-8-
For measurement accuracy it is important to set the calibration point for the second alarm threshold
accurately.
When using dry test gas a slower response time of the sensor has to be taken into account.
Adhere to the following procedure to check the sensitivity:
Suppress alarm transmission.
Screw calibration adapter onto transmitter.
Supply test gas to CI21 without pressure and
with a flow rate of approx. 0.5 l/min.
Within about 2-5 minutes a voltage corresponding to the gas concentration must
be measured at the test contacts.
Alternatively the calibration can be checked at the display of the GMA controller.
Is the measured value equivalent to the calibration curve?
Re-adjust calibration by turning potentiometer C.
Alternatively: Re-adjust calibration point at GMA
controller. Please refer to the operation manual of
your GMA.
Disconnect test gas and calibration adapter.
Disconnect alarm suppression.
Service
According to the German "Regulation for Explosion Protection", "UVV-Gase" and DIN 31051, service
stands for maintenance, inspection and repair of gas warning equipment. Regulation T023 of "BG
Chemie" describes the relating measures, among others requesting a regular function check.
This function check means:
Check with alarm gas concentration
Check of zeropoint and sensitivity (calibration)
Check of response time
Check of gas sampling and gas processing system (if any)
Check of alarm signal activation
Check of failure alarm
The check must be done by an expert, who has to report the result in written form. Generally the
function check intervals should not exceed 16 weeks. The function check shall be done before putting the
system into operation, and it should be repeated at least once a year.
no yes