+GF+ 4630 User guide

Page 1
GF 4630 Free Chlorine Analyzer System GF 4632 Chlorine Dioxide Analyzer System
*3-463X.090*
3-4630-X3.090 Rev 0 01/21
The GF 4630 System is EPA 334.0 Compliant.
Description
The GF Chlorine Analyzer Systems are an integrated, all-in-one Chlorine Panel System designed to measure Free Chlorine or Chlorine Dioxide in drinking water and clean, fresh water treatment applications. This manual applies to the 4630 Free Chlorine and 4632 Chlorine Dioxide Analyzer Systems.
Features:
• EPA 334.0 Compliant: The 4630 Free Chlorine System can be used for reporting chlorine residuals in accordance with EPA Method 334.0.
• Complete Chlorine Analyzer System allows quick setup and easy installation. Connect to a DC power supply and water source.
• Unique integrated clear fl ow cell combines sensors, fl ow regulator, fi lter and variable area fl ow indicator in one compact unit.
• Built-in variable area fl ow indicator facilitates fl ow rate confi rmation at a glance.
• Integrated fl ow regulator with removable fi lter accepts inlet pressures of 1 to 8 bar (15 to 120 psi) while maintaining constant fl ow and minimal pressure to the sensors.
• Water fl ows vertically into sensor tip, eliminating bubble entrapment. Raised exit in fl ow cell sensor chamber ensures sensors stay submerged even when system and fl ow is turned off.
• Flow cell accommodates two sensors; one chlorine and an optional pH sensor.
• Automatic pH and temperature compensation or manual pH value input capability for accurate free chlorine readings.
• Easy viewing of the transmitter via the bright backlit LCD display.
• Moisture-proof NEMA 4X wiring enclosure.
English
Operating Instructions
English
Deutsch
Français
Español
Italiano
中文
Do not use the 4630 in systems that could have the absence of Chlorine.
Additional information can be found in the individual product manuals, refer to www.gfps.com.
Safety
CAUTION!
1. Follow instructions carefully to avoid personal injury.
2. Do not exceed the maximum pressure or temperature specifi cations.
3. In areas with elevated temperatures, mounting the Chlorine System in an outdoor box
4. Do not alter product construction.
5. For use with clean fresh water only.
6. Disconnect from external power source before opening wiring enclosure.
7. This panel system may use AC voltages. Wiring should be done by qualifi ed personnel only.
WARNING: The GF Chlorine Analyzer (electrode) is designed to be used in a clean,
chlorinated fl owing stream at all times. DO NOT use in applications where electrode could be exposed to periods without chlorine.
Page 2
Warranty Information Table of Contents
Refer to your local Georg Fischer Sales offi ce for the most current warranty statement.
All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales offi ce or distributor. Product returned without a Service Form may not be warranty replaced or repaired.
GF products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, or mishandling (e.g. broken glass,
chemical poisoning, dry-out)
damaged membrane,
freezing and/or extreme temperatures).
Safety Information
Caution / Warning / Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death.
Electrocution Danger
DO NOT FREEZE
Alerts user to risk of electrocution.
Electrostatic Discharge (ESD)
Alerts user to risk of potential damage to product by ESD.
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of GF products.
Pressurized System Warning
Sensor may be under pressure, take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.
Hand Tighten Only
Overtightening may permanently damage product threads and lead to failure of the retaining nut.
Do Not Use Tools
Use of tool(s) may damage product beyond repair and potentially void product warranty.
Note / Technical Notes
Highlights additional information or detailed procedure.
Do Not Freeze
Products are temperature sensitive and may contain freezable liquids. Freezing damage to pH, ORP, and Chlorine electrodes voids product warranty.
potential of injury or death via
Description ..................................................................................... 1
Warranty Information ...................................................................... 2
Safety Information .......................................................................... 2
Chlorine System Dimensions / Specifi cations ............................... 3
Mounting / System Inventory ......................................................... 4
Quick Start ..................................................................................... 4
9950-3 Operation / Terminal Identifi cation ..................................... 5
Panel Assembly.............................................................................. 6
Component Identifi cation: Flow Cell .............................................. 7
Wiring
3-9950-3 Input Power Wiring ..................................................... 8
4 to 20 mA Wiring Loop 1 & 2 .................................................... 8
9950 Dual Channel 4 to 20 mA Module ..................................... 8
3-9950.393-3 Standard with 9950-3 .......................................... 9
3-9950.393-3 with internally powered fl ow switch ..................... 9
Sensor Wiring ............................................................................. 9
Chlorine Sensor Preparation / Maintenance ............................... 10
Sensor Installation.........................................................................11
Tubing Connections ..................................................................... 12
Water Flow ................................................................................... 12
9950-3 Chlorine Controller
Calibration ........................................................................... 13-15
Signal Type: Digital (S³L) ......................................................... 16
Module Installation ................................................................... 16
Plug-In Modules ....................................................................... 17
Loop and Modbus Modules ..................................................... 17
Relay Modes ....................................................................... 18-21
Chemical Guard ....................................................................... 19
Standard Relay Mode .............................................................. 20
Menu System
View Mode Overview ............................................................... 22
Menu Mode Overview .............................................................. 22
Password Overview ................................................................. 22
View Mode ..............................................................................23-24
Calibration .................................................................................... 25
Chlorine Electrode Calibration ................................................. 25
pH Electrode Calibration .......................................................... 26
Input Menu .............................................................................. 27-28
4 to 20 mA Output (Loop Output)................................................. 29
Relay Menu
Chemical Guard Menu ........................................................ 30-31
Standard Relay Menu .............................................................. 32
Option Menu ........................................................................ 33-34
Modbus Menu .........................................................................35-36
9950 Field Software Upgrade / Updating the 9950 ..................... 37
Troubleshooting ......................................................................38-39
Calibration Error Messages ......................................................... 39
2630 Free Cl / 2632 ClO2 Electrode .......................................40-42
2724 DryLoc pH Electrode ........................................................... 43
pH System Calibration ............................................................ 44-45
Buffer Solutions ............................................................................ 45
Electrode Storage ........................................................................ 45
463X Chlorine Analyzer Flow Cell ............................................... 46
O-ring Installation ......................................................................... 47
Flow Cell Assembly ...................................................................... 48
Troubleshooting
463X Chlorine Analyzer ........................................................... 49
9950-3 Chlorine Controller ....................................................... 50
9950-3 Error Messages ........................................................... 51
2630 Free Cl / 2632 ClO2 Electrodes ...................................... 52
2751-7 pH Electronics .............................................................. 52
Specifi cations
9950-3 Chlorine Controller ....................................................... 53
2650 DryLoc® Amperometric Electronics ................................. 54
2751-7 pH Electronics .............................................................. 54
2630 Free Cl / 2632 ClO2 Electrodes ...................................... 55
2724 DryLoc® pH Electrode ..................................................... 56
Notes ............................................................................................ 57
Ordering Information .................................................................... 58
2
463X Chlorine Systems Manual
Page 3
Dimensions System Specifi cations
279.4 mm (11 in)
305 mm
(12 in)
2
3
4
1
CHLORINE Analyzer
1-8 bar
(15-20 psi)
254 mm
(10 in)
(3 in.)
76 mm
330 mm (13 in.)
171 mm (6.75 in.)
76 mm (3 in)
330 mm
(13 in)
305 mm (12 in.)
254 mm (10 in.)
Hand Holds
148 mm (5.83 in)
25 mm (1.0 in.)
Ø 7 mm (0.27 in.)
Ø 13 mm (0.52 in.)
General
Compatible GF Sensors 3-2630-1 Free Chlorine Sensor 0.02 to 2 ppm 3-2630-2 Free Chlorine Sensor 0.02 to 5 ppm 3-2630-3 Free Chlorine Sensor 0.05 to 20 ppm 3-2632-1 Chlorine Dioxide Electrode 0.02 to 2 ppm 3-2724-00 Flat pH Electrode
Free Chlorine .....................5.5 to 8.2 pH
(consult factory for pH above 8.2)
Chlorine Dioxide ................4.0 to 11 pH
Materials
Panel .................................Black acrylic
Flow Cell ...........................Acrylic
Wiring Enclosure ...............Polycarbonate
Wetted Materials
Flow cell, spacer rings .......Acrylic
Flow regulator housing.......Polycarbonate
Strainer, e-clip, regulator
spring, float .......................Stainless steel
Valves, vent .......................Polypropylene
Flow Cell O-rings,
diaphragm .........................EPR (EPDM), FKM
Chlorine electrode .............PVC, PTFE, FKM, Nylon, Silicone
pH electrode ......................PPS, Glass, UHMWPE, FKM
Sealing tape
(valves, plug, vent) ............PTFE
Plug ...................................Polyethylene
Flow Switch .......................Polypropylene
Performance
System Inlet
Pressure Rating ................. 1 to 8 bar (15 to 120 psi)
Pressure Regulator ............ < 0.69 bar (10 psi) variation over
all ranges of flow and pressure
Flow tolerance ................... ±15% or rated specification above
Flow rate limits ..................8 to 12 gph (US)
(30.24 to 45.36 LPH)
Electrical
DC Input (standard) ...........24 VDC nominal
(12 to 32 VDC, ± 10% regulated) Environmental Requirements
Storage Temperature .........0 °C to 65 °C (32 °F to 149 °F)
Operating Temperature ......5 °C to 45 °C (41 °F to 113 °F)
Relative Humidity ...............0 to 95%
Maximum Altitude ..............4000 m (13,123 ft)
Shipping Weight ................10 kg (22 lbs)
Enclosure ..........................NEMA 4X (with output wire
glands sealed)
Standards and Approvals
CE, UL, CUL, WEEE RoHS Compliant Manufactured under ISO 9001, ISO 14001 and ISO 45001. China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and, (2) This device must accept any interference received, including interference that may cause undesired operation.
Safety Data Sheets (SDS) are available online at www.gfps.com.
463X Chlorine Systems Manual
3
Page 4
Mounting
System Inventory
Do not mount in direct sunlight.
• Bright light can promote algae growth. Indoor mounting is recommended.
• If the system is mounted outdoors, an outdoor enclosure for the whole system must be used to protect the electronics and fl ow cell from light, rodents, insects and dirt.
• Mount the panel according to local electrical, building, and/or plumbing codes and seismic requirements.
• Use four 6 mm (¼ in.) diameter screws or bolts of suffi cient length to mount the panel to a sturdy vertical surface. A mounting template is provided.
• Allow clearance on the sides and bottom for service to the unit.
• Keep panel system electronics and enclosure away from dripping water.
• The panel must be mounted vertically in an upright position.
1. Chlorine panel assembly a. Built-in pressure regulator 15 to 120 psi. b. 3/8 inch hose barb connectors. c. Flow switch
2. 1 each 2630 series Free Chlorine sensor or 2632 Chlorine Dioxide sensor; each with protective cap a. 1 spare membrane cap b. 2 bottles of electrolyte solution c. 1 syringe needle (taped to bottle) d. 1 syringe
3. 1 each Flat pH sensor 3-2724-00 (159 001 545) (3-4630-13, -23, -33 and 3-4632-13 models only)
4. Manual package a. English and Chinese panel assembly manuals b. Wall mounting hardware c. Drill template
5. Customer supplied a. 3/8" hose input and drain b. Hose clamps
Quick Start
Follow the steps below to set up a new Chlorine Analyzer System. Refer to the individual component manuals for detailed information.
Step 1. Mount the panel on a vertical at surface using appropriate hardware.
Do not turn on power at this time.
Step 2. Open the wiring enclosure and wire input power.
(see page 8: Wiring Input)
Step 3. Wire any 4 to 20 mA and relay output. Step 4. Remove sensor access plugs from the ow cell, see page 7.
If the optional pH sensor is NOT used, do not remove the left-side plug from the fl ow cell.
Step 5. Remove the protective cap from the chlorine electrode.
(Keep the electrode cap in a safe place for future use. It is recommended to use the cap to protect the sensor during the removal of the electrode for cleaning or maintenance of the fl ow cell)
Step 6. Complete Sensor Preparation (see page 10) and install the chlorine sensor into the electronics.
(see page 11: Sensor Installation) Install the chlorine electrode into the fl ow cell. The chlorine sensor is installed in the right-side access port, optional pH sensor is installed in the left-side access port.
NOTE:
Step 7. Repeat step 5 and 6 for the pH electrode. Step 8. Remove the ow switch from the bottom of the ow cell and install into the ow cell , see page 7 Step 9. Install the in uent water source to the "Inlet Port" nipple assembly of the ow switch. Install 3/8-inch tubing and secure
Step 10. Install 3/8-inch tubing and secure with a hose clamp on the "Drain" port and direct the tube to a proper drain.
NOTE: Drain line (zero pressure) must be lower than in uent line to prevent back ow into the ow cell. Step 11. Verify the inlet and drain ball valves are in the open position and the sample port is in the off position.
Step 12. Turn on the in uent water source and check the system for leaks. Step 13. Apply power to the system and allow system to initialize. Calibrate per instructions.
Both the Chlorine and the pH electrodes are required to be calibrated for accuracy. Upon inital set up the electrodes will need to be placed in a chlorinated fl ow stream for a minimum of 4 hours in a conditioning period. Calibration of the Chlorine electrode is typically required every 14 to 21 days to comply with the EPA standard 334.0. The calibration frequency of the pH electrode is application dependent and the intervals between calibrations must be determined by the user of the system.
with a hose clamp. (Not included. See page 12: Tubing Connections)
(Not included)
(See page 12: Tubing Connections)
(See page 10: Chlorine Sensor Calibration)
See page 10: Chlorine Sensor Calibration, for chlorine calibration and set up procedure.
4
463X Chlorine Systems Manual
Page 5
9950-3 Operation
Channel 1
display area
Channel 2
display area
Numeric Values or labels
Bar Graph
Numeric Values or labels
Bar Graph
Relay 1 Indicator LED
CH1
Backlight Sensor (do not block)
Relay 2 Indicator LED
9999.9
9999.9
00.00
CH2
9999.9
9999.9
00.00
UP, DOWN keys
Scroll through Menu options or adjust values during editing Press both together to exit a menu or escape without saving
RIGHT key
Select item or character to edit
Relay 3 Indicator LED
UOM UOM
99.99
UOM UOM
99.99
ENTER
Relay 4 Indicator LED
4
Units of Measure (mg/l/ppm)
Units of Measure (pH)
Menu Navigation Keys
9950-3 Terminal Identifi cation
Relay Module 3-9950.393-3
USB Port
ENTER
DC Power Terminals
ENTER key
Access menus Save changes
Current Loop Outputs 1 and 2
Current Loop Outputs 3 and 4 3-9950.398-2
Channel 1 Input Chlorine Sensor Electrode
Optional Modbus Module 3-9950.395-M
Channel 2 Input pH Electrode
463X Chlorine Systems Manual
5
Page 6
Panel Assembly
pH sensor
electronics
3-2751-7 (159 001 957)
pH electrode
3-2724-00 (159 001 545)
2
1
3
4
ENTER
Wiring enclosure
Chlorine Controller
3-9950-3 (159 001 955)
Vent
Cap
Amperometric chlorine sensor electronics
3-2650-7 (159 001 670)
Chlorine electrode
3-2630-X (Free Chlorine) 3-2632-1 (Chlorine Dioxide)
Flow range limits
Flow rate indicator
Flow regulator with strainer
Inlet port
Quick disconnect
3-4630.395 (159 001 960)
2000-2525*
Flow switch
Quick disconnect plug
9.5 mm (3/8 inch hose)
2000-3825*
1-8 bar
(15-20 psi)
Drain port
9.5 mm
(3/8 inch hose)
Flow cell block
Flow cell O-rings
Ball valve
9.5 mm (3/8 inch) 4300-5225*
Hose barb
9.5 mm (3/8 inch) 2000-3804*
Sampling port
* Contact factory for quote
6
463X Chlorine Systems Manual
Page 7
Component Identifi cation: Flow Cell
Spacer ring
Stainless steel fi lter
Pressure regulator
Flow cell nut
Flat washer
Flow cell back block
Regulator Assembly (3-4630.391)
(159 001 689)
Pressure regulator outer O-ring (large)
Pressure regulator outer O-ring (small)
Flow cell O-rings (4 segments)
Flow cell front block
Flat washer*
Disc spring (2700-0003)*
Knurled thumb nut
Flow cell bolt (2490-0554)
*Not included with 3-4630.392 (159 001 690)
Flow Cell Rebuild Kit (3-4630.390)
(159 001 688)
(1700-2512)*
3-4630.392
(159 001 690)
Flat washer
Disc spring (curved washer)
Included in kit 3-4630.390
Item Quantity
Flow cell nut 6
Flat washer 15
Disc spring 9
Knurled thumb nut 3
Flow cell bolt 6
463X Chlorine Systems Manual
7
Page 8
Wiring
3-9950-3 Input Power Wiring
WARNING
DO NOT connect AC power to the DC version THE 3-9950-3 MUST BE POWERED BY 12 - 32 VDC at 0.5 A.
DC POWER
Red
Red
Red
Black
Blue
Blue
DC POWER
+
+
+
Power Supply
12 to 32 VDC
PLC or
Recorder
PLC or
Recorder
Red
Black
3-9950-3
+
Power Supply
12 to 32 VDC
Caution:
For DC power input and loop current regulated voltage, use UL60950-1 or UL61010-1 certifi ed power supply. Power supply shall also be rated for operation at 4000 m altitude.
9950 Dual Channel 4 to 20 mA Module4 to 20 mA Wiring Loop 1 & 2
PLC
MODULE 1
Loop A+
Loop A– Loop B+
Loop B–
MODULE 2
Recorder
Red
Blue
Red
Blue
or
+
Recorder
PLC
or
Power
Supply
+
+
8
463X Chlorine Systems Manual
Page 9
Wiring continued
3-9950.393-3 Standard with 9950-3
The alarm is OFF during normal operation, and will go ON when the relay energizes according to 9950 Relay settings.
RELAY MODULE
Blue
Red
Blue
Red
Black
Black
Relay 1
Relay 2
No Connection
Binary Input 1 Binary Input 2 Binary Input 3 Binary Input 4
N/O 1 N/C 1
COM 1
N/O 2 N/C 2
COM 2
X SW1 SW2 SW3 SW4
SW COM
Chemical Dosing pump
Chemical Dosing pump
DC
Power Supply
No Connection
BINARY INPUT 1 BINARY INPUT 2
BINARY INPUT 3 BINARY INPUT 4
Flow Switch Assembly
(159 001 960)
3-9950.393-3 with internally powered fl ow switch from factory
Black
RELAY 1
RELAY 2
N/O 1
N/C 1
COM 1
N/O 2 N/C 2
COM 2
SW1
SW2
SW3
SW4
SW COM
Red
X
Black
Black
ALARM!
AC or DC
power
Sensor Wiring
CH1CH2
DC POWER
Technical Notes:
• Route sensor cables away from AC power lines.
• Connect the silver (shield) wire to earth ground in case of EMI noise interference.
Black
Red
White
Black
Red
White
2650-7 with 263X Chlorine Electrode
2751-7 with 2724 pH Electrode
463X Chlorine Systems Manual
9
Page 10
Chlorine Sensor Preparation
2630 Free Chlorine Electrode 2632 Chlorine Dioxide Electrode
NOTE: See page 40 for additional 2630, 2632 information.
• Chlorine sensors are shipped without internal electrolyte solution.
• Prior to installation and supplying power, Chlorine sensors must be fi lled with the appropriate internal electrolyte solution.
• Verify the correct electrolyte solution is utilized with the corresponding sensor.
• Free Chlorine and Chlorine Dioxide sensors require different electrolyte solutions.
Avoid skin or eye contact with electrolyte solution. Wear rubber gloves and goggles.
*Safety Data Sheets (SDS) are available online at www.gfps.com.
CAUTION! DO NOT touch the gold tip or the membrane of the sensor.
Initial Fill Procedure:
When adding electrolyte, be prepared for an accidental spill. Working near a sink is recommended.
1. Remove the protective bottle from the end of the electrode.
2. Remove the membrane cap from the front of the sensor.
Note: When new sensors are shipped, the membrane cap is not tightened to
the sensor.
3. Fill supplied syringe with electrolyte solution.
Additional caution should be taken when handling Chlorine Dioxide electrolyte solution.
4. Place the electrode on a level surface.
5. Insert syringe needle fully into one of the eight electrode holes while injecting with
electrolyte solution. Slowly injecting the electrolyte solution into the sensor to avoid introducing air bubbles. The electrode holds approximately 14 milliliters of solution. Slowly fi ll until solution begins to fl ow out of holes. Do not allow the solution to run down the electrode and wet the electrical contacts in the DryLoc connector.
6. Slowly screw on the membrane cap fi nger tight. Do not use tools. To avoid damage
and contamination, do not touch the white membrane surface on the membrane cap.
ml
10
9
8
7
6
5
4
3
2
1
CAUTION:
DO NOT touch the gold-plated tip or the membrane of the sensor.
Chlorine Sensor Calibration
A new chlorine electrode or one that has had the membrane cap changed must be calibrated. See page 13, 9950-3 information on calibration of the chlorine sensor. A diethyl-p-phenylenediamine (DPD) colorimeter test kit (not included) is required for sensor calibration. A sample is taken and analyzed with the DPD test kit, then this value is entered into the 9950-3.
• A new system requires an inital 4 hour conditioning.
• Calibrate after a membrane cap change (requires 2 hour stabilization time).
• Calibrate after the internal electrolyte is replaced (requires 2 hour stabilization time).
• Check calibration 24 hours after sensor is placed in service.
• Check calibration weekly to monthly depending on process requirements.
Chlorine Sensor Maintenance
The sensor membrane and internal electrolyte solution must be replaced over the life of the electrode. To maintain accurate chlorine measurements, GF recommends that the internal electrolyte be replaced every 3 to 6 months, or when Chlorine readings drift low and/or cannot maintain a calibration longer than 5 days. Actual interval between maintenance of the sensor will be dependent on the application, chlorine level and contaminates in the water. See page 41; Maintenance and Storage.
Keep spare membrane caps available. Membrane caps carry no warranty.
To maintain accurate chlorine measurements, GF recommends the internal electrolyte be replaced and the gold-plated sensor electrode be properly polished every 3 to 6 months. See page 41.
NOTE: Inspect and change the membrane cap if damaged. See page 41, Maintenance, for cleaning procedure.
10
463X Chlorine Systems Manual
Page 11
Sensor Installation
• Remove sensor access plugs from the fl ow cell.
Note: Chlorine Sensor Preparation must be completed prior to installation, see 10.
• Holding the 3-2751-7 (159 001 957) or 3-2650-7 (159 001 670) electronics inverted, open the DryLoc turning the upper locking ring ¼-turn counter-clockwise.
• Insert the electrode facing up. Turn the locking ring ¼-turn clockwise to lock the electronics in place.
• The mechanism will “click” when it is locked.
• Install the complete electrode assembly into the fl ow cell and ensure the key on the electrode aligns with the key slot on the fl ow cell.
Avoid skin or eye contact with electrolyte solution. Wear rubber gloves and goggles. Safety Data
Sheets (SDS) are available online at www.gfps.com.
Lubricate the DryLoc Do not lubricate the sensors body O-ring. Turn by hand. Do Not Use Tools. Do Not Drop or Strike the Membrane.
DryLoc® Connection to Sensor Electronics
®
connector by
®
O-ring only.
Chlorine Electrode
Lubricate O-rings with a
non-petroleum based, viscous
lubricant (grease) compatible
with the system.
(159 001 670)
Chlorine Electronics
3-2630-X
3-2632
3-2650-7
Do not lubricate this O-Ring
Locking Ring
Lock
GF 3-2630-X GF 3-2632 Chlorine Electrode
GF 3-2724-00 (159 001 545)
Flat pH Electrode
3-2724-00 (159 001 545) pH Electrode
Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system.
Lock
3-2751-7 (159 001 957) pH Electronics
pH Cl
IN
OUT
GF 3-2650-7 pH Electronics
GF 3-2751-7
Electronics
pH Cl
IN
463X Chlorine Systems Manual
OUT
11
Page 12
X
A
Flow Switch and Tubing Connections
Use suitable 9.5 mm (3/8 in) ID tubing that is rated for your inlet pressure.
Use hose clamps.
Remove the Flow Switch from the three ball valves located on the bottom of the fl ow cell.
Insert inlet quick connect into the mating connector on the fl ow cell.
Inlet Quick Release Connection
Press down gray plunger to release. Water fl ow is automatically shut off when disconnected.
Valve position for start up and normal use.
Note: Turn off inlet valve rst when stopping water ow.
• The drain tube must be positioned lower than the infl uent water source to allow proper fl ow through the fl ow cell. The fl ow cell must drain by gravity, not system pressure.
• When testing, allow the sample to fl ow for a few seconds before collection.
IN
Inlet Drain
OUT
(to outlet)
Sampling port (flow off)
Water Flow
Flow Rate
The fl ow rate is in the proper range when the fl oat is between the Min. and Max. markers on the fl ow cell. The fl ow range limits are 30.24 to 45.36 L/h (8 to 12 gal/h).
OK
For Low Flow applications (fl ow less than 1 bar/15 psi), the Flow Regulator can be removed and the fl ow adjusted using the Flow Cell ball valve. NOTE: The inlet hose barb will also need to be changed. (Contact factory for details.)
ClpH
IN
OUT
12
463X Chlorine Systems Manual
Page 13
9950-3 Calibration
Both the Chlorine and the pH electrodes are required to be calibrated for accuracy. Upon inital set up the electrodes will need to be placed in a chlorinated fl ow steam for a minimum of 4 hours in a conditioning period. Calibration of the Chlorine electrode is typically required every 14 to 21 days to comply with the EPA standard 334.0. The calibration frequency of the pH electrode is application dependent and the intervals between calibrations must be determined by the user of the system.
In order to achieve the most accurate chlorine measurement, after the minimum 4 hour conditioning period, calibration must be performed in the following specifi c order:
• Temperature
pH electrode (Free Chlorine only)
• Chlorine electrode
NOTE: The Temperature element only requires calibration one time during initial commissioning. Once calibrated the Temperature element is calibrated for the life of the chlorine electrode.
Chlorine sensor temperature element calibration
The temperature element inside the chlorine sensor needs to be calibrated. Use a reference thermometer to verify the actual temperature of the sample. This value is then entered into the calibration screen of the Temperature Calibration menu. Tip: Remove the pH electrode from the fl ow cell and insert the reference thermometer. If no pH sensor is being used
system), remove the cell plug to insert the thermometer. Replace the plug after calibration.
(ClO
2
pH electrode calibration It is highly recommended to perform a two point calibration of the pH electrode using a pH buffer 7 and pH buffer 4 or 10. The pH buffers' temperatures should be close to the system water temperature for an accurate calibration. Once a successful calibration is performed, the 9950 will automatically time stamp the "Last CAL" display with the date of calibration. The user can add a date to the "NEXT CAL" display which will generate a "Calibration Due" icon across the pH channel (CH2).
Chlorine sensor
Calibration must be performed to every new chlorine sensor (FCl or ClO2) and any time a membrane cap or internal electrolyte is replaced. With the chlorine sensor attached to the 2650 electronics (with system power on) and placed in a chlorinated fl ow steam for the minimum of 4 hour conditioning period, the temperature and chlorine In-Process Calibration needs to be performed. Once a successful calibration is performed the 9950 will automatically time stamp the "Last CAL" display with the date of calibration. The user can add a date to the "NEXT CAL" display which will generate a "Calibration Due" icon across the chlorine channel (CH1).
NOTE: It is advised to return 24 hours after the initial Chlorine calibration to verify and recalibrate if necessary.
Temperature Calibration
4
CAL
INPUT LOOP RELAY OPTION
ENTER
CH 1 FREE CI
25.00
C
5s
ENTER
CH 1 FREE CI
CH 2 pH
TEXT 3
+25.30
+25.00
C
CH 1 FREE CI
OFF
HOLD OUTPUTS
to Save
ENTER
+
Display returns to calibration display
CH 1 FREE CI
SET
TEMPERATURE
CH 1 FREE CI
CH 2 pH
TEXT 3
463X Chlorine Systems Manual
13
Page 14
Calibration
pH Electrode Calibration
CH 1 FREE CI
CH 2 pH
TEXT 3
CH 2 pH
to edit to edit
CH 2 pH
OFF
HOLD OUTPUTS
CH 2 pH
SET
SET
pH STANDARD
pH STANDARD
CH 2 pH
6.80
RAW - 3 mV
Place electrode tip in saturated pH 7.0 buf
fer.
to edit
to edit
CH 2 pH
NEXT CAL
MM/DD/YYYY
+07.01
6.80
RAW - 3 mV
to Save
ENTER
pH STANDARD
Use the arrows to modify the pH value, matching the buffer solution.
+04.80 CH 2 pH
4.80
RAW 173 mV
to Save
ENTER
Use the arrows to modify the pH value, matching the buffer solution.
to edit
MM/DD/YYYY
NEXT CAL
MM/DD/YYYY
to Save
ENTER
CH 2 pH
SET
SET
pH SLOPE
CH 2 pH
NEXT CAL
MM/DD/YYYY
CH 2 pH
SET
pH SLOPE
Place electrode tip in saturated pH 4.0 buf
CH 2 pH
LAST CAL
MM/DD/YYYY
Display returns to calibration display
+
fer.
CH 1 FREE CI
CH 2 pH
TEXT 3
14
463X Chlorine Systems Manual
Page 15
Calibration
Chlorine Electrode Calibration
CH 1 FREE CI
CH 2 pH
TEXT 3
CH 1 FREE CI
CH 2 pH
TEXT 3
to edit
CH 2 pH
CH 1 FREE CI
SET
OFF
pH STANDARD
HOLD OUTPUTS
x3
CH 2 pH
CH 1 FREE CI
SET
SET
pH STANDARD
ZERO CAL
Place electrode tip into DI water.
- .01
nA
0.0 PPM
to Save
ENTER
ENTER
CH 2 pH
CH 1 FREE CI
SET
SET
pH STANDARD
IN PROCESS
to Save
ENTER
CH 1 FREE Cl
NEXT CAL
MM/DD/YYYY
CH 2 pH
CH 1 FREE CI
SET
SET
pH STANDARD
IN PROCESS
CH 1 FREE Cl
MM/DD/YYYY
Test sample with an external DPD test instrument.
LAST CAL
+
Display returns to calibration display
to edit
CH 1 FREE Cl
NEXT CAL
MM/DD/YYYY
+ 002.25
+ 002.25
RAW 100.5 nA
Modify Chlorine PPM value to match DPD reference.
MM/DD/YYYY
NEXT CAL
MM/DD/YYYY
463X Chlorine Systems Manual
15
Page 16
Signal Type: Digital (S3L)
• The input terminals on the 9950-3 carry Digital (S3L) serial data from the sensor.
• Do not route sensor or output cables in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
• The TOTAL cable length from I/O devices to the transmitter must not exceed 305 m (1000 ft).
• In case of noise interference, ground the sensor SHIELD wire to a local earth ground at a point near the sensor.
• Consult the sensor manual for additional wiring information.
• The maximum cable length of the Digital (S
3
L) bus varies depending on the types of sensors connected and the size of the conductors in the cable. For best results, determine the maximum cable length for the system before routing cables.
Module Installation
In case of noise interference, connect the cable shield to earth ground.
Maximum total cable length of the Digital (S
3
L) Bus:
The quality of the cable used in the bus determines the maximum length. The maximum cable length may not exceed 305 m (1,000 ft), regardless of current requirements.
CAUTION
Avoid Electrostatic Discharge (ESD).
The plug-in modules may be installed either before or after the base unit is mounted. If the 9950-3 Base Unit will be mounted using the provided quick clip mounting bracket, install plug-in modules fi rst with power disconnected.
CAUTION
Exercise care when installing modules. Do not bend connecting pins.
To install modules:
Remove power from the 9950-3. Carefully align pins and connectors (do not bend connecting pins) and push module fi rmly into place, then attach with screw(s) for Relay Modules only.
9950
MODULE 1
MODULE 2
Rounded Corners
Important!
Ensure the rounded corner of the module is alligned with the rounded corner of the module slot before installing
Dual Channel 4 to 20 mA Module
To move modules:
Remove power from the 9950-3.
For Relay Modules:
Unplug connectors, remove screw(s), and carefully pull module straight out from the base unit. Do not bend the connecting pins.
For Dual Channel 4 to 20 mA, and Modbus Modules:
Unplug connectors and squeeze the two retaining tabs while pulling module out of 9950-3.
WARNING
Relays may be connected to external high-voltage power sources or multiple power sources creating an electrocution hazard.
16
463X Chlorine Systems Manual
Page 17
Plug-In Modules
a
c
d
Standard modules for the 9950-3:
a. Base Unit (required) 3-9950-3 159 001 841 Base Controller, no modules b. Relay Module 3-9950.393-3 159 310 270 Relay Module c. Module 1 3-9950.398-2 159 001 848 Dual Channel 4 to 20 mA
Output Module
Optional module for the 9950-3:
d. Module 2 3-9950.395-M 159 001 905 Modbus Module (optional)
Each item is ordered separately. Modules are fi eld-replaceable at any time. See Module Installation (pg. 11) and Ordering Information (pg. 68) sections for more details.
CAUTION
Avoid Electrostatic Discharge (ESD).
Minimize handling of the plug-in modules to reduce the possibility of damage due to ESD.
Handle modules by the edges.
Never touch any exposed circuitry or contacts.
Wear an anti-static wristband or stand on an anti-static mat, or keep one hand touching a
properly grounded pipe or other piece of properly grounded metal when handling modules.
Loop and Modbus Modules
Mfr. Part No. Code Description 3-9950.398-2 159 001 848 Dual Channel 4 to 20 mA Current Loop Output Module 3-9950.395-M 159 001 905 Modbus Module
Dual Channel 4 to 20 mA Module
Two additional 4 to 20 mA passive current loop outputs in a single module, allows up to six current loop modules in the 9950-3 Chlorine Controller.
Modbus Module
The Modbus Module connects the 9950-3 to serial RS485 Modbus automation networks. The Modbus Module will allow for PLCs and SCADA Systems to query the current readings, channel status, state of the relay and 4 to 20 mA outputs, along with status of the 9950-3 Chlorine Controller and Modbus communications. Detail programming information is available in the 9950 Modbus Programming Manual, available at www.gfps.com.
Default Communication Parameters ModBus Address 95 ModBus Termination Off ModBus Mode RTU ModBus Baud Rate 19.2K ModBus Parity Even ModBus Endian Big
463X Chlorine Systems Manual
17
Page 18
Relay Modes
The 9950-3 has a slot for an optional Relay Module for the module listed on page 17.
• The four Binary Inputs can detect if an external switch is open or closed. The switch state can be used to enable or disable the relays in the module control relay activation and are not considered an input measurement. This is standard on the 9950-3.
• A fl ow switch must be wired into Binary input #1 to allow access to the Chemical Guard mode of the 9950-3.
• Dry-contact relays are electromechanical switches with a moving contact armature. They are suitable for many general-purpose applications, AC or DC, including AC loads up to 250 V.
For wiring information, refer to the Relay Wiring section, page 8.
Binary Input Ratings
Maximum Input voltage (without damage)
. The four binary inputs only
30 VDC
3-9950.393-3
Power
Switch
EXT
INT
Minimum Input voltage (without damage)
Maximum input voltage for signal “Off” (low or "0")
Minimum input voltage for signal “On” (high or "1")
Maximum current draw for signal “0” (low)
Minimum current draw for signal “1” (high)
Typical current draw for signal “1” (high)
-5 VDC (no operation below 0 VDC)
1.5 VDC
3.0 VDC
<500 ADC
500 A
6.0 mA at 30 VDC,
4.8 mA at 24 VDC,
2.4 mA at 12 VDC,
1.0 mA at 5 VDC
The 3-9950.393-3 module can be set to supply power to the external switch/ sensor. If an externally powered sensor is connected with the 3-9950.393-3 module, set the power switch to the EXT position. The module can power external sensors by switching the power switch to the INT position. Maximum power that can be drawn from the module is 6mA at 30 VDC. Important for externally powered switches/sensors to ensure the output signal is within the allowable range for the 3-9950.393-3.
Int / Ext
Switch
EXT
INT
CAUTION
Keep relay wiring separate from electrode and signal wiring to prevent interference and damage to the 9950-3 Transmitter.
NOTE:
• The four red Indicator LEDs on the front panel of the 9950-3 show the status of relays 1, 2, 3, & 4. The LEDs will activate with or without a relay module installed.
• Hysteresis and time delay are adjustable for each relay.
CAUTION
Switching active loads (usually inductive) can cause contact arcing suffi cient to damage the relays.
The RC Filter Kit or “snubber” (part number 3-8050.396) is available as an accessory to reduce or eliminate these damaging effects. Recommended for inductive loads greater than 50 VAC (remote relays, solenoids, pumps, etc.)
18
463X Chlorine Systems Manual
Page 19
Relay Modes
Chemical Guard
The Chemical Guard prevents overdosing of oxidant type chemicals (Chlorine) whose concentration varies based on the pH level of the application.
The Chemical Guard is only accessible in the relay menu when used in conjunction with a fl ow switch (GF option) that is wired to the Binary Input #1 of the 3-9950.393-3 relay module.
When the Chemical Guard function is selected, Relay 1 is automatically defaulted to control and adjust the Oxidant chemical (Chlorine) and Relay 2 is defaulted to control and adjust the pH level of the application.
The control logic (Fig #) will always interrupt (disable) Relay 1 if the pH requires adjusting. Once the pH is corrected, Relay 1 is then available to add additional oxidants to the application.
The Chemical Guard allows Relay 1 and Relay 2 to be independently programmed as a low set point, high set point or a proportional pulse output.
Flow Switch Enabled
YES
Chemical Guard ON
YES
No
Is there flow?
YES
Is the pH correct?
YES
Is the Chlorine level
correct?
YES
No
No
Disable R1 and R2
Turn on R2
Disable R1
Turn on R1 Disable R2
1. Press and hold ENTER for 3 seconds
2. Go to the Relay Menu by pressing then ENTER
3. Display shows "FLOW SWITCH" NO, press and change to YES, Press ENTER to save
4. Press  screen should show "RELAY CHEMICAL GUARD" OFF
5. Press  and change to YES, press ENTER to save
6. Press  screen should show "Oxidizer" Mode OFF
7. Press  and select LOW
8. Press  x2 to set POINT DISPLAY
9. Press  to enter the LOW SET POINT value 3.0 ppm, press ENTER to save
10. Press  then press  to enter ON DELAY time of 15 seconds, press ENTER to save
11. Press  then press  to enter an OXIDIER HYSTERESIS of 1.0 ppm, Press ENTER to save
12. Press  x3 screen should show pH ADJUST Mode OFF
13. Press  to change the Mode to HIGH
14. Press  x2 then press  to enter HIGH SET POINT of 7.5
15. Press  then press  to enter ON DELAY time of 15 seconds, press ENTER to save
16. Press  then press  to enter pH ADJUST HYSTERESIS of 0.3 ppm
17. Press + at the same time to go back to the VIEW Mode
Example: Maintain a pool's chemistry using two dosing pumps at pH 7.2 and chlorine at
3.0 ppm
463X Chlorine Systems Manual
19
Page 20
Standard Relay Mode
Set your relay functions to your own application requirements.
Once a setting is saved it becomes immediately active.
1. Press and hold ENTER for 3 seconds
2. Go to the Relay Menu by pressing then ENTER
3. Press the  to select desired source. You will be asked to enter a code or password. Select source and press ENTER to confi rm
4. Press  to enter the relay MODE selection screen
5. Press  and then  to select R1 MODE LOW. Press ENTER to confirm
6. Press  to R1 SET LOW. Press  to enter PPM value of 2.0
7. Use the  and  to change Setpoint. Press ENTER to save
8. Scroll  to the R1 HYSTERESIS menu
9. Press  to edit
10. Set the hysteresis for this relay. Set the value to 1.0 ppm
11. Press ENTER
12. Scroll down  to the R1 ON DELAY menu
13. Press  to edit
14. Set the turn-on delay in seconds for the relay: 15.0
15. Press ENTER Then
16. Exit to View Mode
Relay function can be tested in the RELAY menu
Example: Set a relay R1 to turn on at a low setpoint of 2.0 ppm with a time delay of 15 seconds and turn off at
3.0 ppm.
Remember,
SET LOW + hysteresis = OFF
Relay 1 Source = CH1 Primary Relay 1 Mode = Low R1 Set Low = 2.0 R1 Hysteresis = 1.0 R1 on delay = 15.0 sec
Relay Modes
• Binary
When the relay source is set to “Binary", the Binary Inputs can sense if the input is ON (Closed) or OFF (Open). The four binary inputs only control relay activation and are not considered a measurement source. Each binary mode is independent and each of them can be assigned to a relay mode.
• Error Mode
Relay activated when any related failures occur for the system.
Example: Binary Input
A 3-4630.395 Flow Switch is mounted in the Chlorine fl ow cell. The fl ow switch will prevent the relay(s) from activating if the ow to the cell is interrupted.
20
463X Chlorine Systems Manual
Page 21
10
5
Relay Output Rate:
0 to 100 Pulses/min.
Pulse rate
0 pulses
100 pulses
7.5
Relay Modes
Hysteresis
Time
Low Setpoint
Process
Low Setpoint + Hysteresis
Hysteresis
Time
High Setpoint
Process
High Setpoint - Hysteresis
Relay energized
Relay de-energized
The 9950-3 relays are selectable and confi gurable and can be used as switches that respond when the process value moves above or below a user-defi ned setpoint or it can be used to generate a pulse at a rate proportional to the process value. They can be used for Low Alarm, High Alarm or Proportional Pulse triggering related to the process value. All relay functions are set up in the RELAY menus.
The 9950-3 supports the ability to activate the Red Backlight when a relay is activated. An optional check box is displayed during the programming of a relay that will turn the Red Backlight on when the relay is activated. The Red Backlight can be activated by any relay or relay mode.
Low Setpoint:
Relay is on when the measured value is less than or equal to the setpoint. Relay turns off when the measured value is equal to or greater than the Low Setpoint + Hysteresis.
High Setpoint:
Relay is on when the measured value is greater than or equal to the setpoint. The relay turns off when the measured value is less than or equal to the High Setpoint - Hysteresis.
Proportional Pulse Operation:
The transmitter can output a pulse at the rate defi ned by the settings in the CAL menu and the sensor input. The maximum pulse output is 300 pulses per minute.
As the process value rises above the setpoint, the output will start pulsing in relation to the process value, the maximum pulse endpoint and the programmed pulses/minute. The pulse rate will change as the process value changes and approaches the programmed endpoint. This functionality can be used to precisely control the process.
The starting point, endpoint and maximum pulse rate are select able in the RELAY menus.
Typical usage would be to control solenoid-operated dosing pumps.
NOTE: Relay LEDs will ash in PULSE mode.
100 pulses
Starting point Endpoint
In the example:
• The output will be 0 pulses/min. when value is less than 5.
• The output will be 50 pulses/min. when value is 7.5.
• The output will be 100 pulses/min. when value is 10 or greater.
463X Chlorine Systems Manual
21
Page 22
Menu System
VIEW Mode Overview
The top level of screens are referred to as the VIEW Mode. The VIEW Mode displays the sensor measurement values, sensor raw measurement data, current output and relay status. The horizontal bar graph represents the primary measurement value that is also displayed in the numeric fi eld above the bar graph. The bar graph is primarily used to display the full scale range of the sensor, but can be scaled via the OPTION menu item.
During normal operation, the 9950-3 displays the VIEW mode.
• To select a display, press the or arrow keys. The display selections scroll in a continuous loop.
• Changing the display selection does not interrupt system operations.
• No password is necessary to change display selection.
• Output settings cannot be edited from the View Mode.
• The display will return to the VIEW Mode if no button is pressed for 10 minutes.
MENU Mode Overview
The MENU Mode enables the user to view and confi gure all menu items. The fi ve menus available are: CAL, INPUT, LOOP, RELAY and OPTION.
MENU Mode is entered by pressing and holding ENTER for three seconds.
To select a menu, use the and arrow keys to highlight the desired menu and press ENTER to select the menu.
Error Handling
Errors occurring while on the measurement screens show a specifi c message (e.g., Wrong Sensor). Once the error is resolved or cleared, the error message stops.
_ _ _ _ ._
CH1
MISSING SENSOR
9999.9
00.00
CH2
6 2 .8
00.00
_ _ _ _ ._
CH1
WRONG SENSOR
9999.9
00.00
CH2
6 2 .8
00.00
7 .0
7.0
mV
GPMS
00.00
pH
15.00
mV
GPMS
00.00
pH
15.00
°F
°F
In the selected menu, use the and keys to navigate through the menu. Use the , and keys to edit the selected item.
To save the new selection, press the ENTER key. A message displaying “Saving” will be displayed for 3 seconds. After this message is displayed, the newly selected value will be displayed, if applicable. To abandon the changes press the and arrows simultaneously.
Password Overview
The password is often required to start editing. Once entered correctly, this password will not be needed for subsequent edits, until the menu system is exited. The password is required when the menu system is re-entered.
Your choice of password (STD or CODE) is selected in the Options Menu.
• STD
The standard (STD) password is , pressed in sequence. This password is designed to protect the 9950 from unintentional changes. It is best suited for systems
where a group of people need to be able to change settings.
• CODE
The CODE default setting is 0000, adjustable to any 4-digit numerical code up to 9999. Using a personal code provides the maximum degree of security. This code can be modifi ed in the Options menu.
Password Reset
Turn off 9950-3 Press and hold all 4 keys on keypad while turning on 9950 and unit will revert back to STD password.
Scrolling
In some cases, more than one message or measurement may need to be displayed. This is accomplished by alternating the message portions across the screen.
PASSWORD TYPE
WRONG CODE
TEXT 3
In the MENU mode, if the wrong code or password is entered, an ERROR message is displayed.
LANGUAGE
0
0000
LANGUAGE
ENTER CODE
LANGUAGE
PASSWORD
SAVING
TEXT 3
To change your CODE, go to OPTIONS mode, enter your desired code and press ENTER. (The STD password cannot be changed).
22
463X Chlorine Systems Manual
Page 23
VIEW Mode
9950-3 VIEW Mode
FCI
CH2
-1.0
CH1
FREE CL
CH2
pH
-1.0
0.0
25 C
0.0
7.2
0.0
2.43
0.5
7.0
2.43
0.5
7.0
RAW
CH1 115.00 nA +
CH2 +117 mV
GLASS IMPEDANCE
CH2 1049 MOhm ->
mg/l
TEMP
mg/l
15.0
mg/l
15.0
View Measurement Display 1 Displays the primary (top line) and secondary (bottom line) values from the sensors assigned to Channel 1
5.0
and 2. If the Bar Graph feature is enabled in OPTION menu, the bar graphs will also be visible.
pH
If Channel 2 doesn't have a sensor assigned to it, the area will be blank.
14
This is the default VIEW display and does not time out.
View Measurement Display 2
5.0
Channel 1 electrode information: Chlorine sensor primary value and visual bar graph on top line.
pH
Channel 2 electrode information: pH primary value and visual bar graph on bottom line.
View Measurement Display 3
5.0
Channel 1 electrode information: primary value, bar graph and custom name on top line.
pH
Channel 2 electrode information: primary value, bar graph and custom name on bottom line.
Electrode Wellness Information Channel 1 electrode information: Chlorine sensor raw nanoamps (nA) value Channel 2 electrode information: pH electrode millivolt (mV) value
pH only, last measured glass impedance, press key to manually update reading.
SENSOR DATA
CH1 FC1 ->
CH2 pH ->
SENSOR DATA
CH1 FC1 ->
CH2 pH ->
INTERNAL LOOPS FCI
TEMP
12.00
4.00
Select which sensor data to be reviewed by using the and arrows. CH 1 Using the key, you will see the following screens:
[Cl Sensor S/N] [Cl Type and Range] [ZeroCAL: ppm & nA] [In Proc: ppm & nA ] [Temp at CAL] [pH at CAL] [Temp Offset]
Electrode Serial Number Identify the chlorine sensor type connected and its ppm range View user entered Zero Calibration data in ppm and nA Users in process calibration value when it is entered in the Calibration Menu Temperature recorded during user In-Process Calibration pH value recorded during user In-Process Calibration Temperature offset calculated from user-entered temperature calibration from
Calibrate Menu [Elapsed Time] [Low & High: C]
Total hours of operation
Lowest and highest temperatures the Cl sensor has been subjected to during
operation
CH 2 Using the key, you will see the following screens:
[Serial Number] [Model Number] [pH Sensor Effi ciency] [pH Offset Calibration] [Factory Impedance] [Usage Time] [Minimum pH] [Maximum pH] [Minimum Temperature] [Maximum Temperature]
Electrode Serial Number
Electrode Model Number
Calibration Slope Value
Offset
pH only, Factory pH glass Impedance
Electrode Runtime in Hours
Minimum Measured Value
Maximum Measured Value
Minimum Measured Temperature
Maximum Measured Temperature
Displays the internal 4 to 20 mA LOOP output for each channel assigned primary and secondary sources. (3.6, 22.00 are error output values).
MODULE LOOPS pH FCI MOD2 A MOD2 B
18.00
12.00
8.00
4.00
Displays the Module 4 to 20 mA LOOP output for each channel assigned primary and secondary sources. (3.6, 22.00 are error output values).
463X Chlorine Systems Manual
23
Page 24
VIEW Mode
RELAY STATUS
RELAY 1 RELAY 2 RELAY 3 RELAY 4
BINARY STATUS FLOW SWITCH BINARY 2 BINARY 3 BINARY 4
MODBUS STATUS ADDRESS BUS MSG CNT 9950 MSG CNT MSG ERR CNT
SYS CONFIG CH1 CH2 RELAY MOD1 MOD2
GENERATION 2
SW VER 9950 S/N
TEXT 1
MENU ->
TEXT 3
OFF OFF OFF
OFF
OFF
999 99999 99999 99999
FREE CI
399504033
39950408
4-20 OUT
00-06
0123456789
ON
ON
pH
ON
Relay status is stated to the right of the relay number as (OFF, ON) for each relay.
When the fl ow switch is enabled, the label for Binary Input will automatically change to FLOW SWITCH. The Binary labels can be manually confi gured. Binary labels can be edited by pressing on the Binary Status screen and scrolling through the Binary Label submenu. Pressing Cancel ( and simultaneously) returns to the Binary Status screen in View Mode. English, French, German, Italian and Spanish only. Binary status is stated to the right of the binary channel number as (OFF, ON) for each of the four binary inputs.
If there is no module installed, the error code "Missing Module" will appear and blink.
SYS CONFIG shows channel, parameter and relay module part number.
Generation Screen shows version of software and serial number of unit. Note: Top shows the Generation of 9950
MENU-> will allow access to top level menu by pressing the key.
DATE
09/01/2016
09:06
DATE and TIME display show the current system date and time
Important
The Relay, Mod1 and Mod2 number returned by the Sys Confi g Screen is not the Mfr. Part No. To convert from the displayed number to the GF Mfr. Part No. use the chart below.
Module RELAY, MOD1 or MOD2 Mfr. Part No.
Dual Channel 4 to 20 Out 39950408 3-9950.398-2
Modbus Module 39950409 3-9950-395M
Relay Four Mechanical 399504031 3-9950.393-1
Relay Two SS and Two Mechanical 399504032 3-9950.393-2
Relay Two Mech and Four Binary 399504033 3-9950.393-3
24
463X Chlorine Systems Manual
Page 25
Calibration
For greater accuracy the initial calibration of the system must be performed in the following order:
1. Temperature (only required on the initial start up of the FCl electrode).
4
CAL
INPUT LOOP RELAY OPTION
ENTER
ENTER
WARNING Temperature Calibration
Calibrate temperature element: Must be performed on new chlorine electrodes. The temperature element is located in Chlorine Electrode (CH 1), see screens 1 and 2 below. Once calibrated the temperature element does not have to be calibrated for the life of the chlorine electrode.
Chlorine Electrode Calibration
2. pH electrode (2 point calibration required using pH buffer 7 and pH buffer 4 or 10).
3. Zero calibration, single point calibration in DI water.
4. Chlorine sensor. (Single point calibration using a DPD reference).
• Before initial calibration, the FCl electrode must be connected in the 3-2650-7 electronics and system powered. The Chlorine electrode must be installed in the fl ow cell with chlorinated water fl owing through the fl ow cell for at least 4 hours.
• It is recommended that the FCl electrode be process calibrated 24 hours after the initial "Start up" calibration.
• Chlorine and pH calibration screens will be shown only when a valid sensor is detected.
• Chlorine calibration should be stable per EPA 334.0 and should require additional calibration every 14 to 21 days.
• pH calibration is required over a period of time established by the customer based on the application requirement.
CH 1 FREE CI
OFF
HOLD OUTPUTS
CH 1 TEMP
SET
TEMPERATURE ->
CH 1 FREE CI
SET
ZERO CAL
CH 1 FREE CI
SET
IN PROCESS ->
CH 1 FREE CI
LAST CAL
MM - DD - YYYY
CH 1 FREE CI
NEXT CAL
MM - DD - YYYY
ON prevents relays from activating while making adjustments and relays in PULSE mode will suspend pulsing. 4 to 20 mA output is held until the user exits the CAL menu or turns it OFF. Select ON/OFF. Default = OFF
Enter process temperature from a reference thermometer: °C or °F. Units are set up in the INPUT Menu. CAUTION: Refer to the calibration procedure on page 13
The ZERO CAL is only required to be performed when a chlorine electrode is initially installed. The chlorine sensor should be placed in a beaker of non-chlorinated water (DI water). When is pressed, the LIVE readings in nA is displayed, but cannot be edited. Once the nA are stable press ENTER to save the ZERO OFFSET.
Enter process chlorine value determined by customer supplied DPD test kit.
After a successful chlorine calibration the LAST CAL display is automatically updated and date stamped with the date of the calibration.
Operator manually enters the next required calibration date. To avoid an immediate error code, this date must be at least one day after the LAST CAL date. A warning will show up on the chlorine sensor channel showing "Calibration is Due". To override this warning the operator can input a year that is several years away (ie 2050).
CH FREE FCI
RESET CAL
This menu resets Cl readings and Zero Calibration back to Factory Calibration. WARNING! User entered Cl calibration settings will be lost, the Temperature, Zero Point and Process Calibration will need to be done again.
463X Chlorine Systems Manual
25
Page 26
Calibration
pH Electrode Calibration
CH 2 pH
OFF
HOLD OUTPUTS
CH 2 pH
°SETF
pH STANDARD ->
CH2 pH
SET
pH SLOPE ->
CH 2 pH
100LAST CAL.00
12 - 31 - 2015
CH 2 pH
NEXT CAL
MM - DD - YYYY
CH2 pH
SET
RESET pH CAL ->
ON prevents relays from activating while making adjustments and relays in PULSE mode will suspend pulsing. 4 to 20 mA output is held until the user exits the CAL menu or turns it OFF. Select ON/OFF. Default = OFF
Applies an offset to the pH measurement. The slope value and the standard value must be at least 2 pH units apart. The ideal values are the minimum and maximum values of your process. An error message is displayed if offset is too high.
Applies a slope to the pH measurement. The slope value and the standard value must be at least 2 pH units apart. The ideal values are the minimum and maximum values of your process. An error message is displayed if the slope is too high or low.
After a successful pH calibration, the LAST CAL display is automatically updated and date stamped with the date of the calibration.
Operator manually enters the next required calibration date. A warning will show up on the pH sensor channel showing "Calibration is Due". To override this warning, the operator can input a date that is several years away (ie 2050).
Press to reset pH calibration to factory default. This is useful to wipe out suspected corrupted calibration process prior to performing a new calibration.
26
463X Chlorine Systems Manual
Page 27
INPUT Menu
CAL INPUT
INPUT
LOOP RELAY OPTION
Allows manual selection of sensor type assigned to the channel. Channel 1 is dedicated to Chlorine electrodes (FCl and Cl02) Channel 2 is dedicated to pH
4
electrode.
Allows user to reset 9950-3 Transmitter to Factory settings or select "NONE" for sensor input. This feature is used when a pH electrode is not being used (Cl02).
ENTER
Allows user to confi gure electrode specifi c parameters such as selecting engineering units, Temperature (°C or °F) and Chlorine (mg/l or PPM).
ENTER
Note: User is strongly discouraged from changing the sensor type away from the correct sensor.
CH 1 FREE CI ->
CH 2 pH ->
CHANNEL 1
NAME
FREE Cl
CHANNEL 1
PPM
UNITS
CH 1 FREE CI
°C
TEMP UNITS
CH 1 FREE CI
AVERAGE
OFF
CH 1 FREE CI
TYPE
FREE Cl
Select the desired channel. Channel 1 is dedicated to Chlorine electrodes (FCl and Cl02). Choose FACTORY, FREE Cl or Cl02.
Assign a specifi c name to the measurement point. Default name is the electrode type.
Select the Engineering units to be displayed. Select PPM or MG/L (Default PPM)
Select the Engineering units of the Temperature on the display. Select °C or °F. (Default is °C)
Dampens Display, Output and Relay Response Rate. Select LOW, MED, HIGH, OFF Default = OFF
Current Channel type is listed. Manually change the channel type to match sensor type if sensor is not found by S3L. Select FACTORY, FREE Cl, Cl02 DEFAULT = FACTORY, the 2751-7 (pH Electronics) and 2650 (Chlorine Electronics) will automatically be found by the controller.
463X Chlorine Systems Manual
27
Page 28
INPUT Menu
pH INPUT Menu
CH 1 FREE CI ->
CH 2 pH ->
CHANNEL 2
NAME
pH
MANUAL pH
OFF
MANUAL pH VALUE
7.00
pH
CH 2 pH
AVERAGE
OFF
Select sensor type assigned to the channel. Channel 2 is dedicated to pH electrode. Choose FACTORY, NONE (use with Cl02) or pH electrode.
Assign a specifi c name to the measurement point. Default name is the electrode type (Free Chlorine or NONE for Chlorine dioxide).
Select MANUAL pH or Electrode. If manual pH is set to "OFF", the pH value from a connected pH electrode will be used for FREE CHLORINE calculations and pH measurement. When manual pH is selected "ON" , a pH value manually entered is used for all Chlorine calculations.
Enter your pH value here if a pH electrode is not connected. Default value = 7.00
Dampens Display, Output and Relay Response Rate. Select LOW, MED, HIGH, OFF Default = OFF
CH 2 pH
TYPE
pH
Current Channel type is listed. Manually change the channel type to match sensor type if sensor is not found by S³L. Select NONE, FACTORY, pH DEFAULT = FACTORY, the 2751-7 (pH Electronics) will automatically be found by the controller. Select NONE when using Chlorine dioxide with NO pH electrode.
28
463X Chlorine Systems Manual
Page 29
4 to 20mA Output (LOOP Output)
The following can individually be set for each loop on the 9950-3.
CAL INPUT
LOOP
RELAY OPTION
4
ENTER
ENTER
Loops 1 and 2 are in the 9950-3 base unit. Loops MOD1 A and MOD1 B are present when a loop module is in the Module 1 Slot. Loops MOD2 A and MOD2 B are present when a loop module is in the Module 2 Slot.
LOOP 1
SOURCE
NONE
LOOP 1
LABEL
FCI
L1 4 mA SETPOINT
0.00
PPM
L1 20 mA SETPOINT
5.00
PPM
LOOP 1
ERROR VALUE
22 mA
L1 ADJUST 4 mA
4.00
mA
L1 ADJUST 20 mA
21.0
mA
L1 TEST LOOP
12.05
mA
Select source for each loop. Choose NONE, FCL, ClO2, pH or Temperature. Temperature values are from the Chlorine sensor.
English, French, German, Italian and Spanish languages only. A custom label can be assigned to the loop for ease of identifi cation. Default name will be the SOURCE sensor type chosen above.
(ALL) Set value corresponding to desired 4 mA output. 6 digits max. Default = 0 ppm (Chlorine), 0 (pH), 0 °C (32 °F) (Temperature)
Set value corresponding to desired 20 mA output. 6 digits max. Defaults = 5.0 ppm (Chlorine), 14 (pH), 45 °C (113 °F) (Temperature)
(ALL) Set desired LOOP output value when sensor error (e.g., bad sensor, broken wire) is detected. Select (3.6 mA, 22 mA,
or NONE). Default = 22
(ALL) Allows fi ne-tuning to compensate for errors in other equipment connected to the 9950-3. Adjust the current output at 4 mA. The display value represents the precise current output. Adjustment limits: from 3.80 mA minimum to 5.00 mA maximum. Default = 4.00 mA
(ALL) Allows fi ne-tuning to compensate for errors in other equipment connected to the 9900. Adjust the current output at 20 mA. The display value represents the precise current output. Adjustment limits: from 19.00 mA minimum to 21.00 mA maximum. Default = 20 mA.
(ALL) Press or to manually control the output current value from 3.8 mA to 21.00 mA to test the output of LOOP. The current loop will revert to automatic control when this menu item is left.
463X Chlorine Systems Manual
29
Page 30
Relay Menu
CAL INPUT LOOP
RELAY
OPTION
ENTER
CHEMICAL GUARD Menu
4
ENTER
FLOW SWITCH?
NO
RELAY 1
CHEMICAL GUARD
OFF
OXIDIZER
MODE
OFF
RELAY 1
LABEL
OXIDIZER
OXIDIZER SET LOW
0.00
PPM
OXIDIZER SET HIGH
0.00
PPM
OXIDIZER MODE
MODE
PROP PULSE
OXIDIZER PULSE MIN
5.00
PPM
OXIDIZER PULSE MAX
0.10
PPM
R1 MAX RATE
120.00
mPULSES/MINA
A fl ow switch is necessary to enable the Chemical Guard feature. Verify the use of a fl ow switch by selecting YES. If NO is selected Chemical Guard will not be accessed and will go to Standard Relay Mode.
Enable the Chemical Guard feature by changing to ON. If Chemical Guard is not enabled go to RELAY 1 source by pressing once. See Standard Relay information on page 20.
RELAY 1 will automatically be assigned to control Oxidizing chemicals using the Input source of the Chlorine Sensor (CH1). MODE available are LOW, HIGH or PROPORTIONAL PULSE. NOTE: Chemical Guard must use a pH electrode. If the pH input is changed to MANUAL pH input, the Chemical Guard feature is disabled and the standard relay modes must be used control the system until the pH electrode is replaced.
English, French, German,Italian and Spanish languages only. A custom label can be assigned to the relay for ease of identifi cation. Default name will be the "Oxidizer".
Relay turns on if the process measurement is equal to or lower than this value and the pH value is correct. Set to desired value. See Chemical Guard overview.
Relay turns on if the process measurement is equal to or higher than this value and the pH value is correct. Set to desired value. See Chemical Guard overview.
Proportional Pulse Mode varies the frequency of pulses in direct proportion to minimum and maximum set points.
Enter the desired minimum set point (END POINT) of the application. The number of pulses will be reduced to the chemcial dosing pump as the set point is reached to prevent overdosing of chemicals. Default = 5.00 ppm
Enter the maximum off set (START POINT) from desired set point (SEND POINT) of the application. This is usually the maximum allowable OFF SET value from the set point of the application. The closer to the starting point the more pulses will be sent to the chemical pump. Default = 0.01 ppm
Set desired maximum pulse rate (300 max) NOTE: Pulse width fi xed at 100 ms.
30
463X Chlorine Systems Manual
Page 31
Relay Menu
CHEMICAL GUARD Menu
RELAY 1 ACTIVATE
RED BACKLIGHT
NO
pH ADJUST
MODE
OFF
pH ADJUST
SET LOW
00.00
RELAY 2
LABEL
OXIDIZER
Select YES to have the Red Backlight illuminate during relay activation, NO to not have the Red Backlight activated by the relay activation.
RELAY 2 will automatically be assigned to control pH adjusting chemicals using the Input source of the pH Electrode (CH2). MODE available are LOW, HIGH or PROPORTIONAL PULSE. NOTE: Chemical Guard must use a pH electrode. If the pH input is changed to MANUAL pH input, the Chemical Guard is disabled and the standard relay modes must be used to control the system until the pH electrode is replaced.
Relay turns on if the process measurement is equal to or lower than this value. Set to desired value. See Chemical Guard overview.
English, French, German, Italian and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
pH ADJUST
SET HIGH
0.00
pH ADJUST
PULSE MIN
7.00 pH
pH ADJUST
PULSE MAX
8.00 pH
R1 MAX RATE
120.00
mPULSES/MINA
RELAY 2 ACTIVE
RED BACKLIGHT
NO
RELAY 2
TEST RELAY
mAA
Relay turns on if the process measurement is equal to or higher than this value. Set to desired value. See Chemical Guard overview.
Enter the desired minimum set point (END POINT) of the application. The number of pulses will be reduced to the chemcial dosing pump as the set point is reached to prevent overdosing of chemicals. Default = 8.00 pH
Enter the maximum offset (START POINT) from desired set point (SEND POINT) of the application. This is usually the maximum allowable OFFSET value from the set point of the application. The closer to the starting point the more pulses will be sent to the chemical pump. Default = 7.00 pH
Set desired maximum pulse rate (300 max) NOTE: Pulse width fi xed at 100 ms.
Select YES to have the Red Backlight illuminate during relay activation, NO to not have the Red Backlight activated by the relay activation.
(ALL) Press or to turn relay on or off for testing purposes.
463X Chlorine Systems Manual
31
Page 32
Relay Menu
STANDARD RELAY Menu
RELAY 1
SOURCE
mNONEA
RELAY 1
MODE
OFF
RELAY 1
LABEL
RELAY 1
R1 SET LOW
0.10
PPM
R1 SET HIGH
5.00
PPMA
R1 HYSTERSIS
0.50
PPM
R1 ON DELAY
All relay sources can be assigned to Relay 1 – 4. Choose NONE, FCl, ClO2, pH, Temperature, Binary 1 – 4. If CHANNEL 1 is assigned a chlorine sensor the temperature source comes from the chlorine sensor.
(ALL) Select the desired mode of operation for the (R1) output (OFF, LOW, HIGH, PROP PULSE, ERROR mode) (See page 18, Relay Modes). Default = OFF. Continue stepping through to select R2, R3 and R4 output modes. When MODE is set to ERROR, relay turns on if sensor problem is detected on either channel. ON DELAY delays the turning on of the relay until after the programmed delay time.
English, French, German, Italian and Spanish only. A custom label can be assigned to the relay for ease of identifi cation.
(ALL) (Shown if LOW mode) Relay turns on if process measurement is equal to or lower than this value. Set desired value. Default = Chlorine 0.10 PPM, Temperature 25 C, pH 4.0
Relay turns on if process measurement is equal to or higher than this value. Set desired value. Default = Chlorine 5.00 PPM, Temperature 45 C, pH 8.0
Hysteresis prevents the mechanical relays from chattering around the set point. Set amount (in units of measure from INPUT Mode) to add to SET LOW or SET HIGH values.
5.0
SEC
Set seconds (up to 9999.9) to wait before activating relay once the relay condition is true.
NOTE: Screens below are only visible when Porportional Pulse Mode is selected.
R1 PULSE MIN
5.00
PPMIPLE
R1 PULSE MAX
0.10
PPMF
R1 MAX RATE
120.00
mPULSES/MINA
RELAY 1 ACTIVATE
RED BACKLIGHT
NO
RELAY 1
TEST RELAY
mA
Enter the desired minimum set point (END POINT) of the application. The number of pulses will be reduced to the chemcial dosing pump as the set point is reached to prevent overdosing of chemicals. Default = Chlorine 5.00 PPM, Temperature 45 C, pH 8.0
Enter the maximum off set (START POINT) from desired set point (SEND POINT) of the application. This is usually the maximum allowable OFF SET value from the set point of the application. The closer to the starting point the more pulses will be sent to the chemical pump. Default = Chlorine 0.10 PPM, Temperature 25 C, pH 4.0
Set desired maximum pulse rate (300 max) NOTE: Pulse width is fi xed at 100 ms.
Select Yes to have the Red Backlight illuminate during relay activation, No to not have the Red Backlight activated by the relay activation.
(ALL) Press or to turn relay on or off for testing purposes. Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.
32
463X Chlorine Systems Manual
Page 33
OPTION Menu
CAL INPUT LOOP RELAY
OPTION
ENGLISH FRANÇ
DEUTCH
ESPAÑOL
Зݧ
TIME FORMAT
4
ENTER
ENTER
Choose desired the language for the 9950-3. Default = English
24 HR
BINARY ON
SET TIME
08:56
TIMER
DATE FORMAT
MM/DD/YYYY
MULTIPLE
SET DATE
02/25/2016
OFF
SET DECIMAL MARK
999.99
mA
SELECT UNITS
METRIC
mA
CONTRAST
Choose a time format between a 24 Hour or an AM/PM clock. Default = 24 HR
Set the time according to the format chosen 00:00
Date Format; MM/DD/YYYY, DD/MM/YYYY, YYYY/MM/DD
Set the date according to the format chosen 00/00/0000
Choose either comma or point according to local practices. Default = 999.99 (point).
Choose METRIC or US CUSTOMARY
mA
3
Adjust the displays contrast for best viewing. A setting of 1 is lower contrast, 5 is higher and 3 is default.
463X Chlorine Systems Manual
33
Page 34
OPTION Menu
WHITE BACKLIGHT
AUTO LOW
mA
RED BACKLIGHT
ON
DISPLAY
BAR GRAPH
ON
PASSWORD TYPE
STD
%
PASSWORD TYPE
CODE
%
CH2 BAR GRAPH MIN
0.00
PPM
CH2 BAR GRAPH MAX
100.00
PPM
CH1
_ _ _ _._
PPM DECIMAL
CH1
_ _ _ _._
°F DECIMAL
White Backlight: OFF, LOW, HIGH, AUTO LOW, AUTO HIGH
The Red Backlight illuminates when an error condition is detected or user selected relay(s) are activated. Default = ON
Display Bar Graph: ON, OFF When enabled, user enters range to be displayed.
(ALL) Select STD, CODE. Default = STD (Refer to page 22 for further detail)
(CODE) If code is selected Default Code = 0000 (Refer to page 22 for further detail)
Enter a value to represent bar at minimum. Default = 0
Enter a value to represent bar at maximum. Defaults = 5.00 ppm (Chlorine), 14 (pH)
(ALL) Set the decimal to the best resolution for your application for both CH1 and or CH2. The display will automatically scale up to this resolution for each channel. Select 0, 1, 2, 3, or 4 decimal places, (varies by parameter). Default = 1 Decimal Place
(Chlorine electrodes) Set the Temperature decimal to the best resolution for your application for CH1. The display will automatically scale up to this resolution. Select 0, 1, or 2. Default = 1 Decimal Place
34
MEMO
MEMO
SEC
(ALL) Enter 17-character string, if desired. Default = MEMO
463X Chlorine Systems Manual
Page 35
Modbus Menus
MODBUS
MODULE
NONE
MODBUS
COMMUNICATION
SETTINGS ->
MODBUS
REMOTE UPDATE
NO
9950
Select NONE, MOD 1 or MOD 2. The MODBUS module must be installed into Module 2 of the 9950.
Press the key to change the settings of the Modbus Module. See page 36 for Modbus Communication Settings Menus.
Setting Modbus Remote update to YES allows remote Modbus device to change the settings on the 9950-3. Setting Modbus Remote Update to No prevents remote devices from make any changes to the 9950-3. Remote device will still be able to read all of the Modbus parameters.
GENERATION 3
SEC
USER MANUAL
100.00
SEC
Displays Transmitter Generation Version.
QR code for user manual.
463X Chlorine Systems Manual
35
Page 36
Modbus Menus
OPTION Menu
MODBUS
COMMUNICATION
SETTINGS ->
Press the key to change the settings of the Modbus Module
Modbus Communication Settings Menus
MODBUS
ADDRESS
95
NETWORK
Set the Modbus address of the 9950-3. The default address is 95
The 9950 Modbus Module has an electronic termination network.
TERMINATION
NO
MODBUS
BAUD RATE
19.2 K
MODBUS
PARITY
EVEN
MODBUS
ENDIAN
BIG
RESET TO
The Termination can be activated by setting the value to Yes. The default value is No (the termination is off).
Set the baud rate of the Modbus network. The default value is 19.2K baud, 19200 baud.
Set the parity of the Modbus network. The default value is Even parity.
Set the byte order in which the 9950-3 transmits fl oating point numbers in the Modbus protocol. The default value is Big Endian.
DEFAULT
NO
ACTIVATE
SETTINGS
NO
Reset all Modbus setting to the default values.
ACTIVATE SETTINGS: YES sends the changes to the Modbus Module, NO keeps the current settings
36
463X Chlorine Systems Manual
Page 37
9950 Field Software Upgrade
The 9950 upgrade fi le will be available on the Georg Fischer website. You will need a USB fl ash drive that is formatted, using Microsoft Windows, in either FAT16 or FAT32 format. Do not use exFAT or NTSF formats. These are incompatible with the 9950.
The upgrade fi le is named Update.Fwc. Copy the fi le to the root directory of the Flash Drive. Do not change the name of the fi le or store the fi le in a sub directory on the Flash Drive. The 9950 will only look in the root directory for the specifi c le, Update.Fwc.
Important! Do not use a USB extension cable. The USB ash drive must be directly connected to the 9950 USB port.
Sample Flash Drive Directory with Update File
Updating the 9950
Updates available on www.gfps.com.
1. Disconnect power from the 9950.
2. Insert the Flash Drive into the 9950 USB slot.
3. Reconnect power to the 9950.
4. LED 1 on the 9950 will fl ash quickly as the unit searches for a Flash Drive and the correct fi le.
5. When the 9950 detects a Flash Drive and a valid update fi le, LED 2 will slowly ash during the update process.
6. The update process takes approximately 30 seconds.
7. After a successful update the 9950 will boot to the normal screen.
8. If the update is successful, disconnect power, remove USB drive, reconnect power
and review application settings.
Troubleshooting
If, after 10 seconds, the unit boots to the normal screen, the 9950 was unable to fi nd the Flash Drive or the fi le.
a. Verify the Flash Drive has been formatted in either FAT16 or FAT32 format, the upgrade fi le is in the root directory of the Flash Drive, and the fi le name is Update.Fwc. You may also need to try a different fl ash drive.
b. If LED 4 on the 9950 is continuously illuminated, this
indicates either the fi le found on the Flash Drive is corrupted, or power was interrupted during the upgrade
process. Remove power from the 9950. Remove the Flash Drive. Apply power to the 9950.
c. If the 9950 starts normally, the fi le on the Flash Drive was
corrupted. Download a new copy of the update fi le and copy it to the Flash Drive. Repeat the update instruction with the new fi le.
d. If the 9950 starts up and stops with LED 4 on and a
blank screen, this indicates that the update process was
interrupted and the 9950 cannot start. Disconnect power from the 9950. Reattach the Flash Drive to the 9950. Repeat the upgrade procedure.
e. If the 9950 still does not respond after the second
upgrade attempt to: Download a new copy of the update fi le and copy it to a different Flash Drive. Repeat the update instruction with the new Flash Drive.
463X Chlorine Systems Manual
37
Page 38
Troubleshooting
Condition Possible Causes Suggested Solution
Incorrect sensor installed on channel Connect correct sensor to channel
Wrong Sensor
Sensor Type set incorrectly in 9950-3
Wrong Code Wrong password entered Enter correct password (see page 22)
Set correct sensor TYPE in INPUT menu (see page 27)
Backlight inoperative
Relays inoperative
Relay always ON
– – – – – Measurement exceeds display capability Change unit of measure
Missing Module
Wrong Module
Calibration Due
Backlight turned OFF (NOTE: Backlight can turn off automatically in AUTO mode)
Relay Module installed incorrectly Remove and reseat Relay Module
Wrong settings in RELAY menu
Flow Switch OFF Turn Flow Switch ON
Hysteresis value too large Change the hysteresis value
Defective Relay Module Replace Relay Module
4 to 20 mA Module or Modbus Module is not installed in the Module Slot
The wrong Module 4 to 20 mA or Modbus Module) is installed in the slot
Calibration is required for all new sensors
Calibration was reset to factory
Next Calibration date reached
Set BACKLIGHT to LOW, HIGH, AUTO LOW or AUTO HIGH in OPTION menu.
Use test relay to verify relay operation then check relay settings.
Install the module in the correct slot
Replace module with the correct module
Do the Calibration
Next Calibration date was not set
No pH value pH electrode faulty
No Flow No ow detected through the Flow Switch
Manual pH compensation has been selected
38
Man pH
463X Chlorine Systems Manual
Replace pH electrode
Place pH electrode input to Manual
Verify water source is turned ON
Verify the fl ow cell outlet ball valve is open
Disable Flow Switch input via Relay Menu
Replace pH electrode
Set Manual pH to OFF
Page 39
Condition Possible Causes Suggested Solution
No Probe
(Chlorine or pH)
Check Preamp
(Chlorine or pH)
Troubleshooting
9950 "cannot communicate" with Chlorine or pH electrode
Missing sensor or bad temperature element
9950 "cannot communicate" with the Chlorine or pH electronics
• Check wiring
• Install or replace electrode
Check wiring
Replace 2751-7 or 2650 Electronics
Backlight of Screen is RED
Broken Glass
Hi Impedance
Check Cal
(pH only)
Error is detected or Relay is activating the red backlight
pH electrode glass has been damaged and caused very low impedance
The measured pH electrode impedance is above the high impedance level.
Electrode could be in air Ensure electrode is submersed at all times
Slope and/or Offset are out of range
Calibration Error Messages
Message Cause Solution
Out Of Range Use Manual Calibration
Err Too Large To Calibrate
Cal Error Out Of Range
Slope Too Close To Standard (pH) Difference in calibration values
Standard Too Close To Slope
(pH) Buffer not found; Error > ±1.5 pH units
(pH) Offset > 1.3 pH units; Slope error > 100%
Chlorine - nA out of range when compared to the DPD sample
pH - Calibration exceeds the maximum offset of the reference junction allowed
must be > 2 pH units
(pH) Difference in calibration values must be > 2 pH units
Correct error condition
Visually inspect pH electrode for cracked and/or chipped glass
Visually inspect the pH electrode and clean if necessary
Set pH Slope or Standard (pg. 32 & 60)
Reset pH CAL (pg. 32)
• Use 4 and 7 pH buffers
• Clean sensor
• Use manual calibration method
• Check reference
• Clean sensor
• Replace sensor
• Check sensor range
• Check reference
• Replace sensor
• Check sensor
• Use fresh buffer
• Use two different buffer values
• Clean sensor
• Clean sensor
• Use fresh 4, 7 and 10 pH buffers
• Use two different buffer values
463X Chlorine Systems Manual
39
Page 40
2630 Amperometric Chlorine Electrodes
Overview
pH Compensation for Free Chlorine
Amperometric free chlorine sensors measure only hypochlorous acid. As noted in the text above and in Figure 1, the ratio of hypochlorous acid and hypochlorite is pH dependent. The GF Free Chlorine Panel assemblies comes complete with a pH electrode to ensure accurate free chlorine measurement if the pH changes.
100
90
80
70
60
50
40
30
% free chlorine
20
10
0
5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0
HOCl OClˉ
pH at 25 °C
Figure 1
Automatic pH Compensation and Free Chlorine
In many applications, the process pH does not signi cantly uctuate and only a chlorine sensor and instrument are
necessary for accurate chlorine measurement. When the pH varies the free chlorine concentration cannot accurately determined without the use of automatic pH compensation.
The addition of the 3-2724-00 (159 001 545) pH electrode along with its 3-2751-7 (159 001 957) sensor electronics to the system makes pH compensation extremely easy and automatic even with wide fl uctuations or high pH.
See Figure 2 for pH variation recommendations.
Chlorine dioxide is not pH dependant and does not require a pH electrode or electronics.
±0.3
±0.2
pH variation
±0.1
0
6.5 7.0 7.5 8.0 8.5 9.0 Sample pH
Automatic pH compensation recommended
=
in ranges within shaded area
Figure 2
40
463X Chlorine Systems Manual
Page 41
Maintenance
2630 Free Chlorine Electrode 2632 Chlorine Dioxide Electrode
It is required to calibrate the sensor after servicing the membrane and electrolyte. See page 10.
Verifying the sensor's accuracy using the DPD method should be performed to determine if the sensor requires maintenance.
1. Inspect the membrane for dirt or damage. Replace the membrane if it's torn or if the gold cathode is visible.
2. If the membrane is dirty clean the membrane by soaking it in 1 - 5% HCL and gently wash with a stream of DI water (do not use any mechanical device on the membrane).
Refi ll Procedure:
When adding electrolyte, be prepared for an accidental spill. Working near a sink is recommended.
1. Remove the membrane cap from the front of the sensor.
2. Turn the sensor upside down and shake the sensor
vigorously to remove the internal electrolyte.
3. Fill supplied syringe with electrolyte solution. Additional caution should be taken when handling Chlorine Dioxide electrolyte solution.
4. Place the electrode on a level surface.
5. Insert syringe needle fully into one of the eight electrode
holes while injecting with electrolyte solution. Slowly injecting the electrolyte solution into the sensor to avoid introducing air bubbles. The electrode holds approximately 14 milliliters of solution. Slowly fi ll until solution begins to fl ow out of holes. Do not allow the solution to run down the electrode and wet the electrical contacts in the DryLoc connector.
6. Slowly screw on the membrane cap fi nger tight. Do not use tools.
To avoid damage and contamination, do not touch the white membrane surface on the membrane cap.
Diluted HCl can irritate the eyes and skin, use proper safety equipment. Do not use surface tension reducing chemicals, detergents or solvents on the membrane.
• If a fresh water rinse does not clean the membrane, it will need to be replaced.
• Keep spare membrane caps available. Membrane caps carry no warranty.
Avoid skin or eye contact with electrolyte solution. Wear rubber gloves and goggles.
* Safety Data Sheets (SDS) are available online at www.gfps.com.
ml
10
9
8
7
6
5
4
3
2
1
CAUTION:
DO NOT touch the gold-plated tip or the membrane of the sensor.
Storage
If the sensor or panel assembly is to be removed from service for a period of time the sensor must be properly prepared for storage and may need to be recommissioned.
Storage Periods:
1 week or less:
• Close the drain valve, then close the inlet valve of the ow cell, to maintain water inside the fl ow cell to keep the membrane wet.
• If draining the fl ow cell is required, remove the sensor from the cell and the 2650 electronics and store in the
shipping bottle with tap water added.
Continue to page 42 for additional storage periods
DO NOT FREEZE
Dry electrodes are shipped and need to be stored between –10 ºC to 60 ºC (14 ºF to 140 ºF) at a relative humidity that does not exceed 95%.
The primary concerns when storing the electrode is damage to the membrane.
463X Chlorine Systems Manual
41
Page 42
Storage continued and Reconditioning
Storage periods more than 1 week:
Remove the membrane cap and internal electrolyte solution.
Rinse the sensor internal chamber with DI water or cold tap water; drain and allow to dry.
Place the membrane cap back onto the sensor. INSTALL LOOSE, DO NOT COMPLETELY TIGHTEN THE CAP. WHEN STORED DRY, THE MEMBRANE CAP MUST BE STORED RELAXED AND UNSTRESSED.
Store sensor DRY in the shipping bottle, DO NOT ADD WATER.
Recommissioning Procedure:
1. Fill the sensor with the electrolyte as outlined in Maintenance, see page 10, and install the membrane cap.
2. The sensor will have to be polarized in the fl ow cell with fl owing Chlorinated water before being used. Note: This may take 30 to 120 minutes before calibration can be performed.
3. If the sensor does not recover after recommission, follow steps 1-10 outlined in Reconditioning.
4. If the sensor still fails to work properly, continue with steps 11-17.
CAUTION!
Bleach and Chlorine Dioxide solutions are very corrosive and may release dangerous gases if it comes in contact with acids.
Wear proper protective clothing (gloves and eye protection)
• Avoid contact with skin and eyes
• Observe all warnings on safety data sheets
• Avoid spilling bleach and possible contact with acids
Figure 1
Figure 2
Curved Surface
Required to recondition a sensor:
• DI Water
• Beaker (any size available)
• Polishing Sheets (Included in sensor maintenance kit)
• Free Chlorine: Chlorine Bleach (13% concentration)
• Chlorine Dioxide: Aqueous Chlorine Dioxide solution
Reconditioning Procedure (Steps 1-10):
1. Remove the sensor from the 2650 electronics.
2. Remove the membrane cap.
3. Place the sensor on a fi rm, fl at surface with gold cathode pointing upward.
4. Apply a small amount of water to the blue (coarse) polishing paper (dull side).
5. Polish the gold electrode by moving the paper in a circular pattern for 30 seconds. DO NOT go back and forth in a single direction. See Figure 1.
6. Rinse the sensor tip with DI water.
7. Apply a small amount of DI water to the white (fi ne) polishing paper (dull side) and polish the gold electrode by moving the paper in a circular pattern for 30 seconds. DO NOT go back and forth in a single direction. See Figure 1.
8. Rinse the sensor tip with DI water.
9. Top off the sensor with electrolyte and inspect membrane for dirt or damage. Replace if necessary.
10. Insert the sensor into the 2650-7 electronics and apply power.
Reconditioning Procedure (Steps 11-17):
11. Fill beaker with a 12 mm (½ inch) of the appropriate solution.
12. Position or suspend the sensor 6 mm to 12 mm (¼ in. to ½ in.) above the appropriate solution. DO NOT SUBMERGE THE SENSOR. See Figure 2.
13. Apply power to the system.
14. Monitor the nA of the sensor (press the down arrow three times on the 9950-3 Chlorine Controller). The nA reading should start to rise. Response time and nA reading will depend upon the temperature of the bleach.
15. Once the sensor's nA reading reaches approximately 300-360 nA, allow the sensor to remain in the beaker, suspended over the appropriate solution, for an additional 20 minutes.
• If sensor does not recover, cover the beaker to avoid air contamination.
• Contact the factory for assistance (www.gfps.com)
16. After 20 minutes, remove the sensor from the beaker and install it into the fl ow cell and restore ow to the system.
17. Calibrate the sensor after the system has become stable.
42
463X Chlorine Systems Manual
Page 43
2724 DryLoc pH Electrode
Electrode Date Code
The electrode date code indicates the manufacturing date of the electrode.
Electrodes should be put into service as soon as possible and should not remain in the box for more than two years.
Over time, the storage solution (found in the "boot" covering the electrode tip) will evaporate or leak, allowing the delicate sensing tip and reference junction to dry.
To rehydrate a dry electrode, soak the tip in 3.0 mol/L KCl solution for 24 to 48 hours. Preheating the KCl solution to <140 °F may speed up the re-hydration process.
Electrodes more than 2 years old may still be functional, but will take longer to rehydrate.
Restoration may not be effective for severely dehydrated electrodes.
First Digit = Month
N = January
M = February
L = March
K = April
J = May
H = June
G = July
F = August
E = September
D = October
C = November
B = December
Example: L6 = manufactured in March 2021
K2
Second Digit = Year
6 = 2021
7 = 2022
8 = 2023
9 = 2024
0 = 2025
1 = 2026
2 = 2027
3 = 2028
4 = 2029
5 = 2030
6 = 2031
7 = 2032
Electrode Care
Depending on the type of application and the accuracy needed, pH and ORP electrodes will require routine maintenance. Contaminated/dirty electrodes should be swapped with properly cleaned conditioned and calibrated electrodes.
The solution used to clean an electrode varies depending on the type of coating:
Caution: Always wear appropriate safety gear and exercise proper safety practices when working with or near chemicals
General Cleaning – Soak the electrode for 5 minutes in a mixture of warm tap water (< 140° F), and mild detergent such as
liquid dish detergent. A soft brush, like a toothbrush may also be used to clean off particulates.
• Bio lm and bacterial growth – Soak the electrode in a diluted household laundry bleach solution (pH electrodes only), 1part bleach to 10 parts water for 10 minutes.
Alkaline or Scale – Lime/calcium and mineral buildup deposits may be removed by soaking the electrode in a solution of 2% to 5% hydrochloric acid (HCl) or vinegar for no more than 5 minutes.
Acidic contaminants – May be removed by soaking the electrode in a weak caustic (less than 2% to 5% NaOH) for a few minutes only. Use the least harsh chemical which will remove the contaminant within 2-5 minutes of soaking without attacking the materials of construction.
Oily or organic coatings – Soak the probe for 5 to 10 minutes in a solution of warm tap water < 140° F, and dish detergent. Alternatively, use an appropriate solvent (isopropyl alcohol or similar) that will not attack the materials of construction. If required gently clean junction and/or glass with a very soft tooth brush, and follow up with brief soaking (2-5 minutes) in 2% to 5% NaOH (sodium hydroxide). In the case of pH (not ORP) a diffi cult organic fouling can be treated with household bleach (4-6%) as a substitute for the dish detergent.
ORP Platinum Coating – Gently wipe the electrode surface with a soft lint-free paper towel. If the various forms of chemical cleaning are not suffi cient to achieve an accurate measurement and response time, a last resort would be to polish the platinum surface with a 0.3-1.0 micron alumina polishing slurry to remove any stubborn coatings.
After cleaning, rinse pH/ORP electrode in distilled or tap water, then re-examine the electrode for damage that may have been hidden from view. Sensor may not initially behave properly; and have a temporary offset due to the cleaning chemistry. Proper conditioning is always recommended.
Electrode Conditioning
Soak sensor tip (pH or ORP) in 3 mol/L KCl (potassium chloride), for at least 45 minutes to regenerate/condition. The more aggressive the cleaning, the longer the soak in potassium chloride. Failure to soak the electrode will cause the electrode to drift for a time after the cleaning procedure, and could adversely affect the calibration process/results.
If the cleaning and conditioning yield results within operational tolerance, the pH/ORP sensor is once again suitable for use. However, if the results do not bring the readings of the pH/ORP sensor within tolerance, it is time to replace the sensor.
463X Chlorine Systems Manual
43
Page 44
pH System Calibration
There are two functions in a pH electrode that require the system to be calibrated:
Temperature Calibration
The temperature output of the electrode (measured by a 3K Balco or Pt1000 RTD) must be calibrated only once. When a new electrode is installed, it does not need to be repeated.
Because the temperature measurement has a signifi cant infl uence on the electrochemical measurement, the temperature output in new pH electrodes should always be calibrated before the pH/mV calibration.
NOTE: All GF transmitters and controllers incorporate automatic temperature compensation.
pH error due to temperature changes in fl uid
°C pH 2 pH 3 pH 4 pH 5 pH 6 pH 7 pH 8 pH 9 pH 10 pH 11 pH 12
0.15 0.12 0.09 **0.06 0.03 0 0.03 0.06 0.09 0.12 0.15
15 25 35 45
Table 1
As the pH value moves away from neutral (7 pH) or the temperature moves away from 25 ºC, the electrochemical output is affected.
• * Example: At pH 5 the mV output of the electrode is not affected if the temperature is at 25 °C.
• ** The electrode output will be shifted by 0.06 pH units if the temperature is reduced to 15 °C.
55
Offset and Slope Calibration
The pH electrode calibration is an important function that must be done routinely to standardize the electrode due to its ever-changing characteristics. The electrode contains a gel that depleates over time, so the instrument must be readjusted periodically to maintain system accuracy. There are many infl uences, including aging, temperature, coatings and chemicals used that will affect the probe characteristics. After cleaning and conditioning, GF pH electrodes may be calibrated in two pH buffer solutions of different pH values (a two point calibration).
000* 0 0000000
0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15
0.3 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.3
0.45 0.36 0.27 0.18 0.09 0 0.09 0.18 0.27 0.36 0.45
The pH electrode is calibrated by placing it in a buffer solution of known pH and measuring the cell potential. This is a linear function of pH in the range of pH 2-11, therefore two calibration points are needed. Most common calibration buffers are pH 4.01 and pH 7.00 (@ 25 °C). If the pH measurements are made in an acid to neutral range, we recommend using buffer pH 4.01 and buffer pH 7.00 for calibration. If an alkaline sample is measured, we recommend using buffer 7.00 and buffer 10.01 for calibration.
The response of the glass electrode is characterized by two parameters, offset and slope. The theoretical output of a pH electrode at pH 7 is 0 mV. The offset, or zero pH is the deviation of the pH from the nominal value. In the GF transmitters, the offset calibration is performed under STANDARD.
The glass electrode slope is the mV developed per each pH unit (mV/pH). At 25 °C, the theoretical slope is 59.16 mV/pH. So, at pH 4.01, +177 mV will be generated, while at pH 10.01 (with a lower H+ concentration), a potential of -177 mV will be generated, (see Table 2). A new electrode however, will generate 0 15 mV in pH 7.00 and will have a slope effi ciency between 93.2 and 103%. In the GF transmitters, manual slope calibration is performed under SLOPE.
Theoretical mV
Values @ 25 °C
pH mV
2 +295.8
3 +236.64
4 +177.48
5 +118.32
6 +59.16
70
8 –59.16
9 –118.32
10 –177.48
11 –236.64
12 –295.8
Table 2
44
463X Chlorine Systems Manual
Page 45
pH System Calibration continued
How to calculate a pH electrode slope effi ciency
Two pH buffers are selected to check the pH probe performance, pH 4.01 and pH 7.00. The ratio of the measured potentials (E2-E1) to the difference of pH (7.00 – 4.01) gives the slope of the straight line.
1. Read the mV potential generated by the electrode in two calibration buffers
2. Determine the slope (the mV potential generated per pH unit)
3. Divide this number by the theoretical slope, 59.16 mV/pH @ 25 °C and multiply by 100.
Example 1
pH electrode generates -12 mV in pH 7.00 buffer and +162 mV in pH 4.01 buffer.
160 mV - (-12 mV) = 172 mV 172 mV/ 3 = 57.33 mV/pH unit
57.33/59.16 x 100 = 96.9 % slope effi ciency
In Example 2, slope effi ciency of 90.1% is acceptable but the offset of -45 mV is not. An offset shift can be the result of the reference electrode being contaminated or poisoned. A slope shift can be the result of the glass being etched or coated with a hard build-up.
As the electrode ages, the slope decreases gradually, generally a slope between 85% and 105% is acceptable. If the offset value (pH 7.00) is > +/- 45 mV, electrode should be replaced. When a great accuracy is required from the electrode, the offset and the slope values need to be as close as possible to the theoretical values.
Refer to the Calibration Procedure section of the instrument manual for a 2 point pH calibration. If the sensor is used in a 4 to 20 mA ‘blind application’ refer to the Calibration section of the 2751 Smart Sensor Electronics for a 2-point EasyCal pH calibration.
Example 2
pH electrode generates -45 mV in pH 7.00 buffer and +115 mV in pH 4.01 buffer.
115 mV - (-45 mV) = 160 mV 160 mV/3 = 53.33 mV/pH unit
53.33/59.16 x 100 = 90.1%
Buffer Solutions
Proper handling, storage and use of buffers is very important on the pH measurement accuracy. If the buffers are contaminated or used improperly, the calibration will be inaccurate and all subsequent measurements will be wrong.
When calibrating pH electrodes, always use pH buffers in the range of your sample.
To achieve an accurate calibration make sure the electrode and the buffer are at the same temperature.
Buffers have limited shelf life. Do not use a buffer if the expiration date has passed.
Never return used buffers to the buffer bottle. Discard it.
Do not leave the buffer bottle open (exposed to air). The atmospheric carbon dioxide lowers the pH of the alkaline buffers (10.01 pH).
To avoid contamination, rinse the electrode with distilled or tap water before placing it in the buffer. If necessary, the electrode might be blotted dry using a lint-free paper towel (e.g. Kimwipes) to remove excess water. Be extra careful not to touch or rub the surface of the glass. The wiping of the glass can produce a static charge that will interfere with the voltage reading of the electrode. In addition to this, the hydration gel layer is interrupted.
Store buffers at room temperature.
GF offers the following pH buffer solutions, pH 4.01, pH 7.00 and pH 10.01.
Make fresh buffer solutions just before using.
Dispose of all calibration solutions in accordance with local guidelines and regulations.
Electrode Storage
Proper electrode storage maximizes electrode performance and extends electrode life.
Prior to storage - add 3 mol/L KCl to the electrode storage cap (about half full), and carefully secure the cap onto the electrode tip.
Store boxed electrodes wet, fl at or upright (electrode tip down) to maximize hydration of the glass surface.
Store electrodes in stable temperature environment, avoiding freezing conditions (below 0 °C (<33° F) and heat above 100° F).
Never store the electrode in deionized (DI) water. Use 3 mol/L KCl solution to keep the glass wet when out of the process.
Short term storage (up to 24 hours) - soak the electrode tip in either pH 4.01 buffer or 3 mol/L KCl solution.
Drying out of the pH sensitive glass and the junction must be avoided (or damage to the electrode will occur).
Prior to electrode installation, make sure there’s a light fi lm of synthetic grease (such as silicone-free grease) applied to the upper electrode O-ring. If the electrode will be installed into a GF Installation Fitting, then also apply a thin fi lm of synthetic grease to the lower O-ring.
Do not lubricate the bottom O-ring of the electrode if using the 4630 fl ow cell, see page 12.
Long term storage (longer than 24 hrs) - add 3 mol/L KCl solution to the storage cap (about half full) and carefully secure the cap onto the electrode tip. This will ensure probes are always ready to use.
463X Chlorine Systems Manual
45
Page 46
463X Chlorine Analyzer Flow Cell
Maintenance
Sensor Removal
CAUTION: Over time, a sensor can get tight in the ow cell tting. When removing the sensor, avoid
hitting the sensor electronics on the bottom of the wiring enclosure if the sensor suddenly releases.
Take care not to damage the components.
While holding the fl ow cell, grasp the yellow electronics housing as a whole unit and carefully pull upwards, gently rocking back and forth if necessary. Once the sensor is loose, disengage the electronics, then remove the sensor.
TIP: A tool can be used to carefully pry the sensor electronics up and out if it is diffi cult to remove.
Under certain conditions, a dirty fl ow cell and fi lter can create a chlorine demand which could lower the chlorine concentration in the water fl owing past the chlorine sensor. It is recommended to clean the fi lter and fl ow cell on a regular basis. The frequency of necessary cleaning will depend on the application in which the system is being used and the level of accuracy required.
Cleaning
Service the fl ow cell on a clean, dirt free surface to avoid scratches or damage to the fl ow cell.
Figure 1
Step 1. Keep the system powered on. Step 2. Disable the relays and any output loops. Step 3. Turn off the water ow. Step 4. Remove the sensors from the ow cell. NOTE: It is not
necessary to remove the electronics from the sensor.
Step 5. Install the vinyl caps on to the sensor tips or place the tips of
the removed sensors in a cup of sample water for temporary storage while cleaning the fl ow cell.
Keep the sensor tips wet and the sensors energized.
Step 6. Remove the ow cell from the panel by removing the
knurled nuts (Figure 1).
Step 7. Remove the remaining hardware from the fl ow cell.
(Figure 2).
NOTE: It is recommended that an anti-seizing compound appropriate for the application be used during the reassembling of the fl ow cell.
• Wash the surfaces with plain or soapy water only.
• Use gentle, liquid dishwashing soap if necessary.
• DO NOT USE commercial glass cleaning fl uids.
• Use a soft, lint-free cloth.
Servicing the Flow Regulator and Filter
Figure 2
Remove fi lter screen to clean. If the fi lter is not
desired, it may
be removed from
the system.
WARNING: Do not disassemble the fl ow regulator. There are no user serviceable parts inside.
If your inlet pressure is less than 1 bar (15 psi), remove the fl ow regulator and quick disconnect inlet connector.
46
463X Chlorine Systems Manual
Flow Regulator
Page 47
O-ring Installation
CAUTION! The 3-4630.390 (159 001 688) O-ring kit
comes complete with two O-rings (5 and 6) and a single cord of material that must be cut and fi tted into the O-ring groove of the fl ow cell (1 and 4).
• The sealing of the fl ow cell is accomplished by four cut O-ring segments and two round O-rings sealing the fl ow regulator. Refer to the illustration below for special O-ring fi tting instructions. Butt all O-ring joints together so there is no gap.
• Remove the O-rings during fl ow cell disassembly. Both the O-rings and all sealing grooves should be examined for cleanliness.
• Wipe all surfaces carefully with a soft, lint-free cloth to ensure good sealing.
• The cross section of the O-rings should be round and smooth. If they are fl attened, their ability to seal is reduced.
• Do not use liquid or paste sealant.
Do not scratch the sealing surfaces of the fl ow cell block. Scratches to the sealing surfaces can cause irreparable leaks.
Fit all O-ring joints so there is no gap in fi nal assembly.
Butt ends together
1
2
Individual O-rings shown. Butt all segment joints together in fi nal assembly.
New O-ring
cross section
Old fl attened O-ring
cross section
(Replace)
3
4
5
6
Angle cut O-ring
Fit flush with no gap
Straight cut
Fit flush with no gap
Straight cut
Angle cut O-ring
Fit flush with no gap
463X Chlorine Systems Manual
47
Page 48
Flow Cell Assembly
WARNING! Do not over tighten fl ow cell bolts. Maximum torque is 8.1 Nm (72 Lb-In). Over tightening the bolts can damage the fl ow cell. Do not over-tighten the bolts in an attempt to stop a leak.
Install the assembled pressure
1
regulator and the two round O-rings into the fl ow cell.
Install the four O-ring segments.
2
Insert the center bolt into the block.
3
Tighten bolts in 3 passes
2
6
5
4
Torque
8.1 Nm
(72 Lb-In)
MAX!
3
Install the nut and washers.
5
1
Tighten the nut just enough to hold
6
the block together and keep the O-rings in place.
Install the remaining bolts, washers
7
and nuts.
48
Align and place the back block onto
4
the fl ow cell. Check to be sure that the O-rings are seated correctly.
463X Chlorine Systems Manual
Tighten the bolts in three passes to
8
specifi cation.
Install the fl ow cell on the panel.
9
Screw on the thumb nuts fi nger tight.
DO NOT use tools on the thumb nuts.
NOTE: It is recommended that an anti-seizing compound appropriate for
the application be used during the reassembling of the fl ow cell.
Page 49
Troubleshooting
463X Chlorine Analyzer
The troubleshooting table below outlines possible causes and remedies related to the fl ow cell panel system. Refer to the respective transmitter and sensor manuals for specifi c component troubleshooting.
Problem Possible Cause Remedies
Inlet pressure below 1 bar (15 psi)
No water fl ow
Low water fl ow
Excessive fl ow
Flow cell leaks
Leaks around sensor
Water leaks out of top vent hole
Algae growth in fl ow cell
Excessive small bubbles in fl ow cell accumulating on sensors
Transmitter does not turn on
Chlorine reading too high Breached membrane Replace membrane and fi ll solution
Chlorine reading too low
Sensor output drifts
Damaged fl ow regulator Replace fl ow regulator
Clogged fi lter
Clogged fl ow switch Clean or change fl ow switch
Clogged fi lter Clean fi lter
Insuffi cient inlet pressure
Damaged fl ow regulator Inspect or replace
Clogged fl ow switch Clean or change fl ow switch
Excessive pressure over 8 bar (120 psi) Ensure inlet pressure does not exceed 8 bar (120 psi)
Damaged fl ow regulator Replace fl ow regulator
Incorrect assembly Inspect, clean, and reassemble
Loose bolts Tighten bolts (8.1 Nm, 72 Lb-In Max)
Defective or missing O-rings Replace O-rings
Damaged sealing surfaces Replace fl ow cell
Damaged sensor O-ring Inspect or replace O-ring
Damaged fl ow cell Inspect fl ow cell
Drain valve closed or plugged Inspect
Hose connections are backwards Connect water source correctly
Outlet drain not below fl ow cell Route drain line below fl ow cell and vented to atmosphere
Low chlorine concentration in water
Exposure to light
Water source contains bubbles
Water source is saturated with dissolved gasses
No power or external circuit breaker is switched off
Loose or incorrect wire connections Check connections
Electrode needs to be repolished Polish gold-plated electrode
Not enough electrolyte Check electrolyte level
Low fl ow rate Check source pressure
Filter clogged Clean the fi lter
Sensor conditioning not long enough
Electrode needs to be repolished Polish gold-plated electrode
Not enough electrolyte Check electrolyte level
Variable fl ow rate Check fl ow rate
Clogged fi lter Clean the fi lter
Contaminated fi ll solution Change electrolyte fi ll solution
Check source pressure. Remove regulator if pressure is < 1 bar/15 psi
Clean fi lter or remove pressure regulator if inlet pressure is below 1 bar (15 psi)
Increase inlet pressure to specifi cation or remove fl ow regulator
Locate fl ow cell away from light
Check water source
Inspect power and wiring connections
Allow the sensor to condition for 4 hours (See 2630 or 2632 sensor manual for details)
Chlorine Electrode Troubleshooting (see 2630-2 or 2632-1 electrode manual: 3-2630.090)
Transmitter Troubleshooting (see 9950-3 manual: 3-9950-XP.090)
pH Electrode Troubleshooting (see 2724 manual: 3-2724.090)
463X Chlorine Systems Manual
49
Page 50
Troubleshooting
9950-3 Chlorine Controller
Several factors can cause irregular or incorrect readings. The fi rst thing to check is to verify that the controller and sensors have been installed correctly. The list below outlines possible causes and remedies.
Problem Possible Cause Remedies
Incorrect wiring.
Transmitter does not turn on.
Display screen is too dark or too dim.
LCD backlight does not work. Contrast set to low Adjust contrast in Options Menu.
No or low voltage supplied to transmitter. Blown fuse. Bad wire connections or splices.
Contrast set incorrectly or ambient temperature is too high.
Check wiring, power supply and wiring connections.
Adjust contrast in Options Menu.
Incorrect temperature reading.
Display or Current output is erratic.
Output is not zero when electrode is placed in non-chlorinated water.
4 to 20 mA output is incorrect.
Chlorine Sensor Troubleshooting (see 2630 sensor manual)
Incorrect readings
Faulty chlorine sensor. Bad sensor connection.
Electrical noise interfering with the measurement. Sensor malfunction.
Electrode not properly conditioned. Noise interfering with the measurement. Calibration incorrect.
4 to 20 mA is not scaled same as Loop device. Loop device is not scaled same as sensor.
Sensor conditioning time too short. Run for 4 hours before calibrating.
Membrane cap damaged or contaminated. Clean or replace cap.
Interference from water contaminants. See Specifi cations data.
Low fl ow rate. Check fl ow.
Air bubbles on membrane.
pH outside working range (See Specifi cations data).
Low or no electrolyte in sensor. Fill sensor with electrolyte.
Membrane cap loose. Inspect/tighten (do not use tools).
Only combined chlorine present when measuring free chlorine.
Sensor not making good contact with electronics.
Defective sensor. Replace.
No pH compensation being used.
Cl sensor not calibrated. Calibrate Cl sensor.
Check connections or replace sensor.
Ensure system is properly grounded. See Sensor Troubleshooting.
Condition new sensor for 4 hours. Cap replacement or electrolyte refi ll: 2 hours. Properly ground system. Replace sensor.
Re-span loop device to match sensor.
Install fl ow cell vertically with upwards ow.
Check pH.
Validate with DPD test.
Inspect and reconnect.
Manually enter pH value in Input Menu or calibrate pH sensor.
50
463X Chlorine Systems Manual
Page 51
9950-3 Error Messages
The 9950-3 error warnings are self-explanatory. An error message can appear under the following circumstances:
• User input value is out of range
• Poor electrical connection
• Sensor is not connected or detected
Error Messages Problem Possible Cause Remedies
Calibration Due
Out of Range Electrode is exceeding range.
Error too Large pH Calibration issue.
No pH Value pH value not appearing.
Check Preamp
No Probe
Calibration of the electrode is required.
Chlorine or pH electronics not detected.
Flashing over Channel.
Prode not detected in electronics.
• Temperature error
• Incorrect sensor type chosen in the Options Menu
• Two calibration points are too close together when calibrating pH or chlorine.
New electrode. Calibrate the electrode.
Next calibration reminder. See Calibration Menu.
Temperature of the sample is > 45 °C (113 °F).
pH is outside proper level of
5.5 to 8.2.
pH electrode at the end of life.
Dirty pH electrode. Clean and recalibrate electrode.
pH electrode in manual mode.
pH electrode faulty.
Electronics not plugged into the proper channel.
Wiring incorrect in connector. Cycle power Off and On.
Damaged Preamp electronics.
Chlorine and pH electronics are not inserted into correct channel.
Chlorine or pH electrode missing.
Damaged electrode. Replace the electrode.
Verify range of the application.
Replace faulty electrode. Calibrate the electrode.
Replace the electrode.
See Input Menu.
Check connection to pH electrode.
Check Chlorine Controller and electronics wiring.
Replace the electronics.
Verify Channel 1 and Channel 2 have the correct electronics.
Remove and reset the electrodes in proper electronics.
NOTE: If the type of chlorine sensor is changed, i.e. Free Chlorine or Chlorine Dioxide, the measurement type on the Input Menu should be changed.
463X Chlorine Systems Manual
51
Page 52
Troubleshooting
2630 Free Chlorine Electrode 2632 Chlorine Dioxide Electrode
Transmitter error messages related to calibration are detailed in the 9950-3 operation manual.
Problem Possible Causes Remedies
Condition for 4 hours minimum prior to initial calibration
DPD value must be greater than 0.2 ppm to calibrate
Inspect visually. Tap to remove bubbles. Mount at an angle
Condition for 4 hours minimum prior to initial calibration
Check for chloramine with appropriate DPD test
Inspect and reconnect
Inspect visually. Tap to remove bubbles. Mount at an angle
Replace membrane. Condition sensor for at least 2 hours and recalibrate.
Inspect and reconnect
Check 3-2650 Electronics connection to electrode (see 3-2650 manual for instructions). Make sure connections are dry. Check instrument hookup
Sensor cannot be calibrated. Output is lower than DPD test.
Sensor output very low
Unstable output from sensor
Sensor conditioning time too short
Chlorine content too low
Low fl ow rate Check to make sure fl ow rate is suffi cient
Air bubbles on electrode membrane
Low or no electrolyte in electrode Fill electrode with electrolyte
Organic chlorination agents present in water See Specifi cations data
Surfactants in water Remove surfactants and replace cap
Membrane cap coated Clean or replace membrane cap
Membrane cap loose Tighten or replace membrane cap
pH outside working range. See Specifi cations data
Electrode needs to be repolished Polish gold-plated electrode
Not enough electrolyte Check electrolyte level
Sensor conditioning time too short
Chlorine content too low Add chlorine to validate
Only bound chlorine present. No free chlorine
Electrode not making good contact with electronics
Electrode needs to be repolished Polish gold-plated electrode
Not enough electrolyte Check electrolyte level
Air bubbles on electrode membrane
Membrane damaged
Electrode not making good contact with electronics
Non-sensor problem
Troubleshooting - 2751-7 pH Electronics
Problem Possible Cause Remedies
After completing calibration procedure, the output values are inaccurate.
Transmitter Troubleshooting (see 9950-3 manual: 3-9950-XP.090)
pH Electrode Troubleshooting (see 2724 manual: 3-2724.090)
52
463X Chlorine Systems Manual
Insuffi cient time allowed for electrode stabilization during calibration.
Recalibrate, verify that test solutions are at room temperature and wait at least 30 seconds after placing electrode in solution.
Page 53
Specifi cations
9950-3 Chlorine Controller
General
Compatibility:
• Sensors: 3-2630-1 Free Chlorine Sensor 0.02 to 2 ppm 3-2630-2 Free Chlorine Sensor 0.05 to 5 ppm 3-2630-3 Free Chlorine Sensor 0.01 to 20 ppm 3-2632-1 ClO 3-2724-00 Flat pH Electrode,
See FCL and ClO2 system pH specifi cation
• Electronics: 3-2650-7 Amperometric Electronics 3-2751-7 pH Sensor Electronics
Materials:
• Case: PBT
• Panel gasket: Silicone rubber
• Window: Shatter-resistant glass
• Keypad: Silicone rubber
• Display: Backlit alphanumeric dot matrix LCD Update rate: 1 second Contrast: User selected, 5 levels
Electrode, 0.02-2 ppm
2
Performance
System Operational Ranges/Limits:
• 3-2630-1: Free Cl: 0.02 to 2 ppm, pH: 5 to 8.5
• 3-2630-2: Free Cl: 0.05 to 5 ppm, pH: 5 to 8.5
• 3-2630-3: Free Cl: 0.1 to 20 ppm, pH: 5 to 8.5
• 3-2632-1: ClO2: 0.2 to 2 ppm, pH: 4 to 11
• 3-2724-00: pH: 0 to 14 (8.5 when used with the 4630) Cl Temp. Range: 0 °C to 45 °C (32 °F to 113 °F)
Max. Cable Length:
• Digital (S3L): 30 m (100 ft) maximum
Electrical
Input Power: 12 to 24 VDC ±10% regulated,
250 mA max current
Sensor Power: 5 VDC ±1% @ 25 ºC, regulated
(provided by 9950-3)
Input Specifi cations
• One Digital (S3L) input from Free Chlorine or Chlorine Dioxide electrode
• One Digital (S
3
L) input from pH electrode
Output Specifi cations:
• Current Loop (4 loops provided)
• 4 to 20 mA, isolated, adjustable span, reversible with minimum and maximum endpoint adjustment.
• Chlorine, Temperature and pH
• Update Rate: 300 ms
• Max Loop impedance: 50 max. @ 12 V 325  max. @ 18 V 600  max. @ 24 V
Relay Module:
• 2 mechanical SPDT contacts with adjustable hysteresis and programmable High, Low, Off or Proportional and 4 binary inputs
• May be disabled if not used
• Max. voltage rating: 5 A @ 30 VDC 5 A @ 250 VAC, resistive load
• Time delay: Programmable from 0 to 99999 s
Environmental Requirements
• Operating Temperature: -25 °C to 120 °C (-13 °F to 248 °F) (transmitter only)
• Storage Temperature: -15 °C to 80 °C (5 °F to 176 °F)
• Relative Humidity: 0 to 95%, non-condensing
• Maximum Altitude: 4000 m (13,123 ft)
• Enclosure: NEMA 4X
Standards and Approvals
• CE, UL, CUL, WEEE
• RoHS Compliant
• Manufactured under ISO 9001, ISO 14001 and ISO 45001
China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
463X Chlorine Systems Manual
53
Page 54
Specifi cations
2650 DryLoc Amperometric Electronics 2751-7 pH Electronics
General
Compatibility:
• Sensors: All GF Amperometric DryLoc Sensors
• Instruments: GF 9950-3 Chlorine Controller
Mounting: Materials: PBT Cable: 4.6 m (15 ft) 3 conductor shielded, 22 AWG
• Max. Length: 30 m (100 ft)
DryLoc connection
Performance
• Accuracy < 5 nA or 1% of reading, whichever is greater @ 25 °C over full input range
• Temperature: ± 1.0 °C (PT1000) over full operation range (when calibrated at ambient temperature)
• Update Rate: 500 ms
• Range: ± 450 nA
• Resolution: 0.1 nA
Electrical
Input Specifi cations:
• Sensor: Raw signal
• Temperature: PT1000 RTD
Output Specifi cations:
• Digital (S3L): Serial ASCII, TTL level 9600 bps
General
Compatibility:
• Electrode: GF 2724-00 Flat pH Electrode
• Instrument: GF 9950-3 Chlorine Controller
Mounting: DryLoc connection Materials: PBT Cable: 4.6 m (15 ft) 3 conductor shielded, 22 AWG
Performance
• Accuracy: ± 0.03 pH @ 25 °C
• Resolution: 0.02 pH
• Response Time (includes electrode response): < 6 s for 95% of change
• Range: 0.00 to 14.00 pH (8.2 when used with the 4630)
• Resolution: 0.02 pH
Electrical
Input Specifi cations:
• Input Impedance: >1011
• Temperature Drift: ± 0.002 pH per ºC
• Input Resolution: 0.02 pH, 0.3 ºC
Output Specifi cations:
• Digital (S3L): Serial ASCII, TTL level 9600 bps
Power Supply Input:
• Digital (S3L): 5 to 6.5 V ± 10%, 3 mA max (from the 9950-3)
Environmental
• Storage Temp.: -20 °C to 50 °C (-4 °F to 120 °F)
• Operating Temp.: 0 °C to 85 °C (32 °F to 185 °F) (electronics only)
• Relative Humidity: 0 to 95%, non-condensing (no electrode connected)
• Enclosure: NEMA 4X/IP65 with electrode installed
• Shipping Weight: 0.64 kg (1.41 lb)
Standards and Approvals
• CE, WEEE
• RoHS Compliant
• Manufactured under ISO 9001, ISO 14001 and ISO 45001
China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
Power Supply Input:
• Digital (S3L): 5 to 6.5 V ± 10%, 3 mA max (from the 9950-3)
Environmental
• Storage Temp.: -20 °C to 50 °C (-4 °F to 120 °F)
• Operating Temp. (Electronics Only):
0 °C to 85 °C (32 °F to 185 °F)
• Relative Humidity: 0 to 95%, non-condensing
• Enclosure: NEMA 4X/IP65 with electrode installed
• Shipping Weight: 0.64 kg (1.41 lb)
Standards and Approvals
• CE, WEEE
• RoHS Compliant
• Manufactured under ISO 9001, ISO 14001 and ISO 45001
China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
54
463X Chlorine Systems Manual
Page 55
Specifi cations
2630-X Amperometric Free Chlorine Electrode 2632 Amperometric Chlorine Dioxide Electrode
General
Compatibility
Flow Cells ............................. GF 3-4630.392
Electronics ............................GF 3-2650-7
Mounting .................................. GF DryLoc connection
Materials
Housing ................................CPVC
Membrane
Free Chlorine .................... PTFE
Chlorine Dioxide ...............PTFE
O-ring ...................................FKM
Electrode ..............................Gold-plated (working)
Silver halide (counter-reference)
Wetted Materials ...................... PVC, PTFE, FKM, Nylon,
Silicone
Performance
Electrode
Repeatability ........................ ± 0.08 ppm (mg/L) or 3% of
selected range whichever is less Slope
Free Chlorine ....................... 15 to 85 nA/ppm (mg/L) @ 25 ºC
Chlorine Dioxide ..................40 to 200 nA/ppm (mg/L) @ 17 ºC
Response time, T90 ............< 2 minutes
Sensor Conditioning 4 hours max. before calibration (new/fi rst time start) 2 hours max. (subsequent start-ups)
System (including electronics and instrument)
Accuracy ..............................< ±3% of electrode signal after
calibration
Resolution ............................ 0.5% of electrode range
Temp. Element ..................... PT1000, Class B with
measurement range of -50 ºC to
150 ºC (-58 ºF to 302 ºF)
(measurement only, not operating
specifi cation)
Operational Ranges and Limits
Free Chlorine ....................... 3-2630-1: 0.02 to 2 ppm (mg/L)
3-2630-2: 0.05 to 5 ppm (mg/L)
3-2630-3: 0.1 to 20 ppm (mg/L)
Free Chlorine pH .................5.5 to 8.2 pH
Chlorine Dioxide .................. 3-2632-1: 0.02 to 2 ppm (mg/L)
Chlorine Dioxide pH ............. 4.0 to 11.0 pH
Temperature .............................0 °C to 45 °C (32 °F to 113 °F)
Maximum Pressure
Membrane ............................0.48 bar @ 25 °C (7 psi @ 77 °F)
Flow Velocity Across Membrane Surface:
Minimum ..............................15 cm/s (0.49 ft/s)
Maximum .............................30 cm/s (0.98 ft/s)
Cross Sensitivity
Free Chlorine ....................... ClO
Chlorine Dioxide ..................FCl, ozone
Chemical Compatibility ............ < 50% ethanol/water < 50%
Environmental
Temperature
Operating .............................0 ºC to 45 ºC (32 ºF to 113 ºF)
Storage (dry) ........................-10 ºC to 60 ºC (-4 ºF to 140 ºF)
Relative Humidity .....................0 to 95% indoor/outdoor non-
Shipping Weight.......................0.14 kg (0.30 lb)
Standards and Approvals
• CE, WEEE
• RoHS Compliant
• Manufactured under ISO 9001, ISO 14001 and ISO 45001 China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
, ozone, bromine
2
glycerol/water
condensing to rated ambient
463X Chlorine Systems Manual
55
Page 56
Specifi cations
2724 DryLoc pH Electrode
General
Compatibility: ..................... GF 2751-7 pH/ORP Sensor
electronics
Process Connection ........... ¾ in. NPT or ISO 7/1 R¾ threads
Wetted Materials ................ PPS, glass, UHMWPE, FKM
Performance
Effi ciency ........................ >97% @ 25 °C (77 °F)
Range ............................. Free Chlorine 5.5 to 8.2 pH
Environmental
Pressure
Operating ........................ -10 ºC to 65 °C (14 °F to 149°F),
Temperature
Storage ........................... 0 °C to 50 °C (32 °F to 120 °F).
or GF fl ow fi ttings
Effi ciency indicates the "wellness" of a new electrode. Effi ciency is measured by comparing the actual slope (mV/pH) at 25 °C to the theoretical output of 59.16 mV/pH. An effi ciency of 97% to 100% is equivalent to a slope of 57.39 to 59.16 mV/pH.
Chlorine dioxide 4.0 to 11 pH
0 to 6.9 bar (0 to 100 psi) 65 °C to 85 °C (149 °F to 185°F), Linearity derated 6.9 to 4.0 bar (100 to 58 psi)
CAUTION: The electrode glass will be shattered if shipped or stored at temperatures below 0 °C (32 °F). The performance life of the electrode will be shortened if stored at temperatures above 50 °C (120 °F).
Operating ........................ -10 ºC to 85 ºC (14 ºF to 185 ºF)
Rel. Humidity ...................... 0 to 95% indoor/outdoor non-
Shipping Weight ................. 0.25 kg (0.55 lb)
Standards and Approvals
• CE, WEEE
• RoHS Compliant
• Manufactured under ISO 9001, ISO 14001 and ISO 45001 China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
56
463X Chlorine Systems Manual
condensing to rated ambient
Page 57
Notes
463X Chlorine Systems Manual
57
Page 58
Ordering Information
463X Chlorine Analyzer
Mfr. Part No. Code Description
3-4630-13 159 001 949 Chlorine panel, free chlorine sensor (0.02 to 2 ppm) w/ sensor electronics, pH sensor w/ electronics
3-4630-23 159 001 950 Chlorine panel, free chlorine sensor (0.05 to 5 ppm) w/ sensor electronics, pH sensor w/ electronics
3-4630-33 159 001 951 Chlorine panel, free chlorine sensor (0.01 to 20 ppm) w/ sensor electronics, pH sensor w/ electronics
3-4632-12 159 001 952 Chlorine panel, chlorine dioxide sensor (0.02 to 2 ppm), with sensor electronics, no pH sensor
3-4632-13 159 001 958
Accessories and Replacement Parts
3-9950-3 159 001 954 Chlorine Controller
3-9950-5 159 001 956 Chlorine Monitor, no Relay or Output modules
3-2630-1 159 001 746 Free Chlorine electrode, 0.02 to 2 ppm (mg/L)
3-2630-2 159 001 662 Free Chlorine electrode, 0.05 to 5 ppm (mg/L)
3-2630-3 159 001 747 Free Chlorine electrode, 0.01 to 20 ppm (mg/L)
3-2632-1 159 001 767 Chlorine Dioxide electrode, 0.02 to 2 ppm (mg/L)
3-2724-00 159 001 545 pH electrode, Flat Glass, Pt1000 Temp Element, ¾ in. MNPT
3-2751-7 159 001 957 pH - Inline Electronics, Digital (S
3-2650-7 159 001 670 Chlorine - In-line Amperometric Electronics, digital (S3L), 4.6 m (15 ft) cable
3-4630.390 159 001 688 Rebuild kit, O-rings, boots, screws, 1 fi lter screen
3-4630.391 159 001 689 Pressure regulator with 1 spare fi lter screen
3-4630.392 159 001 690 Acrylic fl ow cell complete with all components and connections
3-4630.393 159 310 162 Flow Switch Kit, PP
7300-0024 159 001 693 24 VDC Power Supply
3-2630.391 159 001 674 Electrolyte Kit, 30 ml bottle with syringe and needle
3-2632.391 159 310 160 Chlorine Dioxide electrolyte, 30 mL (2)
3-2630.394 159 310 164
3-2630.398 159 310 166 Free Chlorine Sensor maintenance kit - (2) electrolyte, (2) PTFE membranes, (2) Silicone Bands, and
3-2632.398 159 310 165 Chlorine Dioxide maintenance kit - (2) electrolyte, (2) PTFE membranes, (2) Silicone Bands, and
1220-0021 159 801 182 O-ring FKM
3-0700.390 198 864 403 pH Buffer Kit (1 each 4, 7, 10 pH buffer in powder form, makes 50 mL of each)
3822-7004 159 001 581 pH 4.01 buffer solution, 1 pint (473 mL) bottle
3822-7007 159 001 582 pH 7.00 buffer solution, 1 pint (473 mL) bottle
3822-7010 159 001 583 pH 10.00 buffer solution, 1 pint (473 mL) bottle
3-2700.395 159 001 605 Calibration kit: included 3 polypropylene cups, box used as cup stand, 1 pint pH 4.01, 1 pint pH 7.00
3-2759 159 000 762 pH/ORP System Tester (adapter cable sold separately)
3-2759.391 159 000 764 2759 DryLoc Adapter Cable (for use with 2751-7)
3800-5000 159 838 107
3-2700.397 159 001 870
3-2700.398 159 001 886
Chlorine panel, chlorine dioxide sensor (0.02 to 2 ppm), with sensor electronics, pH sensor w/ electronics
3
L), 4.6 m (15 ft) cable
Free Chlorine and Chlorine Dioxide Replacement PTFE membrane (1)
Polishing Paper
Polishing Paper
3.0M KCl Storage Solution for pH and ORP, 1 pint (473 ml) bottle
Protective Cap for pH/ORP electrodes, 5 pieces
Lubricant Kit
Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfps.com • Or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfps.com
3-4630-X3.090 Rev 0 03/21 English © Georg Fischer Signet LLC 2021
Loading...