Dissolved Oxygen probe, ( YK-200PDO )
to be a professional Conductivity meter, TDS ( Total
Dissolved Solids) meter , or Dissolved Oxygen meter. After
plugging in the new probe, no new calibration procedures are
required. ** PLUG & PLAY FUNCTION**
* The instrument is with mV ( millivolt ) function for
mV measurement ( Be plugged with optinal ORP probe
ORP-04 to be a professional ORP meter)
* Wide manual temperature compensation adjustment
can be easily operated by push button on the front panel.
* Optional ATC (Automatic Temp. Compensation) probe is
available for PH measurement.
* Microprocessor circuit assures high accuracy and reliable
performance.
* The instrument is patented with it's intelligent design concept.
* Large LCD, dual function display.
* Records Maximum and Minimum readings with recall.
* Data hold.
* Auto shut off saves battery life.
* Powered by 006P DC 9V battery.
* RS 232 computer serial interface.
1
2. SPEC
S
circuit.
* 蚓 or 蚌 can be converted by push button on the front panel.
* PH calibration is easily to be done by push button on the
front panel.
* Using the durable, long-lasting components and a strong
lightweight ABS-plastic housing case.
* PH function with high input impedance avoids measuring
error.
* Wide applications: water conditioning, aquariums, beverage,
fish hatcheries, food processing, photography, laboratory,
paper industry, plating industry, quality control, school &
college.
IFICATION
2-1 General Specifications
CircuitCustom one-chip of microprocessor LSI
Display51 mm x 32 mm, dual function LCD display,
15 mm ( 0.6" ) digit size.
MeasurementPH0 to 14 PH
mV-1999 mV to 1999 mV
Input101 ohm
Impedance
TemperatureManual0 to 100 蚓, be adjusted by
Compensation push button on front panel.
for pHAutomaticWith the optional TEMP.
measurement( ATC )probe ( YK-200PATC )
0 to 65 蚓.
pH PH7, PH4, and PH10, 3 points calibration
Calibrationensure the best linearity and accuracy.
2
Intelligent * The instrument can be plugged with the
(
functionsoptional Conductivity and TDS probe
(YK-200PCT), Conductivity probe
(YK-200PCD), Dissolved Oxygen probe
(YK-200PDO) to become a professional
Conductivity meter, TDS(Total Dissolved
Solids) meter, and Dissolved meter.
After plugging in the new probe, no new
calibration procedures are required.
* The instrument is with mV ( millivolt )
function for mV measurement ( Be plugged
with optional ORP probe ORP-04 to be
a professional ORP meter).
Data holdHold the current reading value on the display.
Memory Maximum and Minimum reading values can
Recallbe saved and retrieved by record function.
Power offAuto shut off saves battery life, or manual
off by push button.
Data OutputRS 232 computer serial interface.
Overload"- - - -" symbol on the display.
indication
PH Optional,
ElectrodeAny PH electrode with BNC connector.
Operating 0 to 50 C degree (32 to 122 F degree).
Temperature
Operating Max. 80% RH.
Humidity
Sampling Time Approx. 0.8 second.
Power Supply006P DC 9V battery
Alkaline or Heavy duty type ).
Power Current Approx. DC 7 mA.
3
Weight250 g/0.55 LB ( battery included ).
2
-
5
蚓
)
*
*
h
Size ( meter ) 195 x 68 x 30 mm ( 7.6 x 2.6 x 1.2 inch ).
PH0 to 14 PH0.01 PH± (0.02 PH + 2 d)
mV0 to 1999 mV1 mV± (0.5% + 2 d)
PH accuracy is based on calibrated meter only.
Specification tests under the environment RF Field Strengt
less than 3 V/M & frequency less than 30 MHz only.
7.00 ( 177.4 mV at PH 4 ) and YK-2001PH has been
always calibrated with signals which simulate the most ideal
PH ELECTRODE ( based on 25蚓 ambient environment
). However not every PH ELECTRODE is as accurate as
the most ideal one, so calibration procedures are
necessary to be done before the first time measurement.
In addition to the first time measurement, users are also
recommended to execute the calibration procedures to
ensure the high accuracy measurement.
4-2 Required Equipments for Calibration
1) PH ELECTRODE ( optional ).
2) PH buffer solutions ( optional ).
4-3 Two Points Calibration Procedure
1) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to let meter be
operated under the PH function with a " PH " symbol on the
display.
3) Adjust the " Temperature Compensation Value " to make
it same as the temperature value of the pH buffer
solution.
Manual temperature compensation value adjustment
procedure, please refer to 5-1 (Page 10).
Automatic temperature compensation value adjustment
procedure, please refer to 5-2 (Page 14).
7
4) PH 7 calibration
Connect the PH ELECTRODE with the " BNC socket "
( 3-10, Fig. 1 ) and immerse the electrode in the PH7
buffer solution.
Press the " CAL Button " ( 3-8, Fig. 1 ) then the upper
display shows texts of " CAL " and the lower display
shows the default calibration value.
CAL
7.00
* The texts " CAL " will flash for around 5 seconds.
After that, the meter calibrates itself automatically.
The upper display will show the calibrated value,
the lower display will show the temperature value.
7.00
25.0
5) PH 4 or PH 10 calibration
Rinse the electrode with distilled water.
Immerse the electrode in the PH4 buffer solution ( or PH10
buffer solution ).
Press the " CAL Button " ( 3-8, Fig. 1 ) then the upper
display shows texts of " CAL " and the lower display
shows the default calibration value.
8
CAL
4.00
* The texts " CAL " will flash for around 5 seconds.
After that, the meter calibrate itself automatically.
The upper display will show the calibrated value,
the lower display will show the temperature value.
4.00
25.0
6) Rinse the electrode with distilled water again.
7) Repeat above (4) to (5) procedures two times at least.
8) The instrument and electrode are now finished the
" TWO POINTS CALIBRATION " & ready for the
measurement.
4-4 Singal Point Calibration
If PH4 and PH10 buffer solution are not available, single
point (PH7) calibration can be executed from procedures
of 4-3 (1) to (4). However for more accurate measuring
result and linearity, two points calibration is always
recommended.
9
4-5 Others
*
*
Above calibration procedures effect only when reading value
within 1 PH of the calibration point.
However if the reading value beyonds
1) Connect the PH ELECTRODE to the " PH BNC Input
Socket " ( 3-10, Fig. 1 ).
2) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display
4) Set the " Manual temperature compensation " value to 25 蚓
refer to 5-1 calibration procedures, page 11.
5) Place the electrode into the standard solution ( PH7, PH4 or
PH10 ) , then the instrument will have the PH value on the
display.
6) * For measuring the PH7 standard solution, adjust the
PH 7 VR ( 3-15, Fig. 3 ) until the display value
within PH6 to PH8.
* For measuring the PH4 standard solution, adjust the
PH 4 VR ( 3-16, Fig. 3 ) until the display value
within PH3 to PH5.
* For measuring the PH10 standard solution, adjust the
PH 10 VR ( 3-16, Fig. 3 ) until the display value
within PH9 to PH11.
7) The following calibration procedures will be same as 4-3
( page 7 ) and 4-4 ( page 9 ).
10
5. PH TEMPERATURE COMPENSATION
Enable the meter to gain high accuracy measuring results
from different kinds of solution, the temperature
compensation calibration procedures are necessary to
be executed.
Manual temperature compensation calibration procedures
please refer to 5-1 ( See below).
Automatical temperature compensation calibration procedures
please refer to 5-2 ( Page 15 ).
5-1 Manual temperature compensation procedures
Before the manual temperature compensation calibration
procedures, please make sure there are no ATC probe
( YK-200ACT ) in the " Optional Probe Input Socket"
( 3-11, Fig. 1 ).
1) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a " PH " symbol on the display
3) " TEMP. C Button " ( 3-7, Fig. 1 ) is used to adjust the
following values :
a. Tmperature compensation value
b. PH 4 Default calibartion value
c. PH 7 Default calibartion value
d. PH 10 Default calibartion value
11
a. Adjust the temperature compensation value
* Press the " TEMP. C Button " ( 3-7, Fig. 1 ) first,
The upper display will show the measured PH values.
the lower display will show the manual temperature
compensation value.
PH
7.91
026.1 蚓
@ Use the " Left Button " ( 3-6, Fig. 1 ), " Up Button "
( 3-5, Fig. 1 ) and the " Down Button " ( 3-8, Fig. 1 ) to
adjust the manual temperature compensation value.
b. Adjust the PH 4 default calibration value
* When the manual temp. compensation values
be adjusted. Press the " TEMP. C Button "
( 3-7, Fig. 1 )once again to adjust the PH 4
default calibartion value.
The upper display will show the values of " 4.00 " and
the lower display will show the " PH 4 default calibration
value ".
PH
4.00
04.03
12
@ Use the " Up Button " ( 3-5, Fig. 1 ) and the " Down
Button " ( 3-8, Fig. 1 ) to adjust the PH 4 default
calibration value.
@ The adjustment range of " PH4 default calibration value "
is limited within 4.0 0.20 PH
c. Adjust the PH 7 default value
* When the PH 4 default value be adjusted. Press
the " TEMP. C Button " ( 3-7, Fig. 1 ) once
again to adjust the PH 7 default calibration
value.
The upper display will show the values of " 7.00 " and
the lower display will show the " PH 7 default
calibration value ".
PH
7.00
07.12
@ Use the " Up Button " ( 3-5, Fig. 1 ) and the "
Down Button " ( 3-8, Fig. 1 ) to adjust the PH 7
default calibration value.
@ The adjustment range of " PH 7 default calibration
value " is limited within 7.0 0.20 PH.
d. Adjust the PH 10 Default Value
* When the PH 7 default calibration value be
adjusted. Press the " TEMP. C Button " ( 3-7,
Fig. 1 ) once again to adjust the PH 10 default
calibration value.
13
The upper display will show the values of " 10.00 " and
the lower display will show the " PH 10 default
calibration value ".
PH
10.00
10.02
@ Use the " Up Button " ( 3-5, Fig. 1 ) and the " Down
Button " ( 3-8, Fig. 1 ) to adjust the PH 10 default
calibration value.
@ The adjustment range of " PH10 default calibration
value " is limited within 10.0 0.20 PH
e. Finish the adjustment
* When the PH 10 default calibration value be
adjusted, press the " TEMP. C Button " ( 3-7,
Fig. 1 ) once again to finish the manual
temperature calibration procedures and return to
the measuring mode.
Consideration :
If you want to skip any procedures above, just press
" Temp. C Button " ( 3-7, Fig. 1 ).
Above PH default calibration values you set will
become the default value when you execute the PH
Calibration Procedure. It is very convenient for user
when PH 4.00, PH 7.00, and PH 10.00 buffer
solution are not available.
14
5-2 Automatic temperature compensation
1) Plug the " Optional ATC Temp. Probe, YK-200ACT "
into the " Optional Probe Input Socket " ( 3-12, Fig. 1 ).
Slide the " Lock Switch " ( 3-11, Fig. 1 ) to the
lock ( ) position.
2) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display
4) Place the "Temp. Probe" into the solution, then
temperature will be compensated automatically for PH
measurement.
6. MEASURING PROCEDURE
Make sure the " Lock Switch " ( 3-11, Fig. 1 ) in
the lock ( ) position before measurement.
6-1 PH Measurement
Whenever the calibration procedures are recommended to
be done before PH measurement.
1) Connect the PH ELECTRODE to the " PH BNC Input
Socket " ( 3-10, Fig. 1 ).
2) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display.
15
4) * If the operating is under the " Manual temperature
compensation ", then please refer to the above 5-1
calibration procedures.
* If the operating is under the " Automatic temperature
compensation ", then please refer to the above 5-2
calibration procedures.
5) Place the electrode into the solution, then the instrument
will have the PH value on the display.
6) After the measurement, please rinse the electrode with
distilled water.
6-2 mV Measurement
The instrument build in mV ( millivolt ) measurement
function, which enable you to make ion-selective, ORP
(oxidation-reduction potential), and other precise mV
measurements.
Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the mV function with a " mV " symbol on the display.
6-3 Temp. Measurement
1) Plug the " Optional ATC Temp. Probe, YK-200PATC "
into the " Optional Probe Input Socket " ( 3-12, Fig. 1 ).
* If you intend to measure " 蚓 ", then press the " 蚓/ 蚌
2)
Button " ( 3-5, Fig. 1 ) and select the " 蚓 " unit.
*If you intend to measure " 蚌 ", then press the " 蚓/ 蚌
Button " ( 3-5, Fig. 1 ) and select the " 蚌 " unit.
3) Place the " Temp. Probe " into the solution, and the instrument
will have the temperature value on the display.
16
6-4 Data Hold
Press the " Hold Button " ( 3-4, Fig. 1 ) will hold the
measured value & the LCD will indicate a " HOLD " symbol
on the display during the measuring.
* Press the " Hold Button " again to exit the data hold
function.
6-5 Data Record ( Max., Min. reading )
* The data record function records the maximum and
minimum readings. Press the " REC. Button " ( 3-3, Fig. 1 )
to start the Data Record function and there will be a
" REC " symbol on the display.
* With the " REC " symbol on the display :
a) Press the " REC. Button " ( 3-3, Fig. 1 ) once, the
" REC Max " symbol along with the maximum value
will appear on the display.
If you intend to delete the maximum value, just press
the " Hold Button " ( 3-4, Fig. 1 ) for a while, then
the display will show the " REC " symbol only &
execute the memory function continuously.
b) Press the " REC. Button " ( 3-3, Fig. 1 ) again, the
" REC Min " symbol along with the minimum value
will appear on the display.
If you intend to delete the minimum value, just press
the " Hold Button " ( 3-4, Fig. 1 ) for a while, then
the display will show the " REC " symbol only &
execute the memory function continuously.
c) To exit the memory record function, just press the
" REC " button for 2 seconds at least. The display will
revert to the current reading.
17
6-6 Following are the block diagrams for quick
measuring procedures
Calibration
Set theSingle point calibration
PowermanualPH 7(CAL.)
ONtemp.
valuesTwo points calibration
or
plugPH4 ( PH10, SLOPE)
ATC probe
PH measuring procedures
Setting the Manual temp.M
compensation valuesE
on front panelA
PowerS
ONorU
Automatic temperature M
compensation by optionalE
ATC probeN
PH 7(CAL.) &
R
E
T
Optional measuring procedures
DATA HOLDMEMORY RECORDRS232 OUTPUT
Max., Min.
18
Power management
AUTO POWER OFFMANUAL POWER OFF
(Not activated duringor
Memory Record Selection)
7. DISABLE AUTO POWER OFF
The instrument has " Auto Power Off " function in
order to prolong battery life. The meter will shut
off automatically if none of the buttons are pressed
in approx. 10 min.
To disable this function, Select the memory record
function during the measurement by pressing the
" REC. Button " ( 3-3, Fig. 1 ).
8. RS232 PC SERIAL INTERFACE
The instrument features RS232 output via
3.5 mm Terminal ( 3-13, Fig. 1 ).
The signals output is a 16 digits data stream which
can be utilized for user's specific application.
A RS232 lead with the following connection will
be required to link the instrument with the PC
serial interface.
19
MeterPC
K
4
蚓
蚌
(3.5 mm jack plug)(9W 'D" Connector)
Center Pin.............................. Pin 4Pin 2
For example :
If the display reading is 1234, then D8 to D1 is :
0000123
D9Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D10Polarity
0 = Positive1 = Negative
D11 & D12Annunciator for Display
= 01
mV = 18mS = 14PPM = 19
OF2= 06mg/L = 07
D13When send the upper display data = 1
When send the lower display data = 2
D144
D15Start Word
= 02PH = 05
RS232 FORMAT : 9600, N, 8, 1
20
9. BATTERY REPLACEMENT
e
蚓0 蚓
蚓
蚌32蚌
蚌
1) When the left corner of LCD display show " ", it
is necessary to replace the battery. However, in-spec.
measurement may still be made for several hours after
low battery indicator appears before the instrument
become inaccurate.
2) Slide the " Battery Cover " (3-14, Fig. 1) away from the
instrument and remove the battery.
3) Replace with 9V battery ( Alkaline or Heavy duty type )
and reinstate the cover.
4) Make sure the battery cover is secured after changing the
battery.
10. OPTIONAL PROBES & ACCESSORIES
DISSOLVED * YK-200PDO plug into the
OXYGENYK-2001PH to become a
PROBEprofessional Dissolved Oxygen
Meter.
Model :
YK-200PDO
MeasurementRang
DO0 to 20.0 mg/L
to 50
to 122
21
CONDUCTIVIT
Y
e
蚓0 蚓
蚓
蚌32蚌
蚌
Y
e
蚓0 蚓
蚓
蚌32蚌
蚌
PROBEYK-2001PH to become a
Model :
YK-200PCD
* YK-200PCD plug into the
professional Conductivity Meter.
MeasurementRang
Conductivity2 mS
Conductivity20 mS
to 60
to 140
CONDUCTIVIT
and TDSYK-2001PH to become a
PROBEprofessional Conductivity & TDS
Model :
YK-200PCT
ORP ELECTRODEORP electrode ORP-04 plug into the
Model :function ) to become a professional
ORP-04ORP (oxidation-reduction potential)
BUFFER SOLUTIONPH 7.00 standard buffer solution.
PH-07for calibration purpose.
BUFFER SOLUTIONPH 4.00 standard buffer solution.
PH-04for calibration purpose.
* YK-200PCT plug into the
Meter.
MeasurementRang
Conductivity2 mS
Conductivity20 mS
PPM2,000 PPM
PPM20,000 PPM
to 60
to 140
YK-2001PH ( select to the mV
Meter.
22
ATC* YK-200PATC plug into the
e
蚓0 蚓
蚓
蚌32蚌
蚌
&
&
&
T
PROBEYK-2001PH to be a ATC
(Automatic Temperature
Compensation ) Probe for PH
Model :function.
YK-200PATC
MeasurementRang
to 65
to 149
CARRYING CASEHard carrying case
CA-06
PH ELECTRODEGeneral purpose, laboratory
PE-03field usage. 12.3 mm dia. x 160 mm.
Epoxy body, 1 - 13 pH.
PH ELECTRODEGeneral purpose, laboratory
PE-11field usage. 10 mm dia. x 130 mm.
Epoxy body, 1 - 13 pH.
(0 - 14 pH typical)
PH ELECTRODEProfessional, laboratory
PE-01field usage. 9.5 mm dia. x 130 mm.
Epoxy body, 0 - 14 pH.
SPEAR TIP
PH ELECTRODEperfect for those pH measurements in
PE-04HDrequired. Meat, sausage and cheese
he " Spear Tip pH electrode " is
applications where sample piercing is
are ideal applications.
Measuring range : 0 to 14 pH .
Size 12 mm dia. x 157 mm.
RS232 cableRS232 cable for connecting between
UPCB-02the meter & the computer.
SOFTWAREWindows version application software
SW-U801-WINapplies as the performance of data
logging system & data recorder...