GERATHERM Micromesh GT-7002 Instruction Manual

Page 1
GB
Nebulizer with Micromesh
Technology
Instruction manual
GT-7002
Page 2
24
GB
Thank you for your confi dence in Geratherm and congratu-
lations on your purchase! By purchasing the Geratherm micromesh nebulizer you have acquired a quality product by Geratherm
®
. Please read these instructions for use carefully to use the device without problems and enjoy your Geratherm® neb­ulizer for a long time.
Best Regards,
The Geratherm
®
Team
Page 3
25
GB
Table of Contents
Item Supplied and Packaging ....................................... 26
Features ........................................................................ 26
Descriptions of Parts ...................................................... 27
Safety Precautions ........................................................ 28
Important Information about Inhalation
Therapy and Mesh Technoloy, purpose of use .............. 29
Important Information about
Inhalants/Medications ................................................... 29
Preparation before Use/ Operation ............................... 30
Battery Replacement ............................................. 30
Operating Instructions ................................................... 30
Important Information before Use .......................... 30
Unit Assembly and Usage .......................................31
Inhalation with Mouth Piece and Masks ................ 33
Cleaning and Disinfection .............................................. 34
Before Cleaning ...................................................... 34
Cleaning and Disinfection of Nebulizer
Parts/ Accessories .................................................. 34
Cleaning and Disinfection of Nebulizer Unit ........... 35
Maintenance .................................................................. 36
Changing Accessories .................................................... 37
Troubleshooting ............................................................. 38
Technical Data ............................................................... 39
Symbol Information ....................................................... 40
Disposal ......................................................................... 41
Limited Warranty ............................................................ 41
Warranty Card ................................................................ 42
Page 4
26
GB
Before you start to use the Geratherm® micromesh, please check if the following items are included in the package:
• 1 nebulizer Geratherm
®
micromesh GT-7002 with nebulizer unit, mesh cap, adapter (for masks and mouth piece) and cover
• 1 mouthpiece*
• 1 PVC mask adult*
• 1 PVC mask children*
• 2 “AA” batteries 1.5 V
• 1 storage bag
• 1 instruction manual
* Note
For reasons of hygiene, please use the enclosed masks and mouth piece for one specifi c in- dividual only and do not exchange them. The masks and the mouth piece must be cleaned after each use as described in the section entitled “Cleaning and Disinfection of Nebuliser Parts/ Accessories“. Only silicon masks, which are available from specialist suppliers, can be disinfected by boiling or by means of vaporisation (at 100 °C for approx. 10 minutes).
Use only the parts listed above; use of other parts may con­stitute a hazard.
Items Supplied and Packaging
Discharge-proof - fully operative even when positioned on an inclined surface (up to 45°)
No warming of medicine, thus preventing damage to the structure of the medication
Low power consumption
Convenient design – compact and lightweight, ideal for
travelling
Silent operation – thanks to the micromesh technology
Short treatment time – with high nebulization rate
No medication cups necessary (no disposable items)
Features
Page 5
27
GB
Descriptions of Parts
Unit lidPVC masks
(adult + children)
2 AA Alkaline batteries
Mouth piece
Mesh cap
adapter (for masks
and mouth piece)
Medication
cup
Indicator:
green: usage red: battery needs to be replaced green and red: reset the vibration rate
ON/OFF
button
Battery cover
Nebulizer unit
Page 6
28
GB
• Use the equipment only for its intended use as described in this manual. Do not use attachments not recommend-
ed by the manufacturer.
• Use the equipment with medications only under the in-
struction of your physician.
• Do not use the equipment if it has any damaged parts, or it has fallen into water/ other liquids.
• Remove the batteries when not in use for a long period
of time to avoid circuit damage due to leakage from bat-
tery.
• To use the equipment for the fi rst time, or after storing it for an extended period, be sure to clean and sterilize
all necessary parts as described in the cleaning instruc-
tions.
• Do not use while bathing.
• The unit should not be left unattended while plugged in
or turned on.
• Keep the unit out of reach of small unsupervised children.
The small parts detached from the device may result
children choking from inhaling or swallowing.
• The accessible materials used in the device will not
cause the potential allergic reactions to skin.
• Do not try to modify the device to prevent any dangers.
• Do not place the device in liquid, nor put it where it could fall into liquid. If the device becomes wet, unplug the
device before touching it.
• Do not use the device if it is not working properly, or if it has suffered any damage.
• Do not subject the nebulizer or the mesh cap to any im­pacts and do not drop them.
• Before using the nebulizer for the fi rst time, after a pe- riod of no use, and after every use, please clean all re­movable parts of the nebulizer.
• Before storing, please make sure all parts are clean and dry.
• Do not expose the nebulizer to direct sunlight, high tem­peratures or high humidity.
• The device is intended for private domestic use only.
• To guarantee an effective inhalation, just inhale with mask or mouth piece.
Safety Precautions
Page 7
29
GB
Important Information about Inhalation The­rapy and Mesh Technology, Purpose of Use
Geratherm® micromesh is suitable for the treatment of asth­ma, chronic bronchitis and other respiratory conditions af­fecting the upper and lower airways, e.g. the common cold.
The medication is nebulized by means of an ultra-modern
vibrating membrane or mesh.
This membrane is partially porous, enabling tiny droplets of
the inhalant to penetrate it as soon as it begins vibrating.
The key component of this system is the mesh cap.
We recommend replacing the cap approx. after six months when in regular use in order to ensure effi cient nebuliza- tion.
The device is ideally suitable for inhalation at home or when
travelling. It guarantees highly effective, fast-acting inhala­tion treatment for patients of all ages. Carry out inhalation in a quiet and relaxed state, and inhale slowly and deeply, so that the medication can penetrate to the fi ne, deep bronchial tubes. Exhale normally.
Important Information about Inhalants/ Medications
Please only use the medicinal products/medication pre­scribed for you by your doctor.
The Geratherm
®
micromesh nebulizer is suitable for use with all common aerosol products for medical use with the exception of
- viscous solutions (high viscosity)
- suspensions (products with solid constituents)
- volatile oils
Attention: Use only water-soluble medicines or aerosol products which have an alcohol base.
Page 8
30
GB
The nebulizer is powered by two 1.5V AA alkaline batteries.
Battery Replacement
Replace batteries when the red indicator blinks. Before re­placing the batteries, make sure that the nebulizer is turned off.
1. Press the buckle on the battery cover and lift up to remove the cover.
2. Remove the used batteries. Insert two new ones. Cor­rectly align the polarities (+ and -) with battery indica­tion marks on the device.
3. Close the battery cover by pressing down fi rmly until a click is heard.
Note:
Remove the batteries if not using the device for three months or longer. Discard the used batteries according to your local regula­tions.
Important Information Before Use
• Ensure that all parts are clean or disinfect if necessary and dry.
• Before starting treatment, talk to your physician about the duration, dosage and frequency of use of the neb­ulizer.
• The green indicator is on during nebulization.
• The device turns off automatically after 15 minutes of non-use (without the ON/OFF button being pressed).
• Do not remove the medication cup cover from the unit during nebulization.
Preparation before Use / Operation
Operating Instructions
Page 9
31
GB
Operating Instructions
Unit Assembly and Usage
1. Remove the unit lid and the mesh cap of the nebulizer. Remove the mesh cap by gently pressing on the pro­truding plastic fl ap. The cover can then be pulled upwards and lifted off.
2. Slowly pour the prescribed medication into the medica­tion cup.
The maximum capacity of
the medication cup is 6 ml.
Please note that the medi-
cation must not exceed this maximum level marking (6 ml).
3. Slide the mesh cap into the unit, align the 2 knobs with the cutouts in the direction indicated by arrow 1. Then press the cover down until it clicks into place.
4. Attach the adapter on the mesh cap. Please pay atten­tion that the knobs of mesh cap are in the same position as the cuttings of adapter. Click the adapter into mesh cap. Now attach the unit lid.
1
2
1
2
Page 10
32
GB
5. Attach the mask or the mouthpiece to the adapter:
6. Take the nebulizer in your hand and press the ON/OFF button to begin nebulization. You can select one of two delivery modes. Please read the following section con­cerning this.
7. To end the nebulization, press the ON/OFF button to turn it off.
Delivery mode
The nebulizer provides you with two delivery modes accord-
ing to your needs:
automatic or manual.
Automatic delivery of aerosol
1. Press and release the ON/OFF button. The unit turns on and starts to deliver the prescribed medication continu­ously for 15 minutes. The nebulizer will turn off automati­cally after 15 minutes of non-use (without the ON/OFF button being pressed).
2. To stop the treatment at any time, press the ON/OFF but­ton again to turn off the device.
Operating Instructions
or
Page 11
33
GB
Operating Instructions
Manual delivery of aerosol
In the manual mode, the duration of nebulization is control-
led by holding down the ON/OFF button.
1. Press the ON/OFF button down for more than 2 seconds. The unit will start with nebulization.
2. When you want to fi nish the process, release the ON/ OFF button and the device will stop delivering medica­tions.
Inhalation with Mouth Piece and Masks
Inhalation using the mouthpiece
1. Place the mouthpiece in your mouth with the lips sealed around the mouthpiece.
2. Switch on the unit.
3. During treatment, take slow, deep breaths through your mouth. If possible, hold each breath for two to three sec­onds before breathing out. This allows the medication to settle into the respiratory system.
4. Continue the treatment until the medication has been used up (about 10 to 15 minutes).
5. If you become dizzy, stop the treatment and rest for about fi ve minutes. Then continue the treatment, but try to breathe more slowly. If these symptoms continue with future treatments, inform your physician.
6. Clean and disinfect if necessary like described below.
Inhalation Using the Mask
1. Place the mask over your mouth and nose.
2. Switch on the unit.
3. During treatment, take slow, deep breaths through your nose. If possible, hold each breath for two to three sec­onds before breathing out. This allows the medication to settle into the respiratory system.
4. Follow the procedures described under “Inhalation using the mouthpiece”.
Page 12
34
GB
Please clean your device after each use according to the instructions below.
Note: After two uses (inhalations) please rinse the neb-
ulizer with water by fi lling the medication cup with water in order to clean the membrane.
Before Cleaning
Make sure that the nebulizer is turned off before disas-
sembly.
1. Remove the unit lid, mask, mouthpiece and adapter from the unit. You could remove the adapter from mesh cap by slightly turning in clock­wise direction.
2. Remove the mesh cap by gently pressing the fl ap and lifting up in the direction indicated the arrows.
3. Discard any remaining medications (pay attention to statutory provisions).
Cleaning and Disinfection of Nebulizer Parts/ Accessories
Cleaning after every use
Clean the mask, mouthpiece, unit lid, adapter and mesh cap after each use.
1. Disassemble the components from the unit.
2. Rinse all above-named parts under a small stream of tap water.
3. Gently wipe off excess water with a clean dry cloth and allow to air dry. Possible remained drops on the mesh surface could remove by easy shaking of mesh cap.
Cleaning and Disinfection
1
2
Page 13
35
GB
Cleaning and Disinfection
Daily disinfecting
Disinfection by boiling:
1. Silicon masks and mesh cap can be disinfected by boil­ing in distilled water.
2. After 10-20 minutes of boiling, gently wipe off excess water with a clean dry cloth and allow to air dry on a clean paper towel.
Disinfection by using alcohol:
1. PVC mask, mouthpiece, unit lid and mesh adapter can be disinfected by using medical alcohol (Ethanol 75% dilution with water).
2. Disinfect the above-named parts by wiping with a cloth moistened with medical alcohol (Ethanol 75% dilution with water).
3. Let them air dry on a clean paper towel.
Note: Only silicon mask can be disinfected by boiling, PVC mask just can be disinfected by using medical alcohol 75%
(dilution with water).
Cleaning and Disinfection the Nebulizer Unit
Cleaning after every use
1. Use a cloth moistened with water to clean the medication cup and outer casing of the unit.
2. Wipe off the unit with an unused dry cloth and let it air dry.
Daily disinfecting
1. Use a cloth moistened with medical alcohol (Ethanol 75% dilution with water) to disinfect the medication cup
and outer casing of the unit.
2. Let them air dry on a clean paper towel.
Page 14
36
GB
CAUTION
• Do not immerse the nebulizer unit in any liquid and do
not hold it under running water. Be careful to ensure that no water penetrates the nebulizer unit.
• Do not clean the nebulizer or its component parts in a
dishwasher.
• Do not allow any moisture to come into the battery com-
partment.
• Do not use any other disinfectant like described above to
clean the unit or its components.
• Do not use a hairdryer to dry any parts.
When using a new mesh cap for the fi rst time, you need to reset the vibration rate:
1. Pour water into the medicine cup (pay attention to maxi-
mum marking).
2. Attach the mesh cap to the nebulizer unit.
3. Take the batteries out.
4. While holding down the ON/OFF button, reinsert the bat­teries and close the battery cover.
5. Keep the unit upright and then release the ON/OFF but­ton. The green and red indicator light up, indicating that the nebulizer is resetting the vibration rate.
6. After a few seconds, the nebulizer automatically turns off.
The vibration rate resetting is fi nished. The nebulizer is
ready for use.
The unit must be kept upright during this procedure.
Cleaning and Disinfection
Maintenance
Page 15
37
GB
Changing Accessories
We recommend that the mesh cap be replaced after about
6 months of use if the unit is used daily (once per day), so
as to ensure effi cient nebulization. This recommendation is
based on a content volume of 2 ml and an inhalation time
of about 8-15 minutes.
The possible use time of 6 months could shorten if the unit
is used often daily.
The accessories should be replaced after continuous use.
Accessories of the nebulizer micromesh involve: mesh cap, mesh adapter (for masks and mouth piece), PVC masks (for adult and children), silicon masks (for adult and chil­dren), and mouth piece. Accessories are available in spe­cialist shop.
description article number EAN
mesh cap 9730002017 4018674000215
PVC adult mask
9730002012 4018674000161
PVC children mask
9730002011 4018674000154
silikon adult mask
9730002014 4018674000185
silikon children mask
9730002013 4018674000178
mouth piece 9730002016 4018674000208
connecting piece/ adapter
9730002015 4018674000192
Page 16
38
GB
Troubleshooting
What happened
Causation Correction
The indicator turns red.
The batteries have run out of power.
Replace the bat­teries.
The mesh cap is dirty.
Clean the mesh cap. If the problem persists, please replace with a new mesh cap.
The device does not work when it is turned on
No power.
Check the batter­ies and replace them if neces­sary. Pay atten­tion to polarity.
The indicator is green but no aerosolis produced or the delivery is slower than usual.
The medication cup is empty.
Fill medication cup and turn on the nebulizer.
The mesh cap is broken.
Replace with a new mesh cap.
The medication cup or the mesh cap is dirty.
Clean those parts or replace mesh cap.
There is medication in the medication cup and the ON/OFF but­ton has been pressed, but no aerosols are being produced.
No power supply.
Check the batter­ies and replace them if neces­sary. Pay atten­tion to polarity.
The mesh is covered with liquid.
Check if the mesh cap is cov­ered with liquid. If yes, shake off the liquid and then try again.
Page 17
39
GB
Technical Data
Dimensions 76 mm x 60 mm x 135 mm
Weight 190 g
Power source 2 x AA alkaline batteries (1.5 V) Battery life Over 7 hours operating Power consumption 1.5 W
Vibration frequency 185.5 kHz
Nebulization rate >0.25 ml/min Medication capacity 6 ml
Operating conditions 10 °C to 40 °C,
30 to 85 % relative humidity
Storage conditions -25°C to 70°C,
up to 93% relative humidity
IP code (protection rating): IP22 (effective against bigger solid bodies; protected gainst dripping water)
Automatic function Automatic power shut-off if system idles for 15 minutes
Power consumption Normal mode: less than 1.5 W
Standby mode: less than 0.3 mW
Quality standard The device complies with the Council Directive
93/42/EEC concerning medical devices. A quality manage-
ment system certifi cated under DIN EN ISO 13485 guaran-
tees compliance with the required quality criteria. EN 13544-1 respiratory therapy equipment – part 1: neb-
ulizing systems and their components
Distribution of particle size
(with mask and mouth piece):
1,0
0,8
0,5
0,3
0,0
Partikelgröße/
Particle size [μm]
Volumenanteil/
Volume fraction
0 5 10 15 20
Page 18
Symbol Information
Symbol Referent
Consult instructions for use
Use by/expiry date for masks
Batch code for nebulizer
Manufacturing date of masks
Manufacturer
Serial number
Caution, consult acompanying docu­ments
Do not dispose of as household waste
Type BF Equipment
40
GB
Page 19
GB
Disposal
41
Adhere to the applicable regulations when disposing of the
device and packaging. This product must not be disposed of in domestic waste.
All users are obliged to hand in all electrical or electronic
devices, regardless of whether or not they contain toxic
substances, at a municipal or commercial collection point
so that they can be disposed of in an environmentally ac-
ceptable manner.
Please remove the batteries before disposing of the de-
vice/unit.
Do not dispose of old batteries with your house­hold waste, but at a battery collection station at a
recycling site or in a shop.
This ultrasonic nebulizer is guaranteed for 3 years from the
date of purchase.
If your nebulizer does not function properly due to defective
parts or assembly we will repair it free of charge. All parts
are covered by this warranty, except accessories which are
subject to normal wear and tear. The warranty does not
cover damage to your unit due to improper handling.
Limited Warranty
Page 20
The Geratherm® micromesh nebulizer is guaranteed for
three years from the date of purchase against manufactur­er’s defect if it is returned to the dealer from whom it was purchased.
During this warranty period the unit will be repaired or re­placed free of charge if the defect is due to faulty design or manufacture.
This guarantee does not cover any damage or malfunction-
ing that is caused through improper handling, through use otherwise than in accordance with the instructions in this manual or through unauthorised maintenance attempts.
Your local dealer cannot declare this manufacturer’s war-
ranty null and void, but may extend it by additional guaran-
tees at the discretion and at the expense of the dealer.
Model no.: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Serial no.: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Date of purchase: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dealer’s signature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Invoice no. / Receipt no.: . . . . . . . . . . . . . . . . . . . . . . . . . .
Dealer’s name and address: . . . . . . . . . . . . . . . . . . . . . . .
Geratherm Medical AG Fahrenheitstraße 1 D-98716 Geschwenda Germany
Tel: +49 36 205 980
Fax: +49 36 205 116 www.geratherm.com
42
Warranty Card
Loading...