Gilson 215 User Manual

Quad-Z 215 Liquid Handler
User's Guide
LT2545/©2003 Gilson, Inc. All rights reserved
April 2003

Table of Contents

Declaration of Conformity
1 Introduction
Safety Precautions ...............................................................1-1
Unpacking.............................................................................1-3
Standard Equipment ....................................................1-4
Accessories .....................................................................1-4
Customer Service ................................................................. 1- 5
Technical Specifications .....................................................1-6
2 Installation
Locator Plate Installation ...................................................2-2
Armlock Removal ................................................................ 2- 3
Z-Arm Installation...............................................................2-4
Installing the Z-Arm Cable Support Rod ................. 2 -5
Probe Installation ................................................................. 2 -6
Installing the Probe Guide Inserts.............................. 2 - 6
Installing the Probes .....................................................2 - 6
Plumbing Connections ........................................................2-7
Transfer Tubing Installation ....................................... 2 -7
Installing the Level Sensing Cables............................2 - 8
Rinse Station and Drain Waste Tubing Installation.. . 2 - 9
Rack Setup ............................................................................ 2-11
Code 200-Series and Code 500-Series Racks ...........2-11
Code 200-Series and Code 20-Series or
Code 30-Series Racks .................................................... 2-12
Only Code 20-Series or Code 30-Series Racks......... 2-13
Final Z-Arm Adjustment ................................................... 2-14
Electrical Connections ........................................................ 2-16
Rear Panel.......................................................................2-16
Input/Output Ports...................................................... 2-17
RS-232 Port .....................................................................2-19
GSIOC Port ..................................................................... 2-20
Unit ID and Baud Rate/Mode Selection .................. 2-21
Fuses ................................................................................2-23
Power Cord Connection .............................................. 2-24
3 Operation
Front Panel ............................................................................ 3 -2
Start Button ....................................................................3-2
Stop Button .....................................................................3-2
LED Display ................................................................... 3 -2
Power Indicator Light .................................................. 3 -2
Start Up .................................................................................3-3
Running Programs ..............................................................3-4
Configuring the Liquid Handler....................................... 3 - 5
About...............................................................................3-6
Pump Type .....................................................................3 -7
Syringe Options ............................................................. 3 -7
Rinse Site.........................................................................3 - 8
Liquid Detector..............................................................3 -9
Adjust XY ....................................................................... 3-10
Z Height ..........................................................................3-12
Home Phase ....................................................................3-13
Safety Contact ...............................................................3-14
Bearing Life ....................................................................3-14
Sound and Display ....................................................... 3-15
Emergency Contact ...................................................... 3-15
Testing the Liquid Handler's Contacts ............................3-16
4 Maintenance
Helpful Hints ........................................................................4-2
Cleaning ................................................................................4-3
Cleaning the Liquid Handler ......................................4 - 3
Cleaning the Fluid Path ...............................................4 -3
Horizontal Pitch and Vertical Rods ........................... 4 -7
Replacing Parts ....................................................................4-9
Replacing Tubing .......................................................... 4 -9
Replacing a Probe ......................................................... 4 -9
Replacing a Level Sensing Cable................................4-11
Replacing a Fuse ............................................................4-13
Checking Position Alignment ............................................ 4-14
Transporting the Liquid Handler .....................................4-15
5 Troubleshooting
Error Messages .....................................................................5 -2
Electrical ................................................................................5-6
Liquid Level Detector..........................................................5 - 8
Repair and Return Policies ................................................ 5 - 9
Appendix A Replacement Parts and Accessories
Probes.....................................................................................A-1
Control Software .................................................................A-2
Probe Guide Inserts .............................................................A - 2
Transfer Tubing and Waste Bottle ....................................A -2
Rinse Station ......................................................................... A -3
Rack Accessories..................................................................A-3
Cables and I/O Accessories ............................................... A -4
Miscellaneous .......................................................................A-4
Appendix B Racks
Peltier Racks ...................................................................B-11
Appendix C GSIOC Commands
GSIOC Commands ..............................................................C-2
GSIOC Command List ................................................. C -2
Appendix D Pipetting Techniques
Minimizing Cross-Contamination ....................................D-2
Minimizing Dilution Effect ................................................ D -3

Declaration of Conformity

Application of Council Directives:
89/336/EEC, 73/23/EEC
Standards to which Conformity is Declared:
EN61326, EN61000-3-3, EN61000-3-2, EN61010-1
Manufacturer’s Name ...........................................Gilson, Inc.
Manufacturer’s Address .......................................3000 W. Beltline Highway
Middleton, WI 53562
EC Office Address ..................................................Gilson S.A.S.
19 Avenue des Entrepreneurs, B.P. 145
F-95400 Villiers-le-Bel, France
Type of Equipment.................................................Laboratory Equipment
Model. .......................................................................Quad-Z 215 Liquid Handler
Beginning with Serial Number: 250A1K001 Month and Year of Manufacture: January 2001
I, the undersigned, hereby declare that the equipment
specified above conforms to the above Directives and Standards.
Place: Middleton, WI (USA) Michael Jacquart Issue Date: January 2001 Senior Vice President
Corporate Technology Development

Introduction

1

Safety Precautions

For safe and correct use of this instrument, it is recommended that both operating and service personnel follow the instructions contained in this guide when installing, cleaning, and maintaining this instrument.
Because the probes installed on the Z-arm may contain a dangerous substance, use the safety shield included with the instrument and do not interfere in the work area of the instrument until the liquid handler has completed its procedures. If dangerous liquids are used, adequate protection such as proper ventilation, safety glasses, etc., should be used.
Always switch the power to off when making adjustments to the liquid handler. The potential exists for bodily harm if you interfere with the work area of the instrument while it is running.
1-1
Introduction 1

Description

The Gilson Quad-Z 215 Liquid Handler is an XYZ robot that can automate any number of manual
Description
liquid handling procedures. The Quad-Z has four independently operated probes with variable horizontal spacing (from 9 to 18 mm) allowing access to virtually any tube, vial, or microplate configuration.
The optional 849 Multiple Injection Module can be configured with four sample loops of varying capacities to provide complete flexibility for open access laboratories. The advanced liquid-level detection for each independently operated probe minimizes carryover ensuring accurate and reproducible results.
The external 444 QuadDilutor provides for the accurate and precise handling of liquids.
1-2
Introduction 1

Unpacking

The Quad-Z 215 Liquid Handler is delivered with all major components already assembled except for auxiliary parts such as the Z-arm, probe, racks, tubing, etc. Keep the original container and packing assembly in case the liquid handler must be returned to the factory.
The Quad-Z 215 Liquid Handler and its components are shipped in two containers:
One container holds the auxiliary items, such as locator plate, tubing, probes, rinse station, Z-arm, and any other accessories you may have ordered with your system.
The other container holds the Quad-Z 215 Liquid Handler.
Unpacking
To remove the liquid handler from its container: 1 Cut the metal strapping. 2 Lift the outer box off and away from the liquid
handler.
3 Lift the inner box off and away from the liquid
handler.
4 Lift the unit off its base platform and place it on
a lab bench or cart. Gilson recommends that
two people lift the liquid handler off the base of the packing container. To lift the liquid
handler: a) Using the two cutouts for hand holds, place
a hand at the base of the packing container. b) Grip the liquid handler under the base plate. c) Lift the unit up and out of the foam packing
material. The side containing the electronics
cabinet is the heavier side.
Do not attempt to lift the instrument from the Y-arm (the horizontal arm). Always lift the instrument from its base.
1-3
Introduction 1

Standard Equipment

Once the liquid handler and the accessories containers have been unpacked, you should have
Unpacking
the following: Quad-Z 215 Liquid Handler 444 QuadDilutor with accessories Locator plate with one drain base (includes four
mounting screws) Rinse drain package which includes:
2-liter waste bottle
Cap with quick connect fitting
Rinse station with fittings
5 feet of Tygon waste tubing with quick connect fitting
Z-arm and control cable with retaining clip and level sensing cables
Accessory package which includes:
Fuse drawers, fuses, and power cords
10-pin terminal block connector
8-pin terminal block connector
9/64" ball driver for removal of armlock
Eight tubing retaining clips
Cable support rod with bracket and two Phillips-head attachment screws
Tubing support rod
215 Utility Programs CD-ROM 444 Utility Programs CD-ROM
Quad-Z 215 Liquid Handler User’s Guide 444 QuadDilutor User’s Guide

Accessories

1-4
Based upon your configuration, you’ll also receive additional accessories, such as the probes, transfer tubing, racks, etc. If necessary, refer to Appendix A for part numbers.
Introduction 1

Customer Service

Gilson, Inc. and its worldwide network of authorized representatives provide customers with the following assistance: sales, technical applications, and instrument repair.
If you need assistance, please contact your Gilson representative or if you are in the United States call the Gilson Customer Service Department at 800-445­7661 or 608-836-1551. You can also contact the Customer Service Department via its e-mail address: service@gilson.com. Specific contact information can be found on the Gilson web site at www.gilson.com. To help us serve you quickly and efficiently, please refer to the Before calling us section on page 5-8.
Customer Service
1-5
Introduction 1

Technical Specifications

Please be aware of the following before operating the liquid handler.
Warning: Changes or modifications to the liquid handler not expressly approved by Gilson could void the factory-authorized warranty.
T echnical Specifications
The liquid handler has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC commercial environment. The liquid handler generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. Operation of the liquid handler in a residential area is likely to cause harmful interference; in which case, the user will be required to correct the interference at the user’s own expense.
Shielded cables must be used with the liquid handler to ensure compliance with the Class A FCC limits.
1-6
Introduction 1
T echnical Specifications
1-7
Introduction 1
T echnical Specifications
1-8

Installation

This section takes you through the steps for setting up your Quad-Z 215 Liquid Handler.
2
2-1
Installation 2

Locator Plate Installation

The locator plate serves two functions:
Positions the racks and accessories that fit onto the bed of the liquid handler.
Contains liquid spills, such as those caused by overflowing vessels.
Locator Plate Installation
The locator plate and its four mounting screws are shipped in a separate box with the liquid handler’s accessories. To install the locator plate onto the instrument bed:
1 Make sure the locator plate’s rinsing station
base is at the rear of the instrument. The locator plate will only install in this orientation.
2 Align the four corner holes of the locator plate
with the four holes on the instrument bed and lower the plate onto the bed.
3 Using a Phillips screwdriver, secure the locator
plate using the four mounting screws.
2-2
Installation 2

Armlock Removal

The armlock on the liquid handler secures the Y-arm during shipment. You must remove the armlock prior to installing the Z-arm and operating the instrument. If the armlock is not removed, the liquid handler cannot move in the X-direction. This results in an error state during operation.
If you need to move the liquid handler, always reinstall the armlock. This safeguards against mechanical damage.
To remove the armlock: 1 Remove the cardboard label in front of the
armlock.
Armlock Removal
2 Using the 9/64" ball driver, loosen the
tensioning screw that immobilizes the Y-arm.
3 Using the 9/64" ball driver, remove the two
remaining screws that hold the armlock in place.
4 Remove the armlock and store it and the ball
driver for future use.
2-3
Installation 2

Z-Arm Installation

Follow these steps to install the Z-arm: 1 Using a Phillips screwdriver, loosen the
mounting screw on the Z-arm mounting
Z-Arm Installation
bracket located on the Y-arm. Turn counterclockwise to loosen.
2-4
2 Partially pull out the bracket. Do not remove
completely.
3 Place the Z-arm into the mounting bracket. You
will need to insert one side of the Z-arm into place at a time.
4 Tighten the screw on the mounting bracket
until the Z-arm is secure. You’ll adjust the Z-arm to its proper height
after rack and rinse station installation. This adjustment is described on page 2-14.
Installation 2

Installing the Z-Arm Cable Support Rod

1 Using the two Phillips screws, attach the cable
support rod bracket in the holes located in the rear of the X-arm extrusion.
Z-Arm Installation
2 Extend the arm to the extreme X- and Y-direction
to ensure that the cable will have enough slack. Plug the Z-arm control cable into the back topside of the control cabinet. The control cable should be tucked into the groove located in the top of the X-arm extrusion. The retaining clip that is already on the control cable should be snapped onto the top of the cable support rod. Refer to diagram below.
2-5
Installation 2

Probe Installation

There are different probes available for use on the Quad-Z 215 Liquid Handler. Depending upon your application, you have purchased the appropriate probes and probe guide inserts. When installing the
Probe Installation
probes or custom sized probe guide inserts (1.3 or
1.5 mm) refer to the following procedures and
diagrams.

Installing the Probe Guide Inserts

Your Quad-Z is delivered with 1.5 mm probe guide inserts installed. If you ordered custom sized probe guide inserts for more precise XY accuracy, refer to the installation instruction below.
To install the probe guide inserts, turn the insert clockwise into the probe guide.

Installing the Probes

Insert the probes into the top of the isolation probe holders and pull them through the holders and the probe guide inserts until the tip of the probe is in the probe guides.
2-6
Installation 2

Plumbing Connections

Transfer Tubing Installation

1 Connect one end of each piece of transfer
tubing to the 444 QuadDilutor. Refer to the 444 QuadDilutor User’s Guide for more information.
2 Locate the level sensing cables (LLD cables)
supplied with the Z-arm.
3 Pass the other end of each transfer tubing
through the open end of an LLD cable casing nearest the right-angle connector. The tubing should exit the back of the LLD cable casing approximately 6.5 cm from the end.
4 Connect each tubing to the top of the isolation
probe holder using a 1/4"-28 nut and ferrule supplied with the tubing. Firmly tighten this fitting using the supplied headless nut extender (part number 49041032) since it holds the probe in place.
Plumbing Connections
2-7
Installation 2

Installing the Level Sensing Cables

To install the level sensing cable: 1 Plug the right-angle connector on one end of the
cable into socket one on the back of the Z-arm.
Plumbing Connections
2-8
2 Snap the cable into the LLD cable guide (refer
to the picture above).
3 Loosen the small hexagonal nut on the probe
holder for probe one. Attach the split-tongue connector to the nut. Retighten the nut until snug (approximately one turn after finger tightening).
4 Repeat for probes 2 through 4.
Installation 2

Rinse Station and Drain Waste Tubing Installation

You’ll clean the probe using the rinse station. To eliminate carryover of liquids, the rinsing procedure pumps an excess volume of diluent or probe washing solution through the probe and out into the rinse station. The small diameter of the rinse station inserts allow the outside of the probe to be washed along with the inside.
The rinse station’s design accommodates three kinds of rinses:
Shallow-pocket rinse - Used for level sensing applications where the probe is only immersed in a few millimeters of the sample.
Plumbing Connections
Deep-pocket rinse - Used for non-level sensing applications. This type of insert allows for a deeper insertion of the probe into the rinse well resulting in a greater area of the outside of the probe to be rinsed.
Flow-through rinse - Used in applications where a rigorous wash of the probe’s exterior is required. A second source of liquid is pumped to the rinse station to perform this type of rinse.
It may be necessary to vary the types and volumes of probe wash solutions to most efficiently eliminate carryover of particular compounds. Generally, the smaller the volume of probe wash solution used, the faster your automated liquid handling protocol.
2-9
Installation 2
Installing the rinse station
The base of one rinse station is shipped already secured to the locator plate and is located at the rear of the locator plate. The locator plate can hold an optional second rinse station or you can move the rinse station base to the alternate location.
Before installing the rinse station, make sure the
Plumbing Connections
locator plate has been properly installed with the previously-installed rinse station base located at the rear of the instrument bed.
To install the rinse station onto the base, follow these steps:
1 Align the triangle on the bottom of the rinse
station with the base.
2 With the rinse station’s fittings facing you,
insert the rinse station into the base.
3 Press down and turn the rinse station clockwise.
The rinse station is secure when you feel the rinse station snap into place. When installed correctly, the fittings point toward the control cabinet.
4 If you will be doing shallow- or deep-pocket
rinses, connect waste tubing to the barbed fitting installed on the rinse station.
If you will be doing flow-through rinses, remove the plug installed on the rinse station and replace it with a barbed fitting. Connect tubing between the barbed fitting and the external liquid source.
2-10
Installation 2

Rack Setup

The Quad-Z 215 Liquid Handler is equipped to locate Code 20-, 30-, 200-, and 500-series racks. See
Appendix B for a list of racks available for the liquid
handler. Depending on the racks you’re using, refer to the
appropriate procedures on the following pages.

Code 200-Series and Code 500-Series Racks

If all your racks are Code 200-series or Code 500­series racks, place them directly onto the locator plate:
1 Orient the rack so that the code number (for
example, 200) is facing forward.
Rack Setup
2 Fit the rack on the locator plate so that the slots
and holes on the underside of the rack align with the pins on the locator plate.
2-11
Installation 2

Code 200-Series and Code 20-Series or Code 30-Series Racks

To use a combination of Code 200-series and Code
Rack Setup
20-series or Code 30-series racks on the locator plate:
1 Install each Code 200-series rack as described
on page 2-11.
2 For each Code 20-series or Code 30-series rack,
place an adapter plate onto the locator plate. Fit the plate so that the slots on the adapter plate align with the pins on the locator plate.
3 Place the Code 20-series or Code 30-series rack
onto the adapter plate.
2-12
Installation 2

Only Code 20-Series or Code 30-Series Racks

If all your racks are Code 20- or Code 30-series racks and you do not have the optional 849 Multiple Injection Module installed, follow the steps below.
1 Install the locator tray (part number 2504627,
ordered separately) onto the locator plate of the liquid handler.
For Code 20-series racks, the handles face
the front.
For Code 30-series racks, the hose fittings
should face the back.
2 Position each rack onto the locator tray. You
can install up to seven racks using this tray.
Rack Setup
2-13
Installation 2

Final Z-Arm Adjustment

Follow these steps to adjust the Z-arm to the proper height.
1 Turn off power to the liquid handler. 2 Locate the appropriate Z-height adjustment
tool in the accessory package. Two Z-height
Final Z-Arm Adjustment
adjustment tools are supplied in the package.
Part number Description
25051094 125 mm Z-height adjustment tool.
Use this tool to adjust the Z-arm on the Quad-Z 215 Liquid Handler for liquid handling and injection.
Z-height
adjustment tool
25051095 175 mm Z-height adjustment tool.
Use this tool to adjust the Z-arm on the Quad-Z 215 Liquid Handler for liquid handling.
3 Place the Z-height adjustment tool on one of its
ends near the center of the locator plate (if necessary, remove any racks or accessories before doing this).
2-14
Installation 2
4 Loosen the mounting screw on the Z-arm
mounting bracket and slightly raise the Z-arm.
Final Z-Arm Adjustment
5 Manually move the liquid handler’s arm so the
Z-arm is centered over the Z-height adjustment tool.
6 While holding Z-height adjustment tool flat
against the locator plate, use the other hand to lower the Z-arm until it lightly rests on the adjustment tool.
7 Tighten the mounting screw on the Z-arm
mounting bracket so the Z-arm is secure.
8 While holding the adjustment tool in place,
slide the Z-arm off the tool. Ensure that the bottom of the Z-arm lightly rubs against the adjustment tool as it moves. Repeat steps 4 through 7 until this is true.
9 Store the Z-height adjustment tool.
2-15
Installation 2

Electrical Connections

Rear Panel

1 Input/Output (I/O) ports 2 Gilson Serial Input/Output Channel (GSIOC) port 3 Keypad port (not used on the Quad-Z) 4 RS-232 port
Electrical Connections
5 Fuse drawer 6 Power switch 7 Power receptacle 8 Unit ID selector 9 Baud rate/mode selector
2-16
Installation 2

Input/Output Ports

You can use the input and output contacts found on the rear panel of the liquid handler to control peripheral devices. Refer to the diagram on page
2-16 for the location of the input/output ports.
Contact inputs
The input terminal block of the liquid handler has eight contacts. All of the inputs are paired, and each pair includes a GROUND reference ( ).
The contact input pairs are labeled A, B, C, and D. A contact is connected if it has a short across the
input or is held low by a TTL output or other device.
Electrical Connections
Never connect voltages higher than 5V DC to an input. When using TTL signals, be sure to match GROUND connections.
Contact outputs
The output terminal block has 10 contacts. Pins 1 and 2 supply a +24V DC output. Do not use
this output unless the receiving device can accept 24V power.
Pins 3 through 10 are paired, isolated-relay contact closures and are labeled 1, 2, 3, and 4.
Items you’ll need
To make connections, you’ll need the following:
2-conductor cable (22–30 gauge for each wire)
wire insulation stripper
small-blade screwdriver
You can purchase a 6-foot piece of suitable cable (part number 709910206) or a package of five cables with identification markers (part number
36078155) from Gilson.
2-17
Installation 2
Making connections
To prepare and make connections with the 2­conductor cable:
1 Cut the cable into pieces of appropriate length. 2 Strip about 0.25 cm of insulation from each end
of the cable.
Electrical Connections
3 Remove the terminal block connector from the
liquid handler. Insert each wire into the appropriate slot on the terminal block connector.
Note: When making connections, be sure to maintain the correct orientation of the connector relative to the port.
Push the wire all the way in; then tighten its corresponding pin screw.
4 Reconnect the terminal block connector to the
liquid handler. The wires will be facing left and the pin screws will be facing you as you look at the rear of the instrument. Push the connector in as far as it will go. It is designed to fit snugly into its receptacle.
5 Connect the opposite ends of the wires to the
other device(s). Be sure to match ground connections.
6 Label each cable to identify the purpose of the
connection.
2-18
Installation 2

RS-232 Port

The RS-232 port is used to transfer information between the liquid handler and a computer. For the location of the RS-232 port, refer to the diagram on page 2-16.
Be sure your computer is turned off before making any connections.
To connect your computer to the liquid handler, you’ll need an RS-232 cable. Obtain a cable with D-connectors that are appropriate for the liquid handler and your computer. The liquid handler requires a 25-pin male D-connector. Refer to the back panel of your computer or its documentation to determine which type of D-connector it requires. RS-232 cables are available from Gilson and your local computer store.
Electrical Connections
Connecting an RS-232 cable
Attach the male end of the RS-232 cable to the RS-232 port located on back panel of the Quad-Z 215 Liquid Handler. Tighten the retaining screws.
Attach the female end of the RS-232 cable to the computer’s RS-232 serial communications port. (Do not mistake it for the female 25-pin parallel printer port!) Again, tighten the retaining screws.
2-19
Installation 2

GSIOC Port

Gilson systems feature a two-way communication interface between the computer and most Gilson modules. Communication occurs along the Gilson Serial Input/Output Channel (GSIOC).
The liquid handler can convert the RS-232 signal levels used by computers to the RS-422/485 signal
Electrical Connections
levels required by the GSIOC and vice versa. (See page 2-19 for information on making the RS-232 connection between the liquid handler and computer.)
GSIOC cable
Use the GSIOC cable to link an additional Gilson GSIOC module to the liquid handler and control both devices via a program executed on the computer.
2-20
Connect the female connector, located individually at one end of the cable, into the GSIOC port of the liquid handler. Tighten the retaining screws. (Refer to diagram below.)
Connect the other female connector, located on the same end as the male connector, to the Gilson module. Tighten the retaining screws.
If you’re connecting another Gilson module, use the male connector to join another GSIOC cable and make the necessary connection to the next Gilson module.
Installation 2

Unit ID and Baud Rate/Mode Selection

Use the SW1 selector to choose a different unit ID and the SW2 to choose a different baud rate/mode. If necessary, refer to the diagram on page 2-16 for the location of these selectors.
Unit ID
The unit ID identifies the liquid handler to Gilson software packages that can issue GSIOC commands to the liquid handler.
At the factory, Gilson set the unit ID to 22. There is no need to change this number unless it is the same as that assigned to another Gilson device that’s also connected along the GSIOC.
Electrical Connections
To change the unit ID: 1 Gently insert a small flat blade screwdriver into
the SW1 selector on the rear panel and turn it.
2 Align the white dot with one of the indicated
numbers. The unit ID is 20 plus the selected number.
2-21
Installation 2
Baud rate/mode
As a default, the baud rate/mode is set to 6, indicating that the liquid handler is set for a baud rate of 19200 and is a master device.
You’ll need to change the selection to 0 (zero) if the liquid handler is connected via the GSIOC to a Gilson system and is being controlled by the 506C
Electrical Connections
System Interface from Gilson control software. A setting of 0 indicates the liquid handler is a slave device and the baud rate is being clocked externally.
Note that other selections are available if the liquid handler is being controlled by non-Gilson applications. Refer to the table shown on the liquid handler’s rear panel; see page 2-16.
To change the baud rate/mode: 1 Gently insert a small flat blade screwdriver into
the SW2 selector on the rear panel and turn it.
2 Align the white dot with one of the indicated
numbers.
2-22
Installation 2

Fuses

You received the liquid handler without any fuses installed. To install the fuses:
1 Locate the accessory package containing the
fuse drawer appropriate for your line voltage. Discard the other fuse drawer.
2 Locate the accessory package containing the
5.0A “T” Slo-Blo fuse (5 x 20 mm size) fuses.
3 Install the fuse(s) into the fuse drawer. The fuse
drawer for 100/120V accepts one fuse. The fuse drawer for 220/240V accepts two fuses.
4 Insert the fuse drawer into its receptacle in the
liquid handler. See rear panel diagram on page
2-16.
Fuse installation for 100/120 voltage
Electrical Connections
Fuse installation for 220/240 voltage
2-23
Installation 2

Power Cord Connection

Locate the appropriate power cord for your line voltage. Discard the other power cord.
Use the power cord to connect the liquid handler to an AC power source.
Electrical Connections
2-24

Operation

The Quad-Z 215 Liquid Handler is controlled as follows:
Via a method running on a personal computer. You create the method using 735 Sampler Software (ordered separately, part number 2106773532).
If you have not already done so, install the 735 Sampler Software. Refer to the documentation supplied with the software.
3
3-1
Operation 3

Front Panel

The front panel of the liquid handler contains a Start button, Stop button, LED display, and power
Front Panel
indicator light.

Start Button

The Start button can be used to home the XYZ-arm when the liquid handler is first powered up or when the motors for the XYZ-arm have been relaxed. When pressed, the yellow LED lights.

Stop Button

The Stop button is a large touch-sensitive pad that can be used to terminate a program and stop the liquid handler from responding to any more commands coming from the running program. This button also relaxes the motors for the XYZ-arm so that you can easily lift the probe and move the arm. When pressed, the yellow LED light is turned off.
In a situation where an emergency stop is required, pressing the Stop button immediately stops the liquid handler. The Stop button is designed to be sensitive enough that if you just brush it with your hand it activates.

LED Display

The 8-character LED display shows the current status of the liquid handler and any error codes as they are encountered. Your program can also contain instructions for showing 8-character messages on the display when the program is run.
Refer to Section 5, Troubleshooting for a list of current error codes and required actions.

Power Indicator Light

This indicator becomes lit when you turn on power to the liquid handler using the power switch located on the rear panel. Refer to the rear panel diagram on page 2-16 if necessary.
3-2
Operation 3

St art Up

To start the liquid handler: 1 Make sure the liquid handler is connected to a
power source.
2 Turn on the liquid handler using the power
switch located on the rear panel. Refer to rear panel diagram on page 2-16 if necessary. The power indicator light on the front panel illuminates.
When power is turned on, the liquid handler beeps and displays the current version of its installed firmware. This message appears for about one second before the LED display returns to a blank state.
Start Up
In order to determine what PROM version is installed in your liquid handler, you may need to turn the unit off then on again and watch the display for the version number to appear.
3 After the liquid handler powers up, press the
Start button. This initiates the homing sequence that allows the liquid handler to determine its mechanical reference positions. The sequence takes approximately one minute to complete.
While the homing sequence progresses, the LED display shows Homing. When the sequence completes, it blanks.
Note: If the program being executed by the liquid handler doesn’t include commands for homing the instrument, perform step 3 before starting the program.
The utility programs, supplied with the liquid handler, home the instrument if necessary.
3-3
Operation 3

Running Programs

The liquid handler is controlled by programs executed from a personal computer.
Executable (.EXE) programs can be run from a computer. The computer is connected to the liquid
Running Programs
handler via an RS-232 cable. Refer to Section 2 for correct installation of the RS-232 cable between the liquid handler and the computer.
The following utility programs may be used with the Quad-Z 215 Liquid Handler and are supplied on the 215 Utility Programs CD-ROM supplied with your liquid handler.
215 Setup Utility - Specifies configuration parameters for the liquid handler.
215 Contact Test Utility - Enables you to test contact connections.
3-4
Operation 3

Configuring the Liquid Handler

The liquid handler comes from the factory with its configuration set by Gilson. Configuration information is stored in the non-volatile memory of the liquid handler. Prior to using the liquid handler for the first time, it is important to review and adjust the default configuration to make sure it is correct for your application.
The following pages describe how to use 215 Setup Utility to configure the liquid handler. This information is also available in the on-line help supplied with the program.
When you execute the 215 Setup Utility from the computer, the following tabs appear:
Configuring the Liquid Handler
about
pump type
syringe options
rinse site
liquid detector
• adjust XY
Z height
home phase
safety contact
bearing life
sound & display
emergency contact
Following is a description of each of the tabs.
3-5
Operation 3

About

Configuring the Liquid Handler
The About tab allows you to indicate the unit ID of the instrument being configured so the 215 Setup Utility can communicate with that instrument.
This tab also lists software version and copyright information for the 215 Setup Utility.
Unless the Non-Volatile RAM has been cleared, information about the type of 215 you are using will appear automatically. You will see the following indicating that you are using a Quad-Z 215 Liquid Handler: 215QADvX.XX.
3-6
Operation 3

Pump Type

Because the Quad-Z 215 Liquid Handler uses an external dilutor, Blank Front is selected and all options are inactive.
Configuring the Liquid Handler

Syringe Options

All options on this tab are inactive for the Quad-Z 215 Liquid Handler because it has an external dilutor.
3-7
Operation 3

Rinse Site

Configuring the Liquid Handler
The instrument stores the location of a rinse site in memory. This allows the instrument to move to this location before homing the syringe and prevents the spilling of waste liquid or rinse diluent.
You must modify the rinse site coordinates if you have a Quad-Z 215 Liquid Handler. Refer to the following table to select the correct X, Y, and Z coordinates depending on the kind of probe rinse that will be performed. Coordinates for the rinse site are provided for the X, Y, and Z axis in millimeters. The coordinates identify the rinse site for the left-most probe installed on the Quad-Z 215 Liquid Handler.
XYZ Shallow-pocket rinse 82.9 16.5 100.0 Deep-pocket rinse 82.9 3.8 47.5 Flow-through rinse 82.9 3.8 47.5
3-8
Operation 3

Liquid Detector

The Liquid Detector tab allows you to adjust the default sensitivity for liquid detection.
Note: Gilson control software (such as 735 Sampler Software) will override these settings.
Configuring the Liquid Handler
Raising the percentage suppresses false liquid detection while lowering the percentage increases sensitivity for situations where liquid is harder to detect. By lowering the probe so it touches the liquid for one of your samples, you can manually test the current sensitivity of the liquid detection. In this manner, you can observe the amount of change you might expect with each sample.
The factory default setting is 10% for each probe.
3-9
Operation 3

Adjust XY

Configuring the Liquid Handler
The Adjust XY tab allows you to test whether the instrument is properly adjusted and to make minor adjustments to the X- and Y-axis offsets if needed. You may need to use the options under this tab if the probe is not accessing the injection port of the injection module installed or the vessels in the installed racks.
The X offset and Y offset text boxes display the current offsets stored in the instrument’s memory.
To determine if the probe(s) need(s) to be adjusted in the X or Y direction select the model of the injection module that is on your Quad-Z 215 or select other for a user-defined test point. The default XY coordinates are shown next to the model number.
Select the model of the injection module that is on your 215 or select other for a user-defined test point. The default XY coordinates are shown next to the model number.
3-10
Operation 3
model 819:
X-coordinate: 544.0 mm Y-coordinate: 3.8 mm
If the model 819 is not installed next to the right support use the following coordinates: X-coordinate: 351.1 mm Y-coordinate: 1.3 mm
model 841:
X-coordinate: 475.3 mm Y-coordinate: 3.8 mm
model 889:
X-coordinate: 336.1 mm Y-coordinate: 3.8 mm
model 849:
X-coordinate: 336.1 mm Y-coordinate: 3.8 mm
other:
Configuring the Liquid Handler
3-11
Operation 3

Z Height

Configuring the Liquid Handler
Use the Z Heights tab to identify the size of the installed Z-arm and the height at which the Z-arm is clamped.
The tower height options are inactive for the Quad-Z 215 Liquid Handler because there is only one tower height available.
The Z-arm can be clamped at an adjustable height over the locator plate. You can set this height so that the liquid handler is able to properly find heights that you specify. Type the clamp height in millimeters. A clamp height of 0 mm means the Z-arm is flat on the locator plate.
3-12
Operation 3

Home Phase

Use the Home Phase tab to display the current X­and Y-phase of the instrument.
Clicking Start causes the liquid handler to perform the phase procedure. This procedure consists of the liquid handler homing itself 10 times.
Configuring the Liquid Handler
The liquid handler finds out where home is located by “feeling” for the back and left walls of the unit. The liquid handler expects to find these walls in the same place each time. If it does not, you will get an error. If this error was caused by an obstruction, just clear the obstruction and try again. If the problem does not clear or if a change is made to the mechanics, you will probably need to repeat this option to find the true home location.
Once the process completes, the spreadsheet displays the values generated from each phase procedure.
3-13
Operation 3

Safety Contact

Configuring the Liquid Handler
The instrument has provisions for connecting safety devices that your application may require, as long as they present a contact closure or TTL type interface. The Safety Contact tab allows you to specify which input contact is connected to the safety device and what is the active state of that device. The function of the safety contact is equivalent to pressing the Stop button on the instrument’s front panel.

Bearing Life

The Bearing Life tab displays the XYZ travel in kilometers.
3-14
Operation 3

Sound and Display

Use the options in the Sound and Display tab to adjust the brightness of the display, sound level, and tone.
Clicking Beep tests the sound level and tone that are currently selected.
Configuring the Liquid Handler
For the L.E.D. brightness, you can select a range of 0 through 7 where 0 is the dimmest and 7 is the brightest. Default setting is 5.

Emergency Contact

The emergency stop option provides for sending a signal to a peripheral device (such as a Gilson 818 AutoMix) whenever the liquid handler’s Stop button is pressed or safety input is activated.
Note: Once an emergency output has been activated, reset the contact to its non-emergency state using the 215 Contact Test Utility.
3-15
Operation 3
T esting the Liquid Handler’s Contact s
Following is a description of how to use the 215 Contact Test Utility to toggle output contacts to determine if the correct contact connections have been made to peripheral devices to be controlled by the Quad-Z 215 Liquid Handler. The program also identifies the state of input contacts and lets you test the Start and Stop buttons on the liquid handler.
Before using this software, you need to connect the peripheral device’s inputs to the appropriate output pair on the liquid handler. If necessary,
T esting the Liquid Handler’s Cont acts
refer to Section 2, Installation for information on making contact connections.
3-16

Maintenance

To obtain optimum performance and maximum life from the Quad-Z 215 Liquid Handler, it is important to keep the instrument well-maintained.
This section contains some general guidelines that will help you to maintain your liquid handler.
4
4-1
Maintenance 4

Helpful Hints

In order to keep your liquid handler at peak performance, Gilson recommends that you do the
Helpful Hints
following:
Change or clean the tubing regularly to
maintain maximum performance.
Flush the probe housings and rinse stations
daily with distilled or deionized water. On a weekly basis, flush with a 10% solution of bleach or weak detergent.
Check periodically to ensure that all fittings are
tight.
Wipe up all spills immediately.
Cold fluids may cause leakage; warm fluids to
room temperature before running them through the system.
Lubricate the rods on the Z-arm at least once
every six months (see page 4-7).
4-2
Maintenance 4

Cleaning

Cleaning the Liquid Handler

The liquid handler should be cleaned occasionally using a dry, clean cloth. Or, if necessary, use a cloth dipped in soapy water. If liquid is accidentally spilled on the liquid handler, wipe the instrument using a dry, clean cloth.

Cleaning the Fluid Path

Depending on your use of the liquid handler, it may be necessary to flush the entire fluid path.
It’s important to clean the fluid path if you won’t be using the liquid handler for a while or if you’re using a solution with a high salt concentration for a probe wash or as a diluent.
Cleaning
Prime the system using distilled or deionized water. Check the beaker during the priming sequence to ensure it always has liquid in it.
4-3
Maintenance 4
Cleaning methods
Depending on the samples or reagents that come
Cleaning
into contact with the fluid path, you may need to vary your cleaning methods accordingly. Use the following cleaning protocols as references and make any changes to them as required for the samples and reagents being pumped for your application.
Proteins and peptides - Follow this procedure if the fluid path is in contact with proteins and peptides:
1 Prime the fluid path using a weak detergent
solution. 2 Pause the priming sequence. 3 After 30 minutes, resume priming the fluid path
using distilled or deionized water to pump the
remaining detergent from the tubing into a waste
container. Prime the fluid path a minimum of 10
cycles with distilled or deionized water. 4 When you’re satisfied that the entire fluid path
has been flushed with water, end the priming
sequence.
4-4
Maintenance 4
Acidic and basic compounds - Follow this procedure if the fluid path is in contact with acidic and basic compounds:
1 Prime the fluid path using a 0.1N NaOH
solution. 2 Pause the priming sequence. 3 After 10 minutes, resume priming the fluid path
using distilled or deionized water. Prime until
the fluid path has been flushed with water. 4 Pause the priming sequence. 5 Prime the fluid path using a 0.1N NaOH
solution. Continue to prime until the fluid path
has been flushed with 0.1N NaOH. 6 Pause the priming sequence. 7 After 10 minutes, resume priming the fluid path
using distilled or deionized water. Prime until
the fluid path has been flushed with water.
Cleaning
8 When you’re satisfied that the entire fluid path
has been flushed with water, end the priming
sequence.
4-5
Maintenance 4
Biological fluids - Follow this procedure if the fluid path is in contact with biological fluids such as blood products:
Cleaning
1 Make a solution of 10% bleach by adding one
part of commercial bleach to nine parts of water.
2 Prime the fluid path using the bleach solution
until the entire fluid path has come into contact
with bleach. 3 Pause the priming sequence. 4 After 30 minutes, resume priming the fluid path
using distilled or deionized water to pump the
remaining bleach solution from the tubing into a
waste container. Prime the fluid path a minimum
of 10 cycles with distilled or deionized water. 5 When you’re satisfied that the entire fluid path
has been flushed with water, end the priming
sequence.
4-6
Maintenance 4

Lubricating

Always switch the power to off when making adjustments to (such as lubricating) the liquid handler. The potential exists for bodily harm if you interfere with the work area of the instrument while it is running.

Horizontal Pitch and Vertical Rods

Refer to the instructions and diagrams on the next few pages for information on lubricating the horizontal pitch and vertical rods on the Quad-Z
215.
What you need
Mobile 007 grease (part number 25494101, included in the 215 Alignment Kit, part number
254941)
Applicator (a cotton swab, for example)
Lubricating
Procedures
1 Turn OFF power to the Quad-Z 215. 2 Dispense a small bead of grease on the
applicator.
3 Apply the grease to the visible and accessible
areas on the vertical rods and horizontal pitch rods on the Z-arm (refer to the pictures on the next page).
4 Repeatedly (and slowly), move the
probe holders up and down to distribute the grease along the vertical rods.
5 Using a 2.5 mm hex wrench, adjust the
pitch to 9 or 18 mm (whichever is opposite of the current setting). Refer to the diagram at right.
4-7
Maintenance 4
6 Dispense another small bead of grease on the
applicator.
7 Apply the grease to the visible and accessible
Lubricating
areas on the vertical rods and horizontal pitch rods on the Z-arm (see below).
8 Repeatedly (and slowly), move the probe
holders up and down to distribute the grease along the vertical rods.
4-8
18 mm spacing
9 mm spacing
Maintenance 4

Replacing Parts

Replacing Tubing

It is important to keep all tubing clean and free of crimps. Tubing that has become dirty, blocked or crimped can result in poor accuracy and precision, or loss of air gap.
Replace both the transfer tubing and inlet tubing as needed. See Appendix A for part numbers for replacement tubing. For tubing installation procedures, see Section 2.

Replacing a Probe

Refer to the appropriate instructions below depending on whether you’re replacing a probe with one of the same type or one of a different type.
Replacing Parts
Installing same type of probe
To install a replacement probe of the same type that’s currently installed:
1 Remove the transfer tubing’s
1/4"-28 fitting connected to the top of the isolation probe holder.
2 Grasp the current probe and
push it up through the top of the isolation probe holder.
3 Insert the probe into the top of
the isolation probe holder and pull it through the holder and the probe guide insert until the tip of the probe is in the probe guide.
4 Replace and tighten the 1/4"-28
fitting.
4-9
Maintenance 4
Installing different type of probe
To install a replacement probe of a different type than is currently installed, you may want to obtain a new probe guide insert for precise XY probe accuracy.
Replacing Parts
1 Remove the transfer tubing’s 1/4"-28 fitting
connected to the top of the isolation probe holder.
2 Grasp the current probe and push it up through
the top of the isolation probe holder.
3 Remove the current probe guide insert by
turning it counterclockwise. Then place the new probe guide insert into probe guide and secure it by turning it clockwise.
4-10
4 Insert the new probe into the top of the
isolation probe holder and pull it through the holder and the new probe guide insert until the tip of the probe is in the probe guide.
5 Replace and tighten the 1/4"-28 fitting.
Maintenance 4

Replacing a Level Sensing Cable

Removing the level sensing cable
1 Detach the transfer tubing
from the top of the isolation probe holder by using the supplied headless nut extender (part number 49041032) to loosen and remove the 1/4"-28 nut and ferrule.
2 Unplug the right-angle
connector for the liquid level sensing cable (LLD cable) from the socket on the back of the Z-arm.
3 Detach the LLD cable from the
LLD cable guide.
Replacing Parts
4 Loosen the small hexagonal
nut on the probe holder for probe one. Detach the split­tongue connector from the nut.
5 Remove the transfer tubing
from the LLD cable casing.
4-11
Maintenance 4
Installing the new level sensing cable
1 Locate the new level sensing cable supplied
with the Z-arm.
2 Pass the transfer tubing through the open end
Replacing Parts
of an LLD cable casing nearest the right-angle connector. The tubing should exit the back of the LLD cable casing approximately 6.5 cm from the end.
3 Connect the transfer tubing to the top of the
isolation probe holder using a 1/4"-28 nut and ferrule supplied with the tubing. Firmly tighten this fitting using the supplied headless nut extender (part number 49041032) since it holds the probe in place.
4 Plug the right-angle connector on one end of the
cable into socket one on the back of the Z-arm. 5 Pass the cable through the LLD cable guide. 6 Attach the split-tongue connector to the small
hexagonal nut on the probe holder. Tighten the
nut until snug (approximately one turn after
finger-tightening).
4-12
Maintenance 4

Replacing a Fuse

A blown fuse may indicate the existence of another problem in the instrument. If the replacement fuses blow, don’t try others. Contact your local representative or Gilson. See Before calling us on page 5-8.
To change a fuse, follow these steps. 1 Disconnect the power cord from the power
outlet and from the rear panel receptacle.
2 Locate the fuse drawer on the rear panel. See
page 2-16 if necessary.
3 Insert a small screwdriver into the notch next to
the fuse drawer.
Replacing Parts
4 Twist the screwdriver to open and remove the
fuse drawer. The fuse drawer contains one
5.0A “T” Slo-Blo fuse (5 x 20 mm size) for a 100/120 voltage selection. It contains two 5.0A fuses for a 220/240 voltage selection.
5 Remove the old fuse(s) and insert the new
fuse(s).
6 Insert the fuse drawer into its receptacle in the
liquid handler.
Fuse drawer for 100/120
voltage selection
Fuse drawer for 220/240
voltage selection
4-13
Maintenance 4

Checking Position Alignment

The 215 Setup Utility (Adjust XY tab), described in
Section 3, allows you to test whether the liquid
handler is properly aligned and to make minor adjustments to the X-axis and Y-axis offsets if needed. You may need to use the options under this tab if the probe is not accessing the injection port of the injection module installed or the vessels in the installed racks. This tab is described below.
Checking Position Alignment
The X offset and Y offset text boxes display the current offsets stored in the instrument’s memory.
To determine if the probe(s) need(s) to be adjusted in the X- or Y-direction select the model of the injection module that is on your Quad-Z (probably an 849) or select other for a user-defined test point. The default XY coordinates are shown next to the model number below.
model 819:
X-coordinate: 544.0 mm Y-coordinate: 3.8 mm
If the model 819 is not installed next to the right support use the following coordinates: X-coordinate: 351.1 mm Y-coordinate: 1.3 mm
model 841:
X-coordinate: 475.3 mm Y-coordinate: 3.8 mm
4-14
model 889:
X-coordinate: 336.1 mm Y-coordinate: 3.8 mm
model 849:
X-coordinate: 336.1 mm Y-coordinate: 3.8 mm
other:
Maintenance 4

T ransporting the Liquid Handler

When moving the liquid handler to another location or when sending it back to the factory, do not use the Y-arm as a handle. Reinstall the armlock (see Section 2) and always lift the liquid handler from the base.
T ransporting the Liquid Handler
4-15

Troubleshooting

5
5-1
Troubleshooting 5

Error Messages

Error Description Solution
1 5 NV-RAM checksum is invalid Send the buffered ~9 GSIOC command to
reset the NV-RAM and initialize to the
Error Messages
1 6 X scale factor is invalid Contact the Gilson Customer Service Department. 1 7 Y scale factor is invalid Contact the Gilson Customer Service Department. 2 0 X motor position error Turn power off then on to the liquid handler. 2 1 Y motor position error Turn power off then on to the liquid handler.
default value. Run 215SETUP.EXE or the 215 Setup Utility program
Replace the main board
2 2 Z motor position error Turn power off then on to the liquid handler. 2 4 X target less than minimum X Send the immediate Q command using the
GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.
2 5 X target more than maximum X Send the immediate Q command using the
GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.
2 6 Y target less than minimum Y Send the immediate Q command using the
GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.
2 7 Y target more than maximum Y Send the immediate Q command using the
GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.
2 8 Z target less than minimum Z Send the immediate Q command using the
GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.
5-2
Troubleshooting 5
2 9 Z target more than maximum Z Send the immediate Q command using the
GSIOC Utility Program to read the travel range. Correct the error in the program controlling the
liquid handler. 3 0 X encoder inactive Contact the Gilson Customer Service Department 3 1 Y encoder inactive Contact the Gilson Customer Service Department 3 2 Z position sensor inactive Contact the Gilson Customer Service Department 3 3 Safety contact activated Release contact then restart. 3 4 X home phase is invalid Run 215SETUP.EXE or the 215 Setup Utility
program to correct the problem. 3 5 Y home phase is invalid Run 215SETUP.EXE or the 215 Setup Utility
program to correct the problem. 3 6 X and Y home phases are invalid Run 215SETUP.EXE or the 215 Setup Utility
program to correct the problem.
Error Messages
3 9 Stop button has been pressed Turn power off then on to the liquid handler. 41 GSIOC communication error Contact the Gilson Customer Service Department.
(“Time out”) 42 Undefined GSIOC command Contact the Gilson Customer Service Department. 4 3 GSIOC command sequence Contact the Gilson Customer Service Department.
incorrect 44 Cannot send commands Contact the Gilson Customer Service Department.
(“Unit busy”) 5 5 Probe A motor position error Turn power off then on to the liquid handler. 5 6 Probe B motor position error Turn power off then on to the liquid handler. 5 7 Probe C motor position error Turn power off then on to the liquid handler. 5 8 Probe D motor position error Turn power off then on to the liquid handler. 5 9 Pitch motor position error Turn power off then on to the liquid handler.
5-3
Troubleshooting 5

Mechanical

Probe(s) no longer finding tube center
Mechanical
Probe(s) may be bent. Straighten or replace the probe.
Incorrect tray file defined. Review and if necessary change the tray file.
Liquid handler may be misaligned. Perform the position alignment procedures, described on page 4-14.
5-4
Troubleshooting 5

Electrical

Input functions not operating
Make sure connections into terminal block connector are secure.
Make sure terminal block connector is secure in input/output port.
Check connections for proper pin assignments.
Be sure pins from external devices are assigned correctly.
Check polarity of input. Inputs should be a contact closure. If not, it must be TTL level (logic 0 activates).
Electrical
Confirm that source supplying input to liquid handler is working.
Output functions not operating
Make sure connections into terminal block connector are secure.
Make sure terminal block connector is secure in the input/output port.
Check connections for proper pin assignments.
Output from liquid handler should be compatible with device to which it is interfaced. Outputs are contact closures.
5-5
Troubleshooting 5
Unit not operational
Make sure power is turned on.
Electrical
Check AC power cord connections.
Try different AC outlet.
Check fuse(s); replace if necessary.
Check all liquid handler connections and make sure that the unit is plugged in.
Unit blows fuses
Contact the Gilson Customer Service Department.
5-6
Troubleshooting 5

Liquid Level Detector

Liquid level detector not detecting liquid level
Ensure that the level sensing cables are plugged in.
Check sensitivity setting in the 215 Setup Utility (see page 3-9) and lower the percentage.
Check if liquid is detectable. Liquid level detection works only if there is electrical conductivity in your liquid. Liquid level detecting will not work with most non-polar liquids. For intermediate polarity liquids and polar liquids, check the sensitivity setting in the 215 Setup Utility.
Call Gilson if this is caused by faulty circuitry.
Liquid Level Detector
Liquid level detector falsely detecting liquid level
Ensure that probes are installed correctly.
Check sensitivity setting in the 215 Setup Utility (see page 3-9) and raise the percentage.
Call Gilson if this is caused by faulty circuitry.
5-7
Troubleshooting 5

Repair and Return Policies

Before calling us
Gilson Customer Service personnel will be able to serve you more efficiently if you have the following information:
the serial number and model number of the equipment involved. The serial number is visible
Repair and Return Policies
on the back of the control panel of the liquid handler.
the installation procedure you used
list of concise symptoms
list of operating procedures and conditions you were using when the problem arose
list of other devices connected to the liquid handler and a description of those connections
list of other electrical connections in the room
Warranty repair
Units covered under warranty will be repaired and returned to you at no charge. If you have any questions about applicability, please contact Gilson or your authorized representative.
Non-warranty repair
For out-of-warranty repairs, contact your local Gilson representative or the Gilson Customer Service Department. A Customer Service representative will discuss service options with you and can assist in making arrangements to return the equipment, if necessary.
5-8
Troubleshooting 5
Rebuilt exchange
For some units, rebuilt exchange components are available. Contact Gilson for details.
Return procedure
In the United States, contact the Gilson Customer Service Department to obtain authorization before returning any Gilson equipment. To return a piece of equipment:
Carefully pack the unit to prevent damage in transit. Check with Gilson regarding proper method of shipment. No responsibility is assumed by Gilson for damage caused by improperly packaged instruments. Indicate the authorization on the carton and on the packing slip.
Always insure for the replacement value of the unit.
Repair and Return Policies
Include a description of symptoms, your name, address, phone number and purchase order to cover repair costs, return and shipping charges, if your institution requires it. Ship to:
Gilson, Inc. Attention: Customer Service (indicate the authorization here) 3000 W. Beltline Highway Middleton, WI 53562
Outside the United States, contact your Gilson representative for return procedures.
5-9

Replacement Parts and Accessories

A

Probes

2507214 Non septum-piercing probe; constricted tip, capacitive
level-sensing, stainless steel. Dimensions: 269 x 1.8 x
1.4 mm ID (tip dimensions: 1.5 x 1.2 x 0.8 mm ID). Requires probe guide insert (part number 25064473) for precise XY accuracy.
2507215 Non septum-piercing probe; constricted tip, capacitive
level-sensing, stainless steel. Dimensions: 269 x 1.3 x
0.8 mm ID (constricted tip: 1.5 x 0.9 x 0.45 mm ID). Requires probe guide insert (part number 25064471) for precise XY accuracy.
2507254 Non septum-piercing probe; flat tip, capacitive level-
sensing, stainless steel. Dimensions: 269 x 1.8 x 1.4 mm ID. Requires probe guide insert (part number
25064473) for precise XY accuracy.
25073645 Non septum-piercing probe; beveled tip, capacitive
level-sensing, stainless steel. Dimensions: 269 x 1.3 x
0.8 mm ID. Requires probe guide insert (part number
25064471) for precise XY accuracy.
2507253 Micro septum-piercing probe; constricted 45°
bevel tip, capacitive level sensing, stainless steel. Dimensions: 269 x 1.5 x 1.1 mm ID (tip dimensions: 10 x 0.7 x 0.4 mm ID). Requires probe guide insert (part number 25064472) for precise XY accuracy.
2507216 Non septum-piercing probe; constricted, beveled tip,
capacitive level sensing, stainless steel. Dimensions: 269 x 1.5 x 0.8 mm ID (tip dimensions: 1.5 x 0.9 x 0.45 mm ID). Requires probe guide insert (part number
25064472) for precise XY accuracy.
A-1
Appendix A
2507255 Non septum-piercing probe; beveled tip, capacitive
level-sensing, stainless steel. Dimensions: 269 x 1.5 x 0.4 mm ID. Requires probe guide insert (part number 25064472) for precise XY accuracy.

Control Software

2106773532 735 Sampler Software running directly from PC;
requires PC with Microsoft Windows NT and serial cable.

Probe Guide Inserts

25064473 Probe guide insert for 1.8 mm outer diameter probes.
Replacement Parts and Accessories
25064472 Probe guide insert for 1.5 mm outer diameter probes. 25064471 Probe guide insert for 1.3 mm outer diameter probes. 25064475 Probe guide insert for 2.7 mm outer diameter probes.

Transfer Tubing and Waste Bottle

250531734 1.1 mL, 1.6 mm (1/16") OD transfer tubing
assembly for four probe (100 mL–1.0 mL syringes)
250531744 5.5 mL, 3.0 mm (1/8") OD transfer tubing assembly
for four probe (5.0 mL syringes)
250531754 10.5 mL, 3.0 mm (1/8") OD transfer tubing
assembly for four probe (5, 10, and 25 mL syringes)
49041034 Upchurch P-250 ferrule for 1.6 mm (1/16") OD
tubing, anti-twist
49041050 Upchurch P-350 ferrule for 3.0 mm (1/8") OD
tubing, anti-twist
49041035 Upchurch P-287 bushing for 1.6 mm OD tubing,
headless, anti-twist
49041022 Upchurch P-387 bushing for 3.0 mm OD tubing,
headless, anti-twist
A-2
Appendix A
23077310 Waste bottle (2 liter) with lid and quick-connect
fitting
470343706 Tygon tubing (5/16" ID x 7/16" OD) for
connection between rinse station and waste bottle; per foot
23077332 Quick-connect fitting to connect Tygon tubing to
waste bottle

Rinse Station

25045525 Multiple Probe/Quad-Z 215 rinse station; connects
to rinse station base
25245512 Rinse station base; attaches directly to 215 locator
plate
23077333 Y-connector to connect two rinse stations to one
waste bottle
Replacement Parts and Accessories

Rack Accessories

For part numbers for available racks, refer to Appendix B. To create your own Code 200­style rack, order the blank rack kit (part number 254461) and rivet gun (part number
4391002).
2504621 Adapter plate for installing Code 20- or 30-series
rack on locator plate
2504627 Locator tray for installing up to seven Code 20- or
30-series racks on the locator plate. Note: This locator tray cannot be used when an
optional 849 Multiple Injection Module is installed.
A-3
Appendix A

Cables and I/O Accessories

25061401 Level-sensing cable 36083121 Serial cable, IBM PS/2-type, 25 to 25 pin 36083122 Serial cable, IBM AT-type, 9-pin female to 25-pin male 36083123 Serial cable adapter, 9-pin female to 25-pin male 638308512 Terminal block connector, 8-pin 638310512 Terminal block connector, 10-pin 709910206 2-conductor interconnect wire, 6', for making
contact connections 36078143 Shielded GSIOC cable, 30" 6730504007 5.0A, T-5.0 Slo-Blo fuse

Miscellaneous

Replacement Parts and Accessories
2509211 Armlock with hex screws 4311403 9/64" ball driver (hex wrench for armlock)
A-4

Racks

The Quad-Z 215 Liquid Handler can be configured with a variety of rack types and sizes. The following pages describe the racks that can be purchased for use on the liquid handler. Refer to Section 2 for rack installation procedures.
B
B-1
Appendix B
Code 20 rack
Racks
For 108 vessels Material: polypropylene Vessels and maximum capacity: 10 x 100 mm tubes (4.5 mL) Part number: 150425 Required accessories: adapter plate (part number 2504621)
Code 21 rack
For 60 vessels Material: polypropylene Vessels and maximum capacity: 13 x 100 mm tubes (9 mL) Part number: 150422 Required accessories: adapter plate (part number 2504621)
Code 23 rack
For 44 vessels Material: polypropylene Vessels and maximum capacity: 17 x 55 mm vials (6.8 mL)
17 x 65 mm vials (8 mL) Part number: 150426 Required accessories: adapter plate (part number 2504621)
Code 24 rack
For 14 scintillation vials Material: polypropylene Vessels and maximum capacity: 28 x 60 mm (20 mL) Part number: 150427 Required accessories: adapter plate (part number 2504621)
B-2
Appendix B
Code 28 rack
For 108 vessels Material: polypropylene Vessels and maximum capacity: 10 x 65 mm tubes (3 mL)
10 x 75 mm tubes (3.5 mL) Part number: 150420 Required accessories: adapter plate (part number 2504621)
Code 29 rack
For 60 vessels Material: polypropylene Vessels and maximum capacity: 12 x 75 mm tubes (5 mL)
13 x 75 mm tubes (6 mL) Part number: 150429 Required accessories: adapter plate (part number 2504621)
Racks
Code 30 rack
For 60 vessels Material: aluminum Vessels and maximum capacity: 12 x 32 mm vials (2 mL) Part number: 2704430 Required accessories: adapter plate (part number 2504621) and thermostating cuvette (part number 2704429) or thermostating cuvette (part number 2759550) and 832 Temperature Regulator (part number 2759502)
Code 31 rack
Thermostated rack for 108 vessels Material: aluminum Vessels and maximum capacity: 7 x 40 mm vials (0.7 mL) Part number: 2704431 Required accessories: adapter plate (part number 2504621) and thermostating cuvette (part number 2704429) or thermostating cuvette (part number 2759550) and 832 Temperature Regulator (part number 2759502)
B-3
Appendix B
Code 32 rack
Racks
Thermostated rack for 60 vessels Material: aluminum Vessels and maximum capacity: 13 x 65 mm tubes (6 mL)
13 x 100 mm tubes (9 mL) Part number: 2704432 Required accessories: adapter plate (part number 2504621) and thermostating cuvette (part number 2704429) or thermostating cuvette (part number 2759550) and 832 Temperature Regulator (part number 2759502)
Code 33 rack
Thermostated rack for 14 scintillation vials Material: aluminum Vessels and maximum capacity: 28 x 60 mm (20 mL) Part number: 2704433 Required accessories: adapter plate (part number 2504621) and thermostating cuvette (part number 2704429) or thermostating cuvette (part number 2759550) and 832 Temperature Regulator (part number 2759502)
B-4
Code 200 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 13 x 100 mm (9 mL) Part number: 2504600
Appendix B
Code 201 rack
For two microplates, two Eppendorf vials, and two 13 x 100 mm tubes Material: aluminum Vessels and maximum capacity:96 well microplates
Eppendorf vials (1.5 mL) 13 x 100 mm tubes (9 mL)
Part number: 2504601
Code 201H rack
For two microplates, two Eppendorf vials, and two 13 x 100 mm tubes; with microplate covers Material: aluminum Vessels and maximum capacity: 96 well microplates
Eppendorf vials (1.5 mL) 13 x 100 mm tubes (9 mL)
Part number: 2504601H
Racks
Code 202 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 10.25 x 47 mm (2.5 mL) Part number: 2504602
Code 203 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 10.25 x 64 mm (4 mL) Part number: 2504603
B-5
Appendix B
Code 204 rack
Racks
For 27 scintillation tubes Material: aluminum Vessels and maximum capacity: 28 x 57 mm (20 mL) Part number: 2504604
Code 205 rack
For two deep-well microplates, two Eppendorf vials, and two 13 x 100 mm tubes Material: aluminum Vessels and maximum capacity: 96 deep-well microplates
Eppendorf vials (1.5 mL) 13 x 100 mm tubes (9 mL)
Part number: 2504605
B-6
Code 205H rack
For two deep-well microplates, two Eppendorf vials, and two 13 x 100 mm tubes; with microplate covers Material: aluminum Vessels and maximum capacity: 96 deep-well microplates
Eppendorf vials (1.5 mL) 13 x 100 mm tubes (9 mL)
Part number: 2504605H
Appendix B
Code 206 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 13 x 75 mm (7 mL) Part number: 2504606
Code 207 rack
For 75 vessels Material: aluminum Vessels and maximum capacity: 16 x 100 mm (12 mL) Part number: 2504607
Code 208 rack
Racks
For 70 vessels Material: aluminum Vessels and maximum capacity: 18 x 150 mm (25 mL) Part number: 2504608
Code 209 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 12 x 32 mm Part number: 2504609
B-7
Appendix B
Code 210 rack
Racks
For 75 vessels Material: aluminum Vessels and maximum capacity: 16 x 75 mm (11 mL) Part number: 2504610
Code 211 rack
For 9 Boston round screw-cap bottles Material: aluminum Vessels and maximum capacity: 48 x 113 mm (125 mL) Part number: 2504611
Code 212 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 48 13 x 100 mm (9 mL)
48 13 x 75 mm (7 mL) Part number: 2504612 Note: Place the shorter tubes into the notched tube locations.
Code 213 rack
For 74 vessels Material: aluminum Vessels and maximum capacity:37 16 x 100 mm (32 mL)
37 16 x 75 mm (25 mL) Part number: 2504613 Note: Place the shorter tubes into the notched tube locations.
B-8
Appendix B
Code 214 rack
For 96 vessels Material: aluminum Vessels and maximum capacity: 48 10.25 x 47 mm (2.5 mL)
48 10.25 x 64 mm (44 mL) Part number: 2504614 Note: Place the shorter tubes into the notched tube locations.
Code 216 rack
For 60 Waters WISP vials Material: aluminum Vessels and maximum capacity: 60 15 x 45 mm (4 mL) Part number: 2504616
Racks
Code 217 rack
For 96 tubes Material: aluminum Vessels and maximum capacity: 96 10 x 75 mm culture tubes (3.5 mL) Part number: 2504617
Code 218 rack
For two microplates and eight 10 x 75 culture tubes Material: aluminum Vessels and maximum capacity: two 96-well microplates and eight 10 x 75 mm culture tubes (3.5 mL) Part number: 2504618 Note: Instead of microplate(s), you can install Beckman modular reservoirs.
B-9
Appendix B
Code 222 rack
Racks
For 27 conical bottom tubes Material: aluminum Vessels and maximum capacity: 27 (50 mL) Part number: 2504622
Code 228 rack
For four reagent bottles Material: aluminum Vessels and maximum capacity: 4 (500 or 700 mL) Part number: 2504628
Code 505 rack
For 10 standard or deep-well microplates Material: aluminum Part number: 2504651
Code 505H rack
For 10 standard or deep-well microplates with hold-down covers Material: aluminum Part number: 2504651H
B-10
Appendix B

Peltier Racks

Code 242 Peltier rack
For 2 shallow, flat-bottom, 96-well microplates Material: aluminum Vessels and maximum capacity: 2 shallow, flat­bottom, 96-well microplates Part number: 25146331 Note: Order Peltier Controller (part number 2515331) and junction box (part number 2505332) separately.
Code 542 Peltier rack
Racks
For 10 shallow, flat-bottom, 96-well microplates Material: aluminum Vessels and maximum capacity: 10 shallow, flat­bottom, 96-well microplates Part number: 2514542 Note: Order Peltier Controller (part number 2515331) separately.
B-11
Appendix B
Code 852 Peltier rack
Racks
For two Becton Dickinson Falcon 96-well, shallow-well assay plates. U-bottom style Material: aluminum Part number: 2514852 Note: Order Programmable Peltier Controller (part number
2515850) separately.
Code 853 Peltier rack
For 96 12x32 mm 2 ml, flat-bottom vials Material: aluminum Part number: 2514853 Note: Order Programmable Peltier Controller (part number
2515850) separately.
Code 854 Peltier rack
For two Ritter or Beckman 96-well, deep-well microplates Material: aluminum Part number: 2514854 Note: Order Programmable Peltier Controller (part number
2515850) separately.
B-12

GSIOC Commands

The Gilson Serial Input Output Channel (GSIOC) is an asynchronous serial communications interface that enhances the power of your Gilson equipment.
The GSIOC incorporates an EIA RS-485 interface and allows up to 32 slave devices to be controlled from a single master in a multi-drop configuration.
Each slave device is identified by a unique number which must be known to the device and to the computer. The default ID code of the Quad-Z 215 is 22.
To control the Quad-Z 215 Liquid Handler via the GSIOC interface, you will need the following:
a personal computer with Microsoft® Windows® 98 or Windows NT® and any Gilson
control software or 706 Device Driver Software installed
an unused RS-232 communication port
C
From the controller, you:
specify the Quad-Z 215 as the device you want to control
issue commands that set operating parameters, control operation, or request
information from that instrument.
C-1
Appendix C

GSIOC Commands

There are two kinds of commands that you can send over the GSIOC:
Buffered commands send instructions to the liquid handler. These commands are executed
GSIOC Commands
one at a time.
Immediate commands request status information from the Quad-Z 215. These commands are executed immediately, temporarily interrupting other commands in progress.

GSIOC Command List

In the command list on the following pages, the GSIOC command must be entered in the proper upper or lower case format. If a buffered command requires additional information, you’ll see italicized text next to the command. The description of the command identifies what you need to enter in place of the italicized parameter. Also note that if a parameter is optional, it appears within brackets, [ ].
I - Immediate B - Buffered
Command Type Description
% I Identifies the selected slave device. Returns character
string: “215QADVx.yz”, where x, y, and z represent software version.
$ I Resets the Quad-Z 215. Returns “$” and resets the
instrument to its power-up state which:
clears any error state
relaxes all axes
opens contact closure inputs
C-2
Loading...