General Electric ML_CL 2000_4000_7000 SYSTEM SVC MNL V. 5.2_SM_2007623-079_E University Physics Volume 2

Page 1
Mac-Lab®/CardioLab
2000/4000/7000
System Service Manual
2007623-079 Revision E
®
Page 2
900 SC, ACCUSKETCH, AccuVision, APEX, AQUA-KNOT, ARCHIVIST, Autoseq, BABY MAC, C Qwik Connect, CardioServ, CardioSmart, CardioSys, CardioWindow, CASE, CD TELEMETRY, CENTRA, CHART GUARD, CINE 35, CORO, COROLAN, COROMETRICS, Corometrics Sensor Tip, CRG PLUS, DASH, Digistore, Digital DATAQ, E for M, EAGLE, Event-Link, FMS 101B, FMS 111, HELLIGE, IMAGE STORE, INTELLIMOTION, IQA, LASER SXP, MAC, MAC-LAB, MACTRODE, MANAGED USE, MARQUETTE, MARQUETTE MAC, MARQUETTE MEDICAL SYSTEMS, MARQUETTE UNITY NETWORK, MARS, MAX, MEDITEL, MEI, MEI in the circle logo, MEMOPORT, MEMOPORT C, MINISTORE, MINNOWS, Monarch 8000, MULTI-LINK, MULTISCRIPTOR, MUSE, MUSE CV, Neo-Trak, NEUROSCRIPT, OnlineABG, OXYMONITOR, Pres-R-Cuff, PRESSURE-SCRIBE, QMI, QS, Quantitative Medicine, Quantitative Sentinel, RAC RAMS, RSVP, SAM, SEER, SILVERTRACE, SOLAR, SOLARVIEW, Spectra 400, Spectra-Overview, Spectra-Tel, ST GUARD, TRAM, TRAM-NET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, Trimline, UNION STATION, UNITY logo, UNITY NETWORK, Vari-X, Vari-X Cardiomatic, VariCath, VARIDEX, VAS, and Vision Care Filter are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
12SL, 15SL, Access, AccuSpeak, ADVANTAGE, BAM, BODYTRODE, Cardiomatic, CardioSpeak, CD TELEMETRY Cumulus, Event-Link Nimbus, HI-RES, ICMMS, IMAGE VAULT, IMPACT.wf, INTER-LEAD, IQA, LIFEWATCH, Managed Use, MARQUETTE PRISM, MARQUETTE MicroSmart, MMS, MRT, MUSE CardioWindow, NST PRO, NAUTILUS, O
®
-LAN, CENTRALSCOPE, Corolation, EDIC, EK-Pro, Event-Link Cirrus, Event-Link
®
RESPONDER, MENTOR,
SENSOR, Octanet, OMRS, PHi-
2
Res, Premium, Prism, QUIK CONNECT V, QUICK CONNECT, QT Guard, SMART-PAC, SMARTLOOK, Spiral Lok, S wee the art, UNITY, U niv ers al , Wa te r fall , and Walk mom are tra demar ks of GE Medi cal S yst ems Information Technologies.
© GE Medical Systems Information Technologies, 2003, 2004, 2005. All rights reserved.
T-2 Mac-Lab/CardioLab Revision E
2007623-079 15 April 2005
Page 3

Contents

1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Document Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Related Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Statement of Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Chapter Content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Electromagnetic Compatibility and Interference . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Danger, Warnings, and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-16
Equipment Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-17
Equipment Identifications Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-17
Required Tools and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
Power Input Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
2 Equipment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Acquisition Computer with Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Central Processing Unit (CPU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Additional Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Optional Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
CardioLink Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
MotherBoard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Component Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Revision E Mac-Lab/CardioLab i
2007623-079
Page 4
Integrated Electronics Box (IEB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
CardioLab 2000/4000/7000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Mac-Lab 2000/4000/7000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
ComboLab 2000/4000/7000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Integrated Electronics Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
CardioLink Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Changing the SA Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
On Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
On Acquisition and Review Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Recommended Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Operator’s Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Pre-service Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Environmental Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Daily Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
General Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Cleaning the Exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Cleaning the Monitor Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Cleaning EP Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Monthly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Cleaning the Print Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Annual Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Cleaning the Interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Inspect All Cables, Plugs, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Cleaning the RAU and TRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Domestic Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
AC Line Voltage Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
Ground Continuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Maintenance/Repair Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
ii Mac-Lab/CardioLab Revision E
2007623-079
Page 5
Maintenance/Repair Log Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
General Troubleshooting for the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
Functional Checkout Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
System Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
Front End Data Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
Troubleshooting the RAU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
RAU Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
TRAM modules in the RAU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Troubleshooting the Video Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Video Failure Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
Thermal Writer Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Troubleshooting the Thermal Writer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Thermal Writer Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Repairing the Thermal Writer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
RAU Failure Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
Server UPS Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
During Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18
After Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18
UPS Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-20
TRAM-net 4A LED Troubleshooting Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
CardioLink Utilities Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
6 Recovery Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
CardioLink Server (5.2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Ensure 5.2 backup tapes exist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Stop data replication on the server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Revision E Mac-Lab/CardioLab iii
2007623-079
Page 6
Stop data replication and disable networking at each acquisition system . . . . . . .6-5
Back up custom lists and reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Record system information at the server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Record Microsoft security patches installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Record LMT information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Check for advanced security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
Record job schedules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
Backup LMT Custom Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Replace the Server Hard Drive (if applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Re-Image Server (G3 Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Re-Image Server (G1 and G2 Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
Load AIT tape driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Restore data to the server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
Rename the Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
Configure Network Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-26
Restore the CardioLink Database from BackupTape . . . . . . . . . . . . . . . . . . . . .6-29
Enterprise Backups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-31
Install LMT server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-37
Install LMT client . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-37
Install LMT Backup/Restore utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-39
Re-Install Microsoft security patches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-39
Restore the GEMS Registry Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-39
Restore the clab database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-39
Restore LMT database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-40
Install Serial Number update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-40
Restore LMT Custom Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-41
Verify time and time zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-41
Register Acquisition systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-41
Enable for advanced security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-42
Run the Replication utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-42
Setup file paths at each Acquisition system . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-43
Push emergency cases to the server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-44
Re-link LMT tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-44
Restore ACC access /PedCath/Personal Files . . . . . . . . . . . . . . . . . . . . . . . . . .6-44
Schedule reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-45
Verify the LMT application installed correctly . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-45
Verify the modem connection and PC Anywhere . . . . . . . . . . . . . . . . . . . . . . . .6-46
Reconfigure hospital shares and mapped drives . . . . . . . . . . . . . . . . . . . . . . . . .6-46
Reconfigure SQL Jobs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-47
Run DB Maintenance Job . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-47
Setup Archive Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-48
Run a full tape backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-48
CardioLink Network Functional Checkout Procedure . . . . . . . . . . . . . . . . . . . . .6-49
7 Drawings and Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Diagrams/Drawings/Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Chassis Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Chassis Back View CD-ROM Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
iv Mac-Lab/CardioLab Revision E
2007623-079
Page 7
Optical Drive - 3 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
RS422 PCI Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
Optical Drive - 5 1/4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
Back Panel Showing Slots for CardioLab/Mac-Lab Computers . . . . . . . . . . . . . . .7-9
Slots for CardioLab/Mac-Lab Computer with RS422 or TRAM-net Card . . . . . . .7-10
Back Panel Showing Slots for CardioLab/Mac-Lab Computers . . . . . . . . . . . . .7-11
Slots for CardioLab/Mac-Lab Computer with PSI Card . . . . . . . . . . . . . . . . . . . .7-12
Thermal Writer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
Remote Acquisition Unit (RAU) with TRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
RAU Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
RAU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-16
RAU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-17
Connections for the NT CardioLab/Mac-Lab Computer . . . . . . . . . . . . . . . . . . . .7-18
Rack-Mount Server Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-20
Rack-Mount Server Back View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-21
Rack-Mount Server UPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-22
TRAMnet Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-24
TRAM-RAC 4A Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-24
Ordering Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-28
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-28
Appendix A – Technical Specifications . . . . . . . . . . . . . .A-1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Special Instructions for Using Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-6
CardioLink Server G1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
CardioLink Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
Rack for CardioLink Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-14
CardioLink Server G2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-15
System Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-15
LaserJet 4200 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-18
HP LaserJet 4200 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-18
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
Minimum recommended circuit capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
Environmental specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
Acoustic emissions (per ISO 9296) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
Safety certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
HP 2200 D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21
HP LaserJet 2200 D Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21
Revision E Mac-Lab/CardioLab v
2007623-079
Page 8
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-21
Appendix B – CardioLink Server Rename . . . . . . . . . . . .B-1
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Special Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-3
Equipment needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Customer Notification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-4
New Server Install . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
Re-name Installed Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
System Functional Checkout Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-14
CardioLink Server Rename Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-15
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
vi Mac-Lab/CardioLab Revision E
2007623-079
Page 9

1 Introduction

Revision E Mac-Lab/CardioLab 1-1
2007623-079
Page 10
For your notes
1-2 Mac-Lab/CardioLab Revision E
2007623-079
Page 11
Introduction: Document Information

Document Information

Revision History

Each page of the document has the document part number and revision letter at the bottom of the page. The revision letter identifies the document’s update level.
The revision history of this document is summarized in the table below.
A 31 March 2003 Initial release of manual. B 28 July 2003 Added ML370 G3 server information. C 01 September 2004 Added Disaster Recovery Server Repartitioning
D 13 October 2004 Updated Disaster Recovery Section (Chapter 6)
Table 1. Revision History PN 2007623-079
(+180 GB)

Manual Purpose

Intended Audience

Related Manuals

E 15 May 2005 Updated Per ECR01905-003; Removed
incorrect leakage current info in Chapter 4.
This manual contains the instructions necessary to operate and service the Mac-Lab/CardioLab system safely in accordance with its function and intended use. These instructions i nclude but are not limited to:
Describing fundamental controls and indicators.
Operating procedures.
Connecting and disconnecting detachable parts and accessories.
Troubleshooting an d service procedures.
This manual is int ended for t he perso n who mainta ins and t roublesho ots this equipment.
Part Number Document
2007623-042 Mac-Lab Operator’s Manual 2007623-043 CardioLab Operator’s Manual 2007623-083 Centricity CardioLink Operator’s Manual 2007623-081 Centricity CardioLink Installation/Upgrade Instructions
Revision E Mac-Lab/CardioLab 1-3
2007623-079
Page 12
Introduction: Document Information
Part Number Document
2007623-080 Mac-Lab/CardioLab System Installation Manual 2007623-059 Mac-Lab/CardioLab Advanced Security Guide 2007623-078 Mac-Lab/CardioLab/Centricity CardioLink Pre-Installation Manual 2007623-082 Mac-Lab/CardioLab System Software Upgrade Instructions 2007623-169 Mac-Lab/CardioLab Lab Management Tools (LMT) Installation/Service Manual
1-4 Mac-Lab/CardioLab Revision E
2007623-079
Page 13
Introduction: Document Information

Statement of Indications for Use

Mac-Lab System
The Mac-Lab system is intended for use under the direct supervision of a licensed healthcare practitioner to monitor, calculate, or record cardiovascular data from patients as they undergo cardiac catheterization. Cardiovascular data may be manually entered or acquired via interfaced GE Medical Systems Information Technologies TRAM modules, MUSE cardiovascular system, and other interfaced information syst ems. Data includes: ECG waveforms, heart rate, pulse oximetry (SpO2), respiration rate, valve gradients and areas, cardiac output, hemodynamic measurements, invasive and non-invasive blood pressure, procedural information, and optional intracardiac electroca r di og r a m (I E C G )
This is applicable to pediatric and adult patients requiring cardiac/ circulatory system catheterization. This is intended for use in catheterization and related cardiovascular specialty laboratories.
CardioLab System
The intended use of the CardioLab system is to acquire, filter, digitize, amplify, display, and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals, and pressure recordings. Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be d erived from the s ignal data, dis played and recorded. The system allows the user to monitor the acq uisition of data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. Additionally, the system may acquire, amplify, display, and record data received from other medical devices typically used during these procedures, such as imaging devices and RF generators.
.
The CardioLab EP system does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze physiological data or other data acquired during an EP procedure. It does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities.
Revision E Mac-Lab/CardioLab 1-5
2007623-079
Page 14

Chapter Content

Introduction: Document Information
This manual is organized into chapters and appendices. 1 Introduction – This chapter describes the field service manual and
chapter contents. Provides general information on safety, service requirements, equipment symbols, and serial number identi fication.
2 Equipment Overview – This chapter describes the equipment and its technical characteristics, connector locations, and preparation for use.
3 Installation – This chapter contains the system components for the Mac-Lab/CardioLab system.
4 Maintenance – This chapter contains a preventive maintenance schedule, illustrated instructions for opening the unit, and cleaning guidelines. Instructions for performing electrical safety (leakage) tests are also included.
5 Troubleshooting – This chap ter provides overall and specific troubleshooting help, an d directions for performing functional tests.
6 CardioLink Disaster Recovery – This chapter provides the procedures needed to perform a server disaster recoveries in a CardioLink network environment.
7 Drawings and Parts Lists – This chapter contains upper-level parts lists, photos, and illustrations for reference in repairing the unit at the assembly level.
Appendix A - Technical Specifications – This appendix lists the technical descriptions and specifications for the Mac-Lab/CardioLab system.
Appendix B - CardioLink Server Rename – This appendix explains the procedure used to change the server name in a CardioLink network environment.
Index – The Index lists the terms used in the document.
1-6 Mac-Lab/CardioLab Revision E
2007623-079
Page 15
Introduction: Safety Information

Safety Information

Follow all of the instructions and warnings marked included in this manual, save these instructions for later use. However, these instructions in no way supersede current medical practices regarding patient care and safety.

Electromagnetic Compatibility and Interference

This catheterization system meets the requirements of the European Medical Device Directive, and by reference, the requirements for electromagnetic compatibility specified by EN 60601-1-2. This equipment is not likely to interfere with other equipment, except with certain types of rate adaptive pacemakers that rely on bioelectric impedance measurement (BIM) technology. Also, interference with the operation of this equipment is not likely to result from proximity to general equipment.
This equipment meets or exceeds Class A immunity levels. However, avoid the use of strong int entional radio transmitters in the immediate vicinity, including cellular telephones. The use of strong unintentional radiators, such as electrosurgical equipment, may cause some level of interference. GE Medical Systems Information Technologies offers optional ECG pati ent cables with electrosurgical noise filtering, which reduces the level of interference on ECG traces.
If using electrosurgical or RF ablation equipment in conjunction with this equipment, select an ESU filtered cable. For more information about management of the equip ment in the clinical environment (from an electromagnetic compatibility perspective) consult the Association for the Advancement of Medical Instrumentation’s technical information report
Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical Engineers.

Responsibility of the Manuf acturer

GE Medical Systems Information Technologies is responsible for the effects of safety, reliability, and performance only if:
Installation operations, extensions, readjustments, modifications, or repairs are performed by p erso ns autho rize d by GE M edical S ystems Information Technologies.
The equipment is used in accordance with the instructions provided.
The user attaches only accessory equipment recommended for interaction with the amplifier.
Revision E Mac-Lab/CardioLab 1-7
2007623-079
Page 16

General

Introduction: Safety Information
This system is intended for use under the direct supervision of a licensed health care practitioner.
The Mac-Lab/CardioLab syst em main ta ins le aka ge curre nts be low medical limits with an isolation transformer which is incorporated into the design of the integrated electronics box (IEB). The system power is supplied by a dedicated power supply on an isolated circuit for proper behavior of the system.
To ensure patient safety, use only parts and access ories manufactured or recommended by GE Medical Systems Information Technologies. Parts and accessories used must meet the requiremetns for the applicable IEC 601 series safety standards, and the system configuration must meet the requirements of the IEC 601-1-1 medical electrical systems standard. Contact GE Medical Systems Information Technologies before connecting any device to this sys tem that is not recommended for use in this manual.
If the equipment is installed in the USA and uses 240V rather than 120V, the source must be a center-tapped, 240V single-phase circuit.
To ensure patient safety, use only parts and access ories manufactured or recommended by GE Medical Systems Information Technologies.
This equipment should only be connected as specified by GE Medical Systems Information Technologies. Installation of the equipment should only be performed by qualified GE Medical System s Information Technologies service representatives. Connectio n of the system to the equipment is not recommended in this manual.
Contact GE Medical Systems Information Technologies for information before connecting any device to this equipment that is not recommended in this manual. Only cab les and acces sory equipmen t purchased form GE Medical Systems Information Technologies should be used in conjunction with the system.
It is important that all of the instructions in this manual be followed. However, these instructions in no way supersede current medical practices regarding patient care and safety.
Refer servicing of the equipment to GE Medical Systems Information Technologies authorized service personnel. Any attempt to repair equipment under warranty will void that warranty.
Failure on the part of responsible individuals or institutions to implement a satisfactory equipment maintenance schedule may lead to undue equipment failure and possible health hazards.
1-8 Mac-Lab/CardioLab Revision E
2007623-079
Page 17
Introduction: Safety Information
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice includes:
Use of the accessory in the PATIENT VICINITY.
Evidence that the safety certification of the ACCESSORY is performed in accordance with the appropriate IEC 601-1 and IEC 601-1 harmonized national standard.
Revision E Mac-Lab/CardioLab 1-9
2007623-079
Page 18

Equipment Symbols

Equipment Symbols Descriptions
Introduction: Safety Information
Table 2. Equipment Symbols
Underwrite rs Laboratories, Inc.
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical and other specified hazards, only in accordance with UL 2601-1, CAN/CSA C22.2 No.
601.1, IEC 60601-1, and, if required, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-1-1.
CE Marking
Power Ratings Chart:
V~ = Voltage I = Current Fuse =
Equipotentiality
A nonessential electrical connection which may be used to connect the chassis to other equipment to achieve the same potential.
Type CF Equipment that is Defibrillator Proof
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and internal application to the patient excluding direct cardiac application. The "paddles" outside the box indicate that the part is defibrillator proof. (Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a higher degree of protection against electric shock than that provided by Type B applied parts.
Type CF Equipment
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and internal application to the patient excluding direct cardiac application. (Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a higher degree of protection against electric shock than that provided by Type B applied parts.
Attention
Pay attention to the documents delivered with the equipment.
Input Symbol
1-10 Mac-Lab/CardioLab Revision E
2007623-079
Page 19

Conventions

Safety Messages

Introduction: Conventions
'$1*(5
Safety messages which indicate an imminently hazardous situation which, if not avoided, WILL result in death or serious injury.
:$51,1*
Safety messages which indicate a potentially hazardous situation which, if not avoided, COULD result in death or serious injury.
&$87,21
Safety messages which indicate a potentially hazardous situation which, if not avoided may result in minor or moderate injury, loss of data or system failure.

Definitions

127(
Messages which provide additional user information.
Items shown in Black text are keys on the keyboard, text to be entered, or hardware items such as buttons or switches on the equipment.
Items shown in Italicized text are software terms which identify menu items, buttons, or options in various windows.
T o perf orm an operation which appears with a plus (+)sign between the names of two keys, you press and hold the first key while pressing the second key once. This is called a keystroke combination.
For example, Press Ctrl+Esc means to press and hold down the Ctrl key while pressing the Esc key.
When instructions are given for typing a precise text string with one or more spaces, the point where the spacebar must be pressed is indicated as: <Space>. The purpose of the < > brackets is to ensure you press the spacebar when required.
Enter means to press the Enter or Return key on the keyboard. Do not type enter.
Revision E Mac-Lab/CardioLab 1-11
2007623-079
Page 20
Introduction: Danger, Warnings, and Cautions

Danger, Warnings, and Caut ions

'$1*(5
EXPLOSION HAZARD Do not use in the presence of flammable anesthetics or
other flamma b le gases.
:$51,1*6
ACCESSORIES Only connect UL 544 or UL 2601-1 listed or IEC certified
medical equipment. ACCIDENTAL SPILLS If fluid of any kind should leak into the system,
discontinue use of the eq uipment and conta ct GE Medical Systems Information Technologies immediately.
DO NOT allow hydraulic fluid from any device to contac t any conductive surface. Hydraulic fluid is electrically conductive.
ACCURACY The system may report erroneo us heart rate va lues when
pacemaker spikes or arrhythmias are present. Reported values should be verified by a qualified physician.
CONDUCTION HAZARD Keep the conductive parts of lead electrodes and
associated parts away from other conducting parts, including earth.
DEFIBRILLATOR PRECAUTIONS DO NOT contact the amplifier or patient during
defibrillation. Standby defibrillation equipment is required in the event
that the patient needs to be cardioverted or defibrillated. Disconnect leadwires from patient before defibrillating.
Otherwise, serious injury, death, or damage to the equipment could result.
IEC All equipment not complying with IEC 601-1 should be
placed outside the patient environment.
1-12 Mac-Lab/CardioLab Revision E
2007623-079
Page 21
Introduction: Danger, Warnings, and Cautions
:$51,1*6
INTERPRETATION HAZARD A qualified physician must overread all computer-
generated measurements and tracings. Computerized interpretation is only significant when used in conjunction with clinical findings.
Signal quality should be checked only by a licensed health care practitioner.
All computer assisted measurement functions must be verified by a licensed health care practitioner.
LEAKAGE CURRENT TEST The system should be tested yearly for adequate
grounding on patient leakage current. Keep leakage current within acceptable limits when
connecting auxiliary equipment to the system. When more than two electrical devices are
interconnected, the summation of leakage currents must be checked yearly.
MAINTENANCE/SERVICE High voltages exist in the unit. Use insulated tools.
Remove jewelry and use only one hand when possible. Always disconnect power from the equipment before
attempting service. PINCH POINT HAZARD Keep hands, hair, jewelry and loose cloth ing away from
moving parts. Otherwise, serious injury could result. POWER SUPPLY Total system earth leakage current must not exceed 500
microamperes. Voltages between the three power outlet connections
must be accurate BEFORE measuring power outlet ground-to-neutral loop resistance.
Electrostatic discharge (ESD) can damage electronic components. Be sure you are properly grounded when using or servicing the system.
To reduce the risk of electric shock or damage to the equipment, do not disable the power cord grounding feature. This equipment is designed for conne ction to a grounded (earthed) power outlet.
Revision E Mac-Lab/CardioLab 1-13
2007623-079
Page 22
Introduction: Danger, Warnings, and Cautions
:$51,1*6
The supply line must be fused and capable of delivering 16A.
Route the AC power cable away from moving parts. Connect only to a properly earth grounded outlet. REPLACEMENTS Replace only with the same type and rating of fuse. SHOCK HAZAR D Electrical shock ha zard between chassis ground and
isolated (floating) ground when power is applied. Unplug the unit from the power source before proceeding.
To reduce the risk of electric shock or damage to your equipment, do not disable the power cord grounding feature. This equipment is designed for conne ction to a grounded (earthed) power outlet. The grounding plug is an important safety feature.
SUPERVISED USE This device is intended for use under the direct
supervision of a licensed health care practitioner.
&$87,216
Federal (USA) law restricts this device to sale by or on the order of a physician.
ACCURACY DO NOT display any channels which are based on
unused catheter input modules. Otherwise, erroneous values may occur.
ECG signal quality should only be checked by qualified medical personnel.
CABLING Route optical cables t horough conduit in the ceiling or
floor to avoid damage to the cables of cable connectors. DEVICES DO NOT reuse single use devices .
1-14 Mac-Lab/CardioLab Revision E
2007623-079
Page 23
Introduction: Danger, Warnings, and Cautions
&$87,216
EMC No known electromagnetic or other interference between
the system exists. However, precautions should be taken to avoid the use of cellular telephones, pagers, or ot her transmitters.
OPTICAL DISK Exit the software application properly before turning off
the system. Failure to do so may damage the optical disk. Remove the optical disk before moving the system.
Failure to do so may result in damage to the drive or the disk.
TRANSDUCER Remove air that is trapped in the transducer or
associated tubing by flushing the system according to Cath lab procedures.
REMOTE MONITOR Always power the system through separate isolation
transformers which are independently connected to a wall outlet. You can power the laser printer through the IEB/Workstation power supply. Ot herwise, the printer may draw enough current away from the system and cause it to reboot.
SOFTWARE DO NOT add unauthorized software to the Mac-Lab/
CardioLab system. Doing so may make the Mac-Lab/ CardioLab software unstable.
Revision E Mac-Lab/CardioLab 1-15
2007623-079
Page 24

Service Information

Service Requirements

Follow the service requirements listed below:
Introduction: Service Information
Equipment servicing should be performed by GE Medical Systems Information Technologies authorized service personnel onl y.
Any unauthorized attempt to repair equipment under warr anty voids that warranty.
It is the user’s responsibility to report the need for service to GE Medical Systems Information Technologies or one of their authorized agents.
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the system is always functional when required.
1-16 Mac-Lab/CardioLab Revision E
2007623-079
Page 25

Equipment Identification

Every GE Medical Systems Information Technologies device has a unique serial number for identification. The serial number appears on the back of the CardioLab II Plus Amplifier.
Introduction: Service Information
B
CardioLab Computer
A0XH0001FP
GE Medical Systems Information Technologies, Inc.
A
DEFGHI

Equipment Identifications Table

Table 3. Equipment Identifications
Item Name Description
A Manufacturer GE Medical Systems Information Technologies B Name of device CardioLab computer C Serial number Unique identifier D Month manufactured A = January, B = February, C = March, D = April, E = May, F = June, G = July,
H = August, J = September, K = October, L = November, M = December E Year manufactured 9 = 1999, 0= 2000, 1= 2001, (and so on) F Product code Two-character product descriptor XH= CardioLab computer
C
PK1
G Product sequence number Four-character manufacturing number (of total units manufactured) H Plant designator F = Cardiology G = Monitoring
I Device characteristics One or two characters that further describe the unit, for example: P = prototype not
conforming to marketing specification; R = refurbished equipment; S = special product
documented under Specials part numbers; U = upgraded unit
Revision E Mac-Lab/CardioLab 1-17
2007623-079
Page 26
Introduction: Required Tools and Supplies

Required Tools and Supplies

The following is a list of tools needed to service the Mac-Lab/CardioLab system:
Adjustable wrench with a 1-inch capacity
Phillips screwdriver
Needle nose pliers
Pliers
Socket wrench (3/16 inch to 1 inch size bits)
Phillips head screwdriver (small and large sizes)
Flat head screwdriver (small and large sizes)
Cutting pliers
Power outlet ground tester
Grounding strap
Signal generator with ECG, pressure, cardiac output, and temperature outputs.
Cables to connect simulator to ECG inputs of amplifier and TRAM
Cables to co nnec t the simulator to the catheter i nput modu le s (2 m m shielded)
Adapter cable to connect the simulator to the amplifier and TRAM pressure inputs
Cables to connect the simulator to the TRAM, cardiac output, and temperature inputs

Power Input Requirements

If the equipment is installed in the USA and uses 240V rather than 120V, the source must be a center-tapped, 240V single-phase circuit (120V, 20A; 240V, 10A).
1-18 Mac-Lab/CardioLab Revision E
2007623-079
Page 27

2 Equipment Overview

Revision E Mac-Lab/CardioLab 2-1
2007623-079
Page 28
For your notes
2-2 Mac-Lab/CardioLab Revision E
2007623-079
Page 29
Equipment Overview: Acquisition Computer with Workstation

Acquisition Computer with Workstation

Central Processing Unit (CPU)

The Mac-Lab/CardioLab products are diagnostic systems used for the assessment of the cardiac electrophysiologic (CardioLab) and hemodynamic (Mac-Lab) condition of patients.
The computer hardware consists of a PC motherboard, expansion cards for added functionality, and a SCSI drive system.
See Chapter 7, “Drawings and Parts List” for diagrams of the computer.
PK1
Revision E Mac-Lab/CardioLab 2-3
2007623-079
Page 30
Equipment Overview: Acquisition Computer with Workstation
General Description Standard Equipment
Barcode Scanner 2003417-001 Laser Printer
4200 2200
Monitors:
18" Flat Screen 19" CRT 20" Flat Screen 21" CRT
Table 1. Standard Equipment
Description Part Number
2015205-001 2007078-001
2013547-002 2012980-001 2006144-001 2010311-001
Keyboard: Mac-Lab (English)
CardioLab (English)
ComboLab (English) Scroll Mouse 2007023-001 TRAM Modules (Mac-Lab 7000/Combination systems only)
450 SL
400 SL
Remote Acquisition Unit (RAU) Mac-Lab 7000/Combination
systems only) (400SL or 450 SL) CardioLab II Plus Amplifier (Mac-Lab/CardioLab/Combination systems only)
32 Channel 64 Channel 96 Channel
128 Channel Acquisition Computer 2011646-00X TRAM-Net RAC 4A (Mac-Lab/Combo System only) 900031-006
2003809-001 2003809-002 2003809-003
412654-001 411161-001
415540-001
2003232-001 2003232-002 2003232-003 2003232-004
2-4 Mac-Lab/CardioLab Revision E
2007623-079
Page 31
Equipment Overview: Acquisition Computer with Workstation

Additional Systems

Nurses Station/Pre-Post Op Station/Review Station 2011646-006

Optional Equipment

Strip Chart Recorder (Thermal Writer) 2002480-001 Analog Output Box 2006168-001 Video Printer 2004142-001
Table 2. CardioLink Clients
Description
Table 3. Optional Equipment
Description Part Number

CardioLink Server

G2:
High-End Server
Mid-Level Server
Local Server G3:
High-End Server
Mid-Level Server
Low-End Server Network Switch 10/100 Base-TX 2001084-001 Rack-Mounted UPS 2006397-001 Rack 2006483-001

Software

Table 4. Server
Description Part Number
2009350-001 2009350-002 2009350-003
2016992-001 2016992-002 2016992-003
Operating System
The computer uses the Windows NT 4.0 Operating System.
Application
The Mac-Lab/CardioLab system is a single, custom application which has an EP and a CathLab mode. The ComboLab system includes both Mac-Lab and CardioLab.
Revision E Mac-Lab/CardioLab 2-5
2007623-079
Page 32
Accessory Software
Equipment Overview: Acquisition Computer with Workstation
In addition to the Mac-Lab/CardioLab system, the computer has the following accessory software installed:
Microsoft Office (for generating procedure reports)
PC Anywhere 32 (for remote support via modem)
Microsoft Database Engine (MSDE) for database communication
2-6 Mac-Lab/CardioLab Revision E
2007623-079
Page 33

MotherBoard

The motherboard is a server and has the following features:

Component Boards

The following expansion cards may be added to the computer for additional functionality.
Video PCI Expansion Cards
The video subsystem utilizes a PCI video card. This card supports the high resolutions used for the application (up to 1600 by 1280 per monitor).
The computer can be configured with one or two video cards, depending on the configuration purchased. The use of the extra video card allows the desktop displa y to be shown across two monitors instead of one. The board in the slot labeled Video 1 drives the real-time signal display of the system, as well as non-real-time display elements. The optional board in the slot labeled Video 2 displays only non-real-time d isplay elements.
Equipment Overview: MotherBoard
Pentium III processor support Integrated 10/100 Network card Integrated SCSI controller
Revision E Mac-Lab/CardioLab 2-7
2007623-079
Page 34
RS422 PCI Expansion Card
The computer can be configured with an optional RS422 expansion card. This card allows the computer to communicate with the TRAM RAU module. This card has two independent RS422 DB9 connectors, allowing it to communicate with two TRAM RAUs simultaneously.
Equipment Overview: MotherBoard
127(
The DB9 connectors on the board do not connect directly to the TRAM, but connect instead to the IEB. The IEB routes the dual RS422 connections to a single DB15 connector, which connect to the TRAM RAU cable.
Imaging PCI Expansion Card - Optional
The computer can be configured with an optional Imaging expansion card. This card allows the computer to acquire and display analog video input, such as images from an X-ray system.
2-8 Mac-Lab/CardioLab Revision E
2007623-079
Page 35
COMMS ISA Expansion Card
The computer is configured with a COMMS expansi on card. The purpose of this card is to:
Equipment Overview: MotherBoard
Communicate with the CardioLab II/II Plus amplifier (via 2 ST connectors on the bracket).
Output of analog signals for use by other equipment (output via the DB 44 connector on the bracket).
Input of analog signals from other equipment (input via the DB 44 connector on the bracket).
Provide additional serial ports (COM 5,6,7,8,9,10). Input/Output via the internal 40 pin header for COM 5,6,7,8,9 and internal 10 pin header for COM 10.
Output a video switching signal to toggle the video display of the Switched Video output on the system VID/MUX (video distribution/ amplifier board in the IEB) (output via the DB 44 connector on the bracket).
SCSI PCI Expansion Card - Optional
The computer can be configured with an optional SCSI expansion card. This card allows the computer to review old patient studies from the previous DOS CardioLab EP version of the CardioLab product.
SCSI Drive Subsystem
The computer is configured with an Ultrawide II SCSI interface. A SCSI interface is used to increase performance of the data read/write that is achieved with the SCSI.
The SCSI drives used in the system are:
Hard drive used for the operating system, application software, and accessory software.
CD ROM drive used for loading software as needed.
Optical drive used for recording system studies to portable optical media.
Revision E Mac-Lab/CardioLab 2-9
2007623-079
Page 36
USB Port (part of IEB Box)
Optical Drive 2.6GB
Equipment Overview: MotherBoard
The computer is configured with two USB ports:
The USB port communicates with or prints to the optional Thermal Writer. Only one port is needed for the writer.
The second port is an active port which can be used in the event of a problem with the first port.
The optical drive used for recording system st udies to portable optical media.
Type IV (High-end option):
5 1/4 inch Optical Drive)
Optical Drive 320MB
The optical drive used for recording system st udies to portable optical media.
Type III (low-end option):
3 1/2 inch Optical Drive
TRAM-net to Ethernet conversion board
Converts signals from the TRAM-net RAC to Ethernet. Allows Mac-Lab communications with the TRAM-net RAC.
2-10 Mac-Lab/CardioLab Revision E
2007623-079
Page 37
Equipment Overview: Integrated Electronics Box (IEB)

Integrated Electro n ic s Box (IEB)

Introduction

The IEB serves as the power conditioner/distribution device for the Mac­Lab/CardioLab/ComboLab workstations. The power supplied by the IEB supports all sp ecifi ed eq ui pment tha t i s needed f or sys te m o peratio n. The IEB is designed to provide power to the following equipment:
NT Computer
3- 20" LCD Monitors
Uninterruptible Power Supply (UPS) (Internal)
Laser printer
Video Multiplexer (VID/MUX) amplifier (Internal)
Modem (Internal)
Thermal Writer
Two TRAM modules
There are three differen t versions of the IEB. They are:
IEB 100V, 50/60 Hz
IEB 120V, 50/60 Hz
IEB 240V, 50/60 Hz
See Chapter 7, “Drawings and Parts List” for part numbers.
Isolation Transformer (part of IEB Box)
The isolation transformer is used to provide isolation as required by IEC601-1. The transformer is a 2.0kVA medical grade high isolation transformer. It has the appropriate primary and secondary taps to achieve 1:1 ratings for 100V, 120V, and 240V.
Revision E Mac-Lab/CardioLab 2-11
2007623-079
Page 38
Equipment Overview: Integrated Electronics Box (IEB)
Uninterruptible Power Supply (UPS)
There are three different versions of the UPS, one for each rated voltage version of the IEB. The UPS is only required to power the computer (no monitors) to prevent momentary loss of power. The UPS is powered directly from the unswitched power on the secondary of the transformer. This prevents the UPS from detecting a power fault condition when the power is switched off to other components.The computer is turned off using the power switch on ei ther the UPS or the computer.
The UPS is mounted on the upper internal shelf and is fastened in place using the UPS Hold Bracket. The front panel of the UPS is visible through the opening in the front p anel of th e IEB. Th is is nee ded for user access to the UPS power switch.
Modem
The universal modem is used for remote dialing to the system f or service.
The IEB is suppli e d wi t h an i n te rn al mod e m that is co nn ec t ed t o th e computer.
There are two versions of modems. One for use in the 120V IEB and one that is used in the 100V and 220V versions. Each version of modem is supplied with a power supply.
2-12 Mac-Lab/CardioLab Revision E
2007623-079
Page 39
Equipment Overview: Integrated Electronics Box (IEB)
Modem Power Supply (part of IEB Box)
The IEB is supp li e d wi t h an i n te rn al mod e m t hat is co nn ec ted t o th e computer.
There are two versions of modems. One for use in the 120V IEB and one that is used in the 100V and 220V versions. Each version of modem is supplied with a power supply.
The modem is attached to the lower internal shelf a nd the power is connected to the appropriate AC o utlet on the same shelf. The modem is secured with Velcro. The serial interface to the computer is connected through the DB Connect Circuit Board. The phone line in and out is connected through the DB Connect Panel.
Video Multiplexer (VID/MUX) for IEB
The VID/MUX is a video board that is capable of the following functions:
Accepts three video inputs
Two buffered outputs for each input (one for a local monitor and one for remote)
One buffered output that is software selectable to view input 1 or 2
VID/MUX Power Supply (part of IEB Box)
The VID/MUX board requires ±5V @ 1.25A each. The power supply for the VID/MUX board is mounted in the lower power compartment of the IEB. Switched AC power is delivered by the power supply.
Revision E Mac-Lab/CardioLab 2-13
2007623-079
Page 40
Equipment Overview: Integrated Electronics Box (IEB)
DB Connect Circuit Board
The DB Connect Panel has the following I/O connections.
Analog IN and OUT from computer on DB44 The DB44 is the new connector on the COMMS board. It combines
the functions of Analog In and Analog OUT.
Analog OUT to breakout box on DB37 This connection routes the Analog OUT from the DB44 to the DB37.
This allows for interface to the current analog output breakout cable.
Analog OUT Channel 1-2 on BNC These connections are individual channel outputs from the DB44
connector.
Analog IN CH 1-4 on BNC These connections are analog inputs that connect to the DB44.
Digital IN on 1/4 inch phone jack. These connections are digital inputs tha t connect to the DB44.
Modem Line IN on pass through RJ12. Connects modem to the phone line.
Modem Phone OUT on pass through RJ12. Port allows for the connection of the externa l phone to the phone line.
Serial IN for the modem. Serial connection for the internal modem.
RS422 from the computer on 2 DB9. There are two RS422 ports from the computer for interface to 2
TRAM modules
RS422 to TRAM on DB15 (DB9 shell). This connector has integrated the two RS422 ports and power for the
TRAM. This connects to an RAU.
Power from TRAM power supply for TRAM (solder side conne ctor).
2-14 Mac-Lab/CardioLab Revision E
2007623-079
Page 41
Equipment Overview: Integrated Electronics Box (IEB)
Thermal Overcurrent Circuit Breaker
The device is mounted on the lower half of the rea r panel. It has two functions:
Main Power Switch This switch controls power to all internal components and all
external components that are connected to the system.
Overcurrent Protection This is an overcurrent condition to trip the breaker. Once the
overcurrent condition is removed, the breaker can be turned on.
ON/Standby Switch
This switch controls power to all components connected to the 4 outlets on the rear of the unit. It does not switch power to the sin gle unswit ched outlet on the rear of the unit . The main purpos e for this switch i s to allow for power control of all the components except the UPS that delivers power to the computer.
Rack-Mounted UPS for CardioLink Server
The UPS features a maximum of 2700 Watts of true power. This new rack-space saving design allows more support for critical rack mountable equipment.
LED and switch menbrane integrated into the front panel with four button controls (three buttons for UPS power control and one button under the front bezel for configuration).
Revision E Mac-Lab/CardioLab 2-15
2007623-079
Page 42
For your notes
Equipment Overview: Integrated Electronics Box (IEB)
2-16 Mac-Lab/CardioLab Revision E
2007623-079
Page 43

3 Installation

Revision E Mac-Lab/CardioLab 3-1
2007623-079
Page 44
For your notes
3-2 Mac-Lab/CardioLab Revision E
2007623-079
Page 45

Installation

CardioLab 2000/4000/7000

Components
The components may vary depend ing on the options in use. These variations are covered in instructions that are provided with each opti on.
Catheterization Procedure Room
The components that are found in the catheterization procedure room are:
Patient leads, ECG cable
Catheter input
Analog output cable
Analog input cable
Remote monitors
Video cables
Video power cables
Amplifier
Stim input cables
Fiber optic cable
Amplifier power cable
RF generator
Stimulator
Integrated Electronics Box (IEB) power cable
Radio Frequency (RF) cable
IEB
Computer
Workstation
Printer
Barcode scanner
Network cable (with Network option)
Mouse
Keyboard
Image input cable (with Image option)
Analog phone line
Installation: Installation
Revision E Mac-Lab/CardioLab 3-3
2007623-079
Page 46
Reviewing the Assembly
Installation: Installation
Before beginning to service the system, rev iew the existing assembly to ensure that it is correctly installed and connected. This should be done prior to turning on the system.
Ensure that the power is off.
Ensure that the components are conveniently positioned in the room and have adequate room for access and ventilation.
Have the monitors been mounted properly?
Use a line analyzer to check the wall outlet s for power and ground. Is the wall outlet in both the control room an d procedure room a dupl ex 20 Amp circuit on the hospital emergency power?
Is the conduit sized correctly and correctly installed?
Make sure that all connections are complete and secure.
If a line conditioner is to be used, make sure that it is connected now.
Route all cables so that they will not be walked on, tripped over, or rolled on.
Ensure that Network cables are correctly installed (Network option).
Ensure that the floppy disk drive is empty.
Turn on the power to the system at the IEB.
Check the screen adjustments on each monitor. Is the picture centered, shaped properly, sized correctly, and the right color? Is there any color distortion indicating a need for degaussing the monitor? Moving the monitor may cause color distortion.
3-4 Mac-Lab/CardioLab Revision E
2007623-079
Page 47
Installation: Installation
PK1
Revision E Mac-Lab/CardioLab 3-5
2007623-079
Page 48

Mac-Lab 2000/4000/7000

Connecting System Components
For complete information on planning for and assembling the system, mounting brackets, and options, see the following manual:
Mac-Lab System EX RAU Mounting Kits Installation Manual
This chapter is limited to an overview of the assembly/disassembly as it pertains to the repair of a unit already in the field.
Components
The components may vary depend ing on the options in use. These variations are covered in instructions that are provided with each opti on.
Catheterization Procedure Room
The components that are found in the catheterization procedure room are:
Remote Acquisition Unit (RAU) or TRAM-net RAC 4a
TRAM
RAU mounting bracket
Patient leads
BP cuff SPO2 probe
Analog output cable
Analog input cable
RAU data acquisition/power cable (from RA unit DB-15 to buffer plate DB-15 connection)
Remote monitors
Video cable
Video power cable
Installation: Installation
Control Room
The components that are found in the control room are:
IEB power cable
Video power cables to IEB
Video cables
IEB
Computer
Workstation
Printer
Thermal Writer
Image input cable (with Image option)
Barcode scanner
Network cable (with Network option)
Local system monitors
Mouse
3-6 Mac-Lab/CardioLab Revision E
2007623-079
Page 49
Reviewing the Assembly
Installation: Installation
Keyboard
Analog phone line
Before beginning to service the system, rev iew the existing assembly to ensure that it is correctly installed and connected. This should be done prior to turning on the system.
Ensure that the power is off.
Ensure that the components are conveniently positioned in the room and have adequate room for access and ventilation.
Is the RA unit pos i ti one d so th at the T R AM ope ni ng i s no t fa ci ng th e ceiling or floor?
Have the monitors been mounted properly?
Use a line analyzer to check the wall outlet s for power and ground. Is the wall outlet in both the control room an d procedure room a dupl ex 20 Amp circuit on the hospital emergency power?
Is the conduit sized correctly and correctly installed?
Make sure that all connections are complete and secure.
If a line conditioner is to be used, make sure that it is connected now.
Route all cables so that they will not be walked on, tripped over, or rolled on.
Ensure that Network cables are correctly installed (Network option).
Ensure that the floppy disk drive is empty.
Turn on the power to the system at the IEB.
Check the screen adjustments on each monitor. Is the picture centered, shaped properly, sized correctly, and the right color? Is there any color distortion indicating a need for degaussing the monitor? Moving the monitor may cause color distortion.
Revision E Mac-Lab/CardioLab 3-7
2007623-079
Page 50
Installation: Installation
PK2
3-8 Mac-Lab/CardioLab Revision E
2007623-079
Page 51

ComboLab 2000/4000/7000

Connecting System Components
For complete information on planning for and assembling the system, mounting brackets, and options, see the following manual:
Mac-Lab System EX RAU Mounting Kits Installation Manual
This chapter is limited to an overview of the assembly/disassembly as it pertains to the repair of a unit already in the field.
Components
The components may vary depend ing on the options in use. These variations are covered in instructions that are provided with each opti on.
Catheterization Procedure Room
The components that are found in the catheterization procedure room are:
RAU or TRAM-net RAC 4A
TRAM
RAU mounting bracket
Patient leads
BP cuff SPO2 probe
ECG cable
Catheter input modules,
Analog output cable, and
Analog input cable.
Simulator and cables
RAU data acquisition/power cable (from RA unit DB-15 to buffer Plate DB-15 connection)
Remote monitors
Video cables
Video power cables
Amplifier
IEB
Workstation
RF generator
Fiber optic cable
Amplifier power cable
Computer
Thermal Writer
IEB power cable
Printer
Barcode scanner
Network cable (with Network option)
Analog phone line
Installation: Installation
Revision E Mac-Lab/CardioLab 3-9
2007623-079
Page 52
Reviewing the Assembly
Installation: Installation
Mouse
Keyboard
Image input cable (with Image option)
RF cable
Before beginning to service the system, rev iew the existing assembly to ensure that it is correctly installed and connected. This should be done prior to turning on the system.
Ensure that the power is off.
Ensure that the components are conveniently positioned in the room and have adequate room for access and ventilation.
Ensure that the RA unit is positioned so that the TRAM opening is not facing the ceiling or floor.
Ensure that the monitors are mounted properly.
Use a line analyzer to check the wall outlet s for power and ground. Is the wall outlet in both the control room an d procedure room a dupl ex 20 Amp circuit on the hospital emergency power?
Is the conduit sized correctly and correctly installed?
Make sure that all connections are complete and secure.
Ensure that the line conditioner being used is connected.
Route all cables so that they will not be walked on, tripped over, or rolled on.
Ensure that LAN cables are correctly installed (LAN option).
Ensure that the floppy disk drive is empty.
Turn on the power to the system at the IEB.
Check the screen adjustments on each monitor. Is the picture centered, shaped properly, sized correctly, and the right color? Is there any color distortion indicating a need for degaussing the monitor? Moving the monitor may cause color distortion.
3-10 Mac-Lab/CardioLab Revision E
2007623-079
Page 53
Installation: Installation
PK3
Revision E Mac-Lab/CardioLab 3-11
2007623-079
Page 54

Integrated Electronics Box

Installation: Installation
PK4
3-12 Mac-Lab/CardioLab Revision E
2007623-079
Page 55

CardioLink Diagram

Installation: Installation
Revision E Mac-Lab/CardioLab 3-13
2007623-079
Page 56
Installation: Changing the SA Password

Changing the SA Password

If desired the system administrator can sa password.

On Server

127(
Record the new password, as it will be needed for any future software upgrades.
To change the sa password on the server, follow these steps:
1. Login to the system as an admini strator
127(
See system administrator for login information.
2. Select Start > Programs > Enterprise Manager
3. Select Server > Security > Logins
4. Double-click on SA and change the password.
5. Click OK.
6. At the Confirm Password window, re-enter the new password and click OK.
7. Click OK.
8. Close Enterprise Manager.

On Acquisition and Review Systems

127(
Record the new password, as it will be needed for any future software upgrades.
To change the sa password on an acquisition or review system, follow these steps:
1. Login to the system as an administrator
127(
See system administrator for login information.
2. Select Start > Microsoft Access
3. At the Microsoft Access opening window, click Cancel.
4. Select File > New.
5. Double-click the Project (Existing Database) icon.
6. Enter a file name and click Create.
7. In the Connection tab, from the Select or Enter a server name dropdown box, select the local system name as the server name.
8. Select the Use a specific user radio button.
3-14 Mac-Lab/CardioLab Revision E
2007623-079
Page 57
Installation: Changing the SA Password
9. Type the current username in the Username textbox.
127(
The default user for version 5.2 i s sa; for ve rs ion 5.1D is clabsys.
10. Uncheck the Blank Password and Allow saving password checkboxes.
11. Type the current password in the Password textbox.
127(
The default password for version 5.2 is gemssqladmin; for version 5.1D is clab.admin
12. From the Select the database on the server dropdown box, choose clab.
13. Click the Test Connection button.
14. At the Test Connection Succeeded window, click OK.
15. Once in the database, Select Tools > Security > Database Security.
16. Single-click sa.
17. Click the Edit button.
18. Type the new password in the Password textbox.
19. Click OK.
20. Type the new password in the Confirm New Password textbox and click OK.
21. At the SQL Server Login Properties window, click OK.
22. Close Microsoft Access.
Revision E Mac-Lab/CardioLab 3-15
2007623-079
Page 58
For your notes
Installation: Changing the SA Password
3-16 Mac-Lab/CardioLab Revision E
2007623-079
Page 59

4 Maintenance

Revision E Mac-Lab/CardioLab 4-1
2007623-079
Page 60
For your notes
4-2 Mac-Lab/CardioLab Revision E
2007623-079
Page 61
Maintenance: Introduction

Introduction

Recommended Maintenance

This chapter provides preventive maintenance instructions for the system and its equipment and supplies. These mai ntenance guidelines are followed to prevent unnecess ary equipment and supply failure and possible health hazards.
This chapter contains instructions for the following recommended maintenance:
Inspecting and cleaning the unit
AC line voltage tests to verify that the power outlet is properly wired
Leakage tests to verify that the equipment does not pose a microshock hazard
Ground continuity test to verify that all exposed metal is properly grounded

Operator’s Maintenance Schedule

This chapter contains maintenance schedules for users. The CathLab supervisor and assigned operators should perform the mai n tenance described.

Pre-service Procedures

Before you open the unit for maintenance or service, perform these required preparations:
Turn OFF the system at the console and at the IEB, and unplug the IEB before servicing the equipment.
Check the power cord, but do not remove it if you will be doing anything that requires a ground. The power cord should be removed for cleaning purposes.
Verify that the line, neutral, and ground wires are all connected firmly to the plug and that they are not shorted together.

Environmental Protection

Although the system operates in many different environments, some precautions are observed.
Take care to av oid ext remes of temper ature, moisture, and humidit y. The system components are not waterproof.
Protect the various system components from spills or other debris. Eating, drinking and smoking should not be permitted near the system during a study or during maintenance procedures.
Revision E Mac-Lab/CardioLab 4-3
2007623-079
Page 62
Maintenance: Inspection and Cleaning

Inspection and Cleaning

Precautions

Turn off the Integrated Electronics Box (IEB) and remove the power cord from the wall outlet before inspecting or cleaning.
Wear latex surgical gloves and follow h ospital regulations when cleaning equipment that has been exposed to bodily fluids.
Do not immerse any part of the equipment in water.
Do not use organic solvents, strong ammonia-based solutions, or abrasive cleaning agents which may damage equipment surfaces.
Do not use a cleaning solution or solvent on the diskettes or disk drives. Doing so can cause loss of data or damage to the data and/or drive.
Do not use metal articles to clean the disk drives.
Do not allow any magnetic material near the system unit.

Daily Visual Inspection

Operators should perform a general overall inspection daily.
Turn off the console and the IEB, and remove the power cord from the wall outlet before making an inspection or cleaning the unit.
Check the case and display screens for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Inspect all plugs, cables, and connectors for bent prongs or pins.
Verify that all cords, socketed components, and connectors are securely seated.
Inspect keys and controls for proper operation.
Toggle keys should not stick in one position. Diskette drives should be fully functional.
Perform a visual exterior and interior inspection during preventive maintenance checks.
Inspect interior for debris, oil, and moisture (liquid or condensation). Determine the source of the contamination and correct both the source and resulting problems.
Inspect for missing hardware. Be sure that all components are present and intact. Socketed components should be firmly seated in their respective sockets. Check that all board components are properly attached, especially soldere d compo nent s added as part of a repair or modification.
Check for burn damage from overheated components. The familiar smell of burned components indicates that something has failed, even when there is no discoloration or disformation.
Check that all printed circuit boards are seated properly if mounted in a card cage or connected to any type of edge connector. Make sure that all ground straps are connected.
4-4 Mac-Lab/CardioLab Revision E
2007623-079
Page 63
Maintenance: Inspection and Cleaning
Check that all power cords and interface cables are firmly connected, with no sign of excessive wear or excessive tension.
Check that screws and other hardware connectors are in place and properly installe d. Check the se curity of t he Remote A cquisi tion Unit (RAU) mounting bracket. The knobs holding the RAU to the bracket should be tight. However, be careful not to strip the threads.
Verify that the remote video monitors are properly secured to the mounting plates. Verify that the mounting plates are properly secured.
Verify that the local monitors are properly placed or secured.
Revision E Mac-Lab/CardioLab 4-5
2007623-079
Page 64
Maintenance: General Maintenance

General Maintenance

This section contains preventive maintenance instructions for the system and its equipment and supplies. These maintenance guidelines should be followed by the operator to prevent unnecessary eq uipment failure or possible health hazards. Bef ore starting any maintenance, turn the system power off.

Cleaning the Exterior

Clean the exterior of the syst em on a reg ular basis with a cl ean, soft cl oth and a solution consisting of mild detergent diluted in water. However, avoid coming in contact with open vents, plugs and connectors of any kind, the keyboard, mouse, and th e thermal printer. Thorou ghly wipe the exterior of the unit with a lint-free cloth dampened with one of the following approved solutions, then dry it with a clean, soft cloth or paper towel. The cleaning agents include:
Ammonia (diluted to 10% strength)
Cidex
Mild soap
Sodium hypochlorite bleach (diluted to 10% strength)
To avoid damage to the equipment, follow these rules:
Wring the excess cleaning solution from the cloth.
If exterior surface is swabbed with bleach or 10% ammonia solution, follow by swabbing with water.
Always immediately wipe off all of the cleaning solution with a dry cloth or paper towel after cleaning.
Never pour water or any cleaning solution on the equipment or permit fluids to run into the Thermal Writer, Console, AC Distribution Unit, RAU, monitors, connectors, or ventilation openings in the equipment.
Never use these cleaning agents:
Abrasive cleaners or solvents of any kind
Alcohol-based cleaning agents
Wax containing a cleaning substance
Acetone
Betadine
Failure to follow these guidelines may melt, distort, or dull the finish of the case, blur lettering on the labels, or cause equipment failures.
4-6 Mac-Lab/CardioLab Revision E
2007623-079
Page 65
Maintenance: General Maintenance

Cleaning the Monitor Screens

Clean the monitor’s screen using a gentle commercial glass cleaner. Apply a small amount of cleaner to a clean, soft, anti-static, lint-free cloth, then use on screen. This method prevents excess clea ner from running inside of the monitor.
Use a gentle commercial cleaner.
Follow directions for use on product.
Do not spray directly on the glass.
Use a clean, non-abrasive, soft, anti-static, lint-free cloth.
Wipe glass carefully and allow to dry.

Cleaning EP Cables

GE’s electrophysiology cables are typically used outside the sterile zone and therefore are not routinely sterilized. The cables may, however, be cleaned using one of the following solutions:
Ammonia (diluted to 10% strength)
Cidex
Sodium hypochlorite bleach (diluted to 10% strength)
Mild soap
Isopropyl alcohol
Revision E Mac-Lab/CardioLab 4-7
2007623-079
Page 66
Maintenance: Monthly Maintenance

Monthly Maintenance

Perform the following preventive maintenance on a monthly basis. Before starting any maintenance, turn off the system.

Cleaning the Print Head

Clean the thermal print head using isopropyl alcohol and a clean, soft cloth. Non­abrasive alcohol wipes may also be used. To access the print head, open the writer door and wipe the print head where it makes contact with the capstan roller. Clean the roller when necessary with a small amount of alcohol. Excessive use of alcohol over time may cause the roller to crack.
PK1
4-8 Mac-Lab/CardioLab Revision E
2007623-079
Page 67

Annual Maintenance

Perform the following preventive maintenance on an annual basis. Before starting any maintenance, turn off the Console and AC Distribution Unit, and remove the power cord from the wall outlet.

Cleaning the Interior

Generally, it is no t necessary t o clean the interior o f the syst em; however , if dust is an environmental problem, use a commercially available, electronically-safe, dust remover (compressed air, difluoroethane, or other). Follow the manufacturer’s directions for use. Note caution s about flammability, the possibility of ignition, and the possibility of causing frostbite. Be sure to clear the no zzle before use, hold t he can uprigh t, and do not allow liquid propellant to contact skin or electronics. Be careful not to drive dust into connections or disk drives.
Whenever an exterior spill occurs, you should completely turn off the IEB and remove power from the wall outlet. Clean up the spill, and check to be sure the spill has not penetrated the CPU, Thermal Writer, RAU, or IEB. If it has, carefully blot and dry the affected area. Do not reconnect power or turn the system on until you are sure no moisture remains.
Maintenance: Annual Maintenance

Inspect All Cables, Plugs, and Connectors

Inspect all cords for fraying or other damage. Inspect all plugs and connectors for bent prongs or pins. Verify that all cords and connectors are securely seated. Repair or replacement must be performed by qualified service personnel.

Cleaning the RAU and TRAM

Check the RAU and TRAM modules for buildup of moisture, fluids, or dust. Clean the units in the same way as the exterior of the console, being careful not to allow any moisture into the unit.
Revision E Mac-Lab/CardioLab 4-9
2007623-079
Page 68
Maintenance: Preventative Maintenance

Preventative Maintenance

Perform the following general preventative maintenance.
Make a backup copy of the hard drive onto a blank optical disk. This way, if any files ever become corrupt, a backup copy will exist to replace them.
Test the integrity of the disk using scandisk or a similar utility found at the local computer store. DO NOT ATTEMPT TO REPAIR ANY PROBLEMS FOUND. Only diagnose the errors. If errors are found, contact technical support for assistance.
Scan the hard disk for viruses using the recommended validated anti-virus software.
Remove the fiber optic cable from the amplifier and computer and clean the contacts with rubbing alcohol and a clean, lint-free cloth. When finished, verify that you have connected the cables back correctly by acquiring a test patient in the software.
Remove any dust or lint from inside the computer and IEB using electronics-safe compressed air. Do not use compressed air to clean the inside of the optical drive.
Ensure that all cables connected to the monitors, IEB and computer are securely fastened.
The standard life of the UPS ranges from 3 to 6 years. Test the UPS by turning the main power switch on, letting the computer boot completely and then unplugging the IEB from the wall. The computer should maintain power and the UPS will beep to indica te that power was lost to the system. If computer mainta ins power, the UPS is functioning normally.
4-10 Mac-Lab/CardioLab Revision E
2007623-079
Page 69
Maintenance: Domestic Electrical Safety Tests

Domestic Electrical Safety Tests

AC Line Voltage Test

This test verifies that the domestic wall outlet supplying power to the equipment is properly wired. For international wiring tests, refer to the internal standards agencies of that particular country.
120 VAC, 50/60 Hz
Use a digital voltmeter to check the voltages of the 120-volt AC wall outlet (dedicated circuit recommended). If the measurements are significantly out of range, have a qualified electrician repair the outlet. The voltage measurements should be as follow s:
120 VAC (± 10%) (108 to 132 VAC) between the line contact and neutral and between the line contact and ground.
Less than 3 VAC between neutral and ground.
PK2
240 VAC, 50/60 Hz
Use a digital voltmeter, set to measure at least 300 VAC, to check the voltages of the AC wall outlet (dedicated circuit recommended). If the measurements are signifi can tly out of ra nge, ha ve a q uali fi ed elec trici an repair the outlet. Th e voltage measurements should be as follows:
240 VAC (± 10%) (216 to 264 VAC) between either “hot” contact and ground.
216 to 264 VAC between the two “hot” contacts.
Revision E Mac-Lab/CardioLab 4-11
2007623-079
Page 70
Leakage Current Tests
HOT
GROU ND
2
1
1
HOT
Maintenance: Domestic Electrical Safety Tests
PK3
The leakage tests are safety tests to ensure that the equipment poses no electrical microshock hazards. The specific leakage current tests and the frequency at which they should be performed should be determined by your facility.
Special Test Considerations
If the unit under test fails th e le akage t ests, do no t allow the custome r to use the equipment. Call Tech Support for assistance. (See the “How to Reach Us” page in the front of the manual.)
GE Medical Systems recommends that you perform these tests:
Before applying power for the first time
Every year as part of routine maintenance
Whenever internal assemblies are serviced
You need a leakage tester to perform the le akage tests.
127(
The accuracy of the leakage tests depends on a properly wired wall outlet. Do not proceed until you verify the integrity of the power source.
When patient sink current is measured, it may be necessary to manipulate the patient cable to avoid excessive stray capacitance to ground, which may result in artificially high measurements.
4-12 Mac-Lab/CardioLab Revision E
2007623-079
Page 71

Ground Continuity

Special Considerations
Maintenance: Domestic Electrical Safety Tests
This test verifies that there is continuity (less than 100 mΩ resistance) between all the exposed metal surfaces on the console that have the potential to become energized, and the ground prong on the Mains AC power cord.
Measure the resistance between the equipotential lug on the back of the console and the ground pin o n the AC inlet of the AC Distri bution Unit. Repeat between the shell of the I/O port on the Thermal Writer and the ground pin on the AC inlet of the AC Distribution Uni t.
If the measurements are significantly out of range, check for breaks in the power cord or in the internal connections within the unit.
To perform this test accurately, it is necessary to use a special ground impedance measuring device that can accurately measure low resis tance s while prov idin g a sig nific ant a mount o f cur rent (10 to 30 A). This test cannot be performed accurately with a typical volt ohm meter.
Because of the resistance i n the power cables between the console or writer and the AC Distribution Unit, it is essential that special care is taken to obtain precise measurements. Resistance in test leads must be subtracted from an overall reading to accurately isolate the resistance in the equipment under test.
For a cursory field service check, a volt ohm meter may be used to verify continuity and that resistance is reasonable low (less than 0.3 to 0.4 ohms, including test leads).

Maintenance/Repair Log

The Maintenance/Repair Log is included in this manual to aid in keeping a record of the work performed on the system.
127(
Unless you have an Equipment Maintenance Contract, GE Medical Systems Information Technologies does not in any manner assume the responsibility for performing the recommended maintenance procedures. The sole responsibility rests wit the individual or institution using the equipment. GE Medical Systems Information Technologies service personnel may, at their discretion, follow the procedures provided in this manual as a guide during visits to the equipment site.
Revision E Mac-Lab/CardioLab 4-13
2007623-079
Page 72
Maintenance: Domestic Electrical Safety Tests

Maintenance/Repair Log Table

Unit Serial Number: Institution Name:
Date Maintenance/Repair Technician
4-14 Mac-Lab/CardioLab Revision E
2007623-079
Page 73
Maintenance: Domestic Electrical Safety Tests
Unit Serial Number: Institution Name:
Date Maintenance/Repair Technician
Revision E Mac-Lab/CardioLab 4-15
2007623-079
Page 74
Maintenance: Domestic Electrical Safety Tests
4-16 Mac-Lab/CardioLab Revision E
2007623-079
Page 75

5 Troubleshooting

Revision E Mac-Lab/CardioLab 5-1
2007623-079
Page 76
For your notes
Troubleshooting:
5-2 Mac-Lab/CardioLab Revision E
2007623-079
Page 77

Introduction

Initial Inspection

Troubleshooting: Introduction
This chapter includes the following sections:
General Troubleshooting for the system
Troubleshooting the Remote Acquisition Unit (RAU)
Troubleshooting the Video Processing
Troubleshooting the Thermal Writer
The following are the most common causes for system failure:
Ensure that the IEB, CPU, RAU, monitors, keyboard, and optional LAN are properly connected. If not, power down the system, connect the cables properly, and turn on the IEB.
Ensure that the wall outlet is live. With a line analyzer, check the outlet for both power and ground.
Ensure that a disk is not in the floppy disk drive. No disk should be present.
Ensure that the IEB is turned on.
Ensure that the monitors are turned on.
Ensure that the CPU is turned on.
Ensure that nothing (other than a patient lead on the TRAM) is connected or disconnected while the system is under power. If so, turn off the IEB, reconnect the cables, and then turn the IEB on again.

Visual Inspection

A thorough visual inspection of the equipment can save time. Small things — disconnected cables, foreign debris on circuit boards, missing hardware, loose components — can frequently cause symptoms and equipment failures that may appear to be unrelated and which are difficult to track.
Revision E Mac-Lab/CardioLab 5-3
2007623-079
Page 78
Troubleshooting: General Troubleshooting for the System

General Troubleshooting for the System

Problems

No Boot Up
If the computer does not boot:
Verify that the power cable from the workstation is plugged into the wall.
Verify that the power cable is secure in the back of the IEB.
Check the power cable from the IEB to the computer to verify that it is secure on both ends.
With the IEB power switch turned on, try pressing the power switch on the CPU. This ca n b e acc essed by op eni ng t he right fro nt pa nel o n the workstation.
Replace the power cables.
System Does Not Boot
Disk in Disk Drive
Check to see if you left a disk in the floppy disk drive. If you did, remove the disk. Turn the system off and wait 10 seconds before turning the system on again.
Check the IEB and then the other hardware.
System Locked Up
The system locked up and there is a patient on the table.
If you are still able to monitor the patient, continue, but get assistance from Technical Support. See the How to Reach Us insert at the front of the manual.
If you are not able to monitor the patient, recycle power on the system. Have hospital ready to put patient on standby equipment. Get assistance from Technical Support.
If there is no patient on the table when the system locks up:
Turn the system of f, wa i t te n s econd s, then tu rn t he s ys te m on ag ai n.
If the system still does not work correctly, get assistance from technical support.
Multiple Monitors Change Colors
If more than one monitor changed colors, there is a possible VID/MUX problem. Get assistance from Technical Support. See the How to Reach Us insert at the front of the manual.
5-4 Mac-Lab/CardioLab Revision E
2007623-079
Page 79
Troubleshooting: General Troubleshooting for the System
Monitor Went Black (No Signal)
One Monitor Went Black (No Signal)
If only one monitor went black, check to see if the power light is on.
If the power light is not on, check power to the monitor.
If the power is OK, the fuses are OK, and the power light is still off, the monitor needs to be replaced.
If the power light is on, hook the monitor up to another monitor signal cable.
If the monitor still does not work with a good known signal, the monitor must be replaced.
Multiple Monitors Went Black (No Signal)
If more than one monitor goes black, there is a possible VID/MUX problem. Get assistance from Te chnical Support.See the How to Reach Us insert at the front of the manual.
Take the time to make all the recommended visual chec ks (refer to the visual inspection chart on the next page) before starting any detailed troubleshooting procedures
Revision E Mac-Lab/CardioLab 5-5
2007623-079
Page 80
Inspection List
Area Look for the following problems
I/O Connectors and Cables
Troubleshooting: General Troubleshooting for the System
Table 1. Visual Inspection List (Continued)
Fraying or other damage
Bent prongs or pins
Damaged housing
Loose screws in plugs Fuses Interface Cables
Circuit Boards
Ground Wires/Wiring
Mounting Hardware
IEB
Type and rating. Replace as necessary.
Excessive tension or wear
Loose connection
Strain reliefs out of place
Moisture, dust, or debris (top and bottom)
Loose or missing components
Burn damage or smell of over-heated components
Socketed components not firmly seated
PCB not seated properly in edge connectors
Solder problems: cracks, splashes on board, incomplete feedthrough, prior modifications or
repairs
Loose wires or ground strap connections
Faulty wiring
Wires pinched or in vulnerable position
Loose or missing screws or other hardware, especially fasteners used as connections to
ground panes on PCBs
IEB not plugged in to working wall outlet.
CPU or peripheral device not properly connected to AC Distribution Unit. Power Source
5-6 Mac-Lab/CardioLab Revision E
Faulty wiring, especially AC outlet
Circuit not dedicated to system
Power source problems cause static discharge, resetting problems, and noise.
2007623-079
Page 81
Troubleshooting: General Troubleshooting for the System
Power On Failure Modes
No display, system appears not to initialize
Most errors will be displa yed on the window. Th e following are addi tional things to check.
Can you hear the floppy disk drive being accessed several seconds after Power On? The “in use” LED will come on at the same time. The keyboard LEDs will turn on at Power On and turn off just prior to the floppy disk drive’s LED turning on. These LEDs indica te that the PC/AT system (EPC-9) is functioning properly.
Is the floppy disk drive active?
If you do not hear the floppy disk drive, check the fuses and power.
Several different failures modes may exist. If the floppy drive LED is never active and there is brief activity on the hard disk drive (listen for the drive activity or watch for the hard disk drive light), the problem is probably the floppy disk drive, and vice versa.
If the two drives appear inactive at initialization, there is probably a controller failure.
Is the Hard Disk Drive Active?
If the system appears to initialize, but there is still no display, listen for the hard drive or watch the hard drive light to see if the hard disk drive is active several seconds after turning the syste m on. This activation should occur shortly after you hear the floppy disk drive spin up.
If the hard disk drive is not acti ve , com mu ni cat io n be twee n the disk drive and the control board does not exist, and the software is not being load ed into the s yst em fro m the dri ve . Th e hard disk drive should be active for about 17 seconds, indicating that it is loading software for the entire system. If it is on for only about 2 to 3 seconds, then it is not loading the software.
Is the COMMS board active? Check for the green LED on the COMMS board.
If the system ap pears inact ive, o bserv e the LEDs on th e COMMS board. After Power On, the green LED sh ould be on.
Is the software present?
Revision E Mac-Lab/CardioLab 5-7
2007623-079
Page 82
Troubleshooting: Functional Checkout Procedure

Functional Checkout Procedure

Introduction

As part of the system checkout procedure, hook up a simulator to the TRAM modul e in the RAU or the CardioLab II Plus Amplifier located in the procedure area.
1. If a GE or Bio-Tek simulator is being used, connect the P1 output to the P1 input of the TRAM module in the slot A position.
2. Connect the P2 output to the P2 input of the same TRAM module.
3. Define the pressures to be Arterial and Venous. These two pressure simulations provide a re alisti c represent ation of pressure monitori ng waveforms.
4. Also connect the ECG leadwires.
See the “TRAM 100-600 A and SL Modules Service Manual” for additional information.

System Initialization

After you are su r e th at the sy st em is properl y co nnected, t he periphe ra ls are turned on, and there is no disk in the floppy disk drive, you can Power On the syst em .
1. Turn on the IEB and CPU power switches.
2. Verify that the image fills the entire window and that the colors are correct. If there appear to be video problems, refer to the “Video
Failure Modes” on page 5-13. If the system does not successfully
initialize, refer to “General Troubleshooting for the System” on page 5-4.

Front End Data Acquisition

1. Adjust the simulator so that the BP waveforms are zero.
2. Re-adjust the simulator so that it is simulating BP waveforms. If you have hooked up the simulator as recommended, you should see pressures in the window.
3. Verify that the correct ECG waveforms are present at the top of the window.
4. Verify that intracardiac inputs are present. Hook simulator ECG outputs to catheter input modules.
5-8 Mac-Lab/CardioLab Revision E
2007623-079
Page 83
Troubleshooting: Troubleshooting the RAU

Troubleshooting the RAU

RAU Description

This section is designed to clarify operation and troubleshooting of the RAU and its interfacing with the Mac-Lab/CardioLab system. The RAU is also called the “Front End”. Introductory material on the system configuration is followed by specific symptom-dependent troubleshooting techniques. Front end data, ECG, and invasive blood pressures, are acquired by using a TRAM module installed in an RAU.
Each system is shipped wi th ei t her o ne or two T RA M modul es cap able o f providing up to 12 ECG waveforms and up to 4 press ure waveforms. Note that the system RAUs for earlier MAC-LAB units are different from the Mac-Lab/CardioLab system RAU. They cannot be interchanged. Older RAUs have room for three TRAMs (one was not functional).
The system RAUs have room for only two. The system RAU receives its power and data via a single cable rather than via separate power and data cables. The system RAU does not have an on/off switch.

TRAM modules in the RAU

The TRAM modules are installed in the RAU that is mounted in the procedure area. Power for the RAU comes from the Console.
RAU Interconnection
The TRAM series of modules, used with the system RAU, has only one indicator, the power-on indicator.
If the system is operati ng with one TRAM module, it must be in the slot A position (top slot) in the RAU. Otherwise, there are no ECG waveforms.
The RAU receives its power from the IEB.
Data Transmission
The serial data from and to the TRAM module is communicated via a DB-15 connector on the rear DB connect board on the IEB. The cable also carries power for the RAU. The data is then communicated to the COMMS board.
The COMMS board must be working in order for the system to initialize. The error message “No TRAM module” appears on the review moni tor window if communication with the TRAM fails.
RS-422 serial communication at a rate of 115 K baud is used between the RAU and the COMMS board.
Revision E Mac-Lab/CardioLab 5-9
2007623-079
Page 84
Pressure Waveforms
Troubleshooting: Troubleshooting the RAU
Individual waveforms may be removed from the monitor window by selecting that pressure and turning off the pressure. This removes the trace.
Channel is labeled, but no pressure values are listed:
If there is a label, but no pressure numbers, and a transducer is connected, a zeroing procedure may be required.
Channel is labeled, but no pressure values or waveforms:
If a pressure site label is present, but there are no pressure numbers, a waveform may or may not be present. Check the pressure gain of the system.
5-10 Mac-Lab/CardioLab Revision E
2007623-079
Page 85
Troubleshooting: Troubleshooting the Video Processing
MENU
+
-
POWER
POWER SWITCH
POWER INDICATOR + BUTTON (Next Choice)
- BUTTON (Previous Choice) MENU (also ACCEPT)
D-SUB Mini 15 Pin Connector Red
Green/Sync Blue H/HV V AC IN Connector
V
H/HV B G/SYNC
R
AC IN
D-SUB

Troubleshooting the Video Processing

General Information

The system video consists of one or more color monitors. For most configurations, two are mounted on the workstation and at least one is mounted in the catheterization procedure lab . At the desk are the real­time and review monitors. In the catheterization procedure lab is a remote monitor, which is a slave of the other monitors.
The review monitor displays waveforms and system-related informati on and allows the technician to operate the system via the menu bar at the top of the window.
The real-time monitor displays real-time data, such as ECG and blood pressure waveforms. The monit or in the catheterization procedure lab can display either workstation moni tor signal, allowing the doctor to track monitoring and measurement functions.
Revision E Mac-Lab/CardioLab 5-11
2007623-079
PK1
Page 86
Monitor Description
Installation Checklist
Troubleshooting: Troubleshooting the Video Processing
All monitors are color monitors that are controlled by the video boards in the computer. The monitors themselves are not serviced by GE Medical Systems Service, but are sent back to the manufacturer for repairs. However, there is a window calibration procedure that includes brightness and position of the tracings. Several brands of monitors have been used. All are essentially the same, but have different controls.
Be sure that the ventilation openings are not blocked or covered.
Do not place the monitor near a heating appliance or in direct sunlight.
Do not expose th e mo nitor to m oistu re of any k ind, as th is may re sult in electrical shock or fi re hazard.
Route the power cord and cables so that they will not be walked on, tripped on, or rolled over. The monit ors should be powered vi a the AC outlets in the lab or should be connected to one of the outlets on the IEB.
Never use a monitor if the power cord has been damaged.
Unplug the monitor and clean the window as needed with a small amount of a mild glass cleaner sprayed on a soft cloth. Never spray cleaner directly on the window as it may penetrate the monitor enclosure.
Never open the enclosure. The re are no user-se rviceable parts i nside. Risk of serious shock hazard! Refer all monitor problems to GE Service personnel.
5-12 Mac-Lab/CardioLab Revision E
2007623-079
Page 87
Troubleshooting: Troubleshooting the Video Processing

Video Failure Modes

Color Distortion
Demagnetize the mo nitor window by pressing the degauss switch, which generally is located just below the monitor window or can be found by pressing the MENU button.
One or More Colors Missing on Monitor
1. Confirm that cabling for all three colors is connected to the monitor and that the cabling is properly connected to the rear interconn ection panel and to the Video/Writer board in the console.
2. If the problem moves to the other moni to r, then repl ace t he def ect iv e board.
3. Otherwise, replace the monitor.
Video Signal on One Monitor is Distorted or Missing:
1. If video signals are not present, verify that monitor is receiving power.
2. Take the video cable from the back of the working monitor an d plug it into the monitor with the bad signal. This will verify that the monitor works.
3. If the step above p roduces a good ima ge, try repla cing t he video ca ble from the VIDMUX to the monitor.
4. If problem continues, replace cable from the associated video signal input on the VIDMUX to correct video board the computer.
5. If the problem persists, the problem may be with the system hardware.
Revision E Mac-Lab/CardioLab 5-13
2007623-079
Page 88
Troubleshooting: Thermal Writer Problems

Thermal Writer Problems

Thermal Writer Starts to Print, then Stops on its Own

Check and clean dust from the paper sensor by using canned air. Paper sensor is located on the left side of the top of the paper slot. If the thermal writer still fails to work properly, get assistance from Technical Support.

Thermal Writer Outputs Blank Sheets of Paper

Check and clean dust from the paper sensor by using canned air. Paper sensor is located on the left side of the top of the paper slot. If the thermal writer still fails to work properly, get assistance from Technical Support.

Thermal Writer Does Not Print (Blank Sheets Only)

If the thermal wri te r pr int s onl y blank s hee ts of p ap er, t he paper ma y be installed upside down. Try fl ippi ng p aper en d fo r en d. The t op side of the paper should mark easily if scratched. Verify that the correct paper is being used. If the thermal writer still does not work correctly, get assistance from Technical Support.

Paper Jams During Printing

Verify that correct paper is being used. If the thermal writer still does not work correctly, get assistance from Technical Support.

Thermal Writer Does Not Print, Makes Machine Gun Noise

A machine gun noise indicates a possible hardware p roblem. Get assistance from Technica l Support. Replace the thermal writer.
5-14 Mac-Lab/CardioLab Revision E
2007623-079
Page 89
Troubleshooting: Troubleshooting the Thermal Writer

Troubleshooting the Thermal Writer

Thermal Writer Description

The system uses a direct digital writer with an 8.5-inch thermal print head. The paper is a plain, fan-fold style with no pre-printe d grid. The system prints the necessary grid at the same time as it prints the waveforms and text. The thermal writer can run at four speeds: 5, 25, 50, and 100 mm/sec. Paper is stored in a slot backed by a paper strap that slides in and out for ease of replacement.
The thermal writer release button, which disengages the print head and opens the door, must be pres sed before c losing the door a s well. When t he door is open the 26.5 volt power is removed from the print head, but not from the thermal writer control board. The 5 volt power is still on the print head. The print head assembly is loc ated inside and above t he door. This allows easy access for cleaning and replacement.

Repairing the Thermal Writer

Generally, the only in-field repair of the thermal writer is to replace the print head or anti-static brush. Replacing the printer engine in its bracket requires the use of a fixture, and so is performed only at the factory. Likewise, the electrical adjustment is only done at the factory. This is different from older MAC-LAB procedures.
Thermal Writer Failure Modes
Chart paper will not queue. Thermal writer runs extra sheets of blank paper.
Erratic or no chart paper drive
The queuing sensor could be dirty. Clean the queuing sensor by blowing air across it.
If the paper does not advance, or it advances erratically and there is no error message, check the fixed and movable cables that connect along the side of the writer drawer. Confirm that there is a solid connection.
Verify that all required conditions are met to enable the thermal writer.
Revision E Mac-Lab/CardioLab 5-15
2007623-079
Page 90
Troubleshooting: Troubleshooting the Thermal Writer
Printing too light or too dark
If the thermal print head voltage is too low, the printing may be too light, and the traces may fade prematurely. If the voltage is too high, the printing may be too dark and the life of the print head will be shortened. Voltage adjustments must be performed at the factory.
Check the thermal paper quality. Some non-GE thermal papers are more heat-sensitive than others.
Check print head voltage to ensure that it is in range: 24V to 27V.
If there is no other reason, replace the Thermal Writer.
White line running across the printing
Clean the print head and the print roller. If the lines still occur and they are obscuring the tracings, replace the print head.

RAU Failure Modes

ECG labeled, but no ECG data
These symptoms indicate no communications with the TRAM modules. Confirm that each TRAM module is installed and that the RAU is plugged in.
Check that the cable is connected on the bac k of the RAU (DB-15 connector) and is connected on the rear DB connect board on the IEB (DB-15 connector).
Is the power-on LED on the TRAM module illuminated? Try switching the position of the TRAM modules.
BP sites are labeled, but no pressure numbers or waveforms
This condit ion indi cates t hat a tra nsduc er needs t o be zer oed or rep laced. Rezero this BP by selecting the correct channel and pressing the Zero button. This could also be a TRAM module problem.
BP waveforms are present, but no ECG waveforms
Some patients require skin preparation, including abrasion, depending on their skin type.
If the BP waveforms are present, and the ECG cable is properly connected, ensure that the ECG waveforms have not been turned off.
If the ECG waveforms are selected, are not turned off, and still are not present, suspect a TRAM module failure.
ECG waveform is present, but one or both BPs are missing
Has the BP trace been turned off at the keyboard? Try rezeroing and replacing the corresponding transducer. This situation also could be a TRAM module failure.
5-16 Mac-Lab/CardioLab Revision E
2007623-079
Page 91
Troubleshooting: Server UPS Troubleshooting

Server UPS Troubleshooting

The following tables provide troubleshooting information for the Tower UPS. For more detailed information, see the OEM manuals incl uded with the UPS.
For troubleshooting recommendations for the rack-mount server UPS, see the OEM manuals.
Revision E Mac-Lab/CardioLab 5-17
2007623-079
Page 92

During Start

Troubleshooting: Server UPS Troubleshooting
Symptom Possible Cause Suggested Action
UPS will not start
LED 5 is red (Site wiring fault)
LED 10 is red (Battery Service)
No Utility Power Check Power at the Utility power receptacle or
contact a qualified electrician.
UPS Power cord disconnected
UPS input circuit breaker open
Utility power receptacle ungrounded or no ground wire in UPS power cord
Line and neutral wires reversed in utility power receptacle or in UPS power cord
Battery voltage is low because the UPS has been out of service for a long period
The UPS was powered on with the Sleep mode disabled, draining the batteries
Connect the power cord.
Press the circuit breaker button to reset. If the breaker trips repeatedly, contact an authorized service provider (high models only).
Contact a qualified electrician.
For units factory-configured for 208V, the Site Wiring Fault function has been disabled. If reconfiguring a 230V unit to operate at 208V, the Site Wiring Fault function must be manually disabled (high models only).
Allow the UPS to charge the batteries for 24 hours. Initiate a self-test: if LED 10 does not turn off, replace batteries.
Utility voltage has returned to the voltage range for which the UPS has been configured. The UPS is supplying utility power. The audio alarm should be reset.

After Start

Symptom Possible Cause Suggested Action
Audio Alarm Alarm condition exists Identify the red LED associated with this alarm
condition. Check this troubleshooting guide to determine the cause of the alarm.
LED 1 is red (Input power)
LED 1 is flashing red
5-18 Mac-Lab/CardioLab Revision E
Utility voltage is too high
Alarm may need to be reset
2007623-079
The utility voltage is higher than the UPS operating range. The UPS switches to battery power. If this happens repeatedly, update the configuration.
The utility voltage has returned within the UPS operating range. Press the TEST/ALARM RESET button.
Page 93
Troubleshooting: Server UPS Troubleshooting
Symptom Possible Cause Suggested Action
LED 4 is red
Utility voltage is too low The utility voltage is lower than the UPS
(Input power)
LED 4 is flashing red
UPS frequently
Alarm may need to be reset
Utility power variations The utility voltage is frequently outside the UPS switches between utility and battery power
LED 11 is red Protected devices are
exceeding the UPS
power rating
(UPS may switch from
utility to battery power) LED 9 is red
Low battery voltage If the UPS is supplying battery power, save (Battery Charge)
operating range. The UPS switches to battery power. If this happens repeatedly, update the configuration. Contact a qualified electrician to make sure that utility power is suitable for the UPS.
The utility voltage has returned within the UPS operating range. Press the TEST/ALARM RESET button.
operating range. Update the configuration. Contact a qualified electrician to make sure that utility power is suitable for the UPS.
Remove one or more devices to reduce the power requirements.
Make sure that the devices are not defective.
current work and shut down the system. Allow the batteries to charge. If the UPS is supplying utility power, no user intervention is required. Allow the batteries to charge.
Insufficient warning of low batteries
LED 10 is red (Battery Service)
LED 10 is flashing red; audio alarm cannot be silenced
Battery service
required
Shutdown Delay
configuration
inappropriate
Potential battery failure
detected
New batteries
improperly connected
Internal UPS over-
voltage condition exists
Allow batteries to charge for 24 hours, then initiate self-test. If LED 10 is red, replace batteries.
Update the Shutdown Delay from 5-seconds to 3-minutes. Use Compaq Power Management Software to specify a suitable delay.
Allow batteries to charge for 24 hours, then initiate self-test. If LED 10 is red, replace batteries.
Check connections.
Shut down the UPS. Contact an authorized service provider.
Revision E Mac-Lab/CardioLab 5-19
2007623-079
Page 94

UPS Alarm Conditions

Troubleshooting: Server UPS Troubleshooting
Silencing an Audio Alarm
Alarm
Condition
Utility power failure
Site wiring faulty LED 5 on red Yes Battery service LED 10 on red Yes Internal UPS
over voltage
To silence the alarm, pres s the TE S T/ A LAR M RESE T butt on.
Even though an audio alarm may be silenced, the condition that caused the alarm may still exist. For information on procedures to follow if the UPS detects an alarm condition, see the Troubleshooting chapter or the UPS manual.
If a utility power failure caused the alarm (AC Input LED 1 or LED 4 red), the alarm will be silenced after utility power is restored.
LED Activity Can be disabled?
LED 1 or LED 4 red Yes
LED 10 flashing red No
5-20 Mac-Lab/CardioLab Revision E
2007623-079
Page 95
Troubleshooting: TRAM-net 4A LED Troubleshooting Chart

TRAM-net 4A LED Troubleshooting Chart

Use the following chart to determine if the TRAM-net 4A is functioning properly. This chart refers to LEDs on the Processor Acquisition PCB. For problems, refer to the appropriate manual for more information about troubleshooting a patient monitor and TRAM-net housing together.
Processor Acquisition PCB Indicators
LED/Color Signal Name Function Condition
DS6/Green Power Indicator Proper Operation
Visible on TRAM-net 4A front bezel (lower left)
DS5/Red Network Activity I hear talking on TRAM-net
DS4/Green Transmit Enable I am talking on TRAM-net
DS3/Yellow Error Detect I detect an error on this PCB
DS2/Red RAC COMM 80C152 COMM Processor is OK
DS1/Red RAC DAS 80C31 DAS Processor is OK
On Continuous: normal operation
Flashing Quickly (2Hz): ADC Error
Flashing Slowly (0.5Hz): watchdog time-outs
LED Off: power off, supply fault, RAC FPGA configuration fault
On mostly steady: TRAM-RAC connected
Flickers low: TRAM-net not connected (I hear myself talk)
Flickers occasionally: when talking
On continuous: while graphing
Off: normal operation
Flashes with DS1 (twice per second): normal operation
Flashes with DS2 (twice per second): normal operation
127(
“I hear talking on my TRAM-net” LED flashes:
with the “I am talking on TRAM-net” LED and
alone when someone else is talking.
Revision E Mac-Lab/CardioLab 5-21
2007623-079
Page 96
Troubleshooting: CardioLink Utilities Troubleshooting

CardioLink Utilities Troubleshooting

The informat io n be lo w sho u ld be u s ed wh e n er ror s are e nc ount er e d w ith the CardioLink Utilities software. Currently, the components are Archive, Restore and Purge. Most errors are logged in the Windows NT Application Log. Errors documented in the Windows NT Application log
will be identified in the error box displayed in the Utilities user interface.
Error Solution
DB Version Error Archive utility has detected a Clab database version
different that expected. Upgrade Mac-Lab/CardioLab software to 5.1D or 5.2 as applicable.
One of the study paths is invalid (viewed on Mac-Lab/ CardioLab machine)
Study skipped because it is in use Study is currently being accessed from an acquisition or
Study on Acquisition System Study has been skipped because it has not been
Error writing to purgelog or archivelog tables Utilities software could not write to the archivelog or
Minimum free space exceeded There is not enough free space in the archive area to
Study skipped because is has just be restored Recently restored studies will be skipped in the archive
Insufficient disk space Archive area has run out of disks space. Free up disks
Attempting to review study while archive is in progress. Wait for archive to complete, then restore study if necessary.
review system. Study will be skipped for this archive
archived to the server. Run archive generator (AG.exe) on the acquisition system before attempting to archive patient.
purgelog files. Files are either missing or corrupt. Contact technical support.
continue the archive process. Free up space in the archive area, or reduce the amount of free space required in the utility.
process.
space in the archive area or select a different archive area in the utility.
Invalid Archive Path Archive path no longer exists or is inaccessible. Contact
your network administrator.
5-22 Mac-Lab/CardioLab Revision E
2007623-079
Page 97

6 Recovery Procedures

Revision E Mac-Lab/CardioLab 6-1
2007623-079
Page 98
For your notes
Recovery Procedures:
6-2 Mac-Lab/CardioLab Revision E
2007623-079
Page 99
Recovery Procedures: CardioLink Server (5.2)

CardioLink Server (5.2)

Introduction

This section contains instructions for re-imaging the Server for Disaster Recovery and Restoring Data from Backup for a CardioLink 5.2 Server. The re-image procedure should be used in the event of catastrophic server failure to restore the server to its state previous to problems occurring. The “Restore data to the server” on page 6-22 should be used when there is a need to only restore the database or patient data in the event of data corruption or suspected data corruption.
127(
This procedure can only be run if there is a current 5.2 tape backup.
Sections “Ensure 5.2 backup tapes exist” through “Backup LMT Custom
Reports” assume that the server operating system is accessible. If the
server cannot be booted completely, skip to section “Re-Image Server (G3
Only)” on page 6-9, or “Re-Image Server (G1 and G2 Only)” on page 6-15
depending on the type of server being reimaged.

Required Parts

Part Number Description Required for
2017591-001 G3 Server Image CD G3 Servers 2024636-001 MEGA SERVER PARTITION UPGR TOOL 5.2 CD G3 Server 2007041-001 Image Cast Boot Floppy G1/G2 Servers 2007004-002 Low-End Server Image CD v1B G1 Servers 2007005-002 Mid-Level Server Image CD v1B G1 Servers 2007006-002 High-End Server Image CD v1B G1 Servers 2009336-001 Low-End Server Image CD v1B G2 Servers 2009338-001 Mid-Level Server Image CD v1B G2 Servers 2009340-001 High-End Server Image CD v1B G2 Servers 2006556-008 Prucka CardioLink version 1.1D CD G1/G2 Servers
Not all of the software listed below may be installed on a particular server.
2015220-001 Core Update CD G1/G2 Servers 2015222-001 CardioLink 5.2 CD G1/G2/G3 Servers 2023718-001 HP/Sony AIT Driver CD G1/G2/G3 Servers 2015474-001 LMT 1.3C Server CD G1/G2/G3 Servers 2015472-001 LMT 1.3C Client CD G1/G2/G3 Servers
Revision E Mac-Lab/CardioLab 6-3
2007623-079
Page 100
Recovery Procedures: CardioLink Server (5.2)
Part Number Description Required for
2017867-001 LMT Backup/Restore Utility CD G1/G2/G3 Servers 2006961-001 LMT Options Floppy Disk G1/G2/G3 Servers 2017869-001 LMT and ML-CL Serial Number Update CD G1/G2/G3 Servers N/A GEMS Advanced Security Floppy Disk (should be on-site) G1/G2/G3 Servers

Ensure 5.2 backup tapes exist

Before reimaging, ensure a full set of 5.2 (both full and incremental) backup tapes are available.
127(
Any studies performed since the last backup must be reviewed from the optical drive after the disaster recovery procedure is completed.

Stop data replication on the server

Perform this procedure if the server is still running. If not, skip to
“Replace the Server Hard Drive (if applicable)” on page 6-9
1. At the server, log in as mlcltechuser.
2. Select Start > Programs > Microsoft SQL Server 7.0 > Enterprise Manager.
3. Under SQL Server Group, navigate to and highlight the server.
4. Select Tools > Replication > Disable Publishing.
5. Click Next.
6. Select Yes, disable publishing on [Server name] and click Next.
7. Click Next in the Confirm Dropping of Publications window.
8. Click Finish in the Completing the Disable Publishing and Distribution Wizard window.
9. Click OK.
10. Right-click the server name and select Refresh.
11. Expand the server.
12. Verify the Replication Monitor is NOT visible.
13. Close the SQL Server Enterprise Manager window.
6-4 Mac-Lab/CardioLab Revision E
2007623-079
Loading...