Geotech ET WLM Installation And Operation Manual

Rev 04/22/2016 Part # 12050062
ET Water
Installation and Operation
Manual
1
Table of Contents
Section 1: System Description .................................................................................... 4
Section 2: System Installation ..................................................................................... 7
Section 3: System Operation .................................................................................... 10
Section 4: System Maintenance ................................................................................ 11
Section 5: System Troubleshooting........................................................................... 12
Section 6: System Specifications .............................................................................. 13
Section 7: System
Schematic
.................................................................................. 15
Section 8: Replacement Parts List ............................................................................ 17
EC Declaration of Conformity .................................................................................... 20
The Warranty .......................................................................................................... 201
2
DOCUMENTATION CONVENTIONS
This document uses the following conventions to present information:
WARNING
CAUTION
NOTE
An exclamation point icon indicates a WARNING of a situation or condition that could lead to personal injury or death. You should not proceed until you read and thoroughly understand the WARNING message.
A raised hand icon indicates CAUTION information that relates to a situation or condition that could lead to equipment malfunction or damage. You should not proceed until you read and thoroughly understand the CAUTION message.
A note icon indicates NOTE information. Notes provide additional or supplementary information about an activity or concept.
3
Notice for consumers in Europe: This symbol indicates that this product is to be collected
separately.
The following applies only to users in European countries:
This product is designated for separate collection at an appropriate
collection point. Do not dispose of as household waste.
For more information, contact the seller or the local authorities in charge
of waste management.
4
Section 1: System Description
Function and Theory
The Geotech ET Water Level Meter (ET WLM) is a portable instrument used to accurately measure water levels in monitoring wells and bore holes. The sensor consists of a stainless steel and PTFE probe attached to a reel-mounted, polyethylene-coated engineers tape. The engineer's tape comes in standard or metric graduations, and is accurate to 1/100th of a foot, or 1 millimeter.
The sensor relies on fluid conductivity to determine the presence of water. An audible signal and visible red LED activate when the probe contacts water. It features adjustable sensitivity, used to prevent false triggering.
The ET WLM is intended for use as a depth to water measuring device. Using the ET WLM for any other purpose may compromise safety of the operator and/or void manufacturers warranty.
To avoid damage to tape and strain relief, do not over tighten reel with probe in storage position.
System Components
Figure 1-1
5
Figure 1-2
Figure 1-3
6
Figure
1-4
7
Section 2: System Installation
The reel frame has a metal loop called the hanger. The hanger can be used to hang the reel frame onto the well casing (as shown in Figure 2-1). The tape can then slide easily over the side of the reel leg.
Figure 2-1 Reel frame on well casing
8
If you are not able to hang the frame onto the well head, then either use the white plastic leader guard (standard with all units), or the optional Tape Guide, to prevent the edge of the well head from damaging the tape. Figure 2-2 is an example of the two parts.
Figure 2-1 Tape Guide (optional) , Tape Weight (optional), Tape Leader Guard
Optional Tape Weight Installation Attach tape weight to tape near probe end being lowered into well.
1) Loosen the screws using an Allen wrench.
2) Slide the weight onto the tape (Figure 2-2).
3) Tighten the screws (Figure 2-3).
Do not use the tape leader guard in wells larger than 4” (10cm), as it may fall down the well. The optional tape weight helps sink probe into a deep or crooked well
9
Figure 2-2 Fitting (Optional) Tape Weight
Figure 2-3 Tightening (Optional) Tape Weight
10
Section 3: System Operation
Turn the ET WLM on with the ON/SENSITIVITY switch. If the buzzer makes a loud signal and a red light is visible, the battery is adequate for normal operation.
Lower the probe down the well to the water surface. The light and buzzer will activate. At this point, adjust the probe dial by rotating it counter-clockwise until the light and buzzer turn off.
With the probe still in contact with the water, adjust the probe sensitivity dial clockwise until the light and buzzer barely activate. In this setting, the probe will detect the water level and avoid false triggering.
Water level measurements can now be taken from the top of the casing or any reference point.
The ET WLM should be stored with the ON/SENSITIVITY switch in the OFF position. To turn the unit off, rotate the switch fully counter-clockwise until the switch clicks. If the ET WLM is not used and is stored for longer than three (3) months, remove the battery to prevent battery leakage, which can cause internal damage.
To avoid damage to the tape and strain relief, do not over tighten the reel with the probe in storage position.
This unit is not rated for use with hydrocarbons or flammable liquids. If measuring depth to fluid of wells containing hydrocarbons, use Geotechs line of interface probe products.
Operation of Optional Draw Down Control
1. Lower the probe to desired depth of maximum allowable draw down with the unit in standard ET WLM mode (switch at “Right”) – see Figure 1-3.
2. Move the mode switch to the Left “, for Draw Down mode.
3. Connect the ET WLM to a controller; such as the Geotech GeoControl Pro or Geo Controller, which is equipped with a Draw Down feature, as shown in Figure 1-3.
4. Operate your controller in accordance with the controllers operation manual. The ET WLM will stop the controller when the water level dips below it. It will allow the controller to re-start when the water level rises and makes contact with the probe
11
Section 4: System Maintenance
Battery Replacement
Replace the battery when the audible and visible signals become weak or the unit does not operate.
1. Gently remove the battery tray.
2. Remove the old battery and replace it with a new one.
Be aware of the pola rity (+, -) of the battery when placing the new battery in the tray. Use a 9V alkaline battery only.
Cleaning The ET WLM can be cleaned with mild detergents such as trisodium phosphate
(TSP). If other detergents are used, take care to select detergents that are compatible with PTFE, polypropylene, and stainless steel. The reel should not be submerged in any liquid, but may be cleaned with a damp cloth.
Cleaning the Conductivity Contact
The conductivity contact of the probe should be periodically cleaned with a non­abrasive cleaner such as isopropyl alcohol, xylene, methanol, or phosphate free type cleaner. To clean the conductivity contact, place a small amount of the cleaner on a cotton swab; rub the conductivity contact to remove all foreign matter. Repeat this process until all foreign matter has been removed.
12
Section 5: System Troubleshooting
Problem: No signal (audible or visible) when unit is turned on. Solutions:
The battery is discharged. Check or change battery (Section 4).  The circuit is malfunctioning. Contact Geotech Service.
Problem: No indication of water. Solutions:
The conductive contact is dirty. Clean the contact (Section 4).  There is an open connection in the tape. Replace tape and/or probe.  The circuit is malfunctioning. Contact Geotech Service.
Problem: The signal (audible or visible) is intermittent. Solutions:
There is an open connection in the tape. Replace tape and/or probe.  There is a loose connection in the circuit or the probe. Repair the
connection.
Problem: The signal (audible or visible) is continuous when not in water. Solutions:
Make sure the meter is in standard mode. Place the mode switch to the
“Right“, for normal operation.
The conductive contact is dirty (causing bridging). Clean the contact
(Section 4).
There is a short in the tape and/or probe. Replace tape and/or probe.  The circuit is malfunctioning. Contact Geotech Service.
Problem: Controller shuts down automatically when ET WLM has draw down feature connected.
Solutions:
ET WLM is in standard ET WLM mode. Lower the probe until it contacts
the water level and switch to draw down mode.
ET WLM is in draw down mode, but water level dropped below probe.Wait
for well to recharge and make contact with probe, or lower probe to contact water.
For technical assistance, call Geotech Environmental Equipment at 1-303-320-
4764 or 1-800-833-7958
13
Section 6: System Specifications
Length/Weight: 100 foot (30 meters) = 9 lbs (4 kg)
200 foot (60 meters) = 11 lbs (5 kg) 300 foot (100 meters) = 14 lbs (6.5 kg)
5/8 Probe Material: Stainless steel, PTFE, Viton
Weight: 4.68 oz (132.5 g) Diameter: .625 (1.59 cm) Length: 6.125” (19.7 cm) Minimum Conductivity Threshold (detects water at): 10µS
3/8 Probe (optional) Material: Stainless steel, PTFE, Viton
Weight: 1 oz (30.5 g) Diameter: .375 (1 cm) Length: 6.5” (16.5 cm) Minimum Conductivity Threshold (detects water at): 10µS
Tape Material: Polyethylene coated steel tape w/ stainless steel conductors
Accuracy: 100th of a foot/100 (3 mm/30.5
meters) Per Federal Specification:
GGG-T-106E or Material: MDPE tape w/ stainless steel conductors Accuracy: 100th of a foot/100 (3 mm/30.5 meters)
Reel/Frame Material: Polypropylene & aluminum
Size: 13” H x 11 W x 7” D (33 cm H x 28 cm W x 18 cm D)
14
Unit Battery: 9 volt alkaline
Max power: < 1 Watt Max sound pressure output: 75 dB @ 30 cm Insulation rating: Insulation rating is considered double as
there are no hazardous voltages or power circuits to pose a threat to the operator.
Battery life: Continuously detecting: 8 hours
On but not detecting: >1 year
Output tone: 5 kHz
Operating temperature range: 32 – 140 °F (0 – 60 °C) Operating environment: Rated for Outdoor use.
Storage temperature range: -40 – 158 °F (-40 – 70 °C) Response time: <10 milliseconds Humidity: 0-100% Elevation: -328 – 13,124 feet (-100 – 4000 meters) Tape Weight (Optional) 1” (2.54cm) OD, 4” (10.16cm) Long,
12.74oz (361g)
15
Section 7: System
Schematic
-- -
Figure 7-1 Water Level Meter (front
view)
16
Figure 7-2 Water Level Meter (side
view)
17
Section 8: Replacement Parts List
Item #
Parts Description
Part #
1
ASSY,FRAME,GEOWLM/GIP
52050002
2
ASSY,REEL,100FT,GEOWLM
52050029
‘’
ASSY,REEL,200FT,GEOWLM
52050030
‘’
ASSY,REEL,300FT,GEOWLM
52050031
Replacement Tape Assemblies (5/8" Probe and reel not Included)
3
ASSY,TAPE,GEOWLM,POLY,100FT
52050013
‘’
ASSY,TAPE,GEOWLM,POLY,200FT
52050012
‘’
ASSY,TAPE,GEOWLM,POLY,300FT
52050014
‘’
ASSY,TAPE,GEOWLM,POLY,30M
52050072
‘’
ASSY,TAPE,GEOWLM,POLY,60M
52050042
‘’
ASSY,TAPE,GEOWLM,POLY,100M
52050053
‘’
ASSY,TAPE,GEOWLM,MDPE,100'
52050181
‘’
ASSY,TAPE,GEOWLM,MDPE,200'
52050182
‘’
ASSY,TAPE,GEOWLM,MDPE,300'
52050183
‘’
ASSY,TAPE,GEOWLM,MDPE,30M
52050230
‘’
ASSY,TAPE,GEOWLM,MDPE,60M
52050231
‘’
ASSY,TAPE,GEOWLM,MDPE,100M
52050232
18
Item #
Parts Description
Part #
Replacement Tape Assemblies (3/8" Probe Included, reel not included)
3 + 4
ASSY,TAPE,GEOWLM,100FT,3/8"
52050126
‘’
ASSY,TAPE,GEOWLM,200FT,3/8"
52050127
‘’
ASSY,TAPE,GEOWLM,300FT,3/8"
52050128
‘’
ASSY,TAPE,GEOWLM,30M,3/8"
52050129
‘’
ASSY,TAPE,GEOWLM,60M,3/8"
52050130
‘’
ASSY,TAPE,GEOWLM,100M,3/8"
52050131
‘’
ASSY,TAPE,GEOWLM,MDPE,100',3/8
52050226
‘’
ASSY,TAPE,GEOWLM,MDPE,200',3/8
52050227
‘’
ASSY,TAPE,GEOWLM,MDPE,300',3/8
52050228
‘’
ASSY,TAPE,GEOWLM,MDPE,30M,3/8
52050233
‘’
ASSY,TAPE,GEOWLM,MDPE,60M,3/8
52050234
‘’
ASSY,TAPE,GEOWLM,MDPE,100M,3/8
52050235
4
ASSY,PROBE,GEOWLM (5/8")
52050052
Reel and Control only (does not include tape or probe)
5
ASSY,CONTROL,GEOWLM TAPE LENGTH REQUIRED
52050003
6
PROBE HOLDER, 5/8", UNI-FRAME
52050063
‘’
PROBE HOLDER, 3/8" ,UNI-FRAME
52050263
7
KNOB,KNURLED,3/4x5/16",BLK (USE WITH #12050525)
12050524
8
SCREW,SS8,1/4-20X1.375",SHCS
12050525
9
NUT,NYL,1/4-20,HEX
17500129
10
KNOB,PHENOLIC,OVAL/TAPERED REEL HANDLE
12050002
11
BOLT,SS8,KNOB HANDLE STRIPPER 5/16x1.5" (1/4-20 THRD)
17500123
12
O-RING,VITON,.480X.039,BROWN
17500202
13
WASHER, WAVE
12050255
14
KNOB,INDICATING,1/4",RND (SENSITIVITY)
12050090
15
HANDLE, GRIP, VINYL, 3/4 x 5-1/16”
12050007
Not shown:
GUARD,LEADER,PROPAMIDE,NATURAL
12050060
GUIDE,TAPE,PVC
22050601
CASE,GEOWLM,100-300'
12050059
MANUAL,INSTRUCTION,GEOWLM
12050062
TAPE WEIGHT (OPTIONAL)
52050277
19
DOCUMENT REVISIONS
EDCF#
DESCRIPTION
REV/DATE
-
Previous Release
08/31/2012
-
Updated 3/8” probe specs, updated EC Declaration of
Conformity, updated international office address, SP
03/14/2014
-
Updated EC Declaration of Conformity, correct backpage
footer information, SP
03/03/2015
EDCF#
1972
Updated Replacements Parts List & Diagram with
sensititivty knob (#12050090) information, SP
04/22/2016
20
EC Declaration of Conformity
Manufacturer:
Geotech Environmental Equipment, Inc. 2650 E 40th Avenue
Denver, CO 80205 Declares that the following products, Product Name: Water Level Meter, manual rewind only Model(s): ET Water Level Meter
ETL Water Level Meter Year of manufacture: 2010 Conform to the principle safety objectives of 2006/95/EC Low Voltage Directive by application
of the following standards:
EN 61010-1: 2010
EN 809-1 + A1:2010
Year of affixation of the CE Marking: 2008 Conform to the protection requirements of 2004/108/EC Electromagnetic Compatibility (EMC)
by application of the following standards:
EN 61000-6-1: 2007
EN 61000-6-3: 2012
EN 61326-1: 2013
EMC conformity established 11/28/2008 Production control follows the ISO 9001:2008 regulations and includes required safety routine
tests. This declaration issued under the sole responsibility of Geotech Environmental Equipment, Inc.
Joe Leonard Product Development
Serial number ________________
21
The Warranty
For a period of one (1) year from date of first sale, product is warranted to be free from defects in materials and workmanship. Geotech agrees to repair or replace, at Geotechs option, the portion proving defective, or at our option to refund the purchase price thereof. Geotech will have no warranty obligation if the product is subjected to abnormal operating conditions, accident, abuse, misuse, unauthorized modification, alteration, repair, or replacement of wear parts. User assumes all other risk, if any, including the risk of injury, loss, or damage, direct or consequential, arising out of the use, misuse, or inability to use this product. User agrees to use, maintain and install product in accordance with recommendations and instructions. User is responsible for transportation charges connected to the repair or replacement of product under this warranty.
Equipment Return Policy
A Return Material Authorization number (RMA #) is required prior to return of any equipment to our facilities, please call our 800 number for appropriate location. An RMA # will be issued upon receipt of your request to return equipment, which should include reasons for the return. Your return shipment to us must have this RMA # clearly marked on the outside of the package. Proof of date of purchase is required for processing of all warranty requests.
This policy applies to both equipment sales and repair orders.
FOR A RETURN MATERIAL AUTHORIZATION, PLEASE CALL OUR
SERVICE DEPARTMENT AT 1-800-833-7958.
Model Number:
Serial Number:
Date of Purchase:
Equipment Decontamination
Prior to return, all equipment must be thoroughly cleaned and decontaminated. Please make note on RMA form, the use of equipment, contaminants equipment was exposed to, and decontamination solutions/methods used. Geotech reserves the right to refuse any equipment not properly decontaminated. Geotech may also choose to decontaminate the equipment for a fee, which will be applied to the repair order invoice.
Geotech Environmental Equipment, Inc
2650 East 40th Avenue Denver, Colorado 80205
(303) 320-4764 ● (800) 833-7958 ● FAX (303) 322-7242
email: sales@geotechenv.com website: www.geotechenv.com
In the EU
Geotech Equipos Ambientales
Calle Francesc I Ferrer, Guardia Local 19, Mollet del Valles, Barcelona 08100, España
Tlf: (34)93 5445937
email: ventas@geotechenv.com
website: http://spainish.geotechenv.com