GEHealthcare
MARSAmbulatoryECGSystem
UNIXtoPCConversionInstructions
SoftwareVersion1.0
2027879-040RevisionB
MARSAmbulatoryECGSystem
UNIXtoPCConversionInstructions
English
©2010GeneralElectricCompany.
AllRightsReserved.
PublicationInformation
TheinformationinthismanualappliesonlytoMARSAmbulatoryECGSystemVersion8.0.Itdoesnotapplytoearlierproductversions.Due
tocontinuingproductinnovation,specicationsinthismanualaresubjecttochangewithoutnotice.
MARS,MUSE,andSambaSystemsaretrademarksownedbyGEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompany
goingtomarketasGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheirrespectiveowners.
Thedocumentpartnumberandrevisionappearatthebottomofeachpage.Therevisionidentiesthedocument’supdatelevel.The
revisionhistoryofthisdocumentissummarizedinthetablebelow.
Revision
A
B
Date
12March2010 Initialreleaseofthisdocument.
7June2010
Comment
ReviseSambaServiceschaptertoreectchangesfromSPR.
2
MARSAmbulatoryECGSystem
2027879-040B
7June2010
Contents
1 Introduction
Conventions................................................................................................6
2 Conversion
PlanningYourConversion...........................................................................7
ConversionWorkow..................................................................................7
3 SambaServices
StoppingtheSambaDaemon.....................................................................9
SharingFoldersontheMARSUNIXSystem...............................................10
TypographicalConventions.........................................................................6
SafetyConventions...................................................................................6
IPAddressofUNIXMachineonenterpriseNetwork.................................11
4 RestoringArchiveDatatoUNIXWorkstation
RestoringtheData....................................................................................13
EjectingtheTape.......................................................................................14
CopyingtheRestoredDatatoaNetworkDrive........................................14
5 MARSUNIXMigrationTool
InstallingMARSUNIXMigrationTool........................................................15
6 ConvertingData
UnderstandingConvertandArchive.........................................................17
ConvertandArchive(Recommended).......................................................18
ConvertOnly.............................................................................................19
ArchiveOnly..............................................................................................20
Viewing/SavingtheLog.............................................................................20
7 TheConversionLog
2027879-040B
MARSAmbulatoryECGSystem
3
ReadingtheConversionLog.....................................................................21
8 CheckoutProcedure
CheckingtheConversionontheMARSPC................................................23
A VIEditor
StartingtheVIEditor.................................................................................25
ExitingVI...................................................................................................25
VICommands............................................................................................25
B UsageInstructionsUsingCommandLines
UNIXtoPCCommandLine........................................................................27
ListingMARSUNIXSlot/ReportFile(s).......................................................28
ConvertingMARSUNIXSlot/ReportFile(s).................................................28
ArchivingConvertedMARSUNIXSlot/ReportFile(s)..................................28
ConvertingandArchivingMARSUNIXSlot/ReportFile(s)..........................29
ViewingtheLog.........................................................................................29
SavingtheLog...........................................................................................29
4
MARSAmbulatoryECGSystem
2027879-040B
1
Introduction
TheMARSUNIXPCMigrationToolconvertsMARSUNIXslotandreportlesforuseby
MARSPCversion8.0.Tousethetool,youmustrstrestorearchivedlestotheMARS
UNIXsystemfromthetapearchive.Thetoolcannotdirectlyconvertlesfromtape.
Thetoolsupportsthreeworkows:
Workow
ConvertOnly
ArchiveOnly
ConvertandArchive
Toconvertslotandreportles,youmustrestorethemonthecustomer’sexisting
MARS3000,MARS5000,MARS8000system,runningMARSUNIXversion4.1.
YoucanruntheconversiontoolonasystemwithorwithoutMARSPCinstalled.
SupportedOperatingSystemsforthetoolincludeWindows2003Server ,Windows
2000,andWindowsXP.BecausethetoolisWindows-based,youmustsetupSamba
Services ontheMARSUNIXsystem.ThisallowstheWindowssystemtoaccessthe
MARSdirectory,wheretherestoredlesreside.
NOTE:
SambaallowsnetworksharesforchosenUNIXdirectoriestoappeartoMicrosoft
WindowsasnormalWindowsfoldersaccessibleviathenetwork.
Instructionsincludeacheckoutproceduretoverifythattheconversionwassuccessful.
Youmayconvertthefollowingletypes:
FileExtension
.rpt
.nat
.nrt
Description
ConvertUNIXslot/reportles.Youcansaveandarchive
theconvertedlesatalaterdate.
ArchivealreadyconvertedUNIXslot/reportles.
Archiveslot/reportlesimmediatelyafterconversion.
DataType
Reports
Holter(SEER,TAPE,monitoring)
Monitoring
Theinstructionsprovidedinthisdocumentassumeyouarerunningthetoolusingits
userinterface.Youcanalsorunthetoolusingcommandlines.Theinstructionsfor
usingcommandlinesareinAppendixB.
2027879-040B
MARSAmbulatoryECGSystem
5
Introduction
Conventions
Thefollowingconventionsareusedinthismanual.
TypographicalConventions
Style Denition
BoldText Indicateskeysonthekeyboard,texttoenter,orhardwareitemssuch
Italicized-BoldText Indicatessoftwaretermsthatidentifymenuitems,buttonsoroptions
Ctrl+EscIndicatesakeyboardoperation.A(+)signbetweenthenamesof
<Space>
Enter
asbuttonsorswitchesontheequipment.
invariouswindows.
twokeysindicatesyoupressandholdtherstkeywhilepressing
thesecondkeyonce.
Forexample,“PressCtrl+Esc”meanstopressandholddowntheCtrl
keywhilepressingtheEsckey.
Indicatesthatyoumustpressthespacebar.
Wheninstructionsaregivenfortypingaprecisetextstringwithone
ormorespaces,thepointwherethespacebarmustbepressedis
indicatedas:<Space >.Thepurposeofthe<>bracketsistoensure
youpressthespacebarwhenrequired.
Indicatesthatyoumustpressthe“Enter ”or“Return ”keyonthe
keyboard.Donottype“Enter ”.
SafetyConventions
ThetermsDANGER,WARNING,andCAUTIONareusedthroughoutthismanualtopoint
outhazards,andtodesignateadegreeorlevelofseriousness.Familiarizeyourself
withtheirdenitionsandsignicance.
Safety
Convention
Hazard Asourceofpotentialinjurytoaperson.
DANGER
WARNING
CAUTION
NOTE
Denition
Indicatesanimminenthazardwhich,ifnotavoided,willresultindeath
orseriousinjury.
Indicatesapotentialhazardorunsafepracticewhich,ifnotavoided,
couldresultindeathorseriousinjury.
Indicatesapotentialhazardorunsafepracticewhich,ifnotavoided,
couldresultinminorpersonalinjuryorproduct/propertydamage.
Providesapplicationtipsorotherusefulinformationtoassurethatyou
getthemostfromyourequipment.
Additionalsafetymessagesthatprovideappropriatesafeoperationinformationmay
befoundthroughoutthismanual.
6
MARSAmbulatoryECGSystem
2027879-040B
Conversion
PlanningYourConversion
RestoringtherawslotlesfromthetapearchivetotheMARSUNIXsystemcantake
uptothreeminutesperle.Thisisduetothesizeoftheslotleanddatatransferrate
ofthetapedrive.Pleaseconsiderthetimeinvolvedwhenplanningyourconversion.
Forexample,asingletapearchiving50slotlescouldtakeaslongas2.5hoursto
restore.OncerestoredontheMARSUNIXsystem,theconversionprocessitselfruns
quickly,takingonly15-30secondstoconverteachslotle.
Toshortenthetimeneeded,youmaywanttoconsiderconvertingonlyreportles.
Theselesaremuchsmallerandtakemuchlesstimetorestore,andonlyafew
secondstoconvert.
Whenplanningyourconversion,rememberthatyoumusthave:
2
•AworkingMARSUNIXsystem.
•AnoperatingDATTapedriveavailable.
Makesurethatalldataconversionsarecompletedbeforeremovingthesystemfrom
thenetworkorretiringthesystemfromservice.
ConversionWorkow
Followingisthegeneralconversionworkow:
1. Ifyouareconvertingtherawholterdata,deletepatientsontheUNIXsystemto
openasmanyslotsaspossible.
2. RestorethearchivedatafromDATtotheUNIXworkstation.
3. Chooseoneofthefollowingoptions:
•RuntheconversiondirectlyontherestoredslotlesontheUNIXsystem.
•Copytherestoredslotlestoanetworkshareandconvertthelesfromthere.
NOTE:
4. RuntheUNIXtoPCConversionToolonaWindows-basedsystem
Thesecondoptionallowsyoutorestoreslotleswithouttheneedto
immediatelyconvertthem.
2027879-040B
MARSAmbulatoryECGSystem
7
Conversion
5. SelectConvertorArchive.
•ConvertsavestheconvertedlesintheMARSPCslotleformat(thatis,
seerslot1s20.nat )
•ArchivesavestheconvertedlesintheMARSPCarchiveformat(thatis,
archive.nat )
6. Setthesourcelelocation.
Thisiswheretherestoredslotlesarelocated.
7. Setthedestinationlelocation.
Thisiswhereyouaresavingtheconvertedles.
8. Selectthelesorreportsyouwanttoconvert.
9. Runtheconversion.
10. Repeatthesestepsasrequired.
8
MARSAmbulatoryECGSystem
2027879-040B
SambaServices
ThischapterdescribeshowtosetuptheSambaservicestodisplayUNIXfolders
onaPC.
StoppingtheSambaDaemon
ThefollowingstepsdescribehowtostoptheSambadaemon.
1. FromtheMARSapplication,holdthemiddlemousebutton,andclickanddrag
anysystemiconontothesystemmenuheader,locatedintheupper-leftcorner
ofthescreen.
AcmdtoolwindowopensandtheMARSxx-yyyycommandpromptisdisplayed.
2. Attheprompt,typesuroot.
3. Atthenextprompt,enterthepasswordfortherootuser.
3
Thisgivesyourootaccountpermissions.
NOTE:
Ifyoudonotknowthepassword,contactGEHealthcareTechnicalSupport.
4. CheckifSambaisrunning:
a. Attheprompt,typeps[space]—eandpressEnter.
b. Reviewthelistandidentifyifthesmbdprocessispresent.
Ifitispresent,makenoteoftheProcessID(PID)andcontinuetostep5.
Ifsmbdisnotdisplayedinthelist,itisnotrunning.Skipstep5andgoto
“SharingFoldersontheMARSUNIXSystem”onpage10.
5. KilltheSambaprocess:
a. Atthepasswordprompt,enterthepasswordfortherootuser.
NOTE:
Ifyoudonotknowthepassword,contactGEHealthcareTechnical
Support.
b. Typekill<PID>andpressEnter.
Where<PID>istheprocessnumberassociatedwiththesmbdprocess.
2027879-040B
MARSAmbulatoryECGSystem
9
SambaServices
SharingFoldersontheMARSUNIXSystem
UsethecommandsinthissectiontosharefoldersontheMARSUNIXsystem.
1. Openthesmb.conflefromthecmdtoolusingVIEditororfromtheCommon
DesktopEnvironment(CDE) usingtheTextEditor.
UsingtheVIEditor:
1. Openacmdtoolwindowandlogonasrootuser.
2. Typevi[space]/etc/smb.conf.
Thesmb.confleopensintheeditor.
NOTE:
IfyouareunfamiliarwiththeVIEditor,refertoAppendixA“VIEditor”
onpage25.
UsingtheTextEditor
1. LogontoMARSasarootuser.
TheCDEisdisplayed.
2. ClickontheTextEditoricon.
TheTextEditorwindowopens.
3. Browsetothe/etcfolderandopenthesmb.confle.
2. Addthefollowingentriestothesmb.confles:
[slots]
comment=UNIXslotfiles
path=/slots
public=yes
writeable=yes
createmask=UNIXpermissions
3. Saveyourchanges.
IfyouareusingtheVIEditor,type:wqtosaveyourchangesandexittheeditor .
IfyouareusingtheTextEditor,selectSavefromthelemenu.
4. Fromacmdtoolwindow,makesureyouareloggedonasrootuserandtype
exec[space]/opt/MarsSamba/bin/smbd .ThisstartstheSambadaemon.
5. Typeps[space]—eandcheckthatthesmbdprocessislisted.
NOTE:
Ifyouarehavingaproblemgettingtheprocesstostart,youmayneedto
doaSystemShutdownandboot.
10
MARSAmbulatoryECGSystem
2027879-040B
IPAddressofUNIXMachineonenterprise
Network
TheIPaddressoftheUNIXmachineisneededlatertogainaccesstothefolder
sharesonthenetwork.
1. FromtheMARSapplication,holdthemiddlemousebutton,andclickanddrag
anysystemiconontothesystemmenuheader,locatedintheupper-leftcorner
ofthescreen.
AcommandtoolwindowopensandtheMARSxx-yyyycommandpromptis
displayed.
2. IdentifytheIPaddresstotheEnterpriseNetwork:
a. Attheprompt,typesuroot.Thischangestheuseraccountforthe
commandlinetooltotherootuser.
b. Atthepasswordprompt,enterthepasswordfortherootuser.
NOTE:
Ifyoudonotknowthepassword,contactGEHealthcareTechnical
Support.
SambaServices
c. Atthecommandlineprompt,typeifcong[space]—aandpressEnter.
AlistofIPaddressesisdisplayed.
d. MakenoteoftheIPaddressconnectedtotheEnterpriseNetwork.
NOTE:
MARSUNIXsystemsconnectedtothemonitoringreal-timenetwork
havemorethanonenetworkcard.TheaddressfortheEnterprise
Network typicallyisidentiedasthehighesthmeXnumberinthelist.
Forexample,ifbothhme1andhme2arelisted,hme2isnormallythe
IPaddressfortheEnterpriseNetwork.
2027879-040B
MARSAmbulatoryECGSystem
11
SambaServices
12
MARSAmbulatoryECGSystem
2027879-040B
RestoringArchiveDatatoUNIX
Workstation
YouneedalloftheslotlesontheMARSworkstationopentorestoreasmanyslot
lesaspossible.Toopenslotles,youneedtodeletepatients,asdescribedinthe
MARSOperator’sManual .
YoumaywanttoconvertexistingslotlesandreportsontheMARSUNIXworkstation
astherststepoftheconversionprocessbeforedeletingthem.Ifyoupreferto
archivethecurrentdataandconvertitlater ,copyittoDATTape,asdescribedinthe
Operator’sManual,“StoringDATAtoTAPE”intheMARSUser’sManual.
RestoringtheData
1. InserttheDATtapeintheDATtapedrive.
4
Pushthetapeintothedriveuntilthedrivedrawsitin.Whenready,the DAT
driveactivitylightglowsgreen.
2. SelectthePatientSelectmenuicon.
3. ClickDATintheDataTypelistbox.
4. Selectthepatientorpatientsyouwanttorestore.
5. ClickTools.
6. ClickRestorefromDAT.
Thefollowingmessageisdisplayed:YouhaveselectedXle(s)forDATrestore.
Areyousureyouwanttorestoretheselectedle(s)?
7. ClickYestorestoreasinglepatient,orclickYestoALLtorestoremorethan
onepatient.
TheDATactivitylightblinksforafewsecondsandthefollowingmessageis
displayed:Xle(s)queuedforDATrestore.
8. ClickOK.
Theworkstationrestoreslestothefollowingdatatypes:
2027879-040B
MARSAmbulatoryECGSystem
13
RestoringArchiveDatatoUNIXWorkstation
EjectingtheTape
1. Checkthattheactivitylightisnotglowing.
2. PressEject.
3. Thetapecartridgeejects.
4. Removethetapedrive.
Repeatthisprocedureforeachsetofslotlesyouarerecovering.
Files
Tape,SEER,orMonitoringHolter Holter
DischargedMonitoringPatientFiles
Report
Thestatusoftheselectedle(s)changesfromQueuedforAcqto%Acquired
andnallytoReadytoEdit.
NOTE:
PressingEjectwhiletheActivitylightisblinkingmayinterruptatapeaccess
command.
TheActivitylightashesforabout30seconds
DataTypes
Monitoring
Storedreports
CopyingtheRestoredDatatoaNetworkDrive
IfyouareconvertingthedatafromanetworkdriveinsteadoftheMARSUNIX
workstation,
1. Accessthe\slotsor\reportsfoldersharesyousetupinChapter3“Samba
Services”onpage9.
2. Copythefolderstothenetworkdriveyouareusing.
14
MARSAmbulatoryECGSystem
2027879-040B
MARSUNIXMigrationTool
InstallingMARSUNIXMigrationTool
NOTE:
CustomerswhodonothaveaninstallationCDavailablecanaccessthe UNIXto
PCDataConversiontooldirectlyfromaMARSv8Workstation.UseWindows
Explorerandbrowsetoc:gemsit\opt\MarsNT\bin.Double-clickUIUnix2PC.exe
tolaunchthetool.
1. InserttheUNIXtoPCDataConversionToolCDintothePCwhereyouare
installingthetool.
2. NavigatetotheCD-ROMdriveranddouble-clickSetup.exeinstaller.
3. SelectNextattheWelcomescreen.
4. SelecttheIacceptthetermsofthelicenseagreementcheckboxandclickNext.
5
5. AttheChooseDestinationLocationdialogbox,dooneofthefollowing:
•Toacceptthedefaultinstallationpath,clickNext.
•Tochangetheinstallationpath,clickChange,browsetothelocationwhere
youwanttoinstallthetool,andclickNext.
2027879-040B
MARSAmbulatoryECGSystem
15
MARSUNIXMigrationTool
6. AttheReadytoInstalltheProgramdialogbox,selectInstall.
7. ClickFinishwhenpromptedtocompletetheinstallation.
16
MARSAmbulatoryECGSystem
2027879-040B
ConvertingData
YouhavethefollowingoptionswhenusingtheMARSUNIXtoPCConversionTool:
•ConvertandarchiveMARSUNIXslot/reportle(s)
•ConvertMARSUNIXslot/reportles
•ArchiveconvertedMARSUNIXslot/reportle(s)
6
Therecommendedoption,thisconvertstheexistingMARSUNIXslot/report
lesandarchivesitintheMARSPCformat.Thisoptionsavestimeandavoids
confusionlateintheprocessandisdiscussedinthesection“ConvertandArchive
(Recommended)”onpage18.
ThisoptionconvertstheexsitingMARSUNIXslot/reportlesintotheMARSPC
slot/reportleformat.
ThisoptionarchivesMARSUNIXslot/reportlesthatwerepreviouslyconvertedinto
theMARSPCarchiveformat.
UnderstandingConvertandArchive
Theconversiontoolsavestheconvertedlesusingthestandardslotlenaming
convention.Toavoidgeneratingalargenumberofconvertedlesintoasinglefolder ,
GEHealthcarerecommendsyoucreateafolderstructurefortheconvertedlesso
thatitaidsinidentifyingthelesinit.Forexample,ifyoualreadyhaveanaming
schemeforthetapes,youmaywanttousesomethingsimilarasafolderstructure
fortheconverteddata.
Whendataisconvertedtothearchivestate,anindex.datleiscreatedinthefolder
andupdatedeachtimenewlesareadded.Theindex.datprovidesinformationtothe
MARSPCthatidentiestheslotorreportletothesystemtorestoreitifneeded.Ifthe
destinationfolderischanged,anewindex.datleiscreated.Youcannotcombine
archivedleslateronfromtwodifferentfolderssincetheindex.datleisindexing
onlythearchivelesforonefolder .
Eachtimeyouruntheconversiontool,itaddstheconvertedorarchivedlestothe
destinationfolder .Iftherearepreviouslyarchivedorconvertedles,newlesare
createdusingthenextnumericalsequence.Forexample,ifthelastarchivedfolderis
2027879-040B
MARSAmbulatoryECGSystem
17
ConvertingData
archive5.nat ,thenextlecreatedisarchive6.nat.Ifthedestinationfolderischanged,
thenumbersequencestartsoverat0.
NOTE:
ConvertandArchive(Recommended)
Toconvertandarchivedatainasinglestep,usethefollowingprocedure:
1. LaunchtheUIUNIXtoPCDataConversiontool.
2. ClickChangeundersourcefolderframeorselectFile->Open.Thisopensthe
3. Ifconvertingdatadirectlyfromthe\slotsor\reportsfolderontheMARSUNIX
IfyouconvertaMARSUNIXslotlethatcontainsSiteandLocationinformation,
thatinformationwillnotbeconvertedtotheMARSPCleunlesstheSiteand
LocationalreadyexistontheMARSPCsystem.Therefore,beforebeginningthe
conversion,youmaywanttoverifythatallSitesandLocationslistedontheMARS
UNIXsystemexistintheMARSPCsystem.
FileOpen dialogbox.
workstation,givetheIPaddressoftheUNIXmachineprecededby\\intheFile
Name textbox.Thatis:\\192.43.12.3.
NOTE:
Youcanalsouseamappeddrive,providedyoualreadymappedittothe
MARSUNIXmachine.
•Ifyoucopiedtherestoredslotlesorreportstoanetworkshare,providethe
UNC pathtotheshare,\\<computername>\<foldershare>.
•Ifyoucopiedtherestoredlestothesystemrunningtheconversiontool,you
needtoprovidethepathtothefolder .
4. Ifthefoldershareisrestricted,youmaybepromptedforaUsernameand
Password .
ForaMARSUNIXworkstation,thesemaybetheUsernameandPasswordfor
theMARSUNIXmachine.
5. Thetooldisplaysalltheslotlesandreportsavailableintheselectedfolder.
6. SelecttheUNIXslotsorreportlestoarchive.
7. ClickOpen.
8. Selectthedestinationfolderfortheconvertedslot/reportle(s).
9. ClickArchive.
Thetooldisplaysaprogressionbarfortheconversionoperationasitconverts
theUNIXslot/reportle(s)toPCslot/reportle(s)andthenarchivesthe
convertedslot/reportle(s)tothespecieddestinationfolder,driveorCD.
18
MARSAmbulatoryECGSystem
2027879-040B
ConvertingData
ConvertOnly
ToconvertlesonlyandsavetheminthenativeMARSPCslotorreportleformat,
usethefollowingprocedure:
1. LaunchtheUIUNIXtoPCDataConversiontool.
2. ClickChangeundersourcefolderframeorselectFile->Open.Thisopensthe
FileOpen dialogbox.
3. Ifconvertingdatadirectlyfromthe\slotsor\reportsfolderontheMARSUNIX
workstation,givetheIPaddressoftheUNIXmachineprecededby\\intheFile
Name textbox.Thatis:\\192.43.12.3.
NOTE:
Youcanalsouseamappeddrive,providedyoualreadymappedittothe
MARSUNIXmachine.
•Ifyoucopiedtherestoredslotlesorreportstoanetworkshare,providethe
UNC pathtotheshare,\\<computername>\<foldershare>.
•Ifyoucopiedtherestoredlestothesystemrunningtheconversiontool,you
needtoprovidethepathtothefolder .
4. Ifthefoldershareisrestricted,youmaybepromptedforaUsernameand
Password .
ForaMARSUNIXworkstation,thesemaybetheUsernameandPasswordfor
theMARSUNIXmachine.
5. Thetooldisplaysalltheslotlesandreportsavailableintheselectedfolder.
6. SelecttheUNIXslotsorreportlesforconversion.
7. ClickOpen.
2027879-040B
MARSAmbulatoryECGSystem
19
ConvertingData
ArchiveOnly
8. Selectthedestinationfolderfortheconvertedslot/reportle(s).
9. ClickConvert.
Thetooldisplaysaprogressionbarfortheconversionoperationasitconverts
theslot/reportle(s)andputstheconvertedlesinthespecieddestination
folder.
YoucanperformtheArchiveOnlyprocedureonlesthatwerepreviouslyconverted.
Toarchivepreviouslyconvertedles,usethefollowingprocedure:
1. InvoketheUIUNIXtoPCDataConversiontool.
2. Selecttheconvertedslot/reportle(s)intheFile->Opendialogbox.
3. ClickOpen.
4. Selectthedestinationfolderforthearchivedslot/reportles.
5. ClickArchive.
Thetooldisplaysaprogressionbarshowingthearchivalprogressasitarchives
theslot/reportle(s)tothespecieddestinationfolder.
Viewing/SavingtheLog
Thesystemcreatesanewlogleeachtimethetoolislaunched.Theloglecontinues
toappendentriesaslongasitremainsopen.Oncethetoolisclosed,thelogleis
deleted.Ifyouwanttokeeparecordoftheconversion,savetheloglebeforeclosing
thetool.
1. Duringconversionorarchiving,clickDetails.
Thetoolexpandsdownwardanddisplaysthestatusloginaneditbox.
Thetoolalsodisplaysthenumberoflesselected,thenumberoflesprocessed,
thenumberoflesremainingtoprocess,andthenumberoflesfailedin
process.
2. SelectFile->SaveLog.
ASavedialogboxopenswithadefaultname<LogleName-Date-Time>.login
intheFileNametextbox.(Forexample:Log13-May-08-10-44-48.login).
NOTE:
Thelenameanddatearetheactualname,date,andtimeofsaving.
3. ClickSavetosavethelogle.
20
MARSAmbulatoryECGSystem
2027879-040B
TheConversionLog
ReadingtheConversionLog
Aftercompletingtheconversion,checkthelogleordisplayfordetailsonthe
conversion/archivalprocess.Theendofthelogdisplaysthesummary:
Totalxle(s).
SuccessfulConversionxle(s).
FailedConversion0le(s).
Basedonthisyougetanoverviewoftheconversion’ssuccess.Fordetails,lookatthe
logforeachconversion.Eachleshouldlooksomethinglikethis:
Copyingfrom\\3.36.24.203\reports\report12.rpttoD:\DOCUME~2\romits\
LOCALS~1\Temp\\report12.rpt...Done
ConvertingfromD:\DOCUME~2\romits\LOCALS~1\Temp\\report12.rptto
D:\DOCUME~2\romits\LOCALS~1\Temp\\pcreport.rpt...Done
MovingfromD:\DOCUME~2\romits\LOCALS~1\Temp\\pcreport.rptto
C:\gemsit\unix2pc\NewFolder\report0.rpt...Done
7
Ifthereanyerrors,theyareloggedforeachle.Thefollowingtabledisplayssomeof
thecommonerrorsyoumayndinthelog:
Error
SelectedleisotherthanMARSles InvalidMARSdatale
Selectedslotleiscorrupted Filemaybeemptyorcorrupted
Diskspaceisnotavailableinthedestination
folder
SelectedleisalreadyconvertedtoPC
format
2027879-040B
MARSAmbulatoryECGSystem
LogReads:
Insufcientdiskspace
AlreadyConverted
21
TheConversionLog
22
MARSAmbulatoryECGSystem
2027879-040B
CheckoutProcedure
Thissectiondescribestheproceduretofollowifyouwanttochecktheconversion
aftertheConvertonlyworkow.(Incaseyoudonotwanttochecktheconversion
immediately,youcanfollowtheConvertandArchiveworkowandlaterrestorefrom
thearchivetoMARSandchecktheles.)
CheckingtheConversionontheMARSPC
1. ClosetheMARSapplicationifitisrunning.
2. StopallMARSservicesfromtheservicecontrolpanel.
3. Converttheslotlesand/orreportlestoPCformatusingeitherthecommand
lineorgraphicaluserinterface.YoucandothisoneithertheMARSPCorona
separateconversionsystem.
•Itishighlyrecommendedthatyouconvertslotstoadestinationfolder(other
thangemsit\slots).Thoughtheconvertedslotsarenotimmediatelyviewable
inMARS,ithelpsavoidalotofconfusion.
8
•Theinstructionsforusingthetooltoconvertslotlesarein“ConvertingData”
onpage17.
4. Checktheconversionlogforsuccessfulconversionandforanyerrors.
5. Savethelogle.See“TheConversionLog”onpage21formoredetailsonthis.
6. Renametheexisting\gemsit\slotsfolderto\gemsit\slots_backupandthe
\gemsit\reports folderto\gemsit\reports_backup.
7. Createanother\gemsit\slotsfolderanda\gemsit\reportsfolder .
8. Copytheconvertedslotandreportlestotheirrespectivefoldersundergemsit.
9. StartalltheMARSservicesfromthecontrolpanel.
10. LaunchtheMARSapplication.
11. Gotopatientselectandcheckifalltheconvertedslots/reportsarepresent.
2027879-040B
MARSAmbulatoryECGSystem
23
CheckoutProcedure
12. Foreachconvertedslot,selectitandrunthrough Patientinformation,Patient
diary,Stripreview,Pagereview,Eventreview,ShapereviewandEpisode
reviewtocheckiftheslotisviewable.
NOTE:
WhenMARSPCwasintroduced,the DetectHorizontalorDownsloping
OnlyeldwasaddedtoeventdenitionsforSTsegments.Ifthiseldis
checkedontheMARSPC,slotlesthatareconvertedfromMARSUNIX
willerroneouslydisplayanSTDepressionvalueof0.Toresolvethis,
cleartheDetectHorizontalorDownslopingOnlyeldonthePatient:
EventDenitionsSettingsandAnalysisResults window.ThecorrectST
Depressionvalueshouldthenbedisplayed.
13. Foreachconvertedreport,clickonReportreviewtocheckifthereportis
viewable.
14. Afterverifyingthattheleswereconverted,doarollbacktoreturnthesystemto
itsformerstate.
a. CloseMARS.
b. StopallMARSservices.
c. Deletetheslotsandreportsfolderingemsit.
d. Renamethe\gemsit\slots_backupto\gemsit\slots.
e. Renamethe\gemsit\reports_backupto\gemsit\reports.
f. RestartMARSservices.
24
MARSAmbulatoryECGSystem
2027879-040B
VIEditor
StartingtheVIEditor
Thecommandtostarttheeditorisvi,followedbylename.Forexample,toeditale
calledtemporary,typevitemporaryandthenpressreturn.
ExitingVI
ToexitVI:
1. PressEscandthentype:q.
2. Type:wtosavecontentsoftheeditor .
3. Type:wqtosaveandexitfromtheeditor.
A
VICommands
Command
h Movesthecursortotheleftonecharacterposition
I
j
k Movesthecursoruponeline
l Movesthecursortotherightonecharacterposition
r
u
x
dd Deletesthewholeline
2027879-040B
Action
Entersinsertmode;thecharacterstypedinareinsertedbeforethecurrent
cursorposition.Ifyouspecifyacount,allthetextthatwasinsertedis
repeatedthatmanytimes.
Movesthecursordownoneline
Replacesonecharacterunderthecursor;specifythecounttoreplacea
numberofcharacters
Undoesthelastchangetothele;typinguagainwillre-dothechange.
Deletesthecharacterunderthecursor;thecountspecieshowmany
characterstodelete;thecharactersaredeletedafterthecursor .
MARSAmbulatoryECGSystem
25
VIEditor
26
MARSAmbulatoryECGSystem
2027879-040B
UsageInstructionsUsingCommand
Lines
UNIXtoPCCommandLine
YoucanrunallavailablefunctionsintheUserInterfaceforthetoolfromaCommand
Line.ItisrecommendedthatcustomersusetheUserInterfaceprovidedwiththe
tooltoavoidconfusionandreducetheriskoferrors.Thefollowingparametersfor
commandlineusageisforGEHealthcareServicepersonnel.
<>
B
Usage
Requiredparameters
[]
cmdunix2pc[<-src:Source-Path><-usr:Username><-pwd:Password>|<-src:MapDrive>][<-ls>|[<-conv>|<-arc>[-dst:<Destination-Path>][-log][-s:[Saveas
Location]
Parameters
-src:Source-Path PathtoremoteUNIXsystem.
-src:Map-Drive
-usr:Username
-pwd:Password PasswordoftheUNIXsystem.
-dst:Destination-Path PathtostoreconvertedMARSdatales.
-ls ListsallMARSdatales.
-conv
-arc
-log Displaysthelogincommandlinewindow.
-s
-s:SaveasLocation
NOTE:
Use*.natforslotlesand*.rptforreportles.
Optionalparameters
Description
Mapsthedriveofaremotesystem.
UsernameoftheUNIXsystem.
Flagforconversion.
Flagforarchiving.
Saveslogleatcurrentlocation.
Saveslogleatthespeciedpath.
2027879-040B
MARSAmbulatoryECGSystem
27
UsageInstructionsUsingCommandLines
ListingMARSUNIXSlot/ReportFile(s)
1. OpenacommandlinewindowandgotothedrivewheretheUnix2PCtoolis
installed.
2. RuntheCMDUnix2PC.exewiththe-lsoptionalongwith–src,-usr,-pwdoptions.
Forexample:
CMDUnix2PC.exe–src:\\3.36.24.203\slots–usr:mei–pwd:mei–ls
Thiscommandlistsalltheslotandreportlesundertheslotsfolder.
ConvertingMARSUNIXSlot/ReportFile(s)
1. OpenacommandlinewindowandgotothedrivewheretheUnix2PCtoolis
installed.
2. RuntheCMDUnix2PC.exewiththe-convoption:
Forexample:
CMDUnix2PC.exe-src:\\3.36.24.203\slots\seerslot3.s24.nat–usr:mei
–pwd:mei–conv–log–dst:<destinationpath>
Thisconvertstheslotleandputsitinthedestinationfolder.
3. Forconvertingmultipleslotles,givethefollowingcommand:
Forexample:
CMDUnix2PC.exe–src:\\3.36.24.203\slots\*.nat–usr:mei–pwd:mei–conv
–log–dst:<destinationpath>
Thisconvertsallslotlesintheslotsfolderandputstheminthespecied
destinationfolder .
NOTE:
Theconvertedslotlename(s)arethenextavailablesequencenumberfor
thesameslottypeinthespecieddestinationfolder.
ArchivingConvertedMARSUNIXSlot/Report
File(s)
1. OpenacommandlinewindowandgotothedrivewheretheUnix2PCtoolis
installed.
2. RuntheCMDUnix2PC.exewiththe-arcoptionandthearchivelocationinthe
–dst option.The–srclocationpointstoalocalfolderwhereconvertedslotsare
present.Usethe*.*optionforarchivingbothreportsandslots.
28
MARSAmbulatoryECGSystem
2027879-040B
UsageInstructionsUsingCommandLines
ConvertingandArchivingMARSUNIXSlot/Report
File(s)
Forexample:CMDUnix2PC.exe–src:<Convertedslotspath>\*.*-arc–log–dst:<archive
path>
1. Thisarchivesalltheslotlespresentin<Convertedslotsfolder>.
2. OpenacommandlinewindowandgotothedrivewheretheUnix2PCtoolis
installed.
3. RuntheCMDUnix2PC.exewiththe-arcoptionandthearchivelocationinthe
–dst option.The–srclocationpointstotheUNIXfolderwheretheslots/reports
arepresent.
Forexample:
CMDUnix2PC.exe–src:\\3.36.24.203\slots\*.nat–usr:mei–pwd:mei–arc–log
–dst:<archivepath>
Thisconvertstheslotle(s)fromUNIXformattoPCformatandthenarchives
theconvertedslotle(s)inthespeciedarchivefolder.
ViewingtheLog
1. OpenacommandlinewindowandgotothedrivewheretheUnix2PCtoolis
2. RuntheCMDUnix2PC.exewiththe-logoption.
SavingtheLog
1. OpenacommandlinewindowandgotothedrivewheretheUnix2PCtoolis
2. RuntheCMDUnix2PC.exewiththeoption-logoptionand-soption.
installed.
Forexample:
CMDUnix2PC.exe–src:\\3.36.24.203\slots\seerslot2.s48.nat–usr:mei
–pwd:mei–conv–log-dst:c:\gemsit\slots
Thisdisplaysthestatuslog,thetotalnumberoflesselected,thetotalnumber
oflesconvertedsuccessfully,andthenumberoflesthatfailed.
installed
Forexample:
CMDUnix2PC.exe–src:\\3.36.24.203\slots\seerslot2.s48.nat–usr:mei
–pwd:mei–conv–log-dst:c:\gemsit\slots–s
Thissavestheloglewiththelename
Unix2pc<dd>-<mmm>-<yy>-<hour>-<min>-<sec>.log (timestamp)inthe
currentlocation.
3. Tosavethelogletoaspeciedlocation,usetheoption-s:<folderpath>.
Forexample:
CMDUnix2PC.exe–src:\\3.36.24.203\slots\seerslot2.s48.nat–usr:mei
–pwd:mei–conv–log-dst:c:\gemsit\slots–s:d:\log
2027879-040B
MARSAmbulatoryECGSystem
29
UsageInstructionsUsingCommandLines
30
MARSAmbulatoryECGSystem
2027879-040B
GEMedicalSystems
InformationTechnologies
8200WestTowerAvenue
Milwaukee,WI53223USA
Tel: +14143555000
+18005587044(USOnly)
Fax:+14143553790
GEMedicalSystems
InformationTechnologies GMβH
MunzingerStraße5
D-79111FreiburgGermany
Tel: +497614543-0
Fax:+497614543-233
AsiaHeadquarters
GEMedicalSystems
InformationTechnologies
Asia;GE(China)Co.,Ltd.
No.1HuatuoRoad
ZhangjiangHi-techParkPudong
Shanghai,People’sRepublicofChina201203
Tel: +862152574650
Fax:+862152082008
GEMedicalSystemsInformationTechnologies,aGeneralElectricCompany,goingtomarketas
GEHealthcare.