Geokon, Inc. warrants its products to be free of defects in materials and workmanship, under
normal use and service for a period of 13 months from date of purchase. If the unit should
malfunction, it must be returned to the factory for evaluation, freight prepaid. Upon examination
by Geokon, if the unit is found to be defective, it will be repaired or replaced at no charge.
However, the WARRANTY is VOID if the unit shows evidence of having been tampered with
or shows evidence of being damaged as a result of excessive corrosion or current, heat, moisture
or vibration, improper specification, misapplication, misuse or other operating conditions outside
of Geokon's control. Components which wear or which are damaged by misuse are not
warranted. This includes fuses and batteries.
Geokon manufactures scientific instruments whose misuse is potentially dangerous. The
instruments are intended to be installed and used only by qualified personnel. There are no
warranties except as stated herein. There are no other warranties, expressed or implied, including
but not limited to the implied warranties of merchantability and of fitness for a particular
purpose. Geokon, Inc. is not responsible for any damages or losses caused to other equipment,
whether direct, indirect, incidental, special or consequential which the purchaser may experience
as a result of the installation or use of the product. The buyer's sole remedy for any breach of this
agreement by Geokon, Inc. or any breach of any warranty by Geokon, Inc. shall not exceed the
purchase price paid by the purchaser to Geokon, Inc. for the unit or units, or equipment directly
affected by such breach. Under no circumstances will Geokon reimburse the claimant for loss
incurred in removing and/or reinstalling equipment.
Every precaution for accuracy has been taken in the preparation of manuals and/or software,
however, Geokon, Inc. neither assumes responsibility for any omissions or errors that may
appear nor assumes liability for any damages or losses that result from the use of the products in
accordance with the information contained in the manual or software.
TABLE OF CONTENTS
1. THEORY OF OPERATION .................................................................................................................................. 1
2. PRIOR TO INSTALLATION ................................................................................................................................ 3
SATURATING THE FILTER STONE .......................................................................................................................... 3
2.2.2 Reassembly Instr uct i ons ............................................................................................................................... 4
ESTABLIS HING AN INITIAL ZERO READING ........................................................................................................... 5
2.3
3. INSTALLATIO N .................................................................................................................................................... 7
3.2.1 The Drop Weight Method ............................................................................................................................. 8
SPLICIN G AN D JUNCTION BOXES ........................................................................................................................ 12
CHECKING THE CALIBRATION ............................................................................................................................ 18
4.5
5. DATA REDUCTION ............................................................................................................................................ 20
TEMPERATURE CORRECTION .............................................................................................................................. 21
5.2
BAROMETRIC CORRECTION (REQUIRED ONLY ON UNVENTED TRANSDUCERS) .................................................... 22
APPENDIX A. SPECIFICATIONS ......................................................................................................................... 26
A.1
STANDARD PIEZOMETER WIRING ...................................................................................................................... 26
APPENDIX B. THERMISTOR TEMPERATURE DERIVATION ..................................................................... 27
APPENDIX C. HIGH TEMPERATURE TH ERMISTOR LINEARIZATION .................................................. 28
APPENDIX D. IMPROVING THE ACCURACY OF THE CALCULATED PRE SSURE ............................... 29
FIGURE 2-CLOSE UP OF 4500MLP ............................................................................................................................... 2
FIGURE 5-MAKE A MARK AT SIX O’CLOCK ................................................................................................................. 4
FIGURE 6-TIGHTEN ONE AND ONE-QUARTER TURNS .................................................................................................. 4
FIGURE 7-MARKS FO R REASSEMBLY ........................................................................................................................... 4
FIGURE 8-FERRULES SEATED AGAINST FITTING BODY ................................................................................................ 5
TABLE 4-RESISTANCE WORK SHEET ...........................................................................................................................25
Geokon Model 4500 Vibrating Wire Piezometers are intended primarily for long-term
measurements of fluid, and/or pore pressures in standpipes, boreholes, embankments, pipelines
and pressure vessels. The Model 4500MLP is designed to permit the easy installation of several
piezometers inside a single borehole. It eliminates the need for alternating sand and bentonite
zones by allowing the entire hole to be backfilled with bentonite grout by tremie piping it in
through a grout pipe.
The basic Model 4500S piezometer (Figure 1) utilizes a sensitive stainless steel diaphragm to
which a vibrating wire element is connected. In use, changing pressures on the diaphragm cause
it to deflect. This deflection is measured as a change in tension and frequency of vibration in the
vibrating wire element. The square of the vibration frequency is directly proportional to the
pressure applied to the diaphragm.
Two coils, one with a magnet insert, the other with a pole piece insert, are located close to the
vibrating wire. In use, a pulse of varying frequency (swept frequency) is applied to the coils
causing the wire to vibrate primarily at its resonant frequency. When the excitation ends, the
wire continues to vibrate. During vibration, a sinusoidal signal is induced in the coils and
transmitted to the readout box where it is conditioned and displayed.
1
Figure 1 - Model 4500S Vibrating Wire Piezometer
Care needs to be exercised when choosing the pressure range of the piezometer. During
installation, the full pressure of the wet bentonite grout will be felt by the piezometer. Geokon
piezometers can withstand overranging up to 100 percent of the calibrated range without
affecting the calibration, shifting the zero reading, or damaging the unit. At higher overrange
pressures, the piezometer may temporarily cease reading until the grout sets up and the pressure
returns to normal.
2
Figure 2 - Close up of 4500MLP
For the Model 4500MLP, the filter stone shown in Figure 1 is replaced by a tube leading to a
much larger, curved, and porous filter stone, which is forced against the walls of the borehole by
a spring mechanism. The spring mechanism can be actuated remotely after the piezometer has
been lowered in its closed configuration to its desired location. This is shown schematically in
Figure 2 and Figure 3.
With the filter stones pressed against the wall of the borehole in this manner, the borehole can be
filled completely with bentonite cement grout without running the risk of getting the filter stone
plugged with bentonite, or having substantial amounts of the impervious bentonite interposed
between the piezometer diaphragm and the ground water in the surrounding soil.
Figure 3 - 4500MLP System
2. PRIOR TO INSTALLATION
2.1 Saturating the Filter Stone
1) Disconnect the Swagelok union at the transducer.
2) Immerse the whole assembly, upside down, in a bucket of water.
3) Apply a vacuum to the nylon filter tube using the large syringe supplied with the system.
This removes the entrapped air from the large filter stone.
4) Purge the transducer of air by injecting water into the Swagelok fitting as follows:
a) Fill the syringe with water.
b) Attach the small diameter tube supplied with the system to the syringe.
c) Push the tube all the way down inside the Swagelok fitting and into the piezometer
cavity.
d) Inject water from the syringe into the piezometer cavity.
5) With the assembly still under water, reconnect the nylon tube from the filter stone to the
transducer. Tighten the Swagelok per the instructions in Section 2.2.
6) Keep the assembly under water until ready to lower into the borehole.
Caution! Do not allow the piezometer to freeze once it has been filled with water.
3
2.2 Swagelok Tube Fitting Instructions
These instructions apply to one inch (25 mm) and smaller fittings.
2.2.1 Installation
1) Fully insert the tube into the fitting until it bumps against the shoulder.
Figure 4 - Tube Insertion
2) Rotate the nut until it is finger-tight. (For high-pressure applications as well as high-
safety-factor systems, further tighten the nut until the tube will not turn by hand or
move axially in the fitting.)
3) Mark the nut at the six o’clock position.
4
Figure 5 - Make a Mark at Six O’clock
4) While holding the fitting body steady, tighten the nut one and one-quarter turns until
the mark is at the nine o’clock position. (Note: For 1/16”, 1/8”, 3/16”, and 2, 3, and 4
mm fittings, tighten the nut three-quarters of a turn until the mark is at the three
o’clock position.)
Figure 6 - Tighten One and One-Quarter Turns
2.2.2 Reassembly Instructions
Swagelok tube fittings may be disassembled and reassembled many times.
Warning: Always depressurize the system before disassembling a Swagelok tube
fitting.
1) Prior to disassembly, mark the tube at the back of the nut, then make a line along the
nut and fitting body flats. These marks will be used during reassembly to ensure the nut is returned to its current position.
2) Disassemble the fitting.
3) Inspect the ferrules for damage and replace if necessary. If the ferrules are replaced
the connector should be treated as a new assembly. Refer to the section above for
installation instructions.
Figure 7 - Marks for Rea ssembly
5
4) Reassemble the fitting by inserting the tube with preswaged ferrules into the fitting
until the front ferrule seats against the fitting body.
Figure 8 - Ferrules Seated Against Fitting Body
5) While holding the fitting body steady, rotate the nut with a wrench to the previous
position as indicated by the marks on the tube and the connector. At this point, there
will be a significant increase in resistance.
6) Tighten the nut slightly.
Figure 9 - Tighten Nut Slightly
2.3 Establishing an Initial Zero Reading
It is imperative that an accurate initial zero reading be obtained for each piezometer as this
reading will be used for all subsequent data reduction.
Generally, the initial zero reading is obtained by reading the instrument prior to installation.
Vibrating Wire Piezometers differ from other types of pressure sensors in that they indicate a
reading when no pressure is exerted on the sensor.
Calibration data is supplied with each gage, a factory zero reading taken at a specific temperature
and absolute barometric pressure is included. (See Figure 18 for a sample calibration sheet.) Zero
readings at the site should coincide with the factory readings within 20 digits after barometric
and temperature corrections are made. (Barometric pressures change with elevation at a rate of
about 3.45 kPa (1/2 psi) per 300 meters (1,000 ft.) A thermistor is included inside the body of the
piezometer for the measurement of temperature.
6
To take an initial zero reading, complete the following:
1) Saturate the filter stone per the instructions in Section 2.1.
2) Allow the assembly to sit under water in a bucket for 5 to 15 minutes while the temperature
stabilizes. (If possible, the temperature of the water in the bucket should be the same as the
temperature of the water in the borehole. Use the thermistor inside the piezometer to measure
the water temperature.)
3) Once the temperature has stabilized, take a zero reading. (See Section 4 for readout
instructions.)
4) Note the barometric pressure if possible, or record the time for later referral to local weather
station data.
5) For maximum accuracy, record the depth of the piezometer below the water surface at the
time of the reading. (The piezometer diaphragm is at the same level as the lowest leaf
spring.)
50 PSI Grout for Medium
to Hard Soils
Ratio by
Weight
Ratio by
Weight
Water
30 gallons
2.5
75 gallons
6.6
Portland
Cement
The 28 day compressive strength of this mix
clay. The modulus is about 10,000 psi
The 28-day strength of this
similar to very soft clay.
3. INSTALLATION
Borehole sizes are not critical but they should be at least 100 mm (four inch) in diameter and
not more than 30 mm larger than the nominal size for which the spring-loaded mechanism was
designed.
3.1 Grouting requirements
For more details on grouting, refer to “Piezometers in Fully Grouted Boreholes” by Mikkelson
and Green, FMGM proceedings Oslo 2003. Copies are available from Geokon.
The general rule for grouting multilevel piezometers is to mimic the strength of the surrounding
soil. The emphasis should be on controlling the water to cement ratio. This is accomplished by
mixing the cement with the water first. The most effective way of mixing is in a 50 to 200
gallon barrel or tub, using the drill rig pump to circulate the mix.
Any kind of bentonite powder used to make drilling mud, combined with a Type 1 or Type 2
Portland cement can be used. The exact amount of bentonite added will vary somewhat. Table 1
shows two possible mixes for strengths of 50 psi and 4 psi.
7
Amount
94 lb. (one sack) 1 94 lb. (one sack) 1
Bentonite
NOTES:
Add the measured amount of clean water to the barrel then gradually add the cement in the
correct weight ratio. Next add the bentonite powder, slowly, so clumps do not form. Keep adding
bentonite until the watery mix turns to an oily/slimy consistency. Let the grout thicken for
another five to ten minutes. Add more bentonite as required until it is a smooth, thick cream,
similar to pancake batter. It is now as heavy as it is feasible to pump.
When pumping grout (unless the tremie pipe is to be left in place,) withdraw the tremie pipe after
each batch, by an amount corresponding to the grout level in the borehole.
25 lb. (as required) 0.3 39 lb. (as required) 0.4
is about 50 psi, similar to very stiff to hard
Table 1 - Cement/Bentonite/Water ratios
4 PSI Grout for Soft Soils
Amount
mix is about four psi,
8
3.2 Borehole Installation
In steeply inclined boreholes directed downwards, the drop weight method of installation is
recommended. See Section 3.2.1.
(Note that the drop weight method cannot be used in holes that will not stay open, and in holes
which require the casing to be removed as the piezometers are installed. This is because when
using the drop weight method the top piezometers are released first.)
For shallow inclinations and upward directed holes, (and for other situations where the drop
weight method is not desired or possible,) the pneumatic cutter method or the pull-pin method
can be used. See Sections 3.2.2 and 3.2.3.
3.2.1 The Drop Weight Method
4500MLP sensors are designed to be installed around flush coupled, one inch, Schd 80
PVC grout pipe; however, other grouting arrangements may be used.
Preparing the grout pipe:
1) Connect the sections of grout pipe together and lay them out along the ground.
2) Mark the desired locations of the piezometers on the grout pipe at the calculated
depths.
3) Drill a single 7/32 or 1/4 inch drill hole diametrically through the pipe in each marked
location.
Attaching the piezometers:
Each piezometer assembly is held to the grout pipe by means of a single 50 lb. nylon zip
tie (supplied), which passes through the two holes on opposite sides of the pipe. The zip
tie also holds the piezometer assembly in its closed position.
Pass the zip tie through the grout pipe, through the spaces between the leaf springs,
around the two platens, and then back to itself. Keep the cables from lower piezometers
inside the zip tie and leaf springs. The zip tie should pass just below the center leaf
spring so that the bottom of the platen assemblies will be held tight to the grout pipe.
Pull the zip tie tight around the grout pipe (Figure 10).
Figure 10 - 4500MLP zip Tied for Drop Weight Me thod
9
As the grout pipe is assembled and pushed into the borehole, the piezometer assemblies
are added, and the electrical cables are fed into the borehole. When the grout tube has
reached its final position and the assembly is complete, a special weight (provided with
the equipment,) is tied to a length of aircraft cable and allowed to fall freely down the
inside of the grout pipe. As the weight hits each of the zip ties stretching across the pipe,
the zip tie is snapped, allowing the leaf springs to expand and force the filter stones
against the walls of the borehole.
3.2.2 Pneumatic Cutter Method
4500MLP sensors are designed to be installed around flush coupled, one inch, Schd 80
PVC grout pipe; however, other grouting arrangements may be used.
The assembly is held in its closed position during installation by two nylon zip ties
(supplied). Making sure the grout pipe is between the platens, and the cables from the
lower piezometers are kept inside the zip tie and leaf springs, attach the body of the
piezometer to the grout pipe by passing the zip ties through the eyebolts in the platens
and through the holes in the cutting tool. Orient the tool in such a way that it will be
above the assembly when pushed into the borehole. See Figure 11.
If the grout pipe is going to be removed from the borehole the grout pipe is assembled
and pushed into the borehole first, then each piezometer assembly is pushed around the
grout pipe (and any lower piezometer cables) to the desired elevation. Piezometers should
be installed sequentially, from the base of the borehole to the mouth.
If the grout pipe is to be left in place, the piezometers can be taped to the grout pipe.
(Make sure not to put the tape around the outside of the filter stone and spring
mechanism.)
Figure 11 - 4500MLP P re-release
10
When the desired elevation is reached, the cutting tool is activated by connecting the
pneumatic tube from the piezometer to a bottle of CO2. (CO2 can be obtained locally from
any welding supply outlet.) The CO2 bottle should have a pressure regulator that is set to
a pressure of at least 2.5 MPa (350 psi), with the shutoff valve closed (Figure 12).
Figure 12 - Pneumatic Cutter Equipment
11
The shut off valve is then opened suddenly, allowing the pressure to reach the cutting
tool. The cutting tool cuts the zip ties, releasing the spring-loaded platens against the
borehole walls (Figure 13).
Figure 13 - 4500MLP Post-release
If installed in drill casing:
The sensor is lowered to the proposed elevation and the casing is pulled just above this
elevation before the assembly is released. The cutting tool is then removed from the drill
hole and the next assembly is prepared for installation.
When lowering the subsequent piezometers down the hole, feed the cables from the lower
piezometers through the middle of the assembled piezometer rather than around the
outside. (This will prevent the cables from interfering with the filter contacting with the
borehole wall).
When all the assemblies are installed, the hole can be grouted from the bottom up using
bentonite cement grout.
12
3.2.3 Pull-Pin Method
In this method, the piezometer assembly is held in its closed position by means of a pullpin. After the filter stone and the platen are squeezed together, the pull-pin passes through
two sets of three eyebolts each, which are mounted on the filter stone and the platen.
If the grout pipe is going to be removed from the borehole:
The piezometer assembly must be pushed around the grout pipe, down to the desired
elevation. This is accomplished using a second pipe, which can be another length of the
grout tube. While holding the piezometer in position by the second pipe, the pull-pin
cable should be pulled gently until all the slack is taken out. With a sudden and strong
jerk, pull on the pull-pin cable. This will release the platens without changing the position
of the piezometer relative to the borehole. (A bit of practice pulling the pins before the
actual installation will give some "feeling" and confidence as to how the system works.)
Piezometers should be installed sequentially, from the base of the borehole to the mouth.
If the grout pipe is going to be left in the borehole:
The piezometer assemblies should be attached to the grout pipe in a manner that does not
restrict the movement of the platens. The grout pipe is assembled, length by length, while
the piezometer assemblies, (each with its own pull-pin,) are attached to it by taping the
electrical cable to the grout pipe near the piezometer. The grout pipe and piezometer
assemblies can then be pushed down the hole as a unit. When the final position is reached
the pull-pins are pulled, activating the platens.
If installed in drill casing:
The sensor is lowered to the proposed elevation. The casing is pulled just above this
elevation before the assembly is released. The tool is then removed from the drill hole
and the next assembly prepared for installation. When lowering (or pushing) the
subsequent piezometers down the hole, feed the cables from the lower piezometers
through the middle of the assembled piezometer rather than around the outside. This will
prevent the cables from coming between the filter and the borehole wall.
When all the assemblies are installed, the hole can be grouted from the bottom up using a
bentonite cement grout. The grout pipe can either be removed from the hole or left in
place.
3.3 Splicing and Junction Boxes
Because the vibrating wire output signal is a frequency rather than a current or voltage,
variations in cable resistance have little effect on gage readings. Therefore, splicing of cables has
no effect and in some cases may in fact be beneficial.
13
For example, if multiple piezometers are installed in a borehole, and the distance from the
borehole to the terminal box or datalogger is great, a splice (or junction box) could be made to
connect the individual cables to a single multi-conductor cable. (See Figure 14.) This multiconductor cable would then be run to the readout station. For these installations, it is
recommended that the piezometer be supplied with enough cable to reach the installation depth
plus extra cable to pass through drilling equipment (rods, casing, etc.).
The cable used for making splices should be a high quality twisted pair type with 100% shielding
and an integral shield drain wire. When splicing, it is very important that the shield drain wires be spliced together.
Splice kits recommended by Geokon incorporate casts that are placed around the splice and then
filled with epoxy to waterproof the connections. When properly made, this type of splice is equal
or superior to the cable itself in strength and electrical properties. Contact Geokon for splicing
materials and additional cable splicing instructions.
Junction boxes and terminal boxes are available from Geokon for all types of applications. In
addition, portable readouts and dataloggers are available. Contact Geokon for specific
application information.
Figure 14 - Typical Multi-Piezometer Installation
14
3.4 Lightning Protection
In exposed locations, it is vital that the piezometer be protected against lightning strikes. A
tripolar plasma surge arrestor, which protects against voltage spikes across the input leads, is
built into the body of the piezometer. (See Figure 15.)
Additional lightning protection measures available include:
• Placing a Lightning Arrestor Board (LAB-3) in line with the cable, as close as possible to
the installed piezometer. (See Figure 15.) These units utilize surge arrestors and
transzorbs to further protect the piezometer. This is the recommended method of
lightning protection.
• Terminal boxes available from Geokon can be ordered with lightning protection built in.
The terminal board used to make the gage connections has provision for the installation
of plasma surge arrestors. Lightning Arrestor Boards (LAB-3) can also be incorporated
into the terminal box. The terminal box must be connected to an earth ground for these
levels of protection to be effective.
• If the instruments will be read manually with a portable readout (no terminal box) a
simple way to help protect against lightning damage is to connect the cable leads to a
good earth ground when not in use. This will help shunt transients induced in the cable to
ground, away from the instrument.
The Model GK-404 Vibrating Wire Readout is a portable, low-power, handheld unit that is
capable of running for more than 20 hours continuously on two AA batteries. It is designed for
the readout of all Geokon vibrating wire gages and transducers, and is capable of displaying the
reading in either digits, frequency (Hz), period (µs), or microstrain (µε). The GK-404 also
displays the temperature of the transducer (embedded thermistor) with a resolution of 0.1 °C.
4.1.1 Operating the GK-404
Before use, attach the flying leads to the GK-404 by aligning the red circle on the silver
“Lemo” connector of the flying leads with the red line on the top of the GK-404 (Figure
16). Insert the Lemo connector into the GK-404 until it locks into place.
15
Figure 16 - Lemo Connector to GK-404
Connect each of the clips on the leads to the matching colors of the sensor conductors,
with blue representing the shield (bare).
To turn the GK-404 on, press the “ON/OFF” button on the front panel of the unit. The
initial startup screen will display:
Geokon Inc.
GK-404 verX.XX
After approximately one second, the GK-404 will start taking readings and display them
based on the settings of the POS and MODE buttons.
The unit display (from left to right) is as follows:
• The current Position: Set by the POS button. Displayed as a letter A through F.
• The current Reading: Set by the MODE button. Displayed as a numeric value
followed by the unit of measure.
• Temperature reading of the attached gage in degrees Celsius.
Use the POS button to select position B and the MODE button to select Dg (digits).
(Other functions can be selected as described in the GK-404 Manual.)
The GK-404 will continue to take measurements and display readings until the unit is
turned off, either manually, or if enabled, by the Auto-Off timer.
For more information, consult the GK-404 manual.
16
4.2 GK-405 Readout Box
The GK-405 Vibrating Wire Readout is made up of two components: The Readout Unit,
consisting of a Windows Mobile handheld PC running the GK-405 Vibrating Wire Readout
Application; and the GK-405 Remote Module, which is housed in a weatherproof enclosure and
connects to the vibrating wire gage to be measured. The two components communicate
wirelessly using Bluetooth®, a reliable digital communications protocol. The Readout Unit can
operate from the cradle of the Remote Module, or, if more convenient, can be removed and
operated up to 20 meters from the Remote Module.
4.2.1 Connecting Sensors
Connecting Sensors with 10-pin Bulkhead Connectors Attached:
Align the grooves on the sensor connector (male), with the appropriate connector on the
readout (female connector labeled senor or load cell). Push the connector into place, and
then twist the outer ring of the male connector until it locks into place.
Connecting Sensors with Bare Leads:
Attach the GK-403-2 flying leads to the bare leads of a Geokon vibrating wire sensor by
connecting each of the clips on the leads to the matching colors of the sensor conductors,
with blue representing the shield (bare).
4.2.2 Operating the GK-405
Press the button labeled “POWER ON (BLUETOOTH)”. A blue light will begin
blinking, signifying that the Remote Module is waiting to connect to the handheld unit.
Launch the GK-405 VWRA program on the handheld PC by tapping on “Start”, then
“Programs”, then the GK-405 VWRA icon. After a few seconds, the blue light on the
Remote Module should stop flashing and remain lit, indicating that the remote module
has successfully paired with the handheld PC. The Live Readings Window will be
displayed on the handheld PC. Figure 17 shows a typical vibrating wire piezometer
output in digits and thermistor output in degrees Celsius.
If the no reading displays or the reading is unstable, see Section 6 for troubleshooting
suggestions. For further information consult the GK-405 Instruction Manual.
Figure 17 - Live Readings – Raw Readings
4.3 GK-403 Readout Box (Obsolete Model)
The GK-403 can store gage readings as well as apply calibration factors to convert readings to
engineering units. The following instructions explain taking gage measurements using Modes
"B" and "F" (similar to the GK-401 switch positions "B" and "F"). Consult the GK-403
Instruction Manual for additional information.
4.3.1 Connecting Sensors
Connecting Sensors with 10-pin Bulkhead Connectors Attached:
Align the grooves on the sensor connector (male), with the appropriate connector on the
readout (female connector labeled senor or load cell). Push the connector into place, and
then twist the outer ring of the male connector until it locks into place.
Connecting Sensors with Bare Leads:
Attach the GK-403-2 flying leads to the bare leads of a Geokon vibrating wire sensor by
connecting each of the clips on the leads to the matching colors of the sensor conductors,
with blue representing the shield (bare).
4.3.2 Operating the GK-403
1) Turn the display selector to position "B" (or "F").
2) Turn the unit on.
3) The readout will display the vibrating wire output in digits (See Equation 1 in
Section 5.1.) The last digit may change one or two digits while reading.
4) The thermistor reading will be displayed above the gage reading in degrees
centigrade.
5) Press the "Store" button to record the value displayed.
If the no reading displays or the reading is unstable, see Section 6 for troubleshooting
suggestions.
The unit will automatically turn off after approximately two minutes to conserve power.
17
18
4.4 Measuring Temperatures
All vibrating wire piezometers are equipped with a thermistor that gives a varying resistance
output as the temperature changes. The white and green leads of the instrument cable are
normally connected to the internal thermistor.
The GK-403, GK-404, and GK-405 readout boxes will read the thermistor and display the
temperature in degrees C. (High temperature versions use a different thermistor, which must be
read using an ohmmeter.)
To read temperatures using an ohmmeter:
1) Connect an ohmmeter to the green and white thermistor leads coming from the strain gage.
(Since the resistance changes with temperature are large, the effect of cable resistance is
usually insignificant. For long cables a correction can be applied, equal to 14.7 ohms per
one thousand feet. Multiply this factor by two to account for both directions.)
2) Look up the temperature for the measured resistance in Appendix B, Table 7. For high
temperature models use Appendix C, Table 8.
4.5 Checking the Calibration
The following procedure can be used to verify the calibration factor as supplied on the
calibration sheet:
1) Saturate the filter stone and fill the space between it and the diaphragm with water (see
Section 2.1).
2) Lower the piezometer to the bottom of a water-filled borehole using the cable to measure the
actual depth.
3) Allow 15 to 20 minutes for the piezometer to come to thermal equilibrium.
4) Using a readout box record the reading at that level (see Section 4 for readout instructions).
5) Raise the piezometer a known amount and record the reading.
6) Calculate the calibration factor using the change in pressure and the reading.
7) Compare to the calibration sheet value. The two values should agree within ±0.5%.
When doing this test please be aware that the actual water level inside the borehole might change
due to displacement of water by the different lengths of submerged cable. This is especially
critical where the cable length is long and the borehole diameter small. Allowing sufficient time
for the water level to equilibrate may solve this problem, or keep the borehole filled to the top.
19
Figure 18 - Sample Calibration She e t
20
5. DATA REDUCTION
5.1 Pressure Calculation
The digits displayed by the Geokon Models GK-403, GK-404, and GK-405 Readout Boxes on
channel B are based on the equation:
Digits =
Period
2
1
x 10-3 or Digits=
Equation 1 - Digits Calculation
For example, a piezometer reading 8000 digits corresponds to a period of 354 µs and a frequency
of 2828 Hz. Note that in the above equation, the period is in seconds. Geokon readout boxes
display microseconds.
Digits are directly proportional to the applied pressure.
Since the linearity of most sensors is within ±0.2% FS, the errors associated with nonlinearity are
of minor consequence. However, for those situations requiring the highest degree of accuracy, it
may be desirable to use a second order polynomial to get a better fit of the data points. The use
of a second order polynomial is explained in Appendix D.
The instrument’s calibration report, a typical example of which is shown in Figure 18, shows the
data from which the linear gage factor and the second order polynomial coefficients are derived.
Columns on the right show the size of the error incurred by assuming a linear coefficient and the
improvement that can be expected by going to a second order polynomial. In many cases, the
difference is minor.
The calibration report gives the pressure in certain engineering units. These can be converted to
other engineering units using the multiplication factors shown in Table 2.
21
From →
psi
1
.036127
.43275
.0014223
1.4223
.49116
.019337
14.696
.014503
14.5039
.14503
145.03
"H2O
27.730 1 12
.039372
39.372
13.596
.53525
406.78
.40147
401.47
4.0147
4016.1
'H2O
2.3108
.08333
1
.003281
3.281
1.133
.044604
33.8983
.033456
33.4558
.3346
334.6
mm H20
704.32
25.399
304.788
1
1000
345.32
13.595
10332
10.197
10197
101.97
101970
m H20
.70432
.025399
.304788
.001
1
.34532
.013595
10.332
.010197
10.197
.10197
101.97
"HG
2.036
.073552
.882624
.0028959
2.8959
1
.03937
29.920
.029529
29.529
.2953
295.3
mm HG
51.706
1.8683
22.4196
.073558
73.558
25.4 1 760
.75008
750.08
7.5008
7500.8
atm
.06805
.002458
.029499
.0000968
.0968
.03342
.001315
1
.000986
.98692
.009869
9.869
mbar
68.947
2.4908
29.8896
.098068
98.068
33.863
1.3332
1013.2
1
1000
10
10000
bar
.068947
.002490
.029889
.0000981
.098068
.033863
.001333
1.0132
.001 1 .01
10
kPa
6.8947
.24908
2.98896
.0098068
9.8068
3.3863
.13332
101.320
.1
100
1
1000
MPa
.006895
.000249
.002988
.0000098
.009807
.003386
.000133
.101320
.0001
.1
.001
1
To
↓
psi
"H2O
'H2O
mm H20
m H20
"HG
mm HG
atm
mbar
bar
kPa
MPa
Table 2 - Engineering Units Multiplication Factors
(Note: Due to changes in specific gravity with temperature, the factors for mercury and water in
the above table are approxi m ati ons.)
5.2 Temperature Correction
The materials used in the construction of Geokon’s vibrating wire piezometers have been
carefully selected to minimize thermal effects; however, most units still have a slight temperature
coefficient. Consult the calibration sheet supplied with the instrument to obtain the coefficient
for the individual piezometer.
Since piezometers are normally installed in a tranquil and constant temperature environment,
corrections are normally not required. If this is not the case for the selected installation,
corrections can be made using the internal thermistor for temperature measurement. See Section
4.4 for instructions regarding obtaining the piezometer temperature.
The temperature correction equation is as follows:
Temperatur e Correction =
(Current Temperature - Initial Temperature) × Therm a l Factor
The calculated correction would then be added to the pressure calculated using Equation 2. If the
engineering units were converted, remember to apply the same conversion to the calculated
temperature correction.
For example: If the initial temperature was 22° C, and the current temperature is 15° C, and the
thermal factor (K on the calibration report,) is +0.1319 kPa per °C rise. The temperature
correction is +0.1319(15-22) = -0.92 kPa. Refer to the calibration report provided with the
instrument for the initial temperature and thermal factor.
Or
PT = (T1-T0) x K
Equation 3 - Temperature Correction
22
5.3 Barometric Correction (required only on unvented transducers)
Since the standard piezometer is hermetically sealed and unvented, it responds to changes in
atmospheric pressure. Corrections may be necessary, particularly for the sensitive, low-pressure
models. For example, a barometric pressure change from 29 to 31 inches of mercury would
result in ≈1 PSI of error (or ≈2.3 feet if monitoring water level in a well). Thus, it is advisable to
read and record the barometric pressure every time the piezometer is read. A separate pressure
transducer (piezometer), kept out of the water, may be used for this purpose.
The barometric correction equation is as follows:
Barometric Correction =
(Current Barometer - Initial Barometer) × Convers ion Factor
Or
PB = (S1-S0) x F
Equation 4 - Barometric Correction
The calculated barometric correction is subtracted from the pressure calculated using Equation 2.
If the engineering units were converted, remember to apply the same conversion to the
calculated barometric correction.
Barometric pressure is usually recorded in inches of mercury. The Conversion Factor for inches
of mercury to PSI is 0.491, and from inches of mercury to kPa is 3.386. Table 2 lists other
common Conversion Factors.
Equation 5 shows the pressure calculation with temperature and barometric correction applied.
P
corrected
The user should be cautioned that this correction scheme assumes ideal conditions. In reality,
conditions are not always ideal. For example, if the well is sealed, barometric effects at the
piezometer level may be minimal or attenuated from the actual changes at the surface. Thus,
errors may result from applying a correction that is not required. In these cases, Geokon
recommends independently recording the barometric pressure changes and correlating them with
the observed pressure changes in order to arrive at a correction factor.
An alternative to making barometric correction is to use piezometers that are vented to the
atmosphere. See Section 5.4.
= (R1 – R0)G + (T1-T0) K - (S1-S0) F
Equation 5 - Corrected Pressure Calculation
23
5.4 Vented Piezometers
Vented piezometers are designed to eliminate barometric effects. The space inside the transducer
is not hermetically sealed and evacuated; instead, it is connected via a tube (integral with the
cable) to the atmosphere. A chamber containing desiccant capsules is attached to the end of the
tube to prevent moisture from entering the transducer cavity. Vented piezometers require more
maintenance then unvented types, and there is always a danger that water can find its way into
the inside of the transducer and ruin it.
In order to keep the desiccant fresh during storage and transportation, the outer end of the
desiccant chamber is closed by means of a seal screw when shipped from the factory. THIS
SEAL SCREW MUST BE REMOVED BEFORE THE PIEZOMETER IS PUT INTO
SERVICE!
The desiccant capsules are blue when fresh. They will gradually turn pink as they absorb
moisture. When they have turned light pink in color, they should be replaced. Contact Geokon
for replacement capsules.
5.5 Environmental Factors
Since the purpose of the piezometer installation is to monitor site conditions, factors that may
affect these conditions should always be observed and recorded. Seemingly minor effects may
have a real influence on the behavior of the structure being monitored, and may give an early
indication of potential problems. Some of these factors include, but are not limited to: blasting,
rainfall, tidal levels, traffic, temperature and barometric changes, weather conditions, changes in
personnel, nearby construction activities, excavation and fill level sequences, seasonal changes,
etc.
24
6. TROUBLESHOOTING
Maintenance and troubleshooting of vibrating wire piezometers is confined to periodic checks of
cable connections and maintenance of terminals. The transducers themselves are sealed and are
not user serviceable. Gages should not be opened in the field.
Should difficulties arise, consult the following list of problems and possible solutions. For
additional troubleshooting and support, contact Geokon.
Symptom: Thermistor resistance is too high
Likely, there is an open circuit. Check all connections, terminals, and plugs. If a cut is
located in the cable, splice according to instructions in Section 3.3.
Symptom: Thermistor resistance is too low
A short is likely. Check all connections, terminals, and plugs. If a short is located in the
cable, splice according to instructions in Section 3.3.
Water may have penetrated the interior of the piezometer. There is no remedial action.
Symptom: Piezometer reading unstable
Make sure the shield drain wire is connected to the blue clip on the flying leads. (Green for
the GK-401.)
Isolate the readout from the ground by placing it on a piece of wood or other insulator.
Check for sources of nearby electrical noise such as motors, generators, antennas, or
electrical cables. Move the piezometer cable away from these sources if possible. Contact the
factory for available filtering and shielding equipment.
The Piezometer may have been damaged by overranging or shock. Inspect the diaphragm and
housing for damage.
The body of the Piezometer may be shorted to the shield. Check the resistance between the
shield drain wire and the Piezometer housing. If the resistance is very low, the gage
conductors may be shorted.
Symptom: Piezometer fails to give a reading
Check the resistance of the cable by connecting an ohmmeter to the sensor leads. Table 3
shows the expected resistance for the various wire combinations; Table 4 is provided for the
customer to fill in the actual resistance found. Cable resistance is approximately 14.7Ω per
1000' of 22 AWG wire. Multiply this factor by two to account for both directions.
If the resistance is very high or infinite, the cable is probably broken or cut. If the resistance
is very low, the gage conductors may be shorted. If a cut or a short is located in the cable,
splice according to instructions in Section 3.3.
Check the readout with another gage to ensure it is functioning properly.
The Piezometer may have been overranged or shocked. Inspect the diaphragm and housing
for damage.
25
≅180Ω
≅180Ω
3000Ω
25°C
3000Ω
25°C
Vibrating Wire Sensor Lead Grid - SAMPLE VALUES
Red Black White Green Shield
Red N/A
Black
N/A infinite infinite infinite
White infinite infinite N/A
Green infinite infinite
infinite infinite infinite
at
at
N/A infinite
Shield infinite infinite infinite infinite N/A
Table 3 - Sample Resistance
Vibrating Wire Sensor Lead Grid - SENSOR NAME/## :
Red Black White Green Shield
Red
Black
White
infinite
Green
Shield
Table 4 - Resistance Work Sheet
26
Model
4500S
4500AL1
4500B
4500C
45802
Available
0-50
0-15000
0-5
0-50
0-50
0-1
Resolution
0.025% FS
0.025% FS
0.025% FS
0.05% FS
0.01% FS
Linearity
0.5% FS3
0.5% FS3
0.5% FS3
0.5% FS3
0.5% FS3
Accuracy
0.1% FS4
0.1% FS4
0.1% FS4
0.1% FS4
0.1% FS
Overrange
1.5 x Rated Pressure
Thermal
Coefficient
<0.025% FS/
°C
<0.05% FS/
°C
<0.025% FS/
°C
<0.05% FS/
°C
<0.025% FS/
°C
.75"
19.05 mm
1"
25.40 mm
.687"
17.45 mm
.437"
11.10 mm
1.5"
38.10 mm
5.25"
133.35 mm
5.25"
133.35 mm
5.25"
133.35 mm
6.5"
165.10 mm
6.5"
165.10 mm
Frequency
Range Hz
A
Vibrating Wire Gage +
Red
B
Vibrating Wire Gage -
Black
C
Thermistor
White
D
Thermistor
Green
E
Cable Shield
Shield
F-K
Not Used
APPENDIX A. SPECIFICATIONS
Ranges
(psi)
OD
Length
0-100
0-150
0-250
0-500
0-750
0-1000
0-1500
0-3000
0-5000
0-10000
0-10
0-25
0-100
0-250
0-100
0-250
0-5
Accuracy of Geokon test apparatus: 0.1%
Contact Geokon for specific application information.
Notes:
Accuracy of test apparatus: 0.05%
Other ranges available upon request.
0.1% FS linearity available upon request.
Derived using second order polynomial.
Pin Function Wire Color
Table 6 - Standard Piezometer Wiring
Ohms
Temp
Ohms
Temp
Ohms
Temp
Ohms
Temp
Ohms
Temp
201.1K
-50
16.60K
-10
2417
+30
525.4
+70
153.2
+110
187.3K
-49
15.72K
-9
2317
31
507.8
71
149.0
111
174.5K
-48
14.90K
-8
2221
32
490.9
72
145.0
112
162.7K
-47
14.12K
-7
2130
33
474.7
73
141.1
113
151.7K
-46
13.39K
-6
2042
34
459.0
74
137.2
114
141.6K
-45
12.70K
-5
1959
35
444.0
75
133.6
115
132.2K
-44
12.05K
-4
1880
36
429.5
76
130.0
116
123.5K
-43
11.44K
-3
1805
37
415.6
77
126.5
117
115.4K
-42
10.86K
-2
1733
38
402.2
78
123.2
118
107.9K
-41
10.31K
-1
1664
39
389.3
79
119.9
119
101.0K
-40
9796 0 1598
40
376.9
80
116.8
120
94.48K
-39
9310
+1
1535
41
364.9
81
113.8
121
88.46K
-38
8851 2 1475
42
353.4
82
110.8
122
82.87K
-37
8417 3 1418
43
342.2
83
107.9
123
77.66K
-36
8006 4 1363
44
331.5
84
105.2
124
72.81K
-35
7618 5 1310
45
321.2
85
102.5
125
68.30K
-34
7252 6 1260
46
311.3
86
99.9
126
64.09K
-33
6905 7 1212
47
301.7
87
97.3
127
60.17K
-32
6576 8 1167
48
292.4
88
94.9
128
56.51K
-31
6265 9 1123
49
283.5
89
92.5
129
53.10K
-30
5971
10
1081
50
274.9
90
90.2
130
49.91K
-29
5692
11
1040
51
266.6
91
87.9
131
46.94K
-28
5427
12
1002
52
258.6
92
85.7
132
44.16K
-27
5177
13
965.0
53
250.9
93
83.6
133
41.56K
-26
4939
14
929.6
54
243.4
94
81.6
134
39.13K
-25
4714
15
895.8
55
236.2
95
79.6
135
36.86K
-24
4500
16
863.3
56
229.3
96
77.6
136
34.73K
-23
4297
17
832.2
57
222.6
97
75.8
137
32.74K
-22
4105
18
802.3
58
216.1
98
73.9
138
30.87K
-21
3922
19
773.7
59
209.8
99
72.2
139
29.13K
-20
3748
20
746.3
60
203.8
100
70.4
140
27.49K
-19
3583
21
719.9
61
197.9
101
68.8
141
25.95K
-18
3426
22
694.7
62
192.2
102
67.1
142
24.51K
-17
3277
23
670.4
63
186.8
103
65.5
143
23.16K
-16
3135
24
647.1
64
181.5
104
64.0
144
21.89K
-15
3000
25
624.7
65
176.4
105
62.5
145
20.70K
-14
2872
26
603.3
66
171.4
106
61.1
146
19.58K
-13
2750
27
582.6
67
166.7
107
59.6
147
18.52K
-12
2633
28
562.8
68
162.0
108
58.3
148
17.53K
-11
2523
29
543.7
69
157.6
109
56.8
149
Table 7 - Thermistor Resistance versus Temperature
55.6
150
APPENDIX B. THERMISTOR TEMPERATURE DERIVATION
Thermistor Type: YSI 44005, Dale #1C3001-B3, Alpha #13A3001-B3
Resistance to Temperature Equation:
1
T=
A+B(LnR)+C(LnR)
Equation 6 - Resistance to Temperature
Where;
T = Temperature in °C.
LnR = Natural Log of Thermistor Resistance
A = 1.4051 × 10-3
B = 2.369 × 10-4
C = 1.019 × 10-7
Note: Coefficients calculated over the −50 to +150° C. span.
-273.2
3
27
28
Temp
R
(ohms)
LnR
LnR3
Calculated
Temp
Diff
FS
Error
Temp
R
(ohms)
LnR
LnR3
Calculated
Temp
Diff
FS
Error
-30
113898
11.643
1578.342
-30.17
0.17
0.06
120
407.62
6.010
217.118
120.00
0.00
0.00
-25
86182
11.364
1467.637
-25.14
0.14
0.05
125
360.8
5.888
204.162
125.00
0.00
0.00
-20
65805
11.094
1365.581
-20.12
0.12
0.04
130
320.21
5.769
191.998
130.00
0.00
0.00
-15
50684.2
10.833
1271.425
-15.10
0.10
0.03
135
284.95
5.652
180.584
135.00
0.00
0.00
-10
39360
10.581
1184.457
-10.08
0.08
0.03
140
254.2
5.538
169.859
140.01
-0.01
0.00
-5
30807.4
10.336
1104.068
-5.07
0.07
0.02
145
227.3
5.426
159.773
145.02
-0.02
-0.01
0
24288.4
10.098
1029.614
-0.05
0.05
0.02
150
203.77
5.317
150.314
150.03
-0.03
-0.01
5
19294.6
9.868
960.798
4.96
0.04
0.01
155
183.11
5.210
141.428
155.04
-0.04
-0.01
10
15424.2
9.644
896.871
9.98
0.02
0.01
160
164.9
5.105
133.068
160.06
-0.06
-0.02
15
12423
9.427
837.843
14.98
0.02
0.01
165
148.83
5.003
125.210
165.08
-0.08
-0.03
20
10061.4
9.216
782.875
19.99
0.01
0.00
170
134.64
4.903
117.837
170.09
-0.09
-0.03
25
8200
9.012
731.893
25.00
0.00
0.00
175
122.1
4.805
110.927
175.08
-0.08
-0.03
30
6721.54
8.813
684.514
30.01
-0.01
0.00
180
110.95
4.709
104.426
180.07
-0.07
-0.02
35
5540.74
8.620
640.478
35.01
-0.01
0.00
185
100.94
4.615
98.261
185.10
-0.10
-0.04
40
4592
8.432
599.519
40.02
-0.02
-0.01
190
92.086
4.523
92.512
190.09
-0.09
-0.03
45
3825.3
8.249
561.392
45.02
-0.02
-0.01
195
84.214
4.433
87.136
195.05
-0.05
-0.02
50
3202.92
8.072
525.913
50.01
-0.01
-0.01
200
77.088
4.345
82.026
200.05
-0.05
-0.02
55
2693.7
7.899
492.790
55.02
-0.02
-0.01
205
70.717
4.259
77.237
205.02
-0.02
-0.01
60
2276.32
7.730
461.946
60.02
-0.02
-0.01
210
64.985
4.174
72.729
210.00
0.00
0.00
65
1931.92
7.566
433.157
65.02
-0.02
-0.01
215
59.819
4.091
68.484
214.97
0.03
0.01
70
1646.56
7.406
406.283
70.02
-0.02
-0.01
220
55.161
4.010
64.494
219.93
0.07
0.02
75
1409.58
7.251
381.243
75.01
-0.01
0.00
225
50.955
3.931
60.742
224.88
0.12
0.04
80
1211.14
7.099
357.808
80.00
0.00
0.00
230
47.142
3.853
57.207
229.82
0.18
0.06
85
1044.68
6.951
335.915
85.00
0.00
0.00
235
43.673
3.777
53.870
234.77
0.23
0.08
90
903.64
6.806
315.325
90.02
-0.02
-0.01
240
40.533
3.702
50.740
239.69
0.31
0.11
95
785.15
6.666
296.191
95.01
-0.01
0.00
245
37.671
3.629
47.788
244.62
0.38
0.13
100
684.37
6.528
278.253
100.00
0.00
0.00
250
35.055
3.557
45.001
249.54
0.46
0.16
105
598.44
6.394
261.447
105.00
0.00
0.00
255
32.677
3.487
42.387
254.44
0.56
0.19
110
524.96
6.263
245.705
110.00
0.00
0.00
260
30.496
3.418
39.917
259.34
0.66
0.23
115
461.91
6.135
230.952
115.00
0.00
0.00
APPENDIX C. HIGH TEMPERATURE THERMISTOR LINEARIZATION
High Temperature Thermistor Linearization using Stein-Hart Log Equation
Thermistor Type: Thermometrics BR55KA822J
Resistance to Temperature Equation:
1
T=
A+B(LnR)+C(LnR)
Equation 7 - High Temperature Resistance to Temperature
Where;
T = Temperature in °C.
LnR = Natural Log of Thermistor Resistance
A = 1.02569 × 10-3
B = 2.478265 × 10-4
C = 1.289498 × 10-7
Note: Coefficients calculated over the -30° to +260° C. span.
-273.2
3
Table 8- Thermistor Resistance versus Temperature for High Temperature Models
29
APPENDIX D. IMPROVING THE ACCUR ACY OF THE CALCULATED
PRESSURE
Most vibrating wire pressure transducers are sufficiently linear (±0.2 % FS) that the use of the
linear calibration factor satisfies normal requirements. However, it should be noted that the
accuracy of the calibration data, which is dictated by the accuracy of the calibration apparatus, is
always ±0.1 % FS.
This level of accuracy can be recaptured, even where the transducer is nonlinear, by the use of a
second order polynomial expression, which gives a better fit to the data then does a straight line.
The polynomial expression has the form:
Pressure = AR2 + BR + C
Equation 8 - Second Or de r Polynomial Expression
Where;
R is the reading (digits channel B)
A, B, and C are coefficients
Figure 18 shows a typical calibration sheet of a transducer that has fairly normal nonlinearity.
The figure under the “Linearity (%FS)” column is
1-R0
F.S.
)
-P
x 100%
Calculated Pressure-True Pressure
G(R
x 100%=
Full Scale Pre ssure
Equation 9 - Linearity Calculation
Note: The linearity is calculated using the regression zero for R0 shown on the sheet.
For example when P= 420 kPa, G (R1 – R0) = - 0.1795(6749-9082), gives a calculated pressure
of 418.8 kPa. The error is 1.2 kPa equal to 122 mm of water.
Whereas the polynomial expression gives a calculated pressure of A (6749)2 + B (6749) + 1595.7
= 420.02 kPa and the actual error is only 0.02 kPa or two millimeters of water.
Note: If the polynomial equation is used it is important that the value of C be taken in the field,
following the procedures described in Section 2.3. The field value of C is calculated by inserting
the initial field zero reading into the polynomial equation with the pressure, P, set to zero.
If the field zero reading is not available, the value of C can be calculated by using the zero
pressure reading on the calibration sheet. In the above example the value of C would be derived
from the equation 0 = A(9074)2 + B(9074) from which C = 1595.7
It should be noted that where changes of water levels are being monitored it makes little
difference whether the linear coefficient or the polynomial expression is used.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.