Genevri OGB Polycare 2 Service Manual

“OGB Poly Care 2” – sevice manual Rev. 05
OGB Poly Care 2
Incubator
Service Manual
Ginevri s.r.l. 7459bmt November 2009 1
Rev. 05 “OGB Poly Care 2” – service manual
RESPECTING THE ENFORCED NORMS
2 7459bmt November 2009 Ginevri s.r.l.
Index
1 PRESENTATION ...................................................................................................................................................................4
2 “EC” CERTIFICATE ............................................................................................................................................................5
3 CALIBRATION CHECK.......................................................................................................................................................6
3.1 TEMPERATURE PROBES CALIBRATION...................................................................................................................6
4. ELECTRICAL INTERCONNECTION DIAGRAM .........................................................................................................8
5. SPARE PART.......................................................................................................................................................................10
5.1 INCUBATOR POLY CARE 2 ........................................................................................................................................10
5.2 LOCKED POWER CONTROL PANEL .........................................................................................................................12
5.3 ASSEMBLED BODY INCUBATOR..............................................................................................................................14
5.4 ASSEMBLED WATER BOX .........................................................................................................................................16
5.5 ASSEMBLED CONTROL PANEL ................................................................................................................................18
5.6 ASSEMBLED HOOD .....................................................................................................................................................20
5.7 ASSEMBLED BASAMENT ...........................................................................................................................................21
5.8 CONTROL PANEL LOCKED KEYBOARD.................................................................................................................23
5.9 ASSEMBLED WOOD DOOR ........................................................................................................................................25
6. MAINTENANCE SERVICE ..............................................................................................................................................26
7. PROGRAMMED AND PREVENTIVE MAINTENANCE .............................................................................................27
8. FINAL CHECK....................................................................................................................................................................28
9. TECHNICAL DATA ...........................................................................................................................................................29
10. OGB POLY CARE 2 COMPLIES WITH THE FOLLOWING STANDARDS:.........................................................30
Rev. 05 “OGB Poly Care 2” – service manual
1 PRESENTATION
Dear customer, we would like to thank you for choosing our Incubator “OGB Polycare 2”. Ginevri’s quality, long experience, safety and easy maintenance will prove to be the best reward for your continuous interest in our products line.
Giorgio Ginevri
GINEVRI s.r.l. – Registered office
Via Giacomo Boni - 00162 Roma
Before using the device, all personnel, who will be working with the unit, should read and thoroughly understand this manual.
This manual provides instructions for calibration and operator maintenance. Ginevri cannot be responsible for the performance of the Incubator “OGB Polycare 2” if the user does not operate the unit in accordance with the instructions, fails to follow the maintenance recommendations in this manual, or effect any repairs with unauthorized components.
Calibration and repair should be performed only by qualified service personnel. Technical information is available through your local distributor.
If there is anything you do not understand, please contact your Ginevri’s representative for further information.
Warning
GINEVRI srl - Servizio Assistenza Clienti
Via Cancelliera, 25/B 00041 Cecchina (Roma) - ITALY
Tel. ++39 06 93459331 Fax ++39 06 93459393
E-mail: export@ginevri.com
www.ginevri.com
This manual refers to the enclosed equipment:
4 7459bmt November 2009 Ginevri s.r.l.
Incubator OGB Polycare 2
S/N
2 “EC” CERTIFICATE
Rev. 05 “OGB Poly Care 2” – service manual
3 CALIBRATION CHECK
Every three months check the Temperature, probes correct operation. To check the probes reading use the optional Ginevri Reference Simulator P/N 10487A60. To check the reading of the probes perform the following steps:
Temperature probes
a) Switch off the equipment and disconnect it from the mains b) Disconnect from the control panel the probe assembly connector and replace it with the
Reference Simulator c) Connect the equipment to the mains and switch on d) Select on the Reference Simulator: PROBES on AIR 1 TEMP °C on 25°C SELECT on MUTE
The AIR LED display should show the selected reading with a ±0.2°C tolerance for
temperatures lower then 30°C and with a ±0.1°C tolerance for temperatures higher than
30°C.
Repeat this step for all the temperature settings
e) Check the AIR2 and SKIN Temperature probes selecting AIR2 first and then SKIN on
the simulator dial and performing the same operations of the preceeding step f) Select on the Reference Simulator : PROBES MAX TEMP TEMP °C on 39.8°C SELECT on H.T.
check that the MAX TEMP alarm does not activate.
Select on the Reference Simulator: TEMP °C on 40.1°C And check that the MAX
TEMP alarm activates.
If the readings of the probes are not correct, within the specified tolerances, the probes should be
calibrated.
The calibration is performed by means of the Reference Simulator as described in the following
chapters.
3.1 TEMPERATURE PROBES CALIBRATION
ATTENTION: Unless specified, the temperature probes calibration is performed on the
microprocessor board (PN 7477).
To calibrate the temperature probes perform the following steps:
a) Switch off the equipment
b) Open the control panel to be able to access the boards
c) Disconnect from the control panel the probe assembly connector and replace it with the
Reference Simulator
d) Connect the equipment to the mains and switch on, if an alarm activates mute it by MUTE
push button.
AIR 1
e) Select on the Reference Simulator: PROBES on AIR 1
6 7459bmt November 2009 Ginevri s.r.l.
AIR 2
SKIN
MAX TEMP
TEMP °C on 25°C SELECT 1 on MUTE
f) connect the negative lead of the multimeter to PIN 2 of J5, connect the positive lead of the
multimeter to TP1
g) Adjust the trimmer R29 to read on the multimeter 1060 mV ±1
h) Select on the Reference Simulator: TEMP °C su 36°C
i) Disconnect the multimeter
j) Adjust the trimmer R 28 to read 36°C ±0.1 on the AIR display of the control panel
k) Check the calibration Selecting on the Reference Simulator dial all the settable values,
with the exception of 0°C and MAX TEMP, verifying that the set value is displayed on the AIR display (to obtain a correct reading each set temperature value should be kept for at
least 10 seconds) the reading should be between the following tolerances < 30°C ±0.2°C > 30°C ±0.1°C
l) Select on the Reference Simulator: PROBES on AIR 2 TEMP °C on 36°C
m) Adjust the trimmer R31 to read 36°C ±0.1 on the AIR display of the control panel. Follow
the instructions of step k) setting the PROBES dial of the reference simulator first on AIR1
And then on AIR2.
n) Select on the Reference Simulator: PROBES on SKIN TEMP °C on 36.5°C
o) Adjust the trimmer R30 to read 36.5°C ±0.1 on the SKIN display of the control panel. .
Follow the instructions of step k) setting the PROBES dial of the reference simulator first
on AIR1, AIR2 And then on SKIN.
p) Select on the Reference Simulator: PROBES on MAX TEMP TEMP °C on 40°C SELECT on H.T. The alarm should be active.
q) Adjust the trimmer PT 101 placed on display board half way between alarm activated and
not activated. Adjust the trimmer PT 101 to have the MAX TEMP alarm activated. r) Select on the Reference Simulator VAR condition and adjust the potentiometer placed in
the simulator to have the MAX TEMP alarm activated. Select on the Reference Simulator
AIR 1 and verify that it is possible to read on incubator air display a value of 40.0°C ±
0.1°C , if necessary repeat the q) point.
Rev. 05 “OGB Poly Care 2” – service manual
4. ELECTRICAL INTERCONNECTION DIAGRAM
8 7459bmt November 2009 Ginevri s.r.l.