Geokon 8600-2, 8600-1 Instruction Manual

Instruction Manual
Model 8600-2
Micro-6000 Multi-Channel Datalogger
No part of this instruction manual may be reproduced, by any means, without the written consent of Geokon®.
The information contained herein is believed to be accurate and reliable. However, Geokon
omissions or misinterpretation. The information herein is subject to change without notification.
Copyright © 2016-2018 by Geokon
(Doc Rev D, 12/17/2018)
®
®
assumes no responsibility for errors,
Warranty Statement
Geokon warrants its products to be free of defects in materials and workmanship, under normal use and service for a period of 13 months from date of purchase. If the unit should malfunction, it must be returned to the factory for evaluation, freight prepaid. Upon examination by Geokon, if the unit is found to be defective, it will be repaired or replaced at no charge. However, the WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion or current, heat, moisture or vibration, improper specification, misapplication, misuse or other operating conditions outside of Geokon's control. Components which wear or which are damaged by misuse are not warranted. This includes fuses and batteries.
Geokon manufactures scientific instruments whose misuse is potentially dangerous. The instruments are intended to be installed and used only by qualified personnel. There are no warranties except as stated herein. There are no other warranties, expressed or implied, including but not limited to the implied warranties of merchantability and of fitness for a particular purpose. Geokon is not responsible for any damages or losses caused to other equipment, whether direct, indirect, incidental, special or consequential which the purchaser may experience as a result of the installation or use of the product. The buyer's sole remedy for any breach of this agreement by Geokon or any breach of any warranty by Geokon shall not exceed the purchase price paid by the purchaser to Geokon for the unit or units, or equipment directly affected by such breach. Under no circumstances will Geokon reimburse the claimant for loss incurred in removing and/or reinstalling equipment.
Every precaution for accuracy has been taken in the preparation of manuals and/or software, however, Geokon neither assumes responsibility for any omissions or errors that may appear nor assumes liability for any damages or losses that result from the use of the products in accordance with the information contained in the manual or software.
TABLE of CONTENTS
1. OVERVIEW ............................................................................................................................................................ 1
2. SOFTWARE ............................................................................................................................................................ 1
3. HARDWARE ........................................................................................................................................................... 1
INSTALLATION ..................................................................................................................................................... 2
3.1
POWER ................................................................................................................................................................. 2
3.2
EARTH GROUND .................................................................................................................................................. 3
3.3
GAUGES ............................................................................................................................................................... 3
3.4
COMMUNICATIONS .............................................................................................................................................. 3
3.5
4. BATTERY MAINTENANCE ................................................................................................................................ 4
4.1
AC POWER .......................................................................................................................................................... 4
SOLAR POWER ..................................................................................................................................................... 4
4.2
EXTERNAL BATTERY ........................................................................................................................................... 4
4.3
BATTERY REPLACEMENT ..................................................................................................................................... 4
4.4
FUSES .................................................................................................................................................................. 4
4.5
5. TROUBLESHOOTING .......................................................................................................................................... 5
APPENDIX A. SPECIFICATIONS ........................................................................................................................... 7
A.1
MODEL 8600-2 DATALOGGER ............................................................................................................................ 7
CR6 MEASUREMENT AND CONTROL MODULE ................................................................................................... 7
A.2
MODEL 8032 MULTIPLEXER ............................................................................................................................... 7
A.3
APPENDIX B. SHIP LIST ......................................................................................................................................... 8
B.1
HARDWARE ......................................................................................................................................................... 8
APPENDIX C. DATA STORAGE ............................................................................................................................. 9
C.1
DATA STORAGE .................................................................................................................................................. 9
DATA STORAGE EXAMPLE .................................................................................................................................. 9
C.2
C.2.1 Overwrite Times when Data is Stored Once a Minute ............................................................................... 9
C.2.2 Overwrite Times when Data is Stored Once an Hour ................................................................................ 9
C.2.3 Overwrite Times when Data is Stored Once a Day .................................................................................... 9
APPENDIX D. SYSTEM WIRING ......................................................................................................................... 10
CR6 WIRING ..................................................................................................................................................... 10
D.1
CHARGER PORT WIR ING ................................................................................................................................... 10
D.2
UNREGULATED CHARGER WIR ING ................................................................................................................... 10
D.3
REGULATED CHARGER (DEFAULT) ................................................................................................................... 11
D.4
EXTERNAL POWER CABLE ................................................................................................................................ 11
D.5
FUSE DESIGNATIONS FOR POWER DISTRIBUTION BOARD ................................................................................. 11
D.6
APPENDIX E. LOGGERNET QUICK START GUIDE ...................................................................................... 12
E.1
OVERVIEW ........................................................................................................................................................ 12
GETTING STARTED ............................................................................................................................................ 12
E.2
E.2.1 Program Files ........................................................................................................................................... 12
E.2.2 Connection Setup ...................................................................................................................................... 12
E.2.3 Connecting to the Datalogger .................................................................................................................. 16
MODIFYING THE DATALOGGER PROGRAM ........................................................................................................ 17
E.3
E.3.1 Scan Interval ............................................................................................................................................. 17
E.3.2 Zero Readings and Gauge Factors ........................................................................................................... 17
E.3.3 Temperature Zero Readings and Thermal Factors .................................................................................. 18
E.3.4 Output Labels ........................................................................................................................................... 18
UPLOADING FILES ............................................................................................................................................. 19
E.4
E.4.1 Sending the Include file (.DLD) ................................................................................................................ 19
E.4.2 Sending the Main program file (.CR6) ..................................................................................................... 21
E.5 DATA HANDLING ...............................................................................................................................................21
E.5.1 Data Collection Configuration .................................................................................................................21
E.5.2 Live Monitoring ........................................................................................................................................22
E.5.3 Collecting Data .........................................................................................................................................23
APPENDIX F. EXCERPTS FROM THE 8032 MULTIPLEXER MANUAL .....................................................24
THEORY OF OPERATION .....................................................................................................................................24
F.1
MICRO-6000 MODE OF OPERATION .................................................................................................................26
F.2
WIRIN G ..............................................................................................................................................................28
F.3
F.3.1 Gauge Wiring ............................................................................................................................................28
F.3.2 Connector and Cable Wiring ....................................................................................................................30
F.3.3 Model 8032-27 and Load Cell Wiring ......................................................................................................31
TROUBLESHOOTING ...........................................................................................................................................32
F.4
8032 MULTIPLEXER SPECIFICATIONS ................................................................................................................32
F.5
DAISY-CHAIN OPERATION .................................................................................................................................33
F.6
MAXIM UM 8032-5 (TAN CABLE) CAB LE LENGTHS ............................................................................................35
F.7
APPENDIX G. CELLULAR MODEM COMMUNICATIONS ............................................................................36
G.1
SUPPLY .............................................................................................................................................................36
INSTALL ............................................................................................................................................................37
G.2
COMMUNICATIONS ............................................................................................................................................38
G.3
FIGURES
IGURE 1 - TYPICAL MICRO-6000 CONFIGURATION ....................................................................................................... 2
F
IGURE 2 - POWER DISTR IBUTION CIRCUIT BOARD ........................................................................................................ 2
F
IGURE 3 - TYPICAL BOTTOM VIEW MICRO-6000 ........................................................................................................ 3
F
IGURE 4 - LAUNCH MENU, DEVICE CONFIG UTILITY .................................................................................................. 12
F
IGURE 5 - INSTALL DEVICE DRIVER............................................................................................................................ 13
F
IGURE 6 - LAUNCH MENU, SETUP ............................................................................................................................... 13
F
IGURE 7 - ADD COMMUNICATIONS PORT (ROOT) ....................................................................................................... 14
F
IGURE 8 - SELECT DEVICE TO ADD ............................................................................................................................. 14
F
IGURE 9 - SELECT COMMUNICATIONS PORT ............................................................................................................... 15
F
IGURE 10 - SELECT MAXIMUM BAUD RATE ............................................................................................................... 15
F
IGURE 11 - LAUNCH MENU, CONNECT ....................................................................................................................... 16
F
IGURE 12 - CONNECT TO THE CR6 .............................................................................................................................. 16
F
IGURE 13 - SET THE SCAN INTERVAL .......................................................................................................................... 17
F
IGURE 14 - SET ZERO READINGS AND GAUGE FACTORS ............................................................................................. 17
F
IGURE 15 - SET TEMPERATURE ZERO AND THERMAL FACTOR ................................................................................... 18
F
IGURE 16 - UNCOMMENT SPECIFIED LINE .................................................................................................................. 18
F
IGURE 17 - ALIAS NAMES ........................................................................................................................................... 19
F
IGURE 18 - TRANSFER FILE TO DATALOGGER............................................................................................................. 19
F
IGURE 19 - SELECT DLD FILE TYPE ........................................................................................................................... 20
F
IGURE 20 - UNCHECK “RUN NOW” ............................................................................................................................. 20
F
IGURE 21 - SEND THE MAIN CRBASIC PROGRAM FILE ............................................................................................... 21
F
IGURE 22 - LAUNCH MENU, SETUP ............................................................................................................................. 21
F
IGURE 23 - DATA FILES TAB....................................................................................................................................... 22
F
IGURE 24 - LAUNCH MENU, CONNECT ....................................................................................................................... 22
F
IGURE 25 - LIVE MONITORING .................................................................................................................................... 23
F
IGURE 26 - COLLECTING DATA ................................................................................................................................... 23
F
IGURE 27 - 16-CHANNEL SWITCHING BLOCK DIAGRAM ............................................................................................. 25
F
IGURE 28 - 32-CHANNEL SWITCHING BLOCK DIAGRAM ............................................................................................. 25
F
IGURE 29 - 16 OR 32-CHANNEL SELECTION ................................................................................................................ 26
F
IGURE 30 - GK-403/DATALOGGER SELECTION .......................................................................................................... 26
F
IGURE 31 - 16-CHANNEL MICRO-6000 CHANNEL SELECTION TIMING ..................................................................... 27
F
IGURE 32 - 32-CHANNEL MICRO-6000 CHANNEL SELECTION TIMING ..................................................................... 27
F
IGURE 33 - TERMINAL BOARD LAYOUT ...................................................................................................................... 29
F
IGURE 34 - MODEL 8032-27 JUMPER WIRE ASSEMBLY .............................................................................................. 31
F
IGURE 35 - DAISY-CHAIN CONFIGURATION (DOES NOT APPLY TO MODEL 8600-2) ................................................... 34
F
IGURE 36 - RECOMMENDED MAXIMUM CABLE LENGTH (DOES NOT APPLY TO 8600-2) ............................................. 35
F
IGURE 37 - MODEL 8600-2 WITH CELLULAR MODEM INSTALLED .............................................................................. 37
F
IGURE 38 - ADD COMMUNICATIONS PORT (ROOT) ..................................................................................................... 38
F
IGURE 39 - SELECT “PAKBUSPORT” ........................................................................................................................... 38
F
IGURE 40 - SELECT DEVICE TO ADD ........................................................................................................................... 38
F
IGURE 41 - ENTER INTERNET IP ADDRESS .................................................................................................................. 39
F
TABLES
ABLE 1 - 8600-2 SPECIFICATIONS ................................................................................................................................ 7
T
ABLE 2 - TOTAL NUMBER OF ARRAYS AVAILABLE PER MULTIPLEXER ....................................................................... 9
T
ABLE 3 - AMOUNT OF TIME UNTIL DATA OVERWRITE OCCURS WHEN DATA IS STORED EVERY 60 SECONDS ............ 9
T
ABLE 4 - AMOUNT OF TIME UNTIL DATA OVERWRITE OCCURS WHEN DATA IS STORED EVERY 60 MINUTES ............. 9
T
ABLE 5 - AMOUNT OF TIME UNTIL DATA OVERWRITE OCCURS WHEN DATA IS STORED ONCE PER DAY.................... 9
T
ABLE 6 - CR6 WIRING ................................................................................................................................................10
T
ABLE 7 - CHARGER PORT WIRING ..............................................................................................................................10
T
ABLE 8 - UNREGULATED CHARGER WIRING ...............................................................................................................10
T
ABLE 9 - REGULATED CHARGER DEFAULT WIRING ...................................................................................................11
T
ABLE 10 - EXTERNAL POWER CAB LE WIRIN G ............................................................................................................11
T
ABLE 11 - FUSES DESIGNATIONS FOR POWER DISTRIBUTION BOARD .........................................................................11
T
ABLE 12 - 16-CHANNEL MULTIPLEXER/TERMIN AL BOARD WIRING ..........................................................................28
T
ABLE 13 - 32-CHANNEL MULTIPLEXER/TERMIN AL BOARD WIRING ..........................................................................28
T
ABLE 14 - J4 (I/O) CONNECTOR ..................................................................................................................................30
T
ABLE 15 - P1 (I/O) CONNECTOR .................................................................................................................................30
T
ABLE 16 -COMMON CONDUCTOR CHART ...................................................................................................................31
T
ABLE 17 - STANDARD VW LOAD CELL WIR ING WHEN USING 8032-27 .....................................................................31
T
ABLE 18 - MULTIPLEXER SPECIFICATIONS .................................................................................................................33
T
ABLE 19 - DAISY-CHAIN OPERATION/CHANNELS ACCESSED .....................................................................................34
T
1

1. OVERVIEW

The MICRO-6000 Datalogger is designed to support the reading of a large number of Geokon Vibrating Wire Instruments for various unattended data collection applications through the use of an internal Geokon Model 8032 Multiplexer. Weatherproof packaging allows the unit to be installed in field environments where inhospitable conditions prevail. The Nema 4X enclosure also has a provision for locking to limit access to responsible field personnel.
A basic datalogging system consists of the Campbell Scientific CR6 Datalogger and an internal Model 8032 Multiplexer. Multiplexers expand the system by increments of 16 or 32 channels depending on the instrument type and configuration (see Appendix F for more information).

2. SOFTWARE

The Micro-6000 Datalogger is operated by a download file that is generated by a software program. Geokon supplies LoggerNet software (developed by Campbell Scientific, Logan, Utah). Please refer to the manual supplied with the software for instructions on installing the software and creating the required download file for Datalogger operation. A basic starter CRBasic program (STARTPROG) can be provided upon request. LoggerNet 4.3+ is required for the CR6 Datalogger.

3. HARDWARE

The controller portion of the MICRO-6000 Datalogger consists of a Campbell Scientific CR6 Measurement and Control Datalogger. To gain an understanding on the workings and capabilities of the CR6 it is necessary to read the Overview section of the CR6 User Manual. The CR6 User Manual is available in pdf format at:
https://s.campbellsci.com/documents/us/manuals/cr6.pdf
The CR6 Datalogger provides the required excitation and signal processing for the Vibrating Wire sensors connected to the Datalogger. The CR6 uses Campbell Scientific’s VSpect spectral analysis to find the resonant frequency of the Geokon Vibrating Wire Gauge. The CR6 is also used to measure the thermistor used in Geokon’s Vibrating Wire instruments. For complete specifications of the Vibrating Wire Measurements please refer to the CR6 Datalogger User Manual.
A provided IP68 rated Regulated AC charger is used to provide the charging voltage for proper maintenance of the installed lead acid battery. An internal power distribution circuit board contains fused terminal blocks used to provide a nominal 12 VDC supply for the CR6 Datalogger and accessories. A regulated Solar panel can be used in place of the regulated AC charger assuming the maximum short circuit current does not exceed six amps.
The Model 8032 Multiplexer expands the number of channels that can be read by the MICRO­6000 Datalogger. The channel switching is accomplished by mechanical relays mounted on the underside of the circuit board and the transducer connections are accomplished by friction locking terminals. Power, Reset and Clocking for the Multiplexer are supplied by the MICRO-
6000. The Model 8032 Multiplexer can be configured for 16x4 or 32x2 channels depending on the application. Please see the Model 8032 Instruction Manual for more information.
2
A 12V – 7Ah lead acid battery is used to provide power for the Datalogger. The battery supports operating power for a limited period of time should the AC or solar power, used to maintain the battery, be interrupted. Under normal operating conditions and proper maintenance, the life expectancy of the battery is approximately five years

3.1 Installation

The recommended method of installation involves attaching the various enclosures (MICRO­6000 and Multiplexers) to a fixed structure, such as a wall, in an upright position (Figure 1). Mounting dimensions shown in inches.
Figure 1 - Typical Micro-6000 Configuration

3.2 Power

After the Datalogger is installed the Geokon Regulated AC Charger or Regulated Solar Panel can be connected to the three-pin charger port on the Datalogger. The supply voltage to the Micro­6000 will be “Off” by default and the On/Off switch SW1 on the Power Distribution circuit board should be switched to the “On” position (Figure 2). The Internal Lead Acid battery will charge in either position. It is recommended that the charger be left plugged in at all times (Section 4.1). Additionally, a Geokon Unregulated AC Charger or Solar Panel can be used as the three-pin charger port is also wired to the CR6 built in Charge Regulator.
Figure 2 - Power Distribution Circuit Board
3

3.3 Earth Ground

An earth grounding lug is supplied on the exterior of the Micro-6000 enclosure to connect the system to earth ground (Figure 3). A grounding rod can be driven (or other suitable attachment to earth utilized) to ground the system and provide a path to earth for protection against a lightning strike or other transient voltage. A two to two and a half meter (six to eight foot) copper grounding rod connected to the Dataloggers grounding lug with a large gauge wire (12 AWG or larger) is recommended. The earth connection should be made as close to the Datalogger as possible.

3.4 Gauges

Following enclosure installation, Geokon Vibrating Wire sensors can be connected to the internal Multiplexer. Please refer to the Multiplexer instruction manual for the appropriate connection description. Addition Start Guides can be provided for other types of gauges or custom configurations.

3.5 Communications

Standard direct communications with the Datalogger is established by connecting the Micro­6000 USB port (Figure 3) and an available USB port on the computer. A COM-113 USB A-A Male cable is supplied with every new order. A USB Driver for the CR6 must be installed prior to communications. The Device Configuration Utility can be found under Utilities on the main LoggerNet Screen.
Figure 3 - Typical Bottom View MICRO-6000
4

4. BATTERY MAINTENANCE

4.1 AC Power

The MICRO-6000 is supplied with an external AC to DC Rain Proof Regulated Smart Charger for maintaining the charge of the internal battery which provides power to the Datalogger and peripherals. It is imperative that a regulated charging source remain connected to the Datalogger as the battery installed in the Datalogger is only provided as a temporary source of power should the power supply be disconnected or mains power interrupted. Actual run time, solely on battery power, will vary for each Datalogger system and depends on the hardware configuration and sensor scan interval. If the battery voltage drops below 9.6 volts operation of the Datalogger will become erratic as evidenced by communication problems and possible improper measurements. Additionally, a Geokon unregulated AC charger can be used as the three pin Charger port is also wired to the CR6 built in Charge Regulator.

4.2 Solar Power

If AC mains power is not available, a properly sized regulated solar panel can be used to provide power to maintain the charge state of the battery. The size of the solar panel is determined by geographic location of the Datalogger, hardware configuration, communications interval and the sensor scan interval. Additionally, a Geokon Unregulated Solar Panel can be used as the three pin Charger port is also wired to the CR6 built in Charge Regulator.

4.3 External Battery

Additionally, a large external battery, such as a deep cycle marine battery, can be connected to the Datalogger via the supplied external power cable to provide power to the Datalogger and peripherals. When using this type of battery, the internal battery should be manually disconnected from the system to prevent parasitic drain on the external battery.

4.4 Battery Replacement

If the internal Lead Acid Battery has failed it is recommended that the unit be returned to the factory for service by Geokon personnel. However, with skilled personnel and appropriate tools, it is possible for the user to replace the battery. Consult the factory for information.

4.5 Fuses

There are two fuses on the internal Power Distribution Board of the MICRO-6000 (Figure 2). Once removed and with the power switch off, a fuse can be checked visually and with an ohmmeter. A gap may be evident (with some discoloration) if the fuse needs replacing. This can be verified by a high resistance measurement (megaohms) with an ohmmeter. If fuse needs replacing, insert one of the supplied replacement fuses. If there are no replacement fuses available, consult the factory or they can be purchased from an electrical supply house. The F1 charger fuse is an eight amp SLO-BLO 5x20 mm and the F2 fuse is a four amp SLO-BLO 5x20 mm. Consult Appendix D.6 for fuse assignments.

5. TROUBLESHOOTING

This section will NOT attempt to cover all possible problems that could be encountered in the course of Datalogger operations. Consult the factory if other problems arise or remain unresolved.
Symptom: Cannot communicate with the Datalogger:
The USB Driver for the CR6 has not been installed on the host computer. See LoggerNet
- Utilities - Device Config Utility.
Verify the CR6 USB Driver has been properly installed.
The wrong USB communication port is being selected in the LoggerNet Setup Screen
(refer to Device Manager of PC for assigned COM Port). Consult the appropriate software manual for instructions on changing the communication port.
The USB communication port on the host computer is defective. Verify the functioning and configuration of the COM port by using it with another USB device, such as a camera or cellular phone.
The Datalogger Load Fuse (F2) on the Power Distribution Board is blown. Consult Section 4.5 for checking and/or replacement.
Symptom: The system battery voltage and panel temperature read odd numbers:
The system battery could be low. Charge and check again (see Section 4, Battery Maintenance).
A disruptive current loop may be operating as a result of improper grounding or excessive noise. Consult the factory for more information.
Symptom: The internal battery measurement does not increase when charging:
The AC Regulated Charger may be damaged. Check the output pins of the adaptor with a voltmeter.
The Charger fuse (F1) of the Power Distribution Board is blown. Consult Section 4.5 for checking and/or replacement.
The internal battery is no good. Consult the factory.
The Regulated Solar Panel may be damaged, obstructed or have poor sun exposure.
Check the output pins of the adaptor with a voltmeter. Verify Solar Panel is not obstructed.
5
6
Symptom: The Datalogger will not operate on external power:
The external voltage supply is below operating limits. If the external source is a battery, charge it. If it is a power supply, check the output with a voltmeter.
Verify both Load and Charger Fuses on Power Distribution Board (Refer to Section 4.5).
Symptom: Loss of Micro-6000 program and/or data:
The system has experienced a voltage dropout or surge which disrupted operations.
The surge originated as a result of lightning. Install appropriate grounding. Install
lightning protection devices on all incoming and outgoing lines (consult factory).
Verify .CR6 Program is running on the Datalogger (See Appendix E).
Symptom: Sensor readings show “NAN” or are unstable:
The wrong sensor type has been connected. Check the model number of the sensor against the requested sensor list.
The cable(s) to the sensor(s) have been damaged permitting moisture and debris to enter the jacket. Wires may be shorted together. Inspect the cable and perform Ohm test to check for short circuits.
If all sensors on a particular Multiplexer are erratic or returning “NAN” perhaps the Multiplexer or cable has been damaged. Inspect the cable. Try another cable to verify proper functioning.
The sensor(s) have been damaged. For example, overrange on a vibrating wire sensor can cause erratic readings.
There is an electrical noise source nearby. Move the sensor, cables, and Datalogger away from the noise source. Install grounding devices. Consult the factory.
Analog Inputs
12 single-ended or 6 differential with ±5000 mV, ±1000 mV, ±200 mV ranges 24 bit ADC
Analog Outputs
±2.5 V or ±2.5 mA ranges 12 bit DAC
Accuracy
±(0.04% of reading + 2 microvolts), 0-40 °C
Resolution
50 mV (±200 mV range, differential measurement, input reversal, 5 Hz fN1)
Static Frequency-Analyzed Vibrating Wire
12 V p-p 100-6000 Hz (Spectral analysis technique)
Accuracy
±0.013% of reading
Resolution
0.001 Hz RMS
Thermistor Accuracy
±0.25% of reading
Temperature Range
-40 to +80 °C
Battery
12 V, 7 Ah Gel Cell
LxWxH
502 x 461 x 263 mm

APPENDIX A. SPECIFICATIONS

A.1 Model 8600-2 Datalogger

Table 1 - 8600-2 Specifications

A.2 CR6 Measurement and Control Module

Please refer to Campbell Scientifics Specification for the CR6 Datalogger
https://s.campbellsci.com/documents/us/product-brochures/s_cr6.pdf
7

A.3 Model 8032 Multiplexer

Please refer to the Geokon Model 8032 Multiplexer Manual.
http://www.geokon.com/content/manuals/8032_Terminal_Board_and_Multiplexer.pdf
8

APPENDIX B. SHIP LIST

B.1 Hardware

The following equipment is included with the system:
External Power cable
USB A Type Male to Male 1M
Small regular screwdriver
Three spare Slo-Blo fuses, four and eight amp
AC Smart Regulated Charger Rain Proof
The following manuals are included:
MICRO-6000 Instruction Manual
Campbell Scientific Resource disk
8032 Multiplexer Instruction Manual (if Multiplexers provided)
Optional accessories:
Cellular Phone Modem Install Kit with Start Guide (Modem Provided by Customer)
Micro-SD memory Card
Ethernet Bulkhead for communications
Regulated Solar Panel with mounting hardware
RF Modem with manuals (wireless Datalogger communication)
Datalogger Starter CRBasic Program
Custom CRBasic Programming
WIFI Communications Modem
RF Whip or High Gain Antennas with mounting hardware
Consult the factory for additional information on any of the optional accessories.
Total Arrays
1 Multiplexer Array Storage:
49,6652
Total
Arrays
1 Multiplexer Overwrite Time:
496,652
8,277
344
496,652
Total
Arrays
1 Multiplexer Overwrite Time:
496,652
20,693
56.6
496,652
Total
Arrays
1 Multiplexer Overwrite Time:
571,200
496,652
1,359.7
496,652

APPENDIX C. DATA STORAGE

C.1 Data Storage

Total Arrays of Data that can be stored per 16-channel Multiplexer. Each array contains all Data stored at each read interval. The 8600-2 has a single internal Multiplexer.
Table 2 - Total Number of Arrays Available per Multiplexer

C.2 Data Storage Example

C.2.1 Overwrite Times when Data is Stored Once a Minute

9
Table 3 - Amount of Time Until Data Overwrite Occurs when Data is Stored Every 60 Seconds
Minutes Hours Days

C.2.2 Overwrite Times when Data is Stored Once an Hour

Table 4 - Amount of Time Until Data Overwrite Occurs when Data is Stored Every 60 minutes
Hours Days Years

C.2.3 Overwrite Times when Data is Stored Once a Day

Table 5 - Amount of Time Until Data Overwrite Occurs when Data is Stored Once per Day
Hours Days Years
10
Internal
Ribbon Cable
CR6
Cable
U1
Brown
NC
Vibrating Wire +
U2
Red
NC
Vibrating Wire -
U3
Orange
NC
Thermistor +
U4
Yellow
NC
Thermistor -
U5
Gray
NC
MUX Enable 1
C4
White
NC
MUX Clock
G
Violet
NC
Ground
Analog Ground
CR6 12 VDC
CR6 Power Ground
Pin
Description
Wire Color
A
Unregulated Charger
Gray
B
Ground
Blue
C
Regulated Charger / External
Battery
Violet
Pin
Description
Wire Color
A
Unregulated Charger
Red (Black Smooth)
B
Ground
Black (Black Rough)
C
No Contact
NC

APPENDIX D. SYSTEM WIRING

D.1 CR6 Wiring

CR6 Connections
MUX 1 Interface
Power
Description
12V Blue NC MUX 12 VDC
AG Black and Green NC
PWR IN 12V NC Red
PWR IN G NC Black
Table 6 - CR6 Wiring
NOTE: Ports U1 and U2 are used for Vibrating Wire when the 8600-2 is configured for 32 ea
Vibrating Wire Gauges. U3 and U4 are used when the 8600-2 is configured for 32 ea Thermistors.

D.2 Charger Port Wiring

D.3 Unregulated Charger Wiring

Table 7 - Charger Port Wiring
Table 8 - Unregulated Charger Wiring
11
Pin
Description
Wire Color
A
No Contact
NC B Ground
Black(Black Rough)
C
Regulated Charger/External Battery
Red(Black Smooth)
Pin
Description
Wire Color
Clip
A
No Connection
NC
No Connection
B
Ground
Black
Black
C
Battery+(12VDC)
Red
Red
Fuse
Description
F1
Eight Amp Charger/External Battery
F2
Four Amp Battery/Load

D.4 Regulated Charger (Default)

Table 9 - Regulated Charger Default Wiring

D.5 External Power Cable

Table 10 - External Power Cable Wiring
NOTE: Disconnecting the Internal Battery is recommended when powered by External Battery.

D.6 Fuse Designations for Power Distribution Board

Table 11 - Fuses Designations for Power Distribution Board
12

APPENDIX E. LOGGERNET QUICK START GUIDE

E.1 Overview

Campbell Scientific’s LoggerNet Software is used to communicate with the Datalogger, program user specific settings and collect Data stored in the Datalogger memory. LoggerNet is designed to be used with Geokon Micro-6000 Dataloggers reading Vibrating Wire and MEMS tilt sensors; other configurations and sensor outputs are available upon request. Upon request Geokon provides the user with a generic Start Program to simplify acquiring Data without the need to be proficient in Campbell Scientific CRBasic programming code. The Start Program(s) will be preconfigured with a default Scan Interval and will store Raw Units by default. Each Start Program will have two associated files on the factory supplied USB flash drive. The first is the Main CRBasic program (.CR6) and includes all programming code required to read and store Data according to the Datalogger configuration. The Main CRBasic program does not require modification by the user in most cases. The second file is the Include (.DLD) text file and is used by the Main program to import user specific parameters. The Include file allows program parameters such as the Scan Interval, Zero Readings, Linear Gauge Factors, Data File Header Labels and Temperature Correction factors to be modified by the user.

E.2 Getting Started

E.2.1 Program Files

After installing LoggerNet (must be Version 4.3 +), the CRBasic program files are located on the Geokon USB Thumb Drive. The file types are “.DLD” and “.CR6”, these files should first be copied to the end users computer in C:\Campbellsci\CRBasicEditor.

E.2.2 Connection Setup

The Launch Menu provides a list of all the categories on the LoggerNet toolbar. Hovering over a category will display a list of applications related to that category.
Prior to the first time any computer connects to a Micro-6000/CR6 Datalogger via USB the USB drivers must be installed. Device Config must be Version 2.10 +
With the Launch Menu open, hover over “Utilities” and press “Device Config Utility” from the resulting list of applications.
Figure 4 - Launch Menu, Device Config Utility
13
In Device Type, Select “CR6” from the datalogger group. Within the CR6 pane use the “install the USB driver” hyperlink to install the USB drivers. Follow the steps in the Device Driver Installation Wizard.
Figure 5 - Install Device Driver
With the Launch Menu open, hover over “Main” and press “Setup” from the resulting list of applications.
Figure 6 - Launch Menu, Setup
By default, the EZ View of the Setup Screen is displayed. To change the viewing method, press the “Std View” icon at the top right of the Setup Screen. This Quick Start will show the “Std View” of the Setup Screen.
14
To add a communication port, press “Add Root” button.
Figure 7 - Add Communications Port (Root)
Each Micro-6000 Datalogger will communicate via USB. Select connection type “ComPort”, “PakBusPort (Other Loggers)”, “CR6Series and press “Close”.
Select “ComPort” under the “Network Map”, select the communications port to connect to the Datalogger in “ComPort Connection”.
CONDITIONAL NOTE: IT MAY BE NECESSARY TO PRESS “COMMUNICATIONS ENABLED” TO
OPEN THE
COMMUNICATIONS PORT.
Figure 8 - Select Device to Add
15
Figure 9 - Select Communications Port
COM NUMBERS VARY WITH COMPUTER, USB DEVICES AND SERIAL ADAPTERS. REFER TO THE COMPUTER DEVICE MANAGER IF UNSURE OF WHICH COM NUMBER
TO USE
.
Select “PakBusPort” under the “Network Map”, in “Maximum Baud Rate” for direct connection to the Datalogger select “115200”.
WHEN FINISHED, PRESS THE “APPLY BUTTON AT THE BOTTOM LEFT OF THE SETUP SCREEN TO SAVE ALL SETTINGS.
Figure 10 - Select Maximum Baud Rate
16

E.2.3 Connecting to the Datalogger

With the LoggerNet Launch Menu open, hover over “Main” and press “Connect”.
Figure 11 - Launch Menu, Connect
Select the Datalogger to communicate with under “Stations” and press the “Connect” button.
Figure 12 - Connect to the CR6

E.3 Modifying the Datalogger Program

Navigate to the Include (.DLD) file on the end user’s computer C:\Campbellsci\CRBasicEditor. Optional changes to Scan Interval, Zero Readings, Gauge Factors and Alias names can be made by opening the Include (.DLD) file with Notepad. Each Datalogger configuration has a unique Include (.DLD) file.
WARNING: ONLY CHANGE VALUES AFTER THE “=” SIGN.

E.3.1 Scan Interval

The number entered in the Scan Interval section determines how often the Datalogger will read the gauges and store data. The Scan Interval is in seconds.
17
Figure 13 - Set the Scan Interval

E.3.2 Zero Readings and Gauge Factors

Default values for Zero Readings are 0 and Linear Gauge Factors are 1. The default settings read in Digits for Vibrating Wire sensors and Volts for MEMS tilt sensors. Zero Readings and Gauge Factors can be added if the user requires the Datalogger to store engineering units.
Figure 14 - Set Zero Readings and Gauge Factors
18

E.3.3 Temperature Zero Readings and Thermal Factors

NOTE: TEMPERATURE ZERO AND THERMAL FACTORS ARE USED FOR TEMPERATURE
CORRECTION
A GAUGE
VARIES ON DIFFERENT MODELS OF GAUGES AND TEMPERATURE CORRECTION IS OFTEN NOT
REQUIRED
AVAILABLE
. TEMPERATURE CORRECTION CAN BE USED TO COMPENSATE FOR CHANGES IN
S READINGS DUE TO TEMPERATURE CHANGE. TEMPERATURE EFFECT ON READINGS
. DEPENDING ON THE GAUGE MODEL, TEMPERATURE CORRECTION MAY NOT BE
; SOME GAUGE MODELS REQUIRE A DIFFERENT FORMULA TO COMPENSATE.
TEMPERATURE CORRECTION IS COMMENTED OUT IN THE MAIN PROGRAM FILE BY DEFAULT.
Default values for Temperature Zero Readings and Thermal Factors are 0. Temperature Zero Readings and Thermal Factors can be added if the user requires the Datalogger to use temperature compensation.
Figure 15 - Set Temperature Zero and Thermal Factor
Navigate to the Main (.CR6) file C:\Campbellsci\CRBasicEditor. Open the Main (.CR6) file with notepad and uncomment the line specified in the code for all model gauges that temperature correction is required.

E.3.4 Output Labels

Default Alias names correspond with the Direct or Multiplexer channel of each gauge and are displayed in the Header of the Data file. Alias names can be modified by the user.
WARNINGS: ALIAS NAMES CANNOT BE MORE THAN 35 CHARACTERS LONG AND
MUST ONLY CONTAIN ALPHANUMERIC VALUES AND UNDERSCORES
SYMBOLS
). ALIAS NAMES MUST BEGIN WITH AN ALPHA CHARACTER
Figure 16 - Uncomment Specified Line
(NO SPACES OR
19
Figure 17 - Alias Names
SAVE THE INCLUDE FILE IF MODIFIED.

E.4 Uploading Files

The Include.DLD file must be saved when modified. After modification, the Include.DLD must be sent to the Datalogger using the File Control Menu in the LoggerNet Connect Screen. The Main (.CR6) program file must be sent to the Datalogger each time the Include.DLD file is updated.

E.4.1 Sending the Include file (.DLD)

From the “Connect Screen” press “File Control” and press “Send” to transfer the Include (.DLD) file to the Datalogger.
NOTE: EACH DATALOGGER HAS A UNIQUE INCLUDE (.DLD) AND MAIN (.CR6) FILE
Figure 18 - Transfer File to Datalogger
20
Select DLD file type to make the Include (.DLD) file visible in C:\Campbellsci\CRBasicEditor.
Uncheck “Run Now”, the Include (.DLD) file cannot run the Datalogger.
Figure 19 - Select DLD File Type
Figure 20 - Uncheck “Run Now”
21

E.4.2 Sending the Main program file (.CR6)

To send the Main (.CR6) CRBasic Program from the “Connect Screen” press “Send” and select the Main (.CR6) program file in C:\Campbellsci\CRBasicEditor.
NOTE: EACH DATALOGGER HAS A UNIQUE INCLUDE (.DLD) AND MAIN (.CR6) FILE
Figure 21 - Send the Main CRBasic Program File
When sending the Main (.CR6) program file through the “Connect Screen” the program runs automatically.

E.5 Data Handling

E.5.1 Data Collection Configuration

With the Launch Menu open, hover over “Main” and press “Setup” from the resulting list of applications.
Figure 22 - Launch Menu, Setup
22
Select the Datalogger “CR6Series”, “Data Files” tab then select “Table1”. Dataloggers can be renamed by pressing “Rename” button. Renaming the Datalogger affects the name of the Data file. “Output File Name” option allows changing the file name, where to collect and store the Data file. “File Output Option” determines whether new Data collected is appended into one file, overwrites old Data, or creates a new file every time Data is collected.
COLLECTED DATA TABLES ARE SIGNIFIED WITH A GREEN CHECK MARK.
Figure 23 - Data Files Tab
WHEN FINISHED, PRESS THE “APPLY BUTTON AT THE BOTTOM LEFT OF THE SETUP SCREEN TO SAVE ALL SETTINGS.

E.5.2 Live Monitoring

With the LoggerNet Launch Menu open, hover over “Main” and press “Connect”.
Figure 24 - Launch Menu, Connect
Once the program is running, most current readings can be viewed under “Table1” of the “Table Monitor” and “Num Display”. “Graphs” can also be used to view live and historical Data.
CONDITIONAL NOTE: IT MAY BE NECESSARY TO START MONITORING BY PRESSING “START”.
23
If unexpected “NAN” readings appear, refer to Section 5 (Troubleshooting).
Figure 25 - Live Monitoring

E.5.3 Collecting Data

Pressing “Collect Now” collects and stores Data on the computer. After collecting Data a “Data Collection Results” screen will open.
Figure 26 - Collecting Data
PLEASE SEE LOGGERNET MANUAL FOR MORE SPECIFIC DETAILS REGARDING USE OF THE
SOFTWARE
.
24

APPENDIX F. EXCERPTS FROM THE 8032 MULTIPLEXER MANUAL

The following Appendix includes excerpts from the Model 8032 Multiplexer manual that apply to the Micro-6000. A complete version of the Model 8032 Multiplexer Manual can be found at
WWW.GEOKON.COM.

F.1 Theory of Operation

The Model 8032 Multiplexer expands the number of channels that can be read by the MICRO­6000 Datalogger. Channel switching is accomplished by mechanical relays mounted on the underside of the circuit board and the transducer connections are accomplished by friction locking spring-loaded terminals mounted on the top side of the circuit board.
To protect against lightning or EMI/RFI induced transients, each channel is protected by an integrated lightning protection system, incorporating 230V tripolar plasma surge arrestors, 150V bipolar plasma surge arrestors, 10uH inductors and 16V transient voltage protection diodes. See Appendix A for complete specifications on these components.
The multiplexer is powered by a nominal 12 VDC supply. Two control lines (RESET and CLOCK) determine how channel selection is accomplished. Two schemes are supported - one when connected to the GK-403 Vibrating Wire Readout Box and the other when connected to Micro-6000 Datalogger.
Two switching configurations are supported, 16-channels of four conductors (Figure 27) or 32­channels of two conductors (Figure 28). The 16-channel, four wire switching configuration is typically used to multiplex four wire sensors such as resistance strain gauge load cells. It is also used to switch connections for instruments which have more than one sensor integral to them, such as vibrating wire pressure transducers with an integral thermistor for measuring temperature. The 32-channel two wire switching configuration is typically used to multiplex two wire sensors such as a vibrating wire pressure transducers, thermistors or thermocouples.
25
Relay
Control
Microcontroller
12V
GROUND
RESET
CLOCK
U1
U2
CR6 Sensor #1
12V
GROUND
RESET
CLOCK
Multiplexer/Terminal Board
AG
Sensor
Shield
U3
U4
Sensor #16
Sensor
Shield
1H
2H
1L
2L
S1
31H
32H
32L
31L
S16
Relay
Control
Microcontroller
12V
GROUND
RESET
CLOCK
U1
U2
CR6
Sensor #1
Sensor #32
12V
GROUND
RESET
CLOCK
Multiplexer/Terminal Bo
AG
Sensor
Shield
Sensor
Shield
1H
1L S1
32H
32L
S16
Figure 27 - 16-Channel Switching Block Diagram
Figure 28 - 32-Channel Switching Block Diagram
26
ON
OFF
SW1
1 2 3
4
GK-403
DATALOGGER
SW2
For the 8032C, these configurations are set by a Dipswitch on the top side of the circuit board. A second Dipswitch selects whether the 8032C is being used with a Datalogger or GK-403 Readout Box.
Position one the DIP switch SW1 is used for switching between 16 and 32-channel operation. Switch 1 ON = 32-channel; Switch 1 OFF = 16-channel. In Figure 29, 32-channel mode is chosen:
Figure 29 - 16 or 32-Channel Selection
The DIP switch SW2 is used for switching between a GK-403 or Datalogger application. “DATALOGGER” is the default SW2 position, as shown in Figure 30.
Figure 30 - GK-403/Datalogger Selection

F.2 MICRO-6000 Mode of Operation

The MICRO-6000 (which utilizes a CR6 controller, manufactured by Campbell Scientific, Inc. of Logan, Utah) mode of operation uses two control lines to operate the multiplexer. The RESET line enables the multiplexer and activates the Micro-6000 mode of clocking. Pulses received on the CLOCK line sequentially increment the channels while the RESET line is held high. See the timing diagram below.
The CLOCK line sequentially advances the channels beginning with channel 1. Note the timing diagrams below for 16 and 32-channel switching, respectively.
27
Reset
Clock
Channel 1
T1
T2 T3
No Channel
Selected
No Channel
Selected
Channel 16
Pulse 1
Pulse 2
Pulse 31
Timing:
T2 = 2
mSsec(min)
T3 = 2 mSsec(min)
T1 = 50
mSec(min.)
Selected
Selected
Pulse 32
Reset
Clock
Channel 1
Channel 2
T1
T2
T3
No Channel
Selected
No Channel
Selected
Channel 32
Pulse 1
Pulse 2
Pulse 32
Selected
Selected
Selected
Timing:
T2 = 2 mSec (min)
T3 = 2 mSec (min)
T1 = 50 mSec (min)
Figure 31 - 16-Channel MICRO-6000 Channel Selection Timing
Figure 32 - 32-Channel MICRO-6000 Channel Selection Timing
28
Terminal
Board
Vibrating Wire
with Thermistor
Resistance Strain
Gauge Bridge
Linear Potentiometer
(with Remote Sense)
1H
VW Sensor #1
S+ from Bridge #1
Excitation Pot #1
1L
VW Sensor #1
S- from Bridge #1
Wiper Output Pot #1
2H
Thermistor #1
P+ to Bridge #1
Remote Sense Pot #1
2L
Thermistor #1
P- to Bridge #1
Ground Pot #1
S1
Shield Drain Wire
from Sensor #1
Shield Drain Wire
from Bridge #1
Shield Drain Wire
from Pot #1
3H
VW Sensor #2
S+ from Bridge #2
Excitation Pot #2
3L
VW Sensor #2
S- from Bridge #2
Wiper Output Pot #2
4H
Thermistor #2
P+ to Bridge #2
Remote Sense Pot #2
4L
Thermistor #2
P- to Bridge #2
Ground Pot #2
S2
Shield Drain Wire
from Sensor #2
Shield Drain Wire
from Bridge #2
Shield Drain Wire
from Pot #2
• • •
• • •
• • • • •
31H
VW Sensor #16
S+ from Bridge #16
Excitation Pot #16
31L
VW Sensor #16
S- from Bridge #16
Wiper Output Pot #16
32H
Thermistor #16
P+ to Bridge #16
Remote Sense Pot #16
32L
Thermistor #16
P- to Bridge #16
Ground Pot #16
S16
Shield Drain Wire
from Sensor #16
Shield Drain Wire
from Bridge #16
Shield Drain Wire
from Pot #16
Terminal
Board
Vibrating Wire
Thermistor
Thermocouple
1H
VW Sensor #1
Thermistor #1
Thermocouple #1
1L
VW Sensor #1
Thermistor #1
Thermocouple #1
2H
VW Sensor #2
Thermistor #2
Thermocouple #2
2L
VW Sensor #2
Thermistor #2
Thermocouple #2
S1
Shield Drain Wires
from Sensors 1 and 2
Shield Drain Wires
from Thermistors 1 and 2
3H
VW Sensor #3
Thermistor #3
Thermocouple #3
3L
VW Sensor #3
Thermistor #3
Thermocouple #3
4H
VW Sensor #4
Thermistor #4
Thermocouple #4
4L
VW Sensor #4
Thermistor #4
Thermocouple #4
S2
Shield Drain Wires
from Sensors 3 and 4
Shield Drain Wires
from Thermistors 3 and 4
• • • • • • •
• • •
31H
VW Sensor #31
Thermistor #31
Thermocouple #31
31L
VW Sensor #31
Thermistor #31
Thermocouple #31
32H
VW Sensor #32
Thermistor #32
Thermocouple #32
32L
VW Sensor #32
Thermistor #32
Thermocouple #32
S16
Shield Drain Wires
from Sensors 31 and 32
Shield Drain Wires
from Thermistors 31 and 32

F.3 Wiring

F.3.1 Gauge Wiring

Actual gauge connections to the terminal board will vary depending on the instrument type and cable used. Note the following tables to get the general idea.
Table 12 - 16-Channel Multiplexer/Terminal Board Wiring
Table 13 - 32-Channel Multiplexer/Terminal Board Wiring
29
J1
J2
1H
1L
2H
2L
S1 3H
3L
4H
4L
S2 5H
5L
6H
6L
S3 7H
7L
8H
8L
S4
9H
9L
10H
10L
S5 11H
11L
12H
12L
S6 13H
13L
14H
14L
S7 15H
15L
16H
16L
S8
17H
17L
18H
18L
S9 19H
19L
20H
20L
S10 21H
21L
22H
22L
S11 23H
23L
24H
24L
S12
25H
25L
26H
26L
S13
27H
27L
28H
28L
S14 29H
29L
30H
30L
S15 31H
31L
32H
32L
S16
EARTH
J4
T1/2
T3/4
T5/6
T7/8
T9/10
T11/12
T13/14
T15/16
T17/18
T19/20
T21/22
T23/24
T25/26
T27/28
T29/30
T31/32
SHLD
SHLD
C1H
C1L
C2H
C2L
SHLD
+12V
GND
RESET
CLOCK
SHLD
P1
GK-403
DATALOGGER
SW2
SW1
ON
OFF
Figure 33 depicts the terminal board to which gauge connections are made. If the terminal board is equipped with manual switches, connectors J1 and J2 will have ribbon cables that are connected to the switch boards.
Terminal Blocks T1/2 to T31/32 are for the gauge connections.
Figure 33 - Terminal Board Layout
30
Inside
Color
10 Pin
Bendix
8032-5 (TAN) Cable
Wire Color
1
Brown A COM HI 1
White
2
Red B COM LO 1
White's Black
3
Orange
C
COM HI 2 ( 16-channel)
Red
4
Yellow
D
COM LO 2 ( 16-channel)
Red's Black
Shield Drain Wires – all pairs plus
overall
6
Blue F +12 Volt Power
Yellow
7
Purple G Power Ground
Yellow's Black
RESET (DATALOGGER)
SENSE (GK-403)
9
White J CLOCK
Green's Black
10 E
No Connection
Blue and Blue’s Black (unused)
P1 Terminal
Label
Description
1,2
SHLD
SHIELD
3,4
SHLD
SHIELD
5,6
C1H
COM HI 1
7,8
C1L
COM LO 1
9,10
C2H
COM HI 2 ( 16-channel)
11,12
C2L
COM LO 2 ( 16-channel)
13,14
SHLD
SHIELD
15,16
+12V
+12 Volt Power
17,18
GND
Power Ground
19,20
RESET
RESET
21,22
CLOCK
CLOCK
23,24
SHLD
SHIELD

F.3.2 Connector and Cable Wiring

J4
Description
5 Green K Analog Ground
8 Grey H
Table 14 - J4 (I/O) Connector
Green
Note: P1 is a “stacking” type connector. Odd number terminals are read left to right on the bottom row. Even numbered terminals are read left to right on the top row.
Table 15 - P1 (I/O) Connector
31
10 pin Bendix
PT06A-12-10P
Function
3 Gauge VW Load
Cell, Purple Cable
4 Gauge VW Load
Cell, Purple Cable
6 Gauge VW Load
Cell, Orange Cable
* White's black and Green wires are switched on Geokon 3 gauge VW load cells prior to serial number 3313.
Multiplexer
Board
Vibrating Wire
with Thermistor
1H
VW Sensor #1
1L
8032-27
2H
Thermistor
2L
Thermistor
S1
Shield Drain Wire
3H
VW Sensor #2
3L
8032-27
4H - 4L - S2 - 5H
VW Sensor #3
5L
8032-27
6H - 6L - S3
-
11H
VW Sensor #6
11L
8032-7
12H
-
12L - S6
-

F.3.3 Model 8032-27 and Load Cell Wiring

Connect the “common” VW- conductor from the load cell to the 8032-27 by lifting up on the orange tab located on the opposite side of the six black conductors, inserting the common conductor fully into the 8032-27 (Figure 34), and then pushing down on the orange tab until it snaps into place. Refer to Table 16 to identify which conductor carries the common VW- signal.
Figure 34 - Model 8032-27 Jumper Wire Assembly
H Common White's Black* Green Blue
Table 16 -Common Conductor Chart
Table 17 details the connections between the load cell and 8032-27 with the multiplexer board.
Table 17 - Standard VW Load Cell Wiring When Using 8032-27
32
GENERAL:
Power Requirements
10-16 VDC (unregulated)
Quiescent Current
80 µA (16CH mode) 130µA (32CH mode)
Quiescent Current (GK-403 mode)
12 mA
Channel Activated Current
30 mA
Control Line Input Impedance
100 k (CLOCK), 100 k (RESET)
Control Line Input Levels
TTL or RS-232 (±9 VDC)
Transient Protection
16 VDC Transzorbs
Operating Temperature
40 to +60° C
RELAYS:
Type
NAIS TXS2SA-4.5V DPDT non-latching
Power
11.1 mA @ 5VDC (55.5 mW)
Contact Type
Gold clad silver alloy
On Resistance
100 m
Coil Resistance
405
Maximum Switching Power
30W (resistive)
Maximum Switching Voltage
110 VDC
Maximum Switching Current
1 A
Operate Time
5 msec Max.

F.4 Troubleshooting

Below are some commonly experienced problems along with possible remedial action. Contact the factory if any problem remains unresolved or additional help is required.
Symptom: A particular channel on the multiplexer appears to be malfunctioning:
Check sensor connections on the terminal board. Clean if corrosion exists.
Try moving the sensor wired to the suspect channel to another channel to verify the
malfunctioning of the channel (as opposed to the sensor).
Symptom: No channels are working:
Inspect circuit board for shorts, opens, or other damage.
Is moisture present on circuit board? If so, install desiccant to absorb.
Symptom: Channel selection appears to be random:
Has corrosion built up on the circuit board? Clean if necessary.
Is there a source of electrical noise nearby? Move multiplexer or noise source if possible.

F.5 8032 Multiplexer Specifications

(MICRO-800/MICRO-1000 mode)
33
Release Time
5 msec
Switching Life
Mechanical: 5 x 107 operations Electrical @ 30W: 2 x 105 operations
Ambient Temperature
40 to +70°C
TRIPOLAR PLASMA SURGE ARRESTOR:
Nominal DC Breakdown Voltage
230V
Surge Life
100 (10/1000 ms pulse @ 200 A)
Maximum Surge Current
5 kA per side (8/20 µs pulse)
Insulation Resistance
109
Operating Temperature
65 to +125° C
BIPOLAR PLASMA SURGE ARRESTOR:
Nominal DC Breakdown Voltage
150V
Maximum Surge Current
1 kA (8/20 µs pulse)
Insulation Resistance
1010
Operating Temperature
65 to +125° C
INDUCTOR:
Rated Current
4A
Inductance
10µH (±20%)
D.C.R.
25m MAX (at 20°C)
TRANSIENT VOLTAGE SUPPRESSOR (TRANSORB)
Rated Power
1500W
Peak forward Surge Current
200A
Reverse Standoff Voltage
16.0V
TRANSDUCER CONNECTION MAXIMUM OPERATING VOLTAGE LEVELS:
Common-mode Voltage/Earth Ground
16V(max)
Differential-mode Voltage (Channel # ‘H’ – Channel # ‘L’)
16V(max) RESET AND CLOCK MAXIMUM OPERATING VOLTAGE LEVELS:
Single-ended Control Voltage/System Ground
16V(max)
Table 18 - Multiplexer Specifications

F.6 Daisy-Chain Operation

Up to eight 8032’s can be “daisy-chained” together using a common RESET and CLOCK control line. This may be advantageous in situations where either there are not enough control ports available on the Micro-6000 Datalogger for the number of multiplexers desired, or to reduce the number of cables required to implement a large multi-channel system.
SW1 located on the Terminal Block side of the circuit board determines the address of each multiplexer and the corresponding signal channels. As many as 256 two-conductor channels or 128 four-conductor channels can be accessed per RESET line.
34
SW1 Setting
Channels Accessed
2 3 4
OFF
OFF
OFF
1-32 (32-channel mode), 1-16 (16-channel mode) DEFAULT MUX1
OFF
OFF
ON
33-64 (32-channel mode), 17-32 (16-channel mode) MUX2
OFF
ON
OFF
65-96 (32-channel mode), 33-48 (16-channel mode) MUX3
OFF
ON
ON
97-128 (32-channel mode), 49-64 (16-channel mode) MUX4
ON
OFF
OFF
129-160 (32-channel mode), 65-80 (16-channel mode) MUX5
ON
OFF
ON
161-192 (32-channel mode), 81-96 (16-channel mode) MUX6
ON
ON
OFF
193-224 (32-channel mode), 97-112 (16-channel mode) MUX7
ON
ON
ON
225-256 (32-channel mode), 113-128 (16-channel mode) MUX8
Reset
Clock
Micro-6000 Datalogger
U5
C4
Multiplexer #1
MUX1
Reset
Clock
Reset
Reset
Clock
Clock
Multiplexer #2
Multiplexer #3
H
J H H J J
VW Gages 1-32
VW Gages 65-96
VW Gages 33-64 H J H
J
IN
IN
IN
OUT
OUT
SW1 SETTING:
OFF OFF OFF
SW1 SETTING:
OFF OFF ON
SW1 SETTING:
OFF ON OFF
Table 19 - Daisy-Chain Operation/Channels Accessed
Figure 35 is a schematic representations of the daisy-chain configuration with multiplexers configured for 32-channels. It shows three multiplexers sharing the same control ports, and a single cable is used to interconnect them.
2 3 4
MUX2
2 3 4
MUX3
2 3 4
Figure 35 - Daisy-Chain Configuration (Does not apply to Model 8600-2)
35

F.7 Maximum 8032-5 (Tan Cable) Cable Lengths

The 8032 Multiplexer is a low power device, that when combined with a 6000 Datalogger can be physically located at a considerable distance from that Datalogger.
Still, there are limits to the maximum distance – mostly due to the voltage dropped by the 8032-5 MUX cable over its length. Factors such as ambient temperature, number of 8032 Multiplexers (Daisy-Chain configuration) and system battery voltage need to be considered in determining the maximum 8032-5 cable length.
For a single 8032 Multiplexer under normal operating conditions (System Battery = 12V, ambient temperature = 20 °C), the maximum recommended cable length from the Datalogger to the 8032 Multiplexer is 4588 feet (1400m). Figure 36 displays the maximum recommended cable length for various Datalogger, Battery Voltage and ambient Temperature combinations.
Feet
Figure 36 - Recommended Maximum Cable Length (Does not apply to 8600-2)
12V @ 20C
9.6V @ 20C
12V @ 50C
9.6V @ 50C
# of 8032 MUX's
36

APPENDIX G. CELLULAR MODEM COMMUNICATIONS

G.1 Supply

As there are many providers, and options, available with respect to the choice of cellular modems, Geokon requires that the customer choose his/her modem, and provider of choice (or to add another modem to an existing account).
For customers with little or no experience sourcing cellular modems, Geokon suggests one of the following models, available from Sierra Wireless Intelligent Gateways: (A list of modem distributors can be found on the Sierra Wireless homepage:
http://www.sierrawireless.com)
RV50: The RV50 modem is the lowest power Modem available from Sierra Wireless and is compatible with CDMA, GSM and International Networks according to SIM card supplied by user. This modem is recommended in Solar Applications.
Raven XT: The Raven XT modem is available for Verizon Wireless only.
LS300: The LS300 modem can be used with AT&T when Verizon service is unavailable.
FX400: The FX400 series modems are preferred for International applications.
In most static applications, a 250 MB data plan will be sufficient to program the Datalogger, communicate regularly and collect the Text based data files. An Unrestricted Static IP address is recommended for most applications. When required, a Dynamic IP address can be used with a free IP Manager Service such as provided by Sierra Wireless (Airlink).
For standard, transparent RS-232 Communications, a template must be loaded after the modem is provisioned. The Modem supplier will load this template for you.
Verizon Raven XT: http://www.geokon.com/content/software/Raven_XT_Verizon_CSI.zip AT&T LS300: http://www.geokon.com/content/software/LS300G_ATT_CSI.zip
RV50: Please contact Geokon Directly depending on Network
For customers with little or no experience configuring cellular modems, Geokon suggests contacting one of the following companies, who can provide and provision modems for use in North America:
Access Wireless Data Solutions
http://www.accesswds.com orders@accesswds.com
(813) 751-2039
Industrial Networking Solutions
http://www.industrialnetworking.com
(972) 248-7466
If there are any doubts, concerns, or questions with respect to any of the above, please contact Geokon. Any communication and or programming issues related to the operation of your Datalogger will be fully supported by Geokon.
37

G.2 Install

Cellular modem install kits can be provided upon request. These install kits can accommodate most cellular modems and whip or high gain antenna options. A mounting bracket, antenna bulkhead / cable and RS-232 null modem adapter are preinstalled for plug and play of the antenna and customer supplied modem.
1) Modem is secured in mounting bracket using the Velcro mounting strap
2) SMA bulkhead cable is secured to modem SMA connector and antenna secured on outside of enclosure.
3) Modem RS-232 port is connected through a Null Modem cable to the Datalogger CPI port using the provided CPI to RS-232 Null Modem cable.
4) The power supply cable provided with the modem is connected to an empty Load terminal on the Power Distribution PCB board or directly to 12 V and G on the CR6 module. Switch 12 VDC power (SW12) can also be used to control the modem power at a given interval (programming required).
NOTE: Verify power cable wiring in the cellular modem manufacturer provided user manual.
When using a Sierra Wireless Raven XT modem, the red wire is 12 VDC and Black is Ground (white is not used). When using the Sierra Wireless LS-300 or FX400 series, the Red and White wires are both tied to +12 VDC and Black is Ground.
Figure 37 - Model 8600-2 with Cellular Modem Installed
38

G.3 Communications

To add a communication port based off the Geokon Cell Modem Template, press the “Add Root” button and select a device. Continue selecting devices from the Add Box as shown until the communication link to the Datalogger is represented. Once the Datalogger (CR6Series) is added to the Network Map, press “Apply”. Select “IPPort” (Figure 38) and then Select “PakBusPort (PakBus Loggers)” (Figure 39)
Figure 38 - Add Communications Port (Root)
Figure 39 - Select “PakBusPort”
Select “CR6Series”. When finished, press the “Apply” button at the bottom left of the Setup Screen.
Figure 40 - Select Device to Add
While still within the Setup Screen, IP Address must be entered. The Cell Modem IP Address must be entered as:
XXX.XXX.XXX.XXX:3001
3001 represents the default port of the Cell Modem
39
Figure 41 - Enter Internet IP Address
When finished, press the “Apply” button at the bottom left of the Setup Screen. Minimize or close the Setup Screen and go back to the LoggerNet Launch Menu.
Loading...