Gilson Multi-channel microlitre pipette Pipetman® 8-channel, 2 to 20 µl, P8x20 User guide

Page 1
User’s Guide
EN
G
Page 2
NOTICE
TABLE OF CONTENTS
INTRODUCTION | 3
Parts Checklist | 3
Good Laboratory Practice (GLP) Compliance | 4
Description | 5
Specifications | 6
SETTING THE VOLUME | 8
PIPETTING | 9
Fitting the Tips | 9
TABLE OF CONTENTS
Pre-Wetting the Tips | 11
Aspirating | 11
Dispensing | 12
Guidelines for Good Pipetting | 12
ACCESSORIES | 14
TROUBLESHOOTING | 15
Leak Test | 16
MAINTENANCE | 17
Maintenance for Single Channel Models | 17
Maintenance for Multichannel Models | 21
CLEANING AND DECONTAMINATION | 21
SPARE PARTS | 25
WARRANTY | 28
Decreased pipetting forces are due to the redesigned piston assembly including the use of a very high quality lubricants. The use of other lubricant cancels the warranty of this pipette. The approved lubricant is available under the part number 5440011070. Please contact your Gilson distributor.
2
TABLE OF CONTENTS | PIPETMAN® G
Page 3
Chapter 1
NOTICE
INTRODUCTION
PIPETMAN® G is a fully adjustable air-displacement pipette and is used with disposable tips.
To meet pipette user’s needs, PIPETMAN G relies on the legendary accuracy, precision, and reliability of PIPETMAN, with pipetting and purge forces dramatically decreased for hours of comfortable pipetting.
Eight single channel models cover a volume range from 0.2 µL to 10 mL.
P2G and P10G: measurement and transfer of microliter volumes, DNA sequencing, and enzyme assays
P20G, P100G, P200G, and P1000G: measurement and transfer of general aqueous solutions, acids, and bases
P5000G and P10mLG: measurement and transfer of large volumes
Eight multichannel models cover a volume range from 1 µL to 300 µL.
PIPETMAN G is identified by a visual blue ring.
Parts Checklist
After unpacking the pipette, verify that the following items were included and are undamaged:
Single Models
PIPETMAN G
PIPETMAN G Quick Guide
Safety bag
Dual-position adapter (for P2G and P10G models only)
Certificate of Conformity (including barcode sticker)
Multichannel Models
PIPETMAN G
PIPETMAN G Quick Guide
Safety bag
Ejector spacers for D10 tips (for PX10G model only)
Certificate of Conformity (including barcode sticker)
INTRODUCTION
PIPETMAN® G | INTRODUCTION
3
Page 4
Good Laboratory Practice (GLP) Compliance
The serial number is engraved on the body of the pipette. It provides unique identification of your pipette and the date of manufacture.
Ex: AA10369
The barcode on the box and the Certificate of Conformity provide traceability of your pipette.
INTRODUCTION
Serial number
Figure 1
Serial number location
4
INTRODUCTION | PIPETMAN® G
Page 5
Description
Color-coded push button
Tip ejector button
Thumbwheel
Blue ring
Body or
handle
Clip tip ejector Connecting nut
Ejector support
Ejector lock
Cover
Tip ejector
Tip holder
Ejector spacer
PIPETMAN®
DIAMOND tip
Figure 2
PIPETMAN® G multichannel and single channel models
INTRODUCTION
Please refer to the following chapters for a full description of the different parts and functions of the pipette.
PIPETMAN® G | USER’S GUIDE
5
Page 6
Specifications
NOTICE
PIPETMAN G is a high quality pipette that offers excellent accuracy and precision. The data provided in the Gilson Maximum Permissible Errors table were obtained using PIPETMAN® DIAMOND Tips. These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used.
Each pipette is inspected and validated by qualified technicians in accordance with the Gilson Quality System. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655 standard by type testing.
SPECIFICATIONS
The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6).
The data given in the tables conform to the ISO 8655-2 standard. With a precise pipetting technique, the P2G model may be used to aspirate volumes as low as 0.1 µL and the P10G model as low as 0.5 µL .
PIPETMAN G is equipped with a stainless steel tip ejector and can also be used with a plastic tip ejector, which is available as an accessory (refer to ACCESSORIES on page 14).
6
INTRODUCTION | PIPETMAN® G
Page 7
Maximum Permissible Errors
MAXIMUM PERMISSIBLE ERRORS
VOLUM E (µL)
PIPE TMAN G SI NGLE CH ANNE L
P2G (P/N F1 44054 M) WITH D10 O R DL10 TIP S
0.2
0.5
1 2
P10G (P/N F 14405 5M) WITH D1 0 OR DL10 TI PS
1 5 10
P20G (P/N F 14405 6M) WITH D2 00 TIPS
2 10 20
P100G ( P/N F1440 57M) WITH D2 00 TIPS
10 50 100
P200G ( P/N F1440 58M) WIT H D200 TI PS
20 100 200
P100 0G (P/N F144 059M) WI TH D100 0 TIPS
100 500 1000
P5000 G* (P/N F144 066) WI TH D500 0 TIPS
500 2500 5000
P10M LG* (P/N F14 4067) WI TH D10M L TIPS
1000 5000 10000
PIPETMAN G MULTICHANNEL
P8X10 G (P/N F1440 68) P12X10G (P/N F 144069)
WITH D10 A ND DL10 TI PS
1 5 10
P8X20 G (P/N F14407 0) P12X2 0G (P/N F14 4071)
WITH DL1 0 TIPS
2 10 20
P8X20 0G (P/N F14 4072) P12 X200G ( P/N F14407 3)
WITH D20 0 TIPS
20 100 200
P8X300 G (P/N F144 074) P12X300 G (P/N F144 075)
WITH D30 0 TIPS
30 150 300
* P5000G and P10mLG do not have tip ejectors
GILSON ISO 865 5
SYSTEMATIC ERROR (µL)
± 0.024 ± 0.025 ± 0.025 ± 0.030
± 0.025 ± 0.075 ± 0.100
± 0.10 ± 0.10 ± 0.20
± 0.35 ± 0.40 ± 0.80
± 0.50 ± 0.80 ± 1.60
± 3.0 ± 4.0 ± 8.0
± 12 ± 15 ± 30
± 30 ± 40 ± 60
± 0.08 ± 0.20 ± 0.20
± 0.10 ± 0.20 ± 0.40
± 0.50 ± 1.00 ± 2.00
± 1.00 ± 1.50 ± 3.00
RANDOM ERROR (µL)
0.012 0.012 0.012 0.014
0.012 0.030 0.040
0.030 0.050 0.060
0.10 0.12 0.15
0.20 0.25 0.30
0.6 1.0 1.5
3 5 8
6 10 16
0.05 0.10 0.10
0.08 0.10 0.20
0.25 0.40 0.50
0.35 0.60 1.00
SYSTEMATIC ERROR (µL)
± 0.08 ± 0.08 ± 0.08 ± 0.08
± 0.12 ± 0.12 ± 0.12
± 0.2 ± 0.2 ± 0.2
± 0.8 ± 0.8 ± 0.8
± 1.6 ± 1.6 ± 1.6
± 8.0 ± 8.0 ± 8.0
± 40 ± 40 ± 40
± 60 ± 60 ± 60
± 0.24 ± 0.24 ± 0.24
± 0.4 ± 0.4 ± 0.4
± 3.2 ± 3.2 ± 3.2
± 8.0 ± 8.0 ± 8.0
PIPETMAN® G | USER’S GUIDE
RANDOM ERROR (µL)
0.04 0.04 0.04 0.04
0.08 0.08 0.08
0.1 0.1 0.1
0.3 0.3 0.3
0.6 0.6 0.6
3.0 3.0 3.0
15.0 15.0 15.0
30.0 30.0 30.0
0.16 0.16 0.16
0.20 0.20 0.20
1.2 1.2 1.2
3.0 3.0 3.0
SPECIFICATIONS
7
Page 8
Chapter 2
Px10G
Px20G
Chapter 2
SETTING THE VOLUME
The volume of liquid to be aspirated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, depending on the model (refer to Figure 3).
P2G
P10G
The volume is set by turning the thumbwheel or the push button (refer to Figure 4). The push button makes it easier and quicker to
1.25 µL
P100G
75 µL
set volumes, especially when wearing gloves.
SETTING THE VOLUME
The thumbwheel may be turned using only one hand to slowly reach the
Figure 3
Dial colors by model
required setting.
MODEL
P2G µL 0.01 µL 0.002 µL
P10G - P20G - PX10G - PX20G µL 0.1 µL 0.02 µL
P100G - P200G - PX200G - PX300G µL - 0.2 µL
P1000G 0.01 mL mL 0.002 mL
P5000G 0.01 mL mL 0.002 mL
P10mLG mL 0.1 mL 0.02 mL
COLOR OF VOLUMETER NUMBERS
BLAC K RED INCREMENT
1
2 5
0 7 5
P5000G
1
2 5
1.25 mL
7.5 µL
P200G
125 µL
0 7
5
1 2 5
P20G
12.5 µL
P1000G
0.75 mL
P10mLG
0 7
5
7.5 mL
1 2
0 7
5
7.5 µL
Px200G
1 2 5
125 µL
1 2
5
12.5 µL
Px300G
1 2 5
125 µL
5
0
7 5
To achieve maximum accuracy when setting the volume, proceed as follows:
When decreasing the volume setting, slowly reach the required setting, making sure to not overshoot the mark.
When increasing the volume setting, pass the required value by 1/3 of a turn and then slowly decrease to reach the volume, making sure to not overshoot the mark.
8
SETTING THE VOLUM E | PIPETMAN® G
Push button
Thumbwheel
Figure 4
Location of push button and thumbwheel
Page 9
Chapter 3
NOTE
NOTE
PIPETTING
While many brands of tips can be used with the PIPETMAN G, it is recommended to use PIPETMAN DIAMOND Tips with the PIPETMAN G for optimum performance. These tips are made from pure polypropylene. Plastic tips are for a single application—they should not be cleaned for reuse.
Fitting the Tips
To fit a new PIPETMAN DIAMOND Tip, push the tip holder into the tip using a slight twisting motion to ensure a firm, airtight seal.
Fitting the Tips on P2G and P10G Models
For the P2G and P10G models, a dual­position adapter (plastic) is required to fit DL10 tips (longtips) or D10 tips (short tips). The metallic rod of the tip ejector is shaped so the adapter may be clipped to it in e it he r position.
P2G and P10G models are delivered with the adapter in place, ready to use DL10 tips. If D10 tips are used, the adapter must be repositioned in the shorter slot as follows:
1. Pull the adapter down from the metallic rod.
2. Turn the adapter 180°.
3. Refit the adapter so the end of the metallic rod engages the shorter slot of the adapter.
The dual-position adapter is autoclavable.
Short tips
Figure 5
Dual-position adapter for P2G and P10G with stainless steel tip ejector
Long tips
PIPETTING
PIPETMAN® G | USER’S GUIDE
9
Page 10
An autoclavable tip ejector extension can be ordered as an accessory for P2 and P10 equipped with a plastic tip ejector (part number F107027; refer to ACCESSORIES on page 14).
To fit a tip ejector extension:
1. Slide the extension over the tip holder.
2. Push the extension firmly onto the end of the tip ejector until it clicks into place.
PIPETTING
To remove a tip ejector extension:
1. Gently twist the extension.
2. Pull it away from the pipette.
Ejector Spacer for PIPETMAN G 10 µL Multichannel Models
According to the tip used, D10 or DL10, you may have to exchange the ejector spacer. The broad one is dedicated to D10, and the small one is dedicated to DL10.
Remove the tip ejector, keeping both ejector locks depressed. Pull the tip ejector down.
Gently press the tabs from the ejector spacer and remove it from the tip ejector.
Insert the alternative ejector spacer and click it to the tip ejector.
To refit the tip ejector, gently re-insert the tip ejector vertically into the rails of the ejector support.
spacer
Ejector
Figure 6
Ejector spacer location
Ejector spacer
D10
Short tipsLong tips
PIPETMAN DIAMOND Tips are best fitted with the ROCK Y RACK technique, invented by Gilson, available only in our TIPACK and TOWERPACK.
PIPETMAN DIAMOND Tip Compatibility
for Multichannel Models
Figure 7
ROCKY RACK technique
10
PIPETTING | PIPETMAN® G
P8x10G, P12x10G D10*, DL10, DF10, DFL10
P8x20G, P12x20G DL10, DFL10, DF30, D200
P8x200G, P12x200G D200, D300, DF200, DF300
P8x300G, P12x300G D200, D300, DF200, DF300
* Using the broad ejector spacer D10, you can adapt a D10 tip (see above).
Figure 8
PIPETMAN DIAMOND Tips compatibility charts for multichannel models
Page 11
Pre-Wetting the Tips
Pre-wetting the tips before pipetting helps prepare the tips for the best pipetting performance. Ideally, the pre-wet includes both immersing the tip in the liquid and performing one pipetting step.
Pre-wetting the tips helps ensure that volumes that you pipette will achieve accuracy and precision within specifications.
1 2
Aspirating
1. Press the push button to the first stop (this corresponds to the set volume of liquid).
2. Hold the pipette vertically and immerse the tip in the liquid (refer to Guideline for Good
Pipetting on page 12).
3. Release the push button slowly and smoothly (to the top position) to aspirate the set volume of liquid. Wait one second (refer to
Guidelines for Good Pipetting on page 12)
and then withdraw the pipette-tip from the liquid.
4. You may wipe any droplets away from the outside of the tip using a medical wipe, however, if you do so take care to avoid touching the tip’s orifice.
PIPETTING
Rest position
First stop
Second stop
Figure 9
Pipetting motion - aspirate and dispense
PIPETMAN® G® | USER’S GUIDE
11
Page 12
Dispensing
1. Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
2. Press the push button slowly and smoothly to the first stop.
3. Wait for at least one second, then press the push button to the second stop to expel any residual liquid from the tip. Keep the push button pressed fully down and (while
PIPETTING
removing the pipette) draw the tip along the inside surface of the vessel.
4. Release the push button smoothly. Eject the tip by pressing firmly on the tip ejector button.
Guidelines for Good Pipetting
Make sure that you operate the push button slowly and smoothly.
When aspirating, keep the tip at a constant depth below the surface of the liquid (refer to the table below).
IMMERSIO N DEPT H AND WAIT TIME
MODEL
P2G 1 1
P10G - Px10G 1 1
P20G - Px20G 2-3 1
P100G - Px200G - Px300G 2-4 1
P200G 2-4
P1000G 2-4
P5000G 3-6
P10mLG 5-7 4-5
IMMERSION DEPTH (MM)
WAIT TIME (SECONDS)
1
2-3
4-5
12
PIPETTING | PPIPETMAN® G
Page 13
Change the tip before aspirating a different
NOTICE
liquid, sample, or reagent.
Change the tip if a droplet remains at the end of the tip from the previous pipetting operation.
Pre-wet each new tip with the liquid to be pipetted
Liquid should never enter the tip holder.
To prevent this:
Press and release the push button slowly
and smoothly. Never turn the pipette upside down. Never lay the pipette on its side when
there is liquid in the tip.
If you use the same tip with a larger volume, pre-rinse the tip.
For volatile solvents, you should saturate the air cushion of your pipette by aspirating and dispensing the solvent repeatedly before aspirating the sample.
When the temperature of the liquid is different from the ambient temperature, pre-wet the tip several times before use.
You may remove the tip ejector (refer to
MAINTENANCE on page 17) to aspirate from
very narrow tubes.
After pipetting acids or other corrosive liquids that emit vapors, remove the tip ejector and the tip holder; and then rinse, dry, and lubricate the piston (refer to MAINTENANCE on page 17). For the model P1000G, you can increase the lifetime of the piston by using a specific tip holder equipped with a filter (refer to ACCESSORIES on page 14).
Do not pipette liquids with temperatures above 70°C or below 4°C. The pipette can be used between 4°and 40°C, but the specifications may vary according to the temperature (refer to Maximum Permissible
Errors on page 7).
PIPETTING
PIPETTING
It is recommended to regularly verify the volume displayed on the dial while pipetting.
PIPETMAN® G | USER’S GUIDE
13
13PIPETTING | PPIPETMAN® G
Page 14
Chapter 4
ACCESSORIES
To make pipetting more comfortable and more secure, Gilson has developed several accessories.
DESCRIPTION PART NUMBER
Plastic ejector P2/P10 with adapter F107027
Plastic ejector P20 F107028
Plastic ejector P100 F107029
ACCESSORIES
Plastic ejector P200 F107030
Plastic ejector P1000 F107031
Pipette stands allow users to store pipettes vertically to avoid the possibility of liquid running back into the pipette.
DESCRIPTION PART NUMBER
CARROUSEL™ Pipet te Stand ( 7 pipet tes) F161401
TRIO™ Stand (3 pipet tes) F161405
SINGLE™ Pip ette Holder F16140 6
COLORIS™ Identification Clips identify or personalize your pipette. Write your name or application to avoid mix-ups and then snap into place.
DESCRIPTION PART NUMBER
COLORIS clips (mixed colors set of 10) F161301
COLORIS clips (red, set of 10)
COLORIS clips (ye llow, set of 1 0) F1 61303
COLORIS clips (green , set of 10) F161304
COLORIS clips (blue, s et of 10) F161305
COLORIS clips (white, set of 1 0) F161306
F161302
With The JIMMY ™, hands-free microtube opener, you can open both snap-cap and screw-cap microtubes to increase pipetting efficiency and reduce the fatigue associated with manual tube opening.
DESCRIPTION PART NUMBER
THE JIMMY (set of 3) F144983
To protect the piston when pipetting corrosive liquids, you can use a specific tip holder and filter for the model P1000G.
DESCRIPTION PART NUMBER
Corrosion Protection Kit (tip holder + a bag of 10 filte rs)
14
ACCESSORIES | PIPETMAN® G
F144570
Page 15
Chapter 5
NOTE
TROUBLESHOOTING
A quick inspection of the pipette may help you to detect a problem.
You can download the Two Minute Inspection Poster from the Gilson website (www.gilson.com). The poster describes how to perform a quick diagnosis of your pipette.
The following table can help you to identify and correct the problem you might encounter.
SYM PTOM POSSIBLE CAUSE ACTION
SINGLE CHANNEL
Pipette is leaking sample
Pipette won’t aspirate
Pipette is inaccurate
Pipette is not precise
Tips fall o or do not fit correctly
Pipetting seize up Piston need lubricant Lubricate piston assembly
MULTICHANNEL MODELS
Tips fall o or do not fit correctly
Pipette won’t aspirate
Pipette is inaccurate
Pipette is not precise
If you can not solve the prob lem, contact y our local Gils on representa tive.
Damaged tip holder Worn O-ring or seal
Worn O-ring and seal Damaged tip holder Connecting nut is loose Damaged or corroded piston Improper repair or assembly
Improper repair or assembly Unscrewed tip holder Connecting nut is loose
Tip holder is loose Connecting nut is loose Incorrect operator technique Damaged or corroded piston(s) Damaged tip holder(s) Worn O-ring or seal
Low quality tips Dirty tip holder Damaged tip holder(s) Damaged tip ejector Ejector spacer is missing The ejector spacer is damaged The tip ejector is loose The ejector lock is misaligned
Low quality tips Damaged tip ejector Damaged ejector spacer Dirty tip holder(s) Damaged tip holder
Connecting nut is loose Tighten the connecting nut
Connecting nut is loose Tighten the connecting nut
Connecting nut is loose Incorrect operator technique
Replace the tip holder Replace both parts and lubricate
Replace both parts and lubricate Replace the tip holder Tighten connecting nut Return pipette to supplier (refer to
MAINTENANCE on page 17)
Refer to page 17 - MAINTENANCE Tighten connecting nut Tighten connecting nut
Tighten connecting nut Tighten connecting nut Operator training Return pipette to supplier Replace the tip holder Replace both parts and lubricate
Use PIPETMAN DIAMOND Tips Clean the tip holder with alcohol Replace the tip holder Replace tip ejector Mount the spacer on the tip ejector Replace the ejector spacer Assemble the tip ejector properly Align the ejector lock
Use PIPETMAN DIAMOND Tips Replace the tip ejector Replace the ejector spacer Clean the tip holder(s) with isopropanol or ethanol Contact your local authorized Gilson Service Center
Tighten the connecting nut Operator training
TROUBLESHOOTING
PIPETMAN® G | USER’S GUIDE
15
Page 16
CAUTION
Before returning any pipette to your local Gilson Service Center, ensure that it is completely free of chemical, biological, or radioactive contamination (refer to CLEANING
AND DECONTAMINATION on page
22). Please use the included safety bag to return the pipette to your local Gilson Service Center.
Leak Test
This test may be performed at any time to check that the pipette does not leak, especially after performing a maintenance or decontamination procedure. If a pipette fails this test, replace the
TROUBLESHOOTING
O-ring and seal. After making sure that the pipette is correctly reassembled, repeat this test.
For the P2G to P200G Models
1. Fit a PIPETMAN DIAMOND Tip.
2. Set the pipette to the maximum volume given in the specifications and pre-wet.
3. Aspirate the set volume from a beaker of dist il led water.
4. Maintain the pipette in the vertical position and wait for 20 seconds.
If a water droplet appears at the end of the tip, there is a leak.
If you see no droplet, re-immerse the tip below the surface of water. The water level inside the tip should remain constant. If the level goes down, there is a leak.
For the P1000G, P5000G, and P10mLG Models
1. Fit a PIPETMAN DIAMOND Tip.
2. Set the pipette to the maximum volume given in the specifications.
3. Aspirate the set volume from a beaker of dist il led water.
4. Maintain the pipette in the vertical position and wait for 20 seconds.
If a water droplet appears at the end of the tip, there is a leak.
TROUBLESHOOTING | PIPETMAN® G
16
Page 17
NOTICE
For the Multichannel Models (PX8G PX12G)
1. Fit the PIPETMAN DIAMOND Tip.
2. Set the pipette to the maximum volume given in the specifications and pre-wet.
3. Aspirate the set volume from a reagent reservoir of distilled water.
4. Maintain the pipette in the vertical position and wait for 20 seconds. Fluid level in the tips should remain constant.
If a water droplet appears at the end of the tip, there is a leak.
For volumes below 200 μL, if you see no droplets, re-immerse the tip below the surface of water. The water level inside the tip should remain constant. If the level goes down, there is a leak.
Chapter 6
MAINTENANCE
Routine maintenance will help keep your pipette in good condition, ensuring a continued high level of performance.
Maintenance is limited to:
Cleaning or decontamination (refer to
CLEANING AND DECONTAMINATION on
page 22).
Replacing spare parts
Greasing the piston assembly
MAINTENANCE
MAINTENANCE
Maintenance for Single Channel Models
PIPETMAN P2G and P10G should not be disassembled, so you can only replace the push
button, tip ejector, and its dual position adapter. With these pipettes if the tip holder is damaged, the piston may also be damaged.
After replacing any parts, you should verify the performance of your pipette following the verification procedure available on the Gilson website (www.gilson.com). If the pipette needs to be readjusted, please contact your local Gilson Service Center.
PIPETMAN® G | USER’S GUIDE
17
Page 18
Changing the Tip Ejector for Single Channel Models
To remove:
1
Push the ejection button.
Laterally push the tip ejector.
2
Slide and remove the
3
tip eje ctor.
To r efit :
Push the ejection
1
MAINTENANCE
button.
Slide the tip ejector
2
along the tip holder.
Clip the tip ejector on
3
the body of the pipette.
Changing the Tip Holder – No Tools Required
1. Remove the tip ejector (refer to Changing
the Tip Ejector for Single Channel Models
on page 18).
2. Unscrew the connecting nut by turning it counterclockwise.
3. Carefully separate the lower and upper parts.
4. Remove the piston assembly and the seals.
5. Clean, autoclave, or replace the tip holder.
6. If necessary, lightly lubricate the piston and its seals (refer to Servicing the Piston
Assembly on page 19).
7. Reassemble the pipette (refer to the Figure 10 on page 19).
8. Tighten the connecting nut (turn clockwise)
9. Refit the tip ejector (refer to Changing the Tip
Ejector for Single Channel Models on page 18).
1
2
3
MAINTENANCE | PIPETMAN® G
18
Page 19
NOTICE
P10G, P20G
P100G, P200G
Tip holder
P1000G
Color-coded push button
Blue ring
Connecting nut
Piston
assembly
P5000G P10mLG
Tip ejector
MAINTENANCE
Stainless steel
Plastic tip ejector available as an acessory
Figure 10
Disassembled PIPETMAN G, with dif ferent piston assemblies
Servicing the Piston Assembly
You may remove the piston assembly for cleaning purposes only. If the piston assembly is changed, the pipette must be adjusted and calibrated at a local Gilson Service Center. Because P2G and P10G models contain miniaturized parts, it is best not to disassemble these pipettes yourself.
The piston assembly must not be autoclaved.
PIPETMAN® G | USER’S GUIDE
19
Page 20
1. Remove the tip ejector (refer to
NOTICE
Changing the Tip Ejector for Single Channel Models on page 18).
2. Unscrew the connecting nut by turning it counterclockwise.
3. Carefully separate the lower and upper parts.
4. Remove the piston assembly and the seals.
5. Leave the piston exposed, clean it with isopropanol or ethanol, and lubricate lightly. For P100G, P200G,
MAINTENANCE
and P1000G models, lubricate only the internal part of the piston and the seals. For P5000G and P10mLG, lubricate the tip of the piston, disassemble the seals, and lubricate their internal part.
6. Reassemble the pipette (refer to Figure 10 on page 19).
7. Tighten the connecting nut (turn clockwise).
8. Refit the tip ejector (refer to Changing the Tip
Ejector for Single Channel Models on page 18).
Changing the Seal Guide (or Seal)
The O-ring and seal on the piston must not be autoclaved. If worn or damaged in any way (chemical or mechanical), they must be replaced. Because the models P2G and P10G contain miniaturized parts, it is best not to disassemble these pipettes yourself. Please contact your local Gilson Service Center. The dimensions of the O-ring vary depending on the model of pipette.
1. Remove the tip ejector (refer to Changing the Tip
Ejector for Single Channel Models on page 18).
2. Unscrew the connecting nut by turning it counterclockwise.
3. Carefully separate the lower and upper parts.
4. Remove the piston assembly, O-ring, and seal.
5. If necessary, clean the piston and replace the seal. Lightly lubricate both seal and piston. Remember to replace the seal and O-ring in the correct order.
6. Reassemble the pipette (refer to the Figure 10 on page 18).
7. Tighten the connecting nut (turn clockwise).
8. Refit the tip ejector (refer to Changing the Tip
Ejector for Single Channel Models on page 18).
MAINTENANCE | PIPETMAN® G
20
Decreased pipetting forces are due to the design of the piston assembly, which includes a very high-quality lubricant (part number 5440011070). The use of any other lubricant voids the warranty of this pipette.
Page 21
Maintenance for Multichannel Models
Routine maintenance will help keep your pipette in good condition, ensuring a continued high level of performance.
Maintenance is limited to:
Cleaning or decontamination (refer to
CLEANING AND DECONTAMINATION on
page 22)
Replacing spare parts
Greasing the piston assembly
Changing the Tip Ejector for Mutichannel Model
To remove the tip ejector, keep both ejector locks depressed. Pull the tip ejector down.
To refit the tip ejector, gently re-insert the tip ejector vertically into the rails of the ejector support. Pull lightly on the tip ejector to check the position.
MAINTENANCE
Tip ejector
Ejector lock
Ejector spacer
Figure 11
PIPETMAN® G multichannel tip ejector and ejector spacers
PIPETMAN® G | USER’S GUIDE
21
Page 22
Chapter 7
NOTICE
CLEANING AND
DECONTAMINATION
PIPETMAN G is designed so that the parts normally in contact with liquid contaminants can easily be cleaned and decontaminated. However, because the models P2G and P10G contain miniaturized parts, it is best not to disassemble these pipettes yourself. Please contact your local Gilson Service Center.
You can refer to the decontamination procedure available on the Gilson website (www.gilson.com). Liquid must
never enter the upper part (handle) of any pipette.
Cleaning for Single Channel Models
The pipette must be cleaned, as described below, before it is decontaminated. Soap solution is
CLEANING AND DECONTAMINATION
recommended for cleaning PIPETMAN G.
External
1. Remove the tip ejector (refer to Changing the Tip
Ejector for Single Channel Models on page 18).
2. Wipe the tip ejector with a soft cloth or lint-free tissue soaked with soap solution.
3. Wipe the entire pipette with a soft cloth or lint-free tissue soaked with soap solution to remove all dirty marks. If the pipette is very dirty, a brush with soft plastic bristles may be used.
4. Wipe the entire pipette and the tip ejector with a soft cloth or lint-free tissue soaked with dist il led water.
5. Refit the tip ejector and allow the pipette to dry.
Internal
Only the following components can be immersed in a cleaning solution: connecting nut, tip ejector, tip holder, piston assembly, seal, and O-ring.
1. Disassemble the pipette (refer to Figure 10 on page 19).
2. Set aside the upper part in a clean, dry place. Clean the individual components of the lower part of the pipette using an ultrasonic bath (20 minutes at 50°C) or with a soft cloth and brushes.
CLEANING AND DECONTAMINATION | PIPETMAN® G
22
Page 23
NOTE
The piston assembly and seals must be degreased with isopropanol or ethanol before being immersed in the ultrasonic bath. Small round brushes with sof t plastic bristles may be used to clean the interior of the tip holder.
3. Rinse the individual components with distilled water.
4. Let the parts dry by evaporation or wipe them with a clean soft cloth or lint-free tissue.
5. Reassemble the pipette as described (refer to Figure 10 on page 19).
Cleaning for Multichannel Models
Only the following components can be immersed in a cleaning solution: tip ejector, ejector locks, and ejector spacer.
1. Remove the tip ejector and the ejector spacer.
2. Immerse the tip ejector, ejector locks, and
ejector spacer in the cleaning solution or wipe them with a soft cloth or lint-free tissue soaked with the cleaning solution.
3. Rinse the components with distilled water.
4. Wipe the entire pipette with a soft cloth
or lint-free tissue soaked with the cleaning solution.
5. Wipe it with distilled water.
6. Let the parts dry by evaporation or wipe them
with a clean soft cloth or lint-free tissue.
7. Refit the tip ejector (refer to Changing the Tip
Ejector for Multichannel Models on page 21).
CLEANING AND DECONTAMINATION
Autoclaving
The upper part (body) and the piston assembly of the pipette are not autoclavable. Only the following parts may be autoclaved: tip ejector, tip holder, and connecting nut. The O-ring and seal are not autoclavable. They may be cleaned or replaced with the ones found in SPARE PARTS on page 25.
1. Clean the parts to be autoclaved, especially
the tip holder.
2. Put the parts in an autoclaving sack.
3. Autoclave for 20 minutes at 121°C, 0.1 MPa.
4. Check that the parts are dry before
reassembling the pipette.
5. Set the pipette aside to equilibrate to room
temperature.
6. Reassemble the pipette (refer to Figure 10 on
page 19).
PIPETMAN® G | USER’S GUIDE
23
Page 24
Chemical Decontamination
You may choose to decontaminate your pipette chemically, in accordance with your own procedures. Whatever decontaminant you use, check with the supplier of the decontaminant to verify that it is compatible with stainless steel and the plastics used in the construction of the pipette: PA (Polyamide), PBT (Polybutylene Terephtalate), PC (Polycarbonate), PC/PBT (Polycarbonate/Polybutylene Terephtalate), POM (Polyoxymethylene), or PVDF (Polyvinylidene Fluoride).
Chemical Decontamination for Single Channel Models
Upper Part (handle)
1. Wipe the upper part (handle) of the pipette with a soft cloth or lint-free tissue soaked with the chosen decontaminant.
2. Wipe the upper part of the pipette with a soft cloth or lint-free tissue soaked with distilled water or sterile water.
Lower Part (volumetric module)
CLEANING AND DECONTAMINATION
Only the following components can be immersed in a decontaminant solution: connecting nut, tip ejector, and tip holder.
The piston assembly and seals must be degreased with isopropanol or
NOTE
1. Remove the tip ejector, the tip holder, and the connecting nut.
2. Immerse them in the decontamination solution.
3. Degrease the piston assembly and the seals. Then immerse them in another vessel of decontamination solution.
4. Rinse each component with distilled water.
5. Let the parts dry by evaporation. The tip ejector, tip holder, and connecting nut can be wiped dry with a soft cloth.
6. Lubricate the piston assembly and the seals.
7. Reassemble the piston assembly, the tip holder, and the tip ejector.
ethanol in a separate vessel before being immersed in decontamination solution.
Chemical Decontamination for Multichannel Models
Only the following components be immersed in a decontamination solution: tip ejector, ejector locks, and ejector spacer.
24
CLEANING AND DECONTAMINATION | PIPETMAN® G
Page 25
1. Remove the tip ejector and the ejector spacer.
2. Immerse the tip ejector, ejector locks and ejector spacer in the decontamination solution or wipe them with a soft cloth or lint-free tissue soaked with the decontamination solution.
3. Rinse the components with distilled water.
4. Wipe the entire pipette with a soft cloth or lint-free tissue soaked with the decontamination solution.
5. Wipe the entire pipette with distilled water.
6. Let the parts dry by evaporation or wipe them with a clean soft cloth or lint-free tissue.
7. Refit the tip ejector (refer to Changing the Tip
Ejector for Multichannel Models, page 21).
Chapter 8
SPARE PARTS
Service Kit 1
st
level includes:
3 piston seals* or seal guides
3 O-rings*
1 tip holder
C
D
E
Service Kit 2nd level includes:
1 push button
1 connecting nut
A
B
Stainless steel tip ejector
F2
Plastic tip ejector F1 available separately as an accessory
For the P2G and P10G models only:
5 dual-position adapter for stainless steel tip ejector
1 extension for plastic tip ejector
F4
F3
A
B
C
D
SPARE PARTS
* For P50 00G, t wo piston
seals an d O-rings are
provide d on the Ser vice
Kit 1st level.
F4
Figure 12
Spare parts identified
F2
F3
F1
PIPETMAN® G | USER’S GUIDE
F2
E
25
Page 26
Single Channel Models
P2G (F 144054M) AND P1 0G (F 144055M)
LABEL DESCRIPTION
C+D+E Service Kit 1st level F144501 FA07001
A+B+F1 to F4 Service Kit 2nd level F1619702 F1619712
C+D
F3
SPARE PARTS
F4 Tip ejector extension F2070903 F2070903
P20G ( F14 4056 M) AND P 100 G (F14 4057M)
LABEL DESCRIPTION
C+D+E Service Kit 1st level FA07002 FA07003
A+B+F Service Kit 2nd level F1619722 F1619732
C+D
P200 G (F14405 8M) AND P1000G ( F14 4059 M)
LABEL
LABEL DESCRIPTION
C+D+E Service Kit 1st level FA07004 FA07005
A+B+F Service Kit 2nd level F1619742 F1619782
C+D
Seal Guide + O-ring (5 sets)
Dual-Position Adapter (x5)
Seal Guide + O-ring (5 sets)
Seal Guide + O-ring (5 sets)
P/N F OR P2G
F144861 FA07012
F144879 F144879
P/N F OR P20G
FA07013 FA07014
P/N F OR P200G
FA07015 FA07016
P/N F OR P10G
P/N F OR P100G
P/N F OR P1000G
P500 0G (F 144066) AND P10MLG (F 144067)
LABEL DESCRIPTION
C+D+E Service Kit 1st level FA07021 FA07022
C+D
E Tip holder F123608 F161263
DESCRIPTION PART NUMBER
Lubricant 5440011070
SPARE PART S | PIPETMAN® G
26
Seal Guide + O-ring (2 sets)
P/N F OR P5000G
FA07017 FA07018
P/N F OR P10MLG
Page 27
Multichannel Models
F4
Tip ejector (12 channels)
Tip ejector (8 channels)
F3
Ejector lock
G
SPARE PARTS
Ejector spacer (8 channels)
Ejector spacer D10
H4
(12 channels)
Ejector spacer
H2
(12 channels)
H3
Ejector spacer D10 (8 channels)
P8X1 0G (F1440 68) AN D P12X10G( F14 4069)
LABEL DESCRIPTION
F3-F4 Tip ejector F507005 F507006
G Ejector lock F507008 F507008
H1 - H2 Ejector spacer F507001 F507003
H3 - H4 Ejector spacer D10 F807114 F807115
P/N F OR P8X10G
P/N F OR P12X10G
P8X20G (F1 44070) AN D P12 X20G(F14 4071)
LABEL DESCRIPTION
F3-F4 Tip ejector F507005 F507006
G Ejector lock F507008 F507008
H1 - H2 Ejector spacer F507001 F507003
P/N F OR P8X20G
P/N F OR P12X20G
P8X200G (F 144 072) AND P 12X 200G(F14 4073)
LABEL DESCRIPTION
F3-F4 Tip ejector F507005 F507006
G Ejector lock F507008 F507008
H1 - H2 Ejector spacer F507001 F507003
P/N F OR P8X200 G
P/N F OR P12X200G
H1
P8X30 0G (F 144074) AND P12X300 G(F14 4075)
LABEL DESCRIPTION
F3-F4 Tip ejector F507005 F507006
G Ejector lock F507008 F507008
H1-H2 Ejector spacer F507001 F507003
P/N F OR P8X300G
P/N F OR P12X300G
PIPETMAN ® G | USER ’S GUIDE
27
Page 28
Chapter 9
NOTICE
WARRANTY
Gilson warrants this pipette against defects in material under normal use and service for a period of three years for both single channel and multichannel models from the date of purchase.
This warranty shall not apply to pipettes which
WARRANTY
are subject to abnormal use and/or improper or inadequate maintenance (contrary to the recommendations given in the user’s guide), including, but not limited to pipettes which have been subjected to physical damage, improper handling, or spillage or exposure to any corrosive environment. This warranty shall also be void in the event pipettes are altered or modified by any party other than Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes. Routine cleaning, control,
and recalibration are not covered under the warranty. The replacement of wearing parts such as seals, O-rings, broken pistons assembly, and broken tip holders are not covered under the warranty.
THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE, UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE, INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR IMPLIED WARRANTY.
28
WARRANTY | PIPETMAN® G
Yearly rou tine ma inten ance is highly recommended to keep your pipette in good condition, ensuring a continued high level of performance.
Page 29
NOTES
NOTES
PIPETMAN® G | USER’S GUIDE
29
Page 30
gilson.com/contactus
LT801122/F | ©2021 Gilson, Inc. All rights reser ved.