• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
Contents
1.Installation
2.Specifi cations
3.Electrical Connections
4.Menu Functions
Signet pH/ORP
Transmitter
Relay 1
1. Installation
ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary
hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires a separate mounting kit. The 3-8050 Universal kit enables the transmitter to be installed virtually
anywhere. Detailed instructions for fi eld installation options are included with the 3-8050 Universal kit.
1.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive
template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
panel
terminals
d
n
sr G
)
en
S
D
L
IE
H
(S
7
r IN
s
en
S
)
D
E
(R
t -
6
u
r
tp
+
u
w
O
P
tem
ys
S
-
p
o
Lo
2
P
m
ste
y
S
+
p
o
o
L
1
r V
s
n
e
)
S
K
4
C
A
L
(B
5
r
w
t +
u
tp
u
O
3
quick-clips
96 mm
(3.8 in.)
FRONT VIEW
Field Mount &
Panel Mount
96 mm
(3.8 in.)
42 mm
(1.7 in.)
106 mm (4.2 in.)
SIDE VIEW
Field Mount
8050
64 mm
(2.5 in.)
82 mm
(3.23 in.)
41 mm
(1.6 in.
)
SIDE VIEW
Panel Mount
Optional
Rear
Cover
56 mm
(2.2 in.)
97 mm
(3.8 in.)
92 mm
(3.6 in.)
gasket
latch
Panel Mount
Installation Detail
10.20 pH
25.0 C
mounting
Relay 2
ENTER
bracket
2. Specifi cations
General
Compatible electrodes: Signet 2714-2717 Electrodes with 2720
pH/ORP Preamplifi er, 2764-2767 and
2774-2777 Electrodes with 2760-1,-2
pH/ORP Preamplifi er
Accuracy: ±0.03 pH, ± 2 mV's ORP
Enclosure:
• Rating: NEMA 4X/IP65 front
• Case: PBT
• Panel case gasket: Neoprene
• Window: Polyurethane coated polycarbonate
• Keypad: Sealed 4-key silicone rubber
• Weight: Approx. 325g (12 oz.)
Display:
• Alphanumeric 2 x 16 LCD
• Contrast: User selected, 5 levels
• Update rate: 1 second
Electrical
Power: 12 to 24 VDC ±10%, regulated, 220 mA max.
Sensor input range:
• pH: 0.00 to 14.00 pH
• Temp. (pH only) 3K Balco, -25 to 120°C (-13 to 248°F)
• 4 to 20 mA, isolated, fully adjustable and reversible
• Max loop impedance: 50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
• Update rate: 0.5 seconds
• Accuracy: ±0.03 mA @ 25°C, 24 V
Relay outputs (2 sets mechanical SPDT contacts):
• Maximum voltage rating: 5 A @ 30 VDC, or 5 A @ 250 VAC
resistive load
• Hi or Lo programmable with adjustable hysteresis
• Pulse programmable (maximum 400 pulses/minute)
Environmental
• Operating temperature: -10 to 70°C (14 to 158°F)
• Storage temperature: -15 to 80°C (5 to 176°F)
• Relative humidity: 0 to 95%, non-condensing
• Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
Standards and Approvals
• CE, UL listed
• Immunity: EN50082-2
• Emissions: EN55011 Class B
• Manufactured under ISO 9001 and ISO 14001
Page 2
3. Electrical Connections
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
Terminals 5-10: Relay Outputs
Two sets mechanical SPDT contacts,
programmable (see CALIBRATE menu) as:
• High or Low with adjustable hysteresis
• proportional pulse based on pH or ORP
• May be disabled (Off) if not used.
Terminals 15-18: Sensor connections
15-16 are mV (pH or ORP) input from the electrode.
17-18 are temperature input from the pH electrode or
10KΩ identifying resistor from the ORP electrode.
2
1
Terminals 3 and 4: Loop Power
12-24 VDC ±10% system power
and current loop output.
Max. loop impedance:
50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
4
3
2
1
Terminals 1 and 2: AUXILIARY power
Provides power for relay operation.
System Pwr
Loop -
System Pwr
Loop +
AUX
Power -
AUX
Power +
Tem p -
18
(WHITE)
Tem p +
17
(GREEN)
Iso. GND
(BLUE)
16
Relay 2
10
(NO)
Relay 2
9
(COM)
Relay 2
8
(NC)
Relay 1
7
(NO)
Relay 1
6
(COM)
Relay 1
5
(NC)
Terminals 11-14: Preamplifi er power and grounds
11-12 are DC voltages from the 8750 to power the
preamplifi er.
13-14 are ground terminals for the sensor and for
earth ground
15
14
13
12
11
mV Input
(BROWN)
Earth
GND
Sensr Gnd
(SHIELD)
V(BLACK)
V+
(RED)
2
8750-2 pH/ORP Transmitter
Page 3
3.1 Sensor Input Connections
Terminals
*
* (connection of terminal 14 to Earth GND
White (Temp)
18
Green (Temp)
17
Blue (ISO GND)
16
Brown (mV input)
15
14
Silver (Shield)
13
Black (V-)
12
Red (V+)
11
may reduce electrical interference)
3.2 System Power/Loop Connections
Stand-alone application, no current loop used
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Preamp:
Signet 2720
Twist-Lock
Electrodes
Transmitter
Terminals
4
4
3
3
2
2
1
1
Preamp:
Signet 2760-1
Signet 2760-2
DryLoc
Electrodes
+
Transmitter
10.20 pH
Signet pH/ORP
Transmitter
Relay 1
Power
Supply
12-24 VDC
25.0 C
Relay 2
ENTER
≤ 400 ft. (122 m)
Connection to a PLC/Recorder, separate supply
Connection to a PLC with built-in power supply
PLC or Recorder
Transmitter
Terminals
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Transmitter
Terminals
4
4
3
3
2
2
1
1
+
-
-
12-24 VDC
+
Loop Input
4-20 mA
Power
Supply
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
AUX power required for all 8750-2 systems
Example: Two transmitters connected to PLC/Recorder with separate power supply
PLC or Recorder
+
Channel 1
4-20 mA in
-
+
Channel 2
4-20 mA in
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
Transmitter 1
Terminals
4
3
2
1
4
4
3
3
2
2
1
1
Transmitter 2
Terminals
4
4
3
3
2
2
1
1
Internal
PLC
Connection
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
PLC Terminals
Loop Input
+
4-20 mA
-
+
Power
Supply
12-24 VDC
-
12-24 VDC
+
Power
Supply
38750-2 pH/ORP Transmitter
Page 4
3.3 Relay Output
The relay outputs can be programmed to respond when the pH/
ORP value moves above or below a setpoint, or it can be used to
generate a pulse that is proportional to the pH/ORP value.
The relays can be disabled if not in use.
• Low
Relay triggers when the pH/ORP is less than the setpoint.
The relay will relax when the pH/ORP moves above the setpoint
plus the hysteresis value.
• High
Relay triggers when the pH/ORPis greater than the setpoint.
The relay will relax when the pH/ORP drops below the setpoint
plus the hysteresis value.
Process
• Proportional Pulsing
The relay output will generate a 100 mS pulse at the rate defi ned
by settings in the CALIBRATE menu (see page 6)
In the example below:
• The output will be 0 pulses/min. at pH values less than 5.0.
• The output will be 50 pulses/min. at 7.5 pH.
• The output will be 100 pulses/min. at pH values above 10 pH.
100 pulses
Hysteresis
Low Setpoint
Relay Output Rate:
0 to 100 Pulses/min.
0 pulses
Relay energized
Time
5
Starting point
10
endpoint
Relay relaxed
Process
High Setpoint
Hysteresis
Time
VIEW menu
• During normal operation, ProcessPro displays the VIEW menu.
• When using the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no activity
occurs for 10 minutes.
• To select the item you want displayed, press the UP or DOWN arrow keys. The items will scroll in a continuous
loop. Changing the display selection does not interrupt system operations.
• No key code is necessary to change display selection.
• Output settings cannot be edited from the VIEW menu.
Pulse rate
10.20 pH
25.0 C
Signet pH/ORP
Transmitter
Relay 1
Relay 2
ENTER
View Menu for pH
DisplayDescription
7.00 pH
12.6 ºC
All of the VIEW displays below are temporary. The permanent display will return after ten minutes.
Input:
307 mV
Loop Output:
14.16 mA
Last CAL:
03-17-66 >
EASY
CAL: >
4
Monitor the pH and the Temperature input from the sensor.
This is the Permanent display.
Monitor the millivolt input from the electrode. Use this display to determine the
relative condition of your electrode during periodic calibration.
(7 pH buffer = 0 mV, ±50 mV)
Monitor the 4-20 mA Loop output.
Monitor date for scheduled maintenance or date of last calibration.
Easy Cal is the fastest and simplest periodic calibration method. Requires
4 pH, 7 pH and 10 pH. (Any two)
8750-2 pH/ORP Transmitter
Page 5
ProcessPro Editing Procedure:
Step 1. Press and hold ENTER key:
• 2 seconds to select the CALIBRATE menu
• 5 seconds to select the OPTIONS menu.
Step 2. The Key Code is UP-UP-UP-DOWN keys in sequence.
• After entering the Key Code, the display will show the fi rst item in the selected menu.
Step 3. Scroll menu with UP or DOWN arrow keys.
Step 4. Press RIGHT ARROW key to select menu item to be edited.
• The fi rst display element will begin fl ashing.
Step 5. Press UP or DOWN keys to edit the fl ashing element.
• RIGHT ARROW key advances the fl ashing element.
Step 6. Press ENTER key to save the new setting and return to Step 3.
Notes on Step 1:
• The View Menu is normally displayed.
• The CALIBRATE and OPTIONS menus require a KEY CODE.
Press &
hold for
access:
VIEW
ENTER
Notes on Step 2:
If no key is pressed for 5 minutes while display is showing "Enter
Key Code", the display will return to the VIEW menu.
2s5s
CALIBRATE
OPTIONS
Notes on Steps 3 and 4:
• Refer to pages 6 and 7 for complete listing of menu items and their use.
• From the Step 3 display, pressing the UP and DOWN keys simultaneously will
return the display to the VIEW menu.
• If no key is pressed for 10 minutes, display will also return to the VIEW menu.
Step 3: Finished Editing?
Press the UP and DOWN keys simultaneously after
saving the last setting to return to normal operation.
CALIBRATE: ----
Enter Key Code
CALIBRATE: *---
Enter Key Code
CALIBRATE: **--
Enter Key Code
CALIBRATE: ***-
Enter Key Code
Temperature >
Temperature >
Relay1 Setpnt:
10.0 pH >
Set:
Set:
Step 3
First item in CALIBRATE menu
Step 4
Notes on Steps 5 and 6:
• All output functions remain active during editing.
Relay1 Setpnt:
• Only the fl ashing element can be edited.
• RIGHT ARROW key advances the fl ashing element in a continuous loop.
• Edited value is effective immediately after pressing ENTER key.
• If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.
• Step 6 (pressing ENTER key) always returns you to Step 3.
• Repeat steps 3-6 until all editing is completed.
Step 5: Made an Error?
Press the UP and DOWN keys simultaneously
while any element is fl ashing. This will recall the
last saved value of the item being edited and
return you to Step 3.
1
0.00 pH
Relay1 Setpnt:
0
0.00 pH
Relay1 Setpnt:
0
Step 6
Relay1 Setpnt:
9.00 pH >
0
.00 pH
Relay1 Setpnt:
Step 5
9
.00 pH
0
Relay1 Setpnt:
ENTER
Saving
58750-2 pH/ORP Transmitter
Page 6
Calibrate Menu for pH
Display
(Factory settings shown)
Description
Set:
Temperature >
Provides a maximum 25ºC offset to match temperature measurement to external reference.
Set:
Applies a linear offset to the pH measurement. The ideal value is the average pH of your
application. (A sample of your application at process temperature is recommended.)
Standard >
Set:
Slope >
Applies a slope to the pH measurement. The slope value and the standard value must be
at least 2 pH units apart. The ideal values are the minimum and maximum values of your
process.
Loop Range: pH
Select the minimum and maximum values for the 4-20 mA Current loop output.
0.00 › 14.00 >
Relay1 Source:
pH >
Select pH or Temperature as the source for this relay.
Relay1 Mode:
Select the mode of operation for this relay: High, Low or proportional Pulse.
The signal may be disabled if not in use.
Off >
Relay1 Setpnt:
In Low or High Mode, the relay will be activated when the pH reaches this value.
4.00 pH >
Relay1 Hys:
0.51 pH >
In Low or High mode, the relay will be deactivated at Setpoint ± Hysteresis, depending on
High or Low Setpoint selection. (See details on page 4.)
Relay1 Range:
4.00 → 8.00 >
Relay1 PlsRate:
120 Pulses/min >
Last CAL:
09-24-05 >
Settings repeat for Relay 2.
If Relay 1 is in PULSE mode, set the start and end point of the range and also set the
maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.)
The combined Relay1 Range and Relay1 Pulse rate settings shown here indicate:
"Start pulsing when the pH value is 4 and increase the pulse rate up to the maximum of
120 pulses per minute when the pH value reaches 8"
Use this “note pad” to record important dates, such as annual recertifi cation or scheduled
maintenance.
6
8750-2 pH/ORP Transmitter
Page 7
EASY CAL Procedure - pH
• This procedure simplifi es system calibration using standard 4.0, 7.0, 10.0 pH buffers only. If these pH buffers are not available,
calibrate the system via the CALIBRATE menu, using the STANDARD and SLOPE settings.
• Access the CALIBRATE menu and set sensor temperature before performing EASY CAL for new electrode installations.
• Access EASY CAL menu from the view menu.
EASY CAL: ---Enter Key Code
Press UP, UP, UP, DOWN buttons in sequence to enter menu,
_ _ _ _
XXXX will appear during code entry.
To Calibrate:
Place Sensor in
pH Buffer #1
Place electrode tip in
first pH buffer
pH 7.0 = 0 mV
pH 4.0 = 177
pH 10 = -177
Limit ± 50 mV
Place Sensor in
pH Buffer #2
Place electrode tip
in second pH buffer.
1
2
To exit menus and return to
VIEW press UP and DOWN
button at the same time
Response:
* 6.90 pH
-005 mV
Allow for stabilization
30 seconds*
* 3.93 pH
+179 mV
Allow for stabilization
30 seconds*
Display returns to VIEW
Menu in 10 minutes or
when ENTER is pressed
Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher.
Select lower contrast if the display is in warmer ambient surroundings.
OFF provides the most instantaneous output response to changes in input value.
LOW averaging = 4 seconds, HIGH averaging = 8 seconds of input signal.
Select temperature units: ºC or ºF.
Adjust the minimum and maximum current output. The display value represents the
precise current output.
Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 ma
Use this setting to match the 8750 loop output to any external device.
Press UP or DOWN keys to manually order any output current value from 3.6 mA to
21.00 mA to test current loop output.
Press UP or DOWN keys to manually toggle the state of the relay output.
Press UP or DOWN keys to manually toggle the state of the relay output.
Description
78750-2 pH/ORP Transmitter
Page 8
Calibrate Menu for ORP
Display
(Factory settings shown)
Set:
Standard >
Set:
Slope >
Loop Range: mV
-1000 > +1000 >
Relay1 Mode:
Off >
Relay1 Setpnt:
-500 mV >
Relay1 Hys:
10 mV >
Relay1 Range: mV
-500 → +500 >
Relay1 PlsRate:
120 pulses/min >
Last Cal:
09-24-05 >
Description
Applies a linear offset to the ORP measurement. If you are using a single-point calibration,
use this function. The ideal value is the average pH of your application. (A sample of your
application at process temperature is recommended.)
Applies a slope to the ORP measurement. This function should only be used in a two-point
calibration. The slope value and the standard value must be at least 2 pH units apart. The
ideal values are the minimum and maximum values of your application.
Select the minimum and maximum ORP values for the 4-20 mA Current loop output.
Maximum settings are -1000 mV to +2000 mV.
Select the mode of operation for this relay: High, Low or proportional Pulse.
The signal may also be disabled if not in use.
In Low or High Mode, the relay will be activated when the ORP reaches this value.
In Low or High mode, the relay will be deactivated at Setpoint ± Hysteresis, depending on
High or Low Setpoint selection. (See details on page 4.)
If Relay 1 is in PULSE mode, set the start and end point of the range and also set the
maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.)
The combined Relay1 Range and Relay1 Pulse rate settings shown here indicate:
"Start pulsing when the ORP value is -500 and increase the pulse rate up to the maximum
of 120 pulses per minute when the ORP value reaches +500"
Use this “note pad” to record important dates, such as annual recertifi cation or scheduled
maintenance.
8
8750-2 pH/ORP Transmitter
Page 9
EASY CAL Procedure - ORP
• This procedure simplifi es system calibration using standard 4.0 pH and 7.0 pH buffers saturated with Quinhydrone. To calibrate
using any other ORP buffer solutions, use the Standard and Slope functions in the CALIBRATE menu.
• Access EASY CAL menu from the view menu.
EASY CAL: ---Enter Key Code
To Calibrate:
Place Sensor in
ORP Buffer #1
Place electrode tip
in first pH buffer;
pH 7.0 Å 87 mV
pH 4.0 Å 246 mV
Place Sensor in
ORP Buffer #2
Place electrode tip
in second (different)
pH buffer.
pH 4.0 Å 246 mV
pH 7.0 Å 87 mV
To exit menus and return to
VIEW press UP and DOWN
button at the same time
Press UP, UP, UP, DOWN buttons in sequence to enter menu,
will appear during code entry.
1
2
Response:
* ORP: + 84 mV
Input: + 82 mV
Allow for stabilizationto accept
30 seconds*
* ORP: +262 mV
Input: +260 mV
Allow for stabilization
30 seconds*
Display returns to VIEW
Menu in 10 minutes or
when ENTER is pressed
To Accept:
* ORP: + 84 mV
Input: + 82 mV
ENTER
* ORP: + 87 mV
Input: + 82 mV
* ORP: +262 mV
Input: +260 mV
ENTER
Press to accept
second buffer calibration.
* ORP: +264 mV
Input: +260 mV
Good Easy Cal
Press <ENTER>
*
For best results, gently stir the
submerged electrode for approximately 5
seconds during the stabilization period.
Large temperature differences from pro-
cess fl uids to buffers may require longer
stabilization time.
Technical notes:
The difference between the actual mV and
value shown is a good indication of the
condition of the electrode.
Differences in excess of 50 mV may indicate a need to service the electrode.
Options Menu for ORP
Display
(Factory settings shown)
Contrast:
3 >
Averaging:
Off >
Loop Adjust:
4.00 mA >
Loop Adjust:
20.00 mA >
Test Loop:
>
Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher.
Select lower contrast if the display is in warmer ambient surroundings.
OFF provides the most instantaneous output response to changes in input value.
LOW averaging = 4 seconds, HIGH averaging produces more stable display and
output response.
Adjust the minimum and maximum current output. The display value represents the
precise current output.
Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 ma
Use this setting to match the 8750 loop output to any external device.
Press UP or DOWN keys to manually order any output current value from 3.6 mA to
21.00 mA to test current loop output.
Description
Test Relay 1:
Press UP or DOWN keys to manually toggle the state of the relay.
>
Test Relay 2:
>
Press UP or DOWN keys to manually toggle the state of the relay.
98750-2 pH/ORP Transmitter
Page 10
Troubleshooting - pH
Display ConditionPossible causes
During EasyCal:
"Out of Range
Use Manual CAL"
During EasyCal:
"Same Buffer"
During CALIBRATE Std:
"Standard too close to
Slope!"
During CALIBRATE Slope:
"Slope too close to
Standard!"
During CALIBRATE:
"Out of Range
Check Sensor"
1. Required 4, 7 or 10 buffers not being
used.
2. Sensor is depleted too severely to use
EasyCal.
1. Sensor was not moved from buffer #1 to
buffer #2.
2. Buffer is contaminated
1. pH Standard value within 2 pH units of
Slope value.
2. pH Sensor effi ciency is inadequate.
1. pH Slope value within 2 pH units of
Standard value.
2. pH Sensor effi ciency is inadequate.
1. No temperature or mV signal from
sensor detected.
2. No connection between pH sensor and
preamplifi er.
Suggested Solutions
1. Use pH 4, 7, 10 buffers.
2. Clean probe and retry EASY CAL.
3. Use Manual calibration for Standard and
Slope if mV offset exceeds 50 mV.
1. Place sensor in correct buffer solution.
2. Use fresh buffer.
1. Use pH values at least 2 pH units apart.
2. Clean pH sensor; replace if necessary.
3. Use fresh buffer.
1. Use pH values at least 2 pH units apart.
2. Clean pH sensor; replace if necessary
3. Use fresh buffer.
1. Check all wiring, contacts in preamplifi er.
2. Verify sensor is securely installed.
3. Replace electrode pH sensor.
During normal operation:
Constant "15.00 pH"
or
constant "0.00 pH"
with good temp value
During normal operation:
"Check Sensor?"
The mV value from the sensor when placed in a 7 pH buffer represents the sensor offset. Signet recommends servicing/replacing the
sensor when the offset exceeds 50 mV.
The 3-2759 pH/ORP system tester allows simple system troubleshooting.
mV input is less than 0 pH or greater than
15 pH.
1. No temperature or mV signal from
sensor detected.
2. No connection between pH sensor
and preamplifi er.
During CALIBRATE Std:
"Standard too close to
Slope!"
During CALIBRATE Slope:
"Slope too close to
Standard!"
During CALIBRATE:
"Out of Range
Check Sensor"
During normal operation:
Constant "-1000"
or
constant ""+2000"
1. Required 4, 7 buffers with
quinhydrone not being used.
2. Sensor is depleted too severely to
use EasyCal.
1. Sensor was not moved from buffer #1
to buffer #2.
2. Buffer is contaminated
1. ORP Standard value within 120 mV
of Slope value.
2. ORP Sensor effi ciency is inadequate.
1. ORP Slope value within 120 mV of
Standard value.
2. ORP Sensor effi ciency is inadequate.
1. No mV signal or sensor id from
sensor detected.
2. No connection between ORP sensor
and preamplifi er.
mV input is less than -999 or greater
than +1999.
1. Use pH 4, 7 buffers saturated with
quinhydrone.
2. Clean probe and retry EASY CAL.
3. Use Manual calibration for Standard and
Slope if mV offset exceeds 50 mV.
1. Place sensor in correct buffer solution.
2. Use fresh buffer.
1. Use ORP values at least 120 mV apart.
2. Clean ORP sensor; replace if necessary.
3. Use fresh buffer.
1. Use ORP values at least 120 mV apart.
2. Clean ORP sensor; replace if necessary
3. Use fresh buffer.
1. Check all wiring, contacts in preamplifi er.
2. Verify sensor is securely installed.
3. Replace ORP sensor.
1. Recalibrate system.
2. Replace pH sensor.
3. Replace preamplifi er.
During normal operation:
"Check Sensor?"
1. No temperature or mV signal from
sensor detected.
2. No connection between ORP sensor
and preamplifi er.
1. Check all wiring, contacts in preamplifi er.
2. Verify sensor is securely installed.
3. Replace ORP sensor.
118750-2 pH/ORP Transmitter
Page 12
Ordering Information
Mfr. Part No. Code Description
3-8750-1 159 000 053 pH/ORP transmitter Field mount
3-8750-1P 159 000 054 pH/ORP transmitter Panel mount
3-8750-2 159 000 055 pH/ORP transmitter Field mount with relays
3-8750-2P 159 000 056 pH/ORP transmitter Panel mount with relays
3-8750-3 159 000 057 pH/ORP transmitter Field mount with single input/dual output
3-8750-3P 159 000 058 pH/ORP transmitter Panel mount with single input/dual output
3-8050 159 000 184 Universal mounting kit
3-8050.395 159 000 186 Splashproof rear cover
3-8050.396 159 000 617 RC Filter kit (for relay use)
3-8052 159 000 188 3/4” Integral mounting kit
3-8052-1 159 000 755 3/4" NPT mount junction box
3-0000.596 159 000 641 Heavy duty wall mount bracket
3-0700.390 198 864 403 pH buffer kit
3-8050.392 159 000 640 Model 200 retrofi t adapter
3-5000.598 198 840 225 Surface mount bracket
3-5000.399 198 840 224 5 x 5 inch adapter plate for Signet retrofi t
3-9000.392 159 000 368 Liquid-tight connector kit, 3 sets, 1/2 in. NPT
3-9000.392-1 159 000 839 Liquid-tight connector kit, 1 set, 1/2 in. NPT
3-9000.392-2 159 000 841 Liquid-tight connector kit, 1 set, PG 13.5
7300-7524 159 000 687 24 VDC Power Supply 7.5 Ω, 300mA
7300-1524 159 000 688 24 VDC Power Supply 15 Ω, 600mA
7300-3024 159 000 689 24 VDC Power Supply 30 Ω, 1.3 A
7300-5024 159 000 690 24 VDC Power Supply 50 Ω, 2.1 A
7300-1024 159 000 691 24 VDC Power Supply 100 Ω, 4.2 A
3-3719-11 159 000 804 pH/ORP Wet-Tap, 11/2 in. NPT
3-3719-21 159 000 805 pH/ORP Wet-Tap, 2 in. NPT
3-3719-12 159 000 806 pH/ORP Wet-Tap, ISO 7/1-R 1.5
3-3719-22 159 000 807 Wet-Tap Assembly, ISO 7/1-R 2
3-2716-WT 159 000 809 Twist-Lock pH Electrode (use with 3719 Wet-Tap)
3-2717-WT 159 000 811 Twist-Lock ORP Electrode (use with 3719 Wet-Tap)
2007-0225 159 000 812 PP Clamp-on Saddle, 2.5 in. x 11/2 in. (ASTM, NPT)
2007-0230 159 000 813 PP Clamp-on Saddle, 3 in. x 11/2 in. (ASTM, NPT
2007-0240 159 000 814 PP Clamp-on Saddle, 4 in. x 11/2 in. (ASTM, NPT)
2007-0260 159 000 815 PP Clamp-on Saddle, 6 in. x 2 in. (ASTM, NPT)
2007-0280 159 000 816 PP Clamp-on Saddle, 8 in. x 2 in. (ASTM, NPT)
2007-0210 159 000 817 PP Clamp-on Saddle, 10 in. x 2 in. (ASTM, NPT)
2007-0212 159 000 818 PP Clamp-on Saddle, 12 in. x 2 in. (ASTM, NPT)
1
/4 in. NPT
George Fischer Signet Inc., 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090