General Technologies CM600 User Manual

600V CATII 300V CATIII
600A
H
HOLD
OFF
A
A
V
V
ZERO
P
H
A
600V CATII 300V CATIII
600V
COM
1
2
3
4
5
8
9
7
6
P
PEAK
Press 2 Sec
General Technologies Corp.
600VCATII 300VCATIII
600A
H
HOLD
OFF
A
A
V
V
ZERO
P
H
A
600VCATII 300VCATIII
600V

P
PEAK
Press 2Sec
600VCATII 300VCATIII
600A
H
HOLD
OFF
A
A
V
V
ZERO
P
H
A
600VCATII 300VCATIII
600V

P
PEAK
Press 2Sec
AUTO
P
H
A
V
Relative Indicator
AC/DC Indicator
Low Battery
Digital Reading
Ohm Indicator
Voltage Indicator
Ampere Indicator
Continuity
Data Hold
Peak Hold
CM600
600 Amp AC/DC Clamp Meter
1. SAFE TY RULES
This meter is designed for indoor use at temperatures between 0°C to 40°C and
altitudes up to 2,000m.
To ensure that the meter is used safely, follow all safety and operating instructions in
this operation manual. If the meter is not used as described in this operation manual, the safety features of this meter might be impaired.
the meter is not operating properly.
When using the instrument, keep your  ngers behind the  nger guards on the plastic
casing and probes.
Disconnect the live test lead before disconnecting the common test lead.
Make sure power is off before cutting, desoldering, or breaking the circuit wires. Small
amounts of current can be dangerous.
Do not apply more than 600 VDC or 600V AC rms between a terminal and ground.
rms. Such voltages pose a shock hazard.
Never make measurements with the battery cover off.
To avoid electrical shock or damage to the meter, do not exceed the input limits.
2. INTERNATIONAL SYMBOLS
Important information
see manual
AC
DC Double Insulation
Dangerous Voltages
Continuity
Ground
3. TECHNICAL SPECIFICATIONS
3.1 General Specifications
Display: 4 digits LCD, 9999 counts and en gi neer ing units annunciator display Polarity: Automatic, (-) negative polarity indication Zero adjustment: One touch button Sample rate: 0.5 Sec. Peak hold sample time: 10 mS for DCV and DCA Over range indication: Only the “OL” is displayed Power: 2x UM-4 or AAA 1.5 Volt battery Auto power off: Automatically powers off after 30 min. of no operation Battery life: Approx. 50 hours. (w/ alkaline batteries) Clamp opening size: 25 mm maximum. Dimension: 189 x 70 x 34 mm. Weight: Approx. 220 g. (Including battery) Accesories: User’s Manual, Soft Carrying Pouch, Test Leads and 2 x AAA Alkaline Batteries
4. OPERATION
4.1 INSTRUMENT DESCRIPTION
1) Current sensing clamp
2) Safety protection ring
3) Clamp opening handle
4) LCD Display
5) COM input terminal
6) Positive input terminal
7) Zero button
8) Function selection knob
9) Peak/Data hold button
4.2 LCD Screen Description
4.3 Current Measurement
CAUTION: Remove all test leads connected to the instru-
ment before proceeding with this test.
4.3.1 DC Current Measurement
Set the function selector knob to (DCA) and press the Zero button, the will appear in the display. Open the clamp by pressing the clamp opening handle, insert the cable or wire inside the current sensing clamp, and then close the clamp and read from the LCD screen.
3.2 Electrical Specifications
The mark next to the probe tip, is a warning that the input voltage
should not exceed the indicated values. This is to prevent damage to the internal circuitry.
The function switches should be set to the function which you want to
test before operation.
Accuracies are ±(% of reading + number of least signi cant digits) at
23°C ±5°C, less than 75% RH.
Function Range Resolution Accuracy Remarks Overload Protection
DC Current 600 A 0.1 A ±(2.5%+10) - 660 Arms
AC Current 600 A 0.1 A ±(2%+10) 50~500Hz 660 Arms
DC Voltage 600 V 0.1 V ±(1%+2) - 660 Vrms
AC Voltage 600 V 0.1 V ±(1.5%+5) 50~500Hz 660 Vrms
Resistance
Continuity
9.999 K 1.0
Beeps when < 100
±(1%+3)
3VDC Max. Test Voltage 660 Vrms
-
660 Vrms
4.3.2 AC Current Measurement
Set the function selector knob to
(ACA). Open the clamp by pressing the clamp opening handle, insert the cable or wire inside the current sensing clamp, and then close the clamp and read from the LCD screen.
Wrong Right
4.4 Voltage Measurement
600VCATII 300VCATIII
600A
H
HOLD
OFF
A
A
V
V
ZERO
P
H
600VCATII 300VCATIII
600V

P
PEAK
Press 2Sec
600VCATII 300VCATIII
600A
H
HOLD
OFF
A
A
V
V
ZERO
P
H
600VCATII 300VCATIII
600V

P
PEAK
Press 2Sec
CAUTION: Maximum Input Voltage is 600Vrms, do not
exceed this rating to avoid personal injuries or damage to the instrument.
Battery
4.4.1 DC Voltage Measurement
Set the function selector knob to
positive input terminal (+), and the black test lead into the COM terminal on the front of the instrument. Proceed to connect the test leads across the source or load under measurement. The polarity will be indicated when the tip probe is connected. Read voltage in the display.
(DCV). Insert the red test lead into the
4.5.2 Continuity Test
Set the function selector knob to
(Resistance). Insert the red test lead into the positive input terminal (+), and the black test lead into the COM terminal on the front of the instrument. Proceed to connect the test leads tips across the component or circuit under test. If the resistance less is than 100Ω,, the buzzer will emit a continuous tone.
4.6 Other Functions
4.6.1 Relative Measurements
By pressing the ZERO button, the instrument is entered into zero mode and the indicator will appear on the LCD screen. The value in the display immediately before pressing the button is stored into memory as a reference value for future tests. To exit the Zero mode press the ZERO button again.
4.6.2 Data Hold and Peak Hold
Pressing the HOLD/PEAK button once, the present reading is held and appears on the display. Press the switch again to return to normal mode (continuous readings) and to release the data hold feature. The PEAK HOLD function is enabled by pressing the HOLD/PEAK button until the
indicator appears on the screen. When in this mode, the instrument will hold and display the highest reading obtained during the measurement. To exit the peak/hold mode press the HOLD button for approx. 2 sec. until the display indicator turns off. This function works with DCV, ACV, DCA and ACA.
4.4.2 AC Voltage Measurement
Set the function selector knob to
(ACV). Insert the red test lead into the positive input terminal (+), and the black test lead into the COM terminal on the front of the instrument. Proceed to connect the test leads across the source or load under measurement. Read voltage in the LCD screen.
4.5 Resistance and Continuity Measurement
CAUTION: Maximum Input Voltage is 600Vrms, do not
exceed this rating to avoid personal injuries or damages to the instrument. Also ensure there is no power applied to the circuit and that all capacitors are discharged.
4.6.3 Auto Power Off
The meter will turn off automatically after approximately 30 minutes of no operation to extend the battery life. Auto power off can be disabled by pressing and holding down the ZERO button while turning on the instru­ment, when auto power is off, the symbol on the screen will not appear. Auto power off will be re-enable after turning off and on the instrument.
5. MAINTENANCE
CAUTION: Before attempting battery removal or replace-
ment, disconnect test leads and remove the instrument from any energized circuit to avoid shock hazard.
5.1 Battery Replacement
When the battery needs replacement, the battery symbol will be ap­pear in the upper left hand side of the LCD display. To replace the battery, remove the screws at the back cover and replace with a 2 AAA 1.5 V alkaline batteries.
5.2 Cleaning
Periodically wipe the case with a soft damp cloth and mild household cleanser. Do not use abrasives or solvents. Ensure that no water gets inside the instrument to prevent possible shorts and damage.
4.5.1 Resistance Measurement
Set the function selector knob to
into the positive input terminal (+), and the black test lead into the COM terminal on the front of the instrument. Proceed to connect the test leads across the component or circuit under test. Read resistance in Ohms on the LCD screen.
(Resistance). Insert the red test lead
6. WARRANTY
One year limited warranty, excluding batteries and fuses. For details see Standard Warranty Information in our webpage or you may request a printed copy.
General Technologies Corp.
#121 - 7350 72nd Street Tel.: (604) 952-6699 Delta, BC Fax: (604) 952-6690 Canada V4G 1H9 www.generaltechnologies.net
© Copyright 2010 General Technologies
Loading...