General Life Biotechnology BeneCheck User Manual

BeneCheck Meter Kit Meter (Front Side & Back Side) / BeneCheck Messgerät
Messgerät (Vorderseite & Rückseite) / Kit Medidor BeneCheck Medidor (Parte delantera y parte trasera) / Kit de l’Appareil BeneCheck Appareil (Face Avant et Face Arrière)
Serial Number Sticker Seriennummer Pegatina de número de serie Vignette du Numéro de Série
Battery Batterie Batería Pile
Battery Lid Batterieabdeckung Tapa de batería Couvercle de la pile
Meter Label Messgeräteetikett Etiqueta de medidor Étiquette de l’Appareil
LCD Screen LCD-Display Pantalla LCD Écran LCD
Right “S” Button Rechte S-Taste Botón “S” Derecha Bouton “S” droit
Left “M” Button Linke M-Taste Botón “M” Izquierda Bouton “M” gauche
Test Port Test-Port Puerto de análisis
Orice de Test
Ejection Button Auswurftaste Botón de expulsión Bouton d’Éjection
Alert Tones: • Normal Alert: a short “beep”
• Warning Alert: 3 short “beeps”
• Turning On/ Off: a long “beep” Alarmtöne: • Normaler Alarm: ein kurzer “Piep”
• Warnalarm: 3 kurze “Piepser”
• An-/Ausschalten: ein langer “Piep” Tonos de alerta: • Alerta normal: un pitido corto
• Alerta de advertencia: 3 pitidos cortos
• Encender/ Apagar: un pitido largo Tonalités d’Alerte : • Alerte Normale: un “bip” court
• Alerte de Mise en Garde: 3 “bips” courts
• Marche/ Arrêt: un “bip” long
LCD Screen Display: Information and test result display LCD-Displayanzeige: Anzeige von Informationen und Testergebnissen Visualización de pantalla LCD: Visualización de información y resultados de prueba
Afchage de l’Écran LCD: Afchage de l’information et du résultat de test
① ② ③ ④ ⑤ ⑥
⑧ ⑨ ⑩ ⑪
Result Area / Messergebnisse/ Zona de resultado/ Zone du Résultat
Low Battery / Schwache Batterie / Batería baja / Batterie Faible
Temperature Icon / Temperatursymbol / Icono de
temperatura / Icône de Température
Average (GLU) / Durchschnitt (GLU) / Media (GLU) / Moyenne (GLU)
Unit Icon / Einheitssymbol / Icono de unidad / Icône
d’Unité
System Check, Data upload successful / Systemüberprüfung, Daten-Upload war erfolgreich / Control del sistema, Subida de datos correcta
/ Vérication du Système, Téléchargement des données réussi
Strip Loading Icon / Streifenbeladungssymbol / Icono de tira cargando / Icône de Chargement de la Bandelette
Blood Loading Icon / Blutentnahmesymbol / Icono de sangre cargando / Icône de Chargement du San
Bluetooth on / Bluetooth an / Bluetooth activado /
Bluetooth activé
Test Mode Icon / Testmodussymbol / Icono de modo de prueba / Icône de Mode de Test
Memory Mode Icon /Speichermodussymbol / Icono
de modo de memoria / Icône de Mode Mémoire
Code Number Icon / Codenummersymbol / Icono de
código numérico / Icône de Numéro de Code
Test Strip / Teststreifen / Tira reactiva / Bandelette de Test: GLU (GDH-FAD)- Glucose / Glucose / Glucosa / Glucose CHOL- Total Cholesterol / Gesamtcholesterin / Colesterol Total / Cholestérol Total UA- Uric Acid / Harnsäure / Ácido úrico / Acide urique
Electronic Contact Bars Elektronische Kontaktleisten Rodillos de contacto electrónicos Barres de Contact Electroniques
Sample Inlet Probeneinlass Entrada de muestra Entrée de l’échantillon
Code Strip: Contact Bar Code-Streifen: Kontaktleiste Código de tira: Rodillo de contacto Bandelette de Code: Barre de Contact
Test Mode Testmodus Modo de prueba Mode de Test
Code Code Código Code
UA
CHOL
7142
Labelling and Information / Etikettierung und Informationen / Etiquetado e información/ Etiquetage et Information
Caution, consult accompanying documents
Vorsicht, Begleitdokumente lesen
Cuidado, consulte los documentos adjuntos
Attention, consulter les documents joints
• Do not re-use • Nicht wiederverwenden
• No reutilizar • Ne pas réutiliser
• Keep dry • Trocken halten
• Mantener seco • Garder au sec
• Operation temperature limitation
• Temperaturbegrenzung Arbeitsbereich
Límite de temperatura de funcionamiento
Température de Fonctionnement Limitée
In-vitro diagnostic • In-Vitro-Diagnostik
• Diagnóstico in vitro
Diagnostic In-vitro
• Lot number • Chargennummer
• Número de lote • Numéro de lot
• CE certication • CE Zertikat
• Certicado CE • Certication CE
Read instructions • Anweisungen lesen
Leer instrucciones • Lire les instructions
• Storage & Transportation Condition
• Temperaturbegrenzung Lagerung und Transport
Condiciones de almacenamiento y transporte
• Condition de Transport et de Stockage
• Use by • Verwenden bis
• Fecha de caducidad • Utiliser avant
EU Representative
• EU Repräsentant
• Representante UE •
Représenté dans l’UE
• Date of Manufacture • Herstellungsdatum
• Fecha de fabricación • Date de fabrication
• Manufactured by • Hersteller
• Fabricado por • Fabriqué par
• Keep away from sunlight
• Vom Sonnenlicht fernhalten
• Mantener alejado de la luz solar
Conserver à l’abri de la lumière
Operation
90%
10%
Storage &
Transportation
• Serial number • Seriennummer
• Número de serie • Numéro de série’
Storage & Transportation Relative Humidity
Lagerung & Transport Relative Luftfeuchtigkeit
• Humedad relativa de almacenamiento y transporte
Humidité Relative pour le Transport et le Stockage
Storage &
Transportation
4
Comply with WEEE Directive 2012/19/EU
Entspricht der WEEE Richtlinie 2012/19/EU
Eento con la Directiva WEEE 2012/09/EU
Conforme à la Directive DEEE 2012/19/EU
EN
Chapter 1 Introduction ................................................................................. 1
Chapter 2 Setting the Meter
2.1 Installing/ Replacing the Batteries ...................................................... 2
2.2 Set the Date and Time .................................................................... 2
2.3 Code the Meter................................................................................ 2
Chapter 3 How to Perform a Test
3.1 Perform the Test ........................................................................... 3-5
3.2 Alternate Site Testing (AST) .......................................................... 5-6
Chapter 4 Meter Memory Function ................................................................ 7
Chapter 5 Control Solution Test .................................................................... 8
Chapter 6 Transmission Function ................................................................ 9
Chapter 7 Care and Maintenance
7.1 Storing Your Meter and Strip ............................................................10
7.2 Cleaning and Caring for Your Meter ...................................................10
Chapter 8 Error Message and Trouble Shooting .......................................11-12
Chapter 9 Specication ................................................................................13
1
Chapter 1
Introduction
Introduction
Please read carefully before using meter kit, and consult healthcare professional before making any important medical decision. Please contact your local customer service for further assistance with the product.
Normal Environmental Conditions
This meter designed under the following conditions:
• Indoor use • Overvoltage category II • Pollution degree 2
Electromagnetic Compatibility
This meter meets the electromagnetic compatibility, emission, and immunity, and the requirements of IEC 61326-2-6, ISO 18113-5 and IEC 61010-1.
Intended Use
In vitro diagnostic use only.
• Do not inhale or swallow
• Measuring blood glucose in fresh capillary whole blood from ngertip, palm, or forearm/
venous blood /arterial blood.
• Measuring total cholesterol/ uric acid in fresh capillary whole blood from ngertip.
• The meter can be used by laypersons or healthcare professionals.
Principles of the Examination Method
• Electrochemical biosensor technology. The meter is plasma-calibrated by reference instruments, which are traceable to the following
standard reference materials and methods.
Test Standard Method Glucose NIST SRM 917 Glucose Dehydrogenase Total Cholesterol NIST SRM 911 Abell ⁄ Kendall Uric Acid NIST SRM 913 Uricase ⁄ UV
2
2.1 Installing/ Replacing the Batteries
This meter uses AAA battery * 2. Please remove the plastic tab under the battery before using. Note: Dispose the batteries according to your local environmental regulations.
2.2 Set the Date and Time
Press “S” Button (3 secs)→One “Beep” Sound→Setting Mode→Turn Off Automatically after Setting
• Setting order: Year/ Month/ Date/ Hour/ Minute
• Press “M” button to advance one unit, “S” button to enter next setting.
Note: Correct setting is important while managing your health records.
2.3 Code the Meter (Total Cholesterol/ Uric Acid)
• Code your meter when you rst use it or open a new vial of strips.
• Make sure the meter is off before you insert the code strip.
Chapter 2
Setting the Meter
1 2 3
AAA
AAA
3
• Make sure the codes on screen, code strip, and strip vial label are the same.
Note: With Auto Strip Recognition function, once you code your meter, the meter will switch to the test mode automatically when you insert a strip.
Materials you need to perform a test:
BeneCheck Meter/ BeneCheck Test Strip/ Lancing Device/ Lancets/ Tissue or Cotton Ball with 75% Ethanol or Disinfection Wipes
3.1 Perform the Test
• Wash and clean your hands with disinfection wipes, and make sure your hands are dry before testing.
Chapter 3
How to Perform a Test
2
Insert lancet
rmly.
Remove and save the protective cap.
Recap lancing device and adjust penetration depth.
Take a strip and close the vial immediately.
Insert the strip.
1 2 3 4 5
Operation
Test Strip
Control Range
7142
1
C
H
O
L
7
1
4
2
Test Strip
7142
Operation
Storage & Transportation
Storage & Transportation
90%
10%
109-164 mg/dL
6.1-9.1 mmol/L
G1303190 2016/01
7142
C H O L 7 1 4 2
3
C
H
O
L
7
1
4
2
4
10
Note: Insufcient ll up of blood sample could lead to inaccurate or failed test result.
Do not rell the test strip.
Glucose/ Uric Acid/ Total Cholesterol
Fill up ok
Fill up FAILED
GLU UA CHOL
0.8
6 7 8 9
Make sure the code number is correct.
Pull the sliding barrel till it clicks.
Press the release button to sample.
Wipe off rst drop
of blood and start sampling.
Result will show after countdown. Then eject the used strip to biohazard container.
11 13 14 15
Pierce used lancet into protective cap.
Discard used lancet to biohazard container.
Recap lancing device and storage.
Note: Please refer to Lancing Device Instructions for
detail procedure.
Touch the blood sample with strip.
The meter will turn off automatically.
12
7
1
4
2
5
Note:
• Please nish the test within 3 minutes or the meter will turn off automatically.
• The meter will not turn on if you insert the wrong end or wrong side of strip.
• Marked open date on new open strip vial, do not use expired strip.
• Dropping, bumping or other violent impact will damage the meter or cause malfunction.
• Do not use the meter in an environment with possible magnetic, electromagnetic, and radioactive interferences.
3.2 Alternate Site Testing (AST)
You can test your glucose from ngertip, palm or forearm. Taking blood from palm or forearm
could reduce the pain, but the glucose level changes faster. These differences may cause wrong medical decision.
Note: Please consult healthcare professional before AST sampling.
Suitable timing to acquire blood sample from alternate sites:
• Routinely before meal.
• Prior or 2 hours after meal/ short-acting or rapid-acting insulin analogue/ exercise.
Warning:
● Please follow local regulations to discard used test strips and lancets.
● Used test strips, lancets and any other material that has been in contact
with blood should be treated as potential biohazards.
● If user has infectious disease, the used test materials could be sources of
infection.
● Lancets cannot be reused.
● Always use certied lancets to ensure safety.
● Keep the system away from children and pets.
6
DO NOT test from alternate sites:
• During or less than 2 hours after meal/ short-acting or rapid-acting insulin analogue/ exercise.
• When you think your glucose level is low or unaware of your low blood glucose condition.
• When you are examined for hypoglycemia or hyperglycemia.
• Your AST test result does not match your health condition.
• When you are ill, or you are operating machinery or driving a car.
Sampling from an Alternative Site:
1.Repeat the steps 1-7 in Chapter 3.1.(Replace the lacing device tip with adjustable AST tip.)
2.Hold the lancing device against sampling site, and press the release button.
Keep holding the lancing device against sampling site until sufcient sample formed.
3.Then repeat steps 10-15 in Chapter 3.1.
Note:
• Sampling from ngertip if your AST test result does not match your health condition.
• Repeat puncturing the same spot may cause soreness and calluses.
• Do not squeeze the site excessively. It may take longer for sufcient blood sample to form.
• Do not use smeared blood sample, please acquire new blood sample.
• If you continue failing in getting enough blood samples, please try to get lancets in lower
gauge or sample from ngertip instead.
Palm sampling
• No visible veins.
• Away from deep palm prints.
Forearm sampling
• Away from bones, visible veins and hair.
7
• Glucose - 360 results, capable in counting 7-, 14-, 21- and 28- days average
• Total Cholesterol - 50 results
• Uric Acid - 50 results The latest test result will replace the oldest when the records exceed maximum memory capacity. The memories start record from M1 to M360 or M1 to M50, include test results and control solution test results.
Note: The control results are not included in the average.
Directions for Checking Memories:
No Test Strip in the Meter→Press “M” Button (3 secs)→A Short “Beep”→Enter the Memory Mode→Full Display on Screen→Press “M” Button to Switch Mode (GLUC/ CHOL/ UA)→Press “S” Button to Conrm→Press “S” Button for Next Test Record/ Press “M” Button for Previous Record→Press “M” Button (3 secs) to Turn Off
• In GLU memory mode, it will display 7-, 14-, 21-, 28- days average rst.
• Once you enter one memory mode, you cannot switch. You need to turn off the meter (Press “M” button for 3 seconds) and enter the memory mode again.
Memory Records are shown as follows:
Chapter 4
Meter Memory Function
7-, 14-, 21-, 28­days average
Test Record
Blink Blink
Control Record
8
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
Perform a Control Test: Insert an Unused Strip→Make Sure the Code Number is Correct (Total Cholesterol/ Uric Acid)→Press “M” Button (3 secs)→Enter Control Solution Mode (Shown CL on Screen)→Shake the Control Solution Well→Discard First Three Drops→Put One Drop onto a Clean Surface→ Touch the Control Solution with Strip Sample Inlet→Test Result Display after Count Down→Compare the Result with the Range Listed on Test Strip Vial
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.
Chapter 5
Control Solution Test
98-164 mg/dL
5.4-9.1 mmol/L
EA13120901
Control Range LV1
2017-01
Test Strip
142
Operation
Storage &
Transportation
Storage &
Transportation
90%
10%
98-164 mg/dL
5.4-9.1 mmol/L
G1303190 2017/01
EA13120901 2017-01
9
Chapter 6
Transmission Function
The meter provides Bluetooth transmission function. It can transmit test results to connected device wirelessly via Bluetooth.
The screen show " " means Bluetooth transmission function is working.
• The meter with Bluetooth 4.0 can apply to IOS, Android 2.3.3 and above.
• The meter complies with IEC 60601-1-2 and the relevant EMC (electromagnetic compatibility) and RF (radio transmission) requirements regulated by US Federal Communications Commission. The purpose of these requirements is to ensure that meter does not affect or being affected by other devices during operation.
• The meter and the transmission function may be interfered while other device is operating nearby. Ex: mobile phone, wireless internet, etc.
• If the transmission has been interfered. Please keep the meter away from the source of interference or turn off the interfering device.
• Please make sure the meter and the receiving device are placed within a reasonable
distance (less than ve meters) during transmission via Bluetooth.
• Please do not share the meter with other people if you are using Bluetooth transmission function. The test results from other people will also transmit to the receiving device, and it
will inuence your test record.
• The Bluetooth transmission function may not work on certain types of mobile phone due to the compatibility of Android systems.
• Frequency range: 2402 – 2480MHz. RF power: 3 dBm @ room temperature Typical.
10
7.1 Storing Your Meter and Strip
Meter:
• Avoid bump or violent behavior.
• Do not use in extremely dry environment. It may cause static discharges.
• Do not use under electromagnetic radiation, ex: electrical equipment.
• Do not disassemble the meter for any reason.
• No modication of this equipment is allowed.
• Keep meter clean by wiping the exterior appearance with tissues or lint-free cloth.
• Do not expose the meter under lint, dust, sunlight, heat or humid environment.
Strip:
• Do not store in high humidity environment, or expose directly to sunlight.
• Do not freeze or refrigerate the meter and strips.
• Keep your hand dry and clean while handing the strips and performing the test.
• Do not bend, cut or fold the strips.
7.2 Cleaning and Caring for Your Meter
Please use soft cloth slightly damp with one of the following solution to clean meter surface.
• 75% alcohol
• Super Sani-Cloth disposable wipes
• Mild dishwashing liquid with water
• 10% household bleach solution and 90% water
Note:
• Do not allow any other wet cloth or liquid.
• Do not allow any liquid run in or around the test port and battery cover.
• Make sure the meter is completely dry before use.
• Protection impairment if used in a manner not specied by the manufacturer.
Chapter 7
Care and Maintenance
11
• Incorrect meter operating temperature.
Repeat the test after meter return to operating temperature. If the problem persists, please contact local distributor for service.
Chapter 8
Error Message and Trouble Shooting
Message Cause Solution
• Problem with code strip.
• Problem with test strip.
• Insert strip improperly.
Repeat the coding procedure (Chapter 2.3) and insert the strip again. If the problem persists, please contact local distributor for service.
• Low battery. Replace with new battery.
• Problem with code strip
or meter.
• Insert code strip improperly.
Repeat the coding procedure (Chapter 2.3). If the problem persists, please contact local distributor for service.
• Used strip.
• Damped strip.
Follow Chapter 3.1 and repeat the test with a new strip. If the problem persists, please contact local distributor for service.
12
• Test incomplete due to
removing the strip during measuring.
Follow Chapter 3.1 and repeat the test with a new strip. Do not remove the strip before the test is completed.
• Improper code strip. Repeat the coding procedure (Chapter 2.3).
If the problem persists, please contact local distributor for service.
• Test result is higher than
the range listed on Chapter 9.
Follow Chapter 3.1 and repeat the test with the same strip. If the problem persists, please contact local distributor for service.
• Test result is lower than
the range listed on Chapter 9.
Follow Chapter 3.1 and repeat the test with the same strip. If the problem persists, please contact local distributor for service.
• Sample volume not
enough.
Repeat the test with a new strip, and make sure the sample volume is enough. If the problem persists, please contact local distributor for service.
13
Chapter 9
Specication
Note:
• Please refer to the strip insert for accuracy, precision, limitation, and other important information.
Test Sample Fresh Capillary Whole Blood
Measuring Time GLU:5 seconds; CHOL: 26 seconds; UA: 15 seconds
Measuring Range GLU: 20-600 mg/dL (1.1-33.3 mmol/L);
CHOL: 100-400 mg/dL (2.59-10.35 mmol/L ); UA: 3-20 mg/dL (0.18-1.19 mmol/L )
Sample Volume GLU: 0.7 µL; CHOL: 0.8 μL; UA: 1 μL
Storage & Transportation Condition
4-30˚C (39-86˚F)
Operation Temperature 10-40˚C (50-104˚F)
Storage & Transportation Relative Humidity
10-90%
Memory 460 Test Results (GLU: 360 ; CHOL: 50 ; UA: 50)
Battery Type AAA battery * 2
Battery Life Approximately 1,000 tests
Dimensions 95*60*20mm
Weight 71g (with battery)
Altitude 10,000 feet (3048 m)
Expected Service Life 5 years
Transmission Function Bluetooth 4.0
Contents of the Kit (please check the meter outer box for exact detail)
• BeneCheck Supreme Multi-Monitoring Meter (BX-M00D) (with AAA battery * 2)
• User’s Manual
• Quick Guide
• Pouch
• Lancing Device
Optional
(not included in the standard kit package, please contact your local distributor for
ordering)
• BeneCheck Supreme Glucose Test Strips (BK-SG1)(with Insert)
• BeneCheck Supreme Total Cholesterol Test Strips (BK-SC2)(with code card & Insert)
• BeneCheck Supreme Uric Acid Test Strips (BK-SU1)(with code card & Insert)
• BeneCheck III Glucose Control Solution
• BeneCheck II Total Cholesterol Control Solution
• BeneCheck Uric Acid Control Solution
• Lancets (Please refer to package for manufacture information.)
Note:
After purchasing, if the contents are damaged, please contact authorized distributor immediately.
Electromagnetic Compatibility
This meter meets the electrostatic discharge immunity testing was basic standard IEC6100-4-2. In addition, the meter meets the electromagnetic emissions requirements as per EN61326. The purpose of these requirements is to ensure that meter does not affect or being affected by other devices during operation.
Warning:
● Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. The BeneCheck Multi-Monitoring Meter may provide a description or list of equipment with which the BeneCheck Multi-Monitoring
Meter has been tested in a stacked or adjacent conguration.
● Use of accessories, transducers and cables other than those specied or
provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
● Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the BeneCheck Multi-Monitoring Meter, including cables specied
by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
FCC Statement:
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may
cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
● Reorient or relocate the receiving antenna.
● Increase the separation between the equipment and receiver.
● Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
● Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.