.
2) Insert the ejector spacer and click it to the tip­ejector.
15 CLEANING AND DECONTAMINATION
PIPETMAN M is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated.
As part of your quality system, you may have procedures for pipette decontamination. We recommend that you refer to “Decontamination Procedures for Gilson Pi­pettes”, which you can download from Gilson’s website.
I
f you use chemical solutions for decontamination or detergents for cleaning, other than those speci­fied below, you should check with your supplier that the solution or detergent is safe for use with one of the f
ollowing materials: PP (Polypropylene), PBT (Polybutylene Terephthalate), PC (Polycarbonate),
POM (Polyoxymethylene), PVDF (Polyvinylidene Fluoride), PEI (Polyetherimide) and stainless steel.
Cleaning
The pipette must be cleaned, as described below, before it is decontaminated. For cleaning your PIPETMAN M you may use a simple soap solution or any of the solutions mentioned in “Decontamination Procedures for Gilson Pipettes”.
Single channel models
External
1) Remove the tip-ejector for cleaning (see Chapter 14).
Liquid must not enter the handle of the pipette!
Use alcohol (ethanol or isopropanol) to clean up the dirt and dust on the tip-holder!
2) Replace the elastomer plug to protect battery charging port.
3) Wipe the entire pipette with a soft-cloth or lint-free tissue impregnated with soap solution, to remove all dirty marks. If the pipette is very dirty, a brush with soft plastic bristles may be used.
4) To rinse, wipe the entire pipette with a soft-cloth or lint-free tissue impregnated with distilled water.
5) Leave to air dry.
Internal
The following components only can be immersed in a cleaning solution: tip-ejector, tip-holder, connecting nut and piston.
1) Disassemble the pipette as described in Chapter 14.
2) Set aside the handle in a dry and secure location.
3) Clean the individual components using an ultrasonic bath (20 minutes at 50 °C) or with a soft-cloth and brushes.
4) Rinse the individual components with distilled water.
5) Leave the parts to dry by evaporation or wipe them with a clean soft-cloth or lint-free tissue.
6) Lubricate the piston and reassemble the pipette according to the instructions given in Chapter 14.
Multichannel models
The following components only can be immersed in a cleaning solution: tip-ejector, ejector locks and ejector spacer.
1) Remove the tip-ejector and the ejector spacer (see Chapter 14).
2) Immerse the tip-ejector, ejector locks and ejector spacer in the cleaning solution or wipe them with a soft cloth or lint-free tissue impregnated with the cleaning solution.
3) Rinse the components with distilled water.
4) Leave the parts to dry by evaporation or wipe them with a clean soft-cloth or lint-free tissue.
5) Refit the tip-ejector.
Please note that although the lower part of PIPETMAN M Multichannel can withstand a few number of autoclaving cycles, we do not recommend autoclaving it.
Decontamination
Autoclaving - Single channel models
However, after separation from the body, any of the following components of the volumetric module may be autoclaved individually: tip-ejector, connecting nut, tip-holders, pistons and seals (except O-rings).
1) Clean the parts to be autoclaved, especially the tip-holders.
2) Put the parts in an autoclaving bag.
3) Autoclave for 20 minutes at 121°C and 0.1 MPa.
4) Check that the parts are dry before reassembling the pipette.
5) Set the pipette aside to stabilize at room tempera
-
ture for at least 6 hours.
The body (handle) of the pipette is not autoclavable!
The specications of the pipette must be checked after autoclaving/ disassembling!
- 27 -- 26 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
Chemical decontamination - Single and multichannel models
You may choose to decontaminate your pipette chemically, in accordance with your own procedures. Whatever decontaminant you use, check that it is compatible with the plastics used in the construc­tion of the pipette (see page 22).
Parts that may not be immersed
1) Wipe the handle of the pipette with a soft-cloth or lint-free tissue covered with the chosen decon­taminant.
2)
Wipe the handle of the pipette with a soft-cloth or lint-free tissue covered with distilled water.
Parts that may be immersed
The following components only can be immersed in a decontaminant solution:
• Single channel models ->tip-ejector, tip holder (both parts), connecting nut, piston (including
holder) and guide.
• Multichannel models -> tip-ejector, ejector locks and ejector spacer.
Please note that although the lower part of PIPETMAN M multichannel can withstand a few number of autoclaving cycles, we do not recommend autoclaving it.
1) Disassemble the pipette as described in Chapter 14.
2) Immerse the components in the decontaminant solution or wipe them according the instructions given by the manufacturer or supplier of the decontaminant.
3) Rinse the individual components with distilled or sterile water.
4) Leave the parts to dry by evaporation or wipe them with a clean, lint-free tissue or soft-cloth.
5) Lubricate the piston and reassemble the pipette according to the instructions given in this chapter.
16 LEAK TEST
This test may be performed at any time to check that the pipette does not leak, especially after performing a maintenance or decontamination procedure. If a pipette fails this test, you should replace the faulty part (e.g. O-ring, tip-holder...) and repeat this test, after making sure that the pipette is correctly reassembled.
• Fit a PIPETMAN DIAMOND Tip
• Set the pipette to the nominal volume
• Aspirate the nominal volume from a beaker of distilled water
• Hold the pipette in the vertical position and wait for 20 seconds
• If a water droplet appears at the end of the tip there is a leak (see Chapter 17)
• If you see no droplet, re-immerse the tip below the surface of water
• The water level inside the tip should remain constant; if the level goes down there is a leak (see
Chapter 17)
• For Multichannel models, check if the water level between each tip is the same
17 TROUBLESHOOTING
In case of malfunction, first reset the pipette by pressing on the push-button and the Mode-button simultaneously at least for 8 seconds.
If the problem persists, you may consult the fol­lowing table which identifies potential problems and their solutions
.
Problem Possible Cause Refer to page
Pipette is leaking sample. Worn O-ring. 22
Pipette won’t aspirate. Worn O-ring. 22 Unscrewed lower part of tip-holder. 22 Damaged or corroded piston. 23 Damaged tip-holder. 22 Improper repair or assembly. 21-24 Connecting nut is loose. 23 Software needs to be reset. 5
Noisy operation. Piston needs lubricating. 23
Pipette is inaccurate. Improper repair or assembly. 21-24 Unscrewed lower part of tip-holder. 22 Pipette is out of adjustment. 20 Connecting nut is loose. 23
Problem Possible Cause Refer to page
Pipette is not precise. Unscrewed lower part of tip-holder. 22 Incorrect operator technique. 14 Worn O-ring. 22 Connecting nut is loose. 23 Damaged or corroded piston. 23 Damaged tip-holder. 22
Tips fall or don’t t. Low quality tips. 12-13 Damaged tip-holder. 22 Damaged tip-ejector. 22-24 Ejector spacer is damaged. 24 Tip-ejector is loose. 22-24 Dirty tip-holder. 21
No OLED display. Pipette is in sleep mode. 4 Battery needs recharging. 12 Software needs to be reset. 5
No operation possible. Batter y needs recharging. 12 Software needs to be reset. 5
Calibration impossible. S oftware needs to be reset. 5
Before returning any pipette to your local Gilson Service Center, ensure that it is completely free of chemical, biological, or radioactive contamination. Please use the included safety bag.
18 SPECIFICATIONS
PIPETMAN M is a high quality pipette that offers excellent accuracy and precision; it is fully compliant with ISO 8655 and is CE marked.
The figures given in the “Gilson Maximum Permissible Errors” table were obtained using “PIPETMAN TIPS”. These figures are only guaranteed when using genuine PIPETMAN TIPS.
Each pipette is inspected and validated by qualified technicians according to the Gilson Quality System. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655 Standard, by type testing. The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6):
• Basis of adjustment, Ex.
• Reference temperature, 20 °C
• Relative humidity, 50 %
• Barometric pressure, 101 kPa
• Use of distilled water grade 3 (ISO 3696)
• Ten measurements for each test volume, which are Nominal Volume, 50 % of Nominal Volume and 10 % of Nominal Volume.
(tested mode: PIPET, speed 6 with PIPETMAN Tips)
12-13 22 22-24 24 22-24 21 4 12 5 12 5
- 29 -- 28 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
19 SAFETY PRECAUTIONS AND LIMITATIONS OF USE
For safety reasons, it is important to observe the following instructions:
1. Battery and electrical specifications:
Use AC-adaptor and stand adapter indoors. PIPETMAN M uses Lithium-Ion battery. Operation on the battery is done at user’s risk only. Opening
the pipette voids the warranty.
Dispose of used batteries in accordance with legal regulations. Batteries may not be disposed of
with household waste and may explode if disposed in fire !
• Li-ion battery pack; 1 Ah / 3.6 V
Charging time: approx. 3 hours for a fully discharged battery (80% in one hour).
• AC adaptor
Input voltage, country-specific: 100-240 V, 50/60 Hz (0.5A max.) Output voltage: +5 V DC, 3.5A (17.5W max.)
• Class II
this device is double insulated.
2. Do not use PIPETMAN M in a potentially explosive environment
or with potentially explosive chemicals.
3. When pipetting infectious, radioactive, toxic and other hazardous solutions,
please observe all the safety precautions (e.g. wear protective clothing, goggles and gloves) and
regulations appropriate for your country.
4. Do not allow the liquid to enter the body of the pipette.
5. Storage conditions
Temperature: -20°C to 50°C - Humidity max: 80%
6. Temperature of use
Between 4°C and 40°C (specifications may vary).
7. Use only genuine PIPETMAN Tips and original Gilson’s accessories and spare parts.
If the equipment is used in a manner not specified by Gilson in the user’s guide, the protection
provided in the equipment may be impaired.
8. Equipment disposal
This equipment must not be disposed of with unsorted municipal waste. Instead, it is your responsibility
to correctly dispose of your waste equipment by handing it over to an authorised facility for separate collection and recycling. It is also your responsibility to decontaminate the equip
­ment in case of biological, chemical, and/or radiological contamination so as to protect from health hazar
ds the persons involved in the disposal and recycling of equipment. For more information about where you can drop off your waste equipment for recycling, please contact your local dealer from whom you originally purchased the product or your local council. By doing so, you will help conserve natural resources and you will ensure that your waste equipment is recycled in a manner that protects human health and the environment. Thank you.
9. PIPETMAN M can be used indoor and outdoor,
if precautions are respected as described above in this handbook and in the GLP.
PIPETMAN M
model
Ordering reference
PIPETMAN DIAMOND
Tips compatibility
PIPET Mode REPETITIVE Mode Volume
Systematic error
(µL)
Random error
(µL)
Systematic error
(µL)
Random error
(µL)
P10M F81022
D10
DL10
DF10
DFL10
0.5-10 µL 0.5-10 µL
0.5 ± 0.040 ≤ 0.013 ± 0.12 ≤ 0.08
1 ± 0.025 ≤ 0.012 ± 0.12 ≤ 0.08
5 ± 0.060 ≤ 0.020 ± 0.12 ≤ 0.08
10 ± 0.080 ≤ 0.025 ± 0.12 ≤ 0.08
P20M F81023
D200
DF30
2-20 µL 2-20 µL
2 ± 0.075 ≤ 0.025 ± 0.2 ≤ 0.1
10 ± 0.100 ≤ 0.035 ± 0.2 ≤ 0.1
20 ± 0.150 ≤ 0.050 ± 0.2 ≤ 0.1
P100M F81013
D200
DF100
5-100 µL 5-100 µL
5 ± 0.35 ≤ 0.10 ± 0.8 ≤ 0.3
10 ± 0.30 ≤ 0.10 ± 0.8 ≤ 0.3
50 ± 0.38 ≤ 0.12 ± 0.8 ≤ 0.3
100 ± 0.40 ≤ 0.15 ± 0.8 ≤ 0.3
P200M F81024
D200
D300
DF200
DF300
20-200 µL 5-200 µL
20 ± 0.40 ≤ 0.15 ± 1.6 ≤ 0.6
100 ± 0.80 ≤ 0.22 ± 1.6 ≤ 0.6
200 ± 1.00 ≤ 0.26 ± 1.6 ≤ 0.6
P300M F81014
D200
D300
DF200
DF300
20-300 µL 10-300 µL
20 ± 0.80 ≤ 0.16 ± 4.0 ≤ 1.5
30 ± 0.70 ≤ 0.20 ± 4.0 ≤ 1.5
150 ± 0.90 ≤ 0.23 ± 4.0 ≤ 1.5
300 ± 1.05 ≤ 0.30 ± 4.0 ≤ 1.5
P1200M F81015
D1000 D1200
DF1000 DF1200
100-1200 µL 20-1200 µL
100 ± 2.5 ≤ 0.4 ± 16 ≤ 6.0
120 ± 2.4 ≤ 0.4 ± 16 ≤ 6.0
600 ± 3.6 ≤ 0.8 ± 16 ≤ 6.0
1200 ± 6.0 ≤ 1.2 ± 16 ≤ 6.0
P5000M F81016 D5000 500-5000 µL 100-5000 µL
500 ± 10 ≤ 2 ± 40 ≤ 15.0
2500 ± 15 ≤ 4 ± 40 ≤ 15.0
5000 ± 25 ≤ 7 ± 40 ≤ 15.0
P10 mL M F81017 D10 mL 1-10 mL 200 µL-10 mL
1 mL ± 25 ≤ 4 ± 60 ≤ 30.0
5 mL ± 30 ≤ 8 ± 60 ≤ 30.0
10 mL ± 50 ≤ 12 ± 60 ≤ 30.0
P8x10M
P12x10M
F81025
F81026
D10
DL10
DF10
DFL10
0.5-10 µL 0.5-10 µL
0.5 ± 0.05 ≤ 0.02 ± 0.24 ≤ 0.16
1 ± 0.04 ≤ 0.02 ± 0.24 ≤ 0.16
5 ± 0.08 ≤ 0.04 ± 0.24 ≤ 0.16
10 ± 0.10 ≤ 0.06 ± 0.24 ≤ 0.16
P8x20M
P12x20M
F81027
F81028
DL10
D200
DFL10
DF30
1-20 µL 1-20 µL
1 ± 0.08 ≤ 0.05 ± 0.4 ≤ 0.2
10 ± 0.15 ≤ 0.10 ± 0.4 ≤ 0.2
20 ± 0.25 ≤ 0.12 ± 0.4 ≤ 0.2
P8x100M
P12x100M
F81018
F81019
D200
DF100
10-100 µL 5-100 µL
10 ± 0.25 ≤ 0.14 ± 1.6 ≤ 0.6
50 ± 0.50 ≤ 0.20 ± 1.6 ≤ 0.6
100 ± 0.80 ≤ 0.25 ± 1.6 ≤ 0.6
P8x200M
P12x200M
F81029
F81030
D200
D300
DF100 DF200 DF300
20-200 µL 5-200 µL
20 ± 0.50 ≤ 0.16 ± 3.2 ≤ 1.2
100 ± 1.00 ≤ 0.30 ± 3.2 ≤ 1.2
200 ± 2.00 ≤ 0.50 ± 3.2 ≤ 1.2
P8x300M
P12x300M
F81031
F81032
D200
D300
DF200
DF300
10-300 µL 10-300 µL
10 ± 1.00 ≤ 0.18 ± 8.0 ≤ 3.0
30 ± 1.00 ≤ 0.18 ± 8.0 ≤ 3.0
150 ± 1.50 ≤ 0.375 ± 8.0 ≤ 3.0
300 ± 2.40 ≤ 0.45 ± 8.0 ≤ 3.0
P8x1200M
P12x1200M
F81020
F81021
D1200 DF1200 50-1200 µL 50-1200 µL
50 ± 4.0 ≤ 0.7 ± 32 ≤ 12
120 ± 4.0 ≤ 0.7 ± 32 ≤ 12
600 ± 6.0 ≤ 1.5 ± 32 ≤ 12
1200 ± 9.6 ≤ 1.8 ± 32 ≤ 12
Gilson ISO 8655-2
NEW
NEW
NEW
NEW
NEW
NEW
NEW
Volume range
Maximum Permissible Errors
(standard pipetting mode)
We strongly recommend you fully charge the battery before using the pipette. PIPETMAN M is supplied with an AC adaptor that is suitable for your country. You must only use an original Gilson AC adaptor specic to this product. Charge the battery in the pipette, using the AC-adapter or the stand adapter.
- 31 -- 30 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
20 REPLACEMENT PARTS
Single channel models
Multichannel models
Description P8x10M P12x10M P8x20M P12x20M P8x100M P12x100M P8x200M P12x200M P8x300M P12x300M P8x1200M P12x1200M
Battery tag, 1x4 F807013 F807013 F807013 F807013 F807013 F807013 F807013 F807013 F807013 F807013 F807013 F807013
Ejector tag, 1x4 F807014 F807014 F807014 F807014 F807014 F807014 F807014 F807014 F807014 F807014 F807014 F807014
Power supply USB F807015 F807015 F807015 F807015 F807015 F807015 F807015 F807015 F807015 F807015 F807015 F807015
Power tub cover F807022 F807022 F807022 F807022 F807022 F807022 F807022 F807022 F807022 F807022 F807022 F807022
Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005
Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006 F807006 F807006 F807006 F807006 F807006
Button assembly F807141 F807141 F807144 F807144 F807124 F807124 F807143 F807143 F807145 F807145 F807125 F807125
Ejector spacer F507001 F507003 F507001 F507003 F507001 F507003 F507001 F507003 F507001 F507003 F507139 F507140
Ejector spacer D10 F807117 F807118 - - - - - - - - - -
Ejector lock F507008 F507008 F507008 F507008 F507008 F507008 F507008 F507008 F507008 F507008 F507008 F507008
Tip-ejector assembly F507005 F507006 F507005 F507006 F507005 F507006 F507005 F507006 F507005 F507006 F507137 F507138
Description P10M P20M P100M P200M P300M P1200M P5000M P10mLM
Connecting nut F807012 F807012 F807012 F807012 F807012 F807012 F807012 F807012
Battery tag, 1x4 F807013 F807013 F807013 F807013 F807013 F807013 F807013 F807013
Ejector tag, 1x4 F807014 F807014 F807014 F807014 F807014 F807014 F807014 F807014
Power supply USB F807015 F807015 F807015 F807015 F807015 F807015 F807015 F807015
Power tub cover F807022 F807022 F807022 F807022 F807022 F807022 F807022 F807022
Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005
Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006 F807006
Lubrican tube 3.5g - F3070902 F3070902 F3070902 F3070902 F3070902 F3070902 F3070902
Seal, 1x5 F161902 - - - - - - -
Seal & O-R ing, 5 sets - F144863 - - - - - -
O-ring, 1x5 - - F807146 F2070501 F807134 F2070601 F807148 F807149
Tip-holder, upper part F2070117 F2070117 F807135 F2070517 F2070517 F2070617 - -
Tip-holder, lower part F2070218 F2070318 F807136 F2070518 F2070518 F2070618 F2070719 F807147
Button assembly F807141 F807142 F807119 F807143 F807120 F807121 F807122 F807123
Tip-ejector assembly F807008 F807009 F807130 F807010 F807010 F807011 F807131 F807132
Piston assembly F807017 F807018 F807126 F807019 F807127 F807020 F807128 F807129
Ejector spacer
Ejector lock
Tip-ejector assembly
1. Determine the mean value e of the evapora­tion loss ei that occurs during your pipetting
cycles. Proceed as described in Appendix 3 to determine e
i
2. Change the pipette tip and perform the first weighing. Then, keep a regular
cycle and perform the 10 following measurements.
3. Calculate the mean weight
4. Calculate the mean volume
For a temperature of 21.5 °C and an air pressure of 1013 hPa, the Z factor is equal to 1.0032 µL/ mg (see table in Appendix 2).
5. Evaluate accuracy
6. Evaluate precision (repeatability)
Standard Deviation (SD
W
)
wr = 0.957 mg w1 = 0.968 mg
w6 = 0.966 mg w7 = 0.955 mgw2 = 0.960 mg
w8 = 0.972 mgw3 = 0.984 mg w4 = 0.942 mg w5 = 0.969 mg
w9 = 0.958 mg
wr rinsing measurement which is
disregarded for the calculation
w10 = 0.967 mg
W =
_
W = (0.968+0.960+0.984+0.942
+ 0.969+0.966+0.955+0.972 + 0.958+0.967) / 10
_
W =
0.964 mg
_
n number of weighings W
i
weighing results
Wi
1
n
n
i=1
V = ( W+e ) x Z
V = (0.964+0.018) x 1.0032
_
V =
0.985 µL
_
E = V - V0
V0 true value set on the instrument E = 0.985 - 1 = - 0.015 µL
E% = (V - V0) x 100 / V0
E% = (-0.015 x 100) / 1 = - 1.50 %
Systematic error (E):
Relative error (E%):
(
W
i - W
)
n - 1
SDw =
2
n
i=1
n
i=1
SD
w
2
= ( Wi - W )
2
1
n - 1
1
9
SD
w
2
=
SD
w
= 0.011 mg
(0.968-0.964)2+ (0.960-0.964)2+ (0.984-0.964)2+ (0.942-0.964)
2
+ (0.969-0.964)2+ (0.966-0.964)2+
(0.955-0.964)
2
+ (0.972-0.964)2+ (0.958-0.964)2+
(0.967-0.964)
2
SDv = SD
w
x Z
SDv = 0.011 x 1.0032 = 0.011 µL
Random error (SDv) :
APPENDIX 1  EXAMPLE OF A PERFORMANCE CHECK
Below is an example of how to evaluate the performance of PIPETMAN P10M at 1 µL.
- 33 -- 32 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
APPENDIX 2  Z FACTOR
The reference calculation equation is :
Where : PA = density of air at t°C.
PW = density of the test liquid at t°C. PB = density of the balance weights. Use 8 g/cc for P
B
Weights conforming to International Recommendation N°33 of
OIML have been adjusted to give results when weighing in air as if the density of the weights were 8.0 g/mL.
Values of the conversion factor Z (µL/mg) as a function of temperature and pressure for distilled water.
Temperature Air pressure (hPa) (°C) 800 853 907 960 1013 1067
15 1.0018 1.0018 1.0019 1.0019 1.0020 1.0020
15.5 1.0018 1.0019 1.0019 1.0020 1.0020 1.0021
16 1.0019 1.0020 1.0020 1.0021 1.0021 1.0022
16.5 1.0020 1.0020 1.0021 1.0022 1.0022 1.0023
17 1.0021 1.0021 1.0022 1.0022 1.0023 1.0023
17.5 1.0022 1.0022 1.0023 1.0023 1.0024 1.0024
18 1.0022 1.0023 1.0024 1.0024 1.0025 1.0025
18.5 1.0023 1.0024 1.0025 1.0025 1.0026 1.0026
19 1.0024 1.0025 1.0025 1.0026 1.0027 1.0027
19.5 1.0025 1.0026 1.0026 1.0027 1.0028 1.0028
20 1.0026 1.0027 1.0027 1.0028 1.0029 1.0029
20.5 1.0027 1.0028 1.0028 1.0029 1.0030 1.0030
21 1.0028 1.0029 1.0030 1.0030 1.0031 1.0031
21.5 1.0030 1.0030 1.0031 1.0031 1.0032 1.0032
22 1.0031 1.0031 1.0032 1.0032 1.0033 1.0033
22.5 1.0032 1.0032 1.0033 1.0033 1.0034 1.0035
23 1.0033 1.0033 1.0034 1.0035 1.0035 1.0036
23.5 1.0034 1.0035 1.0035 1.0036 1.0036 1.0037
24 1.0035 1.0036 1.0036 1.0037 1.0038 1.0038
24.5 1.0037 1.0037 1.0038 1.0038 1.0039 1.0039
25 1.0038 1.0038 1.0039 1.0039 1.0040 1.0041
25.5 1.0039 1.0040 1.0040 1.0041 1.0041 1.0042
26 1.0040 1.0041 1.0042 1.0042 1.0043 1.0043
26.5 1.0042 1.0042 1.0043 1.0043 1.0044 1.0045
27 1.0043 1.0044 1.0044 1.0045 1.0045 1.0046
27.5 1.0044 1.0045 1.0046 1.0046 1.0047 1.0047
28 1.0046 1.0046 1.0047 1.0048 1.0048 1.0049
28.5 1.0047 1.0048 1.0048 1.0049 1.0050 1.0050
29 1.0049 1.0049 1.0050 1.0050 1.0051 1.0052
29.5 1.0050 1.0051 1.0051 1.0052 1.0052 1.0053
30 1.0052 1.0052 1.0053 1.0053 1.0054 1.0055
Z = [1/(PW-PA)] [1-(PA/PB)]
APPENDIX 3  EVAPORATION LOSS
Procedure for the determination of evaporation loss
Use the same distilled water, weighing vessel and balance as you will be using for the gravimetric check.
1. Half fill the weighing vessel with distilled water.
2. Cover the weighing vessel with its lid and place it on the balance using a pair of tweezers.
3. Aspirate a sample.
4. Tare the balance and take the weighing vessel out of the balance.
5. Take off the lid with tweezers.
6. Dispense the sample into a dummy vessel.
7. Replace the lid on the weighing vessel and, using tweezers, replace the vessel on the balance.
8. Read the negative result e
1
(record the absolute value).
9. Repeat steps 3 to 8, three times to obtain e
2
, e3, and e4.
10. Calculate the evaporation loss e using the formula:
In normal conditions, this value is usually between 0.01 mg
and 0.03 mg.
APPENDIX 4  PIPETMAN M SOFTWARE
e = (e1 + e2 + e3 + e4)
1 4
Using the Custom Mode
USER MANUAL
3534
EN
1. Introduction ........................................................................................................3
1.1 What is PIPETMAN M Software? ............................................................................ 3
1.2 Passwords and user information ...........................................................................3
1.3 Forgotten password ...................................................................................................3
2. Home screen .......................................................................................................4
3. Installing PIPETMAN M Software ...................................... 5
3.1 Computer requirements for PIPETMAN M Software .....................................5
3.2 Installing PIPETMAN M software .......................................................................... 5
3.3 Installing the drivers...................................................................................................6
3.4 Software files ................................................................................................................ 6
4. Uninstalling PIPETMAN M Software ............................. 6
5. Creating and managing custom protocols ...........6
5.1 Creating a custom protocol .....................................................................................6
5.1.1 Default task parameters.................................................................................. 8
5.1.2 Editing, cloning and deleting tasks ............................................................11
5.1.3 Pipette volume indicator ...............................................................................11
5.1.4 Warning and error messages .......................................................................12
5.2 Managing existing protocols in Manage custom protocols ........................12
5.3 Importing a protocol................................................................................................13
5.4 Practical tips for creating custom protocols....................................................13
6. Transferring custom protocols to the pipette 13
7. Updating pipette settings ...........................................................15
8. Other administration ..........................................................................17
8.1 Pipette management ...............................................................................................17
8.2 View history .................................................................................................................18
8.3 Change password ......................................................................................................19
8.4 Institution/Company profile..................................................................................19
1. Introduction
1.1 What is PIPETMAN M Software?
PIPETMAN M Software is an easy-to-use software for creating custom pipetting protocols for PIPETMAN M pipettes. Use it to create personalized pipetting protocols on your computer, and to transfer these protocols to PIPETMAN M pipettes in the Custom Mode, where you can execute them.
PIPETMAN M Software runs on Windows 7, Windows 8 and Windows 8.1. The software file is downloadable on www.gilson.com/pipetmanm.
1.2 Passwords and User information
When running the PIPETMAN M Software for the first time (e.g. after installation), you are prompted to enter some information regarding your password and your institution.
User password.
Enter user password and select a Hidden question / Hidden answer (used to recover forgotten passwords). You can also select I don’t want to use passwords. You can still switch to start using passwords later (in main menu select Other administration Change password).
User profile.
Provide Institution/Company details. Providing this information is optional and can also be done later (in main menu select Other administration Institution/Company profile).
During the installation you are prompted to enter a Recovery password that will be used to retrieve forgotten passwords.
Password must contain 8-30 characters, start with a letter, contain at least one numerical digit and one uppercase letter (no special symbols allowed).
Note: When using a password, please make sure you remember or save your Recovery
password and/or the answer to the Hidden question – those are the only two means of
retrieving your login information in the event of forgetting the password (see following section).
1.3 Forgotten password
Users who use password protection have to provide the password to run PIPETMAN M Software and to modify custom protocols in PIPETMAN M.
In the case of a forgotten password, click the Forgotten Password button on the PIPETMAN M login window. Select one of the following options to recover the forgotten password:
• I know the answer to the hidden question
• I have the recovery password
And click Next. Provide the answer to the Hidden question or the Recovery password on the next screen and click Next. On the following window, provide a new password and click
Finish.
2. Home screen
After launching PIPETMAN M Software you are directed to the home screen.
The Home screen offers the following options:
Manage custom protocols - Create, edit, import or export custom protocols.
Transfer custom protocols to pipette – Connect a PIPETMAN M pipette via USB cable and
transfer a selected custom protocol to the PIPETMAN M pipette.
Update pipette settings - Access additional pipette settings such as locking functionality,
sound alerts, maximum volume limit, or pipette firmware updates.
Other administration – Manage the inventory of your PIPETMAN M pipettes, view history
of custom protocol uploads to specific PIPETMAN M pipettes, manage passwords and the Company/Institution profile.
By clicking the home icon
you return to the home screen.
Menu bar offers the following options:
File Exit; exit the PIPETMAN M Software
Help Help; access the user’s guide
Help About; access basic information on the PIPETMAN M Software
3. Installing PIPETMAN M Software
3.1 Computer requirements for PIPETMAN M Software
PIPETMAN M Software is supported on PCs with Windows 7, Windows 8 and Windows 8.1.
Table below lists the computer system requirements for the software.
System Minimum/Allowed Recommended
Operating system Windows 7 Windows 7 and above
Hard drive 1 GB 10 GB
Processor speed 2.0 GHz 2.0 GHz
RAM 1 GB 2 GB
Screen resolution 480x800, 768x1024, 720x1280,
768x1280, 800x1280, 768x1366
768x1024
Text size (Windows only) 100%, 125%, 150% 100%
USB USB 2.0 High-speed port USB 2.0 High-speed port
3.2 Installing PIPETMAN M Software
i. Download the appropriate PIPETMAN M Software installation file for your operating
system to the computer.
ii. Double-click the installer and follow the instructions on the screen to complete the
installation.
iii. When completed, the PIPETMAN M Software icon will appear on desktop of the
computer if Create Desktop Shortcut is checked.
USER MANUAL
3736
EN
PIPETMAN M Software home screen
USER MANUAL
3938
EN
iv. If PIPETMAN M Software is not launched automatically, double-click the
PIPETMAN M icon on the desktop.
v. Once installed, the PIPETMAN M Software can be run by any user on the installed
computer.
3.3 Installing the drivers
On Windows 7 systems, drivers are installed along with the installer. On some Windows 8 and Windows 8.1 systems, drivers might need to be installed manually.
3.4 Software files
PIPETMAN M Software stores information about custom pipetting protocols in specific files with the extension .pmp.
4. Uninstalling PIPETMAN M Software
When uninstalling PIPETMAN M, you can select to remove the user data (select checkbox Remove user data during uninstall process). This will remove all user data, custom protocols, pipette data and history related to pipettes and protocols.
5. Creating and managing custom protocols
In Manage custom protocols you can create, edit, import or export custom protocols.
5.1 Creating a custom protocol
Start creating a new custom protocol from scratch by clicking the Create a new protocol button.
Specify the Protocol name (e.g. DNA-isolation) and click Create. The edit custom protocol screen appears with the Protocol info and Protocol setup tabs.
In the Protocol info tab, select a pipette from a list of PIPETMAN M models. Enter the name of the user who created the protocol (protocol created by: field) and enter any comments into the Comments box, if required.
In the Protocol setup tab, build a pipetting protocol by adding tasks from the Available tasks list (on the left of the screen) to the Protocol pane (on the right of the screen) and modifying the task’s default parameters:
1. Select a task in the Available task list.
2. Drag-and-drop it to the Protocol pane, use the arrow button
or simply double-
click on the task).
3. Modify the task’s parameters in the Edit task window that opens automatically after you add the task to protocol, and click Create to add the task.
4. Add further tasks to the protocol if desired.
5. When the protocol contains all desired tasks, click the Save button to store the protocol. If you are using password protection, you will be asked to confirm this action by entering your password.
The protocol is now listed in Manage custom protocols.
Protocols can contain up to 40 tasks.
Tasks listed in the protocol pane will be executed on the PIPETMAN M from top to bottom. You can move the task’s position in the protocol by drag-and-drop movement or by using up/down arrows next to the task
.
Building a custom protocol
USER MANUAL
4140
EN
Default tasks available:
Aspirate: aspirates specified volume into the tip.
Dispense: dispenses specified volume from the tip.
Dispense all: dispenses all remaining volume from the tip followed by a purge action.
Repeat
(1)
: enables repetition of any task once or more.
Mix: mixes the specific volume in one or more cycles.
Purge: dispenses all remaining volume from the tip followed by purging
extra volume.
Wait: introduces a wait for a specified number of seconds.
Wait for click: pauses the protocol until Push-button is clicked.
Beep: makes pipette produce a beeping sound while selected tasks are executed.
Add tasks to the protocol through drag-and-drop actions, by selecting a task and clicking the arrow button
or by double-clicking the task. Drag-and-drop actions position the tasks in the following way: the selected task will be placed directly above the task highlighted in the protocol section. If you are moving a task upwards or creating a new task, the task will be positioned above the highlighted task. If you are moving the task downwards, the task will be dropped after the highlighted task.
After you add a task to the protocol, you can modify task parameters.
(1)
Tasks within Repeat will be executed in a sequence for the specified number of iterations. Example: If Repeat contains two tasks: Aspirate 20 µL followed by Dispense 20 µL and the number of iterations is set to 3, the tasks will be executed on PIPETMAN M in the following order:
1. Aspirate 20 µL, 2. Dispense 20 µL, 3. Aspirate 20 µL, 4. Dispense 20 µL, 5. Aspirate 20 µL, and 6. Dispense 20 µL.
5.1.1. Default task parameters
Each task has modifiable parameters, defined as follows:
Aspirate and Dispense tasks have the following parameters:
Task name
(2)
: specify the name of the task.
Volume (µL)
(2)
: specify the volume for the task in µL.
Speed
(2)
: specifiy the speed for aspirating and dispensing.
Comments: enter any comments related to the task.
Display text
(2)
: enter text that is displayed on the pipette’s LCD screen when the task is
activated (note that the number of characters for display text is limited
(3)
).
Execute automatically (checkbox)
(4)
: specify whether the task should be executed
automatically after the previous task is completed.
Dispense all task has the following parameters:
Task name
(2)
: specify the name of the task.
Speed
(2)
: specify the speed for the dispense.
Comments: enter any comments related to the task.
Display text
(2)
: enter text that is displayed on the pipette’s LCD screen when the task is
activated (note that the number of characters for display text is limited
(3)
).
Execute automatically (checkbox)
(4)
: specify whether the task should be executed
automatically after the previous task is completed.
Repeat task has the following parameters:
Task name
(2)
: specify the name of the task.
Number of iterations
(2)
: specify how many times the pipette should repeat the tasks
that are placed inside the repeat task.
Comments: enter any comments related to the task.
Note: for certain PIPETMAN M models the minimal volumes in Aspirate and Dispense
tasks might be different if these tasks are located inside Repeat loop.
Mix task has the following parameters:
Task name
(2)
: specify the name of the task.
Volume (µL)
(2)
: specify the volume for the task in µL.
Aspiration speed
(2)
: specify the speed for the aspiration.
Dispense speed
(2)
: specify the speed for the dispense.
Use Push-button mixing (checkbox): when checked, mixing will be performed
continuously for as long as the user holds the Push-button on the pipette; once the Push-button is released, mixing stops and the next task in the protocol is executed.
Number of cycles
(2)
: specify how many times the volume should be mixed.
Comments: enter any comments related to the task.
Display text
(2)
: enter text that is displayed on the pipette’s LCD screen when the task is
activated (note that the number of characters for display text is limited
(3)
).
Execute automatically (checkbox)
(4)
: specify whether the task should be executed
automatically after the previous task is completed.
USER MANUAL
4342
EN
Purge task has the following parameters:
Task name
(2)
: specify the name of the task.
Comments: enter any comments related to the task.
Display text
(2)
: enter text that is displayed on the pipette’s LCD screen when the task is
activated (note that the number of characters for Display text is limited
(3)
).
Execute automatically (checkbox)
(4)
: specify whether the task should be executed
automatically after the previous task is completed.
Wait task has the following parameters:
Task name
(2)
: specify the name of the task.
Duration(s)
(2)
: specify the duration of the pause in seconds.
Comments: enter any comments related to the task.
Display text
(2)
: enter text that is displayed on the pipette’s LCD screen when the task is
activated (note that the number of characters for Display text is limited
(3)
).
Wait for click task has the following parameters:
Task name
(2)
: specify the name of the task.
Comments: enter any comments related to the task.
Display text
(2)
: enter text that is displayed on the pipette’s LCD screen when the task is
activated (note that the number of characters for Display text is limited
(3)
).
Beep task has the following parameters:
Task name
(2)
: specify the name of the task.
Comments: enter any comments related to the task.
Example of use: in order to produce a beep sound with duration of 1 second on the pipette before the Aspirate task, position the following set of tasks before the Aspirate task:
- Beep task (creates the beep start and end tasks).
- Wait task (with duration 1 s), which should replace the beep placeholder task between the two beep tasks.
(2)
Parameters marked with an asterisk are mandatory.
(3)
Pipettes are able to display a limited number of characters on their LCD screen. If you use
longer strings of characters in Display text not all may be displayed.
(4)
Display text field is disabled when Execute automatically is checked. Additionally, the
Use Push-button mixing option for mix mode is also disabled for tasks that are
automatically executed.
5.1.2. Editing, cloning and deleting tasks
Once the task is listed in the protocol, you can edit its parameters by selecting the task, right-clicking on it and selecting Edit task, or by double-clicking the task. The Edit task window opens, where you can modify the task’s parameters and store them.
You can clone a task by right-clicking on the desired task and selecting Clone task. Add task window opens, where you can rename the task and change the parameters if desired. When done, add the task to the protocol by clicking Create button. The task is added immediately after the original task.
You can delete tasks from the protocol by right-clicking on the desired task and selecting Delete task. Alternatively, you can select a task in the protocol and click the left-arrow
.
When deleting a task, a pop-up window appears, where you can confirm task deletion.
5.1.3. Pipette volume indicator
As you build a protocol, the pipette volume indicator will display how much of the pipette’s maximum volume is used at the end of the protocol with indicator bars (green to red color). The exact volume is also displayed next to the color bars (displayed in microliters). If you click on a particular task, a rectangle will mark the volume of the current task (blue rectangle indicates aspirating, and red rectangle indicates dispensing of volume).
If the volume in pipette exceeds the maximal volume, or is below the minimal volume of the pipette, the pipette volume indicator will be surrounded with a red rectangle.
Numbers displayed next to colored bars represents
volume remaining in the pipette at the end of protocol
(displayed in microliters).
100%
full pipette
0%
full pipette
50%
full pipette
USER MANUAL
4544
EN
5.1.4. Warning and error messages
PIPETMAN M Software verifies the protocol after the addition of each task, or after modification of a task, to determine whether the volume of the pipette is still within the volume specifications for that particular PIPETMAN M model. Should the volume in pipette model fall outside specifications, an error sign
will appear next to the task that caused this. More details about error are available if you position the mouse over the error sign. Protocols that contain one or more error signs
may be saved, however they cannot be transferred to a pipette and cannot be
exported.
In certain cases a yellow warning sign
will appear next to a protocol in Manage custom protocols and Transfer custom protocols to pipette. More details about the warning are available if you position the mouse over the warning sign. Such protocols may be transferred to a pipette and can be exported. However, executing these protocols on the pipette repeatedly may cause the pipette to have an error.
5.2 Managing existing protocols in Manage custom protocols
All protocols that have been created are listed in the Manage custom protocols window with information on pipette model, date of creation, date of last modification, person that created the protocol, and person that last modified the protocol. If a protocol contains a warning or an error, the warning or error icon is displayed.
For each of the protocols listed you can perform the following actions. Simply select the protocol and click on one of the buttons:
View: view a protocol. No parameters can be changed in the “View mode”.
Edit: edit a protocol (change the PIPETMAN M model, rename protocol, add/edit/
remove tasks).
Clone: clone a protocol (if you are creating a similar protocol, cloning will save time).
Export: export a protocol to a .pmp file and store it on the computer.
Export to pdf: create a summary of the protocol in the form of a print-ready pdf
document.
Delete: delete a protocol.
Alternatively, you can perform these actions by right-clicking on a desired protocol. Editing protocol is also available by double-clicking the selected protocol.
5.3 Importing a protocol
You can add protocols to the Manage custom protocols by importing a previously created protocol (.pmp files).
Click on Import a protocol from file button, navigate to the .pmp file and click Open. The protocol is then available in Manage custom protocols.
5.4 Practical tips for creating custom protocols
If you have existing protocols, cloning and modifying them will save you time.
PIPETMAN M Software enables you to easily create a similar protocol on a different PIPETMAN M model. Simply clone the protocol and change the PIPETMAN M model. If necessary, modify the volume of transfers to fit volume specifications of the new PIPETMAN M model.
6. Transferring custom protocols to the pipette
To transfer a protocol to a PIPETMAN M pipette, click the Transfer custom protocols to pipette button and follow the instructions on the screen:
1. Connect pipette to the PC via USB cable and click Next. Note: should the pipette not be recognized immediately, disconnect the USB cable, reconnect it and click Tr y again button. If a pipette is already correctly connected when Transfer custom protocols to pipette button is clicked, this step is skipped.
2. By default, the protocols for the connected PIPETMAN M model are listed. You can access the protocols created for other PIPETMAN M models by selecting the model from the Show protocols for pipette model
(2)
: drop-down list.
Custom protocols are listed in Manage custom protocol window
USER MANUAL
4746
EN
3. Select a protocol in the list and click Finish.
4. If a pipette is locked, transferring protocols will unlock it. For that purpose, a pop-up window appears, where you can confirm that you would like to transfer the selected protocol, unlocking the pipette in the process.
5. You can also check “Mute beeps” if you want to disable any pipette sounds during beep tasks.
6. Disconnect the pipette. The Custom protocol is now available on the pipette.
Note: Protocols that have a red error sign
cannot be transferred to pipettes.
• You can transfer a protocol created for a different PIPETMAN M model to the connected
pipette, providing the protocol is using volumes within the connected pipette’s volume specifications. Protocols that contain volumes outside connected pipette’s volume will have a red error sign
.
• The pipette must be registered in the software prior to transferring custom protocols. If the
connected pipette is not yet registered, a pop-up appears which informs you that the pipette is not yet registered. Here, you can click on Add pipette now, which takes you to the pipette management screen, or you can proceed to/stay at step 1 by clicking Connect different pipette.
• On the bottom left of Transfer custom protocols to pipette wizard, you can find the Pipette
connection information - An icon accompanied with text which reports the current connectivity status. The icon is green
if there is a pipette connected to the computer
or gray
when there are no pipettes connected. The connectivity status is refreshed
when any of the buttons is clicked.
7. Updating pipette settings
Here you can set the following configurations on the connected PIPETMAN M pipette:
Set the Beep to On or Off (switches the tone of the pipette to On or Off). Beep On and
Off only affects the built-in pipette protocols, as well as Aspirate, Dispense, Mix, Purge tasks (but NOT beep task!).
Set the Lock pipette On or Off (Lock pipette On disables access to Configuration,
Speed and Mode on PIPETMAN M pipette)
Set Maximal volume limit: this sets the maximal allowed volume on the particular
pipette; it cannot be higher than pipette’s specified maximal volume. Simply enter the volume (in µL) in the field.
Update Firmware.
To modify these configurations, follow these steps:
1. Click Update pipette settings in the main menu.
2. If a pipette is already correctly connected, proceed to step 4.
3. Connect pipette to the PC via USB cable and click Next. Should the pipette not be recognized immediately, disconnect the USB cable, reconnect it and click Tr y again button.
Selecting a protocol for transfer to a PIPETMAN M P200M pipette.
Update pipette settings.
USER MANUAL
4948
EN
4. Check the checkbox for the configuration that you would like to change and then modify it (e.g. change Beep from On to Off; or modify the Maximal volume limit); Unchecked configurations will not be updated.
5. Click Finish.
6. Disconnect pipette.
You can refresh the pipette’s Maximal volume limit to pipette’s specified maximal volume by clicking the Reset to pipette max.
Note: the pipette must be registered in the software prior to transferring custom
protocols. If the connected pipette is not yet registered, a pop-up appears which informs you that the pipette is not yet registered. Here, you can click on Add pipette now, which takes you to the pipette management screen, or you can proceed to/stay at step 1 by clicking Connect different pipette.
• On the bottom left of Update pipette settings wizard, you can find the Pipette connection
information – An icon accompanied with text which reports the current connectivity status. The icon is green
if there is a pipette connected to the computer or gray when there are no pipettes connected. The connectivity status is refreshed when any of the buttons is clicked.
Updating pipette firmware
To update firmware on PIPETMAN M pipette follow these steps:
1. Click Update pipette settings in the main menu.
2. If a pipette is already correctly connected, proceed to step 4.
3. Connect pipette to the PC via USB cable and click Next. Should the pipette not be recognized immediately, disconnect the USB cable, reconnect it and click Tr y again button.
4. Click the Update firmware button.
5. Navigate to the firmware file and click Open.
6. On the next dialog, click OK to confirm the firmware update. IMPORTANT: the pipette
must not be disconnected during firmware update. Disconnecting pipette could result in pipette malfunction.
7. Wait for the firmware update, click OK and disconnect pipette.
8. Other administration
Here you can perform the following administrative tasks:
• Pipette management - Manage the inventory of your PIPETMAN M pipettes.
• View history - View the history of protocol uploads to PIPETMAN M pipettes and the
history of protocol modifications by user.
• Change password - Change the password to access PIPETMAN M Software.
Institution/Company profile - Manage the information related to your Institution/Company.
8.1 Pipette management
Here you can register your individual PIPETMAN M pipettes. Registering a pipette is required to allow for transfer of your personalized protocols onto the pipette, as well as updating the pipette settings.
Register a new pipette by clicking Register new pipette button and follow the instructions on the screen:
1. Connect a pipette to the PC via USB cable and click Next. Should the pipette not be recognized immediately, disconnect the USB cable, reconnect it and click Try again button. If a pipette is already correctly connected when Register new pipette button is clicked, this step is skipped.
2. Type in the name of the pipette and click Finish.
3. Pipette is now listed in the Registered pipettes list along with the basic information
(Model, Firmware version, Serial number, Minimal volume (µL), Maximal volume (µL), Last maintenance date, Next maintenance date).
You can rename a pipette in the Registered pipettes list by selecting the pipette, clicking the Rename button, entering a new name and clicking Rename.
You can reset the maintenance dates (last and next) by clicking the Refresh dates button. This process requires you to connect the selected pipette to the computer.
You can delete a pipette in the Registered pipettes list by selecting the pipette, and clicking Delete this pipette button.
Note: on the bottom left of Register new pipette wizard, you can find the Pipette connection
information – An icon accompanied with text which reports the current connectivity
status. The icon is green
if there is a pipette connected to the computer or gray when there are no pipettes connected. The connectivity status is refreshed when any of the buttons is clicked.
USER MANUAL
5150
EN
8.2 View history
Here you can view the history of protocol uploads to PIPETMAN M pipettes (tab Protocol upload) and history of protocol modifications (tab Protocol management).
To display the history of protocol uploads to PIPETMAN M pipettes, navigate to the Protocol upload tab, select the time interval and click Show history. The protocols that have been uploaded to any PIPETMAN M pipette within the selected time interval will be listed in the table below, displaying the following information for each protocol: Protocole name, Pipette name, Pipette model, Pipette serial, Uploaded on, and Result.
To display history of protocol modifications, navigate to Protocol management tab, select the time interval, and click Show history. The protocols that have been modified within selected time interval will be listed in the table below, displaying the following information for each protocol: protocol ID, Original protocol name, Modified by, Modified on, and Action.
8.3 Change password
Here you can change a password, change the Hidden question and Hidden answer, or disable/enable the use of passwords (checkbox: Don’t use password).
Note: remember or store the answer to the hidden question in case of password recovery.
8.4 Institution/Company profile
Enter the information related to your Institution/Company here or modify it. Simply enter the information in provided fields and click Save.
None of the information is mandatory.
Display history of protocol uploads to PIPETMAN M pipettes
Change password, Hidden question, Hidden answer
Enter the information related to your Institution/Company
Display history of protocol modification
- 53 -- 52 -
USER’S GUIDE
EN
PIPETMAN M
USER’S GUIDE
EN
PIPETMAN M
EC DECLARATION OF CONFORMITY
The company
Hereby certifies on its sole responsibility that the products listed below:
comply with the requirements of the following European Directives:
GILSON S.A.S.
19, avenue des Entrepreneurs BP 145 - 95400 VILLIERS LE BEL FRANCE Tel.: +33(0)1 34 29 50 00 Fax: +33(0)1 34 29 50 20
PIPETMAN® M
P10M, P20M, P100M, P200M, P300M, P1200M, P5000M, P10mLM P8x10M, P12x10M, P8x20M, P12x20M P8x100M, P12x100M, P8x200M, P12x200M, P8x300M, P12x300M, P8x1200M, P12x1200M
* Annex III, self declared
98/79/EC*
In Vitro Diagnostic Medical Devices
2004/108/EC
Electromagnetic Compatibility, EMC
2006/95/EC
Low Voltage Directive, LVD
NOTES
- 54 -
USER’S GUIDE
EN
PIPETMAN M
WARRANTY
Gilson warrants this pipette against defects in material under normal use and service for a period of 24 months from the date of purchase.
This warranty shall not apply to pipettes which are subject to abnormal use and/or improper or inadequate maintenance (contrary to the recommendations given in the User’s guide), including, but not limited to pipettes which have been subjected to physical damage, improper handling, or spillage or exposure to any corrosive environment. This warranty shall also be void in the event pipettes are altered or modified by any party other than Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes.
THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE, UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE, INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR IMPLIED WARRANTY.
Gilson, Inc. World Headquarters
3000 Parmenter Street I P.O. Box 628098 Middleton, WI 53562 I USA Tel: (1) 800-445-7661 I Fax: (1) 608-821-4403
Gilson S.A.S. I 19, avenue des Entrepreneurs
BP 145 I F-95400 Villiers-le-Bel I FRANCE Tel: +33(0)1 34 29 50 00 I Fax: +33 (0)1 34 29 50 20
LT801527/E © 11/2015 Gilson SAS.
All rights reserved. All trademarks
are the property of Gilson Inc.
and its subsidiaries.
Specifica
tions subject to change
without notifications - error omitted.
P
rinted in France.
www.gilson.com I sales@gilson.com
service@gilson.com I training@gilson.com
Loading...