GEHealthcare
MAC™800
RestingECGAnalysisSystem
ServiceManual
SoftwareVersion2.0.x
2060321-001RevisionL
MAC800RestingECGAnalysisSystem
English
© 2012-2017GeneralElectricCompany.
AllRightsReserved.
PublicationInformation
TheinformationinthismanualappliesonlytoMAC™800SoftwareVersion2.0.x.Itdoesnotapplytoearlierproductversions.Dueto
continuingproductinnovation,specicationsinthismanualaresubjecttochangewithoutnotice.
CardioSoft,CASE,HookupAdvisor ,MAC,Mactrode,MUSEand12SLaretrademarksownedbyGEMedicalSystemsInformationTechnologies,
Inc.,aGeneralElectricCompanygoingtomarketasGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheir
respectiveowners.
Thedocumentpartnumberandrevisionappearatthebottomofeachpage.Therevisionidentiesthedocument’supdatelevel.The
revisionhistoryofthisdocumentissummarizedinthefollowingtable.
Revision
A
B
C 31May2012
Date
1March2012 Initialreleaseofdocument.
21March2012 UpdatedtheProductSpecications:Certicationssection.
Comments
SPRHCSDM00135660thefollowingupdatesare:
•IntheIntroductionChapter,underSafetyHazards,awarningregardingthe
maintenanceschedulewasadded.
•IntheTroubleshootingChapter ,undertheEquipmentProblemssection,
SystemDate/TimeSettingssectionwasadded.
D
16July2012 PerSPRHCSDM00149939thefollowingchangesweremade:
UpdatedPartNumbersoffollowingparts:
•MAC800FRULCDASSEMBL Y
•MAC800V2FRUCDMANUAL
•MAC800V2FRUPROGRAMMEDSDCARD
•AllMAC800DataMatrixBarcodeScannerKitpartnumberswerechanged.
FRUWiChannelsectionwasaddedtothePartsListchapter.
IntheTroubleshootingChaptertheStart-upscreenwasupdated.
E
F
G
13June2013
4August2014
5February2015 Thefollowingchangesweremade:
PerSPRHCSDM00206148,thefollowingchangeismade:
Replaced2061821-001MAC800V2FRUPROGRAMMEDSDCARDwith
2061821-002MAC800V2FRUPROGRAMMEDSDCARD-SW2.0.4
Thefollowingchangeismade:
Replaced2061821-002MAC800V2FRUPROGRAMMEDSDCARD-SW2.0.4with
2061821-003MAC800V2FRUPROGRAMMEDSDCARD-SW2.0.5
•Replaced2061821-003MAC800V2FRUPROGRAMMEDSDCARD-SW2.0.5
with2061821-004MAC800V2FRUPROGRAMMEDSDCARDwithSW2.0.6
•Removed”whereverECGtestingisperformedtorecordECGsignals
fromsurfaceelectrodes”fromtheIndicationsforUsesectioninChapter
Instruction.
H
J
K
L
27June2016
9October2016 Replaced2061821-004MAC800V2FRUPROGRAMMEDSDCARDwithSW2.0.6
17April2017 Addedthenewsuppliesandaccessoriesmanualinformation.
6November2017 UpdatedtophaseinSilexwirelessmodule.
UpdatetheWiparameterinformation.
with2061821-005MAC800V2FRUPROGRAMMEDSDCARD-SW2.0.8.
ToaccessotherGEHealthcaremanuals,gototheCommonDocumentationLibrary(CDL),locatedathttp://www.gehealthcare.com/usen/
service/biomed_tech_selfservice/services_user_doc/products/support.htmlandclickCardiology.
ToaccessOriginalEquipmentManufacturer(OEM)manuals,gotothedevicemanufacturer'swebsite.
2 MAC™800 2060321-001L
6November2017
ServiceManualLanguageInformation
WARNING
(EN)
ПРЕДУПРЕЖДЕНИЕ
(BG)
警告
(ZH-CN)
警告
(ZH-TW)
UPOZORENJE
(HR)
VAROVÁNÍ
(CS)
ThisservicemanualisavailableinEnglishonly.
•Ifacustomer'sserviceproviderrequiresalanguageotherthanEnglish,itisthecustomer's
responsibilitytoprovidetranslationservices.
•Donotattempttoservicetheequipmentunlessthisservicemanualhasbeenconsulted
andisunderstood.
•Failuretoheedthiswarningmayresultininjurytotheserviceprovider,operator,orpatient,
fromelectricshock,mechanicalorotherhazards.
Товаупътванезаработаеналичносамонаанглийскиезик.
•Акодоставчикътнауслугатанаклиентаизискадругезик,задължениенаклиентае
даосигурипревод.
•Неизползвайтеоборудването,предидастесеконсултиралииразбралиупътването
заработа.
•Неспазванетонатовапредупреждениеможедадоведедонараняваненадоставчика
науслугата,оператораилипациентврезултатнатоковударилимеханичнаили
другаопасност.
本维修手册仅提供英文版本。
•如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
•未详细阅读和完全理解本维修手册之前,不得进行维修。
•忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
本維修手冊只提供英文版。
•如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責提供翻
譯服務。
•除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維修。
•不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或其他因素
而受到傷害。
Oveuputezaservisiranjedostupnesusamonaengleskomjeziku.
•Ukolikokorisničkiserviszahtijevanekidrugijezik,korisnikovajeodgovornostosigurati
odgovarajućiprijevod.
•Nemojtepokušavatiservisiratiopremuukolikonistekonzultiraliirazumjelioveupute.
•Nepoštivanjeovogupozorenjamožerezultiratiozljedamaservisnogosoblja,korisnikaili
pacijentaprouzročenimelektričnimudaromtemehaničkimilinekimdrugimopasnostima.
Tentoprovoznínávodexistujepouzevanglickémjazyce.
•Vpřípadě,žeexterníslužbazákazníkůmpotřebujenávodvjinémjazyce,jezajištění
překladudoodpovídajícíhojazykaúkolemzákazníka.
•Nesnažteseoúdržbutohotozařízení,anižbystesipřečetlitentoprovoznínávoda
pochopilijehoobsah.
•Vpřípaděnedodržovánítétovarovánímůžedojítkporaněnípracovníkaprodejního
servisu,obslužnéhopersonálunebopacientůvlivemelektrickéhoproudu,respektive
vlivemmechanickýchčijinýchrizik.
2060321-001L MAC™800
3
ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(DA)
Denneservicemanualndeskunpåengelsk.
•Hvisenkundesteknikerharbrugforetandetsprogendengelsk,erdetkundensansvar
atsørgeforoversættelse.
•Forsøgikkeatservicereudstyretmedmindredenneservicemanualharværetkonsulteret
ogerforstået.
•Manglendeoverholdelseafdenneadvarselkanmedføreskadepågrundafelektrisk,
mekaniskellerandenfareforteknikeren,operatørenellerpatienten.
WAARSCHUWING
(NL)
DezeservicemanualisalleeninhetEngelsverkrijgbaar.
•Indienhetonderhoudspersoneeleenanderetaalnodigheeft,danisdeklant
verantwoordelijkvoordevertalingervan.
•Probeerdeapparatuurnietteonderhoudenvoordatdezeservicemanualgeraadpleegd
enbegrepenis.
•Indiendezewaarschuwingnietwordtopgevolgd,zouhetonderhoudspersoneel,de
gebruikerofeenpatiëntgewondkunnenrakenalsgevolgvaneenelektrischeschok,
mechanischeofanderegevaren.
HOIATUS
(ET)
Käesolevteenindusjuhendonsaadavalainultinglisekeeles.
•Kuiklienditeeninduseosutajanõuabjuhenditinglisekeelesterinevaskeeles,vastutab
klienttõlketeenuseosutamiseeest.
•Ärgeüritageseadmeidteenindadaenneeelnevaltkäesolevateenindusjuhendiga
tutvumistjasellestarusaamist.
•Käesolevahoiatuseeiraminevõibpõhjustadateenuseosutaja,operaatorivõipatsiendi
vigastamistelektrilöögi,mehaanilisevõimuuohutagajärjel.
VAROITUS
(FI)
Tämähuolto-ohjeonsaatavillavainenglanniksi.
•Josasiakkaanhuoltohenkilöstövaatiimuutakuinenglanninkielistämateriaalia,tarvittavan
käännöksenhankkiminenonasiakkaanvastuulla.
•Äläyritäkorjatalaitteistoaennenkuinoletvarmastilukenutjaymmärtänyttämän
huolto-ohjeen.
•Mikälitätävaroitustaeinoudateta,seurauksenavoiollahuoltohenkilöstön,laitteiston
käyttäjäntaipotilaanvahingoittuminensähköiskun,mekaanisenviantaimuun
vaaratilanteenvuoksi.
ATTENTION
(FR)
Cemanueltechniquen'estdisponiblequ'enanglais.
•Siunservicetechniqueclientsouhaiteobtenircemanueldansuneautrelangueque
l'anglais,ildevraprendreenchargelatraductionetlaresponsabilitéducontenu.
•Nepastenterd'intervenirsurleséquipementstantquelemanueltechniquen'apasété
consultéetcompris.
•Lenon-respectdecetavertissementpeutentraînerchezletechnicien,l'opérateuroule
patientdesblessuresduesàdesdangersélectriques,mécaniquesouautres.
4
MAC™800 2060321-001L
6November2017
ServiceManualLanguageInformation(cont'd.)
WARNUNG
(DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
AÐVÖRUN
(IS)
PERINGATAN
(ID)
DieseServiceanleitungistnurinenglischerSpracheverfügbar.
•FallsderKundendiensteineandereSprachebenötigt,musserfüreineentsprechende
Übersetzungsorgen.
•KeineWartungdurchführen,ohnedieseServiceanleitunggelesenundverstandenzu
haben.
•BeiZuwiderhandlungkanneszuVerletzungendesKundendiensttechnikers,des
AnwendersoderdesPatientendurchStromschläge,mechanischeodersonstigeGefahren
kommen.
Τοπαρόνεγχειρίδιοσέρβιςδιατίθεταιστααγγλικάμόνο.
•Εάντοάτομοπαροχήςσέρβιςενόςπελάτηαπαιτείτοπαρόνεγχειρίδιοσεγλώσσαεκτός
τωναγγλικών,αποτελείευθύνητουπελάτηναπαρέχειυπηρεσίεςμετάφρασης.
•Μηνεπιχειρήσετετηνεκτέλεσηεργασιώνσέρβιςστονεξοπλισμόεκτόςεάνέχετε
συμβουλευτείκαιέχετεκατανοήσειτοπαρόνεγχειρίδιοσέρβις.
•Εάνδενλάβετευπόψητηνπροειδοποίησηαυτή,ενδέχεταιναπροκληθείτραυματισμός
στοάτομοπαροχήςσέρβις,στοχειριστήήστονασθενήαπόηλεκτροπληξία,μηχανικούς
ήάλλουςκινδύνους.
Ezaszervizkézikönyvkizárólagangolnyelvenérhetőel.
•Haavevőszervizellátójaangoltóleltérőnyelvretartigényt,akkoravevőfelelőssége
afordításelkészíttetése.
•Nepróbáljaelkezdenihasználniaberendezést,amígaszervizkézikönyvbenleírtakat
nemértelmeztékésértettékmeg.
•Ezengyelmeztetésgyelmenkívülhagyásaaszervizellátó,aműködtetővagyapáciens
áramütés,mechanikaivagyegyébveszélyhelyzetmiattisérüléséteredményezheti.
Þessiþjónustuhandbókereingöngufáanlegáensku.
•Efaðþjónustuveitandiviðskiptamannsþarfnastannarstungumálsenensku,erþaðskylda
viðskiptamannsaðskaffatungumálaþjónustu.
•Reyniðekkiaðafgreiðatækiðnemaþessiþjónustuhandbókhefurveriðskoðuðogskilin.
•Brotáaðsinnaþessariaðvörungeturleitttilmeiðslaáþjónustuveitanda,stjórnandaeða
sjúklingifráraosti,vélrænumeðaöðrumáhættum.
ManualservisinihanyatersediadalambahasaInggris.
•JikapenyediajasaservispelangganmemerlukanbahasalainselaindariBahasaInggris,
merupakantanggungjawabdaripenyediajasaservistersebutuntukmenyediakan
terjemahannya.
•Janganmencobamelakukanservisterhadapperlengkapankecualitelahmembacadan
memahamimanualservisini.
•Mengabaikanperingataninibisamengakibatkancederapadapenyediaservis,operator ,
ataupasien,karenaterkenakejutlistrik,bahayamekanisataubahayalainnya.
2060321-001L MAC™800
5
ServiceManualLanguageInformation(cont'd.)
AVVERTENZA
(IT)
警告
(JA)
CẢ NHBÁO
(VI)
ЕСКЕРТУ
(KK)
BRĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
IlpresentemanualedimanutenzioneèdisponibilesoltantoinInglese.
•Seunaddettoallamanutenzionerichiedeilmanualeinunalinguadiversa,ilclienteè
tenutoaprovvederedirettamenteallatraduzione.
•Siprocedaallamanutenzionedell'apparecchiaturasolodopoaverconsultatoilpresente
manualeedavernecompresoilcontenuto.
•Ilnonrispettodellapresenteavvertenzapotrebbefarcompiereoperazionidacuiderivino
lesioniall'addetto,allamanutenzione,all'utilizzatoreedalpazienteperfolgorazione
elettrica,perurtimeccaniciodaltririschi.
このサービスマニュアルは英語版しかありません。
•サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
•このサービスマニュアルを熟読し、十分に理解をした上で装置のサービスを
行ってください。
•この警告に従わない場合、サービスを担当される方、操作員あるいは患者が、感電
や機械的又はその他の危険により負傷する可能性があります。
TàiLiệ uHướ ngDẫ nSử aChữ achỉcóbả ntiế ngAnh.
•Nế ucácđơ nvịcungcấ pdị chvụchokháchhàngyêucầ umộ tngônngữnàokháctiế ng
Anh,thìkháchhàngsẽcótráchnhiệ mcungcấ pcácdị chvụdị chthuậ t.
•Khôngđượ csử achữ athiế tbịtrừkhiđãthamkhả ovàhiể uTàiliệ uHướ ngdẫ nSử achữ a.
•Khôngtuânthủnhữ ngcả nhbáonàycóthểdẫ nđế ncáctổ nthươ ngchongườ ithự c
hiệ nsử achữ a,ngườ ivậ nhànhhaybệ nhnhân,dosố cđiệ n,cácrủ irovềcơkhíhay
cácrủ irokhác.
Бұ лқызметкө рсетубойыншанұ сқ аулығ ытекағ ылшынтіліндеқолжетімді.
•Тұ тынушыныңқызметпровайдеріағ ылшынтіліненбасқ атілдегінұ сқ аныталапетсе,
аударубойыншақызметтеріменқамтамасызетутұ тынушыжауапкершілігіндеболуы
тиіс.
•Бұ лқызметкө рсетубойыншанұ сқ аулығ ынназарғ аалып,тү сінбегенше,жабдыққ а
қ ызметкө рсетуденбастарты ң ыз.
•Бұ лескертудіелемеуқызметпровайдері,операторнемесеемделушініңэлектр
шогынан,механикалықнемесебасқ ақауіптернә тижесіндежарақ аталуынаәкелуі
мү мкін.
Šīapkalpotājurokasgrāmatairpieejamatikaiangļuvalodā.
•Jaapkalpošanassniedzējamnepieciešamainformācijacitā,nevisangļu,valodā,klienta
pienākumsirnodrošināttāstulkošanu.
•Neveicietaprīkojumaapkopi,neizlasotunnesaprototapkalpotājurokasgrāmatu.
•Šībrīdinājumaneievērošanavarradītelektriskāsstrāvastrieciena,mehāniskuvaiciturisku
izraisītutraumuapkopessniedzējam,operatoramvaipacientam.
Šiseksploatavimovadovasyraprieinamastikanglųkalba.
•Jeiklientopaslaugųtiekėjasreikalaujavadovokitakalba-neanglų,numatytivertimo
paslaugasyraklientoatsakomybė.
•Nemėginkiteatliktiįrangostechninėspriežiūros,nebentatsižvelgėteįšįeksploatavimo
vadovąirjįsupratote.
•Jeineatkreipsitedėmesioįšįperspėjimą,galimisužalojimaidėlelektrosšoko,mechaninių
arkitųpaslaugųtiekėjui,operatoriuiarpacientui.
6 MAC™800 2060321-001L
6November2017
ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(NO)
Denneservicehåndbokennnesbarepåengelsk.
•Hviskundensserviceleverandørtrengeretannetspråk,erdetkundensansvaråsørge
foroversettelse.
•Ikkeforsøkåreparereutstyretutenatdenneservicehåndbokenerlestogforstått.
•Manglendehensyntildenneadvarselenkanføretilatserviceleverandøren,operatøren
ellerpasientenskadespågrunnavelektriskstøt,mekaniskeellerandrefarer .
OSTRZEŻENIE
(PL)
Niniejszypodręcznikserwisowydostępnyjestjedyniewjęzykuangielskim.
•Jeślidostawcausługklientawymagajęzykainnegoniżangielski,zapewnienieusługi
tłumaczeniajestobowiązkiemklienta.
•Nienależyserwisowaćwyposażeniabezzapoznaniasięizrozumienianiniejszego
podręcznikaserwisowego.
•Niezastosowaniesiędotegoostrzeżeniamożespowodowaćurazydostawcyusług,
operatoralubpacjentawwynikuporażeniaelektrycznego,zagrożeniamechanicznego
bądźinnego.
AVISO
(PT-BR)
Estemanualdeassistênciatécnicasóseencontradisponíveleminglês.
•SeoserviçodeassistênciatécnicadoclientenãoforGE,eprecisardeoutroidioma,será
daresponsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanualde
assistênciatécnica.
•Onãocumprimentodesteavisopodeporemperigoasegurançadotécnico,operadorou
pacientedevidoachoqueselétricos,mecânicosououtros.
AVISO
(PT-PT)
Estemanualtécnicosóseencontradisponíveleminglês.
•Seaassistênciatécnicadoclientesolicitarestesmanuaisnoutroidioma,éda
responsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanual
técnico.
•Onãocumprimentodesteavisopodeprovocarlesõesaotécnico,aoutilizadorouao
pacientedevidoachoqueseléctricos,mecânicosououtros.
AVERTISMENT
(RO)
Acestmanualdeserviceestedisponibilnumaiînlimbaengleză.
•Dacăunfurnizordeserviciipentruclienţinecesităoaltălimbădecâtceaengleză,estede
datoriaclientuluisăfurnizezeotraducere.
•Nuîncercaţisăreparaţiechipamentuldecâtulteriorconsultăriişiînţelegeriiacestui
manualdeservice.
•Ignorareaacestuiavertismentarputeaducelarănireadepanatorului,operatoruluisau
pacientuluiînurmapericolelordeelectrocutare,mecanicesaudealtănatură.
ПРЕДУПРЕЖДЕНИЕ
(RU)
Настоящееруководствопообслуживаниюпредлагаетсятольконаанглийскомязыке.
•Еслисервисномуперсоналуклиентанеобходиморуководствоненаанглийском,ана
каком-тодругомязыке,клиентуследуетобеспечитьпереводсамостоятельно.
•Преждечемприступатькобслуживаниюоборудования,обязательнообратитеськ
настоящемуруководствуивнимательноизучитеизложенныевнемсведения.
•Несоблюдениетребованийданногопредупрежденияможетпривестиктому,что
специалистыпообслуживанию,операторыилипациентыполучатударэлектрическим
током,механическуютравмуилидругоеповреждение.
2060321-001L MAC™800
7
ServiceManualLanguageInformation(cont'd.)
UPOZORENJE
(SR)
VAROVANIE
(SK)
OPOZORILO
(SL)
ADVERTENCIA
(ES)
VARNING
(SV)
UYARI
(TR)
Ovoservisnouputstvojedostupnosamonaengleskomjeziku.
•Akoklijentovserviserzahtevanekidrugijezik,klijentjedužandaobezbediprevodilačke
usluge.
•Nepokušavajtedaopraviteuređajakonistepročitaliirazumeliovoservisnouputstvo.
•Zanemarivanjeovogupozorenjamožedovestidopovređivanjaservisera,rukovaocaili
pacijentausledstrujnogudara,ilimehaničkihidrugihopasnosti.
Tentonávodnaobsluhujekdispozíciilenvangličtine.
•Akzákazníkovposkytovateľslužiebvyžadujeinýjazykakoangličtinu,poskytnutie
prekladateľskýchslužiebjezodpovednosťouzákazníka.
•Nepokúšajtesaoobsluhuzariadeniaskôr,akosineprečítatenávodnaobsluhua
neporozumietemu.
•Zanedbanietohtovarovaniamôževyústiťdozraneniaposkytovateľaslužieb,obsluhujúcej
osobyalebopacientaelektrickýmprúdom,mechanickýmaleboinýmnebezpečenstvom.
Taservisnipriročnikjenavoljosamovangleškemjeziku.
•Čeponudnikstoritvestrankepotrebujepriročnikvdrugemjeziku,morastrankazagotoviti
prevod.
•Neposkušajteservisiratiopreme,četegapriročnikanistevcelotiprebraliinrazumeli.
•Četegaopozorilaneupoštevate,selahkozaradielektričnegaudara,mehanskihalidrugih
nevarnostipoškodujeponudnikstoritev,operateralibolnik.
Estemanualdeserviciosóloexisteeninglés.
•Sielencargadodemantenimientodeunclientenecesitaunidiomaquenoseaelinglés,el
clientedeberáencargarsedelatraduccióndelmanual.
•Nosedeberádarserviciotécnicoalequipo,sinhaberconsultadoycomprendidoeste
manualdeservicio.
•Lanoobservanciadelpresenteavisopuededarlugaraqueelproveedordeservicios,el
operadoroelpacientesufranlesionesprovocadasporcausaseléctricas,mecánicaso
deotranaturaleza.
Denhärservicehandbokennnsbaratillgängligpåengelska.
•Omenkundsserviceteknikerharbehovavettannatspråkänengelskaansvararkunden
föratttillhandahållaöversättningstjänster .
•Försökinteutföraservicepåutrustningenomduinteharlästochförstårdenhär
servicehandboken.
•Omduintetarhänsyntilldenhärvarningenkandetresulteraiskadorpåserviceteknikern,
operatörenellerpatiententillföljdavelektriskastötar ,mekaniskafarorellerandrafaror .
Buserviskılavuzununsadeceİngilizcesimevcuttur.
•Eğermüşteriteknisyenibukılavuzuİngilizcedışındabirbaşkalisandantalepederse,bunu
tercümeettirmekmüşteriyedüşer .
•Serviskılavuzunuokuyupanlamadanekipmanlaramüdahaleetmeyiniz.
•Buuyarıyauyulmaması,elektrik,mekanikveyadiğertehlikelerdendolayıteknisyen,
operatörveyahastanınyaralanmasınayolaçabilir .
8 MAC™800 2060321-001L
6November2017
ServiceManualLanguageInformation(cont'd.)
ЗАСТЕРЕЖЕННЯ
(UK)
CẢ NHBÁO
(VI)
Данекерівництвозсервісногообслуговуванняпостачаєтьсявиключноанглійськоюмовою.
•Якщосервіснийінженерпотребуєкерівництвоіншоюмовою,користувачзобов'язаний
забезпечитипослугиперекладача.
•Ненамагайтесяздійснюватитехнічнеобслуговуванняданогообладнання,якщо
винечитали,абонезрозумілиінформацію,наданувкерівництвізсервісного
обслуговування.
•Недотриманняцьогозастереженняможепризвестидотравмуваннясервісного
інженера,користувачаданогообладнанняабопацієнтавнаслідокелектричногошоку,
механічногоушкодженняабозіншихпричинневірногообслуговуванняобладнання.
TàiLiệ uHướ ngDẫ nSử aChữ achỉcóbả ntiế ngAnh.
•Nế ucácđơ nvịcungcấ pdị chvụchokháchhàngyêucầ umộ tngônngữnàokháctiế ng
Anh,thìkháchhàngsẽcótráchnhiệ mcungcấ pcácdị chvụdị chthuậ t.
•Khôngđượ csử achữ athiế tbịtrừkhiđãthamkhả ovàhiể uTàiliệ uHướ ngdẫ nSử achữ a.
•Khôngtuânthủnhữ ngcả nhbáonàycóthểdẫ nđế ncáctổ nthươ ngchongườ ithự c
hiệ nsử achữ a,ngườ ivậ nhànhhaybệ nhnhân,dosố cđiệ n,cácrủ irovềcơkhíhay
cácrủ irokhác.
2060321-001L MAC™800 9
10 MAC™800 2060321-001L
6November2017
Contents
1 Introduction
IndicationsforUse....................................................................................15
Contraindications.....................................................................................15
RegulatoryandSafetyInformation...........................................................16
SafetyConventions..................................................................................16
SafetyHazards........................................................................................17
PartsandAccessoriesInformation................................................................17
ResponsibilityoftheManufacturer................................................................18
ResponsibilityofthePurchaser/Customer.......................................................18
EquipmentSymbols..................................................................................18
Training.....................................................................................................18
EquipmentIdentication...........................................................................18
ProductLabel.........................................................................................19
SerialNumberFormat...............................................................................19
ProductCodes........................................................................................20
ServiceInformation...................................................................................20
ServiceRequirements...............................................................................20
AdditionalAssistance................................................................................20
ManualInformation..................................................................................21
IntendedAudience...................................................................................21
ManualPurpose......................................................................................21
DocumentConventions.............................................................................21
RelatedDocuments...................................................................................22
2 EquipmentOverview
EquipmentDescription.............................................................................23
FrontView.............................................................................................23
SideView...............................................................................................24
RearView..............................................................................................24
BottomView...........................................................................................25
KeypadLayout........................................................................................26
SystemArchitecture..................................................................................27
HardwareBlockDiagram...........................................................................27
Hardware/FirmwareArchitecture.................................................................28
ProductInterfaces....................................................................................28
SoftwareArchitecture...............................................................................29
3 InstallingtheUSBPoweredSilexWirelessBridge
2060321-001L MAC™800 11
InstallingtheSilexWirelessBridge...........................................................32
4 Troubleshooting
GeneralFaultIsolation..............................................................................35
Power-UpSelf-Test...................................................................................35
PoorQualityECGs....................................................................................36
VisualInspection......................................................................................36
EventLogging...........................................................................................37
SettingUpEventLogging...........................................................................37
ExportingtheEventLog.............................................................................39
PerformingDiagnosticTests.....................................................................39
AccessingtheSystemDiagnosticsFunction....................................................39
DisplayTest............................................................................................41
SpeakerTest...........................................................................................42
KeyboardTest.........................................................................................43
AcquisitionModuleTest.............................................................................44
BatteryTest............................................................................................44
WriterTest.............................................................................................45
RS232Test.............................................................................................47
LANTest................................................................................................47
ModemTest...........................................................................................49
USBTest................................................................................................50
PatientLeadWireTest...............................................................................51
EquipmentProblems.................................................................................52
ECGDataNoise.......................................................................................52
SystemDate/TimeTroubleshooting...............................................................52
5 Maintenance
ErrorCodes...............................................................................................52
AcquisitionErrorCodes..............................................................................53
PrinterErrorCodes...................................................................................53
FrequentlyAskedQuestions.....................................................................53
Maintenance..........................................................................................53
SystemSetup..........................................................................................56
SettingtheWIFIParameter.......................................................................60
RequiredToolsandSupplies.....................................................................63
High-LevelFRUIdentication....................................................................64
FRUReplacementProcedures...................................................................66
PreparingSystemforFRUReplacement.........................................................66
ReplacingthePatientCable........................................................................66
ReplacingtheBatteryAssembly...................................................................67
ReplacingtheReal-timeClock(RTC)Battery.....................................................68
ReplacingtheTopCoverAssembly................................................................69
ReplacingtheKeypadAssembly...................................................................72
ReplacingtheLCDAssembly.......................................................................72
ReplacingthePrinterAssembly....................................................................73
ReplacingtheMainboardAssembly..............................................................76
ReplacingtheInternalModem(option)...........................................................80
ReplacingthePowerSupplyAssembly...........................................................81
ReplacingtheBottomCoverAssembly...........................................................85
12 MAC™800 2060321-001L
FunctionalCheckout.................................................................................87
ElectricalSafetyChecks............................................................................90
6 PartsList
OrderingParts...........................................................................................95
FieldReplaceableUnits(FRUs)..................................................................95
ReplacingtheFuse...................................................................................86
VisualInspection......................................................................................88
FunctionalCheckoutProcedures..................................................................89
UpdatingSoftware...................................................................................91
ConditioningtheBatteryPack.....................................................................93
Real-timeClockCheck...............................................................................94
UpperLevelAssemblyDiagrams..................................................................95
UpperLevelAssemblyPartList....................................................................98
FRUTopCoverAssembly,PN2039939-001...................................................103
FRUPrinterAssembly,PN2039941-001.......................................................104
FRUMainboardAssembly,PN2061575–001..................................................104
FRUPowerSupplyAssembly,PN2040052-001...............................................104
FRUBottomCoverAssembly,PN2039943-001..............................................105
ModelDataMatrixBarcodeScannerKits......................................................106
Keypads..............................................................................................107
FRUCLOCKBATTERY,PN2040761-001.........................................................107
USBWirelessDongle...............................................................................107
USBWirelessDongleUpgradeKit...............................................................108
SilexWirelessBridge...............................................................................108
SilexWirelessBridgeUpgradeKit...............................................................108
FRUKits,PN2039945-001........................................................................108
7 SystemConguration
SetupFunctions......................................................................................111
BasicSetup..........................................................................................111
RestingECGSetup..................................................................................115
ArrhythmiaSetup...................................................................................122
CommunicationSetup.............................................................................123
CountrySetup.........................................................................................131
PatientSetup...........................................................................................132
UserSetup...............................................................................................136
OptionsSetup..........................................................................................138
ServiceSetup...........................................................................................139
Date/TimeSetup.....................................................................................139
OrderManagerSetup..............................................................................140
SetupUtilities..........................................................................................141
PrintSetupReport..................................................................................141
SelectSetup.........................................................................................141
2060321-001L MAC™800 13
ExportSetup.........................................................................................142
ImportSetup........................................................................................143
ExportingAuditTrail...............................................................................143
A ProductSpecications
Specications..........................................................................................145
SilexWirelessBridge...............................................................................150
B WiFiCountryList
C ElectromagneticCompatibility(EMC)
GuidanceandManufacturer’sDeclaration—Electromagnetic
Emissions................................................................................................156
GuidanceandManufacturer’sDeclaration—Electromagnetic
Immunity.................................................................................................157
GuidanceandManufacturer’sDeclaration—Electromagnetic
Immunity.................................................................................................158
RecommendedSeparationDistances.....................................................160
14 MAC™800 2060321-001L
Introduction
ThisdocumentdescribestheMAC™800RestingECGAnalysisSystem,alsoreferredto
as“thesystem”or“thedevice”.
Thischapterprovidesgeneralinformationrequiredfortheproperuseofthesystem
andthemanual.Familiarizeyourselfwiththisinformationbeforeusingthesystem.
IndicationsforUse
TheMAC™800RestingECGAnalysisSystemisaportableECGacquisition,analysis,
andrecordingsystemintendedto:
•acquire,analyze,display,andrecordinformationfromadultandpediatric
populations,
•beusedunderthedirectsupervisionofalicensedhealthcarepractitioner,
1
•beusedbytrainedoperatorsinahospitalormedicalprofessional'sfacility
environment,aswellasusedinclinics,physicianofces,outreachcenters,
•offertwobasicmodesofoperation:(1)restingECGmodeand(2)arrhythmiamode
•print3-and6leadsofECG,
•beupgradeabletoprovidesoftwareoptionssuchas12-leadECGmeasurement
andinterpretiveanalysis,
•providefortheoptionaltransmissionandreceptionofECGdatatoandfroma
centralECGcardiovascularinformationsystem.
NOTE:
Pediatricpopulationisdenedaspatientsbetweentheagesof0and15years.
Arrhythmiadetectionisprovidedfortheconvenienceofautomaticdocumentation.
Contraindications
ThisMAC™800deviceisNOTintended:
•tobeusedduringpatienttransport,
•tobeusedforintracardiacapplications,
•tobeusedasavitalsignsphysiologicalmonitor,
•toprovidealarmsforarrhythmiadetection.
2060321-001L MAC™800 15
Introduction
RegulatoryandSafetyInformation
Thissectionprovidesinformationaboutthesafeuseandregulatorycomplianceof
thisdevice.Familiarizeyourselfwiththisinformationandreadandunderstandall
instructionsbeforeattemptingtousethisdevice.Thesystemsoftwareisconsidered
medicalsoftware.Assuch,itwasdesignedandmanufacturedtotheappropriate
medicalregulationsandcontrols.AnyexceptionsarenotedintheCompliance
Information-Exceptionssection.
NOTE:
Disregardingthesafetyinformationprovidedisconsideredabnormaluseof
thisdeviceandcouldresultininjury,lossofdata,andvoidanyexistingproduct
warranties.
SafetyConventions
AHazardisasourceofpotentialinjurytoaperson,property,ortheproduct.
ThismanualusesthetermsDANGER,WARNING,andCAUTIONtopointouthazards
andtodesignateadegreeorlevelofseriousness.Familiarizeyourselfwiththe
followingdenitionsandtheirsignicance.
DenitionsofSafetyConventions
Safety
Convention
DANGER
WARNING
CAUTION
Denition
Indicatesanimminenthazard,which,ifnotavoided,willresultindeath
orseriousinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided,
couldresultindeathorseriousinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided,
couldresultinminorpersonalinjuryorproduct/propertydamage.
16 MAC™800 2060321-001L
SafetyHazards
Thefollowingmessagesapplytothesystemasawhole.Specicmessagesmayalso
beprovidedelsewhereinthemanual.
DANGER:
WARNING:
WARNING:
WARNING:
Introduction
Donotuseinthepresenceofammableanesthetics.
CONNECTIONTOMAINS.ThisisclassIequipment.
Themainsplugmustbeconnectedtoanappropriatelygroundedpowersupply.
BATTERYOPERATION.Iftheintegrityoftheprotectiveearthconductorisindoubt,
operatetheunitfromitsbattery.
Failureonthepartofallresponsibleindividuals,orinstitutions,employingtheuse
ofthisdevicetoimplementtherecommendedmaintenanceschedulemaycause
equipmentfailureandpossiblehealthhazards.
CAUTION:
PartsandAccessoriesInformation
WARNING:
Thisequipmentcontainsnouserserviceableparts.Referservicingtoqualied
servicepersonnel.
U.S.Federallawrestrictsthisdevicetothesalebyorontheorderofaphysician.
PATIENTSAFETY—Toensurepatientsafety,useonlypartsandaccessories
manufacturedorrecommendedbyGEHealthcare.
ContactGEHealthcareforinformationbeforeconnectinganydevicestothis
equipmentthatarenotrecommendedinthismanual.
IftheinstallationofthisequipmentintheU.S.A.uses240Vratherthan120V,the
sourcemustbeacenter-tapped,240V,single-phasecircuit.
Partsandaccessoriesmustmeettherequirementsoftheapplicable60601safety
standards,and/orthesystemcongurationmustmeettherequirementsofthe
60601-1-1MedicalElectricalSystemsstandard.
Usingaccessoryequipmentthatdoesnotcomplywiththeequivalentsafety
requirementsofthisequipmentmayleadtoareducedlevelofsafetyofthe
resultingsystem.Considerationrelatingtothechoiceshallinclude:
•UseoftheaccessoryinthePatientVicinity.
Patientvicinityisdenedasaspace,withinalocationintendedforthe
examinationandtreatmentofpatients,extending6ft.(1.83m)beyondthe
normallocationofthebed,chair,table,treadmill,orotherdevice(s)supporting
thepatientduringexaminationandtreatment,andextendingverticallyto8ft.
2.4in.(2.5m)abovetheoor.
•Evidencethatthesafetycerticationoftheaccessorywasperformedin
accordancewiththeappropriate60601-1and/or60601-1-1standard(s).
2060321-001L MAC™800 17
Introduction
ResponsibilityoftheManufacturer
GEHealthcareisresponsiblefortheeffectsofsafety,reliability,andperformanceon
GE-suppliedhardwareonlyifthefollowingconditionsaremet:
•Assemblyoperations,extensions,readjustments,modications,orrepairsare
carriedoutbypersonsauthorizedbyGEHealthcare.
•Theelectricalinstallationoftherelevantroomcomplieswiththerequirementsof
theappropriatelocal,state,andothergovernmentregulations.
•Theequipmentisusedinaccordancewiththeinstructionsforuse.
ResponsibilityofthePurchaser/Customer
Thecustomerisresponsibleforprovidingappropriatedesks,chairs,electricalwall
outlets,networkconnections,analogphonelines,andforlocatinganyofthesystem
componentsdescribedinthismanualincompliancewithalllocal,state,andnational
codes.
EquipmentSymbols
SeetheMAC™800RestingECGAnalysisSystemOperator’sManualforinformation
aboutthesymbolsusedonthisproductanditspackaging.
Training
Thismanualisintendedasasupplementto,notasubstitutefor,thoroughproduct
training.Ifyouhavenotreceivedtrainingontheuseofthedevice,youshouldrequest
trainingassistancefromGEHealthcare.
Toseeavailabletraining,gototheGEHealthcaretrainingwebsite
(http://www.gehealthcare.com/usen/education/index.html )andselectDiagnostic
Cardiology undertheTechnicalServiceEducationsection.
Formoreself-pacedcourseofferings,tools,andreferenceguidesyou
maynduseful,pleasevisittheGEHealthcareEducationStoreat
www.gehealthcare.com/educationstore
.
EquipmentIdentication
EveryGEHealthcaredevicehasaproductlabelthatidentiestheproductname,part
number,manufacturinginformation,anduniqueserialnumber.Thisinformationis
requiredwhencontactingGEHealthcareforsupport.
18 MAC™800 2060321-001L
ProductLabel
Introduction
Theproductlabelislaidoutinthefollowingformat.Dependingontheproduct,the
labelmayvaryslightlyinformat,butitcontainsthesameinformation.
ProductLabelFormat
SerialNumberFormat
Item
1
2
3
4
5
Description
Productpartnumber
Productdescription
DateofmanufactureinYYYY-MMformat
Unitserialnumber(See“SerialNumberFormat”onpage19formore
information.)
Productbarcode
Eachdevicehasaserialnumberthatuniquelyidentiesthedeviceandprovides
importantinformationaboutthedevice.Theserialnumberformatisshowninthe
followingillustration:
SerialNumberFormat
Item Name
1
2
2060321-001L MAC™800 19
ProductCode Three-lettercodethatuniquelyidentiesthe
YearManufactured Two-digitcodeidentifyingtheyearthedevice
Description
productline.Referto“ProductCodes”onpage
20formoreinformation.
wasmanufactured.Valuesrangefrom00to99
Forexample:00=2000,04=2004,05=2005
(andsoon).
Introduction
SerialNumberFormat(cont'd.)
ProductCodes
Item Name
3
4
5
6
Theproductcodeidentiesspecicsystemplatforms.Youneedtheproductcode
beforeservicingorrequestingsupportforyourdevice.
Youcanidentifytheproductcodeusingtheserialnumberlistedontheproductlabel
locatedonthebaseofthesystem.
FiscalWeek
Manufactured
ProductSequence Four-digitnumberidentifyingtheorderinwhich
ManufacturingSite One-lettercodeidentifyingthesitewherethe
Miscellaneous
Characteristic
Description
Two-digitcodeidentifyingtheweekthedevice
wasmanufactured.
Valuesrangefrom01to52.GEHealthcare'sscal
weekscorrespondtothecalendarweek.For
example,01=rstweekinJanuary.
thisdevicewasmanufactured.Valuesrangefrom
000to9999.
devicewasmanufactured.Forexample,F=
Milwaukee,N=Freiburg,P=Bangalore
Forexample,P=unitisaprototype,R=unitwas
refurbished,U=unitwasupgradedtomeetthe
specicationsofanotherproductcode.
ServiceInformation
ServiceRequirements
AdditionalAssistance
Thissectionprovidesinformationpertainingtothemaintenanceandservicingofthe
device.FamiliarizeyourselfwiththisinformationbeforerequestingservicefromGE
Healthcareoritsauthorizedrepresentatives.
Failureonthepartoftheresponsibleindividual,hospital,orinstitutionusingthis
equipmenttoimplementasatisfactorymaintenanceschedulemaycauseundue
equipmentfailureandpossiblesafetyhazards.
Regularmaintenance,irrespectiveofusage,isessentialtoensurethatthecomponents
ofthissystemarealwaysfunctionalwhenrequired.
GEHealthcaremaintainsatrainedstaffofapplicationandtechnicalexpertstoanswer
questionsandrespondtoissuesandproblemsthatmayariseduringtheinstallation,
maintenance,anduseofthisproduct.
ContactyourlocalGEHealthcarerepresentativetorequestadditionalassistance.
20 MAC™800 2060321-001L
Introduction
ManualInformation
Thissectionprovidesinformationforthecorrectuseofthismanual.
Keepthismanualwiththeequipmentatalltimesandperiodicallyreviewit.You
shouldrequesttrainingassistancefromGEHealthcare,ifneeded.
IntendedAudience
Thismanualisintendedforthepersonwhouses,maintains,ortroubleshootsthis
equipment.
ManualPurpose
Thismanualsuppliestechnicalinformationforservicerepresentativesandtechnical
personnelsotheycanmaintaintheequipmenttotheassemblylevel.Useitasa
guideformaintenanceandelectricalrepairsconsideredeldrepairable.Where
necessary,themanualidentiesadditionalsourcesofrelevantinformationand/or
technicalassistance.
DocumentConventions
TypographicalConventions
SeetheMAC™800RestingECGAnalysisSystemOperator’sManualfortheinstructions
necessarytooperatetheequipmentsafelyinaccordancewithitsfunctionand
intendeduse.
Thismanualusesthefollowingconventions.
Convention Description
BoldText Indicateskeysonthekeyboard,texttoenter ,orhardwareitemssuchas
buttonsorswitchesontheequipment.
Italicized-Bold
Text
CTRL+ESC
<space>
Indicatessoftwaretermsthatidentifymenuitems,buttonsoroptionsin
variouswindows.
Indicatesakeyboardoperation.Aplus(+)signbetweenthenamesoftwo
keysindicatesthatwhileholdingtherstkey,youshouldpressandrelease
thesecondkey.Forexample,PressCTRL+ESCmeanstopressandholdthe
CTRL keyandthenpressandreleasetheESCkey.
Indicatesthatyoumustpressthespacebar.Wheninstructionsaregiven
fortypingaprecisetextstringwithoneormorespaces,thepointwhere
youmustpressthespacebarisindicatedas:<space>.Thisensuresthat
thecorrectnumberofspacesareinsertedinthecorrectpositionswithin
theliteraltextstring.Thepurposeofthe<>bracketsistodistinguishthe
commandfromtheliteraltextwithinthestring.
2060321-001L MAC™800 21
Introduction
Convention Description
Enter
>
IndicatesthatyoumustpresstheEnterorReturnkeyonthekeyboard.Do
nottypeEnter.
Thegreaterthansymbol,orrightanglebracket,isaconcisemethodto
indicateasequenceofmenuselections.
Forexample,thestatement“Fromthemainmenu,selectSystem>Setup>
Options toopentheOptionActivationwindow”replacesthefollowing:
1. Fromthemainmenu,selectSystemtoopentheSystemmenu.
2. FromtheSystemmenu,selectSetuptoopentheSetupmenu.
3. FromtheSetupmenu,selectOptionstoopentheOptionActivation
window.
Illustrations
Allillustrationsinthemanualareprovidedasexamplesonly.Dependingonsystem
conguration,screensthatappearinthemanualmaydifferfromthescreensasthey
appearonyoursystem.
Notes
RelatedDocuments
Allpatientnamesanddataarectitious.Anysimilaritytoactualpersonsis
coincidental.
Notesprovideapplicationtipsoradditionalinformationthat,whileuseful,arenot
essentialtothecorrectoperationoftheproduct.Theyarecalledoutfromthebody
textthroughaagwordandindentation,asfollows:
NOTE:
ThetiporadditionalinformationappearsindentedbelowtheNOTEagword.
Youcanndadditionalinformationinthefollowingdocuments:
DocumentsRelatedtotheMAC800RestingECGAnalysisSystemServiceManual
PartNumber DocumentTitle
2060026-001
2102946-001
MAC™800RestingECGAnalysisSystemOperator’sManual
SuppliesandAccessoriesGuide,DiagnosticCardiology
22 MAC™800 2060321-001L
EquipmentOverview
Thedeviceisa3-and6-leadprint,12-channeldisplaysystemwitha7inch(17.78
cm)diagonaldisplay,activepatientcable,batteryoperation,andoptionsfor
communicationcapabilities.
EquipmentDescription
FrontView
2
Item Name
1 Screen
2 ACLED
2060321-001L MAC™800 23
Description
Displayswaveformandtextdata.
IndicateswhentheunitisconnectedtoACpower.
EquipmentOverview
Item Name
3 BatteryLED
4 PowerLED
5
6
7
Keypad Controlsthesystemorentersdata.See“Keypad
WriterDoorButton
Writer Printsreports.
Description
Indicatescurrentbatterystatus.
•Solidindicatesthebatteryischarging.
•Flashingindicatesthebatteryislow.
•Offindicatesthebatteryisfully-chargedoris
dischargingbutnotatalowstate.
Indicateswhentheunitispoweredon.
Layout”onpage26formoreinformation.
Opensprinterdoor.
SideView
Item Name
1
ECGsignalinputconnector D-sub15–pinfemaleconnectorforthe
acquisitioncable.
2
SDcardslot SecureDigitalcardslot.Insertcard
asindicatedbytheicon.Thesystem
supportsonlySDcardsformattedforthe
FAT16orFAT32lesystems.
RearView
24 MAC™800 2060321-001L
Description
EquipmentOverview
Item Name
1
ModemPort RJ11connectorfromtheoptionalinternal
2
LANconnection
3
USBconnector
4
COMMPort
5
ACpowerconnector
6
Equipotentialgroundinglug Usedtoconnectnon-groundedperipheral
Description
modemtoananalogphoneline.
RJ45networkconnector .
TheLEDsindicateLANstatus.
•ThesteadygreenLEDtotherightof
thisportindicatesagoodethernet
connection.
•TheashingamberLEDtotheleftof
thisportindicatesnetworktrafc.
UniversalSerialBusconnectorforUSB
devices,suchasanoptionalbarcode
reader,amagneticcardreader,an
externalUSBkeyboard,alaserprinter,or
aUSBWiFidevice.
Serialconnectorfordatacommunication
withCASE/CardioSoftortheMUSEsystem
withaserialcable.
StandardconnectorfortheACpower
cable.
devicestoensureequipotential.
BottomView
Item Name
Battery
1
2
Carryinghandle Handleforcarryingthedevice.
Description
Rechargeablelithium-ionbattery.
2060321-001L MAC™800 25
EquipmentOverview
KeypadLayout
Item Name
1
2
3 PowerButton
4
5
6
7
8
9SpaceKey
10T9Key
FunctionKeys(F1throughF6) Selectsmenuoptionsonthescreen.
Leadskey Changestheleadswhenthescreenis
ECGkey AcquiresarestingECGandprintsa
Trimpad Thearrowsmovethecursorleft,right,
Rhythmkey Printsacontinuous,real-timerhythm
Stopkey Stopsthewriterfromprinting.
BackspaceKey Deletescharacters.
Description
displayingwaveforms.
Turnsthesystemonandoff.
10-secondreportinArrhythmiamode.
up,ordowntohighlightamenuor
screenitem.Pressingthecenterbutton
selectsthehighlighteditem.
strip.PressStoptostoptherhythmstrip
fromprinting.(TheRhythmreportisnot
storedandcannotbetransmitted.)
Addsaspacebetweentypedcharacters.
Switchesbetweendifferentinput
methods.Formoreinformationon
usingtheT9key,refertotheMAC™800
RestingECGAnalysisSystemOperator's
Manual.
26 MAC™800 2060321-001L
SystemArchitecture
HardwareBlockDiagram
EquipmentOverview
2060321-001L MAC™800 27
EquipmentOverview
Hardware/FirmwareArchitecture
Thehardwareandrmwaresubsystemsincludethefollowing:
HardwareSubsystems
•CPUcore
•Display
•Keyboard
•ECGAcquisitionsubsystem
•Thermalprinter
•Powersupply
•Housing
FirmwareSubsystems
•CEOS(Board-Support-Package)
•FPGA(FirmwareforthePrinter)
ProductInterfaces
Thesystemoffersthefollowinginterfacesforconnectingtoexternaldevicesfordata
communication,softwareupdates,andthecontrolofworkloaddevices:
InterfaceDevice
RS232port(1)
RJ-45port(1)
USBconnector(2)
SecureDigital(SD)Cardslot InterfaceswithaSecureDigitalcard,which
RJ-11port(1)
Connectstoexternalsystems,suchasthe
MUSEsystemortheCardioSoftsystem.
Connectstonetworksvia100baseTethernet
connectorviaanexternalmedicalisolator .
ConnectstoUSB-capabledevices,suchas
anoptionalbarcodereaderoranexternal
USBkeyboard.
storesECGs,toashthedevicewithsoftware
updates,andtoconnectmemory/futureI/O
extensions.
ConnectsaninternalmedicalgradeAnalog
Modem(optional)toaphoneline.
Description
28 MAC™800 2060321-001L
SoftwareArchitecture
LayeredStructureofApplicationSoftware
EquipmentOverview
2060321-001L MAC™800 29
EquipmentOverview
ECGDataFlowwithSamplingRates
30 MAC™800 2060321-001L
InstallingtheUSBPoweredSilex
WirelessBridge
Thischapteronlydescribesthehardwareinstallation.Fordetailedguidanceon
congurationandsetup,refertoMobileLinkWirelessCommunicationInstallation
ManualForGEH-BR-4600WirelessBridge(2053535-080) .
ThenewSilexwirelessbridgekit(PN2098761-001forglobalversion,PN2095170-001
forUS)comeswithSilexwirelessmodule,ethernetcable,USBpowercableanddual
locks(2pcs).
3
2060321-001L MAC™800 31
InstallingtheUSBPoweredSilexWirelessBridge
InstallingtheSilexWirelessBridge
1. SelecttheDualLocksfromtheSilexWirelessBridgeKitandengagethehook
andloopsurfaceoftheDualLocksagainsteachothertoformapair.
2. Removetheadhesivelinerfromonesideoftheduallockandadhereeachpair
totheatsurfaceofSilexwirelessbridgemodule.
3. RemovetheadhesivelinerandcarefullypositiontheSilexModuleontheMAC
800unitbyplacingittotherearoftheprinterside.TheLANportandpower
switchshouldbevisiblefromtherearoftheMAC800device.
32 MAC™800 2060321-001L
InstallingtheUSBPoweredSilexWirelessBridge
4. RouteandconnecttheEthernetandUSBpowercablesfromtheSilexwireless
moduletotheMAC800unit.
5. ContinuewiththeappropriatefunctionalcheckoutprocedureforthisFRU.
See“FunctionalCheckout”onpage87formoreinformation.
6. Verifytheinstallationbyperformingthefollowingsteps:
a. TurnontheMAC800device.
b. PressFileManager.
c. SelecttheECGyouwouldliketotransmit.
NOTE:
IfnoECGsareavailableonthedevice,downloadonefromanSDcard
ortakeaatlineECGwithnopatientIDnumberonitanstoreiton
theSDCard.
d. VerifythattheECGtransmitssuccessfullyandcanbefoundintheMUSE
Edit list.
FormoreinformationonhowtotransmitanECGfromthecart,referto
theMAC800ECGAnalysisSystemOperator’sManual(2060026–001).
2060321-001L MAC™800
33
InstallingtheUSBPoweredSilexWirelessBridge
34
MAC™800 2060321-001L
Troubleshooting
GeneralFaultIsolation
RefertotheMAC™800Operator’sManual,Chapter2,“EquipmentOverview:Getting
Started”toverifyoperationofthedevice.
Power-UpSelf-Test
Onpower-up,thesystemautomaticallyrunsaninternalself-test.Ifalltestspass,
youseethefollowingstart-upscreen.
4
Thenextscreentoopendependsonthe PowerUpmodeselectedinSystem
Conguration.TheRestingECGmodeisthedefaultPowerUpmode.
Iftheequipmentisnotworkingproperly,considerthefollowing:
•Isthedeviceturnedon?
•Havetherebeenanychangesintheuse,location,orenvironmentoftheequipment
thatcouldcausethefailure?
•Hastheequipmenthardwareorsoftwarebeenmodiedsincethelastuse?
•Isoperatorerrorthecauseoftheproblem?
Trytorepeatthescenarioexactlyandcomparethattotheproperoperationofthe
equipmentdescribedinthemanual.
•Isthebatteryinstalled?
•WhenconnectedtotheACwalloutlet,doesthegreenACpowerlightglow?
2060321-001L MAC™800
35
Troubleshooting
PoorQualityECGs
SeveralfactorscancausepoorECGsincluding:
•Factorsintheenvironment.
•Inadequatepatientpreparation.
•Hardwarefailuresrelatedtotheacquisitionmodule.
•Leadwires,cables,orproblemsinthedevice.
VisualInspection
Athoroughvisualinspectionoftheequipmentcansavetime.Smallthings,suchas
disconnectedcables,foreigndebrisoncircuitboards,missinghardware,orloose
components,canfrequentlycausesymptomsandequipmentfailuresthatmayseem
tobeunrelatedanddifculttotrack.
NOTE:
Takethetimetomakealltherecommendedvisualchecksbeforestartingany
detailedtroubleshootingprocedures.
VisualInspectionChecklist
I/Oconnectorsand
cables
ACpowercord
Interfacecables
Circuitboards
Area
•Frayingorotherdamage
•Bentprongsorpins
•Crackedhousing
•Loosescrewsinplugs
•Excessivetensionorwear
•Looseconnection
•Strainreliefsoutofplace
•Moisture,dust,ordebris(topandbottom)
•Looseormissingcomponents
•Burndamageorsmellofover-heatedcomponents
•Socketedcomponentsnotrmlyseated
•PCBnotseatedproperlyinedgeconnectors
•Solderproblems:cracks,splashesonboard,incomplete
feedthrough,priormodicationsorrepairs
Lookforthefollowingproblems
Groundwires/wiring
Fasteners
36 MAC™800 2060321-001L
•Loosewiresorgroundstrapconnections
•Faultywiring
•Wirespinchedorinvulnerableposition
Looseormissingscrewsorotherhardware,especiallyfasteners
usedasconnectionstogroundplanesonPCBs
VisualInspectionChecklist(cont'd.)
Troubleshooting
Powersource
Area
•Faultywiring,especiallyACoutlet
•Circuitnotdedicatedtosystem
Lookforthefollowingproblems
NOTE:
Powersourceproblemscancausestaticdischarge,resetting
problems,andnoise.
Keyboard
LCDdisplaylter Scratches,cracks,oranopaquedisplaylter(transparentpartof
Batterypack
SDcard
•Cutsorcracksinkeyboardmembrane
•Illegiblelabels
thekeyboardbezel)thatimpairviewing
•Cracked,swollen,orleakybatterypackenclosure
•Debrisonbatterypackelectricalcontacts
•CrackedSDcard
•Brokengoldcontacts
•Dirt,scratches,ordebrisoncontacts
EventLogging
SettingUpEventLogging
YoucansetupthesystemtocreateanEventLoginXMLformat.Tocongurethe
systemforthelevelofseverityofmessageswrittentotheEventLog,usethefollowing
steps:
1. PoweronthesystembypressingthePowerbutton.
2. OntheMainMenu,pressF4toselectSystemConguration.
3. PressMore>More>ServiceSetup.
AwindowopenspromptingyoutoentertheServicepassword.
ContactGEHealthcaresupportifyoudonotknowtheservicepassword.
2060321-001L MAC™800
37
Troubleshooting
4. TypetheservicepasswordandpressF6toselectOKtoopentheServiceSetup
menu.
5. UsethetrimpadtohighlightEventLogandpressthecenterbutton.
6. Enableordisableeventlogging.
•Toenableeventlogging,selecttheKeyEventLoggingcheckbox.
or
•Todisableeventlogging,deselecttheKeyEventLoggingcheckbox.
7. SelectalevelofseveritytologfromtheLogLevellist:
•SelectNonetolognothingtotheEventLog.
•SelectErrortologonlyerrorstotheEventLog.
•SelectWarningtologerrorsandwarningstotheEventLog.
•SelectInformationtologerrors,warnings,andinformationtotheEventLog.
8. PressSavetosaveyoursettings.
38 MAC™800 2060321-001L
ExportingtheEventLog
1. Repeatsteps1through5in“SettingUpEventLogging”.
2. InserttheSDcardintotheSDcardslotinrightsideasshown.
Thegoldcontactsareface-down.
3. PressExportLogFiles.
Troubleshooting
ThecurrentEventLogle,log_0.log,iscopiedtoalogdirectory
ontheSDcard.
NOTE:
Toaccessthelogle,inserttheSDcardintoanSDcardreaderthatis
connectedtoacomputerwithaWindowsoperatingsystemandatext
editorsuchasNotepadorWordPad.IfGEHealthcaretechnicalservice
requeststheEventLogfortroubleshootinganissue,sendtheleasan
emailattachment.
PerformingDiagnosticTests
AccessingtheSystemDiagnosticsFunction
UsetheSystemDiagnosticsmenutoperformfunctionaldiagnostictests.Usethe
followingproceduretoaccesstheSystemDiagnosticsmenu.
1. PoweronthesystembypressingthePowerbutton.
2. OntheMainMenu,pressF4toselectSystemConguration.
3. PressMore>More>ServiceSetup.
AwindowopenspromptingyoutoentertheServicepassword.
ContactGEHealthcaresupportifyoudonotknowtheservicepassword.
2060321-001L MAC™800 39
Troubleshooting
4. TypetheservicepasswordandpressF6toselectOKtoopentheServiceSetup
menu.
5. UsethetrimpadtohighlightSystemDiagnosticsandpressthecenterbuttonto
opentheDiagnosticTestswindow.
Thefollowingsectionsdescribehowtoperformthespecicdiagnostictests.
Proceedtotheappropriatesectionforthetestyouneedtoperform.
40 MAC™800 2060321-001L
DisplayTest
Troubleshooting
UsetheDisplayTesttodetermineifthedisplaypixelsareworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectDisplayTest.
TheStartTestwindowopens.
3. SelectStartTest.
Thefollowingwindowopens.
4. PresstherightarrowontheTrimpadrepeatedlytomovethecolorbars
horizontallyacrossthescreen.
5. Verifythatthecolorbandpattern(red,green,blue,white)scrollsacrossthe
screen.
Passthetestifthepatternisreplicatedwithoutdiscoloration.
6. PressF1toswitchtohorizontalcolorbars.
7. PressthedownarrowontheTrimpadrepeatedly.
8. Verifythatthecolorbandpattern(red,green,blue,white)scrollsdownthe
screen.
Passthetestifthepatternisreplicatedwithoutdiscoloration.
9. PressF1tocyclethroughthesolidcolorpane(red,green,blue,white).
Foreachpane,checkforblackpixels.
2060321-001L MAC™800 41
Troubleshooting
Passthetestifnomorethan4blackpixelsareobservedonanysinglecolor
pane.
NOTE:
Ablackpixelobservedononepanewillprobablybeobservedoneverypane.
10. Pressthecenterbuttonofthetrimpadwhenthetestiscomplete.
Thefollowingwindowopens.
11. Selectpassorfail:
•Ifthetestpassed,pressYes.
•Ifthetestfailed,pressNo.
Ifthedisplaytestfailed,replacethedisplayassemblyasdescribedin
“ReplacingtheLCDAssembly”onpage72.
SpeakerTest
UsetheSpeakerTesttodetermineifthespeakerisworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectSpeakerTest.
3. Listenforabriefaudibletonecomingfromthespeaker.
Thefollowingwindowopens.
4. Selectpassorfail:
•Ifyouheardanaudibletone,pressYes.
•Ifyoudidnothearanaudibletone,pressNo.
Ifthespeakertestfailed,replacethemainboardassemblyasdescribed
in“ReplacingtheMainboardAssembly”onpage76.
42 MAC™800 2060321-001L
KeyboardTest
Troubleshooting
UsetheKeyboardTesttodetermineifthekeyboardisworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectKeyboardTest.
Thefollowingwindowopens.
3. Presseachkeyonthekeyboardandverifythevalueappearsinthe
correspondingrepresentationofthatkeyonthescreen.
Akeypassesthetestifitsvalueappearsonthescreenwhenthecorresponding
keyispressed.
4. Totestforstickykeys,continuetopresskeysandverifythatthescreen
representationofthekeyisrefreshingwitheachsubsequentkeypress.
Akeypassesifthekeyonthescreenrefresheswitheachrepeatedkeypress.
5. Whenthetestiscomplete,pressF1>Stop.
Thefollowingwindowopens.
6. Selectpassorfail:
•Ifeverykeypassesthetests,pressYes.
•Ifanykeyfailsthetests,pressNo.
Ifthekeyboardtestfailed,replacethekeyboardassemblyasdescribedin
“ReplacingtheKeypadAssembly”onpage72.
2060321-001L MAC™800
43
Troubleshooting
AcquisitionModuleTest
UsetheAcquisitionModuleTesttodetermineiftheacquisitionboardisworking
properly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectAcquisitionModuleTest.
Awindowsimilartotheoneshowninthefollowingillustrationopens.
BatteryTest
3. NotethetestresultandpressCancel.
IftheresultoftheAcquisitionModuleTestResultisFailed,replacethemainboard
assemblyasdescribedin“ReplacingtheMainboardAssembly”onpage76.
UsetheBatteryTesttodeterminethestatusoftheLithium-Ionbattery.Youmust
performthistestwhilerunningonbatterypower.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectBatteryTest.
Awindowsimilartotheoneshowninthefollowingillustrationopens.
3. NotethebatterystatusinformationandpressCanceltoclosetheBATTERY
STATUS window.
IftheBatteryStatuswasFailed,replacethebatteryasdescribedin“Replacingthe
BatteryAssembly”onpage67.
44
MAC™800 2060321-001L
WriterTest
Troubleshooting
UsetheWriterTesttodetermineifthewriterisworkingproperly.
NOTE:
BeforeperformingtheWriterTest,besurethatthermalpaperisproperlyloadedin
thewritertray.
Refertoyoursystem’sOperatorManualforinstructionsonloadingpaper.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectWriterTest.
Thefollowingwindowopens.
3. Performthe50mm/sSpeedTest.
a. Select50mm/sSpeedTest.
Thewriterprintsthe50mm/sspeedtestreport.
b. Whenonepageofthereporthasprinted,pressStop.
Thefollowingwindowopens.
2060321-001L MAC™800
45
Troubleshooting
c. Examinetheprintedreport.
Usethefollowingcriteriatodetermineifthewriterpassedorfailedthe
50mm/sspeedtest.
•Ifonecycleofthesquarewavespans50mmonpaper,measuredfrom
cornertocornerofthewave,withanallowabletoleranceof1.0mm,
thetestpasses.
•Ifthiscriteriaisnotmet,thetestfails.
d. Ifthetestpassed,pressYes.
Ifthetestfailed,pressNo.
4. Repeatthepreviousstepfortheotherspeedtests.Thepass-failcriteriaforeach
oftheremainingtestsareasfollows:
•25mm/sSpeedTest
Ifonecycleofthesquarewavespans25mmonpaper,measuredfrom
cornertocornerofthewave,withanallowabletoleranceof0.5mm(2%),
thetestpasses.
Ifthiscriteriaisnotmet,thetestfails.
•5mm/sSpeedTest
Ifonecycleofthesquarewavespans5mmonpaper,measuredfromcorner
tocornerofwave,withallowabletoleranceof0.25mm,thetestpasses.If
thiscriteriaisnotmet,thetestfails.
5. PerformthePrintHeadTest.
a. SelectPrintHeadTest.
Thewriterprintsa1–pageprintheadtestreport.
b. Verifythattherearenogapsinanyofthelinesprinted.
•Iftherearenogapsinthelinesontheprintedreport,pressYes.
•Iftherearegapsinthelinesontheprintedreport,pressNo.
Replacetheprinterasdescribedin“ReplacingthePrinterAssembly”on
page73.
6. Whenallwritertestsarecompleted,pressCanceltoclosethewindow.
46 MAC™800 2060321-001L
RS232Test
Troubleshooting
UsetheRS232TesttodetermineiftheCOMportsareworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. Useapapercliptoshortpins2and3intheCOMport.
3. SelectRS232Test.
Thefollowingwindowopens:
4. PerformtheCOMPortLoopBackTestonCOMA.
a. SelectCOMA.
TheresultsoftheCOMPortLoopBackTestaredisplayed.
b. NotetheresultsoftheCOMPortLoopBackTest.
Ifthetestfailed,replacethemainboardassemblyasdescribedin
“ReplacingtheMainboardAssembly”onpage76.
5. Whenthetestisdone,pressEscorCanceltoclosetheresultswindow.
LANTest
UsetheLANTesttotestconnectivitybetweennetworksorIPaddresses.
NetworkConnectivity
1. ConnectthedevicetoanactiveLAN.
EnsurethattheLANisanactivenetwork.Ifyouconnecttoaninactivenetwork
tap,thetestresultmaybeafalsenegative.
2. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2060321-001L MAC™800
47
Troubleshooting
3. SelectLANTest.
Thefollowingwindowopens:
4. SelectTestnetworkconnectivity.
Thefollowingmessageisdisplayed:Checkingconnectivity.Pleasewait.
Thenthetestresultsaredisplayed.
•Ifthefollowingmessageisdisplayedinthewindow,thetestpasses:System
ConnectedtoNetwork .
•Ifthefollowingmessageisdisplayedinthewindow,andyouaresurethe
systemisconnectedtoanactivenetwork,thetestfails:NetworkUnavailable.
Replacethemainboardassemblyasdescribedin“ReplacingtheMainboard
Assembly”onpage76.
5. Whenthetestisdone,pressEscorCanceltoclosetheresultswindow.
IPAddressConnectivity
1. ConnectthedevicetoanactiveLAN.
EnsurethattheLANisanactivenetwork.Ifyouconnecttoaninactivenetwork
tap,thetestresultmaybeafalsenegative.
2. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
48 MAC™800 2060321-001L
3. SelectLANTest.
Thefollowingwindowopens:
Troubleshooting
ModemTest
4. UsethetrimpadtomovethecursortotheIPAddresseld.
5. TypetheIPAddressandpressthecenterbuttononthetrimpadtoenter.
6. SelectPingandpressthecenterbuttononthetrimpadtostarttest.
•IftheIPaddressconnectedsuccessfully,theIPaddressconnectionstatus
isdisplayedinthewindow
•IftheIPaddresspingisunsuccessful,thefollowingmessageisdisplayed
inthewindow:PINGfailed.
TheModemTestcanbeusedtotesttheinternalmodem.
1. Connectthedevicetoanactiveanalogphoneline.
Ensurethatthephonelineisactive.Ifyouconnecttoaninactivephoneline,the
testresultmaybeafalsenegative.
2. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2060321-001L MAC™800 49
Troubleshooting
3. SelecttheModemTestbutton.
Thefollowingwindowopens:
4. SelecttheInternalModemTestbutton.
Thefollowingmessageisdisplayedinthewindow:TestinProgress.Please
wait.
Thentheresultsofthetestaredisplayed.
•Thetestpassesifthefollowingmessageisdisplayedinthewindow:Passed.
•Thetestfailsifthefollowingmessageisdisplayedinthewindow:Failed.
IftheInternalModemTestfails,replacetheinternalmodemasdescribedin
“ReplacingtheInternalModem(option)”onpage80.
USBTest
UsetheUSBTesttotesttheUSBport.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. ConnectaUSBkeyboardtotheUSBportontherearpanelofthedevice.
NOTE:
TheUSBkeyboardusedforthistestmustmatchthelanguagethatis
selectedinsetup.
3. SelectUSBTest.
Thefollowingwindowopens:
4. PressanykeyontheUSBkeyboardandverifypassorfail:
•IfthecharacterthatappearsintheCharacterInputeldmatchesthekey
youpressed,thetestpassed.
•Ifthecharacterdoesnotmatchthekeyyoupressed,ornocharacterappears
intheCharacterInputeld,thetestfailed.
50 MAC™800 2060321-001L
5. Whenthetestisdone,pressEscorCancel.
Thefollowingwindowopens:
6. Dooneofthefollowing:
•Ifthetestpassed,pressYes
•Ifthetestfailed,pressNo.
Replacethemainboardassemblyasdescribedin“ReplacingtheMainboard
Assembly”onpage76.
PatientLeadWireTest
Usethefollowingproceduretotestthepatientleadwires:
Troubleshooting
1. OpentheDiagnosticTestswindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. Connectapatientcablewithleadwirestothedevice’spatientcableconnector.
3. Connectallleadstoapatientsimulatororshortingbar.
4. SelectPatientLeadWireCheck.
Thefollowingwindowopens:
5. SelectStartTest.
Thetestresultsaredisplayedforeachleadwire.
•Ifthefollowingmessageisdisplayed,theleadwirepassedthetest:
Connected .
•Ifthefollowingmessageisdisplayed,theleadwirefailedthetest:
Disconnected .
6. PressF6(Cancel )whenthetestiscomplete.
2060321-001L MAC™800 51
Troubleshooting
EquipmentProblems
ECGDataNoise
7. Replaceeveryleadwirethatfailedthetest.
8. Repeatthetest.
Iftheleadwirestillfailsthetest,replacethemainboardassemblyasdescribed
in“ReplacingtheMainboardAssembly”onpage76.
IftheacquiredECGdatadisplaysunacceptablenoiselevels:
•Besuretheproblemisnotcausedbypoorskinpreparation,placement,orcondition
oftheelectrodeswhentroubleshootingnoiseorsignalquality.
Carefulskinpreparationisthekeytoaninterference-freeECG.RefertothePatient
Preparation chapteroftheOperator’sManual.
SignalqualityisindicatedusingHookupAdvisor.
HookupAdvisorcanbeturnedonoroffintheECGmenu.SelectMainMenu>
SystemConguration >RestingECGSetup>PageDown.
SystemDate/TimeTroubleshooting
ErrorCodes
•Checkfordefectiveordate-expiredelectrodes.
•Checkfordefective,broken,ordisconnectedleadwires.
•RuntheAcquisitionModuleTestsintheDiagnosticmenuandmakesurealllead
wirespassthenoisetest.
Refertothe“AcquisitionModuleTest”onpage44.
TheDateand/orTimeiswronginsystem.
Therecommendedmethodisasfollows:
1. CorrectthesettingsinDate/TimeSetup.
2. Removethepowercordandbattery;thenpowerOFFthedevice.
3. Pluginpowercordandre-installbattery,powerONthedeviceandcheckthe
DateandTime.
IftheDateandTimeisnotcorrectedand/orthedeviceisveormoreyears-old;
replacetheReal-TimeClock(RTC)Battery.See“ReplacingtheReal-timeClock(RTC)
Battery”onpage68.
Noactionisnecessaryforisolatederroroccurrences.However,ifthesystem
ismalfunctioningandanyofthefollowingerrormessagesarerepeatingand
unrecoverable,replacetheFRUsintheorderlisted.
52 MAC™800 2060321-001L
AcquisitionErrorCodes
Ifyourepeatedlyreceiveanyofthefollowingacquisitionerrorcodes,replacethe
mainboardassemblyasdescribedin“ReplacingtheMainboardAssembly”onpage76.
Troubleshooting
AcquisitionErrorCodes
ErrorCode
Cause
PrinterErrorCodes
AcquisitionError-1
AcquisitionError3
AcquisitionError9
Ifyourepeatedlyreceiveanyofthefollowingprintererrorcodes,replacetheprinter
assemblyasdescribedin“ReplacingthePrinterAssembly”onpage73.
PrinterErrorCodes
ErrorCodes
PrinterInternalError2 Printheadtemperatureistoohotortoocoldtoprint
PrinterInternalError3 Printerdrivercouldnotbeopened
PrinterInternalError4 Printerdrivercommunicationerror
PrinterInternalError5 Printerdrivertimeouterror
PrinterInternalError6 Printerdrivermiscellaneouserror
PrinterInternalError7 Undenedprinterstatuswasreceived
Generalacquisitionerror
Sequencenumbererrorin100msECGPacket
Acquisitionselftesterror
Cause
FrequentlyAskedQuestions
Maintenance
NOTE:
RefertotheSystemCongurationinformationintheOperator’sManualforthis
system.
SaveSystemSetupstoSDCard
Q:HowdoIsavechangesIhavemadetotheSystemConguration?
2060321-001L MAC™800
53
Troubleshooting
A:Performthefollowingsteps:
1. InserttheSDcardintotheSDcardslotinrightsideasshown.
2. PushtheSDcardintotheslottoseatitinplace.
3. OntheMainMenu,pressF4toselectSystemConguration.
4. PressMore>More>ExportSetup.
StoringECGs
5. Onthelistonleftsideofthewindow,highlightthesetupleyouwanttosaveto
theSDcard.
6. PressExport.
Thefollowingwindowopens.
7. PressOK.
8. EjecttheSDcardbypushingitinonce.
Storeitinasecurelocation.
Q:Whywon'tanyoftheECGsIperformsavetotheSDcard?
A:Verifythefollowing:
•TheSDcardisfullyinsertedintothedrive.
•TheSDcardis128MBorgreater.
54
MAC™800 2060321-001L
Troubleshooting
•TheSDcardisnotwrite-protected.
•TryanewSDcard.
•Yoursystemissetuptoautomaticallysaverecords.
•Ifyoursystemisnotsetuptoautomaticallysaverecords,youpressedStore.
Cleaning
Q:ShouldIcleanthedevice?
A:Cleantheexteriorsurfacesofalltheequipmentandperipheraldevicesmonthly,or
morefrequentlyifneeded.
•Useaclean,softclothandamilddishwashingdetergentdilutedinwater.
•Wringtheexcesswaterfromthecloth.
Donotdripwater,oranyliquid,onthewriterassembly,andavoidcontactwith
openvents,plugs,andconnectors.
•Drythesurfaceswithacleanclothorpapertowel.
RefertotheSuppliesandAccessoriesGuide,DiagnosticCardiologyfordetailson
cleaningthedevice.
MACAddress
Calibration
Q:Ineedtoprovidetheaddressofthedevicetothenetworkadministratortoenable
theLANcommunicationoption.HowdoIobtaintheaddress?
A:Followthesestepstoobtaintheaddressofthedevice:
1. OpentheDIAGNOSTICTESTSwindowasdescribedin“AccessingtheSystem
DiagnosticsFunction”onpage39.
2. SelectServiceReportandpressthecenterbuttononthetrimpad.
3. Findtheaddressontheprintedservicereport.
Q:HowdoIcalibratethedevice?
A:Whenitbecomesnecessary,youcancalibratethedeviceusingthefollowing
procedure:
1. Usingastandardizingwaveformgenerator,producea1.00±0.01-mVpulse
signalwitharise-timenogreaterthan5msandawidthnogreaterthan100ms.
2. Connectthepulsesignaltoallavailablechannelsandsetthegainto10mm/mV.
3. Verifythatthedisplaypulseshaveanamplitudewithin±5%oftheamplitude
obtainedwhenthe1.00±0.01-mVsignalisapplied.
4. Repeatthetestforallxedgainsettingstoverifythatthestandardizationpulse
correctlyreectsthegainsetting.
NOTE:
Theerrormustbelessthan±5%oftheexpectedvalueor0.5mm,whichever
isgreater.
5. Verifythatthestandardizationsignalappearsonallchannels.
2060321-001L MAC™800
55
Troubleshooting
SystemSetup
LocationNumber
PatientQuestions
Q:Whenenteringinthepatientdata,howdoIgettheLocationeldtoautomatically
populatewiththesamenumber?
A:YoucansettheLocationnumberinBasicSetupsoyoudonotneedtoenterit
foreachtest.
1. OntheMainMenu,pressF4toselectSystemConguration.
2. PressBasicSetup.
3. UsetheTrimpadtomovethecursortotheLocationeld.
4. TypethedesiredLocationnumber.
5. PressSave>MainMenu.
Q:HowdoIchangethequestionsIseewhenIamenteringthepatientdata?
A:ThepatientquestionsyouseeonthePatientDatawindowwhenstartingatest
weresetupinPatientSetup.
1. OntheMainMenu,pressF4toselectSystemConguration.
2. PressF6(More )>F4(PatientSetup)>F4toselectPageDown.
3. UsetheTrimpadtomovethecursortoExtraQuestions...andpressthecenter
button.
Thefollowingwindowopens.
4. ForeachextraquestionyouwishtoaskinthePatientDatawindow:
a. TypethePrompt.
b. SelectthetypeofquestionfromtheTypelist:
•Alphanumeric
•Numeric
•Yes/No/Unknown
5. IntheExtraQuestions...window,pressSave.
6. IntheTestInformationSetupwindow,pressSave.
7. ReturntotheMainMenu.
56 MAC™800 2060321-001L
Passwords
Clinical
Troubleshooting
Q:ThesystemwassetupforHighSecurityModeandIforgotmypassword.HowdoI
accessthesystem?
A:Usethefollowingstepstoaccessthesystem:
1. ContactGEHealthcareTechnicalSupportandprovidetheserialnumberof
thedeviceyouwanttoaccess.
GEHealthcareTechnicalSupportwillgenerateatemporary,device-specic
nameandpasswordthatyoucanusefor24hours.
2. LogintothesystemwiththepasswordprovidedbyGEHealthcareTechSupport.
3. Immediatelyafterloggingintothesystem,verifyyourdevice’susernameand
password.
4. Recordthisinformationandstoreitinasecurelocationforfuturereference.
RestingECGReportFormat
Q:HowdoIchangethewayanECGlooks(format)whenitprintsout?
A:FollowthesestepstochangetheECGformat:
1. OntheMainMenu,pressF4toselectSystemConguration.
2. PressRestingECG.
3. PressF4(PageDown)threetimes.
4. Onthe10sECTGReportFormatlist,selectwhichtypeofECGreportyouwant
tochange.
5. OntheReportCopieslist,selectthenumberofcopiesyouwant.
6. Ifyouwantthe12SLInterpretationincludedontheECG,selectthePrint
Interpretation checkbox.
7. Ifyoudonotwantthe12SLInterpretationtoprintontheECG,clearthePrint
Interpretation checkbox.
8. PressF6tosavethesetup.
Editing
Q:Canyouedittheinterpretationonthedevice,andthentransmittheeditedrecord
totheMUSEsystemasanunconrmedrecord?
A:Thisdevicedoesnotsupporteditinterpretation.
NavigatingtheUserInterface
Q:HowdoInavigatefromthestartupscreentotheMainMenu?
A:Thesystemcanbeconguredinanumberofdifferentways.Someofthese
congurationchoicesdeterminetheactionsthatneedtobeperformedinorderto
proceedfromthepowerupdisplaytotheMainMenu.
Therearethreecongurationsthatdeterminetheinitialwindowthatappearsatpower
upandwhatactionstheuserwillneedtoperformtonavigatetotheMainMenu.
2060321-001L MAC™800
57
Troubleshooting
PowerUpmodecurrentlyselectedinBasicSetup:
HighSecuritymodeenabledinBasicSetup:
USBBarcodeReadersupportoptionactivated-yesorno.
Thevariousstepsinthissectiondescribehowtonavigatefromthepowerupscreen
totheMainMenuforthevarioussystemcongurations.Usethestepsthatapply
toyoursystemcongurationsettings.
•IfyoursystemisconguredtopowerupintheRestingECGmode,goto“Resting
ECGPowerUpMode”onpage58.
•IfyoursystemisconguredtopowerupintheArrhythmiamode,goto“Arrhythmia
ModePowerUpMode”onpage59.
•IfyoursystemisconguredtopowerupintheMainScreenmode,goto“Main
ScreenPowerUpMode”onpage59.
RestingECGPowerUpMode
ThesestepsdescribehowtonavigatetotheMainMenuafterpoweringonthesystem
whenRestingECGisselectedforPowerupmodeinBasicSetup.
NOTE:
Ifyouneedtoperformsystemsetupfunctions,besureyouloginasauserwhois
assignedsetupeditingprivileges.
1. IftheHighSecurityModeisenabled,proceedwithstepsathroughdafterthe
windowopenspromptingforaUserIDandPassword.
Ifthepasswordpromptisnotdisplayed,gotostep2.
a. TypeyouruserIDintheUserIDeld.
b. PressthedownarrowontheTrimpadtomovethecursortothePassword
eld.
58 MAC™800 2060321-001L
Troubleshooting
c. TypeyourpasswordinthePasswordeld.
d. PressLogin.
2. PressCancel>More>MainMenu.
ArrhythmiaModePowerUpMode
ThesestepsdescribehowtonavigatetotheMainMenuafterpoweringonthesystem
whenArrhythmiaisselectedforPowerupmodeinBasicSetup.
NOTE:
Ifyouneedtoperformsystemsetupfunctions,besureyouloginasauserwhois
assignedsetupeditingprivileges.
1. IftheHighSecurityModeisenabled,proceedwithstepsathroughdafterthe
windowopenspromptingforaUserIDandPassword.
Ifthepasswordpromptisnotdisplayed,gotostep2.
a. TypeyouruserIDintheUserIDeld.
b. PressthedownarrowontheTrimpadtomovetothePasswordeld.
c. TypeyourpasswordinthePasswordeld.
d. PressLogin.
Ifthebarcodereaderoptionisenabled,awindowopensprompting
youtoScanthePatientbarcode.
NOTE:
Ifthebarcodepromptisnotdisplayed,gotostep3.
2. PressCancel.
3. PressCancel>More>MainMenu.
MainScreenPowerUpMode
ThesestepsdescribehowtonavigatetotheMainMenuafterpoweringonthesystem
whenMainScreenisselectedforPowerupmodeinBasicSetup.
NOTE:
Ifyouneedtoperformsystemsetupfunctions,besureyouloginasauserwhois
assignedsetupeditingprivileges.
1. IfHighSecurityModeisenabled,proceedwithstepsathroughdafterthe
windowopenspromptingforaUserIDandPassword.
Ifthepasswordpromptisnotdisplayed,gotostep2.
a. TypeyouruserIDintheUserIDeld.
b. PressEnterorthedownarrowontheTrimpadtomovethecursorto
thePasswordeld.
c. TypeyourpasswordinthePasswordeld.
d. PressLogin.
TheMainMenuisdisplayed.
2. IfthesystemisconguredforMainScreenPowerupmodeanddoesnothave
theHighSecurityModeenabled,theMainMenuisdisplayedafterpowering
upthesystem.Youdonotneedtopressanyotherkeysinordertodisplaythe
MainMenu .
2060321-001L MAC™800 59
Troubleshooting
SettingtheWIFIParameter
UsingthefollowingprocedurestosetuptheWIFIparameter.
1. OntheMainMenu,pressF4toselectSystemConguration.
2. PressMore>More>ServiceSetup.
AwindowopenspromptingyoutoentertheServicepassword.
ContactGEHealthcaresupportifyoudonotknowtheservicepassword.
3. TypetheservicepasswordandpressF6toselectOKtoopentheServiceSetup
menu.
4. UsethetrimpadtohighlightWIFIParameterSettingsandpressthecenter
buttontoopentheWIFIParameterSettingswindow.
5. ThefollowingtablesdescribeeachsettingavailableonWIFIParameterSettings.
Field
Enable
Roaming
Description
Thiseldallowstheroamingfunctionofthesystemtobeenabled
ordisabled.
Roaming
Signal
Threshold
60 MAC™800 2060321-001L
Identiestheroamingsignalthresholdtoenableroaming.
ThiseldisavailableonlyiftheEnableRoamingeldischecked.
Thedefaultvalueis70(-dbm).
Valuesrangefrom0-150-dbm.
Troubleshooting
Field
Roaming
Interval
Roaming
Channels
WIFI
Connection
Rate
Description
Identiestheroamingintervaltoscanandconnecttowireless
accesspoint.
ThiseldisavailableonlyiftheEnableRoamingeldischecked.
Thedefaultvalueis5(s).
Valuesrangefrom2-255s.
Identiestheroamingchannelsthesystemscanfor ,forexample,
ThiseldisavailableonlyiftheEnableRoamingeldischecked.
ThedefaultvalueisAllwhichmeansthesystemscansallthe
roamingchannels.
Thevalueischangeddependingontheregion.Therangeof1to13
isusedinthecountriessupportETSI;therangeof1to11isusedin
thecountriessupportFCC.
NOTE:
Thiseldshouldbeconguredaccordingtothelabelonthe
WiFiDongle.
DeterminestheWIFIconnectionrateofthesystem.Optionsare:
•Auto
•1Mbps
•2Mbps
•5Mbps
•6Mbps
•9Mbps
•11Mbps
•12Mbps
•18Mbps
•24Mbps
•36Mbps
•48Mbps
•54Mbps
NOTE:
Iftheconnectionrateis≤11Mbps,theUSBdongleworksat
802.11bmode.
RTSThreshold DeterminesthethresholdforWIFIdongletosendtheRTSrequest.
Ifthepacketsizeissmallerthanthethreshold,theWIFIdonglewill
notsendtheRTSrequest.
Valuesrangefrom0-2347.Select0toallowtheWIFIdongleto
sendtheRTSrequesttowirelessaccesspointforeverypackage,
andselect2347tostoptheWIFIdongletosendtheRTSrequest
towirelessaccesspoint.
Thedefaultvalueis2347.
2060321-001L MAC™800 61
Troubleshooting
Field
Fragment
Threshold
Periodic
Scanwhen
Connected
Periodic
Scanwhen
Disconnected
Description
Determinesthemaximumsizeforperdatapacketbeforebeing
fragmentedintomultiplepackets.
Thedefaultvalueis2346.
Valuesrangefrom0-2346.
Determinestime-outvalueinmillisecondstoretryavalid
conguration.
Thedefaultvalueis60000ms.
Determinestime-outvalueinmillisecondstorecoverfromafailed
conguration.
Thedefaultvalueis5000ms.
6. ToresetthevaluesintheWIFIParameterSettingswindowtodefaultvalues,
pressResetToDefault.
7. TosaveallthesettingsintheWIFIParameterSettingswindow,pressSave.
Thefollowingmessagedisplays.
PressOKandthenre-pluginWIFIUSBdongletobringthesettingsintoeffect.
8. TodiscardcurrentchangedsettingsandclosetheWIFIParameterSettings
window,pressCancel.
62 MAC™800 2060321-001L
5
Maintenance
Regularmaintenance,irrespectiveofusage,isessentialtoensurethattheequipment
willalwaysbefunctionalwhenrequired.SeetheSuppliesandAccessoriesGuide,
DiagnosticCardiology forcleaningprocedures.GErecommendsthatelectricalsafety
checksbeperformedannually.
WARNING:
MAINTENANCERESPONSIBILITIES—Failureonthepartofallresponsible
individuals,hospitalsorinstitutionsemployingtheuseofthisdevicetoimplement
therecommendedmaintenanceschedulemaycauseequipmentfailureand
possiblehealthhazards.Themanufacturerdoesnot,inanymanner,assumethe
responsibilityforperformingtherecommendedmaintenanceschedule,unlessan
EquipmentMaintenanceAgreementexists.
Thesoleresponsibilityforperformingtherecommendedmaintenanceschedule
restswiththeindividuals,hospitals,orinstitutionsutilizingthedevice.
RequiredToolsandSupplies
Thefollowingtoolsarerequiredtoperformtheproceduresdescribedinthischapter:
•ECGsimulator
•Phillips#1screwdriver
•Hexagonalscrewdrivers
•Currentleakagetester
•Anti-staticwriststrap
•MAC™800RestingECGAnalysisSystemServiceManual
•MAC™800RestingECGAnalysisSystemOperator’sManual
NOTE:
Alwaysuseananti-staticwriststrapwhileopeningthedevicetoavoidpossible
damageduetostaticelectricity.
2060321-001L MAC™800 63
Maintenance
High-LevelFRUIdentication
TopCoverAssembly
BottomAssembly
Battery
LCDAssembly
InternalModem(1,option)andMainboard(2)
64 MAC™800 2060321-001L
PowerSupplyAssembly
WriterAssembly
KeypadAssembly
Maintenance
BarcodeReader(option)
2060321-001L MAC™800 65
Maintenance
PatientCable
Real-timeClockBattery
FRUReplacementProcedures
PreparingSystemforFRUReplacement
Priortoperforminganydisassemblyprocedures,performthefollowingsteps:
NOTE:
Takestrictprecautionsagainstelectrostaticdischargedamagewhilereplacingeld
replaceableunits.
1. Poweroffthesystem.
2. DisconnectthedevicefromtheACwalloutlet.
3. Disconnectthepowercordfromtherearpanelconnector.
4. Disconnectthepatientcablefromthedeviceasdescribedin“Replacingthe
PatientCable”onpage66.
5. Removethebatteryasdescribedin“ReplacingtheBatteryAssembly”onpage
67.
ReplacingthePatientCable
1. DisconnectthedevicefromACpower.
2. Disconnectthepatientcablefromthesidepanelconnectorasshowninthe
followingphotograph.
66 MAC™800 2060321-001L
3. Connectanewpatientcabletosidepanelconnector.
4. ContinuewiththefollowingfunctionalcheckoutproceduresforthisFRU.
ReplacingtheBatteryAssembly
1. DisconnectthedevicefromACpower.
Maintenance
2. Turnthedeviceover.
3. Pressthebatteryreleasetab(1)andraisethebatteryfromitscompartment
toremoveit.
WARNING:
ENVIRONMENTALHAZARDImproperdisposalofthebatterycancause
environmentalandhealthhazards.
DoNOTdisposeofthebatterybyburning.
Followlocalenvironmentalguidelinesconcerningdisposalandrecycling.
4. Insertthenewbatteryandpressthelatchuntilitsnapsintoplace.
NOTE:
Afully-chargedbatteryiscapableofprintingapproximately1000
single-pagereportsor2hoursofcontinuousoperation(withoutprinting).
5. Continuewiththefollowingfunctionalcheckoutprocedures.
2060321-001L MAC™800 67
Maintenance
ReplacingtheReal-timeClock(RTC)Battery
TheRTC4574hasaninternal32.768KHzcrystalunit.Serialcommunicationbetween
theCPUandtheRTC4574,throughtheI/OportoftheCPU,exchangestimeand
dateinformationbetweentheCPUandRTC.TheRTCwillnotdriftbymorethan320
secondsperyear.WhenthedeviceisturnedonwithACpower,theRTCispowered
by+3.3Vsupply.Whenthedeviceisturnedoff,theRTCbatterypowerstheRTCto
keeptrackofthetimeanddate.Thebatteryisa3VLithiumIoncoinbatterywith
1000mAHcapacity,whichcansufcientlysupplytheclockforaminimumof5years
(worstcasecalculation).Inthenormalcase,thefunctioningperiodisassumedto
bemorethan10years.
NOTE:
Alwaysuseananti-staticwriststrapwhileopeningthedevicetoavoidpossible
damageduetostaticelectricity.
1. RemovetheMainboardAssemblyasinstructedin“RemovingtheMainboard
Assembly”onpage77.
2. Withthedisplayconnectoratthetopandfacingupwards,locatetheRTC
batteryBTontherightsideoftheMainboard.
68 MAC™800 2060321-001L
3. Pullupthebatteryfromthebatteryholdertoremovethebattery.
4. Insertthenewbatteryintothebatteryholderonthemainboard.
5. ReassembletheMainboardassemblyasinstructedin“Reassemblingthe
MainboardAssembly”onpage78.
Maintenance
6. Adjustthetimeanddate.
7. Performthefunctionalcheckoutandvisualinspectionprocedures.
Referto“FunctionalCheckout”onpage87.
TochecktheRTCbatteryfunction,see“Real-timeClockCheck”onpage94.
ReplacingtheTopCoverAssembly
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
onpage67.
3. Removethesixscrewsfromthebottomofthedevice.
4. Turnthedevicerightsideup.
2060321-001L MAC™800 69
Maintenance
5. Presstheprinterbutton.
6. Opentheprinterdoor.
7. Liftthetopassemblyapproximately1inchattherearpanelside.
8. Pullupthelock-releasetabonmainboardkeypadconnector.
70 MAC™800 2060321-001L
9. Disconnectthekeypadcableasshowninthefollowingillustration.
10. Removetheeightscrewsfromthebottomofthetopcoverassembly.
Maintenance
11. Separatethekeypadfromthetopcoverassembly.
2060321-001L MAC™800 71
Maintenance
12. Reassembleanewtopcoverassemblybyreversingthestepsforremoval.
13. Performtheapplicablecheckoutprocedures.
ReplacingtheKeypadAssembly
1. Performstep1tostep11asdescribedin“ReplacingtheTopCoverAssembly”
2. Reassembleanewkeypadassemblybyreversingthestepsforremoval.
3. Performtheapplicablecheckoutprocedures.
ReplacingtheLCDAssembly
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
Referto“FunctionalCheckout”onpage87.
onpage69.
Referto“FunctionalCheckout”onpage87.
onpage67.
3. Removethetopcoverassemblyasdescribedin“ReplacingtheTopCover
Assembly”onpage69.
4. RemovethetwoscrewsthatholdtheLCDassemblyinplace.
5. PushtheLCDassemblyforwardandawayfromtherearpanel.
72 MAC™800 2060321-001L
6. DisconnecttheLCDcablefromthemainboard.
7. LifttheLCDassemblyoutofthebottomassembly.
8. ReassembleanewLCDassemblybyreversingthestepsforremoval.
9. Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage87.
Maintenance
ReplacingthePrinterAssembly
RemovingthePrinterAssembly
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
3. Removethetopcoverassemblyasdescribedin“ReplacingtheTopCover
4. RemovetheLCDAssemblyasdescribedin“ReplacingtheLCDAssembly”on
5. Removetheprinterdoorfromthebottomcoverassemblyasshowninthe
onpage67.
Assembly”onpage69.
page72.
followingillustration.
2060321-001L MAC™800
73
Maintenance
6. Disconnecttheprintercablefromthemainboard.
74
MAC™800 2060321-001L
7. Removethetwoscrewsfromtheprintermountingbaseasshowninthe
followingillustration.
Maintenance
8. Removetheprintermotorfromtheprintermountingbase.
2060321-001L MAC™800
75
Maintenance
ReassemblingthePrinterAssembly
1. Replaceanewprintermotoronthebottomassemblyasshowninthefollowing
illustration.
2. Replacethetwomountingscrews.
3. Reconnecttheprintercabletothemainboard.
4. Replacetheprinterdoor.
5. ReassembletheLCDassembly.
6. Reassemblethetopcoverassembly.
7. Reassemblethebatteryassembly.
8. Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage87.
ReplacingtheMainboardAssembly
ProcessingECGsinInternalStorage
Beforereplacingthemainboard,ifthesystemhastheinternalstorageoption,transmit
anyECGsremaininginstoragetoyourarchivalsystemand/orprintthemtoensure
youhaveaprintedrecord.
SavingSystemCongurationSettings
1. StoretheSystemCongurationsettingstoanSDcard.
a. InsertanSDcardintheSDcardslotofthedevice.
b. OntheMainMenu,pressF4toselectSystemConguration.
c. PressMore>More>ExportSetup.
d. HighlightthesystemsetupleyouwanttoexporttotheSDcard.
e. PressExport.
76 MAC™800 2060321-001L
2. PrinttheSystemSetupReportifyoufeelyoumayneeditforadditional
RemovingtheMainboardAssembly
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
Maintenance
f. Whenthefollowingmessageisdisplayed,press OK:Congurationwas
successfullyexported
g. RemovetheSDcardandstoreitinasecurelocation.
referenceaftertheFRUreplacementprocedure.
a. OntheMainMenu,pressF4toselectSystemConguration.
b. PressMore>PrintSetupReport.
c. UsethetrimpadtohighlightCompleteSetupandpressthecenterbutton.
d. Savetheprintedsetupreportinasecurelocation.
Youcanusethereportasareferenceifyouneedtorestorethesystem
setupmanually.
onpage67.
3. Removethetopcoverassemblyasdescribedin“ReplacingtheTopCover
Assembly”onpage69.
4. RemovetheLCDAssemblyasdescribedin“ReplacingtheLCDAssembly”on
page72.
5. Disconnecttheprintercablefromthemainboard.
2060321-001L MAC™800
77
Maintenance
6. Removethe10screwsthatholdthemainboardinplace.
7. Liftthemainboardassemblyapproximately1.5inches(3.81cm).
8. Disconnectthebatterycablefromthebottomsideofthemainboard.
9. DisconnectthePowerSupplycablefromthebottomsideofthemainboard.
ReassemblingtheMainboardAssembly
1. ReconnectthePowerSupplycabletothebottomofthenewmainboard.
2. Reconnectthebatterycabletothebottomofthenewmainboard.
78 MAC™800 2060321-001L
Maintenance
3. Replacethenewthemainboardassemblyonthebottomcoverassemblyas
showninthefollowingillustration.
4. Replacethe10screwsthatwereremovedin6.
5. Reconnecttheprintercabletothenewmainboard.
6. ReassembletheLCDassembly
7. Reassemblethetopcoverassembly.
8. Reassemblethebatteryassembly.
9. ConnectthepowercordtoACpower.
10. RestoresystemsetupsthatweresavedtotheSDcard.
11. Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage87.
SerialNumber
Whenthemainboardisreplaced,theserialnumberofthedeviceneedstobeentered
inthesystem.
Thedevice’sserialnumberislocatedontheproductlabelonthebottomofthedevice.
1. OntheMainMenu,selectF4(SystemConguration)>F6(More )>F6(More )>
F5 (ServiceSetup).
Thesystempromptsfortheservicepassword.
ContactGEHealthcaresupportifyoudonotknowtheservicepassword.
2. TypetheservicepasswordandpressthecenterbuttonontheTrimpad.
TheServicewindowopens.
3. UsetheTrimpadtomovethecursortoDeviceSettingsandpressthecenter
button.
TheDeviceSettingswindowopens.
4. Typethedevice’sserialnumberandpressF6(Save ).
2060321-001L MAC™800 79
Maintenance
Externalkeyboardlanguage
ThedefaultkeyboardlanguageofthedeviceisEnglish.Youcanalsochooseyour
preferencekeyboardlanguage.
1. OntheMainMenu,selectF4(SystemConguration)>F6(More )>F6(More )>
F5 (Service ).
Thesystempromptsfortheservicepassword.
ContactGEHealthcaretechnicalsupportifyoudonotknowtheservice
password.
2. TypetheservicepasswordandpressthecenterbuttonontheTrimpad.
TheServicewindowopens.
3. UsetheTrimpadtomovethecursortoDeviceSettingsandpressthecenter
button
TheDeviceSettingswindowopens.
4. UsetheTrimpadtomovehighlighttoKeyboardLanguage.
5. SelectyourlanguageandpressthecenterbuttonoftheTrimpad.
Thedevicesupportsthefollowinglanguages:
•English
•French
•German
•Italian
•Spanish
6. PressF6(Save ).
ReplacingtheInternalModem(option)
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
onpage67.
3. Removethetopcoverassemblyasdescribedin“ReplacingtheTopCover
Assembly”onpage69.
4. RemovetheLCDAssemblyasdescribedin“ReplacingtheLCDAssembly”on
page72.
80 MAC™800 2060321-001L
5. Removetheinternalmodemfromitssocketonthemainboard.
6. Reassembletheinternalmodembyreversingthestepsforremoval.
Alignthecontactpinswiththesocketsandaligntheholewiththeplasticpin
beforepushingitintothesockets.
Maintenance
7. Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage87.
ReplacingthePowerSupplyAssembly
RemovingthePowerSupplyAssembly
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
onpage67.
3. Removethetopcoverassemblyasdescribedin“ReplacingtheTopCover
Assembly”onpage69.
4. RemovetheLCDAssemblyasdescribedin“ReplacingtheLCDAssembly”on
page72.
5. Removethemainboardassemblyasdescribedin“RemovingtheMainboard
Assembly”onpage77.
2060321-001L MAC™800 81
Maintenance
6. RemovethesixM3X8screwsfromtheshieldplateasshowninthefollowing
illustration.
7. RemovethetwoM3X12atscrewsfromtheshieldplateasshowninthe
followingillustration.
82 MAC™800 2060321-001L
8. Removethefourhexagonscrewsfromtheshieldplateasshowninthefollowing
illustration.
9. Removetheshieldplatefromthebottomcoverassembly.
Maintenance
2060321-001L MAC™800 83
Maintenance
10. Removethefourhexagonscrewsofpowersupplyassemblyasshowninthe
followingillustration.
11. LiftthepowersupplyassemblyanddisconnecttheACcableasshowninthe
followingillustration.
ReassemblingthePowerSupplyAssembly
1. ConnecttheACcabletothenewpowersupplyassembly.
2. Placethenewpowersupplyassemblyinthebottomcoverassembly.
3. Replacethefourhexagonscrews.
4. Routethepowersupplycableforthemainboardasshowninthefollowing
illustration.
5. Replacetheshieldplate.
84 MAC™800 2060321-001L
6. ReplacethesixM3X8screws,twoM3X12atscrews,andfourhexagonscrews
ontheshieldplate.
7. Reassemblethemainboardassembly.
8. Reassembletheprinterassembly.
9. ReassembletheLCDassembly.
10. Reassemblethetopcoverassembly.
11. Reassemblethebatteryassembly.
12. Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage87.
ReplacingtheBottomCoverAssembly
1. DisconnectthesystemfromACpower.
2. Removethebatteryassemblyasdescribedin“ReplacingtheBatteryAssembly”
onpage67.
Maintenance
3. Removethetopcoverassemblyasdescribedin“ReplacingtheTopCover
Assembly”onpage69.
4. RemovetheLCDAssemblyasdescribedin“ReplacingtheLCDAssembly”on
page72.
5. Removetheprinterassemblyasdescribedin“RemovingthePrinterAssembly”
onpage73.
6. Removethemainboardassemblyasdescribedin“RemovingtheMainboard
Assembly”onpage77.
7. Removethepowersupplyassemblyasdescribedin“RemovingthePower
SupplyAssembly”onpage81.
8. Installthenewbottomcoverassembly.
9. Reassemblethepowersupplyassembly.
10. Reassemblethemainboardassembly.
11. Reassembletheprinterassembly.
12. ReassembletheLCDassembly.
13. Reassemblethetopcoverassembly.
14. Reassemblethebatteryassembly.
15. Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage87.
2060321-001L MAC™800 85
Maintenance
ReplacingtheFuse
1. DisconnectthesystemfromACpower.
2. Usingascrewdriver,takeoutthefuseholderfromtheACpluginletasshownin
thefollowingillustration.
3. Replacetwonewfusesinthefuseholder.
4. ReassemblethefuseholderintotheACpluginlet.
86 MAC™800 2060321-001L
FunctionalCheckout
Thefollowingcheckoutproceduresapplytoyoursystem.
NOTE:
TheFRUcheckoutprocedureforanylistedFRUalsoappliestoitsinternalPCBs
andcomponents.
Ifthereisanasterisk(*)listedinthefollowingtables,performtheapplicable
componentorsystemcongurationprocedures.
FRUreplacementproceduresareinprevioussectionsofthischapterofthe
manual.
Maintenance
BasicSystemFRURepairs
FRUDescription
PatientCable
KeypadAssembly
TopCoverAssembly
LCDAssembly
PrinterAssembly
MainboardAssembly
PowerSupplyAssembly
BottomCoverAssembly
BatteryAssembly
Real-timeClock(RTC) RefertotheMainboard
ACPowerCord
NonListedFRUs
VisualInspection FunctionalCheckout
1,2,7 1,2,3
3,6,7 1,2,3,7
6,7 1,2,3,14
3,6,7 1,2,3,6
6,7 1,2,3,8
6,7 1,2,3,4,5,6,7,8,9,10,11,
6,7 1,2,3
6,7 1,2,3
5
Assembly
4
6,7 1,2,3,*14
Procedures
12,13
1,2,3,11
RefertotheMainboard
Assembly
1,2,3
1
2060321-001L MAC™800 87
Maintenance
1
WhenACPowerMainsdisturbed.
OptionalSystemFRURepairs
FRUDescription
InternalModem
BarcodeReader
FRUDescription
Nopartsreplaced
Softwareupdate
Hardwareupgrade
AnnualElectricalSafety
Checkout
VisualInspection FunctionalCheckout
Procedures
6,7 1,2,3,5,13
6,7 1,2,3
Non-FRURepairs
VisualInspection FunctionalCheckout
Procedures
4,5,6,7 1,2,3,*4,5
n/a
6,7 1,2,3,*4,5
1,3,4,5 1,2,3,6,7,8,9,10,11,12,
1,2,3,*4,5
14,13
VisualInspection
Inspectthefollowingforexcessivewearandanysignsofdamage.
1. Checkforbrokenpatientcable/leadwiresandout-of-dateelectrodes.
a. Verifythatpatientcable/leadwirespassinspection.
b. Verifythattheelectrodespassinspection.
2. Discusselectrodeplacement,skinprep,andpatient-relatedrequirementswith
theECGtechnician.
Verifythatthecustomerisfollowingtheproceduresrecommendedinthe
operator’smanual.
Formoreinformation,referto“PreparingthePatient”intheoperator’smanual
foryoursystem.
3. Verifythatthekeyboard/LCDdisplaylterpassesinspection.
Referto“VisualInspection”onpage36formoreinformation.
4. VerifythattheACpowercordpassesinspection.
Referto“VisualInspection”onpage36formoreinformation.
5. Verifythatthebatterypackpassesinspection.
Referto“VisualInspection”onpage36formoreinformation.
6. Verifyallharnessesandinternalwiringaresecure.
Referto“VisualInspection”onpage36formoreinformation.
7. Verifyfastenersarereplacedandsecure.
Referto“VisualInspection”onpage36formoreinformation.
88 MAC™800 2060321-001L
FunctionalCheckoutProcedures
Performthefunctionalcheckoutproceduresthatareapplicabletothereplacement
procedureperformed.
OperationalChecks
1. Verifythatthesystempassesthepower-upself-test.
Referto“Power-UpSelf-Test”onpage35formoreinformation.Ifasoftware
updatewasperformed,verifythatthenewversionofsoftwareisdisplayedon
theMainMenuscreen.
2. Verifytherhythmstriprecordedsuccessfully.
Formoreinformation,referto“RecordingaRestingECG”intheoperator’s
manualforyoursystem.
3. VerifyanECGrecordedsuccessfully.
Formoreinformation,referto“RecordingaRestingECG”intheoperator’s
manualforyoursystem.
Maintenance
DiagnosticTests
4. VerifytheECGwasstoredsuccessfully.
Formoreinformation,referto“ManagingInternalStorage”intheoperator’s
manualforyoursystem.
5. VerifythatsimulatedECGdatawastransmittedsuccessfullytoareceiving
product.
Formoreinformation,referto“ManagingInternalStorage”intheoperator’s
manualforyoursystem.
1. Verifythatthedisplaytestwassuccessful.
Referto“DisplayTest”onpage41formoreinformation.
2. Verifythatthekeyboardtestwassuccessful.
Referto“KeyboardTest”onpage43formoreinformation.
3. Verifythatthewritertestwassuccessful.
Referto“WriterTest”onpage45formoreinformation.
4. Verifythattheacquisitionmoduletestwassuccessful.
Referto“AcquisitionModuleTest”onpage44formoreinformation.
5. Verifythatthepatientleadwirecheckwassuccessful.
Referto“PatientLeadWireTest”onpage51formoreinformation.
6. Verifythebatterytestwassuccessful.
Referto“BatteryTest”onpage44formoreinformation.
7. VerifythattheLANtestwassuccessful.
Referto“LANTest”onpage47formoreinformation.
8. Verifythatthemodemtestwassuccessful.
Referto“ModemTest”onpage49formoreinformation.
2060321-001L MAC™800 89
Maintenance
ElectricalSafetyChecks
Verifythatthecurrentleakagetestresultsmeetrequirements.
Performelectricalsafetycheckswhenindicated.Allindicatedelectricalsafetychecks
requireapass/failindicationforthestepsperformed.Recordthemeasurement
valuesinyourdebrief.
Step
EarthLeakageCurrent
1
Forward
Polarity
2
NeutralOpen,
Forward
Polarity
3
NeutralOpen,
Reverse
Polarity
4 Reverse
Polarity
Condition
ElectricalSafetyChecks
NC
SFC
SFC
NC
1
UUT—ON
______µA
______µA
______µA
______µA
2
Result LeakageCurrent
Limits
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
500µA
1,000µA
1,000µA
500µA
EnclosureLeakageCurrent
1
Forward
Polarity
2
NeutralOpen,
Forward
Polarity
3
GroundOpen,
Forward
Polarity
4
GroundOpen,
Reverse
Polarity
5
NeutralOpen,
Reverse
Polarity
Reverse
6
Polarity
PatientLeakageCurrentToGround
1
Forward
Polarity
2
NeutralOpen,
Forward
Polarity
NC
SFC
SFC
SFC
SFC
NC
NC
SFC
______µA
______µA
______µA
______µA
______µA
______µA
______µA
______µA
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
100µA
500µA
500µA
500µA
500µA
100µA
10µA
50µA
90 MAC™800 2060321-001L
Maintenance
ElectricalSafetyChecks
Step
3
GroundOpen,
Forward
Polarity
4
GroundOpen,
Reverse
Polarity
5
NeutralOpen,
Reverse
Polarity
Reverse
6
Polarity
Condition
SFC
SFC
SFC
NC
1
UUT—ON
______µA
______µA
______µA
______µA
2
Result LeakageCurrent
Limits
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
50µA
50µA
50µA
10µA
GroundContinuity
1
ACpowercord
groundprong
to
exposed
metalsurface
(groundlug)
1
NC=NormalCondition;SFC=SingleFaultCondition;N/A=NotApplicable
2
UUT=UnitUnderTest
UpdatingSoftware
SoftwareupdatesareprovidedonanSDcard.Performasoftwareupdateas
describedinthissection.
1. InserttheSDcardwiththesoftwareupdate(goldcontactsdown)intheSDcard
slotintherightsideofthedevice,asshowninthefollowingillustration.
Resistance
N/A
______Ω
Pass/Fail
Lessthan200mΩ
2060321-001L MAC™800 91
Maintenance
2. PowerupthesystemintoBootloaderbypressingtheF1+T9+Powerkeysat
thesametime,asshowninthefollowingillustration.
3. OntheBootloaderMainMenu,usethetrimpadtohighlight1=OSUpdateand
pressthecenterbuttonofthetrimpadtostartthesoftwareupdate.
92 MAC™800 2060321-001L
Maintenance
4. TheBootloaderreadsthesoftwarefromtheSDcardandwritestoFlash.Donot
pressanykeyuntilthefollowingmessagedisplays:
Writingimagecompleted.
5. PressandholdthePowerkeyuntilthesystemshutsdown.
6. PressthePowerkeyagaintorebootthesystem.
Thesystemisnowupdated.
ConditioningtheBatteryPack
Tomaintainthestoragecapacityofthebatterypackinstalledinyourdevice,GE
Healthcarerecommendsthatyouconditionthedevice’sbatterypackonceevery
sixmonthstoresettheelectronicfuelgaugeinsidethebattery.Aconditioncycle
consistsofanuninterrupted“charge-discharge-charge”cycle.
Youcanconditionthebatterypackwhileitisinstalledinadevicethatyouarenot
currentlyusingtorecordtestsonapatient.
1. DisconnecttheACpowerfromthedevice.
2. DisplaytheBATTERYSTATUSwindow:
a. OntheMainMenu,pressF4toselectSystemConguration.
b. PressMore>More>ServiceSetup.
c. Whenprompted,entertheServicepasswordandpressOK.
Ifyoudonotknowtheservicepassword,contactGEHealthcareTechnical
Support.
2060321-001L MAC™800 93
Maintenance
d. SelectSystemDiagnostics.
e. SelectBatteryTest.
TheBATTERYSTATUSwindowopens.
3. AllowthebatterytodischargeuntiltheBatteryChargeRemaining[%]isless
than5%.
4. TurnoffthedeviceandreconnectittotheACpower.
5. Allowthebatterytofullycharge.
6. RemovetheACpowerandturnonthedevice.
7. Leavethedeviceonandallowthebatterytodischargeuntilthedeviceshutsoff.
8. ReconnecttheACpowertothedevice,leavingitturnedoff,andallowthe
Real-timeClockCheck
Tocheckthatthetimeisfunctioningproperly,performthestepsdescribedinthis
section.
1. Recordthetimedisplayedonthedevice.
2. Poweroffthedeviceforveminutesandthenpoweritonagain.
3. Checktomakesurethetimehasadvancedbyveminutes.
NOTE:
AsolidamberbatteryLEDindicatesthebatteryischarging.Whenthe
batteryLEDturnsoff,thisindicatesthatthebatteryisfullycharged.
batterytofullyrecharge.
WhentheamberbatteryLEDindicatorstopsashingandturnsasolidamber,
thebatteryisfullychargedandtheconditioningcycleiscomplete.
94 MAC™800 2060321-001L
PartsList
OrderingParts
TheFRUpartslistsinthischaptersupplyenoughdetailforyoutoorderpartsforthe
assemblies,stand-aloneFRUs,andFRUkitsconsideredeldserviceable.Onlyitems,
assemblies,andkitswhichhavepartnumbersgiveninthischapterareavailablefor
purchaseasFRUs.Toorderparts,contactGEHealthcareServiceParts.
FieldReplaceableUnits(FRUs)
UpperLevelAssemblyDiagrams
Thefollowingdiagramsidentifytheeldreplaceableunitsofthesystem.Thenumbers
inthecall-outsreferencepartdescriptionsfoundin“UpperLevelAssemblyPartList”
onpage98.
6
TopCoverandKeypadAssemblies
2060321-001L MAC™800 95
PartsList
LCDandPrinterAssemblies
Mainboard,PowerSupplyandBatteryCoinAssemblies
96 MAC™800 2060321-001L
PartsList
BottomCoverAssemblies,1of3
BottomCoverAssemblies,2of3
2060321-001L MAC™800 97
PartsList
BottomCoverAssemblies,3of3
UpperLevelAssemblyPartList
Thefollowingtableidentiesthepartsavailableforthesystem.
ThenumbersintheItemcolumnrefertothecall-outsfromthediagramsfoundin
“UpperLevelAssemblyDiagrams”onpage95.
ThenumbersinthePartNumbercolumnidentifytheGEHealthcarepartnumberfor
orderableparts.Itemswithoutpartnumberscannotbepurchasedindependentlyof
aFRUkit.
UpperLevelAssembly
Item
1
2 LCDPOLYFOAM
3
4
PartNumber
ItemDescription
M3X8MACHINESCREWPHILIPSPANNIIncludedwith:
•“FRUMainboardAssembly,PN2061575–001”onpage104
•“FRUBottomCoverAssembly,PN2039943-001”onpage
105
•“FRUKits,PN2039945-001”onpage108
Includedwith“FRUTopCoverAssembly,PN2039939-001”
onpage103
LCDLENS
Includedwith“FRUTopCoverAssembly,PN2039939-001”
onpage103
PRINTERBUTTON
includedwith“FRUTopCoverAssembly,PN2039939-001”
onpage103
98 MAC™800 2060321-001L
UpperLevelAssembly(cont'd.)
PartsList
Item
5
PartNumber
TOPCOVER
ItemDescription
Includedwith“FRUTopCoverAssembly,PN2039939-001”
onpage103
6 NAMEPLATE15MMGELOGO
Includedwith“FRUTopCoverAssembly,PN2039939-001”
onpage103
7 KEYPAD
Referto“Keypads”onpage107.
8 2061574–001 MAC800V2FRULCDASSEMBLY
9 PRINTERDOOR
includedwith“FRUPrinterAssembly,PN2039941-001”on
page104
10 THERMALPRINTERHEAD-MAC800
includedwith“FRUPrinterAssembly,PN2039941-001”on
page104
11 M2X6MACHINESCREWPHILIPSPAN
Includedwith:
•“FRUPrinterAssembly,PN2039941-001”onpage104
•“FRUKits,PN2039945-001”onpage108
12 PRINTERHOLDER
includedwith“FRUBottomCoverAssembly,PN2039943-001”
onpage105
13 PCBMAC800MAINBOARD
includedwith“FRUMainboardAssembly,PN2061575–001”
onpage104
14
REARPANEL
includedwith“FRUMainboardAssembly,PN2061575–001”
onpage104
15 ACQCOVER1
includedwith“FRUMainboardAssembly,PN2061575–001”
onpage104
16 ACQCOVER2
includedwith“FRUMainboardAssembly,PN2061575–001”
onpage104
17 PATIENTCABLECONNECTORHOLDER
includedwith“FRUMainboardAssembly,PN2061575–001”
onpage104
2060321-001L MAC™800 99
PartsList
UpperLevelAssembly(cont'd.)
Item
18
PartNumber
AN003884-40x3/8SCREWNI
ItemDescription
includedwith:
•“FRUMainboardAssembly,PN2061575–001”onpage104
•“FRUKits,PN2039945-001”onpage108
19 SDCARDHOUSE
includedwith“FRUMainboardAssembly,PN2061575–001”
onpage104
20 60W12VOUTPUTAC-DCMODULEBIGHEATSINK
includedwith“FRUPowerSupplyAssembly,PN2040052-001”
onpage104
21 FLATSCREWM3X12
includedwith:
•“FRUBottomCoverAssembly,PN2039943-001”onpage
105
•“FRUKits,PN2039945-001”onpage108
22 CRIMPRING.217
includedwith:
•“FRUBottomCoverAssembly,PN2039943-001”onpage
105
•“FRUKits,PN2039945-001”onpage108
24 M4X16MACHINESCREWPHILIPSPANNI
includedwith:
•“FRUBottomCoverAssembly,PN2039943-001”onpage
105
•“FRUKits,PN2039945-001”onpage108
25 2039946-001 FUSE2.0AQUI-BLO250V5X20MM
AvailableasanindependentFRU.Alsoincludedwith“FRU
BottomCoverAssembly,PN2039943-001”onpage105.
26 HEXAGONSCREW1
includedwith:
•“FRUBottomCoverAssembly,PN2039943-001”onpage
105
•“FRUKits,PN2039945-001”onpage108
27 HEXAGONSCREW2
includedwith:
•“FRUBottomCoverAssembly,PN2039943-001”onpage
105
•“FRUKits,PN2039945-001”onpage108
100 MAC™800 2060321-001L