Your personally designed MyPIPETMAN® pipette has an air-displacement mechanism and is used with
disposable pipette tips.
This Gilson pipette line provides:
●Light and comfortable body
●Unique design and configuration
●Low pipetting forces ensuring ergonomics for users’ well-being
Eight single channel models cover a volume range from 0.2 μL to 10 mL.
Eight multichannel models (4 models of 8 and 12 multichannel) cover a volume range from 0.5 μL to 300 μL.
Parts Checklist
Carefully unpack your pipette and take a moment to verify that the following items are included and
undamaged:
SINGLE MODELS
●Your personalized MyPIPETMAN
●Quick Guide
●Certificate of Conformity
●Safety bag
●Tip ejector adapter equipped on the
P2 μL and P10 μL models (fully autoclavable)
●Bag of ten filters
channel models
(only for P10mL
)
single
MULTICHANNEL MODELS
●Your personalized MyPIPETMAN
●Quick Guide
●Ejector spacer for short collar tips (only for
P10 μL models)
●Certificate of Conformity
●Safety bag
INTRODUCTION
GLP Features
The serial number is engraved on the body of the pipette. It provides unique identification of your pipette
and the date of manufacture. Example: MY524CQ
The barcode on the box and the Certificate of Conformity provide traceability of your pipette.
RANDOM NUMBER
231
354
324
522
457
120
876
489
542
MONTHCODE
JanuaryA
FebruaryB
MarchC
AprilD
MayE
JuneG
JulyH
AugustJ
SeptemberK
OctoberL
NovemberM
DecemberN
YEARCODE
2019Q
2020R
2021S
2022T
2023U
2024W
2025X
2026Y
2027Z
MY 524 C Q
MODEL
RANDOM
MONTH YEAR
NUMBER
MYPIPETMAN 524 MARCH 2019
Figure 1
Serial number
MyPIPETMAN®|USER’S GUIDE
4
Page 4
Description
Please refer to the following chapters for a full description of the different parts and functions of the
pipette.
INTRODUCTION
Color
window
Single channel
Multichannel
Color-coded
push button
Locking system
Tip ejector
button
Personalized
handle
Color ring
Figure 2
MYPIPETMAN™ single channel and multichannel models
Connecting nut
Ejector clip
Tip holder
Ejector lock
Cover
Tip ejector
5
INTRODUCTION| MyPIPETMAN®
Page 5
Specifications
NOTE
MyPIPETMAN is a high-quality pipette that offers excellent accuracy and precision especially with
PIPETMAN DIAMOND Tips. The data given in the “Gilson Maximum Permissible Errors” table were
obtained using PIPETMAN DIAMOND Tips. These values are guaranteed only when genuine
PIPETMAN DIAMOND Tips are used.
Each pipette is inspected and validated by qualified technicians in accordance with the Gilson quality
system. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655
standard by type testing.
The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6).
INTRODUCTION
MODELVOLUME (µL)
MyPIPETMAN SINGLE CHANNEL
0.2
MyPIPETMAN P2
MyPIPETMAN P10
MyPIPETMAN P20
MyPIPETMAN P100
MyPIPETMAN P200
MyPIPETMAN P1000
0.5
1
2
1
5
10
2
10
20
10
50
100
20
100
200
100
500
1000
GILSON MAXIMUM
PERMISSIBLE ERROR
Systematic
Error (µL)
± 0.024
± 0.025
± 0.027
± 0.030
± 0.025
± 0.075
± 0.100
± 0.10
± 0.10
± 0.20
± 0.35
± 0.40
± 0.80
± 0.50
± 0.80
± 1.60
± 3
± 4
± 8
Random
Error (µL)
≤ 0.012
≤ 0.012
≤ 0.013
≤ 0.014
≤ 0.012
≤ 0.030
≤ 0.040
≤ 0.03
≤ 0.05
≤ 0.06
≤ 0.10
≤ 0.12
≤ 0.15
≤ 0.20
≤ 0.25
≤ 0.30
≤ 0.6
≤ 1.0
≤ 1.5
ISO-8655 MAXIMUM
PERMISSIBLE ERROR
Systematic
Error (µL)
± 0.08
± 0.08
± 0.08
± 0.08
± 0.12
± 0.12
± 0.12
± 0.2
± 0.2
± 0.2
± 0.8
± 0.8
± 0.8
± 1.6
± 1.6
± 1.6
± 8.0
± 8.0
± 8.0
Random
Error (µL)
≤ 0.04
≤ 0.04
≤ 0.04
≤ 0.04
≤ 0.08
≤ 0.08
≤ 0.08
≤ 0.1
≤ 0.1
≤ 0.1
≤ 0.3
≤ 0.3
≤ 0.3
≤ 0.6
≤ 0.6
≤ 0.6
≤ 3.0
≤ 3.0
≤ 3.0
MyPIPETMAN P5000
MyPIPETMAN P10mL
MyPIPETMAN MULTICHANNEL
MyPIPETMAN P8x10
MyPIPETMAN P12x10
MyPIPETMAN P8x20
MyPIPETMAN P12x20
MyPIPETMAN P8x200
MyPIPETMAN P12x200
MyPIPETMAN P8x300
MyPIPETMAN P12x300
500
2500
5000
1000
5000
10000
0.5
1
5
10
2
10
20
20
100
200
20
30
150
300
± 12
± 15
± 30
± 30
± 40
± 60
± 0.08
± 0.08
± 0.20
± 0.20
± 0.10
± 0.20
± 0.40
± 0.50
± 1.00
± 2.00
± 1.00
± 1.00
± 1.50
± 3.00
≤ 3
≤ 5
≤ 8
≤ 6
≤ 10
≤ 16
≤ 0.04
≤ 0.05
≤ 0.10
≤ 0.10
≤ 0.08
≤ 0.10
≤ 0.15
≤ 0.25
≤ 0.40
≤ 0.50
≤ 0.35
≤ 0.35
≤ 0.60
≤ 1.00
± 40
± 40
± 40
± 60
± 60
± 60
± 0.24
± 0.24
± 0.24
± 0.24
± 0.4
± 0.4
± 0.4
± 3.2
± 3.2
± 3.2
± 8.0
± 8.0
± 8.0
± 8.0
≤ 15.0
≤ 15.0
≤ 15.0
≤ 30.0
≤ 30.0
≤ 30.0
≤ 0.16
≤ 0.16
≤ 0.16
≤ 0.16
≤ 0.20
≤ 0.20
≤ 0.20
≤ 1.2
≤ 1.2
≤ 1.2
≤ 3.0
≤ 3.0
≤ 3.0
≤ 3.0
The data given in this table conform to the ISO 8655-2 standard. With a precise
pipetting technique, the P2 model can be used to aspirate volumes as low as 0.1 μL
and the P10 model as low as 0.5 μL.
MyPIPETMAN®|USER’S GUIDE
6
Page 6
Chapter 2
NOTE
SETTING THE VOLUME
The volume of liquid to be aspirated is set using the volumeter. The dials are colored either black or red to
indicate the position of the decimal point, depending on the model (refer to Figure 3 below).
Volume selection differs according to your configuration. To obtain maximum accuracy when setting the
volume, proceed as follows:
●When decreasing the volume setting, slowly reach the required setting, making sure not to overshoot
the volume mark.
●When increasing the volume setting, pass the required value by 1/3 of a turn and then slowly decrease
to reach the volume, making sure not to overshoot the volume mark.
SETTING THE VOLUME
MODEL
MyPIPETMAN SINGLE CHANNEL
MyPIPETMAN P2µL0.01 µL0.002 µL
MyPIPETMAN P10µL0.1 µL0.02 µL
MyPIPETMAN P20µL0.1 µL0.02 µL
MyPIPETMAN P100µL-0.2 µL
MyPIPETMAN P200µL-0.2 µL
MyPIPETMAN P10000.01 mL mL0.002 mL
MyPIPETMAN P50000.01 mL mL0.002 mL
MyPIPETMAN P10mLmL0.1 mL0.02 mL
MyPIPETMAN MULTICHANNEL
MyPIPETMAN P8x10µL0.1 µL0.02 µL
MyPIPETMAN P12x10µL0.1 µL0.02 µL
MyPIPETMAN P8x200µL-0.2 µL
MyPIPETMAN P12x200µL-0.2 µL
MyPIPETMAN P8x300µL-0.2 µL
MyPIPETMAN P12x300µL-0.2 µL
COLOR OF VOLUMETER NUMBERS
BLACKREDINCREMENT
P2
1
2
5
1.25 µL
P200
1
2
5
125 µL
Figure 3
Dial colors by model
Simple Ratchet
For pipettes mounted with a simple ratchet, select the volume by
turning the push button (refer to Figure 4).
3-Position Locking System
For pipettes equipped with a 3-position locking system, turn the
selector to your desired position (refer to Figure 5).
Single channel
model
P10
7.5 µL
P1000
P20
0
7
5
1
2
5
12.5 µL
P5000
0
7
5
0.75 mL
1.25 mL
Multichannel
model
P100
x10
0
7
5
75 µL
1
2
5
P10mL
0
7
5
7.5 mL
7.5 µL
x200
125 µL
x20
0
7
5
1
2
5
12.5 µL
1
2
5
x300
1
2
5
125 µL
Free volume selection position
1. Turn the selector to the left until the cursor reaches the
unlocking icon 1.
2. Select the volume by turning the push button.
Light click volume selection position
1. Turn the selector until the cursor reaches the middle icon 2.
2. Select the volume by turning the push button.
When selecting the volume,
you will hear click sounds,
allowing for more precise
settings. Use these sounds for
final fine volume selection.
Secure Pipetting
After selecting your volume, turn the selector to
the right until the cursor reaches the locking icon
for secure pipetting 3.
7
SETTING THE VOLUME| MyPIPETMAN®
Figure 4
Simple ratchet
1
Figure 5
3-position locking
system
2
3
Page 7
Chapter 3
NOTE
PIPETTING
Fitting the Tip Ejector Adapter
For P2 and P10 pipettes equipped with a stainless steel tip ejector, a dual-position
adapter is required to fit and eject DL10 tips (long tips) and D10 tips (short tips).
To fit the adapter for short and long tips (refer to Figure 6):
1. Pull the adapter down from the stainless steel tip ejector.
2. Turn the adapter 180° 2.
3. Refit the adapter so that the end of the stainless steel tip ejector engages the
shorter or longer slot of the adapter 1.
For P2 and P10 pipettes equipped with a plastic tip ejector, a dual-position adapter
(refer to Figure 7) is required to fit and eject D10 tips (short tips).
To fit a tip ejector adapter:
1. Slide the adapter over the tip holder 1.
2. Push the adapter firmly onto the end of the tip ejector until it clicks into place.
To remove a tip ejector adapter:
1. Gently twist the adapter 2.
2. Pull it away from the pipette 3.
Fitting the Ejector Adapter for P8x10 and P12x10
Multichannel Models
According to the tip used, D10 or DL10, you might have to fit the ejector adapter;
long collar tips can be inserted without it.
1. Remove the tip ejector, keep both ejector locks depressed 1, and then pull the tip ejector down 2.
2. Fit the ejector adapter 3 and click it to the tip ejector.
3. To refit the tip ejector, gently re-insert the tip ejector vertically into the rails of the ejector support 4.
Figure 6
Adapter for short
and long tips
Figure 7
Dual-position
adapter
Short
tips
1
Long
tips
2
3
2
PIPETTING
1
Long tips can be used without tip ejector adapter.
1
2
3
4
MyPIPETMAN®|USER’S GUIDE
8
Page 8
PIPETTING
Fitting the Tips
It is recommended to use PIPETMAN® DIAMOND Tips with MyPIPETMAN for optimum performance. These
tips are made from pure polypropylene. Plastic tips are for a single application—they should not be cleaned
for reuse.
Single Channel Models
To fit a new PIPETMAN DIAMOND Tip, push the tip holder into the tip using a slight twisting motion to ensure
a firm and airtight seal.
PIPETMAN DIAMOND Tip Compatibility
for Single Channel Models
P2, P10 D10, DL10, DF10, DFL10
P20 D200, DF30
P100 D200, DF100
P200 D200, DF200, D300, DF300
P1000 D1000, D1200, DF1000, DF1200
P5000 D5000
P10mL D10mL
Multichannel Models
PIPETMAN DIAMOND Tips are best fitted onto multichannel pipettes
from the patented ROCKY RACK available only in our TI PACK S and
TOWERPACKS.
PIPETMAN DIAMOND Tip Compatibility
for Multichannel Models
P8x10L, P12x10L D10*, DL10, DF10, DFL10
P8x20L, P12x20L DL10, DFL10, DF30, D200
P8x200L, P12x200L D200, D300, DF200, DF300
P8x300L, P12x300L D200, D300, DF200, DF300
*Using an ejector adapter, you can adapt a D10 tip
(refer to page 8)
Figure 8
The patented ROCK Y RACK system
available on TOWERPACK and
TIPACK makes it extremely easy to fit
tips on a multichannel pipette
Pre-Rinse the Tips
Some liquids (e.g., protein-containing solutions and organic solvents)
can leave a film of liquid on the inside wall of the tip. Pre-rinse the tip to minimize any errors that may be
related to this phenomenon.
Pre-rinsing consists of aspirating the first volume of liquid and then dispensing it to waste.
Tips will not fall off nor will they have to be manually positioned. Subsequent volumes that you pipette will
have levels of accuracy and precision within specifications.
9
PIPETTING| MyPIPETMAN®
Page 9
Aspirate
NOTE
1. Press the push button to the first stop (this corresponds to the set volume of liquid).
2. Hold the pipette vertically and immerse the tip in the liquid (refer to Guidelines for Good Pipetting on
page 11.
3. Release the push button slowly and smoothly (to top position) to aspirate the set volume of liquid.
4. Wait one second (refer to Guidelines for Good Pipetting on page 11) and then withdraw the pipette
tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe;
however, if you do so take care to avoid touching the tip’s orifice.
For the multichannel models, use a reagent reservoir.
Dispense
1. Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
2. Press the push button slowly and smoothly to the first stop.
3. Wait for at least a second, then press the push button to the second stop to expel any residual liquid
from the tip. Keep the push button pressed fully down and (while removing the pipette) draw the tip
along the inside surface of the vessel.
4. Release the push button smoothly. Eject the tip by pressing firmly on the tip ejector button.
PIPETTING
Rest position
First stop
Second stop
Prepare
AspirateDispense
PurgeHome
Figure 9
Pipetting motion - aspirate and dispense
MyPIPETMAN®|USER’S GUIDE
10
Page 10
Guidelines for Good Pipetting
●Make sure that you operate the push button slowly and smoothly.
●When aspirating, keep the tip at a constant depth below the surface of the liquid (refer to the table).
PIPETTING
MODEL
P2, P10, P8x10, P12x1011
P20, P8x20, P12x20, P8x200, P12x2002-31
P8x300, P12x3002-41
P10002-42-3
P50003-64-5
P10mL5-74-5
●Change the tip before aspirating a different liquid, sample, or reagent.
●Change the tip if a droplet remains at the end of the tip from the previous pipetting operation.
●Each new tip should be pre-rinsed with the liquid to be pipetted.
●Liquid should never enter the tip holder; to prevent this:
●Press and release the push button slowly and smoothly.
●Never turn the pipette upside down.
●Never lay the pipette on its side when there is liquid in the tip.
●If you use the same tip with a larger volume, pre-rinse the tip.
●For volatile solvents, you should saturate the air cushion in your pipette by aspirating and dispensing
the solvent repeatedly before aspirating the sample.
●When the temperature of the liquid is different from the ambient temperature, pre-rinse the tip
several times before use.
●You may remove the tip ejector (refer to Fitting the Tip Ejector Adapter on page 8) to aspirate from
very narrow tubes.
●After pipetting acids or other corrosive liquids that emit vapors, remove the tip ejector and tip holder,
rinse with distilled water, dry, and lubricate the piston (refer to MAINTENANCE on page 16). For the
P1000L model, use a specific tip holder equipped with a filter to increase the lifetime of the piston
(refer to ACCESSORIES on page 12).
●Do not pipette liquids with temperatures above 70°C or below 4°C. The pipette can be used between
+4°C and +40°C, but the gravimetric performance may vary according to the temperature (refer to
the ISO 8655-2 standard for conditions of use).
IMMERSION DEPTH
(MILLIMETERS)
WAIT TI ME
(SECONDS)
11
PIPETTING| MyPIPETMAN®
Page 11
Chapter 4
ACCESSORIES
To make pipetting more comfortable and more secure, Gilson has developed several accessories.
Pipette stands allow users to store pipettes vertically to
avoid the possibility of liquid running back into the pipette.
DESCRIPTION PART NUMBER
CARROUSEL Pipette Stand
(7 pipettes)F161401
TRIO Pipette Stand (3 pipettes)F161405
Universal Multichannel Stand*F161417
SINGLE Pipette HolderF161406
* Universal Multichannel Stand is for multichannel models only.
COLORIS™ Identification Clips identify or personalize your
pipette. Write your name or application to avoid mix-ups
and then snap into place.
DESCRIPTION PART NUMBER
COLORIS clips (mixed colors set of 10) F161301
COLORIS clips (red, set of 10) F161302
COLORIS clips (yellow, set of 10) F161303
COLORIS clips (green, set of 10) F161304
COLORIS clips (blue, set of 10) F161305
COLORIS clips (white, set of 10) F161306
Universal Multichannel
Stand
COLORIS™
Identification Clip
ACCESSORIES
CARROUSEL™
TRIO™
Pipette Stand
SINGLE™
Pipette Holder
With the hands-free JIMMY™ Microtube Opener, you can
open both snap-cap and screw-cap microtubes to increase
pipetting efficiency and reduce the fatigue associated with
manual tube opening.
DESCRIPTION PART NUMBER
JIMMY Microtube Opener (set of 3)F144983
To protect the piston when pipetting corrosive liquids,
you can use a specific tip holder and filter for the model
P1000L.
DESCRIPTION PART NUMBER
Corrosion Protection Kit
(tip holder + a bag of 10 filters)F144570
Corrosion Protection Kit Filter (Bag of 10) F144571
Corrosion Protection Kit Filter (Bag of 50) F144572
Disposable reagent reservoirs are ideal for dispensing
reagent with multichannel pipettes.
DESCRIPTION PART NUMBER
Reagent reservoir 25 mL (x100)F267660
Reagent reservoir 50 mL (x100)F267670
Corrosion Protection Kit
Reagent Reservoir
MyPIPETMAN®|USER’S GUIDE
12
Page 12
Chapter 5
NOTE
WARNING
TROUBLESHOOTING
A quick inspection of the pipette may help you to detect a potential problem.
You may download the Two Minute Inspection Poster from the Gilson website
(www.gilson.com), which shows how to perform a quick diagnosis of your pipette.
Before returning any pipette to your local Gilson Service Center, ensure that it is
completely free of chemical, biological, or radioactive contamination. Refer to
CLEANING AND DECONTAMINATION on page 14. Please use the included safety bag
to return the pipette to your local Gilson Service Center.
Troubleshooting Table
The following table may help you to identify and correct the potential problem you might encounter.
Worn O-ring or sealReplace both parts and lubricateS
Worn O-ringReplace O-Ring and seal and lubricateS
Damaged tip holderReplace the tip holderS
Connecting nut is looseTighten connecting nutS+M
Damaged or corroded
piston
Improper repair or assembly Refer to page 16 - MAINTENANCES
Improper repair or assembly Refer to page 16 - MAINTENANCES
Unscrew tip holderTighten connecting nutS
Connecting nut is looseTighten connecting nutS+M
Tip holder is looseTighten connecting nutS
Connecting nut is looseTighten connecting nutS+M
Incorrect operator
technique
Damaged or corroded
piston(s)
Damaged tip holder(s)Replace the tip holder(s)S
Worn O-ring or sealReplace both parts and lubricateS
Low quality tipsUse PIPETMAN DIAMOND TipsS+M
Damaged tip holder(s)
Damaged tip ejector
adapter
Damaged ejectorReplace the tip ejectorS+M
Dirty tip holder
The tip ejector is looseAssemble the tip ejector properlyS
The ejector lock is
misaligned
Return pipette to supplierS
Operator trainingS+M
Return pipette to supplierS
Replace the tip holder(s) for single
channel model
Contact your local Gilson Service
Center for multichannel model
Replace the tip ejector adapterS+M
Clean the tip holder with
isopropanol or ethanol
Align the ejector lockS
S+M
S+M
*S = Sing le Chann el
M = Multichannel
13
TROUBLESHOOTING| MyPIPETMAN®
Page 13
Leak Test
NOTE
This test may be performed at any time to check that the pipette does not leak, especially after performing
a maintenance or decontamination procedure. If a pipette fails this test, replace the O-ring and seal. After
making sure that the pipette is correctly reassembled, repeat this test.
1. Fit a PIPETMAN DIAMOND Tip.
2. For the variable volume models, set the pipette to the maximum volume given in the specifications,
and pre-rinse.
For P2 to P200, pre-rinse the tip.
3. Aspirate the set volume from a beaker of distilled water.
4. Maintain the pipette in the vertical position and wait for
20 seconds.
●If a water droplet appears at the end of the tip, there is a leak.
●If you see no droplets, re-immerse the tip below the
surface of water.
The water level inside the tip should remain constant; if the level
goes down there is a leak.
no leak
Figure 10
MyPIPETMAN® multichannel leak test
Chapter 6
CLEANING AND DECONTAMINATION
MyPIPETMAN pipettes are fully autoclavable without disassembly for maximum convenience as well as
protection from contamination.
CLEANING AND DECONTAMINATION
Autoclaving
1. Before autoclaving
Set the pipette volume to the lowest designated usable volume of the model before placing it in the autoclave.
2. Procedure
MyPIPETMAN can be sterilized by steam autoclaving at 121°C (252°F), 1 bar relative pressure for
20 minutes without disassembly.
When autoclaving, the pipette will dry better and faster without using a bag.
3. After autoclaving
Following the autoclaving cycle, let the pipette cool down to room temperature and dry completely before use.
Verify pipette’s functionality according to your procedure.
Chemical Decontamination
You may choose to decontaminate your pipette chemically, in accordance with your own procedures.
Whatever decontaminant you use, check with the supplier to verify that the decontaminant is compatible
with stainless steel and the plastics used in the construction of the pipette: PA (Polyamide), PBT
(Polybutylene Terephtalate), PC (Polycarbonate), PC/PBT (Polycarbonate/ Polybutylene Terephtalate),
POM (Polyoxymethylene), PVDF (Polyvinylidene Fluoride), PP (Polypropylene), PPA (Polyphthalamide), or
PPS (Polyphenylene Sulfide).
MyPIPETMAN®|USER’S GUIDE
14
Page 14
CLEANING AND DECONTAMINATION
NOTE
Lower Part (Volumetric Module)
For Single Channel Models
The following components only can be immersed in a decontaminant solution: connecting nut, tip ejector,
and tip holder.
The piston assembly and seals must be degreased with isopropanol or ethanol in a separate
vessel before being immersed in decontamination solution.
1. Disassemble the pipette
2. Immerse tip ejector, tip holder, and connecting nut in the decontamination solution.
3. Disassemble and degrease the piston assembly (Refer to Servicing Piston Assembly, on page 17)
4. Rinse the connecting nut, tip ejector, and tip holder with distilled water.
5. Let the parts dry by evaporation (or wipe the tip ejector, the tip holder, and connecting nut with a
soft cloth).
6. Lubricate the piston assembly and the seals (Refer to Servicing Piston Assembly, on page 17).
7. Reassemble the piston assembly, the tip holder, and the tip ejector.
For Multichannel Models
The following components only can be immersed in a decontamination solution: tip ejector, ejector locks,
and ejector spacer.
1. Remove the tip ejector and the ejector adapter.
2. Immerse the tip ejector, ejector locks, and ejector adapter in the decontamination solution or wipe
them with a soft cloth or lint-free tissue soaked with the decontamination solution.
3. Rinse the components with distilled water.
4. Wipe the entire pipette with a soft cloth or lint-free tissue soaked with the decontamination solution.
5. Wipe it with distilled water.
6. Let the parts dry by evaporation or wipe them with a clean soft cloth or lint-free tissue.
7. Refit the tip ejector (refer to Fitting the Tip Ejector Adapter on page 8).
15
CLEANING AND DECONTAMINATION| MyPIPETMAN®
Page 15
Chapter 7
MAINTENANCE
Routine maintenance will help keep your pipette in good condition, ensuring a continued high level of
performance.
Visit the Gilson website at www.gilson.com for:
●Complete technical service contact information
●Access to the documentation and information about accessories and related products
For information or technical assistance contact techsupport@pipetman.com
MAINTENANCE
Changing the Tip Ejector for Single Channel
To remove
1. Push the ejection button 1.
2. Turn the tip ejector clockwise 2.
3. Slide and remove the tip ejector 3.
To refit
1. Push the ejection button.
2. Slide the tip ejector along the tip holder.
3. Clip the tip ejector on the body of the pipette.
Changing the Tip Holder for Single Channel
1. Push the ejection button.
2. Turn the tip ejector clockwise.
3. Slide and remove the tip ejector.
4. Unscrew the connecting nut.
Changing the Tip Ejector for Multichannel
To remove
1. Keep both ejector locks depressed 1.
2. Pull the tip ejector down 2.
1
2
3
To refit
1. Gently re-insert the tip ejector vertically into the rails of the ejector support 3.
2. Pull lightly on the tip ejector to check the position.
1
2
3
MyPIPETMAN®|USER’S GUIDE
16
Page 16
NOTE
NOTE
Color-coded
push buton
Locking system
1
Tip ejector
button
P10, P20,
P100, P200
Connecting nut
MAINTENANCE
P1000
2
4
P5000
P10mL
Piston
assembly
2
3
Tip
holder
Tip ejector
Figure 11
MyPIPETMAN® piston assembly (disassembled)
Servicing the Piston Assembly
You may remove the piston assembly for cleaning purposes only. If the piston assembly is changed, the
pipette must be adjusted and calibrated at an authorized Gilson Service Center. As the models P2 and P10
contain miniaturized parts, it is best not to disassemble these pipettes yourself.
Separate the upper part and the lower part
1. Push the ejection button 1.
2. Turn the tip ejector clockwise 2.
3. Slide and remove the tip ejector 3.
4. Unscrew the connecting nut 4.
Carefully separate the tip holder from the upper part (refer to Figure 11)
1. Remove the piston assembly, O-ring, and the seals (refer
to Changing the Tip Ejector for Single Channel on page 16).
2. Leave the piston exposed, clean it with isopropanol or ethanol, and lubricate lightly.
3. Clean, autoclave, or replace the tip holder. If necessary, lightly lubricate the piston and its seals.
For P20, P100, P200, P20 models: lubricate only the useful part of the
piston (20 ±5 mm length) and the O-ring (refer to Figure 10).
For P1000 model: lubricate the piston.
For P5000, P10000 models: disassemble the seals,
lubricate the internal part of the seal, and then lubricate the
useful part of the the piston. Do not lubricate the O-ring.
For P2 and P10 models, contain miniaturized parts.
It is best not to disassemble these pipettes yourself.
4. Reassemble the pipette.
5. Tighten the connecting nut (turn counterclockwise).
6. Refit the tip ejector.
If the piston assembly is changed, the pipette must be adjusted and calibrated in at an
authorized Gilson Service Center.
Gilson warrants this pipette against defects in material under normal use and service for a period of three
years from the date of purchase.
This warranty shall not apply to pipettes which are subject to abnormal use and/or improper or inadequate
maintenance (contrary to the recommendations given in the user’s guide), including, but not limited to
pipettes which have been subjected to physical damage, improper handling, spillage or exposure to any
corrosive environment, abuse misuse or misapplication, or from ordinary wear and tear (including seal and
O-rings). This warranty shall also be void in the event pipettes are altered or modified by any party other
than Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole
option, repair or replacement of any defective components of pipettes or refund of the purchase price paid
for such pipettes.
THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER
WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF
MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE,
UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE,
INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR
IMPLIED WARRANTY.
19
WARRANTY| MyPIPETMAN®
Page 19
NOTES
NOTES
MyPIPETMAN®|USER’S GUIDE
20
Page 20
Specifications subject to change without notification — errors excepted.