A.3 Example of File in EM61MK2win Format ............................................. 52
Appendix B ....................................................................................................... 53
B.1 Using the EM61MK2win with a GPS Receiver ....................................... 53
B.2 Description of GGA and GSA Data Messages ...................................... 54
1. Introduction
e Geonics EM61MK2win Data Logging System (DAS70-win) consists of a tablet/laptop eld
computer, data logging program EM61MK2win, and associated cable to connect the computer
to the Geonics EM61-MK2 instrument. e program EM61MK2win is designed for a IBM PC
compatible computer and MS Windows 10 (or 7) operating system.
e program is prepared for the MS Windows 10 tablet Mesa2 (manufactured by Juniper Systems)
operated in the vertical mode display with the default setup. However the program can be used in
other Windows 10( or 7) based eld computers with vertical and horizontal oriented displays. e
minimum supported display resolution is 800 x 600 pixels. In case high resolution display is employed, it is advised to set fonts to 125% or 150% to increase legibility of the screen. e maximum
size of program main windows is 1024 x 762 pixels.
e EM61MK2win program acquires and records survey data from the EM61-MK2 system, under
the control of the operator. It also records various eld information such as survey line number (line
name), starting station, increment, comments, etc. Readings are displayed in graphic and numeric
mode. Readings are displayed in real time in mV. In addition, the program allows you to monitor
the instrument output while data is not recorded. e EM61MK2win program continuously monitors the condition of the instrument battery, without leaving the program. e EM61MK2 also
provides the possibility of automatic nulling of the instrument output at any time during the survey.
e program allows the user to set the EM61-MK2 into a specic instrument mode of operation:
AUTO, Wheel, or Manual modes. In AUTO mode readings can be automatically recorded in desired
time intervals. In WHEEL mode readings are triggered by a counter installed at the EM61-MK2
wheel assembly, and in MANUAL mode readings are triggered manually by the operator.
e program supports the standard EM61-MK2 instrument as well as EM61-MK2 High Power
unit. It also recognizes all versions of instrument rmware automatically. e program allows you
to record data while using various EM61-MK2 antenna (1 x 0.5 m, 1 x 1 m, or 0.5 x 0.5 m sensors)
and Geonics EM61HH-MK2 Hand Held sensor.
e EM61MK2win will accept NMEA-0183 compatible data from a GPS receiver directly connected to an Allegro eld computer. GPS data which are embedded in the EM61MK2win data le
can be processed later in the Geonics DAT61MK2 program. e connected GPS must be able to
stream NMEA-0183 compatible messages. e EM61MK2win uses two NMEA messages GGA
and GSA. While message GGA is mandatory, the GSA string is used only to provide information
related to the GPS signal quality during data collection.
e EM61MK2win program records data together with a time stamp at each station. Data les created with this program can be used to position a survey according to locations recorded separately
by a Global Positioning System (GPS).
Survey setup parameters are saved in a conguration le, therefore they can be automatically used
during subsequent data collection sessions.
Data les are saved to the user selected data directory. Data le names, which can be set by the
program based on the computer clock or user specied, have extension names R61. Files can be
converted to M61 format and viewed in DAT61MK2 program. Number of readings in a data le
is limited only by the hard drive capacity, however it is strongly recommended to avoid huge data
les (i..e one to two hours long data les are adequate). e maximum speed of data collection is
approximately 18 readings per second assuming 5 Hz (or less) GPS input.
EM61MK2win Data Logging System (DAS70-M2)
1
1.1 Program Requirements
To successfully use this software, you will need :
Computer
- Mesa2 eld computer (or similar) operating under Windows 10 or 7,
- minimum 4 Gb of RAM memory,
- USB port or other mean to transfer les,
- minimum 800 x 600 pixel display resolution,
- serial Ports:
one serial port for instrument plus one serial port if directly connected GPS receiver
is to be used (USB to multiple RS-232 ports adapters can be used) or available
Bluetooth connection in serial port mode.
Geonics EM61-MK2
e EM61-MK2 instrument with associated cables (or Bluetooth option).
e EM61MK2win program may be stored on DAT61MK2 CD disk and it is available for download from the Geonics web site. All necessary conguration les (with extension names .INI) are
created in your eld computer after the program is run for the rst time.
1.2 Installing EM61MK2win
e program consists of a standalone le EM61MK2win.exe, no DDLs nor other libraries are
required to run this program.
If the program le is included separately on the DAT61MK2 CD disk, or downloaded from the
web site it can be installed by copying the EM61MK2win.exe le to any directory on your hard
disk. is can be performed by using Windows Explorer.
After you run the program for the rst time it will create permanent conguration le (.INI) which
contain some of the program settings. If the above le does not exist on the default drive they will
be created with default parameters during the rst execution of the program.
e EM61MK2win data les contain extension name R61 and their base names should be limited
to 8 characters.
1.3 Running EM61MK2win Program
Start EM61MK2win by double clicking the EM61MK2win icon in the Start|Programs menu, in
Windows Explorer, or on the desktop if a shortcut was created. At the start, EM61MK2win displays
the following screen.
e EM61MK2win is a command button and dialog driven program. Command buttons can be
executed by taping (if touch screen is used) or clicking with the left mouse button, or by pressing
the indicated (underlined) character on the keyboard, or by using TAB to scroll through the buttons
and ENTER to execute. In the Main Screen of the program Up and Down arrow keys can be also
used to scroll through the buttons.
2
e Main Screen appears always as the rst window after the program is started. It contains the
name of the program, its version number, and list of command buttons with available options on
Geonics Limited
the right side. e major , left portion of the window contains information about current survey
and system settings. is section is updated in real time as soon as the program parameters are
changed in dialogs.
1.4 Main Screen
e Main Screen always appears after the program is started. It contains the name of the program,
its version number, and a list of buttons representing the available options. e EM61MK2win
Main Screen is displayed below.
ese options are selected by clicking on buttons, or from keyboard using TAB/ENTER or shortcuts
indicated by underline label characters. In case of Main Screen buttons the Down and Up arrow
keys and <ENTER> can be used as well.
Short description for each of the options follows.
Short description of Main Screen options
Logging
is option allows to monitor and log the EM61-MK2 output. e Logging screen starts
always in Monitor mode, and then after a data le is created Logging mode is available.
Monitor mode provides several options: the instrument nulling, performing Internal and
External QC coil calibration, monitoring EM61-MK2 battery level, monitoring quality
of GPS signal, etc.
EM61MK2win Data Logging System (DAS70-M2)
3
Survey Setup
e Survey Setup dialog will be displayed. All survey settings (survey line name, increments,
etc.) can be specied in this dialog.
System Setup
Dialog window that is associated with this option is used to set the instrument type, antenna
size, serial port numbers, battery warning level, and type of pause key.
4
Set GPS Port
e GPS Port Setup dialog allows to disable and enable GPS data acquisition. is option is
also used to set the serial port number used for GPS input and to specify necessary serial port
communication settings. GPS monitoring window can be accessed from this dialog as well.
GPS Monitoring
e program enters terminal mode and displays input from GPS. is function is very
convenient to test communication with GPS receiver and examine GPS stream type.
Display Options
e dialog which will allow you to specify colour and thickness of prole lines will appear.
is dialog provides also choice of linear or compressed amplitude used for prole display.
View Files
is option can be used to open collected data le and display survey line proles.
Exit
is option will terminate the program execution.
Geonics Limited
2. Survey Setup
e Survey Setup dialog, presented below, contains several parameters which aect two important
procedures: instrument settings (instrument mode, frequency of data collection, etc.) and survey
geometry layout (survey line names, line spacing, start station, station increment, etc.).
To select any option click on the corresponding drop-down list box or text box, or use TAB key to
scroll to the option and then use mouse or the Down and Up keys in drop-down box (parameters
that have only a few possible options), or keyboard for text box entry.
Description of Options and Parameters:
EM61-MK2 Mode
is option allows you to select the type of instrument used. In the drop-down list box
select Standard for the EM61-MK2 or High Power if EM61-MK2 High Power (HP)
modication is used.
Set the EM61-MK2 mode of operation by selecting an item from the drop-down listbox
labeled EM61 Mode. Available modes are: Auto, Wheel, and Manual. ese modes of
operation are described below.
EM61MK2win Data Logging System (DAS70-M2)
5
Auto Mode
Readings will be triggered automatically at a specied frequency (see option
Readings/s).
Please note that while in the former data acquisition system (DAS70) Stn Increment (station increment) had only two options: Positive or Negative, in the current
program this value is optional.
Wheel Mode
Readings will be triggered automatically by a counter attached to the wheel. Wheel
increment is approximately 20 cm (or 0.64 foot) for the EM61-MK2 equipped
with 1 x 0.5 m or 1 x 1 m antennas. e Hand Held EM61HH-MK2 has two
wheel increments 0.1 and 0.2 m. Check the wheel increment setting on the antenna assembly.
When Wheel mode is selected option Reading/second is not available, see gure
below. Please note that while in former data acquisition system (DAS70) Stn
Increment (station increment) had only two options: Positive or Negative, in the
current program this parameter is optional and its value has to be entered according to used wheel increment.
Manual Mode
Readings will be taken only while the manual trigger (switch on the logger cable) is
pressed. is mode may be used only in very specic applications since the highly
6
dynamic EM61-MK2 response requires nely spaced data points.
Please note that when Manual mode is selected option Reading/second is not
available.
Geonics Limited
Readings/second
When this option is available (only in Auto mode) activate text box by tapping or using
TAB key and then enter desired value. is parameter describes number of readings per
second that will be taken.
ere is a dierence between EM61-MK2 units whether they include the communication
option or not.
In cases of older units not using Bluetooth any number larger than zero can be entered,
however the EM61-MK2 maximum frequency of data output is approximately 15 readings per second. For values lower than 15 readings/s the real rate of data collection will be
relatively close to specied value.
e EM61-MK2 with Bluetooth communication has a slightly dierent way of streaming data. Again any number larger than zero can be entered, however in this EM61-MK2
model the hardware xed frequency of data output is 18 readings per second. Any entered
value is rounded to the nearest possible base frequency of the instrument, i.e. user specied
frequency of 14 or more readings/s will result in the eective data acquisition rate of 18
readings/s (every reading is logged), 10 readings/s will cause the program to log at a rate of
9 readings/s (every second reading from instrument output is logged), etc. At slower rates
the dierence between specied and eective frequencies is negligible.
Survey Line (survey line name)
Activate text box by clicking with a mouse left button or using TAB key and then enter
desired name (number) for the survey line.
is is a user’s tag number/name for the prole line. e length of the name can not exceed
8 characters. e line name is usually used as a coordinate perpendicular to the survey lines
direction. For example, when survey lines are laid out along W-E direction stations describe
W-E coordinate, while Line names may describe S-N (vertical on a map) coordinate.
Line Incr. (survey line name)
Activate text box by clicking with a mouse or using TAB key and then enter desired for
the survey line increment.
is parameter species the distance by which survey lines will be separated. is setting
will be used to determine number (name) of the next survey line.
Sequence
When this option is highlighted and drop-down box is expanded use mouse or use Down
or Up cursor key to toggle between two available settings: Alternate and One Way.
EM61MK2win Data Logging System (DAS70-M2)
7
Alternate is used when neighboring lines are surveyed in the opposite direction, which is
the most common procedure during eld surveys.
One Way is used when each survey line is traversed in the same direction.
e choice of this parameter will aect the default start station, a signature of the station
increment, and line direction when parameters for the next survey lines is determined.
Direction
Clicking on the down arrow next to the text box opens a drop-down box showing the
available options, or when the keyboard is used activate the text box and then by using
up or down arrow keys select one of four available settings: East, West, South, and North.
is parameter indicates the heading of the survey line.
Start Station (start station of a survey line)
Activate text box by clicking or using TAB key and then enter the desired value for the
start station.
is parameter species the starting station number for the selected survey line. is value
is used in conjunction with Station Increment to calculate the current station number for
display purposes.
Stn Increment (station increment)
Activate text box by clicking with a mouse or using TAB key and then enter the desired
value for the station increment.
is parameter species the station increment for the selected survey line. is value is used
in conjunction with Start Station to calculate the current station number for display purposes.
After all the parameters in the Survey Setup dialog are updated click on the button OK or press
ENTER key to accept the displayed settings. e program will return to the Main Screen. Updated
settings will be written to the initial le and they will be given as default parameters in the subsequent Survey Setup dialog.
To return to original settings (state before this dialog was selected) click on the Cancel (X) button
or press Esc key. All parameters will be reset to initial settings and the program will return to the
Main Screen.
8
Geonics Limited
3. System Setup
is dialog allows you to specify used type of the instrument, EM61-MK2 antenna size and set
several parameters in the logging program. e System Setup dialog is presented below.
Description of Parameters:
EM61-MK2
is option allows you to select type of used instrument. In the drop-down list box select
Standard for the EM61-MK2 or High Power if EM61-MK2 High Power (HP) modication is used.
Sensor Size
e Sensor Size option allows you to select the size of the EM61-MK2 sensor. If the
EM61HH-MK2 Hand Held antenna is used this option can be ignored, the program recognizes Hand Held antenna connection automatically. If the standard antennas are used (1
x 0.5 m, 1 x 1 m, or 0.5 x 0.5 m) the size of the sensor must be specied before monitoring
the instrument output. Select proper size of the EM61-MK2 sensor in the drop-down list
box labeled Sensor Size.
EM61MK2win Data Logging System (DAS70-M2)
9
Battery Warning
Specify instrument battery warning level. When the EM61-MK2 battery voltage will
reach and drop below specied value the program will display battery voltage on the red
background.
EM61 Port
e number of serial port that is assigned to the EM61-MK2. Available selections: COM1
to COM60. e program default is COM1. Communication parameters for the selected
10
Geonics Limited
serial port are set by the program, since the EM61-MK2 operates at xed settings: Baud
Rate (9600), Parity (N), Data Bits (8), and Bit Stop (1).
is port must be dierent than the port specied in the list box "GPS Port:" (see below) or
in the GPS Port Setup dialog window (see chapter 4), otherwise a message will be displayed
and ports will have to be reassigned.
Select port number in the drop-down list box.
GPS Port
e number of serial port that is assigned to the GPS receiver. Available selections: COM1
to COM60. e program default is COM2. Communication parameters for the selected
serial port can be set in GPS Port Setup dialog. e GPS port can be also assigned in the
latter dialog.
is port must be dierent than the port specied as the EM61-MK2 serial port.
Select port number in the drop-down list box similarly to EM61-MK2 port assignment
procedure.
Audio
Two selections are available: Yes or No. e audible click will be generated at each reading
when this option is enabled. At fast rate (above 5 readings/s) the sound may appear not
uniform, however this does not aect quality of data acquisition.
Display Units
ree selections are available: Meters, Feet or US Survey Feet.
Pause key
Four selections are available: any key, Enter, Space Bar, and P is feature is used to
pause data recording during logging session. Default setting any key can be changed to
one of the three specic keys for eld conditions where a logger key can be accidentally
pushed causing unwanted stop of data logging. When Windows based tablet computer
Mesa2 is used the best setting is <ENTER> key, since user specied keys P1, P2, and P3
may have various assignment, and when <any key> is selected then only four arrow keys
and Enter key are reliable.
EM61MK2win Data Logging System (DAS70-M2)
11
Disable Pause Button (for Touch Screen)
is option is useful when Mesa2 and similar computer with enable touch screen are used.
It helps avoiding logging procedure by accidental touch of the computer display. When this
option is checked the touch operation is disabled for the Pause button and logging can be
paused only by pressing computer key (for example <Enter>).
In case of a conicting Pause Key (lack of specied key in the computer) a soft keyboard
can be opened and used at any moment. is option aects only sensitivity of the Pause
button, other touch enabled functions are not aected.
After all the parameters in the System Setup dialog are updated click on the button OK or press
ENTER key to accept the displayed settings. e program will return to the Main Screen. Updated
settings will be written to the initial le and they will be given as default parameters in the subsequent System Setup dialog.
To return to original settings (state before this dialog was selected) click on the Cancel (or window
X) button or press Esc key. All parameters will be reset to initial settings and the program will return
to the Main Screen.
12
Geonics Limited
4. GPS Port Setup & Monitoring
e GPS Port Setup dialog allows for enabling GPS input, and setting communication parameters
for serial port associated with GPS input. GPS Monitoring window allows you monitoring the GPS
output in terminal mode and to send NMEA commands to GPS receivers.
4.1 GPS Port Setup
After the Set Port for GPS button was clicked on (or executed from the keyboard) in the Main
Screen the Set Port for GPS dialog window appears on the screen. is dialog allows you to enable
and disable the GPS input, select serial port for GPS input, and to specify serial communication
parameters matching GPS receiver settings. e dialog is presented below.
Description of Options and Parameters:
GPS Input
is option allows you to Enable/Disable a serial port for GPS input. When Disabled is
chosen logging and monitoring screens will display message “GPS disabled” in place of
GPS parameters.
e GPS Input can be Enabled even if there is no GPS system connected to the computer.
In such case data le will contain proper sequence of EM61-MK2 readings without any
GPS input.
Clicking on the down arrow next to the text box labeled GPS Input opens a drop-down
box showing the available settings, or when the keyboard is used activate the text box by
pressing TAB key and then by using up or down arrow keys select one of two available
items: Enabled and Disabled.
EM61MK2win Data Logging System (DAS70-M2)
13
Serial Port
e number of serial port that is assigned to the GPS input. Available selections: COM1
to COM60. e program default is COM2. Communication parameters for the selected
serial port can be determined in options described below.
14
is port must be dierent than the port specied in the System Setup menu (for EM61MK2), otherwise a message will be displayed and ports will have to be reassigned.
Clicking on the down arrow next to the text box labeled Serial Port opens a drop-down box
showing the available ports, or when the keyboard is used activate the text box by pressing
TAB key and then by using up or down arrow keys select one of available items.
Baud Rate
Specify Baud Rate for the output port, the selected value should match the Baud Rate of
the GPS system, default is 9600. Available settings are: 4800, 9600, 19200, 38400, 57600,
and 115200.
Clicking on the down arrow next to the text box labeled Baud Rate opens a drop-down
box showing the available parameters, or when the keyboard is used activate this text box
by pressing TAB key and then by using up or down arrow keys select one of available items.
Geonics Limited
Parity
Select Parity for the output port, the parameter should much the Parity set in the GPS serial
port settings. Available settings are None, Even, and Odd; default is N.
Clicking on the down arrow next to the text box labeled Parity opens a drop-down box
showing the available parameters, or when the keyboard is used activate this text box by
pressing TAB key and then by using up or down arrow keys select one of available items.
Data Bits
Specify Data Bits for the output port, the selected value should much settings in the GPS
receiver, default is 8. Other available selection is 7.
Clicking on the down arrow next to the text box labeled Data Bits opens a drop-down box
showing the available parameters, or when the keyboard is used activate this text box by
pressing TAB key and then by using up or down arrow keys select one of available items.
Stop Bits
Specify Stop Bits for the output port, the selected value should much settings in the GPS
receiver, default is 1. Available selections are 1 or 2.
Clicking on the down arrow next to the text box labeled Stop Bits opens a drop-down box
showing the available parameters, or when the keyboard is used activate this text box by
pressing TAB key and then by using up or down arrow keys select one of available items.
After all the parameters in the GPS Port Setup dialog are updated click on the button OK or press
ENTER key to accept the displayed settings. e program will return to the Main Screen. Updated
settings will be written to the initial le and they will be given as default parameters in the subsequent Survey Setup dialog.
To return to original settings (state before this dialog was selected) click on the Cancel (X) button
or press Esc key. All parameters will be reset to initial settings and the dialog window will disappear.
EM61MK2win Data Logging System (DAS70-M2)
15
4.2 Monitoring GPS Receiver Output
After the Main Screen command button GPS Monitoring is clicked or executed by the keyboard
the program will display the GPS Monitoring window in terminal mode. In this mode the screen
is divided into three parts. e terminal window in the top portion of the screen displays the GPS
receiver output. e middle portion labeled NMEA Command is used to display entered NMEA
commands previously sent to the GPS receiver, and at the bottom, command buttons with available options (Pause/Go, Send, and Exit) is displayed. ese buttons can be clicked on, executed by
TAB key and ENTER keys, or by using shortcuts (pressing underlined letter keys). is screen is
shown below.
As soon as the EM61MK2win screen is in terminal mode and the GPS is streaming data, each message transmitted by GPS receiver will appear in the top portion of the display (the end may be cut
o if an NMEA message is longer than screen). e display is updated with the frequency the GPS
receiver outputs data. is allows you to recognize the GPS update rate and type of messages being
sent by the connected GPS. In cases where the GPS data is not received by the logger a message NO
DATA and current time will appear in the top window of the display, as shown below.
16
e message NO DATA is normally updated with a rate of 7 seconds. is may indicate the following: serial port number not correctly specied in Set Port for GPS dialog, GPS receiver not sending
any data, and not working or not connected GPS receiver. If the message is updated more often than
7 seconds (i.e. every 1 or 2 seconds) or the display does not show legible characters, it is possible
that the GPS is working correctly and is connected to the proper serial port, however communication parameters are not specied correctly. In most cases the Baud Rate or Parity must be adjusted.
Geonics Limited
e NO DATA message may also appear if the GPS data are received correctly, but the GPS receiver
was set to send data with a time interval longer than 7 seconds. In this case the NO DATA message
will be displayed in between GPS messages. is indicates that the GPS is working correctly, however
the operator should consider adjustment of the GPS receiver output update rate. Most high resolution geophysical surveys require positioning update of 1 or 2 seconds, and a 5 seconds interval can
be used only when the survey is carried out at an even pace and along relatively straight survey lines.
e monitoring display can be stopped any time by using button labeled Pause. At that time scrolling
of the GPS output will be stopped, and the button will be labeled Go. e next click (or keyboard
action) on this button will activate receiving and display of GPS data.
e button labeled Send allows you to send a NMEA command to the GPS receiver. It is preferable
if the GPS receiver parameters are set using the GPS manufacturer software or controller (GPS logger
or panel keys). However, when the operator is familiar with NMEA protocol and structure of commands for a given GPS system, this function can be very convenient and useful when the
update rate and enabling or disabling messages in the data stream is required.
After the button Send is clicked a dialog titled GPS Message is displayed and the beginning of the
standard NMEA command, $PASHS, or the last entered command is displayed. After the entire
NMEA command is typed in, click OK button or press the key <ENTER> to send the command
to the GPS receiver. Clicking the Cancel button or pressing the <Esc> key will cancel the command
and hide the NMEA Message dialog. An example of a command that will enable the NMEA message SAT is given in the below gure (it is assumed that the GPS receiver output serial port is A).
EM61MK2win Data Logging System (DAS70-M2)
17
After this command is received by the GPS receiver, the conrmation message will be send by the
receiver ($PASHR, ACK*3D) and data stream will contain the message SAT if it is available in the
connected GPS receiver.
Please note, that not every GPS system accepts and uses the same standard set of NMEA commands
and messages. In addition, some GPS systems do not accept commands sent by the serial port at all.
e conguration of these type of receivers can be updated only by the controlling device (usually
GPS logger, controller, or the receiver panel keys).
18
Geonics Limited
5. Display Options
After the Display Options button was clicked (or executed from the keyboard) in the Main Screen
the Setup Display Options dialog window appears on the screen. is dialog allows you to enable
and disable the display of each channel prole, specify color and thickness of proles, and select
linear or compressed amplitude for proles. e dialog is presented below.
Description of Options and Parameters:
Visible
Color
To enable or disable displaying of each channel prole click on the corresponding button
labeled Visible. Proles of all channels with checked buttons will be displayed during data
logging.
Regardless of which channels are chosen to be displayed as proles, data for all four channels will be displayed in numeric form below prole display.
To change colour of each prole click on the corresponding button (with colour line) labeled
Color (click to change). e following dialog titled Color will appear.
Select desired colour by clicking on a colour box (the selected colour box will be highlighted).
EM61MK2win Data Logging System (DAS70-M2)
19
Click on the button OK or press ENTER key to accept the highlighted colour. e Color
dialog will disappear and the colour of the appropriate channel button will updated. To
cancel colour selection click Cancel (X) button or press Esc key.
Thickness
Specify thickness of a prole for a desired channel by using one of four drop-down boxes
labeled thickness. ickness of a prole curve is expressed in pixels. Available settings are:
1, 2, 3, or 4 pixels.
Clicking on the down arrow next to the text box (labeled by number of pixels) opens a
drop-down box showing available selection (see Figure below). Select thickness by clicking
on the desired selection. If keyboard is used activate text box by pressing TAB key (till the
box is highlighted) and then by using up or down arrow keys select one of available items.
Prole Amplitude
is option allows you to select Linear or Compressed amplitude scale for proles. e
compressed amplitude (squre root function) allows you to display the high dynamic range
of the EM61-MK2 data in a legible way and it is recommended.
Clicking on the down arrow next to the text box labeled Prole Amplitude opens a dropdown box showing available parameters, or when keybord is used activate this text box by
pressing TAB key and then by using up or down arrow keys select one of available items.
Please note, that readings in numeric form are always given in mV (linear scale) regardless of the Prole Amplitude selection. Further, graphic display in the compressed
scale does not aect readings saved in the data le. Data are always written to le in
original form.
After all the parameters in the Display Options dialog are updated click the button OK or press
ENTER key to accept the displayed settings. e program will return to the Main Screen. Updated
settings will be written to the initial le and they will be given as default parameters in the subsequent Survey Setup dialog.
To return to original settings (state before this dialog was selected) click the Cancel (X) button or
press Esc key. All parameters will be reset to initial settings and the dialog window will disappear.
20
Geonics Limited
6. Logging Data
After the Logging button (in Main Screen) is clicked or executed by the keyboard, the program
enters the logging session which contains three modes: Monitoring, Stand By and Logging. e
program starts the logging session always in Monitoring mode. In this mode EM61-MK2 readings
and GPS parameters can be quickly examined and some functions (creating data le, nuling, QC
coil check) can be performed. After a data le is created in Monitoring mode, the program switches
to Stand By mode and allows access to Logging mode. In Stand By mode instrument output can
be monitored and some survey and logging parameters can be changed, and Logging mode is used
only to record data. Two modes Stand By and Logging are toggled by GO and Pause buttons.
6.1 Monitoring Mode
e Monitor mode allows initial inspection of the range of the instrument readings at a particular
site, monitoring the instrument performance, quick inspection of the condition of the instrument
battery, perform QC Coil check, and setting zero level of the instrument.
It is assumed that the instrument is turned ON prior to using this option. If the instrument is OFF
or the instrument console is not connected to the computer the following message shown below
will appear.
EM61MK2win Data Logging System (DAS70-M2)
21
Check the connection or turn the instrument ON and select the Monitor/Log option again.
Assuming that the instrument works properly the program will display the Monitor mode window
and will start normalization session which lasts less than 10 seconds. After the normalization is
nished the layout of the Monitor window is as follows (left - during Normalization, right - after
Normalization):
22
e Monitor (as well as Stand by and Logging) window is divided into ve sections.
One line of text in the top section provides survey parameters (used in Stand By and Logging modes)
and instrument information. In Monitor mode this section displays: instrument mode (4 or D),
current display mode (Monitor, Stand By or Logging), and the EM61-MK2 battery level in Volts
(following label B:). When ducial marker is pressed then label Marker is displayed.
e second, largest graphic section will be used to display proles during logging data. Numeric
labels on the right side of prole window correspond to scale minimum (below the prole window)
and maximum (above the prole window).
e remaining bottom section contains a window with current EM61-MK2 readings, three frames
(File Parameters, Line Parameters, and GPS information), and command buttons located on the
right side of the screen.
Geonics Limited
If Mode 4 was selected on the instrument panel (it is displayed at the top of the screen) channels 1,
2, 3, and 4 will be displayed under prole window, and when Mode D was selected (for Top and
Bottom antennas) channels 1, 2, 3, and T will be given as labels in numeric window. e EM61MK2 readings are updated approximately 10 times per second during the monitoring session.
When GPS is Disabled in the Set GPS Port dialog a message GPS Disabled will be displayed. If the
GPS port is Enabled and a working GPS receiver is connected to the eld computer the Monitoring
screen will display GPS parameters, as presented below.
In the above Figure one line of the display is dedicated to show the GPS status. A label Positions
will be used during data acquisition, it will provide the number of GPS positions saved in the data
le. Below, two lines display the parameters describing the quality of GPS positions. A label DGPS
(Dierential Global Positioning System) indicates that GPS readings are dierentially corrected in
real time, while label AGPS (Autonomous Global Positioning System) indicates lack of dierential
correction. On the right side of DGPS or AGPS label two green and white circles alternate. ese
two circles should alternate colours with the frequency of GPS update rate (usually 1 second intervals). If the circles do not change colours for long periods of time it means that the GPS system is
not working or that it is not connected to the eld computer. e next label PDOP with a value
varying between 0 and 99.9 represents an index called Position Dilution of Precision (PDOP). is
value is available only when messages both, GGA and GSA are received from GPS. e last label
Sat and following number shows number of currently tracked satellites. Refer to section 4 (Set Port
for GPS), Appendix B, and to GPS manuals for more information about GPS parameters. Two
bottom lines of the frame give current Latitude and Longitude of the instrument position (datum
WGS1984). Position is updated with frequency of GPS receiver stream.
EM61MK2win Data Logging System (DAS70-M2)
23
Six command buttons located in the right bottom corner of the Monitor window lists the program
functions available directly in the Monitor mode. Available options are as follows: File (create data
le), Scale Readings (change scale for EM61-MK2 readings in prole window), Exit (exits the
Monitor mode and returns the program to Main Screen, Null (perform nulling function), Int. QC
(internal QC-coi calibration), Ext.QC (external QC-coil calibration).
In case of any problems with instrument connection, battery failure, or accidental instrument Turn
O a message "No Communication, trying" will be displayed in numeric values window. e message is highlighted by red background to alert operator as fast as possible. e EM61MK2win will
try to re-establish communication and after the problem is corrected the program will start communication automaticaly. Alternatively, the Monitor mode can be stopped by clicking on the Exit
button or executing this function by a keyboard. e Monitoring window with alerting message
is shown below.
Description of Monitoring Mode Options
File (create data le)
e log data le can be created in any directory. e name of the le is given by the eld
computer clock and it consists of month (2 digits), day (2 digits), hour (2 digits), and one
alphabetic character A, B, C, etc. (If all letters during one hour are used use the Overwrite
option). e Create Data File dialog is presented below.
e le name can be specied in the Create Data File dialog using the Windows standard
interface procedure. e EM61MK2win data les cannot be appended.
Each data le in the eld computer (raw data le) has an extension name R61 and it is
created in the directory specied in Create Data File dialog. e R61 les are created in the
24
Geonics Limited
Null
instrument binary format. ese les can be also converted to ASCII format and processed
in the Geonics DAT61MK2 program.
After the le is specied click on the button OK or press ENTER key to accept and create
data le. e program will switch Monitor mode to Stand By mode and the data le will
be displayed on the screen.
To cancel selection and return to Monitoring mode tap the Cancel (X) button or press
Esc key.
To perform a null of the instrument click on the button labeled Null or execute this option
using the keyboard (navigate to the button by pressing TAB key and when highlighted
press ENTER key, or use shortcut - press key U). At this moment the computer takes 50
readings and calculates the instrument oset.
EM61MK2win Data Logging System (DAS70-M2)
25
e calculated oset is applied to all the readings that follow this operation. If needed, this
procedure can be repeated several times until satisfactory results are obtained. However,
there is no associated “Undo” function. If original values (without calculated osets) of
the EM61-MK2 readings are needed, exit the EM61MK2 program and run it again. e
EM61-MK2 instrument does not have to be turned OFF. Null operation can be also performed later during data collection.
Int.QC (Internal QC Coil Calibration)
e Internal QC coil calibration is described in detail in the EM61-MK2 Operating Instructions.
To start the calibration using Internal QC coil click on the key labeled Int. QC or execute
this option by keyboard (navigate to the button by pressing TAB key and when highlighted
press ENTER key, or use shortcut - press key I). is action begins by performing the
automatic nulling which in this case is only in eect for the duration of the calibration
process. It will not aect EM61-MK2 readings. After the nulling is completed (less than
5 seconds) the Internal QC coil Calibration window is displayed. All command buttons
in the bottom right corner of logging windows are disabled during Calibration procedure.
26
When Calibration window is displayed please use keyboard keys (O or C) rather than a mouse
to execute buttons labeled OK and Cancel. Clicking on buttons by mouse will work, however
several mouse button clicks may be required.
After OK button is clicked (or shortcut key O is pressed) the window displays timer with
elapsed seconds and EM61-MK2 readings. is display lasts for 60 seconds. During this
time please follow the instructions provided on the EM61MK2win display.
Geonics Limited
Internal QC coil calibration can end in the following four ways:
e process can be stopped by an operator at any time by clicking on the Cancel
button or pressing shortcut key C.
e calibration may end by time out (60 seconds) if QC coil button was not pressed
or this action did not activate QC coil. e following screen will then be displayed:
After pressing any key the program returns to Monitoring window. To repeat Internal QC coil calibration please click on the Int. QC button again.
e logging program will determine if the instrument has passed test. If after activating the QC coil, the program determines that the reading is inside the standard
values within +/- 10% tolerances then a message “CALIBRATION OK” will be
displayed (see below).
Otherwise, if the program determines that the reading is outside of the test values
range the corresponding message “CALIBRATION FAILED” is displayed.
EM61MK2win Data Logging System (DAS70-M2)
27
Regardless of the test result (OK or FAILED) the readings will be saved on screen. e program returns to Monitoring window after any button is tapped or ENTER key is pressed.
en the Internal QC coil calibration process can be repeated.
Ext. QC (External QC Coil Calibration)
e External QC coil calibration is described in detail in the EM61-MK2 Operating
Instructions.
To start the calibration using the External QC coil tap the button labeled Ext. QC or execute
this option by keyboard (navigate to the button by pressing TAB key and when highlighted
press ENTER key, or use shortcut - press key E). is action begins by performing the
automatic nulling which in this case is only in eect for the duration of the calibration
process. It will not aect EM61-MK2 readings. After the nulling is completed (less than 5
seconds) the External QC coil Calibration window is displayed (Figure below).
28
Geonics Limited
When Calibration window is displayed please use keyboard keys (O or C) rather than a mouse
to execute buttons labeled OK and Cancel. Clicking on buttons by mouse will work, however
several mouse button clicks may be required.
After OK button is tapped (or shortcut key O is pressed) the window displays timer with
elapsed seconds and EM61-MK2 readings. is display lasts for 60 seconds. During this
time please follow the instructions provided on the computer display screen.
External QC coil calibration can end in the following four ways:
e process can be stopped by an operator at any time by tapping Cancel button
or pressing shortcut key C.
e calibration may end by time out (60 seconds) if the External QC coil button
was not pressed or this action did not activate QC coil. e following window will
then be displayed (see below).
After pressing any key the program returns to Monitoring window. To repeat External QC coil calibration please tap the Ext. QC. button again.
e logger will determine if the instrument has passed test. If after activating the
QC coil, the program determines that the reading is inside the standard values
within +/- 10% tolerances then a message “CALIBRATION OK” will be displayed.
EM61MK2win Data Logging System (DAS70-M2)
29
Otherwise, if the program determines that the reading is outside of the test values
range the corresponding message “CALIBRATION FAILED” is displayed.
Regardless of the test result (OK or FAILED) the readings will be saved on screen. e program returns to Monitoring window after any button is clicked or ENTER key is pressed.
en the External QC coil calibration process can be repeated.
Scale (New Scale for Prole Plot)
Selecting New Scale option allows the operator to enter new scale parameters for the prole
plot. is option is available by clicking on the command button labeled Scale Rdgs. or
directly from keyboard by using shortcut key <R>.
e Readings Scale dialog will be displayed. It contains two text boxes for Minimum and
Maximum values of a new scale for prole plot. Current settings (minimum and maximum
30
Geonics Limited
values) for the scale are displayed in the bottom right and top right corners of the plot area.
After minimum and maximum values are specied click on the button OK or press ENTER
key to accept to accept new values and the prole plot area will be redrawn.
To ignore an entry and return to Stand By mode click on the Cancel (X) button or press
Esc key, the dialog window will disappear and measurements can be continued.
Exit (exit Monitor mode)
After clicking on the Exit button (or executing this option using keyboard) the program
immediately returns to Main Screen.
6.2 Stand By Mode
After data le is created in Monitoring mode the program automatically switches logging window
to Stand By mode. is windows is similar to Monitor mode however it contains survey parameters
(le name, survey line name, etc.) and dierent set of options represented by buttons displayed at
the bottom of the screen. e logging window in Stand By mode in initial state (before any data is
recorded) is presented below.
e main portion of the screen is occupied by the plot area which displays proles (see Figure
below which shows Stand By mode after some data were collected). Readings in the plot area can
be displayed in compressed amplitude scale, which corresponds to Square Root of the amplitude.
In compressed amplitude scale 0 to 100 corresponds to 0 to 10,000 mV, and compressed range 0
to 200 corresponds to 0 to 40,000 mV. Plotting data in compressed amplitude allows you to show
EM61MK2win Data Logging System (DAS70-M2)
31
details in the low range of amplitude, as well as relatively good resolution in the high range of data
on the small screen. Please note, that data displayed in the numeric form are always given in the standard, linear scale. e amplitude scale is divided by four or ve grey grid lines. In the
case where the amplitude scale starts with a negative value, then the grid line corresponding to zero
is always plotted as a thicker line.
32
Readings for Channels 1, 2, 3, 4 or T are shown in numeric form in a window below the plot area.
Values of readings are labeled: Ch1, Ch2, Ch3, Ch4 or ChT. e program checks instrument mode
continuously, therefore label Ch4 indicates mode 4 while label ChT corresponds to instrument
mode D. EM61-MK2 readings (in numeric form) are updated in Stand By mode approximately
10 times per second. Proles in the plot area are updated only during Logging mode when data are
actually saved in the data le.
A section of the window at the top of the screen contains one line of text displaying survey and
instrument related parameters. ese are (from top left): logging mode Stand By or Logging, survey mode (Auto, Wheel or Manual), instrument mode 4 or D (labeled Mode:)), and EM61-MK2
battery level in Volts (labeled Battery:). When ducial marker is active a label Marker is displayed
in this line.
File and Survey Line parameters, and GPS information are displayed in three framed boxes at the
bottom of the Stand By window. File Parameters frame displays: name of the data le, total number
of survey lines, total number of stations collected in the le, and total time of surveying. e Line
Parameters frame contains: name of current survey line, current station, number of stations in the
current survey line, and time of surveying current line. In the Stand By mode the station number
will not change. Stations in the EM61MK2win program are incremented by station increment value
after each reading is written to the data le.
Geonics Limited
e third frame is dedicated to showing the GPS status. A number following the label Position
provides the number of GPS positions saved in the current data le. In Stand By mode the number
of positions is not incremented (it is updated only when GPS position is written to the le). Next
two lines in this frame display parameters describing quality of GPS positions. A label DGPS (Differential Global Positioning System) indicates that GPS readings are dierentially corrected in real
time, while label AGPS (Autonomous Global Positioning System) indicates lack of dierential correction. On the right side of DGPS or AGPS label two green and white circles alternate. ese two
circles should alternate colours with the frequency of GPS update rate (usually 1 second intervals).
If the circles do not change colours for long periods of time it means that the GPS system is not
working or that it is not connected to the eld computer. e next label PDOP with a value varying
between 0 and 99.9 represents an index called Position Dilution of Precision (PDOP). is value
is available only when messages both, GGA and GSA are received from GPS. e last label Sat and
following number shows the number of currently tracked satellites. Two bottom lines of the frame
give current Latitude and Longitude of the instrument position (datum WGS1984). Position is
updated with frequency of GPS receiver stream.
In case of any problems with instrument connection, battery failure, or accidental instrument Turn
O a message "No Communication, trying" will be displayed in numeric values window. e message is highlighted by red background to alert operator as fast as possible. e EM61MK2win will
try to re-establish communication and after the problem is corrected the program will start communication automatically. Alternatively, click on the Exit button or execute this command from a
keyboard. e program window in Stand By mode and the alert message is shown below.
Several available eld options (in form of command buttons) are listed at the bottom right corner
of the Stand By mode window. ey will be described in the following section 6.4.
EM61MK2win Data Logging System (DAS70-M2)
33
6.3 Logging Mode
e Logging mode is enabled by clicking on the button labeled GO (log data) or pressing the
shortcut key G (or the <ENTER> key if button GO is highlighted) in Stand By mode. After this
button is pressed the list of buttons at bottom of the screen will be replaced by one PAUSE button,
label Stand By will be replaced by label Logging (at the top of the display) and data will be logged
in the mode corresponding to the selected EM61-MK2 mode in the Survey Setup menu. All labels
and parameters (with the exception of buttons representing Stan By mode options) are the same as
in Stand By mode and they are described in the preceding section 6.2. e screen in Logging mode
is presented below.
34
After the screen changes to Logging mode the current station (labeled Station:) is updated according
to the station interval. Similarly, if GPS Input was enabled, total number of GPS positions (label
Positions) in the data le is incremented every time (usually once a second) GPS position is written
to the le. Prole curves (for channels selected in Display Options window) are updated after each
reading is written to the data le.
In case of any problems with instrument connection, battery failure, or accidental instrument Turn
O a message "No Communication, trying" will be displayed in numeric values window. e message is highlighted by red background to alert operator as fast as possible. e EM61MK2win will
stop collecting data and will try to re-establish communication. After the problem is corrected the
program will start communication automaticaly. Alternatively, click on the Pause button to return
to Stand By mode. e program window in Stand By mode and the alert message is shown below.
Geonics Limited
ere is only one option available in the Logging mode - pause logging. After a Pause button is
clicked or Pause key (any key or other Pause Key selected in Logger Setup window) is pressed the
recording is stopped and the Logging mode returns to the Stand By mode. In Stand By mode the
EM61-MK2 data will be displayed with the update rate approximately 10 readings per second,
however data will not be saved in the log le and prole curves will be not updated.
Exiting from the logging session and access to various eld options are available only from the logging window in Stand By mode.
6.4 Stand By Mode Field Options
Several eld options are available while the Logging window is in the Stand By mode. Options can
be accessed directly from command buttons. Command buttons can be used by clicking on the
desired button, or from the keyboard by pressing one of the shortcut keys (underlined characters on
button labels) or by navigating using <TAB> key (sets button as a default button - default button
is highlighted) and pressing <ENTER> key.
GO (log data)(start data logging)
Click on the GO button, or while using the keyboard press shortcut key <G> or if the
button is a default button (highlighted) press <ENTER>. e logging window in Stand By
mode will change to Logging mode and logging data starts immediately.
EM61MK2win Data Logging System (DAS70-M2)
35
New Line (New Survey Line)
e New Survey Line dialog is displayed (see Figure below). Selecting this option allows
the operator to enter a new survey line number (name) and associated line parameters (Line
Increment, Line Sequence, Direction, Start Station, and Station Increment). e new line
number and associated parameters are prompted by the program based on parameters
specied in the Survey Setup dialog.
36
At the top of the dialog the last survey line name and the last logged station are displayed.
Default name for the new line is given based on the Line Increment parameter. e default
Start Station, direction of the Station Increment, and Direction are determined based on
Sequence selection. All these parameters can be overwritten by the user as described in the
Survey Setup menu description (chapter 2).
After all the parameters in the New Line dialog are updated click on the button OK or
press ENTER key to accept the displayed settings. e program will return to the Logging
window in Stand By mode. Survey line name (Curr. Line) name and current station (Sta-tion) value will be updated and prole curves plot for former survey line will disappear.
To return to Stand By mode and current survey line settings (state before this dialog was
selected) click on the Cancel (X) button or press Esc key.
New Stn.(New Station)
Selecting this option allows the operator to enter a new station number (within the same
survey line). e New Station dialog is displayed (see Figure below on the left).
Current survey line, Start and Current station are displayed at the top of the dialog. e
New Station can be entered in the provided edit box labeled New Station.
Geonics Limited
Click on the button OK or press ENTER key to accept the new value. e program will
return to the Logging window in Stand By mode. Current station (Station:) value will be
updated and after data logging is activated the prole curves will have a small gap (two
pixels) showing the new station entry (see Figure above on the right).
To return to Stand By mode and current survey line settings (state before this dialog was
selected) click on the Cancel (X) button or press Esc key, the dialog window will disappear
and measurements can be continued.
Comment (Field Comment)
Selecting this option allows the operator to enter a comment at any point of the survey.
A maximum of 11 characters can be entered as a comment. e Enter Comment dialog
is displayed.
Click on the button OK or press ENTER key to accept the comment entered in a text box
of the dialog. e text of the comment is saved in the le with a corresponding time stamp
and the program will return to the Logging window in Stand By mode.
To ignore an entry and return to Stand By mode click on the Cancel (X) button or press
Esc key, the dialog window will disappear and measurements can be continued.
EM61MK2win Data Logging System (DAS70-M2)
37
Scale Rdgs (New Scale for Prole Plot)
Selecting New Scale option allows the operator to enter new scale parameters for the prole
plot.
e Enter Scale dialog will be displayed (see Figure below on the left). It contains two text
boxes for Minimum and Maximum values of a new scale for prole plot. Current settings
(minimum and maximum values) for the scale are displayed in the bottom right and top
right corners of the plot area.
38
After minimum and maximum values are specied click on the button OK or press ENTER key to accept new values and the prole plot area will be redrawn (see Figure above
on the right).
To ignore an entry and return to Stand By mode tap the Cancel (X) button or press Esc
key, the dialog window will disappear and measurements can be continued.
Proles (new options for prole plot display)
After the Proles button was clicked or executed from the keyboard the Display Options
dialog window appears on the screen. is dialog allows you to enable and disable the
display of each channel prole, specify color and thickness of proles, and select linear or
compressed amplitude for proles. e dialog is presented below.
Please refer to chapter 5 where this dialog and its parameters are described in detail.
Click on the button OK or press ENTER key to accept updated display parameters. e
dialog will disappear and prole plot area will be redrawn in Stand By mode.
To return to Stand By mode and current display options (state before this dialog was se-
lected) click on the Cancel (X) button or press Esc key, the dialog window will disappear.
Geonics Limited
Null
During data collection (in Stand By mode) a conrmation message will be displayed (there
is no such message if Nulling is performed in Monitoring mode) before actual operation
is applied (see below Figure on the left).
EM61MK2win Data Logging System (DAS70-M2)
39
After above message is conrmed the program collects 50 readings and calculates the instrument oset.
e calculated oset is applied to all the readings that follow this operation (see Figure
above on the right). If needed, this procedure can be repeated several times until satisfactory
results are obtained. However, there is no associated “Undo” function. If original values
(without calculated osets) of the EM61-MK2 readings are needed, exit the EM61MK2
program and run it again. e EM61-MK2 instrument does not have to be turned OFF.
Int. QC (Internal QC coil calibration)
e Internal QC coil calibration is described in detail in the EM61-MK2 Operating Instructions.
e Internal QC coil procedure in Stand By mode is identical as in Monitoring mode.
e Please refer to section 6.2 (Description of Monitoring Mode Options) for more screen
shots examples.
To start the calibration using Internal QC coil click on the key labeled Int. QC or execute
this option by keyboard (navigate to the button by pressing TAB key and when highlighted
press ENTER key, or use shortcut - press key I). is action begins by performing the auto-
matic nulling which in this case is only in eect for the duration of the calibration process.
It will not aect EM61-MK2 readings during later data collection. After the nulling is
completed (less than 5 seconds) the Internal QC coil Calibration window is displayed. All
command buttons in the bottom right corner of the logging window in Stand By mode
are disabled during Calibration procedure.
When Calibration window is displayed please use keyboard keys (O or C) rather than a mouse
to execute buttons labeled OK and Cancel. Clicking on buttons by mouse will work, however
several mouse button clicks may be required.
After OK button is clicked (or shortcut key O is pressed) the window displays timer with
elapsed seconds and EM61-MK2 readings. is display lasts for 60 seconds. During this
time please follow the instructions provided on the computer screen.
Internal QC coil calibration can end in the following four ways:
e process can be stopped by an operator at any time by clicking on the Cancel
button or pressing shortcut key C.
e calibration may end by time out (60 seconds) if QC coil button was not pressed
or this action did not activate QC coil.
40
After pressing any key the program returns to Monitoring window. To repeat Internal QC coil calibration please click on the Int. QC button again.
e logging program will determine if the instrument has passed test. If after activating the QC coil, the program determines that the reading is inside the standard
Geonics Limited
values within +/- 10% tolerances then a message “CALIBRATION OK” will be
displayed.
Otherwise, if the program determines that the reading is outside of the test values
range the corresponding message “CALIBRATION FAILED” is displayed (see
Figure below).
Regardless of the test result (OK or FAILED) the readings will be saved on screen. e program returns to Monitoring window after any button is tapped or ENTER key is pressed.
en the Internal QC coil calibration process can be repeated.
Ext. QC (External QC Coil Calibration)
e External QC coil calibration is described in detail in the EM61-MK2 Operating
Instructions.
e External QC coil procedure in Stand By mode is identical as in Monitoring mode.
e Please refer to section 6.2 (Description of Monitoring Mode Options) for more screen
shots examples.
To start the calibration using the External QC coil click on the button labeled Ext. QC
or execute this option by keyboard (navigate to the button by pressing TAB key and when
highlighted press ENTER key, or use shortcut - press key E). is action begins by performing the automatic nulling which in this case is only in eect for the duration of the
calibration process. It will not aect EM61-MK2 readings. After the nulling is completed
(less than 5 seconds) the External QC coil Calibration window is displayed (Figure below).
When External QC Coil Calibration window is displayed please use keyboard keys (O or C)
rather than a mouse to execute buttons labeled OK and Cancel. Clicking on buttons by mouse
will work, however several mouse button clicks may be required.
After OK button is tapped (or shortcut key O is pressed) the window displays timer with
elapsed seconds and EM61-MK2 readings. is display lasts for 60 seconds. During this
time please follow the instructions provided on the computer display.
EM61MK2win Data Logging System (DAS70-M2)
41
External QC coil calibration can end in the following four ways:
e process can be stopped by an operator at any time by tapping Cancel button
or pressing shortcut key C.
e calibration may end by time out (60 seconds) if the External QC coil button
was not pressed or this action did not activate QC coil.
After pressing any key the program returns to Monitoring window. To repeat External QC coil calibration please tap the Ext. QC. button again.
e logger will determine if the instrument has passed test. If after activating the
QC coil, the program determines that the reading is inside the standard values
within +/- 10% tolerances then a message “CALIBRATION OK” will be displayed.
Otherwise, if the program determines that the reading is outside of the test values
range the corresponding message “CALIBRATION FAILED” is displayed.
Regardless of the test result (OK or FAILED) the readings will be saved on screen. e program returns to Monitoring window after any button is clicked or ENTER key is pressed.
en the External QC coil calibration process can be repeated.
Exit(exit data logging)
During data collection (in Stand By mode) a conrmation message will be displayed (there
is no such message if Exit is performed in Monitoring mode) before program exits logging
window.
42
After above message is conrmed the program stops logging, closes data le and returns
to Main Screen.
Geonics Limited
7. View Data Files
is option allows you to view recorded data les. After the View Files button is clicked or tapped
(or executed from the keyboard) in the Main Screen the View Data Files window will be displayed.
To open the data le click or tap on the button labeled File or execute this option using the keyboard (navigate to the button by pressing TAB key and when highlighted press ENTER key, or use
shortcut - press key F). e Open Data File dialog is presented below.
EM61MK2win Data Logging System (DAS70-M2)
43
e le name can be selected in the Open Data File dialog using the Windows standard interface
procedure. After the le is selected (highlighted) tap the button OK or press ENTER key to accept
and display data le (to cancel selection tap Cancel (X) button or press Esc key). After the le is
opened readings taken at the beginning of the rst survey line in the le are displayed, as shown below.
44
e main portion of the screen is occupied by prole plot of currently displayed survey line. Readings in the plot area can be displayed in linear or compressed amplitude scale. Amplitude display
range (minimum and maximum value) is given by labels located at the right side of the plot area.
Range of scale display can be changed by tapping the button labeled Scale Rdgs. e plot area is
divided by four or ve grey grid lines. In cases where the amplitude scale starts with a negative value,
then the grid line corresponding to zero is always plotted as a thicker line. Proles for each channel
are plotted with colours and line thickness as selected in the Plot Options (see section Chapter 5).
A vertical bar (grey vertical line) indicates currently displayed station. e station is displayed in the
window labeled Station Parameters (see description in the next paragraph) next to label Curr.Stn:
and readings at this station for channels 1, 2, 3, and 4 (or T) are shown in numeric form below the
plot area. e EM61-MK2 readings (in numeric form) are updated every time vertical bar is moved.
ere are three windows describing le, survey line, and station parameters. e left window labeled
File Parameters lists le name, number of survey lines in the le, total number of stations, and
total number of GPS positions if GPS positioning was used during the survey. e center window
Geonics Limited
labeled Line Parameters contains parameters associated with currently displayed survey line: line
name, number of stations in the current line, minimum and maximum stations. e right window
labeled Station Parameters displays currently indicated station parameters: station number, reading
mode, marker presence, and station increment within line.
To change the currently displayed survey line click or tap on the button labeled Select Line. A dialog
labeled Select Survey Line will be displayed.
Tapping on the down arrow next to the text box labeled Line opens a drop-down box showing all
available survey lines. When the keyboard is being used you can activate the text box by pressing the
TAB key until the drop down box is highlighted and then by using Up or Down arrow keys select
one of available lines (as shown above). After the survey line is selected and the button labeled OK
is clicked on the dialog disappears and selected survey line is plotted. Please see next Figure where
Survey Line 10 is displayed.
To change range of displayed values click on the Scale Rdgs button. A dialog labeled Reading Scale
will be displayed. Specify new Minimum and Maximum values and then press OK. e plot area
will be re-drawn with new range values.
EM61MK2win Data Logging System (DAS70-M2)
45
To shift the range of the displayed stations and to move the vertical bar indicating the currently
displayed station, use the buttons located along the bottom of the View Files window. ese buttons
can be activated by tapping or by pressing the keyboard keys. ey are divided in two groups: four
buttons labeled Shift Prole Image are used to move image while four buttons labeled Shift Station
Bar move grey vertical bar selecting current station and displaying its numeric values.
Shift Prole Image buttons:
button labeled |<< moves display to the minimum station of the currently displayed survey line,
button labeled >>| moves display to the maximum station of the survey line,
button labeled < moves display about 40% of the displayed range to the left,
button labeled > moves display about 40% of the displayed range to the right.
When using Shift Station Bar buttons:
button labeled << moves the bar 50 stations to the left,
button labeled >> moves the bar 50 stations to the right,
button labeled < moves the bar one station to the left,
button labeled > moves the bar one station to the right.
To exit the View Data Files window tap or click on the button labeled Exit. e program will return
to Main Screen.
46
Geonics Limited
Appendix A
A.1 Description of Data File in EM61MK2win (R61) Format
Each record contains 24 characters, including line feed at the end of each record.
Header of the le (contains six records starting with characters E, H, O, O, O, and O)
12345678910 11 12 13 14 15 16 17 18 19 20
M 61
E
HFile NameTime Increment [s] (F7.3)10
O
O
O
O
EM61MK2 - identication of program le
W101 - version number (1.00, W - indicates le created by
Windows CE program)
Survey Type - GPS (if GPS Input Enabled) or GRD (grid)
UT - unit type (0 = meters, 1 = feet)
IT - instrument type
IM IP - not used (=1 xed)
File Name - le name, maximum 8 characters
Time Increment - time increment (Auto Mode) in seconds
Oset - oset for indicated channels in mV
IC QC coil value 10 - Line Feed character
QC coil calibration (=N not performed, <>N otherwise)
value of QC coil calibration for indicated channels in mV
21
22
23 24
IP
10
10
10
10
Header at the start of survey line (eight records starting with L, B, A, Z, O, O, O, O)
12345678910 11 12 13 14 15 16 17 18 19 20
Line Name (8 characters)
L
BStart Station (Format F11.2)10
A
Dir
Z DDMMYYYY HH:MM:SS10
O
O
O
O
Offset for Ch1 (F9.2)
Offset for Ch2 (F9.2)
Offset for Ch3 (F9.2)
Offset for Ch4 (F9.2)
Station Increment (Format F11.2)10
EM61MK2win Data Logging System (DAS70-M2)
Former Offset for Ch1 (F9.2)
Former Offset for Ch2 (F9.2)
Former Offset for Ch3 (F9.2)
Former Offset for Ch4/T (F9.2)
21
22
23 24
10
10
10
10
10
47
Line Name - Line Name, maximum 8 characters
Start Station - Start Station for the Line, format F11.2
Dir - Direction of the Line (E, W, N, or S)
Station Inc. - Station Increment, format F11.3
Date - Date when Line was created, format DD-MM-YYYY
Time - Real Time when Line was created, format HH:MM:SS
Oset - Oset for indicated channels in mV at the start of Line
Former Oset - Former oset for indicated channels in mV
10 - Line Feed character
Timer Relation
12345678910 11 12 13 14 15 16 17 18 19 20 21
Computer Time (Format HH:MM:SS.sss)
*
Indicates relation between computer clock and the program timer. is record links timer
in milliseconds and computer time (local time) in format HH:MM:SS.sss. is record is
written to the le each time a new new line is entered.
Reading
12345678910 11 12 13 14 15 16 17 18 19 20 21
I
Gn
1h 1l
I - indicator T, D, E, F, M, N, P, or Q. Each record representing reading
starts with one of the following character, which indicates type of
reading:
T - Standard, Mode 4, channels 1, 2, 3, 4 D - Standard, Mode D, channels 1, 2, 3, TE - Hand Held, Mode 4, channels 1, 2, 3, 4F - Hand Held, Mode D, channels 1, 2, 3, TM - Standard, Mode 4, channels 1, 2, 3, 4 +MarkerN - Standard, Mode D, channels 1, 2, 3, T +MarkerP - Hand Held, Mode 4, channels 1, 2, 3, 4 +MarkerQ - Hand Held, Mode D, channels 1, 2, 3, T +Marker
Gn - one character parameter (Hex format), contains Gain,
see table of ranges at the end of this section.
1h - higher byte of the 2’s complement Hex number of Channel 1
1l - lower byte of Channel 1
2h - higher byte of the 2’s complement Hex number of Channel 2
2l - lower byte of Channel 2
3h - higher byte of the 2’s complement Hex number of Channel 3
3l - lower byte of Channel 3
4h - higher byte of the 2’s complement Hex number of Channel 4
4l - lower byte of Channel 4
5h - higher byte of the 2’s complement Hex number of TX current
5l - lower byte of TX current
6 - fraction of current (5h 5l), Hex number
Time Stamp - time in ms from the Windows start (resets every 49.7 days)
10 - Line Feed character
3h
2l2h
3l
4h
5h 5l6
4l
Time Stamp in milliseconds
Time Stamp in ms (10 digits)
22
22
23
23 24
24
10
10
48
Geonics Limited
Comment
12345678910 11 12 13 14 15 16 17 18 19 20 21
Comment (maximum 11 characters)
C
Time Stamp in ms (10 digits)
New Station
12345678910 11 12 13 14 15 16 17 18 19 20 21
S
New Station (Format 11.2)
Time Stamp in ms (10 digits)
Nulling
12345678910 11 12 13 14 15 16 17 18 19 20 21
O
O
O
O
Offset for Ch1 (F9.2)
Offset for Ch2 (F9.2)
Offset for Ch3 (F9.2)
Offset for Ch4 (F9.2)
Former Offset for Ch1 (F9.2)
Former Offset for Ch2 (F9.2)
Former Offset for Ch3 (F9.2)
Former Offset for Ch4/T (F9.2)
QC Coil Calibration
12345678910 11 12 13 14 15 16 17 18 19 20 21
C
I
I
C
Offset for Ch1 (F9.2)
Offset for Ch2 (F9.2)
23 24
22
10
23 24
22
10
22 23 24
10
10
10
10
22 23 24
10
10
C
I
I
C
Offset for Ch3 (F9.2)
Offset for Ch4 (F9.2)
10
10
IC - II indicates Inrenal QC coil calibration,IE indicates External QC coil calibration.
GPS Data Message Records
Each GPS record (GGA Message) is broken in to several 22 characters strings and placed in the
EM61MK2 data le which contains 24 characters records, including one character indicator and
line feed at the end of each record. e GPS sequence starts at the line which contains character
@ as the rst character, then records that contain continuation of the same message start with
character #. e GPS sequence ends with a line starting with the character !. e last line contains
logger time stamp in milliseconds for given GPS reading. A sample of the GPS message written in
EM61MK2 format is given below.
12345678910 11 12 13 14 15 16 17 18 19 20 21
@
#
#
#
m
+
$
G P
m
mm
q
q
,
n
x
xx.x,M,s
x
A
GG
,
m
pp.p,saa
,
h
,
,ddd mm.m
s
m mss.ss,
h
ss,
aaa
a
m m m
.
aa
ddm
m
aa
c*
,s
c
22
m
u,
CR
LF
23
24
.
10
,
10
,
10
10
!
e GPS sequence may contain 4 to 6 records. Component of the GGA message may dier in
length, however they are placed in the same number of columns. Refer to Appendix B (section B.2)
for denition of each component of GGA data message.
EM61MK2win Data Logging System (DAS70-M2)
Time Stamp in milliseconds
10
49
A.2 Conversion Factors
EM61MK2 has four channels. Channels 1, 2, and 3 are common for Mode 4 and Mode D. Channel 4 in Mode D is named Channel T (it corresponds to Top coil).
e instrument response is converted to output voltage in mV for each sampling channel as given
below.
Channel 1 to 4 - converted data
DATA1 (to 4) - instrument output for each channel as recorded in logger
RANGE - range is controlled by the EM61MK2, it can be 1, 10,
100
Standard Unit - Mode 4 (One Sensor 1 x 0.5 m or 1 x 1 m)
Channel 1=(DATA1 x 4.8333 x 2)/RANGE
Channel 2=(DATA2 x 4.8333 x 2)/RANGE
Channel 3=(DATA3 x 4.8333 x 2)/RANGE
Channel 4=(DATA4 x 4.8333 x 2)/RANGE
Standard Unit - Mode D (Two Sensors 1 x 0.5 m or 1 x 1 m, Top and Bottom coils))
Channel 1=(DATA1 x 4.8333 x 2)/RANGE
Channel 2=(DATA2 x 4.8333 x 2)/RANGE
Channel 3=(DATA3 x 4.8333 x 2)/RANGE
Channel T=(DATA4 x 4.8333 x 4)/RANGE
if coil is 1 x 0.5 m Channel T is further multiplied by a factor 1.114.
Hand Held Unit - Mode 4 (One Sensor)
Channel 1 = 0.902500 x (DATA1 x 4.8333)/RANGE
Channel 2 = 1.363000 x (DATA2 x 4.8333)/RANGE
Channel 3 = 2.026795 x (DATA3 x 4.8333)/RANGE
Channel 4 = 3.018856 x (DATA4 x 4.8333)/RANGE
Hand Held Unit - Mode D (Two Sensors, Top and Bottom coils))
Channel 1 = 0.9025 x (DATA1 x 4.8333)/RANGE
Channel 2 = 1.3630 x (DATA2 x 4.8333)/RANGE
Channel 3 = 2.0430 x (DATA3 x 4.8333)/RANGE
Channel T = 12.152 x (DATA4 x 4.8333)/RANGE
Further each channel is normalized by current following formula:
Standard Unit
Channel = Channel x 3000/Current
Hand Held Unit
Channel = Channel x 1800/Current
50
where, current is a value represented by 5h, 5l, and 6 in EM61MK2 data le (see section B.1)
Geonics Limited
Table of Ranges Determined by the EM61-MK2 Microprocessor
A.3 Example of Data File in EM61MK2win (R61) Format
e Em61MK2 data le records are written in binary format, therefore the le may have dierent
shape when displayed or printed, depending on particular video or pronter settings.
EM61MK2 W100GPS000 1
H 042320A 0.200
O 0.00E 0.00
O 0.00E 0.00
O 0.00E 0.00
O 0.00E 0.00
L0
B 0.00
AS 1.000
Z23042005 20:02:51
O 0.00 0.00
O 0.00 0.00
O 0.00 0.00
O 0.00 0.00
*20:02:51.000 4179829
@$GPGGA,020412.00,4336.
#59410,N,07936.64856,W,
#2,6,2,141.56,M,-35,M,5
#,119*55
! 4287423
@$GPGSA,A,3,30,,,25,,20
#,,14,,06,,01,02.8,02.1
#,02.0*0D
! 4287441
Eÿ Ø `/^ ”N 4287673
Eÿ †/“ ”Q 4287879
C comment text 4287983
Eÿ ; G/yï ”O 4288089
Eÿ § h/¼þ ”W 4288279
@$GPGGA,020413.00,4336.
#59408,N,07936.64857,W,
#2,6,2,141.51,M,-35,M,6
#,119*58
! 4288338
Eÿ ¶ 5/ # ”S 4288488
@$GPGSA,A,3,30,,,25,,20
#,,14,,06,,01,02.8,02.1
#,02.0*0D
! 4288419
Eÿ þ º/j ”O 4288677
Eÿ L t/k ”Q 4288885
@$GPGGA,020414.00,4336.
#59411,N,07936.64846,W,
#2,6,2,141.68,M,-35,M,6
#,119*5D
! 4289337
@$GPGGA,020416.00,4336.
#59420,N,07936.64848,W,
#2,6,2,142.11,M,-35,M,4
#,119*5C
! 4291335
E
tsÝ3 • 4291502
@$GPGSA,A,3,30,,,25,,20
#,,14,,06,,01,02.8,02.1
#,02.0*0D
! 4291414
Eÿ*‚ b Œ ”Y 4291691
C building 4291793
E
# *â ”V 4291900
E¸/D ”C 4292299
52
Geonics Limited
Appendix B
B.1 Using the EM61MK2win with a GPS Receiver
e EM61MK2win program accepts input from GPS receiver that stream NMEA-0183 compatible
data through their output port. e program uses two NMEA messages: GGA and GSA. e entire
GGA message is used later by the DAT61MK2 program, while the GSA message is used only to
display PDOP index on the logger screen.
e GPS system means (control device, receiver panel, or manufacturer software) must be used to
set GPS receiver communication parameters, to specify frequency of GPS output, and number and
type of NMEA messages sent by the GPS system output port. Any GPS system can send various
NMEA messages. It is important to select only two messages (GGA and GSA) that are actually used by EM61MK2. e program will accept any GPS string sent by the GPS receiver, however it
uses time to process GPS data that is not being used. erefore, selecting a larger number of NMEA
messages for GPS output will result in slower data acquisition of EM61MK2win. Normally, the
EM61MK2win running in a Windows XP based computer uses less than 100 ms to process and
record GPS data from two NMEA messages, GGA and GSA.
Only message GGA is necessary to position EM61-MK2 data. If message GSA is not available in a
particular system, the EM61MK2win will function and record position data based on GGA message.
Lack of GSA message will result in PDOP index displayed as Not Available (N/A) on the logger
display. Using message GGA alone will also result in slightly faster operation of the program. e
speed can be further improved by setting higher Baud Rate (if it is supported by the employed GPS
receiver) in Set GPS Port menu.
e EM61MK2win dedicates portion of the display and shows GPS status in a frame box. A number
following the label Position provides the number of GPS positions saved in the current data le.
Next two lines in this frame display parameters describing quality of GPS positions. A label DGPS
(Dierential Global Positioning System) indicates that GPS readings are dierentially corrected in
real time, while label AGPS (Autonomous Global Positioning System) indicates lack of dierential
correction. On the right side of DGPS or AGPS label two green and white circles alternate. ese
two circles should alternate colours with the frequency of GPS update rate (usually 1 second intervals). If the circles do not change colours for long periods of time it means that the GPS system is
not working or that it is not connected to the eld computer.
e next label PDOP with a value varying between 0 and 99.9 represents an index called Position
Dilution of Precision (PDOP). is value is available only when messages both, GGA and GSA are
received from GPS. is index measures the strength of satellite coverage for a given area. PDOP
is aected by the number of satellites visible and their relative positions in the sky. e smaller the
number of PDOP the stronger the satellite coverage is. When there are more than 5 satellites widely
spaced visible, the PDOP is 4 or less. However, when there are less satellites visible, or they are
unevenly spaced in the sky, PDOP values can be 6 or higher. In most cases, the PDOP in an open
sky is less than 3, and most accuracies given for many GPS systems are given for this norm. Refer to
GPS documentation and literature for more information related to error sources of GPS positioning.
e last label Sat and following number shows number of currently tracked satellites.
Two bottom lines of the frame give current Latitude and Longitude of the instrument position
(datum WGS1984). Position is updated with frequency of GPS receiver stream.
EM61MK2win Data Logging System (DAS70-M2)
53
B.2 Description of GGA and GSA Data Messages
GGA Data Message
e GGA message contains the GPS position information and it is the most widely used NMEA
data message. is message takes the following form:
hhmmss.ss UTC time in hours, minutes, seconds of the GPS position
ddmm.mmmmm Latitude in degrees, minutes, and decimal minutes
s s=N or s=S, for North and South latitude
dddmm.mmmmm Longitude in degrees, minutes, and decimal minutes
s s=E or s=W, for East and West longitude
n Quality indicator, 0 = no position, 1 = raw, no dierentially corrected
position, 2 = dierentially corrected position, 9 = position computed
using almanac information
qq Number of satellites used in position computation
pp.p HDOP = 0.0 to 99.9
saaaaa.aa Antenna altitude
u Altitude units, M=meters
+xxxx.x Geoidal separation (requires geoidal height option)
M Geoidal separation units, M = meters
sss Age of dierential corrections in seconds
aaaa Base station identication
*cc Checksum
<CR><LF> Carriage return and Line feed
GSA Data Message
e GSA message contains active satellites and PDOP value. e GSA message is given in the following form:
c1 Mode, M = manual, A = automatic
d1 Mode, 2 = 2D, 3 = 3D
d2-d13 Satellites used in position computation (range 0 to 32)
f1 PDOP (range 0 to 99.9)
f2 HDOP (range 0 to 99.9)
f3 VDOP (range 0 to 99.9)
*cc Checksum
<CR><LF> Carriage return and Line Feed
54
Geonics Limited
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.