General Tools DPHC2001 User Manual

TABLE OF CONTENTS
.
.
1. FEATURES............................................................ 1
2. SPECIFICATIONS................................................... 2
3. FRONT PANEL DESCRIPTION.................................. 5
4. PH CALIBRATING PROCEDURE...............................
4-1 Calibrating Consideration..................................
4-2 Requiring Equipment for calibration................... 7
4-3 Two Points Calibration...................................... 7
4-4 Singal Point Calibration..................................... 9
4-5 Others............................................................. 10
5. PH TEMPERATURE COMPENSATION........................ 11
5-1 Manual temperature compensation.................... 11
5-2 Automatic temperature compensation................15
6. MEASURING PROCEDURE.......................................15
6-1 PH Measurement..............................................15
6-2 mV Measurement............................................. 16
6-3 Temp. Measurement........................................
7. AUTO POWER DISABLE.......................................... 19
8. RS232 PC SERIAL INTERFACE.................................19
9. BATTERY REPLACEMENT........................................ 21
10. OPTIONAL PROBES & ACCESSORIES....................... 21
7 7
16
1. FEATURES
* Master meter is a professional PH/mV METER.
PH range : 0 to 14 PH x 0.01 PH. mV range : -1999 mV to 1999 mV.
* DPHC2001 can be plugged with the optional
Conductivity and TDS probe, ( YK-200PCT ) Conductivity probe, ( YK-200PCD )
Dissolved Oxygen probe, ( YK-200PDO ) to be a professional Conductivity meter, TDS ( Total Dissolved Solids) meter , or Dissolved Oxygen meter. After plugging in the new probe, no new calibration procedures are required. ** PLUG & PLAY FUNCTION**
* The instrument is with mV ( millivolt ) function for
mV measurement ( Be plugged with optinal ORP probe ORP-04 to be a professional ORP meter)
* Wide manual temperature compensation adjustment
can be easily operated by push button on the front panel.
* Optional ATC (Automatic Temp. Compensation) probe is
available for PH measurement.
* Microprocessor circuit assures high accuracy and reliable
performance.
* The instrument is patented with it's intelligent design concept. * Large LCD, dual function display. * Records Maximum and Minimum readings with recall. * Data hold. * Auto shut off saves battery life. * Powered by 006P DC 9V battery. * RS 232 computer serial interface.
1
2. SPEC
S
circuit.
* or can be converted by push button on the front panel. * PH calibration is easily to be done by push button on the
front panel.
* Using the durable, long-lasting components and a strong
lightweight ABS-plastic housing case.
* PH function with high input impedance avoids measuring
error.
* Wide applications: water conditioning, aquariums, beverage,
fish hatcheries, food processing, photography, laboratory, paper industry, plating industry, quality control, school & college.
IFICATION
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
Display 51 mm x 32 mm, dual function LCD display,
15 mm ( 0.6" ) digit size.
Measurement PH 0 to 14 PH
mV -1999 mV to 1999 mV Input 101  ohm Impedance Temperature Manual 0 to 100 , be adjusted by Compensation push button on front panel. for pH Automatic With the optional TEMP. measurement ( ATC ) probe ( YK-200PATC )
0 to 65 蚓.
pH PH7, PH4, and PH10, 3 points calibration Calibration ensure the best linearity and accuracy.
2
Intelligent * The instrument can be plugged with the
(
functions optional Conductivity and TDS probe
(YK-200PCT), Conductivity probe (YK-200PCD), Dissolved Oxygen probe (YK-200PDO) to become a professional Conductivity meter, TDS(Total Dissolved Solids) meter, and Dissolved meter. After plugging in the new probe, no new calibration procedures are required.
* The instrument is with mV ( millivolt )
function for mV measurement ( Be plugged with optional ORP probe ORP-04 to be a professional ORP meter).
Data hold Hold the current reading value on the display. Memory Maximum and Minimum reading values can Recall be saved and retrieved by record function.
Power off Auto shut off saves battery life, or manual
off by push button.
Data Output RS 232 computer serial interface. Overload "- - - -" symbol on the display. indication PH Optional, Electrode Any PH electrode with BNC connector.
Operating 0 to 50 C degree (32 to 122 F degree). Temperature Operating Max. 80% RH. Humidity
Sampling Time Approx. 0.8 second. Power Supply 006P DC 9V battery
Alkaline or Heavy duty type ).
Power Current Approx. DC 7 mA.
3
Weight 250 g/0.55 LB ( battery included ).
2
-
5
)
*
*
h
Size ( meter ) 195 x 68 x 30 mm ( 7.6 x 2.6 x 1.2 inch ).
Standard Instruction manual...........1 PC.
Accessories
Optional PH electrode, PE-03, PE-01, PE-11
Probes & ATC temp. probe, YK-200PATC Accessories ORP electrode, ORP-04
PH 7 buffer solution, PH-07
PH 4 buffer solution, PH-04
Conductivity and TDS probe, YK-200PCT
Conductivity probe, YK-200PCD
Dissolved oxygen probe, YK-200PDO
RS232 cable, UPCB-02
Application Software, SW-U801-WIN
2 Electrical Specifications ( 23
Measurement Range Resolution Accuracy
PH 0 to 14 PH 0.01 PH ± (0.02 PH + 2 d) mV 0 to 1999 mV 1 mV ± (0.5% + 2 d)
PH accuracy is based on calibrated meter only. Specification tests under the environment RF Field Strengt less than 3 V/M & frequency less than 30 MHz only.
4
3. FRONT PANEL DESCRIPTION
Fig. 1
5
Fig. 2 Fig. 3
T
3-1 Display 3-2 Power Button 3-3 REC. Button 3-4 Hold Button 3-5 C, F Button/A Button/up Button 3-6 PH/mV Button/B Button/left Button
3-7
EMP. C Button/D Button 3-8 CAL Button/C Button/down Button 3-9 Battery Compartment/Cover 3-10 PH BNC Input Socket 3-11 Lock Switch 3-12 Optional Probe Input Socket 3-13 RS-232 Out Terminal 3-14 Stand 3-15 PH 7 VR 3-16 PH 4/PH 10 VR
6
4. PH CALIBRATION PROCEDURE
4-1 Calibration Consideration
The most ideal PH ELECTRODE generates 0 mV at PH
7.00 ( 177.4 mV at PH 4 ) and YK-2001PH has been always calibrated with signals which simulate the most ideal PH ELECTRODE ( based on 25 ambient environment ). However not every PH ELECTRODE is as accurate as the most ideal one, so calibration procedures are necessary to be done before the first time measurement. In addition to the first time measurement, users are also recommended to execute the calibration procedures to ensure the high accuracy measurement.
4-2 Required Equipments for Calibration
1) PH ELECTRODE ( optional ).
2) PH buffer solutions ( optional ).
4-3 Two Points Calibration Procedure
1) Power on the instrument by pressing the " Power Button " ( 3-2, Fig. 1 ).
2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to let meter be operated under the PH function with a " PH " symbol on the display.
3) Adjust the " Temperature Compensation Value " to make it same as the temperature value of the pH buffer solution.
Manual temperature compensation value adjustment procedure, please refer to 5-1 (Page 10). Automatic temperature compensation value adjustment procedure, please refer to 5-2 (Page 14).
7
4) PH 7 calibration
Connect the PH ELECTRODE with the " BNC socket " ( 3-10, Fig. 1 ) and immerse the electrode in the PH7
buffer solution. Press the " CAL Button " ( 3-8, Fig. 1 ) then the upper display shows texts of " CAL " and the lower display shows the default calibration value.
CAL
7.00
* The texts " CAL " will flash for around 5 seconds.
After that, the meter calibrates itself automatically. The upper display will show the calibrated value, the lower display will show the temperature value.
7.00
25.0
5) PH 4 or PH 10 calibration
Rinse the electrode with distilled water.
Immerse the electrode in the PH4 buffer solution ( or PH10
buffer solution ). Press the " CAL Button " ( 3-8, Fig. 1 ) then the upper display shows texts of " CAL " and the lower display shows the default calibration value.
8
CAL
4.00
* The texts " CAL " will flash for around 5 seconds.
After that, the meter calibrate itself automatically. The upper display will show the calibrated value, the lower display will show the temperature value.
4.00
25.0
6) Rinse the electrode with distilled water again.
7) Repeat above (4) to (5) procedures two times at least.
8) The instrument and electrode are now finished the " TWO POINTS CALIBRATION " & ready for the measurement.
4-4 Singal Point Calibration
If PH4 and PH10 buffer solution are not available, single point (PH7) calibration can be executed from procedures of 4-3 (1) to (4). However for more accurate measuring result and linearity, two points calibration is always recommended.
9
4-5 Others
*
*
Above calibration procedures effect only when reading value within 1 PH of the calibration point. However if the reading value beyonds
*1 PH of PH 7 ( > PH8, < PH6 )
1 PH of PH 4 ( > PH5, < PH3 ) 1 PH of PH 10 ( > PH11, < PH9 )
the calibration procedures are :
1) Connect the PH ELECTRODE to the " PH BNC Input Socket " ( 3-10, Fig. 1 ).
2) Power on the instrument by pressing the " Power Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display
4) Set the " Manual temperature compensation " value to 25 refer to 5-1 calibration procedures, page 11.
5) Place the electrode into the standard solution ( PH7, PH4 or
PH10 ) , then the instrument will have the PH value on the
display.
6) * For measuring the PH7 standard solution, adjust the
PH 7 VR ( 3-15, Fig. 3 ) until the display value within PH6 to PH8.
* For measuring the PH4 standard solution, adjust the
PH 4 VR ( 3-16, Fig. 3 ) until the display value within PH3 to PH5.
* For measuring the PH10 standard solution, adjust the
PH 10 VR ( 3-16, Fig. 3 ) until the display value within PH9 to PH11.
7) The following calibration procedures will be same as 4-3 ( page 7 ) and 4-4 ( page 9 ).
10
5. PH TEMPERATURE COMPENSATION
Enable the meter to gain high accuracy measuring results from different kinds of solution, the temperature compensation calibration procedures are necessary to be executed. Manual temperature compensation calibration procedures
please refer to 5-1 ( See below).
Automatical temperature compensation calibration procedures please refer to 5-2 ( Page 15 ).
5-1 Manual temperature compensation procedures
Before the manual temperature compensation calibration procedures, please make sure there are no ATC probe ( YK-200ACT ) in the " Optional Probe Input Socket" ( 3-11, Fig. 1 ).
1) Power on the instrument by pressing the " Power Button " ( 3-2, Fig. 1 ).
2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a " PH " symbol on the display
3) " TEMP. C Button " ( 3-7, Fig. 1 ) is used to adjust the following values :
a. Tmperature compensation value b. PH 4 Default calibartion value c. PH 7 Default calibartion value d. PH 10 Default calibartion value
11
a. Adjust the temperature compensation value
* Press the " TEMP. C Button " ( 3-7, Fig. 1 ) first,
The upper display will show the measured PH values. the lower display will show the manual temperature compensation value.
PH
7.91
026.1
@ Use the " Left Button " ( 3-6, Fig. 1 ), " Up Button "
( 3-5, Fig. 1 ) and the " Down Button " ( 3-8, Fig. 1 ) to adjust the manual temperature compensation value.
b. Adjust the PH 4 default calibration value
* When the manual temp. compensation values
be adjusted. Press the " TEMP. C Button " ( 3-7, Fig. 1 )once again to adjust the PH 4 default calibartion value.
The upper display will show the values of " 4.00 " and the lower display will show the " PH 4 default calibration value ".
PH
4.00
04.03
12
@ Use the " Up Button " ( 3-5, Fig. 1 ) and the " Down
Button " ( 3-8, Fig. 1 ) to adjust the PH 4 default calibration value.
@ The adjustment range of " PH4 default calibration value "
is limited within 4.0 0.20 PH
c. Adjust the PH 7 default value
* When the PH 4 default value be adjusted. Press
the " TEMP. C Button " ( 3-7, Fig. 1 ) once again to adjust the PH 7 default calibration value.
The upper display will show the values of " 7.00 " and the lower display will show the " PH 7 default calibration value ".
PH
7.00
07.12
@ Use the " Up Button " ( 3-5, Fig. 1 ) and the "
Down Button " ( 3-8, Fig. 1 ) to adjust the PH 7 default calibration value.
@ The adjustment range of " PH 7 default calibration
value " is limited within 7.0 0.20 PH.
d. Adjust the PH 10 Default Value
* When the PH 7 default calibration value be
adjusted. Press the " TEMP. C Button " ( 3-7, Fig. 1 ) once again to adjust the PH 10 default calibration value.
13
The upper display will show the values of " 10.00 " and the lower display will show the " PH 10 default calibration value ".
PH
10.00
10.02
@ Use the " Up Button " ( 3-5, Fig. 1 ) and the " Down
Button " ( 3-8, Fig. 1 ) to adjust the PH 10 default calibration value.
@ The adjustment range of " PH10 default calibration
value " is limited within 10.0 0.20 PH
e. Finish the adjustment
* When the PH 10 default calibration value be
adjusted, press the " TEMP. C Button " ( 3-7, Fig. 1 ) once again to finish the manual temperature calibration procedures and return to the measuring mode.
Consideration :
If you want to skip any procedures above, just press " Temp. C Button " ( 3-7, Fig. 1 ). Above PH default calibration values you set will become the default value when you execute the PH Calibration Procedure. It is very convenient for user when PH 4.00, PH 7.00, and PH 10.00 buffer solution are not available.
14
5-2 Automatic temperature compensation
1) Plug the " Optional ATC Temp. Probe, YK-200ACT " into the " Optional Probe Input Socket " ( 3-12, Fig. 1 ).
Slide the " Lock Switch " ( 3-11, Fig. 1 ) to the lock ( ) position.
2) Power on the instrument by pressing the " Power Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display
4) Place the "Temp. Probe" into the solution, then
temperature will be compensated automatically for PH
measurement.
6. MEASURING PROCEDURE
Make sure the " Lock Switch " ( 3-11, Fig. 1 ) in the lock ( ) position before measurement.
6-1 PH Measurement
Whenever the calibration procedures are recommended to be done before PH measurement.
1) Connect the PH ELECTRODE to the " PH BNC Input Socket " ( 3-10, Fig. 1 ).
2) Power on the instrument by pressing the " Power Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display.
15
4) * If the operating is under the " Manual temperature
compensation ", then please refer to the above 5-1 calibration procedures.
* If the operating is under the " Automatic temperature
compensation ", then please refer to the above 5-2 calibration procedures.
5) Place the electrode into the solution, then the instrument
will have the PH value on the display.
6) After the measurement, please rinse the electrode with
distilled water.
6-2 mV Measurement
The instrument build in mV ( millivolt ) measurement function, which enable you to make ion-selective, ORP (oxidation-reduction potential), and other precise mV measurements. Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select the mV function with a " mV " symbol on the display.
6-3 Temp. Measurement
1) Plug the " Optional ATC Temp. Probe, YK-200PATC " into the " Optional Probe Input Socket " ( 3-12, Fig. 1 ).
* If you intend to measure " ", then press the " 蚓/ 蚌
2)
Button " ( 3-5, Fig. 1 ) and select the " " unit.
*If you intend to measure " ", then press the " 蚓/ 蚌
Button " ( 3-5, Fig. 1 ) and select the " " unit.
3) Place the " Temp. Probe " into the solution, and the instrument will have the temperature value on the display.
16
6-4 Data Hold
Press the " Hold Button " ( 3-4, Fig. 1 ) will hold the measured value & the LCD will indicate a " HOLD " symbol on the display during the measuring.
* Press the " Hold Button " again to exit the data hold
function.
6-5 Data Record ( Max., Min. reading )
* The data record function records the maximum and
minimum readings. Press the " REC. Button " ( 3-3, Fig. 1 ) to start the Data Record function and there will be a " REC " symbol on the display.
* With the " REC " symbol on the display :
a) Press the " REC. Button " ( 3-3, Fig. 1 ) once, the
" REC Max " symbol along with the maximum value will appear on the display.
If you intend to delete the maximum value, just press the " Hold Button " ( 3-4, Fig. 1 ) for a while, then the display will show the " REC " symbol only & execute the memory function continuously.
b) Press the " REC. Button " ( 3-3, Fig. 1 ) again, the
" REC Min " symbol along with the minimum value will appear on the display.
If you intend to delete the minimum value, just press the " Hold Button " ( 3-4, Fig. 1 ) for a while, then the display will show the " REC " symbol only & execute the memory function continuously.
c) To exit the memory record function, just press the
" REC " button for 2 seconds at least. The display will revert to the current reading.
17
6-6 Following are the block diagrams for quick
measuring procedures
Calibration
Set the Single point calibration Power manual PH 7(CAL.) ON temp.
values Two points calibration
or
plug PH4 ( PH10, SLOPE)
ATC probe
PH measuring procedures
Setting the Manual temp. M compensation values E
on front panel A Power S ON or U
Automatic temperature M
compensation by optional E
ATC probe N
PH 7(CAL.) &
R E
T
Optional measuring procedures
DATA HOLD MEMORY RECORD RS232 OUTPUT
Max., Min.
18
Power management
AUTO POWER OFF MANUAL POWER OFF
(Not activated during or
Memory Record Selection)
7. DISABLE AUTO POWER OFF
The instrument has " Auto Power Off " function in order to prolong battery life. The meter will shut off automatically if none of the buttons are pressed in approx. 10 min.
To disable this function, Select the memory record function during the measurement by pressing the " REC. Button " ( 3-3, Fig. 1 ).
8. RS232 PC SERIAL INTERFACE
The instrument features RS232 output via
3.5 mm Terminal ( 3-13, Fig. 1 ).
The signals output is a 16 digits data stream which can be utilized for user's specific application.
A RS232 lead with the following connection will be required to link the instrument with the PC serial interface.
19
Meter PC
K
4
(3.5 mm jack plug) (9W 'D" Connector)
Center Pin.............................. Pin 4 Pin 2
2.2
Ground/shield............................Pin 2 Pin 5 resistor
The 16 digits data stream will be displayed in the following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
Each digit indicates the following status :
D0 End Word D1 & D8 Display reading, D1 = LSD, D8 = MSD
For example : If the display reading is 1234, then D8 to D1 is :
0000123
D9 Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D10 Polarity
0 = Positive 1 = Negative
D11 & D12 Annunciator for Display
= 01 mV = 18 mS = 14 PPM = 19 OF2= 06 mg/L = 07
D13 When send the upper display data = 1
When send the lower display data = 2
D14 4 D15 Start Word
= 02 PH = 05
RS232 FORMAT : 9600, N, 8, 1
20
9. BATTERY REPLACEMENT
e
蚓0 蚓
蚌32蚌
1) When the left corner of LCD display show " ", it is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
2) Slide the " Battery Cover " (3-14, Fig. 1) away from the instrument and remove the battery.
3) Replace with 9V battery ( Alkaline or Heavy duty type ) and reinstate the cover.
4) Make sure the battery cover is secured after changing the battery.
10. OPTIONAL PROBES & ACCESSORIES
DISSOLVED * YK-200PDO plug into the OXYGEN YK-2001PH to become a PROBE professional Dissolved Oxygen
Meter. Model : YK-200PDO
MeasurementRang
DO 0 to 20.0 mg/L
to 50
to 122
21
CONDUCTIVIT
Y
e
蚓0 蚓
蚌32蚌
Y
e
蚓0 蚓
蚌32蚌
PROBE YK-2001PH to become a
Model : YK-200PCD
* YK-200PCD plug into the
professional Conductivity Meter.
MeasurementRang
Conductivity 2 mS
Conductivity 20 mS
to 60
to 140
CONDUCTIVIT and TDS YK-2001PH to become a PROBE professional Conductivity & TDS
Model : YK-200PCT
ORP ELECTRODE ORP electrode ORP-04 plug into the
Model : function ) to become a professional ORP-04 ORP (oxidation-reduction potential)
BUFFER SOLUTION PH 7.00 standard buffer solution. PH-07 for calibration purpose. BUFFER SOLUTION PH 4.00 standard buffer solution. PH-04 for calibration purpose.
* YK-200PCT plug into the
Meter.
MeasurementRang
Conductivity 2 mS
Conductivity 20 mS
PPM 2,000 PPM
PPM 20,000 PPM
to 60
to 140
YK-2001PH ( select to the mV
Meter.
22
ATC * YK-200PATC plug into the
e
蚓0 蚓
蚌32蚌
&
&
&
T
PROBE YK-2001PH to be a ATC
(Automatic Temperature
Compensation ) Probe for PH Model : function. YK-200PATC
MeasurementRang
to 65
to 149
CARRYING CASE Hard carrying case CA-06
PH ELECTRODE General purpose, laboratory PE-03 field usage. 12.3 mm dia. x 160 mm.
Epoxy body, 1 - 13 pH. PH ELECTRODE General purpose, laboratory PE-11 field usage. 10 mm dia. x 130 mm.
Epoxy body, 1 - 13 pH.
(0 - 14 pH typical) PH ELECTRODE Professional, laboratory PE-01 field usage. 9.5 mm dia. x 130 mm.
Epoxy body, 0 - 14 pH. SPEAR TIP PH ELECTRODE perfect for those pH measurements in
PE-04HD required. Meat, sausage and cheese
he " Spear Tip pH electrode " is
applications where sample piercing is
are ideal applications.
Measuring range : 0 to 14 pH .
Size 12 mm dia. x 157 mm.
RS232 cable RS232 cable for connecting between UPCB-02 the meter & the computer. SOFTWARE Windows version application software SW-U801-WIN applies as the performance of data
logging system & data recorder...
23