General Electric SPIROSOFT SVC MNL_SM_2014847-009_C UniCel DxC Synchron Access Clinical System Integrated Workstations For In Vitro Diagnostic Use

SpiroSoft
Service Manual
2014847-009 Revision C
NOTE: The information in this manual only applies to the SpiroSoft. Due to continuing product innovation, specifications in this manual are subject to change without notice.
Caution:
During repairs/service interventions, observe the protective measures against damage due to ESD.
- assembly operations, extensions, readjustments, modifications, or repairs are carried out by GE Medical Systems Information Technologi or by persons authorized by GE Medical Systems Information Technologies,
- the electrical installation of the relevant room complies with the applicable national and local requirements, and
- the instrument is used in accordance with the instructions for use. This manual contains service information; operating instructions are provided in the operator’s manual of the instrument. This manual is in conformity with the instrument at printing date. All rights are reserved for instruments, circuits, techniques, and names appearing in the manual.
CardioSoft, CardioSys and CASE are trademarks owned by GE Medical Systems Information Technologies, a General Electric Company going to market as GE Healthcare. All other marks are owned by their respective owners.
© 2005 General Electric Company. All rights reserved.
T-2 SpiroSoft Revision C
2014847-009
SpiroSoft: Contents
Contents
1 Block Diagram ............................................................................................................................3
2 Functional Description ................................................................................................................3
3 Names of Signals ........................................................................................................................3
4 Pin Connections ..........................................................................................................................4
5 Instrument Options ......................................................................................................................4
6 Instrument Version ......................................................................................................................4
7 Pictures ........................................................................................................................................4
7.1 Serial Number ....................................................................................................................4
8 Firmware Loading Procedure .....................................................................................................5
8.1 Standard Update .................................................................................................................5
8.2 Update of a Device Without Functioning Firmware .........................................................6
9 Diagnosis Codes / Error Codes ...................................................................................................7
10 Adjustment Instructions ..............................................................................................................8
10.1 Interior View ......................................................................................................................8
11 Jumper Table ...............................................................................................................................8
12 Technical Specifications .............................................................................................................9
13 Technical Inspections ..................................................................................................................9
14 Spare Parts ............................................ ....................................................................................10
Revision C SpiroSoft Service Manual 1
2014847-009
SpiroSoft: Revision History
Revision History
This manual is subject to the GE Medical Systems Information Technologies (GEMS IT) change order service. The revision code, a letter that follows the document part number, changes with every update of the manual. The initial version of the man ual has the letter A.
Table 1. Revision History
Part No./Revision Code Date Comment
2014847-009 A 2003-07 Initial version 2014847-009 B 2003-12 ECO 075536 2014847-009 C 2005-05 ECO 080514
-
2 SpiroSoft Service Manual Revision C
2014847-009
1 Block Diagram
SpiroSoft: Block Diagram
2 Functional Description
The SpiroSoft System consists of a 16-bit microcontroller with an integrated Flash program memory. The peripheral components are a real-time clock backed up by GoldCap, which, when the device is turned off, retains the time
for a minimum of 5 days, an EEPROM which stores the data of the pressure sensor used and the SpiroSoft settings, as well as a Flash memory which stores the texts, the bitmaps and the database.
SpiroSoft has a reset generator which resets SpiroSoft, if the voltage drops below the 2.5-Volt threshold. The voltage is supplied from a 3.0 Volt low-dropout regulator; a 6.1-Volt Transil diode provides transient overvoltage and reverse polarity protection.
A piezoresistive pressure sensor sampled by a serial 16-bit A/D converter with integrated measuring bridge determines the mea­suring value by acquiring the pressure drop across the PT.
3 Names of Signals
none
Revision C SpiroSoft Service Manual 3
2014847-009
SpiroSoft: Pin Connections
4 Pin Connections
Both the communication and the voltage supply take place via the 6-pin connector of SpiroSoft. The pin assignment is as follows.
6 5 4 3 2 1
1 VCC supply (3.6 Volt to 5 Volt) 2 RXD V24 data line 3 TXD V24 data line 4 GND ground 5 IO/1 IO line TTL 6 IO/2 IO line TTL
5 Instrument Options
None
6 Instrument Version
The current hardware is version HW1.0. The current software is version FW1.0.1.
7Pictures
7.1 Serial Number
The serial number of the device is located under the lid of the battery compartment.
4 SpiroSoft Service Manual Revision C
2014847-009
SpiroSoft: Firmware Loading Procedure
8 Firmware Loading Procedure
8.1 Standard Update
The firmware updated is supplied in the form of a zipped file named “SW FIRMWARE SPIROSOFT FW102.zip” (2014848-
002). To begin with, unpack the zipped file. Connect SpiroSoft to the PC, using the connection cable. Then start the update procedure with the program SW FIRMWARE SPIROSOFT FW102.exe.
Select the COM port where SpiroSoft is connected and click the Start button to initiate the update. If the firmware version of the SpiroSoft is newer than the update version, the following window appears.
Pressing “OK” starts the update, pressing “Abbrechen” stopps the update. The update procedure is fully automatic, no further user interventions are required.
ATTENTION!!!
During the update, do not disconnect SpiroSoft from the PC and do not shut down the PC.
During the update, the following message appears on the display of SpiroSoft: “Download Microcontroller Program”. After the update, you will see the message “Download OK! Please remove the PS/2 conector for at least
3 Seconds”.
Revision C SpiroSoft Service Manual 5
2014847-009
SpiroSoft: Firmware Loading Procedure
Press “OK” and click “End” to terminate the update program. Remove the PS/2 connector for at least 3 seconds Then reconnect: SpiroSoft will start up with the new firmware version.
8.2 Update of a Device Without Functioning Firmware
If SpiroSoft has been disconnected from the PC while an update was in progress or the update has been interrupted for one rea­son or another, the SpiroSoft memory may be without a functioning firmware.
In this case, a bridge has to be closed inside SpiroSoft to allow SpiroSoft to start up in the firmware update mode. For this purpose, disconnect the PS/2 connector of SpiroSoft from the PC.
Then take off the lid from the battery compartment and open SpiroSoft by removing the four screws.
Next, remove the enclosure and connect the two contacts to the left of the Western connector (marking), using a metal object (tweezers, or similar object).
Then reconnect the PS/2 connector to the PC and start the update procedure by starting the program SW FIRMWARE SPIROSOFT FW102.exe
6 SpiroSoft Service Manual Revision C
2014847-009
SpiroSoft: Diagnosis Codes / Error Codes
Select the COM port where SpiroSoft is connected, select the “ROM Update” option and click the Start button to initiate the update. The update procedure is fully automatic, no further user interventions are required.
After the update, you will see the message “Download OK! Please remove the PS/2 conector for at least 3 Seconds”. Press “OK” and click “End” to terminate the update program. Remove the PS/2 connector for at least 3 seconds.
Then reconnect: SpiroSoft will start up with the new firmware version.
9 Diagnosis Codes / Error Codes
The following diagnosis/error codes can be displayed on the PC (CardioSoft).
Message Cause
Error in spirometer setup! Please check the settings!
Zeroing of the spirometer was not successful, probably you inhaled prematurely. Please repeat the test!
Data obtained cannot be analyzed. Please repeat the test!
The ambient conditions entered are incorrect; therefore the correction factor is out of range.
The system could not be zeroed before a measurement. This may be due to one of the following reasons
– The patient breathed into the sensor while the system was zeroed. – The tubing to the pressure transducer was squeezed between the
parts of the enclosure after a repair.
The test was aborted; therefore the system is unable to find data that can be analyzed.
In forced tests, this will be the case when no expiration with a volume greater than 100 ml has been found.
In slow spirometry tests, this will be the case when – less than 5 respiratory cycles have been recorded
– the max. VC is found in the first 5 respiratory cycles (before the
beep) – the measured VC is less than 0.5 l – the maneuver took less than 8 seconds.
Revision C SpiroSoft Service Manual 7
2014847-009
Other possible problems:
Problem Cause
SpiroSoft: Adjustment Instructions
No information displayed on SpiroSoft
No values are displayed during the measurement, or the displayed values are only inspiratory or only expiratory values.
SpiroSoft does not respond – If SpiroSoft does not respond any more, disconnect the PS/2 cable
– PC not turned on­– SpiroSoft not correctly connected to the PC – No functioning firmware in SpiroSoft (see 8.2)
– The PT is not properly locked in SpiroSoft – The silicone tubing to the pressure transducer was squeezed between
the parts of the enclosure after a repair – The seals of the tubing inside SpiroSoft are damaged.
of SpiroSoft from the PC for about 3 seconds. Afterwards, SpiroSoft
will restart the firmware.
10 Adjustment Instructions
SpiroSoft does not contain components that require maintenance or can be repaired. Defective devices should be returned to GEMS IT Repair Center for repair.
10.1 Interior View
11 Jumper Table
None
8 SpiroSoft Service Manual Revision C
2014847-009
SpiroSoft: Technical Specifications
12 Technical Specifications
Dimensions: 150 x 94 x 40 mm (h x w x d) Weight: 150 g Operating voltage: 3.0 V to 5.0 V Power consumption: 60 mA max. Measuring principle: JAEGER “High-Quality” pneumotach Interface: serial (RS 232) Measuring range:
PEF 0.1 to 16 l/s FEV1 and FVC 0.1 … 8 l
Resolution:
PEF 5 ml/s FEV1 and FVC 1 ml
Accuracy:
PEF +/-3 % / 0.4 l/s FEV1 and FVC +/-3 % / 0.05 l
Resistance: 2 kPa max. at 15 l/s Environment:
Operation
temperature +10 to +40 °C humidity 10 to 90 %, no condensation atmospheric pressure 600 to 1200 hPa
Transport and Storage
temperature -20 to +50 °C humidity 10 to 95 %, no condensation atmospheric pressure 500 to 1200 hPa
Medical device: CE0123
13 Technical Inspections
None
Revision C SpiroSoft Service Manual 9
2014847-009
SpiroSoft: Spare Parts
14 Spare Parts
The following spare parts and accessories can be ordered for SpiroSoft.
Item Description Item Number
SPARE SPIROSOFT BATTERY COVER 2014846-009 CABLE SPIROSOFT - PC 2014846-002 SPLY PNEUMATACH W MOUTHPIECE #10 2014846-003 SPLY NOSECLIP #25 2014846-008 SPLY MOUTHPIECE #25 2014846-006 SPLY PAD FOR NOSECLIP #100 2014846-007 MNL OPR SPIROSOFT 2014847-001 RCDR SPIROSOFT 2014846-001 RCDR SPIROSOFT EXCHANGE 2014846-010 PACKAGING BOX SPIROSOFT 2014846-005
10 SpiroSoft Service Manual Revision C
2014847-009
gemedical.com
0123
World Headquarters GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI 53223 USA
Tel. +1 414 355 5000
1 800 558 5120 (US only)
Fax +1 414 355 3790
European Representative GE Medical Systems Information Technologies GmbH Munzinger Straße 3 D-79111 Freiburg Germany Tel. +49 761 45 43 - 0 Fax +49 761 45 43 - 233
Asia Headquarters GE Medical Systems
Information Technologies Asia;
24th Floor Shanghai MAXDO Centre, 8 Xing Yi Road, Hong Qiao Development Zone Shanghai 200336, P.R. China Tel: +86-21-5257-4650 Fax: +86-21-5208-2008
GE (China)Co., Ltd.
2019524-001 Rev B
Loading...