Rev 04/22/2016 Part # 12050062
ET Water
Level Meter
Installation and Operation
Manual

1
Table of Contents
Section 1: System Description .................................................................................... 4
Section 2: System Installation ..................................................................................... 7
Section 3: System Operation .................................................................................... 10
Section 4: System Maintenance ................................................................................ 11
Section 5: System Troubleshooting........................................................................... 12
Section 6: System Specifications .............................................................................. 13
Section 7: System
Schematic
.................................................................................. 15
Section 8: Replacement Parts List ............................................................................ 17
EC Declaration of Conformity .................................................................................... 20
The Warranty .......................................................................................................... 201

2
DOCUMENTATION CONVENTIONS
This document uses the following conventions to present information:
WARNING
CAUTION
NOTE
An exclamation point icon indicates a WARNING of a situation or
condition that could lead to personal injury or death. You should
not proceed until you read and thoroughly understand the
WARNING message.
A raised hand icon indicates CAUTION information that relates to
a situation or condition that could lead to equipment malfunction
or damage. You should not proceed until you read and thoroughly
understand the CAUTION message.
A note icon indicates NOTE information. Notes provide additional
or supplementary information about an activity or concept.

3
Notice for consumers in Europe:
This symbol indicates that this product is to be collected
separately.
The following applies only to users in European countries:
This product is designated for separate collection at an appropriate
collection point. Do not dispose of as household waste.
For more information, contact the seller or the local authorities in charge
of waste management.

4
Section 1: System Description
Function and Theory
The Geotech ET Water Level Meter (ET WLM) is a portable instrument used to
accurately measure water levels in monitoring wells and bore holes. The sensor
consists of a stainless steel and PTFE probe attached to a reel-mounted,
polyethylene-coated engineer’s tape. The engineer's tape comes in standard or
metric graduations, and is accurate to 1/100th of a foot, or 1 millimeter.
The sensor relies on fluid conductivity to determine the presence of water. An
audible signal and visible red LED activate when the probe contacts water. It
features adjustable sensitivity, used to prevent false triggering.
The ET WLM is intended for use as a depth to water measuring device. Using
the ET WLM for any other purpose may compromise safety of the operator
and/or void manufacturer’s warranty.
To avoid damage to tape and strain relief, do not over tighten reel
with probe in storage position.
System Components
Figure 1-1

7
Section 2: System Installation
The reel frame has a metal loop called the “hanger”. The hanger can be used to
hang the reel frame onto the well casing (as shown in Figure 2-1). The tape can
then slide easily over the side of the reel leg.
Figure 2-1 Reel frame on well casing

8
If you are not able to hang the frame onto the well head, then either use the white
plastic leader guard (standard with all units), or the optional Tape Guide, to
prevent the edge of the well head from damaging the tape. Figure 2-2 is an
example of the two parts.
Figure 2-1 Tape Guide (optional) , Tape Weight (optional), Tape Leader Guard
Optional Tape Weight Installation
Attach tape weight to tape near probe end being lowered into well.
1) Loosen the screws using an Allen wrench.
2) Slide the weight onto the tape (Figure 2-2).
3) Tighten the screws (Figure 2-3).
Do not use the tape leader guard in wells larger than 4” (10cm), as it
may fall down the well. The optional tape weight helps sink probe
into a deep or crooked well

9
Figure 2-2 Fitting (Optional) Tape Weight
Figure 2-3 Tightening (Optional) Tape Weight

10
Section 3: System Operation
Turn the ET WLM on with the ON/SENSITIVITY switch. If the buzzer makes a
loud signal and a red light is visible, the battery is adequate for normal operation.
Lower the probe down the well to the water surface. The light and buzzer will
activate. At this point, adjust the probe dial by rotating it counter-clockwise until
the light and buzzer turn off.
With the probe still in contact with the water, adjust the probe sensitivity dial
clockwise until the light and buzzer barely activate. In this setting, the probe will
detect the water level and avoid false triggering.
Water level measurements can now be taken from the top of the casing or any
reference point.
The ET WLM should be stored with the ON/SENSITIVITY switch in the OFF
position. To turn the unit off, rotate the switch fully counter-clockwise until the
switch clicks. If the ET WLM is not used and is stored for longer than three (3)
months, remove the battery to prevent battery leakage, which can cause internal
damage.
To avoid damage to the tape and strain relief, do not over tighten
the reel with the probe in storage position.
This unit is not rated for use with hydrocarbons or flammable liquids.
If measuring depth to fluid of wells containing hydrocarbons, use
Geotech’s line of interface probe products.
Operation of Optional Draw Down Control
1. Lower the probe to desired depth of maximum allowable draw down
with the unit in standard ET WLM mode (switch at “Right”) – see Figure
1-3.
2. Move the mode switch to the “Left “, for Draw Down mode.
3. Connect the ET WLM to a controller; such as the Geotech GeoControl
Pro or Geo Controller, which is equipped with a Draw Down feature, as
shown in Figure 1-3.
4. Operate your controller in accordance with the controller’s operation
manual. The ET WLM will stop the controller when the water level dips
below it. It will allow the controller to re-start when the water level rises
and makes contact with the probe

11
Section 4: System Maintenance
Battery Replacement
Replace the battery when the audible and visible signals become weak or the
unit does not operate.
1. Gently remove the battery tray.
2. Remove the old battery and replace it with a new one.
Be aware of the pola rity (+, -) of the battery when placing the new
battery in the tray. Use a 9V alkaline battery only.
Cleaning
The ET WLM can be cleaned with mild detergents such as trisodium phosphate
(TSP). If other detergents are used, take care to select detergents that are
compatible with PTFE, polypropylene, and stainless steel. The reel should not be
submerged in any liquid, but may be cleaned with a damp cloth.
Cleaning the Conductivity Contact
The conductivity contact of the probe should be periodically cleaned with a nonabrasive cleaner such as isopropyl alcohol, xylene, methanol, or phosphate free type
cleaner. To clean the conductivity contact, place a small amount of the cleaner on a
cotton swab; rub the conductivity contact to remove all foreign matter. Repeat this
process until all foreign matter has been removed.

12
Section 5: System Troubleshooting
Problem: No signal (audible or visible) when unit is turned on.
Solutions:
The battery is discharged. Check or change battery (Section 4).
The circuit is malfunctioning. Contact Geotech Service.
Problem: No indication of water.
Solutions:
The conductive contact is dirty. Clean the contact (Section 4).
There is an open connection in the tape. Replace tape and/or probe.
The circuit is malfunctioning. Contact Geotech Service.
Problem: The signal (audible or visible) is intermittent.
Solutions:
There is an open connection in the tape. Replace tape and/or probe.
There is a loose connection in the circuit or the probe. Repair the
connection.
Problem: The signal (audible or visible) is continuous when not in water.
Solutions:
Make sure the meter is in standard mode. Place the mode switch to the
“Right“, for normal operation.
The conductive contact is dirty (causing bridging). Clean the contact
(Section 4).
There is a short in the tape and/or probe. Replace tape and/or probe.
The circuit is malfunctioning. Contact Geotech Service.
Problem: Controller shuts down automatically when ET WLM has draw down
feature connected.
Solutions:
ET WLM is in standard ET WLM mode. Lower the probe until it contacts
the water level and switch to draw down mode.
ET WLM is in draw down mode, but water level dropped below probe.Wait
for well to recharge and make contact with probe, or lower probe to contact
water.
For technical assistance, call Geotech Environmental Equipment at 1-303-320-
4764 or 1-800-833-7958

13
Section 6: System Specifications
Length/Weight: 100 foot (30 meters) = 9 lbs (4 kg)
200 foot (60 meters) = 11 lbs (5 kg)
300 foot (100 meters) = 14 lbs (6.5 kg)
5/8” Probe
Material: Stainless steel, PTFE, Viton
Weight: 4.68 oz (132.5 g)
Diameter: .625” (1.59 cm)
Length: 6.125” (19.7 cm)
Minimum Conductivity
Threshold (detects water at): 10µS
3/8” Probe (optional)
Material: Stainless steel, PTFE, Viton
Weight: 1 oz (30.5 g)
Diameter: .375” (1 cm)
Length: 6.5” (16.5 cm)
Minimum Conductivity
Threshold (detects water at): 10µS
Tape
Material: Polyethylene coated steel tape w/ stainless steel conductors
Accuracy: 100th of a foot/100’ (3 mm/30.5
meters) Per Federal Specification:
GGG-T-106E or
Material: MDPE tape w/ stainless steel conductors
Accuracy: 100th of a foot/100’ (3 mm/30.5 meters)
Reel/Frame
Material: Polypropylene & aluminum
Size: 13” H x 11” W x 7” D (33 cm H x 28 cm W x 18 cm D)

14
Unit
Battery: 9 volt alkaline
Max power: < 1 Watt
Max sound pressure output: 75 dB @ 30 cm
Insulation rating: Insulation rating is considered double as
there are no hazardous voltages or power
circuits to pose a threat to the operator.
Battery life: Continuously detecting: 8 hours
On but not detecting: >1 year
Output tone: 5 kHz
Operating temperature range: 32 – 140 °F (0 – 60 °C)
Operating environment: Rated for Outdoor use.
Storage temperature range: -40 – 158 °F (-40 – 70 °C)
Response time: <10 milliseconds
Humidity: 0-100%
Elevation: -328 – 13,124 feet (-100 – 4000 meters)
Tape Weight (Optional) 1” (2.54cm) OD, 4” (10.16cm) Long,
12.74oz (361g)

15
Section 7: System
Schematic
-- -
Figure 7-1 Water Level Meter (front
view)

16
Figure 7-2 Water Level Meter (side
view)

17
Section 8: Replacement Parts List
Replacement Tape Assemblies (5/8" Probe and reel not Included)
ASSY,TAPE,GEOWLM,POLY,100FT
ASSY,TAPE,GEOWLM,POLY,200FT
ASSY,TAPE,GEOWLM,POLY,300FT
ASSY,TAPE,GEOWLM,POLY,30M
ASSY,TAPE,GEOWLM,POLY,60M
ASSY,TAPE,GEOWLM,POLY,100M
ASSY,TAPE,GEOWLM,MDPE,100'
ASSY,TAPE,GEOWLM,MDPE,200'
ASSY,TAPE,GEOWLM,MDPE,300'
ASSY,TAPE,GEOWLM,MDPE,30M
ASSY,TAPE,GEOWLM,MDPE,60M
ASSY,TAPE,GEOWLM,MDPE,100M

Replacement Tape Assemblies (3/8" Probe Included, reel not included)
ASSY,TAPE,GEOWLM,100FT,3/8"
ASSY,TAPE,GEOWLM,200FT,3/8"
ASSY,TAPE,GEOWLM,300FT,3/8"
ASSY,TAPE,GEOWLM,30M,3/8"
ASSY,TAPE,GEOWLM,60M,3/8"
ASSY,TAPE,GEOWLM,100M,3/8"
ASSY,TAPE,GEOWLM,MDPE,100',3/8
ASSY,TAPE,GEOWLM,MDPE,200',3/8
ASSY,TAPE,GEOWLM,MDPE,300',3/8
ASSY,TAPE,GEOWLM,MDPE,30M,3/8
ASSY,TAPE,GEOWLM,MDPE,60M,3/8
ASSY,TAPE,GEOWLM,MDPE,100M,3/8
Reel and Control only (does not include tape or probe)
ASSY,CONTROL,GEOWLM TAPE LENGTH REQUIRED
PROBE HOLDER, 5/8", UNI-FRAME
PROBE HOLDER, 3/8" ,UNI-FRAME
KNOB,KNURLED,3/4x5/16",BLK (USE WITH #12050525)
SCREW,SS8,1/4-20X1.375",SHCS
KNOB,PHENOLIC,OVAL/TAPERED REEL HANDLE
BOLT,SS8,KNOB HANDLE STRIPPER 5/16x1.5" (1/4-20 THRD)
O-RING,VITON,.480X.039,BROWN
KNOB,INDICATING,1/4",RND (SENSITIVITY)
HANDLE, GRIP, VINYL, 3/4 x 5-1/16”
GUARD,LEADER,PROPAMIDE,NATURAL
MANUAL,INSTRUCTION,GEOWLM

Updated 3/8” probe specs, updated EC Declaration of
Conformity, updated international office address, SP
Updated EC Declaration of Conformity, correct backpage
footer information, SP
Updated Replacements Parts List & Diagram with
sensititivty knob (#12050090) information, SP

20
EC Declaration of Conformity
Manufacturer:
Geotech Environmental Equipment, Inc.
2650 E 40th Avenue
Denver, CO 80205
Declares that the following products,
Product Name: Water Level Meter, manual rewind only
Model(s): ET Water Level Meter
ETL Water Level Meter
Year of manufacture: 2010
Conform to the principle safety objectives of 2006/95/EC Low Voltage Directive by application
of the following standards:
EN 61010-1: 2010
EN 809-1 + A1:2010
Year of affixation of the CE Marking: 2008
Conform to the protection requirements of 2004/108/EC Electromagnetic Compatibility (EMC)
by application of the following standards:
EN 61000-6-1: 2007
EN 61000-6-3: 2012
EN 61326-1: 2013
EMC conformity established 11/28/2008
Production control follows the ISO 9001:2008 regulations and includes required safety routine
tests.
This declaration issued under the sole responsibility of Geotech Environmental Equipment, Inc.
Joe Leonard
Product Development
Serial number ________________

21
The Warranty
For a period of one (1) year from date of first sale, product is warranted to be free
from defects in materials and workmanship. Geotech agrees to repair or replace,
at Geotech’s option, the portion proving defective, or at our option to refund the
purchase price thereof. Geotech will have no warranty obligation if the product is
subjected to abnormal operating conditions, accident, abuse, misuse,
unauthorized modification, alteration, repair, or replacement of wear parts. User
assumes all other risk, if any, including the risk of injury, loss, or damage, direct
or consequential, arising out of the use, misuse, or inability to use this product.
User agrees to use, maintain and install product in accordance with
recommendations and instructions. User is responsible for transportation charges
connected to the repair or replacement of product under this warranty.
Equipment Return Policy
A Return Material Authorization number (RMA #) is required prior to return of any
equipment to our facilities, please call our 800 number for appropriate location.
An RMA # will be issued upon receipt of your request to return equipment, which
should include reasons for the return. Your return shipment to us must have this
RMA # clearly marked on the outside of the package. Proof of date of purchase
is required for processing of all warranty requests.
This policy applies to both equipment sales and repair orders.
FOR A RETURN MATERIAL AUTHORIZATION, PLEASE CALL OUR
SERVICE DEPARTMENT AT 1-800-833-7958.
Equipment Decontamination
Prior to return, all equipment must be thoroughly cleaned and decontaminated.
Please make note on RMA form, the use of equipment, contaminants equipment
was exposed to, and decontamination solutions/methods used. Geotech
reserves the right to refuse any equipment not properly decontaminated.
Geotech may also choose to decontaminate the equipment for a fee, which will
be applied to the repair order invoice.

Geotech Environmental Equipment, Inc
2650 East 40th Avenue Denver, Colorado 80205
(303) 320-4764 ● (800) 833-7958 ● FAX (303) 322-7242
email: sales@geotechenv.com
website: www.geotechenv.com
In the EU
Geotech Equipos Ambientales
Calle Francesc I Ferrer, Guardia Local 19, Mollet del Valles, Barcelona 08100, España
Tlf: (34)93 5445937
email: ventas@geotechenv.com
website: http://spainish.geotechenv.com
Printed in the United States of America