• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
Contents
1.Installation
2.Specifi cations
3.Electrical Connections
4.Menu Functions
Signet pH/ORP
Transmitter
Relay 1
1. Installation
ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary
hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires a separate mounting kit. The 3-8050 Universal kit enables the transmitter to be installed virtually
anywhere. Detailed instructions for fi eld installation options are included with the 3-8050 Universal kit.
1.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive
template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
panel
terminals
d
n
sr G
)
en
S
D
L
IE
H
(S
7
r IN
s
en
S
)
D
E
(R
t -
6
u
r
tp
+
u
w
O
P
tem
ys
S
-
p
o
Lo
2
P
m
ste
y
S
+
p
o
o
L
1
r V
s
n
e
)
S
K
4
C
A
L
(B
5
r
w
t +
u
tp
u
O
3
quick-clips
96 mm
(3.8 in.)
FRONT VIEW
Field Mount &
Panel Mount
96 mm
(3.8 in.)
42 mm
(1.7 in.)
106 mm (4.2 in.)
SIDE VIEW
Field Mount
8050
64 mm
(2.5 in.)
82 mm
(3.23 in.)
41 mm
(1.6 in.
)
SIDE VIEW
Panel Mount
Optional
Rear
Cover
56 mm
(2.2 in.)
97 mm
(3.8 in.)
92 mm
(3.6 in.)
gasket
latch
Panel Mount
Installation Detail
10.20 pH
25.0 C
mounting
Relay 2
ENTER
bracket
2. Specifi cations
General
Compatible electrodes: Signet 2714-2717 Electrodes with 2720
pH/ORP Preamplifi er, 2764-2767 and
2774-2777 Electrodes with 2760-1,-2
pH/ORP Preamplifi er
Accuracy: ±0.03 pH, ± 2 mV's ORP
Enclosure:
• Rating: NEMA 4X/IP65 front
• Case: PBT
• Panel case gasket: Neoprene
• Window: Polyurethane coated polycarbonate
• Keypad: Sealed 4-key silicone rubber
• Weight: Approx. 325g (12 oz.)
Display:
• Alphanumeric 2 x 16 LCD
• Contrast: User selected, 5 levels
• Update rate: 1 second
Electrical
Power: 12 to 24 VDC ±10%, regulated, 220 mA max.
Sensor input range:
• pH: 0.00 to 14.00 pH
• Temp. (pH only) 3K Balco, -25 to 120°C (-13 to 248°F)
• 4 to 20 mA, isolated, fully adjustable and reversible
• Max loop impedance: 50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
• Update rate: 0.5 seconds
• Accuracy: ±0.03 mA @ 25°C, 24 V
Relay outputs (2 sets mechanical SPDT contacts):
• Maximum voltage rating: 5 A @ 30 VDC, or 5 A @ 250 VAC
resistive load
• Hi or Lo programmable with adjustable hysteresis
• Pulse programmable (maximum 400 pulses/minute)
Environmental
• Operating temperature: -10 to 70°C (14 to 158°F)
• Storage temperature: -15 to 80°C (5 to 176°F)
• Relative humidity: 0 to 95%, non-condensing
• Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
Standards and Approvals
• CE, UL listed
• Immunity: EN50082-2
• Emissions: EN55011 Class B
• Manufactured under ISO 9001 and ISO 14001
3. Electrical Connections
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
Terminals 5-10: Relay Outputs
Two sets mechanical SPDT contacts,
programmable (see CALIBRATE menu) as:
• High or Low with adjustable hysteresis
• proportional pulse based on pH or ORP
• May be disabled (Off) if not used.
Terminals 15-18: Sensor connections
15-16 are mV (pH or ORP) input from the electrode.
17-18 are temperature input from the pH electrode or
10KΩ identifying resistor from the ORP electrode.
2
1
Terminals 3 and 4: Loop Power
12-24 VDC ±10% system power
and current loop output.
Max. loop impedance:
50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
4
3
2
1
Terminals 1 and 2: AUXILIARY power
Provides power for relay operation.
System Pwr
Loop -
System Pwr
Loop +
AUX
Power -
AUX
Power +
Tem p -
18
(WHITE)
Tem p +
17
(GREEN)
Iso. GND
(BLUE)
16
Relay 2
10
(NO)
Relay 2
9
(COM)
Relay 2
8
(NC)
Relay 1
7
(NO)
Relay 1
6
(COM)
Relay 1
5
(NC)
Terminals 11-14: Preamplifi er power and grounds
11-12 are DC voltages from the 8750 to power the
preamplifi er.
13-14 are ground terminals for the sensor and for
earth ground
15
14
13
12
11
mV Input
(BROWN)
Earth
GND
Sensr Gnd
(SHIELD)
V(BLACK)
V+
(RED)
2
8750-2 pH/ORP Transmitter
3.1 Sensor Input Connections
Terminals
*
* (connection of terminal 14 to Earth GND
White (Temp)
18
Green (Temp)
17
Blue (ISO GND)
16
Brown (mV input)
15
14
Silver (Shield)
13
Black (V-)
12
Red (V+)
11
may reduce electrical interference)
3.2 System Power/Loop Connections
Stand-alone application, no current loop used
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Preamp:
Signet 2720
Twist-Lock
Electrodes
Transmitter
Terminals
4
4
3
3
2
2
1
1
Preamp:
Signet 2760-1
Signet 2760-2
DryLoc
Electrodes
+
Transmitter
10.20 pH
Signet pH/ORP
Transmitter
Relay 1
Power
Supply
12-24 VDC
25.0 C
Relay 2
ENTER
≤ 400 ft. (122 m)
Connection to a PLC/Recorder, separate supply
Connection to a PLC with built-in power supply
PLC or Recorder
Transmitter
Terminals
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Transmitter
Terminals
4
4
3
3
2
2
1
1
+
-
-
12-24 VDC
+
Loop Input
4-20 mA
Power
Supply
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
AUX power required for all 8750-2 systems
Example: Two transmitters connected to PLC/Recorder with separate power supply
PLC or Recorder
+
Channel 1
4-20 mA in
-
+
Channel 2
4-20 mA in
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Transmitter 1
Terminals
4
3
2
1
4
4
3
3
2
2
1
1
Transmitter 2
Terminals
4
4
3
3
2
2
1
1
Internal
PLC
Connection
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
PLC Terminals
Loop Input
+
4-20 mA
-
+
Power
Supply
12-24 VDC
-
12-24 VDC
+
Power
Supply
38750-2 pH/ORP Transmitter
3.3 Relay Output
The relay outputs can be programmed to respond when the pH/
ORP value moves above or below a setpoint, or it can be used to
generate a pulse that is proportional to the pH/ORP value.
The relays can be disabled if not in use.
• Low
Relay triggers when the pH/ORP is less than the setpoint.
The relay will relax when the pH/ORP moves above the setpoint
plus the hysteresis value.
• High
Relay triggers when the pH/ORPis greater than the setpoint.
The relay will relax when the pH/ORP drops below the setpoint
plus the hysteresis value.
Process
• Proportional Pulsing
The relay output will generate a 100 mS pulse at the rate defi ned
by settings in the CALIBRATE menu (see page 6)
In the example below:
• The output will be 0 pulses/min. at pH values less than 5.0.
• The output will be 50 pulses/min. at 7.5 pH.
• The output will be 100 pulses/min. at pH values above 10 pH.
100 pulses
Hysteresis
Low Setpoint
Relay Output Rate:
0 to 100 Pulses/min.
0 pulses
Relay energized
Time
5
Starting point
10
endpoint
Relay relaxed
Process
High Setpoint
Hysteresis
Time
VIEW menu
• During normal operation, ProcessPro displays the VIEW menu.
• When using the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no activity
occurs for 10 minutes.
• To select the item you want displayed, press the UP or DOWN arrow keys. The items will scroll in a continuous
loop. Changing the display selection does not interrupt system operations.
• No key code is necessary to change display selection.
• Output settings cannot be edited from the VIEW menu.
Pulse rate
10.20 pH
25.0 C
Signet pH/ORP
Transmitter
Relay 1
Relay 2
ENTER
View Menu for pH
DisplayDescription
7.00 pH
12.6 ºC
All of the VIEW displays below are temporary. The permanent display will return after ten minutes.
Input:
307 mV
Loop Output:
14.16 mA
Last CAL:
03-17-66 >
EASY
CAL: >
4
Monitor the pH and the Temperature input from the sensor.
This is the Permanent display.
Monitor the millivolt input from the electrode. Use this display to determine the
relative condition of your electrode during periodic calibration.
(7 pH buffer = 0 mV, ±50 mV)
Monitor the 4-20 mA Loop output.
Monitor date for scheduled maintenance or date of last calibration.
Easy Cal is the fastest and simplest periodic calibration method. Requires
4 pH, 7 pH and 10 pH. (Any two)
8750-2 pH/ORP Transmitter