Giga-tronics 8650A Series Operation Manual

Page 1
Manual Part Number:
Print Date:
31470
E
April 2001
Series 8650A Universal Power Meters
Operation Manual
8650A
............................................Certified Product
Registrar: BSI, Certification No. FM 34226 Registered 04 June 1996 ❖ Amended 01 March 2000
Giga-tronics Incorporated 4650 Norris Canyon Road San Ramon, California 94583
925.328.4650 or 800.726.4442 925.328.4700 (Fax) 800.444.2878 (Customer Service) 925.328.4702 (CS Fax)
www.gigatronics.com
ISO 9001
............................................ Certified Process
Page 2
All technical data and specifications in this manual are subject to change without prior notice and do not represent a commitment on the part of Giga-tronics Incorporated.
© 2001 Giga-tronics Incorporated. All Rights Reserved.
Printed in the USA
WARRANTY
Giga-tronics Series 8650A instruments are warranted against defective materials and workmanship for two years from date of shipment. Giga-tronics will at its option repair or replace products that are proven defective during the warranty period. This warranty DOES NOT cover damage resulting from improper use, nor workmanship other than Giga-tronics service. There is no implied warranty of fitness for a particular purpose, nor is Giga-tronics liable for any consequential damages. Specification and price change privileges are reserved by Giga-tronics.
Model Numbers
The series 8650Aincludes two models: The single-channel Model 8651A and the dual-channel 8652A. Apart from the number of sensors they support, the two models are identical. Both models are referred to in this manual by the general term 8650A, except where it is necessary to make a distinction between the models.
Page 3
DECLARATION OF CONFORMITY
Application of Council Directive(s)
EMC Directive and Low Voltage Directive
Standard(s) to which Conformity is Declared:
EN50081-1 (1992) EN50082-1 (1997) EN61010-1 (1993)
Manufacturer’s Name
Manufacturer’s Address
Type of Equipment
Model Series Number
Model Numbers in Series
Giga-tronics Incorporated
4650 Norris Canyon Road
San Ramon, California 94583
U.S.A.
Universal Power Meters
Series 8650A
Models 8651A and 8652A
89/336/EEC and 73/23/EEC
EMC - Emissions EMC - Immunity Electrical Safety
I, the undersigned, hereby declare that the equipment specified above conforms to the above Directive(s) and Standard(s)
San Ramon, California
(Signature)
Thomas A. Kramer
(Full Name)
Director of Quality Assurance
(Position)
June 25, 1999
(Place)
(Date)
Page 4
Page 5
About This Manual ...................................................................................................... ix
Conventions ................................................................................................................ xi
Record of Manual Changes ........................................................................................ xiii
Special Configurations ................................................................................................ xv
1
Introduction
1.1 Description....................................................................................................1-1
1.2 Installation ....................................................................................................1-4
Contents
1.1.1 Features .......................................................................................1-1
1.1.2 Power Requirements ....................................................................1-2
1.1.3 Environmental Requirements ........................................................1-2
1.1.4 Items Furnished ............................................................................1-2
1.1.5 Items Required .............................................................................1-2
1.1.6 Tools and Test Equipment ............................................................1-2
1.1.7 Cooling .........................................................................................1-2
1.1.8 Cleaning .......................................................................................1-3
1.1.9 Receiving Inspection .....................................................................1-3
1.1.10 Preparation for Reshipment ..........................................................1-3
1.2.1 Safety Precautions ........................................................................1-4
1.2.2 Line Voltage and Fuse Selection ...................................................1-5
1.2.3 Power Sensor Precautions ............................................................1-5
1.2.4 The Rear Panel .............................................................................1-6
1.3 8650A System Specifications ........................................................................1-7
2
Front Panel Operation
2.1 Introduction ..................................................................................................2-1
2.1.1 The Front Panel ............................................................................2-1
2.2 8650A Configuration .....................................................................................2-4
2.2.1 Meter Setup .................................................................................2-5
2.2.2 Display Setup .............................................................................2-14
2.2.3 Rel ..............................................................................................2-17
2.2.4 Sensor Setup ..............................................................................2-17
2.3 Measurement Guide .................................................................................... 2-21
2.3.1 Using the Power Sweep Calibrator ............................................. 2-21
2.3.2 80701A Sensor Operation .......................................................... 2-21
2.3.3 Sensor Zero and Calibration ........................................................2-22
2.3.4 Measuring Source Output Power ................................................2-25
2.3.5 Using the Peaking Meter ............................................................2-25
2.3.6 High Power Level Measurements ...............................................2-26
2.3.7 Modulated Measurement Modes ................................................2-26
2.3.8 Measurement Collection Modes .................................................2-29
2.3.9 Mode Restrictions .......................................................................2-31
2.3.10 When to use CW, MAP and BAP ................................................2-31
2.3.11 Multi-Tone Tests ........................................................................2-31
2.3.12 Peak Hold ...................................................................................2-32
2.3.3.1 Calibration & Zeroing for High Power Sensors with Remov-
able Attenuators ....................................................... 2-23
2.3.3.2 Low Level Performance Check................................... 2-24
Manual 31470, Rev. E, April 2001 i
Page 6
Series 8650A Universal Power Meters
2.3.13 Crest Factor ................................................................................2-33
2.3.14 Burst Signal Measurements .......................................................2-35
2.3.15 Burst Start Exclude, Burst End Exclude ......................................2-36
2.3.16 Burst Dropout .............................................................................2-37
2.3.17 Optimizing Measurement Speed .................................................2-38
2.3.18 Peak Power Measurements ........................................................2-39
2.3.19 Measuring an Attenuator (Single Channel Method) ....................2-39
2.3.20 Improving Accuracy ....................................................................2-40
2.3.21 Performance Verification .............................................................2-41
2.3.22 Sources of Error ..........................................................................2-42
3
Remote Operation
3.1 Introduction...................................................................................................3-1
3.1.1 Sending Commands to the 8650A ................................................3-1
3.1.2 Polling ..........................................................................................3-2
3.1.3 Data Output Formats ....................................................................3-3
3.1.4 Power-On Default Conditions .......................................................3-4
3.2 SCPI Command Interface...............................................................................3-5
3.2.1 Sensor Calibration and Zero ..........................................................3-5
3.2.2 Sensor and Channel Configuration ................................................3-5
3.2.3 Measurement Triggering ..............................................................3-6
3.2.4 Memory Functions ........................................................................3-6
3.2.5 IEEE 488.2 Required Commands ..................................................3-6
3.2.6 Calculate Subsystem Commands ..................................................3-7
3.2.7 Sense Subsystem Commands ......................................................3-8
3.2.8 Trigger Subsystem Commands .....................................................3-9
3.2.9 GPIB Command Syntax ..............................................................3-10
3.2.10 Sensor Calibration and Zeroing ...................................................3-14
3.2.11 Reading Power Measurements ...................................................3-16
3.2.12 Instrument Triggering .................................................................3-18
3.2.13 Arming the Triggering Cycle .......................................................3-22
3.2.14 Channel Configuration ................................................................3-23
3.2.15 Cal Factor Correction ..................................................................3-25
3.2.16 High Speed Measurements .........................................................3-27
3.2.17 Peak Power Sensor Triggering ....................................................3-31
3.2.18 Averaging ...................................................................................3-33
3.2.19 Relative or Referenced Measurements ........................................3-34
3.2.20 Offsets ........................................................................................3-35
3.2.21 SRQ and Status Monitoring ........................................................3-36
3.2.22 Min/Max Configuration and Monitoring ......................................3-40
3.2.23 Limit Line Configuration and Monitoring .....................................3-42
3.2.24 Analog Output ............................................................................3-44
3.2.25 Saving and Recalling Configurations ...........................................3-46
3.2.26 Halting Operation .......................................................................3-46
3.2.27 Preset Configuration ...................................................................3-47
3.2.28 Identification Commands ............................................................3-49
3.2.29 Calibrator Controls ......................................................................3-50
3.2.30 Sensor EEPROM Commands ......................................................3-50
3.2.31 Self-Test .....................................................................................3-51
3.2.32 Error Messages ..........................................................................3-51
3.3 IEEE 488.2 Interface Command Codes.........................................................3-54
3.3.1 Command Syntax .......................................................................3-54
3.3.2 8650A Command Code Set ........................................................3-57
3.3.3 8540C Emulation Command Code Set ........................................3-60
3.3.4 HP436 Emulation Command Code Set ........................................3-63
3.3.5 HP437 Emulation Command Code Set ........................................3-64
3.3.6 HP438 Emulation Command Code Set ........................................3-66
ii Manual 31470, Rev. E, April 2001
Page 7
Preface
3.4 Analog Output.............................................................................................3-68
3.5 Averaging....................................................................................................3-69
3.5.1 Auto Averaging ..........................................................................3-69
3.5.2 Manual Averaging ...................................................................... 3-70
3.6 Cal Factors ..................................................................................................3-71
3.7 Calibration...................................................................................................3-72
3.7.1 Calibration Routine .....................................................................3-72
3.7.2 Calibrator Source ........................................................................3-73
3.7.3 Calibrator Test ............................................................................3-73
3.7.4 Channel Designation .................................................................. 3-74
3.8 Crest Factor.................................................................................................3-75
3.8.1 Enabling the Crest Factor Feature ............................................... 3-75
3.8.2 Reading the Crest Factor Value ...................................................3-75
3.9 Display Control............................................................................................3-76
3.10 Duty Cycle Commands................................................................................3-77
3.10.1 Activating or Deactivating a Duty Cycle ......................................3-77
3.10.2 Specifying a Duty Cycle ..............................................................3-77
3.10.3 Reading Duty Cycle Status ......................................................... 3-77
3.11 Time Gating Measurement ..........................................................................3-78
3.11.1 Description ................................................................................. 3-78
3.11.2 Time Gating Mode ......................................................................3-79
3.12 EEPROM ..................................................................................................... 3-81
3.12.1 EEPROM Cal Factors ..................................................................3-81
3.12.2 Frequency ...................................................................................3-82
3.13 Instrument Identification .............................................................................3-83
3.14 Learn Modes ...............................................................................................3-84
3.14.1 Learn Mode #1 ...........................................................................3-85
3.14.2 Learn Mode #2 ...........................................................................3-86
3.15 Limits .......................................................................................................... 3-87
3.15.1 Setting Limits ............................................................................. 3-87
3.15.2 Activating Limits ......................................................................... 3-87
3.15.3 Measuring with Limits ................................................................3-88
3.16 Measurement Collection Modes (Standard).................................................3-89
3.16.1 Measurement Triggering ............................................................3-89
3.16.2 Group Execute Trigger ................................................................ 3-90
3.17 Measurement Collection Modes (Fast) ........................................................3-91
3.17.1 General .......................................................................................3-91
3.17.2 Data Output Formats for Fast Modes .........................................3-92
3.17.3 Fast Buffered Mode .................................................................... 3-93
3.17.4 Swift Mode .................................................................................3-95
3.17.5 Fast Modulated Mode ................................................................ 3-97
3.18 Measurement Mode Commands .................................................................3-98
3.18.1 CW Mode ................................................................................... 3-98
3.18.2 MAP Mode ................................................................................. 3-98
3.18.3 PAP Mode ..................................................................................3-99
3.18.4 BAP Mode ..................................................................................3-99
3.18.5 Peak Mode .................................................................................3-99
3.18.6 Measurement Mode Query .......................................................3-100
Manual 31470, Rev. E, April 2001 iii
Page 8
Series 8650A Universal Power Meters
3.19 Advanced Features....................................................................................3-101
3.19.1 Burst Start Exclude ...................................................................3-101
3.19.2 Burst End Exclude ....................................................................3-101
3.19.3 Burst Dropout Tolerance ...........................................................3-102
3.19.4 Min/Max Power Value ..............................................................3-102
3.19.5 Offset Commands .....................................................................3-103
3.19.6 Measured Offset Entry ..............................................................3-105
3.19.7 Peak Hold .................................................................................3-106
3.19.8 Peak Power Sensor Commands ................................................3-107
3.19.9 Preset .......................................................................................3-109
3.19.10 Relative Measurements ............................................................3-110
3.19.11 Resolution ................................................................................3-110
3.19.12 Sensor Selection .......................................................................3-111
3.19.13 Status .......................................................................................3-111
3.19.14 Setup ........................................................................................3-117
3.19.15 Store and Recall ........................................................................3-117
3.19.16 Units .........................................................................................3-118
3.19.17 VPROPF Feature .......................................................................3-118
3.19.18 Zeroing .....................................................................................3-120
3.19.19 Histograms, CDF and CCDF ......................................................3-121
3.19.20 Strip Chart ................................................................................3-122
3.19.21 Statistics ..................................................................................3-123
4
Performance Verification
4.1 Introduction...................................................................................................4-1
4.2 Equipment Required......................................................................................4-1
4.3 Calibrator Verification Procedure....................................................................4-2
4.3.1 Calibrator Output Power Reference Level ......................................4-2
4.3.2 Calibrator Frequency Check ..........................................................4-4
4.4 Performance Verification Tests ......................................................................4-5
4.4.1 Equipment Required .....................................................................4-5
4.4.2 Instrument Plus Power Sensor Linearity .......................................4-6
4.4.3 GPIB Port Check ...........................................................................4-8
A
Typical Applications Programs
A.1 Continuous Data Reading............................................................................. A-1
A.2 Remote Calibration of a Sensor..................................................................... A-1
A.3 Speed Tests: Normal and Swift .................................................................... A-2
A.4 Swift Demo 1: FREERUN............................................................................. A-4
A.5 Swift Demo 2: GET...................................................................................... A-5
A.6 Fast Buffered Demo: POST GET .................................................................. A-6
A.6.1 Fast Buffered Demo: POST TTL.................................................... A-7
iv Manual 31470, Rev. E, April 2001
Page 9
B
Power Sensors
B.1 Introduction ................................................................................................. B-1
B.2 Power Sensor Selection................................................................................ B-1
C
Options
C.1 Introduction ................................................................................................. C-1
C.2 Option 01: Rack Mount Kit ........................................................................... C-1
C.3 Option 03: Rear Panel Sensor In & Calibrator Out Connectors, Model 8651A... C-1
C.4 Option 04: Rear Panel Sensor In & Calibrator Out Connectors, Model 8652A... C-1
C.5 Option 05: Soft Carrying Case ...................................................................... C-2
C.6 Option 07: Side-mounted Carry Handle ........................................................ C-2
Preface
B.2.1 Power Sensor Selection Charts ..................................................... B-2
B.2.2 Modulated Sensor Specifications .................................................. B-7
B.2.3 Directional Bridges...................................................................... B-11
C.7 Option 08: Transit Case................................................................................ C-2
C.8 Option 09: Dual Power Meter Rack Mount Kit.............................................. C-2
C.9 Option 10: Assembled Dual Power Meter Rack Mount Kit............................ C-2
C.10 Option 12: 1 GHz, 50 MHz Switchable Calibrator......................................... C-2
C.11 Option 13: Rear Panel Sensor In Connectors, Model 8651A ......................... C-2
C.12 Option 14: Rear Panel Sensor In Connectors, Model 8652A ......................... C-2
D
Menu Structure
D.1 Introduction ................................................................................................. D-1
Index
Series 8650A Universal Power Meters Index ....................................................... Index-1
Manual 31470, Rev. E, April 2001 v
Page 10
Series 8650A Universal Power Meters
Illustrations
Figure 1-1: AC Power Connector & Fuse Holding...................................................1-5
Figure 1-2: The 8650A Rear Panel ..........................................................................1-6
Figure 2-1: The 8652A Front Panel.........................................................................2-1
Figure 2-2: The Main Menu ....................................................................................2-4
Figure 2-3: Power Meter Configuration Display ......................................................2-6
Figure 2-4: Illustration of the Strip Chart.................................................................2-8
Figure 2-5: Illustration of External Gating................................................................2-9
Figure 2-6: Illustration of External Triggering........................................................2-10
Figure 2-7: Illustration of Burst Edge Configuration ..............................................2-11
Figure 2-8: Illustration of the Histogram ...............................................................2-12
Figure 2-9: Illustration of the CDF Curve...............................................................2-13
Figure 2-10: Illustration of the CCDF Curve............................................................2-13
Figure 2-11: The Configuration Display Menu.........................................................2-14
Figure 2-12: The Data Line Configuration Menu .....................................................2-15
Figure 2-13: The Sensor Setup Menu .....................................................................2-17
Figure 2-14: Setup for Sensor Calibration ...............................................................2-22
Figure 2-15: The Peaking Meter..............................................................................2-25
Figure 2-16: Burst Measurement ............................................................................2-27
Figure 2-17: Delay and Delay Offsets......................................................................2-30
Figure 2-18: Peak Hold ...........................................................................................2-32
Figure 2-19: Crest Factor ........................................................................................2-33
Figure 2-20: Burst Start Exclude & Burst End Exclude............................................2-36
Figure 2-21: Burst Dropout.....................................................................................2-37
Figure 3-1: SCPI Subsystem Model ........................................................................3-5
Figure 3-2: CALCulat Subsystem Commands .........................................................3-7
Figure 3-3: SENSe Subsystem Command Tree .......................................................3-8
Figure 3-4: TRIGger Subsystem Command Tree .....................................................3-9
Figure 3-5: TRIGger Subsystem Command Tree ...................................................3-18
Figure 3-6: The SCPI Status Structure Registers ...................................................3-38
Figure 4-1: Calibrator Output Test Setup................................................................4-2
Figure 4-2: Power Linearity Test Setup...................................................................4-6
Figure B-1: 80401A Modulation-Related Uncertainty ............................................ B-8
Figure B-2: 80601A Modulation-Related Uncertainty ............................................ B-9
Figure B-3: 80701A Modulation-Related Uncertainty .......................................... B-10
Figure D-1: The Series 8650A Menu Tree ............................................................. D-1
Figure D-2: Peak Sensor Setup A Menu Structure................................................. D-2
Figure D-3: Modulation Sensor B Setup Menu Structure....................................... D-3
Figure D-4: Meter Setup Menu Structure .............................................................. D-5
Figure D-5: Display Setup Menu Structure ............................................................ D-6
vi Manual 31470, Rev. E, April 2001
Page 11
Preface
Tables
Table 1-1: Measurement Rates........................................................................... 1-10
Table 3-1: Implemented IEEE Standards............................................................... 3-1
Table 3-2: IEEE Required Command Codes........................................................... 3-6
Table 3-3: SCPI Command Syntax...................................................................... 3-11
Table 3-4: Reset and Power on Default Commands............................................ 3-47
Table 3-5: SCPI Standard Error Messages .......................................................... 3-51
Table 3-6: Device Specific Error Messages ......................................................... 3-52
Table 3-7: 8650A IEEE 488.2 Command Set....................................................... 3-57
Table 3-8: 8540C Emulation Command Set........................................................ 3-60
Table 3-9: HP436 Emulation Command Set........................................................ 3-63
Table 3-10: HP437 Emulation Command Set........................................................ 3-64
Table 3-11: HP438 Emulation Command Set........................................................ 3-66
Table 3-12: Measurement Setting Target Default Values...................................... 3-69
Table 3-13: Numbering Averaging........................................................................ 3-70
Table 3-14: Learn Mode #1 Output Format .......................................................... 3-85
Table 3-15: Preset (Default) Conditions .............................................................. 3-109
Table 3-16: Status Byte and Service Request Mark ............................................ 3-111
Table 3-17: Error Code Returned in Position AA ................................................. 3-114
Table 3-18: Error Code Returned in Position aa .................................................. 3-115
Table 3-19: Other Codes in the Status Message................................................. 3-116
Table 4-1: Required Equipment List...................................................................... 4-1
Table B-1: Power Sensor Selection Guide ............................................................ B-2
Table B-2: Power Sensor Cal Factor Uncertainties ............................................... B-6
Table B-3: Modulated Sensor Specifications ........................................................ B-7
Table B-4: Directional Bridge Selection Guide .................................................... B-11
Manual 31470, Rev. E, April 2001 vii
Page 12
Series 8650A Universal Power Meters
viii Manual 31470, Rev. E, April 2001
Page 13

About This Manual

About This Manual
About This ManualAbout This Manual
This operation manual covers the operation and performance verification of the Giga-tronics Series 8650A Universal Power Meters:
Preface:
In addition to a comprehensive Contents and general information about the manual, the Preface also contains a record of changes made to the manual since its publication, and a description of Special Configurations. If you have ordered a user-specific manual, please refer to page xv for a description of the special configuration.
Chapters:
1 – Introduction
This chapter contains a brief introduction to the instrument and its performance parameters.
2 – Front Panel Operation
This chapter is a guide to the instrument’s front panel keys, display and configuration menus.
3 – Remote Operation
This chapter is a guide to the instrument’s GPIB remote control interface.
4 – Performance Verification
This chapter defines the procedures to verify the performance of the 8650A Meter.
Appendices:
A - Sample Programs
This appendix provides you with examples of programs for controlling the 8650A remotely over the GPIB.
B – Power Sensors
This appendix provides selection data, specifications, and calibration procedures for power sensors.
C - Options
This appendix describes options available for the Series 8650A.
D – Menu Structure
This is a pictorial mapping of the various menus used in the Series 8650A. Use this appendix as a map to locate the function you want to perform.
Manual 31470, Rev. E, April 2001 ix
Page 14
Series 8650A Universal Power Meters
Index:
A comprehensive word index of the various elements of the 8650A manual.
Changes that occur after publication of the manual, and Special Configuration data will be inserted as loose pages in the manual binder. Please insert and/or replace the indicated pages as detailed in the Technical Publication Change Instructions included with new and replacement pages.
x Manual 31470, Rev. E, April 2001
Page 15

Conventions

Conventions
ConventionsConventions
The following conventions are used in this product manual. Additional conventions not included here will be defined at the time of usage.
Warning
WARNING
The WARNING statement is encased in gray and centered in the page. This calls attention to a situation, or an operating or maintenance procedure, or practice, which if not strictly corrected or observed, could result in injury or death of personnel. An example is the proximity of high voltage.
Caution
CAUTION
Notes
☛☛☛☛
The CAUTION statement is enclosed with single lines and centered in the page. This calls attention to a situation, or an operating or maintenance procedure, or practice, which if not strictly corrected or observed, could result in temporary or permanent damage to the equipment, or loss of effectiveness.
NOTE: A NOTE Highlights or amplifies an essential operating or maintenance procedure, practice, condition or statement.
Manual 31470, Rev. E, April 2001 xi
Page 16
Series 8650A Universal Power Meters
xii Manual 31470, Rev. E, April 2001
Page 17

Record of Manual Changes

Record of Manual Changes
Record of Manual ChangesRecord of Manual Changes
This table is provided for your convenience to maintain a permanent record of manual change data. Corrected replacement pages will be issued as Technical Publication Change Instructions, and will be inserted at the front of the binder. Remove the corresponding old pages, insert the new pages, and record the changes here.
Change
Instruction
Number
Change
Instruction
Date
Date
Entered Comments
Manual 31470, Rev. E, April 2001 xiii
Page 18
Series 8650A Universal Power Meters
xiv Manual 31470, Rev. E, April 2001
Page 19

Special Configurations

Special Configurations
Special ConfigurationsSpecial Configurations
When the accompanying product has been configured for user-specific application(s), supplemental pages will be inserted at the front of the manual binder. Remove the indicated page(s) and replace it (them) with the furnished Special Configuration supplemental page(s).
Manual 31470, Rev. E, April 2001 xv
Page 20
Series 8650A Universal Power Meters
xvi Manual 31470, Rev. E, April 2001
Page 21

1.1 Description

This chapter contains the Description, Installation procedures, and System Specifications.
The Series 8650A Universal Power Meters are digital-controlled, self-calibrating instruments that can measure RF and microwave signal power over a wide range of frequencies and levels in a variety of measurement modes. They can be operated locally from the front panel or remotely over the General Purpose Interface Bus (GPIB). See Section 1.3 for performance specifications.
The Series 8650A is available as the single-channel Model 8651A or the dual-channel Model 8652A, which can simultaneously measure and display signal data for two channels.
The 8650A and the Series 80601A and 80701A power sensors offer enhanced performance in the measurement of complex modulation signals in the communication industry. The 8650A maintains the functionality of Giga-tronics 8540B and 8540C power meters, and compatibility with all existing power sensor models.
1

Introduction

NOTE: The optional 1 GHz Calibrator is required for operation with Series 80701A power
☛☛☛☛
sensors (see Option 12 in Appendix C).

1.1.1 Features

CW, peak, and modulation power meter with burst mask testing
More than 26,000 readings/second in the Fast Buffered Mode (GPIB only)
90 dB dynamic range CW sensors
+0.3% linearly per degree Centigrade of temperature change
True dual-channel display
SCPI Command modes (GPIB only)
HP 438A, 437B, and 436 native mode emulation (GPIB only)
Giga-tronics 8540B and 8540C native mode emulation (GPIB only)
EEPROM based CAL FACTOR correction sensors
Modulated Average Power (MAP) mode
Pulse Average Power (PAP) mode
Burst Average Power (BAP) mode
Triggered (time-gated) measurement mode
Wide modulation bandwidth – The 8650A is capable of accurately measuring signals with modulation frequencies up to 10 MHz with the 80701A sensors
Manual 31470, Rev. E, April 2001 1-1
Page 22
Series 8650A Universal Power Meters
Dual-channel modulated measurements with the 8652A and the 80601A or 80701A power sensors
Strip Chart function to view the power over a selective period of 40 seconds to 200 minutes
Statistical functions including mean and standard deviation, and graphical displays of Histogram, Cumulative Distribution Function (CDF) and Complementary CDF (CCDF)
Upgradable firmware via the RS-232 port

1.1.2 Power Requirements

100/120/220/240 Vac ±10%, 48-440 Hz, 20 W, typical. See Section 1.2.1 for details to set the voltage and install the correct fuse for the area in which the instrument will be used.

1.1.3 Environmental Requirements

The Series 8650A instruments are type tested to MIL-PRF-28800F, Class 3 for all departments and agencies of the Department of Defense applications except as follows:
Operating temperature range is 0 °C to 55 °C (calibrator operating temperature range is 5 °C to 35 °C)
Operating the 8651A/8652A Power Meters in a high level RF field (approximately 3 V/m) may degrade performance, this degradation occurs at measured levels below -36 dBm and when the frequency of the field is nominally between 50 and 1000 MHz
Non-operating (storage) temperature range is -40 °C to +70 °C
Relative humidity is limited to 95% non-condensing

1.1.4 Items Furnished

In addition to options and/or accessories specifically ordered, items furnished with the instrument are:
1 ea. - Power Cord 1 ea. - Detachable Sensor Cable (for Model 8651A), or 2 ea. - Detachable Sensor Cables (for Model 8652A) 1 ea. - Operation Manual (P/N 31470)

1.1.5 Items Required

The 8650A requires an external power sensor; see Appendix B for Power Sensor Specifications.
NOTE: The optional 1 GHz Calibrator is required for operation with Series 80701A power
☛☛☛☛
sensors (see Option 12 in Appendix C).

1.1.6 Tools and Test Equipment

No special tools are required to operate the 8650A.

1.1.7 Cooling

No cooling is required if the instrument is operated within its specified operating temperature range (0 to 50 ° C).
1-2 Manual 31470, Rev. E, April 2001
Page 23

1.1.8 Cleaning

The front panel can be cleaned using a cloth dampened with a mild detergent; wipe off the detergent residue with a damp cloth and dry with a dry cloth. Solvents and abrasive cleaners should not be used.

1.1.9 Receiving Inspection

Use care in removing the instrument from the carton and check immediately for physical damage, such as bent or broken connectors on the front and rear panels, dents or scratches on the panels, broken extractor handles, etc. Check the shipping carton for evidence of physical damage and immediately report any damage to the shipping carrier.
Each Giga-tronics instrument must pass rigorous inspections and tests prior to shipment. Upon receipt, its performance should be verified to ensure that operation has not been impaired during shipment. Follow the installation instructions in Section 1.2 and the operating instructions in Chapter 2 or 3.

1.1.10 Preparation for Reshipment

Follow these instructions if it is necessary to return the product to the factory.
To protect the instrument during reshipment, use the best packaging materials available. If possible use the original shipping container. If this is not possible, a strong carton or a wooden box should be used Wrap the instrument in heavy paper or plastic before placing it in the shipping container. Completely fill the areas on all sides of the instrument with packaging material. Take extra precautions to protect the front and rear panels.
Introduction
Seal the package with strong tape or metal bands. Mark the outside of the package
DELICATE INSTRUMENT”
regarding reshipment, please reference the full model number and serial number. If the instrument is being reshipped for repair, enclose all available pertinent data regarding the problem that has been found.
NOTE:
☛☛☛☛
Customer Service so that a return authorization number (RMA) can be assigned via e-mail at repairs@gigatronics.com or at 800.444.2878 (The 800 number is only valid within the US). You may also try our domestic line at 925.328.4650 or Fax at 925.328.4702.
If you are returning an instrument to Giga-tronics for service, first contact
. If corresponding with the factory or local Giga-tronics sales office
“FRAGILE —
Manual 31470, Rev. E, April 2001 1-3
Page 24
Series 8650A Universal Power Meters

1.2 Installation

Select the correct operating voltage and install the proper fuse in this housing. Refer to Section 1.2.2, Line Voltage and Fuse Selection for instructions on how to select the voltage and replace the fuse. Observe the following Safety Precautions when installing the 8650A Power Meter. See Section 1.2.4 for connecting to the rear panel. Also see Section 2.3.3 for instructions on how to connect and calibrate power sensors.
CAUTION
Do not connect main power to the unit until you have checked the required operating voltage and fuse rating. The instrument can be damaged if connected to a source voltage with the line voltage se­lector set incorrectly.

1.2.1 Safety Precautions

This 8650A has a 3-wire power cord with a 3-terminal polarized plug for connection to the power source and safety-ground. The ground (or safety ground) is connected directly to the chassis.
WARNING
If a 3-to-2 wire adapter is used, connect the ground lead from the adapter to earth ground. Failure to do this can cause the instru­ment to float above earth ground, posing a shock hazard.
The 8650A is designed for international use with source voltages of 100, 120, 220, or 240 Vac, ±10% at 48 to 440 Hz. The 8650A uses an internationally approved connector that includes voltage selection, fuse, and filter for RFI protection (see Figure 1-1).
1-4 Manual 31470, Rev. E, April 2001
Page 25

1.2.2 Line Voltage and Fuse Selection

The instrument is shipped in an operational condition and no special installation procedures are required except to check and/or set the operating voltage and fuse selection as described in the following.
When the instrument is shipped from the factory, it is set for a power line voltage (120 Vac for domestic destinations). The power line fuse for this setting is 0.50 A Slo-Blo. If the source voltage is to be 220 to 240 Vac, the fuse must be changed to 0.35 A Slo-Blo (see Figure 1-1).
1
1
VOLTAGE SELECTION WHEEL
COVER
FUSE AND FUSE HOLDER
0
1
2
0
Introduction
AC POWER INPUT
Figure 1-1: AC Power Connector & Fuse Housing
The voltage selector and fuse holder are both contained in the covered housing directly above the AC power connector on the rear panel. To gain access to them, use a small screwdriver or similar tool to snap open the cover and proceed as follows:
1. To change the voltage setting:
Use the same tool to remove the voltage selector (a small barrel-shaped component marked with voltage settings). Rotate the selector so that the desired voltage faces outward and replace the selector back in its slot. Close the housing cover; the appropriate voltage should be visible through the window (see Figure 1-1).
2. To replace the fuse:
Pull out the small drawer on the right side of the housing (marked with an arrow) and remove the old fuse. Replace with a new fuse, insert the drawer and close the housing cover (see Figure 1-1).

1.2.3 Power Sensor Precautions

Power sensor safety precautions, selection, specifications, and calibration are detailed in Appendix B to this manual.
Manual 31470, Rev. E, April 2001 1-5
Page 26
Series 8650A Universal Power Meters

1.2.4 The Rear Panel

The rear panels for the Models 8651A and 8652A are identical and are illustrated in Figure 1-2. Any options that have been installed in the unit will be noted on the serial number tag. Refer to the Special Configurations section in the preface of this manual for detailed information about installed options or other special configurations. Appendix C contains information on all available options for the 8650A.
U.S.Patent 4,794,325
!
WARNING
For continued fire protection replace fuse with same type and rating
WARNING
No operator serviceable parts inside. Refer servicing to service trained personnel
OPTION 01
Fuse
110-120V
T250 .50A
220-240V
T250 .35A
-- Line
25VA MAX
LINE VOLTAGE
SELECTION
120Vac
Analog
Out A
GPIB
Analog
Out B
Trigger
RS-232
.
Input
Cal
V
F
In
A
B
Figure 1-2: The 8650A Rear Panel
Line Voltage Selection and Fuse
Select the correct operating voltage and install the proper fuse in this housing. Refer to Section 1.2.2 for instructions and precautions on how to select the voltage and replace the fuse.
Inputs/Outputs
Four BNC-type connectors interface the 8650A to other equipment
Trigger Input
accepts a TTL input for triggering and gating measurements as defined by the
Gate/Trigger menu selections, or under GPIB control. Maximum input without damage is 15 V.
V
PROP
accepts a voltage input that is proportional to frequency and causes the 8650A to
F In
apply appropriate frequency-related cal factors. Maximum input without damage is 15 V.
Analog Out A and Analog Out B
each provide an output voltage that is proportional to
the measured power level of the respective sensors connected to the front panel.
NOTE:
☛☛☛☛
8651A) or Option 04 (Model 8652A) is installed. The Sense In and the Calibrator Out
The Sense In connectors will be relocated to the rear panel when Option 03 (Model
connectors will be relocated to the rear panel when Option 13 (Model 8651A) or Option 14 (Model 8652A) is installed (descriptions of these options are in Appendix C).
Remote Interface
1-6 Manual 31470, Rev. E, April 2001
is a DB-24 connector to interface the 8650A to a host computer over the GPIB.
GPIB RS-232
is a DB-9 connector for interfacing the meter with serial communication equipment.
Page 27

1.3 8650A System Specifications

Power Meter
Introduction
Frequency Range:
Power Range:
Single Senso r
Dynamic Range:
CW Power Sensors: 90 dB
Peak Power Sensors: 40 dB Peak, 50 dB CW
Modulation Sensors: 87 dB CW; 80 dB MAP/PAP; 60 dB BAP
Display Resolution:
Measurement Modes:
Averaging:
dB Rel and Offset:
Configuration Storage
Registers:
Power Measurements and
Display Configuration:
10 MHz to 40 GHz
-70 dBm to +47 dBm (100 pW to 50 Watt)
1
User-selective from 1 dB to 0.001 dB in Log mode and from 1 to
4 digits of display resolution in Linear mode.
CW, Peak, MAP, BAP, PAP
User-selective auto-averaging or manual, 1 to 512 readings.
Timed averaging from 20 ms to 20 seconds. Automatic noise compensation in auto-averaging mode.
Allows both relative and offset readings. Power display can be
offset by -99.999 dB to +99.999 dB to account for external loss/ gain.
Up to 20 front panel setups plus a last instrument state at power-
down to be stored and recalled from non-volatile memory.
Any four of the following channel configurations simultaneously: A, B, A/B, B/A, A-B, B-A, DLYA, DLYB, Min/Max, Bar Graph/ Peaking Meter, Peak Hold, Crest Factor, or Mean & Std Deviation. Alternately, full-screen graphic display of Histogram, Strip Chart, Cumulative Distribution Function (CDF) and Complementary CDF (CCDF) functions.
1
1
Sampling:
CW Mode and Modulation
Mode: 2.5 to 5.0 Mhz, asynchronous
Analog Bandwidth:
CW Mode:
Modulation Mode: >10 MHz
Manual 31470, Rev. E, April 2001 1-7
3 kHz
Page 28
Series 8650A Universal Power Meters
Time Gated Measurements:
Gate Polarity: Specifies the external signal TTL high or low level as true for
Trigger Delay: 0 to 327 ms
Gate Time: 10
Holdoff Time: 0 to 327 ms
External Trigger Polarity: Positive or negative leading edge
Delay & Range Accuracy: +1.5
Settability: 5
Trigger Signal: Standard TTL levels
defining the gated time.
µ
s to 327 ms
µ
s or 100 ppm of the set time, whichever is greater
µ
s steps or selective by cursoring to specific digits
Accuracy
50 MHz Calibrator
Calibrator: +20 to -30 dBm power sweep calibration signal to dynamically
Frequency: 50 MHz nominal
0.0dBm Accuracy: ±1.2% worst case for one year over a temperature range of 5 to
VSWR: <1.05 (Return Loss >33 dB) @ 0 dBm
Connector: Type N, 50
Standard
linearize the sensors.
35 °C.
1 GHz Calibrator
(Option 12)
Calibrator: +20 to -30 dBm power sweep calibration signal to dynamically
Frequency: 1 GHz, nominal
0.0dBm Accuracy: ± 1.2% worst case for one year over a temperature range of 5 to
Connector: Type N, 50
800 MHz to 1 GHz
Synthesizer (Option 12)
Power Range: +15 to -30 dBm, settable in 1 dB steps
Frequency: 800 MHz to 1 GHz, settable in 1 MHz steps
Power Stability: <0.1 dB/hour
Frequency Accuracy: ±0.05%
☛☛☛☛
NOTE:
0 dBm.
Power accuracy for Option 12 is only guaranteed while in calibration mode at 1 GHz,
Required for 80701A Series Sensors (see Option 12 in Appendix C).
linearize the sensors.
35 °C.
VSWR: <1.07 (Return Loss >30 dB) @ 0 dBm
1-8 Manual 31470, Rev. E, April 2001
Page 29
Instrumentation Linearity
3
2
1
0
-1
-2
-3
80301A 80310A 80320A 80321A 80322A 80325A 80330A
80401A,80601A (CW)
80701 (CW)
-70
-64
-60
-50
-40
-40
-30
-67
-64
-60
-54
-50
-40
-30
-30
-20
-57
-54
-50
-44
-40
-30
-20
-20
-10
-47
-44
-40
-34
-30
-20
-10
-10 0
-37
-34
-30
-24
-20
-10 0 0
10
-27
-25
-20
-14
-10 0 0
10 20
-17
-16
-10
-4 0
10 20 20
-7
-7
0
6 10 20 20 30
3
3
10 16 20 30 40 40
13 13
20 26 30 40 44 50
20 20
Input (dBM)
SENSORS
Typical Error (dB)
System Linearity at 50 MHz: ±0.02 dB over any 20 dB range from -70 to +16 dBm
±0.02 dB (±.05 dB/dB) from +16 to +20 dBm ±0.04 dB from -70 to +16 dBm
The following chart shows linearity plus worst case zero set and
noise vs. input power.
Introduction
1
Temperature Coefficient of
Linearity: <0.3%/ °C temperature change following Power Sweep
calibration. 24-hour warm-up required.
Zeroing Accuracy: (CW)
Zero Set: <±50 pW, <± 100 pW with 80400A, 80600A series Modulation
Power Sensors. <±200 pW with 80700A Series Sensors.
<±100 pW during 1 hour
Zero Drift (during 1 hour): <±200 pW with 80400A and 80600A Series Modulation Sensors
<±400 pW with 80700A Series Sensors
Noise: <±50 pW, <±100 pW with 80400A and 80600A Series
modulation power sensors. <±200 pW with 80700A Series Sensors. Measurable over any 1 minute interval after zeroing, three standard deviations.
Notes:
1. Depending on sensor used (see Power Sensor details in Appendix B).
2. Zero Drift Measurement a. Set the meters Average to 512. Perform Calibration. Connect a 50-ohm load to the sensor after Calibration and
Zero meter.
b. Temperature stabilize at 25 °C for 24 hours. c. After the 24 hour stabilization, perform a Zero Drift test. d. Zero the meter and take an initial measurement reading. e. Continue taking one reading every 10 minutes until 6 readings have been taken.
Plot the 6 readings, Zero Drift should be ±200pW or ±400pW, depending on the sensor.
Manual 31470, Rev. E, April 2001 1-9
Page 30
Series 8650A Universal Power Meters
Measurement Rates
Table 1-1 illustrates typical maximum measurement rates for different measurement collection modes. The rate of measurement depends on several factors including the controller speed and the number of averages. The Fast Buffered Mode speed does not include bus communication time. Measurement speed increases significantly using the 8650A Fast Buffered Mode. Storing data in the power meter’s memory for later downloading to your controller reduces GPIB protocol overhead. Up to 5000 readings can be buffered.

Table 1-1: Measurement Rates

Measurement
Collection Mode
Normal (TR3), Continuous Single Readings
Swift Mode, Continuous or Buffered, Bus/TTL triggered >1750 800
Fast Buffered Mode, Buffered Data, Time Interval = 0
Fast Modulated Mode, Continuous Single Readings
Readings per Second
(CW Measurement)
>300 150
26,000 N/A
N/A 800
Readings per Second
(MAP, PAP, BAP
Measurement)
Individual data points are read immediately after measurement in the Normal mode. The Normal mode and the Swift mode both slow down at low power levels (<-37 dBm for Standard Sensors) to average the effects of noise. The Swift mode allows triggering of individual data points and can store the data in the 8650A memory. Measurement timing of individual data points is controlled by setting the time interval (0 to 5000 ms) between the data points following a trigger.
Remote Operation
GPIB Interface: All front panel operations and some GPIB-only operations to be
Interrupts: SRQs are generated for the following conditions:
remotely programmed in IEEE 488.2 or IEC-625 formats.
Power Up, Front Panel key actuation, Operation Complete and Illegal Command.
Fast Buffered Mode Controls
Trigger Source: TTL or GPIB
Data Buffer Control: Pre- or Post-measurement data is collected immediately either
Time Interval: TIME ### - controls time interval in milliseconds between
1-10 Manual 31470, Rev. E, April 2001
before or after receipt of the TTL or GPIB trigger.
measurements. Accurate to 5%, typical.
Page 31
Inputs/Outputs
V
F Input (BNC): Recalls cal factors using source V
PROP
Introduction
F output. Corrects power readings for sensor frequency response using sweeper voltage output. Input resistance = 50K. Does not operate in the fast measurement collection modes (normal mode only).
PROP
Analog Output (BNC): Provides an output voltage of 0 to 10V from either Channel A or
Trigger Input (BNC): Accepts a TTL trigger input signal for swift and fast measurement
GPIB connector: Interfaces the power meter to a host computer for remote
RS-232 connector: Interfaces the power meter to serial communications equipment,
Channel B in either Log or Lin units. Does not operate in the swift and fast measurement buffered modes.
buffered modes, and time gating mode.
programming using SCPI, IEEE 488.2 and IEC-625 coding, also emulation modes.
using RS-232 format.
General Specifications
Temperature Range:
Operating: 0 to 55 °C (32 to 13 2 °F)
Storage: -40° to 70 °C (-40° to 158 °F)
Power Requirements:
100/120/220/240Vac <± 100 pW 10%, 48 to 440 Hz, 20 VA
typical
Physical Characteristics:
Dimensions: 215 mm (8.4 in) wide, 89 mm (3.5 in) high, 368 mm (14.5 in)
Weight: 4.55 kg (10 lbs)
deep
Options
Refer to Appendix C for descriptions of options.
Power Sensors
See Appendix B for power sensor selection, specifications and calibration data.
Manual 31470, Rev. E, April 2001 1-11
Page 32
Series 8650A Universal Power Meters
1-12 Manual 31470, Rev. E, April 2001
Page 33

2.1 Introduction

This chapter describes how to operate the Series 8650A Universal Power Meter using the display and controls on the front panel. It includes descriptions of the front and rear panels, configuration, display menus and practical applications.
See Chapter 3 for instructions for remote operation of the 8650A over the General Purpose Interface Bus (GPIB) and RS-232 serial communication devices.

2.1.1 The Front Panel

Although the 8650A has different modes of operation, the front panel is simple in design and easy to use. The instrument is configured and controlled by means of the dedicated hardkeys and the display menus, which are accessed and controlled with the data interaction softkeys.
The dual-channel Model 8652A front panel is illustrated in Figure 2-1. The single-channel Model 8651A is the same in appearance but does not include the Sensor B connector.
2

Front Panel Operation

Figure 2-1: The 8652A Front Panel
Manual 31470, Rev. E, April 2001 2-1
Page 34
Series 8650A Universal Power Meters
Dedicated Hardkeys
The dedicated hardkeys are located on the right side of the front panel and function as described below. In this manual, instructions to press a dedicated hardkey are with the appropriate key title in bold uppercase enclosed in brackets, such as ‘press
Cursor Keys:
These four keys are arranged in a diamond pattern and move the highlighted item (cursor) in the display to the desired location. A small diamond symbolizing the cursor keys appears in menus, with an arrow in the keys that are effective for that menu. If more than one field is highlighted, the cur­sor pattern will be located nearest the one that will be activated when the cursor keys are pressed.
CAL/ZERO:
This key is for zeroing and calibration of a power sensor. All active sensors should be zeroed when­ever any sensor is added or removed, whether it is calibrated or not.
FREQ:
This key specifies the frequency of an input signal so that the power meter can apply the frequency­specific cal factor from the sensor EEPROM to the measurement.
[CAL/ZERO]
to calibrate a power sensor’.
If the frequency of the input signal changes so often that it is impractical to keep entering the fre­quency with the FREQ key, the frequency information can be conveyed to the 8650A by the use of a voltage input that is proportional to frequency (see the V
F connector on the rear panel).
PROP
When the 8650A is controlled remotely over the GPIB, the frequency information can be sent over the bus.
HELP:
Help screens describe the current display and present your options. From any menu, press
[HELP]
to display the associated Help screen. Press the up or down cursor key, if shown, to scroll the help screen up or down one line at a time. Press [Exit Help] to leave the Help screen and return to the current menu.
Power:
This push-push switch turns main power on and off. The LED above the power switch will be green when the unit is ON, and will turn yellow indicating the standby mode when the power switch is OFF with the main ac power still connected. The LED will be OFF when the ac power is discon­nected. All user setting and sensor calibrations will be automatically stored in non-volatile memory for recall when ac power is turned off with the front panel on/off switch.
CAUTION
To avoid losing some current settings and ensuring that the meter powers up properly, observe the following procedures:
1.) Remove AC Power to the rear panel only after at least 1 second has lapsed from the time the front panel ON/OFF switch has been pressed to turn off the meter.
2-2 Manual 31470, Rev. E, April 2001
Page 35
Front Panel Operation
2.) Cycling the ON/OFF switch should be no faster than once a second. Pressing the ON/OFF switch in rapid succession is not recommended.
3.) If the current setup is critical, it is recommended to save it as a stored setup to prevent accidental loss.
Display Screen
The liquid crystal screen displays measurements and configuration data. It also displays the operating and control functions that can be selected with the data interaction softkeys. Figure 2-1 illustrates a typical display screen with the menu control functions aligned with the data interaction softkeys.
The menus to calibrate and perform measurements with the 8650A are many and varied. A complete menu map is provided in Appendix D of this manual. Until you are thoroughly familiar with the operation of the 8650A, referring to the appropriate menu map to reach the function you intend to perform should speed up your operation.
When you turn on the power the model number and software version will display for five seconds and then be replaced with the Main display (see Figure 2-2).
Data Interaction Softkeys
The data interaction softkeys select the options for configuration and measurement control. In this manual, instructions to press a softkey are identified with the selected option in italics enclosed in brackets, such as ‘press [Sensor Setup] to configure a new power sensor’. Softkey statements will be in the same case as in the actual displays.
Sensor Inputs
and B connectors interface Channel A and Channel B sensors to the power meter. Dual-
AAAA
The channel Model 8652A has A and B sensor inputs; single-channel Model 8651A has only sensor A input. Refer to the Sensor Configuration details in Section 2.2.4.
CAUTION
When connecting sensor cables to these inputs, the cable pins must be aligned properly. Orient the cable so that the guide on the end of it aligns with the notch on the sensor input. If the connec­tor does not seem to fit, forcing it will only damage the connector pins.
Calibrator Output
The Calibrator connector is a reference power output for calibrating the amplitude response of a power sensor (see Section 2.3.3 for instructions on how to calibrate and zero sensors). The frequency of the output is fixed at 50 MHz or 1 GHz, depending on the sensor in use. The Calibrator output can also be used as a programmable low-level power output (see Section 2.2.1). During a calibration run, the output level automatically sweeps from -30 dBm to +20 dBm in 1-dB steps.
Manual 31470, Rev. E, April 2001 2-3
Page 36
Series 8650A Universal Power Meters

2.2 8650A Configuration

The 8650A screen normally displays measurement data, but it also displays the setup and configuration menus for the meter, display, and sensor. The setup menus are dynamic; the display adapts to the current operating mode and the type of sensors and other peripheral connections. For example, the sensor identification in Figure 2-2 will display only when a sensor is connected. If a sensor is not connected, the menu will indicate ‘No Sensor A’. The menu displays these other peripheral data only when applicable.
Figure 2-2 illustrates the Main Menu; other screens will display when sensors are connected or removed. Note that this is the only display that does not contain a mode or identification title in reverse video across the top of the screen.
The options available from the Main menu are:
Meter Setup for configuring the meter (see Section 2.2.1)
Display Setup for configuring the display (see Section 2.2.2)
Sensor Setup for configuring installed sensors (see Section 2.2.4)
dB/mW for selecting dB or power as the measurement unit (see below)
Reset Menu to refresh a min/max value (see below)
Rel for recording Relative Measurements (see Section 2.2.3)
CAL ON
A - 46.01
B - 50.52
A/B A
From the Main Menu, press [dB/mW] at any time to toggle between dB and mW as the measurement units. The values on the displayed lines will also change to match the units.
The Main Menu will display a Reset Menu selection only after you have configured one or more lines to display the data gathering functions, such as Min/Max (illustrated in line 4 in Figure 2-2), Bar Graph/ Peaking Meter, Peak Hold, Crest Factor, and Mean & Standard Deviation. When the selection is displayed, press [Reset Menu] to reset the accumulated data to it’s current value. Refer to Section 2.2.2 for detailed instructions on how to configure the display lines.
Most screens contain ‘OK’ and ‘Cancel’ options. After you have entered changes into the display, press [OK] to store the changes (this equates to Enter on a keyboard), or [Cancel] to cancel and return to the previous display without storing any changes (this equates to Escape on a keyboard).
- 4.49
-Min -45.95 dBm
-Max -46.04 dBm

Figure 2-2: The Main Menu

CW dB
CW dBm
dBm
MAP
Rel
dB
mW
Reset
Menu
Sensor
Setup
Display
Setup
Meter
Setup
2-4 Manual 31470, Rev. E, April 2001
Page 37

2.2.1 Meter Setup

The Meter Setup menu provides the means to configure the meter operating mode, and to store and recall setups. From the Main menu, press [Meter Setup] to display the Setup Menu. Press the softkey for the function you want to perform. The options are:
Calibrator to Configure and to turn the Calibrator output ON or OFF
Config to configure the operating mode and parameters
Sto/Rcl to store and recall meter setups
Service to clear all memory (additional service routines may be added to future revisions of this product)
Display Return to return to the Main menu without selecting a configuration option
Calibrator
The procedures in this section describe only how to use the Calibrator output as an RF output; see Section 2.3.3 for procedures to zero and calibrate power sensors. The Calibrator function must be turned ON to use the output as an RF source, indicated by the words Main Menu (see Figure 2-2).
Turn on the 8650A for 15 minutes then cycle the power OFF and ON, this establishes the 1 mW reference for the instrument firmware.
Front Panel Operation
CAL ON
in the upper left corner of the
To turn the Calibrator ON or OFF, press [Meter Setup] from the Main Menu. This will display the Setup Menu. From the Setup Menu, press [Calibrator] to display the Calibrator menu. Toggle the Calibrator On or OFF by pressing [On/Off].
☛☛☛☛
1. Press [Level dBm] to set the RF output level in dBm. Enter the level with the cursor keys.
2. Press [Level mW] to set the RF output level in power. Enter the level with the cursor keys.
3. Press [Frequency] to set the output reference frequency. Enter the frequency with the cursor keys.
4. Press [OK] to accept the changes or [Cancel] to cancel the changes, and return to the Power Meter
NOTE:
displayed, but it will be changed to reflect your selection when you return to the Main Menu.
Frequencies other than 50 MHz are available only with Option 12 (see Appendix C).
Configuration menu.
The wording at the top of the screen will not change while this menu is
Manual 31470, Rev. E, April 2001 2-5
Page 38
Series 8650A Universal Power Meters
Config
Press [Config] from the Setup menu to display the Power Meter Configuration menu (see Figure 2-3). From this menu you can configure the following:
GPIB mode and address
•V/F (V
Analog A and B outputs
RS-232 interface
Time Gate trigger mode and time parameters
Strip Chart graphic display
Histogram graphic display
Sound
F) input frequency and scale factor
PROP
Power Meter Configuration
GPIB
V/F In
Analog Out A
Strip Chart
Sound
Mode: 8600 Addr: 13
Figure 2-3: Power Meter Configuration Display
RS-232
T-Gate
Analog Out B
Histogram
OK
Config
Cancel
Select the item to be configured with the cursor keys, then press [Config]. The following options will be available for configuring the power meter:
GPIB:
This option assigns the GPIB protocols and address for control of the 8650A over the GPIB interface.
1. From the Power Meter Configuration menu, move the cursor to the GPIB field and press [Config]. The menu to set the GPIB mode and address will display.
2. Press [Mode] to highlight the operating mode and use the cursor keys to select the desired mode of operation. The options are:
SCPI to use the SCPI IEEE 488 command set
8600 to use the IEEE 488.2 8650 native mode
8542 to use the dual-channel 8542 emulation mode (Model 8652A only)
8541 to use the single-channel 8541 emulation mode
438A to use the HP-438A emulation mode
437A to use the HP-437A emulation mode
436B to use the HP-436B emulation mode
3. Press [Address] to move the cursor to the Address field. Step the address value up and down with the cursor keys. Valid addresses are:
00 through 30 in the Listen & Talk mode
40 in the Listen-Only mode
50 in the Talk-Only mode.
2-6 Manual 31470, Rev. E, April 2001
Page 39
Front Panel Operation
4. Press [OK] to accept the changes or [Cancel] to cancel the changes and return to the Power Meter Configuration menu.
V/F In:
The V/F (V
F) input accepts a frequency referenced to 0 Vdc, which the power meter uses to
PROP
determine and apply the appropriate correction factors (stored in the sensor EEPROM). The voltage input is supplied by a V/GHz output from the signal source. Two values must be defined for V
PROP
F: the frequency at 0 Volts, specified in Hz, MHz or GHz, and the scale factor specified in V/GHz. The V/GHz output connector on the frequency source is usually labeled with the scale factor.
From the Power Meter Configuration menu, move the cursor to the V/F In field and press [Config]. The V/F Config menu will display.
1. Frequency:
a. Press [Zero] to highlight the first digit of the frequency.
b. Press the up/down cursor keys to select the value of the first digit of the frequency.
c. Continue in the same manner to set all of the digits for the desired frequency. The units adjust
automatically as you enter the frequency.
2. Scale:
a. Set the scale factor by pressing [Scale] and setting the value in the same manner described for
the frequency.
b. Press [OK] to accept the changes or [Cancel] to cancel the changes, and return to the Power
Meter Configuration menu.
Analog Out A and Analog Out B:
The Analog A and B outputs are voltage proportional to measured power that can be applied to auxiliary test equipment such as a data recorder.
1. Select Source:
The choices of output source are Line 1 through Line 4. The configuration of the Analog outputs is interactive with the Display Configuration. If any line is set for accumulated data, such as Min/ Max, PKHold, CrFact, or Mean & Std Dev, it will not be displayed here for configuration.
2. Log/Lin:
The mode choices are Log and Linear. Press [Log/Lin] to toggle between the Log and Linear modes.
3. Set Scale:
The Set Scale menu adds the options to set the High and Low range in values of volts and power. Press [Set Scale] to enter this menu.
a. Press [Set High] to select the High range or [Set Low] to select the low range.
b. Press [Power] or [Volts] to toggle between the volts and power values.
Manual 31470, Rev. E, April 2001 2-7
Page 40
Series 8650A Universal Power Meters
c. Use the cursor keys to set the desired value in either the High or Low range.
d. Press [Set Volts/Set Power] to toggle between each value. Set the digit value of each with the
cursor keys.
e. Press [OK] to accept the changes or [Cancel] to cancel the changes and return to the Power
Meter Configuration menu.
Strip Chart
The Strip Chart function plots measurements on the screen over a fixed period or continuously. The X-axis displays time from the start of a measurement to a selective period of 1 to 200 minutes. While running, the adaptive autoscaled high and low plotting limits are displayed. While paused, the cursor can be used to review plotted values.
1. From the Power Meter Configuration menu, move the cursor to the
Strip Chart
field and press
[Config] (see Figure 2-4).
2. Press [Select Source]. The choices for measuring sources are Lines 1 through 4. If any line is set for data gathering, it will not be displayed here.
3. Press [Sample Rate]. The choices are from 1 sample per minute to 5 samples per second.
4. Toggle between [Hold Bfr] to stop sampling when the buffer is full, or [Rot Bfr] to set the Strip Chart display to update continuously with new samples after the buffer is initially full while discarding the same number of beginning samples.
5. Press [OK] to accept the changes or [Cancel] to cancel the changes, and return to the Power Meter Configuration menu.
CAL ON
-7
-27
STRIP CHART
Reset
Run
Pause
Cur on
Cur off
-15.75 dBm 33.5 sec (crsr)
A
Figure 2-4: Illustration of the Strip Chart
Exit
Sound:
A speaker within the chassis produces audible clicks and tones to register keystrokes and to draw attention to certain conditions. For example, if a limit has been exceeded or a calibration process has been completed. Press [On] to enable the speaker, or [Off] to disable it. There is no sub menu associated with this feature.
2-8 Manual 31470, Rev. E, April 2001
Page 41
Front Panel Operation
RS-232:
Highlight the RS-232 option in the Power Meter Configuration menu and press [Config]. The RS-232 Config menu will display for configuration of the RS-232 interface.
1. Press [Baud] to set the Baud Rate. Select the desired rate with the up/down cursor keys. The default value is 9600.
2. Press [Length] to set the Data Bits. Select the desired data bit length with the up/down cursor keys. The default value is Press [Parity] and turn the parity on or off with the up/down cursor keys. Default parity is off.
3. Press [Stop bits] to set the stop bit value. Select the desired stop bit value with the up/down cursor keys. The default value is 1.
4. Press [OK] to accept the changes or [Cancel] to cancel the changes and return to the Power Meter Configuration menu.
T Gate:
The Time Gate mode supplies a digital control voltage to enable or disable readings from any one or all sensors. Thus, the mode limits a power measurement to a defined interval that is controlled by a start time and a duration. The start time begins after a programmable delay following an external or internal trigger.
☛☛☛☛
NOTE:
The Time Gate feature operates only with modulation sensors.
Figure 2-5 illustrates the time gated measurement with an external time gated pulse applied to the trigger input. In this mode, the time gate starts and ends with the input of a high or low TTL level input. The duration of the measurement corresponds to the duration of the gated pulse.
RF RF
Ext Gating (TTL = High)
Gate Time
Ext Gating (TTL = Low)
Gate Time
Figure 2-5: Illustration of External Gating
Manual 31470, Rev. E, April 2001 2-9
Page 42
Series 8650A Universal Power Meters
Proceed as follows to configure the power meter for gated measurements by the sensors:
1. From the Power Meter Configuration menu, select [T Gate] then [Config] to display the Gate/ Trigger menu.
2. From the Gate/Trigger menu, select the sensor with the cursor keys, then select the mode of gated operation for that sensor. The options are External Gate, Ext. Trigger Rising, External Trigger Falling, and Burst Edge.
a. Gate:
This configures the gate to measure only during a high or low input. Gate Low enables power meter readings on a low logic input, and Gate High enables readings on a high (+5 V) logic voltage input (see Figure 2-5).
1.) Select the sensor to be configured with the cursor keys. The current configuration of the selected sensor will be indicated in the window at the bottom of the screen.
2.) Press [External Gate] to set the sensor gate measurement to high or low.
3.) Press [Gate High] or [Gate Low] to set the external gating to a high or low logic state.
4.) Press [OK] to return to the Gate/Trigger menu, or <Cancel> to abort the configuration.
b. External Trigger:
Figure 2-6 illustrates the Time Gated measurement parameters with an external trigger. When an external trigger is input (point A in Figure 2-6), it starts the Trigger Delay. At the end of the Trigger Delay, the Gated Time measurement starts and lasts until its preselected time expires. The Holdoff Time then prevents any further trigger inputs (such as point B below): from start­ing a new gated measurement until it has timed out.:
Holdout
Time
B
RF
EXT TRIG
A
Trig
Delay
Gate Time
Figure 2-6: Illustration of External Triggering
1.) Press [Ext Trigger Rising] or [Ext Trigger Falling] to set the parameters of the external trigger. The settings are:
Delay
Gate Time
Holdoff
2-10 Manual 31470, Rev. E, April 2001
Page 43
Front Panel Operation
Upon entering the External Trigger setup, a burst profile will display at the bottom of the screen. If the following specifications are not met, the profile may be inaccu­rate or not displayed at all:
a.) The time from between trigger edges is the period of the trigger. The period
must be equal to or less than 50 ms.
b.) The burst must be at least 0.1 ms wide and have at least 0.1ms off time.
c.) The trigger edge must occur before the leading edge of the burst.
d.) The points on the profile represent a resolution in time, which will vary
according to the requirement that the burst fit on the screen, to a minimum of approximately 10 µs per displayed point.
µ
2.) Set the value of the delay with the cursor keys. The range is 0
µ
3.) Select [Gate Time] and set its value. The range is 5
4.) Select [Holdoff] and set its value. The range is 0
5.) Press [OK] to accept the changes or [Cancel] to cancel the changes and return to the Power Meter Configuration menu.
s to 327.625ms.
µ
s to 327.625ms.
s to 327.625ms.
c. Burst Edge:
The Burst Edge mode uses the burst edges detected by the sensor to trigger power measure­ments, and is configured essentially the same as External Triggering described above. The setup screen for this mode also displays a burst graphic profile with the same conditions as described for External Triggering.
1.) Set the Delay and Gate time as described for External Trigger in steps 2) and 3) above.
2.) Press [OK] to accept the changes or [Cancel] to cancel the changes, and return to the Power Meter Configuration menu.
OK
Delay
Gate
Time
Numeric
Graphic
Cancel
Delay
Gate Time:
Burst Edge
0.070 ms
0.400 ms
Figure 2-7: Illustration of Burst Edge Configuration
Manual 31470, Rev. E, April 2001 2-11
Page 44
Series 8650A Universal Power Meters
Histogram
The Histogram feature displays graphical views of accumulated data distribution over a selective period of time (see Figure 2-8). Highlight the Histogram option in the Power Meter Configuration menu and press [Config].
1. Press [Select Source] and select the input channel (8652A only) or display the line number.
2. Toggle Log or Lin with the [Log/Lin] softkey.
3. Press [Interval] and select the duration the data is to be acquired. The range is from 1 second to 96 hours, or continuous.
4. Press [OK] to accept these settings or press [Cancel] to cancel the settings, and return to the Power Meter Configuration menu.
5. Press [OK] from the Power Meter Configuration menu. The Histogram will display full screen as shown in Figure 2-8.
CAL ON
-48.62 dBm -43.62 dBm Cursor: - 46.89 dBm 1.45%
Interval: 1 min (sampling complete)
Sensor A Histogram
2%
0%
Figure 2-8: Illustration of the Histogram
Reset
Select
Mode
Vert
Horiz
Zoom
Zoom
Out
Exit
In
There are additional functions available from the Histogram screen:
1. Press [Reset] to restart sampling.
2. Press [Select Mode] to toggle between distribution display functions.
a. The Histogram (see Figure 2-8) shows the percentage of time each power level is measured.
b. The CDF (Cumulative Distribution Function) shows the percentage of time a signal is below a
selected power level (see Figure 2-9). The x-axis displays the power level and the y-axis displays the percentage of time the power is at or below the power level specified by the x-axis.
c. The CCDF (Complementary CDF) (see Figure 2-10) shows the percentage of time a signal is
above a selected power level for more accustomed viewing of a descending slope.
3. Press the [Vert/Horiz] toggle key to select the zoom axis scale, then press [Zoom In] or [Zoom Out] for the selected axis.
2-12 Manual 31470, Rev. E, April 2001
Page 45
Front Panel Operation
4. Move the cursor along the slope of any curve to display, below the graph, the cursor position, the power level and the corresponding percentage of time the signal is above or below that level.
CAL ON
-48.62 dBm -43.62 dBm Cursor: - 46.32 dBm 44.05%
Interval: 1 min (sampling complete)
CAL ON
-48.62 dBm -42.62 dBm Cursor: - 64.90 dBm 46.32%
Interval: 1 min (sampling complete)
Line 1 (A) CDF
Figure 2-9: Illustration of the CDF Curve
Line 1 (A) CCDF
100%
0%
100%
0%
Reset
Select
Mode
Vert
Horiz
Zoom
Zoom
Out
Exit
Reset
Select Mode
Vert
Horiz
Zoom
Zoom
Out
Exit
In
In
Figure 2-10: Illustration of the CCDF Curve
Sto/Rcl
Up to 20 setup configurations can be stored in and recalled from non-volatile memory.
Recall:
1. From the Setup Menu press [Sto/Rcl].
2. Select [Recall] to recall a previously stored setup.
3. Step through the stored setup addresses with the cursor keys until the setup to recall is displayed.
4. Press [OK] to recall the selected setup.
5. Press [Recall Preset] to recall the default setup for the connected sensors.
Store:
1. To store the current configuration, press [Sto/Rcl] from the Setup Menu.
2. Select [Store] to store the current setup.
3. Select the setup ID (1 through 20) that you want to assign to this setup.
Manual 31470, Rev. E, April 2001 2-13
Page 46
Series 8650A Universal Power Meters
4. A name of up to 16 characters can be used as an additional means of identifying the setup. Press [Name} to open the display containing this option. Set the name by moving the cursor to the first position desired with the < and > softkeys, then select the character for that position with the cursor keys. Repeat this process for each character in the name.
5. Press [OK] to accept the stored data or press [Cancel] TWICE to cancel the changes, and return to the Power Meter Configuration menu.
Service
This is presently used only to clear all memory in the 8650A. When you press [Service], you will be prompted to press [Clear Mem] to clear all calibration, configuration, and measurement data currently stored in the 8650A RAM. Data stored in sensor EEPROMs are not affected.

2.2.2 Display Setup

Use this option to configure the data and format that will display on each line of the screen.
From the Main Menu, press [Display Setup] to select the Configure Display menu (see Figure 2-11). This screen is the same as the Main Menu except the control items are changed to display line configuration options. The configuration of each line includes the source and data format, data resolution, and peaking meter (bar graph) when configuration for use. In addition, the control to set the screen brightness is included in this display (Other Options).
One to four lines of data can be configured, and any attached sensor can be mapped to any of the four lines. Other information items are displayed at the top of the screen. These include
,
displays whenever the calibration RF source is on; meter is being operated by remote control (GPIB or RS-232); and configured inputs
B/A
☛☛☛☛
and
.
B-A
NOTE:
channel Model 8651A will have input A only.
The above input configurations are available only in the Model 8652A. The single-
CAL ON
Configure Display
A - 70.999
B - 54.245
A/B
A
- 16.754
-B
-24.246
Figure 2-11: The Configuration Display Menu
REMOTE
CW dB
CW dBm
dB
CW dBm
which displays whenever the power
Line 1
Line 2
Line 3
Line 4
More
Options
Cancel
CAL ON
A, B, A/B, A-B,
, which
2-14 Manual 31470, Rev. E, April 2001
Page 47
Front Panel Operation
Line Configuration
Select the line to be configured by pressing the corresponding [Line n] softkey. The menu in Figure 2-12 will display. The current configuration of the selected source displays in the large window at the bottom of the screen.
Data Line 1 Configuration
Source
A B
Line 2 3 4 Other
Off
A 11.11
Figure 2-12: The Data Line Configuration Menu
Data
A A/B
A-B
Statistics
CW dBm
OK
Select
Source
Select
Data
dB
mW
Config
Cancel
Source:
1. Press [Select Source] to enable the cursor in the Source field.
2. Move the highlight with the cursor keys to select the desired source.
a. A B:
This corresponds to sensor A and B inputs. The Data field will show the line configuration possibilities when you select the corresponding source input. When you select source Data field will read
B, B/A
and
B-A
A, A/B
.
and
. When you select source B, the Data field will read
A-B
A-B
A-BA-B
A
, the
b. Line n
:
When you select Line n in the Source field, the Data field will change to
Min/Max
. You will be configuring the selected line number to be the source. For example, you can use the selected line to display the minimum/maximum values of another line (Line n). The only line numbers that will be available for selection are those which have already been config­ured and are not configured to show min/max values of other lines. If you want the selected line number to show the min/max values of another line but the line is not avail­able in the source list, you will have to go back and reconfigure the missing line. It may not be configured at all, or it may be configured to display the min/max of another line.
c. Other:
This selection changes the Data field to copy another line into the line number you are configuring. The Data field will change to indicate the lines, such as n1—>n2, meaning that line n1 will be copied into line n2 (the line you are configuring). You will find this option useful, for example, if you want to change a line but do not want to lose its configu­ration; you can select Other and copy the configuration to a different line and retain the configuration on the new line.
Manual 31470, Rev. E, April 2001 2-15
Page 48
Series 8650A Universal Power Meters
d. Off:
Select this option to turn the line off. When a line is turned off, it will not be available as a source Line n. For example, if Line 4 is turned off, you cannot set Line 3 to read the min­imum/maximum values of line 4.
Data:
1. After you have selected the source for the line being configured, press [Select Data] to enable the cursor in the Data field.
2. Move the highlight with the cursor keys to the configuration of your choice.
a. When you select source
, the Data field will read B,
B
source
, the Data field will read A,
A
and
B/A
B-A
.
A/B
and
. When you select
A-B
b. Other measurements that can be selected for Modulation Sensors are Peak Hold, Crest
Factor, and Mean & Standard Deviation.
3. DB/mW:
This selection toggles the line units between dB and mW. It can be changed at any time to dis­play the opposite units.
Config:
Use the Config selection to set the measurement resolution, upper and lower limits, and utilization of the bar graph. Lines configured with Statistics can be configured only for resolution; the Set Limits and Bar Graph options will not display for these lines.
1. Press [Config] and use the cursor keys to set the resolution of the input. The resolution will be displayed in the large window at the bottom of the screen. Adjust the resolution with the cursor keys to the desired level. The range is 0 to 3 decimal places.
2. Press [Set Limits] to set the upper and lower measurement limits, or to turn the limits off.
a. Press [High] and set each digit of the high limit with the cursor keys.
b. Press [Low] and set each digit of the low limit with the cursor keys.
c. Press [Limits Off] to disable use of the limits. Press [High] or [Low] again to enable and set
the limits.
3. Press [Bar Graph] to set the utilization of the bar graph scale.
a. The menu will display as described above for Set Limits.
b. Follow the Set Limits procedure above to set the bar graph scale, or to disable the bar graph.
c. The upper and lower limits of the bar graph will be displayed in the Main menu as brackets
with the current measurement shown as a horizontal bar between the brackets (e.g.,[ ]).
4. Press [OK] to save all changes and return to the Configure Display screen, or [Cancel] to leave the configuration without saving any changes.
2-16 Manual 31470, Rev. E, April 2001
Page 49
More Options
1. Select [More Options] from the Configure Display menu (see Figure 2-11) to set the brightness of the display screen. The adjustment, made with the cursor keys, is interactive so you can observe the brightness as you adjust it. The range is 0% (off) to 100%. The default is 70%.
2. Press [OK] to save all changes and return to the Configure Display screen, or [Cancel] to leave the configuration without saving any changes.

2.2.3 Rel

Normally, each line of the display shows a sensor measurement value. When you press [Rel] from the Main menu, the present measured value of each line is recorded and all subsequent measurements are expressed in dB or percentage relative to that recorded value. The Relative Measurement menu provides a means of selectively enabling or disabling the relative measurement mode for channels A or B.
Press [Select Line] repeatedly to select the line to be used for relative measurements. Use the cursor keys in this menu only to change the value of the selected line.
The selected line can be switched in and out of the Relative Mode by pressing the [Rel On/Off] key.

2.2.4 Sensor Setup

Front Panel Operation
Sensor configuration enables you to select the operating mode of sensors, or to select the parameters for averaging time-based measurements. From the Main Menu, press [Sensor Setup] to display the installed sensor data and configuration options (see Figure 2-13). The power meter reads the contents of the EEPROMs for installed sensors and displays each type of sensor and its connecting input.
The [Sensor Setup] softkey will not function if there are no sensors attached, or if attached sensors have not been calibrated. The Sensor Setup options are dependent on the type of sensor. All sensors options are
Avg
and
Offset
. Peak and Modulation sensors add a
Sensor Setup
A: Peak Sensor
B: Modulation Sensor
First select sensor above, Then configuration mode at right
Figure 2-13: The Sensor Setup Menu
Config
Offset
Config
Cancel
option (see Figure 2-13).
OK
Avg
Manual 31470, Rev. E, April 2001 2-17
Page 50
Series 8650A Universal Power Meters
Avg
Use this option to set the averaging period of power measurements, including modulated, CW and peak sensors. Press [Avg] from the Sensor Setup menu to display the
menu
. This menu offers three options: Auto, Averages, and Time.
:
Auto
1. Select this setup option if you want the time to be ‘adaptive’ averaged. That is, the time window changes dynamically to reduce the effects of transient changes in power levels.
2. Press [OK] to store the changes, or [Cancel] to exit without saving changes.
Averages:
1. Select this setup option to set the number of periods for averaging measurement signals over a fixed period of time. Each unit is approximately 20 ms in duration. You have a choice of ‘powers of two’ multiples of units from 1 to 512. Set the number of periods with the up/down cursor keys.
2. Press [OK] to store the changes, or [Cancel] to exit without saving changes.
Sensor Setup - Averaging
Time:
1. Select this option to average measurement signals over a fixed period. A menu will display for you to set the time in seconds between 20 ms and 10 seconds. The meter will automatically change the selection to the nearest of 20 ms times powers of 2 multiples of 1 to 512 (e.g., entering 1.000 seconds will be changed to 1.280 seconds). Set the time digits with the cursor keys.
2. Press [OK] to store the changes, or [Cancel] to exit without saving changes.
Offset
Select this option to enter an offset value in dB when an attenuator or amplifier is installed before the sensor input. From the Sensor Setup menu.
1. Press [Offset] and enter the value of the offset with the cursor keys.
2. Press [OK] to store the changes, or [Cancel] to exit without saving changes.
Config
This option is available with Peak and Modulation sensors only.
Peak Sensors:
1. From the Sensor Setup menu select the Peak sensor with the cursor keys.
2. Press [Config] to display the Setup Mode menu. This menu offers two options: CW and Peak.
3. Press [CW] to select the CW mode for the peak sensor. There is no configuration for peak sensors operating in the CW mode and the screen will return to the Setup Menu to configure other sensors.
2-18 Manual 31470, Rev. E, April 2001
Page 51
Front Panel Operation
4. Press [Tr i g L e v e l] to configure the trigger level in either the Internal or External Trigger mode.
a. Internal Trigger:
Peak power will be sampled at a point defined by a Trigger Level, a Delay, and a Delay Off­set. The delay-offset feature is a convenience in some applications (for example, when measuring pulse width from a point other than the trigger level, or when comparing the levels of various pulses within a pulse train).
1.) Press [Int/Ext Trig] to select Internal Triggering.
2.) Set the Trigger Level value with the cursor keys.
3.) Press [Delay] and use the cursor keys to adjust the Trigger Delay (0 to 100 ms). Units will change automatically.
4.) Press [Delay Offset] and use the cursor keys to adjust the Trigger Delay Offset (0 to 100 ms). Units will change automatically.
5.) Press [OK] to store the changes, or [Cancel] to exit without saving changes.
b. External Trigger:
The External Trigger mode is configured the same as the Internal Trigger mode described above, except that the Trigger Level is specified in volts rather than dBm.
1.) Press [Int/Ext] Trig to toggle to the External Trigger mode and set the Trigger Level value in volts with the cursor keys.
2.) Press [OK] to store the changes, or [Cancel] to exit without saving changes.
Modulation Sensors:
Modulation sensors can be used in CW, Modulated Average Power (MAP), Peak Average Power (PAP) and Burst Average Power (BAP) modes.
1. From the Sensor Setup menu select the Modulation sensor with the cursor keys.
2. Press [Config] to select and configure a modulation sensor. The options for this configuration a r e C W, M A P, PA P, a n d B A P.
a. Press [CW] to select the CW mode. There is no configuration for modulation sensors
operating in the CW mode and the screen will return to the Setup Menu to configure other sensors.
b. Press [MAP] to select the MAP mode. There is no configuration required and the screen
will return to the Setup Menu to configure other sensors.
c. Press PAP to select the PAP mode. You will be prompted to enter a duty cycle for pulse
average inputs. Use the cursor keys to adjust the Duty Cycle. The range is 0.001% to
99.999%.
d. Press [BAP] to select the BAP mode. The options are Auto Config and User Config.
1.) Press [Auto Config] for the meter to automatically configure itself and return to the Setup Menu.
Manual 31470, Rev. E, April 2001 2-19
Page 52
Series 8650A Universal Power Meters
2.) Press [User Config] to configure the meter manually. You will be prompted to enter the values for the Burst Start Exclude, Burst End Exclude, and Dropout. Enter these values by pressing the corresponding softkey and using the cursor keys to change the values.
3.) Press [OK] to store the changes, or [Cancel] to exit without saving changes, and return to the Sensor Setup menu to configure the next sensor, or return to the Main menu.
2-20 Manual 31470, Rev. E, April 2001
Page 53

2.3 Measurement Guide

This section presents guidelines for practical applications of the 8650A. See Section 2.3.9 for mode restrictions.

2.3.1 Using the Power Sweep Calibrator

The Power Sweep Calibrator automatically calibrates the power sensor to the power meter. The power sweep operates from -30 to +20 dBm (the complete, non-square-law operating region) and transfers the inherent linearity of an internal, thermal-based detector to the balanced diode sensors. Output is NIST­traceable at 50 MHz, 0 dBm to an accuracy of ±0.7% (±1.2% over one year). (Note: NIST is the National Institute of Standards and Technology).

2.3.2 80701A Sensor Operation

The Series 80701A power sensors are designed for the precise measurement of signals with wide modulation bandwidths (up to 10 MHz). In terms of the various measurement modes (i.e., MAP, BAP, etc.), the 80701A sensors are operated exactly as the Series 804XX and 806XX sensors described in Appendix B.
Front Panel Operation
☛☛☛☛
There is one distinction regarding the operation of the 80701 sensors. Below 200 MHz, the modulation bandwidth of the sensor is limited by a filter which is electronically switched in the sensor. This is done to keep the RF signal out of the base band signal processing circuitry. When a 80701A sensor is calibrated on the meter for the first time (the meter reads UNCALIBRATED before calibration), the unit is set to the default setting of MAP mode with frequency correction set to 1 GHz. This allows the sensor to measure signals with wide-bandwidth modulation. For frequencies of 200 MHz or below, the frequency correction must be set to the measurement frequency to avoid measurement error.
The Series 80701A sensors are compatible with the Models 8651A and 8652A Power Meters.
NOTE:
Series 80701A sensors require the installation of Option 12 (See Appendix C).
CAUTION
Sensor diodes can be destroyed by momentary or continuous ex­posure to excess input power. The maximum power (peak or aver­age) that can be applied to the detector elements without damage is printed on the side of the sensor housing. For standard CW and peak power sensors, the maximum level is +23 dBm (200 mW). Standard sensors should not be used above +20 dBm (100 mW) because this may degrade the sensor’s performance even if it does not burn out the diodes.
When measuring pulsed signals, it is important to remember that the peak power may be much greater than the average power (it depends upon the duty cycle). It is possible to overload the sensor with a pulsed signal even though the average power of the signal is far below the maximum level.
To measure higher power levels, use a high power sensor; or reduce the signal amplitude using a directional coupler or a precision attenuator.
Manual 31470, Rev. E, April 2001 2-21
Page 54
Series 8650A Universal Power Meters

2.3.3 Sensor Zero and Calibration

All sensors must be calibrated and zeroed before making measurements.
Calibration and Zeroing
1. Connect the sensor to be calibrated from Channel A or B to the Calibrator output (see Figure 2-14).
☛☛☛☛
NOTE:
is turned on. Allow a 15-minute thermal equilibrium period and calibrator connector to minimize temperature differences.
Connector sensor to the Calibrator connector after the power meter
for the sensor
2. Calibrate the sensor(s) within the 5 °C to 35 °C (41 to 95 °F) operating temperature range of the Calibrator.
40%
Figure 2-14: Setup for Sensor Calibration
3. Press the
[CAL/ZERO]
hardkey. The meter will automatically zero and calibrate the sensor. A sensor requires several minutes to fully zero and calibrate. After the zero/calibration routine is complete, the display will return to the Main menu.
☛☛☛☛
NOTE:
immediately after calibration, it is possible to measure approximately ±0.02 dBm from the calibrator output instead of 0.00 dBm (0.4% difference). This is normal and within specification due to a combination of tolerances resulting from calibrator accuracy (±1.2% worst case for one year, over temperature range of 5 °C to 35 °C) and connection tolerance.
If the calibrator is set to 0.00 dBm and turned on before or
Zeroing Only
1. If you need to only zero one or both sensors, make sure both sensors are connected to the sensor inputs and that neither sensor is connected to the Calibrate output. When zeroing a sensor, it is best to connect the sensor to the device under test exactly as it will be used in measurement, and deactivate the RF output of that device. Zeroing the sensor in place is the best way to counteract system noise which could significantly effect low-level measurements.
2-22 Manual 31470, Rev. E, April 2001
Page 55
Front Panel Operation
2. Turn off the source output before you zero the sensor. The microwave source must output less than -74 dBm of total noise power during RF Blanking for proper zeroing. The source signal power should be less than -90 dBm.
☛☛☛☛
3. Press it will display a menu for you to select Sensor A, B or both (All) to be zeroed. Select the sensor(s) to start the zeroing process, or press [Cancel] to abort zeroing and return to the Main menu.
4. The sensor should be zeroed just before recording final readings in the lower 15 dB of the power sensor’s dynamic range.
5. The sensor should remain connected to the signal source during zeroing. By turning off the source instead of disconnecting the detector, the zeroing process automatically accounts for ground line voltages and connector interface EMF.
☛☛☛☛
2.3.3.1 Calibration & Zeroing for High Power Sensors with Removable
Attenuators
NOTE:
equilibrium with the source. This could be up to 15 minutes for moderate initial temperature differences.
[CAL/ZERO]
NOTE:
equilibrium with the source. This could be up to 15 minutes for moderate initial temperature differences.
Sufficient time must be allowed for the module to reach thermal
. The meter will check that the sensors are installed and calibrated, then
Sufficient time must be allowed for the module to reach thermal
When using a Giga-tronics high power sensor with external attenuator, Giga-tronics power meters automatically recognizes the sensor type and compensates for the attenuation factor. However, when performing the front panel calibration, be sure to remove the attenuator before connecting to the calibrator port. During measurement, you do not need to enter an offset factor to account for the attenuator loss. The sensor frequency calibration factors automatically correct for the combined frequency response of the sensor and attenuator. Because the sensor and attenuator are a matched set, the serial numbers of the sensor and attenuator are identical. Do not use attenuators from other high power sensors.
1. Remove the high power attenuator from the sensor.
2. Connect the sensor to be calibrated from Channel A or B to the Calibrator Output (see Figure 2-14).
3. Press the
4. Reconnect the high power attenuator to the sensor.
☛☛☛☛
[CAL/ZERO]
NOTE:
measurement uncertainty, align the arrows when reconnecting the attenuator to the sensor.
There are alignment marks (arrows) on the sensor and attenuator. To reduce
hardkey. The meter will automatically zero and calibrate the sensor.
Manual 31470, Rev. E, April 2001 2-23
Page 56
Series 8650A Universal Power Meters
2.3.3.2 Low Level Performance Check
This procedure provides a quick-check list for evaluating meter/sensor performance for low-level measurements. It is not intended to verify performance of specifications such as Noise, Temperature Coefficient and Zero Set. For complete verification, please refer to Chapter 4.
1. This test is meant to check the low level performance of the meter and sensor. In order to do so, the meter and sensor should first be separated from any external amplifiers, test systems, etc. Turn the meter on and allow stabilization at ambient for a minimum of 30 minutes. Connect the sensor cable to the meter and the sensor to the calibrator output port.
Calibration.
2.
☛☛☛☛
3.
NOTE:
This zero is not valid for actual measurements and can limit the measurement range as high as –50 dBm. For proper low-level measurements, the sensor must be zeroed at the test port of the system being tested. Zeroing at the test port provides corrections for ground line voltages and connector interface EMF.
Zeroing.
Calibrate the power meter by pressing the Zero/Cal button.
During calibration an approximate zero is established for calibration purposes only.
Validation of meter and sensor noise floor will be checked using an attenuator or termination. Connect the attenuator or termination to the sensor and allow the unit to stabilize for 3 minutes. The sensor must be thermally stabilized for proper zeroing. If the thermal conditions of the sensor varies during the zero procedure, the zero will not be valid.
4. Set averaging to 512 and configure for CW operation. After the unit has thermally stabilized, push the Cal/Zero button.
5. Immediately after zeroing, confirm that the meter reading is at least 3 dB below the minimum CW operating range of the sensor. This checks the noise floor and zero set capabilities of the meter and sensor.
Zero Drift.
6.
Zero Drift is a measure of the change in noise over time. Each family sensor will have a specified expectation of drift over a one-hour period. To confirm, set the meter to linear display (Watts) after verifying noise floor and check that the display does not drift beyond specification over a one-hour period.
Verification for specifications such as noise, zero drift and temperature coefficient of linearity are difficult, time consuming tests. This checklist is useful to quickly determine if there is a catastrophic system failure. Failure to meet the above guidelines is not necessarily an indication of specification failure. Final confirmation of system specification performance is achieved using the verification procedures found in Chapter 4.
2-24 Manual 31470, Rev. E, April 2001
Page 57

2.3.4 Measuring Source Output Power

To measure the source output power:
1. Connect the power sensor to the RF output of the microwave source.
2. Verify that the microwave source RF output is ON.
Front Panel Operation
3. Press
4. The 8650A will now display the microwave source output power. Adjust the source amplitude to
The 8650A responds rapidly to amplitude changes. Ranging is automatically performed in real time through a 90 dB dynamic range using CW or modulated sensors. The peak sensor dynamic range is 40 dB Peak and 50 dB CW. Entering the operating frequency enables the 8650A to automatically apply frequency calibration factors appropriate to the sensor being used. The operating frequency can be communicated to the 8650A using the front panel menus, the GPIB, or the V
[FREQ]
].
OK
OK
[
OKOK
the desired level.
; enter the operating frequency (use the cursor keys to adjust the value), and press

2.3.5 Using the Peaking Meter

The line bar graph provides a linear display of power level on a decade range basis. For example, a power level of 3 dBm produces an approximate 50% response on the peaking meter. The peaking meter displays on the configured line as a solid horizontal bar between two brackets. The bracket on the left of the screen represents the lower limit of the peaking meter, and the bracket on the right of the screen represents the upper limit (see Figure 2-15). The procedure to turn the peaking meter on and off, and to set it’s limits is detailed in Section 2.2.2.
Lower Limit
Bracket
Peaking Meter
(Bar Graph)
CAL ON
Configure Display
Upper Limit
Bracket
Line 1
F voltage input.
PROP
A - 70.999
[
B - 54.245
A/B
A
Manual 31470, Rev. E, April 2001 2-25
- 16.754
-B
-24.246
Figure 2-15: The Peaking Meter
CW dB
CW dBm
dB
CW dBm
]
Options
Line 2
Line 3
Line 4
More
Cancel
Page 58
Series 8650A Universal Power Meters

2.3.6 High Power Level Measurements

High power amplifiers and transmitters can damage standard sensors. Use only high power sensors to measure these devices without using attenuators and measurements.
For example, if the output of an RF source is amplified to +30 dBm (1 Watt), this signal cannot be measured directly using a standard sensor because the sensor’s maximum input level is +23 dBm (and any level above +20 dBm is potentially harmful to a standard sensor). The signal would have to be attenuated, and the attenuation would have to be corrected for by means of a measurement offset. However, if a 5 Watt high power sensor is used, any power level up to +37 dBm can be measured directly without the use of an attenuator.

2.3.7 Modulated Measurement Modes

The 8650A series of power meters expands upon the capabilities of the previous 8540 power meters in a number of ways. In the past, power measurements of modulated signals (pulse, multi-tone, AM, etc.) required that the signals be attenuated to levels less than -20 dBm to avoid errors due to sensor nonlinearity. The 8650A eliminates this restriction when used with a modulation sensor, and brings the speed and accuracy of diode sensors to the power measurement of modulated signals. Basic measurement procedures are presented below, along with some useful tips on how to get the most out of the modulated measurement modes.
The modulated measurement modes are available through the sensor setup menu when the active sensor a modulated series. The 8650A features three modulated measurement modes:
Modulated Average Power (MAP)
Pulse Average Power (PAP)
Burst Average Power (BAP)
MAP and PAP modes measure the true average power of modulated and pulsed signals. PAP mode differs from MAP mode only in that it allows you to specify a duty cycle figure, which is automatically factored into the measurement. In BAP mode, the true average power within the pulse is measured (the pulse pattern is detected automatically, so there is no need for you to specify the duty cycle).
MAP Mode
The Modulated Average Power (MAP) mode measures RF signals, which are amplitude modulated, pulse modulated, or both. In the MAP mode the 8650A calculates the average RF power received by the sensor over a period of time controlled by the time constant of the internal digital filter. The result is comparable to measurement by a thermal power sensor.
In this mode, the 8650A measures the average power of CW and modulated signals, such as:
•AM
Two-tone
Multi-carrier
Pulse modulation
Digital modulation (QPSK, QAM, etc...)
For example, if an RF signal pulse modulated at 50 Hz with a 10% duty cycle is measured with the averaging factor set to 128, the measured power reading will be 10% of the peak power during pulse ON periods. Because the signal is modulated at a low pulse rate (below about 1 kHz), the 8650A will synchronize the readings precisely with the start of a pulse so that each displayed reading is averaged over a whole number of pulses (that is, there are no fractional pulses included in the measurement). This eliminates a significant amount of noise from the readings. It is important to remember that even
2-26 Manual 31470, Rev. E, April 2001
Page 59
Front Panel Operation
Start of Burst
End of Burst
Noise Threshold
Powe r
Time
though the filter settling time has been set to a long time constant of 2.56 seconds, the update rate of the meter will be much faster, and even the first reading will be very close to the fully settled value.
PAP Mo de
The Pulse Average Power (PAP) mode is similar to the MAP mode, but it measures pulse-modulated signals having a known duty cycle. You can specify this duty cycle and the 8650A will automatically correct the measurements so that the displayed readings indicate the peak RF power during pulse ON periods.
For example, when measuring a pulse modulated signal with 50% duty cycle, MAP mode would give a reading 3 dB lower than the reading that would be given by PAP mode with the duty cycle factor set to 50%.
☛☛☛☛
NOTE:
pulse shape. Any abnormality such as overshoot, undershoot, slow rise time or fall time, inaccuracy of the duty cycle, or deviation from a flat pulse response will cause errors in the indicated reading.
The duty cycle correction presumes a perfectly rectangular profile for the RF
BAP Mode
The Burst Average Power (BAP) mode measures the average power during an RF burst (see Figure 2-16). This mode is very useful for measurement of pulse modulated signals which are not flat or have amplitude modulation during the pulse ON period, as in the case of TDMA (Time Division Multiple Access) communications signals. In this mode, the 8650A recognizes the beginning and end of a burst of RF power and takes an average of the power during that burst. The RF level can vary over a wide range during the burst as long as it remains above a noise threshold, which is automatically calculated by the 8650A. As soon as the RF power drops below the noise threshold, the RF burst is complete and all further readings are discarded until the next burst starts.
Figure 2-16: Burst Measurement
Manual 31470, Rev. E, April 2001 2-27
Page 60
Series 8650A Universal Power Meters
In BAP mode, the 8650A automatically determines which portions of the signal are in the pulse and which are not. In computing the average power, the 8650A uses only those portions that are within the pulse. The result is that, independent of the signal’s pulse duty cycle, the meter always reads the average power in the pulse or burst. As with the PAP mode, when measuring a pulse modulated signal with 50% duty cycle, the reading in the BAP mode would be 3 dB higher than in the MAP mode. However, in the BAP mode, the signal’s duty cycle can change dynamically in time without affecting the meter reading. In the PAP mode, the duty cycle factor must be entered to match the duty cycle of the pulsed signal.
☛☛☛☛
NOTE:
sensor pulse width specification.
BAP Mode requires a minimum pulse on or off time as determined by the power
2-28 Manual 31470, Rev. E, April 2001
Page 61

2.3.8 Measurement Collection Modes

Using a wide range of CW and Peak Power Sensors and the GPIB fast measurement collection modes, the Series 8650A meters provide typical reading speeds of >1750 readings per second in the free-run Swift mode, 800 readings per second in the Fast Modulated mode, and >26,000 readings per second in the Fast Buffered mode. Three Swift mode triggering controls are available: Fast free-run, bus triggered, and TTL triggered modes. Bus and TTL allow triggering control of individual measurement points. Data can be stored in an internal data buffer or read immediately.
Fast buffered power readings are internally buffered for readout at the completion of the fast buffered interval. Maximum measurement rate is about 26,000 readings per second. Data conversion and GPIB communication time are not included in this figure. The maximum buffer size is 5000 readings, or about
2.1 seconds at the maximum reading rate.
CW Mode
This mode is for measuring an unmodulated Continuous Wave (CW) signal. In this mode the RF signal level must be constant for accurate readings to be made. If the signal level changes, a settling time for the internal digital filter is required in order for measurements to be made to the specified accuracy.
The settling time (the time required for a measurement based on an averaging of samples to adapt to a changed condition and become accurate again) is affected by various factors. The maximum settling time is equal to 20 ms multiplied by the averaging factor (for example, if the averaging factor is 128, the maximum settling time is 2.56 seconds). In most situations the actual settling time is well below the maximum.
Front Panel Operation
PEAK Mode (80350A Peak Power Sensor)
The Peak mode is for instantaneous peak measurements of the RF power level of a pulse modulated signal during pulse ON periods. The measurement is based on an instantaneous sample taken at a particular point in time. Sampling is triggered by a pulse rising edge either in the modulated signal itself or in a supplied trigger input signal, followed by a programmable delay. The trigger/delay combination makes it possible for you to specify exactly what part of the pulse is sampled.
In the peak mode, each displayed reading can consist of a single sample or of an average of multiple samples, each taken at the exact same time relative to the pulse’s rising edge. If the averaging factor is set to 1, single samples are used. If it is other than 1, the averaging factor will determine the filter settling time over which the multiple samples will be taken and averaged.
Because the peak mode measures the RF power instantaneously (at the top of the pulse, provided that the delay has been set correctly), no assumptions are made about the pulse shape or duty cycle. In fact, it is possible to profile the pulse by sweeping the delay time over a range of values to reveal the pulse shape from start to finish.
☛☛☛☛
NOTE:
pulse where it is told to sample the pulse whether or not the point sampled is really the peak point. This mode is therefore less intelligent than the BAP mode and must be used carefully, but its flexibility makes it a powerful tool for studying modulated signals.
In the peak mode the 8650A does not know where the peak is. It samples the
Manual 31470, Rev. E, April 2001 2-29
Page 62
Series 8650A Universal Power Meters
Peak power measurements are made by sampling the RF input at a point which is defined by a trigger level, a delay, and a delay offset (see Figure 2-17). The initial triggering event occurs when the power input (or in the case of external triggering, a voltage input) reaches a threshold, which you have defined as the trigger level. The sample is then taken after a delay, which you have defined. To this delay can be added a positive or negative delay offset.
The delay offset is not necessary for peak measurement, but in some applications it is a convenience. For example, a small offset (even a negative offset) might compensate for the difference between the trigger point and some other point of interest (such as the half-power point) especially in applications where pulse width is being measured. Or if it is necessary to measure the levels of various pulses within a pulse train, the pulses can be sampled successively by changing the delay offset. A fixed delay insures that each pulse is sampled at the same point in its cycle.
Peak Power, Sampled After a 120 ns Delay
Sample
120 ns
Delay
Power
Trigger
Level
Trigger Point
Time
Peak Power, Sampled With a Fixed Delay
Sample
Trigger
Power
(No Offset)
2.8 µs
Offset
Delay Offset
Peak Power, Sampled After a 120 ns Delay
Power
Trigger
Level
But Various Delay Offsets
Sample
(11
µs Offset)
and a 10 ns Delay Offset
Sample
10 ns Delay
Offset
Delay Offset
120 ns
Delay
Half-Power Point
Trigger Point
Time
Sample
(22
µs Offset)
6
0
4
2
810
12
16
14
Time (Microseconds)
18
20
22
24
26 28
30
32
Figure 2-17: Delay and Delay Offsets
2-30 Manual 31470, Rev. E, April 2001
Page 63

2.3.9 Mode Restrictions

In certain modes the 8650A has highly specific restrictions on its operation:
In the fast measurement collection modes (swift and fast buffered), it is not possible to make measurements which compare the two channels. In other words, it is possible to make measurements using sensor A, or B, or both, but measurements such as A/B and A-B are not permitted.
In GPIB remote operation, only one reading can be sent over the bus (it can be A, or B, or a comparative measurement such as A/B, but it is not possible for separate measurements of A and B to be sent over the bus). The exception is that in the swift and fast buffered measurement collection modes, it is possible for both A and B to be sent over the bus.

2.3.10 When to use CW, MAP and BAP

For measuring signals with any kind of modulation, MAP mode should be used. In this mode, the 8650A makes use of its digital signal processing algorithms to ensure that the reading is the correct average power level regardless of modulation type (see Section B.2.2 for limits on modulation rate, etc.).
Of course, CW signals may also be accurately measured in MAP mode. This raises the question, why use CW mode? CW mode offers a few more dB of dynamic range at low power levels when using a CW power sensor, such as the 80301A. In addition, in CW mode the 8650A is form, fit, and function compatible with its predecessor, Model 8540.
Front Panel Operation
BAP mode should be used only for the measurement of signals which are pulse modulated. In this mode the meter will accurately measure the average power of the signal during the on-time of the pulse. This mode works equally well regardless of whether the signal is modulated during the pulse on time.

2.3.11 Multi-Tone Tests

Multi-tone testing refers to more than one RF carrier combined into one signal to be measured. Two­tone intermodulation testing, for example, is a common test performed on a wide variety of RF components and subsystems. MAP mode should be selected for these applications. The 8650A test procedure is as follows:
1. Calibrate the sensor according to the procedure outlined earlier in Section 2.3.3.
2. From the Main Menu press [Sensor Setup]. From the Sensor Setup menu, press [Modulated Sensor] and then select the MAP mode by pressing [MAP].
3. Press [FREQ] and enter the operating carrier frequency.
4. Connect the sensor to the multi-tone source and record the power level.
For two-tone testing, small errors in the measurement will result when the carriers are separated by more than the specified bandwidth of the sensor. The amount of error is also a function of average power level. For average power less than about -20 dBm, there is no modulation-induced measurement error at any tone separation. Consult the error charts found in Section B.2.2.
Multi-carrier testing usually refers to more than two carriers combined into one signal. Common multi­carrier tests combine multiple carriers. In determining expected measurement error for these types of signals, the maximum difference in frequency between any two carriers should be used as the tone separation when applying the error charts in the manual.
Another important feature of multi-carrier signals is that they can have a high peak-to-average power ratio. This ratio can be as high as 10 dB for multiple carriers. The significance of this in terms of making
Manual 31470, Rev. E, April 2001 2-31
Page 64
Series 8650A Universal Power Meters
power measurements is two-fold. First, care should be taken to keep the peak power level applied to the sensor below the maximum recommended level. Second, when trying to minimize modulation-induced measurement error for carriers separated by more the specified bandwidth of the sensor, it is the peak power level that should be kept below about -20 dBm.

2.3.12 Peak Hold

When the Peak Hold feature is selected, the 8650A displays the highest instantaneous power measured from the time the feature is enabled until it is reset by the user. In other words, the displayed value tracks the measured value only when the measured value is rising to a new maximum. When the measured value falls, the displayed value holds at the maximum. When the peak hold feature is reset, the displayed value falls to the current measured value and the process begins again.
When the Peak Hold feature is selected, the minimum average power level that can be accurately measured is -20 dBm. Beginning in firmware version 1.51, this has been implemented to increase accuracy and maintain wide bandwidth of Peak Hold measurement over the peak power range of -20 dBm to +20 dBm. The Peak Hold selection will also affect Histogram measurement if there is power below -20 dBm. To obtain accurate average power measurements below -20 dBm, remove the display line containing Peak Hold (or, in remote control, by sending the command which disables the Peak Hold feature).
The Peak Hold feature is available in the MAP, PAP, and BAP measurement modes; it may be enabled from the front panel under the Display Data Line Configuration setup menu, or over the GPIB. Peak Hold is reset by pressing [Reset Line n] (or, in remote control, by sending the command which activates the Peak Hold feature).
The reset function controls the time resolution of the reading (that is, for finer resolution, reset more frequently).
☛☛☛☛
Power
NOTE:
Hold
[Reset Line n] for Peak Hold also resets the Crest Factor.
Peak Hold
Hold
Hold
Hold
(Reset)
Time
Figure 2-18: Peak Hold
Hold
(Reset)
Peak Hold
Instantaneous
2-32 Manual 31470, Rev. E, April 2001
Page 65

2.3.13 Crest Factor

The Crest Factor feature is very similar to the peak hold feature, in that it holds on to the maximum level until a reset occurs, but in this case the displayed value is expressed (in dB) as a ratio of the held maximum power to the average power.
The Crest Factor feature is available in the CW, MAP, PAP, or BAP modes only. It can be enabled from the front panel under the Display Data Line Configuration setup menu, or over the GPIB. The Crest Factor feature is reset by pressing [Reset Line n] of the appropriate line or, in remote control, by sending the GPIB command which activates the Crest Factor feature (see Section 3.8).
In Figure 2-19, the same power input trace is used in two graphs to illustrate the effect of a drop in average power, with and without a reset. In the top graph, the power drop is followed by a reset. The held value drops to the current measured value, and the crest factor represents the ratio between the new maximum level and the new average level. In the bottom graph, there is no reset after the power drop, and the crest factor represents the ratio between the old maximum level and the new average level. For this reason, the crest factor feature should be reset after an input power level change.
Front Panel Operation
Crest Factor With a Power Drop Followed by a Reset
Power
Power
Hold
Hold
Hold
Crest Factor (dB)
Hold
Hold
(Reset)
Time
Avg.
Crest Factor (dB)
Crest Factor With a Power Drop But No Reset
Hold
Hold
Hold
Crest Factor (dB)
Avg.
Crest Factor (dB)
Hold
Avg.
Hold
Avg.
Time
Figure 2-19: Crest Factor
Manual 31470, Rev. E, April 2001 2-33
Page 66
Series 8650A Universal Power Meters
Like the Peak Hold feature, the Crest Factor feature also has the same minimum average power level restriction of -20 dBm. Beginning in firmware version 1.51, this has been implemented to increase accuracy and maintain wide bandwidth of Crest Factor measurement over the peak power range of
-20 dBm to +20 dBm. The Crest Factor selection will also affect Histogram measurement if there is power below -20 dBm. To obtain accurate average power measurements below -20 dBm, remove the display line containing Crest Factor (or, in remote control, by sending the command which disables the Crest Factor feature).
☛☛☛☛
NOTE:
[Reset Line n] for the Crest Factor also resets Peak Hold.
2-34 Manual 31470, Rev. E, April 2001
Page 67

2.3.14 Burst Signal Measurements

In a burst signal, the RF is pulsed on and off (i.e., pulse modulated). Often, the RF is modulated during the pulse on time. Typical examples are TDMA digital cellular telephone formats such as NADC, PDC, and GSM. These formats and many others produce amplitude modulation of the RF during bursts.
Two types of power measurement can be made on these types of signals. If the total average power is desired, MAP mode should be used. Total average power includes both the off and on time of the pulses in the averaging. Often it is desired to know the average power just during the bursts. BAP mode makes this type of measurement very easy. The procedure is as follows:
1. Calibrate the sensor according to the procedure outlined earlier in this section.
2. From the Main Menu press [Sensor Setup]. From the Sensor Setup menu, press [Modulated Sensor] and then select the BAP mode by pressing [BAP].
Front Panel Operation
3. Press
4. Connect the sensor to the burst signal source and record the power level.
The 8650A will automatically find the portions of the signal which are in the burst and include only those portions in the average.
Burst signals can have a high peak power-to-average power ratio depending on duty cycle. This ratio is proportional to the duty cycle and is given by:
This assumes no modulation during the burst. Modulation during the burst will increase this ratio by its own peak-to-average ratio. Due to this characteristic of burst signals, care must be taken to keep the peak power below the maximum rated input power of the sensor.
☛☛☛☛
[FREQ]
NOTE:
bursts are too narrow, the 8650A may lose sync with the bursts and fail to display the burst average power. When this happens, NoSync will flash on the screen to the right of the sensor power units, and the meter will display total average power as in MAP mode. The conditions under which the 8650A may lose sync are listed in the following Sections 2.3.15 and 2.3.16 and also in the minimum pulse width specifications in Section B.2.2 for the modulated sensor used.
and enter the operating carrier frequency.
If the burst average power is too low or if the bursts on time or off time between the
Manual 31470, Rev. E, April 2001 2-35
Page 68
Series 8650A Universal Power Meters

2.3.15 Burst Start Exclude, Burst End Exclude

When measuring burst signals, it is sometimes desirable to mask the beginning or the end of a burst so that overshoot and other distortions do not affect the reading.
The Burst Start Exclude and Burst End Exclude features make it possible for BAP mode measurements to exclude the beginning or the end of a burst in this way. Both features can be used simultaneously, but this requires caution: if the excluded periods overlap, there is nothing left of the burst to be measured. If the entire burst is excluded,
NoSync
the meter will revert to average measurement in the style of the MAP mode.
The start exclude time is selected from a series of discrete values (0.000 ms, 0.030 ms, 0.059 ms,
0.089 ms, and so on up to 45.548 ms). The end exclude time has the same discrete values available except the maximum time value is limited to 31.949 ms.
will flash on the screen to the right of the sensor power units, and
Burst Start Exclude
Burst Width
Exclude
Power
Burst End Exclude
Include
Power
Figure 2-20: Burst Start Exclude & Burst End Exclude
Include
Time
Burst Width
Exclude
Time
2-36 Manual 31470, Rev. E, April 2001
Page 69

2.3.16 Burst Dropout

In the BAP mode, average power is measured only during bursts. Because, in this mode, the bursts are automatically detected by the power meter, the user need not be aware of the burst repetition rate in order to make the measurement.
However, the BAP measurement algorithm defines bursts in a way which may be considered undesirable in some applications. In the example illustrated below, a 3.5 ms burst is followed by an OFF period of the same duration. During the burst, two brief dropouts occur. Normally, in BAP mode, each dropout would be interpreted as the end of a burst; the BAP algorithm would interpret the burst as three separate bursts, and the dropouts would be excluded from the average power measurement. As a result, the average power reading would be artificially raised.
When the Burst Dropout feature is enabled, the BAP algorithm is modified so that a dropout of sufficiently brief duration is not interpreted as the end of a burst. In the example below, dropout time is
µ
specified at 350 therefore the entire burst is interpreted as a single burst, and the dropouts are included in the average power measurement. The 3.5 ms OFF period following the burst is interpreted as the end of the burst, because it exceeds 350
This feature must be configured and interpreted with care. The dropout time is selected from a series of discrete values (.030 ms, .059 ms, .089 ms, and so on up to 3.747 ms); however, these are only the guaranteed minimum values. In practice, the BAP algorithm may tolerate dropouts up to 2.15 times as long as the minimum value. Therefore, the time between bursts must be at least 2.2 times as long as the selected dropout time (because, if the time between bursts is less than the tolerated dropout time, the BAP algorithm never recognizes the end of a burst, and the signal is simply averaged, as if the MAP mode had been selected). Also, dropouts occurring at the end of a burst are a problem, because the BAP algorithm cannot distinguish them from the end of the burst itself; there should be at least 250 burst remaining after the last dropout within that burst.
s. The two dropouts, which occur during the burst have a duration of less than 350 µs;
µ
s in duration.
Front Panel Operation
µ
s of
350
(dropout time)
Power
Burst definition covers this entire time period, including
µs
the two dropouts because they are <350 µs
245 µs
Dropout
Figure 2-21: Burst Dropout
Burst Dropout
(Dropped Time = 350 µs)
Burst definition does not cover this 3.5 ms period because it exceeds 350 µs
280
µs
Dropout
Time
Manual 31470, Rev. E, April 2001 2-37
Page 70
Series 8650A Universal Power Meters

2.3.17 Optimizing Measurement Speed

In many power measurement situations, measurement speed is defined in terms of settling time following a step change in average power. In other words, it is desired to know the average power level within some specified tolerance as quickly as possible following a power level change. This is often accomplished by setting up the power meter in free-run mode over the GPIB and monitoring the collected measurement data with the host computer until it falls within the predetermined tolerance window.
The Auto average feature of the 8650A eliminates the need for the host computer to do any data monitoring and can be set up to automatically output measurement data when it has settled to within the specified tolerance. This is done by triggering each measurement with a TR2 command and waiting for the meter to signal the host with an SRQ. The SRQ is asserted and the data is put on the bus as soon as the power measurement has averaged long enough to be within the specified tolerance.
The tolerance is specified by including the measurement settling tolerance parameter with an FA command (Auto average on). This parameter is specified in terms of percentage. For example, if a measurement settling tolerance of 1% is specified, the 8650A Auto average algorithm will specify an averaging time just long enough so that the result put on the bus is within ±0.5% (that is, ±0.02 dB) of the average power. Thus, the settled measurement data is available on the bus in the minimum time necessary to be within the specified tolerance.
The tolerance specified in the FA command is a target tolerance. For example, it is possible that the peak-to-peak power variation of the signal being measured is so great that the maximum averaging time of 20 seconds is not long enough to reduce the variation to within the specified tolerance. It is also possible that the rate of power variation is so slow that more than 20 seconds of averaging is required. In these cases, further averaging would have to be done by the host computer.
The following example program shows how to set up a triggered measurement, optimized for speed using the auto averaging feature:
Tr2: ! Read using TR2 command ON INTR 7 GOSUB Srq_interrupt ! Set up SRQ interrupt ENABLE INTR 7 ! Enable SRQ interrupt OUTPUT 713;*SRE41 ! Set service request mask OUTPUT 713;CS ! Clear status byte OUTPUT 713;TR2 ! Trigger measurement Data_ready=0 ! Clear flag WHILE Data_ready=0 ! Wait for data ready END WHILE RETURN Srq_interrupt: ! SRQ jumps here State=SPOLL(713) ! Get status byte IF BIT(State,0) THEN ! If the Data Ready bit is set...
Data_ready=1 ! Set the flag ENTER 713;Tr2_reading ! Read the measurement OUTPUT 713;CS ! Clear the status byte OUTPUT 713;*SRE0 ! Clear the service request mask END IF
RETURN
2-38 Manual 31470, Rev. E, April 2001
Page 71

2.3.18 Peak Power Measurements

Peak power sensors directly measure the amplitude of pulsed microwave signals. The direct sampling technique is more accurate than traditional duty cycle correction methods. The sample position can be displayed on an oscilloscope.
1. Calibrate a peak power sensor and connect it to a pulsed microwave source.
2. From the Main Menu, press [Sensor Setup]. From the Sensor Setup menu, select the Peak sensor with the cursor keys.
3. Press [Config] to display the Setup mode menu and select [Peak].
4. Select the desired trigger level (for internal or external triggering).
5. Select the desired sample delay (for internal or external triggering).
6. Optionally, set the desired delay offset (for internal or external triggering).
7. Connect the peak power sensor’s Detector Out to an oscilloscope to view the sample position. For 80350A Peak Power Sensors, also connect the sensor’s Sample Delay output to the oscilloscope and trigger on that channel.
Front Panel Operation

2.3.19 Measuring an Attenuator (Single Channel Method)

Attenuators are useful for many applications. With the 8650A, attenuators can be calibrated quickly and accurately. The single channel calibration procedure outlined below is efficient for calibrating at a single frequency or at a limited number of frequencies.
1. Connect the power sensor to the signal source through a 6 dB attenuator (a matching pad) and adjust the source output power to about 0 dBm. Verify that the source output is stable.
2. Press
3. From the Main menu, press [Rel] to set the reference level.
4. Insert the attenuator to be calibrated between the matching pad and the power sensor.
5. Record the attenuator value.
[FREQ]
and enter the operating frequency (this step is optional).
Manual 31470, Rev. E, April 2001 2-39
Page 72
Series 8650A Universal Power Meters

2.3.20 Improving Accuracy

Mismatch uncertainty is the largest source of error in power measurement. The 6 dB pad that is used in the attenuator calibration procedure above reduces mismatch uncertainty by effectively improving the return loss (or reducing the SWR) of the source. Mismatch uncertainty is large when a device has a
poor impedance match relative to 50
Poorly matched devices reflect a large proportion of incident signals and create standing waves along the transmission line. At various points along the transmission line, the standing wave will be at maximum or minimum amplitude. Mismatch uncertainty is a measure of the deviation between these amplitude levels.
Inserting an attenuator into the transmission line reduces mismatch uncertainty by reducing the amplitude of the reflected signal, thereby reducing the difference between a standing wave’s maximum and minimum levels.
Compared to an attenuator, most microwave sources have poor impedance matching. Using the 6 dB attenuator during the calibration has the effect of lowering the SWR of the microwave source. The only compromise is a corresponding 6 dB reduction in the source’s dynamic range when the 6 dB attenuator is attached.
.
2-40 Manual 31470, Rev. E, April 2001
Page 73

2.3.21 Performance Verification

Verifying accuracy and calibrating test equipment are essential to microwave engineers and technicians. Accurate, repeatable measurements are required for validating designs, certifying calibrations, making engineering decisions, approving product components, certifying standards, and verifying performance specifications (also see Chapter 4 of this manual).
1. A 6 dB attenuator is placed at the input port of a power splitter to provide a good impedance match from the source. This effectively reduces the VSWR of the source. Depending on the signal quality of your source over frequency, additional attenuation may be desirable. A two-resistor power splitter provides consistently matched power levels at its output ports, X and Y. The largest sources of error are power splitter tracking errors and mismatch uncertainty.
2. Connect the reference standard power meter to power splitter output X, and the power meter to be verified to splitter output Y.
3. Adjust the source frequency to a standard reference frequency (50 MHz for 803XX and 804XX series senors, or 1GHz for 806XX and 807XX series sensors).
4. Enter the operating frequency or frequency cal factors into the power meters.
5. Adjust the source amplitude to the maximum sensor operating level (+20 dBm for standard sensors).
Front Panel Operation
6. Zero each power meter and record the measurement values immediately after settling.
7. Adjust the source for +19 dBm output level and repeat Step 6.
8. Continue testing at 1 dB increments through the rest of the standard sensor’s 90 dB dynamic range.
9. Calculate measurement uncertainty and compare the measured results to the specified tolerances.
At low power levels, be sure to zero the sensor prior to taking measurements. At levels below -55 dBm, the measurements should be recorded just after zeroing is completed. The zeroing process must be repeated periodically, depending on the operating level, due to drift characteristics.
Manual 31470, Rev. E, April 2001 2-41
Page 74
Series 8650A Universal Power Meters

2.3.22 Sources of Error

In the previous accuracy verification procedure, there are four sources of error:
Source output level variation
Power splitter output tracking
Power meter X total measurement uncertainty
Power meter Y total measurement uncertainty
Worst case uncertainty, which should be used for calibration purposes, is the arithmetic sum of all four of these sources of error.
Source output level variation occurs in all microwave sources. This happens when the signal source output level changes during the time it takes to record the displayed value on power meter X and then to read the displayed value on power meter Y. This source of error can be minimized by using a laboratory grade signal source.
Power splitter output tracking errors are the maximum signal level variation at the splitter X output as compared to the splitter Y output.
Total measurement uncertainty for each of the power meters is the worst case combination of mismatch uncertainty, instrument accuracy, and sensor accuracy.
Mismatch uncertainty is calculated from the reflection coefficients of the sensor and the splitter (source) according to the following formula:
For a source mismatch specified in terms of return loss (RL), the equation should be modified according to:
The following factors affect instrument accuracy:
Instrument linearity or instrumentation uncertainty
Reference calibrator setability or power reference uncertainty
The following factors affect sensor accuracy:
Calibration factor uncertainty
Calibrator to sensor (or power reference to sensor) mismatch uncertainty
•Noise
Zero set
Calibration pad uncertainty (for thermal-based power meters only)
Sensor linearity
2-42 Manual 31470, Rev. E, April 2001
Page 75

3.1 Introduction

The Series 8650A can be operated from a remote host over the General Purpose Interface Bus (GPIB) using either Standard Commands for Programmable Instruments (SCPI) or IEEE Standard 488-1978 (Digital Interface for Programmable Instruments) commands.
Table 3-1 shows which functions of the IEEE 488 standards are implemented in the 8650A.

Table 3-1: Implemented IEEE Standards

Function 8650A Implementation
Source Handshake SH1 (complete capability)
Acceptor Handshake AH1 (complete capability)
Talker T5 (basic talker, serial poll, talk only mode, unaddressed if MLA)
Extended Talker TE0 (no capability)
Listener L3 (basic listener, listen only mode, unaddressed if MTA)
Extended Listener LE0 (no capability)
Service Request SR1 (complete capability)
Remote/Local RL1 (complete capability)
Parallel Poll PP1 (remote configuration)
Device Clear DC1 (complete capability)
Device Trigger DT1 (complete capability)
Controller C0 (no capability)
3

Remote Operation

3.1.1 Sending Commands to the 8650A
The 8650A power meter uses standard protocols for communication over the GPIB. Commands conform to IEEE 488.1 or IEEE 488.2 guidelines. Four emulation modes (Giga-tronics 8540, HP436, HP437 and HP438) are available for users of power meters who cannot rewrite their application software.
The program examples in this chapter are written in HTBasic™ format (HTBasic is a trademark of TransEra Corporation). Other languages would use different commands but the string that is sent or received will always be the same. In HTBasic, the number after
The factory-set default address of the 8650A is 13 and the address of the GPIB is assumed to be 7; therefore, examples of command strings in this manual are preceded by
Manual 31470, Rev. E, April 2001 3-1
OUTPUT
is the GPIB address of the instrument.
OUTPUT
command sends a string to the GPIB. The
OUTPUT 713
.
Page 76
Series 8650A Universal Power Meters
Clear Device
The interface command CLEAR 713 resets the GPIB and sets the 8650A to its preset condition.
Clear Interface
The interface command ABORT 7 resets the GPIB without resetting the 8650A to its preset condition. The 8650A will not be addressed after the abort.
Local and Remote Control
The interface command LOCAL 713 places the 8650A into the local control mode.
The interface command REMOTE 713 places the 8650A into the remote control mode. Enter LOCAL 713 to return the instrument to local mode.
The interface command LOCAL LOCKOUT 7 places the 8650A in the local lockout mode. This is a remote control mode in which all of the 8650A front panel keys are disabled. The GPIB LOCAL command must be issued to return the 8650A to local mode (disconnecting the GPIB cable will also return the instrument to local mode).
Sensor Selection and Calibration
Power sensor selection data, specifications, and calibration (local and remote) are contained in Appendix B of this manual.
3.1.2 Polling
The GPIB supports parallel and serial polling. The example programs below show how to use the parallel and serial poll capabilities of the 8650A to determine when a requested zeroing operation is completed.
Parallel Polling
Ppoll_zero ! zero using parallel poll PRINT entering parallel poll zero routine PPOLL CONFIGURE 713;8 ! configure response on bit zero OUTPUT 713;CS AEZE ! clear status byte, zero channel A State=0 ! initialize variable WHILE State 1 ! stay here until zero done
PRINT parallel zero done RETURN
State=PPOLL(7) ! read the poll END WHILE PPOLL UNCONFIGURE 713 ! cancel parallel poll mode
3-2 Manual 31470, Rev. E, April 2001
Page 77
Serial Polling
Srq_zero: ! zero with an srq interrupt PRINT entering SRQ interrupt zero routine ON INTR 7 GOSUB Srq_interrupt OUTPUT 713;CS ! clear status byte ENABLE INTR 7;2 ! enable srq interrupts OUTPUT 713;@1;CHR$(2) ! enable srq handshake OUTPUT 713;AEZE ! execute zero command Flag=0 ! test flag reset to false
WHILE Flag=0 ! stay here until test flag set true WAIT 1
PRINT Still inside while loop END WHILE PRINT SRQ interrupt zero done RETURN Srq_interrupt: ! SRQ interrupts jump here PRINT an SRQ interrupt has occurred Example:OUTPUT 713;CS ! clear status byte Flag=1 ! set control flag true RETURN
Remote Operation
3.1.3 Data Output Formats
The data output format for the standard measurement collection mode is:
±±±±
D.DDDDE
±
: Sign of the Mantissa D.DDDD: Mantissa (5 digits) E: Exponent (indicates that an exponent follows)
±
: Sign of the Exponent NN: Magnitude of the Exponent CR: Carriage Return LF: Line Feed
Data output formats for the swift and fast buffered modes are expressed in the form of a signed five-digit number with two digits to the right of the decimal and no exponents. In some cases multiple values are sent:
One sensor swift mode:±DDD.DD CRLF Two sensor swift mode: Fast buffered mode:
±±±±
NNCRLF
±
DDD.DD,±DDD.DD CRLF
±
DDD.DD, . . . . .±DDD.DD CRLF
Manual 31470, Rev. E, April 2001 3-3
Page 78
Series 8650A Universal Power Meters
3.1.4 Power-On Default Conditions
The interface wake-up state is:
GPIB Local Mode
Unaddressed, Service Request Mask Cleared
Status Byte Cleared
TR3 Free Run Trigger Mode Set
GT2 Group Execute Trigger Mode Set
Parallel Poll Data Line Unassigned
Display Enabled
Service Request Mask Cleared
Event Status Register = 128
Event Status Mask Clear
3-4 Manual 31470, Rev. E, April 2001
Page 79

3.2 SCPI Command Interface

This section details operation of the 8650A power meter using the SCPI (Standard Communications for Programmable Instruments) interface commands. A SCPI command reference is presented in Table 3-3 and the sections that follow.
SCPI commands promote consistency in definition of a common instrument control and measurement command language. The structured approach of the SCPI standard offers test system design engineers a number of system integration advantages that achieve considerable efficiency gains during control program development.
SCPI compatible instrument commands are structured from a common functional organization or model of a test instrument (see Figure 3-1). Most of the power meter configuration and measurement functions fall within the Measurement Function Block and the Trigger Subsystem of the SCPI instrument model.
The 8650A uses the Sense Subsystem of the Measurement Function Block to implement commands that apply specifically to the individual power sensors; sensor 1, and sensor 2. For example, the SENSe2:CORRection:OFFSet command corrects for the attenuation of a signal that passes through an attenuator or coupler before it is measured by power sensor 2. Figure 3-1 illustrates the SCPI subsystem model.
Remote Operation
NOTE:
☛☛☛☛
as CALCulate and MEMory. The uppercase is the required input. The whole word can be used if desired. If the whole word is used, it must be used in it’s entirety. For example, CALCulat will not be recognized as a valid command.
Throughout this manual, some commands will be in both upper- and lower-case, such
CAL
ibrate

Figure 3-1: SCPI Subsystem Model

SENS
TRIG

3.2.1 Sensor Calibration and Zero

The Calibration subsystem performs sensor calibration and zeroing. CALibrate:STATen selects if the calibration data is applied or not. If STATe is ON, then the instrument uses the calibration data for correction. If STATe:OFF is selected, then no calibration using the calibration data will be made. The CALibrate:ZERO subsystem controls the autozero calibration of the sensor.
e
ger
CALC
MEM
ulate
ory

3.2.2 Sensor and Channel Configuration

The Calculate subsystem of the Measurement Function Block contains commands that define the form of the measured data from sensor 1 and sensor 2. Calculate commands define the configuration of the two Software Calculation Channels. For example, the CALC1:POW 1 command configures channel 1 to report the power level as measured by sensor 1. CALC2: RAT 2,1 configures channel 2 to report the ratio of power levels, sensor 2 over sensor 1.
Manual 31470, Rev. E, April 2001 3-5
Page 80
Series 8650A Universal Power Meters

3.2.3 Measurement Triggering

The power meter uses the Trigger Subsystem to trigger measurements in two different operational modes
- a normal mode which maximizes the instrument’s functionality, and swift and burst modes that maximize the power measurement rate.

3.2.4 Memory Functions

The MEMory commands control the configuration of the automated Voltage-Proportional-to­Frequency (V
F or VF), sensor Cal Factor correction, and the analog outputs on the rear panel. Each
PROP
of these connectors is used with external devices. The V output of your microwave source. The analog outputs are used with a variety of devices including chart recorders, oscilloscopes, voltmeters, and microwave source leveling inputs.

3.2.5 IEEE 488.2 Required Commands

Consistent with SCPI compliance criteria, the power meter implements all the common commands declared mandatory by IEEE 488.2 (see Table 3-2).

Table 3-2: IEEE Required Command Codes

Mnemonic Name
*CLS Clear Status Command
*ESE Standard Event Status Enable Command
*ESR? Standard Event Status Register Query
*IDN? Identification Query
*IST? Individual Status Query
*OPC Operation Complete Command
*PRE Parallel Poll Enable Register Command
*RCL Recall Command
*RST Reset Command
*SAV Save Command
*SRE Service Request Enable Command
*STB? Read Status Byte Query
*TRG Trigger Command
*TST? Self-Test Query
*WAI Wait-to-Continue Command
F can be configured to match the V
PROP
PROP
F
3-6 Manual 31470, Rev. E, April 2001
Page 81
3.2.6 Calculate Subsystem Commands
Calculate commands specify and query the configuration of power measurement channels, known in SCPI references as Software Configuration Channels, and in this manual as “channels”. See Section 3.2.7 for sensor-specific configuration and measurement function control.
The query form of CALC#?, with # replaced by the channel modifier (1 or 2), returns the current configuration status for that channel. This verifies configuration commands or return current status information following data acquisition or power measurements.
CALC
Channel Configuration
POW
RAT
er
io
LIM
it
UPP
ulate
REF
er
erence
Remote Operation
MAX
imum
STAT
e
STAT
e
STAT
e
DIFF
erence
[:CHAN:FUNC]?
UNIT
Figure 3-2: CALCulat Subsystem Commands
LOW
FAIL
FCO
CLE
STAT
?
?
ar
er
COLL
ect
MIN
imum
STAT
e
e
Limit Lines are set on a channel basis. These commands set limits, monitor the number of violations, and allow the violation counter to be cleared.
The REFerence command allows channel-based offset values. For example, using CALC#:REF:COLL automatically converts the inverse of the current channel measurement value to an offset — simplifying the 1 dB compression testing of amplifiers.
MIN and MAX commands monitor deviation of measured values over a user controllable time period.
The two software calculation channels can individually and simultaneously perform the internal instrument functions that calculate final measurement data. The final measurement data is calculated from Sense subsystem sensor data as well as the Calculate subsystem channel configuration data.
This means that only two measurement configurations can be obtained from the 8650A simultaneously. For example, the controller can obtain measurements for sensor 1 plus sensor 2/sensor 1 simultaneously, but not sensor 1 plus sensor 2 plus sensor 2/sensor 1 simultaneously.
Manual 31470, Rev. E, April 2001 3-7
Page 82
Series 8650A Universal Power Meters
3.2.7 Sense Subsystem Commands
The Sense subsystem configuration commands, illustrated in Figure 3-3, apply to individual sensors. These commands alter the value of the measured power level according to the sensor’s characteristics. For example, measured power levels can be offset for attenuators or couplers in the measurement path so that the power data reading reflects the power level at the measurement point of interest.
Use the Sense subsystem commands for:
Averaging power measurements in the 8650A
Offsetting power measurements for attenuation or amplification
Entering the operating frequency of the measured signal computes and applies sensor specific Calibration Factor corrections, which compensates for sensor frequency response characteristics
Controlling Peak Power Sensor triggering
Sense subsystem commands control functions that are related directly to the individual power sensors. For example, these commands control items that would not apply to numerical alteration of a ratio measurement of Sensor1/Sensor2. Controls that would apply to that type of a configuration are channel functions, not sensor functions, and would therefore be located in the Calculate subsystem.
COUNt
STATe
AVERage
TCONtrol
SENSe
OFFSet
STATe
COLLect
FREQuency
VPROpf
EEPROM
TYPE?
FREQuency
CALFactor
For Peak Power Sensors Only
TRIGgerCORRection
DELay
LEVel
SOURce
OFFSet
TOTAl?
Figure 3-3: SENSe Subsystem Command Tree
SENSe:AVERage functions control the number of data samples for each measurement and the manner in which those numbers are accumulated. COUNt determines the averaging number or AUTO­averaging. TCONtrol determines whether each new sample is added to previous COUNt # of samples or if COUNt # of samples are taken each time the 8650A is triggered. Please note that the SENSe:TRIGger commands are not instrument triggers, but Peak Power Sensor configuration controls.
3-8 Manual 31470, Rev. E, April 2001
Page 83
SENSe:TRIGger functions apply only to Giga-tronics Peak Power Sensors. The DELay and LEVel
General Instrument Measurement Triggering
Burst or Swift Mode Data Acquisition Controls
TRIGger
COUNt
DELay
MODESOUR ce
HOLD
IMMediate
EXTernal
BUS
functions of these Peak Power Sensor controls apply to the 80350A Peak Power Sensors, not the 80340 Peak Power (Triggerable Pulse) Sensors. The AVERage and CORRection commands apply to all 80300 Series CW & Peak Power Sensors.
STATe ON OFF controls for COUNt and DELay, above, are not shown.
3.2.8 Trigger Subsystem Commands
The Trigger Subsystem is divided into two sections; Instrument Measurement Event Triggering, and Special Triggering Configuration commands for the fast-reading buffered data modes (Burst and Swift modes). The Trigger command tree is illustrated in Figure 3-4.
Remote Operation
Figure 3-4: TRIGger Subsystem Command Tree
The query form of these commands, TRIGger? with the appropriate modifier inserted ahead of the ?, will return the instrument’s current configuration status. This can be used to verify triggering configuration or return status information following command errors which are commonly caused by using illegal configuration commands during Swift or Burst Modes.
SOURce:IMMediate triggering allows the 8650A to control measurement triggering; this is the default configuration. External triggering is performed using a TTL signal input. BUS allows software controlled triggering.
COUNt refers to the number of data points to store in the meter’s 5000 reading buffer (128,000 with option 02) before the measurement data is requested by the controller.
DELay controls the time interval between Burst Mode data samples, and the MODE command controls whether the data is taken after receipt of the instrument trigger or if data is collected (in a FIFO buffer) immediately preceding receipt of an instrument trigger.
Manual 31470, Rev. E, April 2001 3-9
Page 84
Series 8650A Universal Power Meters
3.2.9 GPIB Command Syntax
The following conventions are used with the GPIB commands in this manual. Some commands will be in both upper- and lower-case, such as CALCulate and MEMory. The uppercase is the required input. The whole word can be used if desired. Table 3-3 lists in alphabetical order all of the GPIB commands supported by the power meter. The basic function performed by that command is given, and the page number of this chapter where a description of each command can be located.
Commands in Brackets
Commands and command separators within brackets, such as [COMMand:], are optional. These portions of the commands may be used in your program command strings, but are not required for proper operation of the power meter.
Programmer Selective Parameters
Command descriptions enclosed in angle brackets (< and >) show the syntax placement of configurable parameters. A description of the necessary parameter and the range of values or mnemonics valid for that parameter are enclosed in angle brackets.
Italics in Syntax Descriptions
Some command syntax descriptions show certain words in italics such as space and comma to indicate where the character must be included within a command string.
Query Format
Except where specifically noted, all query commands are formed by adding a question mark (?) to the command header. Be sure to omit command parameters when you are using the query format. Some commands have only a query format. With the exception of the Calibration queries, query commands will not change the status of the power meter. The CALibration1? and CALibration1:ZERO? commands query respectively to automatically begin the calibration and zeroing process.
Linking Command Strings
The 8650A uses ASCII strings for commands. When sending more than one command in a single string, a semicolon must be used as a delimiter between commands. No spaces or other characters are necessary. Use only a semicolon to link commands in a string.
Measurement Data Output Format
The examples in this chapter are written in HTBasic™ format. Different languages will use different commands, but the string sent or received will always be the same. In HTBasic, the OUTPUT command sends a string to the GPIB bus. The number or variable after the word OUTPUT is the GPIB address of the power meter.
™HTBasic is a trademark of TransEra Corporation.
3-10 Manual 31470, Rev. E, April 2001
Page 85
Table 3-3: SCPI Command Syntax
Remote Operation
Command Synt ax Function
SCPI Required Codes
*CLS Clear SRQ and status byte registers
*ESE Enable bit 4 of event status register mask
*ESR? Return event status register value
*IDN? Query instrument mfgr and model number
*IST? Individual Status Query
*OPC Allows a one-time SRQ enable
*PRE Parallel Poll Enable Register Command
*RCL Recall 8650A register n
*RST Reset 8650A configuration
*SAV Save at 8650A register n
*SRE Service Request Enable
*STB? Query status byte register
*TRG BUS Trigger Event
*TST? Query self-test result
*WAI Wait, following command completion
SCPI Function Codes
ABORt Halts measurement & triggering
CALCulate<channel 1 or 2>:DATA? Query channel 1 & 2 burst mode data
CALCulate<channel 1 or 2>[:CHANnel]: differencespace<sensor 1 or 2commasensor 2 or 1>
CALCulate<channel 1 or 2>[:FUNCtion]? Quer y channel 2 sensor mode
CALCulate<channel 1 or 2>:LIMit:CLEar[:IMMediate] Reset channel 1 limit violation indicator to 0
CALCulate<channel 1 or 2>:LIMit:FCOunt? Query number of channel 1 limit failures
CALCulate<channel 1 or 2>:LIMit:FAIL? Check for channel 1 limit line violation
CALCulate<channel 1 or 2>:LIMit:LOWerspace<numeric value in dB from -299.99 to 299.99>
CALCulate<channel 1 or 2>:LIMit:STATespace<ON or OFF> Enable channel 1 upper and lower limit
CALCulate<channel 1 or 2>:LIMit:UPPerspace<numeric value in dB from -299.99 to 299.99> Set channel 1 upper limit line to 17 dBm
CALCulate<channel 1 or 2>[:CHANnel]:POWerspace<sensor 1 or 2>Configures channel 2 to measure Sensor 1
CALCulate<channel 1 or 2>[:CHANnel]:RATiospace< sensor 1 or 2commasensor 2 or 1>
CALCulate<channel 1 or 2>:MODEspace<NORMal, BURSt or SWIFt Mode Selection>
CALCulate<channel 1 or 2>:MAXimum:STATespace<ON or OFF>
CALCulate<channel 1 or 2>:MAXimum[:MAGnitude]? Query channel 2 maximum value in dBm
CALCulate<channel 1 or 2>:MINimum:STATespace<ON or OFF>
Configures channel 1 to measure Sensor 2 minus Sensor 1 power
(Output 0 = OK; 1 = fail)
Set channel 1 lower limit line to -50 dBm
checking. (0 = off; 1 = on)
power
Configures channel 1 to measure Sensor 2 over Sensor 1 power
Set measurement mode to SWIFt
Enable channel 2 max. value monitoring
Enable channel 1 min. value monitoring
Pag e
3-36
3-36
3-36
3-49
TBA
3-36
TBA
3-46
3-47
3-46
3-36
3-38
3-19
3-51
3-21
3-46
3-27
3-23
3-24
3-42
3-42
3-43
3-43
3-43
3-43
3-23
3-23
3-23
3-40
3-40
3-40
Manual 31470, Rev. E, April 2001 3-11
Page 86
Series 8650A Universal Power Meters
Table 3-3: SCPI Command Syntax (Continued)
Command Synt ax Function Pa ge
CALCulate<channel 1 or 2>:MINimum[:MAGnitude]? Query channel minimum value in dBm
CALCulate<channel 1 or 2>:REFerence:STATespace<ON or OFF>
CALCulate<channel 1 or 2>:REFerence[:MAGnitude]<dB Offset value from -299.999 to 299.999>
CALCulate<channel 1or 2>:REFerence:COLLect Take channel 2 reading as reference value.
CALCulate<channel 1 or 2>:STATespace<ON or OFF> Disable channel 1 measurement
CALCulate<channel 1 or 2>:UNIT[:POWer]space<data units selection, dBm or Watt>
CALibrate<sensor 1 or 2>[:SENSor] Calibrate Sensor 1
CALibrate<sensor 1 or 2>:STATe? Query sensor for calibration status
CALibrate<sensor 1 or 2>:ZERO Zero Sensor 1
FETCh<channel 1 or 2>? Read previously triggered channel 1 power,
INITiate[:IMMediate] Initiate instrument triggering cycle
INITiate:CONTinuousspace<ON or OFF> 8650A self-triggers continuously
MEASure<channel 1 or 2>[:SCALar:POWer]? Configure triggering and measure Channel 2
MEMory[:TABLe]:CHANnelspace<channel 1 or 2> Set analog out to channel 2
MEMory[:TABLe]:FREQuency < Star t frequency> Set V
MEMory[:TABLe]:POWerspace<start value from -80 to +20 dBm>comma<stop value from -80 to +20 dBm>
MEMory[:TABLe]:POWerspace<start value from 0 to
0.01W>comma<stop value from 0 to 0.01W>
MEMory[:TABLe]:SELectspace<ANALOGout,VPROPF1, VPROPF2>
MEMory[:TABLe]:SLOPespace<Volts per Hz> Set V
MEMory[:TABLe]:UNITspace<Units for analog output configuration, dBm or Watt>
MEMory[:TABLe]:VOLTagespace<start value from -10 to +10 V>comma<stop value from -10 to +10 V>
OUTPut[:BNC]:ANAlog[:STATe]space<ON or OFF> Enable analog out function
OUTPut:ROSCillator[:STATe]space<ON or OFF> Turn ON 0.0 dBm Calibrator Oscillator
READ<channel 1 or 2>[:POWer]? Trigger measurement and read Channel 2
SENSe<sensor 1 or 2>:AVERage:COUNtspace<averaging value 1,2,4,8,16,32,64,128,256,or512>
SENSe<sensor 1 or 2>:AVERage:COUNt:AUTOspace<ON or OFF>
SENSe<sensor 1 or 2>:AVERage:TCONtrolspace<data acquisition averaging method, MOVing or REPeat>
SENSe<sensor 1 or 2>:CORRection:EEPROM:TYPE? Query sensor 1 eeprom type
SENSe<sensor 1 or 2>:CORRection:EEPROM:FREQuency? Query sensor 1 eeprom frequency table
SENSe<sensor 1 or 2>:CORRection:EEPROM:CALFactor? Query sensor 1 eeprom cal factor table
SENSe<sensor 1 or 2>:CORRection:FREQuency[:CW-FIXed] 5e7
SENSe<sensor 1 or 2>:CORRection:OFFSet:COLLect Take sensor 1 current reading as offset value
Activate level reference for relative measurements
Set channel 2 reference offset value in dB Reset channel 2 Relative meas. operation
Reset with CALC:REF 0.0
Selects channel 1 linear units in Watts Selects channel 1 Log Units in dBm
Normal, Swift or Burst mode data
power in NORMal mode
F start frequency to 1 GHz
PROP
Set analog out power range (in dBm) from -80 to 20 dBm
Set analog out power range (in Watts) from 0 to
0.01 Watts
Select memory table V following lines
F slope in 1V/GHz
PROP
F1 for editing in
PROP
Set analog out power unit
Set analog out voltage range from -8 to 2 volts corresponding to power
Set sensor 1 average number to 16
Select sensor 2 Auto-average mode
Set sensor 1 to acquire fresh measurement data before averaging
Set sensor 2 frequency to 50 MHz
3-40
3-34
3-34
3-34
3-24
3-24
3-14
3-14
3-15
3-16
3-22
3-22
3-16
3-44
3-26
3-44
3-44
3-25 3-44
3-26
3-44
3-45
3-45
3-50
3-17
3-33
3-33
3-33
3-50
3-50
3-50
3-26
3-35
3-12 Manual 31470, Rev. E, April 2001
Page 87
Remote Operation
Table 3-3: SCPI Command Syntax (Continued)
Command Synt ax Function Pa ge
SENSe<sensor 1 or 2>:CORRection:OFFSet [:MAGnitude]space<offset value in dB -99.99 to 99.99>
SENSe<sensor 1 or 2>:CORRection:OFFSet:STATespace<ON or OFF>
SENSe<sensor 1 or 2>:CORRection:VPROpf[:STATe]space<ON or OFF>
SENSe<sensor 1 or 2>:TEMPerature? Quer y sensor 1 temperature (in centigrade)
SENSe<sensor 1 or 2>:TRIGger:SOURcespace<peak sensor triggering INTernal, EXTernal or CW>
SENSe<sensor 1 or2>:TRIGger:DELay[:MAGnitude]:space <Peak Sensor sample delay in seconds from -20e-9 to 104e-3> Set sensor 2 peak delay value to 1e6 seconds
SENSe<sensor 1 or 2>:TRIGger:DELay:STATespace<ON or OFF>
SENSe<sensor 1or2>:TRIGger:LEVel[:MAGnitude]space <trigger level of peak power sensor>
SENSe<sensor 1 or 2>:TRIGger :OFFSet[:MAGnitude] Set peak sensor 1 trigger offset time to 0
SENSe<sensor 1 or 2>:TRIGger :TOTAl[:MAGnitude]? Query peak sensor 1 total trigger delay time
*SREspace<event status register value, 128, 64, 32, 16, 8, 4, 2, 1>Enable bit 5 of Status Byte register
Compensate sensor 2 for 10.2 dB attenuation
Enable sensor 2 offset correction
Enable sensor 1 v_prop_f function
Set sensor 1 peak trigger mode to CW mode
Enable peak sensor 1 delay function
Set peak sensor 1 trigger level to 1.7 volts
second
3-35
3-35
3-25
3-15
3-31
3-31
3-31
3-32
3-32
3-32
3-36
STATus:OPERation[:ENABlet] Set Status Enable Mask
STATus:OPERation[:EVENt]? Query operation event register result
STATus:PRESet Clears all the status register value
SYSTem:COMMunicate:GPIB Collect Configuration of Control Interface
SYSTem:COMMunicate:SERial:RECeive:BAUDspace<baud rate>
SYSTem:COMMunicate:SERial:TRANsmit:BAUDspace<baud rate>
SYSTem:ERRor? Query system error message
SYSTem:LANGuage NATIVE Switch to native language
SYSTem:PRESet Reset 8650A configuration
SYSTem:VERSion? Query SCPI version
TRIGger[:IMMediate] Trigger measurement cycle
TRIGger:SOURcespace<instrument triggering source IMMediateBUS, HOLD, or EXTernal>
TRIGger:MODEspace<burst mode data gathering POST trigger receipt or PRE trigger receipt>
TRIGger:DELayspace<delay time between buffered readings in sec 0.000 to 5.000>
TRIGger :COUNtspace<number of data values to buffer in memory from 1 to 5000 for standard 8650s from 1 to 128,000 on one channel with option 02; from 1 to 64000 for two channels with option 02>
Set the receive baud rate
Set the transmit baud rate
Set instrument measurement trigger source to IMMediate
Set Burst or Swift mode data collection relative to Timing of event trigger post
Set Burst mode measurement time to 5 ms between measurements (resolution, one msec)
Set Burst or Swift mode buffer reading number to 100
TBA
3-37
3-37 3-47
TBA
TBA
TBA
3-51
TBA
3-47
3-49
3-19
3-19
3-20
3-20
3-21
Manual 31470, Rev. E, April 2001 3-13
Page 88
Series 8650A Universal Power Meters
3.2.10 Sensor Calibration and Zeroing
CALibrate<sensor 1 or 2> CALibrate<sensor 1 or 2>:STATe? CALibrate<sensor 1 or 2>:ZERO SENSe<sensor 1 or 2>:TEMPerature?
Sensor Calibration
The CALibration commands for sensor calibration and zeroing are important for accurate power measurement results. Be sure to perform the sensor calibration prior to beginning measurement operation or channel configuration. Sensors must be calibrated to the meter before performing measurements.
Zeroing of all active sensors should always be performed whenever a second sensor (whether calibrated or not) is added or removed. Zeroing should also be performed prior to measurement of low signal levels, generally within the lower 15 dB of a sensor’s dynamic range. For standard sensors, this is -55 dBm.
CAL#
Syntax: CALibration<sensor 1 or 2>[:SENSor]
Example: OUTPUT @Pwr_mtr;CAL1 ! Calibrate Sensor 1
Description: This command begins the sensor power sweep calibration process. Power sweep calibra-
tion can be performed only during Normal mode. The sensor must be attached to the front panel CALIBRATOR (use an adapter for high frequency sensors with Type K connectors). If the sensor is not connected or if the sensor is disconnected during the power sweep calibration procedure, the calibration automatically fails and the sensor power sweep calibration table is restored to its previous values.
CALibrate: STATe?:
Syntax: CALibrate<sensor 1 or 2>:STATe?
Example: OUTPUT @Pwr_mtr;CAL1:STAT? ! Query sensor for calibration status
Response: 1 if sensor is calibrated, or 0 if sensor is not calibrated.
Description: This command queries whether or not a sensor has been calibrated.
3-14 Manual 31470, Rev. E, April 2001
Page 89
Remote Operation
Sensor Zero
Zeroing automatically accounts for ground noise and other noise in your measurement system. Measurements will be sensitive to noise-induced errors only in the lowest 15 dB of the sensor dynamic range. Be sure to turn off the signal going into the sensor during zeroing, otherwise a failure will be indicated.
CAL#:ZERO
Syntax: CALibration<sensor 1 or 2>:ZERO
Example: OUTPUT @Pwr_mtr;CAL1:ZERO ! Zero Sensor 1
Description: This command begins the zeroing process. Since you want to offset for measurement
circuit noise during zeroing, you want the sensor to remain attached to the measurement circuit during zeroing. Disable the signal source into the sensor. If you cannot disable the signal source, connect the sensor to a grounded connector (preferably RF grounded) or leave the sensor disconnected. Do not connect the sensor to the power meter front panel calibrator port. If the meter detects excessive RF power at the start of the zeroing process (generally above -50 dBm for standard sensors), zeroing will automatically fail.
SENS#:TEMP?
Syntax: SENSe<sensor 1 or 2>:TEMPerature?
Example: OUTPUT @Pwr_mtr;SENS1:TEMP? ! Query sensor 1 temperature (in centigrade)
Response: Current sensor temperature in degrees centigrade
Description: This command reports sensor temperature. If the temperature varies in your operating
environment, monitor the relative temperature variations. If the variation since the previous
±5
power sweep calibration exceeds that sensor.
°C (±9 °F), perform the power sweep calibration for
Example programs for Sensor Calibration and Zeroing are contained in Appendix A.
Manual 31470, Rev. E, April 2001 3-15
Page 90
Series 8650A Universal Power Meters
3.2.11 Reading Power Measurements
FETCh? MEASure? READ? *OPC
These commands return measurement data from the 8650A. During Normal mode the data will be single measurement values. During Swift or Burst modes, the data will be an array of values. Generally, it is a single array if one sensor is connected and calibrated, and a dual array if two sensors are connected and calibrated.
FETC?
Syntax: FETCh<channel 1 or 2>?
Example: OUTPUT @Pwr_Mtr;FETC1? ! Read previously triggered channel 1 power
Response: Whatever is in the output buffer whether it is a new measurement or an old one.
Description: The FETCh? query command returns post-processed measurement data from the active
channels of the 8650A. After receiving FETCh? the 8650A will output the contents of it’s active data output buffer. The data size and output format of the buffer are dependent upon channel and measurement mode configuration.
! Normal, Swift, or Burst mode data.
For example, if channel 1 is defined as sensor 1 and channel 2 is defined as sensor 2/1, the FETCh? query command will return the power level incident upon sensor 1 and the ratio of the power levels incident upon sensor 2 and 1. FETCh? is used with NORMal mode measurements when INITiate:IMMediate is active, and for reading SWIFt and BURSt Mode measurement data from the meter’s data buffer.
MEAS?
Syntax: MEASure<channel 1 or 2>[:SCALar:POWer]?
Example: OUTPUT @Pwr_Mtr;MEAS2? ! Arm, Trigger, and Measure Channel 2
! power in NORMal mode.
Response: A power measurement.
Description: The MEAS? command returns measured data from the active software calculation
channels of the 8650A. The MEAS? command will also initiate the trigger cycle and will turn on the Auto Averaging mode; that is, measurement will be triggered and the data will be transmitted from the power meter. This is in contrast to the FETCh? command, which is capable of causing immediate output of the measurement data but not initiating the triggering cycle.
3-16 Manual 31470, Rev. E, April 2001
Page 91
Remote Operation
READ?
Syntax: READ<channel 1 or 2>[:POWer]?
Example: OUTPUT @Pwr_Mtr;READ2? ! Trigger measurement and read Channel 2
Response: A power measurement. A
Description: The READ[:POWer]? query command triggers measurement and sends the data to the
controller. The READ? command enables the power meter to acquire data at the next instrument trigger and return post-processed measurement data from the active channels of the 8650A. READ uses the current sensor averaging setup.
*OPC
Example: OUTPUT @Pwr_mtr;*OPC ! *OPC allows one-time SRQ enable
Description: The *OPC command determines when an operation is completed. This command is
generally used to monitor the completion of long measurement sequences. It sets the operation complete bit in the event status register upon completion of operation.
Special Errors
An unusual or non-sense numeric response, such as 9e+40, indicates that you are getting an error response. For instance, if you have not performed the power sweep calibration procedure to calibrate the sensor to the power meter, the response to a MEAS#? command will be 9.0000e+40.
Selected basic SCPI syntax and execution errors apply to these commands.
If you are using the READ#? measurement query and INIT:CONT is ON, you will get a bad value returned, 9e+40. If you send SYST:ERR? error query, -213, Init ignored will be returned. READ#? contains the low level function INIT, since INIT:CONT is ON the INIT within READ#? generates an error. Set INIT:CONT to OFF when using READ#?.
There are no device-specific errors for the preset configuration or status reset commands except the high level -300, Device-specific error response.
Example programs for Reading Power Measurements are contained in Appendix A.
Manual 31470, Rev. E, April 2001 3-17
Page 92
Series 8650A Universal Power Meters
3.2.12 Instrument Triggering
*TRG TRIGger[:IMMediate] TRIGger:SOURce TRIGger:MODE TRIGger:DELay TRIGger:COUNt INITiate:CONTinuous INITiate[:IMMediate] *WAI
These SCPI commands trigger the measurement cycle. They do not configure or provide triggering for Peak Power Sensors. Those commands are defined in Section 3.2.17. For power meter operation, the TRIGger Subsystem is divided into two sections; Instrument Measurement Event Triggering, and Special Triggering Configuration commands for the fast reading buffered data, Burst, and Swift modes.
General Instrument Measurement Triggering
HOLD
IMMediate
EXTernal
BUS
Figure 3-5: TRIGger Subsystem Command Tree
TRIGger
COUNt
Burst or Swift Mode Data Acquisition Controls
DELay
MODESOUR ce
EXT triggering is performed with the rear panel BNC connector and functions only in the BURSt and SWIFt Modes. EXT triggering is not available in the NORMal Mode. Provision has been made for a hardware ready for new trigger type handshaking capability using the analog output.
BUS triggering is available for all operating modes, BURSt, SWIFt, and NORMal.
IMMediate triggering allows the 8650A to free run and perform continuous measurements. This is the default setting. You may also want to INITiate on IMMediate to increase measurement speed. During Normal Mode, with both of these controls set to IMM, power measurements can be read with MEAS, READ, or FETCh. IMMediate triggering is not compatible with the BURSt Mode. If you send the CALC#:MODE BURS command to enter the BURSt Mode, IMM instrument triggering for NORMal Mode will automatically be switched to TRIG:SOUR BUS. If you send the CALC#: MODE SWIF command to enter the SWIFt Mode, IMM instrument triggering will remain IMM but a device-specific error will be generated whenever you specify a TRIG:COUN# higher than 1.
3-18 Manual 31470, Rev. E, April 2001
Page 93
Remote Operation
*TRG
Syntax:*TRG
Example: OUTPUT @Pwr_mtr;*TRG ! BUS Trigger Event
Description: *TRG is an IEEE 488.2 compatible programming command to initiate BUS triggered
measurements. The power meter interprets *TRG as a BUS source for instrument trigger­ing events. This command will not trigger Peak Power Sensors. It is the same as the TRIG command.
TRIG
Syntax: TRIGger[:IMMediate]
Example: OUTPUT @Pwr_mtr;TRIG ! Trigger measurement cycle
Description: TRIG is the SCPI command form of *TRG. The power meter interprets TRIG as a BUS
source for instrument triggering events. If INIT:CONT is OFF, measurements using FETCh#? will not proceed until triggering is armed and instrument triggering is actuated.
TRIG:SOUR
Syntax: TRIGger:SOURcespace<instrument triggering sourc IMMediate, BUS, HOLD, or
EXTernal>
Example: OUTPUT @Pwr_mtr;TRIG:SOUR IMM ! Set instrument measurement trigger
! source to IMMediate
Description: TRIG:SOUR IMM sets trigger control to the 8650A. BUS triggering sets the triggering
control to the controller software using TRIG or *TRG commands. HOLD halts triggering sequences. EXTernal sets triggering control to the front panel BNC connector. The follow­ing data shows triggering and operating mode compatibility.
Operating Mode IMM BUS HOLD EXT
NORMal X X X,
SWIFt X X X
BURSt X X
If the power meter has been configured for EXT triggering in the BURSt or SWIFt modes and the operating mode is switched to NORMal, the triggering configuration remains EXT. However, if the configuration is IMM in NORMal mode and the meter is switched to BURSt, it automatically switches to BUS. As a general rule, when a measurement subrou­tine switches operating modes, you should send the triggering configuration.
Manual 31470, Rev. E, April 2001 3-19
Page 94
Series 8650A Universal Power Meters
TRIG:MODE
Syntax: TRIGger:MODEspace<burst mode data gathering POST trigger receipt or PRE
trigger receipt>
Example: OUTPUT @Pwr_mtr;TRIG:MODE POST ! Set Burst or Swift mode data collection
Description: This command is only for BURSt Mode measurements. Send this command only immedi-
ately after CALC#:MODE BURS. TRIG:MODE controls the operation of the FIFO data buffer inside the power meter. When set to POST, the burst of data points is taken after the receipt of a valid instrument trigger. When set to PRE, the burst of data is assumed to be those data points that have arrived immediately preceding the valid instrument trigger event. Using PRE requires some caution: The single processor in the 8650A is constantly performing data acquisition. If you send any configuration commands while it is collecting the PRE trigger data, you may interrupt the timing of the data point collection.
TRIG:DEL
Syntax: TRIGger:DELayspace<delay time between buffered readings in sec, 0.000 to 5.000>
Example 1: OUTPUT @Pwr_mtr;TRIG:DEL 5e-3 ! Set Burst mode measurement time to
! relative to Timing of event trigger post
! 5 ms between measurements.
Example 2: OUTPUT @Pwr_mtr;TRIG:DEL 0 ! Max. measurement rate (resolution,
! 1 msec)
Description: This command applies only to the BURSt Mode. It sets the time interval between the
measurements and is accurate to about 5%. This is the only way to control the data acquisition of individual data points in a burst measurement. Burst measurement data is taken following or just prior to a single instrument trigger event. The number of measure­ments taken is controlled by TRIG:COUNt. Timing can be set in 0.001 second increments from 0.000 to 5.000 seconds, inclusive. Setting this control to 0.000 activates the highest reading rate (5100 rdgs/sec) possible with the 8650A.
Triggering of the Peak Power Sensor occurs independently from instrument triggering. Therefore, there is a moderate ability to control triggering at the Peak Power Sensors. Generally, if you require control of individual data point triggering, the SWIFt mode will provide the highest measurement rates, with or without SRQ or hardware handshaking.
3-20 Manual 31470, Rev. E, April 2001
Page 95
Remote Operation
TRIG:COUN
Syntax: TRIGger:COUNtspace<number of data values to buffer in memory from 1 to 5000 for
the standard 8650; from 1 to 128,000 on one channel with option 02; and 1 to 64000 for two channels with option 02>
Example: OUTPUT @Pwr_mtr;TRIG:COUN 100 ! Set Burst or Swift mode buffer reading
! number to 100
Description: The power meter has an internal and expandable data buffer for storing measurement
data until the slot 0 controller/resource manager is ready to read the data. During the SWIFt or BURSt modes, this command controls the number of power readings that are stored in the 8650A data buffer.
The value specified here applies to each individual channel. That is, if two sensors are attached and calibrated, the 8650A will perform the specified measurement and buffering for both channels. In the example shown below for this command, that would be 100 points per channel. Option 02 adds additional memory for 128,000 data point buffering. When two sensors are attached and calibrated, the maximum is 64,000 readings per channel.
During BURSt mode, the COUNt# of readings will be taken with just one instrument trigger. During the SWIFt Mode however, you can control the triggering of individual data points. If you set triggering to EXT, the power meter will perform one measurement each time it detects a TTL level signal. The value is then stored in the data buffer. This contin­ues until the meter has received and processed a COUNt# of triggers and also read in a COUNt# of data points to the buffer.
*WAI
Example: OUTPUT @Pwr_mtr;*WAI ! Wait, following command completion
Description: The *WAI command causes the power meter to wait until all previous commands and
queries are completed before continuing operation. It functions only when using the Normal mode of operation.
Example programs for Instrument Triggering are contained in Appendix A.
Manual 31470, Rev. E, April 2001 3-21
Page 96
Series 8650A Universal Power Meters
3.2.13 Arming the Triggering Cycle
INITiate:CONTinuous INITiate[:IMMediate]
The Initiate commands enable the power meter to acquire measurement data at the next instrument trigger. In the absence of an instrument signal, the 8650A is placed in the waiting-for-trigger-state. The default configuration is continuous initiation, INIT:CONT ON.
INITiate[:IMMediate] causes the power meter to exit the idle state and causes the trigger system to initiate and complete one full trigger cycle, returning to the idle state upon completion. For example, INITiate[:IMMediate] can be used with Peak Power Sensors to measure the power level of transient or one-shot pulsed microwave signals. After execution of the triggering sequence, send the FETCh? query command to return the measurement data from the 8650A.
Perform triggering configuration with the TRIGger Subsystem commands. When TRIGger:SOURce is IMMediate, the measurement will start as soon as INITiate is sent to the 8650A and executed (or INITiate:CONTinuous ON sent and executed).
INIT:CONT
Syntax: INITiate:CONTinuousspace<ON or OFF>
Example: OUTPUT @Pwr_mtr;INIT:CONT ON ! Continuous Triggering
Description: This allows user control of measurement cycle arming. When INIT:CONT is set to ON,
measurement cycle arming occurs automatically. The default setting is OFF. When you need to take control of arming the triggering of the measurement cycle, set this control to OFF and use MEAS#? or INIT with READ#? to perform the measurement. Another legal measurement control option would be INIT with TRIG: IMM and FETC#?
When INITiate:CONTinuous is ON, the instrument triggering cycle resets continuously. That is, upon instrument triggering, data is constantly being acquired and updated. Upon completion of a trigger cycle, the meter will immediately enter the wait-for-trigger state rather than the idle state, as is the case with INITiate:CONT OFF.
INIT:IMM
Syntax: INITiate[:IMMediate]
Example: OUTPUT @Pwr_mtr;INIT ! Initiate instrument triggering cycle
Description: This command enables the triggering cycle. Instrument triggering will not occur unless
INIT has been enabled, even when you are sending BUS or EXT triggers to the 8650A. Either this command or INIT:CONT ON performs measurements with FETCh? and READ?. MEAS? is a high level function which contains the INITiation enabling function. Arming occurs automatically when using the MEAS#? command to automatically INITiate, TRIGger and FETCh the measurement data. However, if you are using READ#? or FETCh#? to read data, this command arms the 8650A for triggering upon the occurrence of the next valid trigger.
The INITiate[:IMMediate] command places the power meter in the wait-for-trigger-state. When used with EXT (TTL levels) instrument triggering, this command is useful for measuring signals that are asynchronous with programming (BUS triggered) sequences. When INITiate:CONTinuous is set to OFF, INITiate:[IMMediate] enables the instrument triggering cycle. Sending INITiate:[IMMediate] when INITiate:CONTinuous is set to ON will cause a -123 command error. The INITiate [:IMMediate] command is an event; no query form exists for this command.
3-22 Manual 31470, Rev. E, April 2001
Page 97
3.2.14 Channel Configuration
CALCulate:POWer CALCulate:RATio CALCulate:DIFFerence CALCulate? CALCulate:UNIT CALCulate:STATe CALCulate:STATe?
CALC#:POW
Syntax: CALCulate<channel 1 or 2>[:CHANnel]:POWerspace <sensor 1 or 2>
Example: OUTPUT @Pwr_mtr;CALC2:POW 1 ! Configures channel 2 to
Description: This command configures a sensor to an individual channel, and the channel measures
the sensor power level. This command does not have a query format. For the example above, the response to a CALC2? query is POW 1. This command should be executed only after the designated sensor has been calibrated.
Remote Operation
! measure Sensor 1 power
CALC#:RAT
Syntax: CALCulate<channel 1 or 2>[:CHANnel]:RATiospace <sensor 1 or 2commasensor 2 or 1>
Example: OUTPUT @Pwr_mtr;CALC1:RAT 2,1 ! Configures channel 1 to measure
! Sensor 2 over Sensor 1 power
Description: This command configures a channel as a ratio of the power levels present at the two
sensors. When UNITs are dBm, RATio configuration produces measurements in dB. When UNITS are Watts, RATio configuration measurements are percentage. This command does not have a query format. For the above example, the response to a CALC1? query is RAT 2,1. This command should be executed only after the designated sensor has been calibrated.
CALC#:DIFF
Syntax: CALCulate<channel 1 or 2>[:CHANnel]:DIFFerencespace <sensor 1 or
2commasensor 2 or 1>
Example: OUTPUT @Pwr_mtr;CALC1:DIFF 2,1 ! Configures channel 1 to measure
! Sensor 2 minus Sensor 1 power
Description: This command configures a channel as the difference between the power levels present
at the two sensors. UNITs are dBm or Watts. This command does not have a query format. For the above example, the response to a CALC1? query is DIFF 2,1. This command should be executed only after the designated sensor has been calibrated.
Manual 31470, Rev. E, April 2001 3-23
Page 98
Series 8650A Universal Power Meters
CALC#?
Syntax: CALCulate<channel 1 or 2>[:FUNCtion]?
Example: OUTPUT @Pwr_mtr;CALC2? ! Query channel 2 sensor Configuration
Response: POW #, RAT #,#, or DIFF #,#
Description: This command returns the current channel configuration. This command is a query format
only.
CALC#:UNIT
Syntax: CALCulate<channel 1 or 2>:UNIT[:POWer]space<data units selection, dBm or Watt>
Example 1: OUTPUT @Pwr_mtr;CALC1:UNIT W ! Selects channel 1 Linear Units, Watts
Example 2: OUTPUT @Pwr_mtr;CALC1:UNIT dBm ! Selects channel 1 Log Units, dBm
Description: This command configures a channel to report power measurements in either linear Watts
units or logarithmic dBm units. When UNIT is dBm, the RATio configuration produces measurements in dB, and POWer and DIFF measurements in dBm. When UNIT is Watts, the RATio configuration measurements are percentages, and POWer and DIFF measure­ments are in Watts.
CALC#:STAT
Syntax: CALCulate<Channel 1 or 2>:STATespace<ON or OFF>
Example: OUTPUT @Pwr_mtr;CALC1:STAT OFF ! Disable channel 1 measurement
Description: This command turns a channel off or on. Default is ON. During Normal and Swift modes,
measurement speed increases when only one sensor (Channel configuration is CALC#:POW) is performing measurements.
CALC#:STAT?
Syntax: CALCulate<channel 1 or 2>:STATe?
Example: OUTPUT @Pwr_mtr;CALC1:STAT? ! Query channel 1 measurement status
Response: 0 = off, 1 = on.
Description: This command is the query format of the command above.
Example programs for Channel Configuration are contained in Appendix A.
3-24 Manual 31470, Rev. E, April 2001
Page 99
3.2.15 Cal Factor Correction
SENS#:CORR:FREQ MEMory:FREQuency MEMory:SELect MEMory:SLOPe
Power Sensors have a measurable frequency response. During manufacture, this response is calibrated at 1 GHz intervals. Instead of printing the data on the sensor label, each Giga-tronics power sensor includes a built-in EEPROM which has been programmed with the frequency calibration factor data for that particular power sensor.
The following SCPI commands tell the 8650A the operating frequency of the measured microwave or RF signal. Thus, the meter automatically interpolates the correct Cal Factor and applies that value to the measurement data. By performing this automatically, you do not need to read Cal Factor data from the side of sensor housings and program the data into tables within your ATE programming. Except for advanced measurement techniques using Burst Mode, you are not required to write cal factor interpolation routines; this is all done by the power meter. As shown below, sensor Cal Factor data can be read into your computer. Also see the Sensor EEPROM Commands in Section 3.2.30 for Cal Factor programming.
SENS#:CORR:FREQ
Remote Operation
Syntax: SENSe<sensor 1 or 2>:CORRection:FREQuency[:CW-FIXed] 5e7
Example: OUTPUT @Pwr_mtr;SENS2:CORR:FREQ 5e7 ! Set sensor 2 frequency to 50 MHz
Description This command allows you to correct for sensor frequency response characteristics. The
power meter uses this frequency to automatically interpolate and apply the Cal Factor for that sensor (computed from Cal Factor data points stored within each sensor’s EEPROM). Unless the EEPROM has been modified subsequent to shipment, the Cal Factor is always
0.0 dB at 50 MHz, 5e7. Be sure to correct for sensor cal factor by using this command, using automated V
correcting for Cal Factor in your ATE system by downloading Cal Factor points for the 8650A.
F correction, sending your own cal factor REFerence value, or
PROP
MEM:SEL
Syntax: MEMory[:TABLe]:SELectspace<ANALOGout,VPROpf1,VPROpf2>
Example: OUTPUT @Pwr_mtr;MEM:SEL VPROpf1 ! Select memory table V
! for editing in following lines
Description: This command selects between one of three editable tables. Using the meters automated
F capability to correct for Cal Factor requires that you configure the V
V
PROP
connector to match the signal that your source will output. This is performed by setting the frequency so that your source’s V
entered in the form of a value representing the V/Hz relationship. If your source is a Giga-tronics GT 9000 Series Signal Generator, this relationship is 0.5 V/GHz or 5e-8 V/Hz.
F output is 0.0 V. Then the slope relationship is
PROP
PROP
PROP
F1
F IN
Manual 31470, Rev. E, April 2001 3-25
Page 100
Series 8650A Universal Power Meters
MEM:SLOP
Syntax: MEMory[:TABLe]:SLOPespace<Volts per Hz>
Example: OUTPUT @Pwr_mtr;MEM:SLOP 1e-9 ! Set V
Description: This command sets the V/Hz relationship for automated Cal Factor correction.
F slope to 1 V/GHz
PROP
MEM:FREQ
Syntax: MEMory[:TABLe]:FREQuencyspace<Start frequency>
Example: OUTPUT @Pwr_mtr;MEM:FREQ 1e9 ! Set V
Description: This command sets the automated V
which the voltage input into the 8650A is 0.0 V.
F correction start frequency. The frequency at
PROP
F frequency to 1 GHz.
PROP
SENS#:CORR:FREQ
Syntax: SENSe<sensor 1 or 2>:CORRection:VPROpf[:STATe]space<ON-OFF>
Example: OUTPUT @Pwr_mtr;SENS1:CORR:VPRO ON ! Enable sensor 1 V
Description: Setting this control to ON activates V
☛☛☛☛
NOTE:
Appendix A of this manual.
Example programs for Cal Factor Corrections are contained in
F and deactivates SENS#:CORR:FREQ.
PROP
PROP
F function
Cal Factor Error Control
See Error Messages in Section 3.2.32.
Selected basic SCPI syntax and execution errors apply to these commands.
Device-specific errors include the following and other -300 level errors.
A common device-specific error occurs when the frequency sent to the 8650A in the SENS#:CORR:FREQ ### command is outside the sensor operating frequency range. For example, sending SENS1:CORR:FREQ 18.4e9 when an 18 GHz (max) 80301A CW Power Sensor is attached will yield a device-specific error.
3-26 Manual 31470, Rev. E, April 2001
Loading...