Geotech Geopump I Series, Geopump II Series Installation And Operation Manual

Installation and Operation Manual
Geopump
Peristaltic Pump
Rev. 2 5/24/02 Part # 81350020
1
TABLE OF CONTENTS
CHAPTER 1: SYSTEM DESCRIPTION............................................................... 2
F
UNCTION AND THEORY
S
YSTEM COMPONENTS
CHAPTER 2: SYSTEM INSTALLATION............................................................. 3
CHAPTER 3: SYSTEM OPERATION.................................................................. 4
CHAPTER 4: SYSTEM MAINTENANCE............................................................. 5
CHAPTER 5: SYSTEM TROUBLESHOOTING................................................... 6
CHAPTER 6: SYSTEM SPECIFICATIONS ......................................................... 7
CHAPTER 7: SYSTEM SCHEMATIC.................................................................. 8
CHAPTER 8: REPLACEMENT PARTS LIST...................................................... 9
THE WARRANTY................................................................................................ 10
EQUIPMENT RETURN POLICY......................................................................... 10
EQUIPMENT DECONTAMINATION................................................................... 10
....................................................................................... 2
........................................................................................ 2
Chapter 1: System Description
Function and Theory
The Geotech Series I and II Peristaltic Pumps were designed for single and multi-stage pressure or vacuum pumping of liquids for field or laboratory use. Because they operate to a depth of 27 feet at sea level, they are ideally suited for sample removal from shallow wells and all surface water sources. They operate by mechanical peristalsis, so the sample comes in contact only with the tubing. This allows for sample integrity as well as easy cleaning and replacement. With the optional stainless steel tubing weight, tubing can be lowered to a specific depth without curling or floating on the surface of the water. Differences between the two models affect only the number of pump heads which may be used with the Geopump at one time, and the speed or speeds at which the pump heads operate. Therefore, instructions for general operations will be covered only once.
System Components
SERIES I Peristaltic Pumps are available in AC only, DC only, or an AC/DC
combination. These units have one pumping station which can also be piggy­backed for multi-station pumping. They are variable speed from 60 RPM to 350 RPM.
SERIES II Peristaltic Pumps are available in AC only, DC only, or an AC/DC combination. They have two pumping stations which can also be piggy-backed. The first pumping station is rated at 30 RPM to 300 RPM and the second station at 60 RPM to 600 RPM. Each pumping station works in conjunction with the other.
2
Chapter 2: System Installation
Standard Pump Head Instructions
Separate pump halves. Hold the pump head as shown, with the rollers in the 2, 6, and 10 o’clock positions and the rotor shaft facing down.
Place the tubing around the rollers.
Turn rotor counterclockwise until tubing has completely surrounded the rotor.
Tubing is now in place. Position other pump half onto the motor shaft and snap shut. Be careful not to pinch tubing between plastic halves.
3
Chapter 3: System Operation
The pump kit arrives packed in the hard shelled peristaltic pump carry case with the pump head properly attached to the pump (purchased separately). See section on pump parts.
To put into service:
Remove the pump from the case, be sure pump is turned “OFF” (red, open circle)
For AC/DC combination units, plug in the appropriate power cord into the outlet in the back of the pump & other end of the power cord into the power source.
Insert the tubing into the pump head.
Put one end of the tuning into the sample source (well, river, ditch, lagoon, etc.) and the other end into the sample container.
Determine the desired direction of flow and turn the direction toggle switch in that direction.
Turn the pump “ON” (the black filled circle).
When pumping has begun, the speed dial can be adjusted to the fluid pumping speed desired.
4
Chapter 4: System Maintenance
Peristaltic Pump: Maintenance Tips
The Geotech Peristaltic Pump has a strong reputation for durability and being virtually maintenance free. These simple maintenance steps will assure your pump’s long term reliability:
Pump Tubing:
Geotech recommends regular tubing replacement for optimum performance, use of the proper size and type of tubing for the pump head is essential. If you are unsure of tubing type for your application, please call Geotech. Depending on the pump head design, different sizes of tubing may be used. Use of the incorrect tubing size or type, will cause damage to the pump and/or the pump head and void the warranty.
Pump:
Keep your Geopump clean and dry. In the event that the Geopump is subjected to significant splashing or immersion, discontinue use and wipe the unit down immediately with a clean dry cloth. To keep your Geopump reliable follow these simple guidelines. Do not drop your Geopump Do not immerse your Geopump Do not subject your Geopump to poor power supplies Do not subject your Geopump to extreme heat or cold when in use
Power Cords:
Always replace any kinked or damaged power cord. Units with hard-wired AC or DC power cords should be sent back to Geotech for proper repair (see pg. 12). However, replacement power cords are available for AD/DC combination units.
Pump Head:
Your Geopump pump-head needs to be periodically cleaned with a Phosphate free cleaning detergent and water solution.
5
Chapter 5: System Troubleshooting
Problem: Unit will not turn on Solution: No power to unit:(rollers not moving)
-check power source and compatibility
-check connections
Speed control not set fast enough to overcome tubing resistance:
-check speed setting; if too low turn it up
Check tubing size and type-make sure it is the correct size and type for the pump head
Check circuit breaker; if tripped press it in to reset.
Problem: Unit turns on, but not pumping…(Pump head rollers are moving) Solution: Verify fluid level in well (maxsuction lift unit can pump from 27’ below ground at sea level).
Water level is below down well tubing intake-Increase tubing length.
If using a combination of flexible and rigid tubing check connection, between tubing. A poor connection may cause a vacuum leak. Secure tubing connection.
Flexible tubing in pump head compromised or worn out: replace flexible tubing regularly
Obstruction in tubing: check for clogs and kinks clear any obstructions
Using incorrect tubing type for pump head: tubing may be collapsed. replace with proper tubing type
Problem: Pump head rollers are not moving Solution: Pump head is loose from the pump housing: tighten pump head screws to engage pump head to gear Possible internal damage - call Geotech for consultation.
6
Chapter 6: System Specifications
Operating range 27 feet at sea level Principle of Operation Mechanical peristalsis Dimensions 3.5 x 8 x 8 inches Power source Any external 12 VDC or 120 VAC Power cord 12 VDC adapter cord or standard AC power cord.
Range of speed: Series I 60 to 350 RPM Range of speed: Series II First pumping station 30 to 300 RPM Second pumping station 60 to 600 RPM
Speed control Stepless variable speed control
Liquid delivery rate 1.67 ml per revolution (for size 15 tubing) Pumping options Pressure or vacuum (reversible flow)
Pump head rotor cold rolled steel
OPTIONS
Models: Geopump 1, Geopump 2 Tubing: Silicone, Tygon, Viton, C-Flex Pump Heads: Standard, Easyload, Quickload
7
Chapter 7: System Schematic
8
Chapter 8: Replacement Parts List
Parts Number Part Description
17500035 Adaptor, Cigarette to Clips 57500007 Assy, Power Cord, AC (AC/DC Unit) 57500008 Assy, Power Cord, DC (AC/DC Unit) 51350001 Assy, Gear Housing, Series I 51350002 Assy, Gear Housing, Series II 51350012 Assy, Housing, Bottom, PP 51350003 Assy, Motor, PP 51350007 Assy, Diode, PP 57500009 Assy, Rectifier Bridge, PP-Logic 51350011 Assy, Rheostat 51350013 Assy, Transformer, PP 51350004 Assy, Wiring Harness, PP 17500037 Boot, RBR, Toggle Switch Grey 11350005 Breaker, Thermal, 5amp, Circuit 250V 51350015 Case, Peristaltic Pump with foam 51350026 Faceplate, Gear Hsng, Series I Aluminum 51350025 Faceplate, Gear Hsng, Series II Aluminum 17500042 Foot, Rubber, 9/32 Hole DIA 17200079 F, Screw, Short, Peristaltic Pump SS8, 8-32X1.25, filter 17200199 F, Screw, Thumb, Peristaltic Pump 8-32X2.5, ZN 71350030 F, Screw, Ezload II Pumphead Long Shaft 11350009 Handle, PE, NI 51350022 Housing, Top, Slk Scrn, Series I 51350023 Housing, Top, Slk, Scrn, Series II 11350010 Knob, Plastic, Rheostat 11350020 Switch, Toggle, Dpdt, Frwd/Rvrs 11350021 Switch, Toggle, Dpst, On/Off
9
The Warranty
For a period of one (1) year from date of first sale, product is warranted to be free from defects in materials and workmanship. Geotech agrees to repair or replace, at Geotech’s option, the portion proving defective, or at our option to refund the purchase price thereof. Geotech will have no warranty obligation if the product is subjected to abnormal operating conditions, accident, abuse, misuse, unauthorized modification, alteration, repair, or replacement of wear parts. User assumes all other risk, if any, including the risk of injury, loss, or damage, direct or consequential, arising out of the use, misuse, or inability to use this product. User agrees to use, maintain and install product in accordance with recommendations and instructions. User is responsible for transportation charges connected to the repair or replacement of product under this warranty
A Return Material Authorization number (RMA #) is required prior to return of any equipment to our facilities, please call 800 number for appropriate location. An RMA # will be issued upon receipt of your request to return equipment, which should include reasons for the return. Your return shipment to us must have this RMA # clearly marked on the outside of the package. Proof of date of purchase is required for processing of all warranty requests.
This policy applies to both equipment sales and repair orders.
FOR A RETURN MATERIAL AUTHORIZATION, PLEASE CALL OUR SERVICE DEPARTMENT AT 1-800-833-7958 OR 1-800-275-5325.
Model Number: ________________
Serial Number: ________________
Date: ________________
.
Equipment Return Policy
Equipment Decontamination
Prior to return, all equipment must be thoroughly cleaned and decontaminated. Please make note on RMA form, the use of equipment, contaminants equipment was exposed to, and decontamination solutions/methods used. Geotech reserves the right to refuse any equipment not properly decontaminated. Geotech may also choose to decontaminate equipment for a fee, which will be applied to the repair order invoice.
10