1110-A Interpolating Frequency Standard is from
to 1010 kc. The output frequencies of the 1110-PI
Multivibrator Unit are 1.0- and 0.1-Mc fundamentals
with harmonics up to 200 or more.
Calibration
1000 divisions corresponding to 0.001 per cent or 10
parts per million per division.
A list of check settings is provided on the panel. This
check can be made at any time by simply plugging a
set of headphones into the jack or binding posts provided
on the panel. A trimmer control on the panel provides
for adjusting the oscillator to agreement with the crystal.
To facilitate conversion of the dial readings from
their basic percentage or parts per million values of
frequency increment to fractions of a megacycle or of
0.1 Mc (100 kc), a table listing the number of dial divisions for frequency increments of 1.0 Mc and 0.1 Mc at
each harmonic from 100 to 220 is given on the panel. A
simple slide-rule ratio then gives the desired frequency
increment.
Crystal Oscillator:
within
1
temperature. It should be reliable to within
per million at ordinary room temperatures. The crystal
frequency can be checked and adjhstedin terms of standard frequency transmissions from
part in a million of correct frequency at room
The output frequency range of the
:
The variable frequency oscillator dial has
The crystal oscillator is adjusted to
f
10 parts
WWV
using an
1000
external receiver, maintaining the variable oscillator at
exactly 50 kc in terms of the crystal.
Accuracy of Measurement:
measurement is
lator dial directly.
in terms of the crystal, the over-all accuracy is limited
principally by the error of the crystal.
Vacuum Tubes:
2
-
6AC7
4 - 6SN7-GT 1 - 5R4GY
1 - 6557
1 - 6SA7 1 - 2LAP-430 (Bridge Circuit Lamp)
Power Supply:
cycles.
Power Input:
Mounting:
(attached to 1110-A hy cable) small metal cabinet.
Accessories Supplied:
1110-PI Multivihrator Unit with connecting cable.
Accessories Required
Dimensions:
behind panel, (length)
inches. 1110-P1 (length)
inches.
Net Weights:
TYPE 1110-P1 Multivihrator Unit,
+Z5 parts per million using the oscil-
If
the oscillator is carefully trimmed
The following tubes are supplied:
3
-
6J5GT /G
1
-
9001
Either 105-125 or 210-250 volts, 50-60
85 watts from 115-volt, 60-cycle line.
TYPE 1110-A Relay Rack; TYPE 1110-P1
:
1110-A Panel (length) 19 x (height) 8%;
Type 1110-A assembly, 40 pounds;
The over-all accuracy of
Line connector cord, and TYPE
Head t,elephones.
17% x (height) 8% x (depth) 14
9%
x (height)
5%
x (depth)
735 pounds.
5%
Manufactured and sold under United States Letters Patent:
United States Letters
Patent owned or controlled by American Telephone and Telegraph
Company, and under all patents and patent applications of Dr.
piezo-electric crystals and their associated circuits.
G.
W.
Pierce pertaining to
Page 4
OPERATING INSTRUCTIONS
for
Type
The Type 1110-A Interpolating Frequency Standard
is
designed as a means to improve greatly the accuracy
of frequency measurements made with the heterodynefrequency meters, such as the Types 720-A and 620-A.
The Type 1110 -A Interpolating Frequency Standard
comprises a standard frequency source having a range of
1
per cent, from 1000 to 1010 kc, and
Type
1110-PI, a harmonic generator of the multivibrator
type which generates multiples of
adjustment of the frequency
coverage
1
Mc harmonics and above lOMc, using 100 kc harmonics.
Standard
1010 kc, and isthe sum of a crystal oscillator frequency
of
950
50 to 60 kc. The oscillator
1000 divisions.
is
obtained at frequencies above 100 Mc, usirg
The output of the Type 1110-A Interpolating
is
a
single frequency, lying between 1000 and
kc and of a bridge-type tuned-circuit oscillator of
1110-A
1Mc and 100kc. By
of the standard, complete
is
controlled by a
Interpolating Frequency Standard
is
used with the
PRINCIPLES OF OPERATION
dial
having
SECTION 1.0
INTRODUCTION
SECTION 2.0
Thus integral standard frequencies can be generated for
calibration
can he generated to match any desired frequency,
the approximate range of measurement
200 Mc. With the Type 720-A Heterodyne Frequency
Meter, the range
setting for zero beat
damental or harmonic,
on the heterodyne frequency meter. The
Interbolating Frequency Standard provides a means of
evaluating more precisely the frequency to which the dial
is
set.
trol the
Unit. The multivibrator fundamental frequencies are 1.0
MC and 100 kc. A large number of harmonics are available but in
utilized
quency Meters,
purpses, or an accurately known frequency
With the Type 620-A Heterodyne Frequency Meter,
is
from 10 Mc to
is
from 100 Mc to 2000 Mc. The dial
with
the unknown signal, either fun-
is
obtained in the usual manner
Typ 1110-A
of
The output frequency
multivibrators of the
particular the range from the 100th to 200th
with
the Typs 720-A and 620-A Heterodyne Fre-
the standard
Type
1110-P1 Multivibrator
is
used to con-
is
CRYSTAL
OSCILLATOR
950
kc
fc
-
-
d
L- C
OSCILLATOR
50-60
8
SUPPLY
kc
-
fo
m
------------------
MODULATOR
MAIN
ASSEMBLY
Figure
*
FILTER
/fc + fo
fc
Jc - fo
1.
Block Diagram
and Type 1110-P1
I
I
I
I
I
fc + fo
1
I
)
!
+
MULTNBRAm
/
1Mc
I
I I
I
I
.*
I
I
I
I
I
I
'
_1
for
Type 1110-A
I
I
MULTNIBRATOR
1
100
I
1110-PI
I
MULTIVIBRATOR
I------J
UNIT
kc
I
I
1
I
I
I
I
<I
-1
I
I
Mc
100
yc
720A
HET
0.1
10
M
c
620A
HET. FREQ. METER
I
HARMONICS
200
IIIII
FREQ.
METER
HARMONICS
IIIIII
lqc
20
Mc
Page 5
GENERAL RADIO COMPANY
Figure
Interpolation
66
sliding harmonics." This process
1
is
a block diagram of the system.
is
accomplished by the method of
is
illustrated in Fi-
gure 2. When the dial of the interpolating frequency
standard
fs,
When n
is
that
the range
dial
difference,
UlAL
isset
at
zero, harmonics d the output frequency,
are spaced at frequency intervals fs in the spectrum.
is
100, the available
1%
range of variation of fs
just sufficient to cover the range from nfs to (n+l)fs,
is,
Af
is
equal to f
is
therefore 2fs. To measure a frequency, the
is
advanced to move nfs to
A,
from which
ADVANCED;
Af
f,
I
=
=
A
IN
DIAL DIVISIONS
DIVISIONS
nf, + Af
s.
At the 200th harmonic (n=230),
riffs,
giving a scalc
A£
can be calculated.
I
PER
MEGACYCLE
I
I
100
<n
I
5
I
<
200
Figure 2.
As the oscillator control
scale, the outplt frequency of the standard
1000 to 1010 kc, a range
harmonics of the
multivibrators controlled by these out-
put frequencies are similarly changed by
100th harmonic, the increase
times the original
is
moved through the range from 100' to
1
Mc multivibrator, or from 10.0 to 10.1 Mc for the 100
fundam
of
1%.
mtal;
is
moved from zero to
is
varied from
The frequencies of the
1%.
of
1%
gives a frequency 101
that
is,
the 100th harmonic
101
Mc for the
full
At the
kc multivibrator. For
variation
is
greater than the fundamental frequency so
that complete coverage
above that of the
While complete coverage
quencies below that
is
coverage
each harmonic
coverage
the coverage
Mc. And
often very useful. The range covered at
is
is
readily estimated. For example, at 95 Mc
is
again,
at 67 Mc the coverage
all
higher harmonics the range of
is
obtained at
all
100th harmonic.
is
not obtained at fre-
d
the 100th harmonic, the substantial
I%, so
0.95 Mc, that
that
the extent of the available
is,
from 95,00 Mc to 95.95
is
0.67 Mc., that is,
frequencies
from 67.00 Mc to 67.67 Mc.
11
When the oscillator control of the Type
Interpolating Frequency Standard
precisely, when the oscillator
is
is
set at
set at zero, or more
50
kc, (which can
10-A
be checked in terms of the internal crystal frequency)
the
output frequency
is
just 1000 kc,
and
all
multivibrator
harmonics are then integral standard frequencies which
can be conveniently used in calibrating, or in
checking
the calibration of, the heterodyne frequency meters.
Such calibrations are needed to identify the multivibrator
harmonic used in a measurement. The accuracy of the
frequency meter calibration does not otherwise enter
into the final result.
The basic method
1.
Set the heterodyne frequency meter to zero beat
with the frequency being measured.
sult
is
then given by the dial reading of the heterodyne
of
measurement
is
then as follows:
An
approximate .re-
frequency meter.
2.
Couple the output of the multivibrator to the frequency meter. Advance the control of the interpolating
standard from zero until the
first
strong beat between
a multivibrator harmonic and the frequency meter is obtained. Set to zero beat.
3.
The frequency being measured
is
then given
by
the
frequency of the multivibrator harmonic being used
(identified automatically by the frequency meter
plus
an increment in frequency (determined from the
readmg),
dial
of the interpolating frequency standard).
Detailed
examples are given under Operation, Section
4,
.
SECTION 3.0
INSTALLATION
1.
Plug the power cord into the instrument (rear) and
connect to power line of voltage and frequency
as
marked
3.
put terminal of the multivibrator unit to the frequency
on nameplate beside the socket. meter.
2. Plug the cable of the Type 1110-P1 Multivibrator
Unit into the multipoint socket on the panel of the
11
10 -A Interpolating Frequency Standard.
Type
4. A pir
into the heterodyne frequency meter, or the jack on the
panel of the interpolating frequency standard, should be
provided.
Provide a short coupling wire from the desired out
of telephone receivers, which can be plugged
-
Page 6
TYPE
1110-A
INTERPOLATING FREQUENCY STANDARD
SECTION
OPERATION
4.1
Use of Instrument
4.11
Throw on power switch
frequency
of the multivibrator unit, according to whether a
0.1
mended that both multivibrators be used continuously.
zero. Adjust compensator control for zero beat with
phones plugged into jack on interpolating frequency
standard.
meter output.
vary its frequency.
at every
0.1
(or
cy meter to zero beat with the frequency to be
Note reading d frequency meter. Couple the output of the
multivibrator unit to the frequency meter and listen in
telephones plugged into the heterodyne frequency meter
output. Advance the dial of the interpolating frequency
standard from zero until the first strong beat
within the dial range corresponding to the harmonic used
(see
interval from zero to
standard dial for zero beat. Note the final dial reading.
(For example,
sults,
the final dial reading
meter determine which multivibrator harmonic was used.
(Example: Reading
162.
the number
162
frequency standard divided by the number of divisions
from the table gives the fraction
which must
to obtain the
163
divisions. The final result
standard, Throw on one
Mc harmonics output
4.12
Set
dial
of interpolating frequency standard to
4.13
Plug telephones into heterodyne frequency
Turn on heterodyne frequency meter and
1
Mc in the range
Mc in the range
4.14
In
making
known harmonic) frequency of the heterodyne frequen-
4.16).
In this example, the first strong beat in the
4.15
Carefully set the interpolating frequency'
199.8
the oscillator calibration should be checked before
4.16
From the reading of the heterodyne frequency
Enter the table on the panel at
of
divisions required to cover 1 Mc between
and
163
MC.)
The final
be
added
final
result. Example: Table value at
Mc
is
617.3
divisions. Final
is
desired. It
Zero beat points should be obtained
100
10
to
20
measurements,
617.3
divisions
divisions). For the most precise re-
is
taken. See Section
is
162.3
Mc. The harmonic used
dial
tothe frequency of the used harmonic
is
of
the interpolating
of
the power switches
1
Mc or
is
not recom-
to
200
Mc or at every
Mc.
set
the fundamental
measured.
is
obtained
is
used.
4.2
following.
is
162-163
reading ofthe interpolating
of
a megacycle
Mc and find
br
O.lMd
162-
dial
reading was
1988
4.0
initial frequency
The changes in frequency can then
in dial reading from the reading for the initial condition.
Since each division very closely represents a change of
10
parts per million (see Section
dial reading are easily converted to frequency changes in
parts per million. Also, since each dial division represents a change in the control frequency
changes in dial
at the test oscillator frequency by multiplying by
by the number
of the initial frequency.
4.18
oscillator whose frequency
multiple of
readings corresponds almost exactly with the parts per
million change in the oscillator frequency. However,
the oscillator frequency lies just below the next higher
multiple of
1
part per million
that
would be measured by raising the
162
Mc to
corresponds to
cycles. One prt per million would be
stead of
purposes.
162.98
4.2
Checking the Interpolation Oscillator
4.21
pnel of the Type
ard, the calibration of the
checked in terms
check points, over the entire range of the oscillator dial,
are listed in the table on the
4.22
any small errors in calibration can be made zero at any
of the check settings by use
this,
To do
the tableand adjust the compensator knobto get zero beat.
4.23
ment
is
desired, the oscillator should be checked, and
corrected,
obtained between the standard and the frequency meter,
Paragraph
is
measured as previously described.
be
read as the changes
4.18),
the changes in
of
10
cycles, the
readirg are convertible into cycles change
10
and
of
the harmonic used in the measurement
When observing changes in frequency of an
is
only slightly higher
1
Mc, the parts per million change by dial
1
Mc for example
is
162.98
Mc. On the dial, each division then
10
parts per million
162.98
162.98
Mc, it
cycles. This frequency
162
harmonic from
of
162
Mc or
162.00
cycles, which error is negligible for many
By
plugging the telephones into the jack on the
1110-A
of
Assuming the oscillator alignment
set the
When the best possible accuracy of measure-
4.14
dial
if
necessary, immediately after the beat
above. The final reading
Interpolating Frequency Stand-
50-60
kc oscillator can be
the crystal oscillator. A number of
of
the compensator control.
to the number
of
divisions givenin
is
than
is
evident
1620.0
cycles, in-
is
normal,
is
then taken.
if
a
fx
=
162,324
In other words, the frequency of the
increased
in frequency of an oscillator resulting from arbitrary
changes in the conditions of operation. In this case, the
0.324
4.17
Many times it
Mc in going from
is
Mc
16hd
harmonic was
162
Mc to fx.
desired to measure changes
4,3
Checkiw the Crvstal Oscillator
4.31
Since the multivibrator unit provides output
frequencies at all of the frequencies used for the Bureau
of Sandards' Sandard Frequency Transmissions, it
possible to check the crystal oscillator by obtaining abeat
between the standard frequency and multivibrator har-
If
monic in a short-wave receiver.
made to oscillate or
tor, the check can
if
it includes a heterodyning oscilla-
be
made very precisely.
the receiver can be
is
Page 7
GENERAL RADIO COMPANY
4.32 After picking up the standard frequency transmission, couple the appropriate multivibrator output into
the receiver. The dial of the interpolating frequency
standard may be advanced slightly from zero so as to
produce an audible beat in the receiver. This will facilitate
making
the coupling adjustments.
is
curve, the change in shape
be
can thus
made to fit the ideal curve at any mint where
a means for comparison
negligible. The dial reading
is
available.
In
this instrument
the crystal oscillator furnishes this means at a suffi-
of
ciently large number
points covering the entire range
of the oscillator.
4.33 Plug the telephones into the panel jack on the
interpolating frequency standard and carefully set the
"0"
oscillator to obtain zero beat at the check pint
dial.
It
the
does not matter
actly zero as long as the zero beat
if
the
dial
reading
is
maintained. Check
is
on
not ex-
in the short wave receiver to determine the difference in
frequency between the standard frequency and the output
of the interpolating frequency standard. This difference
should be only a few parts in a million.
4.34 To correct the crystal frequency, remove the
carer in the lower left
snap
of
the interpolating frequency
standard panel and use a screw driver to adjust C-17.
NOTE: At each adjustment of C-17, carefully reset the
oscillator dial to zero beat
against the crystal at the
"0"
reference point. Then check the output frequency of the
standard against the standard frequency transmission.
is
The normal crystal frequency
is
lator
set to zero beat against the crystal at the
reference point, its frequency
950 kc. When the oscil-
"0"
is
close to 50 kc and
is
actually 1/19th of the crystal frequency. The control
is
frequency for the multivibrator unit
and
is
close to 1000 kc.
The output frequency matched
against the standard frequency transmission
quency multiplied by the number
if
(For example,
is
sion
used, the outputafrequencp must be five times the
the 5 Mc standard frequency transmis-
the sum of these,
is
this fre-
of
the harmonic used.
control frequency.)
4.4 Checking the Oscillator Alignment
4.41 To obtain a direct-reading linear calibration
of the oscillator several conditions must be met simultaneously.
be correct.
First, the plate shape of the condensermust
Second, the initial circuit capacitance must
have the correct value, appropriate to the change in ca-
pacitance produced by the shaped rotor plates. When
these conditions are met, the calibration will be linear
and the correct frequency ratio (maximum frequency
ded by minimum frequency)
the correct ratio
is
have the oscillator cover the range from
will
be obtained.
60/50 = 1.20, since it
50
In
is
to 60 kc,
tween the zero and 1000 division points on the dial.
divi-
this case,
desired to
be-
In
order to have the oscillator cover exactly the specified
is
range it
so that the initial frequency
then necessary to adjust the circuit inductance
is
exactly 50 kc. Manufactur-
ing tolerances impose a limit on just how closely this
In
ideal can be approached.
to within 5 two divisions
4.42
Small
corrections can be made by means of the
this instrument, alignment
is
considered satisfactory.
compensator condenser (part of C-7) panel control. For
corrections, the primary effect
tial capcitance
is
to move the calibration curve bodily up
or down, and the secondary effect
changing the
is
to changethe shape
ini-
of
of the calibration curve. Over small portions of the
4.43 The list
of
oscillator check points on the panel
plate includes the principal points. Many other points can
be found
these can be identified
and
used to advantage
and
in many instances. The oscillator frequency at any zero
is
given
by
beat point
integers. The principal pints are given when
has
such value
n
60 kc.
cillator beats
tal. The
subtracting
In
this
case, the 16th to 19th harmonic of the os-
with
dial
reading of the ideal oscillator
50 kc from
(in kc) by 100. For example, a check point
fo = T950 kc, where m and n are
as
to give
fo
values lying between 50 atd
the fundamental frequency
fo
and multiplying the remainder
m
=
of
the crys-
is
given by
is
foundat
1
and
52.055 kc, (where n = 73, m = 4). Subtracting 50 kc,
2.055 kc gives a
reading
of
205.5 divisions (as listed
dial
in the table).
If
4.44
the oscillator requires realignment, it
is
evident from the foregoing that the misalignment might
of
be due to a change
capacitance, or of inductance or a
combination of both. The first test, then, should be to
check the
cillator frequency at equal steps in
serving
latter
standard
1
kc (or 500 cycles) from 50 to 60 kc and
noted.
constant, the capacitance adjustment
differences are not constant readjustment
capacitance must be made.
toward the
"uphill"
to correct it.
ings
capacitance.
dial
readings for equal steps in frequency. The
is
generally more convenient where a frequency
is
available. The oscillator
If
the differences
high
frequency end
and
an increase
(Part
d
the top
dial.)
This
is
done
by
measuring the
dial
reading, or ob-
is
checked at every
dial
of
successive
If
the differences increase
of
the
of
initial capacitance
of
C-7, accessible through the open-
dial
readings are
is
correct.
of
dial,
the curve slopes
is
the initial
os-
readings
If
these
required
4.45 When the capacitance has been checked, the
next step
removing the shield can from
is
to adjust the inductance. This
L-2
(to right
is
done by
of
C-7 con-
denser assembly) and loosening the set screw holding
the upper coil just sufficiently to permit the
coil to be
moved. Replace the shield can. Set the condenser dial
at zero and check the oscillator frequency against a
by
frequency standard. Adjust L-2
carefully moving the
the upper coil (upward to raise the frequency). When
is
the frequency
exactly 50 kc, lock the upper coil in
position.
4.46
If
the calibration of the oscillator
is
not badly
in error, the frequency can be checked against the inter-
is
nal crystal, but there
frequency if the error
setting to the
50 kc point, zero on the dial, a quick check
danger of setting to the wrong
is
appreciable. However, if, after
of the principal check points at 277.8, 588.3 and 937.5
is
divisions
made, any error will immediately be dis-
closed because these additional points will not then agree
all
settings listed in the table.
at
Page 8
TYPE 1110 -A INTERPOLATING FREQUENCY
STANDARD
4.5 Checking Control of Multivibrators
4.51 For
oscillating receiver set to obtain a beat against a low
harmonic of the multivibrator, at 2,
causse of the direct signal from the control oscillator,
is
not recommended that a setting of 1 Mc be used. No
attemI;t to adjust multivibrators should be made
have been in operation at least 15 minutes.
4.52
quency of the harmonic will vary erratically and the beat
tone heard in the receiver will be very unsteady. The
multivibrator adjustments are accessible from the top of
the Type 1110-P1 Unit, on removing the dust cover.
Check that R-108, in the Type 1110-P1 Multivibrator,
is in the
an insulated screw driver or alignment tool) to either
side of its previous setting, checking with the receiver
to obtain a clear and steady beat tone. When such a point
has been obtained, leaving the receiver adjusted, note the
range over which C-108 can be turned while obtaining a
steady beat tone in the receiver. Make the final setting
of C-108 at the center of this range.
an oscillating receiver set to obtain a beat against a low
harmonic
of the direct signal from the control oscillator,
recommended that a setting of
to adjust multivibrators should be made until they have
been in operation at least 15 minutes.
quency of the harmonic will vary erratically and the beat
tone heard in the receiver will be very unsteady. The
multivibrator adjustments are accessible from the top
of the Type 1110-P1 Unit, on removing
Check that R-101, in the Type
in full clockwise position. Then adjust C-101 (using an
insulated screwdriver or alignment tool) to either side
of
obtain a clear and steady beat tone. When such a point
has been obtained, leaving the receiver adjusted, note the
range over which C-101 can be turned while obtaining a
steady beat tone in the receiver. Make the final setting
of C-101 at the center of
metal tip, the
full
4.53 For checking the 100 kc Multivibrator, use
4.54
its
previous setting, chetking with the receiver to
4.55 Unless the screwdriver has only a very small
checking the 1 Mc Multivibrator, use an
3,
4, or 5 Mc.
until
*If
the multivibrator
cloclwise position. Then adjust C-108 (wing
of
the multivibrator, at 200 to 900 kc. Because
is
not in control, the fre-
it
1
Mc be used.
Tf
the multivibrator
capcity added when the screwdriver
is
not in control, the fre-
1110-P1 Multivibrator
this
range.
No attempt
'che dust cover,
is
Be-
they
not
is
it
is
in
the slot of the condenser will shift the
frequency slightly. The range of adjustment of the condenser must then be determined by lifting the screwdriver out of the slot after each adjustment.
4.6
Chkcking Calibration of Frequency Meter
4.61 As'described, the use of the
quency standard depends on the calibration of the frequency meter for the identification of the harmonic used.
If
the calibration were in error, the final result would be
in error in that an incorrect harmonic number would be
used,
4.62 The calibration
quency meter may be calibrated, as follows: Set the dial
of the interpolating frequency standard at zero. Set the
frequency meter at the reading believed to be 101 Mc,
and carefully adjust to zero beat. Leaving the frequency
meter alone, move the dial of the interpolating frequency
standard to
tained. If it
101 Mc, and from this key the other calibration points
are identified.
divisions, then the frequency was set at 102 or 100 Mc
respectively.
process is the same as before, using the 0.1 Mc output of
the interpolating frequency standard. Set the dial of the
interpolating frequency standard to zero. Set the frequency meter to the reading believed to
carefully adjust to zero beat. Leaving the frequency
meter alone, move the dial
standard to
obtained.
10.1 Mc, and from this key the other calibration points
are identified.
divisions, then the frequency meter was set at 10.2 or
10.0 Mc respectively.
identifying the key pint
lating frequency standard for zero beat at the
tal check
ly 1.0 or 0.1 Mc.
range and
be made.
1000 divisions. Zero beat should again be ob-
is,
the frequency meter has been set to
If
zero beat
If
4.63
4,64
the frequency meter covers 10 to 20 Mc, the
1000 divisions. Zero beat should again be
If
it is, the frequency meter has been set to
If
zero beat
In
either case, a calibration can be made after
wint. The outpt frequencies are then accurate-
l3y
marking each point, a complete calibration can
is
readily checked, or a fre-
is
reached at 990 or 1010
of
the interpolating frequency
is
reached at 990 or 1010
by
setting the dial of the interp-
tuning the frequency meter over its
mnltivibrator
interwlating fre-
be
10.1 Mc and
"0"
crys-
Page 9
GENERAL RADIO COMPANY
Type
Service and Maintenance Instructions
1.0 FOREWORD
1.1
This Service Information together with the information given in the Operating Instructions should
enable the user to locate and correct ordinary difficulties resulting from normal usage.
1.2 Most of the components mentioned in these in-
be
structions can
1.3 Major service problems should be referred to
the Service Department which will cooperate as far as
possible by furnishing information and instructions, as
well as by shipping any replacements
be required.
old, a reasonable charge may be expected for replacement
parts
is
returned.
located by referring to the photographs.
parts
If
the instrument
or for complete reconditioning
is
more than bne year
if
which may
the standard
1110-A
3.3 NO OUTPUT SIGNAL AT SOCKET SO-1; refer
to Section 6.0.
J-1;
3.4 NO OUTPUT SIGNAL AT TEL JACK,
to Section 7.0.
3.5 NO OUTPUT
refer to Section 8.0.
3.6 NO OUTPUT FROM THE VARIABLE OSCILLA-
TOR; refer to Section 9.0.
3.7 OUTPUT SIGNAL IS ROUGH OR UNSTEADY;
refer to Section 10.0.
3.8
OUTPUT BEATS INDISTINGUISHABLE FROM
THOSE PLUS OR MINUS 50 KC AWAY; refer to Sec-
tion 11.0.
FROM THE
CRYWAL
OSCILLATOR;
refer
1.4 Detailed facts giving type and serial numbers of
parts,
as
the instrument and
should always be included in your report to the Service
Department.
2.0 GENERAL
If
the standard becomes inoperative, a few simple
checks should be made before any circuit checks are
made.
2.1 Make certain the voltage and frequency of the
power line source are correct.
2.2 Test the power supply cord for open circuits or
for poor contacts in the power outlets.
2.3 Check fuses
of the instrument, for open circuits and be sure they are
tight in their holders.
2.4 Be sure that all tubes are completely seated in
their sockets.
3.0 INSTRUMENT INOPERATIVE
3.1 NO OUTPUT SIGNAL AT THE
JACK; refer to Section 4.0.
F-1 and F-2, mounted at the rear
well as operating conditions,
1
MC OUTPUT
3.9 MAIN DIAL IS OFF CALIBRATION; refer to
Section 12.0.
3.10 POWER SUPPLY INOPERATNE; refer to Sec-
tion 13.0.
3.11 VACUUM-TUBE DATA; refer to Section 14.0.
4.0 NO OUTPUT SIGNAL AT THE
This condition can be readily detected by a lack
beat notes in the heterodyne-frequency meter when attempting to identify
4.1 Check switch S-102 for proper operation.
4.2 Make sure that rheostat R-108
clockwise position.
4.3 Check tube V-103 and operating voltages; refer
to Section 14.0.
4.4 Check resistors R-108 through R-112 for open
and short circuits and proper values.
4.5 Check capacitors C-108 through C-112 for open
and short circuits. Capacitors C-108 and C-110 should
have any accumulated dust blown out of their plates.
an
unknown frequency.
1
MC OUTPUT JACK
is
at
its
maximum
of
3.2 NO OUTPUT SIGNAL AT THE 100 KC OUTPUT
JACK; refer to Section 5.0. circuit.
4.6 Check inductor
L-101 for an open or a short
Page 10
TYPE 1110-A INTERPOLATING FREQUENCY STANDARD
4.7 Check tube V-104 and operating voltages; refer
to Section 14.0.
4,8 Check resistors R-113, R-114 and R-115 for
6.6 Check capacitors C-25 through C-29 for open
and short circuits.
6.7 Check inductor L-6 for an open or a short circuit.
open and short circuits and proper values.
6.8 Check the elements of the filter sections LC-1,
4.9 Check capacitors C-113, C-114 and C-115 for
LC-2, LC-3 and LC-4 for open and short circuits,
open and short circuits,
4.10 Check L-102 for an open or
a
short circuit.
7.0 NO OUTPUT SIGNAL AT TEL JACK,
3-1
This condition can be detected by plugging a pair
5.0 NO
OUI'PVT SIGNAL AT TEE 100 KC OUTPUT JACK
of headphones into the TEL jack and listening for the
beats listed on the left hand plate on the panel.
5.1 Check switch S-101 for proper operation.
The voltage from the high potential terminal of
5.2 Make sure that rheostat
mum clockwise position.
R-101
is
set to its maxi-
TEL jack,
with the main dial set to 0. (Tone will be heard for all
J-1, to ground should be about 5.0 volts +20%
positions of the COMPENSATOR control.)
5.3 Check tube V-101 and operating voltages; refer
to Section 14.0.
7.1 Check that jack
3-1
is
making proper contacts.
5.4 Check resistors R-101 through R-105 for open 7.2 Check tube V-3 and operating voltages; refer to
and short circuits and proper values.
5.5 Check
and short circuits,
capcitors C-101 through C-105 for open
Cawcitors C-101 and C-103 should
have any accumulated dust blown out of their plates.
Section 14.0.
7.3 Check resistors R-15, R-16 and R-17 for open
and short circuits and proper values.
5.6 Check tube V-102 and operating voltages; refer 7.4 Check capacitors C-12 through C-16 for open
to Section
14,0,
and short circuits. Turn the main
dial
to settings both
above and below 10 divisions which is necessary in the
has
5.7 Check resistors R-106 and R-107 for open and event that the variable oscillator
gotten out of align-
short circuits. ment.
5,8 Check capacitor C-106 for an open circuit.
6.0 NO OUTPUT SIGNAL AT SOCKET
SO-1
This condition can be detected by the following
procedure. Disconnect the
Type
1110-P1 Multivibrator
from the Type 1110-A Standard by pulling out plug
PL-101. Connect a 1000-ohm resistor between ter-
minals
#7 or #8 and #12 of socket SO-1. Measure the
voltage across this resistor using a vacuum-tube volt-
meter. This should be 6
-
11
volts ax.
A low voltage
or no voltage will cause the multivibrators to fall out
of control.
6.1 Check tubes V-6, V-7 and V-8 and operating
voltages; refer to Section 14.0.
6.2 Check resistors R-24 and R-30 for open and
short circuits and proper values.
6.3 Check capacitors C-37 through C-44 for open
and short circuits and leakage. Capacitors C -40 and
C-42 should have any
accbmulated dust blown out of
their plates.
6.4 Check tube V-5 and operating voltages; refer
to Section 14.0.
6.5 Check resistors
open and short
circuits and proper values.
R-14A, R-20 through R-23 for
7.5 Check inductor L-4 for
an
open or a short circuit.
7.6 'Refer to Sections 8.0 and 9.0.
8.0 NO OUTPUT FROM THE CRYSTAL OSCILLATOR
The output measured from the arm of rheostat
R-14A to ground should be 0.9 - 1.5 volts ax. as measured with a vacuum-tube voltmeter.
8.1 Check tube V-4 and operating voltages; refer to
Section 14.0.
8.2 Check resistors R-13, R-18 and R-19
for open
and short circuits and proper values.
'
Check capacitors C-17 through C-24 for open
8.3
and short circuits.
8,4 Check inductor L-5 for an open or a short circuit.
8,5 The
oscillating.
Quartz
Plate, Type 376-L, may have stopped
If
so a very light tap on its case should
start it again,
In
8.6
case this crystal
is
defective it should be reported to the Service Department immediately. Under
no circumstances should the holder be opened or the
seals broken
the user unless authorized
by
the Service
by
Department as this procedure will void the guarantee,
Page 11
GENERAL RADIO COMPANY
The General Radio Company will not be responsible for
the proper operation of the standard
if
the crystal seals
have been broken.
9.0 NO OUTPUT FROM THE VARIABLE OSCILLATOR
The output of this oscillator, measured from pin
#8
of tube V-5 to ground, should be approximately 1.4
volts
a.c. measured with a vacuum-tube voltmeter, (The
main dial should be set at 500
divisions,)
9.1 Check tubes V-1, V-2 and V-10 and operating
voltages; refer to Section 14.0.
9,2 Check resistors R-1 through R-12 and R-14 for
open and short circuits and proper values.
9.3 Check
capcitors C-lthrough C-5 and C-7 through
C12 for open and short circuits.
9.4 Check inductors
L-1, L-2 and L-3 for open and
short circuits.
9.5 Set dial at 500, adjust rheostat
R-7A so that the
voltage measured across R-10 (or pin #4, V-2 to ground)
is
3.0 volts.
10.0 OUTPUT SIGNAL IS ROUGH OR UNSTEADY
12.0 MAIN DIAL IS OFF CALIBRATION
The dial readings should be within
the settings shown on the
is
checked in terms of the crystal oscillator fre-
tor
pel when
52 divisions of
the
variable oscilla-
quency. The COMPENSATOR control should be set for
zero beat with the main dial set to
0.
12.1 The frequency of the crystal oscillator should
be checked against a frequency standard or
by
the method
outlined in Paragraph 4.3 of the Operating Instructions
by comparing it with the standard frequency transmissions of the National Bureau of Standards, radio station
WWV.
The crystal frequency can be adjusted
The crystal oscillator frequency
is
950 kilocycles.
by
resetting capa-
citor C-17, available from the front panel.
12.2 Check the frequency alignment of the variable
oscillator as outlined in Paragraphs
Operating Instructions.
If
realignment
4,42 and 4.43
is
of
the
necessary,
follow the procedure of Paragraphs 4.44, 4.45, 4.46 of
the Operating Instructions.
12.3
If
the above procedure does not serve to adjust
the frequency and span of the variable oscillator correctly, then the main capacitor
this
is
the case, it will be necessary to return the com-
is
probably at fault.
If
plete instrument to the factory for repair.
Type
Themultivibratorsin the
1110-P1 are probably out of control. Check that a control voltage for the
multivibrators
is
available by following the procedure
of Section 6.0. The multivibrators can be checked and
set by following the procedure of Paragraph 4.5 of the
Operating Instructions.
10.1 Refer to Sections 4.0 and 5.0.
11.0 OUTPUT BEATS INDISTINGUISHABLE FROM
THOSE PLUS OR MINUS 50 KC AWAY
This
is
an abnormal situation but little can be
done to correct it. Normally if such spurious beat
notes are present, they are the result of a modulation
of the multivibrator control voltage. The multivibrator
harmonics are modulated and each
appars as a
carrier
frequency with upper and lower side bands, where the
deslred harmonic frequency
is
the carrier.
If
each beat
has the same amplitude or intensity the correct one can
always
instrument the middle beat