SAFETY INSTRUCTIONS FOR IN-LINE ELECTRODE INSTALLATION
1. Do not remove from pressurized lines.
2. Do not exceed maximum temperature/pressure specifi cations.
3. Wear safety goggles or face shield during installation/service.
4. Do not alter product construction. Failure to follow safety instructions may result in severe personal injury!
1. Wiring
Preamplifier:
pH Electrodes: 2714 2714-HF 2716 2716-DI
ORP Electrodes: 2715 2717
Silver (earth ground)
Brown Blue
Green
White
Red
Black
(mV Input)
(Iso. Gnd)
(T+)
(T-)
(V+)
(V-)
ALARM 1
4
2
0
Signet
7
6
pH
8
10
ALARM 2
12
14
ENTER
Preamplifier:
pH Electrodes: 2714 2714-HF 2716 2716-DI
ORP Electrodes: 2715 2717
Silver (earth ground)
Brown Blue
Black
Red
Green
White
(mV Input)
(Iso. Gnd)
(V-)
(V+)
(T+)
(T-)
English
Signet
ENTER
Signet
Preamplifier:
pH Electrodes:
Silver (earth ground)
Red Black
Green
Brown
White
Blue
• Use X1 gain selection
(#10) V+
(#11) V-
(#12) T+
(#13) Ain
(#14) ISO
(#14) ISO
RLY 1
RLY 2
RLY 3
RLY 4
ENTER MOD
Technical Notes:
• Use 6-conductor shielded cable for cable extensions to 120 m (400 ft)
• Shield must be maintained through cable splice
2. Signet Fittings
Type Description
Plastic tees
PVC saddles
• 0.5 to 4 in.
• PVC or CPVC
• Mounts via glue-on fittings
• 2 to 4 in., use 1-7/16 in. hole in pipe
• Align wedge arrows with saddle arrows during assembly.
pH CONTROLLER
CAL RELAY OUTPUT
Signet
Preamplifier:
mV
pH
TEMP
ORP Electrodes:
Silver (earth ground)
Red Black
Green
White
Brown
Blue
• Use X1 gain selection
(#10) V+
(#11) V-
N/C
N/C
(#13) Ain
(#14) ISO
RLY 1 RLY 2 RLY 3
RLY 4
ENTER MOD
CAL RELAY
ORP CONTROLLER
OUTPUT
Signet
mV
ORP
Type Description
Carbon steel weld-on weldolets
• 2 to 4 in, use 1-7/16 in. hole in pipe
• Remove insert before welding
• Installed by certified welder only
Carbon steel threaded tees
• 0.5 to 2 in.
• Mounts on threaded pipe ends
Iron
• 2 to 4 in., use 1-7/16 in. hole in pipe strap-on saddles
Consult your local Signet distributor for additional fi tting information.
Universal pipe adapter for large pipes, #P31515-OV200
• For existing pipe fittings 2 in. and up
• External 1-1/4 inch NPT male threads
3. Recommended Position
3/4 in. NPT OR ISO 7/1-Rc 3/4 female threads
Customer supplied:
• 3/4 in. NPT nipple (male)
• J-box
• flex conduit
Customer supplied:
• 3/4 in. NPT threaded
C.
B.
D.
A.
pipe (A)
• J-box (B)
• Flex conduit (C)
• Quick release pipe clamp (D)
• Tank flange (E)
B.
C.
A.
0
+60
-60
In-Line
Process pipe
• Vertical (0 ) position optimum
• 3m/s (10ft/s) or less for max. performance and sensor life
4. Installation
In-line Applications
1
Preamplifier
A. O-ring lubricant recommended (factory applied)
C.
B. Align marks
Hand
tighten
only!
E.
2 in.
min.
Open
tanks
Closed
tanks
Caution: If liquid level is not constant, always ensure liquid contact with electrode tip
Submersible Applications
2
Preamplifier
A. O-ring lubricant recommended (factory applied)
1
Customer supplied pipe
C.
B. Align marks
2
Sealant
3/4 in. NPT or ISO 7/1-Rc 3/4
Sealant
#P31542 Sensor Cap (purchased separately)
Fitting inlet
3
Fill with 3 to 4 in. of sealant
Attach 3/4 in. watertight pipe to the top of the 2720. Secure the threaded connection to prevent any leakage. For additional defense against possible accu­mulation of condensation at the back seal area of the sensor, fi ll the lower 3 to 4 inches of conduit or extension pipe with a fl exible sealant such as silicone.