• Remove power to unit before wiring input and output
connections.
• Follow instructions carefully to avoid personal injury.
6. Menu Functions
7. Parts and Accessories
8. Specifi cations
9. Quick Reference Menu Parameters
10. Troubleshooting
11. Maintenance
3.
1 2
Std. Sensor
U
L
1 2
1 2 3 4 5 6
Freq. IN
LISTED
77CJ
Freq. IN
Sen. Pwr.
Open Collector
Sensor
12-24 V
10 W
+
+
Iso. Gnd
output
1 2 3 4
1
Gnd
+
Gnd
Total
AUX
PLS
reset
output
-
2
Flow
Patent No. D376,328
1. Power Connections2. Compatible Electrode/Preamp Wiring
CAUTION!
Never connect 110 VAC or 220 VAC to
rear power terminals. High voltage AC
will damage instrument and void warranty.
5700
12-24 V
10 W
Terminals
External power
+
-
= Double Insulate
= DC or AC power
supply
-
+
12 - 24 VDC
OR
12 - 24 VAC
Technical Notes:
• To reduce the possibility of noise interference, isolate AC power lines from signal lines.
• Maximum 4-20 mA loop impedance (sec. 3) is affected by the supply voltage.
• Use six conductor shielded cable for cable extensions up to 120 m (400 ft) max. (#5523-0064)
• Cable shield must be maintained through cable splice.
Silver (earth ground)
Brown
Blue
Green
White
Red
Black
T -
T +
Green wire
V +
White wire
Red Wire
Blue wire
Brown wire
Iso. Gnd
mV Input
V -
Black wire
+
4-20
mA
Preamp:
Signet 2720
Twist-Lock
-
Electrodes
pH/ORP
5700
Terminals
Preamp:
Signet 2760-1
Signet 2760-2
DryLoc
Electrodes
3. 4 - 20 mA Current Output Connections
Technical Notes:
**1/8A fuse recommended (customer supplied)
* 4-20 mA output is internally powered (non-isolated), maximum
loop impedance 350 Ω with a 12 V instrument supply voltage,
950 Ω with a 24 V instrument supply voltage.
To isolate output and prevent ground loop problems:
1. Use monitor device with isolated inputs, or
2. Use separate DC supply for 5700 and monitor device, or
3. Power 5700 with12 - 24 VAC step down transformer
Fuse**
1/8A
+
4-20
mA
4 - 20 mA
+
*
-
-
pH/ORP
5700
Terminals
4. Relay Connections
Two internal relay contact sets (COM, NO, and NC) may be used
for external device control. Front panel LED annunciators indicate
the activation status of each relay. Each relay can control up to
two devices simultaneously, as shown. Relay operation modes
include Low alarm, High alarm, and Proportional Pulse (sec. 5).
Common device connections include:
• Pulse mode - metering pump control
• Pulse mode - solenoid valve control
• Low or High mode - warning lamps
• Low or High mode - bells or sirens
• Low or High mode - external heavy-duty relay
Wiring Example Right
Device A IS powered when relay 2 is de-energized (front panel
LED “off”). Power is discontinued when the relay 2 setpoint is
reached (front panel LED “on”). Device B IS NOT powered when
the relay 2 is de-energized. Power is applied after the relay 2
setpoint is reached.
5. Relay Operation
A. LOW alarm mode
The relay is energized when the ORP drops
below the setpoint, and is de-energized when
the ORP rises above the setpoint plus
hysteresis (sec. 6.3D, 6.3E).
Low setpoint=
Adjustable hysteresis=
Relay energized=
Relay de-energized=
ORP
Relay
Energized
C
NO
NC
5700
Terminals
NO
C
NC
+
External AC/DC
Relay 1Relay 2
NO
power
Relay
De-energized
C
NO
NC
C
NC
Device
+
A
-
-
-
Device
B
+
Technical Notes:
• Maximum relay contact ratings: 5 A @ 30 VDC, 5 A @ 125
VAC, or 3 A @ 250 VAC
• An external heavy-duty relay must be used for devices with
surge currents or operating currents that exceed the above
specifi cations.
B. HIGH alarm mode
The relay is energized when the ORP rises
above the setpoint and is de-engergized
when the ORP falls below the setpoint plus
hysteresis (sec. 6.3D, 6.3E).
High setpoint=
Adjustable hysteresis=
Relay energized=
Relay de-energized=
ORP
C. Proportional PULSE mode
The proportional pulse relay configuration is primarily designed for metering pump control. The
operator is prompted to enter a minimum and maximum ORP setpoint and maximum pulse rate for
the assigned relay (sec. 6.3F, 6.3G). Relay pulse width is fixed at 130 ms. Refer to the operation
examples below.
• Metering pump chemical addition (dry contact activation type required)
Example 1 (right):
As the process ORP rises above the minimum pulse setpoint (-500 mV) the relay
begins pulsing; triggering the metering pump for chemical addition. As the ORP
continues to rise, pulsing accelerates proportionally until the maximum programmed
pulse rate of 300 pulses/minute and setpoint (+500 mV) are reached, forcing the
process ORP back down to intended levels (e.g. ≤ -500 mV).
ORP
-1000
mV
Example 2 (right):
As the process ORP falls below the minimum pulse setpoint (+500 mV) the relay
begins pulsing; triggering the metering pump for chemical addition. As the ORP
continues to decrease, pulsing accelerates proportionally until the maximum
programmed pulse rate of 300 pulses/minute and setpoint (-500 mV) are
reached, forcing the process ORP back up to intended levels (e.g. ≥ +500 mV).
-1000
mV
2
300 pulses
0 pulses
Relay 1 Rate:
300 Pulses/min >
Relay 1 or
Relay 2
-500 mV
3
300 pulses
Relay 1 or
Relay 2
-500 mV
Signet 5700 pH/ORP Monitor - ORP Instructions
+500 mV
0 pulses
+500 mV
+1000
ORP
+1000
mV
mV
6. Menu Functions
ENTER
To access either CALIBRATE or OPTIONS menus, press and hold
the ENTER key as illustrated below:
Menus:
• VIEW menu (sec. 6.1): The VIEW menu is displayed during
standard operation. The operator can navigate freely through
the menu by pressing either UP or DOWN arrow keys.
VIEW
(sec. 6.1)
CALIBRATE
(sec. 6.3)
OPTIONS
6.1 VIEW Menu
1
Choose:
↓G
A.
ORP mV/Raw mV
(sec. 6.4)
Press & hold ENTER key
for access:
ENTER
2
seconds
seconds
Change:
• CALIBRATE Menu (sec. 6.3): The CALIBRATE menu contains
all critical display setup and output parameters. A simple security code feature prevents unauthorized tampering. The operator
is required to enter a simple access code for menu access. The
same code also unlocks OPTIONS menus.
• OPTIONS Menu (sec. 6.4): The OPTIONS menu contains
setup and display features that are seldom accessed for minor
5
display or output adjustments.
• Reversible Dial
Your 5700 includes a
reversible dial face for
pH use. See enclosed
pH manual for opera-
BA
0 - 14 pH
ALARM 1
pH
ALARM 2
0
D
14
C
ALARM 1
ORP
ALARM 2
-
1000
+1000
tion details.
2
3
Save:
Menu Displays A - G:
(Factory default displays shown in
menu column 1)
A. Shows calibrated solution ORP
mV and Raw (uncalibrated) input
mV from ORP electrode/preamplifi e r
B.
C.
D.
E.
F.
Last calibration
G.
Buffer calibration
Loop output
Range
Relay 1
Relay 2
Press and release to access
EASY CAL Procedure
↑A
See EASY CAL Procedure (Section 6.2)
B. Loop output display: shows the
loop current output level.
C. ORP range display: shows the
programmed min and max meter
range (sec. 6.3A)
D. Relay 1 display: this display
shows the programmed operation mode and setpoint for relay 1
(sec.
6.3D-E)
E. Relay 2 display: this display
shows the programmed operation mode and setpoint for relay 2
(sec.
6.3D-E)
F. Last calibration: this display
shows a user defi ned setup date
for
maintenance records. This
feature is not an internal timer or
calender.
G. EASY CAL: press the right arrow
key to access the EASY CAL
buffer calibration procedure (sec.
6.2).
Signet 5700 pH/ORP Monitor - ORP Instructions
3
6.2 EASY CAL Procedure
Requirements
• This procedure simplifi es system calibration using standard pH 4.0 and 7.0 buffers saturated with
Quinhydrone (customer supplied). If standard pH 4.0 and 7.0 buffers and Quinhydrone are not
available, the system can be calibrated using the CALIBRATE menu standard and slope settings
(sec. 6.3B, 6.3C).
• Quinhydrone (customer supplied) is the oxidizer measured by the ORP electrode and is essential for EASY CAL calibration. To guarantee saturation with pH 4.0 and 7.0 buffers, mix 1/8 g (1/4
mL) of Quinhydrone powder for each 50 mL of pH buffer solution (sec. 7).
1
To Calibrate:
Place electrode tip in
first pH buffer saturated
with Quinhydrone;
pH 7.0 ≈ 87 mV
pH 4.0 ≈ 264 mV
Place electrode tip in
second (different) pH
buffer saturated with
Quinhydrone;
• Input mV: Shows the actual
Raw (uncalibrated) preamplifi er
ORP electrode input
For best results, gently stir the
submerged electrode for
approximately 5 seconds during
the stabilization period (step 2).
pH 4.0 ≈ 264 mV
pH 7.0 ≈ 87 mV
To exit EASY CAL at any time
without saving changes:
2
30 seconds*
quick
press
Display returns to VIEW
Menu in 3s
Technical Note:
The differences between input mV and ORP mV is a good indication of the electrode's condition.
Differences in excess of 50 mV may indicate a need to service the electrode. Refer to electrode
manual for maintenance recommendations.