Service Manual - Test Procedure
VitaGuard®
Projekt ID: 0027
VG 3100 / VG 2100 / VG 310
8 Equipment and Materials
The equipment listed below marked R is the equipment used by getemed AG. Similar
models from other manufacturers can be used as long as the performance is similar.
The model numbers of the equipment listed with getemed AG as manufacturer
correspond with the ordering (REF) numbers, should the distributor wish to purchase
the equipment from getemed.
Equipment Model / REF Manufacturer
Power adapter NA 3000-2 7344 1101 getemed AG
ECG patient cable 7341 1001 getemed AG
Masimo SpO2 cable PC08 70257 getemed AG, Masimo
Masimo SpO2 Sensor DC I 70254 getemed AG, Masimo
PC Interface cable (USB-Kabel) 7341 2001 getemed AG
VitaWin software VitaWin 3 getemed AG
R 1A / 3-6V power supply HM7042 Hameg
R Digital Multimeter Model 87 Fluke
R ECG/Respiration patient simulator 214B DNI Nevada Inc.
R SpO2 patient simulator Oxitest Plus DNI Nevada Inc.
R Stop watch - R Safety tester Unimet 1000 ST Bender
R Test PC with Windows 2000 or XP - -
0027-TD-0027-Rev C
9 Procedure
The procedure is divided into 4 sections:
· Visual inspection of the monitor
· Functional measurements
· Safety inspection
· Visual inspection of accessories
For easy reference, the numbers of this procedure correspond with the numbers in
the test report.
Before beginning the test procedure, the revision stand of the monitor should be
checked. The distributor should contact getemed on a regular basis at
service@getemed.de in order to determine if changes are necessary.
Should the monitor under test show any signs of malfunction, then it must be returned
immediately to getemed AG for inspection.
The test reports must be archived for at least 10 years and made available to
getemed AG on request.
The distributor shall maintain a database showing when a monitor was last tested and
by whom.
The remaining part of section 9 explains how to perform the test procedure. To
correspond with the test report, numbering will begin with 9.7.1.
0027-TD-0027-Rev C-B5.1-Service Manual 18 Month Regular Inspection Test Procedure English.doc
Page 2 of 11
Service Manual - Test Procedure
VitaGuard®
Projekt ID: 0027
VG 3100 / VG 2100 / VG 310
9.7.1 Visual inspection of the monitor
The first section of the test report is the visual inspection of the monitor.
9.7.1-1 Labels
All three labels listed in box 1 must be readable and free from damage. If this is the
case, then box 1 can be ticked-off with o.k.
The following labels are readable and not damaged:
· Top label with keys (key panel)
1
· Label on the back of the monitor with serial number and
safety warnings; warning label in battery compartment.
· Label stripe at the front marking all inputs
9.7.1-2 Connectors
The connectors listed in box 2 must be fitted tightly. The connector pins must be
visually inspected to ensure that none of them are damaged. If this is the case, then
box 2 can be ticked-off with o.k.
The following connectors are fitted tightly and not damaged:
· ECG input (VG 3100 and VG 2100 only)
· NA 3000-2 power supply input
2
· SpO2 input (VG 3100 and VG 310 only)
· USB input
· AUX input
9.7.1-3 Casing and LCD display
0027-TD-0027-Rev C
o.k.
o.k.
The monitor casing must not be cracked or have any significant scratches. The LCDdisplay must also be free from damage. If this is the case, then box 3 can be tickedoff with o.k.
3 The monitor casing and the LCD display are not damaged. o.k.
9.7.2 Functional measurements
The second section of the test report involves measuring parameters and performing
functional tests.
The external power adapter NA 3000-2 used must be modified to measure the current
flowing to the monitor. To do this, cut the positive wire going to the monitor and attach
two 4 mm connector jack plugs to the loose ends. These can then be inserted into a
multimeter.
It is assumed that the default settings are selected before beginning the test
procedure.
0027-TD-0027-Rev C-B5.1-Service Manual 18 Month Regular Inspection Test Procedure English.doc
Page 3 of 11
Service Manual - Test Procedure
VitaGuard®
Projekt ID: 0027
VG 3100 / VG 2100 / VG 310
0027-TD-0027-Rev C
VitaGuard
Monitor
Comp.
Battery
Digital-
Multimeter
(1)
Power supply
3-6V / 2A
Digital-
Multimeter
NA 3000-2
SpO2Sensor
and
simulator
(3)
SpO2 ECG
Power
AUX
(4)
Patient
simulator
(2)
Figure 1 – Current Measurement Set-Up
VitaGuard
Monitor
Battery
Comp.
SpO2 ECG Power
AUX
Power supply
3-6V / 1A
SpO2
Simulator
PatientSimulator
Figure 2 – Alarm Measurement Set-Up
VitaGuard
Monitor
Battery
Comp.
SpO2 ECG
USB
Power AUX
Power supply
3-6V / 1A
PC
VitaWin
Figure 3 – PC Measurement Set-Up
0027-TD-0027-Rev C-B5.1-Service Manual 18 Month Regular Inspection Test Procedure English.doc
Page 4 of 11