+GF+ 9950 User guide

Page 1
GF 9950 Six-Channel Transmitter
*3-9950-10.090*
3-9950-10.090 Rev. 1 01/23
Description
The 9950 is a multi-channel, multi-sensor transmitter designed to meet and exceed the industry standards, and expectations for a small, compact, ¼” DIN transmitter. The power and versatility of the 9950 allows the use of up to six GF sensors to manage complex water treatment applications.
The 9950 analyzer supports all like sensors or a mix of any GF sensors. Sensor types and accessories supported by the 9950 are GF Flow (frequency and/or digital S³L), pH/ORP, Conductivity/Resistivity, Salinity, Temperature, Pressure, Level, Dissolved Oxygen and any device that transmits a 4 to 20 mA signal when used with the single channel 3-8058 iGo® Signal Converter.
The 9950 base unit comes complete with two each 4 to 20 mA output, two additional dual 4 to 20 mA output modules can be installed to increase the number of 4 to 20 mA outputs to a total of six outputs.
3-9950-10, -11
Four Conductivity sensor measurements are supported with either a single or dual channel conductivity module. If six conductivity sensors are required, the use of a 3-2850-51-XX can be added to the main S³L input terminals.
English
Operating Instructions (Refer to pg. 11)
English
C
US
E171559
The 3-9950.393-3 Relay Module provides four binary inputs that are compatible with any open collector or mechanical contacts, such as level switches, fl ow switches, pressure switches or other devices. The 9950 offers advanced features such as derived functions, advanced multiple relay modes (Boolean logic) and timer-based relay functions.
The 9950 Modbus Module allows for remote access to primary and secondary measurements, derived functions, status of current loop outputs and relays, over a serial RS485 Modbus automation network.
The 9950 supports the following modules:
Four Channel Mechanical Relay Module
Two Mechanical and Two Solid State Relay Module
Two Mechanical Relays and Four Binary Inputs Module
Single Channel Conductivity Resistivity Module
Dual Channel Conductivity Module
Dual Channel 4 to 20 mA Current Loop Output Module
• Modbus Module
Compatibility
The 9950 is compatible with all GF products listed in the column to the right.
• pH/ORP electrodes require the GF 2751
DryLoc® Sensor Electronics (sold separately).
• Conductivity/Resistivity measurement
requires the GF 2850 Conductivity/Resistivity electronics or a single or dual conductivity module and proper conductivity sensor (sold separately).
Sensor
Model
515 X
525 X 2000 X 2100 X 2250 X 2350 X 2450 X 2507 X 2536 X
2537-5 X
2540 X 2551 XX 2552 XX 258X XX
U1000 XX U3000 XX U4000 XX
2260 X 2270 X 2290 X 2291 X
2610-51 X
2751 X
2850-51-XX* X
2850-61* X 2850-63* X
Freq
Output
Digital (S3L)
Output
Requires
8058-1
* No conductivity module required
Page 2
Start-Up Guide
Safety Information
• Please read entire manual before unpacking, setting up or operating this equipment. Adhere to all danger, warning and caution statements. Failure to do so could result in serious injury to the operator or damage the equipment. Make sure the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than specifi ed in this manual.
• This unit is designed to be connected to equipment which can be hazardous to persons and property if used incorrectly.
• Read and understand all associated equipment manuals and
Warning / Caution / Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death
Electrostatic Discharge (ESD) / Electrocution Danger
Alerts user to risk of potential damage to product by ESD
Electrocution Danger
Alerts user to risk of potential of injury or death via electrocution
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of GF products
safety warnings before using with this product.
• Remove power to unit before wiring connections.
• Wiring connections to this product should only be performed by
NOTE / Technical Notes
Highlights additional information or detailed procedure
qualifi ed personnel.
• Do not use unit if front panel is cracked or broken.
Before commissioning the device, make sure the supply voltage matches the voltage specifi cations on the nameplate. For DC power input and Loop current regulated voltage, use UL60950-1 or UL61010-1 certifi ed power supply. Power supply shall also be rated for operation at 4000 m altitude. Provide a suitable switch or circuit breaker the installation. This switch must be located close to the device (easily reached), and marked as a circuit breaker. The switch or circuit breaker to be used for power disconnect shall be certifi ed to IEC 60947-1 and IEC 60947-3, per IEC 61010-1, Clause 6.11.4.2. Overcurrent protection (rated > 10 A) is required for the power cable.
Installation
For future reference, for each installation, it is recommended to record the part number and serial number of each of the components listed here:
Facility Tag Number or System ID (user assigned):__________________
_ Base unit 3-9950-____ S/N ___________________
Relay Module 3-9950.393-____ S/N ___________________
Single Channel Cond./Resist. Module 3-9950.394-____ S/N ___________________ Dual Channel 4-20 Current Loop Out. Module 3-9950.398-____ S/N ___________________ Dual Channel Conductivity Module 3-9950.394-2 S/N ___________________ Modbus Module 3-9950.395-M S/N ___________________
Panel Mount Installation
Panel Cutout
92 x 92 mm
(+ 0.8, - 0 mm)
3.6 x 3.6 in.
(+0.031, -0 in.)
Panel Cutout
92 x 92 mm
(+ 0.8, - 0 mm)
3.6 x 3.6 in.
(+0.031, -0 in.)
Panel Cutout
92 x 92 mm
(+ 0.8, - 0 mm)
3.6 x 3.6 in.
(+0.031, -0 in.)
minimum
clearance
25 mm
(1 in.)
Panel Cutout
92 x 92 mm
(+ 0.8, - 0 mm)
3.6 x 3.6 in.
(+0.031, -0 in.)
panel
gasket on front side of panel
MODULE 1
DC POWER
USB
RELAY MODULE
PWR+
PWR–
Loop1+
DC Power
12-32 VDC
Loop1
0.50A Max
Loop2+
Loop Voltage
12-32 VDC
Loop2
0.03A Max
V+
FREQ
DATA
CH1
GND
V+
CH2
FREQ
DATA
N
GND
L
50-60 Hz, 24VA Max
100-240 VAC~
Made in U.S. from U.S. and imported parts
MODULE 2
quick-clip
mounting
bracket
Allowable panel thickness 2.36 mm (0.093 in.) to 33 mm (1.31 in.)
2
GF 9950-10 Six Channel Transmitter
Page 3
Start-Up Guide
Terminal Identifi cation
Prepare the transmitter installation location. If the back of the transmitter is diffi cult to access when installed, wire the removable terminal blocks fi rst, then install it completely.
DC Power Terminals
RELAY MODULE
PWR+
PWR–
Loop 1+
Loop 1–
Loop 2+
Loop 2–
FREQ
DATA
GND
CH2 CH1
FREQ
DATA
GND
11-24 VDC
+V
+V
USB
N
L
100-240 VAC~
Made in U.S. from U.S. and imported parts
MODULE 1
MODULE 2
Sensor Input 2 (Frequency/S
AC Power Terminals
Sensor Input 1 (Frequency/S
3
L)
3
L)
The 9950 requires regulated
12 to 32 VDC, ±10% regulated (24 VDC nominal) or 100­240 VAC at 50-60 Hz from an
external power supply (not supplied).
Maximum current draw is:
500 mA with DC power 24 VA with AC power
DC Power Terminal 3-9950-10 or 3-9950-11
Required by the instrument
• 12 to 32 VDC, ±10% regulated, 0.5 A MAX. DC power input and Loop current regulated voltage require the use of a UL60950-1 or UL61010-1 certifi ed power supply. Power supply shall also be rated for operation at 4000 m altitude.
Current Loop Outputs 1 and 2
AC Power Terminal 3-9950-11 Only
Required by the instrument
• 100-240 VAC at 50-60 Hz, 24 VA MAX
Relay Module
Dependent on model
• 3-9950.393-1 Four mechanical relays rated at 5A, 250 VAC or 5A, 30 VDC
• 3-9950.393-2 Two mechanical relays, 5A 250 VAC or 30 VDC, and two solid state relays rated at 50 mA, 30 VAC or 30 VDC
• 3-9950.393-3 Two mechanical relays, 5A 250 VAC or 30 VDC, and four binary inputs rated at 6 mA, 10 VDC to 24 VDC
Modules 1 and 2
• 3-9950.394-1 Single Channel Direct Conductivity/Resistivity Module
• 3-9950.394-2 Dual channel Conductivity Module
• 3-9950.395-M Modbus Module
• 3-9950.398-2 Dual Channel 4 to 20 mA Current Loop Output Module
Sensor 1 and Sensor 2: Digital (S3L)/Frequency Input
• V+: +5 VDC out to sensor (black wire)
• FREQ: Frequency input signal from sensor (red wire)
• S3L DATA: Digital input signal from sensor (red wire)
• GND: Sensor ground (white wire or silver wire from paddle wheel fl ow sensor)
Loop Outputs
• Two Passive 4 to 20 mA current loop outputs 12 - 32 VDC, ±10% regulated (30 mA Max)
• Up to 4 additional Loop Outputs via the Dual Channel 4 to 20 mA Output Module
ModBus Module
• Six Channel Primary and Secondary values
GF 9950-10 Six Channel Transmitter
3
Page 4
Sensor Wiring
515/8510
525 2536/8512 2540
FLOW
2000 2100 2507
Start-Up Guide
9950 Frequency
DC POWER
Loop
Black
Red
No connection
White
FLOW
Loop
+V
FREQ
X
DATA
GND
+V
CH2 CH1
FREQ
+V
FREQ
DATA
GND
Technical Notes:
• See corresponding product manuals for maximum cable length.
• Maintain cable shield through cable splice.
• Route sensor cable away from AC power lines.
• Select "Yes" to CH # Flow 515/525 when setting up the input on the 9950. On the 515/8510 and 525 installations, connect the silver (shield) wire to earth ground in case of EMI noise interference.
2537-5
S1
2250 2350 2450
2551* 2751 2850
* 2551-XX-21, -41 Display Magmeter
NOTE: The 2850 has no SHIELD wire.
Blk Red Shld
9950 S3L Inputs
Technical Notes:
• Wiring terminals on the 2537 are rated for 16 to 22 AWG wires.
DC POWER
-
+
Black
Red
White
S2
No connection
Loop
Loop
+V
X
FREQ
DATA
GND
+V
CH2 CH1
FREQ
+V
FREQ
DATA
GND
• The cable must be 7 mm to 10 mm in diameter (0.275 in. to 0.394 in.) to seal properly in the liquid-tight connector.
• The conduit ports have ½ inch NPT threads. After routing the cables, seal the port with a liquid-tight conduit connector (3-9000.392-1) or with conduit.
• The 2537 models connect to the 9950 via the Digital (S3L) output.
Technical Notes:
• Sensor dependent, see sensor manual for maximum
9950 S3L Inputs
cable length.
• Maintain cable shield through cable splice.
DC POWER
• Route sensor cable away from AC power lines.
• Connect the silver (shield) wire to earth ground in case of EMI noise interference.
• The 9950 is not compatible with the 3-2580-63, dual channel conductivity device.
Sensor Terminal Connections
Sensor
Terminal
1 +V Shield +V 2 S3L/Freq GND S3L 3 GND S3L GND 4 Not Used +V Not Used
2551 2751 2850
Black
No connection
White
Shield
Loop
Loop
+V
X
FREQ
DATA
GND
+V
CH2 CH1
FREQ
+V
FREQ
DATA
GND
3-2610-51
Red
Blue
Green Brown
+
No connection
Power Supply
12 to 32 VDC
White
Black
Shield
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
FREQ
DATA
GND
CH2 CH1
FREQ
DATA
GND
DC POWER
+V
+V
PWR +
PWR –
+V
FREQ
DATA
GND
Technical Notes:
The wiring of the 3-2610-51 is non-standard:
• RED is 12 to 24 VDC (PWR +)
• WHITE is Data
• BLACK is VDC Ground (PWR -)
• A jumper MUST be installed between PWR- and
3
L GND.
S
WARNING: When using more than one DO sensor it is recommended to:
1. Not to use daisy chain wiring
2. Wire directly to CH1 or CH2 sensor input terminals
3. Share the loads across both input channels
NOTE: 3-2610-51 only.
9950 S3L Inputs
4
GF 9950-10 Six Channel Transmitter
Page 5
Start-Up Guide
Sensor Wiring
2551-XX-11
Blind Magmeter
Sensor
Terminal
1 Black (+V)
3
S L
Signal
Black
No connection
Red
White
Shield
9950 S3L Inputs
DC POWER
PWR+
PWR–
4
Loop1+
Loop1–
3
Loop2+
Loop2–
+V
X
FREQ
DATA
GND
+V
CH2 CH1
FREQ
DATA
GND
+V
FREQ
DATA
GND
Frequency
Black
Red
No connection
White
Shield
Only in case of EMI interference. Refer to Frequency Output Technical Notes (2551 & 2552) at right.
2 Red (S3L/Freq) 3 White (GND) 4 Not Used
Input Wiring for 2551 and 2552 sensors
• Either Frequency or Digital (S3L) may be used.
GF recommends con guring these sensors with the Digital (S3L) output to display
reverse fl ow (negative numbers).
• Input type is selected by choosing between “SENSOR FREQ” and "SENSOR S3L" in the FLOW sensor type INPUT menu (Refer to page 11).
2552
9950 Frequency
PWR+
PWR–
4
Loop1+
3
Loop1–
Loop2+
Loop2–
FREQ
X
DATA
GND
CH2 CH1
FREQ
DATA
GND
DC POWER
+V
+V
+V
FREQ
DATA
GND
2551 Technical Notes:
• When the blue jumper illustrated here is placed over both pins, the 2551-XX-11 (Blind Magmeter) outputs an open collector frequency signal. When the jumper is removed (or placed over one pin for storage) the 2551-XX-11 outputs a digital (S
3
L)
signal (recommended).
Frequency Output Technical Notes (2551 & 2552):
• The frequency output will be displayed as positive
ow regardless of the ow direction.
• 5 VDC power required by the 2551 & 2552 is supplied by the 9950. No additional power is required.
• Connect the silver wire (shield) to earth ground in case of EMI noise interference.
Black
No connection
Brown White
Blue
Shield
9950 S3L Inputs
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
X
FREQ
DATA
GND
+V
CH2 CH1
FREQ
DATA
GND
+V
FREQ
DATA
GND
Black Brown Blue
White
Shield
No
X
connection
9950 Frequency
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
FREQ
DATA
GND
FREQ
CH2 CH1
FREQ
+V
DATA
GND
+V
DATA
GND
2552 Technical Notes:
• The 2552 outputs an open collector frequency signal that can be connected to the 9950.
• Route sensor cable away from AC power lines.
• Connect the silver (shield)
wire to earth ground in case of EMI noise interference.
GF 9950-10 Six Channel Transmitter
5
Page 6
Sensor Wiring
Start-Up Guide
No connection No connection
2581
X X
+V FREQ DATA GND
9950
Digital (S3L)
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
FREQ
DATA
GND
Ground (White)
+V
CH2 CH1
FREQ
DATA
GND
Data (Red)
Power cable
Output cable
Blue not used
Brown Black
Not used
Shield
2581
Black
Red
24 VDC @ 1A min. Power Supply
+
Shield (EMC enhancement)
No connection
No connection
X
X
+V FREQ DATA GND
9950
Frequency
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
FREQ
DATA
GND
+V
CH2 CH1
FREQ
DATA
GND
Frequency (Red)
Ground (White)
Blue not used
Brown Black Shield
2581
Output cable
Not used
6
GF 9950-10 Six Channel Transmitter
Page 7
Power Wiring
Start-Up Guide
WARNING
DO NOT connect AC power to the DC version THE 3-9950-10 MUST BE POWERED BY 12 - 32 VDC at 0.5 A.
3-9950-10
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
FREQ
DATA
GND
CH2 CH1
FREQ
DATA
GND
and imported par
Red
Red
+V
+V
Red
Black
Black
Black
+
Power Supply
12 to 32 VDC
PLC or Recorder
Loop Input
+
4 to 20 mA
PLC or Recorder
Loop Input
+
4 to 20 mA
Caution:
For DC power input and loop current regulated voltage, use UL60950-1 or UL61010-1 certifi ed power supply. Power supply shall also be rated for operation at 4000 m altitude.
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
FREQ
DATA
GND
+V
CH2 CH1
FREQ
N
L
100-240 VAC~
50-60 Hz, 24VA Max
Made in U.S. from U.S. and imported par
DATA
GND
DC power optional
Red
Red
Caution: Electrical shock hazard exists!
Never connect live AC lines to the instrument.
3-9950-11
Red
Black
Black
Black
Neutral
Line
+
Power Supply
12 to 32VDC
PLC or Recorder
Loop Input
+
4 to 20 mA
PLC or Recorder
Loop Input
+
4 to 20 mA
100-240 AC
External DC Power is required for the current loops.
CAUTION
Keep AC power separate from sensor and signal wiring to prevent interference and damage to the 9950 Transmitter.
GF 9950-10 Six Channel Transmitter
7
Page 8
Start-Up Guide
Relay Module Wiring
Terminals accept 12 to 22 AWG wire.
Strip 10 to 12 mm (0.4 to 0.5 in.) of insulation from wire tips and tin bare ends to eliminate fraying.
Insert wire tip or ferrule completely into the terminal and secure with the screw.
Do not allow any AC leads that may be connected to the internal relays to come in contact with low voltage wiring.
3-9950.393-1 and 3-9950.393-2
Relay 1: The alarm is OFF during normal operation, and will
go ON when relay energizes according to 9950 Relay settings.
Relay 4: The valve is ON during normal operation, and will
go OFF when relay energizes according to 9950 Relay settings.
NO = normally open (closes when energized) NC = normally closed (opens when energized)
Mechanical Relays Rating: 5A 250 VAC, 5A 30 VDC Solid State Relays Rating: 50 mA 30V AC/DC
(for 309950.393-2 relay 1 & 2)
3-9950.393-3
The alarm is OFF during normal operation, and will go ON when the relay energizes according to 9950 Relay settings.
Binary input 4 will be ON when the tank level is above the level switch, Binary Input 4 will be OFF when the tank level is below the level switch.
Binary Input Ratings
Maximum Input voltage (without damage)
Minimum Input voltage (without damage)
Maximum input voltage for signal “Off” (low or "0")
30 VDC
-5 VDC (no operation below 0 VDC)
1.5 VDC
Minimum input voltage for signal “On” (high or "1")
Maximum current draw for signal “0” (low)
Minimum current draw for signal “1” (high)
Typical current draw for signal “1” (high)
3.0 VDC
<500 ADC
500 A
6.0 mA at 30 VDC, 4.8 mA at 24 VDC,
2.4 mA at 12 VDC, 1.0 mA at 5 VDC
3-9950.393-1
NO 1
RELAY 1
RELAY 2
RELAY 3
RELAY 4
3-9950.393-3 Internal Power
RELAY 1
RELAY 2
No Connection
BINARY INPUT 1 BINARY INPUT 2
BINARY INPUT 3 BINARY INPUT 4
3-9950.393-3 External Power
RELAY 1
RELAY 2
No Connection
BINARY INPUT 1 BINARY INPUT 2
BINARY INPUT 3 BINARY INPUT 4
N/C 1
COM 1
NO 2
N/C 2
COM 2
NO 3
N/C 3
COM 3
NO 4
N/C 4
COM 4
Flow
9950.393 Internal Power
N/O 1
N/C 1
COM 1
N/O 2 N/C 2
COM 2
SW1
SW2
SW3
SW4
SW COM
9950.393 External Power
N/O 1
N/C 1
COM 1
N/O 2
N/C 2
COM 2
SW1
SW2
SW3
SW4
SW COM
DC power
AC or DC
power
X
X
+
-
Level Switch
Level Switch
ALARM!
AC or DC
power
ALARM!
AC or DC
ALARM!
Valve
power
Level
AC or DC
power
Level
Power
Switch
INT
EXT
EXT
If an externally powered sensor is connected with the 3-9950.393-3 module, set the power switch to the EXT position. The module can power external sensors by
switching the power switch to the INT position. Maximum
INT
8
GF 9950-10 Six Channel Transmitter
power that can be drawn from the module is 6mA at 30 VDC.
Page 9
Start-Up Guide
Relay Module Wiring
8059 External Relay Module
3-8059-4AC Four external dry-contact relays with power supply 3-8059-4 Four external dry-contact relays 3-8050.396 RC Filter Kit 6205-0002 DIN Rail, 1m 6205-0003 End Clips, DIN Rail
The 9950-10/11 will support up to eight (8) relays, though only four actually fi t directly inside its compact ¼ DIN enclosure. 8059 External Relay Modules of four relays may be connected to the 9950 via digital (S³L) at the I/O Module (this does NOT­consume a sensor input channel).
External relays are completely independent of internal relays, so it is not necessary to install an internal relay module into the 9950 to use the external relays. In the 9950 menus, external relays will always be identifi ed as Relays A, B, C, and D.
Digital (S³L) pass-thru terminals on the 8059 can help to reduce sensor wiring congestion at the 9950. See the 8059 instruction manual for details.
External Relays mount on DIN rail inside a panel and can be tested remotely from the 9950, or locally with switches built-in for convenience. They also include LED relay status indicators, but the status of these and all relays is available at all times in a single screen in the 9950 View mode.
Any and all relays are freely assignable to any channel. Hysteresis and time delay are adjustable for each relay.
It is perfectly acceptable to mix and match relay types between internal and external modules. However, NOTE: The 9950 only allows external relays to be used in modes Off, Low, Hi, Window, Total, and Error.
8059 External Relay Modules contain dry-contact relays only. These electromechanical switches are suitable for many general­purpose applications, AC or DC, including AC loads up to 250 V.
Switching active loads (most commonly inductive) can cause contact arcing suffi cient to damage relays. The RC Filter Kit or “snubber” is available as an accessory to reduce or eliminate these damaging effects.
NC COM NO
RELAY A
Signet 3-8059 Relay Module
NC COM NO
RELAY B
COM
32:(5
LISTED
®
E171559
PASS-THRU
INPUT
+
S3L
-
+
S3L
-
NC COM NO
RELAY C
Test A
Test B
Test C
Test D
Relay A
Relay B
Relay C
Relay D
DC OUTPUT
+ -
NC C NO
NC C NO
RELAY B
RELAY A
Signet 3-8059 Relay Module
COM
POWER
LISTED
®
E171559
PASS-THRU
INPUT
+
S3L
-
+
S3L
S³L Connector{DC{AC
NC COM NO
RELAY D
AC or DC
Power
Test A
Test B
Test C
Test D
-
DC INPUT
- +
NC C NO
RELAY C
OUTPUT 24VDC
+ -
Relay A
Relay B
Relay C
Relay D
NC C NO
RELAY D
AC INPUT
L N
{
8059-4AC
Shield
NC C NO
RELAY A
Signet 3-8059 Relay Module
LISTED
®
E171559
INPUT
+
S3L
-
9950 S3L Inputs
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
FREQ
DATA
GND
+V
CH2 CH1
FREQ
DATA
GND
NC C NO
RELAY B
COM
32:(5
PASS-THRU
+
PWR +
PWR –
+V
FREQ
DATA
GND
8059-4AC8059-4
+ -
NC C NO
RELAY D
Relay A
Relay B
Relay C
Relay D
AC INPUT
L N
NC C NO
RELAY C
Test A
Test B
Test C
Test D
S3L
OUTPUT 24VDC
-
NOTE: 9950-1 and 9950-2 do not support the 8059 external relay.
GF 9950-10 Six Channel Transmitter
9
Page 10
Conductivity Module Wiring
Start-Up Guide
Loop Wiring
Red
Black
White
Shield
SHLD
TEMP
ISO GND
SIGNAL
3-9950.394-1
Single Channel
Conductivity Module
SHIELD (silver) TEMP (white)
ISO GND (black) SIGNAL (red)
3-9950.394-1
Single Channel Cond.
Shield
White
Black
Red
Shield
White Black
Red
ISO GND
ISO GND
SHLD
TEMP
SIGNAL
SHLD
TEMP
SIGNAL
3-9950.394-2
Dual Channel
Conductivity Module
PLC
or
Recorder
+
PLC
or
Recorder
+
Power
Supply
3-9950.398-2
Dual Channel
4 to 20 mA Module
+
Loop A+
Loop A–
Loop B+
Loop B–
Loop A + Loop A ­Loop B + Loop B -
3-9950.398-2
Dual Channel 4-20mA
3-9950.395-M
Modbus Module
RS485
RCVR
PLC
+ –
Yellow
Brown
(Optional) Shield
D1 + RS485
D0 + RS485
GND
10
Shield recomended in noisy environment
GF 9950-10 Six Channel Transmitter
Page 11
Start-Up Guide
System Setup Menu
Keypad Functions
The four buttons of the keypad ( ENTER) are used to navigate display modes according to the descriptions in this table. Notice the function of each button may change depending on the display mode.
This menu operation sets up the 9950 for basic function when initially powered up or when the 9950 has ALL six channels set to FACTORY.
1. Select desired language using the and arrows.
Press ENTER to save.
2. Press  to select the TIME FORMAT.
Press or to select 24 hour or AM/PM. Press ENTER to save the format selection.
6. Press  to move to the next selection,
SET DECIMAL MARK. Press and use or to highlight desired decimal separator (comma or decimal mark). Press ENTER to save the decimal mark selection.
7. Press  to move to the next selection, SELECT UNITS. Press and scroll or to choose Metric or U.S. Customary. Press ENTER save unit selection.
3. Press the  move to the next selection, SET TIME. Press to scroll through hours, and minutes place values. Use or to adjust the value. Press ENTER to save the time selection.
4. Press  to move to the next selection, SET DATE FORMAT. Press to select the MM/DD/YYYY date format and or to scroll through other format options. Press ENTER to save the date format selection.
5. Press  to move to the next selection, SET DATE. Press to scroll through the days, months, and years. Use or to adjust the number, and to select the next number. Press ENTER to save the Date selection.
4
ENGLISH FRANÇ
DEUTCH
ESPAÑOL
ENTER
TIME FORMAT
24 HOUR
TEXT 3
SET TIME
08:56:47
DATE FORMAT
TEXT 3
MM/DD/YYYY
SET DATE
TEXT 3
07/04/2022
SET DECIMAL MARK
TEXT 3
999.99
SELECT UNITS
TEXT 3
US CUSTOMARY
SETUP
TEXT 3
CHANNEL 1
PRESS ENTER
3
SETUP
CHANNEL 2
PRESS ENTER
SETUP
CHANNEL 3
SETUP
CHANNEL 4
SETUP
CHANNEL 5
SETUP
CHANNEL 6
EXIT STARTUP
24 HOUR
AM/PM
GERMAN LANGUAGE
00:00:040
LANGUAGE
AM/PM
MM/DD/YYYY
DD/MM/YYYY YYYY/MM/DD
LANGUAGE LANGUAGE
ENTER
09:11:04
02/25/20162
AM/PM
02/25/2016
99.999
99,999
METRIC
U.S. CUSTOMARY
NONE FACTORY
FLOW
pH ORP COND/RES
MM/DD/YYYY
DD/MM/YYYY
YYYY/MM/DD
OR
ENTER
saves changes
8. Press  to move to the next selection SETUP CHANNEL
Press ENTER. Use or to scroll through the sensor list and select another sensor type to assign to the channel. Press ENTER to save sensor type selected.
Press  to bypass channel 1 and scroll to any specific channel (1-6) for set up. NOTE: If no sensor is selected, it is recommended to select NONE as sensor choice.
Press to move to CHANNEL SOURCE.
Use or to scroll through the list of available sensors or modules for the channel TYPE selected in step 8. Press ENTER to save sensor source selected. Repeat steps in number 8 to set up channels 2 through 6. When done, press +
together one time exits to the
previous menu. Press + together a second time to exit Easy Start Up.
DATE FORMAT
NAME
SAVING
100% SOLUTION
ENTER
saves changes
LOOKING FOR
SENSOR TYPE
TEXT 3
NONE S3L 2209291234
S3L 2209292345
MOD1A MOD1B FREQ1 FREQ1
ENTER
saves changes
GF 9950-10 Six Channel Transmitter
11
Page 12
Start-Up Guide
Password
Password Overview
The password is required to start editing. Once entered correctly, this password will not be needed for subsequent uses, until the menu system is exited. The password will be required when the menu system is re-entered.
Your choice of password (STD or CODE) is selected in the Options Mode.
 STD
The standard (STD) password is , pressed in sequence. This password is designed to protect the 9950 from unintentional changes. It is best suited for systems where a group of people need to be able to change settings.
 CODE
The CODE default setting is 0000, adjustable to any 4-digit numerical code up to 9999. Using a personal code provides the maximum degree of security. This code can be modifi ed in the Options mode.
Operation
Channel 1,3,5
display area
Channel 2,4,6
display area
Numeric Values or labels
Bar Graph
Numeric Values or labels
Bar Graph
Relay 1 Indicator LED
CH5
00.0 0
CH6
00.0 0
Backlight Sensor (do not block)
Relay 2 Indicator LED
9999.9
9999.9
9999.9
9999.9
Relay 3 Indicator LED
UOM UOM
99.9 9
UOM UOM
99.9 9
ENTER
Relay 4 Indicator LED
4
Units of Measure (GPM, pH, sec, %, etc.)
Units of Measure (GPM, pH, sec, %, etc.)
Menu Navigation Keys
12
UP, DOWN keys
Scroll through Menu options or adjust values during editing Press both together to exit a menu or escape without saving
GF 9950-10 Six Channel Transmitter
RIGHT key
Select item or character to edit
ENTER
ENTER key
Access menus Save changes
Page 13
Table of Contents
Description .......................................................................1
Compatibility .....................................................................1
Safety Information ............................................................2
Installation ........................................................................2
Terminal Identifi cation ......................................................3
Wiring ......................................................................... 4-10
Sensor Wiring ......................................................... 4-6
Power Wiring ..............................................................7
Modules Wiring .................................................... 8-10
System Setup Menu .......................................................11
Password .......................................................................12
Keypad Operation ..........................................................12
Warranty Information ......................................................13
Dimensions ....................................................................14
Module Installation .........................................................14
Plug-In Modules .............................................................15
Relay Modules ...............................................................15
Conductivity Loop and Modbus Modules ......................16
Signal Type: Frequency ................................................16
Signal Type: Digital (S3L) .............................................16
Signal Type: 4 to 20 mA ................................................17
Relay Functions .............................................................17
Relay Modes ............................................................ 17-19
Derived Functions ..........................................................20
Sensor-Specifi c Menus ............................................ 21-50
Flow .................................................................... 21-22
pH ....................................................................... 23-25
ORP .................................................................... 26-27
Conductivity/Resistivity ...................................... 28-29
Pressure ............................................................. 30-31
Level/Volume ...................................................... 31-34
Temperature ....................................................... 34-35
4 to 20 mA .......................................................... 36-37
Salinity ................................................................ 38-40
Dissolved Oxygen .............................................. 41-42
Relay Menus ............................................................ 43-57
Low ..........................................................................43
High .........................................................................44
Window IN or Window OUT ....................................45
Proportional Pulse ...................................................46
Pulse Width Modulation ..........................................47
Volumetric Pulse .....................................................48
Total Pulse ...............................................................49
United States Pharmacopeia ..................................50
Error Mode ..............................................................51
Binary Input .............................................................52
Timer ................................................................. 53-55
Multiple .............................................................. 56-57
Menu System .................................................................58
Common Menus ....................................................... 59-64
Loop Menus .............................................................59
Option Menu ........................................................ 60-63
View Mode ...............................................................64
Troubleshooting ....................................................... 65-66
Averaging .......................................................................67
LOG Current Loop Output .............................................67
Custom Measurements ............................................ 68-71
Calibration Procedures - pH .................................... 72-73
Calibration Procedure - ORP ................................... 74-75
Calibration Procedure - Conductivity/Resistivity ............76
Calibration Procedure - Salinity .....................................77
Calibration Procedure - Flow .........................................78
Calibration Error Messages ...........................................79
USP Limits......................................................................80
Modbus .........................................................................80
Modbus Communication Setting Menus ........................81
Maintenance ...................................................................81
Specifi cations ........................................................... 82-83
Ordering Information ......................................................84
Warranty Information
Refer to your local Georg Fischer Sales offi ce for the most current warranty statement.
All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales offi ce or distributor. Product returned without a Service Form may not be warranty replaced or repaired.
GF products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, chemical poisoning, dry-out) or mishandling (e.g. broken glass, damaged membrane, freezing and/or extreme temperatures).
GF 9950-10 Six Channel Transmitter
13
Page 14
Dimensions
y
3-9950-10/-11
99.06 mm (3.90 in.)
12 34
Side ViewFront View
90.73 mm (3.57 in.)
82.60 mm (3.25 in.)
99.06 mm (3.90 in.)
Module Installation
ENTER
9950
MODULE 1
MODULE 2
102.37 mm (4.03 in.)
Dual Channel 4 to 20 mA Module
Rounded Corners
Important!
Ensure the rounded corner of the module is alligned with the rounded corner of the module slot before installing
91.44 mm (3.60 in.)
CAUTION
Avoid Electrostatic Discharge (ESD).
The plug-in modules may be installed either before or after the base unit is mounted. If the 9950 Base Unit will be mounted using the provided quick clip mounting bracket, install plug-in modules fi rst with power disconnected.
CAUTION
Exercise care when installing modules. Do not bend connecting pins.
To install modules:
Remove power from the 9950. Carefully align pins and connectors (do not bend connecting pins) and push module fi rmly into place, then attach with screw(s) for Relay Modules only.
Note: Modules can be installed in either slot.
14
GF 9950-10 Six Channel Transmitter
Shown with Single Channel Conductivit
Module installed in Module Slot 2
To move modules:
Remove power from the 9950.
For Relay Modules:
Unplug connectors, remove screw(s), and carefully pull module straight out from the base unit. Do not bend the connecting pins.
For Conductivity, Dual Channel 4 to 20 mA, and Modbus Modules:
Unplug connectors and squeeze the two retaining tabs while pulling module out of 9950.
WARNING
Relays may be connected to external high-voltage power sources or multiple power sources creating an electrocution hazard.
Page 15
Plug-In Modules
a
bcd
Optional modules and accessories are available for the 9950:
a. Optional Relay Module b. Optional Module 1 c. Optional Module 2
Each item is ordered separately. Modules are fi eld-replaceable at any time. Refer to Module Installation (page 8) and Ordering Information (page 82) for more details.
CAUTION
Avoid Electrostatic Discharge (ESD).
Minimize handling of the plug-in modules to reduce the possibility of damage due to ESD.
Handle modules by the edges.
Never touch any exposed circuitry or contacts.
Wear an anti-static wristband or stand on an anti-static mat, or keep one hand touching a
properly grounded pipe or other piece of properly grounded metal when handling modules.
Relay Modules
Mfr. Part No. Code Description
3-9950.393-1 159 310 268 Relay Module - 4 Mechanical Relays
3-9950.393-2 159 310 269 Relay Module - 2 Mechanical Relays, 2 Solid State Relays
3-9950.393-3 159 310 270 Relay Module - 2 Mechanical Relays , 4 Binary Inputs
The 9950 has a slot for an optional Relay Module for any of the modules listed above.
• Dry-contact relays are electromechanical switches with a moving contact armature. They are suitable for many general-purpose applications, AC or DC, including AC loads up to 250 V.
• Solid-state relays are electronic switches with no moving parts. They may be used with AC or DC loads, but have lower current handling capability and voltage ratings than Dry-contact relays. Solid-state relays will outlast Dry-contact relays in pulsing applications. 30 VDC/ VAC maximum.
• The four Binary Inputs can detect if an external sensor or switch is open or closed. The switch state can be used to control the relays in the module relay activation and are not considered an input measurement.
. The four binary inputs only control
3-9950.393-3
For wiring information, refer to the Relay Wiring section, page 8.
The 3-9950.393-3 module can be set to supply power to the external switch/ sensor. For switches/sensors that require power set the Int/Ext switch to the Int positions. If the external Switches/Sensors are powered set the Int/Ext switch to Ext. Important for externally powered switches/sensors ensure the output signal is within the allowable range for the 3-9950.393-3.
Int / Ext
Switch
EXT
INT
NOTE:
• The four red Indicator LEDs on the front panel of the 9950 show the status of relays 1, 2, 3, & 4. The LEDs will activate with or without a relay module installed.
• Hysteresis and time delay are adjustable for each relay.
CAUTION
CAUTION
Switching active loads (usually inductive) can cause contact arcing suffi cient to damage the relays.
The RC Filter Kit or “snubber” (part number 3-8050.396) is available as an accessory to reduce or eliminate these damaging effects. Recommended for inductive loads greater than 50 VAC (remote relays, solenoids, pumps, etc.)
Keep relay wiring separate from sensor and signal wiring to prevent interference and damage to the 9950 Transmitter.
GF 9950-10 Six Channel Transmitter
15
Page 16
Conductivity, Loop and Modbus Modules
Mfr. Part No. Code Description 3-9950.394-1 159 001 894 Single Channel Direct Conductivity Resistivity Module 3-9950.394-2 159 001 847 Dual Channel Conductivity/Resistivty Module 3-9950.395-M 159 001 905 Modbus Module 3-9950.398-2 159 001 848 Dual Channel 4 to 20 mA Current Loop Output Module
Direct Conductivity Resistivity Module
Compatible with all GF Conductivity Electrodes. Maximum Conductivity cable length is 30 m (100 ft). Up to two modules can be installed in a 9950 Transmitter.For longer distances use the 3-2850 Conductivity Sensor Electronics.
Dual Channel 4 to 20 mA Module
Two additional 4 to 20 mA passive current loop outputs in a single module, allows up to six current loop modules in the 9950 Transmitter.
Modbus Module
The Modbus Module connects the 9950 to serial RS485 Modbus automation networks.
Signal Type: Frequency
GF fl ow sensors 515/8510, 525, 2000, 2100, 2507, 2536/8512 and 2540 provide a frequency output. (Flow sensors 2537, 2551 and 2552 can be confi gured with either Digital (S
The maximum allowable cable length for sensors with frequency output is dependent upon the output signal strength of the sensors themselves, and the degree to which the signals are susceptible to EMI or "noise." This is largely a function of whether the sensors are self-powered (515/8510 and 525), or powered by an external source (-1).
3
L) or Frequency outputs).
• The input terminals on the 9950 carry frequency data signals from the sensor.
• The 9950 has a selection for 515/525 under the input menu. Select "Yes" for optimum signal performance.
• Do not route sensor or output cables in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
• In case of noise interference, ground the sensor SHIELD wire to a local earth ground at a point near the sensor.
• Consult the sensor manual for additional wiring information.
Flow sensor models with
Frequency
Output
515/8510 X
525 X
2000 X
2100 X
2507 X
2536/8512 X
2537 X
2540 X
2551 X
2552 X
Maximum
Cable Length
60 m
(200 ft)
305 m
(1000 ft)
Signal Type: Digital (S3L)
• The input terminals on the 9950 carry Digital (S3L) serial data from the sensor.
• Do not route sensor or output cables in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
• The TOTAL cable length from I/O devices to the transmitter must not exceed 305 m (1000 ft).
• In case of noise interference, ground the sensor SHIELD wire to a local earth ground at a point near the sensor.
• Consult the sensor manual for additional wiring information.
• The maximum cable length of the Digital (S of sensors connected and the size of the conductors in the cable. For best results, determine the maximum cable length for the system before routing cables.
16
GF 9950-10 Six Channel Transmitter
3
L) bus varies depending on the types
In case of noise interference, connect the cable shield to earth ground.
Maximum total cable length of the Digital (S3L) Bus:
The quality of the cable used in the bus determines the maximum length. The maximum cable length may not exceed 305 m (1000 ft), regardless of current requirements.
Page 17
Signal Type: 4 to 20 mA
When connecting a non-GF sensor to the 9950, the sensor’s 4 to 20 mA signal must be converted to Digital (S and converts it into Digital (S
1. Wire the 8058 between the 4 to 20 mA loop source and the 9950 Digital (S
3
L). The 8058 i-Go Signal Converter accepts any 4 to 20 mA signal
3
L).
3
L) input
terminals.
2. In the 9950 INPUT menu, sensor TYPE screen, select the '4 to 20 INPUT' sensor
(Refer to System Setup Menu, page 11).
9950 S3L Inputs
8058-1
8058-1
4-to-20 mA
Input
+GF+
4-20 mA input S3L Output
Signet 8058 i-Go™
4-20 mA to S
3
L Converter
Black
No connection
Red
White
Shield
3
Output
L S
X
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
FREQ
DATA
GND
CH2 CH1
FREQ
DATA
GND
DC POWER
+V
+V
+V
FREQ
DATA
GND
8058-2
+GF
+
3
2 6 5
L o P W
L o PWR
S DATA
S i gn
e
t 805
N/C 7 8
1
4
o p1
R
o p2
3
L
8
9
4-20 mA
L Converter
3
3
S L
4-20mA to S
8058 i-Go™
+GF+
5VDC MAX
3
Loop2+ 4
Loop2 - 5
Input
N/C 6
Output
Loop1 + 1
Loop1 - 2
N/C 3
3-8058.610D
3
9 GND
8 S L
7 +5VDC
8058-2 close-up
7 8 9
N/C
3
S L
DATA
No connection
Black
Red
White
X
9950 S3L Inputs
DC POWER
PWR+
PWR–
Loop1+
Loop1–
Loop2+
Loop2–
+V
FREQ
DATA
GND
+V
CH2 CH1
FREQ
DATA
GND
+V
FREQ
DATA
GND
Relay Functions
Fail-Safe Behavior
No matter the setting, the Open Collector output turns off if the 9950 looses power. This must be taken into account when evaluating system failure consequences. If the system layout requires a "closed" or "on" condition for the output in case of power loss, a mechanical dry-contact relay (NC contacts) must be used instead of the Open Collector (R1) output.
Set your relay functions to your own application requirements.
Once a setting is saved it becomes immediately active.
1. Press and hold ENTER for 3 seconds
2. Go to the Relay Menu by pressing then ENTER
3. Press the  to select desired source. You will be asked to enter a code or
password. Select source and press ENTER to confi rm
4. Press  to enter the relay MODE selection screen
5. Press  and then  to select R1 MODE LOW. Press ENTER to confirm
6. Press  to R1 SET LOW. Press  to enter GPM value of 5.5
7. Use the
and to change Set Point. Press ENTER to save
8. Scroll  to the R1 HYSTERESIS menu
9. Press  to edit
10. Set the hysteresis for this relay. Set the value to 2.5 gpm
11. Press ENTER
12. Scroll down  to the R1 ON DELAY menu
13. Press  to edit
14. Set the turn-on delay in seconds for the relay: 15.0
15. Press ENTER Then
16. Exit to View Mode
Example: Set a relay R1 to turn on at a low setpoint of
5.5 gpm with a time delay of 15 seconds and turn off at
8.0 gpm.
Remember,
SET LOW + hysteresis = OFF
Relay 1 Source = CH1 Primary Relay 1 Mode = Low R1 Set Low = 5.5 R1 Hysteresis = 2.50 R1 on delay = 15.0 sec
Relay function can be tested in the RELAY menu
GF 9950-10 Six Channel Transmitter
17
Page 18
Hysteresis
Time
Low Setpoint
Process
Low Setpoint + Hysteresis
Hysteresis
Time
High Setpoint
Process
High Setpoint - Hysteresis
Relay energized
Relay de-energized
Relay Modes
Time
High Limit
Hysteresis
Low Limit
Process
Window
10
5
Relay Output Rate:
0 to 100 Pulses/min.
Pulse rate
0 pulses
100 pulses
7.5
Time
High Limit
Hysteresis
Low Limit
Process
Window
Hysteresis
The 9950 relays are selectable and confi gurable and can be used as switches that respond when the process value moves above or below a user-defi ned setpoint or it can be used to generate a pulse at a rate proportional to the process value. They can be used for Low Alarm, High Alarm or Proportional Pulse triggering related to the process value. All relay functions are set up in the RELAY menus.
Low Setpoint:
Relay is on when the measured value is less than or equal to the setpoint. Relay turns off when the measured value is equal to or greater than the Low Setpoint + Hysteresis.
High Setpoint:
Relay is on when the measured value is greater than or equal to the setpoint. The relay turns off when the measured value is less than or equal to the High Setpoint - Hysteresis.
Proportional Pulse Operation:
The transmitter can output a pulse at the rate defi ned by the settings in the CAL menu and the sensor input. The maximum pulse output is 300 pulses per minute.
Example usage would be to control
solenoid-operated dosing pumps.
For example: As the process value rises above the setpoint, the output will start pulsing in relation to the process value, the maximum pulse endpoint and the programmed pulses/minute. The pulse rate will change as the process value changes and approaches the programmed endpoint. This functionality can be used to precisely control the process.
The starting point, endpoint and maximum pulse rate are select ­able in the RELAY menus.
NOTE: Relay LEDs will ash in PULSE mode.
100 pulses
Window In/Out:
Relay is on when the value is equal to or higher or lower than the high or low setpoint. WINDow IN = relay on if measurement is inside the window
of two setpoints.
WINDow OUT = relay on if measurement is outside the
window of two setpoints.
Window IN example
Starting point Endpoint
In the example:
• The output will be 0 pulses/min. when value is less than 5.
• The output will be 50 pulses/min. when value is 7.5.
• The output will be 100 pulses/min. when value is 10 or greater.
18
Window OUT example
GF 9950-10 Six Channel Transmitter
Page 19
y
Relay Modes
• Pulse Width Modulation
PWM automatically varies the ratio of ON time to OFF time proportional to minimum and maximum range settings.
The relay period is the sum of the time a relay is ON and the time it is OFF.
Relay pulse width is the time the relay is ON.
The 9950 must be programmed with the relay period, and with the low and high setpoints.
NOTE: The PWM mode is not used for Pressure applications. Relay LEDs will fl ash in PWM mode.
PWM Maximum
100%
80%
60%
40%
Relay Pulse Width
20%
(as a % of Relay period)
0%
alwa
Relay is
s OFF
PWM Minimum
Process value
Range
Relay is
always ON
NOTE: The pulse width will be 100% of the relay period (relay always ON) when the process value is greater than the maximum range.
• Totalizer Volume
Relay activates and latches when a specifi ed volume of fl uid is registered on the resettable totalizer. For Flow inputs only.
Total Volume mode counts the TOTALIZER Units until the setpoint volume is reached, then turns on the relay until the resettable totalizer is reset.
If the Resettable Totalizer reading is greater than the setpoint, the relay will be turned on immediately. The relay will be off when the totalizer is below the set point or the resettable totalizer is reset to zero.
This mode is useful to trigger a reminder when a process is due, as for a backwash cycle or fi lter change.
Relay Mode - ERROR
Example:
• The pulse width will be 0% of the relay period (relay always OFF) when the process value is less than the minimum range.
• The pulse width will be 100% of the relay period (relay always ON) when the process value is greater than the maximum range.
• The pulse width will be 60% of the relay period when the process value is at 60% of the span between the minimum and maximum range.
• Volumetric Pulse
The relay will activate for the set pulse width once the specifi c volume of fl uid is registered. Flow Inputs only.
NOTE: Relay LEDs will turn on when the relay is active
CH1
00.0 0
CH2
00.0 0
_ _ _ _ ._
NO SENSOR
9999.9
7 .0
6 2 .8
mV
GPMS
00.00
pH
15.00
°F
Error Mode
When the relay "Error" mode is assigned to a sensor, any of the following sensor-related errors will activate the relay and turn the screen "RED" with the appropriate error message.
Wrong sensor connected
No response from a sensor, e.g. a missing sensor or a broken wire
Sensor returning a measurement error (for pH and ORP, a missing probe will cause this)
An out-of-range measurement value from the sensor
An expired Dissolved Oxygen cap
A missing Dissolved Oxygen cap
A pH electrode whose glass impedance is less than or equal to the user-specifi ed broken glass limit
A conductivity module that is compensating the wire length and is not returning a measurement
A missing module
The wrong module inserted into the slot
GF 9950-10 Six Channel Transmitter
19
Page 20
R embrane
R embrane
R embrane
Feed
Concentrate
Permeate
R embrane
R embrane
R embrane
Feed
Concentrate
Permeate
Feed
Concentrate
R embrane
R embrane
R embrane
Permeate
Derived Functions
RO Membrane
RO Membrane
RO Membrane
Feed
Concentrate
Permeate
When two of the same type of measurement (either primary or secondary) are present, the 9950 can calculate several derived functions from like pairs of measurements. Set-up menu located in the Options menu, refer to page 59. Up to four derived Functions can be defi ned and used as the source for a loop output or relay control. Derived calculations can be viewed in the View Mode.
• Flow, Temperature, Pressure, Conductivity and Level measurements must have matching units. (Flow channels must also have same time base).
• Conductivity channels will automatically scale to S/cm before the function calculation is made.
Three types of derived measurements can be applied to any set of sensors, regardless of type.
• Ratio: Measurement 1 ÷ Measurement 2 or Measurement 2 ÷ Measurement 1
• Delta (Difference): Measurement 1 - Measurement 2 or Measurement 2 - Measurement 1
• Sum: Measurement 1 + Measurement 2
• Custom labels can be assigned to Derived Functions for ease of identifi cation.
% Passage and % Reject
• % Passage and % Reject are derived functions based on conductivity measurements only, specifi cally for use in reverse osmosis systems.
• % Passage is the amount of contaminates remaining in the product water compared to the level of contaminates in the feed water. For example, if the feed water measures 375 S and the product water measures 18.75 S, the % Passage is
(18.75/375) x 100= 5%
• % Reject is the amount of contaminates rejected to the concentrate water compared to the amount of contaminates in the feed water. For example, if the feed water measures 375 uS and the product water measures 18.75 uS the % reject is [1-(18.75/375)] x 100 = 95%
• Decreasing Reject values and increasing Passage values usually indicate a problem with the RO membrane.
% Passage: (Permeate ÷ Feed) x 100
% Reject: [1 - (Permeate ÷ Feed)] x 100
% Recovery
• % Recovery is a derived function based on fl ow rate, in a reverse osmosis system.
• To measure % Recovery, the 9950 must have two fl ow sensors connected. They may be located in the Feed line, the Concentrate line or the Permeate line.
• The 9950 provides 3 different methods for calculating Recovery to accommodate any confi guration.
• Both fl ow sensors must use the same time base and units of measure.
% Recovery A: (Permeate ÷ Feed) x 100 In the Setup menu, select the option that states % Recovery A,
FEED: PERMEATE
20
GF 9950-10 Six Channel Transmitter
% Recovery B: Permeate ÷ (Permeate + Concentrate) x 100 In the Setup menu, select the option that states % Recovery B,
PERMEATE: CONC
% Recovery C: [(Feed - Concentrate) ÷ Feed] x 100 In the Setup menu, select the option that states % Recovery C, FEED: CONC
Page 21
Sensor-Specifi c Menus
Flow
The following pages list the sensor-specifi c settings for each sensor type. The user can confi gure the unit to run each channel on different parameters if needed.
VIEW Mode Menu
CH1
50.30
40560.0
0.00
CH2
987.9
65.5
0.00
Displays the fl ow rate and the resettable or permanent totalizer below. This is the normal display and does not time out.
CH1
50.30
FLOW
0.00
CH2
987.9
FLOW
0.00
CH 1 TOTALIZER
P 40560.0
R 40560.0 ->
LITERS
GPM
100.00
GPM
100.00
GPM
R
100.00
GPM
R
100.00
View Measurement 2 displays the primary value and the channel name. This screen does not time out.
Displays the Permanent Totalizer and Resettable Totalizer for each FLOW channel (note the "P" indicating Permanent and "R" for Resettable) Pressing will prompt you to reset. Pressing again will verify YES or NO. Pressing ENTER will reset the totalizer.
FLOW Setup Checklist
1. Make sure FLOW sensor type is selected (Refer to System Setup Menu, page 9).
2. Set the Units of Measurement.
3. Set Sensor Type (Freq or S3L) and if 515 or 525, select Yes.
4. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
5. Set K-Factor (pulses per Unit Volume) from Flow Sensor manual.
6. Set Totalizer factor.
7. Set Last Cal Date and initials.
8. If applicable, set up relay functions for your application.
INPUT Menu
SETUP
CHANNEL 1
ENTER
CH1 FACTORY
TYPE
FACTORY
CH1 FLOW
SOURCE
NONE
CHANNEL 1
NAME
FLOW
CH 1 FLOW
515/2536
YES
CH 1 FLOW
GPM
FLOW UNITS
CH 1 FLOW
CUST/S
CUSTOM UNITS
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Press ENTER to save sensor type.
Press the arrow, using the or to select the INPUT source to the channel. To change sensor selection after its been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = FLOW
If the fl ow sensor is self powered such as a 515/525, select YES. Select NO if an open connector such as 2536/2540/2100 fl ow sensor is being used. Default = YES
Select the FLOW RATE UNIT from the pull down menu. Custom units can be created and used. The last character sets the timebase: S (seconds) M (minutes) H (hours) D (days). Default = GPM
NOTE: Screen will only appear if CUSTOM is selected from previous menu. English, French, German, and Spanish Only. Display visible only when FLOW UNITS CUST/S, CUST/M, CUST/H, CUST/D is selected. Allows user to set custom fl ow rate units for primary measurement. CUST/S, M, H, D label can be adjusted to refl ect to user's custom units.
GF 9950-10 Six Channel Transmitter
21
Page 22
Flow
INPUT Menu
TOTALIZER
TOTAL UNITS
GALLONS
(LABEL only) Identifi es the totalizer units, has no effect on any calculations. Default = LITERS
TOTALIZER
DISPLAY
R TOTALIZER
CH1 FLOW
100.0
SENSITIVITY
CH1 FLOW
AVERAGE
OFF
CAL Menu
CH 1 FLOW
OFF
HOLD OUTPUTS
CH 1 FLOW
KF
60.00
CH 1 FLOW
TF
1.00
CH 1 FLOW
Set the Totalizer displayed in view mode. Select R TOTALIZER for resettable or select PERM TOTALIZER for permanent. Default = R TOTALIZER
The Sensitivity setting determines how the 9950 responds to sudden changes in the fl ow rate for a frequency sensor. The value is expressed in units of measurement. If the setting is exceeded, it "overrides" the Averaging function briefl y to allow for the actual change in fl ow rate to be displayed. Averaging resumes shortly after. The result is a smooth fl ow display and a quick response to large shifts in the fl ow rate. To disable the sensitivity function set it to 99999.
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF
ON prevents relays from activating while making adjustments and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it off. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Set K-Factor (pulses per unit volume) from Flow Sensor manual. Min: 0.0001, max. 9999999. Cannot be zero. Default = 60.0000
Sets the volume of each count of the Totalizer as a multiple of the volume unit of the K-Factor. Min: 0.0001, max. 9999999. Cannot be zero. Default = 1.0000
22
RATE CAL ->
MULTIPLE
CH 1 FLOW
VOLUME CAL ->
OFF
CH1 FLOW
LAST CAL ->
12-31-2015 XX
GF 9950-10 Six Channel Transmitter
Select to calibrate using rate method. (Refer to page 76)
Select to calibrate using Volume method. (Refer to page 76)
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
Page 23
pH
VIEW Mode Menu
CH1
7.0
77.0
-1.00
CH2
4.0
99.0
-1.00
Displays the pH value and the temperature below. This is the normal display and does not time out.
pH
°F
15.00
pH
°F
15.00
pH Setup Checklist
1. Make sure pH sensor type is selected. (Refer to System Setup Menu, page 11)
2. Set the Temperature Units (°C or °F).
3. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
4. Perform calibration (EasyCal, Standard or Standard and Slope).
5. Set Last Cal Date and initials.
6. Select source for Relay output (pH or Temp).
7. If applicable, set up relay functions for your application.
CH1
pH
-1.00
CH2
pH
-1.00
GLASS IMPEDANCE
CH1 1049 MOhm ->
7.0
7.0
RAW
CH 1 264.10 mV
CH2 182.10 mV
REF IMPEDANCE
CH1 27 MOhm ->
Sensor Data
CH1->
pH
15.00
pH
15.00
View Measurement 2 displays the primary value and the channel name. This screen does not time out.
Displays the RAW millivolt input from the electrode. Use this display to determine the relative condition of your electrode during periodic calibration.
[Glass Impedance] 2751 only, pH only, last measured glass Impedance, press key to manually update reading.
2751-X, 274X electrodes Only, last reference impedance measurement, press key to update reading.
2571-X only. Using the key, you will see the following screens: [Serial Number] Electrode Serial Number, [Model Number] Electrode Model Number, [pH Calibration Slope] Calibration Slope Value, [pH Offset] Calibration Offset, [Temperature Offset] pH only, Temperature Offset Value, [Factory Impedance] pH only, Factory pH glass Impedance, [Usage Time] Electrode Runtime in Hours, [Minimum pH] Minimum Measured Value, [Maximum pH] Maximum Measured Value, [Minimum Temperature] pH only, Minimum Measured Temperature, [Maximum Temperature] pH only, <Maximum Measured Temperature.
EASYCAL
CH1 pH ->
CH2 pH ->
pH EASYCAL looks for buffers of 4, 7, or 10 pH buffers. (Refer to page 71)
GF 9950-10 Six Channel Transmitter
23
Page 24
pH
INPUT Menu
SETUP
CHANNEL 1
ENTER
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly.
NOTE: Display only available when the 9950 is in FACTORY mode.
CH1 FACTORY
TYPE
FACTORY
CH1 pH
SOURCE
NONE
CHANNEL 2
NAME
pH
CH 2 pH
°F
AT INSTRUMENT
CH 2 pH
AVERAGE
CUST CEOFFL CONST
CH 1 pH
60
IMP UPDATE TIME
Press the arrow, and use the and arrow to select which TYPE of the sensor will be assigned to the channel(s) (1-6). Choose pH, and press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain pH sensors electronics (2751-X) To change sensor selection after its been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = pH
Select °F or °C. Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF
NOTE: GF Strongly recommends leaving Averaging OFF for pH, ORP, and Pressure measurements.
2751, pH Only. Enter the time between automatic impedance measurement updates in 15 minute increments, 0 = Off - manual updates only. Default = 60
MOhm
03
BROKEN GLASS
CH 1 pH
3X
HI IMP WARN
2751, pH Only. Enter the impedance value for Broken pH Glass, 0 = Off, Default = 3
2751, pH only. Enter the high Impedance warning set point as NONE, 2, 3, 4, or 5 times the factory Impedance of the electrode. Default = NONE
24
GF 9950-10 Six Channel Transmitter
Page 25
pH
CAL Menu
CH 2 pH
OFF
HOLD OUTPUTS
CH 2 pH
CAL
AT INSTRUMENT
CH 2 pH
EASYCAL ->
CH 2 pH
SET
TEMPERATURE ->
CH 2 pH
°SETF
pH STANDARD ->
CH2 pH
SET
pH SLOPE
CH2 pH
Setting to ON prevents relays from activating while making adjustments or calibrating. Relays in PULSE mode will suspend pulsing. 4 to 20 mA outputs are held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Select AT SENSOR to perform calibration using the GF 2751-2 Sensor Electronics. Select AT INSTRUMENT to perform calibrations at the 9950 using the 3-2751-1, 3-2751-3, or 3-2751-4 electronics. (Refer to pH Calibration procedures, page 71). Default = AT INSTRUMENT
(CAL AT INSTRUMENT only) Press to start the EasyCal process. The password prompt is the same as the other parameters. (Refer to pH EasyCal procedure, page 71)
(CAL AT INSTRUMENT only) Applies a linear offset to the temperature measurement. "SAVING" will appear if offset is acceptable, "ERR TOO LARGE TO CALIBRATE" if offset is outside of range.
(CAL AT INSTRUMENT only) Applies an offset to the pH measurement. The slope value and the standard value must be at least 2 pH units apart. The ideal values are the minimum and maximum values of your process. (Refer to pH Calibration procedures, page 71). Shows error message if offset is too high.
(CAL AT INSTRUMENT only) Applies a slope to the pH measurement. The slope value and the standard value must be at least 2 pH units apart. The ideal values are the minimum and maximum values of your process. (Refer to pH Calibration procedures, page 71). Shows error message if slope is too low or high.
SET
RESET pH CAL ->
CH2 pH
100.0
RESET TEMPCAL ->
CH 2 pH
100LAST CAL.00
12 - 31 - 2015 XX
(CAL AT INSTRUMENT only) Press  to reset pH calibration to factory default.
(CAL AT INSTRUMENT only) Press  to reset temperature calibration to factory default.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
GF 9950-10 Six Channel Transmitter
25
Page 26
ORP
VIEW Mode Menu
CH1
CUSTOM LABEL
-999.0
CH2
CUSTOM LABEL
-999.0
Displays the ORP millivolt value. This is the normal display and does not time out.
84.0
1000.0
74.0
1000.0
mV
mV
ORP Setup Checklist
1. Make sure ORP sensor type is selected (Refer to System Setup Menu, page 11).
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Averaging.
4. Perform calibration or set Standard (and Slope if desired).
5. Set Last Cal Date and initials.
6. If applicable, set up relay functions for your application.
CH1
ORP
-999.0
CH2
ORP
-999.0
76.0
50.0
RAW
CH 1 264.10 mV
CH2 182.10 mV
REF IMPEDANCE
CH1 27 MOhm ->
Sensor Data
CH1->
EASYCAL
CH1 ORP ->
CH2 ORP ->
mV
1000.0
mV
1000.0
View Measurement 2 displays the primary value along with the channel name. The screen does not time out.
Displays the RAW millivolt input from the electrode. Use this display to determine the relative condition of your electrode during periodic calibration.
2751, 276x electrodes only, last reference impedance measurement, press to update reading.
2571 only. Using the key, you will see the following screens: [Serial Number] Electrode Serial Number, [Model Number] Electrode Model Number, [ORP Calibration Slope] Calibration Slope Value, [ORP Calibration Offset] Calibration Offset, [Usage Time] Electrode Runtime in Hours, [Minimum ORP] Minimum Measured Value, [Maximum ORP] Maximum Measured Value.
ORP EASYCAL looks for buffer of 87, 264, or 476 mV. Press to begin EASYCAL. (Refer to page 73)
26
GF 9950-10 Six Channel Transmitter
Page 27
ORP
INPUT Menu
SETUP
CHANNEL 2
ENTER
CH2 FACTORY
TYPE
FACTORY
CH2 ORP
SOURCE
NONE
CHANNEL 2
NAME
ORP
CH 2 ORP
AVERAGE
CUST CEOFFL CONST
CAL Menu
CH 2 ORP
OFF
HOLD OUTPUTS
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose ORP, press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain ORP sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = ORP
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF NOTE: GF Strongly recommends leaving Averaging OFF for pH, ORP, and Pressure measurements.
Setting to ON prevents relays from activating while making adjustments or calibrating. Relays in PULSE mode will suspend pulsing. 4 to 20 mA outputs are held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
CH 2 ORP
CAL
AT INSTRUMENT
CH 2 ORP
EASYCAL ->
CUST CELL CONST
CH 2 ORP
SET
ORP STANDARD ->
CH 2 ORP
°SETF
ORP SLOPE ->
CH2 ORP
AVERAGE
RESET ORP CAL ->
CH2 ORP
Select AT SENSOR to perform calibration using the GF 2751-2 sensor electronics. Select AT INSTRUMENT to perform calibrations at the 9950 using the 3-2751-1, 3-2751-3, or 3-2751-4 electronics. (Refer to ORP Calibration procedures, page 73). Default = AT INSTRUMENT
(CAL AT INSTRUMENT only) Press  to start the EasyCal process. You will be prompted to enter your password. (Refer to ORP EasyCal procedure, page 73)
(CAL AT INSTRUMENT only) Applies a linear offset to the ORP measurement. For two-point calibrations, assign the min or max value of your process to ORP STANDARD. (Refer to ORP Calibration procedures, page 73)
(CAL AT INSTRUMENT only) Applies a slope to the ORP measurement. The ORP SLOPE is used for two­point calibration along with the ORP STANDARD. If you applied the min value of your process to the ORP STANDARD, then apply the max value to the ORP SLOPE. Also, apply the min value to the ORP SLOPE. The slope value and the standard value must be at least 80 mV apart. (Refer to ORP Calibration, page 73)
(CAL AT INSTRUMENT only) Resets calibration to factory settings. After pressing , select YES/NO.
LAST CAL
12-31-2015 XX
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
GF 9950-10 Six Channel Transmitter
27
Page 28
Conductivity / Resistivity
VIEW Mode Menu
CH1
1205.0 0
77.0
0.0000
CH2
987.9
65.5
0.0000
Displays the conductivity value and the temperature below. This is the normal display and does not time out.
uS
°F
100.00
uS
°F
100.00
Cond/Res Setup Checklist
1. Make sure COND/RES sensor type is selected. (Refer to System Setup Menu, page 11)
2. Set Cell Constant.
3. Set the Temperature Units (°C or °F).
4. Set Conductivity units.
5. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
6. Set Temperature Compensation.
7. Set Last Cal Date and initials.
8. Select source for Relay output (COND or TEMP).
9. If applicable, setup relay functions for your application.
1205.00
CH1
COND/RES
0.0000
9999.9
CH2
COND/RES
0.0000
INPUT Menu
SETUP
CHANNEL 1
ENTER
CH1 FACTORY
TYPE
FACTORY
CH1 COND/RES
SOURCE
NONE
CHANNEL 2
NAME
COND/RES
uS
100.00
uS
100.00
View Measurement 2 displays the primary value and the channel name. The screen does not time out.
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose COND/RES, press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain Conductivity/Resistivity sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = COND/RES
CH 2 COND/RES
CELL CONSTANT
CH 2 COND/RES
SENSOR 1.00000
CUST CELL CONST
CH 2 COND/RES
28
SENSOR 1.0
uS
COND UNITS
GF 9950-10 Six Channel Transmitter
Press the arrow, using the or to select the cell constant - 20.0, 10.0, 1.0, 0.1, 0.01, or CUSTOM. Default = 1.0 (See NOTE below)
Display visible only when CUST is selected in the CELL CONSTANT selection of the menu. Press the arrow, using the or to enter the precise cell constant from the certifi cate provided with your sensor or from the information label on the sensor if CUST (custom) cell constant is selected.
Press the arrow, using the or to select the measurement units S, mS, PPM, PPB, KOhm, or MOhm. Default = S NOTE: In USP Relay Mode, TEMP COMP must be set to NONE and Unit Of Measure must be set to S.
Page 29
Conductivity / Resistivity
INPUT Menu
CH 2 COND/RES
0.50
TDS FACTOR
CH2 COND/RES
°F
TEMP UNITS
CH2 COND/RES
AVERAGE
OFF
CH2 COND/RES
TEMP COMP
LINEAR
CH 2 COND/RES
2.0
ADJ TEMP COMP
CAL Menu
CH 2 COND/RES
OFF
HOLD OUTPUTS
Display only visible when PPM or PPB units are selected. Press the arrow, using the or to select the TDS factor, a TDS factor for converting the conductivity units to TDS is required. Default = 0.50
Select °F or °C. Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF
Press the arrow, using the or to select the temperature compensation (NONE, LINEAR, PURE H2O). Default = LINEAR NOTE: In USP Relay Mode in Conductivity, Relay Source must be set to COND, TEMP COMP must be set to NONE and Unit Of Measure must be set to S.
Adjusted the % of temperature compensation. Maximum slope setting is 9.99 % per ºC. Default = 2.0 (If Temperature Compensation setting is NONE, this item will not be displayed)
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
CH 2 COND/RES
AUTO CAL ->
CS
CH 2 COND/RES
MANUAL CAL ->
CL CONST
CH 2 COND/RES
SET
TEMPERATURE
CH 2 COND/RES
F
RESET CONDCAL ->
CH2 COND/RES
AVERAGE
RESET TEMPCAL ->
CH2 COND/RES
Shows real-time value and automatically selects closest standard. “PLACE SENSOR IN STANDARD”. Unit waits until reading is stable; if bad calibration, returns “ERROR, CANNOT DETERMINE STANDARD”. Refer to buffer values and AUTO CAL Procedure. Refer to page 75.
Shows the live CONDUCTIVITY value on bottom line; when user presses any button the live value is frozen and the user edits that value. If bad calibration, returns “ERR TOO LARGE TO CALIBRATE”. Refer to Manual Cal procedure, page 75
Shows the live TEMPERATURE value on bottom line; when user presses any button the live value is frozen and the user edits that value. If bad calibration, returns “ERR TOO LARGE TO CALIBRATE”.
Resets Conductivity calibration. After pressing , select YES/NO.
Resets Temperature calibration. After pressing , select YES/NO.
LAST CAL
12-31-2015 XX
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
GF 9950-10 Six Channel Transmitter
29
Page 30
Pressure
VIEW Mode Menu
CH1
7.0
77.0
0.0
CH2
4.0
99.0
0.0
Displays the pressure reading. This is the normal display and does not time out.
PSI
°F
10.0
PSI
°F
10.0
PRESSURE Setup Checklist
1. Make sure PRESSURE sensor type is selected. (Refer to System Setup Menu, page 11)
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (PSI, BAR, KPa).
4. Set Last Cal Date and initials.
5. If applicable, set up relay functions for your application.
CH1
PRESSURE
0.0
CH2
PRESSURE
0.0
90.0
89.0
INPUT Menu
SET UP
CHANNEL 1
ENTER
CH 1 FACTORY
TYPE
FACTORY
CH 1 PRESSURE
SOURCE
CUST CENONEL CONST
CHANNEL 2
NAME
PRESSURE
CH 2 PRESSURE
PSI
PRESS UNITS
PSI
10.0
PSI
10.0
View Measurement 2 displays the primary value and the channel name. This screen does not time out.
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose PRESSURE, press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain PRESSURE sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = PRESSURE
Press the arrow, using the or to select the units of pressure measurement - PSI, BAR, or KPa. Default = Determined by the SELECT UNITS screen menu. Metric = Bar, U.S. Customary = PSI.
CH 2 PRESSURE
AVERAGE
CUST CEOFFL CONST
30
GF 9950-10 Six Channel Transmitter
Dampens display, output and relay response rates. Select: Low, Med, High, OFF. Default = OFF. GF strongly recommends leaving averaging OFF for pH and pressure measurements
Page 31
Pressure
CAL Menu
CH 2 PRESSURE
OFF
HOLD OUTPUTS
CH 2 PRESSURE
SET ZERO
CH 2 PRESSURE
SET PRESSURE
CH 2 PRESSURE
RESET CAL ->
CH2 ORP
LAST CAL
12-31-2015 XX
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
With process pressure at zero, press the arrow, using the or to set zero point for measurement. NOTE: Application must be OFF and have zero pressure to perform this operation.
Press the arrow, using the or to enter the pressure reading from a external reference NOTE: Allows a maximum offset of 5 PSI to be entered.
Resets calibration to factory default. After pressing , select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
Level / Volume
VIEW Mode Menu
CH1
25.0
78.5
0.0000
CH2
20.0
99.0
0.0000
Displays the level value and the volume below. This is the normal display and does not time out.
CH1
LEVEL
0.0000
CH2
LEVEL
0.0000
25.0
42.0
FT
10.000
FT
10.000
FT
GAL
10.000
FT
GAL
10.000
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
LEVEL/VOLUME Setup Checklist
1. Make sure LEVEL/VOLUME sensor type is selected. (Refer to System Setup Menu, page 11)
2. Select Main Measurement (Level or Volume).
3. Set Units of Measurement for LEVEL display (FT, IN, M, CM).
4. If desired, set Units of Measurement for VOLUME display.
5. Set the minimum and maximum 4 to 20 mA setpoints.
6. Set Specifi c Gravity.
7. Set Sensor Offset.
8. If VOLUME is used, set Shape.
9. Set Last Cal Date and initials.
10. If applicable, set up relay functions for your application.
GF 9950-10 Six Channel Transmitter
31
Page 32
Level / Volume
INPUT Menu
SETUP
CHANNEL 2
ENTER
CH 2 FACTORY
TYPE
FACTORY
CH 2 LEVEL
SOURCE
NONE
CHANNEL 2
NAME
LEVEL
CH 2 LEVEL
MAIN MEAS
LEVEL
CH 2 LEVEL
M
C LEVEL UNITSL
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose LEVEL, press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain LEVEL/VOLUME sensors. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = LEVEL
Press the arrow, using the or to select between LEVEL and VOLUME. Default = LEVEL
Press the arrow, using the or to select unit of measure for LEVEL display (FT, IN, M, CM). Default = Determined by the SELECT UNITS screen menu. Metric = M, U.S. Customary = FT.
CH 2 LEVEL
DISP LEVEL AS %
OFF
CH2 LEVEL
L
VOLUME UNITS
CH 2 LEVEL
DISP VOL AS %
OFF
CH 2 LEVEL
1.00
SPEC GRAVITYS
CH 2 LEVEL
0.00
SENSOR OFFSET
CH2 LEVEL
AVERAGE
OFF
ON = Measurement will be displayed as a percentage of full scale. OFF = Measurement will be displayed in unit of measure selected in previous setting. Default = OFF
3, in3
Select unit of measure for VOLUME display (GAL, L, Lb, KG, FT
, M3, cm3).
Default = Determined by the SELECT UNITS screen menu. Metric = L, U.S. Customary = GAL.
ON = Measurement will be displayed as a percentage of full scale. OFF = Measurement will be displayed in unit of measure selected in previous setting. Default = OFF
Enter the specifi c gravity of the fl uid at normal operating temperature. Default = 1.0000 (water)
Enter the distance from sensor location to the Zero reference point in the vessel. Displayed in units of measure chosen in LEVEL UNITS. Default = 0.00
Dampens display, output and relay response rates. Select Low, Med, High, OFF Default = OFF
32
GF 9950-10 Six Channel Transmitter
Page 33
Level / Volume
INPUT Menu
CH 2 LEVEL
VERT CYLINDER
SHAPE
M
2.00
TANK DIAMETER
FT
10.00
TANK LENGTH
FT
10.0
TANK WIDTH
CH 2 LEVEL
4
NUM CUST POINTS
CH 2 LEVEL
MANUAL
LEVEL CALC
FT
Select the shape of the vessel where the level sensor is located. VERT CYLINDER, HORIZ CYLINDER, RECTANGLE, or CUSTOM. (To defi ne a custom tank shape, refer to page 67, Defi ning a Custom Tank). Default = VERT CYLINDER
If VERT CYLINDER or HORIZ CYLINDER is selected, enter the diameter of the cylinder. Displayed in units of measure chosen in LEVEL UNITS. Default = 2.0000
If HORIZ CYLINDER is selected, enter the diameter of the cylinder. Set length of tank. Default = 10.00
Display seen only when RECTANGLE shape is selected. Set width of tank. Default = 10.00
If Custom shape is selected, enter the number of measurement points to be used to defi ne the vessel shape (Refer to Level and Volume Calculation in Custom Shaped Vessels, page 68). Minimum 4 points, maximum 32 points. A larger number of points improves accuracy.
Select (AUTO, MANUAL). Manual allows you to edit both the Level and the corresponding Volume for your custom tank. Automatic allows you to edit the Volume measurement (while displaying an automatically calculated Level value).
0.00
POINT X LEVEL
FT
0.00
POINT X VOL
Where (X) is the number of custom points selected in "NUM CUST POINTS" used to calculate LEVEL.
Where (X) is the number of custom points selected in "NUM CUST POINTS" used to calculate VOLUME.
GF 9950-10 Six Channel Transmitter
33
Page 34
Level / Volume
CAL Menu
CH 2 LEVEL
OFF
HOLD OUTPUTS
CH 2 LEVEL
LEVEL CAL
CH 2 LEVEL
REURESET CALELL
CH 2 LEVEL
CH2 LEVEL
SET
LAST CAL
pH STANDARD ->
12-31-2015 XX
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Shows SET LEVEL on middle line. When user presses any key, the live value is frozen and the user edits that value. Returns either GOOD CAL or LEVEL OFFSET TOO LARGE.
Resets calibration to factory default. After pressing , select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
Temperature
VIEW Mode Menu
CH1
0.00
CH2
113.0
9999.9
89.0
GPMS
100.00
9999.9
0.00
Displays the temperature value. This is the normal display and does not time out.
100.00
°F
°F
TEMPERATURE Setup Checklist
1. Make sure TEMPERATURE sensor type is selected. (Refer to System Setup Menu, page 11)
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (°C or °F).
4. Set Last Cal Date and initials.
5. If applicable, set up relay functions for your application.
CH1
TEMPERATURE
CH2
TEMPERATURE
34
0.00
0.00
100.00
100.00
°F
°F
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
113.0
8 9.0
GF 9950-10 Six Channel Transmitter
Page 35
Temperature
INPUT Menu
SET UP
CHANNEL 1
ENTER
CH 1 FACTORY
TYPE
FACTORY
CH1 TEMPERATURE
SOURCE
NONE
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose TEMP, press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain TEMPERATURE sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
CHANNEL 1
NAMEF
TEMPERATURE
CH1 TEMPERATURE
AV°FFE°°
TEMP UNITS
CH1 TEMPERATURE
AVERAGE
OFF
CAL Menu
CH1 TEMPERATURE
OFF
HOLD OUTPUTS
CH1 TEMPERATURE
CSET TEMP->NT
CH1 TEMPERATURE
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = TEMPERATURE
Select °F or °C. Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Provides a maximum 20 °C offset to match to a known standard (external reference).
RESET CAL->T
CH1 TEMPERATURE
LAST CAL
12-31-2015 XX
Resets Temperature Calibration to factory settings. After pressing , select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
GF 9950-10 Six Channel Transmitter
35
Page 36
4 to 20 mA
VIEW Mode Menu
CH1
50.0
CUSTOM LABEL
0.0
CH2
53.0
CUSTOM LABEL
0.0
Displays the input value. This is the normal display and does not time out.
UNIT
5.0
UNIT
5.0
4 to 20 mA Setup Checklist
1. Make sure 4 to 20 mA INPUT sensor type is selected. (Refer to System Setup Menu, page 11)
2. Set 4 mA value. (refer to your 3rd-party sensor manual)
3. Set 20 mA value. (refer to your 3rd-party sensor manual)
4. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
5. Set Last Cal Date and initials.
6. If applicable, set up relay functions for your application.
CH1
4-20 mA INPUT
CH2
4-20 mA INPUT
CH1 12.00 mAh
CH2 4.50 mAh
50.0
0.0
32.0
0.0
4-20 mA INPUT
INPUT Menu
SETUP
CHANNEL 2
ENTER
CH 2 FACTORY
TYPE
FACTORY
CH 2 4-20 mA
SOURCE
NONE
CHANNEL 2
AVNAMEFE°°
4-20mA INPUT
UNIT
UNIT
5.0
5.0
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
4 to 20 mA input live readout.
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose 4-20mA, press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain 4-20mA sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = 4-20 mA
CH2 4-20mA INPUT
SENSOR UNITS
CH2 4-20 mA INPUT
4 mA VALUE
36
Units are selectable from a list, IF the user selects CUST, the CUSTOM UNITS screen is displayed
UNIT
following this SENSOR UNITS screen. Custom units are only available for English, German, French and Spanish languages. Enter up to 4 characters describing unit of measure. Default = UNIT
0.00
GF 9950-10 Six Channel Transmitter
Measurement value of your sensor when its output is 4.00 mA.
Page 37
4 to 20 mA
INPUT Menu
CH2 4-20mA INPUT
100.00
20 mA VALUE
CH2 4-20mA INPUT
AVERAGE
COFF
CAL Menu
CH2 4-20mA INPUT
OFF
HOLD OUTPUTS
CH2 4-20mA INPUT
CSET STANDARDN-
CH2 4-20mA INPUT
SET SLOPEL
CH2 4-20mA INPUT
Measurement value of your sensor when its output is 20.00 mA.
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Applies a linear offset to the measurement. For single point calibrations, assign the average value of your process to STANDARD. For two-point calibrations, assign the min or max value of your process to STANDARD.
Applies a slope to the measurement. The SLOPE is used for two-point calibrations along with the STANDARD above. If you assigned the min value of your process to the STANDARD, then assign the max value to the SLOPE. Also, assign the min value to the SLOPE. The slope and standard values must be at least 0.1 units apart.
RESET CAL
CH2 4-20mA INPUT
LAST CALF
12-31-2015 XX
Resets Standard and Slope calibration to factory settings. After pressing , select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX).
GF 9950-10 Six Channel Transmitter
37
Page 38
Salinity
VIEW Mode Menu
Setup Checklist
1. Make sure SALINITY sensor type is selected
CH1
81.9
PPT
(Refer to System Setup Menu, page 11).
2. Set Cell Constant. Default is 10.0
0.0000
CH2
85.9 76
0.0000
Displays the value and the temperature below. This is the normal display and does not time out.
77
°F
80.000
PPT
°F
80.000
3. Set the Temperature Units (°C or °F).
4. Calibrate the Temperature element.
5. Calibrate the Conductivity sensor.
6. Set Last Cal Date and initials.
7. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
8. Select source for Relay output (SAL or TEMP).
9. If applicable, set up Relay Functions for your application.
Requires prepared calibration solution with a value as close to the applications being measured.
NOTE: The Salinity reading is calculated from the Conductivity and Temperature measurements. Salinity is only as accurate as the measurements used for calculation. Salinity can be displayed in PPT (parts per thousand) or PSU (practical salinity units). The units are equivalent, and both use the Practical Salinity Scale found in Standard Methods for the Examination of Water and Wastewater for calculation. Because they have an impact on the Salinity measurement, Conductivity and Temperature must be calibrated before Salinity.
CH1
SALINITY
0.0000
CH2
SALINITY
0.0000
CONDUCTIVITY
81.9
50.0
PPT
80.00 0
PPT
80.00 0
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
CH1 25 mS
INPUT Menu
SETUP
CHANNEL 2
ENTER
CH 2 FACTORY
TYPE
FACTORYT
CH 2 SALINITY
SOURCE
NONE
CHANNEL 2
NAME
SALINITY
Displays the equivalent conductivity value in millisiemens.
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose SALINITY press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain SALINITY sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = SALINITY
CH 2 SALINITY
CCELL CONSTANT
CH 2 SALINITY
CUST CELL CONSTT
38
10.0
50.0000
GF 9950-10 Six Channel Transmitter
Enter cell constant of sensor. Select 10.0, 1.0 or CUSTOM. Default = 10.0
Enter the precise cell constant from the certifi cate provided with your sensor or from the information label on the sensor. Shown if CELL CONSTANT = CUSTOM.
Page 39
Salinity
INPUT Menu
CH 2 SALINITY
PPT
SAL UNITS
CH 2 SALINITY
°F
TEMP UNITS
Salinity Unit: Choose units of measurement Default = PPT or select PSU
Select °F or °C. Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
CH 2 SALINITY
AVERAGE
COFF
CH 2 SALINITY
TEMP COMP
LINEARA
CH 2 SALINITY
1.91
ADJ TEMP COMP
Dampens display, output and relay response rates. Select Low, Med, High, OFF. Default = OFF
Select temperature compensation (NONE, LINEAR). Default = LINEAR
For LINEAR temperature compensation, select a % per ºC slope. Maximum slope setting is 9.99 % per ºC. (If Temperature Compensation setting is NONE, this item will not be displayed). Default = 1.91
GF 9950-10 Six Channel Transmitter
39
Page 40
Salinity
CAL Menu
CH 2 SALINITY
OFF
HOLD OUTPUTS
CH 2 SALINITY
SET
CCONDUCTIVITYS
CH 2 SALINITY
SET
TEMPERATURE
CH 2 SALINITY
SET
CSALINITYS
CH 2 SALINITY
RESET COND CAL
CH 2 SALINITY
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Manually set value to match to a known liquid standard (external reference, NIST solutions).
Manually set to apply a linear offset to the Temperature measurement element. SAVING will appear if offset is acceptable, VALUE TOO HIGH will appear if offset is outside of range.
Manually enter the known Salinity value of your specifi c Ocean. NOTE: Performing a Salinity calibration does not affect the Conductivity calibration. WARNING: Salinity calibration is not necessary if a standard conductivity calibration is performed and the customer is satisfi ed with the results.
Press to reset the Conductivity calibration to factory default.
°NF
RESET TEMP CAL
CH 2 SALINITY
°NF
RESET SAL CAL
CH 2 SALINITY
AVERLAST CALAGE
12-31-2015 XX
Press to reset Temperature calibration to factory default.
Press to reset the Salinity calibration to factory default.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
40
GF 9950-10 Six Channel Transmitter
Page 41
Dissolved Oxygen
VIEW Mode Menu
CH1
5.0
PPM
20.0
0.00
CH2
10.9
20.00
PPM
20.0
0.00
This is the normal display and does not time out.
5.0
CH1
DO
0.00
10.9
CH2
DO
0.00
CAP EXPIRATION
°NCH1 00/00/0000F
CH2 00/00/0000
PPM
20.00
PPM
20.00
INPUT Menu
20.00
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
Displays Cap Expiration Date MM-DD-YYYY. If sensor cap is missing, CH# 00/00/0000 will be displayed.
°C
°C
DISSOLVED OXYGEN Setup Checklist (3-2610-51)
2610 wiring on page 4.
1. Make sure DISSOLVED OXYGEN sensor type is selected. (Refer to System Setup Menu, page 11)
2. Set Units of Measurement (PPM, %SAT, TOR).
3. Set the Temperature Units (°C or °F).
4. Set Salinity reference value.
5. Set Barometric reference value.
6. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
7. Select source for Relay output (PPM or TEMP).
8. If applicable, set up relay functions for your application.
SETUP
CHANNEL 2
ENTER
CH 2 FACTORY
TYPE
FACTORY
CH 2 DO
SOURCE
DO
CH2 DO
AVE NAMEAGE
DO
CH2 DO
PPM
MEASUREMENT
CH2 DO
0.0
SALINITY (PSU)
Press ENTER to start assigning sensors to each channel. Use the and arrow to go to any of the six channels directly. NOTE: Display only available when the 9950 is in FACTORY mode.
Press the arrow, use the and arrow to select which TYPE of sensor will be assigned to the channel(s) (1-6). Choose Dissolved Oxygen press ENTER to save sensor type.
Press the arrow, using the or to select the input source to the channel. Sources list will only contain Dissolved Oxygen sensors or modules. To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered. Enter 17-character string. Default = DO
Set the units of measurement: PPM = DO in mg/L; % SAT = DO % saturation; TOR = Oxygen partial pressure. Default = PPM
Manually set value to match application (0 - 42 PSU). Practical Salinity Units (1 PSU = 1g/kg, or 1 PPT) (Parts per Thousand). Fresh water = 0.00 PSU. Default = 0.00
GF 9950-10 Six Channel Transmitter
41
Page 42
Dissolved Oxygen
INPUT Menu
CH 2 DO
1013.20
BAROMETRIC
CH 2 DO
°F
TEMP UNITS
Manually set Barometric value to match application altitude above or below sea level (506.62 - 1114.7 mBAR). Default = 1013.2 (sea level)
Select °F or °C. Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
CH 2 DO
AVERAGE
C OFFL CONST
CAL Menu
CH 2 DO
OFF
HOLD OUTPUTS
CH 2 DO
SET ->
C100 % SOLUTIONS
CH 2 DO
SET ->
C0 % SOLUTIONL
CH 2 DO
Dampens display, output and relay response rates. Select Low, Med, High or OFF. Default = OFF
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF. Press the arrow, using the or to select OFF/ON. Press ENTER to save. Default = OFF
Allows user to initiate the calibration process. NOTE: Dissolved Oxygen sensors are calibrated at the factory and do not require regular calibration. Press to begin the calibration process. User will be prompted to place sensor in 100% Solution standard. Press ENTER to save value and establish a calibration point. If a single point calibration is desired press the and keys to exit.
NOTE: 100% calibration must be performed fi rst to allow access to the 0% calibration screen. Dissolved Oxygen sensors are calibrated at the factory and do not require regular calibration. Press to begin the calibration process. User will be prompted to place sensor in 0% Solution standard. Press ENTER to save value and establish a calibration point.
Resets Dissolved Oxygen calibration to factory settings. After pressing , select YES/NO.
RESET DO CAL ->
42
CHANNEL 2
°NLAST CALF
12-31-2015 XX
GF 9950-10 Six Channel Transmitter
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
Page 43
Relay Mode - LOW
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
LOW
RELAY 1
LABEL
SERELAY 1C
R1 SET LOW
0.00
UNITS
R1 HYSTERSIS
0.50
UNITS
R1 ON DELAY
5.0
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
NO
RELAY 1
TEST RELAY
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(ALL) LOW set point. Relay energizes when the measurement value is equal or lower than this value. Set desired value.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(ALL) LOW set point. Relay turns on if process measurement is equal to or lower than this value. Set desired value.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or USP mode) Hysteresis delays the deactivation of the relay and prevents relay chattering around the set point. Set the Hysteresis amount (in units of measure from INPUT Mode) to the LOW or HIGH set point values.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay activating the relay once the relay condition is true.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
GF 9950-10 Six Channel Transmitter
43
Page 44
Relay Mode - HIGH
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
RELAY 1
MODE
HIGH
RELAY 1
LABEL
SERELAY 1C
R1 SET HIGH
0.00
UNITS
R1 HYSTERSIS
0.50
UNITS
R1 ON DELAY
5.0
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
NO
RELAY 1
TEST RELAY
(ALL) HIGH set point. Relay energizes when the measurement value is equal or higher than this value. Set desired value.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(ALL) HIGH set point. Relay turns on if process measurement is equal to or higher than this value. Set desired value.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or USP mode) Hysteresis delays the deactivation of the relay and prevents relay chattering around the set point. Set the Hysteresis amount (in units of measure from INPUT Mode) to the LOW or HIGH set point values.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay activating the relay once the relay condition is true.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
44
GF 9950-10 Six Channel Transmitter
Page 45
Window IN or Window OUT
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
mWIND INA
RELAY 1
LABEL
mRELAY 1A
R1 SET LOW
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(ALL) Window IN or Window OUT Wind IN, relay energizes when the measurement value is INSIDE of two set points, Low and High. WIND OUT, relay energizes when the measurement value is OUTSIDE of two set points, Low and High.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
0.0000
GPM
R1 SET HIGH
100
0GPM
R1 HYSTERESIS
0.5000
GPM
R1 ON DELAY
0.00
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
MULTNOIPLE
RELAY 1
TEST RELAY
OFF
Enter LOW limit of the Window.
Enter HIGH limit of the Window.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or USP mode) Hysteresis delays the deactivation of the relay and prevents relay chattering around the set point. Set the Hysteresis amount (in units of measure from INPUT Mode) to the LOW or HIGH set point values.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay activating the relay once the relay condition is true.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
GF 9950-10 Six Channel Transmitter
45
Page 46
Proportional Pulse
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
mPROP PULSEA
RELAY 1
LABEL
mRELAY 1AA
R1 PULSE MIN
0.0000
mLPMA
R1 PULSE MAX
100.00
LPM
R1 MAX RATE
120.00
PULSE/MIN
RELAY 1 ACTIVATE
RED BACKLIGHT
%NOE
RELAY 1
TEST RELAY
SC
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
Proportional Pulse mode varies the frequency of pulses (relay switching on and off) in direct proportion to the minimum and maximum set points.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
Set the MINIMUM set point value (pulse frequency) for Proportional Pulse. Should be set to the application's normal value.
Set the MAXIMUM set point value (pulse frequency) for Proportional Pulse. Should be set to the absolutely maximum offset allowed by the application.
(ALL except PRESSURE) Set desired maximum pulse rate (300 max) to metering device. NOTE: Pulse width xed at 100 ms.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay activating the relay once the relay condition is true.
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
46
GF 9950-10 Six Channel Transmitter
Page 47
Pulse Width Modulation
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
PWM
RELAY 1
LABEL
SERELAY 1C
R1 PWM MIN
0.0000
LPMA
R1 PWM MIN
10.000
LPM
R1 PWM PERIOD
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
PWM automatically varies the ratio of ON time to OFF time proportional to minimum and maximum range settings. The relay period is the sum of the time a relay is ON and the time it is OFF.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
Set MINIMUM value for pulse width modulation.
Set MAXIMUM value for pulse width modulation.
1.000
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
NO
RELAY 1
TEST RELAY
Set time value for one complete pulse cycle (relay ON time + relay OFF time).
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
GF 9950-10 Six Channel Transmitter
47
Page 48
Volumetric Pulse
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
VOL PULSE
RELAY 1
LABEL
RELAY 1
R1 SET VOLUME
100.00
L
R1 PULSE WIDTH
1.00
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
mNOA
RELAY 1
TEST RELAY
mA
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(FLOW) The relay will activate for the set pulse width once the specifi c volume of fl uid is registered.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
Enter volumetric set point value that triggers the relay.
Set the duration of the relay pulse in seconds. Minimum 0.1 seconds, maximum 999.9 seconds.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
48
GF 9950-10 Six Channel Transmitter
Page 49
Total Pulse
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
mTOTALA
RELAY 1
LABEL
RELAY 1
R1 SET TOTAL
100.00
LITERS
RELAY 1 ACTIVATE
RED BACKLIGHT
%NOE
RELAY 1
TEST RELAY
mA
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(FLOW) The relay will activate for the set pulse width once the specifi c volume of fl uid is registered.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(FLOW) (Shown only if TOTAL) Resettable value that, when exceeded, turns relay on. Must reset Totalizer (in VIEW Mode) to clear relay. Set maximum value.
Set the duration of the relay pulse in seconds. Minimum 0.1 seconds, maximum 999.9 seconds.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
GF 9950-10 Six Channel Transmitter
49
Page 50
United States Pharmacopeia
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
MODE
USP
RELAY 1
LABEL
RELAY 1
R1 HYSTERESIS
0.5000
GPM
R1 USP PERCENT
10.0
%
RELAY 1 ACTIVATE
RED BACKLIGHT
mNOA
RELAY 1
TEST RELAY
mA
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
USP (United States Pharmacopeia) has defi ned a set of conductivity values (limits) to be used for pharmaceutical water.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or USP mode) Hysteresis delays the deactivation of the relay and prevents relay chattering around the set point. Set the Hysteresis amount (in units of measure from INPUT Mode) to the LOW or HIGH set point values.
Enter USP set points as a percentage below the USP limit. USP alarm is always a HIGH alarm.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
50
GF 9950-10 Six Channel Transmitter
Page 51
Error Mode
RELAY Menu
RELAY 1
MODE
ERROR
RELAY 1
LABEL
SERELAY 1C
R1 ON DELAY
0.00
SEC
RELAY 1
TEST RELAY
Relay will activiate when any sensor specifi c error condition occurs. Refer to page 17 for details.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay activating the relay once the relay condition is true.
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
GF 9950-10 Six Channel Transmitter
51
Page 52
Binary Input
When the relay source is set to BINARY. The Binary Inputs can sense if the input is ON (Closed) or OFF (Open). The four binary inputs only control relay activation and are not considered a measurement source. Each binary mode is independent and each of them can be assigned to a relay mode.
Binary Input from any point level switches to the 9950 allows controlling a relay activity.
The 9950.393-3 relay module allows up to four binary (point level sensors or similar) to be used to assist in relay control of an application.
Binary Input signals can be used with the MULTIPLE source input menu in the relay menu.
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
SOURCE
BINARY 1
RELAY 1
MODE
OFF
RELAY 1
LABEL
RELAY 1
R1 ON DELAY
0.00
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
mNOA
RELAY 1
TEST RELAY
mA
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(All Binary) Binary Inputs can be set to active the relay. Requires the 3-9950.393-3 module for use.
(All Binary) Binary Inputs can be set to enable the relay based on whether the Binary input is ON (Closed) or OFF (Open). Requires the 3-9950.393-3 module for use.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay activating the relay once the relay condition is true.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
52
GF 9950-10 Six Channel Transmitter
Page 53
Timer
When the TIMER relay source is selected, the time can be selected using Period or Weekday modes. Both relay modes work directly with the time settings and formats that are entered at start up or adjusted in the options menu. The relay is normally open until the specifi ed time is reached, at which time the relay is closed, triggering the relay for a specifi ed duration. The "Hold while" feature allows the user to hold the current loop output(s) at the present value and deactivate relays that are associated with the measurement channel(s) for the specifi ed recovery time.
Period Mode lets the user set a periodic interval, in days, hours or minutes, at which the relay will activate. A First Start screen lets the user set the day and time for the Period Mode to begin.
Weekday Mode lets the user specifi c the days of the week, Sunday, Monday, Tuesday, Wednesday, etc., at which the relay will activate. A Start Time screen lets the user set the time of day for the weekday activation. Weekday Mode does not have a Start Delay parameter.
Caution: If HOLD WHILE RELAY # ACTIVE setting is changed while an Activation or Recovery is in process, the setting will be applied after the cycle is complete. To apply the change immediately, cycle power to the 9950.
Recovery
Activation
Time
Duration
Hold measurement
during activation
and recovery
Recovery
Activation
Time
Duration
Sun
Hold measurement
during activation
and recovery
Relay Period Mode
Period 1 Period 2
Activation
Recovery
Duration
Time
and recovery
Period 2
Recovery Time
and recovery
^
^
^
Period 1
^
Hold measurement
during activation
Relay Weekday Mode
Sunday, Tuesday and Thursday activation
Activation Duration
Mon Tue Wed Thu Fri
Hold measurement
during activation
Activation Duration
Activation Duration
...
Time
...
Time
Timer - Period
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
SOURCE
mTIMERA
RELAY 1
MODE
mPERIODA
RELAY 1
LABEL
RELAY 1
R1 PERIOD
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module). TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(TIMER) Allows the relay to be enabled based off either a PERIOD of time during the day or WEEKDAY to enable the relay based on a specifi c day during the week.
TIMER in PERIOD mode will allow the user to select a unit of time to enable the relay.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4,
MIN
0UNITS
R1 PERIOD
10
MIN
Select period units in MIN (minute), HOURS, or DAYS, which the relay will automatically enable.
Enter the PERIOD time which would continuously cycle ON the relay in MIN (minute), HOURS, or DAYS.
GF 9950-10 Six Channel Transmitter
53
Page 54
Timer - Period
RELAY Menu
RELAY 1
09/29/2022
START DATE
RELAY 1
11:30
MULTSTART TIMEIPLE
R1 DURATION
Enter the DATE that the timer feature should start. Caution: Verify the correct date is programmed into the 9950 during setup.
Enter the START TIME of the feature. Caution: Verify the correct time of the day is programmed into the 9950 during setup.
30
SEC
HOLD WHILE
R1 ACTIVE
NO
RECOVERY TIME
30
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
MULTNOIPLE
RELAY 1
TEST RELAY
OFF
Enter the duration (Time ON) of the relay in SEC (seconds).
Choose which channels (one or ALL) display and 4 to 20 MA outputs are frozen to prevent relays from activating while this feature is used.
Choose RECOVERY TIME in seconds for the display and 4 to 20 mA outputs to start updating after the TIMER feature has timed out.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
54
GF 9950-10 Six Channel Transmitter
Page 55
Timer - Weekday
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
SOURCE
mTIMERA
RELAY 1
MODE
mWEEKDAYA
RELAY 1
LABEL
RELAY 1
RELAY 1
SUN
9
9
WED
9
9 9
MON
THU
SAT
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(TIMER) Allows the relay to be enabled based off either a PERIOD of time during the day or WEEKDAY to enable the relay based on a specifi c day during the week.
TIMER in WEEKDAY mode will allow the user to select which day of the week wil be enabled in the relay.
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
TUE
9
FRI
9
TIMER in WEEKDAY mode will allow the user to select a day or multiple days of the week with a START TIME, DURATION and a HOLD feature. Default = 10:00, 60 secs, HOLD WHILE RELAY # ACTIVE NO
RELAY 1
11:30
MULTSTART TIMEIPLE
R1 DURATION
30
SEC
RELAY 1 ACTIVATE
RED BACKLIGHT
MULTNOIPLE
RELAY 1
TEST RELAY
OFF
Enter the START TIME of the feature. Caution: Verify the correct time of the day is programmed into the 9950 during setup.
Enter the duration (Time ON) of the relay in SEC (seconds).
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
GF 9950-10 Six Channel Transmitter
55
Page 56
Multiple
When the Relay Source is set to MULTIPLE, four Boolean logic formulas called "Relay Operators" are available. Relay operators monitor three different input sources (sensor or binary input) based on the conditions set up by the end user (HIGH or LOW conditions or binary input ON or OFF) to activate a relay.
The relay will only be activated when the complete formula is satisfi ed.
A TRUE condition means that the sensor's condition is beyond the setpoint.
A FALSE condition means that the sensor's condition is within the setpoint.
If any one of the three conditions shows CHK SENSOR then that condition will be treated as always FALSE.
Example: Set Relay 1 Mode to A | B | C. "This relay will activate IF channel 1 pH falls below 6, OR if channel 2 pH rises above 7.8, OR if binary input 1 is active (On or Closed). Relay #1 will activate 30 seconds after any one the conditions are met". Condition A: The pH on Channel 1 is less than 6 Condition B: The pH on Channel 2 is greater than 7.8 Condition C: Binary Input 1 is On On Delay: 30 When operating properly, all three of these conditions are FALSE. If any one of them becomes TRUE, then Relay 1 will be activated. Relay 1 will remain activated until the TRUE condition becomes FALSE again, including the hysteresis band.
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module. External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY Menu
RELAY 1
SOURCE
CH 1 PRIMARY
RELAY 1
MODE
OFF
RELAY 1
SOURCE
MULTIPLE
RELAY 1
MODE
A | B | C
RELAY 1
MODE
A & B & C
RELAY 1
MODE
A | (B & C)
RELAY 1
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay module).
Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4, TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW, HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
MULTIPLE will use up to 3 sources where the relay will only be activated if these sources meet specied conditions A | B | C, A & B & C, A | (B & C), A & (B | C) Default = OFF
MULTIPLE mode A | B | C will activate if ANY conditions A, B or C are true.
MULTIPLE MODE A & B & C will ONLY activate if all conditions A, B and C are true.
MULTIPLE MODE A (B & C) will activate if A is true OR if B and C are true.
56
MODE
A & (B | C)
GF 9950-10 Six Channel Transmitter
MULTIPLE MODE A (B | C) will activate if A is true AND if B or C are true.
Page 57
Multiple
RELAY 1
LABEL
RELAY 1
RELAY 1
SOURCE A
0NONE
RELAY 1
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
Select Source "A", "B", and "C". Chose any PRIMARY, SECONDARY or BINARY input.
MODE A
OFF
R1 A SET LOW
0.00
LPM
RI A SET HIGH
10.00
MULTLPMPLE
R1 HYSTERESIS
0.5000
GPM
RELAY 1 ACTIVATE
RED BACKLIGHT
NO
RELAY 1
TEST RELAY
Select relay mode, LOW or HIGH alarm.
Select LOW or HIGH alarm for ALL three Sources (A , B, and C).
Enter the set point (LOW or HIGH) for each Source (A, B, and C).
(ALL) HYSTERESIS delays the deactivation of the relay and prevents relay chattering around the set point. Set the Hysteresis amount (in units of measure from INPUT Mode) to the LOW or HIGH set point values.
Select YES to have the Red Backlight illuminate during relay activation. Select NO to not have the Red Backlight activated by the relay activation. Default backlight setting - NO
(ALL) Press or to turn relay ON or OFF for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
GF 9950-10 Six Channel Transmitter
57
Page 58
Menu System
VIEW Mode Overview
The top level of screens are referred to as the VIEW Mode. This view displays measurement values as well as current outputs, derived function values, and relay status. The horizontal bar graph represents the primary measurement value that is also displayed in the numeric fi eld above the bar graph. The bar graph is primarily used to display the full scale range of the sensor, but can be scaled via the OPTION menu item.
During normal operation, the 9950 displays the VIEW mode.
• To select a display, press the or arrow keys.
The display selections scroll in a continuous loop.
• Changing the display selection does not interrupt system operations.
• No password is necessary to change display selection.
• Output settings cannot be edited from the View Mode.
• The display will return to the VIEW mode if no button is pressed for 10 minutes.
MENU Mode Overview
The MENU mode enables the user to view and confi gure all menu items. The fi ve menus available are: CAL, INPUT, LOOP, RELAY, and OPTION.
Error Handling
Errors occurring while on the measurement screens show a specifi c message (e.g., Wrong Sensor). This message will ash and stays on for 3 seconds and fl ash off. Once the error is resolved or cleared, the error message stops.
_ _ _ _ ._
CH1
MISSING SENSOR
9999.9
00.00
CH2
6 2 .8
00.00
7 .0
mV
GPMS
00.00
pH
15.00
°F
MENU Mode is entered by pressing and holding ENTER for three seconds.
To select a menu use the and arrow keys to highlight the desired menu and press ENTER to select the menu.
In the selected menu, use the and keys to navigate through the menu. Use the , and keys to edit the selected item (Refer to Menu Navigation discussion, page 12).
To save the new selection, press the ENTER key. A message displaying “Saving” will be displayed for 3 seconds. After this message is displayed, the newly selected value will be displayed, if applicable. To abandon the changes press the and arrows simultaneously.
Password Overview
The password is often required to start editing. Once entered correctly, this password will not be needed for subsequent edits, until the menu system is exited. The password is required when the menu system is re-entered.
Your choice of password (STD or CODE) is selected in the Options Menu.
• STD
The standard (STD) password is , pressed in sequence. This password is designed to protect the 9950 from unintentional changes. It is best suited for systems
where a group of people need to be able to change settings.
• CODE
The CODE default setting is 0000, adjustable to any 4-digit numerical code up to 9999. Using a personal code provides the maximum degree of security. This code can be modifi ed in the Options menu.
Password Reset
Turn off 9950. Press and hold all 4 keys on keypad while turning on 9950 and unit will revert back to STD password.
_ _ _ _ ._
CH1
WRONG SENSOR
9999.9
00.00
CH2
6 2 .8
00.00
7.0
mV
GPMS
00.00
pH
15.00
°F
Scrolling
In some cases, more than one message or measurement may need to be displayed. This is accomplished by alternating the message portions across the screen.
PASSWORD TYPE
WRONG CODE
TEXT 3
In the MENU mode, if the wrong code or password is entered, an ERROR message is displayed.
LANGUAGE
0
0000
LANGUAGE
ENTER CODE
LANGUAGE
PASSWORD
SAVING
TEXT 3
To change your CODE, go to OPTIONS mode, enter your desired code and press ENTER. (The STD password cannot be changed).
58
GF 9950-10 Six Channel Transmitter
Page 59
Common Menus
Common Menus
The menu system shares certain modes between sensor types. The following describes the menus found in common between most sensor types.
(ALL) Manually select Sensor Type. Source to be assigned to the specifi c loop output. (Refer to page 11 for further instruction).
LOOP Menu
The following can individually be set for each loop on the 9950. Loops 1 and 2 are in the 9950 base unit. Loops MOD1 A and MOD1 B are present when a loop module is in the Module 1 Slot. Loops MOD2 A and MOD2 B are present when a loop module is in the Module 2 Slot.
NOTE: Menu and Mode displays shown are examples only. Your displays may vary.
LOOP 1
SOURCE
CH SECONDARY
LOOP 1
LABEL
LOOP 1
LOOP 1
MODE
LOG
L1 4 mA SETPOINT
0.00
UNITS
L1 20 mA SETPOINT
100.00
UNITS
LOOP 1
ERROR VALUE
22 mA
L1 ADJUST 4 mA
4.00
mA
L1 ADJUST 20 mA
21.0
mA
L1 TEST LOOP
12.05
mA
Select source for each loop. Choose PRIMARY, SECONDARY (CH 1-6) or DERIVED FUNCTION 1 - 4. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity, and Dissolved Oxygen. Derived Functions are only available for confi gured Derived Functions.
English, French, German, and Spanish languages only. A custom label can be assigned to the loop for ease of identifi cation.
(COND/RES only) Select LIN/LOG. Default = LIN Display only visible when Conductivity or Resistivity sensors are available. Refer to LOG Current LOOP Output, page 66.
(ALL) Set value corresponding to desired 4 mA output. 6 digits max. Default = 0 (ORP = -999).
(ALL) (Not shown in COND/RES LOG Mode) Set value corresponding to desired 20 mA output. 6 digits max. Defaults = 100 (Flow, Cond/Res, Temp), 14 (pH), 1000 (ORP), 10 (Lvl/Prs), 5 (4 to 20 mA), 42 (Sal).
(ALL) Set desired LOOP output value when sensor error (e.g., bad sensor, broken wire) is detected. Select (3.6 mA, 22 mA,
or NONE).
Default = 22
(ALL) Allows fi ne-tuning to compensate for errors in other equipment connected to the 9950. Adjust the current output at 4 mA. The display value represents the precise current output. Adjustment limits: from 3.80 mA minimum to 5.00 mA maximum. Default = 4.00 mA
(ALL) Allows fi ne-tuning to compensate for errors in other equipment connected to the 9900. Adjust the current output at 20 mA. The display value represents the precise current output. Adjustment limits: from 19.00 mA minimum to 21.00 mA maximum. Default = 20 mA.
(ALL) Press or to manually control the output current value from 3.8 mA to 21.00 mA to test the output of LOOP. The current loop will revert to automatic control when this menu item is left.
GF 9950-10 Six Channel Transmitter
59
Page 60
Common Menus
Derived functions are an important calculation to determine the effi ciency of a membrane in a reverse osmosis applications. Derived calculations can be assigned as a source for either a loop output or to activate a relay.
OPTIONS Menu
FUNCTION 1
(FUNCTION 1 – 4) Confi gure the derived functions, up to four functions available.
to select the correct derived function.
NONE
FUNCTION 1
LABEL
FUNC 1
F1 SUM
SOURCE A
NONE
F1 SUM
SOURCE B
NONE
F1 RECOVERY A
FEED
NONE
F1
PERMEATE
mNONEA
English, French, German, and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
Select fi rst sensor to be used in the derived function calculation, display visible only when SUM, DELTA, RATIO, % PASSAGE, and % REJECTION are selected. To change sensor selection, source A and B should be changed to NONE then the sensors can be reassigned.
Select second sensor to be used in the derived function calculation, display visible only when SUM, DELTA, RATIO, % PASSAGE, and % REJECTION are selected. To change sensor selection, source A and B should be changed to NONE then the sensors can be reassigned.
Select fi rst sensor to be used in the derived function, display only visible when % PASSAGE, % REJECT, % RECOVERY A, RECOVERY B, and RECOVERY C are selected. To change sensor selection, source A and B should be changed to NONE then the sensors can be reassigned.
Select second sensor to be used in the derived function, display only visible when % PASSAGE, % REJECT, % RECOVERY A, RECOVERY B, and RECOVERY C are selected. To change sensor selection, source A and B should be changed to NONE then the sensors can be reassigned.
60
GF 9950-10 Six Channel Transmitter
Page 61
Common Menus
OPTION Menu
ENGLISH FRANÇ
DEUTCH
ESPAÑOL
中文
TIME FORMAT
Choose desired the language for the 9950. Default = English
24 HR
BINARY ON
SET TIME
08:56
TIMER
DATE FORMAT
MM/DD/YYYY
MULTIPLE
SET DATE
09/29/2022
OFF
SET DECIMAL MARK
999.99
mA
SELECT UNITS
METRIC
mA
CONTRAST
3
mA
Choose a time format between a 24 Hour or an AM/PM clock. Default = 24 HR
Set the time according to the format chosen 00:00
Date Format; MM/DD/YYYY, DD/MM/YYYY, YYYY/MM/DD
Set the date according to the format chosen 00/00/0000
Choose either comma or point according to local practices. Default = 999.99 (point).
Choose METRIC or US CUSTOMARY
Adjust the displays contrast, 1 - 5 Default = 3
WHITE BACKLIGHT
AUTO LOW
mA
RED BACKLIGHT
ON
White Backlight; Off, Low, High, Auto Low, Auto High
The RED BACKLIGHT illuminates when an error condition is detected or user selected relay(s) are activated. Default = ON
NOTE: Defaults for most relay functions are dependent upon sensor type and are not listed here.
GF 9950-10 Six Channel Transmitter
61
Page 62
Common Menus
OPTION Menu
AUTO SCROLLING
OFF
(CH 1-6) Enable AUTO SCROLL for measurement Display screens in VIEW MODE.
SCROLL TIME
4
SEC
DISPLAY
BAR GRAPH
ON
CH2 BAR GRAPH MIN
0.00
GPM
CH2 BAR GRAPH MAX
100.00
GPM
FUNCTION 1
NONE
OFF
CH1
_ _ _ _._
GPM DECIMAL
CH1
_ _ _ _._
°F DECIMAL
CH1
_ _ _ _._
TOTAL DECIMAL
COND/RES
AUTORANGE
OFF
CH1
TOTAL LOCK
OFF
PASSWORD TYPE
STD
%
PASSWORD TYPE
CODE
%
Set the SCROLL TIME interval for each measurement display screen. Display available when scrolling is set to ON. Minimum 4 sec, maximum 99 seconds. Press any key to interrupt autoscroll for 20 seconds, or the scroll time, whichever is greater.
(CH 1-6) Display Bar Graph; ON or OFF
(CH 1-6) Enter a value to represent bar at minimum. Default = 0 (ORP = -999)
(CH 1-6) Enter a value to represent bar at maximum. Defaults = 100 (Flow, Cond/Res, Temp), 14 (pH), 1000 (ORP), 10 (Lvl/Prs), 5 (4 to 20 mA), 42 (Sal), 20 (DO)
Function 1 - 4, Confi gure the derived function, up to four functions available.
(ALL) Set the decimal to the best resolution for your application for CH 1-6. The display will automatically scale up to this resolution for each channel. Select 0, 1, 2, 3, or 4 decimal places, (varies by parameter). Default = 1 Decimal Place
(pH, COND/RES, TEMP, SAL, DO only) Set the Temperature decimal to the best resolution for your application for CH 1-6. The display will automatically scale up to this resolution. Select 0, 1, or 2. Default = 1 Decimal Place
(FLOW only) Set the decimal to the best resolution for the Permanent and Resettable 2 Totalizer display. The display will automatically scale up to this resolution. Select 0, 1, 2, 3, or 4 decimal places. Default = 1 Decimal Place
(CONDUCTIVITY/RESISTIVITY only) Set ON to automatically scale the conductivity/resistivity within range to be displayed. Default = OFF
(CH 1-6, FLOW only) requires password to reset the resettable totalizer. Select OFF, ON (Does not affect Permanent Totalizer). Default = OFF
(ALL) Select STD, CODE. Default = STD (Refer to page 12 for further detail)
(CODE) If code is selected Default Code = 0000 (Refer to page 12 for further detail)
62
GF 9950-10 Six Channel Transmitter
Page 63
Common Menus
OPTION Menu
MEMO
MEMO
SEC
MODBUS
MODULE
NONE
MODBUS
COMMUNICATION
SETTINGS ->
MODBUS
REMOTE UPDATE
NO
RESET TO
(ALL) Enter 17-character string, if desired. Default = MEMO
Select which MODULE port the Modbus Module is installed into, MOD1, MOD 2 or NONE Default = NONE.
Press to change the settings of the Modbus Module. Screen available when Modbus module is enabled. Refer to Modbus Communication Settings Menus, page 78
Setting Modbus Remote update to YES allows remote Modbus device to change the settings on the 9950. Setting Modbus Remote Update to NO prevents remote devices from make any changes to the 9950. Screen available when the Modbus module is enabled. Remote device will still be able to read all of the Modbus parameters.
FACTORY
NO
9950
GENERATION 3
SEC
USER MANUAL
100.00
SEC
VIEW Mode
50.30
CH1
40560.0
0.00
CH2
987.9
65.5
0.00
CH1
50.30
FLOW
0.00
CH2
987.9
FLOW
0.00
CH 1 TOTALIZER
P 40560.0
R 40560.0 ->
LITERS
INTERNAL LOOPS LOOP 1
LOOP 2
Master Reset - Sets the 9950 back to Factory settings.
Displays Transmitter Generation Version.
QR code for user manual.
Measurement View Mode Screens are depicted in the Sensor Specifi c Menus. Below are the screens that all parameters share in common.
GPM
R
View Measurement 1 displays the primary value with the secondary value below.
100.00
GPM
This is the normal display and does not time out.
R
100.00
GPM
100.00
View Measurement 2 displays the primary value and the channel name. This screen does not time out.
GPM
100.00
Parameter specifi c screen. Flow totalizer, pH raw mV, ORP, pH Impedance, pH/ORP Sensor Data, raw mV, 4 - 20 mA input, conductivity, DO cap expiration date.
12.00
4.00
Displays the internal 4 to 20 mA LOOP output for each channel assigned primary and secondary sources. (3.6, 22.00 are error output values).
Func 1
Func 2
Func 3
Func 4
MODULE LOOPS MOD1 A MOD1 B MOD2 A MOD2 B
50.00
100.00
2.00
0.00
18.00
12.00
8.00
4.00
Function values are stated to the right of the function number. Select a derived measurement function, up to four functions available. (SUM, DIFFERENCE, RATIO, % PASSAGE, % REJECT, % RECOVERY A, % RECOVERY B, % RECOVERY C) (Refer to page 20 for further details)
Displays the Module 4 to 20 mA LOOP output for each channel assigned primary and secondary sources. (3.6, 22.00 are error output values).
GF 9950-10 Six Channel Transmitter
63
Page 64
Common Menus
VIEW Mode
RELAY STATUS RELAY 1 RELAY 2 RELAY 3 RELAY 4
BINARY STATUS BINARY 1 BINARY 2 BINARY 3 BINARY 4
MODBUS STATUS ADDRESS BUS MSG CNT 9950 MSG CNT MSG ERR CNT
SYS CONFIG CH1 CH2 RELAY MOD1 MOD2
GENERATION 3
SW VER 9950 S/N
TEXT 1
MENU ->
TEXT 3
12/01 10:00
SUN 10:00
OFF
OFF
OFF
999 99999 99999 99999
FLOW FLOW
399504033
4-20 OUT
30-13
0123456789
ON
ON
ON
Relay status is stated to the right of the relay number as (OFF, ON, DATE/DAY and TIME) for each relay. If Relay is using Timer Source, time and date of next activation will be displayed. while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function.
The relay(s) will remain ON
The timer will restart if the condition still exists.
Binary labels can be edited by pressing on the Binary Status screen and scrolling through the Binary Label submenu. Pressing Cancel (Up and simultaneously) returns to the Binary Status screen in View Mode. English, French, German, and Spanish only. Binary status is stated to the right of the binary channel number as (OFF, ON) for each of the four binary inputs.
The Modbus Status screen is displayed if a Modbus module is installed.
SYS CONFIG shows channel and parameter and relay module part number.
Generation Screen shows version of software and serial number of unit. Note: Top shows the Generation of 9950
MENU-> will allow access to top level menu by pressing the key.
DATE
09/01/2016
09:06
EASYCAL
CH1 pH ->
CH2 ORP ->
DATE and TIME display show the current system date and time
pH EASYCAL looks for 4, 7, or 10 pH buffers. (Refer to page 71)
ORP EASYCAL looks for 87, 264, or 476 mV buffers. Press the  to begin EASYCAL. (Refer to page 73)
Important
The Relay, Mod1, and Mod2 number returned by the Sys Confi g Screen is not the Mfr. Part No. To convert from the displayed number to the GF Signet Mfr. Part No. use the chart below.
Module RELAY, MOD1, or MOD2 Mfr. Part No.
Single Channel Conductivity 399504041 399503941
Dual Channel Conductivity 399504042 399503942
Dual Channel 4 to 20 Out 39950408 399503982
Modbus Module 39950409 39950395M
Relay Four Mechanical 399504031 399503931
Relay Two SS and Two Mechanical 399504032 399503932
Relay Two Mech and Four Binary 399504033 399503933
64
GF 9950-10 Six Channel Transmitter
Page 65
Troubleshooting
Condition Possible Causes Suggested Solution
Incorrect sensor installed on channel Connect correct sensor to channel
Wrong Sensor
Sensor Type set incorrectly in 9950
Wrong Code Wrong password entered Enter correct password (Refer to page 12)
K-Factor Out Of Range K-Factors cannot be set to 0 Enter K-Factor from 0.0001 to 99999
Set correct sensor TYPE in INPUT menu (Refer to page 11)
Backlight inoperative
Relays inoperative
Relay always on
– – – – –
Missing Module
Wrong Module
No Probe
(pH/ORP only)
No Sensor
(Flow, Cond/Res, Press, Level, Temp, 4-20 mA, Sal, Batch, DO)
Check Preamp
Backlight of Screen is RED
Missing Cap
Replace Cap
Backlight turned OFF (NOTE: Backlight can turn off automatically in AUTO mode)
Set BACKLIGHT to LOW, HIGH, AUTO LOW or AUTO HIGH in OPTION menu.
Relay Module installed incorrectly Remove and reseat Relay Module
Wrong settings in RELAY menu
Use test relay to verify relay operation then check relay settings.
Hysteresis value too large Change the hysteresis value
Defective Relay Module Replace Relay or Modbus Module
Measurement exceeds display capability
Increase Flow units time base
Change unit of measure
Conductivity or Dual Channel 4 to 20 mA Module or Modbus Module is not
Install the module in the correct slot
installed in the Module Slot
The wrong Module (Conductivity or Dual Channel 4 to 20 mA or Modbus
Replace module with the correct module
Module) is installed in the slot
9950 "cannot communicate" to sensor
Missing sensor or bad temperature
• Check wiring
• Install or replace sensor
element
• Check wiring
9950 "cannot communicate" to sensor
• Install or replace sensor
9950 "cannot communicate" to the preamp
Error is detected or Relay is activating the red backlight
Dissolved Oxygen sensor is missing the sensor cap
Dissolved Oxygen sensor cap has expired
Check wiring or replace preamp
Correct error condition
Reinstall Dissolved Oxygen sensor cap
Install new Dissolved Oxygen sensor cap
GF 9950-10 Six Channel Transmitter
65
Page 66
Troubleshooting
Condition Possible Causes Suggested Solution
Broken Glass
Hi Impedance
Check Cal
(pH/ORP only)
Hi Salinity Warning
pH sensor glass has been damaged and caused very low impedance
The measured pH sensor impedance is above the high impedance level.
Visually inspect pH/ORP sensor for cracked and/or chipped glass
Visually inspect the pH electrode and clean if necessary
Electrode could be in air Ensure electrode is submersed at all times
Perform pH EasyCal (page 71)
Perform ORP EasyCal (page 73)
Slope and/or Offset are out of range
Set pH Slope or Standard (page 71)
Set ORP Slope or Standard (page 73)
Reset pH CAL (page 71)
Reset ORP CAL page 73)
Clean electrode and recalibrate
Calibration Value entered is out of the range allowed
Enter a lower offset value
Replace sensor
Temperature Warning
Temperature Value entered is out of the range allowed
Enter a lower offset value
Replace sensor
66
GF 9950-10 Six Channel Transmitter
Page 67
Averaging
NO AVERAGING, NO SENSITIVITY
With SENSITIVITY set to a large value and AVERAGING set to OFF (0 seconds), the 9950 responds immediately to every shift in the process. The dashed red line represents the actual output of the sensor in varying conditions.
AVERAGING ONLY
With SENSITIVITY still set to a large value and AVERAGING set to MED or HIGH the rate is stabilized, but a sharp change in rate is not represented for 8 to 32 seconds or longer.
AVERAGING AND SENSITIVITY
With SENSITIVITY at 50 and AVERAGING set to MED or HIGH, the rate is stabilized, while a sudden shift in fl ow rate exceeding 50 units of measure will be displayed immediately.
NOTE: The SENSITIVITY function applies only to FLOW. The SENSITIVITY function has no effect if the AVERAGING function is set to OFF. Averaging will not function if the Sensitivity is set to 0 (Zero).
Important
With Averaging set to Off the Sensitivity function is disabled. With Sensitivity set to 0 (zero) the Averaging function is disabled.
To set the unit to Averaging with no Sensitivity set Sensitivity to a large value, 99999.
5 s0 s 10 s 15 s 20 s 25 s 30 s 35 s
Averaging is different depending on the measurement type. Seconds to 99.5% of Final Value for Low, Med, and High are:
Sensor Type Low Medium High
Flow 10 40 120
pH 2 4 12
ORP 2 4 12
Cond/Res 4 6 12
Pressure 4 10 30
Level/Volume 4 10 30
Temperature 3 10 30
4 to 20 mA 4 10 30
Salinity 4 6 12
LOG Current Loop Output
In Conductivity/Resistivity, the logarithmic (LOG) mode can be used when a very large measurement range is required, yet high resolution is needed at the low end (e.g. in a clean-in-place application where a high-resolution conductivity reading is needed at the low end, while a very high conductivity reading is needed when a cleaning cycle is in progress).
Only two parameters need to be set up, the starting or base conductivity value (4 mA SETPOINT) and the ending or maximum conductivity value (20 mA SETPOINT). The 4 mA setpoint may be larger than the 20 mA point (reverse span).
What equation should be put in the PLC?
Conductivity = 10
n = (mA input – 4.0)
If only fi xed thresholds are required, they can be calculated in mA. Then the mA value can be checked directly.
Inside the 9950 the following equation is used:
(Log
mA =
NOTE:
If ADJUST 4 mA or ADJUST 20 mA is used, the mA value can be affected. To prevent any problems the adjust function should only be used to get exactly 4.0 and 20.0 at the PLC. The 9950 is accurate and the adjust functions are only needed to compensate for an offset due to noise or a not-so-accurate PLC input card.
The error value of either 3.6 mA or 22 mA should be tested fi rst before applying the conductivity equation.
10
n
×
Conductivity – Log
(Log10 20 mA setpnt – Log10 4 mA setpnt)
16 mA
4 mA setpnt)
10
×
(Log10 20 mA setpnt – Log10 4 mA setpnt)
+ Log10 4 mA setpnt
16
+ 4
GF 9950-10 Six Channel Transmitter
67
Page 68
Custom Measurements
Z
For most vessels, the zero reference point (Z) may be designated as any height in the vessel.
For horizontal cylinders only, the zero reference point MUST be the lowest point in the vessel.
L > 0
S
Loc
O(ffset) > 0
Z(ero)
L > 0
O(ffset) < 0
Z(ero)
S
Loc
L < 0
Z
Z(ero)
O(ffset) < 0
S
Loc
Defi ning a Custom Tank
1. Determine where the level measurement should start. This is the zero reference point (Z). Review the diagram to help select the best option.
2. Determine where you will mount the sensor. This is S
. Consult the Sensor manual for
LOC
information regarding the best location for the sensor.
3. Measure the distance between Z and S
LOC
.
This is O(ffset).
4. Enter the Offset into the INPUT Mode menu.
Zero reference point (Z):
The point in the vessel where you want the 9950 to display zero (0 ft, 0 gal. etc.).
• If Z is located below the fl uid surface, the 9950 will display a positive level measurement.
• If Z is located above the fl uid surface, the 9950 will display a negative level measurement.
Sensor Location point (S
LOC
):
The point on the level sensor where the measurement is taken.
• The pressure sensor measures from the center line of the diaphragm.
Offset (O):
The distance from Z to S
LOC
.
• Enter a positive value in the Calibrate menu if the sensor is located above Z.
• Enter a negative value in the Calibrate menu if the sensor is located below Z.
• Enter 0 in the Calibrate menu if the sensor is located at Z.
68
Level (L):
The distance from Z to surface of fl uid (displayed as “Level” by 9950).
GF 9950-10 Six Channel Transmitter
Page 69
Custom Measurements
Level and Volume Calculation in Custom Shaped Vessels
In the LEVEL/VOLUME menu, if Custom Shape is selected in the INPUT menu, you can defi ne from three to ten Custom Points to establish the relationship of level to volume in the vessel.
• Select Manual Level Measurement mode to edit both level and volume data (dry confi guration).
• Select Automatic Level Measurement mode to accept the sensor measurement of the Level, while you assign a volumetric value to each custom point (wet confi guration).
• Enter from 4 to 32 custom points to link level and volume values.
• The fi rst custom point must be the lowest fl uid level in the vessel. Each successive point must be greater than the preceding point.
• The last point must be equal to or greater than the highest fl uid level in the vessel.
• A custom point should be located at all transition points in the vessel shape (for example, at custom point #9, where the shape changes from a cylinder to a cone).
• The more complex sections should be defi ned with more points. NOTE: The conical section of the illustration has been defi ned by custom points 1 through 9. More complex tanks will require additional points to ensure an accurate calculation.
• Simpler sections require fewer defi ning points. NOTE: A cylinder requires only custom points 9 and 10.
Custom point #10
Custom point #9
Custom point #8
Custom point #7
Custom point #6
Custom point #5
Custom point #4
Custom point #3
Custom point #2
Custom point #1
3000 gal.
1300 gal.
1000 gal.
700 gal.
380 gal.
250 gal.
160 gal.
118 gal.
45 gal.
0 gal.
GF 9950-10 Six Channel Transmitter
69
Page 70
Custom Measurements
In the LEVEL/VOLUME INPUT menu (Refer to page 31), if SHAPE is set to HORIZ CYLINDER, RECTANGLE or CUSTOM, the tank shape can be defi ned with the following screens:
CH 2 LEVEL
4
NUM CUST POINTS
CH1 LEVEL
AUTO
LEVEL CALC
FT
0.00
POINT 1 LEVEL
GAL
0.00
POINT 1 VOL
FT
0.00
POINT X LEVEL
If Custom shape is selected, enter the number of measurement points to be used to defi ne the vessel shape (Refer to Level and Volume Calculation in Custom Shaped Vessels, page 68). Minimum 4 points, maximum 32 points. A larger number of points improves accuracy.
Select (AUTO, MAN). Manual allows you to edit both the Level and the corresponding Volume for your custom tank. Automatic allows you to edit the Volume measurement (while displaying an automatically calculated Level value). See example below.
Enter the Level (if MAN measurement is selected) at each custom point in your vessel. If AUTO is selected, indication will read actual tank level in LEVEL UNITS at that point in your tank.
Set the Volume (if MAN measurement is selected) at each custom point in your vessel.
Where (X) is number of custom points used to calculate level.
FT
0.00
POINT X VOL
Where (X) is number of custom points used to calculate volume.
To set AUTO LEVEL MEAS value:
1. Add a known quantity of fl uid into a tank.
2. POINT 1 LEVEL indicates actual tank level.
3. Press  for POINT 1 VOL. Press  to enter quantity of fl uid (in VOLUME UNITS) added to the tank in step 1.
Press ENTER.
4. Repeat for each point set in NUM CUST PNTS. For example, in a 25-gallon conical tank set for three custom points:
1. Add 10 gallons of fl uid into the conical tank. POINT 1 LEVEL will indicate actual tank level.
2. In POINT 1 VOL, enter 10.
3. Add another 10 gallons into the tank. POINT 2 LEVEL will indicate actual tank level.
4. In POINT 2 VOL, enter 10.
5. Add the fi nal 5 gallons into the tank. POINT 3 LEVEL will indicate actual tank level.
6. In POINT 3 VOL, enter 5.
70
GF 9950-10 Six Channel Transmitter
Page 71
()
⎥ ⎦
⎢ ⎣
×⎟⎟
⎜ ⎝
⎟ ⎠
⎜ ⎝
××=
2cos
12
hrhhr
r
hr
rA
2
Custom Measurements
Technical Reference for Level, Volume, and Mass Measurement
The 9950 can automatically perform level, volume and mass calculations:
• Pressure-to-level
• Mass
• Volume
Pressure to level conversion:
Level = P ÷ (SG × D) where P = Pressure SG = Specifi c Gravity of fl uid
D = Density of water
With pressure in psi:
Level (meters) = 0.703069 × (P/SG)
With pressure in bar:
Level (meters) = 1.019715 × (P/SG)
Mass Conversion
m = D × SG × V
A
where m = mass of fl uid D = density of water = 1000 kg/m
3
SG = Specifi c Gravity of fl uid V = Volume of fl uid (m3)
m (kg) = 1000 × SG × V
Volume Calculations
Vertical cylinder:
r
V = π × r
where r = radius of cylinder
2
× h
h = height of fl uid
h
Rectangular vessel:
V = w × l × h
where w = width l = length h = height
Horizontal cylinder:
V = A × L
where A = area of segment L = length of vessel
Length
where r = radius of vessel
h = height of segment
GF 9950-10 Six Channel Transmitter
71
Page 72
Calibration Procedures - pH
EasyCal Procedure - pH
This procedure simplifi es pH calibration using standard 4.0, 7.0, 10.0 pH buffers only. If these pH buffers are not available, use MANUAL CAL and calibrate the system using the STANDARD and SLOPE settings.
Set sensor temperature in the CAL Mode before performing EasyCal for new electrode installations.
+
From any VIEW screen, press
(Press again to access CH1 if necessary)
EASYCAL
1.
2.
CH 1 pH ->
CH 2 pH ->
PLACE SENSOR
IN BUFFER 1
START <ENTER>
Place electrode tip in first pH buffer. pH 4.0 = 177 mV pH 7.0 = 0 mV pH 10 = -177 mV
1
Limit ± 50 mV
then press .
ENTER
Allow for stabilization
_ _ _ _
ENTER CODE
CH1 pH
ENTER
7.00
EASY 13.10 mV
30s
30 seconds
4.
3.
PLACE SENSOR
IN BUFFER 2
START <ENTER>
Place electrode tip in second pH buffer.
2
EASYCAL
SAVING
CH 1 pH ->
CH 2 pH ->
+
Display returns to VIEW Mode
CH1
-0.99
CH2
-0.99
ENTER
EASY 184.33 mV
Allow for stabilization
EASYCAL is complete
7.0
77.0
4.0
99.0
CH1 pH
4.00
30s
30 seconds
EASYCAL
CH 1 pH ->
CH 2 pH ->
pH
°F
15.00
pH
°F
15.00
ENTER
NOTE: The solutions can be used for calibrating more than one sensor; however,
the solution must remain free of debris and must not be diluted by rinse water from previous calibrations.
72
GF 9950-10 Six Channel Transmitter
Page 73
Calibration Procedures - pH
Manual Calibration Procedure - pH
Requires prepared buffers. System calibration is possible with any known pH solutions within 0 to 14 pH range.
Select pH values that are proximate to the application range and at least 2 pH units apart.
Navigate to the CAL Menu
Set HOLD OUTPUTS if necessary
Select CAL AT INSTRUMENT
1.
2.
CH1 pH ->
CH 2 pH ->
To Set Standard:
CH1 pH
SET
pH STANDARD ->
HOLD OUTPUTS
ENTER CODE
Place electrode tip in pH buffer.
CH1 pH
OFF
_ _ _ _
CH1 pH
CAL
AT INSTRUMENT
RAW 13.25 mV
Allow for stabilization 30 seconds to several minutes
CH1 pH
7.13
30s
CH1 pH
EASYCAL ->
CH1 pH
SET
TEMPERATURE ->
To Change Reading:
+07.1
AM/PM
RAW 13.25 mV
3
7.13
to accept
ENTER
pH STANDARD ->
CH1 pH
CH 1 pH ->SAVING
pH STANDARD ->
Theoretical mV values
pH @ 25 °C mV
2 +296
3 +237
4 +177
5+118
6 +59
7+0
8 -59
9-118
10 -177
11 -237
12 -296
CH1 pH
SET
3.
4.
To Set Slope:
CH1 pH
SET
pH SLOPE ->
Place electrode tip in pH buffer two pH units different than standard.
To Set Calibration Date:
CH1 pH
SET
RESET pH CAL ->
CH1 pH
7.13
RAW 13.25 mV
Allow for stabilization 30 seconds to several minutes
CH1 pH
100.0
RESET TEMPCAL ->
+
x4 Display returns to VIEW Mode
30s
CH1 pH
100LAST CAL.00
04-20-2017 XX
To Change Reading:
+07.1
3
AM/PM
pH
13.25
CH1
-0.99
CH2
-0.99
to accept
ENTER
To Change Reading:
04-20-2017 XX
LAST CAL
04-20-2017 XX
7.0
77.0
4.0
99.0
CH1 pH
CH 1 pH ->SAVING
pH SLOPE ->
to accept
ENTER
pH
°F
15.00
pH
°F
15.00
CH1 pH
SET
pH SLOPE ->
CH1 pH
CH 1 pH ->SAVING
05-15-2017 XX
Single-point calibration sets STANDARD only; GF recommends a two-point calibration to set SLOPE in addition to STANDARD.
Quick Manual Calibration Procedures:
1-Point Calibration:
1. Set solution standard.
2-Point Calibration
(recommended):
1. Set solution standard.
2. Set solution slope.
CH1 pH
100LAST CAL.00
05-15-2017 XX
GF 9950-10 Six Channel Transmitter
73
Page 74
Calibration Procedures - ORP
EasyCal Procedure - ORP (one-point calibration)
EasyCal is the fastest and simplest periodic ORP calibration method using any one of three specifi c buffer solutions. system using the STANDARD and SLOPE settings.
Saturate 50 mL of pH 4 (87 mV) or pH 7 (264 mV) buffers with 1/8 g quinhydrone.
Premixed Light's Solution (476 mV) can be used instead of pH buffers with quinhydrone.
Set sensor temperature in the CAL Mode before performing EasyCal for new electrode installations.
1. From any VIEW screen, press
2. Press . (Press again to access CH1 if necessary)
To Calibrate: To Accept:Response:
86.2
AM/PM
Place electrode tip in saturated pH 7.0 buffer pH 7.0 = 87 mV
If the required buffers are not available, use MANUAL CAL and calibrate the
+
CH 1 ORP
SAVING
EASY 87.0 mV
87.0
EASY 87.0mV
PLACE SENSOR
IN BUFFER
START <ENTER>
30s
Allow for stabilization
1
30 seconds
ENTER
to accept
NOTE: ORP solutions made with quinhydrone are very unstable and may not read properly once exposed to air for a
prolonged time. These solutions must be disposed of within an hour.
The solution can be used for calibrating more than one sensor. However, the solution must remain free of debris and must not be diluted by rinse water from previous calibrations.
Acceptable ranges for the readings are ± 80 mV (i.e., 87 ± 80 mV).
To exit menus and return to
9,(:SUHVVŸDQGź
buttons at the same time
+
74
GF 9950-10 Six Channel Transmitter
Page 75
Calibration Procedures - ORP
Manual Calibration Procedure - ORP
Requires prepared buffers and a prepared quinhydrone solution: Saturate 50 mL of pH 4 and 7 buffers with 1/8 g quinhydrone. (System calibration is possible with two known ORP solutions, but use a buffer close to your own process value).
1.
2.
CH1 ORP ->
CH 2 ORP ->
To Set Standard:
CH1 ORP
SET
ORP STANDARD ->
Place electrode tip in saturated pH 7.0 buffer. pH 7.0 = 87 mV
To Set Slope:
3.
CH1 ORP
SET
ORP SLOPE ->
Place electrode tip in saturated pH buffer two pH units different than standard. pH 4.0 = 264 mV
CH1 ORP
OFF
HOLD OUTPUTS
_ _ _ _
ENTER CODE
CH1 ORP
7.13
RAW 13.25 mV
CH1 ORP
CAL
AT INSTRUMENT
CH1 ORP
EASYCAL ->
CH1 ORP
7.13
RAW 13.25 mV
30s
Allow for stabilization 30 seconds to several minutes
To Change Reading:
+07.1
3
AM/PM
257.49
13.25
30s
Allow for stabilization 30 seconds to several minutes
To Change Reading:
+76.1
3
AM/PM
76.13
RAW 73.25 mV
to accept
CH1 ORP
ENTER
CH 1 pH ->
ORP SLOPE ->
to accept
ENTER
SAVING
CH1 ORP
CH 1 pH ->SAVING
ORP STANDARD ->
CH1 ORP
SET
ORP SLOPE ->
Single-point calibration sets STANDARD only.
GF recommends a two-point calibration to set SLOPE in addition to STANDARD.
Quick Manual Calibration Procedures:
1-Point Calibration:
1. Set solution standard.
2-Point Calibration :
1. Set solution standard.
2. Set solution slope.
CH1 ORP
SET
ORP STANDARD ->
4.
To Set Calibration Date:
CH1 ORP
SET
RESET ORP CAL ->
CH1 ORP
100LAST CAL.00
04-20-2017 XX
+
x4 Display returns to VIEW Mode
To Change Reading:
04-20-2017 XX
LAST CAL
04-20-2017 XX
700.0
00.00
120.0
00.00
to accept
ENTER
CH1 ORP
SAVING
CH 1 pH ->
05-15-2017 XX
mV
00.00
mV
00.00
CH1 ORP
100LAST CAL.00
05-15-2017 XX
Acceptable ranges for the readings are ± 80 mV (i.e., 87 ± 80 mV).
NOTE: ORP solutions made with quinhydrone are very unstable and may not read properly once exposed to air for a prolonged time. These solutions must be disposed of within an hour.
The solution can be used for calibrating more than one sensor. However, the solution must remain free of debris and must not be diluted by rinse water from previous calibrations.
GF 9950-10 Six Channel Transmitter
75
Page 76
Calibration Procedures - Conductivity / Resistivity
Calibration Procedure - Conductivity/Resistivity
AutoCal is the fastest and simplest periodic calibration method. Requires prepared buffer of a value appropriate to your process.
AutoCal Procedure
AutoCal is a single-point calibration system. During this procedure, if the measured
value is within ± 10% of any of the test values listed below, the 9950 will automatically recognize the test value and calibrate the output to that value.
NOTE: The fi rst step (Reset) is recommended each time an electrode is replaced, but is
NOT necessary upon initial installation or periodic calibration.
NOTE: Ensure that the buffer solution is within ± 5 °C of 25 °C.
1. Reset the sensor to factory calibration. (refer to sensor manual for procedure)
2. On the 9950, select AUTO CAL from the CAL menu. Press .
3. Place the electrode/sensor assembly into the conductivity test solution appropriate to your operating range. Shake the electrode to dislodge any air bubbles visible on the surface of the electrode.
4. Allow at least 2 minutes for the electrode response to stabilize.
5. When the display stabilizes, press ENTER.
6. If calibration is successful, 9950 will display "SAVING". If error is too large, "OUT OF RANGE USE MANUAL CALIBRATION" will display.
Calibration is complete. Return the system to service.
Manual Cal Procedure
NOTE: The rst step (Reset) is recommended each time an electrode is replaced,
but is NOT necessary upon initial installation or periodic calibration.
NOTE: Ensure that the buffer solution is within ± 5 °C of 25 °C.
1. Reset the sensor to factory calibration. (refer to sensor manual for procedure)
2. On the 9950, select MANUAL CAL from the CAL menu. Press .
3. Place the electrode/sensor assembly into the conductivity test solution appropriate to your operating range. Shake the electrode to dislodge any air bubbles visible on the surface of the electrode.
4. Allow at least 2 minutes for the electrode response to stabilize.
5. When the display stabilizes, press the key, and enter the value of the buffer solution using the , and buttons.
6. Press ENTER.
7. 9950 will display "SAVING". If error is too large, "ERR TOO LARGE TO CALIBRATE" will display.
Conductivity units are displayed as selected in the CALIBRATE menu. Resistivity displayed when K or M ranges are selected.
Available buffer values are:
 10  100  146.93  200  500  1000  1408.8  5000  10,000  12856  50,000  100,000
(all values in S)
Refer to page 77 to troubleshoot Calibration Errors
WARNING: Calibration standards are very unstable below 50S. For resistivity applications it is recommended to use a NIST sensor and enter the sensor data into the Custom Cell menu feature.
Calibration is complete. Return the system to service.
76
GF 9950-10 Six Channel Transmitter
Page 77
Calibration Procedures - Salinity
NOTE: Ensure that the buffer solution is within ± 5 °C of 25 °C.
1. Reset the sensor Conductivity and Temperature calibration to factory calibration (refer to sensor manual for procedure).
2. On the initial commissioning, use an external thermometer, sample the water temperature and calibrate the temperature element (temperature only required on the initial startup of the sensor).
3. Place the electrode/sensor assembly into the conductivity test solution appropriate to your operating range. Shake the electrode to dislodge any air bubbles visible on the surface of the electrode.
4. Allow at least 2 minutes for the electrode response to stabilize.
5. When the display stabilizes, press the key, and enter the value of the buffer solution using the , and buttons.
6. Press ENTER.
7. 9950 will display "SAVING". If value is outside of range error VALUE TOO HIGH or VALUE TOO LOW will display.
Calibration is complete. Return the system to service.
GF 9950-10 Six Channel Transmitter
77
Page 78
Calibration Procedures - Flow
Calibration Procedure - Flow
Select RATE CALIBRATION to match the dynamic fl ow rate to an external reference. Entering a rate will modify the existing K-Factor.
Select VOLUME CALIBRATION if the fl ow rate can be determined by fi lling a vessel of known volume. The 9950 will count the number of pulses generated as the known volume of fl uid passes through the sensor, and then use the information to calculate a new K-Factor.
Rate Calibration Procedure
1. Press the  to enter the rate calibration.
2. Allow the fl ow to stabilize and then press the .
3. Using the , , and keys, set the fl ow rate in the text box to match the fl ow rate of the system.
4. 4. Press
5. The 9950 displays the newly calculated K-Factor for your reference. (If the calculated K-Factor is less than 0.0001 or greater than 999999 (out of range at either extreme), the 9950 displays "ERROR NEW KF OUT OF RANGE" and returns to RATE CAL. If fl ow is too low to accurately calibrate, the 9950 displays "ERROR FLOW RATE TOO LOW and returns to RATE CAL.
6. Press ENTER to accept the new K-Factor (9950 displays "SAVING") or press + keys simultaneously to escape without saving and return to Enter Volume. NOTE: You may enter your own calculated K-Factor in the CAL menu.
ENTER
to complete the fl ow rate calculation.
CH 1 Flow
RATE CAL ->
FLOW UNITS
GPM
SET FLOW
50.00
0
0050.00
AM/PM
SET FLOW
50.00
Volume Calibration Procedure
1. Press ENTER to start the volumetric calibration period. The 9950 starts counting pulses from the fl ow sensor.
2. Press ENTER to stop the volumetric calibration period. The 9950 stops counting pulses from the fl ow sensor.
3. Enter the volume of fl uid known to have fl owed past the sensor during the volumetric calibration period. This will modify the existing Flow K-Factor.
4. The 9950 displays the newly calculated K-Factor for your reference. (If the calculated K-Factor is less than 0.0001 or greater than 999999 (out of range at either extreme), the 9950 displays "ERROR VOLUME TOO HIGH" (or LOW) and returns to VOLUME CAL).
5. Press ENTER to accept the new K-Factor (9950 displays "SAVING") or press + keys simultaneously to escape without saving and return to Enter Volume. NOTE: You may enter your own calculated K-Factor in the CAL menu.
CH 1 Flow
VOLUME CAL ->
CH 1
START <ENTER>
CH 1
RUN
0
0050.00
AM/PM
ENTER VOLUME
FLOW UNITS
Refer to page 77 to troubleshoot Calibration Errors
78
GF 9950-10 Six Channel Transmitter
STOP <ENTER>
50.00
Page 79
Calibration Error Messages
Message Cause Solution
(
Cond/Res) Error > 10% in AutoCal
Out Of Range Use Manual Calibration
Err Too Large To Calibrate
Error Volume Too Low User-entered volume too small to calibrate
Error New KF Out Of Range The calculated K-Factor too low or high
Error Flow Rate Too Low
Cal Error Out Of Range
Slope Too Close To Standard
Standard Too Close To Slope
Level Offset Too Large Offset must be < 1.0 meter
(pH) Buffer not found; Error > ±1.5 pH units
(ORP) No quinhydrone in buffer Error greater than ±80 mV
(
Cond/Res) Manual Cal when error > 100%
(pH) Offset > 1.3 pH units; Slope error > 100%
(Press) Slope must be < ±50% or offset must be < 2.75 PSI or equivalent
(Rate Cal) Flow too low to accurately calibrate
(4 to 20 mA) Slope error > 100 Check input at 4 mA and 20 mA settings
(Temp) Offset must be < ±20 °C or equivalent
(4 to 20 mA) Difference in calibration values must be > 0.1 units
(pH) Difference in calibration values must be > 2 pH units
(ORP) Difference in calibration values must be > 30 mV
(4 to 20 mA) Difference in calibration values must be > 0.1 units
(pH) Difference in calibration values must be > 2 pH units
(ORP) Difference in calibration values must be > 30 mV
Use manual calibration method
• Use 4, & 7 pH buffers (with quinhydrone for ORP calibration) or Light's Solution
• Clean sensor and retry EasyCal
• Use manual calibration method
• Inspect sensor and wiring for damage
• Clean sensor
• Check reference
• Clean sensor
• Replace sensor
• Correct volume entry
• Use longer calibration period
• Verify volume or rate entered
• Verify fl ow is present
Increase fl ow
• Check sensor range
• Check reference
• Replace sensor
• Check sensor
• Use fresh buffer
• Use two different buffer values
• Clean sensor
• Clean sensor
• Use fresh 4, 7, 10 pH buffers
• Use two different buffer values
• Decrease offset
• Replace sensor
Pressure Too High
Pressure Too Close To Zero
Value too High (Salinity)
Value too Low (Salinity)
Pressure must be lower than 2.5 PSI or equivalent to do zero cal
Pressure must be higher than 3 PSI or equivalent to do slope calibration
Salinity value outside of measurement valid range 0 to 42 PPT
Temperature value outside of measurement valid range and -2 to 35 C°
Decrease pressure
• Increase pressure
• Check reference
• Verify application range does not exceed 42 ppt
• Remove, clean and calibrate sensor
• Verify correct cell constant was selected
GF 9950-10 Six Channel Transmitter
79
Page 80
USP Limits
USP (United States Pharmacopeia) has defi ned a set of conductivity values (limits) to be used for pharmaceutical water. The standard requires that conductivity measurement without temperature compensation be used for these applications. The limits vary according to the temperature of the sample. The 9950 has the USP limits stored in memory. It will automatically determine the proper USP limit based on the measured temperature. Important Conductivity TEMP COMP must be set to NONE and Unit Of Measure must be set to S.
Using the USP function
USP setpoints are defi ned as a percentage below the USP limit, so a USP alarm is always a HIGH alarm. The 9950 can be set to warn you if the conductivity approaches within a set percentage of the USP limit.
The following settings and conditions are required for a USP relay function:
1. In the RELAY menu:
• RELAY MODE must be set to USP.
2. In the INPUT menu:
• COND UNITS must be set to μS.
• TEMP COMP must be set to None.
Example:
• The water temperature is 19 ºC, so the USP limit is 1.0 S.
• The USP PERCENT is set to 40%.
• The relay will be activated when the conductivity value reaches 40% below the
1.0 USP limit, or 0.6 S.
• If the water temperature drifts to more than 20 ºC, the 9950 will automatically adjust the USP limit to 1.1. The relay will now be activated when the conductivity value reaches 40% below 1.1 S (0.66 S).
Temperature
Range (ºC)
0 to < 5 0.6
5 to < 10 0.8
10 to < 15 0.9
15 to < 20 1.0
20 to < 25 1.1
25 to < 30 1.3
30 to < 35 1.4
35 to < 40 1.5
40 to < 45 1.7
45 to < 50 1.8
50 to < 55 1.9
55 to < 60 2.1
60 to < 65 2.2
65 to < 70 2.4
70 to < 75 2.5
75 to < 80 2.7
80 to < 85 2.7
85 to < 90 2.7
90 to < 95 2.7
95 to < 100 2.9
100 to < 105 3.1
USP limit
(μS)
NOTE: Relay will activate if USP limit is exceeded, Temperature Comp is set
to Linear or Pure Water, Conductivity Measurement is NOT in uS, or if the Conductivity Sensor Electronics is indicating a measurement error.
Modbus
The 9950 Modbus Module allows the 9950 to be connected to a serial Modbus RS485 network. The Modbus Module will allow for PLCs and Scada Systems to query the current readings, channel status, derived function values, state of the relay and 4 to 20 mA outputs, along with status of the 9950 and Modbus communications. Detail programming information is available in the 9950 Modbus Programming Manual.
Default Communication Parameters ModBus Address 95 ModBus Termination Off ModBus Mode RTU ModBus Baud Rate 19.2K ModBus Parity Even ModBus Endian Big
80
GF 9950-10 Six Channel Transmitter
Page 81
Modbus Menus
OPTION Menu
MODBUS
MODULE
NONE
MODBUS
Select which MODULE port the Modbus module is installed into, MOD1, MOD 2 or NONE Default = NONE.
COMMUNICATION
SETTINGS ->
Press the Arrow key to change the settings of the Modbus Module
Modbus Communication Settings Menus
MODBUS
ADDRESS
95
NETWORK
Set the Modbus address of the 9950. The default address is 95
The 9950 Modbus Module has an electronic termination network.
TERMINATION
NO
MODBUS
BAUD RATE
19.2 K
MODBUS
PARITY
EVEN
MODBUS
ENDIAN
BIG
The Termination can be activated by setting the value to Yes. The default value is No (the termination is off).
Set the baud rate of the Modbus network. The default value is 19.2K baud, 19200 baud.
Set the parity of the Modbus network. The default value is Even parity.
Set the byte order in which the 9950 transmits fl oating point numbers in the Modbus protocol. The default value is Big Endian.
RESET TO
DEFAULT
NO
ACTIVATE
SETTINGS
NO
Reset all Modbus setting to the default values.
Activate setting send all the change to the default parameter to the Modbus module. If you do not activate the setting the confi gured setting will not take effect until after the 9950 is power cycled.
Maintenance
Clean the instrument case and front panel with a soft cotton cloth dampened with a mild liquid soap solution.
Never wipe the front window with static retentive cloths such as wool or polyester which may induce a static charge. If a static charge develops on the window, you may notice temporary blotches form on the screen. When this occurs, clean the front window with an anti-static cloth, or a soft cotton cloth and anti-static spray or a mild liquid soap solution to remove the static charge.
GF 9950-10 Six Channel Transmitter
81
Page 82
Specifi cations
General
Input channels ............Six channels, programmable for
Digital (S
3
L) or frequency input
Outputs.......................Two 4 to 20 mA output in base unit,
Two or four 4 to 20 mA outputs in optional modules,
Two or four programmable relay outputs optional
Enclosure and Display
Case Material .............PBT
Window ....................... Shatter-resistant glass
Keypad ....................... 4 buttons, injection-molded
silicone rubber seal
Display ........................ Dot Matrix LCD
Update Rate ...............1 s
LCD Contrast .............. 5 settings
Indicators .................... Two horizontal digital bar graphs,
four LED relay status indicators
Enclosure
Size .....................¼ DIN
Mounting
Panel ................... ¼ DIN, ribbed on four sides for panel
mounting clip inside panel, silicon gasket included
Wall ...................... Large enclosure (sold as an accessory)
Terminal Blocks
Pluggable screw type: use minimum 105 °C rated wire Torque ratings
Power/Loop..........0.49 Nm (4.4 lb-in.)
Freq/S
3
L ...............0.49 Nm (4.4 lb-in.)
Relay....................0.49 Nm (4.4 lb-in.)
Connector wire gauge:
Power, Loop .........12 to 28 AWG
Freq/S
3
L ...............16 to 28 AWG
Module connector wire gauge:
Relay....................12 to 28 AWG
Environmental
Ambient Temperature:
DC Power ...............-10 °C to 70 °C (14 °F to 158 °F)
AC Power ...............-10 °C to 60 °C (14 °F to 140 °F)
Storage Temp .........-15 °C to 70 °C (5 °F to 158 °F)
Relative Humidity........ 0 to 100% condensing for (front only);
0 to 95% non-condensing (rear panel)
Maximum Altitude .......4,000 m (13,123 ft)
Enclosure Rating .......... NEMA 4X/IP65 (front face only)
Installation Category Cat II Pollution Degree 2
Performance Specifi cations
System Accuracy
• Primarily dependent upon the sensor.
System Response
• Primarily dependent upon the sensor. Transmitter adds a maximum of 150 ms processing delay to the sensor electronics.
• Minimum update period is 100 ms.
• System response is tempered by the display rate, output averaging and sensitivity feature.
Electrical Requirements
Power to Sensors
Voltage...............................
+4.9 to 5.5 VDC @ 25 °C, regulated
Current............................... 30 mA Maximum
Short Circuit .......................Protected
Isolation ............................. Low voltage (< 48 V AC/DC)
Power Requirements
DC (3-9950-10,
3-9950-11) 24 VDC nominal (12 to 32 VDC,
±10% regulated). A UL60950-1 or UL61010-1 certifi ed power supply must be used. Power supply shall also be rated for operation at
4000 m altitude.
AC (3-9950-11).................. 100-240 VAC 50-60 Hz, 24 VA max
Maximum Current
...............200 mA (w/o optional relay module)*
500 mA (with optional relay module)*
Current
Loop ...................... 12 to 32 VDC, ±10% regulated
4 to 20 mA (30 mA max.) Overvoltage protection
....... 48 V Transient Protection Device
(for DC ONLY) Current limiting for circuit protection Reverse-Voltage protection
*The current draw of the other modules and the sensors are minimal
Sensor Input Specifi cations
Digital (S3L) Sensors .......Serial ASCII, TTL level, 9600 bps
Frequency (Flow) Sensors: 0.5 to 1200 Hz
Accuracy ..........................
± 0.5% of reading max error @ 25 °C
Resolution........................ 1 s
Repeatability .................... ± 0.2% of reading
Rejection.......................... No Effect ± 1 A per volt
Short Circuit ..................... Protected
Reverse Polarity .............. Protected
Update Rate ....................(1/frequency) + 100 ms
Shipping Weights
Base Unit ....................0.63 kg (1.38 lb)
Relay Module..............0.19 kg (0.41 lb)
Standards and Approvals
• CE, UL, CUL, WEEE , UKCA
• RoHS Compliant
China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.
• Manufactured under ISO 9001, ISO 14001 and ISO 45001
82
Graphical User Interface created with emWin licensed by SEGGER
GF 9950-10 Six Channel Transmitter
Page 83
Specifi cations
Binary Input Specifi cations (3-9950.393-3)
Input Voltage Range
(without damage) .................. -5 VDC to 30 VDC
(no operation below 0 VDC)
Max. Voltage Rating ............. 30 VDC
Max. Current Rating ............. 6.0 mA
Maximum input voltage for
signal “Off” (low or "0") ........ 1.5 VDC
Minimum input voltage for
signal “On” (high or "1") ........ 3.0 VDC
Maximum current draw for
signal “0” (low): ..................... <500 A DC
Minimum current draw for
signal “1” (high): ................... 500 A
Typical current draw for
signal “1” (high): ................... 6.0 mA at 30 VDC
4.8 mA at 24 VDC
2.4 mA at 12 VDC
1.0 mA at 5 VDC
Current Loop Specifi cations (Internal and Modules)
Analog Outputs .............. Two 4 to 20 mA output in base unit,
Two or four 4 to 20 mA outputs in optional modules
Current Loop Out ........... ANSI-ISA 50.00.01 Class H
(Passive, external voltage required)
Voltage........................... 12 to 32 VDC, ±10% regulated. Use
UL60950-1 or UL61010-1 certifi ed power supply. Power supply shall also be rated for operation at 4000 m altitude.
Max. Impedance: ........... 250 @ 12 VDC
500 @ 18 VDC 750 @ 24 VDC
Span .............................. 3.8 to 21 mA Adjustable, reversible
Accuracy ........................ ± 32 A max. error @ 25 °C @ 24
VDC
Resolution...................... 6 A or better
Temp. Drift ..................... ± 1 A per °C
Isolation ......................... Low voltage (< 48 VAC/DC)
Update Rate ..................100 mS nominal
Zero ............................... 4.0 mA factory set; user programmable
from 3.8 to 5.0 mA
Full Scale .......................
20.00 mA factory set; user programmable from 19.0 to 21.0 mA
Power Supply Rejection ± 1 A per V Actual update rate determined by sensor type Short circuit and reverse polarity protected
Error Condition ..............
Selectable error condition 3.6 or 22 mA or None
Test Mode ...................... Increment to desired current
(range 3.8 to 21.00 mA)
Relay Specifi cations
Dry-Contact Relays
Type ...................................... SPDT
Form ..................................... C
Max. Voltage Rating ............. 30 VDC or 250 VAC
Max. Current Rating ............. 5 A resistive
Solid-State Relays
Type ...................................... SPDT
Form ..................................... C
Max. Voltage Rating ............. 30 VDC or 30 VAC
Max. Current Rating ............. 0.050 A
Hysteresis ............................. Adjustable (absolute in
Engineering Units)
On Delay............................... 9999.9 seconds (max.)
Test Mode ............................. Set On or Off
Maximum Pulse Rate ........... 300 pulses/minute
Proportional Pulse ................ 300 pulses/minute
Volumetric Pulse Width......... 0.1 to 3200 s
PWM Period ......................... 0.1 to 320 s
Input Types
• Digital (S
3
L) or AC frequency
• Open collector
• 4 to 20 mA input via the 8058-1 iGo Signal Converter
• pH/ORP input via the Digital (S3L) output from the 2751 pH/ORP Sensor Electronics
• Raw Conductivity/Resistivity via the Direct Conductivity Module or the Digital (S3L) output from the 2850 Conductivity/Resistivity Sensor Electronics
Sensor Types:
Flow, pH/ORP, Conductivity/Resistivity, Pressure,
Temperature,
Level/Volume, Salinity, Dissolved Oxygen, Other (4 to 20 mA)
Display Ranges:
pH ...............................-1.00 to 15.00 pH
pH Temp. ....................-99 °C to 350 °C (-146 °F to 662 °F)
ORP ............................-1999 to +1999.9 mV
Flow Rate ................... -9999 to 99999 units per second,
minute, hour or day
Totalizer ......................0.00 to 99999999 units
Conductivity ................ 0.0000 to 99999 S, mS, PPM and
PPB (TDS), k, M
Cond. Temp. ...............-99 °C to +350 °C (-146 °F to 662 °F)
Temperature ...............-99 °C to +350 °C (-146 °F to 662 °F)
Pressure .....................-40 to 1000 psi
Level ...........................-9999 to +99999 m, cm, ft, in, %
Volume ......................... 0 to 99999 cm3, m3, in3, ft3, gal, L, lb, kg, %
Salinity ........................0 to 100 PPT*
Salinity Temp Range ...-2 to 35 C° (28.5 to 95 F°) Dissolved O
...............0 to 50 mg/L, 0 to 200%
2
* Salinity measurements are temperature compensated to 42 PPT. Any salinity value above 42 PPT (42.01 to 100 PPT) will not be compensated and should only be used as a reference. Although the 9950 can measure up to 100 PPT, a warning message will be displayed when measured values are above 42 PPT.
GF 9950-10 Six Channel Transmitter
83
Page 84
Ordering Information
9950 Transmitter Base Unit: Dual Channel, Multi-Parameter, AC Power and DC Power
Mfr. Part No Code Description
3-9950-10 159 002 075
3-9950-11 159 002 076
9950 Base Unit – Two Channel Multi-Parameter Inputs, Two 4 to 20 mA Outputs, Panel Mount, DC Power
9950 Base Unit – Two Channel Multi-Parameter Inputs, Two 4 to 20 mA Outputs, Panel Mount, AC or DC Power
Optional Modules
3-9950.393-1 159 310 268 Relay Module with 4 Mechanical Relays 3-9950.393-2 159 310 269 Relay Module with 2 Mechanical and 2 Solid State Relays 3-9950.393-3 159 310 270 Relay Module with 2 Mechanical Relays and 4 Binary Inputs 3-9900.394-1 159 001 846 Single Channel Direct Conductivity/Resistivity Module 3-9950.394-2 159 001 847 Dual Channel Conductivity Module 3-9950.395-M 159 001 905 Modbus Module 3-9900.398-2 159 001 848 Dual Channel 4 to 20 mA Current Loop Output Module
Accessories
3-8050.396 159 000 617 RC Filter Kit (for relay use), 2 per kit 3-8058-1 159 000 966 i-Go® Signal Converter, wire-mount 3-9950.391 159 310 278 Connector Kit, In-Line, 9950 Transmitter 3-9950.392 159 310 279 Relay Module Connector Kit, 9950 Transmitter
3-9900.392 159 001 700 Wall Mount Enclosure Kit 3-9000.392-1 159 000 839 Liquid Tight Connector Kit, NPT (1 pc.) 3-5000.399 198 840 224 5 x 5 inch Retrofi t Adapter 3-8050.392 159 000 640 CR200 ¼ DIN Retrofi t Adapter
Georg Fischer Signet LLC, 5462 N. Irwindale Avenue, Irwindale CA 91706 U.S.A. • Tel. (626) 571-2770 For Worldwide Sales and Service, visit our website: www.gfps.com • or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfps.com
3-9950-10.090 Rev. 1 01/23 English © Georg Fischer Signet LLC 2023
Loading...