* Conductivity/TDS ( Total Dissolved Solids ) Meter
+ PH Meter, multi functions & intelligent features.
* Conductivity includs 2 ranges : 2 mS, 20 mS.
TDS has 2 ranges : 2,000 PPM, 20,000 PPM.
Temperature measurement range from 0 - 60 蚓 /
32 - 140蚌.
* Conductivity/TDS meter build in automatic temperature
compensation circuit, adjustable between 0 to 5.0%
per 蚓.
* Conductivity/TDS meter uses carbon rod electrode with
long life using separated pobe is convenient for remote
measurement.
* PH meter has mV ( millivolt ) function for
mV measurement ( Be plugged with optional ORP probe
ORP-04 to be a professional ORP meter)
* PH meter with wide manual temperature compensation
adjustment and optional ATC (Automatic Temp.
Compensation) probe is available.
* Microprocessor circuit, intelligent function.
* Records maximum and minimum readings with recall.
* Data hold.
* Auto shut off saves battery life.
* RS 232 PC serial interface.
* Temperature function with 蚓 & 蚌 display unit.
1
2. SPECIFICATIONS
2-1 General specifications
CircuitMicroprocessor LSI circuit.
Display51 mm x 32 mm, dual function LCD display,
15 mm ( 0.6" ) digit size.
Measurement* Conductivity
* TDS ( Total Dissolved Solids )
* Temperature ( use the CD/TDS probe )
* PH
* mV
Memory Records maximum and minimum
Recallreadings with recall.
Data holdHold the current reading value on the display.
Memory Maximum and minimum reading values can
Recallbe saved and retrieved by record function.
Power offAuto shut off saves battery life, or manual
off by push button.
Data OutputRS 232 computer serial interface.
Overload"- - - -" symbol on the display.
indication
Operating 0 to 50 蚓 - main instrument.
Temperature0 to 60 蚓 - Conductivity/CT probe only.
Operating Max. 80% RH.
Humidity
Sampling Time Approx. 0.8 second.
Power Supply006P DC 9V battery
( Alkaline or Heavy duty type ).
Power Current Meter onlyApprox. DC 4.2 mA.
Meter + CD/TDS probeApprox. DC 6.0 mA.
Weight250 g/0.55 LB ( battery included ).
AccessoriesATC probe for PH meter.......... YK-200PATC
ORP electrode.........................ORP-04
RS232 cable, UPCB-02............ UPCB-02
Application Software............SW-U801-WIN
2-2. Conductivity/TDS/Temp. specifications
A. Conductivity
Range Measurement Resolution Accuracy
2 mS0.2 to 2.000 mS0.001 mS( 3 % F.S. + 1 d )
20 mS2 to 20.00 mS 0.01 mS
* F.S. - Full scale
* Temperature Compensation :
Automatic from 0 to 60 蚓 ( 32 - 140 蚌 ), with temperature
compensation factor variable between 0 to 5.0% per C.
* mS - milli Simens
* @ 23 5蚓
3
B. TDS ( Total Dissolved Dolids )
e
e
蚓
(
蚌
)
e
n
s
T
蚓
Rang
2,000 132 to 1,3201 PPM
PPMPPM( 3 % F.S. + 1 d )
20,000 1,320 to 13,20010 PPM
PPMPPM
Measurement ResolutionAccuracy
* F.S. - Full scal
* Temperature Compensation :
Automatic from 0 to 60
32 - 140
, with temperatur
compensation factor variable between 0 to 5.0% per C.
* PPM - parts per millio
* @ 23 5蚓
C. Temperature
Function Measuring Range Resolution Accuracy
蚓0 蚓 to 60 蚓0.1 蚓 0.8 蚓
蚌 32 蚌 to 140 蚌0.1 蚌 1.5 蚌
* @ 23 5蚓
2-3 PH meter specification
MeasurementPH0 to 14 PH
mV-1999 mV to 1999 mV
ResolutionPH0.01 PH
mV1 mV
AccuracyPH0.02 PH + 2 d)
* Meter only
Input10 -12 ohm
Impedance
emperatureManual0 to 100 蚓, be adjusted by
Compensation push button on front panel.
for pHAutomaticWith the optional TEMP.
measurement( ATC )probe ( YK-200PATC )
pH PH7, PH4, and PH10, 3 points calibration
Calibrationensure the best linearity and accuracy.
PH Optional,
ElectrodeAny PH electrode with BNC connector.
mV0.5% + 2 d)
0 to 65
.
4
3. FRONT PANEL DESCRIPTION
Fig. 1
5
3-1Display
A
蚓,蚌
R
R
3-2Power Button
3-3 REC. Button
3-4Hold Button
3-5
3-6 B Button ( Range Button )
3-7C Button ( CD/TDS Button )
3-8D Button ( TEMP. C Button )
3-9 Battery Compartment/Cover
3-10PH BNC Input Socket
3-11 Lock Switch
3-12 Conductivity/TDS Input Socket
3-13 RS-232 Out Terminal
3-14Stand
3-15PH 7 V
3-16PH 4/PH 10 V
3-17Conductivity/TDS Probe
3-18Conductivity/TDS Probe pLug
into the " Conductivity/TDS Socket " ( 3-12, Fig. 1 )
Make sure that the probe lock switch ( 3-11, Fig. 1 )
should slide to the lock position ( ).
2) Power on the instrument by pushing the " Power Button "
( 3-2, Fig.1 )
3) Push the " CD/TDS " Button ( C Button ) ( 3-7, Fig. 1 ), and
select the function to the conductivity measurement. The
display will show the unit of " mS ".
4) The instrument will default to 2% per 蚓 Temperature
Compensation factor. The meter has built-in Automatic
Temperature Compensation adjustable between 0 to 5.0 %
per 蚓.
In order to change the default value carry out the
following procedures :
a. Push the " Temp. C Button ( D Button ) " ( 3-8, Fig. 1 ),
the display will show :
o
mS1 蚓
7
b.
)
)
Use the
" up Button ( A Button ) " ( 3-5, Fig. 1 ),
" left Button ( B Button ) " ( 3-6, Fig. 1 )
" down Button ( C Button ) " ( 3-7, Fig. 1 )
to select the desired value of Temperature
Compensation factor.
c. Once the desired value is reached, push the " Temp.
C Button ( D Button ) " ( 3-8, Fig. 1 ) to set the new
value.
5) Select the applicable range, by using the " Range Button
( B Button ) " ( 3-6, Fig. 1 ).
* If the display shows "- - - -", it indicates an overload
condition, select the next higher range.
* If the display shows "_ _ _ _", it indicates an
out-of-range condition, select the next lower range.
6) Immerse the head of " Conductivity/TDS Probe " ( 3-17, Fig. 1
into the solution, up to the immersion level.
7
During the measurement, the lower LCD Display will
show the temperature of the solution.
* Push the " 蚓/蚌 Button " ( 3-5, Fig. 1 ) to change
the temperature display unit from " 蚓 to 蚌 "
or " 蚌 to 蚓 ".
4-2 TDS measurement
The operation procedures are same as the above 4-1,
except that push the " CD/TDS " Button ( C Button ) ( 3 -7,
Fig. 1 ), and select the function to the TDS measurement. The
lower display will show the text of " P ".
8
4-3 Calibration procedures of Conductivity meter
The meter has been calibrated during manufacture.
However, it may be necessary to re-calibrate periodically.
Particularly when the instrument is used for a long period or
when the conductivity electrode is changed.
To re-calibrate the instrument, follow the procedures
shown below :
Range 1 ( 2 mS range ) calibration
1) Prepare a " 1.413 mS Calibration Solution ".
2) Immerse the head of " Conductivity Probe " ( 3-17, Fig. 1 )
into the 1,413 mS solution up to the immersion level.
3) Select the range to the " 2 mS " position.
4) Set the temperature coefficient factor value to " 2.0% per
蚓 "
5) At the same time push the following buttons together :
a. REC.( MAX./MIN. ) Button ( 3-3, Fig. 1 )
b. HOLD Button ( 3-4, Fig. 1 )
The upper display will show the flashing text of " CAL ".
The low display will show the flashing text of " 1.413 ".
6) Please release the two buttons when the display shows as
above and start flashing and then when the text stop
flashing, the range 1 will be calibrated to 1.413 mS
( There maybe a little deviation, and it is normal ).
*
After released two buttons but before the text (CAL,
1.413 ) stop flashing, if push the " HOLD Button " will
cancel the calibration procedures, the meter will return
to previous calibration value.
9
Range 2 ( 20 mS range ) calibration
1) Prepare a " 12.88 mS Calibration Solution ".
2) Immerse the head of " Conductivity Probe " ( 3-17, Fig. 1 )
into the 12.88 mS solution up to the immersion level.
3) Select the range to the " 20 mS " position.
4) Set the temperature coefficient factor value to " 2.0% per
蚓 "
5) At the same time push the following buttons together :
a. REC.( MAX./MIN. ) Button ( 3-3, Fig. 1 )
b. HOLD Button ( 3-4, Fig. 1 )
The upper display will show the flashing text of " CAL ".
The low display will show the flashing text of " 12.88 "
6) Please release the two buttons when the display shows as
above and start flashing and then when the text stop
flashing, the range 2 will be calibrated to 12.88 mS
( There maybe a little deviation, and it is normal ).
*
After released two buttons but before the text
(CAL, 12.88 ) stop flashing, if push the " HOLD Button
" will cancel the calibration procedures, the meter will
return to previous calibration value.
10
4-4 PH meter
* Make sure that the probe lock switch ( 3-11, Fig. 1 )
had switched to the lock position ( ).
* Make sure that the " Conductivity/TDS Plug "
( 3-18, Fig. 1 ) had taken away from the meter.
a. PH Measurement
Whenever the calibration procedures are recommended to
be done before PH measurement.
1) Connect the PH ELECTRODE to the " PH BNC Input
Socket " ( 3-10, Fig. 1 ).
2) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display.
4) * If the operating is under the " Manual temperature
compensation ", then please refer to the " 4-4, d "
calibration procedures.
* If the operating is under the " Automatic temperature
compensation ", then please refer to the " 4-4, e "
calibration procedures.
5) Place the electrode into the solution, and then the instrument
will have the PH value on the display.
6) After the measurement, please rinse the electrode with
distilled water.
11
b. mV Measurement
The instrument built in mV ( millivolt ) measurement
function, which enable you to make ion-selective, ORP
(oxidation-reduction potential), and other precise mV
measurements.
Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the mV function with a " mV " symbol on the display.
c. Temp. Measurement
1) Plug the " Optional ATC Temp. Probe, YK-200PATC "
into the " Probe Input Socket " ( 3-12, Fig. 1 ).
2) * If you intend to measure " 蚓 ", then press the " 蚓/ 蚌
Button " ( 3-5, Fig. 1 ) and select the " 蚓 " unit.
* If you intend to measure " 蚌 ", then press the " 蚓/ 蚌
Button " ( 3-5, Fig. 1 ) and select the " 蚌 " unit.
3) Place the " Temp. Probe " into the solution, and the instrument
will have the temperature value on the display.
d. Manual temperature compensation
Before the manual temperature compensation adjust
procedures, make sure there are no ATC probe or other
probe in the " Probe Input Socket" ( 3-12, Fig. 1 ).
1) Power on the instrument by pressing the " Power
Button " ( 3-2, Fig. 1 ).
12
2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a " PH " symbol on the display
* Press the " TEMP. C Button " ( 3-8, Fig. 1 ) first,
The upper display will show the measured PH values.
the lower display will show the manual temperature
compensation value.
* Use the " Left Button " ( 3-6, Fig. 1 ), " Up Button "
( 3-5, Fig. 1 ) and the " Down Button " ( 3-7, Fig. 1 ) to
adjust the manual temperature compensation value.
e. Automatic temperature compensation
1) Plug the " Optional ATC Temp. Probe, YK-200PACT "
into the " Probe Input Socket " ( 3-12, Fig. 1 ).
Make sure that the probe lock switch ( 3-11, Fig. 1 )
should slide to the lock position ( ).
2) Power on the meter.
3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
the PH function with a "PH" symbol on the display
4) Place the "Temp. Probe" into the solution, then
temperature will be compensated automatically for PH
measurement.
13
f. Two Points Calibration
PH 7 calibration
Connect the PH ELECTRODE with the " BNC socket "
( 3-10, Fig. 1 ) and immerse the electrode in the PH7
buffer solution ( optional ).
Press the " CAL Button " ( 3-7, Fig. 1 ) then the upper
display shows texts of " CAL " and the lower display
shows the default calibration value.
* The texts " CAL " will flash for around 5 seconds.
After that, the meter calibrates itself automatically.
The upper display will show the calibrated value,
the lower display will show the temperature value.
PH 4 or PH 10 calibration
1) Rinse the electrode with distilled water.
2) Immerse the electrode in the PH4 buffer solution ( or PH10
buffer solution ).
3) Press the " CAL Button " ( 3-7, Fig. 1 ) then the upper
display shows texts of " CAL " and the lower display
shows the default calibration value.
* The texts " CAL " will flash for around 5 seconds.
After that, the meter calibrate itself automatically.
The upper display will show the calibrated value,
the lower display will show the temperature value.
4) Rinse the electrode with distilled water again and then
5) Repeat above (2) to (4) procedures two times at least.
6) The instrument and electrode are now finished the
" TWO POINTS CALIBRATION " & ready for the
measurement.
14
4-5 Other functions ( Hold, Memory )
Data Hold
Press the " Hold Button " ( 3-4, Fig. 1 ) will hold the
measured value & the LCD will indicate a " HOLD " symbol
on the display during the measuring.
* Press the " Hold Button " again to exit the data hold
function.
Data Record ( Max., Min. reading )
* The data record function records the maximum and
minimum readings. Press the " REC. Button " ( 3-3, Fig. 1 )
to start the Data Record function and there will be a
" REC " symbol on the display.
* With the " REC " symbol on the display :
a) Press the " REC. Button " ( 3-3, Fig. 1 ) once, the
" REC Max " symbol along with the maximum value
will appear on the display.
If intend to delete the maximum value, just press
the " Hold Button " ( 3-4, Fig. 1 ) for a while, and then
the display will show the " REC " symbol only &
execute the memory function continuously.
b) Press the " REC. Button " ( 3-3, Fig. 1 ) again, the
" REC Min " symbol along with the minimum value
will appear on the display.
If intend to delete the minimum value, just press
the " Hold Button " ( 3-4, Fig. 1 ) for a while, and then
the display will show the " REC " symbol only &
execute the memory function continuously.
c) To exit the memory record function, just press the
" REC " button for 2 seconds at least. The display will
revert to the current reading.
15
5. AUTO POWER OFF DISABLE
The instrument has " Auto Power Off " function in
order to prolong battery life. The meter will shut
off automatically if none of the buttons are pressed
in approx. 10 min.
To disable this function, Select the memory record
function during the measurement by pressing the
" REC. Button " ( 3-3, Fig. 1 ).
6. RS232 PC SERIAL INTERFACE
The instrument features RS232 output via
3.5 mm Terminal ( 3-13, Fig. 1 ).
The signal output is a 16 digits data stream which
can be utilized for user's specific application.
A RS232 lead with the following connection will
be required to link the instrument with the PC
serial interface.
MeterPC
(3.5 mm jack plug)(9W 'D" Connector)
Center Pin................................ Pin 4 Pin 2
For example :
If the display reading is 1234, then D8 to D1 is :
0000123
D9Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D10Polarity
0 = Positive1 = Negative
D11 & D12 Annunciator for Display
= 01
mV = 18mS = 14PPM = 19
OF2= 06mg/L = 07
D13When send the upper display data = 1
When send the lower display data = 2
D144
D15Start Word
= 02PH = 05
RS232 FORMAT : 9600, N, 8, 1
17
7. BATTERY REPLACEMENT
e
蚓0 蚓
蚓
蚌32
蚌
蚌
*
e
1) When the left corner of LCD display show " ", it
is necessary to replace the battery. However, in-spec.
measurement may still be made for several hours after
low battery indicator appears.
2) Slide the " Battery Cover " (3-14, Fig. 1) away from the
instrument and remove the battery.
3) Replace with 9V battery ( Alkaline or Heavy duty type )
and reinstate the cover.
4) Make sure the battery cover is secured after changing the
battery.
8. OPTIONAL PROBES & ACCESSORIES
ORP ELECTRODEORP electrode ORP-04 plug into the
meter ( select to the mV
Model :function ) to become a professional
ORP-04ORP (oxidation-reduction potential)
Meter.
ATC
PROBE
Model :
YK-200PATC
MeasurementRang
to 65
to 149
Automatic Temperatur
Compensation Probe for PH
function.
18
PH ELECTRODEGeneral purpose, laboratory
&
&
&
T
Model :field usage. 12 mm dia. x 130 mm.
PE-03Epoxy body, 1 - 13 pH.
PH ELECTRODEGeneral purpose, laboratory
Model :field usage. 9.5 mm dia. x 120 mm.
PE-11Epoxy body, 1 - 13 pH.
(0 - 14 pH typical)
PH ELECTRODEProfessional, laboratory
Model :field usage. 9.5 mm dia. x 120 mm.
PE-01Epoxy body, 0 - 14 pH.
SPEAR TIP
PH ELECTRODEperfect for those pH measurements in
Model : PE-04HDrequired. Meat, sausage and cheese
BUFFER SOLUTIONPH 7.00 standard buffer solution.
PH-07for calibration purpose.
BUFFER SOLUTIONPH 4.00 standard buffer solution.
PH-04for calibration purpose.
RS232 cableRS232 cable for connecting between
UPCB-02the meter & the computer.
SOFTWAREWindows version application software
SW-U801-WINapplies as the performance of data
he " Spear Tip pH electrode " is
applications where sample piercing is
are ideal applications. The electrode
features a very durable glass
measuring spear packaged in a
rugged virtually unbreakable epoxy
body.
Measuring range : 0 to 14 pH .
Size 12 mm dia. x 150 m.
logging system & data recorder...
19
0111-DCT2001
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.