
INSTRUCTION
MANUAL
TYPE
1608-A
IMPEDANCE
BRIDGE
GENERAL
c
RADIO
COMPANY
-
o-
0
(X)
I
>

OPERATING INSTRUCTIONS
1608-A
IMPEDANCE
Form 1608-0100-C
Jan
uar , 1967
BRIDGE
GENERAL
WEST
Copyright
West
CONCORD,
1962
by
Concord,
General
Massachusetts,
Radio
R A D I 0
MASSACHUSETTS,
Company
USA
COMPANY
USA

TABLE
OF
CONTENTS
Section
Section
Section
1.
1.1
Purpose.
1.2
Description
1.3
Symbols,
1.4
Series
1.5
Accuracy
2. OPERATING
2.1
Installation
2.2
Interpretation
2.3
DC
2.4
AC
2. 5 AC
3.
3.1
Application
3.2
Measurements
3.3
Remote
3.4
Use
3.5
Measurements
3. 6
Limit
3. 7 Measuring
Tuned
3.8
Measurement
INTRODUCTION
. . . . .
• . . .
Abbreviations,
and
Parallel
of
Measurements
PROCEDURE
• . . . . . . . .
of
"X"
Resistance
Measurements
Measurements
SPECIAL
Measurements
of
Type
Testing
Circuits
Measurements
MEASUREMENTS
of
DC
Bias
on
Shielded
1650-Pl
on
Grounded
. . . . . • . . . . . • . . . . . . . . 27
Resonant
of
Rp.
and
Definitions
Parameters
in
Read-out
using
with
Internal
External
to
Unknown
Three-'!
. . . . . . . . . . 26
Test
Jig
Frequency
. . . . . . • . . • . . . • . . . .
•
. . .
.
..
Generator.
Generator
. . . . . .
. . . . . . . . 22
ermmal
• . . . . . . . . . • . .
Components
and
Resonant
•
Components
• • . • . . . . • .
Impedance
. . 19
• . .
of
. . .
1
1
1
1
. 4
. 4
9
9
9
. 9
. 12
• 22
26
27
27
27
.28
Section
4.1
4.2
4.3
4.4
4.5
Section
5.1
5.2
5.3
5.4
5.5
5.6
4.
PRINCIPLES
Bridge
Bridge
Bridge
Centade
Phase-Compensation
5.
General.
Calibration
Adjustments.
Replacing
Trouble-Shooting
Tables
Circuits
Sources
Switching
Operation
SERVICE
. . . • .
Indicator
of
Test
OF
•.....
and
. . . . • . . .
AND
Checks.
. . . •
Voltages
OPERATION
Detectors
. . • . . . .
Techniques.
MAINTENANCE
. . .
.'
Lamps
Suggestions
•.
. •
•
•.
. . • 28
. 28
. 28
. 29
. 29
. 30
. .
31
.
31
.
31
• 32
. . . 33
. 33
• . 34

SPECIFICATIONSI
RANGES
Capacitance:
Inductance: 0.05 l'h
Resistance: (series) 0.05 milliohm
Conductance: (parallel) 0.05
megohms
0: (of series
(of parallel
Q:
(of series
(of parallel
(of
(of parallel
Frequency:
20 kc
ACCURACY
C,
G,
At 1
ranges
full scale.
Additional
C
Rand
Residual
pf.
DC
Resistance
that
to
0.1%
supply
D
(or~)
Q
of
0.05 pf
to
to
to
0.9
ohm).
capacitance)-
capacitance)
inductance)-·
inductance)-1 to
series
resistance)-
conductance)-
1 kc with
with
external
R,
L
kc:
±
0.1% ± 0.005%
and
highest C and
% error terms for high
and
L:
(
±0.001fi,
G:
(±0.002Ji,
Terminal Impedance: H
and
accuracy b limited
and
R or
Conductance:
are
possible from 1
detector.
of
Cor
L:
G:
±0.0005fk, ±
1100
,..r
in seven ranges, series
1100 h in seven ranges, series
nanomhos
0.0005
- 0.02
0.5
0.0005
internal
oscillator.
of full scale
Granges,
± 0.11Jfkc
±O.OOOOOIJZ,
Same
by
sensitivity
ohm
±0.0005 ± 5%
±0.0005fkc
2%.
or
to
1.1 megohms,
to 1 at
to 2 at
to
50
at
2000
to
1.2
0.0005
to
oscillator
where
frequency
::'::
0.5/)2)'/n of
to
1.1 mhos,
1 kc.
at
inductive
1.2
module
ac
1 kc.
1 kc.
1 kc.
at
capacitive
supplied;
except
on lowest R
it
is
±0.2% ± 0.005%
and
large
measured
±0.1Q)%ofmeasuredquantity.
~
0.001
ohm,
L
~
0.15 l'h, C ~ 0.25
m; for I-ke
at
to 1 megohm
at
±
5%
above
the
ran~~:e
I kc
or
1 kc.
measurement,
extremes. Balances
with
lower.
or
parallel.
parallel.
or
ac
or
1 kc.
at
phase
de.
de (20,000
1 kc.
20 cps
and
angle:
quantity.
except
the
internal
GENERATOR AND
Internal
other
for
Internal AC Detector:
of plug-in
imately
Internal
power
External
from
1311-A Audio
~ull
to
DC
Bias: Provision
external
GENERAL
Accessories
L
indicator
of
Accessories
generator
Power
10
watts.
Cabinet:
Dimensions:
(485
inches (485 by 315
Net Weight:
pounds
Shipping
pounds
48
For a complete
March
Oscillator:
frequencies
25
DC
limited
Oscillator
20 cps
Detector
supplies,
lamps.
Requirements: 105
DETECTOR
module
(normally
db;
sensitivity
Supplies:
to
less
and
to
20 kc, TYPE 1210-C
Oscillator,
are
recommended.
and
Supplied:
Available:
and
detector,
Hack-bench
Bench
by
295
Bench
(22 kg).
model--
Bench
description,
by
320
(15.8 kg).
Weight:
1962.
1 kc ±
1%
available
Can
3.5
than
Detector:
is
made
for biasing
TYPE
(see
mm),
mm),
model,
normally
on
request.
be
used
either
1 kc). Second-harmonic rejection
control
provided.
volts, 35 volts, 350 volts,
Ya
watt.
For
measurement
and
CAP-22
TYPE
if used, as listed above.
to
page
over-all; rack
depth
36%
model, 50
Unit
TYPE
for biasing
current
Power
1650-Pl
125 (or 210
210).
width
19,
behind
pounds
pounds
see General Radio Experimenter, 36, 3,
supplied. Plug-in modules
Level control provided.
flat
or
selective
at
frequency
adjustable,
at
othE>r
to
spare
(page
depth
panel,
rack
kg);
frequencies
Amplifier
500 volts
fuses,
35);
to
11!/:1
19
model,
rack
l{-C Oscillator, che
1232-A
Tuned
capacitors
in
inductors.
Cord,
Test
Jig
to
250) volts, 50
height
12~~.
model-
panel
10 inches (255 rnm).
(17
kg);
(22. 7
approx-
and
TYPJo;
and
with
spare
external
60 cps,
inches
by
12~
34%
model,

3
15
4
Figll1'«'
J.J.
The
({or
T)pe
legend
1608-A
Sl'l'
pag"
Impedance
2).
6
llridge
7
t6
8
17
18

'IN
T R 0
DUCT
I 0 N
SECTION
1.1
PURPOSE.
The
1
Type
is a self-contained
includes
tance,
as
1-kc
six
conductance,
the
generators
ac
measurements.
bridges
cs
LOW
O\
(
0
SERIES
TO!)
CAPACITANCE
1608-A Impedance
impedance-measuring
for
the
measure'llent
resistance,
and
detectors
and
PARALLEL
Bridge
system,
inductance,
necessary
Cp
HIGH
(
.02-2
CAPACITANCE
INTRODUCTION
1 2 DESCRIPTION
(Figure
of
for
0\
J
1-1)
which
capacias
well
de
and
1.2.1
Type
Provision
with
generator
positions
AC,
1.2.2
the
of
1.3
this
GENERAL.
1608-A
internal
EXT
CONTROLS
contro
the
Type
SYMBOLS,
Table
manual,
are
shown
is
made
and
external
and
detector
of
the
function
DC)
are
shown
AND
Is
and
connectors
1608-A
Impedance
ABBREVIATIONS,
1-2
lists s mbols
together
The
six
bridges
schematically
for
ac
and
generator
connections
switch
schematically
CONNECTORS.
on
Bridge.
and
with
their
contained
de
measurements,
and
for
(INT
AC,
the
front
AND
abbreviations
definitions.
in
the
in
Figure
1-2.
both
detector.
the
INT
in
Figure
Table
and
four
DC,
rear
1-1
panels
The
"on"
EXT
1-3.
lists
DEFINITIONS.
used
in
ex
·~Cr
A
IIICXRX • IIICTRT
OX •
Gp
CAP())
(
0-1.2
CONOIICTANCE
SERIES
Rs
/NO
(
0-1.2)
RESISTANCE
C\
Figure
G
SERIES
Lx•
RN~CT
Ox"
IIILx •
Rx
1-2.
of
the
Ls
LOW
C\
0.$-
$g/
INOIICTANCE
t~~CrRr
The
Type
six
bridges
1608-A.
Lp
HIGH
(
1-
oo..J
PARALLEL
C\
INOIICTANCE
Lx •
RN~CT
ox.'!.!..:.:_:_
OJLx
.,crRr

~---T_Y_PE
__
16_08_-A
Fig.
1· 1 Ref
1
2 BRIDGE
::>
4
5 DQ
__
Function
t'ULL
CGRL
6
7
8
9
10
Q
GEN
DET
I_M_P_E_D_A_N_C_E
__
B_R_ID_G_E_
TABLEl-1
TABLE
OF
CONTROLS
CONTROLS
Name
5-pos
rotary
SELECTOR
:::>1...1\Lt
LEV
SENS
Kl\l'H:rt
6-pos
rotary
i·pos
rotary
coaxial
Continuous
12-pos
rotary
ll·pos
rotary
Continuous
Continuous
Continuous
1
__________________________
AND
CONNECTORS
Type
switch
switch
switch
rotary
controls
rotary
switch
switch
rotary
rotary
rotary
control
control
control
control
Function
Tum
instrument
selects
ac
Selects
circuit
C
::Selects
indicates
of
Main
knob
digits
controls
DQ
C
Q
and R measurements.
Controls
generator,ac
Controls
detector,
internal
or
de
appropriate
for
,
CP, GP, R
8
measurement
CGRL
balance
controls
of
balance
and L measurements
balance
on or off,
or
operation.
measurement
,
L
8
full-scale
indicator.
control.
two
indicator,
three
left-hand
control,
controls,
output
level
and
de.
sensitivity
ac
and
de.
5
right-hand
external
bridge
of
,
and
range,
value
Small
large
digits.
used
used
of
internal
of
internal
___
LP.
knob
for
in G
2
Fig
1-1
11
12
13
14
15
16
17
18
Ref
Name
EXT
GEN J
DET
OUT
UNKNOWN
EXT
DQ J
BIAS
Power
3RD WIRE
EXT
NECTIONS
GROUND
METER
CON-
CONNECTORS
Type
ack·top
J
ack·top
Jack-top
.
ack·top
Jack-top
Three-terminal
male
Jack-top
Three
pair
pair
pair
pair
pair
connector
pair
jack-top
posts
binding-post
binding·
binding-post
binding-post
binding-post
recessed
binding-post
binding
post
Function
Connection
erator,
ac
Output
connection
ternal
detector.
Connection
component.
Connection
sistance
extend
Connection
bias
Power
Connection
of
Connection
indicator.
or
DQ
supply.
input
three-wire
to
external
and
de.
ac
to
unknown
to
external
capacitance
ranges.
to
external
connector
to
ground
power
to
external
from
only.
wire
line.
gen-
in-
re-
to
de
null
-

INTRODUCTION
c
R
G
GP
z
X
y
B
Q
SYMBOLS
capacitance
series
capacitance
parallel
inductance
series
parallel
resistance
senes
parallel
conductance
series
parallel
impedance,
reactance,
admittance,
the
qua
for
capacitance
(~)
inductance
inductance
(-'\JV\r) R
resistance
resistance
conductance
conductance
Z = R +jX
the
imaginary
. X B
!tty
factor
inductors
TABLE
AND
(
--1
t-)
(-w.,-)
imaginary
Y = G +
part
of
= R = G
or
inductive
1-2
ABBREVIATIONS
=-
G
RP
c;;
1
G = R
1
G
=
--
5
Rs
GP
part
of
an
jB
an
admittance
1
D
resistors
impedance
Q
EXT
AC
OFF
IN
OFF
POWER OFF
EXT
METER
ANO
OC
BUT
TO
POSITION
GEN 0/SCONNECTEO
SHUNTEO
IN
BETWEEN
/NT
OC
BR/OGE
METER
REOUCE
/NT
AC
OFF
POSITIONS
OPERATIVE
SHUNTEO
SENSITIVITY
for
capacitive
d"
D
PF
. . f R G 1
Issipatton
for
capacitors
power
actor
factor
frequency
cu
angular
frequency
0 ohm, a
kO
MO
mO
m 0
f.1
0 micromho 1
n 0
f.1
f
nf
pf
h
mh
f.lh
unit
of
kilohm 1
megohm 1
kO
MO
milliohm 1 mO = 0.001 0
mho, a
unit
of
millimho
1 m 0 = .0010
f.1
nanomho
1 n 0
microfarad, a unit
nanofarad 1 nf
picofarad 1 pf
henry, a unit
millihenry
microhenry
1 mh = 0.001 h
1 f.lh = 1 X
resistors
=x
D
R
~
= 2
nf
resistance,
= 1000 0
=
1,000,000
conductance,
0 1 x
of
capacitance
= 0.001
= 1 x
10-6
of
inductance
Q = cuCPRP
= B = Q
1
cuCp~
reactance
0
susceptance
10-6
0
j.lf
= 1 mj.lf
f.1
f = 1
lQ-6
h
or
impedance
f.lf.l
f
or
admittance
LOWEST THREE R
HIGHEST
RANG£$
FOUR G flANGES
Figure 1-3. Generator
connections.
HIGHEST FOUR R RANGES
LOWEST
THREE GRANGES
and
detector
3

of
nor
fre-
re-
The
as
1
__________________________
Rs
~---T_YP_E_1_6_08_-A
1.4
SERIES
pure
a
quency
sistance
actance
a
series
alent
series-parallel
relationships
follows:
AND
An
impedance
resistance
by
either a series
and
used
or
parallel
circuits
reactance.
in
the
are
conversion
between
I{
eststance
__
I_M_P_E_D_A_N_C_E
PARALLEL
that
is
can
be
represented
or a
The
equivalent
representation
shown
in
is
the
various
and
PARAMETERS.
neither a pure
at
any
parallel
values
circuit
Figure
given
Inductance
combination
of
resistance
depend
is
used.
1-4. A nomograph for
in
Figure
circuit
__
B_R_ID_G_E_
reactance
specific
andre-
on
whether
The
equiv-
1-7.
elements
are
D2 1 1
= 1
+D2
RP
= 1
+Q2
RP
(1
__
+Q2)GP
1 1
1
+D2
Resistance
LP
; L =
and
Capacitance
1c
Ls
Lp
Rp
Gp
Rs
+Q2
--L
P
Q2
= (1 + D2
s s
)L
Figure 1-4.
1.5
ACCURACY
1.5. 1
CGRL
accuracy
the
strument
or
beyond
tion
arm
the
has
fore,
ance
half a digit
in
resolution
an
is
adjustment
is
better
the
check
The
is
the
most
slightly
the
range
The
the
adjustment
over-all
Equivalent
complex
OF
MEASUREMENTS.
ACCURACY
0.1%.
This
and
stability
initially
and
two-year
procedure
lowest-resistance
most
affected
poorer
accuracy
for
fixed
on
of
the
accuracy
should
difficult
each
error
the
and
vernier
by
stability
counter
calibrated
hold
warranty
is
switch
specification
bridge
of
for
at 1 kc
impedance.
AT
is a function
of
the
given
ratio
than
is
±0.005%
read-out
the
rheostat.
(on
5
Cp
Rs
circuits
1 KC.
the
to
0.1%
period. A simple
in
Section
(1-ohm)
arm
and
the
0.
2%.
limitations
all
bridge
an
accuracy
to
lead
other
for
of
allows
This
but
for
The
basic
of
the
arms.
accuracy
5. 2.
resistance
set
ac.furately,
resistance,
arms.
the
lowest
full
scale
for
of
linearity
fixed
the
lowest
bridge
accuracy
The
of
±0.05%
for
calibra-
ratio
Thereimped-
or
backlash
error
gives
range)
Rp
Gp
of
in-
well
is
and
one-
and
4

of
0.105%
scale.
as
many
at
full
Therefore,
digits
on
scale
the
the
and
final
counter
0.15%
balance
as
possible.
at
one-tenth
should
1.5.2 TEMPERATURE COEFFICIENT.
temperature
30
ppm/
change
(50
°F).
the
change
perature
rne
oriage
stabilized
1.5.3 ADDITIONAL
TORS,
The
OQ
C
and L bridges
(high 0 or
resolution
c
or L adjustment
0.
5%02
effect,
extreme
low
enough 0 or a high
ligible
10
coefficient
°C
This
in
reading
This
change
in
the
change.
ana
co
at a temperature
of
means
for a
is
usually
unknown
For
the
"
ERRORS
LOW Q INDUCTORS,
dial
is
but
care.
(see
adjustments
are
low
Q)
of
these
added
to
somewhat
Precision
Figure
components
used
wire-wound
adjustments
to
its
the
specifications
better
enough Q to
1-5).
the
instrument
that
there
temperature
negligible
component
most
accurate
<u
uc
near
23 ° C
FOR
AND
HIGH Q RESISTORS.
for
phase
rheostats.
are
measured,
prohibits
full
resolution.
accuracy
components
for a
(7
HIGH
to
is
generally
make
be
made
The
over-all
is
less
may
be
a 0 03%
change
3 °
of
compared
similar
measurements,
~uvw"u
F).
D CAPACI-
balance
When
the
limited
balance
A
term
allow
for
possible
have
this
term
'lJ
/
of
with
than
10°C
tern-
on
lossy
of
this
with
neg-
full
with
"''-
the
the
of
a
The Q adjustment
sists
of
two
capacitor
capacitors
tively
large,
fore
necessa~y
100
10
~I
TOTAL
tOTAL
0.1
0.01
100
decades
with
appear
and
ERROR
ER!OR
infinite
as
the
(see
0•1
0•
0
0002
Ike
for
the
of
mica
capacitors
resolution.
an R or G error
added
error
Figure
1-6).
~
(f.,)
I I
011%/
--
~
2
-i
~
v;
1/
I
lL
FREQUENCY
'INTRODUCTION
Rs
and
Gp
term
and a variable
Losses
when Q
of
0.1%Q
bridges
in
the
is
is
con-
mica
rela-
there-
1
ff
r;
I
v
J
jj
VII
-;!;
/
/~oo,(ffir
w
I
I
I
I
1/
IOke
100
O.OIL.--L---L--L---L----L_L-
100
Figure 1-5.
ance
_
Ike IOke
FREQUENCY
Capacitance
errors
vs
_,___....._
and
frequency.
__
induct-
.._______,
IOOke
Figure
1.5.4
FREQUENCY
tional
error
on
is
the
inductance
standard
This
the
amounts
kc,
of
error
cies,
of
the
Figure
is
phase-compensated
error
added
is
large
(see
For
there
OQ
2-7.)
capacitor,
is
error
to
a 0.1%
enough
high D (low
is
rheostats.
1-6.
Resistance
ance
errors
vs
ERRORS.
the C and L bridges
of
the
bridge
effectively
proportional
term
Figure
an
added
The
0.001%
error
to
account
1-5).
Q)
This
series
to a large
to
(f/1
at
10
measurements
error
term
rheostat
and
conduct-
frequency.
The
main
at
wiring
increasing
f2,
and
is
kc)2.
kc
and
for
other
term
due
is
0.10
(Cs
degree,
cause
higher
in
series
accounted
This
a 0.4%
smaller
at
high
to
the
(f/
and
but
of
addi-
frequencies
with
its
value.
for
term,
which
error
at
sources
frequen-
inductance
1
kc).
(See
Lp
bridges)
neverthe-
the
in
20
s

~---T_YP_E_1_6_08_-A
__
IM_P
__
E_D_A_N_C_E_B_R_I_D_G_E_I
----------------------------
less
adds
inductance
tor.
The
inductance
bridges)
tor,
capacitance
Ls
more
Figure
the
network
pensate
The
portional w £2
added
(f/1
and
20
1.5.5
curacy
residual
tion
tance
final
more
larly
given
1
D
ments
is
placed
and
at
high
of
this
bridges
resistance
adds
kc
is
useful D and Q ranges
2-7).
A
frequency-dependent
built
into
for
stray
effective
kc)4.
(see
error
an
resistance
terms
The
extra
Figure
RESIDUAL TERMINAL IMPEDANCE.
specifications
terminal
is
considered.
can
be
easily
measured
complicated
at
the
range
in
Table
2-5 .
56
D
AND
Q
and Q accuracy
depends
in
series
of
the
parallel
in
parallel
enough D values
bridge
arm.
somewhat
and
capacitance
and
first
1-6).
impedance
value.
extremes,
less,
conductance
the
standard
of
ont:
prgpertieaal
±0.002
term
is
0.
2%
error
are
valid
The
residual
measured
At
corrections
high
and
ACCURACY.
specifications
upon
the
tracking
with
the
rheostat
with
the
effectively
The
error
and
these
at
high
error
term
bridges
resistance
(refer
to
this
arm
has
te
(f/1
kc)2
more
important
at
10
kc
only
if
of
the
UNKNOWN
resistance
and
subtracted
frequencies
are
necessary,
correction
The 5 percent
for C
accuracy
standard
standard
frequencies
is
paragraph
£4,
and
and
the
and L measure-
capaci-
(Cpand
capaci-
reduces
on
the
Cp
bridges
necessary
because
arm
to
one
term
req11iring
±0.000001
up
to
0.8%
error
The
effect
connec-
and
capaci-
from
somewhat
particu-
formulae
term
of
the
45
of
in
Ls
the
and
have
(see
for
of
com-
4. 5).
pro-
the
kc,
acthe
the
are
the
DQ
a
at
rheostats
±0.0005,
the
quired
This
made
ing.
CGRL
can
error
of
full
factor
achieve
cause
f.
Therefore,
1
kc.
be
improved
ard
bridges
capacitance
term
since
ever,
rk
tremely
not
of
resi~tors,
at
very
with
depends
bridge,
to
achieve
specification
down
Below
adjustment
add
additional
of
0.001
scale.
here.
better D and Q accuracy.
At
high
the
phase
At
frequencies
capacitor.
The
percent
is
is
±0.0005
the
same
for
the
at
higher
good Q accuracy
help
10
high
the
dial
upon
and
many
this
of
to
1/20
of
this
reading,
(R4),
DQ
at
1/100
The
detector
Lower
frequencies
angles
this
fixed
because
term
±2%,
which
decades
at 1 kc,
phase
Rs
and
and
at
the
measurement
which
is
frequencies).
calibration.
the
phase
compensating
accuracy
±0.0005
the
full-scale
the
phase
even
error.
of
full
scale
sensitivity
CGRL
below 1 kc,
it
used
angle
lower
ranges
the
of
the
bridge
error
term
is
limited
in
the Q accuracy
is
limited
for Q adjustment.
just
as
considerations
Gp
bridges,
frequencies.
at
low
of
independent
The
fixed
angle
of
each
components
(refer
to
paragraph
holds
for
measurements
CGRL
counter
angle
of
the
though
This
compensated
could
amount
and
0.005
is
also a Hmiting
should
DQ
error
arms
is
the D accuracy
by
the D of
by
the
in
the L and C bridges,
this
frequencies,
the
time
of
be
increases
increase
0.0005
1
for
accuracy
apply.
term
is
This
constant
frequency
Rs
terrrt,
arm
are
4.5).
read-
vernier
to
at
1/1000
used
~c
above
cannot
the
stand-
and
of
The
fixed
How-
±0.0005
gives
but
does
(Q/
(except
of
re-
for,
an
to
be-
with
Gp
the
ex-
CtJ)
7i

IINTRODUCTION
2 0.5
0.6
0.7
I
Cp
Ls
Rs
0.1
0.2
0.3
0.4
0.5 2
0.6
0.7
0.8
0.9
Cs
Lp
Gp
Rp
10
9
8
7
6
5
4
3
0
0.8
0.9
2
5
00
D
a
a
Figure
1-7. Nomograph for
C,
L,
R,
D,
conversion
and Q at 1 kc.
of
7

~--T_Y_P_E_1_60_8-_A_I_M_P_E_D_A_N_C_E
__
B_R_ID_G_E~I
__________________________
___
8
(ANY
-10>-_._-
XL
FREQ)
mfi
kfi
(I
kc
Q
mfi~-----------------------~
fi fi Ph
kn
Figure
1-8. Nomograph for
Rs
to
Ls
and
vice
conversion
versa.
of
Ls
ONLY)
nh
mh

OPERATING
PROCEDURE
SECTION
2.1
INSTALLATION.
2.1.1
suitable
the
power
or
230v,
plied.
2.1.2
GROUNDING.
erated
cific
cases
supply should
3.5).
The
wire
power
the
rear
nected
bridge
is
not
2.1.3
MOUNT
the
Type
for
relay-rack
equipped
1608-AR
stallation.
brackets,
to
convert
Type
General
pulled
can
be
2
POWER
power
receptacle
50-60
with
the
where
be
ground
cord
of
the
to
the
must
be
used,
this
lNG.
1608-AM, for
with
includes
Instructions
which
from
ZSU-6-7
Radio
out
on
slides
removed
CONNECTIONS.
source
cps). A three-wire
bridge
the
grounded
connection
to
the
instrument.
adjacent
ungrounded.
connection
The
mounting.
aluminum
mounting
may
be
bench
design
for
from
the
2.2 INTERPRETATION
The
main
CGRL
digits
displayed
hand
digits
centric
controlled
an
overlapping
(99)
to
the
extends
two
digits
the
number
simply
mediately
10303;
99X2
The
CGRL
beyond
substitute
are
by
the
next
on
the
10.
to
the
=10002.
letter
in
an
controlled
controls;
smaller
transition
higher
99,
up
same
To
interpret a reading
0 for
left
X
is
OPERATING
Connect
as
indicated
on
the
rear
of
The
bridge
should
chassis
unknown
3RD
instrument
bench
ordered
to
rack
mounting
which
service.
rack
indication
in-line
coarse
to
counter,
the X and add 1 to
of
also
grounded
component
(refer
to
is
made
WIRE GROUND
This
terminal
CHASSIS
If
the
should
mounting,
The
bench-mounting
end
frames,
brackets
for
assembly
separately
use.
brackets
permits
Either
independently
OF
•x•
IN
consists
read-out.
by
the
larger
the
two
(vernier)
from
full-scale
step,
106.
To
avoid
an X
the
X.
For
used
on
paragraphs
three•wire
is
right-hand
the
bridge
on
the
plate
the
instrument
power
cord
generally
except
or a de
through
terminal
should
terminal
be
available
the
unless
power
made
externally.
or
Type
while
for
relay-rack
accompany
(Type
are
of
instrument
chassis
of
READ-OUT.
of
The
of
;he
control.
the
is
two
vernier
vernier
the
ambiguity
used
containing
the
example,
of
the
To
in
to
above
(115
is
sup-
be
op-
in
spebias
3.1. 5 and
the
three-
be
con-
the
cord
as
either
1608-AR,
model
the
Type
in-
these
ZSU-6-7)
a uniquE
to
or
cabinet
the
other.
up
to
five
three
left-
two
con-
digits
provide
reading
read-out
place
an
digit
102X3 =
three
PROCEDURE
dials
used
stitute
0
left
a
on
is
be
are
of
of
X,
im-
Q
of
Users
data
exactly
cludmg
error
in
2.3
DC
2.3.1
PROCEDURE.
a.
check
the
if
necessary,
adjustment
b.
counterclockwise.
.
c.
resistance
sistance
measurements
d.
KNOWN
e.
f.
the
concentric
(center)
LEV
controls
A
meter
known
For
greatest
four
digits
g.
the
counter
the
unit
tor
is
explained
with
the
in
place
of
the
X.
For
may
find
as
it
any
X' s
that
the
interpretation
RESISTANCE
With
the
NULL
center
on
the
meter.
Turn
the
Set
the
BRIDGE
measurements
mea
snrements
from 0
Connect
terminals.
Turn
As
INT
tion
meter
liminary
in
control
Adjust
reading,
deflection
is
larger
The
illuminated
the
the
function
DC,
where
sensitivity
this
position
turned
the
CGRL
for
than
accuracy
showing.
value
with
the
in
the
function
balance
and
Rs
and
Gp
bridges.
the X and add 1 to
example,
it
appears
appear.
.1X4
helpful
on
of
the X can
In
to
the
that
MEASUREMENTS.
function
meter
to
resistor
it
passes
the
FULL
increased
to
If
of
decimal
above.
paragraph
mechanical
the
pointer
DET
SELECTOR
from 0
above
1.1 mho.
switch
NOTE
switch
circuit
is
may
instead
down.
balancing
adjust
the
right
the
indicated
the
reading
not,
turn
the
UNKNOWN
The
switch
SENS
to
1
MD
to
be
is
through
greatly
be
SCALE
the
sensitivity
to
point
meaning
2.2.
Here
=.204;
record
bridge
way,
be
(1,
zero
with
the
control
switch
1.1
MD
and
measured
to
lNT
rotated
an
undetented
is
operative
reduced.
made
of
with
RANGE
controls
DET
indicates
CGRL
should
the
next
is
correctly
again,
the
digit
.2XX
measurement
read-out,
any
possible
rechecked.
Figure
position,
screw-driver
almost
to
and
Gp
for
conductance
to
DC.
from
with
the
the
DET
switch
for a
SENS
and
as
necessary.
that
dial
have
lower
read
directly
located
of
an X indica-
sub-
to
=.310.
1-1)
and,
fully
Rs
for
the
UN-
OFF
posi-
but
A
pre-
switch
SENS
and
zero
GEN
the
setting.
at
least
range.
the
in-
off,
for
re-
to
the
un-
on
and
9

~---T_YP_E_l_6_08_-A
__
IM_P_E_D_A_N_C
__
E_B_R_I_D_G_E
__________________________
___
2.3. 2 ACCURACY.
conductance
sea
e
lowest R and
ficient
the
accuracy
±1
m.O,
For
leads
component.
and
terminals
subtract
known
screw
inserted
2.3.3
KNOWN.
ing a limiting
1/2
watt
en~s
or
GEN
LEV
proximating
tally"
350-volt
SCALE
each
range
Because
vouage
than
the
tual
voltage
ured
on
resistors
2-1
shows
from
the
tance
range.
LEV
control
use
of a high-impedance
w
the
urements,
ratio
between
across
and
therefore
bridge
the
GEN
given
in
2.3.4 EXTERNAL
supplies
from 1 ohm
DET
SENS
urements
sary
to
measurements
h.
h I f h d. · )
(
to
EIA
is
W
lC
highest G ranges
sensitivity.
is
low-resistance
should
the
UNKNOWN
the
input
be
Measure
by
this
amount
1n
place.
binding
in
the
hole.
INTERNAL
There
resistor
or
less
to
the
unknown.
panel
3. 5
the
bridge
supplies
RANGE
as
given
app1iea
open-circuit
applied
the R bridge
specifications
are
summarized
that
these
internal
For
can
where
the
whole
the
(LOW UNKNOWN
LEV
control
Tables
and
the
to
1 megohm
controls
beyond
make
measurements
±1
are
to
control.
volts
switch
of
ro
power
2-2
this
The
accuracy
is
2 0 t e
On
the
lowest R and
limited
used
connecting
are
low-resistance
be
terminals.
the
voltage
bridge
internal
by
measurements,
as
connections
the
zero
from
the
For
best
post
hard
VOLTAGE
three
internal
to
limit
avoid
damage
They
The
open
circuit,
(see
Figure
applied
selects
in
Table
limiting
the
rne
unKnown
value.
to
any
unknown
(with a 115-volt
for
in
standard
supplies
set
for
de
voltmeter
voltage
across
is
fixed
voltmeter
set
to
and
2-3.
DC
DETECTOR.
detector
to
0.1%
are
at
maximum.
range
are
at
of
de
±0.1%, ±0.005%
ast
lglt
as
long
the
sensitivity
resistance
the
free
ends
bridge
reading
connection
enough
lowest
"vertically".
2-1.
Figure
testing
Tables
voltages
the
For
is
can
terminals
give
to
APPLIED
de
supplies,
the
available
to
the
are
all
controlled
voltage
is
applic:d
1-3)
and
the
optimum
resistor,
is
usuauy
2-1
resistor
line
different
2-2
over
most
measurements
desired
connected
high-resistance
applied
the
unknown
over
each
be
placed
the
permit
when
the
If
desired
lower
voltages,
resistance
on
a
as
there
highest G range
short,
to
the
of
together,
with
to
the
notch
TO THE
bridge
supply,
the
35-volt
The
the
or
shows
when
voltage).
and
2-3.
can
be
of
test
vertically,
range
to
chassis)
"bridge
The
internal
measurements
GEN
accurate
or
if
it
of
ll
b h
ut
is
to
±1/2%
heavy
unknown
the
lead!;
the
bridge,
the
each
power
compon-
by
"horizon-
FULL
supply
maximum
mucn
the
meas-
types
Figure
supplied
the
resis-
the
voltage
directly
meas-
and
at
across
voltage"
LEV
meas-
is
neces-
an
exter-
and
full
t e
suf-
wire
UN-
hav-
and
re:s:s
GEN
that
null
and
and
and
un-
to
the
ap-
for
ac-
of
by
the
the
de
000
If
100
I
r
'Y
10
,
li
I
[;
't
o.1
I
Jy
0.01
0
Figure 2-1.
I
I
I
I
I
TO
IM
...,.,.,.
TO
TO
lOOk
IOk
IOOv
~
10
~;-k
"""
I
I
v
1/f
i"""""""
elk
IOk
I
I
I
I
v
~
-
d 100.0.
TO
lk
~~~
I
I
I
~
...,,-
~~o~,o~
/
//
,oo~
b
/
Va
)~
I
I
I
I
I
I
I 2 3 4 5 6
/
V"
De
voltage
resistor
applied
(115-v
30
v
v
.0.
2.5
I I I
o:oL~
1.0.
TO
~
0.1
TO
7
to
line).
TO
3 v
I
10.0.
~
-
1.0.
8
unknown
-
-
,.,.,...
--
10
10

OPERATING
PROCEDURE
FULL
SCALE
Rs
Bridge
GP
Bridge
RESISTANCE BRIDGE
2.7-9.9
10-99
100-999
1000-9999
10-
100kD-1MD
1
VERTICAL
HORIZONTAL
FULL
SCALE RANGE
VERTICAL
HORIZONTAL
D
D
D
D
99k
D
MD-up
HVS HIGH-VOLTAGE
MVS
LVS
DC
SOURCE
RANGE
MEDIUM-VOLTAGE
LOW-VOLTAGE
Rs
Rs
Rs
Rs
Rs
Rs
GP
1100
1100nU
EIA
STANDARD
Fixed
Composition
RANGE EIA
llD
llOD
llOOD
llkD
llOkD
llOOkD
1 n U
mD j 11 D 11100
SUPPLY
SUPPLY
TABLE
AND
DETECTOR
METER
LVS
In
flu
HVS
METER
SUPPLY
TABLE
TEST
2-1
!no
- 35G"
2-2
VOLTAGES
Resistors
TEST
0. 5 - 1 v
0.5
2.5 - 3 v
8-
24 8080-
CONNECTIONS
1100 D j
1100
flu
gp€:n-Giwdt
35v
open-circuit
3.5v
oyen-circuit
(RS172)
VOLTAGE BRIDGE VOLTAGE*
MVS
fLU
MVS
11
In
kD
METER
mU
METER
110 kD
110 mU
f 1100 kD
HVS
j1100
LVS
••
- 1 v
10 v
30
100v
100 v
v
••
19.2-
23v
13.4-16.7v
25.6-
32 v
81 - 101 v
81 - 101 v
mU
RESISTANCE
less
tha"n
10
10-
99
D
100-
999
10oo-
9999
10-
99kD
100kD-
*
* *
This
also
This
1MD
1MD-
is
the
voltage
D
D
D
up
the
voltage
voltage
Low-Power
BRIDGE RANGE
Rs
Rs
Rs
Rs
Rs
Rs
Gp
from
the
LOW
at
the
EXT
GEN
varies
with
the
resistance
EIA
STANDARD
Fixed
Film
Wire-Wound
11.1"2
1100
1100
D
llkD
llOkD
1100kD
1
nU
UNKNOWN
terminals.
of
TABLE
Resistors
Resistors
terminal
the
unknown
2-3
TEST VOLT
(RS-196)
(REC-117
EIA
MAX
0.3
h
3v
10v
30v
100v
100v
to
chassis.
(see
AGES
up
to
VOLTAGE
v
In
paragraph
9999
the
4. 3).
D)
EXT
MAX
BRIDGE VOLTS*
••
••
23v
16.7
32v
101 v
100v
DC
position,
v
this
is
11

~---TY_P_E_l_60_8_-A
EXT
__
IM_P_E_D_A_N_C_E
METER
__
B_R_I_D_G_E
__________________________
EXT
EXT
METER
METER
___
(a)
nal
detector
external
internal
the
meter
CONNECTIONS
in
Figure
with
detector
meter,
by
appropriate
2-2.
increased
in
on
2.3.5 .EXTERNAL
required
may
directly
DC
across
Be
given
components.
measurement
ternal
be
used.
across
position
the
careful
in
Table
When an
supply
on
the
of
horizontal
not
to
procedure
and
The
the
the
exceed
2-4
external
can
be
parallel
the
rear
DC
unknown
EXT
vertical
function
diagonal
in
order
supply
is
detector
sensitivity
connected
with
the
connection
of
the
SUPPLY.
resistor,
GEN
terminals
bridge
switch
on
the
maximum
to
avoid
or
the
same
except
Figure
can
in
series
meter,
to
the
instrument
If
higher
an
external
diagonal
and
the
the
top
voltage
damage
detector
as
that
that
the
2-2.
be
or
in
EXT
are
in
detector
four
to
is
with
used.
voltage
the
GEN
TO
BRIDGE
(b)
External
The
with
the
place
of
METER
as
shown
supply
connected
the
EXT
ranges.
or
current
bridge
used,
the
the
in-
LEV
is
is
meter
control
ply
sitivity
2.4
TOR.
2.4.1
2.4.
Cs
than
Cp
tween
Gp
Cp
connections.
does
not
and
the
OET
SENS
of
an
external
AC
MEASUREMENTS
1-KC
CAPACITANCE
1.1
Procedure.
a.
Set
the
GEN
b.
Set
the
BRIDGE
-
if
the
series
1.
-
if
the
parallel
0.02
and
2.
(Note:
-
if D is
=
Q~e_).
Cs = Cp
greater
~·
(c)
control
the
control
detector.
USING
MEASUREMENT.
LEV
control
SELECTOR
capacitance
capacitance
within
than 2 (measure
level
does
INTERNAL
fully
is
desired
is
0.1%
of
an
external
not
control
the
GENERA-
clockwise.
to:
and
D
is
desired
if
and
D
0(
0.03.)
as a conductance,
is
sup-
sen-
less
be-
12
MAXIMUM
BRIDGE
Rs
Rs
Rs
Rs
Rs
Rs
Rs
GP
EXTERNAL
RANGE
llOOm.O
11.12
110.12
1100.0
llkD
llOkD
llOOkD
lOOOnU
DC
E
14.2v
22v
71v
223v
400v
400v
TABLE
BRIDGE
MAX
1.4v
4.5
v
2-4
VOLTAGE
I
MAX
710ma
223ma
71ma
17.2ma
17.2ma
17.2ma
17.2ma
17.2ma
AND
CURRENT
TERMINALS
EXT
EXT
EXT
EXT
EXT
BIAS
BIAS
BIAS
GEN
GEN
GEN
GEN
GEN

OPERATING
PROCEDURE
c.
Set
the
function
d.
Connect
1/2
means
terminals.
e.
If
the
is
not
reading
control
SCALE
f.
Adjust
control
for minimum
eeatrel
as
g.
The
on
the
and
unit
directly
meaning
2.2.
Accuracy.
the
of
the
last
where
of
full
to
subtract
bridge
terminals,
to
determine
If
this
zero
from
between
of a three-terminal
(refer
The
residual
have
for a
=
0.006
are
used
may
in
Table
When
capacitors
error
(refer
to
Dis
less
1-KC
INDUCTANCE
a.
Set
the
KNOWN
RANGE
for a
SENS
FULL
flection.
DQ
SENS
sensitivity
directly
point
dicated
The
graph
2.4.1.2
±0.1% ·of
±
range,
±0.005%
essary
the
reading
capacitor.
known,
subtracted
itance
by
urement
bridge
except
error
leads
capacitors, a correction
ductance
given
additional
tion
when
2.4.2
2.4.2. 1 Procedure.
the
proper
known,
somewhere
for
an
RANGE
the
may
have
balance
capacitance
counter
illuminated.
on
the
of
an X indicator
reading
digit)
the
accuracy
scale.
the
external
capacitance
the
reading.
the
terminals
to
paragraph
negligible
slight D error
when
when
be
necessary.
2-5.
of
±(0.5%)
paragraph
than
GEN
switch
unknown
range
setting
set
the
concentric
near
upscale
concentric
is
The
meter
switch
meter
te
be
readjust€:d
approached.
of
the
readout
The D of
illuminated
accuracy
±0.005%
on
all
but
is
On
the
residual
approximately
the
correct
leads
are
is
The
and
shielded
3.2).
resistance
effect
on
Cx
= 1000
measurements
for
the
with
high
o2
1.5.3).
0.2.
MEASUREMENT.
LEV
control
to
INT
AC.
capacitor
of
the
FULL
CGRL
5000,
adjust
reading
for a minimum
CGRL
controls
deflection.
t9
unknown
with
the
correct
the
unknown
scale
on
the
is
explained
of
the C reading
of
full
scale
the
highest
±0.2%
of
lowest C range
("zero")
value
used
increased
error
and
on
the
pf).
lead
The
is
added
This
capacitance
0.
25 pf, from
of
the
to
connect
and
caused
leads
may
capacitance
inductance
the
C or D
highest C range
However,
are
made
resistance
correction
D's
are
measured,
to
the
error
fully
clockwise.
to
the
UN-
SCALE
controls
the
DET
and
set
the
meter
de-
and
the
The
DET
gi"€: gr€:at€:r
is
indicated
decimal
is
in-
DQ
dial.
in
para-
(which
capacitance
the
reading
it
is
nec-
of
the
unknown
the
un-
should
by
capac-
be
removed
meas-
of
the
accuracy
(D
if
long
on
large
and
in-
terms
are
specifica-
is
negligible
is
is
be
an
Note:
tance
measured
(refer
to
b.
L s -
if
tween
0.5
Lp
-
if
than
1.
Rs
-
if Q is
Ls
c.
d.
KNOWN
e.
RANGE
for a
SENS
control
SCALE
f.
DQ
control
SENS
control
sensitivity
g.
directly
point
and
dicated
The
meaning
2.
2.
2.4.2.2
±0.1%
±
1/2
of
where
the
full
scale.
of
0.5%
5 or
higher.
On
the
binding
reading
leads
are
bridge,
and
subtracted
inductance,
ance
on
careful
residual
tance
tween
the
For
will
paragraph
Set
the
and
the
parallel
(Note:
=~
Set
Connect
terminals.
If
is
reading
RANGE
Adjust
on
unit
directly
Accuracy.
of
the
the
accuracy
~2,
the
in
then
the
to
inductance
measurement,
2.4.5.4).
the
BRIDGE
series
50.
Ls = Lp
less
than
refer
liJ
the
function
the
the
proper
not
known,
somewhere
for
an
switch
the
for minimum
may
as
balance
The
inductance
the
counter
illuminated.
on
of
an X indicator
reading
last
When Q
which
lowest
posts
order
used
to
the
residual
from
short
Rs
bridge,
keep
the
leads
would
some
iron-cored
depend
SELECTOR
inductance
inductance
within
0. 5
(measure
to
paragraph
switch
inductor
range
set
the
near
upscale
for a minimum
concentric
meter
have
to
be
is
approached.
of
readout
the
illuminated
The
accuracy
±0.005%
digit)
on
is
±0.2%
is
low
is
negligible
range,
(0.14
fLh)
to
obtain
connect
inductance
the L reading.
the
leads
and
lead
configuration
measurement
since
an
increase
inductors
upon
the
is
desired
is
desired
1%
if Q )32)
Rs
2.4.3).
to
INT
to
be
setting
concentric
5000,
reading,
CGRL
deflection.
readjusted
the
unknown
with
The Q of
is
explained
of
all
but
of
the
there
is
when Q
the
residual
must
be
full
accuracy.
the
unknown
together,
calculate
increase
the
excitation
to~
and Q is
and
and
Q;
AC.
measured
of
the
FULL
CGRL
adjust
and
set
meter
controls
to
the
correct
the
unknown
scale
of
in
of
the L reading
full
scale
the
lowest
reading
an
additional
is
approximately
inductance
subtracted
inductor
should
To
measure
mea
sure
Ls
the
same
and
the
in
the
residual
the
induc-
level
Q is
greater
to
the
UN-
SCALE
controls
the
DET
the
FULL
deflection.
and
The
DET
give
greater
is
indicated
decimal
is
the
DQ
dial.
paragraph
(which
ranges,
±0.005%
error
from
If
external
to
be
measured
this
lead
the
imped-
=
Q:s.
for
total
induc-
area
inductance.
be-
the
in-
is
is
of
of
the
the
Be
the
be-
13

~---TY_P_E_1_60_8_-A
The
residual
mately
is
more
capacitance
very
ligible
tor.
0.9mD.
small.
important
large.
compared
Long
This
If
long
{see
of
0.25
However,
leads
resistance
can
leads
pf
with
to
the
__
IM_P_E_D_A_N_C_E
of
the
cause a small Q error
are
used,
Table
2-5).
can
cause
an L error
this
capacitance
the
capacitance
inductor
may
__
bridge
the
Q error
The
residual
when
is
usually
of a large
appreciably change
B_R_I_D_G_E
is
approxi-
when
Lx
becomes
bridge
Lx
is
neg-
induc-
____________________________
the
total
effects
ditional
cations
ible
when Q is
capacitance.
are
given
When
error
(refer
in
inductors
term
to
paragraph
greater
Table
with
of
±0.
than
The
2-5.
low
5%
1. 5. 3).
5.
corrections
Q'
s
are
~
is
added
2
This
for
measured,
to
error
these
the
is
lead
an
ad-
specifi-
neglig-
_
CORRECTIONS
Measured
R0 =0.9mD
Cs
D
cp
D
GP
Q
Rs
Q
FOR
ERRORS
(Add or
Series
subtract
Resistance
+leads
NO
ERROR
-u.;CxRo
+2RoDxu.;Cx2
-wCxRo(l + Dx2)
+Gx2Ro(l
+QxGxRo(l-
+ Qx2)
-Ro
Ro
QxR
X
CAUSED
from
2
)
Qx
TABLE
BY
the
measured
Series
L0=0.14
-w2LoCx2(1-
-w2LoCxDx(l
+u.;2L
+wL0Gx
2-5
TERMINAL
Inductance
,uh
-u:2LoCx2
-w2LoCxDx
2G
3(1-
o x u.;GxLo
{1 +
NO
ERROR
_ cvL
Rx
value
+leads
Q/)
0
AND
as
Dx2)
+ Dx2)
~)
LEAD
indicated.)
Parallel
C0 =
+ D - (1 + D
+u.;2 C 2R
IMPEDANCES
Capacitance
0.25
pf +
-CoO
-Dx2)
co
X
CX
-Co
+Co
Dx
Cx
NO
ERROR
-wC
0
Gx
3(1-
0 X
+
u.;
Co
Rx
( 1 +
leads
2
X
--)
u.;CORX
Qx
2)
)
2Qx
14
Ls
Q
LP
Q
NO
+Qx
+
ERROR
2
wL
2R
+-
Qu.;
Ro
--
wLx
Ro
0
(1
X
+ Q2)
-w2C
-Lo
-
Lo
Qx
LX
(1-
(Qx +
1
-)
Qx2
1
Q}
-L
o
Lo
-
LX
L
o x Qx2
+w2CoLx(Qx
-u;2
CoLx
+w2CoLxQx
2(1-
1
-)
1
+Qx)
2

2.4.3
1-KC RESISTANCE AND CONDUCTANCE MEAS-
UREMENTS.
2.4.3.1
Rs
of
is
Gp
tance
unknown
measurement.)
terminals.
RANGE
for a
SENS
FULL
flection.
three Q controls
The
greater
is
imal
is
pacitive
is
the.
which
of
in
Procedure.
a.
Set
b.
Set
-
if
series
the
unknown
inductive.
-
if
parallel
of
the
is
(Note:
of
the
Rs
tive;
likewise,
quire
use
In
the
range
the
resistor
quired
either
bridge.
vice
versa,
c.
Set
d.
Connect
e.
If
is
reading
control
SCALE
f.
Adjust
DET
SENS
sensitivity
g.
The
indicated
point
read
directly
as
less
than
decimal
makes
0.00
1.
paragraph
the
the
unknown
capacitive.
Any
bridge
unless Q is
the
the
not
directly
and
indicated
0, in
point
The
2. 2.
GEN
LEV
control
BRIDGE
resistance
is
between 0 and 1 MD
conductance
resistor
of
the
between
will
Rs
from
function
the
proper
known,
somewhere
for
an
RANGE
the
for
control
as
resistance
unit
on
which
major
meaning
SELECTOR
is
is
between 0 and
(Refer
small
because
any
resistor
Gp
bridge
1 D
determine
small
may
be
the
formula
switch
unknown
range
setting
set
the
near
upscale
switch
concentric
the
be£t
may
have
balance
or
conductance
on
the
counter
illuminated.
the Q readout
by
the
lights
case
in
the
first
divisions
of
fully
to:
desired,
is
of
and 1 MD
enough
calculated
resistor
concentric
meter
minimum
is
the
an X indicator
and
or
desired,
to
Section
enough
on
to
value
will
large
enough
will
be
the
which
bridge
to
permit
from
Rs
=
_.........,
(1
to
INT
A C.
to
of
the
5000,
adjust
reading
for a minimum
CGRL
controls
meter
to
be
readjusted
approached.
readou~
The Q of
and
is
(unless
opposite
(coarsest)
the
vernier
clockwise.
the
resistance
if
the
unknown
and
the
conduc-
1 mho or
3.8
for
require
capacitive.
t Q2)
the
CGRL
of
inductive
use
be
induc-
to
re-
phase
is
re-
use
Gp,
and
__
_
Gp
UNKNOWN
FULL
SCALE
controls
the
and
set
meter
and
deflection.
to
the
unknown
with
the
the
unknown
the Q balance
is
true).
adjustment,
dial
is
explained
if
of
of
DET
give
dec-
or
Note
steps
the
RP
the
de-
the
ca-
from
the
measured
to
connect
tance
gether,
the Q of
given
measured,
the
specification
is
practically
2.4.4
TOR
AT
cillator-detector
usually
same
fications
an
external
Section
required
to
be
2.4.5
2.4.5. 1 Capacitance
pedance
ponents,
directly
for
measurements
graph
minal
2-3)
enough,
from
the
bridge
ficant
2-6
gives
each
quantity
citance
0.15
p.J,
the
unknown
of
these
leads
and
subtract
Residual
in
MEASUREMENTS USING
supplied
as
2. 5).
calculated
NOTES
3.5).
is
and
cause a reduction
and
if
Note
causes a small
inductance
the
resistor.
Table
2-5. When
an
additional
(refer
negligible
FREQUENCIES
tuning
is
for 1-kc
and D and Q ranges
for
Bridge
since
to
placed
other
the
it
measurements,
generator
The
the
various
(refer
ON
generally
neither
the
panel,
Capacitance
directly
does
not
UNKNOWN
therefore
stray
the
error
measured.
that
for
takes a stray
OPERATING
resistance.
by
this
value
Corrections
error
to
when Q
OTHER
unit
used,
at
plug-in
bridges
to
paragraph
AC MEASUREMENTS.
to
Ground.
UNKNOWN
which
on
grounded
across
cause
causes
capacitance
caused
the
capacitance
capacitance
capacitance
Use
resistor,
connecting
from
and
capacitance
resistors
term
paragraph
is
THAN 1 KC.
other
the
operating
but
are
the
same
unit
gives
so
measures
should
components
from
the
the
an
error,
in
sensitivity.
terminals
an
error
is
by a
short,
measure the
the
the
for
with
of
0.1%Q
1.
less
INTERNAL
thaa
the
the
same
frequency
that
2. 5.1).
The
"ungrounded"
terminal
be
LOW
detector
but
shunts
which
large
stray
bridges
error.
of
PROCEDURE
heavy
leads
resis-
free
ends
to-
measured
these
5. 3).
than
the 1 kc
procedure
accuracy
the
DQ
it
does
Type
grounded
(refer
UNKNOWN
can,
enough.
capacitance
150 pf
value.
affect
only
effects
high
Q' s
is
added
This
0.2.
GENERA-
If
an
is
speci-
as
those
(refer
multiplier
not
have
1608-A
com-
is
connected
except
to
para-
(see
Figure
if
large
Capacitance
an arm of
can
be
signi-
Table
stray
capa-
Since
Ct
to
cause
are
are
to
term
os-
uait
the
for
to
Im-
ter-
the
for
is
2.4.3.2
is
±
est G ranges
±0.005%
the
±0.1%
1/2
Accuracy.
of
the
of
On
bridge
of
the
The
the
reading
last
digit)
on
where
the
accuracy
full
scale.
lowest R range
(approximating
accuracy
±0.005%
all
but
the
0.9
mD)
of
the R or G reading
of
full
scale
the
lowest R and
is
±0.2%
of
the
residual
should
resistance
be
subtracted
(which
high-
reading
is
of
Figure 2-3.
Capacitance
bridge diagrams,
showing
to ground.
and
inductance
capacitances
15