General Electric MACIT_GE_DEVICE_CONFIGURATION_GUIDANCE_SM_2069773-005_D Philips BV 25 T Operation Manual

Page 1
GEHealthcare
OptimaEMSECGManagementSystem GEHealthcareDeviceCongurationGuide
SoftwareVersion1.0 2069773-005RevisionD
OptimaEMS English ©2012-2013GeneralElectricCompany. AllRightsReserved.
Page 2
PublicationInformation
TheinformationinthismanualappliesonlytoOptimaEMSECGManagementSystemVersion1.0.Itdoesnotapplytoearlierproduct versions.Duetocontinuingproductinnovation,specicationsinthismanualaresubjecttochangewithoutnotice.
MUSEandMACaretrademarksownedbyGEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompanygoingtomarket asGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheirrespectiveowners.
Thedocumentpartnumberandrevisionappearatthebottomofeachpage.Therevisionidentiesthedocument’supdatelevel.The revisionhistoryofthisdocumentissummarizedinthefollowingtable.
Revision
A
B
C10January2013
Date
1November2012InternalRelease
10December2012Initialreleaseforcustomer
Comments
PerSPRHCSDM00174744,thefollowingupdatesaremade:
Re-layoutthemanual
AddedinformationaboutCongurationforPDFImport
D
ToaccessotherGEHealthcaremanuals,gototheCommonDocumentationLibrary(CDL),locatedathttp://www.gehealthcare.com/usen/
service/biomed_tech_selfservice/services_user_doc/products/support.html andclickCardiology.
ToaccessOriginalEquipmentManufacturer(OEM)manuals,gotothedevicemanufacturer'sWebsite.
22July2013PerSPRHCSDM00221948,anewchaptertitled:ConnectingtheMAC1200
Systemwasadded.
ServiceManualLanguageInformation
WARNING (EN)
ПРЕДУПРЕЖДЕНИЕ (BG)
警告
ZH-CN
警告
(ZH-TW)
ThisservicemanualisavailableinEnglishonly.
Ifacustomer'sserviceproviderrequiresalanguageotherthanEnglish,itisthecustomer's
responsibilitytoprovidetranslationservices.
Donotattempttoservicetheequipmentunlessthisservicemanualhasbeenconsulted
andisunderstood.
Failuretoheedthiswarningmayresultininjurytotheserviceprovider ,operator ,orpatient,
fromelectricshock,mechanicalorotherhazards.
Товаупътванезаработаеналичносамонаанглийскиезик.
Акодоставчикътнауслугатанаклиентаизискадругезик,задължениенаклиентае
даосигурипревод.
Неизползвайтеоборудването,предидастесеконсултиралииразбралиупътването
заработа.
Неспазванетонатовапредупреждениеможедадоведедонараняваненадоставчика
науслугата,оператораилипациентврезултатнатоковударилимеханичнаили другаопасност.
本维修手册仅提供英文版本。
如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
未详细阅读和完全理解本维修手册之前,不得进行维修。
忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
本維修手冊只提供英文版。
如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責提供翻
譯服務。
除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維修。
不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或其他因素
而受到傷害。
2
OptimaEMSECGManagementSystem
2069773-005D
18July2013
Page 3
ServiceManualLanguageInformation(cont'd.)
UPOZORENJE (HR)
VAROVÁNÍ (CS)
ADVARSEL (DA)
WAARSCHUWING (NL)
HOIATUS (ET)
VAROITUS (FI)
Oveuputezaservisiranjedostupnesusamonaengleskomjeziku.
Ukolikokorisničkiserviszahtijevanekidrugijezik,korisnikovajeodgovornostosigurati
odgovarajućiprijevod.
Nemojtepokušavatiservisiratiopremuukolikonistekonzultiraliirazumjelioveupute.
Nepoštivanjeovogupozorenjamožerezultiratiozljedamaservisnogosoblja,korisnikaili
pacijentaprouzročenimelektričnimudaromtemehaničkimilinekimdrugimopasnostima.
Tentoprovoznínávodexistujepouzevanglickémjazyce.
Vpřípadě,žeexterníslužbazákazníkůmpotřebujenávodvjinémjazyce,jezajištění
překladudoodpovídajícíhojazykaúkolemzákazníka.
Nesnažteseoúdržbutohotozařízení,anižbystesipřečetlitentoprovoznínávoda
pochopilijehoobsah.
Vpřípaděnedodržovánítétovarovánímůžedojítkporaněnípracovníkaprodejního
servisu,obslužnéhopersonálunebopacientůvlivemelektrickéhoproudu,respektive vlivemmechanickýchčijinýchrizik.
Denneservicemanualndeskunpåengelsk.
Hvisenkundesteknikerharbrugforetandetsprogendengelsk,erdetkundensansvar
atsørgeforoversættelse.
Forsøgikkeatservicereudstyretmedmindredenneservicemanualharværetkonsulteret
ogerforstået.
Manglendeoverholdelseafdenneadvarselkanmedføreskadepågrundafelektrisk,
mekaniskellerandenfareforteknikeren,operatørenellerpatienten.
DezeservicemanualisalleeninhetEngelsverkrijgbaar .
Indienhetonderhoudspersoneeleenanderetaalnodigheeft,danisdeklant
verantwoordelijkvoordevertalingervan.
Probeerdeapparatuurnietteonderhoudenvoordatdezeservicemanualgeraadpleegd
enbegrepenis.
Indiendezewaarschuwingnietwordtopgevolgd,zouhetonderhoudspersoneel,de
gebruikerofeenpatiëntgewondkunnenrakenalsgevolgvaneenelektrischeschok, mechanischeofanderegevaren.
Käesolevteenindusjuhendonsaadavalainultinglisekeeles.
Kuiklienditeeninduseosutajanõuabjuhenditinglisekeelesterinevaskeeles,vastutab
klienttõlketeenuseosutamiseeest.
Ärgeüritageseadmeidteenindadaenneeelnevaltkäesolevateenindusjuhendiga
tutvumistjasellestarusaamist.
Käesolevahoiatuseeiraminevõibpõhjustadateenuseosutaja,operaatorivõipatsiendi
vigastamistelektrilöögi,mehaanilisevõimuuohutagajärjel.
Tämähuolto-ohjeonsaatavillavainenglanniksi.
Josasiakkaanhuoltohenkilöstövaatiimuutakuinenglanninkielistämateriaalia,tarvittavan
käännöksenhankkiminenonasiakkaanvastuulla.
Äläyritäkorjatalaitteistoaennenkuinoletvarmastilukenutjaymmärtänyttämän
huolto-ohjeen.
Mikälitätävaroitustaeinoudateta,seurauksenavoiollahuoltohenkilöstön,laitteiston
käyttäjäntaipotilaanvahingoittuminensähköiskun,mekaanisenviantaimuun vaaratilanteenvuoksi.
2069773-005D
OptimaEMSECGManagementSystem
3
Page 4
ServiceManualLanguageInformation(cont'd.)
ATTENTION (FR)
WARNUNG (DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ (GR)
FIGYELMEZTETÉS (HU)
AÐVÖRUN (IS)
PERINGATAN (ID)
Cemanueltechniquen'estdisponiblequ'enanglais.
Siunservicetechniqueclientsouhaiteobtenircemanueldansuneautrelangueque
l'anglais,ildevraprendreenchargelatraductionetlaresponsabilitéducontenu.
Nepastenterd'intervenirsurleséquipementstantquelemanueltechniquen'apasété
consultéetcompris.
Lenon-respectdecetavertissementpeutentraînerchezletechnicien,l'opérateuroule
patientdesblessuresduesàdesdangersélectriques,mécaniquesouautres.
DieseServiceanleitungistnurinenglischerSpracheverfügbar.
FallsderKundendiensteineandereSprachebenötigt,musserfüreineentsprechende
Übersetzungsorgen.
KeineWartungdurchführen,ohnedieseServiceanleitunggelesenundverstandenzu
haben.
BeiZuwiderhandlungkanneszuVerletzungendesKundendiensttechnikers,des
AnwendersoderdesPatientendurchStromschläge,mechanischeodersonstigeGefahren kommen.
Τοπαρόνεγχειρίδιοσέρβιςδιατίθεταιστααγγλικάμόνο.
Εάντοάτομοπαροχήςσέρβιςενόςπελάτηαπαιτείτοπαρόνεγχειρίδιοσεγλώσσαεκτός
τωναγγλικών,αποτελείευθύνητουπελάτηναπαρέχειυπηρεσίεςμετάφρασης.
Μηνεπιχειρήσετετηνεκτέλεσηεργασιώνσέρβιςστονεξοπλισμόεκτόςεάνέχετε
συμβουλευτείκαιέχετεκατανοήσειτοπαρόνεγχειρίδιοσέρβις.
Εάνδενλάβετευπόψητηνπροειδοποίησηαυτή,ενδέχεταιναπροκληθείτραυματισμός
στοάτομοπαροχήςσέρβις,στοχειριστήήστονασθενήαπόηλεκτροπληξία,μηχανικούς ήάλλουςκινδύνους.
Ezaszervizkézikönyvkizárólagangolnyelvenérhetőel.
Haavevőszervizellátójaangoltóleltérőnyelvretartigényt,akkoravevőfelelőssége
afordításelkészíttetése.
Nepróbáljaelkezdenihasználniaberendezést,amígaszervizkézikönyvbenleírtakat
nemértelmeztékésértettékmeg.
Ezengyelmeztetésgyelmenkívülhagyásaaszervizellátó,aműködtetővagyapáciens
áramütés,mechanikaivagyegyébveszélyhelyzetmiattisérüléséteredményezheti.
Þessiþjónustuhandbókereingöngufáanlegáensku.
Efaðþjónustuveitandiviðskiptamannsþarfnastannarstungumálsenensku,erþaðskylda
viðskiptamannsaðskaffatungumálaþjónustu.
Reyniðekkiaðafgreiðatækiðnemaþessiþjónustuhandbókhefurveriðskoðuðogskilin.
Brotáaðsinnaþessariaðvörungeturleitttilmeiðslaáþjónustuveitanda,stjórnandaeða
sjúklingifráraosti,vélrænumeðaöðrumáhættum.
ManualservisinihanyatersediadalambahasaInggris.
JikapenyediajasaservispelangganmemerlukanbahasalainselaindariBahasaInggris,
merupakantanggungjawabdaripenyediajasaservistersebutuntukmenyediakan terjemahannya.
Janganmencobamelakukanservisterhadapperlengkapankecualitelahmembacadan
memahamimanualservisini.
Mengabaikanperingataninibisamengakibatkancederapadapenyediaservis,operator,
ataupasien,karenaterkenakejutlistrik,bahayamekanisataubahayalainnya.
4
OptimaEMSECGManagementSystem
2069773-005D
18July2013
Page 5
ServiceManualLanguageInformation(cont'd.)
AVVERTENZA (IT)
警告
(JA)
경고
(KO)
BRĪDINĀJUMS (LV)
ĮSPĖJIMAS (LT)
ADVARSEL (NO)
IlpresentemanualedimanutenzioneèdisponibilesoltantoinInglese.
Seunaddettoallamanutenzionerichiedeilmanualeinunalinguadiversa,ilclienteè
tenutoaprovvederedirettamenteallatraduzione.
Siprocedaallamanutenzionedell'apparecchiaturasolodopoaverconsultatoilpresente
manualeedavernecompresoilcontenuto.
Ilnonrispettodellapresenteavvertenzapotrebbefarcompiereoperazionidacuiderivino
lesioniall'addetto,allamanutenzione,all'utilizzatoreedalpazienteperfolgorazione elettrica,perurtimeccaniciodaltririschi.
このサービスマニュアルは英語版しかありません。
サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
このサービスマニュアルを熟読し、十分に理解をした上で装置のサービスを
行ってください。
この警告に従わない場合、サービスを担当される方、操作員あるいは患者が、感電
や機械的又はその他の危険により負傷する可能性があります。
본서비스지침서는영어로만이용하실수있습니다.
고객의서비스제공자가영어이외의언어를요구할경우,번역서비스를제공하는것
은고객의책임입니다.
본서비스지침서를참고했고이해하지않는한은해당장비를수리하려고시도하
지마십시오.
이경고에유의하지않으면전기쇼크,기계상의혹은다른위험으로부터서비스제
공자,운영자혹은환자에게위해를가할수있습니다.
Šīapkalpotājurokasgrāmatairpieejamatikaiangļuvalodā.
Jaapkalpošanassniedzējamnepieciešamainformācijacitā,nevisangļu,valodā,klienta
pienākumsirnodrošināttāstulkošanu.
Neveicietaprīkojumaapkopi,neizlasotunnesaprototapkalpotājurokasgrāmatu.
Šībrīdinājumaneievērošanavarradītelektriskāsstrāvastrieciena,mehāniskuvaiciturisku
izraisītutraumuapkopessniedzējam,operatoramvaipacientam.
Šiseksploatavimovadovasyraprieinamastikanglųkalba.
Jeiklientopaslaugųtiekėjasreikalaujavadovokitakalba-neanglų,numatytivertimo
paslaugasyraklientoatsakomybė.
Nemėginkiteatliktiįrangostechninėspriežiūros,nebentatsižvelgėteįšįeksploatavimo
vadovąirjįsupratote.
Jeineatkreipsitedėmesioįšįperspėjimą,galimisužalojimaidėlelektrosšoko,mechaninių
arkitųpaslaugųtiekėjui,operatoriuiarpacientui.
Denneservicehåndbokennnesbarepåengelsk.
Hviskundensserviceleverandørtrengeretannetspråk,erdetkundensansvaråsørge
foroversettelse.
Ikkeforsøkåreparereutstyretutenatdenneservicehåndbokenerlestogforstått.
Manglendehensyntildenneadvarselenkanføretilatserviceleverandøren,operatøren
ellerpasientenskadespågrunnavelektriskstøt,mekaniskeellerandrefarer .
2069773-005D
OptimaEMSECGManagementSystem
5
Page 6
ServiceManualLanguageInformation(cont'd.)
OSTRZEŻENIE (PL)
Niniejszypodręcznikserwisowydostępnyjestjedyniewjęzykuangielskim.
Jeślidostawcausługklientawymagajęzykainnegoniżangielski,zapewnienieusługi
tłumaczeniajestobowiązkiemklienta.
Nienależyserwisowaćwyposażeniabezzapoznaniasięizrozumienianiniejszego
podręcznikaserwisowego.
Niezastosowaniesiędotegoostrzeżeniamożespowodowaćurazydostawcyusług,
operatoralubpacjentawwynikuporażeniaelektrycznego,zagrożeniamechanicznego bądźinnego.
AVISO (PT-BR)
Estemanualdeassistênciatécnicasóseencontradisponíveleminglês.
SeoserviçodeassistênciatécnicadoclientenãoforGE,eprecisardeoutroidioma,será
daresponsabilidadedoclientefornecerosserviçosdetradução.
Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanualde
assistênciatécnica.
Onãocumprimentodesteavisopodeporemperigoasegurançadotécnico,operadorou
pacientedevidoachoqueselétricos,mecânicosououtros.
AVISO (PT-PT)
Estemanualtécnicosóseencontradisponíveleminglês.
Seaassistênciatécnicadoclientesolicitarestesmanuaisnoutroidioma,éda
responsabilidadedoclientefornecerosserviçosdetradução.
Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanual
técnico.
Onãocumprimentodesteavisopodeprovocarlesõesaotécnico,aoutilizadorouao
pacientedevidoachoqueseléctricos,mecânicosououtros.
AVERTISMENT (RO)
Acestmanualdeserviceestedisponibilnumaiînlimbaengleză.
Dacăunfurnizordeserviciipentruclienţinecesităoaltălimbădecâtceaengleză,estede
datoriaclientuluisăfurnizezeotraducere.
Nuîncercaţisăreparaţiechipamentuldecâtulteriorconsultăriişiînţelegeriiacestui
manualdeservice.
Ignorareaacestuiavertismentarputeaducelarănireadepanatorului,operatoruluisau
pacientuluiînurmapericolelordeelectrocutare,mecanicesaudealtănatură.
ПРЕДУПРЕЖДЕНИЕ (RU)
Настоящееруководствопообслуживаниюпредлагаетсятольконаанглийскомязыке.
Еслисервисномуперсоналуклиентанеобходиморуководствоненаанглийском,ана
каком-тодругомязыке,клиентуследуетобеспечитьпереводсамостоятельно.
Преждечемприступатькобслуживаниюоборудования,обязательнообратитеськ
настоящемуруководствуивнимательноизучитеизложенныевнемсведения.
Несоблюдениетребованийданногопредупрежденияможетпривестиктому,что
специалистыпообслуживанию,операторыилипациентыполучатударэлектрическим током,механическуютравмуилидругоеповреждение.
UPOZORENJE (SR)
Ovoservisnouputstvojedostupnosamonaengleskomjeziku.
Akoklijentovserviserzahtevanekidrugijezik,klijentjedužandaobezbediprevodilačke
usluge.
Nepokušavajtedaopraviteuređajakonistepročitaliirazumeliovoservisnouputstvo.
Zanemarivanjeovogupozorenjamožedovestidopovređivanjaservisera,rukovaocaili
pacijentausledstrujnogudara,ilimehaničkihidrugihopasnosti.
6
OptimaEMSECGManagementSystem
2069773-005D
18July2013
Page 7
ServiceManualLanguageInformation(cont'd.)
VAROVANIE (SK)
OPOZORILO (SL)
ADVERTENCIA (ES)
VARNING (SV)
UYARI (TR)
Tentonávodnaobsluhujekdispozíciilenvangličtine.
Akzákazníkovposkytovateľslužiebvyžadujeinýjazykakoangličtinu,poskytnutie
prekladateľskýchslužiebjezodpovednosťouzákazníka.
Nepokúšajtesaoobsluhuzariadeniaskôr ,akosineprečítatenávodnaobsluhua
neporozumietemu.
Zanedbanietohtovarovaniamôževyústiťdozraneniaposkytovateľaslužieb,obsluhujúcej
osobyalebopacientaelektrickýmprúdom,mechanickýmaleboinýmnebezpečenstvom.
Taservisnipriročnikjenavoljosamovangleškemjeziku.
Čeponudnikstoritvestrankepotrebujepriročnikvdrugemjeziku,morastrankazagotoviti
prevod.
Neposkušajteservisiratiopreme,četegapriročnikanistevcelotiprebraliinrazumeli.
Četegaopozorilaneupoštevate,selahkozaradielektričnegaudara,mehanskihalidrugih
nevarnostipoškodujeponudnikstoritev,operateralibolnik.
Estemanualdeserviciosóloexisteeninglés.
Sielencargadodemantenimientodeunclientenecesitaunidiomaquenoseaelinglés,el
clientedeberáencargarsedelatraduccióndelmanual.
Nosedeberádarserviciotécnicoalequipo,sinhaberconsultadoycomprendidoeste
manualdeservicio.
Lanoobservanciadelpresenteavisopuededarlugaraqueelproveedordeservicios,el
operadoroelpacientesufranlesionesprovocadasporcausaseléctricas,mecánicaso deotranaturaleza.
Denhärservicehandbokennnsbaratillgängligpåengelska.
Omenkundsserviceteknikerharbehovavettannatspråkänengelskaansvararkunden
föratttillhandahållaöversättningstjänster.
Försökinteutföraservicepåutrustningenomduinteharlästochförstårdenhär
servicehandboken.
Omduintetarhänsyntilldenhärvarningenkandetresulteraiskadorpåserviceteknikern,
operatörenellerpatiententillföljdavelektriskastötar,mekaniskafarorellerandrafaror .
Buserviskılavuzununsadeceİngilizcesimevcuttur.
Eğermüşteriteknisyenibukılavuzuİngilizcedışındabirbaşkalisandantalepederse,bunu
tercümeettirmekmüşteriyedüşer.
Serviskılavuzunuokuyupanlamadanekipmanlaramüdahaleetmeyiniz.
Buuyarıyauyulmaması,elektrik,mekanikveyadiğertehlikelerdendolayıteknisyen,
operatörveyahastanınyaralanmasınayolaçabilir.
2069773-005D
OptimaEMSECGManagementSystem
7
Page 8
ServiceManualLanguageInformation(cont'd.)
ЗАСТЕРЕЖЕННЯ (UK)
CNHBÁO (VI)
Данекерівництвозсервісногообслуговуванняпостачаєтьсявиключноанглійськоюмовою.
Якщосервіснийінженерпотребуєкерівництвоіншоюмовою,користувачзобов'язаний
забезпечитипослугиперекладача.
Ненамагайтесяздійснюватитехнічнеобслуговуванняданогообладнання,якщо
винечитали,абонезрозумілиінформацію,наданувкерівництвізсервісного обслуговування.
Недотриманняцьогозастереженняможепризвестидотравмуваннясервісного
інженера,користувачаданогообладнанняабопацієнтавнаслідокелектричногошоку, механічногоушкодженняабозіншихпричинневірногообслуговуванняобладнання.
TàiLiuHướngDnSaChachỉcóbntiếngAnh.
Nếucácđơnvịcungcpdchvụchokháchhàngyêucumtngônngữnàokháctiếng
Anh,thìkháchhàngsẽcótráchnhimcungcpcácdchvụdchthut.
KhôngđượcsachathiếtbịtrừkhiđãthamkhovàhiuTàiliuHướngdnSacha.
Khôngtuânthủnhngcnhbáonàycóthểdếncáctnthươngchongườithc
hinsacha,ngườivnhànhhaybnhnhân,doscđin,cácrirovềcơkhíhay cácrirokhác.
8
OptimaEMSECGManagementSystem
2069773-005D
18July2013
Page 9
Contents
1Introduction
ManualInformation..................................................................................11
IntendedAudience...................................................................................11
ManualPurpose......................................................................................11
DocumentConventions.............................................................................11
ResponsibilityofthePurchaser/Customer...............................................12
AdditionalAssistance................................................................................12
MACCartsCongurationOverview..........................................................13
2ConnectingtheMAC1600/MAC800SystemstotheOptima
EMSSystem
SettingUptheAcq.Client.........................................................................15
CSI-LANConnection..................................................................................16
SettingUptheDevice................................................................................17
SettingUptheOptimaEMSSystemClient.......................................................19
CSI-SerialConnection................................................................................20
SettingUptheDevice................................................................................21
SettingUptheOptimaEMSSystemClient.......................................................22
SharedFolderConnection.........................................................................24
SettingUptheDevice................................................................................25
SettingUptheOptimaEMSSystemClient.......................................................26
3DCPConnection
DCPWiredLANSettings............................................................................28
SettingUptheOptimaEMSSystemClient.......................................................28
SettingUptheDevice................................................................................30
DCPWirelessLANSettings........................................................................32
SettingUptheOptimaEMSSystemClient.......................................................32
SettingUptheDevice................................................................................32
4ConnectingtheMAC1200System
ConnectingtheMAC1200SystemtotheOptimaEMSSystemvia
CSI-Serial...................................................................................................35
SettingUptheMAC1200System.................................................................35
RequirementsforConnectingtheSerialLine....................................................36
SettingUptheOptimaEMSSystemClient.......................................................36
VerifytheConnection................................................................................37
2069773-005D
OptimaEMSECGManagementSystem
9
Page 10
5ConnectingtheMAC3500/MAC5000/MAC5500Systems
totheOptimaEMSSystem
CSI-LANConnection..................................................................................39
SettingUptheDevice................................................................................39
SettingUptheOptimaEMSSystemClient.......................................................40
CSI-SerialConnection................................................................................42
SettingUptheDevice................................................................................42
SettingUptheOptimaEMSSystemClient.......................................................42
6CongurationforPDFImport
SettingUpSharedFolder..........................................................................45
ConguringtheStressPDFReportImportintheOptimaEMSSystem
Client.........................................................................................................53
ConguringtheHolterPDFReportImportintheOptimaEMS
System......................................................................................................54
10
OptimaEMSECGManagementSystem
2069773-005D
Page 11
1

Introduction

Thischapterprovidesgeneralinformationrequiredfortheproperuseofthesystem andthismanual.Familiarizeyourselfwiththisinformationbeforeusingthesystem.

ManualInformation

Thissectionprovidesinformationforthecorrectuseofthismanual. Keepthismanualwiththeequipmentatalltimesandperiodicallyreviewit.You
shouldrequesttrainingassistancefromGEHealthcare,ifneeded.

IntendedAudience

Thismanualisintendedforusebythepersonwhoisresponsibleforconguringthe GEHealthcaredevicefortheOptimaEMSsystem(alsoreferredtoas"thesystem" throughoutthismanual).

ManualPurpose

DocumentConventions

TypographicalConventions
2069773-005D
ThisdocumentprovidesadetaileddescriptionofhowtocongureyourGEHealthcare device(MAC800,MAC1600,MAC5500/3500)toconnecttoyoursystemfordata transmission.
Thismanualusesthefollowingconventions.
ConventionDescription BoldTextIndicateskeysonthekeyboard,texttoenter,orhardwareitemssuchas
buttonsorswitchesontheequipment.
Italicized-Bold
Text
CTRL+ESC
Indicatessoftwaretermsthatidentifymenuitems,buttonsoroptionsin variouswindows.
Indicatesakeyboardoperation.Aplus(+)signbetweenthenamesoftwo keysindicatesthatwhileholdingtherstkey,youshouldpressandrelease thesecondkey.Forexample,PressCTRL+ESCmeanstopressandholdthe CTRLkeyandthenpressandreleasetheESCkey.
OptimaEMSECGManagementSystem
11
Page 12
Introduction
ConventionDescription
<space>
Enter
>
Indicatesthatyoumustpressthespacebar.Wheninstructionsaregiven fortypingaprecisetextstringwithoneormorespaces,thepointwhere youmustpressthespacebarisindicatedas<space>.Thisensuresthat thecorrectnumberofspacesisinsertedinthecorrectpositionswithin theliteraltextstring.Thepurposeofthe<>bracketsistodistinguishthe commandfromtheliteraltextwithinthestring.
IndicatesthatyoumustpresstheEnterorReturnkeyonthekeyboard.Do nottypeEnter.
Thegreaterthansymbol,orrightanglebracket,isaconcisemethodto indicateasequenceofmenuselections.
Forexample,thestatement“Fromthemainmenu,selectSystem>Setup> OptionstoopentheOptionActivationwindow”replacesthefollowing:
1.Fromthemainmenu,selectSystemtoopentheSystemmenu.
2.FromtheSystemmenu,selectSetuptoopentheSetupmenu.
3.FromtheSetupmenu,selectOptionstoopentheOptionActivation window.
Illustrations
Allillustrationsinthemanualareprovidedasexamplesonly.Dependingonsystem conguration,screensinthemanualmaydifferfromthescreensonyoursystem.
Allpatientnamesanddataarectitious.Anysimilaritytoactualpersonsis coincidental.
Notes
Notesprovideapplicationtipsoradditionalinformationthat,whileuseful,arenot essentialtothecorrectoperationofthesystem.Theyarecalledoutfromthebody textthroughaagwordandindentation,asfollows:
NOTE:
ThetiporadditionalinformationisindentedbelowtheNOTEagword.

ResponsibilityofthePurchaser/Customer

Thecustomerisresponsibleforprovidingappropriatedesks,chairs,electricalwall outlets,networkconnections,analogphonelines,andforlocatinganyoftheOptima EMSSystemcomponentsdescribedinthismanualincompliancewithalllocal,state ornationalcodes.

AdditionalAssistance

GEHealthcaremaintainsatrainedstaffofapplicationandtechnicalexpertstoanswer questionsandrespondtoissuesandproblemsthatmayariseduringtheinstallation, maintenance,anduseofthissystem.
ContactyourlocalGEHealthcarerepresentativetorequestadditionalassistance.
12
OptimaEMSECGManagementSystem
2069773-005D
Page 13
MACCartsCongurationOverview
Carts,whichsupportOrderDownload,canretrieveordersfromtheOptimaEMS system.YoumustpurchaseOrderDownloadasacartoption.Thesystemrunsa threadforOrderDownloadforeachsite.CartsmustindicatethesiteID,fromwhich theordersaredownloaded.IfthesiteIDdoesnotmatchwiththethread,noorderis returned.
Thefollowingtableprovidesthemethodsavailablefortransferringdatafromthe Carttothesystem.Manyofthemethodslistedinthefollowingtablerequirethe purchaseandactivationofanoptionontheCart.RefertoyourGEHealthcaresales representativeformoreinformation.
ECGCartInterfacestotheSystem
Introduction
ECGCartCSI-LAN MAC800
Software Version1.X.X
MAC800 Software
Version2.X.X MAC1600 MAC3500 MAC5000 MAC5500
1
ThisoptionisavailableontheMAC3500onlyoutsideofNorthAmerica.
1
CSI-SerialSharedFolder
DCP
2069773-005D
OptimaEMSECGManagementSystem
13
Page 14
Introduction
14
OptimaEMSECGManagementSystem
2069773-005D
Page 15
ConnectingtheMAC1600/MAC800 SystemstotheOptimaEMSSystem
TherearethreewaystoconnecttheMAC1600/MAC800devicestotheOptima EMSsystem:
CSI-LAN
CSI-Serial
SharedFolder
ThefollowinginstructionsusetheMAC800asanexample,butboththeMAC1600 andtheMAC800UIarethesame.

SettingUptheAcq.Client

2
BeforesettinguptheconnectionbetweenthedeviceandtheOptimaEMSsystem, performthefollowingprocedureintheOptimaEMSsystemclienttocreateAcq.Client:
1.LogontheOptimaEMSsystem.
2.FromtheMainwindow,clickConguration>SystemConguration. TheSystemCongurationwindowopens.
2069773-005D
OptimaEMSECGManagementSystem
15
Page 16
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
3.SelectAcquisitionClienttoopentheAcquisitionClientpanel.
4.Click+andanewdevicenamedNewAcq.Clientlistedintheleftlist.
5.TypetheclientnameintheNameeld.
6.TypetheclientcomputerIPaddressintheIPAddresseld.
7.ClickOKtosavethemodication.

CSI-LANConnection

BeforeusingaCSI-LANconnection,pleasemakesurethattheLANM(LANtoMUSE Communication)orLANC(LANtoCadiosoftCommunication)optionofyourMAC800 devicehasbeenpurchasedandactivated.
ThefollowingstepsdescribehowtoactivatetheMAC800options.
1.PurchasetheLANM(LANtoMUSECommunication)orLANC(LANtoCardiosoft
Communication)optiontogettheoptioncode. PleasecontactGEHealthcaresupportfordetailedinformation.
2.PowerontheMAC800system.
3.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F6(More)
>F4(OptionsSetup). TheOptionsSetupwindowopens.
16
OptimaEMSECGManagementSystem
2069773-005D
Page 17
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
4.TypetheoptioncodeintheOptionCodeeld.
5.PressEntertoactivatetheoption.

SettingUptheDevice

NOTE:
Beforesettingupthedevice,pleasemakesurethattheEthernetcablehasbeen pluggedintothecablesocketofthedevice.
1.PowerontheMAC800device.
2.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F1 (CommunicationSetup).
TheDataCommunicationSettingswindowopens.
2069773-005D
OptimaEMSECGManagementSystem
17
Page 18
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
3.PressF4(PageDown)repeatedlyuntiltheDataCommunicationLocations windowopens.
4.TypethelocationnameintheLocationeld.
5.AttheDeviceeld,selectLANfromthedrop-downmenu.
6.AttheProtocoleld,selectCSIfromthedrop-downmenu.
7.PressF4(PageDown). TheWiredLANSettingswindowopens.
18
OptimaEMSECGManagementSystem
2069773-005D
Page 19
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
8.RecordthedataintheSerial/IPRedirectorListenPorteldandtheIPAddress eld,whichwillbeusedintheOptimaEMSsystemsettings.
9.PressF6(Save)tocompletethecongurationsoftheMAC800device.

SettingUptheOptimaEMSSystemClient

1.LogontotheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration. TheSystemCongurationwindowopens.
3.SelectDevicestoopentheDevicespanel.
4.Clickthe+icon. AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,MAC800CSI.
DeviceType:selectECG.
Manufacture:selectGE.
Connection:selectCSI.
OnceCSIisselected,theCSICongurationpanelopens.
•SelectLANfortheconnection.
•SelectviaServer/viaAcq.Clientaccordingtothefollowingrules
viaServer:thedeviceissettoconnecttotheOptimaEMSServerviaLAN
viaAcq.Client:thedeviceissettoconnecttotheOptimaEMS AcquisitionClientviaLAN
•SelectoneAcq.Clientfromthedrop-downlistifviaAcq.Clientisselected.
2069773-005D
OptimaEMSECGManagementSystem
19
Page 20
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
Referto“SettingUptheAcq.Client”onpage15fortheinstructionsof creatingAcq.Client.
•TypetheIPAddressandDevicePortyourecordedinStep8in“Setting
UptheDevice”onpage17.
6.ClickOKtosavethemodication. YoumaynowtransmitdatafromtheMAC800devicetotheOptimaEMSsystemvia
aCSI-LANconnection.

CSI-SerialConnection

BeforeusingaCSI-Serialconnection,pleasemakesurethattheMODM(Modemor SerialtoMuse)optionofyourMAC800devicehasbeenpurchasedandactivated.
ThefollowingstepsdescribeshowtoactivatetheMODMoptiononyourMAC800 device.
1.PurchasetheMODM(ModemorSerialtoMuse)optiontogettheoptioncode. PleasecontactGEHealthcaresupportfordetailedinformation.
2.PowerontheMAC800system.
3.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F6(More) >F4(OptionsSetup). TheOptionsSetupwindowopens.
4.TypetheoptioncodeintheOptionCodeeld.
5.PressEntertoactivatetheoption.
20
OptimaEMSECGManagementSystem
2069773-005D
Page 21

SettingUptheDevice

ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
NOTE:
Beforesettingupthedevice,pleasemakesurethattheserialcablehasbeen pluggedintothecablesocketofthedevice.
1.PowerontheMAC800device.
2.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F1
(CommunicationSetup). TheDataCommunicationSettingswindowopens.
3.RecordthedataintheSerialBaudRateeld,whichwillbeusedintheOptima
EMSsystemsettings.
2069773-005D
OptimaEMSECGManagementSystem
21
Page 22
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
4.PressF4(PageDown)repeatedlyuntiltheDataCommunicationLocations
windowopens.
5.TypethelocationnameinLocationeld.
6.AttheDeviceeld,selectSerialfromthedrop-downmenu.
7.AttheProtocoleld,selectCSIfromthedrop-downmenu.
8.PressF6(Save)tocompletethecongurationsintheMAC800device.

SettingUptheOptimaEMSSystemClient

1.LogontotheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration.
TheSystemCongurationwindowopens.
22
OptimaEMSECGManagementSystem
2069773-005D
Page 23
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
3.SelectDevicestoopentheDevicespanel.
4.Clickthe+icon.
AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,CSI.
DeviceType:selectECG.
Manufacture:selectGE.
Connection:selectCSI. OnceCSIisselected,theCSICongurationpanelopens.
•SelectSerialfortheconnection.
•SelectviaServer/viaAcq.Clientaccordingtothefollowingrules
viaServer:thedeviceissettoconnecttotheOptimaEMSServervia COMPort
viaAcq.Client:thedeviceissettoconnecttotheOptimaEMS AcquisitionClientviaCOMPort
•SelectoneAcq.Clientfromthedrop-downlistifviaAcq.Clientisselected. Referto“SettingUptheAcq.Client”onpage15fortheinstructionsof creatingAcq.Client.
•ChoosetheCOMPortwhichdependsontheCOMportoftheMAC800 deviceconnectingtotheOptimaEMSsystem.
•TypetheBaudrateyourecordedinStep3in“SettingUptheDevice”on
page21.
6.ClickOKtosavethemodication. YoumaynowtransmitdatafromtheMAC800devicetotheOptimaEMSsystemvia
aCSI-Serialconnection.
2069773-005D
OptimaEMSECGManagementSystem
23
Page 24
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem

SharedFolderConnection

BeforeusingaSharedFolderconnection,pleasemakesurethattheLANC(LAN toCardiosoftCommunication)optionofyourMAC800devicehasbeenpurchased andactivated.
ThefollowingstepsdescribeshowtoactivatetheLANCoptionattheMAC800device.
1.PurchasetheLANC(LANtoCardiosoftCommunication)optiontogettheoption code. ContactGEHealthcaresupportfordetailedinformation.
2.PowerontheMAC800system.
3.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F6(More) >F4(OptionsSetup). TheOptionsSetupwindowopens.
4.TypetheoptioncodeintheOptionCodeeld.
5.PressEntertoactivatetheoption.
24
OptimaEMSECGManagementSystem
2069773-005D
Page 25

SettingUptheDevice

ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem
1.PowerontheMAC800device.
2.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F1
(CommunicationSetup). TheDataCommunicationSettingswindowopens.
3.SelecttheAllowExportusingSharedDirectorycheckbox.
Thedetailedsettingsareenabled.
4.TypetheShareNameaccordingtothefollowingrules:
•IftheConnectionsettingintheOptimaEMSsystemisviaServer,perform thefollowing:
•Ifyourdevicesoftwareversionis1.X.X,type\\[ServerComputer
Name]\[DeviceName]
•Ifyourdevicesoftwareversionis2.X.X,type\\[ServerComputer
Name]\[SiteName]\[DeviceName]
•IftheConnectionsettingintheOptimaEMSsystemisviaAcq.Client, performthefollowing:
•Ifyourdevicesoftwareversionis1.X.X,type\\[ClientComputer
Name]\[DeviceName]
•Ifyourdevicesoftwareversionis2.X.X,type\\[ClientComputer
Name]\[SiteName]\[DeviceName]
5.TypetheUsernameandPasswordwhichyouusedtologontoyourPC.
6.ConrmyourpasswordintheConrmeld.
7.PressF6(Save)tosavethemodication.
2069773-005D
OptimaEMSECGManagementSystem
25
Page 26
ConnectingtheMAC1600/MAC800SystemstotheOptimaEMSSystem

SettingUptheOptimaEMSSystemClient

1.LogontheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration. TheSystemCongurationwindowopens.
3.SelectDevicetoopentheDevicepanel.
4.Clickthe+icon. AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,Sharetest.
DeviceType:selectECG.
Manufacture:selectGE.
Connection:selectSharedFolder.
OnceSharedFolderisselected,theSharedFolderCongurationpanel opens.
•SelectviaServer/viaAcq.Clientaccordingtothefollowingrules
viaServer:thedevicetransmitsdatadirectlytotheOptimaEMSServer
viaAcq.Client:thedevicetransmitsdatadirectlytotheOptimaEMS AcquisitionClient
•SelectoneAcq.Clientfromthedrop-downlistifviaAcq.Clientisselected. Referto“SettingUptheAcq.Client”onpage15fortheinstructionsof creatingAcq.Client.
6.ClickOKtosavethemodication. YoumaynowtransmitdatafromtheMAC800devicetotheOptimaEMSsystem
viaaSharedFolderconnection.
26
OptimaEMSECGManagementSystem
2069773-005D
Page 27
3

DCPConnection

IftheMAC800devicesoftwareversionis2.X.X,youcanconnectthedevicetothe OptimaEMSsystemviaDCP .
Thereare2connectiontypesforDCP:
WiredLAN(LANM)
WirelessLAN(WIFM)
BeforeusingtheDCPconnection,pleasemakesurethatLANM(LANtoMUSE Communication)orWIFM(WiFitoMUSECommunication)optionofyourMAC800 devicehasbeenpurchasedandactivated.
ThefollowingstepsdescribeshowtoactivatetheLANMandWIFMoptionsofthe MAC800device.
1.PurchasetheLANM(LANtoMUSECommunication)orWIFM(WiFitoMUSE Communication)optiontogettheoptioncode. ContactGEHealthcaresupportfordetailedinformation.
2.PowerontheMAC800system.
3.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F6(More) >F4(OptionsSetup). TheOptionsSetupwindowopens.
2069773-005D
OptimaEMSECGManagementSystem
27
Page 28
DCPConnection
4.TypetheoptioncodeintheOptionCodeeld.
5.PressEntertoactivatetheoption.

DCPWiredLANSettings

SettingUptheOptimaEMSSystemClient

1.LogontheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration.
TheSystemCongurationwindowopens.
3.SelectDevicestoopentheDevicespanel.
28
OptimaEMSECGManagementSystem
2069773-005D
Page 29
4.Clickthe+icon.
AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
DCPConnection
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,DCPServer.
DeviceType:selectECG.
Manufacture:selectGE.
Connection:selectDCP. OnceDCPisselected,theDCPCongurationpanelopensundertheBasic Settingspanel.
•SelectviaServerforLANconnection.
•TypeaportnumberintheDCPPorteld.
NOTE:
Theportnumbershouldnotbeoccupiedbyotherapplicationonthe server.
6.ClickOKtosavethemodication.
7.Onthedesktop,selectStart>Run. TheRunwindowopens.
2069773-005D
OptimaEMSECGManagementSystem
29
Page 30
DCPConnection
8.TypecmdintheOpeneld.
9.ClickOKtoopentheCommandPromptwindow.
10.TypeipcongandpressEnter.
11.RecordtheIPv4Addressdisplayedinthewindowforfurtheruse.

SettingUptheDevice

30
NOTE:
Beforesettingupthedevice,pleasemakesurethattheserialcablehasbeen pluggedintothecablesocketofthedevice.
1.PowerontheMAC800device.
2.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F1 (CommunicationSetup).
TheDataCommunicationSettingswindowopens.
OptimaEMSECGManagementSystem
2069773-005D
Page 31
3.PressF4(PageDown)repeatedlyuntiltheDataCommunicationLocations windowopens.
DCPConnection
4.TypethelocationnameinLocationeld.
5.AttheDeviceeld,selectLANfromthedrop-downmenu.
6.AttheProtocoleld,selectCSIfromthedrop-downmenu.
7.PressF4(PageDown)repeatedlyuntiltheDCPSettingswindowopens.
8.TypetheIPAddressandPortyourecordedinStep11of“SettingUptheOptima
EMSSystemClient”onpage28.
9.PressF6(Save)tosavethemodication.
2069773-005D
OptimaEMSECGManagementSystem
31
Page 32
DCPConnection

DCPWirelessLANSettings

SettingUptheOptimaEMSSystemClient

Pleasereferto“SettingUptheOptimaEMSSystemClient”onpage28in“DCPWired
LANSettings”.

SettingUptheDevice

1.ConnecttheWiFimoduletotheMAC800device.
2.PowerontheMAC800device.
3.FromtheMainMenu,pressF4(SystemConguration)>F6(More)>F1 (CommunicationSetup).
TheDataCommunicationSettingswindowopens.
32
OptimaEMSECGManagementSystem
2069773-005D
Page 33
4.PressF4(PageDown)repeatedlyuntiltheDataCommunicationLocations windowopens.
DCPConnection
5.TypethelocationnameinLocationeld.
6.AttheDeviceeld,selectLANfromthedrop-downmenu.
7.AttheProtocoleld,selectDCPfromthedrop-downmenu.
8.PressF4(PageDown)repeatedlyuntiltheWirelessNetworkingSettings windowopens.
9.SelecttheEnableWirelessLANcheckbox.
10.TypetheNetworkName(SSID). Pleasecontactyournetworkadministratorforadditionaldetails.
11.PressF4(PageDown)repeatedlyuntiltheDCPSettingswindowopens.
2069773-005D
OptimaEMSECGManagementSystem
33
Page 34
DCPConnection
12.TypetheIPAddressandPortyourecordedinStep11of“SettingUptheOptima
EMSSystemClient”onpage28.
13.PressF6(Save)tosavethemodication.
34
OptimaEMSECGManagementSystem
2069773-005D
Page 35

ConnectingtheMAC1200System

ConnectingtheMAC1200SystemtotheOptima EMSSystemviaCSI-Serial
ThesectiondescribestheconnectionmethodbetweenMAC1200deviceandthe OptimaEMSsystem.

SettingUptheMAC1200System

1.PowerontheMAC1200device.
2.FromtheMainMenu,pressCong. TheCongurationwindowopens.
4
3.UsetheDownkeytohighlightAutomaticandpressEnter. TheConguration:Automaticwindowopens.
4.PresstheDownkeyrepeatedlytohighlightManual copy to. MakesuretheHOSTishighlightedforthiseld.Ifnot,usetheLeftkeyto
highlightHOSTandpressEnter.
5.PresstheRecord/StopkeytoreturntotheCongurationwindow. Thesystemwillautomaticallysavethemodication.
6.IntheCongurationwindow,usetheDownkeytohighlightCommunication andpressEnter.
TheConguration:Communicationwindowopens.
7.UsetheArrowkeystomakesuretheeldslisttheexactvaluesasfollows:
Baudrate(PC):57600
Protocol:CSI
Modem:no
2069773-005D
OptimaEMSECGManagementSystem
35
Page 36
ConnectingtheMAC1200System

RequirementsforConnectingtheSerialLine

NOTE:
GEHealthcarerecommendsyoutobuystandardseriallinefromGEHealthcare (PN22336203).
RefertothefollowinggraphicfordetailedinformationaboutconnectingMAC1200to PCwithserialline.

SettingUptheOptimaEMSSystemClient

1.LogontotheOptimaEMSsystemclient.
2.FromtheMainwindow,selectConguration>SystemConguration. TheSystemCongurationwindowopens.
36
OptimaEMSECGManagementSystem
2069773-005D
Page 37
3.SelectDevicestoopentheDevicespanel.
ConnectingtheMAC1200System
4.Clickthe icon. ANewDeviceiscreatedandlistedintheDeviceslist.
5.InBasicSettings,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,MAC1200CSI.
DeviceType:ECG.
Manufacturer:GE.
Connection:CSI.
AfterCSIisselected,theCSICongurationpanelopens,addthefollowing information:
•Connection:SerialLineandviaServer
COMPort:COM1
Baudrate:57.6K
6.ClickOKtosavethemodication. YoumaynowtransmitdatafromtheMAC1200devicetotheOptimaEMS
systemviaaCSI-Serialconnection.

VerifytheConnection

1.SetuptheMAC1200deviceaccordingto“SettingUptheMAC1200System”
onpage35.
2.ConnecttheMAC1220deviceandtheOptimaEMSsystemclientwithaserial lineaccordingto“RequirementsforConnectingtheSerialLine”onpage36.
3.EnterthepatientinformationontheMAC1200device.
4.PresstheRecord/StopkeytoacquiretheECGdata.
2069773-005D
OptimaEMSECGManagementSystem
37
Page 38
ConnectingtheMAC1200System
5.PresstheTransmitkeytostarttransmittingtheECGdata.
6.Duringtransmission,thefollowingprogressbarswillbedisplayed:
Initializingtransmission…
ECGtransmission(CSI)…
Thetransmissionissuccessfullynishedwhentheprogressbarsclose. YoushouldnowbeabletoaccessthetransmittedECGdataontheOptima
EMSsystemclient.
38
OptimaEMSECGManagementSystem
2069773-005D
Page 39
ConnectingtheMAC3500/MAC 5000/MAC5500Systemstothe OptimaEMSSystem
TherearethreewaystoconnecttheMAC3500/MAC5000/MAC5500devicestothe OptimaEMSsystem:
CSI-LAN
NOTE:
CSI-LANConnectionisnotavailableforMAC5000device.
CSI-Serial
ThefollowinginstructionsusetheMAC3500asanexample,buttheMAC3500,MAC 5000andtheMAC5500UIarethesame.
5

CSI-LANConnection

SettingUptheDevice

2069773-005D
BeforeusingaCSI-LANconnection,pleasemakesurethattheELANoptionofyour MAC3500devicehasbeenpurchasedandactivated.
ContactGEHealthcaresupportfordetailedinformationiftheoptionisnotpurchased oractivated.
1.Fromthemainwindow,selectMainMenu>SystemSetup.
2.Typethesystempassword.
3.SelectBasicSystem>Transmission.
OptimaEMSECGManagementSystem
39
Page 40
ConnectingtheMAC3500/MAC5000/MAC5500SystemstotheOptimaEMSSystem
4.IntheDefaultLocationeld,selectEthernet(MUSE).
DefaultLocation:
#1 #2 #3 #4 #5 #6 Serialline(MUSE) Serialline(ASCIIout) Serialline(XMLout)
Ethernet(MUSE)
None
5.PressESCtoreturntothemainwindow.
6.SelectBasicSystem>NetworkSetup. TheNetworkSetupwindowopens.
7.TypetheIPAddress,SubnetMask,GatewayandPortNumber.
NetworkSetup IPAddress SubnetMask Gateway PortNumber Return
:___.___.___.___ :___.___.___.___ :___.___.___.___ :
____
NOTE:
RecordtheIPAddressandPortNumberwhichwillbeusedinOptimaEMS systemsetup.
8.ForMAC3500/5000/5500,youarerequesttosaveallyourmodication.

SettingUptheOptimaEMSSystemClient

1.LogontotheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration. TheSystemCongurationwindowopens.
40
OptimaEMSECGManagementSystem
2069773-005D
Page 41
ConnectingtheMAC3500/MAC5000/MAC5500SystemstotheOptimaEMSSystem
3.SelectDevicestoopentheDevicespanel.
4.Clickthe+icon. AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,MAC3500CSI.
DeviceType:selectECG.
Manufacture:selectGE.
Connection:selectCSI.
OnceCSIisselected,theCSICongurationpanelopens.
•SelectLANfortheconnection.
•SelectviaServer/viaAcq.Clientaccordingtothefollowingrules
viaServer:thedeviceissettoconnecttotheOptimaEMSServerviaLAN
viaAcq.Client:thedeviceissettoconnecttotheOptimaEMS AcquisitionClientviaLAN
•SelectoneAcq.Clientfromthedrop-downlistifviaAcq.Clientisselected. Referto“SettingUptheAcq.Client”onpage15fortheinstructionsof creatingAcq.Client.
•TypetheIPAddressandDevicePortyoutypedinStep7in“SettingUpthe
Device”onpage39.
6.ClickOKtosavethemodication. YoumaynowtransmitdatafromthedevicetotheOptimaEMSsystemviaaCSI-LAN
connection.
NOTE:
MakesureyouhavebothPatientNameandPatientIDforthereport,otherwise thetransmissionwillbefailed.
2069773-005D
OptimaEMSECGManagementSystem
41
Page 42
ConnectingtheMAC3500/MAC5000/MAC5500SystemstotheOptimaEMSSystem

CSI-SerialConnection

SettingUptheDevice

1.Fromthemainwindow,selectMainMenu>SystemSetup.
2.Typethesystempassword.
3.SelectBasicSystem>Transmission.
4.InDefaultLocationeld,selectSerialline(MUSE).
DefaultLocation:
5.ForMAC5500,youarerequesttosaveallyourmodication.

SettingUptheOptimaEMSSystemClient

1.LogontotheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration.
#1 #2 #3 #4 #5 #6
Serialline(MUSE)
Serialline(ASCIIout) Serialline(XMLout) Ethernet(MUSE) None
TheSystemCongurationwindowopens.
3.SelectDevicestoopentheDevicespanel.
42
OptimaEMSECGManagementSystem
2069773-005D
Page 43
ConnectingtheMAC3500/MAC5000/MAC5500SystemstotheOptimaEMSSystem
4.Clickthe+icon. AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,MAC3500CSI.
DeviceType:selectECG.
Manufacture:selectGE.
Connection:selectCSI.
OnceCSIisselected,theCSICongurationpanelopens.
•SelectSerialLinefortheconnection.
•SelectviaServer/viaAcq.Clientaccordingtothefollowingrules
viaServer:thedeviceissettoconnecttotheOptimaEMSServervia COMPort
viaAcq.Client:thedeviceissettoconnecttotheOptimaEMS AcquisitionClientviaCOMPort
•SelectoneAcq.Clientfromthedrop-downlistifviaAcq.Clientisselected. Referto“SettingUptheAcq.Client”onpage15fortheinstructionsof creatingAcq.Client.
•ChoosetheCOMPortwhichdependsontheCOMportoftheMAC3500 deviceconnectingtotheOptimaEMSsystem.
•SelectBaudratewhichisthesameasthebaudrateofMAC3500.
6.ClickOKtosavethemodication. YoumaynowtransmitdatafromthedevicetotheOptimaEMSsystemviaaCSI-Serial
connection.
NOTE:
MakesureyouhavebothPatientNameandPatientIDforthereport,otherwise thetransmissionwillbefailed.
2069773-005D
OptimaEMSECGManagementSystem
43
Page 44
ConnectingtheMAC3500/MAC5000/MAC5500SystemstotheOptimaEMSSystem
44
OptimaEMSECGManagementSystem
2069773-005D
Page 45
CongurationforPDFImport
ThissectiondescribesthecongurationsforGEHealthcareStressandHolterPDF Import.

SettingUpSharedFolder

ToimporttheStress/Holterreport,youneedtosetupasharedfolderontheOptima EMSsystemserverbeforeyousetupinthesystemclient.
1.Beforesettingupthesharedfolder ,youneedtoperformthefollowingprocedure tocreatealocaluseronyourcomputer.
a.Right-clickonMyComputerandselectManage.
TheComputerManagementwindowopens.
b.ClickSystemTools>LocalUsersandGroups>UserstoopentheUsers
window.
6
2069773-005D
OptimaEMSECGManagementSystem
45
Page 46
CongurationforPDFImport
c.Right-clickintheUserswindowandselectNewUser....
TheNewUserwindowopens.
d.CompletethefollowinginformationintheNewUserwindow.
Username:typepdmport.
Password:typePdf!Import.
46
OptimaEMSECGManagementSystem
2069773-005D
Page 47
CongurationforPDFImport
•ConrmPassword:typePdf!Import.
•UnselectthecheckboxofUsermustchangepasswordatnextlogon.
e.ClickCreateandthenewuseriscreatedandlistedintheUserwindow. f.ClickClosetoclosetheNewUserwindowandreturnbacktotheUser
window. Youhavesuccessfullycreatedanewuserforyourcomputer.
2.Createanewfolderonyourcomputer. C:\Stressfolderiscreatedasanexampleinthissection. YoucancreateafoldernamedwithHolterifyouaretransmittingHolterreport.
2069773-005D
OptimaEMSECGManagementSystem
47
Page 48
CongurationforPDFImport
3.Right-clickthefoldericonandselectSharewith>Specicpeople....
TheFileSharingwindowopens.
48
OptimaEMSECGManagementSystem
2069773-005D
Page 49
4.ChooseFindpeople...fromthedrop-downmenu.
CongurationforPDFImport
TheSelectUsersorGroupswindow
5.ClickLocations...attheFromthislocation:eld.
TheLocationswindowopens.
2069773-005D
OptimaEMSECGManagementSystem
49
Page 50
CongurationforPDFImport
6.SelectyourcomputernameintheLocation:eld.
7.ClickOK,thelocationisselectedintheSelectUsersorGroupswindow.
8.TypepdmportintheEntertheobjectnamestoselect(example):eldand clickCheckNames.
50
OptimaEMSECGManagementSystem
2069773-005D
Page 51
Theuserpdmportwillbelistedintheeld.
9.ClickOK.
CongurationforPDFImport
TheuserpdmportwillbelistedintheFileSharingwindow
2069773-005D
OptimaEMSECGManagementSystem
51
Page 52
CongurationforPDFImport
10.AtthePermissionLevelcolumn,SelectRead/Writeforpdmport.
11.ClickShare. TheYourfolderisshared.windowopens,andthefolderisnowsuccessfully
shared.
52
OptimaEMSECGManagementSystem
2069773-005D
Page 53
CongurationforPDFImport
ConguringtheStressPDFReportImportinthe OptimaEMSSystemClient
1.LogontotheOptimaEMSsystemclient.
2.FromtheMainwindow,clickConguration>SystemConguration. TheSystemCongurationwindowopens.
3.SelectDevicestoopentheDevicespanel.
4.Clickthe+icon. AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,Stress.
DeviceType:selectStress.
Manufacture:selectGE.
Connection:selectSharedFolder.
OnceSharedFolderisselected,theSharedFolderCongurationpanel opens.
•SelectviaServerfortheconnection.
•TypethepaththatyouusedtoshareinSharedFolderPatheld. Forexample,\\3.35.24.39\Stress.
2069773-005D
OptimaEMSECGManagementSystem
53
Page 54
CongurationforPDFImport
6.ClickOKtosavethemodication.
7.Toverifythecongurationiscorrect,performthefollowingactions: a.CopythestressPDFreportsampleintheStressfolderfromOptimaEMS
b.OpentheOrderListwindowoftheOptimaEMSsystemtocheckifyou
c.Double-clicktheStressorder,thePDFreportwillbedisplayedinthe
InstallationDVDtothesharedfolderyouhavecongured(forexample, \\3.35.24.39\Stress).
canndtheStressorderinthelist.
Diagnosiswindow.
ConguringtheHolterPDFReportImportinthe OptimaEMSSystem
1.LogontotheOptimaEMSsystem.
2.FromtheMainwindow,clickConguration>SystemConguration. TheSystemCongurationwindowopens.
54
OptimaEMSECGManagementSystem
2069773-005D
Page 55
3.SelectDevicestoopentheDevicespanel.
CongurationforPDFImport
4.Clickthe+icon. AnewdevicenamedNewDeviceiscreatedandlistedinthedeviceslist.
5.AttheBasicSettingseld,addthefollowinginformation:
DeviceName:typetheECGdevicename,forexample,Holter.
DeviceType:selectHolter.
Manufacture:selectGE.
Connection:selectSharedFolder.
OnceSharedFolderisselected,theSharedFolderCongurationpanel opens.
•SelectviaServerfortheconnection.
•TypethepaththatyouusedtoshareinSharedFolderPatheld. Forexample,\\3.35.24.39\Holter.
2069773-005D
OptimaEMSECGManagementSystem
55
Page 56
CongurationforPDFImport
6.ClickOKtosavethemodication.
7.Toverifythecongurationiscorrect,performthefollowingactions: a.CopytheHolterPDFreportsampleintheHolterfolderfromOptimaEMS
b.OpentheOrderListwindowoftheOptimaEMSsystemtocheckifyou
c.Double-clicktheHolterorder,thePDFreportwillbedisplayedinthe
InstallationDVDtothesharedfolderyouhavecongured(forexample, \\3.35.24.39\Holter).
canndtheHolterorderinthelist.
Diagnosiswindow.
56
OptimaEMSECGManagementSystem
2069773-005D
Page 57
Page 58
GEMedicalSystems InformationTechnologies,Inc. 8200WestTowerAvenue Milwaukee,WI53223USA Tel:+14143555000
+18005587044(USOnly)
Fax:+14143553790
AsiaHeadquarters
GEMedicalSystems InformationTechnologies,Inc. Asia;GE(China)Co.,Ltd. 1HuatuoRoad ZhangjiangHi-techParkPudong Shanghai,People'sRepublicofChina201203 Tel:+862138777888 Fax:+862138777451
GEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompany,goingtomarketas GEHealthcare.
www.gehealthcare.com
Loading...