Georg Fischer 8850 Operating Manual

Signet 8850-1 Conductivity/Resistivity Transmitter
*3-8850.090-1*
3-8850.090-1 Rev. G 3/06 English
English
CAUTION!
• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
Contents
1. Installation
3. Electrical Connections
4. Menu Functions
62.50 uS/cm
Conductivity/Resistivity
1. Installation
ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary hardware to install the transmitter. This manual includes complete panel mounting instructions. Field mounting requires a separate mounting kit. The 3-8050 Universal kit enables the transmitter to be installed virtually anywhere. Detailed instructions for fi eld installation options are included with the 3-8050 Universal kit.
1.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
96 mm
(3.8 in.)
FRONT VIEW
Field Mount &
Panel Mount
96 mm (3.8 in.)
41 mm (1.6 in.
)
SIDE VIEW
Panel Mount
Optional Rear Cover
56 mm (2.2 in.)
97 mm (3.8 in.)
92 mm (3.6 in.)
8050
42 mm (1.7 in.)
64 mm (2.5 in.)
106 mm (4.2 in.)
SIDE VIEW
Field Mount w/ 8050 Universal base
82 mm (3.23 in.)
42 mm (1.7 in.)
107 mm (4.2 in.)
SIDE VIEW
Field Mount w/ 8052 Integral kit
8052
65 mm (2.5 in.)
96 mm
(3.8 in.)
25.0 C
Signet
Transmitter
ENTER
2. Specifi cations
General
Compatible electrodes: Signet 3-28XX-1 Standard and Certifi ed
Series Conductivity/Resistivity Electrodes Accuracy: ±2% of reading Enclosure:
• Rating: NEMA 4X/IP65 front
• Case: PBT
• Panel case gasket: Neoprene
• Window: Polyurethane coated polycarbonate
• Keypad: Sealed 4-key silicone rubber
• Weight: Approx. 325g (12 oz.) Display:
• Alphanumeric 2 x 16 LCD
• Contrast: User selected, 5 levels
• Update rate: 1.8 s
Electrical
• Power: 12 to 24 VDC ±10%, regulated, 80 mA max. Sensor input range:
• Conductivity: 0.01 µS/cm to 400,000 µS/cm
• Resistivity: 10 KΩ•cm to 100 MΩ•cm
• TDS: 0.023 to 200 000 PPM nominal (adjustable µS/PPM)
• Temperature: PT 1000, -25 to 120°C (-13 to 248°F) Measurements above 10 MΩ (below 0.1 µS) must be
performed in solution temperatures from 20°C to 100°C.
4 to 20 mA Outputs:
• Passive, isolated, fully adjustable and reversible 4 to 20 mA
outputs are independently source selectable for conductivity or temperature.
• Max loop impedance: 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V
• Update rate: 200 ms
• Accuracy: ±0.03 mA @ 25°C, 24 V Open-collector output, optically isolated:
• 50 mA max. sink, 30 VDC maximum pull-up voltage.
• Programmable for:
• High, Low with adjustable hysteresis
• Pulse operation (max. rate: 400 pulses/min).
Environmental
• Operating temperature: -10 to 70°C (14 to 158°F)
• Storage temperature: -15 to 80°C (5 to 176°F)
• Relative humidity: 0 to 95%, non-condensing
• Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
Standards and Approvals
• CE, UL listed
• Immunity: EN50082-2
• Emissions: EN55011, Class B
• Manufactured under ISO 9001 and ISO 14001
U.S.A. Patent # 5,708,363
3. Electrical Connections
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
2
1
Terminals 3 and 4: Loop Power
12-24 VDC ±10% system power
and current loop output.
Max. loop impedance:
50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V
System Pwr
4
Loop -
System Pwr
3
Loop +
AUX
2
Power -
AUX Power +
1
Terminals 1-2: Auxiliary power
Provides DC power to measurement circuit. Required for all 8850 systems
6
5
Output -
Output +
Internal open-collector
output circuit
S
Terminals 7-10: Sensor input
7 is conductivity input 8 is temperature input 9 is the isolated signal ground 10 is the sensor earth ground
9
8
7
Sensr Gnd (SHIELD)
Iso. Gnd (BLACK)
Temp. IN (WHITE)
Signal IN (RED)
10
15Ω
Output --
Output +
≤ 100 ft. (30 m)
3.1 Sensor Input Connections
Terminals
Shield (Sensor Gnd)
10
Black (Iso Gnd)
9
White (Temp In)
8
Red (Signal In)
7
Isolation
Terminals 5-6: Open-collector Output
• A transistor output, programmable (see CALIBRATE menu) as High or Low setpoint with adjustable hysteresis and proportional Pulse (to 400 pulses per minute maximum).
• May be disabled (Off) if not used.
D
62.50 uS/cm
25.0 C
Signet
Conductivity/Resistivity
Transmitter
ENTER
Signet 28XX-1
Standard and Certified Cells
2 8850-1 Conductivity/Resistivity Transmitter
Select the sensor with the range that best straddles your application requirements.
• 2819-1, 2839-1 (0.01 cell): 0.01 to 100 µS (10 kΩ to 100 MΩ) Use the 2819-1 or the 2839-1 for all resistivity requirements
• 2820-1, 2840-1 (0.1 cell): 1 to 1000 µS
• 2821-1, 2841-1 (1.0 cell): 10 to 10,000 µS
• 2822-1, 2842-1 (10.0 cell): 100 to 200,000 µS
• 2823-1 (20.0 cell): 200 to 400,000 µS
UPW @ 25°C
3-2819, 3-2839
Bottled
USP
Rinse Water
3-2820, 3-2840 ( 0.10 Cell)
Water
3-2822, 3-2842 (10.0 Cell)
3-2821, 3-2841 (1.0 Cell)
(0.01 Cell)
3-2823
Cooling Tower Waste Waters
Brackish Sea Water
(Salinity)
Deionization Regen. Chemicals
Alkali Cleaners, Acids/Bases
(20.0 Cell)
3.2 System Power/Loop Connections
Stand-alone application, no current loop used
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Connection to a PLC/Recorder, separate supply
Transmitter
Terminals
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
AUX power required for all 8850 systems
4
4
3
3
2
2
1
1
Transmitter
Terminals
PLC or Recorder
+
Loop Input
4-20 mA
-
Power
-
Supply
+
12-24 VDC
0.010mS 100 MΩ
4
4
3
3
2
2
1
1
0.054mS
18.3 MΩ
1mS 1MΩ
­+
10mS
100KΩ
Power
Supply
12-24 VDC
100mS
10KΩ
1,000mS 10,000mS
200mS
50,000mS
100,000mS
200,000mS
400,000mS
Connection to a PLC with built-in power supply
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Transmitter
Terminals
4
4
3
3
2
2
1
1
Internal
Connection
PLC
PLC Terminals
Loop Input
+
4-20 mA
-
Power
-
Supply
+
12-24 VDC
Example: Two transmitters connected to PLC/Recorder with separate power supply
PLC or Recorder
+
Channel 1 4-20 mA in
-
­+
­+
Channel 2 4-20 mA in
Power
Supply
12-24 VDC
Transmitter 2
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Transmitter 1
Terminals
4
3
2
1
Terminals
System Pwr Loop -
4
4
System Pwr Loop +
3
3
AUX Power -
2
2
AUX Power +
1
1
38850-1 Conductivity/Resistivity Transmitter
3.3 Open Collector Output
The Open Collector output can be used as a switch that responds when the process value moves above or below a setpoint, or it can be used to generate a pulse at a rate proportional to the process value.
Low: Output triggers when process variable is less than the setpoint. The output will relax when the process variable moves above the setpoint plus the hysteresis value.
High: Output triggers when process variable is greater than the setpoint. The output will relax when the process variable moves below the setpoint plus the hysteresis value.
Process
Hysteresis
Low Setpoint
Time
Relay energized
Relay relaxed
Proportional Pulsing The Open Collector output will generate a 100 mS pulse at the rate defi ned by settings in the CALIBRATE menu.
In the example below:
• The output will be 0 pulses/min. when value is less than 5.
• The output will be 50 pulses/min. when value is 7.5.
• The output will be 100 when value is greater than 10.
The starting point, endpoint and maximum pulse rate are select ­able in the CALIBRATE menu.
100 pulses
0 to 100 Pulses/min.
Open Collector Output Rate:
Pulse rate
0 pulses
Process
5
Starting point
High Setpoint
Hysteresis
Time
VIEW menu
• During normal operation, ProcessPro displays the VIEW menu.
• When editing the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no activity occurs for 10 minutes.
• To select a VIEW display, press the UP or DOWN arrow keys. The selections will scroll in a continuous loop.
• Changing the VIEW display does not interrupt system operations.
• No key code is necessary to change display selection.
• Output settings cannot be edited from the VIEW menu.
View Menu
Display Description
10
endpoint
62.50 uS/cm
Conductivity/Resistivity
Transmitter
25.0 C
Signet
ENTER
123.45 µS/cm +67.89 ºC
Monitor the Conductivity and Temperature values from the sensor. This is the permanent display.
The VIEW displays below are temporary. The permanent display will return after 10 minutes
Loop Output:
13.75 mA
Last CAL: 04-20-07
4 8850-1 Conductivity/Resistivity Transmitter
Monitor the Loop Current output.
Check the Last Calibration date.
Loading...
+ 8 hidden pages