5.2 ASSEMBLED BODY INCUBATOR ........................................................................................................................................18
5.3 ASSEMBLED CONTROL PANEL ...........................................................................................................................................20
5.4 ASSEMBLED WATER BOX .................................................................................................................................................22
5.7 LOCKED POWER CONTROL PANEL ....................................................................................................................................27
5.8 CONTROL PANEL KEYBOARD............................................................................................................................................29
5.9 ASSEMBLED HOOD DOOR .................................................................................................................................................31
6. MAINTENANCE SERVICE .............................................................................................................. 32
7. PROGRAMMED AND PREVENTIVE MAINTENANCE.............................................................33
8. FINAL CHECK ................................................................................................................................... 34
10 COMPLIES WITH FOLLOWING STANDARD........................................................................... 36
Ginevri s.r.l. 7669bmt Novembre 2009 3
Rev.04 “OGB Poly Care 3” - Service Manual
WHITE PAGE
4 7669bmt Novembre 2009 Ginevri s.r.l.
“OGB Poly Care 3” - Service Manual Rev.04
1 PRESENTATION
Dear customer,
we would like to thank you for choosing our Incubator “OGB Polycare 3”.
Ginevri’s quality, long experience, safety and easy maintenance will prove to be
the best reward for your continuous interest in our products line.
Giorgio Ginevri
GINEVRI s.r.l. – Registered office
Via Giacomo Boni - 00162 Roma
Before using the device, all personnel, who will be working with the unit, should read
and thoroughly understand this manual.
This manual provides instructions for calibration and operator maintenance. Ginevri cannot be
responsible for the performance of the Incubator “OGB Polycare 3” if the user does not operate the
unit in accordance with the instructions, fails to follow the maintenance recommendations in this
manual, or effect any repairs with unauthorized components.
Calibration and repair should be performed only by qualified service personnel. Technical
information is available through your local distributor.
If there is anything you do not understand, please contact your Ginevri’s representative for further
information.
Warning
GINEVRI srl - Servizio Assistenza Clienti
Via Cancelliera 25/B 00041 Cecchina (Roma) - ITALY
Tel. ++39 06 93459331 Fax ++39 06 93459393
E-mail: export@ginevri.com
www.ginevri.com
This manual refers to the enclosed equipment:
Ginevri s.r.l. 7669bmt Novembre 2009 5
Incubator OGB Polycare 3
S/N
Rev.04 “OGB Poly Care 3” - Service Manual
2 “EC” CERTIFICATE
6 7669bmt Novembre 2009 Ginevri s.r.l.
“OGB Poly Care 3” - Service Manual Rev.04
3. CALIBRATION CHECK
Every three months check the Temperature, Humidity and Oxygen probes correct operation.
To check the probes reading use the optional Ginevri Reference Simulator P/N 10487A60.
To check the reading of the probes perform the following steps:
Temperature probes
a) Switch off the equipment and disconnect it from the mains
b) Disconnect from the control panel the probe assembly connector and replace it
with the Reference Simulator
c) Connect the equipment to the mains and switch it on
d) Select on the Reference Simulator: PROBES on AIR 1
TEMP °C on 25°C
SELECT on MUTE
The AIR LED display should show the selected reading with a ±0.2°C tolerance
for temperatures lower then 30°C and with a ±0.1°C tolerance for temperatures
higher then 30°C.
Repeat this step for the temperature settings
e) Check the AIR2 and SKIN Temperature probes selecting AIR2 first and then
SKIN on the simulator dial and performing the same operations of the preceding
step
f) Select on the Reference Simulator: PROBES MAX TEMP
TEMP °C on 39.8°C
SELECT on H.T.
check that the MAX TEMP alarm does not activate.
Select on the Reference Simulator: TEMP °C on 40.1°C And check that the
MAX TEMP alarm activates.
Humidity Probe
a) Switch off the equipment and disconnect it from the mains
b) Disconnect from the control panel the probe assembly connector and
replace it with the Reference Simulator
c) Connect the equipment to the mains and switch it on
d) Select on the Reference Simulator: PROBES on HUM
Hum % on 30%
SELECT on MUTE
e) The HUM LED display should show the selected reading
f) Repeat this step for all Hum %settings .
Oxygen probe
a) Switch off the equipment and disconnect it from the mains
b) Disconnect from the control panel the probe assembly connector and
replace it with the Reference Simulator
c) Connect the equipment to the mains and switch it on
d) Select on the Reference Simulator: PROBES on OXY
Oxy % on 21%
SELECT on MUTE
e) The OXY LED display should show the selected reading
f) Repeat this step for all the OXY %settings
If the readings of the probes are not correct, within the specified tolerances, the probes should
be calibrated. The calibration is performed by means of the reference simulator as described in
the following chapters.
Ginevri s.r.l. 7669bmt Novembre 2009 7
Rev.04 “OGB Poly Care 3” - Service Manual
3.1 TEMPERATURE PROBES CALIBRATION
ATTENTION: Unless otherwise stated, the temperature probes calibration is performed
on the microprocessor board ( P/N 10359A15).
To calibrate the temperature probes perform the following steps:
a) Switch off the equipment
b) Remove the four screws securing the control panel and open it to have free
access to the boards
c) Disconnect from the control panel the probe assembly connector and replace it
with the Reference Simulator
d) Connect the equipment to the mains and switch on, if an alarm goes off mute it.
AIR 1
e) Select on the Reference Simulator: PROBES on AIR 1
TEMP °C on 25°C
SELECT 1 on MUTE
f) connect the negative lead of the multimetr to PIN 2 of J5, connect the positive lead
of the multimeter to TP1
g) Adjust the trimmer R29 to read 1060 mV ±1 on the multimeter
AIR 2
SKIN
MAX TEMP
h) Select on the Reference Simulator: TEMP °C su 36°C
i) Disconnect the multimeter
j) Adjust the trimmer R 28 to read 36°C ±0.1 on the AIR display of the control panel
k) Check the calibration selecting on the Reference Simulator dial all the settable
values, with the exception of 0°C and MAX TEMP, verifying that the set value is
displayed on the AIR display (to obtain a correct reading each set temperature
value should be kept stable for at least 10 seconds) the reading should be
between the following tolerances
< 30°C ±0.2°C
> 30°C ±0.1°C
l) Select on the Reference Simulator: PROBES su AIR 2
TEMP °C su 36°C
m) Adjust the trimmer R31 to read 36°C ±0.1 on the AIR display of the control panel.
Follow the instructions of step k) setting the PROBES dial of the Reference
Simulator first on AIR1 and then on AIR2
n) Select on the Reference Simulator: PROBES on SKIN
TEMP °C on 36.5°C
o) Adjust the trimmer trimmer R30 to read 36.5°C ±0.1 on the SKIN display of the
control panel. . Follow the instructions of step k) setting the PROBES dial of the
reference simulator first on AIR1, AIR2 And then on SKIN
p) Select on the Reference Simulator: PROBES on MAX TEMP
TEMP °C on 40°C
SELECT on H.T.
The alarm should be inactive
8 7669bmt Novembre 2009 Ginevri s.r.l.
“OGB Poly Care 3” - Service Manual Rev.04
q) Adjust the trimmer PT 5 placed on display board to reach half-way between alarm
activated and not activated. Adjust the trimmer PT 101 to have the MAX TEMP
alarm activated.
r) Select on the Reference Simulator VAR condition and adjust the potentiometer
placed in the simulator to have the MAX TEMP alarm activated. Select on the
Reference Simulator AIR 1 and verify that it is possible to read on incubator air
display a value of 40.0°C ± 0.1°C , if necessary repeat the q) point.
s) Repeat steps p) and q) to check the MAX TEMP calibration.
3.2 HUMIDITY PROBE CALIBRATION
ATTENTION: Unless otherwise stated, the humidity probe calibration is performed on
the microprocessor board ( P/N 10359A15)
a) Switch off the equipment
b) Remove the four screws securing the control panel and open it to have free access to
the boards
c) Disconnect from the control panel the probe assembly connector and replace it with
the Reference Simulator
d) Connect the equipment to the mains and switch on, if an alarm activates mute it by
MUTE push button.
e) Select on the Reference Simulator: PROBES on HUM
HUM % on 0% (short circuit)
f) On the MICROPROCESSOR BOARD connect the negative lead of a multimeter to
PIN 2 of J5 (black lead) and the positive lead to PIN 5 of U12. Adjust trimmer PT2 on
the DISPLAY BOARD to read 0 mV ±2
g) Select on the reference simulator HUM % on 90%
h) Adjust PT1 to read 90% on the display HUM of the control panel
i) Check the calibration Selecting on the Reference Simulator dial all the settable values,
with the exception of IN, verifying that the set value is displayed on the HUM. For a
further check of the humidity level in the incubator use a wet bulb Hygrometer.
3.3 OXYGEN PROBE CALIBRATION
ATTENTION: Unless other stated, the oxigen probe calibration is performed on the
display board
a) Switch off the equipment
b) Remove the four screws securing the control panel and open it to have free access to
the boards
c) Disconnect from the control panel the O2 probe connector and replace it with the
Reference Simulator
d) Connect the equipment to the mains and switch on, if an alarm goes off mute it.
e) Select on the Reference Simulator: PROBES on OXY
OXY % on 0% (short circuit)
f) On the MICROPROCESSOR BOARD connect the negative lead of a multimeter to
PIN 2 of J5 (black lead), connect the positive lead PIN 6 of U 12. Adjust the trimmer
PT 4 on the DISPLAY BOARD to read 0 mV ±2 . If it is necessary adjust before PT3.
Ginevri s.r.l. 7669bmt Novembre 2009 9
Rev.04 “OGB Poly Care 3” - Service Manual
g) Select OXY % on 21%.
h) Adjust trimmer PT3 to read 980 mV ±5 on the multimeter
i) Select OXI% on CAL and check that the oxi display on the control panel is showing
21%, ( calibration operation ,indicated with CA, takes about 30 sec.).
j) Select OXI % on 98% and adjust PT3 trimmer to obtain the reading of 98% on the oxi
display.
k) Select OXI% on 21% and verify the reading of 21% on the oxi display. If necessary
repeat the previous points i), j) and k).
3.4 HUMIDIFICATOR SYSTEM CALIBRATION.
ATTENTION: Unless specified, the humidificator calibration system is performed on the
power board.
a) Switch off the equipment
b) Remove the four screws securing the control panel and open it to have free
access to the boards
c) Disconnect from the control panel the probe assembly connector and replace it
with the Reference Simulator
d) Connect the equipment to the mains and switch on, if an alarm goes off mute it.
e) Select on the Reference Simulator: PROBES on HUM
HUM % on
0% (short circuit)
f) Disconnect the J5 from Power Board.
g) Verify that D14 led lights on.
h) Connect the negative lead of multimeter to the ground test-point placed near to R31
resistor.
i) Connect the positive lead of multimeter to the pin 3 of U6 and adjust the
trimmer RV1 to 3.83 Vdc.
j) Verify that the leds status does not change.
k) Connect the positive lead of multimeter to the pin 9 of U5 and adjust the
trimmer RV2 to 11 Vdc.
l) Verify that the leds status does not change.
m) Connect the positive lead of multimeter to the pin 6 of U5 and adjust the
trimmer RV5 to 3.8 Vdc.
n) Verify that the leds status does not change.
o) Connect the J5 connector to the Power Board.
p) Verify that D14 led lights off.
q) Select on Reference Simulator HUM = 30%
r) Adjust the HUM SET point value to 66%.
10 7669bmt Novembre 2009 Ginevri s.r.l.
“OGB Poly Care 3” - Service Manual Rev.04
s) Verify that the D19 led lights on.
t) Connect the positive lead of multimeter to pin 3 ( red wire ) of J5 connector
of the power board.
u) Adjust the trimmer RV3 to 3.3Vdc value.
v) Select on Reference Simulator HUM= 66%.
w) Verify that the D19 led lights off.
x) Connect the positive lead of multimeter to pin 3 ( red wire ) of J5 connector
of the power board.
y) Adjust the trimmer RV4 to 2.5Vdc value.
Ginevri s.r.l. 7669bmt Novembre 2009 11