Gigahertz-Optik P-9710 Series, P-9710-1, P-9710-2, P-9710-4 User Manual

E 141203
Users Guide
P - 9710 series Optometer
P - 9710 - 1, P - 9710 - 2, and P - 9710 - 4
Measurement and examination devices for optical radiation detectors
Gigahertz-Optik GmbH
An der Kaelberweide 12, D-82299 Tuerkenfeld. Telephone +49 8193 93700 - 0 Telefax +49 8193 93700 - 50 info@gigahertz-optik.de www.gigahertz-optik.de
Gigahertz-Optik Inc.
5 Perry Way
Newburyport MA 01950-4009 USA Telephone:+ 978 462 1818 Telefax: + 978 462 3677 Email: b.angelo@gigahertz-optik.com Website: www.gigahertz-optik.de
Page 1
Table of contents:
1. Additional information to the operating instructions
1.1 Safety precautions
1.2 Hazard clause
1.3 Liability
1.4 Safety
1.5 Environmental conditions
1.6 Maintenance and cleaning
1.7 Guarantee
1.8 CE Conformity declaration
1.9 Review clause
1.10 Contact details
2. Device description P - 9710 optometer
2.1 Intended use
2.2 Inspection
2.3 Technical specifications P - 9710 series
2.4 Additional technical specifications P - 9710 variants
2.4.1 Optometer P - 9710 - 1
2.4.2 Optometer P - 9710 - 2
2.4.3 Optometer P - 9710 - 4
2.5 Description of the measurement ranges
2.6 Device overview
2.7 Detector interface
2.8 RS232 Interface / Analog Output
2.9 Trigger input (device variant P - 9710 - 4 only)
3. Performing measurements
3.1 Version Info
3.2 Quick start guide
3.3 P - 9710 Menu Structure
3.3.1 General information
3.3.2 Menu Overview
3.3.3 Front panel buttons
3.3.4 Menu description “Mode”
3.3.5 Menu description “Range”
3.3.6 Menu description “Detector”
3.3.7 Menu description “Offset”
3.3.8 Menu description “Reference”
3.3.9 Menu description “Substitution”
3.3.10 Menu description “Setup”
3.3.11 Menu description “Info”
3.4 Remote Interface (RS232)
Page
4
4 4 5 5 5 6 6 6 6 7
8
8 8 9
10
10 10 10 11 12 13 14 14
15
15 15 15 15 16 20 20 23 23 23 23 24 24 26 27
Page 2
4. Accessories
5. Maintenance and services
5.1 Battery
5.2 Calibration
5.3 Service addresses
A. Appendix - Measurement Modes
A.1 I-Effective A.1.1 I-Effective (Schmidt - Clausen, SC) A.1.2 I-Effective (Blondel - Rey, BR) A.2 Relative Logarithmic A.3 Pulse Energy A.3.1 Principle A.3.2 Pulse Energy Measurement
28
29
29 29 30
31
31 31 32 33 33 33 34
Page 3
Additional information to the operating instructions
1
1.1 Safety precautions
Please read this users guide carefully before using the device! Trouble-free functioning of the device as well as the operational safety can only be guar­anteed when the safety precautions and the general safety measures de­scribed in this users guide are considered.
The device should only be used for its intended purpose and according to its technical specifications. Any other use of the device is not allowed.
Immediately after unpacking and before the device start-up, the device should be inspected for any mechanical damages or loose parts. In such cases, the device should not be used.
Safety can also be impaired if the device is not functioning properly.
If during operation it is suspected that the safety has been impaired, the device should be switched off and its use by third parties ruled out.
1.2 Hazard clause
The device described in this instruction manual is designed for optical radi­ation measurements. The device offers no protection against radiation risks affiliated with the measured objects.
Persons in charge of performing measurements and using the device are obligated to read and comply with the instructions given in this users guide before beginning any measurement.
Persons in charge of performing measurements and using the device must be well informed about accident prevention regulations in their field of work.
Handling the device requires intermittent concentration of the operator on the device. The device should therefore not be used in surroundings that might lead to interrupted operator attentiveness. This possible risk must also be included in the briefing of the device operator.
Through this users guide, Gigahertz-Optik GmbH as the device manufac­turer informs the device owner about the possible hazards while using the device. The owner and the responsible device users confirm the receipt and approval of this users guide. The owner and operator of the device have been informed about the diligence required when choosing and brief­ing the device users and operators.
1.3 Liability
The responsibility for all consequences that might arise from the use of the measurement device or its accessories lies solely on the user and/or own­er.
Page 4
Gigahertz-Optik GmbH or its suppliers can under no circumstances be lia­ble for any direct or indirect business interruptions, loss of profits or data loss. This also applies to damages that may arise during the device opera­tion.
The Gigahertz-Optik GmbH Terms and Conditions of Delivery and Pay­ment generally apply.
1.4 Safety
The measurement device is designed for operation in a clean environment and within operating temperatures ranging from +5 to +40 ºC.
The measurement device or its accessories are not to be exposed to direct sunlight or moisture.
The device should be switched off and secured against unintended start-up by third parties in case of indications suggesting unsafe operating condi­tions. Such indicators include:
when the device shows visible damages
when the device contains lose parts, either internal or external
when the device is not functional
Safe operation of the device can be impaired due to one of the following circumstances:
Outdoor storage or storage in moist environment over a long period of time or other adverse conditions.
Excessive strain during transport (e.g., due to improper packaging)
Operation in aggravated environmental conditions such as high temper­atures, humidity, etc.
Operation in dangerous conditions (explosive gas, fumes, dust, etc.)
1.5 Environmental conditions
Optical measurement devices are sensitive to influences resulting from certain environmental conditions, e.g., high temperatures, humidity, and contamination, which might lead to measurement errors!
The measurement device should not be taken from a cold environment and directly put to operation in a warm surrounding. Under certain circumstanc­es, condensation could arise and negatively impact on the electrical and optical functions. Allow the device to adjust to the temperature of the sur­roundings before switching it on.
The measurement device should not be operated within high magnetic, electromagnetic or electrostatic fields.
Page 5
1.6 Maintenance and cleaning
The intervals for the recurrent recalibrations should be observed!
Opening of the device as well as device repair should only be done by au­thorized persons.
Off-the-shelf, residue-free, and non-abrasive detergents can be used for the exterior cleaning of the device.
While cleaning it should be made sure that the detergent doesn’t penetrate the interior of the device. This can lead to short circuiting or device opera­tion errors.
Do not use aggressive cleaning agents.
1.7 Guarantee
Gigahertz-Optik GmbH warrants that the product is free from material and production defects for a period of 12 months beginning from the date of delivery. In case defects are experienced within the warranty period, Giga­hertz-Optik GmbH repairs or replaces the device free of charge.
In order for the warranty to be effective, the customer should submit the product name, serial number, date of purchase as well as the description of the defect in writing to Gigahertz-Optik GmbH after which he/she will re­ceive a service number (RMA) enabling him/her to send back the device.
The user is responsible for the appropriate packaging of the device.
Shipping through the user must be done on a free to buyer’s address basis.
The guarantee does not cover consequential damages.
The Gigahertz-Optik GmbH Terms and Conditions of Delivery and Pay­ment generally apply.
1.8 CE Conformity
Gigahertz-Optik GmbH declares that the measurement device, with re­gards to its design and conception, conforms to the EG-guidelines and corresponding harmonized norms.
The CE Declaration of Conformity can be requested separately.
If any alteration is made on this machine without prior agreement this declaration loses its validity.
1.9 Review clause
Gigahertz-Optik GmbH reserves the right to changes in this users guide without prior notice.
Page 6
1.10 Contact details
Headquarters
Gigahertz-Optik GmbH Kaelberweide 12 D-82299 Tuerkenfeld
Telephone: +49 (0) 8193 93700 - 0 Telefax: +49 (0) 8193 93700 - 50 Email: info@gigahertz-optik.de Website: http://www.gigahertz-optik.de
North and South America
Gigahertz-Optik Inc.
5 Perry Way
Newburyport MA 01950-4009 USA
Telephone: + 978 462 1818 Telefax: + 978 462 3677 Email: b.angelo@gigahertz-optik.com Website: http://www.gigahertz-optik.com
Page 7
2 Device description P - 9710 optometer
2.1 Intended use
The P - 9710 optometers are measurement and examination devices for optical radiation detectors.
The device can be controlled manually via front panel buttons or via a RS232 remote interface.
Measurement results are displayed on the integrated display (LCD) or transmitted via the RS232 remote interface.
Different optical quantities or wavelength bands can be measured by using different detector heads in combination with the P - 9710 devices.
The calibration of the optical quantities is stored in the connector plug of the detector head. This allows using different detector heads with the P ­9710 device or using the detector heads with different P - 9710 devices without the need to change or transfer calibration factors manually.
The P - 9710 devices have built-in batteries that are charged with the sup­plied plug-in power supply. The capacity of the battery allows
5 hours of
operation with display illumination on and 12 hours with illumination off.
2.2 Inspection
Upon receiving your device, please check the package and all its parts for any possible damages that may have occurred during transportation.
Any transport damages should immediately be reported to the shipper.
After unpacking, please check to confirm there are no transport damages or any irregularities with regards to package completeness and correct­ness. The scope of delivery contains:
one measurement device P - 9710
one power supply
a factory calibration certificate
this users guide
one hard-top casing (optional)
one serial interface cable (optional)
different measurement heads and their calibration certificate (optional)
Page 8
2 Device description P - 9710 optometer
2.3 Technical specifications P - 9710 series Power supply
Display
Detector interface Analog output
Units of measurement
Signal amplifier
CW integration time Pulse integration time Offset correction
built-in lead acid battery, 6V / 0.5Ah (Version before
2017) Built-in lithium Ion battery, 3.7V /1400mAh (Version
2017) operating time: 5hr / 12hr with backlight on /off plug 5.5mm/2.5mm/9.5mm for external power supply (7.5VDC - 8.0VDC / min. 250mA)
alphanumeric LCD, 2 rows x 16 chars character height 5mm LED backlight (on / off)
9pin DSUB (female) output voltage proportional to detector input current
connector: TRIAD01 5pin (Tyco)
Amperes (absolute calibration) optical units corresponding to calibration data ratio in percent, logarithmic, and as factor
Transimpedance amplifier 8 ranges (200pA to 2mA) fixed range or autorange
100µs - 5.9999s 10ms - 199.99s correction range transcending
Parameter adjustment
Optical calibration data
Data logger
Front panel control Remote interface
front panel keys or remote control adjusted values permanently stored in EEPROM 10 different overall settings can be stored
stored in detector head connector (EEPROM) max. 250 calibration table entries interpolation between wavelength points possible manual input of a calibration factor possible
max. 12288 entries, permanently stored in flash memory
10 buttons, menu system RS232 (9600 Baud, 8 Data Bits, 1 Stop Bit, No
Parity) Connector: TRIAD01 (Tyco)
Page 9
2 Device description P - 9710 optometer
Temperature range
Size / weight
RoHS
2.4 Additional technical specifications P - 9710 variants
2.4.1 Optometer P - 9710 - 1 Amplifier slew rate
2.4.2 Optometer P - 9710 - 2 Amplifier slew rate
2.4.3 Optometer P - 9710 - 4 Amplifier slew rate Trigger input
operating: 5°C to 40°C storage: -10°C to 50° C
195mm x 100mm x 40mm / 500g
compatible
2ms - 10ms
20ms
20ms starts pulse energy measurement
BNC (female) low active internal pull-up (22 kOhm to +5V)
Page 10
2 Device description P - 9710 optometer
2.5 Description of the measurement ranges
Page 11
2 Device description P - 9710 optometer
2.6 Device overview
P - 9710
Display
Front panel with 10 buttons
Page 12
RS232 interface / analog output
plug for power supply
detector interface
2 Device description P - 9710 optometer
trigger input
(device variant P - 9710 - 4 only)
2.7 Detector interface
DSUB 9 pin
(female)
1 input current 2 GND 3 connection detection (GND by external detector) 4 GND 5 SCL (I2C, EEPROM) 6 GND 7 GND 8 +5V Output (Ri = 100 Ohm) 9 SDA (I2C, EEPROM)
function
Page 13
1
2
3 4 5
2 Device description P - 9710 optometer
2.8 RS232 Interface / Analog Output
Pin (TRIAD01
function
female)
1 Analog out (Ri = 10 kOhm) 2 TXD (has to be connected to RXD of PC) 3 RXD (has to be connected to TXD of PC) 4 GND 5 GND
View of the female connecter at P-9710 (Type TRIAD01 5pin /Tyco)
2.9 Trigger input (device variant P - 9710 - 4 only)
BNC (female) function
inner conductor trigger input, low active, internal pull-up (22 kOhm to +5V) outer conductor GND
Page 14
3 Performing measurements
3.1 Version Info
The following information is related to P - 9710 Firmware Version V5.1.
3.2 Quick start guide
connect the plug-in power supply to the P - 9710 or use the internal battery power
connect the detector head to the P - 9710
press <on/light> button to switch on the P - 9710
first the P - 9710 displays the Firmware Version: “P-9710 Vx.x”.
next the P - 9710 displays "Detector connected" and the detector’s serial number
the meter/detector should now be set-up to default settings, i.e., without further programming measurements can be performed
if necessary the P - 9710 measurement parameters can be reset to the factory default values:
- press the <menu> button
- press <▲> up or <▼> down buttons until "Setup" is displayed, press <enter> button to select
- press <▲> or <▼> buttons choose until "Configuration" is displayed, press <enter> button to select
- press <▲> or <▼> buttons until "Default Init" is displayed, press <enter> button to select
- P - 9710 asks for confirmation: "Change Data?", press <enter> button to confirm or < esc / reset > to abort
now specific measurement parameters can be selected to perform a certain measurement task (see "4.3 P - 9710 Menu Structure") as required
3.3 P - 9710 Menu Structure
3.3.1 General information
The P - 9710 can be manually operated and set-up for measurements using the front panel keypad (10 buttons). Pressing the <menu> button selects the main menu. There are further sub­menus. The up <▲> and down <▼> arrow buttons are used to switch be­tween the different menus. By pressing the <enter> button the menu item dis­played is selected. By pressing the <esc / reset> button one can escape from the menu without changing parameters. Upon selecting a setting, the P - 9710 requests confirmation (“Change Data?”) to change the parameter. Press the <enter> button to confirm that you really want to change the setting.
Page 15
3 Performing measurements
Use the <►> right and <◄> left arrow buttons to select the digit position when changing numbers. The selected digit can be changed by using the <▲> and <▼> arrow buttons. This also applies to the sign. Pressing the <enter> button selects the adjusted number.
3.3.2 Menu Overview
The following table is a quick reference guide to the different menu items / functions available in the P - 9710.
menu item submenu item function
1. Mode CW displays the measurement respective of any offset and calibration factors pro­grammed
CW Maximum displays the highest detected reading
CW Minimum displays the lowest detected reading
Peak Maximum shows the maximum detected peak level
Peak Minimum displays the minimum detected peak level
Peak to Peak displays the difference between the de-
tected peak maximum and peak minimum level
I displays the measurment result regarding
the distance between light source and detector as luminious intensity / radiant intensity
I-Effective (SC) enables the measurement of pulsed light
signals with evaluation of the effective intensity according to the form-factor method defined by Schmidt-Clausen
I-Effective (BR) enables the measurement of pulsed light
signals with evaluation of the effective intensity according to the form-factor method defined by Blondel-Rey
Page 16
3 Performing measurements
menu item submenu item function
Pulse Energy enables the measurement of short and
single light pulses with direct display of exposure for the pulse duration, consider­ing the respectively measured (radiometric) quantity
Relative (%) displays the measurement as percentage
of a reference value
Relative (Log) displays the measurement in dB or dBm in
relation to a reference value
Relative (Fact.) displays the measurement referred to a
reference value
Hold enables or disables the display hold func-
tion for the instrument
CW Level
Check
Dose (C) accumulates the single readings and dis-
Logger (x) stores the single readings, which are taken
Remote RS232 enables the instrument to be controlled by
2. Range selects a fixed measurement range (0 to
3. Detector Detector Infor-
mation
Ampere shows the readings in ampere (current
Manual a custom calibration can be entered
compares actual CW reading with previ­ously defined limit values and indicates status
plays the result as exposure for measured quantity
in predefined intervals up to a number of 12288
a computer via the built in RS232 interface
7) or autorange mode (Auto)
if a detector with calibration data connect­or is attached to the instrument, optionally programmed calibration factors can be selected
measurement)
Page 17
3 Performing measurements
menu item submenu item function
4. Offset
5. Reference sets a reference value, used in different
6. Substitution performs the measurements to get the
7. Setup Integration time sets the integration time
Dose Run Time sets the maximum time duration for expo-
Dose Maximum sets the max. dose level for the exposure
Dose CW Dis-
play
Logger Time sets the sample time interval for the data
Logger Synch. Synchronization of measurement acquire
Display Digits adjusts the number of displayed digits (4,
Zero Adjust
Mode
Pulse Meas.-
Time
IF Time Con-
stant
Pulse Length sets the pulse length in the Blondel-Rey
performs an automatic offset adjustment
display modes
substitution error compensation factor
sure measurement
measurement mode sets dose measurement display mode
logger mode
clock to logger clock
5, or automatic) sets the mode for the internal zero adjust-
ment needs to be set for the measuring modes I
-Effective (SC), I-Effective (BR) and Pulse Energy. It defines the time for collecting
readings and shall be a little longer than the expected pulse width.
sets the time constant C in the Schmidt­Clausen formula
formula
Page 18
3 Performing measurements
menu item submenu item function
Meas.-Distance sets the measurment distance for the cal-
culation of Luminious Intensity / Radiant Intensity ( modes I, I-Effective (SC), I-Effective (BR) ). If set to 0, the distance is not regarded for result calculation.
Pulse Offset determines the method of offset compen-
sation (stray light) for the measuring modes I-Effective (SC), I-Effective (BR) and Pulse Energy
CW Level Min. sets the lower limit value required in the
CW Level Check measuring mode
CW Level Max. sets the upper limit value required in the
CW Level Check measuring mode
Synchronisation selects how the measurement time period
is adapted to the input signal
Substitution enables / disables substitution error com-
Trigger Input enables / disables trigger input for pulse
measurement (only device variant -4)
Code Number sets a four digit lock-out access code Configuration Default Init – sets factory default settings.
Save Config – stores all current settings. Load Config. – loads setting selection (0-
9) for Save Config.
8. Info Battery Status displays the battery charge status as per­centage value
Logger data
displays the recorded logger data
Page 19
3 Performing measurements
3.3.3 Front panel buttons
P - 9710 Switch On
Pressing the <on/light> key turns the P - 9710 on.
P - 9710 Switch Off
Pressing the <off> key turns off the P - 9710.
Display Illumination On / Off
Pressing the <on/light> key (if the P - 9710 is already turned on), switches the display illumination on / off.
Measurement Result Display Update On / Off
Pressing the <run/stop> key stops updating the measurement display (the last display will be frozen). The display‘s upper row will display "Stp" on the right side. Pressing the <run/stop> again continues updating the display.
3.3.4 Menu description “Mode”
„CW“ displays the average measurement value for the pre-set integration time (upper row). The lower row displays the selected wavelength or name of the connected detector.
„CW Maximum“ displays the max. CW value input signal. Reset with <reset/ esc> key.
„CW Minimum“ displays the min. CW value input signal. Reset with <reset/ esc> key.
CW Maximum and CW Minimum should not be confused with Peak Values. Peak Values will be calculated for each integration period (integration time) out of the single ADC samples. CW Maximum and CW Minimum stores the largest or smallest CW value since the last <reset> key press.
„Peak Maximum“ displays the max. ADC sample value of the integration peri­od. It will be calculated new for each integration period (see menu "setup" / "integration time").
„Peak Minimum“ displays the min. ADC sample value of the integration peri­od. It will be calculated new for each integration period (see menu "setup" / "integration time").
„Peak to Peak“ displays the difference of max. and the min. ADC sample val­ue of the integration period. It will be calculated new for each integration period
Page 20
3 Performing measurements
(see menu "setup" / "integration time").
The P - 9710 can only detect peak values of pulses that are longer than the settling time of the input current amplifier - see “2.5 Description of the meas-
urement ranges”
„I“ measures and calculates the result (luminious intensity, radiant intensity) regarding the distance between the light source and the detector. The meas­urement distance (for result calculation) can be adjusted with „Setup / Meas. Distance".
„I-Effective (SC)“ measures single pulses and calculates the result according to the "Schmidt-Clausen" formula. Pressing the <run/stop> key starts the measurement. The measurement time can be adjusted with "Setup \ Pulse Meas. Time". The time constant C (for result calculation) can be adjusted with
„Setup / IF Time Constant". The measurement distance (for result calculation)
can be adjusted with „Setup / Meas. Distance".
„I-Effective (BR)“ measures single pulses and calculates the result according to the "Blondel-Rey" formula. Pressing the <run/stop> key starts the measure­ment. The measurement time can be adjusted with "Setup \ Pulse Meas. Time". The pulse length (for result calculation) can be adjusted with „Setup / IF Pulse Length". The measurement distance (for result calculation) can be ad­justed with „Setup / Meas. Distance".
„Pulse Energy“ measures single pulses and calculates the total energy over the adjusted measurement time. Pressing the <run/stop> key starts the meas­urement. The measurement time can be adjusted with "Setup \ Pulse Meas. Time".
For modes "I-Effective (SC)" "I-Effective (BR)", and "Pulse Energy" the influ­ence of ambient light can be reduced with „Setup / Pulse Offset“. If "Static" is selected an offset value can be set in "Offset” menu. If "Continuous" is select­ed, the last CW measurement before the pulse measurement is started will be used as offset value. If "Auto Range" is set, the P - 9710 will display "OVL" (overload) or "UDL" (underload) if the range of the last pulse measurement was not optimal. On next measurement(s) the range is automatically switched one step in the suitable direction until the optimum range is locked-in.
„Relative (%)“ displays the measurement result in percent of the reference value, „Relative (Fact.)“ as factor to the reference value and „Relative (Log)“ as dB in relation to the reference value. To adjust the reference value
Page 21
3 Performing measurements
see menu "Reference". See Appendix section 16.3 for background infor­mation.
„Hold“ stores and permanently displays the current CW value at the lower display row after Pressing key „reset“.
„CW Level Check“ checks the measured CW value against pre-set min. and max. limits. The lower display line displays the status. The limit values can be adjusted with menu „Setup / CW Level Minimum“ and „Setup / CW Level Maxi­mum“. A relay card (P - 9710 - Z02 available from Gigahertz-Optik GmbH) can be connected to the RS232-Interface of the P - 9710. Relay 1 will be activated if the CW value is within the limits. If the CW value is beyond the min. value re­lay 2 will be activated. If the CW value is above the max. value, relay 3 will be activated.
„Dose“ performs an integration of the measurement results and displays the result as a dose value. Key <run/stop> starts or stops the dose measurement. With key <reset/esc> the dose value can be reset to zero. With menu „Setup / Dose Run Time“ the max. dose measurement time and with „Setup /Dose Maximum“ the max. dose value can be adjusted. If dose run time or dose max­imum is reached, dose measurement is automatically stopped. If the <◄> key is pressed, the pre-set max. dose measurement time (upper row) and the actual dose measurement time (lower row) will be displayed for 5 seconds. If the <►> key is pressed, the pre-set max. dose value (upper row) and the actual dose value (lower row) will be displayed for 5 seconds.
„Logger“ stores the measurement values at the pre-set clock rate (0.1s to 6000s, manual). Max. 12288 values can be stored. Pressing the <run / stop> key starts or stops the accumulation. The sampled data is stored in flash memory (no loss of data if device is switched off). Every ‘start’ produces a da­taset (DSxxx) with additional device parameters (detector, clock, ...) stored in logger memory. Logger data is erased by pressing <reset / esc> key.
„Remote RS232“ sets P - 9710 for control by RS232-Interface (P - 9710 - Z01 cable required, available from Gigahertz-Optik GmbH).
Page 22
3 Performing measurements
3.3.5 Menu description “Range”
Allows the selection of the measurement range. Fixed range adjustment (Range 0 - 7) or Auto (for automatically range switching depending on the in­put signal) can be selected.
If the input signal alters rapidly it could be difficult for the autorange function to adjust the correct range (P - 9710 will frequent display "OVL" or "UDL"). In this case it might be better to set an adequate range manually!
3.3.6 Menu description “Detector”
Allows the selection of the calibration data used to calculate the measurement result.
"Ampere“ displays the measurement result as current (Ampere units). Using the menu item "Manual“ a calibration factor and calibration unit can be adjust­ed manually by key entry. Other adjustments include selection of table entries from the calibration data table of the connected detector. A certain table entry can be selected using the <▲> and key <▼> arrow keys. Using the <►> and <◄> arrow keys interpolation between the calibration table entries can be made (1nm step).
3.3.7 Menu description “Offset”
"0“ sets the offset to zero.
"CW“ selects the actual measurement value as the new offset value. The off-
set value will be subtracted from the following measurement results. Offset is useful to compensate ambient light or dark currents of detectors.
3.3.8 Menu description “Reference”
Selects the reference value used to calculate the measurement results in mode "Relative %“, "Relative (Log) “ and "Relative (Fact)“.
With "One Unit“ the reference value is set to 1 (1A, 1W, 1cd, ... ). "CW“ se­lects the actual measurement value as the new reference value. With
"Manual“ a certain reference value can be set manually with the keys. "1.000mX" sets the reference value to 1/1000 of the selected unit (could be
used to display the result in dBmW).
Page 23
3 Performing measurements
3.3.9 Menu description “Substitution”
Performs the measurements to get the factor for substitution error compensa­tion. The factor is calculated from the ratio of a black standard measurement and the probe measurement.
3.3.10 Menu description “Setup”
"Integrationtime“ sets the measurement time (100µs - 5.9999s). For this peri-
od of time the input signal is sampled (every 100µs) and the average value of the samples is calculated and displayed.
If the input signal has big AC components, to obtain a stable measurement result it may be useful to set the integration time to an integer multiple of the AC time period of the input signal!
"Dose Run Time“ sets the max. time for dose measurement (1s to 1000h). After this time dose measurement is automatically stopped.
"Dose Maximum“ sets the max. dose value for dose measurement. If this dose value is reached, dose measurement is automatically stopped.
"Dose CW Display“ sets the display mode for dose measurement. If "CW Display" is selected, the upper display row shows the actual CW value
and the lower display row shows the dose value. If "No CW Display" is select­ed, the upper display row shows the dose value and the lower display row shows the dose unit.
"Logger Time“ sets the sample time for logger mode (0.1s to 5999.9s and 0). Every elapsed sample time a new actual CW value is stored in the logger memory. If sample time 0 is set, a new CW value is stored in logger memory by Pressing <enter> key.
"Logger Synch.“ enables to lock the measurement acquire clock to the logger clock. If active, the "Integration Time" has to be at least 40ms shorter then the "Logger Time".
“Zero Adjust Mode“ selects the mode for the internal electronic zero adjust­ment (not the external dark current!). If "Range Switch" is selected, every time the measurement range is changed an internal zero adjustment is made. "Each Measurement" performs an internal zero adjustment each measure­ment period (integration time). "Fixed Values" performs a one time zero ad-
Page 24
3 Performing measurements
justment and uses these values for the following measurements.
For fastest measurement range switching "Zero Adjust Mode" should be set to "Fixed Values". For readjusting the zero values, the menu item "Zero Adjust Mode" / "Fixed Values" should be selected again (if environment temperature has changed or after a longer time period).
"Display Digits“ adjusts the number of display digits ("4“, "5“ or "Auto“) for the measurement result display. "Auto" automatically accommodates the resolu­tion to the selected calibration factor.
"Pulse Meas.-Time“ adjusts the integration time (0.01s - 199.99s) for pulse measurement ("I-Effective" or "Pulse Energy"). After Pressing "run / stop" key to start the pulse measurement, an integration of measurement samples is made for this set period.
Pulse Meas. Time should be set long enough, to collect the whole pulse into the measurement time slot. On the other hand, the time should be set as short as possible to minimize offset errors produced by ambient light and detector dark current.
"IF Time Constant“ adjusts the time constant for pulse calculation according to the formula of "Schmidt-Clausen" (0.1s - 5.9999s).
"Pulse Length" adjusts the pulse length for pulse calculation (mode "I-
Effective (BR) according to the formula of "Blondel Rey" (0.001s - 9.999s).
"Meas. Distance" adjusts the measurement distance (0.001m - 0.999m) for the calculation of Luminious Intensity / Radiant Intensity (modes "I", "I-Effective
(SC)" and "I-Effective (BR)"). If set to 0, the distance is not regarded for result
calculation.
"Pulse Offset" selects the method for offset compensation for modes "I­Effective (SC)", "I-Effective (BR)" and "Pulse Energy". The influence of ambi­ent light and detector dark current can be reduced. If "Static" is selected the offset can be adjusted with menu "Offset". If "Continuous" is selected, the last CW measurement value before the pulse measurement is started will be used as offset value. The offset value is subtracted from the measurement result.
"CW Level Minimum“ adjusts the minimum CW level for "CW Level Check"
mode.
Page 25
3 Performing measurements
"CW Level Maximum“ adjusts the maximum CW level for "CW Level Check"
mode.
"Synchronisation" selects how the measurement time period is adapted to
the input signal. If "Not active" is selected, no special synchronisation to the input signal is made. If "Active" is selected, a synchronisation on the AC com- ponent of the input signal is activated that results in a more stable measure­ment display.
"Trigger Input" (only device variant P - 9710 - 4) activates / deactivates the
trigger input. If activated, pulse measurements are only started after the trigger input is set to low level.
"Substitution“ enables / disables substitution error compensation.
"Code Number“ allows a four-digit code number to be set. If a value other
then "0000" is entered, P - 9710 parameters can only be changed after ente­ring this code number. The P - 9710 only asks one time for the code number. For all further parameter setting inputs a new code number is not requested unless the P - 9710 is swit­ched off and switched on again.
"Configuration“ selects an overall device parameter set-up. "Default Init" sets all parameters to factory default values. "Save Config." stores the actual configuration (all adjusted device parameters) permanently at the configuration memory (max. 10 different configurations can be stored). "Load Config." ad­justs the device parameters to certain configuration (before stored with "Save Config."). Numbers 0 - 9 are used to select the configurations.
3.3.11 Menu description “Info”
„Battery Status“ displays the charge state of the P - 9710 battery (0 - 100%).
If the power supply is connected, the „Battery Status“ will always show 100%.
„Logger Data“ shows the logged data. Scroll data with <▲> and <▼> up and
down keys and to select a certain entry (using <►> and <◄> keys the step size is increased to 100). By pressing, <enter> key other parameters
(belonging to that measurement result) can be displayed (sample time, detec-
tor name, start, stop, serial number, calibration table entry number).
Page 26
3 Performing measurements
3.4 Remote Interface (RS232)
The P-9710 can be controlled and values read out via RS232 interface by se­lecting the menu "Mode" / "Remote RS232".
The description of the remote interface is available upon request.
Page 27
4 Accessories
P - 9710 - Z01
RS232 interface adapter cable
for connecting a P - 9710 to a PC or the P - 9710 - Z02 relay card
P - 9710 - Z02
relay card
for controlling signal lamps, etc., when using the level check functionality requires P - 9710 - Z01 for connectivity and an additional power supply
P - 9710 - Z04
plug construction set for the RS232 inter­face / analog output
P - 9710 - Z1/2
adapter cable for measurement heads with BNC connector (-1 detector type)
calibration data has to be stored in the P ­9710 - Z1/2 connector
S - P9710
Application Software for all P-9710 vara­iants: full device control, measurement routines, data display and analysis, report generation, and data export.
Page 28
5 Maintenance and service
5.1 Battery
The device has a built-in rechargeable battery. Make sure that the device is stored only with a fully charged battery (lead acid battery) or at about 30% level for the lithium battery version. After a storage period of three months or longer, the battery should be recharged to avoid a reduction in lifetime. The battery will charge when the plug-in power supply is connected to the de­vice. It is not necessary to switch the device on. An empty battery is complete­ly charged after about 14 hours of charging time.
If the battery needs replacement, please contact Gigahertz-Optik GmbH to obtain a replacement and replacement instructions in case of the lead acid battery. The lithium battery version needs to be repaired by Gigahertz-Optik GmbH.
Obey national environmental laws when disposing the battery or the device.
5.2 Calibration
The calibration of the measuring device is confirmed by the supplied calibra­tion certificate. Gigahertz-Optik GmbH advises its customers to adhere to an­nual calibration of the device. The Gigahertz-Optik calibration lab offers short­term calibration services on the publicized calibration weeks. Please refer to the Gigahertz-Optik GmbH website for more information.
Page 29
6 Maintenance and service
6.3 Service addresses
Headquarters
Gigahertz-Optik GmbH Kaelberweide 12 D-82299 Tuerkenfeld
Telephone: +49 (0) 8193 93700 - 0 Telefax: +49 (0) 8193 93700 - 50 Email: info@gigahertz-optik.de Website: http://www.gigahertz-optik.de
North and South America
Gigahertz-Optik Inc.
5 Perry Way
Newburyport MA 01950-4009 USA
Telephone: + 978 462 1818 Telefax: + 978 462 3677 Email: b.angelo@gigahertz-optik.com Website: http://www.gigahertz-optik.com
Page 30
2
rEI
A Appendix - Measurement Modes
A.1 I-Effective
In order to perceive a signal light or warning beacon, the human eye requires a threshold of illuminance or luminance, which is higher for shorter presentation times. The consequence of this is that for equal peak intensities, a narrow pulse signal light (light-flash) can appear less intense to an observer compared to a steady state signal light: the luminous range of rhythmic light is usually smaller compared to that of a fixed light. Therefore the significant quantity is not the peak intensity but the effective intensity.
The I-Effective measurement modes only apply to measurements of luminous quantities with a photometer head. Generally, the quantity to be evaluated is luminous intensity, which can be derived from an illuminance measurement. Normally, a photometer head is delivered calibrated for illuminance, in order to be universal. But on special order the photometer head can also be calibrated in luminous intensity for a fixed distance. Please refer to the calibration certifi­cate for details regarding the calibration of your specific detector head.
The relationship between both quantities, luminous intensity and illuminance, is described by the inverse square law as shown in the equation below, which only is valid for large measurement distances compared to the extension of the luminous area. As a rule of thumb, one can estimate r to be 10 times the larg­est dimension of the illuminating source area, except for narrow beam sources (half peak angle < 5°).
where
I luminous intensity E illuminance r distance from source to acceptance area of photometer head
A.1.1 I-Effective (Schmidt - Clausen, SC)
The effective (luminous) intensity of a signal light presented as a brief pulse can be calculated by the method of Schmidt-Clausen according to the follow­ing equation
Page 31
dttiCî
dtti
îi
eff
)(
)(
t
dtti
i
eff
2.0
)(
A Appendix - Measurement Modes
where
i
effective intensity
eff
î peak intensity i(t) intensity
C time constant, depending upon the level of adaptation recommended values are
C = 0.1 s for day-time observation C = 0.2 s for night-time observation
A.1.2 I-Effective (Blondel - Rey, BR)
A futher method to calculate the effective (luminious) intensity of a signal light presented as a brief pulse is the method of Blondel-Rey according to the fol­lowing equation:
where
i
effective intensity
eff
i(t) intensity t pulse length
Page 32
ref
x
x
X
10
lg10
T
dttxX )(
A Appendix - Measurement Modes
A.2 Relative Logarithmic
The reading is displayed in dB as the ratio of a defined reference. To define a value, enter the menu item Reference (one unit: reference value is set to 1, CW: reference value is set to actual reading, manual reference: manual entry of reference value, 1.000mW: reference value is set to 1mW, reading is provid­ed in dBm units). If the Relative (Log) measuring mode is entered, the db value according to the following equation is displayed together with the actual read­ing:
where
X displayed quantity in dB or dBm x actual reading of the instrument x
reference value
ref
A.3 Pulse Energy A.3.1 Principle
Pulse energy can be defined as the integral of the input signal over time (which
equates to the pulse area).
X exposure x(t) measured quantity as a function of time
T flash duration time
The P-9710-2 can measure pulse energy for pulse lengths down to 1µs, even though the internal Analog to Digital Converter (ADC) can sample the input signal only every 100µs. To perform this measurement task, the shape of the input signal is changed. A short high pulse is transformed into a long low pulse. But the key point is that the area equal that to the energy of the pulse is preserved.
Page 33
A Appendix - Measurement Modes
original pulse:
transferred pulse:
Short pulses are transformed to pulse length greater 20ms. Therefore, with 100µs ADC sample time, every pulse is sampled at a minimum of 200 times. With optimal offset compensation the total electronic pulse energy measure­ment error is within ±1%. This principle of operation is also valid for a chain of pulses (one pulse chain can be considered to be one pulse).
A.3.2 Pulse Energy Measurement
The pulse measurement must be started manually by pressing <run / stop> key. Next the pulse signal has to be triggered. The P-9710 samples the input signal for the defined time (“Pulse Measurement Time”) or until <run / stop> key is pressed again. The P-9710 calculates the sum (integral over the time) of the complete measurement window (“Pulse Measurement Time”).
input signal:
measurement window:
pressing “run / stop” key starts the measurement
It is important, that the pulse measurement is started (by pressing <run / stop> key) some time (0.5s – 1s) before the pulse signal occurs. The measurement window (“Pulse Measurement Time”) should also be set long enough to catch the whole pulse signal. For device variant -4 it is possible to trigger pulse measurement by an external signal at trigger input (low active). First key "run / stop" has to be pressed. Then the device waits for the trigger signal. After the trigger signal is applied, the measurement is performed.
Page 34
A Appendix - Measurement Modes
Offset Compensation:
The P-9710 not only measures and calculates the sum of the pulse signal, but also electronic offset / ambient light or detector dark current are included in the calculated sum. To minimize the error produced by these offset signals, the measurement window should be set as short as possible (but long enough to catch the whole pulse, practical value: pulse time + 2 seconds).
Any offset error produced can also be minimized by measuring the offset sig­nal beforehand without the applied pulse signal. This offset is subtracted later from the measurement result.
The offset must be measured without applied pulse signal!
There are two methods to measure the offset signal:
If offset compensation “Setup / Pulse Offset” is set to “Static”, the offset can be measured under menu “Offset”. With menu “Offset” the offset value can be set to the actual CW value (no pulse signal has to be ap­plied) or set to zero.
If offset compensation “Setup / Pulse Offset” is set to “Continuous” the P-9710 uses the last measured CW value before pulse measurement is started as offset value. This mode has the advantage, that small chang­es of the offset value are compensated more effectively. For continuous offset compensation it is important, that the pulses are applied only after the pulse measurement is started (not during the preceding CW / offset measurement). This is the reason why for infinite pulse chains “Setup / Pulse Offset” has to be set to “Static”.
The offset signal level can be determined by starting pulse measurement without the presence of the pulses (the measurement range has to be set to fixed range). The displayed measurement value represents the error contri­bution produced by offset signal.
Measurement Range:
The P-9710 measurement range has to be matched to that of the input signal. This can be done manually be selecting a certain range (manual range selec­tion) or by selecting the autorange function (menu “Range” set to “Auto”). If the autorange function is set, the P-9710 will display "OVL" (overload) or "UDL" (underload) if the range of the last pulse measurement was not optimal
Page 35
A Appendix - Measurement Modes
(not well matched to the input signal range). On next measurement(s) the range is automatically switched one range step (up or down), as appropriate, until the optimum range is locked-in. For manual range selection, the optimum range can be found by first selecting the autorange function and following the procedure above. Once this range is identified it can be set as the fixed range.
Infinite Pulse Chains:
For infinite pulse chains “Setup / Pulse Offset” has to be set to “Static”. The offset has to be measured explicitly with pulse signal removed (menu “Offset”). “Setup” / “Puls-Meas.-Time” should be set to a multiple of the pulse period (= 1/frequency). If the pulse period is not known, “Puls-Meas.-Time” should be set as long as possible (more then 100 pulses should be covered by the measure­ment window) to get a stable measurement result.
Example - Pulse Energy Measurement of a Single 10µs Long Pulse
Set P-9710 Parameters:
set menu “Mode” to “Pulse Energy” set menu “Range” to “Auto” set menu “Setup” / “Synchronisation” to “Not active” set menu “Setup” / “Integration Time” to “0.3000s” set menu “Setup” / “Zero Measurement” to “Fixed Values” set menu “Setup” / “Pulse Meas.-Time” to “2.000s” set menu “Setup” / “Pulse Offset” to “Continuous”
Initiate Measurement:
press <run / stop> key to start measurement trigger pulse source (immediately after pressing <run / stop> key read measurement result on P-9710 display if “UDL” or “OVL” is displayed repeat measurement until a valid measure-
ment result is displayed
Example - Pulse Energy Measurement of a Single 10µs Long Pulse with external Trigger Signal (only device variant -4)
Set P-9710 Parameters:
set menu “Mode” to “Pulse Energy” set menu “Range” to “Auto”
Page 36
A Appendix - Measurement Modes
set menu "Setup" / "Trigger Input" to "active" set menu “Setup” / “Synchronisation” to “Not active” set menu “Setup” / “Integration Time” to “0.3000s” set menu “Setup” / “Zero Measurement” to “Fixed Values” set menu “Setup” / “Pulse Meas.-Time” to “0.200s” set menu “Setup” / “Pulse Offset” to “Continuous”
Initiate Measurement:
press <run / stop> key to activate measurement set trigger input to low level to start measurement trigger pulse source (immediately after or simultaneously with setting trig-
ger input to low level) read measurement result on P-9710 display if “UDL” or “OVL” is displayed repeat measurement until a valid measure-
ment result is displayed
Example - Pulse Energy Measurement of an Infinite Pulse Chain (pulse length: 10µs, pulse frequency: 50Hz)
Set P-9710 Parameters:
set menu “Mode” to “Pulse Energy” set menu “Range” to “Auto” set menu “Setup” / “Synchronisation” to “Not active” set menu “Setup” / “Integration Time” to “0.3000s” set menu “Setup” / “Zero Measurement” to “Fixed Values” set menu “Setup” / “Pulse Meas.-Time” to “5.000s” set menu “Setup” / “Pulse Offset” to “Static” remove pulse signal from detector select menu “Offset”, set “Offset = CW” adjust pulse signal to detector
Initiate Measurement:
press <run / stop> key to start measurement read measurement result on P-9710 display if “UDL” or “OVL” is displayed repeat measurement until a valid measure-
ment result is displayed
Page 37
Loading...