‡ SIGNET 2714-2717 Twist-Lock pH and ORP Electrodes
50
3
PSIBAR
50
12275167
25
77
0
10
32
50
25
2
75
5
100
7
°F
°C
2714
2714-HF
ENGLISH
3-2714.090-1
G-1/02 English
SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Confirm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and appropriate eye, face, hand, body,
and/or respiratory protection should be used.
1. Technical Data
1.1 General
Compatibility: +GF+ SIGNET 2720 Twist-Lock Preamplifier
Pipe Size Range:
1
/2 in. and up.
•
• Use +GF+ SIGNET installation fittings to 4 in.
• Use pipe adapter in pipes over 4 in.
• Submersion with 2720 pre-amplifier requires
ISO 7-R 3/4 in. male threaded extensions
Efficiency: >97% @ 25°C (77° F)
Wetted Materials: CPVC Body
Glass
FPM O-rings
Porous UHMW Polyethylene reference
junctions
ORP: Platinum sensing surface
Primary Functions: 2714, 2715: Flat surface resists fouling
2716, 2717: Bulb surface for general use
2714-HF: Extended use in applications
with trace hydrofluoric acid (<2%)
2716-DI: Extended use in pure waters
(<100 µS)
Secondary Junction: Nylon Filament
Element: Ag/AgCl
Shipping Weight: 0.2 kg (0.4 lb)
1.2 2714 and 2716 pH Electrodes
Operating Range: 0 to 14 pH (2714-HF: 0 to 12 pH)
Reference: Electrolyte: Solidified Acrylamide Gel
3.5M KCI (2714, 2714-HF, 2716)
0.1 M KCl (2716-DI)
Temperature Sensor: 3KΩ Balco (3000Ω = 25°C)
ResponseTime: <5 s for 95% of signal change
Response Time, τ: 140 s (2714), 196 s (2716)
1.3 2715/2717 ORP Electrodes
Operating Range: -999 to +999 mV
Reference: Electrolyte: Solidified Acrlyamide Gel
Response Time: Application dependent
3.5M KCL
3
/4 in. NPT or
1.4 Temperature and pressure specifications
ORP electrode operating Temperature/Pressure
• 6.89 bar @ 0°C to 65ºC (100 psi @ 32° to 149° F)
• 4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
2716 and 2716-DI pH electrode operating Temperature/Pressure
• 6.89 bar @ 0°C to 65ºC (100 psi @ 32° to 149° F)
• 4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
2714 pH electrode operating Temperature/Pressure
• 6.89 bar @ 10°C to 65ºC (100 psi @ 50° to 149° F)
• 4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
2714-HF pH electrode operating Temperature/Pressure
• 6.89 bar @ 0°C to 50ºC (100 psi @ 32° to 122° F)
Storage Temperature: > -12° C (11° F)
1.5 Dimensions
1.0 in.
(27 mm)
(104 mm)
3.4 in.
(86 mm)
4.1 in.
1.2 in.
(30.5 mm)
2.6 in.
(65 mm)
1.0 in.
(27 mm)
(97 mm)
3.4 in.
(86 mm)
3.8 in.
1.2 in.
(30.5 mm)
2.6 in.
(65 mm)
0.9 in.
(23 mm)
0.9 in.
(23 mm)
Flat electrode Bulb electrode
page 1 of 4‡ SIGNET 2714-2717 Twist-Lock pH and ORP Electrodes
2. Electrode care and application
pH/ORP electrodes are similar to batteries; they age with time and usage. The following information will help maximize electrode life.
2.1 Conditions to Avoid:
• High temperatures, strong acids or caustics will elevate electrochemical reactions and speed electrode aging.
• Coatings on the glass or junction surfaces (i.e. proteins) cause extended response time and inaccurate measurement.
• Never store the electrode tip in deionized (DI) water.
• Never expose electrode to temperatures below -12 °C (10 °F) or allow it to dehydrate. These conditions will damage the electrode.
• Never scrape or sand the glass electrode surface.
• Treat glass electrode surfaces with care. The glass is very thin and requires care to prevent accidental breakage.
2.2 Submersible Installation Tips:
• Mount electrodes in a location with ample clearance for removal for
periodic cleaning and recalibration. Choose a location that keeps the
YES
YES
electrode glass completely submerged at all times.
• Place the electrode tip in pH 4 buffer during system maintenance or
storage to avoid dehydration.
• Mount the electrode near tank outlet away from reagent addition areas.
2.3 In-Line Installation Tips:
• pH and ORP electrodes respond best in moving fluids.
30°
NO
30°
NO
• The internal measuring electrode chamber contains a wire within a liquid
and a slight amount of air. The electrode must be mounted at least 30°
from horizontal to ensure proper sensing. Mounting angles less than 30°
will impede performance.
NO
3. pH Electrode Calibration
All pH electrodes are designed to ensure linearity during their lifespan. The following
sections define proper electrode operation.
3.1 Offset (STD)
Electrode offsets occur due to:
• Clogged reference junction
• Aged or contaminated reference solution/wire
Check offsets in a pH 7 buffer @ 25 °C. The theoretical output is 0 mV. Any deviation from
0 mV is the pH electrode offset.
pH Electrode Offset pH 7 buffer @ 25°C Theoretical: 7.0 pH (0.0 mV)
New: pH 7 ± 0.25 pH (±15 mV)
Reliable: pH 7 ± 0.85 pH (± 50 mV)
Electrode offsets greater than 0.85 pH (50 mV) indicate the electrode requires cleaning or replacement. See section 5: Maintenance and
cleaning.
3.2 Slope (SLP)
Electrode slope is the number of mV per pH unit. At 25°C the
theoretical slope is 59.16 mV per pH. Temperature has an
appreciable affect on electrode slope. Reliable instrumentation
includes temperature compensation. The graph below illustrates
potential pH error when a temperature compensated instrument
is not used.
°C pH Error
23456789101112
15 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15
25 00000000000
35 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15
45 0.3 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.3
55 0.45 0.36 0.27 0.18 0.09 0 0.09 0.18 0.27 0.36 0.45
Recommendations:
• Calibrate temperature before calibrating the standard and slope.
• The mV offset will track across the entire pH range. The slope is usually not affected by offset changes.
(i.e. pH 7= +10 mV, pH 4= +187 mV); slope = 59 mV
• Coatings on the glass may affect sensor slopes. See section 5 maintenance and cleaning.
•A constant output near 0 mV in all buffer solutions indicates a shorted electrode that must be replaced.
Theoretical mV Values @ 25°C
pH mV
2+ 295 mV
3+ 236 mV
4 +177 mV
5 +118 mV
6+ 59 mV
70 mV
8- 59 mV
9 - 118 mV
10 - 177 mV
11 - 236 mV
12 - 295 mV
3.3 Response Time/Stability
Response time and stability are affected by the condition of the pH electrode's glass surface (ORP electrode - Platinum surface),
reference junction, and reference solution. Restoration to acceptable levels can often be accomplished by cleaning the electrode's glass
surface (ORP electrode - Platinum surface) and reference junction.
Electrode mV values should remain stable ±3 mV. Conditions that may cause fluctuations are:
1. Electrode coating
2. Ground fault: If proper operation is observed in the beaker, but is unstable in the application, a ground fault probably exists.
• Using instrumentation with isolated inputs and outputs may restore stable operation.
• Solution grounding may also restore stable operation.
page 2 of 4 ‡ SIGNET 2714-2717 Twist Lock pH and ORP Electrodes