Geokon, Inc. warrants its products to be free of defects in materials and workmanship, under
normal use and service for a period of 13 months from date of purchase. If the unit should
malfunction, it must be returned to the factory for evaluation, freight prepaid. Upon examination
by Geokon, if the unit is found to be defective, it will be repaired or replaced at no charge.
However, the WARRANTY is VOID if the unit shows evidence of having been tampered with
or shows evidence of being damaged as a result of excessive corrosion or current, heat, moisture
or vibration, improper specification, misapplication, misuse or other operating conditions outside
of Geokon's control. Components which wear or which are damaged by misuse are not
warranted. This includes fuses and batteries.
Geokon manufactures scientific instruments whose misuse is potentially dangerous. The
instruments are intended to be installed and used only by qualified personnel. There are no
warranties except as stated herein. There are no other warranties, expressed or implied, including
but not limited to the implied warranties of merchantability and of fitness for a particular
purpose. Geokon, Inc. is not responsible for any damages or losses caused to other equipment,
whether direct, indirect, incidental, special or consequential which the purchaser may experience
as a result of the installation or use of the product. The buyer's sole remedy for any breach of this
agreement by Geokon, Inc. or any breach of any warranty by Geokon, Inc. shall not exceed the
purchase price paid by the purchaser to Geokon, Inc. for the unit or units, or equipment directly
affected by such breach. Under no circumstances will Geokon reimburse the claimant for loss
incurred in removing and/or reinstalling equipment.
Every precaution for accuracy has been taken in the preparation of manuals and/or software,
however, Geokon, Inc. neither assumes responsibility for any omissions or errors that may
appear nor assumes liability for any damages or losses that result from the use of the products in
accordance with the information contained in the manual or software.
2. INSTALLING AND CONFIGURING THE 8020-59 SOFTWARE APPLICATION ..................................... 2
CONFIGURING THE COMPORT ........................................................................................................................... 5
2.1
3. QUICK STAR T (SINGLE CHANNEL) ............................................................................................................... 7
4. SINGLE AND MULTI-CHANNEL OPERATION ............................................................................................10
4.1
THEORY OF OPERATION .....................................................................................................................................10
4.2.1 Single Channel Mode ................................................................................................................................11
4.2.2 16-Channel Multiplex ed M ode ..................................................................................................................12
4.2.3 32-Channel Multiplex ed M ode ..................................................................................................................13
CONVERSION TO “DIGITS” .................................................................................................................................14
4.3
CONNECTION TO A VOLTAGE INPUT DAS ..........................................................................................................15
4.4
CONNECTION TO A CURRENT INPUT DAS ..........................................................................................................15
4.5
TEMPERATURE MEASUREMENT .........................................................................................................................16
COMMAND LINE INTERFACE ..............................................................................................................................17
6. 8020-59 WINDOWS APPLICA TION ..................................................................................................................20
FIGURE 30-CONNECTION EXAMPLE FOR ONE CHANNEL.8020-59 WITH VOLTAGE OUTPUT,8020-59PLC AND PLC
DAS ................................................................................................................................................................... 35
FIGURE 31-CONNECTION EXAMPLE FOR 16-CHANNEL MULTIPLEXER,8020-59 W/VOLTAGE OUTPUT,8020-59PLC
FIGURE 33-CONNECTION EXAMPLE FOR ONE CHANNEL 8020-59 WITH CURRENT OUTPUT,8020-59PLC AND PLC
DAS ................................................................................................................................................................... 38
FIGURE 34-CONNECTION EXAMPLE FOR 16-CHANNEL MULTIPLEXER,8020-59 W/CURRENT OUTPUT,8020-59PLC
The Model 8020-59 Vibrating Wire (VW) to Analog Converter is a low cost module that
provides a simple way to connect Geokon’s vibrating wire transducers to data acquisition
systems that are not capable of reading frequency signals nor able to generate the proper signals
required to excite VW transducers. The converter can operate with single transducers, as a standalone device, or with multiple transducers in conjunction with the Geokon Model 8032
Multiplexer. The converter is powered using either a 12V or a 24V supply.
The 8020-59 interface is capable of outputting a 0-5V or 4-20mA signal that is directly
proportional to “digits”, a vibrating wire sensor’s native units. These outputs are easily converted
to digits via several simple formulas (see Section 4.3). The analog outputs are automatically
scaled to minimum and maximum parameters (defaults are 0 and 25,000) and sloped to provide
their complete 0-5V or 4-20mA range for each individual transducer. These analog outputs offer
20-bit resolution (one part in 1,048,576) with an accuracy of better than 0.1% of full scale span
(0-5V output) throughout the operating temperature range (-20 ºC min, +80 ºC max). In a similar
fashion, the temperature reading from the transducer’s thermistor is available with 10-bit
resolution (one part in 1024). When configured for single-channel operation, the 8020-59
updates the reading approximately once every second.
1
Signal Description
5V indicates conversion complete
Single Channel Mode: 0V = VW output (digits)
5V = Temperature output (˚C)
16/32-Channel Mode: 0V = Conversion Disabled
Single Channel Mode: 0V = 8020-59 ON
5V = 8020-59 OFF
16-Channel Mode: 0 → 5V transition =
increment channel and select VW Analog or Temperature Analog
Table 1 - Connector Pinout and Signal Description
Page 8
2
2. INSTALLING AND CONFIGURING THE 8 0 20-59 SOFTWARE
APPLICATION
A free setup and configuration application is available for download on Geokon’s website,
providing a user-friendly way to set up the 8020-59 VW to Analog Converter. Follow the steps
below to install the software:
1) After downloading the installer from the Geokon website, right-click on the zip file, “8020-
59_setup.zip”, and select “Extract All…” from the resulting popup menu (see Figure 1).
Figure 1 - Extracting the 8020-59 Installer
2) Another dialog will be displayed showing the editable default destination folder for the
extracted contents (see Figure 2). When satisfied with the destination folder, click on the
Extract button to extract the contents of the zipped installer (see Figure 2).
Figure 2 - Select Destination Folder
3) Figure 3 on the following page shows the extracted installer in the destination folder. Double
click on the file, “setup.exe”, to start the install process.
Page 9
3
Figure 3 - Extracted 8020-59 Installer
4) After a few moments, the dialog shown in Figure 4 will be displayed.
Figure 4 - Install Wizard, Start Screen
5) After clicking on the “Next >” button, the dialog shown in Figure 5 is displayed.
Figure 5 - Install Wizard, Choose Users
Page 10
4
6) Choose whether to install the 8020-59 Software Application for all users or just for the
current user (see Figure 5). NOTE: Without administrative privileges on the PC, the only
option allowed will be “Install just for me”. After making a selection, click “Next >”; this
will bring up the dialog shown in Figure 6.
Figure 6 - Install Wizard, Choose Start Menu Folder
7) The dialog shown in Figure 6 allows the installer to choose where the “shortcut” to launch
the application will be located – the default is the “Geokon” folder. A “shortcut” will also be
located on the desktop. Clicking on “Install” begins the process of copying files and causes
the dialog box shown in Figure 7 to be displayed:
Figure 7 - Install Wizard, Installation Complete
Page 11
8) The installation process is now complete. Clicking “Next >” allows the opportunity to launch
the 8020-59 application when finished by checking the box next to “Launch 8020-59
Application” (see Figure 8). Click on “Finish” (see below) to close the install wizard.
After launching the 8020-59 Software Application, the first thing that must be done is to
configure the COM Port. Click on the “Com Setup” (or press the “Alt” and the “c” key at the
same time) to select a COM port for communication with the 8020-59 VW to Analog Converter
(See Figure 9).
Select a COM port that corresponds to an available RS-232 communication port, whether it is
built-in (such as COM1) or an external one via a USB to Serial Converter (see Figure 10). Some
newer PCs are so fast that a small delay between characters is needed to give the 8020-59 time to
process the characters. The “Inter-character Delay” is specified in milliseconds.
Figure 10 - Communications Parameters Dialog
Page 13
3. QUICK START (SINGLE CHANNEL)
In order to properly set up and use the 8020-59 VW to Analog Converter, a field “Zero Reading”
will need to be taken. This is required so that the 0-5V and 4-20mA outputs are scaled correctly
for the transducer being monitored. This section will allow the user to quickly set up and obtain
valid readings from the 8020-59 Converter:
Equipment Required:
Personal Computer (PC) with RS-232 COM port
8020-59 Software Application
RS-232 Cable (supplied)
VW Transducer
VW Transducer Calibration Report
1) Install the 8020-59 software on the PC that will be used to interface to the 8020-59 unit (see
Section 2).
2) Connect the 8020-59 to the computer’s serial port (typically COM1). If using an external
USB to Serial converter, drivers for this device will need to be installed before this step (see
Section 2.1).
3) Connect +12VDC to connector pin six OR +24VDC to connector pin three (see Table 1 in
Section 1).
4) Connect the power supply return (or Ground) to connector pin seven OR, if using a 24V
supply, connector pin four, 24V GND (see Table 1).
5) Connect the un-pressurized VW transducer to the 8020-59 (see Table 1):
RED wire to connector pin 10 (C+).
BLACK wire to connector pin 11 (C-).
WHITE wire to connector pin 8 (TH+).
GREEN wire to connector pin 7 (TH-)
SHIELD wire to connector pin 1 (GND).
6) Start the 8020-59 program.
7) Click the “Single Channel” button (or press the “Alt” and “s” keys) from the program’s
Startup Dialog (see Section 2.1).
8) The Single Channel Configuration screen should now be displayed (see Figure 11 on the
following page). This screen is used to define the parameters that the 8020-59 uses to read
the attached transducer.
7
Page 14
8
Figure 11 - Single Channel Config ur ation Screen
9) Ensure that the Transducer Model drop-down control is correctly set to the proper VW model
and that the “Use Defaults” checkbox is checked.
10) Click “Activate/Monitor” (or press the “Alt” and “m” keys). The “Upload Confirmation”
dialog will be displayed with the prompt: “Do you want to upload the current settings to the
8020-59?”. Click the “Yes” button. The Single Channel Monitor screen should now appear
(see Figure 12), and after several seconds should update with a reading.
Figure 12 - Single Channel Monitor
Page 15
11) Note the digits that are displayed. Record the VW reading, barometric pressure, and
temperature. This is what is known as the “initial zero” reading. The “initial zero” reading
will be used in future calculations. For example, the VW zero reading will be the R0 value in
the pressure calcul ati on s equation: P = (R1 – R0) x GF, where R1 equals the “Current
Reading” and GF is the Gage Factor obtained from the calibration report for the sensor being
measured. See Appendix D for further information on Data Reduction.
12) Click “Exit” (or press the “Alt” and “e” keys) to return to the Setup Screen.
After setup is complete, the PC is disconnected from the 8020-59 and the sensor is in a
pressurized state, the “Current Reading” (R1) “Digits” will need to be calculated from the value
read from the analog voltage (V OUT, in volts) or current (I-OUT, in milliamperes) output. See
Section 4.3 for more information on how to convert from an analog voltage or current value into
digits. Once a “Digits” value has been obtained, translation into other engineering units is
accomplished by multiplying the “Digits” value by the Gage Factor for the particular sensor
read.
9
Page 16
10
4. SINGLE AND MULTI-CHANNEL OPERATION
In addition to working as a stand-alone device to interface a single transducer to a Data
Acquisition System (DAS), multiple channels may be configured in similar fashion to the Single
Channel. The 8020-59 works in conjunction with the 8032 Multiplexer and the user’s Data
Acquisition System to select either 1 of 16 or 1 of 32 transducers.
The 8020-59 was developed with multiplexing in mind and uses the same control signals as the
Geokon model 8032 Vibrating Wire Multiplexer (for control signal timing requirements, refer to
the model 8032 Multiplexer Instruction Manual). With a simple control program executing on
the host DAS, it is capable of working with 16 vibrating wire transducers (with thermistors), or
32 transducers (without thermistors). All setup parameters for each channel are stored in internal
EEPROM memory, requiring no backup battery and can be retained for years.
The 8020-59 is powered from either a 12V or 24V supply (nominal), and draws approximately
70mA when taking readings. When it is “shut-off”, or between scans when multiplexing, the
total current consumption of the 8020-59 is less than 10μA (12V) or 16.5mA (24V).
The 8020-59 setup and configuration utility runs on a PC operating under the Windows operating
system and communicates to the 8020-59 via the 8020-59’s RS-232 port. A “Command Line”
interface is provided in the 8020-59 internal firmware, with a command set that allows all
functions to be easily set up and calibrated.
4.1 Theory of Operation
The 8020-59 provides excitation for a VW transducer using a swept frequency pluck that is
optimized for the selected gage type. The period of the resulting VW signal is measured using a
25 MHz clock. This measurement is performed 512 times, resulting in an average period being
stored in internal memory. The microprocessor then converts the average period into digits and
scales the output accordingly, for the specific transducer being read. This information is then sent
to a 20-bit D/A converter, which outputs a voltage (0-5V) that is directly proportional to the
average digits. Post-processing circuitry converts this voltage to a non-isolated 4-20mA current
signal. In addition, the 20-bit D/A is put through a self-calibration routine at the start of each
measurement cycle to minimize errors due to temperature and power supply variations. The
resulting 0-5V and 4-20mA signals are brought to the output connector for connection to the
input(s) of the host DAS.
Page 17
4.2 Operating Modes
The 8020-59 has three modes of operation: single channel mode, 16-Channel mode, and 32Channel mode. Each operation mode is detailed in the subsections below.
4.2.1 Single Channel Mode
In this mode (Figure 13), the 8020-59 will maintain continuous excitation of the VW
transducer and provide a continuous 0-5V and 4-20mA output to the host DAS, updated
approximately every second.
By controlling the ENABLE input to the 8020-59, the 8020-59 voltage and current
outputs (VOUT and IOUT) will either be proportional to the transducer’s digits
(ENABLE=0V – or disconnected) or the transducer’s temperature in degrees Centigrade
(ENABLE=5V).
By controlling the CLOCK input to the 8020-59, the 8020-59 will either be ON
(CLOCK=0V − or disconnected) or OFF (CLOCK=5V).
11
Figure 13 - Control Signal Sequence: Single Channel Mode
Page 18
12
4.2.2 16-Channel Multiplexed Mode
In this mode (Figure 14), the 8020-59 will excite and provide output only when told to do
so from the host DAS. As soon as the 8020-59 is powered, if ENABLE and CLOCK are
low (0V), the 8020-59 will go to “sleep” and will wait for commands from the host DAS,
drawing less than 10μA from the +12V system power supply (16.5mA if using a +24V
system power supply).
When a reading is to be taken, the host system first brings ENABLE high (5V) to enable
the 8020-59 and then CLOCK high (5V) to activate channel 1 of 16 (for control signal
timing requirements, refer to the model 8032 Multiplexer Instruction Manual). The 802059 will “wake up” and read the transducer connected to channel one of the multiplexer.
When the reading is ready (approximately one to two seconds later), the 8020-59 will
bring the VALID output high (5V), indicating to the host DAS that the voltage and
current outputs (proportional to the digits) are at VOUT and IOUT and are ready to be
acquired.
Once the reading is acquired, the host system brings CLOCK low (0V) and then high
(5V) again to read the temperature of the transducer. When this reading is ready (<100ms
later), the VALID output will again go high (5V), indicating that the voltage and current
outputs (now proportional to temperature) are at VOUT and IOUT and are ready to be
acquired. When the host DAS again brings CLOCK low (0V) and then HIGH (5V), the
8020-59 and multiplexer increment to channel 2 of 16 and this reading sequence starts
over again. This reading sequence may be used for up to 16 transducers.
When the final transducer has been read, the host DAS brings CLOCK and ENABLE
both low (0V) to reset the system. The 8020-59 will go back to sleep until it is time for
the next reading. Using this feature, up to 16 Vibrating Wire transducers with thermistors
may be multiplexed into a single DAS.
Figure 14 - Control Signal Sequence: 16-Channel Mod e (Channels One and Two Only)
Page 19
13
4.2.3 32-Channel Multiplexed Mode
In this mode (Figure 15), the 8020-59 will also excite and provide output only when told
to do so from the host DAS. As soon as the 8020-59 is powered, if ENABLE and
CLOCK are low (0V), the 8020-59 will go to “sleep” and will wait for commands from
the host DAS, drawing less than 10μA from the +12V system power supply (16.5mA if
using a +24V system power supply).
When a reading is to be taken, the host system first brings ENABLE high (5V) to enable
the 8020-59 and then CLOCK high (5V) to activate channel 1 of 32 (for control signal
timing requirements, refer to the model 8032 Multiplexer Instruction Manual). The 802059 will “wake up” and read the transducer connected to channel one of the multiplexer.
When the reading is ready (approximately one to two seconds later (typ)), the 8020-59
will bring the VALID output high (5V), indicating to the host DAS that the voltage and
current outputs (proportional to the digits of transducer number one) are at VOUT and
IOUT and are ready to be acquired.
Once the reading is acquired, the host system brings CLOCK low (0V), and then high
(5V) again to increment the 8020-59 and multiplexer to the next transducer. When this
reading is ready ( approximately one to two seconds later), the VALID output will again
go high (5V), indicating that the voltage and current outputs (proportional to the digits of
transducer number two) are at VOUT and IOUT and are ready to be acquired. This
reading sequence may be used for up to 32 transducers.
When the final transducer has been read, the host DAS brings CLOCK and ENABLE
both low (0V) to reset the system. The 8020-59 will go back to sleep until it is time for
the next reading. Using this feature, up to 32 Vibrating Wire transducers may be
multiplexed into a single DAS.
Figure 15 - Control Signal Sequence: 32-Channel Mode (Channels One through Four Only)
Page 20
14
4.3 Conversion to “Digits”
A useful frequency related unit that is directly proportional to pressure is the “digit”. These units
are related to vibrating wire frequency as:
Digits = [Frequency (Hz)]² x 0.001
Equation 1 - Vibrating Wire Frequency to Digits Conversion
The 8020-59 provides analog outputs that are directly proportional to digits.
Referring to the sample calibration in Figure 36 of Appendix E, a value in digits can be derived
from the voltage and current outputs as described below.
For the following “positive slope” transducers: 4000, 4100, 4150, 4200, 4210, 4300, 4350, 4360,
4400, 4420, 4425, 4450, 4650 and 4700; use Equation 2 for Analog Voltage Output (assumes
Vout is in volts (V)), or Equation 3 for Analog Current (4-20mA) Output (assumes Iout is in
milliamperes (mA).
For the following “negative slope” transducers: 4500, 4600, 4675, 4800 and 4900; use Equation
4 for Analog Voltage Output (assumes Vout is in Volts (V)), or Equation 5 for Analog Current
(4-20mA) Output (assumes Iout is in milliamperes (mA)).
Digits = Maximum Limit – (Vout * (Digits Span/5V))
The 8020-59 outputs a voltage (V OUT, 0-5V) that is proportional to the digits of the transducer
being read (ENABLE = 0V (or disconnected) in Single Channel Mode, after the first of two
CLOCKS in 16-Channel Mode, or after each CLOCK in 32-Channel Mode). The 8020-59 offers
a digits to voltage output with 20-bit resolution (one part in 1,048,576) of the span of the
transducer (default span is 25,000 digits). This output voltage can be input directly to any Data
Acquisition System capable of reading a 0-5V analog voltage.
The 8020-59 also outputs a voltage proportional to the temperature (also on V OUT) of the
transducer being read (ENABLE = 5V in Single Channel Mode or after the second of two
CLOCKS in 16-Channel Mode). The 8020-59 offers a temperature voltage output with 10-bit
resolution (one part in 1024) over a 100º C span (-20º C to +80º C). See Section 4.6 for specific
calculations.
4.5 Connection to a Current Input DAS
The 8020-59 sources a current (I-OUT, 4-20mA) that is proportional to the digits of the
transducer being read (ENABLE = 0V (or disconnected) in Single Channel Mode, after the first
of two CLOCKS in 16-Channel Mode, or after each CLOCK in 32-Channel Mode). The 8020-59
offers a digits current output with 20-bit resolution (one part in 1,048,576) of the span of the
transducer (default span is 25,000 digits). This output current can be input directly to any Data
Acquisition System capable of reading a non-isolated 4-20mA analog current.
The 8020-59 also outputs a current proportional to the temperature (also on I-OUT) of the
transducer being read (ENABLE = 5V in Single Channel Mode or after the second of two
CLOCKS in 16-Channel Mode). The 8020-59 offers a temperature current output with 10-bit
resolution (one part in 1024) over a 100º C span (-20º C to +80º C). See Section 4.6 for specific
calculations.
See Appendix C for the various voltage and current measurement options.
Page 22
16
4.6 Temperature Measurement
The 8020-59 is capable of reading the thermistor that is part of a VW sensor. The 8020-59
outputs this reading as a voltage (0-5V) and current (4-20mA) that are proportional to
temperature whenever ENABLE is high (5V), in Single Channel Mode, or while the second
clock pulse is high, in 16-Channel mode. See Figure 14 for 16-Channel mode control signal
timing.
Temperature is derived from the voltage output by Equation 6.
Note: Assumes Vout is in volts (V). Minimum Temperature = -20 °C, Degree Span = 100°.
°C = Minimum Temperature + (Vout * (De gree Span / 5V)) =
-20° + (20 * Vout)
Equation 6 - Temperature Derived from Voltage Output
Temperature is derived from the current output by Equation 7.
Note: Assumes Iout is in milliamperes (mA).
° C = Minimum Temperature + ((Iout - 4mA) * (Degree Span / 16mA)) =
-20° + (6.25 * (Iout - 4mA))
Equation 7 - Temperature Derived from Current Output
Page 23
17
5. COMMUNICATIONS
The 8020-59 offers a standard 9-pin RS-232 Serial Port for connection to most desktop and
laptop computers (an optional USB to Serial interface adapter is available).
If using a terminal emulator program such HyperTerminal, Putty, etc., for command line set up
(see Section 5.1) of the 8020-59, configure the program’s communication parameters as follows:
Port: Serial port that 8020-59 is connected to (i.e., COM1, COM2)
Bits per Second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow Control: None
5.1 Command Line Interface
When connected to a computer using a terminal emulator program such as HyperTerminal,
typing <Enter> returns the title screen:
Geokon, Inc
8020-59 VW to Analog Interface
Rev 4.0.2 12/05/2011
Type ? for Help
*
Typing ? <Return> returns the Help Screen:
*?
COMMAND DESCRIPTION
C1 Single VW and Thermistor
C2 16 VW's and 16 Thermistors
C3 32 VW's
S Display 8020-59 Status
R Display Firmware Revision
Z Shutdown and Sleep
<Esc> Exit and Start Taking Readings (C1 only)
*
Commands C1, C2 and C3 configure the 8020-59 for single channel, 16-Channel multiplexing or
32-Channel multiplexing, respectively.
Note: The default value for the lower limit is zero, so for cases where this is acceptable, there is
no need to explicitly set the lower limit.
Typing “D” and the channel # <Enter> displays the limits and slope for that channel. Typing
“D0” <ENTER> displays the limits and slope for all channels.
Typing “E” and the channel # <Enter> ENABLES that channel (16 and 32-Channel mode only)
when multiplexing. Typing “E0” <ENTER> enables all channels.
Commands ME and MD enable and disable the readings monitor.
To set the maximum digits, slope and gage type for each transducer, type “U” and the channel #
(comma) maximum digits (comma) slope (+/-) (comma) gage type <Enter>
Example: *U1,10590,-,1 <Enter>
To set the minimum digits type “L” and the channel # (comma) minimum digits <Enter>
Example: *L1,9990 <Enter>
Command RD disabled the VALID line from activating when a non-successful vibrating wire
sensor conversion takes place. A high level on the VALID line would then indicate that a
successful conversion took place.
Command RE enables the VALID line to go “Ready” (high) after each CLOCK pulse while the
ENABLE line is high. The VALID output then indicates that the CLOCK and ENABLE
sequencing was proper but does not reflect the success of the vibrating wire sensor conversion.
The default state for the 8020-59 is READY ENABLED.
Typing “N” and the channel # <Enter> DISABLES that channel (16 and 32-Channel mode only)
when multiplexing. Typing “N0” <ENTER> disables all channels.
Typing “X4” <Enter> will allow calibration of the 4mA set point.
Typing “X12” <Enter> will allow calibration of the 12mA set point.
Typing “X20” <Enter> will allow calibration of the 20mA set point.
Typing “R” <Enter> will return the firmware revision.
Typing “Z” <Enter> will put the 8020-59 into low power sleep mode (off), drawing less than
10μA from the 12V power source (less than 16.5mA from the 24V power source).
Finally, pressing the <Esc> key (in C1 single channel mode only) will start the readings screen
with the 8020-59 outputting 0-5V and 4-20mA:
Where;
Comma delimited field one is the channel number, followed by the reading in digits, then the
transducer temperature in deg. C, the slope, and finally the gage type. The pound sign (#)
signifies the end of the data string.
19
Page 26
20
6. 8020-59 WINDOWS APPLICATION
All of the commands described in the Command Line Interface Section 5.1 of this manual are
included in a simple Windows program designed to set up, control and monitor the 8020-59.
This software is provided as a convenient way to set and store the various parameters associated
with each transducer, and to monitor the digits and temperature output of each transducer. All
settings such as communications parameters, channel enable, individual transducer type, upper
limits (digits), lower limits (digits), and transducer model are set using a simple and intuitive
user interface.
6.1 Software Installation
Included with the Model 8020-59 VW to Analog Converter is a CD-ROM containing the 802059 application. For more details about installing and configuring the application, please see
Section 2.
6.2 Starting the 8020-59 Application
To start the 8020-59 application, select Start/Programs/8020_59/8020_59. This brings up the
Startup dialog (see Figure 9).
From the Startup dialog, the major functions (Mode, Communications, and Calibration) of the
8020-59 can be accessed. Click Com Setup to select the Com Port that the 8020-59 is connected
to (see Section 2.1 and Figure 10).
After the communication port that the 8020-59 is attached to is selected, click Save/Apply to
return to the Startup dialog. The communication port selected will be saved to a configuration
file and will not need to be set again unless the communication port changes.
6.3 Communications Timeout Error
If communications to the 8020-59 are interrupted for any reason, after approximately 15 seconds
an error message will be displayed with a descriptive error status message in bold font (see
Figure 16).
Figure 16 - Communication Timeout Warning Message
NOTE: In Single Channel Mode, the ENABLE line may indeed be active although the PC
should still be able to communicate with 8020-59 VW to Analog Converter.
Page 27
21
6.4 Single Channel Configuration
If the 8020-59 is to be used with only one transducer, the 8020-59 Mode should be set to Single
Channel. Click “Single Channel” (or press the “Alt” and “s” keys) to display the Single Channel
Configuration window (see Figure 11).
From this window, all of the transducer information may be entered. The transducer limits,
transducer type, and slope are stored in non-volatile EEPROM internal to the 8020-59. All other
information is stored to a configuration file on the host computer’s hard drive, and loaded
automatically when this screen is displayed. Configurations may also be saved by using the
“Save Info” button and opened later by using the “Open Info” button. Opened configurations
must be sent to the 8020-59 by using the “Update Info” button to take effect.
From the Transducer Model drop-down list box, select the model of the transducer being used.
This selection determines the transducer type/slope and these parameters will be set with no
further interaction from the user. Enter the transducer’s limits (in digits) in the text boxes
labeled, Maximum (digits) and Minimum (digits). If desired, enter a Label for the transducer.
NOTE: Minimum and maximum values can be obtained from the transducer’s calibration report
by taking the lowest average reading for the minimum value and the highest average reading for
the maximum value. Another option is to click on “Use Defaults” checkbox in the Limits section
of the setup dialog. As of firmware revision 4.0.0, the default minimum and maximum values are
0 and 25,000, respectively. With its 20-bit Digital to Analog Converter, the 8020-59 has more
than enough resolution to allow the entire voltage or current range to be divided by 25,000. This
can make setup easier than previous firmware revisions.
When complete, click ‘Update Info’ to store this information to the 8020-59. (Note: to
determine the information settings of the currently connected 8020-59, click “Retrieve Info”)
Clicking “Activate/Monitor” (or press the “Alt” and “m” keys) starts the 8020-59 transducer
excitation and reading output functions and displays the Single Channel Monitor screen (see
Figure 12). This screen displays the transducer information along with the reading in digits or
microstrain (μE), the calculated reading voltage, the calculated reading current, the temperature
of the transducer, and the calculated voltage and current that is representative of the temperature.
The screen updates approximately every three to five seconds, providing an almost real-time
display of the transducer output.
NOTE: In Single Channel Configuration, the user selects whether the actual voltage and current
outputs represent the VW reading or the VW temperature by toggling the ENABLE control line
(5V=VW temperature, 0V=VW reading). When left unconnected, the physical voltage and
current outputs default to the VW reading.
If monitoring of the 8020-59 outputs is not of interest, then click Activate/Exit from the
Configuration Window to begin the transducer excitation and reading and return to the Main
Menu.
Page 28
22
6.5 16-Channel Configuration
The 8020-59 can be used in conjunction with the Geokon model 8032 Multiplexer to configure
and monitor up to 16 vibrating wire transducers with thermistors. For wiring details of this
configuration, see Appendix C. The 8020-59 Mode should be set to 16-Channel. Clicking “16-Channel” (or press the “Alt” and “1” keys) from the Startup dialog displays the 16-Channel
Configuration window:
Figure 17 - 16-Channel Configuration Screen
From this window, all of the transducer information for each of the 16-Channels may be entered.
The transducer type, slope, and limits are stored in non-volatile EEPROM internal to the 8020-
59. All other information is stored to a configuration file on the host computer’s hard drive, and
loaded automatically when this screen is displayed. Configurations may also be saved by
clicking the “Save Info” button and may be recalled later by clicking the “Open Info” button.
Page 29
23
From the Channel Selection section of the 16-Channel Configuration screen (see Figure 17),
select the transducer channel to be configured by clicking on the up arrow (▲) or down
(▼)arrow. Alternately, a number from one to 16 can be entered directly. Check “Enable” if the
channel is to be read, otherwise uncheck “Enable”. The circle in the upper right corner indicates
if the channel is currently enabled or not. Green is enabled, and red is disabled.
In the Transducer Information section, select the model of the transducer being used from the
“Transducer Model” drop-down list box. This selection determines the transducer type/slope
and these parameters will be set with no further interaction from the user. If desired, enter a
“Label” for the transducer.
In the Transducer Limits section, the current minimum and maximum settings for the channels
stored in the 8020-59’s on-board EEPROM are in the gray fields on the right. New values in the
configuration are in the white fields on the left. Enter the transducer’s limits (in digits) from the
transducer’s calibration report at “Maximum (digits)” and “Minimum (digits)”. Alternately,
clicking on the “Use Defaults” checkbox will set the minimum and maximum limits to 0 and
25,000, respectively.
When complete, click “Update Info” to store this information to the 8020-59. To update all of
the channels with the current configuration click “Update All Info”. To retrieve the info from
the 8020-59’s EEPROM for the current configuration, click “Retrieve Info” for the current
channel and “Retrieve All Info” for all channels. The “Copy” button may be used to copy the
current channels settings, and the “Paste” button may be used to write the current copied settings
into the currently selected channel. If the “Increment Channel Selection Before Paste”
checkbox is checked, the “Paste” button may be clicked repeatedly to rapidly set the parameters
of all channels to the same values. The “Label” textbox will automatically populate with channel
appropriate text.
Clicking “Activate/Monitor” puts the 8020-59 into low power Standby (Sleep) mode, waiting
for the proper control signals on the ENABLE and CLOCK inputs (refer to Section 4.2.2,16Channel Multiplexed Mode and Figure 22 in Appendix A.2, for signal timing details) and
displays the Monitor screen (see Figure 18).
Figure 18 - 16-Channel Monitor Screen
Page 30
24
The 16-Channel Monitor Screen (see Figure 18) displays the readings and temperatures of any
enabled transducers. Any under-range or over-range conditions are highlighted in red and the
channels that are currently being read is highlighted in green and yellow. (Note: Any channels
that are disabled will be “grayed-out”. Any channels that are enabled but do not have a
transducer connected will display 99999 and be highlighted in red). Hovering the mouse in the
lower-left corner of any channel’s frame (border) displays the tooltip: “Double-click to expand”.
Double-clicking where indicated activates the reading screen (Figure 19) for that channel:
Figure 19 - Channel One Monitor
This screen displays the transducer information along with the reading in digits or
microstrain (μE), the calculated reading voltage, the calculated reading current, the
temperature of the transducer, and the calculated voltage and current that is representative of
the temperature. The screen updates based on ENABLE and CLOCK timing from the Data
Acquisition System.
If monitoring the 8020-59 outputs is not of interest, then click Activate/Exit from the
Configuration Window to put the 8020-59 into low power Standby (Sleep) mode, waiting
for the proper control input signals (refer to Section 4.2.2, for signal timing details).
Page 31
25
6.6 32-Channel Configuration
32-Channel Configuration and Monitor modes work identically to their 16-Channel counterparts
with the only difference being that there are no temperature measurements (see Figure 20).
Figure 20 - 32-Channel Monitor Screen
Figure 21 - Channel Three Monitor
Page 32
26
Physical:
Dimensions (L x W x H):
111.13 mm, 108.36 mm x 36.53 mm
4.375" x 4.266" x 1.438"
Weight:
0.456 lbs.
Operating Temperature:
-20 to +80 degrees Celsius
Power Requirements:
+12V Input Voltage Range:
9 to 15 VDC (12V nominal)
+24V Input Voltage Range:
18 to 30 VDC (24V nominal)
Operating Current:
90mA (max) at 12V @ 25 ºC
75mA (max) at 24V @ 25 ºC
Table 2 - Thermistor Resistance Versus Temperature
55.6
150
APPENDIX B. THERMISTOR TEMPERATURE DERIVATION
Thermistor Type: YSI 44005, Dale #1C3001-B3, Alpha #13A3001-B3
Resistance to Temperature Equation:
1
T=
A+B(LnR)+C(LnR)
Equation 8 - Resistance to Temperature
Where;
T = Temperature in °C.
LnR = Natural Log of Thermistor Resistance
A = 1.4051 × 10-3
B = 2.369 × 10-4
C = 1.019 × 10-7
Note: Coefficients calculated over the −50 to +150° C. span.
-273.2
3
Page 35
APPENDIX C. WIRING CONFI G URATIONS
C.1 Voltage Output – Single Channel Mode (Digital I/O DAS)
29
Figure 24 - Connection Example for One Channel 8020-59 with Voltage Output to Digital I/O DAS
Page 36
30
C.2 Voltage Output – 16-Channel Mode (Digital I/O DAS)
Figure 25 - Connection Example for 16-Channel Multiplexer, 8020-59 with Voltage Output and Digital I/O DAS
Page 37
C.3 Voltage Output – 32-Channel Mode (Digital I/O DAS)
31
Figure 26 - Connection Example for 32-Channel Multiplexer, 8020-59 with Voltag e Output and Digital I/O DAS
Page 38
32
C.4 Current Output – Single Channel Mode (Digital I/O DAS)
Figure 27 - Connection Example for One Channel 8020-59 with Current Output to Digital I/O DAS
Page 39
C.5 Current Output – 16-Channel Mode (Digital I/O DAS):
33
Figure 28 - Connection Example for 16-Channel Multiplexer, 8020-59 with Current Output and Digital I/O DAS
Page 40
34
C.6 Current Output – 32-Channel Mode (Digital DAS)
Figure 29 - Connection Example for 32-Channel Multiplexer, 8020-59 with Current Output and Digital I/O DAS
Page 41
C.7 Voltage Output – Single Channel Mode (PLC DAS)
35
Figure 30 - Connection Example for One Channel. 8020-59 with Voltage Output, 8020-59 PLC and PLC DAS
Page 42
36
C.8 Voltage Output – 16-Channel Mode (PLC DAS)
Figure 31 - Connection Example for 16-Channel Multiplexer, 8020-59 w/ Voltage Output, 8020-59 PLC and PLC DAS
Page 43
C.9 Voltage Output – 32-Channel Mode (PLC DAS)
37
Figure 32 - Connection Example for 32-Channel Multiplexer, 8020-59 w/ Voltage Output, 8020-59 PLC and PLC DAS
Page 44
38
C.10 Current Output – Single Channel Mode (PLC DAS)
Figure 33 - Connection Example for One Channel 8020-59 with Current Out put, 8020-59 PLC and PLC DAS
Page 45
C.11 Current Output – 16-Channel Mode (PLC DAS)
39
Figure 34 - Connection Example for 16-Channel Multiplexer, 8020-59 w/Current Output, 8020-59 PLC and PLC DAS
Page 46
40
C.12 Current Output – 32-Channel Mode (PLC DAS)
Figure 35 - Connection Example for 32-Channel Multiplexer, 8020-59 w/Current Output, 8020-59 PLC and PLC DAS
Page 47
41
APPENDIX D. DATA REDUCTION
D.1 Pressure Calculation
The digits that are calculated from the voltage and current outputs are based on the equation:
2
Digits =
Period
1
x 10-3
Or
2
Digits=
Equation 9 - Digits Calculation
For example, a piezometer reading 8000 digits corresponds to a period of 354 µs and a frequency
of 2828 Hz. Note that in the above equation, the period is in seconds.
Digits are directly proportional to the applied pressure.
Hz
1000
Pressure =
(Current Reading - Initial Zero Reading) × Linear Calibration Factor
Or
P = (R1 – R0) ×G
Equation 10 - Convert Digits to Pressure
Since the linearity of most sensors is within 0.2% FS the errors associated with non-linearity are
of minor consequence. However, for those situations requiring the highest accuracy it may be
desirable to use a second order polynomial to get a better fit of the data points. The use of a
second order polynomial is explained in Appendix F.
The calibration report included with each transducer shows the data from which the linear gage
factor and the second order polynomial coefficients are derived. Columns on the right show the
size of the error incurred by assuming a linear coefficient and the improvement that can be
expected by going to a second order polynomial. In many cases, the difference is minor. The
calibration reports give the pressure in certain engineering units. These can be converted to other
engineering units using the multiplication factors shown in Table 3 below.
Page 48
42
From →
↓
psi "H2
"HG
mm HG
atm
mbar
bar
kPa
MPa
psi
1
.036127
.43275
.0014223
1.4223
.49116
.019337
14.696
.014503
14.5039
.14503
145.03
"H
O
27.730 1 12
.039372
39.372
13.596
.53525
406.78
.40147
401.47
4.0147
4016.1
'H
O
2.3108
.08333
1
.003281
3.281
1.133
.044604
33.8983
.033456
33.4558
.3346
334.6
mm H
0
704.32
25.399
304.788
1
1000
345.32
13.595
10332
10.197
10197
101.97
101970
m H
0
.70432
.025399
.304788
.001
1
.34532
.013595
10.332
.010197
10.197
.10197
101.97
"HG
2.036
.073552
.882624
.0028959
2.8959
1
.03937
29.920
.029529
29.529
.2953
295.3
mm HG
51.706
1.8683
22.4196
.073558
73.558
25.4 1 760
.75008
750.08
7.5008
7500.8
atm
.06805
.0024583
.0294996
.0000968
.0968
.03342
.0013158
1
.0009869
.98692
.009869
9.869
mbar
68.947
2.4908
29.8896
.098068
98.068
33.863
1.3332
1013.2
1
1000
10
10000
bar
.068947
.0024908
.0298896
.0000981
.098068
.033863
.001333
1.0132
.001 1 .01
10
kPa
6.8947
.24908
2.98896
.0098068
9.8068
3.3863
.13332
101.320
.1
100 1 1000
MPa
.006895
.000249
.002988
.00000981
.009807
.003386
.000133
.101320
.0001
.1
.001
1
To
O 'H2O mm H20 m H20
Table 3 - Engineering Units Multiplication Factors
Note: Due to changes in specific gravity with temperature, the factors for mercury and water in
Table 3 are approximations.
D.2 Temperature Correction
Careful selection of materials is made in constructing the vibrating wire piezometer to minimize
thermal effects; however, most units still have a slight temperature coefficient. Consult the
supplied calibration report to obtain the coefficient for a given piezometer.
Since piezometers are normally installed in a tranquil and constant temperature environment,
corrections are not normally required. If however, that is not the case for a selected installation,
corrections can be made using the internal thermistor for temperature measurement.
Temperature correction equation is as follows:
Temperatur e Correction =
(Current Temperature - Initial Zero Temperature) × Therm al Facto r
Or
PT = (T1-T0) x K
Equation 11 - Temperature Correction
The calculated correction would then be added to the Pressure calculated using Equation 10. (If
the engineering units were converted, remember to apply the same conversion to the calculated
temperature correction!) For example, assume the initial temperature was 22° C, the current
temperature is 15° C, and the thermal coefficient is −.01879 PSI per °C rise. The temperature
correction is 0.13 PSI. Adding this to the calculated pressure in the beginning of this appendix
results in a temperature corrected pressure of 19.98 PSI.
Page 49
43
D.3 Barometric Correction (required only on non-vented transducers)
Since the standard piezometer is hermetically sealed and unvented, it responds to changes in
atmospheric pressure. That being the case, corrections may be necessary, particularly for the
sensitive, low-pressure models. For example, a barometric pressure change from 29 to 31 inches
of mercury would result in ≈1 PSI of error (or ≈2.3 feet if monitoring water level in a well!).
Thus, it is advisable to read and record the barometric pressure every time the piezometer is read.
A separate pressure transducer (piezometer), kept out of the water, may be used for this purpose.
Barometric correction equation is as follows;
Barometric Correctio n =
(Current Barometer - Initial Zero Barometer) × Conversion F actor
Or
PB = (S1-S0) x F
Equation 12 - Barometric Correction
Since barometric pressure is usually recorded in inches of mercury, a Conversion Factor is
necessary to convert to PSI. The Conversion Factor for inches of mercury to PSI is .491. Table 3
lists other common Conversion Factors.
The calculated correction is usually subtracted from the Pressure calculated using Equation 10.
If the engineering units were converted, remember to apply the same conversion to the calculated
barometric correcti on!
The user should be cautioned that this correction scheme assumes ideal conditions. In reality,
conditions are not always ideal. For example, if the well is sealed, barometric effects at the
piezometer level may be minimal or attenuated from the actual changes at the surface. Thus,
errors may result when applying a correction that is not required.
An alternative to making barometric correction is to use piezometers that are vented to the
atmosphere.
Equation 13 describes the pressure calculation with temperature and barometric correction
applied.
APPENDIX F. IMPROVING THE ACCUR ACY OF CALCULATED
PRESSURES
Most vibrating wire pressure transducers are sufficiently linear (± 0.2 % FS) that use of the linear
calibration factor satisfies normal requirements. However, it should be noted that the accuracy of
the calibration data, which is dictated by the accuracy of the calibration apparatus, is always ±
0.1 % FS.
This level of accuracy can be recaptured, even where the transducer is non-linear, by the use of a
second order polynomial expression, which gives a better fit to the data then does a straight line.
The polynomial expression has the form:
Pressure = AR2 + BR + C
Equation 14 - Second Order Polynomial Expression
Where;
R is the reading (digits)
A, B, and C, are coefficients
Figure 36 shows a calibration report of a transducer that has a comparatively high non-linearity.
The figure under the “Linearity (%FS)” column is:
1-R0
F.S.
)
-P
x 100%
Calculated Pressure-True Pressure
G(R
x 100%=
Full Scale Pressure
Equation 15 - Linearity Calc ulation
For example when P= 40 psi, G (R0 - R1) gives a calculated pressure of 39.642 psi. The error is
0.357 psi or as much as 9.9 inches of water!
Whereas the polynomial expression gives a calculated pressure of A (7773)2 + B (7773) + C =
39.996 psi and the actual error is only 0.004 psi or 0.1 inch of water.
Note: If the polynomial equation is used, it is important that the value of C, in the polynomial
equation, be taken in the field. The field value of C is calculated by inserting the initial zero
reading into the polynomial equation with the pressure, P, set to zero. As of 8/2011, Geokon no
longer includes the C coefficient on its calibration reports, ensuring that, to properly use the
polynomial equation, users must calculate a C coefficient.
It should be noted that where changes of water levels are being monitored it makes little
difference whether the linear coefficient or the polynomial expression is used.
Page 52
46
APPENDIX G. BAROMETRIC CORRECTI O N ( NO N-VENTED
TRANSDUCERS ONLY)
Since the standard piezometer is hermetically sealed and unvented, it responds to changes in
atmospheric pressure. That being the case, corrections may be necessary, particularly for the
sensitive, low-pressure models.
The digits for the range of the transducer, as shown on the calibration report, may vary
considerably with changes in elevation (offset). It is a simple matter to compensate for this offset
with the 8020-59.
Referring to the example calibration report (Appendix E, Figure 36), note that the average zero
reading (0 psi) is 9139 digits (8020-59 maximum digits) and the average maximum reading (100
psi) is 5691 digits (8020-59 minimum digits). These readings were recorded at an atmospheric
pressure of 1001.4 mbar, corresponding to the conditions and altitude at the Geokon factory. To
compensate for elevation change, take a new zero reading at the site location. This new zero
reading becomes the maximum digits entered into the 8020-59. Add the difference between the
new zero reading and the original zero reading to the maximum reading shown on the calibration
report. This becomes the new minimum digits entered into the 8020-59.
For example:
0 psi Average Reading 100 psi Average Reading
Calibration Report: 9139 5691
Site Location: 9950 (+811 digits) 6502
(new maximum digits) (new minimum digits)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.