NOTE: The information in this manual only applies to the SpiroSoft. Due to continuing product innovation, specifications in
this manual are subject to change without notice.
Caution:
During repairs/service interventions, observe the protective measures against damage due to ESD.
GE Medical Systems Information Technologies considers itself responsible for the effects on safety, reliability, and
performance of the equipment, only if:
- assembly operations, extensions, readjustments, modifications, or repairs are carried out by GE Medical Systems
Information Technologi or by persons authorized by GE Medical Systems Information Technologies,
- the electrical installation of the relevant room complies with the applicable national and local requirements, and
- the instrument is used in accordance with the instructions for use.
This manual contains service information; operating instructions are provided in the operator’s manual of the instrument.
This manual is in conformity with the instrument at printing date.
All rights are reserved for instruments, circuits, techniques, and names appearing in the manual.
CardioSoft, CardioSys and CASE are trademarks owned by GE Medical Systems Information Technologies, a General
Electric Company going to market as GE Healthcare. All other marks are owned by their respective owners.
14Spare Parts ............................................ ....................................................................................10
Revision CSpiroSoft Service Manual1
2014847-009
SpiroSoft: Revision History
Revision History
This manual is subject to the GE Medical Systems Information Technologies (GEMS IT) change order service. The revision
code, a letter that follows the document part number, changes with every update of the manual. The initial version of the man
ual has the letter A.
Table 1. Revision History
Part No./Revision CodeDateComment
2014847-009 A2003-07Initial version
2014847-009 B2003-12ECO 075536
2014847-009 C2005-05ECO 080514
-
2SpiroSoft Service ManualRevision C
2014847-009
1Block Diagram
SpiroSoft: Block Diagram
2Functional Description
The SpiroSoft System consists of a 16-bit microcontroller with an integrated Flash program memory.
The peripheral components are a real-time clock backed up by GoldCap, which, when the device is turned off, retains the time
for a minimum of 5 days, an EEPROM which stores the data of the pressure sensor used and the SpiroSoft settings, as well as
a Flash memory which stores the texts, the bitmaps and the database.
SpiroSoft has a reset generator which resets SpiroSoft, if the voltage drops below the 2.5-Volt threshold. The voltage is supplied
from a 3.0 Volt low-dropout regulator; a 6.1-Volt Transil diode provides transient overvoltage and reverse polarity protection.
A piezoresistive pressure sensor sampled by a serial 16-bit A/D converter with integrated measuring bridge determines the measuring value by acquiring the pressure drop across the PT.
3Names of Signals
none
Revision CSpiroSoft Service Manual3
2014847-009
SpiroSoft: Pin Connections
4Pin Connections
Both the communication and the voltage supply take place via the 6-pin connector of SpiroSoft.
The pin assignment is as follows.
6 5 4 3 2 1
1VCCsupply (3.6 Volt to 5 Volt)
2RXDV24 data line
3TXDV24 data line
4GNDground
5IO/1IO line TTL
6IO/2IO line TTL
5Instrument Options
None
6Instrument Version
The current hardware is version HW1.0.
The current software is version FW1.0.1.
7Pictures
7.1Serial Number
The serial number of the device is located under the lid of the battery compartment.
4SpiroSoft Service ManualRevision C
2014847-009
SpiroSoft: Firmware Loading Procedure
8Firmware Loading Procedure
8.1Standard Update
The firmware updated is supplied in the form of a zipped file named “SW FIRMWARE SPIROSOFT FW102.zip” (2014848-
002).
To begin with, unpack the zipped file.
Connect SpiroSoft to the PC, using the connection cable.
Then start the update procedure with the program SW FIRMWARE SPIROSOFT FW102.exe.
Select the COM port where SpiroSoft is connected and click the Start button to initiate the update.
If the firmware version of the SpiroSoft is newer than the update version, the following window appears.
Pressing “OK” starts the update, pressing “Abbrechen” stopps the update.
The update procedure is fully automatic, no further user interventions are required.
ATTENTION!!!
During the update, do not disconnect SpiroSoft
from the PC and do not shut down the PC.
During the update, the following message appears on the display of SpiroSoft: “Download Microcontroller Program”.
After the update, you will see the message “Download OK! Please remove the PS/2 conector for at least
3 Seconds”.
Revision CSpiroSoft Service Manual5
2014847-009
SpiroSoft: Firmware Loading Procedure
Press “OK” and click “End” to terminate the update program.
Remove the PS/2 connector for at least 3 seconds
Then reconnect: SpiroSoft will start up with the new firmware version.
8.2Update of a Device Without Functioning Firmware
If SpiroSoft has been disconnected from the PC while an update was in progress or the update has been interrupted for one reason or another, the SpiroSoft memory may be without a functioning firmware.
In this case, a bridge has to be closed inside SpiroSoft to allow SpiroSoft to start up in the firmware update mode.
For this purpose, disconnect the PS/2 connector of SpiroSoft from the PC.
Then take off the lid from the battery compartment and open SpiroSoft by removing the four screws.
Next, remove the enclosure and connect the two contacts to the left of the Western connector (marking), using a metal object
(tweezers, or similar object).
Then reconnect the PS/2 connector to the PC and start the update procedure by starting the program SW FIRMWARE
SPIROSOFT FW102.exe
6SpiroSoft Service ManualRevision C
2014847-009
SpiroSoft: Diagnosis Codes / Error Codes
Select the COM port where SpiroSoft is connected, select the “ROM Update” option and click the Start button to initiate the
update.
The update procedure is fully automatic, no further user interventions are required.
After the update, you will see the message “Download OK! Please remove the PS/2 conector for at least 3 Seconds”.
Press “OK” and click “End” to terminate the update program.
Remove the PS/2 connector for at least 3 seconds.
Then reconnect: SpiroSoft will start up with the new firmware version.
9Diagnosis Codes / Error Codes
The following diagnosis/error codes can be displayed on the PC (CardioSoft).
MessageCause
Error in spirometer setup!
Please check the settings!
Zeroing of the spirometer was not
successful, probably you inhaled
prematurely.
Please repeat the test!
Data obtained cannot be analyzed.
Please repeat the test!
The ambient conditions entered are incorrect; therefore the correction
factor is out of range.
The system could not be zeroed before a measurement.
This may be due to one of the following reasons
– The patient breathed into the sensor while the system was zeroed.
– The tubing to the pressure transducer was squeezed between the
parts of the enclosure after a repair.
The test was aborted; therefore the system is unable to find data that can
be analyzed.
In forced tests, this will be the case when no expiration with a volume
greater than 100 ml has been found.
In slow spirometry tests, this will be the case when
– less than 5 respiratory cycles have been recorded
– the max. VC is found in the first 5 respiratory cycles (before the
beep)
– the measured VC is less than 0.5 l
– the maneuver took less than 8 seconds.
Revision CSpiroSoft Service Manual7
2014847-009
Other possible problems:
ProblemCause
SpiroSoft: Adjustment Instructions
No information displayed on
SpiroSoft
No values are displayed during the
measurement, or the displayed values
are only inspiratory or only
expiratory values.
SpiroSoft does not respond– If SpiroSoft does not respond any more, disconnect the PS/2 cable
– PC not turned on– SpiroSoft not correctly connected to the PC
– No functioning firmware in SpiroSoft (see 8.2)
– The PT is not properly locked in SpiroSoft
– The silicone tubing to the pressure transducer was squeezed between
the parts of the enclosure after a repair
– The seals of the tubing inside SpiroSoft are damaged.
of SpiroSoft from the PC for about 3 seconds. Afterwards, SpiroSoft
will restart the firmware.
10Adjustment Instructions
SpiroSoft does not contain components that require maintenance or can be repaired.
Defective devices should be returned to GEMS IT Repair Center for repair.
10.1 Interior View
11Jumper Table
None
8SpiroSoft Service ManualRevision C
2014847-009
SpiroSoft: Technical Specifications
12Technical Specifications
Dimensions:150 x 94 x 40 mm (h x w x d)
Weight:150 g
Operating voltage:3.0 V to 5.0 V
Power consumption:60 mA max.
Measuring principle:JAEGER “High-Quality” pneumotach
Interface:serial (RS 232)
Measuring range:
PEF0.1 to 16 l/s
FEV1 and FVC0.1 … 8 l
Resolution:
PEF5 ml/s
FEV1 and FVC1 ml
Accuracy:
PEF+/-3 % / 0.4 l/s
FEV1 and FVC+/-3 % / 0.05 l
Resistance:2 kPa max. at 15 l/s
Environment:
Operation
temperature+10 to +40 °C
humidity 10 to 90 %, no condensation
atmospheric pressure600 to 1200 hPa
Transport and Storage
temperature-20 to +50 °C
humidity10 to 95 %, no condensation
atmospheric pressure500 to 1200 hPa
Medical device:CE0123
13Technical Inspections
None
Revision CSpiroSoft Service Manual9
2014847-009
SpiroSoft: Spare Parts
14Spare Parts
The following spare parts and accessories can be ordered for SpiroSoft.
World Headquarters
GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel. +1 414 355 5000
1 800 558 5120 (US only)
Fax +1 414 355 3790
European Representative
GE Medical Systems
Information Technologies GmbH
Munzinger Straße 3
D-79111 Freiburg
Germany
Tel. +49 761 45 43 - 0
Fax +49 761 45 43 - 233
Asia Headquarters
GE Medical Systems
Information Technologies Asia;
24th Floor Shanghai MAXDO Centre,
8 Xing Yi Road, Hong Qiao Development Zone
Shanghai 200336, P.R. China
Tel: +86-21-5257-4650
Fax: +86-21-5208-2008
GE (China)Co., Ltd.
2019524-001 Rev B
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.