Mac-Lab/CardioLab HL7 Export Specification
Reference Manual
2047900-111B EN
Version 6.9
Introduction
Document Use
This document is intended for use by systems engineering personnel involved in the transfer of
information between the Hospital Information System (HIS) and the Mac-Lab/CardioLab.
Revision History
Revision Date Comments
A 15 August 2011 Initial release of document.
B 10 May 2012 Updated with serviceability improvements.
Dangers, Warnings, and Cautions
CAUTION: SALE OF DEVICE
U.S. federal law restricts this device to sale by or on the order of a physician.
Message Construction Rules
Message Delimiters
The interface utility must use HL7's v2.x recommended message delimiters:
Segment Terminator - <cr> (hex 0D)
Field Separator - |
Component Separator - ^
Subcomponent Separator - &
Repetition Separator - ~
Escape Character - \
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 1
10 May 2012
Message Header Segment (MSH)
The MSH is used to convey intent, content, source, destination, and some specifics of the syntax
of a message. The following table describes the MSH attributes:
GEHC
SEQ LEN DT
1 1 ST R 00001 Field Separator "|"
2 4 ST R 00002 Encoding Characters "^~\&"
3 180 HD R 00003 Sending Application “MACLAB 6.9” or “CARDIOLAB 6.9”
4 180 HD R 00004 Sending Facility “GEMS”
5 5180 HD O 00005 Receiving Application Configured via DataExportConfig utility - HL7 tab
6 6180 HD O 00006 Receiving Facility Configured via DataExportConfig utility - HL7 tab
7 26 TS R 00007 Date/Time Of Message YYYYMMDDHHMMSS
8 40 ST O 00008 Security Not Used
9 7 CM R 00009 Message Type "ORU^R01"
10 20 ST R 00010 Message Control ID “CATH_YYYYMMDDHHMMSS” or
Required
Field
ITEM # ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
“EP_YYYYMMDDHHMMSS”
11 3 PT R 00011 Processing ID “P”, “D”, or “T”
12 60 VID R 00012 Version ID “2.3”
13 15 NM O 00013 Sequence Number Not Used
14 180 ST O 00014 Continuation Pointer Not Used
15 2 ID O 00015 Accept Acknowledgment
Type
16 2 ID O 00016 Application Acknowledgment
Type
17 2 ID O 00017 Country Code Not Used
18 16 ID O 00692 Character Set Not Used
19 60 CE O 00693 Principal Language Of
Message
20 20 ID O 01317 Alternate Character Set
Handling Scheme
Not Used
Not Used
Language Identifier
Not Used
Example MSH Segment:
MSH|^~\&|MACLAB
6.8|GEMS|RECVAPP|RECVFAC|20020523214333||ORU^R01|CATH_20041108214333|P|2.3|||||
||1033||
2 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Patient Identification Segment (PID)
The PID is used to convey patient identification information. This segment contains permanent
patient identifying and demographic information. The following table describes the PID attributes:
GEHC
SEQ LEN DT
1 4 SI O 00104 Set ID - PID Not Used
2 20 CX O 00105 Patient ID (External) Not Used
Required
Field
ITEM# ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
3 20 CX R 00106 Patient Identifier List
(Internal)
4 20 CX O 00107 Alternate Patient ID -
PID
5 48 XPN R 00108 Patient Name |Patient.LastName^Patient.FirstName^Patient.Mi
6 48 XPN O 00109 Mother's Maiden Name Not Used
7.1 26 TS O 00110 Date/Time of Birth Patient.BirthDate (YYYYMMDD)
7.2 4 ST O Patient's Age at Time of
Study
7.3 8 ST O Patient's Age at Time of
Study Units
8 1 IS R 00111 Sex Gender.Gender (formatted as M, F or U)
9 48 XPN O 00112 Patient Alias Not Used
10 80 CE O 00113 Race RaceType.Race
11 106 XAD O 00114 Patient Address - Home |Patient.PatientAddress1^Patient.PatientAddres
12 4 IS O 00115 County Code Not Used
13 40 XTN O 00116 Phone Number - Home Patient.PhoneHome
Patient.MRN_PatientID
Not Used
ddleName|
DemographicResults.value where indexNum = 4
"Days", "Months", "Years"
s2^Patient.City^Patient.StateOrProvince^Patient
.Zip^Country.Abbrev|
14 40 XTN O 00117 Phone Number -
Business
15 60 CE O 00118 Primary Language Not Used
16 80 CE O 00119 Marital Status Not Used
17 80 CE O 00120 Religion Not Used
18 20 CX O 00121 Patient Account Number Study.AccountNumber
19 16 ST O 00122 SSN Number - Patient Patient.SSN
20 25 DLN O 00123 Driver's License Number
- Patient
21 20 CX O 00124 Mother's Identifier Not Used
22 80 CE O 00125 Ethnic Group Not Used
23 60 ST O 00126 Birth Place Not Used
Patient.PhoneWork
Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 3
GEHC
SEQ LEN DT
24 1 ID O 00127 Multiple Birth Indicator Not Used
25 2 NM O 00128 Birth Order Not Used
26 80 CE O 00129 Citizenship Not Used
27 60 CE O 00130 Veterans Military Status Not Used
28 80 CE O 00739 Nationality Not Used
Required
Field
ITEM# ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
29 26 TS O 00740 Patient Death Date and
Time
30 1 ID O 00741 Patient Death Indicator Not Used
Not Used
Example PID Segment:
PID|||20021986||Hensley^Sonia^D||19650514^37.05^Years|F||Caucasian|241 Kentucky
ST^^Kingston^TX^ 77379^US||281-555-1212|832-496-1212||||66778899|222-33-4444||||||||||||
Observation Request Segment (OBR)
The OBR is the report header for clinical data reporting. It identifies the observation set
represented by the atomic observations that follow. The following table describes the OBR
attributes:
GEHC
SEQ LEN DT
1 4 SI O 00237 Set ID OBR Begin sequential counter
2 22 EI O 00216 Placer Order Number Ord.PlacerOrderNum
3 80 EI R 00217 Filler Order Number Study.StudyUID_CaseDP
Required
Field
ITEM # ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
4.1 4 CE C 00238 Event Phase Number Sequential phase counter - start w/ 0
4.2 128 CE R 00238 Event Phase Name EventPhase.Name
4.3 16 CE O 00238 Phase Datapoint Datapoint for phases
4.4 16 CE R 00238 Universal Svc ID HIS.UsvcID_Code
4.5 64 CE R 00238 Universal Svc ID
Description
4.6 16 CE R 00238 Universal Svc ID Coding
System
5 2 ID O 00239 Priority Not Used
6 26 TS O 00240 Requested Date/Time Not Used
7 26 TS R 00241 Observation Date/Time # Study.StartTime (formatted as
HIS.UsvcID_Desc
HIS.UsvcID_CodeSys
YYYYMMDDHHMM)
4 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
SEQ LEN DT
GEHC
Required
Field
ITEM # ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
8 26 TS O 00242 Observation End Date/
Time #
EventBlock1.EventComment for "Case End"
logic if present (formatted as
YYYYMMDDHHMM)
9 20 CQ O 00243 Collection Volume Not Used
10 60 XCN O 00244 Collector Identifier Not Used
11 1 ID O 00245 Specimen Action Code Not Used
12 60 CE O 00246 Danger Code Not Used
13 300 ST O 00247 Relevant Clinical Info. Not Used
14 26 TS O 00248 Specimen Received
Not Used
Date/Time
15 300 CM O 00249 Specimen Source Not Used
16 80 XCN O 00226 Ordering Provider Not Used
17 40 XTN O 00250 Order Callback Phone
Not Used
Number
18 60 ST O 00251 Placer Field 1 HIS.PlacerField1 (Used by some sites as the
“Accession Number”)
19 60 ST O 00252 Placer Field 2 Not Used
20.1 128 ST O 00253 Registry/Group Node
Name
20.2 40 ST O 00253 Registry/Group Node
GUID
20.3 16 ST O 00253 Registry/Group Node
version
20.4 128 ST O 00253 Registry/Group Node
Comments
Registry/Cust Form Node Name
Registry/Cust Form Node GUID
Registry/Cust Form Node Version
Registry/Cust Form Node Comments
1
2
3
4
21 60 ST O 00254 Filler Field 2 Not Used
22 26 TS O 00255 Results Rpt/Status Chng
Not Used
- Date/Time
23 40 CM O 00256 Charge to Practice Not Used
24 10 ID R 00257 Diagnostic Service Sect ID"CTH" or "EC"
25 1 ID C 00258 Result Status "F"
26 400 CM O 00259 Parent Result Not Used
27 200 TQ O 00221 Quantity/Timing Not Used
28 150 XCN O 00260 Result Copies To Not Used
29 200 CM O 00261 Parent Not Used
30 20 ID O 00262 Transportation Mode Not Used
31 300 CE O 00263 Reason for Study Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 5
SEQ LEN DT
GEHC
Required
Field
ITEM # ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
32 200 CM O 00264 Principal Result
Interpreter
33 200 CM O 00265 Assistant Result
Interpreter
34 200 CM O 00266 Technician Not Used
35 200 CM O 00267 Transcriptionist Not Used
36 26 TS O 00268 Scheduled Date/Time Not Used
37 4 NM O 01028 Number of Sample
Containers
38 60 CE O 01029 Transport Logistics of
Collected Sample
39 200 CE O 01030 Collector's Comment Not Used
40 60 CE O 01031 Transport Arrangement
Responsibility
41 30 ID O 01032 Transport Arranged Not Used
42 1 ID O 01033 Escort Required Not Used
43 200 CE O 01034 Planned Patient
Transport Comment
44 80 CE O 00393 Procedure Code Not Used
|EventStaff.SSN_ID^EventStaff.LastName^Event
Staff.FirstName^EventStaff.MiddleName^StaffTitl
e.Name| where this is the primary physician.
Not Used
Not Used
Not Used
Not Used
Not Used
45 80 CE O 01316 Procedure Code Modifier Not Used
1
This field will store the Registry or custom form node name, like, "ACC 4.0 Forms" or "NASPE Forms" etc.
2
This field will store the Registry or custom form node GUID.
3
This field will store the Registry form node's version, which is in the format major.minor. For example, “4.0000” or “5.0012”. The minor
version will have a precision of 4 points.
4
This field will store the Registry form node's comments, like, “This is version 4.0 release for ACC” or “This is version 4.2 release for ACC”.
Example OBR Segment:
OBR|1||e7c61043-6f7a-11d6-904f-009027f688a1_68909|&-1^Patient
Demographics^^35400^Angioscopy^ANGIO|||
20020524203534|20020524204336||||||||||M2421||||||CTH|F|||||||4777^Stramblow^Bruce^L^M.D.|||
|||||||||||
6 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Observation Result Segment (OBX)
The OBX segment is used to transmit a single observation or observation fragment. It represents
the smallest indivisible unit of a clinical observation or report. The following table describes the
OBX attributes:
GEHC
SEQ LEN DT
1 4 SI O 00569 Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R 00570 Value Type always "ST"
3 200 CE R 00571 Observation Identifier See Mac-Lab/CardioLab Observation ID Tables
4 20 ST C 00572 Observation Sub-ID Not Used
65536
1
5
6 60 CE O 00574 Units Set as the appropriate unit under which OBX.5
7 60 ST O 00575 References Range Not Used
8 5 ID O 00576 Abnormal Flags Not Used
9 5 NM O 00577 Probability Not Used
10 2 ID O 00578 Nature of Abnormal Test Not Used
11 1 ID R 00579 Observation Result Status
Required
Field
* C 00573 Observation Value Set as the observed result or reporting
ITEM# ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
structure
was measured.
“F”
2
3
12 26 TS O 00580 Date Last Obs Normal Values Not Used
13 20 ST O 00581 User Defined Access Checks Not Used
14 26 TS R 00582 Date/Time of the Observation Date and Time formatted as
YYYYMMDDHHMMSS
15 60 CE O 00583 Producer's ID Not Used
16 80 XCN O 00584 Responsible Observer Not Used
17 60 CE O 00936 Observation Method Not Used
1
The length of the observation value field is variable, depending upon value type. See OBX-2-value type.
2
Reporting structures are multi-component observations using component separators in a defined way. See Reporting Structures later in this
document.
3
OBX.11 Observation Result Status: This flag shall be used to signify what action the processing status of this OBX record. Specifically,
whether this record is complete (“F” for Final), is an edit/correction of an existing record (“C” for Correction), or if this record should be deleted
(“D” for Delete) and removed from the receiving system. The default value shall be “F”. This should be considered as future GE Healthcare
functionality that does not yet exist in the feeder/creation system, but that should be planned/designed for in future receiving systems.
Example OBX Segment:
OBX|2|ST|PT-WT-KG||49.00|Kg|||||F|||20020524203534||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 7
Common Order Segment (ORC)
The ORC segment is used to transmit information common to all orders. Each case from a MacLab/CardioLab study will be treated as a separate order. The following table describes the ORC
attributes:
GEHC
SEQ LEN DT
1 2 ID R 00215 Order Control Always "RE"
2 22 EI C 00216 Placer Order Number Not Used
3 80 EI C 00217 Filler Order Number Study.StudyUID_CaseDP
4 22 EI O 00218 Placer Group Number Not Used
5 2 ID O 00219 Order Status Not Used
6 1 ID O 00220 Response Flag Not Used
7.1 32 CQ O 00221 Quantity Not Used
7.2 32 CM O 00221 Interval Not Used
7.3 32 ST O 00221 Duration Not Used
7.4 26 TS O 00221 Start Time EventBlock1.EventTime formatted as
Required
Field
ITEM# ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
YYYYMMDDHHMMSS
7.5 26 TS O 00221 Stop Time EventBlock1.EventTime formatted as
8 200 CM O 00222 Parent Not Used
9 26 TS O 00223 Date/Time of Transaction EventBlock1.EventTime formatted as
10 120 XCN O 00224 Entered By Not Used
11 120 XCN O 00225 Verified By Not Used
12 120 XCN O 00226 Ordering Provider |EventStaff.SSN_ID^EventStaff.LastName^Event
13 80 PL O 00227 Enterer's Location Not Used
14 40 XTN O 00228 Call Back Phone
Number
15 26 TS O 00229 Order Effective Date/
Time
16 200 CE O 00230 Order Control Code
Reason
17 60 CE O 00231 Entering Organization Not Used
18 60 CE O 00232 Entering Device Not Used
YYYYMMDDHHMMSS
YYYYMMDDHHMMSS
Staff.FirstName^EventStaff.MiddleName^^^^Eve
ntStaff.StaffID| where this is the primary
physician.
Not Used
Not Used
Case Type
19 120 XCN O 00233 Action By Not Used
8 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Example ORC Segment:
ORC|RE||e7c61043-6f7a-11d6-904f009027f688a1_68909||||^^^20020524203534^20020524215534||20020524203534|||321444777
^Stramblow^Bruce^L^^^^3214455||||diagnostic cath||||
Sequencing Rules
The OBR shall serve as the header segment for a particular observation set or phase.
Individual OBX segments shall serve as a single observation, observation fragment or a
reporting structure for similar groupings of observations (e.g. pressure measurement).
The OBX represents the smallest indivisible unit of a case report. For the purposes of the GE
Healthcare Mac-Lab/CardioLab data export, the OBR segment shall represent the separate
phases (GE Healthcare customers may use the term ‘condition’).
Examples of customer-named phases might include:
Baseline
Room Air
100% O2
PTCA
Intervention
Phases may be iterative. Each OBX segment, which is subordinate to an OBR, will represent
the clinical events that occurred under and during that particular phase.
OBX’s shall represent the events such as vital signs, administered drugs, pressures, valve
information, supplies, etc. that occurred while in that protocol.
Within a phase, events (OBX’s) shall be in chronological order.
NOTE: Phase naming conventions are user-defined within the Mac-Lab/CardioLab application.
Coordination may be necessary between GE Healthcare customer/clinical support and
the client to standardize their phase names and the conditions for their site’s use as
much as possible. This will ensure best alignment of the client’s data.
Segments Hierarchy
An example hierarchy of segments that creates a pseudo HL7 message follows:
NOTE: The OBR sequence numbers may change in the actual export, depending on:
Number of registry form nodes available in the study
Number of phases available
Number of case type available, etc.
MSH – Identifies this message as unique
PID – Uniquely identifies the patient
OBR “1” – Introduces atomic clinical events which fall under the “Patient Demographics” phase
OBX “1” – PT-HT-CM - 158 cm
OBX “2” – PT-WT-KG – 48 kg
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 9
OBX “3” – PT-BSA – 1.6 meters squared
OBX “4” – PT-SEX - M
OBX “5” – PT-AGE – 72 years
OBR “2”– Introduces atomic clinical events which fall under the “Case Demographics” phase
OBX “1” – STUDYUID – 234c3-199288de5-91184ca-3a9023cf
OBX “2” – HOSPINFO-NAME – Great Medical Center
OBX “3” – TIME-CASETART - 20010307081524
OBX “4” – TIME-CASEEND – 20010307085618
OBX “5” – CVIS-MODE – 1
OBR “3” – Introduces atomic clinical events which fall under the “Event Log” phase
OBX “1” – EVENT - Pt. prepped and draped in usual manner
OBX “2” – EVENT - Pt. informed consent signed
OBR “4” – Introduces atomic clinical events which fall under the “Conscious Sedation” phase
OBX “1” – CS-ACT-PRE -
OBR “5” – Introduces atomic clinical events which fall under the “Xray Summary” phase
OBX “1” – XRAY-FLTIME – 12 min
OBX “2” – XRAY-FLDOSE - 150 cc
OBR “6” – Introduces atomic clinical events which fall under the “DICOM” phase
OBX “1” – DICOM-MWLINFO
OBX “2” – DICOM-RUN
OBX “3” – DICOM-RUN
OBR “7” – Introduces atomic clinical events which fall under the “Reports” phase
OBX “1” – <report1 title> - <report1 name>
OBX “2” – <report2 title> - <report2 name>
OBR “8” – Introduces atomic clinical events which fall under the “Custom Fields” phase
OBX “1” – Custom Field 1
OBX “2” – Custom Field 2
OBR “9” – Introduces atomic clinical events which fall under the “Registry_Fields” phase
OBX “1” – *Admission Status
OBX “2” – Date of Labworks
OBR “10” – Introduces atomic file attachment events which fall under the “Attachments” phase
OBX “1” – <file info 1>
10 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
OBX “2” – <file info 2>
ORC – Diagnostic Cath
OBR “11” – Introduces atomic clinical events which fall under the “Baseline” phase
OBX “1” – Event_Vitals
OBX “2” – Event_Medication
OBX “3” – Event_Inventory
OBX “4” – Event_Procedure – Left Heart Cath
OBX “5” – Event_Personnel – 2633^SMITH^JOHN^WILEY^MD^FELLOW^20010307081722^
OBX “6” – Event_O2SAT – PV^0^96^14.5^^70
OBX “7” – HEMOMEAS_GENERAL - PVR/SVR Ratio^0^2.7^
OBX “8” – Event_COMPLICATION – Tamponade
OBR “12” – Introduces atomic clinical events which fall under the “100% O2” phase
OBX “1” – Event_Vitals -
OBX “2” – Event_Medication -
OBX “3” – Event_Inventory -
OBX “4” – Event_Procedure – AngioGram
OBX “5” – Event_O2SAT – RA^0^97^14.5^^68
OBX “7” – HEMOMEAS_PRESSURE - AO^1^175^mmHg^72^mmHg^110^mmHg^68^BPM
OBX “8” – EVENT_COMPLICATION – CHF
ORC – Interventional Cath
OBR “13” – Introduces atomic clinical events which fall under the “Baseline” phase
OBX “1” – Event_Vitals
OBX “2” – Event_Medication
OBX “3” – Event_Inventory
OBX “4” – Event_Procedure – Left Heart Cath
OBX “5” – Event_Personnel – 2633^SMITH^JOHN^WILEY^MD^FELLOW^20010307081722^
OBX “6” – Event_O2SAT – PV^0^96^14.5^^70
OBX “7” – HEMOMEAS_GENERAL - PVR/SVR Ratio^0^2.7^
OBX “8” – Event_COMPLICATION – Tamponade
OBX “9” – Event_Intervention_Lesion
OBX “10” – Event_Intervention_Treatment
OBX “11” – Event_Intervention_Supply
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 11
OBX “12” – Event_Intervention_Attempt – Attempt 1
OBX “13” – Event_Intervention_Attempt – Attempt 2
OBX “14” – Event_Intervention_Attempt – Attempt 3
OBR “14” – Introduces atomic clinical events which fall under the “100% O2” phase
OBX “1” – Event_Vitals -
OBX “2” – Event_Medication -
OBX “3” – Event_Inventory -
OBX “4” – Event_Procedure – AngioGram
OBX “5” – Event_O2SAT – RA^0^97^14.5^^68
OBX “7” – HEMOMEAS_PRESSURE -AO^1^175^mmHg^72^mmHg^110^mmHg^68^BPM
OBX “8” – EVENT_COMPLICATION – CHF
Mac-Lab/CardioLab Observation Identifier Tables
Static OBR Groupings
Aside from the OBR groupings of OBX segments that are protocol-based, there are special, onetime occurring OBR groups in each message. These OBR groupings will each occur exactly once
per HL7 message (file) in this prescribed order. The special, static OBR segments will not value
subcomponent Event Phase Number.
Patient Demographics OBX Segments (“Patient Demographics”)
OBSERVATION
IDENTIFIER
PT-HT-CM Patient Height in centimeters cm (centimeter) DemographicResults.value where indexNum = 1
PT-WT-KG Patient Weight in kilograms kg (kilogram) DemographicResults.value where indexNum = 2
PT-BSA Patient Body Surface Area m2 (meter2) DemographicResults.value where indexNum = 3
PT-SEX Patient Sex none DemographicResults.value where indexNum = 5 (“M”, “F”,
PT-AGE Patient's Age at time of study SystemUnitType
Example OBX Segment:
OBX|2|ST|PT-WT-KG||49.00|Kg|||||F|||20020524203534||||
DESCRIPTION UNITS MAC-LAB/CARDIOLAB FIELD
or “U”)
DemographicResults.value where indexNum = 4
value
12 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Case Demographics OBX Segments (“Case Demographics”)
Includes Event_CB_Note + several Non-events.
OBSERVATION
IDENTIFIER
STUDYUID Globally unique identifier for a study none Study.StudyUID (a GUID)
HOSPINFO-
NAME
HOSPINFO-
LOC
HOSPINFO-
ADDR
HOSPINFO-
CITY
HOSPINFO-ST Hospital State or Province none HospitalInfo.State
HOSPINFO-ZIP Hospital Zip none HospitalInfo.Zipcode
HOSPINFO-
CNTRY
TIME-
CASESTART
TIME-
CASEEND
TIME-
PTARRIVES
Hospital Name none HospitalInfo.Name
Hospital Location none HospitalInfo.Location
Hospital Address none HospitalInfo.Address
Hospital City none HospitalInfo.City
Hospital Country none Country.Abbrev (ISO 2-character abbreviation)
Time Case Starts YYYYMMDDHHMMStudy.StartTime
Time Case Ends YYYYMMDDHHMMEventBlock1.EventComment when EventTypeID =
Time Patient Arrives YYYYMMDDHHMMEventBlock1.EventComment when EventTypeID =
DESCRIPTION UNITS MAC-LAB/CARDIOLAB FIELD
Event_CB_Note (0x0601) and string = "CASE END"
Event_CB_Note (0x0601) and string = "PATIENT
ARRIVES"
TIME-
PTONTABLE
TIME-
PTSENTTO
TIME-
MDCALLED
TIME-
MDARRIVE
STUDYBINLOC UNC Path to the study binary data
STUDYNUM Study Number None Study.STUDYNUMBER
CVIS-MODE Event list Ids are CVIS generated 1 = CVIS Mode
Time Patient on Table YYYYMMDDHHMMEventBlock1.EventComment when EventTypeID =
Event_CB_Note (0x0601) and string = "PATIENT ON
TABLE"
Time Patient Sent To YYYYMMDDHHMMEventBlock1.EventComment when EventTypeID =
Event_CB_Note (0x0601) and string = "PATIENT
TRANSFERRED TO"
Time Physician Called YYYYMMDDHHMMEventBlock1.EventComment when EventTypeID =
Event_CB_Note (0x0601) and string = "PHYSICIAN
CALLED"
Time Physician Arrived YYYYMMDDHHMMEventBlock1.EventComment when EventTypeID =
Event_CB_Note (0x0601) and string = "PHYSICIAN
ARRIVED"
none OnlineStudyPath using UNC notation
files
CVIS option
0 = No CVIS
Mode
Example OBX Segment:
OBX|9|ST| TIME-PTARRIVES ||||||||F|||20020524203534||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 13
Event Log Events OBX Segments (“Event Log”)
OBSERVATION
IDENTIFIER
EVENT Event Log Window Event none EventBlock1.Summary when
DESCRIPTION UNITS MAC-LAB/CARDIOLAB FIELD
EventBlock1.IncludeInReport = True. Insert
EventBlock1.EventTime in OBX.14 (Observation DT/TM)
Example OBX Segment:
OBX|3|ST|EVENT||Procedure: Left Heart Cath||||||F|||20020524204140||||
Conscious Sedation OBX Segments (“Conscious Sedation”)
OBSERVATION
IDENTIFIER
CS-ACT-PRE Activity - Precase ActivityList.Activity resolved from
CS-ACT-POST Activity - Postcase ActivityList.Activity resolved from
CS-RESP-PRE Respiration - Precase RespirationList.Respiration resolved from
DESCRIPTION UNITS MAC-LAB/CARDIOLAB FIELD
ConsciousSedation.ActivityID where
ConsciousSedation.PrePost = 1
ConsciousSedation.ActivityID where
ConsciousSedation.PrePost = 0
ConsciousSedation.RespirationID where
ConsciousSedation.PrePost = 1
CS-RESP-POST Respiration - Postcase RespirationList.Respiration resolved from
ConsciousSedation.RespirationID where
ConsciousSedation.PrePost = 0
CS-CIRC-PRE Circulation - Precase CirculationList.Circulation resolved from
CS-CIRC-POST Circulation - Postcase CirculationList.Circulation resolved from
CS-CONS-PRE Consciousness - Precase ConsciousnessList.Consciousness resolved from
CS-CONS-POST Consciousness - Postcase ConsciousnessList.Consciousness resolved from
CS-O2SAT-PRE O2 Saturation - Precase O2SaturationList.O2Saturation resolved from
CS-O2SAT-POST O2 Saturation - Postcase O2SaturationList.O2Saturation resolved from
CS-SCORE-PRE Score - Precase ConsciousSedation.Score where
ConsciousSedation.CirculationID where
ConsciousSedation.PrePost = 1
ConsciousSedation.CirculationID where
ConsciousSedation.PrePost = 0
ConsciousSedation.ConsciousnessID where
ConsciousSedation.PrePost = 1
ConsciousSedation.ConsciousnessID where
ConsciousSedation.PrePost = 0
ConsciousSedation.O2SaturationType where
ConsciousSedation.PrePost = 1
ConsciousSedation.O2SaturationType where
ConsciousSedation.PrePost = 0
ConsciousSedation.PrePost = 1
14 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
OBSERVATION
IDENTIFIER
CS-SCORE-POST Score - Postcase ConsciousSedation.Score where
DESCRIPTION UNITS MAC-LAB/CARDIOLAB FIELD
ConsciousSedation.PrePost = 0
CS-COMMENT-PRE Comment - Precase EventBlock1.EventComment where
CS-COMMENT-POST Comment - Postcase EventBlock1.EventComment where
ConsciousSedation.PrePost = 1
ConsciousSedation.PrePost = 0
Example OBX Segment:
OBX|11|ST|CS-SCORE-PRE||9||||||F|||20011003144143||||
Xray Summary OBX Segments (“Xray Summary”)
OBSERVATION
IDENTIFIER
XRAY-FLTIME Fluoro Time min (minutes) XraySummary.FluoroTime (formatted as "x.x")
XRAY-FLDOSE Fluoro Dose
XRAY-CINEDOSE Cine Dose
XRAY-TOTDOSE Total Dose
XRAY- CINEFRAME Cine Frames none XraySummary.CineFrames
XRAY- TOTALRUNS Total Runs none XraySummary.TotalRuns
DESCRIPTION UNITS MAC-LAB/CARDIOLAB FIELD
cGycm
cGycm
cGycm
2
2
2
XraySummary.FluoroDose
XraySummary.CineDose
XraySummary.TotalDose
XRAY- CDID CD ID none XraySummary.CDID
XRAY- CINEID Cine ID none XraySummary.CineID
XRAY- ARCHIVEID Archive ID none XraySummary.ArchiveID
XRAY- INTFLTIME Interventional Total Fluoro Time none XraySummary.intvfluorotime
XRAY- DIAGFLTIME Diagnostic Total Fluoro Time none XraySummary. diagfluorotime
Example OBX Segment:
OBX|1|ST|XRAY-FLTIME||10.00|min|||||F|||20011003144143||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 15
Event_DICOM_RunInfo OBX (Reporting Structure)
One OBX created for each run.
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_DICOM_RunInfo"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 5 ST O Run Number RunInfo.RunNumber (smallint)
5.2 26 ST O Run Time RunInfo RunTime formatted as HHMMSS
5.3 5 ST O Number of Frames RunInfo NumberFrames (smallint)
5.4 8 ST O Frames per Second RunInfo FramesPerSecond (float)
5.5 5 ST O Plane RunInfo Plane (smallint)
5.6 8 ST O kV RunInfo kV (float)
5.7 10 ST O mA RunInfo mA (int)
5.8 10 ST O Mas RunInfo Mas (int)
5.9 10 ST O mS RunInfo ms (int)
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.10 5 ST O Angulation RunInfo Angulation (smallint)
5.11 5 ST O Rotation RunInfo Rotation (smallint)
5.12 5 ST O Focal Distance RunInfo FocalDistance (smallint)
5.13 5 ST O Image Intensifier Mode RunInfo ImageIntensifierMode (smallint)
5.14 64 ST O Sequence Name RunInfo SequenceName
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Study.StartTime formatted as YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
Example OBX Segment:
OBX|8|ST| Event_DICOM_RunInfo ||1^102302^112^32.24^1^24.6^35^42^67^12^24^6^8^Run
number 1||||||F|||20011003144143||||
DICOM Other Information OBX
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_DICOM_Study"
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 64 ST O Study Instance UID Study.DICOMStudyInstanceUID
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Example OBX Segment:
OBX|8|ST| Event_DICOM_Study ||1.2.42567.56798||||||F|||20011003144143||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 17
Reports OBX Segments (“Reports”)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "REPORTS"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 64 ST O Report Name ReportInfo.filename
5.2 64 ST O Report Title ReportInfo.description
5.3 64 ST O Report Time ReportInfo.datecreated
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Study.StartTime formatted as YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Example OBX Segment:
OBX|12|ST| REPORTS || 5e9807dd-746b-43a7-91eefb4bc52ade2940565.8994212963_20972767375757098.doc^EP^20011003144400||||||F|||20011
003144143||||
18 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Custom Fields OBX Segment (“Custom Fields”)
One/more OBX per custom field.
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Custom_Field^FieldID^Field_Name where
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 3000 ST O Field Value CustFormStudyData.valuetext
5.2 36 ST O Field Group ID CustForm.Formno (GUID)
5.3 64 ST O Field Group CustFormStudyData.formname
5.4 3 ST O Value ID CustFormStudyData.Valueid (tinyint) Will be non-
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
Custom_Field is static text always the same FieldID
= CustFormStudyData.labelno (GUID) Field_Name =
CustFormStudyData.labelname
zero for selection based fields only
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Study.StartTime formatted as YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
If the custom field contains a date value it will be of the format YYYYMMDD and a time value will
be of format HHMMSS.
Example OBX Segment:
OBX|1|ST|Custom_Field^f2c30aa2-5ae8-11d7-9068-0010f3030333^Room Number ||243^
E5B36BAC-CA33-47D4-B407-9D43161C8888^Additional Information ||||||F|||20011003144143||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 19
Attachment OBX Segments (“Attachments”)
The Attachments OBR shall be available in the EP and Cath HL7 export. The Attachments
segment shall be the last singleton segment in the HL7 export. The OBX shall contain the
information about the file that is exported along with the study. Number of OBXs under the
singleton OBR will depend on the number of files exported with the study. The SEQ 5.4 Title shall
contain the title of images (Single frame & multi-frame) and report documents.
GEHC
SEQ LEN DT
1 4 SI O Set ID – OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always “ST”
3 200 CE R Observation Identifier Always “ATTACHMENT”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or file information
5.1 64 ST O File Name FileInfo.filename
5.2 64 ST O File Creation Time Not Used
5.3 10 NM C File Size FileInfo.filesize in bytes
5.4 64 ST C Title ReportInfo.Description or Image.Title or
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
Snapshot.Title
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status “F” always
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Date and Time of Data Export
15 60 CE O Producer’s ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
A date value it will be of the format YYYYMMDD and a time value will be of format HHMMSS.
Example OBX Segment:
OBX|1|ST|ATTACHMENT||2788209c-8c2d-41eb-954299375e1486fa38765.5265277778_10485777429162860.doc^^38
24128^Images||||||F|||20110318114020||||
OBX|2|ST|ATTACHMENT||2788209c-8c2d-41eb-954299375e1486faImage0001.AVI^^500224^Imported Image Loop 1||||||F|||20110318114020||||
20 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
OBX|3|ST|ATTACHMENT||2788209c-8c2d-41eb-954299375e1486faImage0002.wmv^^485972^Image Loop 1||||||F|||20110318114020||||
Dynamic OBR Groupings
These are the OBR, which will be available in the HL7 if the registry forms are present on the
system. If no registry forms are present on the system, then the HL7 export will not have these
OBRs. Multiple OBR for registry forms based will be on Number of registry form nodes available
in the system.
Number of OBXs in each OBR will depend on the number of registry fields used in the study
Registry Fields OBX Segments
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Registry_Field^FieldID^Field_name where
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
Registry_Field is static text always the same FieldID
= CUSTFormStudydata.labelno (GUID) Field_name
= CUSTFormStudydata.labelname
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST O Field Value CUSTFormStudydata.valuetext
5.2 36 ST O Field Group ID CUSTForms.Formno
5.3 64 ST O Field Group (See Field Group Tables) CUSTForms.formname
5.4 3 ST O Value ID CUSTFormStudydata.Valueid (tinyint) Will be nonzero for selection based fields only
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Study.StartTime formatted as YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 21
If the registry field contains a date value it will be of the format YYYYMMDD and a time value will
be of format HHMMSS.
Example OBX Segment:
OBX|1|ST|Registry_Field^55716D66-CB65-41C8-BC95-671C427CA473^*Admission
Status:||Outpatient Referal ^ E5B36BAC-CA33-47D4-B4079D43161C7777^Admission^3||||||F|||20011003144143||||
Data Tables - Common / Raw Events Reporting Structures
These are the administrative observations and raw clinical values that are captured during the
course of a study. During any hemodynamic phase, a number of similar measurements can be
made (for example, multiple LV measurements). Generally, only one of these events is used to
compute hemodynamic analysis for that phase. The user, in order to synchronize for inclusion in
the study report, can select those events being reported here. A select, unique, phase-based
subset of these raw clinical events is reported in the Hemodynamic Measurement structures later
in this document. The values included in those Hemodynamic Measurement structures are
synchronized with other values in that section (e.g. a particular cardiac output reported in the
General Structure is the same one used in the Valve Pressure Structure irrespective of the raw
multiple cardiac outputs recorded in a hemodynamic phase) for clinical significance. The raw
Event-based clinical observations & measurements reported in this section should not be used
for hemodynamic analysis.
Event_Medication OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Medication"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST R Medication Description EventMedication.Medication
5.2 8 ST R Medication Amount EventMedication.Amount (real, TJC (JCAHO)
5.3 32 ST R Medication Route MedicationRoute.Name resolved from
5.4 26 ST O Medication Stop Time EventMedication.StopTime formatted as
5.5 128 ST O Medication ACC Code Type EventMedication.ACCCodeType
5.6 32 ST O Medication Billing Code MedicationSummary.BillingCode
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
formatted)
EventMedication.MedicationRouteType
YYYYMMDDHHMMSS
5.7 10 ST O DMS ID EventMedication.medicationID (int)
5.8 10 ST O Medications Datapoint EventMedication.eventdatapoint (int)
5.9 10 ST O Ordered by Staff DMS ID EventStaff.staffID
22 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
5.10 10 ST O Ordered by Staff Datapoint EventStaff.eventdatapoint
5.11 10 ST O Given by Staff DMS ID EventStaff.staffID
5.12 10 ST O Given by Staff Datapoint EventStaff.eventdatapoint
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
6 60 CE R Units SystemUnit.Name resolved from
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Medication Administration or Start
Time)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventMedication.SystemUnitType
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
* Medication Stop Time. If the medication being administered is a drip, the Start Time shall be
denoted using OBX.14 and Stop Time shall be denoted using OBX.5.4. For a single point-in-time
dose (like Valium - IM), the Medication Stop Time shall be null.
Example Medications OBX:
OBX|14|ST|Event_Medication||Ibuprofen^800^PO^^OTHER^389912^222^12322^100^1^121^2|
mg|||||F|||20010307082033||||
NOTE: Medication amounts are formatted per TJC (JCAHO):
Trailing zeros will not be included.
A leading zero before the decimal point will be included.
Event_Inventory OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Inventory"
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 23
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
GEHC
SEQ LEN DT
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST R Inventory Item Description EventSupply.SupplyName
5.2 32 ST O Inventory Item Part Number EventSupply.PartNum
5.3 32 ST O Inventory Item Size EventSupply.Size
5.4 32 ST O Inventory Manufacturer EventSupply.Manufacturer
5.5 64 ST O Inventory Item Serial Number EventSupply.SerialNum
5.6 64 ST O Inventory Item Lot Number EventSupply.LotNum
5.7 64 ST O Inventory Barcode EventSupply.Barcode
5.8 128 ST O Inventory Folder (Category) EventSupply.FolderName
5.9 10 ST R Inventory DataPoint EventSupply.DataPoint (int)
5.10 32 ST O Inventory APC/Billing Code EventSupply.APCCode
5.11 10 ST O DMS ID EventSupply.SupplyID (int)
6 60 CE O Units Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Inventory Consumption)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example Inventory OBX:
OBX|23|ST|Event_Inventory||Quantum Ranger^1479035^7FR^Cordis^SN111^LN222^+H74914790351M^
Balloons^58772^APC488812^333||||||F|||20010307082453||||
24 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Event_Complication OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Complication"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST O Complication description EventComplication.ComplicationName
5.2 10 ST O DMS ID EventComplication.Complicationid (int)
5.3 10 ST O Complication datapoint EventComplication. Eventdatapoint (int)
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Complication)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example Complication OBX:
OBX|72|ST|Event_Complication||Tamponade^333^12322||||||F|||20010307083418||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 25
Event_Procedure OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Procedure"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST O Procedure Description EventProcedure.Procedure
5.2 128 ST O Procedure Code 1 EventProcedure.CPTCode
5.3 128 ST O Procedure Code 2 EventProcedure.Code1
5.4 128 ST O Procedure Code 3 EventProcedure.Code2
5.5 128 ST O Procedure Code 4 EventProcedure.Code3
5.6 10 ST O DMS ID EventProcedure.procedureid (int)
5.7 10 ST O Procedure Datapoint EventProcedure.eventdatapoint (int)
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Procedure)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example Procedure OBX:
OBX|45|ST|Event_Procedure||Left Heart
Cath^CPT9322105^111^222^334^333^12322||||||F|||20010307082141||||
26 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Event_Personnel OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Personnel"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 24 ST O Personnel ID EventStaff.SSN_ID
5.2 32 ST O Personnel Last Name EventStaff.LastName
5.3 32 ST O Personnel First Name EventStaff.FirstName
5.4 16 ST O Personnel Middle Name EventStaff.MiddleName
5.5 32 ST O Personnel Title StaffTitle.Name resolved from EventStaff.Title
5.6 32 ST O Personnel Duty StaffDuty.Name resolved from EventStaff.StaffDuty
5.7 26 ST O Personnel Time In EventBlock1.EventTime (formatted as
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
YYYYMMDDHHMMSS)
5.8 26 ST O Personnel Time Out EventStaff.TimeOut (formatted as
YYYYMMDDHHMMSS)
5.9 10 ST O DMS ID EventStaff.staffed (int)
5.10 10 ST O Staff datapoint EventStaff.eventdatapoint (int)
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Personnel Start Time)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example Personnel OBX:
OBX|9|ST|Event_Personnel||2354^SMITH JOHN
W^MD^FELLOW^20010307081541^20010307085141^333^12322||| |||F|||20010307082141||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 27
Event_Contrast OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Contrast"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST O Contrast Description EventContrast.ContrastName
5.2 8 ST O Contrast Amount EventContrast.Amount (real, TJC (JCAHO)
5.3 10 ST O DMS ID EventContrast.Contrastid (int)
5.4 10 ST O Contrast Datapoint EventContrast.eventdatapoint (int)
6 60 CE O Units Always "ml"
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
formatted)
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Contrast Administration Time)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example Contrast OBX:
OBX|22|ST|Event_Contrast||Omnipaque^85^333^12322|cc|||||F|||20010307082141||||
NOTE: Medication amounts are formatted per TJC (JCAHO):
Trailing zeros will not be included.
A leading zero before the decimal point will be included.
28 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Event_Intervention OBXs
There are several OBX types tied to an Intervention: Lesion, Treatment, Attempt, Intervention
Supply and Graft. Each Treatment OBX will also include information about the Lesion to which
the treatment is being applied. Each Attempt OBX will include information about the treatment to
which the attempt is being made and the lesion to which the treatment is being applied. Each
Intervention Supply OBX will include information about the treatment and lesion with which the
supply is being used.
Lesion OBX
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Intervention_Lesion"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.2 32 ST O Segment Name/Lesion Location INTRVlesion.SegmentID resolve for ACC NCDR
string name from VesselSegment.SegmentName
5.3 5 ST O Lesion Number INTRVlesion.LesionID (smallint)
5.4 16 ST O Lesion Type INTRVlesion.intrvtype resolved for string name
5.5 3 ST O Initial Stenosis INTRVlesion.inistenosis
5.6 3 ST O Residual Stenosis INTRVlesion.resistenosis
5.7 70 ST O Pre TIMI flow INTRVlesion.pretimiflow resolved for string name
5.8 70 ST O Post TIMI flow INTRVlesion.posttimiflow resolved for string name
5.9 32 ST O Previous Dilated Lesion INTRVlesion.prevdilatedlesion resolved for string
5.10 50 ST O In graft to Cited Segment INTRVlesion.basicsegtype resolved for string name
5.11 40 ST O Location in graft INTRVlesion.locid resolved for string name
5.12 32 ST O Lesion Risk INTRVlesion.lesionrisk resolved for string name
5.13 16 ST O Dissection in segment INTRVlesion.dissectinseg resolved for string name
5.14 16 ST O Acute closure INTRVlesion.acuteclosure resolved for string name
5.15 16 ST O Successful reopening INTRVlesion.succreopen resolved for string name
5.16 16 ST O Perforation INTRVlesion.perforation resolved for string name
5.17 32 ST O Graft Type INTRVgraft.grafttype resolved for string name
name
5.18 32 ST O Graft Target Segment INTRVlesion.segmented resolved for string name
5.19 32 ST O Graft Origin INTRVgraft.orig resolved for string name
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 29
GEHC
SEQ LEN DT
5.20 32 ST O Graft Destination INTRVgraft.dest resolved for string name
5.21 32 ST O Guidewire Success INTRVlesion.guidewire resolved for string name
5.22 40 ST O Peripheral Group INTRVlesion.INTRVPeripheralGroup
5.23 5 ST O Segment Name/Lesion Location ID INTRVlesion.SegmentID
5.24 5 ST O Lesion Type ID INTRVlesion.intrvtype
5.25 5 ST O Pre TIMI flow ID INTRVlesion.pretimiflow
5.26 5 ST O Post TIMI flow ID INTRVlesion.posttimiflow
5.27 5 ST O Previous Dilated Lesion ID INTRVlesion.prevdilatedlesion
5.28 5 ST O In graft to Cited Segment ID INTRVlesion.basicsegtype
5.29 5 ST O Location in graft ID INTRVlesion.locid
5.30 5 ST O Lesion Risk ID INTRVlesion.lesionrisk
5.31 5 ST O Dissection in segment ID INTRVlesion.dissectinseg
5.32 5 ST O Acute closure ID INTRVlesion.acuteclosure
5.33 5 ST O Successful reopening ID INTRVlesion.succreopen
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.34 5 ST O Perforation ID INTRVlesion.perforation
5.35 5 ST O Graft Type ID INTRVgraft.grafttype
5.36 5 ST O Graft Target Segment ID INTRVlesion.segmented
5.37 5 ST O Graft Origin ID INTRVgraft.orig
5.38 5 ST O Graft Destination ID INTRVgraft.dest
5.39 5 ST O Guidewire Success ID INTRVlesion.guidewire
5.40 5 ST O Peripheral Group ID INTRVlesion.INTRVPeripheralGroup
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
30 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Example Intervention Lesion OBX:
OBX|27|ST| Event_Intervention_Lesion ||1^pCirc^2^Coronary^75^10^26^11^Balloon^No^No
Graft^Low Risk^No^No^Not
Applicable^No^^^^^Successful^^2^1^1^1^1^1^0^2^0^0^0^0^0^0^0^0^2^0||||||F|||2001030708214
1||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 31
Treatment OBX
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Intervention_Treatment"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 5 ST O Lesion Number INTRVlesion.LesionID (smallint)
5.3 5 ST O Treatment Number INTRVtreatment.treatmentid (smallint)
5.4 32 ST O Treatment Type INTRVtreatment.treatmenttypeid resolved for string
5.5 16 ST O Primary Indication INTRVtreatment.primsec resolved for string name
5.6 16 ST O Success Indication INTRVtreatment.sucfail resolved for string name
5.7 16 ST O Treatment Type ID INTRVtreatment.treatmenttypeid
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
name
5.8 16 ST O Primary Indication ID INTRVtreatment.primsec
5.9 16 ST O Success Indication ID INTRVtreatment.sucfail
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Example Intervention Treatment OBX:
OBX|27|ST| Event_Intervention_Treatment
||1^2^1^Balloon^Yes^Yes^1^1^1||||||F|||20010307082141||||
32 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Attempt OBX
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Intervention_Attempt"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 5 ST O Lesion Number INTRVlesion.LesionID (smallint)
5.3 5 ST O Treatment Number INTRVtreatment.treatmentid (smallint)
5.4 5 ST O Attempt Number EVENTINTRVattempt.attemptid (smallint)
5.5 5 ST O Duration in Seconds EVENTINTRVattempt.Duration (smallint)
5.6 8 ST O Attribute1 EVENTINTRVattempt.Attemptattr1 (real)
5.7 8 ST O Attribute2 EVENTINTRVattempt.Attemptattr2 (real)
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.8 8 ST O Attribute3 EVENTINTRVattempt.Attemptattr3 (real)
5.9 8 ST O Attribute4 EVENTINTRVattempt.Attemptattr4 (real)
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Example Intervention Attempt OBX:
OBX|27|ST| Event_Intervention_ Attempt
||1^2^1^1^50^6.00^0.00^0.00^0.00||||||F|||20010307082141|||
The attribute fields above depend upon the treatment as described in the following table.
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 33
Treatment Attribute Table
Treatment Attribute 1 Attribute 2 Attribute 3 Attribute 4
Balloon Pressure in Atm
Cutting Balloon Pressure in Atm
Bare Metal Stent Pressure in Atm
DCA Pressure in Atm Cuts
Rotational Atherectomy Speed in Rpm Passes Burr size
Atherectomy Speed in Rpm Passes Burr size
AngioJet Volume in Volume out
TEC Volume out
Thrombectomy Volume out
Laser Passes
IVUS
Flowire
Pressure Wire
Sirolimus-Eluting Stent Pressure in Atm
Paclitaxel-Eluting Stent Pressure in Atm
Zotarolimus -Eluting Stent Pressure in Atm
Everolimus-Eluting Stent Pressure in Atm
Heparin Coated Stent Pressure in Atm
Covered Stent Pressure in Atm
Gamma Brachytherapy Dwell time (sec) Dose
Beta Brachytherapy Dwell time (sec) Dose
Distal Embolic Protection
Embolic
Other Device
34 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
Lesion Supply OBX
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Intervention_Supply"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 5 ST O Lesion Number INTRVlesion.LesionID (smallint)
5.3 5 ST O Treatment Number INTRVtreatment.treatmentid (smallint)
5.4 10 ST O Supply Used INTRVsupplyused.Supplyused (int)
5.5 32 ST O Supply Use (pre procedure or during
5.6 16 ST O Supply Use ID INTRVsupplyused.usage (tinyint)
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
INTRVsupplyused.usage (tinyint) resolved to string
procedure)
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Example Intervention Supply OBX:
OBX|27|ST| Event_Intervention_ Supply ||1^2^1^63188^Pre^1||||||F|||20010307082141|||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 35
Event_Vital OBX
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_Vitals"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 5 ST O SpO2 EventVital.SPO2 (smallint)
5.3 5 ST O Heart rate EventVital.HR (smallint)
5.4 8 ST O Systolic Pressure EventVital.SP (real)
5.5 8 ST O Diastolic Pressure EventVital.DP (real)
5.6 8 ST O Mean Pressure EventVital.Mean (real)
5.7 8 ST O Respiration Rate EventVital.RR (real)
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.8 11 ST O Respiration Rate Units Always "breaths/min"
5.9 8 ST O Temperature EventVital.Temp (real)
5.10 3 ST O Temperature Units Recorded as “cel” (degrees Celsius) if temperature
exists
5.11 64 ST O Level of Consciousness EventVital.Loc (smallint)
5.12 5 ST O Pressure Source EventVital.VitalPressureResource (smallint)
5.13 8 ST O Inspired EtCO2 EventVital.inspiredC02
5.14 8 ST O Expired EtCO2 EventVital.expiredC02
5.15 10 ST O Vitals Datapoint EventVital.eventdatapoint
5.16 4 ST O Inspired EtCO2 Unit Units for inspired CO2, always "mmHg"
5.17 4 ST O Expired EtCO2 Unit Units for expired CO2, always "mmHg"
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
36 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
SEQ LEN DT
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation (Vital
Signs Event Time)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example Vital OBX:
OBX|14|ST|Event_Vitals||0^97^71^130^50^89^22^breaths/min^37.1^cel^2 - Fully
Awake^10^12^112334||||||F|| |20010307082141||||
Data Tables - Cath Event Reporting Structures
Event_ThermoDilution_Cardiac_Output OBX (Reporting Structure)
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
3 200 CE R Observation Identifier Always "Event_TDCO"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 8 ST O Cardiac Output EventThermalDilutionCO.CO (real)
5.3 8 ST O Units L/min
5.4 8 ST O Catheter Size EventThermalDilutionCO.CatherSize (tinyint)
5.5 16 ST O Catheter Type EventThermalDilutionCO.CatherType (tinyint)
5.6 8 ST O Cal Factor EventThermalDilutionCO.CalFactor (smallint)
5.7 8 ST O Volume EventThermalDilutionCO.Volume (tinyint)
5.8 16 ST O Method EventThermalDilutionCO.Method (tinyint)
5.9 8 ST O Blood Temperature EventThermalDilutionCO.BloodTemp (real)
5.10 8 ST O Injected Temperature EventThermalDilutionCO.InjectedTemp (real)
5.11 5 ST O Heart Rate EventThermalDilutionCO.HR (smallint)
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 37
GEHC
SEQ LEN DT
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Example TDCO OBX:
OBX|31|ST|Event_TDCO||0^4.70^L/ml^7 F^Edwards^580^10cc^Bath^30.20^21.30^76||||||F|||
20010809122424||||
Event_Manual_Cardiac_Output OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_ManualCO"
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 8 ST O Cardiac Output EventManualCO.CO (real)
5.3 5 ST O Heart Rate EventtManualCO.HR (smallint)
6 60 CE O Units Always “L/min”
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
38 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Example ManualCO OBX:
OBX|22|ST|Event_ManualCO||1^4.65^72|l/min|||||F|||20010307081824|||
Event_PPG_Measurement OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "Event_PPGradient"
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 32 ST O Valve Type EventPPGMeasurement.ValveType resolve from
5.3 32 ST O Site Label Pressure 1 EventPPGMeasurement.SiteLabelPressure1
5.4 8 ST O Mean Systolic Pressure 1 EventPPGMeasurement.MeanSystolicValuePres1
5.5 5 ST O Heart Rate Pressure 1 EventPPGMeasurement.HRPressure1 (smallint)
5.6 16 ST O Label Group Pressure 1 EventPPGMeasurement.LabelGroup_P1 (resolve
5.7 32 ST O Site Label Pressure 2 EventPPGMeasurement.SiteLabelPressure2
5.8 8 ST O Mean Systolic Pressure 2 EventPPGMeasurement.MeanSystolicValuePres2
5.9 5 ST O Heart Rate Pressure 2 EventPPGMeasurement.HRPressure2 (smallint)
5.10 16 ST O Label Group Pressure 2 EventPPGMeasurement.LabelGroup_P2 (resolve
5.11 8 ST O Peak to Peak Gradient EventPPGMeasurement.MeanPtoPGradient (real)
ValveType.Name
(real)
for string)
(real)
for string)
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 39
GEHC
SEQ LEN DT
5.12 1 ST O Manually Edited Flag (0 or 1) EventPPGMeasurement.ManuallyEdited
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Example PPGradient OBX:
OBX|22|ST|Event_PPGradient||2^MITRAL^LV^118^73^^AO^126^73^^8^0||||||F|||200103070829
14
Event_Valve_Area_Analysis OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “Event_ValveAreaAnalysis”
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 32 ST O Value Type EventValveAreaAnalysis.ValveType resolve from
5.2 4 ST O Phase Phase Number
5.3 32 ST O Site Label Pressure 1 EventValveAreaAnalysis.SiteLabelPressure1
5.4 8 ST O Mean Pressure 1 EventValveAreaAnalysis.MeanPressure1 (real)
5.5 8 ST O Mean Diastolic Pressure 1 EventValveAreaAnalysis.MeanDiastolicValue1 (real)
ValveType.Name
40 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
SEQ LEN DT
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.6 8 ST O Mean Systolic Pressure 1 EventValveAreaAnalysis.MeanSystolicValuePres1
5.7 8 ST O Mean A Wave Pressure 1 EventValveAreaAnalysis.MeanAWaveValuePres1
5.8 8 ST O Mean V Wave Pressure 1 EventValveAreaAnalysis.MeanVWaveValuePres1
5.9 5 ST O Heart Rate Pressure 1 EventValveAreaAnalysis.HRPressure1 (smallint)
5.10 16 ST O Label Group Pressure 1 EventValveAreaAnalysis.LabelGroup_P1 (resolve
5.11 32 ST O Site Label Pressure 2 EventValveAreaAnalysis.SiteLabelPressure2
5.12 8 ST O Mean Pressure 2 EventValveAreaAnalysis.MeanPressure2 (real)
5.13 8 ST O Mean Diastolic Pressure 2 EventValveAreaAnalysis.MeanDiastolicValue2 (real)
5.14 8 ST O Mean Systolic Pressure 2 EventValveAreaAnalysis.MeanSystolicValuePres2
5.15 8 ST O Mean A Wave Pressure 2 EventValveAreaAnalysis.MeanAWaveValuePres2
5.16 8 ST O Mean V Wave Pressure 2 EventValveAreaAnalysis.MeanVWaveValuePres2
5.17 5 ST O Heart Rate Pressure 2 EventValveAreaAnalysis.HRPressure2 (smallint)
5.18 16 ST O Label Group Pressure 2 EventValveAreaAnalysis.LabelGroup_P2 (resolve
(real)
(real)
(real)
for string)
(real)
(real)
(real)
for string)
5.19 8 ST O Mean Gradient EventValveAreaAnalysis.MeanGradient (real)
5.20 8 ST O Valve Area EventValveAreaAnalysis.ValveArea (real)
5.21 8 ST O Valve Area Index EventValveAreaAnalysis.ValveAreaIndex (real)
5.22 8 ST O Valve Flow EventValveAreaAnalysis.ValveFlow (real)
5.23 8 ST O DFP/SEP EventValveAreaAnalysis.DFPSEP (real)
5.24 8 ST O Cardiac Output Value Used EventValveAreaAnalysis.COUsed (real)
5.25 8 ST O Ejection Time / Filling Time EventValveAreaAnalysis.ETimeFTime (real)
5.26 10 ST O Phase Shift P1 P2 EventValveAreaAnalysis.PhaseShiftP1P2 (int)
5.27 1 ST O Manually Edited Flag (0 or 1) EventPPGMeasurement.ManuallyEdited
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 41
GEHC
SEQ LEN DT
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Example Valve Area Analysis OBX:
OBX|52|ST|Event_ValveAreaAnalysis||Generic^2^PCW^90.00^0.00^0.00^122.20^123.40^69^P
CW^PA^95.00^79.00^123.00^0.00^0.00^69^PA^2.09^3.99^2.56^255.82^14.85^3.80^0.22^0^0||||
||F|||20011003150217||||
Event_O2_Saturation OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “Event_O2Sat”
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Sample Site EventO2Saturation.Site (resolve for string)
5.2 4 ST O Phase Phase Number
5.3 8 ST O Sample Saturation EventO2Saturation.Saturation (real)
5.4 8 ST O Sample HB EventO2Saturation.HB (real)
5.5 8 ST O Sample PO2 EventO2Saturation.PO2 (real)
5.6 8 ST O Sample Content EventO2Saturation.Content (real)
5.7 16 ST O Sample Group O2SatGroup.Group resolved from
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
EventO2Saturation.GroupType
42 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation
(Sample Time)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
Example O2_Sat OBX:
OBX|41|ST|Event_O2Sat||AO^1^98.00^13.00^120.00^17.68^SA||||||F|||20011003145843||||
Event_Pressure_Measurement OBX (Reporting Structure)
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Event_ValveAreaAnalysis or Event_CathPressure
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 32 ST O Measurement Name EventCathPresMeas.ChannelLabel
5.2 4 ST O Phase Phase Number
5.3 32 ST O Measurement Type EventCathPresMeas.MeasType
5.4 8 ST O Systolic EventCathPresMeas.AvgSystolic
5.5 8 ST O Diastolic EventCathPresMeas.AvgDiastolic
5.6 8 ST O End Diastolic EventCathPresMeas.AvgEndDiastolic
5.7 8 ST O Max dP/dT EventCathPresMeas.MaxDPDT
5.8 8 ST O Mean EventCathPresMeas.Mean
5.9 8 ST O A Wave EventCathPresMeas.AvgA
5.10 8 ST O V Wave EventCathPresMeas.AvgV
5.11 4 ST O Heart Rate EventCathPresMeas.MeanHR
5.12 1 ST O Manually Edited Flag (0 or 1) EventCathPresMeas.ManuallyEdited
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 43
GEHC
SEQ LEN DT
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Change from older versions is that PVW (including LPVW and RPVW) uses s/dm instead of
a/v/m.
Example Event_CathPressure OBX:
OBX|53|ST|Event_CathPressure||LV^0^VENT_TYPE^191^^39^1536^^^^69^0||||||F|||200103070
81824
OBX|76|ST|Event_CathPressure||AO^0^ARTERIAL_TYPE^118^81^^^97^^^84^0||||||F|||2001030
7082157
OBX|83|ST|Event_CathPressure||PCW^1^AWEDGE_TYPE^^^^^7^10^8^69^1||||||F|||200103070
82322
OBX|14|ST|Event_CathPressure||VEN^1^VENOUS_TYPE^^^^^9^^^^1||||||F|||20010307081316
Cath Hemodynamic Measurement Reporting Structures
Mac-Lab can produce over 350 discrete clinical measurements. Although not all measurements
would be provided in a test, any combination can be present. The measurement values taken
during the study are stored in one of six (6) different formats/structures, depending upon the
content of the measurement. The structures and associated observations are listed below. The
formats of the measurement structures are fixed and are not user configurable.
The information reported in these Hemodynamic Measurement reporting structures are the
clinical values used for hemodynamic analysis. The values included in these structures are
synchronized with other values in this section (for example, a particular cardiac output reported in
the General Structure is the same one used in the computation of the Valve Pressure Structure
irrespective of the multiple raw cardiac output events recorded in a hemodynamic phase).
44 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
General Structure
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “HemoMeas_General”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Measurement Name Result.indexNum resolved for a text string
5.2 4 ST O Phase Phase Number
5.3 16 ST O Source Result.SourceType
5.4 8 ST O Value Result.Value (real)
6 60 CE O Units Result.Unit resolved from SystemUnit.Name
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Current Phase Event time formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Source: Each measurement is flagged as to the source of the information. Generally, for all
measurements except for VO2, the source represents whether the value was Calculated or
Manually Entered. For VO2, it represents the specific algorithm used to derive the VO2 value
(second column below). Possible values for this field include:
Manual Entry La Farge
Calculated K * BSA
Measured 125 * BSA
Estimated 3 * Wt(kg)
1.4*Ht+0.8*Wt-36
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 45
The following measurements will be formatted in this type of OBX:
AERP
AGE
ARTCON
ARTSAT
BOX_DIFF
AV_ARE A
AV_FLOW
AV_NDX
AV_SEP
BARPRES
BSA
CSNRT
EXPO2
FICKCI
FICKCO
FICKHR
GENDER (0=M, 1=F)
HB
HEIGHT_CM
INSPO2
LRFLOW
LRFLOWI
LSHNT
LVSW
LVSWI
MANCI
MANCO
MV_AREA
MV_DFP
MV_FLOW
MV_NDX
PACON
PASAT
PO2
PV_AREA
PV_FLOW
PV_NDX
PV_SEP
PVCON
PVR_DSC
PVR_WU
PVRI_DSC
46 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
PVRI_WU
PVRSVR_RATIO
PVSAT
QPEFF
QPEFFI
QPI
QPQS_RATIO
QSI
RLFLOW
RLFLOWI
ROOMO2
ROOMTEMP
RSHNT
RVSW
RVSWI
SCLV
SHNT
SP_AREA
SP_FLOW
SP_NDX
SP_SEP
STROKEI
STROKEV
SVR_DSC
SVR_WU
SVRI_DSC
SVRI_WU
TDCI
TDCO
TPR_DSC
TPR_WU
TPRI_DSC
TPRI_WU
TPRTVR_RATIO
TV_AREA
TV_DFP
TV_FLOW
TV_NDX
TVR_DSC
TVR_WU
TVRI_DSC
TVRI_WU
VEATPS
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 47
VENCON
VENSAT
VERP
VO2
WEIGHT_KG
General Structure Example:
OBX|24|ST|HemoMeas_General||BSA^0^CALCULATED^1.86|m2|||||F|||20010307084420|||
Pressure Structure
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “HemoMeas_Pressure”
4 20 ST C Observation Sub-ID Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Measurement Name Result.indexNum resolved for a text string
5.2 4 ST O Phase Phase Number
5.3 16 ST O Source Result.SourceType
5.4 8 ST O Systolic Result.Value
5.5 32 ST O Systolic Units Result.Unit resolved from SystemUnit.Name
5.6 8 ST O Diastolic Result.Value
5.7 32 ST O Diastolic Units Result.Unit resolved from SystemUnit.Name
5.8 8 ST O Mean Result.Value
5.9 32 ST O Mean Units Result.Unit resolved from SystemUnit.Name
5.10 8 ST O Heart Rate Result.Value
5.11 32 ST O Heart Rate Units Result.Unit resolved from SystemUnit.Name
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
48 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation Current Phase Event time formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
The following measurements will be formatted in this type of OBX:
ABAO
AO
ARCH
ART
ASAO
AXA
BA
CARA
CNDT
COL
COR
DSAO
FA
FCLA
FCLV
FCRA
FCRV
FIS
HABAO
HAO
HARCH
HASAO
HDSAO
HEPA
HNAO
HUA
IMA
INNA
LABAO
LAO
LARCH
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 49
LART
LASAO
LAXA
LBA
LCARA
LCNDT
LCOL
LCOR
LDSAO
LFA
LFCLA
LFCLV
LFCRA
LFCRV
LFIS
LHEPA
LIMA
LINNA
LNAO
LNPA
LPA
LPAB
LPACN
LPAVF
LPDA
LRADA
LRENA
LSBCA
LSP
LTAC
LUA
LVRTA
NAO
NPA
PA
PAB
PACN
PAVF
PDA
RADA
RART
RAXA
RBA
50 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
RCARA
RCNDT
RCOL
RCOR
RENA
RFA
RFCLA
RFCLV
RFCRA
RFCRV
RFIS
RHEPA
RIMA
RINNA
RNPA
RPA
RPAB
RPACN
RPAVF
RPDA
RRADA
RRENA
RSBCA
RSP
RTAC
RVRTA
SBCA
SP
TAC
UA
VRTA
Pressure Structure Example:
OBX|4|ST|HemoMeas_Pressure||AO^0^CALCULATED^175^mmHg^72^mmHg^110^mmHg^64^
beats/min||||||F|| |20010307084420|||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 51
Mean Pressure Structure
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “HemoMeas_Mean_Pressure”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Measurement Name Result.indexNum resolved for a text string
5.2 4 ST O Phase Phase Number
5.3 32 ST O Source Result.SourceType
5.4 8 ST O Value Result.Value
5.5 32 ST O Value Units Result.Unit resolved from SystemUnit.Name
5.6 8 ST O Heart Rate Result.Value
5.7 32 ST O Heart Rate Units Result.Unit resolved from SystemUnit.Name
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation
(Sample Time)
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Current Phase Event time formatted as
YYYYMMDDHHMMSS
The following measurements will be formatted in this type of OBX:
ACV
AXV
AZV
BV
CVP
FV
52 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
HAZ
HCVP
HEPV
HIVC
HSVC
HUV
HVC
INNV
IVC
LACV
LAXV
LAZV
LBV
LCVP
LFV
LHAZ
LHEPV
LINNV
LIVC
LRENV
LSCLV
LSPHV
LSVC
LUV
LVC
LVEN
RACV
RAXV
RAZV
RBV
RENV
RFV
RHAZ
RHEPV
RINNV
RRENV
RSCLV
RSPHV
RVEN
SCLV
SPHV
SVC
UV
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 53
VC
VEN
Mean Pressure Structure Example:
OBX|34|ST|HemoMeas_MeanPressure||VEN^2^MEASURED^175^mmHg^64^beats/
min||||||F|||20010307084420|||
Ventricular Pressure Structure
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “HemoMeas_Ventricular”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Measurement Name Result.indexNum resolved for a text string
5.2 4 ST O Phase Phase Number
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.3 32 ST O Source Result.SourceType
5.4 8 ST O Systolic Result.Value
5.5 32 ST O Systolic Units Result.Unit resolved from SystemUnit.Name
5.6 8 ST O End Diastolic Result.Value
5.7 32 ST O End Diastolic Units Result.Unit resolved from SystemUnit.Name
5.8 8 ST O Heart Rate Result.Value
5.9 32 ST O Heart Rate Units Result.Unit resolved from SystemUnit.Name
5.10 8 ST O dP/dt Result.Value
5.11 32 ST O dP/dt Units Result.Unit resolved from SystemUnit.Name
5.12 8 ST O Diastolic Result.Value
5.13 32 ST O Diastolic Units Result.Unit resolved from SystemUnit.Name
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
54 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
SEQ LEN DT
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
The following measurements will be formatted in this type of OBX:
CV
HLV
HLVA
HLVIN
HLVOC
HLVOT
HRV
HRVA
HRVIN
HRVOC
HRVOT
IB
LCV
LIB
LLV
LLVA
LLVIN
LLVOC
LLVOT
LPVCH
LRV
LRVA
LRVIN
LRVOC
LRVOT
LSB
LV
LVA
LVIN
LVOC
LVOT
PVCH
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 55
RCV
RIB
RPVCH
RSB
RV
RVA
RVIN
RVOC
RVOT
SB
Ventricular Pressure Structure Example:
OBX|34|ST|HemoMeas_Ventricular||LV^1^MEASURED^124^mmHg^18^mmHg^64^beats/
min^1536^mmHg/sec||||||F|| |20010307084420|||
Valve Structure
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “HemoMeas_Valve”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Measurement Name Result.indexNum resolved for a text string
5.2 4 ST O Phase Phase Number
5.3 16 ST O Source Result.SourceType
5.4 4 ST O Heart Rate Result.Value
5.5 32 ST O Heart Rate Units Result.Value
5.6 16 ST O Left Site Label Result.Value
5.7 8 ST O Left Systolic Result.Value
5.8 32 ST O Left Systolic Units Result.Unit resolved from SystemUnit.Name
5.9 8 ST O Left Diastolic Result.Value
5.10 32 ST O Left Diastolic Units Result.Unit resolved from SystemUnit.Name
5.11 16 ST O Right Site Label Result.Value
5.12 8 ST O Right Systolic Result.Value
5.13 32 ST O Right Systolic Units Result.Unit resolved from SystemUnit.Name
5.14 8 ST O Right Diastolic Result.Value
5.15 32 ST O Right Diastolic Units Result.Unit resolved from SystemUnit.Name
56 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
5.16 8 ST O Valve Gradient Result.Value
5.17 32 ST O Valve Gradient Units Result.Unit resolved from SystemUnit.Name
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation Current Phase Event time formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
The following measurements will be formatted in this type of OBX:
AV_DUA L
AV_PULL
AV_SP
NV_DUAL
MV_DUAL_LA
MV_DUAL_PCW
MVPULL_LVLA
MVPULL_LVPCW
MVSP_LVLA
MVSP_LVPCW
NV_PULL
PV_DUAL
PV_PULL
PV_SP
V_SP
TV_DUAL
TV_PULL
TV_SP
Valve Structure Example:
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 57
OBX|30|ST|HemoMeas_Valve||V_SP^2^Measured^69^beats/
min^PCW^0.00^mmHg^0.00^mmHg^PA^123.00^mmHg^79.00^mmHg^2.09^mmHg||||||F|||20011
003150144|||
Atrial Wedge Structure
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “HemoMeas_AtrialWedge”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 16 ST O Measurement Name Result.indexNum resolved for a text string
5.2 4 ST O Phase Phase Number
5.3 32 ST O Source Result.SourceType
5.4 8 ST O A Wave Result.Value
5.5 32 ST O A Wave Units Result.Unit resolved from SystemUnit.Name
5.6 8 ST O V Wave Result.Value
5.7 32 ST O V Wave Units Result.Unit resolved from SystemUnit.Name
5.8 8 ST O Mean Result.Value
5.9 32 ST O Mean Units Result.Unit resolved from SystemUnit.Name
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.10 8 ST O Heart Rate Result.Value
5.11 32 ST O Heart Rate Units Result.Unit resolved from SystemUnit.Name
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Current Phase Event time formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
58 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
The following measurements will be formatted in this type of OBX:
APV
CA
CS
DCS
HLA
HRA
JX
LA
LAPV
LCA
LJX
LLA
LPAW
LPCW
LPV
LPVAT
LPVCN
LPVW
LRA
LSVA
PAW
PCS
PCW
PV
PVAT
PVCN
PVW
RA
RAPV
RCA
RJX
RPAW
RPCW
RPV
RPVAT
RPVCN
RPVW
RSVA
SVA
Atrial Wedge Structure Example:
OBX|23|ST|HemoMeas_AtrialWedge||PCW^2^Measured^74.00^mmHg^123.00^mmHg^93.00^
mmHg^69^beats/min||||||F|||20011003150144||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 59
EP Reporting Structures
The structures and associated observations for EP measurements are listed below. The formats
of the measurement structures are fixed and are not user configurable.
EP SNRT
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “EP_SNRT”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 5 ST O SNRT pacing interval in ms EventSNRT.PacingInterval (smallint)
5.2 5 ST O Corrected SNRT in ms EventSNRT.CSRTInterval (smallint)
5.3 5 ST O Max SNRT in ms EventSNRT.MaxSNRT (smallint)
5.4 5 ST O SNRT sinus cycle length in ms EventSNRT.SinusCycleLength (smallint)
5.5 256 ST O SNRT comment EventBlock1.EventComment
5.6 10 ST O Event Datapoint EventBlock1.eventdatapoint
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation Current Phase Event Time formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
EP SNRT Example:
OBX|2|ST| EP_ SNRT ||525^520^550^100^no comments^34567||||||F|||20011003150144||||
60 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
EP ATGD
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “EP_ATGD”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 32 ST O Refractory region EventATGD.RefractoryRegion
5.2 32 ST O Refractory type EventATGD.Type resolved to string
5.3 5 ST O Refractory period in ms EventATGD.TextRefractoryPeriod (smallint)
5.4 5 ST O S1-S1 Interval in ms EventATGD.S1S1Interval (smallint)
5.5 256 ST O Refractory comment EventBlock1.EventComment
5.6 10 ST O Event Datapoint EventBlock1.eventdatapoint
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
EP ATGD Example:
OBX|3|ST| EP_ ATGD ||AV Nodal^ERP^240^600^no
comments^34567||||||F|||20011003150144||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 61
EP Baseline Conduction
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “EP_BaselineConduction”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 5 ST O AA interval in ms EventBSLC.NAA (smallint)
5.2 5 ST O VV interval in ms EventBSLC.NVV (smallint)
5.3 5 ST O PR interval in ms EventBSLC.NPR (smallint)
5.4 5 ST O QRS Duration in ms EventBSLC.NQRSDUR (smallint)
5.5 5 ST O QT interval in ms EventBSLC.NQT (smallint)
5.6 5 ST O HIS duration in ms EventBSLC.NHISDUR (smallint)
5.7 5 ST O AH interval in ms EventBSLC.NAH (smallint)
5.8 5 ST O PA interval in ms EventBSLC.NPA (smallint)
5.9 5 ST O HV interval in ms EventBSLC.NHV (smallint)
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.10 5 ST O VA interval in ms EventBSLC.NVA (smallint)
5.11 12 ST O Corrected QT interval (for HR) in ms EventBSLC.DQTC (float)
5.12 10 ST O Event Datapoint EventBSLC.eventdatapoint
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
EP Baseline Conduction Example:
OBX|2|ST| EP_BaselineConduction
||645^700^140^94^390^200^145^130^104^110^417^34567||||||F|||20011003150144||||
62 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
EP Arrhythmia
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “EP_Arrhythmia”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 128 ST O Arrhythmia type EventArrhythmia.ArrhythmiaType
5.2 26 TS O Start time EventBlock1.EventTime formatted as
5.3 5 ST O Duration of Arrhythmia in s EventArrhythmia.Duration (smallint)
5.4 16 ST O Sustained? 0=No and 1=Yes YesNo.YesNoText resolved from
5.5 5 ST O Number of Cycles EventArrhythmia.NumOfCycle (smallint)
5.6 5 ST O Ventricular cycle length in ms EventArrhythmia.VerticalCycleLength (smallint)
5.7 5 ST O Atrial cycle length in ms EventArrhythmia.AtrialCycleLength (smallint)
5.8 32 ST O Tolerance EventArrhythmia.ArrhythmiaTolerances
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
YYYYMMDDHHMMSS
EventArrhythmia.Sustained (1=Yes, 0=No)
5.9 128 ST O Initiation EventArrhythmia.Initiation
5.10 128 ST O Termination EventArrhythmia.Termination
5.11 26 TS O Stop time EventArrhythmia.StopTime or Not Used
5.12 10 ST O Event Datapoint EventArrhythmia.eventdatapoint
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 63
EP Arrhythmia Example:
OBX|2|ST| EP_ Arrhythmia ||0^20021219101602^0^N^0^384^1^^^^20021219101602^34567||||||F|||20011003150144||||
Or
OBX|2|ST| EP_ Arrhythmia ||0^20110215183225^0^N^0^384^-
1^^^^^34567||||||F|||20110215183225||||
EP Conduction Block
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “EP_ConductionBlock”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 5 ST O Pacing interval in ms EventConductionBlock.PacingInterval (smallint)
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.2 32 ST O Description EventConductionBlock.BlockDescription
5.3 10 ST O Event Datapoint EventConductionBlock.eventdatapoint
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
EP Conduction Block Example:
OBX|2|ST| EP_ ConductionBlock ||200^2:1 AV Block^34567||||||F|||20011003150144||||
64 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
EP_Ablation
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "EP_Ablation"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 32 ST O Counter for RF Applications “#” where # is sequential counter of RF Ablations
5.2 26 TS O Start time of ablation event EventBlock1.EventTime
5.3 26 TS O Stop time of ablation event EventAblation.StopTime
5.4 5 ST O Duration in s EventAblation.Duration (smallint)
5.5 32 ST O Target arrhythmia EventAblation.TargetArrhythmia
5.6 32 ST O Result EventAblation.Result
5.7 8 ST O Max temperature 1 in deg cel EventAblation.MaxTemp1 (real)
5.8 8 ST O Average temperature 1 in deg cel EventAblation.AvgTemp1 (real)
5.9 8 ST O Max power in watts EventAblation.MaxPower (real)
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
5.10 8 ST O Average power in watts EventAblation.AvgPower (real)
5.11 8 ST O Max impedance in ohms EventAblation.MaxImpedence (real)
5.12 8 ST O Average impedance in ohms EventAblation.AvgImpedence (real)
5.13 8 ST O Max current in mA EventAblation.MaxCurrent (real)
5.14 8 ST O Average current in mA EventAblation.AvgCurrent (real)
5.15 8 ST O Max voltage in V EventAblation.MaxVoltage (real)
5.16 8 ST O Average voltage in V EventAblation.AvgVoltage (real)
5.17 32 ST O Device name EventAblation.DeviceName
5.18 256 ST O Ablation comment EventBlock1.EventComment
5.19 10 ST O Event Datapoint EventBlock1.eventdatapoint
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 65
GEHC
SEQ LEN DT
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Device name can be one of; CryoCath, PEIAtakr, PEIEpt1000, PEIStockert, PEIHat300S,
PEIIbi1500T. This is not a complete list.
EP Ablation Example:
OBX|2|ST| EP_ Ablation
||1^20021219101855^20021219101909^14^^^52.00^44.00^50.00^36.00^149.00^135.00^616.00
^497.00^83.00^67.00^PEIEPT1000^no comment^34567||||||F|||20011003150144||||
EP_3DMap
GEHC
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always "EP_3DMap"
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Phase Phase Number
5.2 26 ST O Reference Annotation Time in ms EventCartoMapPoint.systemtime
5.3 32 ST O Map ID EventCartoMapPoint.mapid
5.4 32 ST O Map Name EventCartoMapPoint.mapname
5.5 10 ST O Point ID EventCartoMapPoint.PointIndex
5.6 10 ST O LAT in ms EventCartoMapPoint.lat
5.7 10 ST O Unipolar in uV EventCartoMapPoint.unipolar
5.8 10 ST O Bipolar in uV EventCartoMapPoint.bipolar
5.9 32 ST O Type EventCartoMapPoint.type
5.10 32 ST O Tag EventCartoMapPoint.tag
5.11 256 ST O Comment EventBlock1.EventComment
6 60 CE O Units Not Used
66 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN
GEHC
SEQ LEN DT
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
13 20 ST O User Defined Access Checks Not Used
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
YYYYMMDDHHMMSS
Example OBX Segment:
OBX|14|ST| EP_3DMap || 0^2^21345^RCA^43255^245^12.3^20.6^Point Type^Point
Tag^comment||||||F|||20011003144143||||
2047900-111B EN Mac-Lab/CardioLab HL7 Export Specification Reference Manual 67
GE Medical Systems Information Technologies , a General Electric Company, going to market as GE Healthcare.
EP_Pacing
GE Medical Systems Information Technologies . • 8200 W. Tower Ave.
Milwaukee, WI 53223 U.S.A.
© 2011 GE Medical Systems Information Technologies . All rights reserved.
SEQ LEN DT
1 4 SI O Set ID - OBX Start sequential counter within an OBR nest.
2 3 ID R Value Type Always "ST"
3 200 CE R Observation Identifier Always “EP_Pacing”
4 20 ST C Observation Sub-ID Not Used
5 65536 * C Observation Value Set as the observed result or reporting structure
5.1 4 ST O Channel Name Channel Name
5.2 26 ST O Channel Number Channel Number (base zero)
6 60 CE O Units Not Used
7 60 ST O References Range Not Used
8 5 ID O Abnormal Flags Not Used
9 5 NM O Probability Not Used
10 2 ID O Nature of Abnormal Test Not Used
11 1 ID R Observation Result Status "F"
12 26 TS O Date Last Obs Normal Values Not Used
GEHC
Required
Field
ELEMENT NAME MAC-LAB/CARDIOLAB FIELD
13 20 ST O User Defined Access Checks Not Used
14 26 TS R Date/Time of the Observation EventBlock1.EventTime formatted as YYYYMMDDHHMMSS
15 60 CE O Producer's ID Not Used
16 80 XCN O Responsible Observer Not Used
17 60 CE O Observation Method Not Used
Example OBX Segment:
OBX|14|ST| EP_Pacing|| A1^63||||||F|||20011003144143||||
68 Mac-Lab/CardioLab HL7 Export Specification Reference Manual 2047900-111B EN