GEHealthcare
MAC™2000
ECGAnalysisSystem
ServiceManual
SoftwareVersion1.1
2053535-003RevisionM
MAC2000ECGAnalysisSystem
English
© 2013-2014,2016-2017GeneralElectric
Company.
AllRightsReserved.
PublicationInformation
TheinformationinthismanualappliesonlytotheMAC™2000ECGAnalysisSystem,Version1.1.Itdoesnotapplytoearlierproductversions.
Duetocontinuingproductinnovation,specicationsinthismanualaresubjecttochangewithoutnotice.
MUSE,MAC,CASE/CardioSoft,and12SLaretrademarksownedbyGEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectric
CompanygoingtomarketasGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheirrespectiveowners.
Thisproductcomplieswiththerequirementsconcerningmedicaldevicesfromthefollowingregulatorybodies.Formoreinformationabout
compliance,refertotheRegulatoryandSafetyGuideforthisproduct.
Thedocumentpartnumberandrevisionareoneachpageofthedocument.Therevisionidentiesthedocument’supdatelevel.Therevision
historyofthisdocumentissummarizedinthefollowingtable.
Revision
Date
Comments
A
B
C 13August2013
D
F
G
H
J
K
L
M
ToaccessotherGEHealthcareDiagnosticCardiologydocuments,gototheCommonDocumentationLibrary(CDL),locatedat
www.gehealthcare.com/documents,andclickCardiology.
ToaccessOriginalEquipmentManufacturer(OEM)documents,gotothedevicemanufacturer'swebsite.
17October2012 Internalrelease
15July2013 CustomerRelease
UpdatedforSPR229270.
8September2014 Updatedwithinformationfornewdisplayreplacingobsoletedisplay.
12July2016 Updateformaintenanceprojects.
25October2016 UpdatetheSilexFRUpartsforSilexWWproject.
26April2017 Addedthenewsuppliesandaccessoriesmanual.
UpdatesmadeforthesimplicationandclaritythroughtheKeepitSimple
ServiceInitiative.
8June2017
5September2017 Updatetheinformationaboutthenewcapabilitybattery.
30September2017 Updatetheinformationaboutembeddedwirelessmodule.
11December2017 Updatetoaddthedetailinformationaboutinsettheantennaoftheembedded
UpdateaccordingtoMAC2000SP5project.
Updateaccordingtoaddthenewprintcablerelatedinformation.
Updatedsomeproductspeechication
wirelessmodule.
2 MAC™2000 2053535-003M
11December2017
ServiceManualLanguageInformation
WARNING
(EN)
ПРЕДУПРЕЖДЕНИЕ
(BG)
警告
(ZH-CN)
警告
(ZH-TW)
UPOZORENJE
(HR)
VAROVÁNÍ
(CS)
ThisservicemanualisavailableinEnglishonly.
•Ifacustomer'sserviceproviderrequiresalanguageotherthanEnglish,itisthecustomer's
responsibilitytoprovidetranslationservices.
•Donotattempttoservicetheequipmentunlessthisservicemanualhasbeenconsulted
andisunderstood.
•Failuretoheedthiswarningmayresultininjurytotheserviceprovider ,operator,orpatient,
fromelectricshock,mechanicalorotherhazards.
Товаупътванезаработаеналичносамонаанглийскиезик.
•Акодоставчикътнауслугатанаклиентаизискадругезик,задължениенаклиентае
даосигурипревод.
•Неизползвайтеоборудването,предидастесеконсултиралииразбралиупътването
заработа.
•Неспазванетонатовапредупреждениеможедадоведедонараняваненадоставчика
науслугата,оператораилипациентврезултатнатоковударилимеханичнаили
другаопасност.
本维修手册仅提供英文版本。
•如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
•未详细阅读和完全理解本维修手册之前,不得进行维修。
•忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
本維修手冊只提供英文版。
•如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責提供翻
譯服務。
•除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維修。
•不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或其他因素
而受到傷害。
Oveuputezaservisiranjedostupnesusamonaengleskomjeziku.
•Ukolikokorisničkiserviszahtijevanekidrugijezik,korisnikovajeodgovornostosigurati
odgovarajućiprijevod.
•Nemojtepokušavatiservisiratiopremuukolikonistekonzultiraliirazumjelioveupute.
•Nepoštivanjeovogupozorenjamožerezultiratiozljedamaservisnogosoblja,korisnikaili
pacijentaprouzročenimelektričnimudaromtemehaničkimilinekimdrugimopasnostima.
Tentoprovoznínávodexistujepouzevanglickémjazyce.
•Vpřípadě,žeexterníslužbazákazníkůmpotřebujenávodvjinémjazyce,jezajištění
překladudoodpovídajícíhojazykaúkolemzákazníka.
•Nesnažteseoúdržbutohotozařízení,anižbystesipřečetlitentoprovoznínávoda
pochopilijehoobsah.
•Vpřípaděnedodržovánítétovarovánímůžedojítkporaněnípracovníkaprodejního
servisu,obslužnéhopersonálunebopacientůvlivemelektrickéhoproudu,respektive
vlivemmechanickýchčijinýchrizik.
2053535-003M MAC™2000
3
ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(DA)
Denneservicemanualndeskunpåengelsk.
•Hvisenkundesteknikerharbrugforetandetsprogendengelsk,erdetkundensansvar
atsørgeforoversættelse.
•Forsøgikkeatservicereudstyretmedmindredenneservicemanualharværetkonsulteret
ogerforstået.
•Manglendeoverholdelseafdenneadvarselkanmedføreskadepågrundafelektrisk,
mekaniskellerandenfareforteknikeren,operatørenellerpatienten.
WAARSCHUWING
(NL)
DezeservicemanualisalleeninhetEngelsverkrijgbaar.
•Indienhetonderhoudspersoneeleenanderetaalnodigheeft,danisdeklant
verantwoordelijkvoordevertalingervan.
•Probeerdeapparatuurnietteonderhoudenvoordatdezeservicemanualgeraadpleegd
enbegrepenis.
•Indiendezewaarschuwingnietwordtopgevolgd,zouhetonderhoudspersoneel,de
gebruikerofeenpatiëntgewondkunnenrakenalsgevolgvaneenelektrischeschok,
mechanischeofanderegevaren.
HOIATUS
(ET)
Käesolevteenindusjuhendonsaadavalainultinglisekeeles.
•Kuiklienditeeninduseosutajanõuabjuhenditinglisekeelesterinevaskeeles,vastutab
klienttõlketeenuseosutamiseeest.
•Ärgeüritageseadmeidteenindadaenneeelnevaltkäesolevateenindusjuhendiga
tutvumistjasellestarusaamist.
•Käesolevahoiatuseeiraminevõibpõhjustadateenuseosutaja,operaatorivõipatsiendi
vigastamistelektrilöögi,mehaanilisevõimuuohutagajärjel.
VAROITUS
(FI)
Tämähuolto-ohjeonsaatavillavainenglanniksi.
•Josasiakkaanhuoltohenkilöstövaatiimuutakuinenglanninkielistämateriaalia,tarvittavan
käännöksenhankkiminenonasiakkaanvastuulla.
•Äläyritäkorjatalaitteistoaennenkuinoletvarmastilukenutjaymmärtänyttämän
huolto-ohjeen.
•Mikälitätävaroitustaeinoudateta,seurauksenavoiollahuoltohenkilöstön,laitteiston
käyttäjäntaipotilaanvahingoittuminensähköiskun,mekaanisenviantaimuun
vaaratilanteenvuoksi.
ATTENTION
(FR)
Cemanueltechniquen'estdisponiblequ'enanglais.
•Siunservicetechniqueclientsouhaiteobtenircemanueldansuneautrelangueque
l'anglais,ildevraprendreenchargelatraductionetlaresponsabilitéducontenu.
•Nepastenterd'intervenirsurleséquipementstantquelemanueltechniquen'apasété
consultéetcompris.
•Lenon-respectdecetavertissementpeutentraînerchezletechnicien,l'opérateuroule
patientdesblessuresduesàdesdangersélectriques,mécaniquesouautres.
4
MAC™2000 2053535-003M
11December2017
ServiceManualLanguageInformation(cont'd.)
WARNUNG
(DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
AÐVÖRUN
(IS)
PERINGATAN
(ID)
DieseServiceanleitungistnurinenglischerSpracheverfügbar.
•FallsderKundendiensteineandereSprachebenötigt,musserfüreineentsprechende
Übersetzungsorgen.
•KeineWartungdurchführen,ohnedieseServiceanleitunggelesenundverstandenzu
haben.
•BeiZuwiderhandlungkanneszuVerletzungendesKundendiensttechnikers,des
AnwendersoderdesPatientendurchStromschläge,mechanischeodersonstigeGefahren
kommen.
Τοπαρόνεγχειρίδιοσέρβιςδιατίθεταιστααγγλικάμόνο.
•Εάντοάτομοπαροχήςσέρβιςενόςπελάτηαπαιτείτοπαρόνεγχειρίδιοσεγλώσσαεκτός
τωναγγλικών,αποτελείευθύνητουπελάτηναπαρέχειυπηρεσίεςμετάφρασης.
•Μηνεπιχειρήσετετηνεκτέλεσηεργασιώνσέρβιςστονεξοπλισμόεκτόςεάνέχετε
συμβουλευτείκαιέχετεκατανοήσειτοπαρόνεγχειρίδιοσέρβις.
•Εάνδενλάβετευπόψητηνπροειδοποίησηαυτή,ενδέχεταιναπροκληθείτραυματισμός
στοάτομοπαροχήςσέρβις,στοχειριστήήστονασθενήαπόηλεκτροπληξία,μηχανικούς
ήάλλουςκινδύνους.
Ezaszervizkézikönyvkizárólagangolnyelvenérhetőel.
•Haavevőszervizellátójaangoltóleltérőnyelvretartigényt,akkoravevőfelelőssége
afordításelkészíttetése.
•Nepróbáljaelkezdenihasználniaberendezést,amígaszervizkézikönyvbenleírtakat
nemértelmeztékésértettékmeg.
•Ezengyelmeztetésgyelmenkívülhagyásaaszervizellátó,aműködtetővagyapáciens
áramütés,mechanikaivagyegyébveszélyhelyzetmiattisérüléséteredményezheti.
Þessiþjónustuhandbókereingöngufáanlegáensku.
•Efaðþjónustuveitandiviðskiptamannsþarfnastannarstungumálsenensku,erþaðskylda
viðskiptamannsaðskaffatungumálaþjónustu.
•Reyniðekkiaðafgreiðatækiðnemaþessiþjónustuhandbókhefurveriðskoðuðogskilin.
•Brotáaðsinnaþessariaðvörungeturleitttilmeiðslaáþjónustuveitanda,stjórnandaeða
sjúklingifráraosti,vélrænumeðaöðrumáhættum.
ManualservisinihanyatersediadalambahasaInggris.
•JikapenyediajasaservispelangganmemerlukanbahasalainselaindariBahasaInggris,
merupakantanggungjawabdaripenyediajasaservistersebutuntukmenyediakan
terjemahannya.
•Janganmencobamelakukanservisterhadapperlengkapankecualitelahmembacadan
memahamimanualservisini.
•Mengabaikanperingataninibisamengakibatkancederapadapenyediaservis,operator,
ataupasien,karenaterkenakejutlistrik,bahayamekanisataubahayalainnya.
2053535-003M MAC™2000
5
ServiceManualLanguageInformation(cont'd.)
AVVERTENZA
(IT)
警告
(JA)
CẢ NHBÁO
(VI)
ЕСКЕРТУ
(KK)
BRĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
IlpresentemanualedimanutenzioneèdisponibilesoltantoinInglese.
•Seunaddettoallamanutenzionerichiedeilmanualeinunalinguadiversa,ilclienteè
tenutoaprovvederedirettamenteallatraduzione.
•Siprocedaallamanutenzionedell'apparecchiaturasolodopoaverconsultatoilpresente
manualeedavernecompresoilcontenuto.
•Ilnonrispettodellapresenteavvertenzapotrebbefarcompiereoperazionidacuiderivino
lesioniall'addetto,allamanutenzione,all'utilizzatoreedalpazienteperfolgorazione
elettrica,perurtimeccaniciodaltririschi.
このサービスマニュアルは英語版しかありません。
•サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
•このサービスマニュアルを熟読し、十分に理解をした上で装置のサービスを
行ってください。
•この警告に従わない場合、サービスを担当される方、操作員あるいは患者が、感電
や機械的又はその他の危険により負傷する可能性があります。
TàiLiệ uHướ ngDẫ nSử aChữ achỉcóbả ntiế ngAnh.
•Nế ucácđơ nvịcungcấ pdị chvụchokháchhàngyêucầ umộ tngônngữnàokháctiế ng
Anh,thìkháchhàngsẽcótráchnhiệ mcungcấ pcácdị chvụdị chthuậ t.
•Khôngđượ csử achữ athiế tbịtrừkhiđãthamkhả ovàhiể uTàiliệ uHướ ngdẫ nSử achữ a.
•Khôngtuânthủnhữ ngcả nhbáonàycóthểdẫ nđế ncáctổ nthươ ngchongườ ithự c
hiệ nsử achữ a,ngườ ivậ nhànhhaybệ nhnhân,dosố cđiệ n,cácrủ irovềcơkhíhay
cácrủ irokhác.
Бұ лқызметкө рсетубойыншанұ сқ аулығ ытекағ ылшынтіліндеқолжетімді.
•Тұ тынушыныңқызметпровайдеріағ ылшынтіліненбасқ атілдегінұ сқ аныталапетсе,
аударубойыншақызметтеріменқамтамасызетутұ тынушыжауапкершілігіндеболуы
тиіс.
•Бұ лқызметкө рсетубойыншанұ сқ аулығ ынназарғ аалып,тү сінбегенше,жабдыққ а
қ ызметкө рсетуденбастарты ң ыз.
•Бұ лескертудіелемеуқызметпровайдері,операторнемесеемделушініңэлектр
шогынан,механикалықнемесебасқ ақауіптернә тижесіндежарақ аталуынаәкелуі
мү мкін.
Šīapkalpotājurokasgrāmatairpieejamatikaiangļuvalodā.
•Jaapkalpošanassniedzējamnepieciešamainformācijacitā,nevisangļu,valodā,klienta
pienākumsirnodrošināttāstulkošanu.
•Neveicietaprīkojumaapkopi,neizlasotunnesaprototapkalpotājurokasgrāmatu.
•Šībrīdinājumaneievērošanavarradītelektriskāsstrāvastrieciena,mehāniskuvaiciturisku
izraisītutraumuapkopessniedzējam,operatoramvaipacientam.
Šiseksploatavimovadovasyraprieinamastikanglųkalba.
•Jeiklientopaslaugųtiekėjasreikalaujavadovokitakalba-neanglų,numatytivertimo
paslaugasyraklientoatsakomybė.
•Nemėginkiteatliktiįrangostechninėspriežiūros,nebentatsižvelgėteįšįeksploatavimo
vadovąirjįsupratote.
•Jeineatkreipsitedėmesioįšįperspėjimą,galimisužalojimaidėlelektrosšoko,mechaninių
arkitųpaslaugųtiekėjui,operatoriuiarpacientui.
6 MAC™2000 2053535-003M
11December2017
ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(NO)
Denneservicehåndbokennnesbarepåengelsk.
•Hviskundensserviceleverandørtrengeretannetspråk,erdetkundensansvaråsørge
foroversettelse.
•Ikkeforsøkåreparereutstyretutenatdenneservicehåndbokenerlestogforstått.
•Manglendehensyntildenneadvarselenkanføretilatserviceleverandøren,operatøren
ellerpasientenskadespågrunnavelektriskstøt,mekaniskeellerandrefarer .
OSTRZEŻENIE
(PL)
Niniejszypodręcznikserwisowydostępnyjestjedyniewjęzykuangielskim.
•Jeślidostawcausługklientawymagajęzykainnegoniżangielski,zapewnienieusługi
tłumaczeniajestobowiązkiemklienta.
•Nienależyserwisowaćwyposażeniabezzapoznaniasięizrozumienianiniejszego
podręcznikaserwisowego.
•Niezastosowaniesiędotegoostrzeżeniamożespowodowaćurazydostawcyusług,
operatoralubpacjentawwynikuporażeniaelektrycznego,zagrożeniamechanicznego
bądźinnego.
AVISO
(PT-BR)
Estemanualdeassistênciatécnicasóseencontradisponíveleminglês.
•SeoserviçodeassistênciatécnicadoclientenãoforGE,eprecisardeoutroidioma,será
daresponsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanualde
assistênciatécnica.
•Onãocumprimentodesteavisopodeporemperigoasegurançadotécnico,operadorou
pacientedevidoachoqueselétricos,mecânicosououtros.
AVISO
(PT-PT)
Estemanualtécnicosóseencontradisponíveleminglês.
•Seaassistênciatécnicadoclientesolicitarestesmanuaisnoutroidioma,éda
responsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanual
técnico.
•Onãocumprimentodesteavisopodeprovocarlesõesaotécnico,aoutilizadorouao
pacientedevidoachoqueseléctricos,mecânicosououtros.
AVERTISMENT
(RO)
Acestmanualdeserviceestedisponibilnumaiînlimbaengleză.
•Dacăunfurnizordeserviciipentruclienţinecesităoaltălimbădecâtceaengleză,estede
datoriaclientuluisăfurnizezeotraducere.
•Nuîncercaţisăreparaţiechipamentuldecâtulteriorconsultăriişiînţelegeriiacestui
manualdeservice.
•Ignorareaacestuiavertismentarputeaducelarănireadepanatorului,operatoruluisau
pacientuluiînurmapericolelordeelectrocutare,mecanicesaudealtănatură.
ПРЕДУПРЕЖДЕНИЕ
(RU)
Настоящееруководствопообслуживаниюпредлагаетсятольконаанглийскомязыке.
•Еслисервисномуперсоналуклиентанеобходиморуководствоненаанглийском,ана
каком-тодругомязыке,клиентуследуетобеспечитьпереводсамостоятельно.
•Преждечемприступатькобслуживаниюоборудования,обязательнообратитеськ
настоящемуруководствуивнимательноизучитеизложенныевнемсведения.
•Несоблюдениетребованийданногопредупрежденияможетпривестиктому,что
специалистыпообслуживанию,операторыилипациентыполучатударэлектрическим
током,механическуютравмуилидругоеповреждение.
2053535-003M MAC™2000
7
ServiceManualLanguageInformation(cont'd.)
UPOZORENJE
(SR)
VAROVANIE
(SK)
OPOZORILO
(SL)
ADVERTENCIA
(ES)
VARNING
(SV)
UYARI
(TR)
Ovoservisnouputstvojedostupnosamonaengleskomjeziku.
•Akoklijentovserviserzahtevanekidrugijezik,klijentjedužandaobezbediprevodilačke
usluge.
•Nepokušavajtedaopraviteuređajakonistepročitaliirazumeliovoservisnouputstvo.
•Zanemarivanjeovogupozorenjamožedovestidopovređivanjaservisera,rukovaocaili
pacijentausledstrujnogudara,ilimehaničkihidrugihopasnosti.
Tentonávodnaobsluhujekdispozíciilenvangličtine.
•Akzákazníkovposkytovateľslužiebvyžadujeinýjazykakoangličtinu,poskytnutie
prekladateľskýchslužiebjezodpovednosťouzákazníka.
•Nepokúšajtesaoobsluhuzariadeniaskôr ,akosineprečítatenávodnaobsluhua
neporozumietemu.
•Zanedbanietohtovarovaniamôževyústiťdozraneniaposkytovateľaslužieb,obsluhujúcej
osobyalebopacientaelektrickýmprúdom,mechanickýmaleboinýmnebezpečenstvom.
Taservisnipriročnikjenavoljosamovangleškemjeziku.
•Čeponudnikstoritvestrankepotrebujepriročnikvdrugemjeziku,morastrankazagotoviti
prevod.
•Neposkušajteservisiratiopreme,četegapriročnikanistevcelotiprebraliinrazumeli.
•Četegaopozorilaneupoštevate,selahkozaradielektričnegaudara,mehanskihalidrugih
nevarnostipoškodujeponudnikstoritev,operateralibolnik.
Estemanualdeserviciosóloexisteeninglés.
•Sielencargadodemantenimientodeunclientenecesitaunidiomaquenoseaelinglés,el
clientedeberáencargarsedelatraduccióndelmanual.
•Nosedeberádarserviciotécnicoalequipo,sinhaberconsultadoycomprendidoeste
manualdeservicio.
•Lanoobservanciadelpresenteavisopuededarlugaraqueelproveedordeservicios,el
operadoroelpacientesufranlesionesprovocadasporcausaseléctricas,mecánicaso
deotranaturaleza.
Denhärservicehandbokennnsbaratillgängligpåengelska.
•Omenkundsserviceteknikerharbehovavettannatspråkänengelskaansvararkunden
föratttillhandahållaöversättningstjänster.
•Försökinteutföraservicepåutrustningenomduinteharlästochförstårdenhär
servicehandboken.
•Omduintetarhänsyntilldenhärvarningenkandetresulteraiskadorpåserviceteknikern,
operatörenellerpatiententillföljdavelektriskastötar,mekaniskafarorellerandrafaror .
Buserviskılavuzununsadeceİngilizcesimevcuttur.
•Eğermüşteriteknisyenibukılavuzuİngilizcedışındabirbaşkalisandantalepederse,bunu
tercümeettirmekmüşteriyedüşer.
•Serviskılavuzunuokuyupanlamadanekipmanlaramüdahaleetmeyiniz.
•Buuyarıyauyulmaması,elektrik,mekanikveyadiğertehlikelerdendolayıteknisyen,
operatörveyahastanınyaralanmasınayolaçabilir .
8 MAC™2000 2053535-003M
11December2017
ServiceManualLanguageInformation(cont'd.)
ЗАСТЕРЕЖЕННЯ
(UK)
CẢ NHBÁO
(VI)
Данекерівництвозсервісногообслуговуванняпостачаєтьсявиключноанглійськоюмовою.
•Якщосервіснийінженерпотребуєкерівництвоіншоюмовою,користувачзобов'язаний
забезпечитипослугиперекладача.
•Ненамагайтесяздійснюватитехнічнеобслуговуванняданогообладнання,якщо
винечитали,абонезрозумілиінформацію,наданувкерівництвізсервісного
обслуговування.
•Недотриманняцьогозастереженняможепризвестидотравмуваннясервісного
інженера,користувачаданогообладнанняабопацієнтавнаслідокелектричногошоку,
механічногоушкодженняабозіншихпричинневірногообслуговуванняобладнання.
TàiLiệ uHướ ngDẫ nSử aChữ achỉcóbả ntiế ngAnh.
•Nế ucácđơ nvịcungcấ pdị chvụchokháchhàngyêucầ umộ tngônngữnàokháctiế ng
Anh,thìkháchhàngsẽcótráchnhiệ mcungcấ pcácdị chvụdị chthuậ t.
•Khôngđượ csử achữ athiế tbịtrừkhiđãthamkhả ovàhiể uTàiliệ uHướ ngdẫ nSử achữ a.
•Khôngtuânthủnhữ ngcả nhbáonàycóthểdẫ nđế ncáctổ nthươ ngchongườ ithự c
hiệ nsử achữ a,ngườ ivậ nhànhhaybệ nhnhân,dosố cđiệ n,cácrủ irovềcơkhíhay
cácrủ irokhác.
2053535-003M MAC™2000 9
10 MAC™2000 2053535-003M
11December2017
Contents
1 Introduction
IntendedUserofthisProduct...................................................................17
IndicationsforUse....................................................................................17
Contraindications.....................................................................................18
PrescriptionDeviceStatement.................................................................18
RegulatoryandSafetyInformation...........................................................18
SafetyConventions..................................................................................18
SafetyHazards........................................................................................19
ClassicationofMedicalDevice...................................................................21
CerticationInformation............................................................................21
RecordingECGsDuringDebrillation.............................................................21
AccuracyofInputSignalReproduction...........................................................22
ModulatingEffectsinDigitalSystems............................................................22
EMI/EMC/RFSafetyInformation...................................................................22
Biocompatibility.......................................................................................23
LegalNotice...........................................................................................23
SuppliesandAccessories...........................................................................24
ResponsibilityoftheManufacturer................................................................24
ResponsibilityofthePurchaser/Customer.......................................................24
ProductandPackagingInformation..............................................................24
Training.................................................................................................32
EquipmentIdentication...........................................................................32
ProductLabel.........................................................................................33
SerialNumberFormat...............................................................................33
DeviceAddressLabelandRatingPlate...........................................................34
ProductCodes........................................................................................34
ServiceInformation...................................................................................35
ServiceRequirements...............................................................................35
WarrantyInformation...............................................................................35
AdditionalAssistance................................................................................35
ManualInformation..................................................................................35
ManualPurpose......................................................................................35
DocumentConventions.............................................................................35
2 ProductOverview
ProductDescription..................................................................................37
HardwareOverview..................................................................................37
FrontView.............................................................................................38
RearView..............................................................................................38
SideView...............................................................................................39
InternalView..........................................................................................40
2053535-003M MAC™2000 11
Keypads................................................................................................41
ArchitectureandTheoryofOperation......................................................44
BlockDiagram........................................................................................44
Hardware/FirmwareArchitecture.................................................................46
ProductInterfaces....................................................................................46
3 InstallingtheUSBPoweredSilexWirelessBridge
InstallingtheSilexWirelessBridge...........................................................50
4 InstallingtheEmbeddedWirelessModule
InstallingtheEmbeddedWirelessModule................................................54
PreparingtheSystem................................................................................54
RemovingtheBattery...............................................................................55
RemovingtheTopCoverAssembly...............................................................56
RemovingtheDisplayBottomCover.............................................................58
InstallingtheAntenna...............................................................................60
RemovingtheMainBoard..........................................................................63
ReplacingtheMainBoard..........................................................................65
ReattachingtheTopCoverAssembly.............................................................66
InstallingtheBattery.................................................................................69
5 ServiceSetup
AccessingServiceSetup............................................................................71
AccessingDeviceSettings.........................................................................72
EventLog..................................................................................................72
SettingUpEventLogging...........................................................................73
ExportingtheEventLog.............................................................................74
AccessingSystemDiagnostics..................................................................74
TestingtheDisplay...................................................................................75
TestingtheSpeaker..................................................................................77
TestingtheKeypad...................................................................................78
TestingtheAcquisitionModule....................................................................79
CheckingtheBatteryStatus........................................................................79
TestingtheWriter....................................................................................80
CommunicationsTests..............................................................................83
TestingthePatientLeadwires......................................................................89
PrintingaServiceReport............................................................................90
UpdatingtheSoftware..............................................................................90
FormatingtheFlashMemory....................................................................98
OpeningaCommandPrompt..................................................................100
SettingtheWIFIParameter.....................................................................101
SettingtheWIFIParameterforDongleWirelessModule....................................101
SettingtheWIFIParameterforEmbeddedWirelessModule...............................103
WirelessCountryofOperation................................................................105
12 MAC™2000 2053535-003M
6 Maintenance
7 Troubleshooting
RecommendedMaintenance..................................................................107
RequiredToolsandSupplies...................................................................107
FRUReplacementProcedures.................................................................108
HighLevelFRUIdentication.....................................................................108
PreparingtheSystemforFRUReplacement..................................................109
Battery................................................................................................109
DeviceCovers.......................................................................................111
PowerSupply........................................................................................134
PowerInletModule.................................................................................136
Printer................................................................................................138
DisplayAssembly...................................................................................155
Keypad...............................................................................................162
InternalModem.....................................................................................169
Antenna..............................................................................................171
EmbeddedWirelessModule......................................................................176
PWA...................................................................................................177
KISSPumpAssembly(Option)....................................................................183
FunctionalCheckout...............................................................................188
VisualInspections..................................................................................190
FunctionalCheckoutProcedures................................................................190
ECGDataNoise.......................................................................................193
GeneralFaultIsolation............................................................................194
ErrorCodes.............................................................................................197
8 PartsList
PartIdentication...................................................................................199
FRULists..................................................................................................211
PowerUpSelfTest..................................................................................194
VisualInspection....................................................................................195
EventLogging.......................................................................................196
DiagnosticTests....................................................................................196
CoverBottom(2066261-001).....................................................................211
Printer(2066261-002/2066261-080)...........................................................211
PrinterMotorandGears(2066261-003)........................................................211
PrinterDoor(2066261-004).......................................................................211
CoverTop,ColorChassis,Hardware(2066261-005).........................................212
MainPWA(2066261-077).........................................................................212
DisplayPlasticParts—AUO(2066261-007)...................................................212
Keypad(StressorNon-Stress)...................................................................213
Hardware—1(2066261-009)....................................................................213
Harness/Cables(2066261-010/2066261-076)................................................213
PrinterDataCables(2066261-066/2066261-076)...........................................214
Hardware3andPower-InModule(2066261-011)...........................................214
PowerSupply(2066261-012).....................................................................214
Battery(2066261-013).............................................................................214
HighCapacityBattery(2066261-085)..........................................................214
DisplayandO-Rings(2066261-014)............................................................215
NLTDisplay,DisplayHolder,andO-Rings(2066261–061)...................................215
2053535-003M MAC™2000 13
DisplayOverlayandO-Rings—AUO(2066261-015).........................................215
DisplayOverlayandO-Rings—NLT(2066261-063)..........................................215
CoverTopandPlasticParts—1(2066261-016)..............................................215
PlasticParts-Printer/PaperTray(2066261-017).............................................216
DOORSENSORPARTS(2066261-070)...........................................................216
PaperSensor(2066261-018).....................................................................216
PrintHeadAssy(2066261-019)..................................................................216
StaticBrush(2066261-020).......................................................................216
KISSPump(2066261-021)........................................................................217
KISSPUMPASSEMBL YWITHBASE(2066261-069)............................................217
InternalModem(2066261-022)..................................................................217
DisplayAssembly—NLTwithNon-stressKeypad............................................217
DisplayAssembly—NLTwithStressKeypad..................................................218
Keypads,Non-Stress...............................................................................219
Keypads,Stress.....................................................................................220
BarcodeScannerKits(USBDataMatrix).......................................................220
PowerSupplyCords................................................................................221
MAC2000DataSDHCCard(2066261-065)....................................................221
SoftwareUpdateSDCard.........................................................................222
USBWi-FiModule..................................................................................222
SilexWirelessBridge-US(2066261-071)......................................................222
SilexWirelessBridgeKit-WW(2066261-075)................................................222
SilexWirelessBridgeCables(2066261-072)...................................................222
SilexWirelessBridge-DualLocks(2066261-073)............................................222
UpgradeEmbeddedWireless(2072532-116).................................................222
FRUMAC2000Antenna(2066261-086)........................................................223
FRUMAC2000Wi-FiModule(2066261-087)..................................................223
FRUMAC2000MainBoardwithEmbeddedWirelessSocket(2066261-088)...........223
FRUSupportingFoamandClip(2066261-089)...............................................223
9 SystemConguration
BasicSetup..............................................................................................225
RestingECGSetup...................................................................................230
ArrhythmiaSetup...................................................................................242
StressECGSetup.....................................................................................245
StressECGSettings.................................................................................245
EditingStressProtocols...........................................................................249
CommunicationSetup.............................................................................252
CountrySetup.........................................................................................265
PrintSetupReport...................................................................................266
PatientSetup...........................................................................................267
UserSetup...............................................................................................273
SelectSetup.............................................................................................275
ImportSetup...........................................................................................276
ExportSetup............................................................................................277
OptionsSetup..........................................................................................277
14 MAC™2000 2053535-003M
ServiceSetup...........................................................................................279
Date/TimeSetup.....................................................................................280
OrderManagerSetup..............................................................................280
RRAnalysisSetup....................................................................................281
PDFFileNamingConvention...................................................................281
DefaultNamingConvention......................................................................281
CustomizingtheNamingConvention...........................................................282
RetrievingYourPassword.......................................................................283
ExportingtheAuditTrail.........................................................................283
A RelatedDocuments
DocumentsRelatedtoMAC™2000ECGAnalysisSystem.......................285
DocumentsAvailableinOtherLanguages..............................................285
B ElectromagneticCompatibility(EMC)
GuidanceandManufacturer’sDeclaration—Electromagnetic
Emissions................................................................................................289
GuidanceandManufacturer’sDeclaration—Electromagnetic
Immunity.................................................................................................290
GuidanceandManufacturer’sDeclaration—Electromagnetic
Immunity.................................................................................................291
RecommendedSeparationDistances.....................................................292
C TechnicalSpecications
SystemSpecications.............................................................................295
SilexWirelessBridge...............................................................................302
D WirelessCountryList
AllowableChannels.................................................................................305
E MainboardandPrintAssemblyReplacementMatrix
................................................................................................................307
2053535-003M MAC™2000 15
16 MAC™2000 2053535-003M
Introduction
ThisdocumentdescribestheMAC™2000ECGAnalysisSystem,alsoreferredtoasthe
“product”,“system”,or“device”.Thedocumentisintendedtobeusedbyanyonewho
maintainsortroubleshootsthisequipment.
Thischapterprovidesgeneralinformationrequiredfortheproperuseofthesystem
andthismanual.Familiarizeyourselfwiththisinformationbeforeusingthesystem.
IntendedUserofthisProduct
TheMAC™2000ECGAnalysisSystemisaportableECGacquisition,analysis,and
recordingsystemthatisintendedforusebytrainedoperatorsinahospitalormedical
professional’sfacilityenvironment,aswellasusedinclinics,physicianofces,
outreachcenters,orwhereverECGtestingisperformed.
1
IndicationsforUse
TheMAC™2000ECGAnalysisSystemisaportabledeviceintendedtobeusedby
orunderthedirectsupervisionofalicensedhealthcarepractitionerusingsurface
electrodestoacquire,analyze,display,andrecordinformationforadultandpediatric
populationsinahospital,medicalprofessional’sfacility,clinics,physician’sofceor
outreachcenters.
NOTE:
Pediatricpopulationsaredenedaspatientsbetweentheagesof0and15years.
TheMAC™2000ECGAnalysisSystemprovidesthefollowingmodesofoperation:
•RestingECGmode
•Arrhythmiamode
•Exercisemodeforexercisestresstesting(optional)
•RRanalysismodeforRRintervalanalysis(optional)
Thebasicsystemprints6or12leadsofECGandisupgradeabletoprovidesoftware
optionssuchas12-leadECGmeasurementandinterpretiveanalysis.
Arrhythmiadetectionisprovidedfortheconvenienceofautomaticdocumentation.
TransmissionandreceptionofECGdatatoandfromacentralECGcardiovascular
informationsystemisoptional.
2053535-003M MAC™2000 17
Introduction
Contraindications
Thissystemisnotintendedforuseinthefollowingmanner:
•Duringpatienttransport
•Withhigh-frequencysurgicalunits
•Asanintra-cardiacapplication
•Asavitalsignsphysiologicalmonitor
PrescriptionDeviceStatement
CAUTION:
UnitedStatesfederallawrestrictsthisdevicetosalebyorontheorderofa
physician.
RegulatoryandSafetyInformation
SafetyConventions
Thissectionprovidesinformationaboutthesafeuseandregulatorycomplianceof
thissystem.Familiarizeyourselfwiththisinformation,andreadandunderstandall
instructionsbeforeattemptingtousethissystem.Thesystemsoftwareisconsidered
medicalsoftware.Assuch,itwasdesignedandmanufacturedtotheappropriate
medicalregulationsandcontrols.
NOTE:
Disregardingthesafetyinformationprovidedinthismanualisconsidered
abnormaluseofthissystemandcouldresultininjury,dataloss,oravoided
warranty.
AHazardisasourceofpotentialinjurytoaperson,property,orthesystem.
ThismanualusesthetermsDANGER,WARNING,CAUTION,andNOTICEtopointout
hazardsandtodesignateadegreeorlevelofseriousness.Familiarizeyourselfwith
thefollowingdenitionsandtheirsignicance.
DenitionsofSafetyConventions
Safety
Convention
DANGER
WARNING
CAUTION
NOTICE
Denition
Indicatesanimminenthazard,which,ifnotavoided,willresultindeath
orseriousinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided,
couldresultindeathorseriousinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided,
couldresultinmoderateorminorinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided,
couldresultinthelossordestructionofpropertyordata.
18 MAC™2000 2053535-003M
SafetyHazards
Thefollowingmessagesapplytothesystemasawhole.Specicmessagesmayalso
beprovidedelsewhereinthemanual.
WARNING:
WARNING:
WARNING:
Introduction
EQUIPMENTMALFUNCTION—Anyattemptbyunauthorizedpersonneltoservice
thedevicecouldresultinequipmentmalfunctionandvoidthewarranty.
Thisequipmentcontainsnouser-serviceableparts.Referservicingtoauthorized
servicepersonnel.
PATIENTINJURY–STRANGULATION—Cablespresentapossiblestrangulation
hazard.
Toavoidpossiblestrangulation,routeallcablesawayfromthepatient’sthroat.
Useashortversionofcableforpediatricpatients.
PERSONALINJURY–STUMBLINGHAZARD—Patientscanbecomeentangledinthe
cablesandleadwiresconnectedtothedevice,whichcouldcausethepatient
tostumbleortrip.
Routecablesandleadwiresinawaytoavoidcreatingastumblinghazard:keep
themofftheoor,androuteleadwiresawayfromthepatient’slegsandthe
healthcareprovider’sworkarea.
WARNING:
MAGNETICANDELECTRICALINTERFERENCE—Magneticandelectriceldscan
interferewiththeacquisitionofECGreadings.
Makesurethatallperipheralcomponentsoperatedinthevicinityofthedevice
complywiththerelevantEMCrequirements.X-rayequipment,MRIdevices,
radiosystems(cellularphones)andsoforth,arepossiblesourcesofinterference
becausetheymayemithigherlevelsofelectromagneticradiation.Verifythe
performanceofthesystembeforeuse.
WARNING:
EXPLOSIONHAZARD—Usingthisdeviceinthepresenceofanestheticvaporsor
liquidscancauseexplosions.
Donotusethisdeviceinthepresenceofanestheticvaporsorliquids.Onlypersons
withadequatetraininginthecorrectuseofthisdevicemayusethisdevice.
WARNING:
EQUIPMENTFAILURE—Polarizingelectrodes(stainlesssteelorsilverconstructed)
maycausetheelectrodestoretainaresidualchargeafterdebrillation.Aresidual
chargeblocksacquisitionoftheECGsignal.
Wheneverpatientdebrillationisapossibility,usenon-polarizingelectrodes
(silver-silverchlorideconstruction)forECGmonitoring.
2053535-003M MAC™2000 19
Introduction
WARNING:
PERSONALINJURY—Contactwithpatientsduringdebrillationcancauseserious
injuryordeath.
Donotcontactpatientsduringdebrillation.Patientsignalinputslabeledwiththe
CFandBFsymbolswithpaddlesareprotectedagainstdamageresultingfrom
debrillationvoltages.
Toensureproperdebrillatorprotection,useonlyGEHealthcarerecommended
cablesandleadwires.Properplacementofdebrillatorpaddlesinrelationtothe
electrodesisrequiredtoensuresuccessfuldebrillation.
WARNING:
INTERPRETATIONHAZARD—ResultsoftheautomatedQTanalysisarenot
consideredadiagnosis.
Aqualiedphysicianorcardiologistmustreviewandconrmthemeasurements
andwaveformsrecordedbythesystem.Itshouldbeusedonlyasanadjunctto
theclinicalhistory,symptoms,andresultsofothertests.
WARNING:
INTERPRETATIONHAZARD—Computerizedinterpretationisonlysignicantwhen
usedinconjunctionwithclinicalndings.
Aqualiedphysicianmustoverreadallcomputer-generatedtracings.
WARNING:
POORSIGNALQUALITY—Improperskinpreparationcancausepoorsignalquality
duringtheECGrecording.
Carefulskinpreparationisthekeytoaninterference-freeECG.
WARNING:
IMPROPERUSE—Thisisaprescriptivedevice.
Thisequipmentisintendedforusebyorunderthedirectsupervisionofalicensed
healthcarepractitioner .
WARNING:
EXPLOSIONHAZARD—Batteriesmayexplodeinres
Donotdisposeofthebatterybyre.Followlocalenvironmentalguidelines
concerningdisposalandrecycling.
WARNING:
ELECTRICSHOCKHAZARD/SYSTEMMALFUNCTIONLiquidsinsideadevicecan
causeelectricshockorsystemmalfunction.
Donotallowliquidstoenterthedevice.Ifliquidsenterthedevice,turnitoffand
informyourservicetechnician.Donotusethedeviceuntilitischeckedbya
servicetechnician.
WARNING:
ELECTRICSHOCK—Improperconnectionofthisequipmentmaycauseelectric
shock.
Toavoidriskofelectricshock,thisequipmentmustonlybeconnectedtoasupply
mainswithprotectiveearth.
20 MAC™2000 2053535-003M
NOTE:
Followtheinstructionsprovided.Donotpositionequipmentinawaythatmakesit
difculttodisconnectthedevicewhenusinganappliancecoupler,mainsplug,or
otherseparableplugasameansofisolation.
ClassicationofMedicalDevice
Thedeviceisclassiedasfollows,accordingtoIEC60601-1:
MedicalDeviceClassications
Introduction
Category
Typeofprotectionagainstelectrical
shock
Degreeofprotectionagainstelectrical
shock
Degreeofprotectionagainstsolids TheIPcodeforthisdeviceisIP20.
Degreeofprotectionagainstharmful
ingressofliquids(IP20)
Degreeofsafetyofapplicationinthe
presenceofammableanesthetic
mixturewithairorwithoxygenorwith
nitrousoxide
Method(s)ofsterilizationordisinfection
recommendedbythemanufacturer
Classication
ClassIinternallypoweredequipment
TypeCFdebrillation-proofappliedpart
Protectedagainstsolidforeignobjectswitha
diameterof12.5mmandgreater
Theobjectprobe,asphere12.5mmdiameter,
shallnotfullypenetrate.Thejointedtestnger
12mmdiameter,80mmlength,shallhave
adequateclearancefromhazardousparts.
TheIPcodeforthisdeviceisIP20.
Non-protected
Thisdeviceisordinaryequipment(enclosed
equipmentwithoutprotectionagainstingress
ofliquids)
Equipmentisnotsuitableforuseinthepresence
ofaammableanestheticmixturewithairor
withoxygenorwithnitrousoxide
Notapplicable
Modeofoperation
Continuousoperation
CerticationInformation
MedicalEquipment
Withrespecttoelectricshock,re,andmechanicalhazardsonlyin
accordancewithUL60601–1,andCAN/CSAC22.2NO.601.1.
RecordingECGsDuringDebrillation
Thisequipmentisprotectedagainsttheeffectsofcardiacdebrillatordischarge
toensurerecovery,asrequiredbyteststandards.Thepatientsignalinputofthe
acquisitionmoduleisdebrillation-proof.Therefore,itisnotnecessarytoremove
theECGelectrodespriortodebrillation.
2053535-003M MAC™2000 21
Introduction
Whenusingstainlesssteelorsilverelectrodes,adebrillatordischargecurrentmay
causetheelectrodestoretainaresidualchargecausingapolarizationorDCoffset
voltage.ThiselectrodepolarizationblocksacquisitionoftheECGsignal.Toavoidthis
condition,ifthereisasituationwhereadebrillationproceduremightbenecessary,
usenon-polarizingelectrodes(whichdonotformaDCoffsetvoltagewhensubjected
toaDCcurrent)suchassilver/silver-chloridetypes.
Ifyouusepolarizingelectrodes,GEHealthcarerecommendsdisconnectingthe
leadwiresfromthepatientbeforedeliveringtheshock.
ElectrodedebrillationrecoveryistheabilityoftheelectrodetoallowtheECGtraceto
returnafterdebrillation.GEHealthcarerecommendsusingnon-polarizingdisposable
electrodeswithdebrillationrecoveryratingsasspeciedinAAMIEC125.2.2.4.AAMI
EC12requiresthatthepolarizationpotentialofanelectrodepairdoesnotexceed100
mV5secondsafteradebrillationdischarge.
Refertothesuppliesandaccessoriesreferenceguideforthissystemforalistof
approvedelectrodes.
AccuracyofInputSignalReproduction
•OverallSystemErrormeetsAAMIEC113.2.7.1requirements.OverallSystemErroris
betweenorwithin±5%or±40µV,whicheverisgreater .
•FrequencyResponsemeetsAAMIEC113.2.7.2requirements,usingtestingmethods
AandD.Frequencyresponseisbetweenorwithin±10%between0.67and40Hz
andbetween+0and-10%for20ms,1.5mVtriangularinput.
ModulatingEffectsinDigitalSystems
Thisdeviceusesdigitalsamplingtechniquesthatmayproducesomevariationin
amplitudesofQ,R,and/orSwavesfromoneheartbeattothenext,whichmaybe
particularlynoticeableinpediatricrecordings.Ifyouobservethisphenomenon,
beawarethattheoriginofamplitudevariationsisnotentirelyphysiological.For
measuringvoltagesofQ,R,andSwaves,GEHealthcareadvisesusingtheQRS
complexeswiththelargestdeectionoftheparticularwaves.
EMI/EMC/RFSafetyInformation
Thissystemisdesignedandtestedtocomplywithapplicableregulationsregarding
EMCandmustbeinstalledandputintoserviceaccordingtotheEMCinformation
statedintheElectromagneticCompatibilityappendixoftheServiceand/orOperator’s
manual.ChangesormodicationstothissystemnotexpresslyapprovedbyGE
HealthcarecouldcauseEMCissueswiththisorotherequipment.Mainspowershould
beastandardcommercialorhospitalenvironment.
Beforeinstallingorusingthedeviceorsystem,beawareoftheproximityofknownRF
sources,suchasthefollowing:
•RadioandTVstations
•PortableandmobileRFcommunicationdevices(cellphones,two-wayradios)
•X-ray,CT,orMRIdevices
22 MAC™2000 2053535-003M
Thesedevicesarealsopossiblesourcesofinterferenceastheymayemithigher
levelsofelectromagneticradiation.
WARNING:
EQUIPMENTMALFUNCTION/INTERFERENCE—Useofportablephonesorother
radiofrequency(RF)emittingequipmentnearthesystemmaycauseunexpected
oradverseoperation.
Donotuseportablephonesorotherelectronicequipmentthatmayemitradio
frequency(RF)nearthissystem.
WARNING:
EQUIPMENTMALFUNCTION/INTERFERENCE—Donotusetheequipmentorsystem
adjacentto,orstackedwith,otherequipment.
Ifadjacentorstackeduseisnecessary,testtheequipmentorsystemtoverify
normaloperationinthecongurationinwhichyouareusingit.
WARNING:
ACCESSORIES/COMPONENTS—Addingaccessoriesorcomponents,ormodifying
themedicaldeviceorsystem,mayresultinincreasedEMISSIONSordecreased
IMMUNITYofthedeviceorsystem.
Introduction
UsethefollowingresourcesformoreinformationonEMI/EMCandRFconcerns:
•TheSuppliesandAccessoriesReferenceGuideforyoursystem
•QualiedGEHealthcareorapprovedthird-partypersonnel
•TheElectromagneticCompatibilityappendixinyoursystemserviceoroperator’s
manual
NOTE:
Complianceprovidesreasonableprotectionagainstradio-frequencyinterference.
However,thereisnoguaranteethatinterferencewillnotoccurinaparticular
installation.Youcantellwhetherthisdeviceorsystemiscausinginterferenceby
turningitoff.Iftheinterferencestops,itwasmostlikelycausedbythedeviceor
system.
Biocompatibility
Thepartsofthesystemdescribedinthismanualthatcomeintocontactwiththe
patientduringtheintendeduse,includingallaccessories,fulllthebiocompatibility
requirementsoftheapplicablestandards.Ifyouhavequestionsinthismatter,contact
yourGEHealthcarerepresentative.
LegalNotice
GEHealthcaresoftwarecontainsseveraleldsthatcanbelledinbeforeperforming
anECG.Someoftheseeldsarerequired,whileothersareoptionalandlefttothe
usertoassesswhethertheyareneededtoperformtheexam.TheeldRaceisone
oftheseoptionalelds.Racehasbeenacknowledgedbythemedicalprofessionas
usefultoanalyzesomepathologies.Youshouldbeawarethat,insomejurisdictions,
theprocessingofdatarevealinganindividual'sracialoriginissubjecttolegal
requirements,suchasobtainingthepatient'spriorconsent.Ifyouelecttocollectthis
typeofdata,itisyourresponsibilitytoensurethatyoucomplywithallapplicable
legalrequirements.
2053535-003M MAC™2000 23
Introduction
SuppliesandAccessories
YoushoulduseonlysuppliesandaccessoriesthatGEHealthcarerecommends.For
alistofrecommendations,refertothesuppliesandaccessoriesreferenceguide
forthissystem
ContactGEHealthcarebeforeusinganythingthatisnotrecommendedforthissystem.
ResponsibilityoftheManufacturer
GEHealthcareisresponsibleforthesafety,reliability,andperformanceofhardware
suppliedbyGEHealthcareonlyifthefollowingconditionsaremet:
•Assemblyoperations,extensions,readjustments,modications,orrepairsare
performedbypersonsauthorizedbyGEHealthcare.
•Theelectricalinstallationoftheroomwherethedeviceisusedcomplieswiththe
requirementsoftheappropriatelocal,state,andothergovernmentregulations.
•Theequipmentisusedinaccordancewiththeinstructionsforuse.
ResponsibilityofthePurchaser/Customer
Thecustomerisresponsibleforprovidingappropriatedesks,chairs,electricalwall
outlets,networkconnections,andanalogphonelines,andforlocatinganyofthe
systemcomponentsdescribedinthismanualincompliancewithalllocal,state,and
nationalcodes.
ProductandPackagingInformation
Thissectionidentiesthefollowing:
•Hardwarelabelsandtheirlocationsonpage24
•SymbolDescriptionsonpage25
HardwareLabelLocations
24 MAC™2000 2053535-003M
Introduction
Item
Label
1
2
3
4
Location Description
Backofthedevice ProductLabel
Identiesthisdevice.See
“ProductLabel”onpage33
foradescriptionofthelabel
contents.
Backofthedevice DeviceAddressLabeland
RatingPlate
Itprovidesregulatoryand
cautionaryinformation.See
“DeviceAddressLabeland
RatingPlate”onpage34for
anexplanationofthelabel.
Bottomcoverofthedevice TheOptionCodelabel.
Usetheoptioncodes
tosetupthepurchased
optionsinyoursystem.
Usetheoptioncodes
tosetupthepurchased
optionsinyoursystem.
Referto“OptionsSetup”in
theoperatormanual.
Ontheshippingpackage Environmentalsymbols
requiredforshipping.
5
6
Ontheshippingpackage BatteryShippingLabel.
FRAGILE—LithiumIon
batteriescancausereif
damaged.
Ontheshippingpackage Theshippinglabel.
SymbolDescriptions
Thefollowingtabledescribessymbolsoriconsthatmaybeonthedeviceorits
packaging.Notallofthesymbolsdenedinthetableapplytoyourdeviceorits
packaging.
Symbolsareusedtoconveywarnings,cautions,prohibitions,mandatoryactions,
orinformation.Anysymbolonyourdeviceorpackagingwithmarkingsincolor
indicatestheremaybeadanger,warning,ormandatoryaction.Anysymbolonyour
deviceorpackagingthatisinblackandwhiteprovidesadditionalinformationormay
2053535-003M MAC™2000 25
Introduction
indicateacaution.Familiaritywiththesesymbolsassistsintheuseanddisposal
oftheequipment.
Forequipmentsymbolsnotshown,refertotheoriginalequipmentmanufacturer
(OEM)manuals.
SymbolDescriptions
Symbol
Description
CatalogorOrderablePartNumber
Indicatesthemanufacturer'scatalogorpartnumber.
SerialNumber
Indicatesthemanufacturer'sserialnumber.
BatchCodeorLotNumber
Indicatesthemanufacturer'sbatchcodeorlotnumber.
DateofManufacture (Year-Month)
Indicatestheoriginalmanufacturedateforthisdevice.
Manufacturer
Indicatesthenameandaddressforthemanufacturerofthisdevice.It
mayalsoincludethedateitwasmanufactured.
AuthorizedRepresentativeintheEuropeanCommunity
Indicatesthenameandaddressoftheauthorizedrepresentativeinthe
EuropeanCommunityforthisdevice.
UniqueDeviceIdenticationisauniquemarkingforidenticationof
themedicaldevice.
RxOnly
USFederallawrestrictsthisdevicetosalebyorontheorderofa
physician.
12SL
IndicatesthedeviceusestheMarquette™12SLECGAnalysisProgram
toanalyzeandinterpretECGreadings.
IPxy
26 MAC™2000 2053535-003M
IPCode(IngressProtectionRating)
Classiesandratesthedegreeofprotectionprovidedagainstthe
intrusionofsolidobjects(suchasbodypartslikehandsandngers,
dust,accidentalcontact),andliquids.
Therstnumeral(x)representsthedegreeofprotectionagainstthe
ingressofsolidobjects.
Thesecondnumeral(y)representsthedegreeofprotectionagainstthe
ingressofliquids.
ForproductswithanIPxyrating,seetheClassicationofMedicalDevice
inthischapterforadescriptionofthatrating.Notallproductshave
anIPxyrating.
SymbolDescriptions(cont'd.)
Introduction
Symbol
Description
ClassIIEquipment
Identiesequipmentthatmeetsthesafetyrequirementsspeciedfor
classIIequipmentbyIEC60601–1.
Thisdevicewasdesignedsothatitdoesnotrequireasafetyconnection
toelectricalearth(USground).Nosinglefailureresultsindangerous
voltagebecomingexposedandcausinganelectricshock.Thisis
achievedwithoutrelyingonanearthedmetalcasing.
ConsultInstructionsforUse
Consulttheoperatinginstructions.
Debrillation-proofTypeBFAppliedPart
Identiesadebrillation-prooftypeBFappliedpartonmedical
equipmentthatcomplieswithIEC60601–1.
Thisdevicemeetstherequirementsforprotectionagainstelectric
shockforanearth-free(oating)appliedpart(oneintendedforcontact
withpatients).
Debrillation-proofTypeCFAppliedPart
Identiesadebrillation-prooftypeCFappliedpartonmedical
equipmentthatcomplieswithIEC60601–1.
Thisdevicemeetstherequirementsforprotectionagainstelectric
shockforanearth-free(oating)appliedpart(oneintendedforcontact
withpatients)forcardiacapplication.
NoUser–orField-serviceableParts
Donotopenordisassemblethedeviceforanyreason.
ProtectiveEarth (ground)
Identiestheterminalofaprotectiveearth(ground)electrodeorany
terminalthatisintendedforconnectiontoanexternalconductorfor
protectionagainstelectricshockincaseofafault.
Non-ionizingElectromagneticRadiation
Indicatesthattheequipmentemitselevated,potentiallyhazardous,
levelsofnon-ionizingradiation(electromagneticenergy)fordiagnosis
ortreatment.
FollowInstructionsForUse
Readandunderstandtheoperator'smanualbeforeusingthedevice
orproduct.
Asamandatoryactionsign,thissymbolisidentiedbyablue
backgroundandwhitesymbol.
CAUTION:
SAFETYGROUNDPRECAUTION
Pullingonthecablecancausethecordtodeteriorateresultingin
electricalproblems.
Removethepowercordfromthemainssourcebygraspingthe
plug.DONOTpullonthecable.
2053535-003M MAC™2000 27
Introduction
SymbolDescriptions(cont'd.)
Symbol
Description
CAUTION:
CONSULTACCOMPANYINGDOCUMENTS
Theremaybespecicwarningsorprecautionsassociatedwiththe
devicethatarenototherwisefoundonthelabel.
Consulttheaccompanyingdocumentationformoreinformation
aboutsafelyusingthisdevice.
CAUTION:
ELECTRICSHOCK
Indicatesthepresenceofhazardousenergycircuitsorelectric
shockhazards.
Toreducetheriskofelectricshockhazards,donotopenthis
enclosure.Referservicingtoqualiedpersonnel.
CAUTION:
HOTSURFACE
Indicatesthatthemarkeditemmaybehot.
Takeappropriateprecautionsbeforetouchingtheitem.
WARNING:
BODILYINJURY
Indicatesthepresenceofmechanicalpartsthatcanresultin
pinching,crushing,orotherbodilyinjury.
Toavoidriskofbodilyinjury,keepawayfrommovingparts.
Disconnectpowerbeforereachingintoareaorservicing.
Asawarningsign,thissymbolisidentiedbyayellowbackground,black
triangularband,andablacksymbol.
WARNING:
HANDCRUSHINGHAZARD
Thisdevicecontainsmovingpartsthatcouldcrushtheuser'shand.
Keephandsclearofthedevicewhileitisinoperation.Disconnect
powerbeforereachingintoorservicingthedevice.
Asawarningsign,thissymbolisidentiedbyayellowbackground,black
triangularband,andablacksymbol.
WARNING:
BODILYINJURY
Indicatesthepresenceofasharpedgeorobjectthatcancause
cutsorotherbodilyinjury.
Topreventcutsorotherbodilyinjury,donotcontactsharpedgeof
object.
Asawarningsign,thissymbolisidentiedbyayellowbackground,black
triangularband,andablacksymbol.
28 MAC™2000 2053535-003M
SymbolDescriptions(cont'd.)
Introduction
Symbol
Description
WARNING:
BODILYINJURY
Indicatesthepresenceofapotentialtip-overhazardthatcanresult
inbodilyinjury.
Toavoidriskofbodilyinjury,followallinstructionsformaintaining
thestabilityoftheequipmentduringtransport,installation,and
maintenance.
Asawarningsign,thissymbolisidentiedbyayellowbackground,black
triangularband,andablacksymbol.
WARNING:
PINCHPOINT
Thisdevicecontainsmovingpartsthatcouldpinchbodyparts.
Keephandsclearofthedevicewhileitisinoperation.Disconnect
thepowerbeforereachingintoorservicingthedevice.
Asageneralprohibitionsign,thissymbolisidentiedbyawhite
background,redcircularbandandslash,andablacksymbol.
WARNING:
PERSONALINJURYDONOTREACHIN
Reachingintotheequipmentcancausepersonalinjury.
Donotplacehandsintoanyopenings.
Asageneralprohibitionsign,thissymbolisidentiedbyawhite
background,redcircularbandandslash,andablacksymbol.
WARNING:
ENVIRONMENTALORHEALTHHAZARD
Incineratingthedeviceorproductcouldpresentarisktothe
environmentorhumanhealth.
Donotincineratethisdeviceorproduct.
Asageneralprohibitionsign,thissymbolisidentiedbyawhite
background,redcircularbandandslash,andablacksymbol.
WARNING:
BREAKAGEDUETOHEAVYLOAD
Heavyobjectsonthesurfacemaycauseittobreak.
Donotloadobjectsheavierthanthemaximumpermissibleload
indicatedforasafeworkingload.
Asageneralprohibitionsign,thissymbolisidentiedbyawhite
background,redcircularbandandslash,andablacksymbol.
CanBeRecycled
Indicatesyoumayrecyclethismaterialordevice.Recycleordisposeof
inaccordancewithlocal,state,orcountrylaws.
2053535-003M MAC™2000 29
Introduction
SymbolDescriptions(cont'd.)
Symbol
Description
WasteElectricalandElectronicEquipment(WEEE)
Indicatesthisequipmentcontainselectricalorelectroniccomponents
thatmustnotbedisposedofasunsortedmunicipalwastebutcollected
separately.Contactanauthorizedrepresentativeofthemanufacturer
forinformationconcerningthedecommissioningofyourequipment.
Contains<heavymetalchemicalsymbol>
Indicatesthisequipmentcontainsheavymetalandmustnotbe
disposedofasunsortedmunicipalwastebutcollectedseparately.
TheexampleshowsLithiumIon.
EnvironmentalFriendlyUsePeriod(EFUP)
PerChinesestandardSJ/T11363–2006,indicatesthenumberofyears
fromthedateofmanufactureduringwhichyoucanusetheproduct
beforeanyrestrictedsubstancesarelikelytoleak,causingapossible
environmentalorhealthhazard.
NOTE:
•Ifthedevicecontainslessthanthemaximumconcentrationof
restrictedsubstances,thesymbolcontainsalowercasee
•ThisisalsoreferredtoasChinaRoHS.
JapanRoHS
Indicatesthedeviceorproductmeetstheregulationslimitorbanfor
specicsubstancesinnewelectronicandelectricequipmentinJapan.
TheGreenMark(withtheG)indicatestheproductiswithinthe
tolerancesofhazardouschemicals.
TheContentMark(withtheRandlettersbelow)indicateswhich
hazardoussubstance(s)wasusedduringthemanufacturingofthe
electricalorelectronicequipmentthatexceedsmaximumtolerances.
Fragile
Indicatesthecontentsarefragile.Handlewithcare.
ThisWayUp
Indicatesthecorrectuprightpositionofthepackage.
DoNotStack
Indicatesthatyoushouldnotstackthecontainerorplacealoadon
thecontainer.
KeepDry
Indicatesthatyouneedtokeepthecontainerawayfromrainandother
sourcesofmoisture.
30 MAC™2000 2053535-003M
SymbolDescriptions(cont'd.)
Introduction
Symbol
Description
HumidityLimits
Indicatesupperandlowerhumiditylimitsforthetransportationand
handlingofthispackage.Theyareindicatednexttotheupperand
lowerhorizontallines.
AtmosphericLimits
Indicatestheupperandlowerbarometricpressurelimitationsforthe
transportationandhandlingofthispackage.Theyareindicatednextto
theupperandlowerhorizontallines.
UpperTemperatureLimit
Indicatesthemaximumtemperaturefortransportationandhandlingof
thispackage.Thelimitisindicatednexttotheupperhorizontalline.
TemperatureLimits
Indicatestheupperandlowertemperaturelimitsforthetransportation
andhandlingofthispackage.Theyareindicatednexttotheupper
andlowerhorizontallines.
Thefollowingtabledescribescerticationsymbolsthatmaybeusedonyourdevice
oritspackaging.Theinclusionofasymbolinthistabledoesnotindicatethatyour
productwascertiedbythatsymbol’sgoverningbodyandislistedforreference
only.Toidentifywhichorganizationshavecertiedyourdevice,refertothelabeling
onyourdeviceoritspackaging.
Certication
Symbol
Description
ULMark
IndicatescompliancewithapplicableUnderwritersLaboratories
requirements.
ULListedMark
Indicatescompliancewithinternationalorregionalstandardsfor
UnderwritersLaboratoriessafetyrequirements.
ULListed,Canada/US
Indicatescompliancewithinternationalorregionalstandardsfor
UnderwritersLaboratoriessafetyrequirementsinCanadaandthe
UnitedStates.
ULClassicationMark
IndicatesthismedicalequipmentisULClassiedwithrespectto
electricshock,re,andmechanicalhazardsonlyinaccordancewithUL
60601-1,CAN/CSAC22.2NO.601.1,andIEC60601-2-25.
ULClassicationMark,Canada/US
IndicatesthismedicalequipmentisULClassiedwithrespectto
electricshock,re,andmechanicalhazardsonlyinaccordancewith
UL60601-1,CAN/CSAC22.2NO.601.1,andIEC60601-2-25forthe
USandCanada.
2053535-003M MAC™2000 31
Introduction
Certication
Symbol
Description
CEMark
IndicatesthedeviceorproductconformswithapplicableEU(European
Union)directives.
PCT(GOST-R)Mark
IndicatesthedeviceorproductconformswithapplicableRussian
Gosstandarttechnicalandsafetystandards.
NRTLCertication
IndicatesthedeviceorproducthasmettheNationalRecognized
TestingLaboratoriescertication.
TheNRTLcerticationattaintedisaddedtothemarkoftheapplicable
testinglaboratory.TheexampledisplaystheNRTLcerticationwith
theMETLaboratoriesmark.
ChinaMetrologyCertication
IndicatesthedeviceorproductcomplieswithapplicableChina
MetrologyCerticationrequirements.
TÜVRheinland
Indicatesthedeviceorproductcomplieswithapplicabletechnical
andsafetyrequirementsfollowingtestingbyTechnischer
Überwachungs-Verein,(TechnicalInspectionsOrganization).
Training
Thismanualisintendedasasupplementto,notasubstitutefor,thoroughproduct
training.Ifyouhavenotreceivedtrainingontheuseofthesystem,youshouldrequest
trainingassistancefromGEHealthcare.
Toseeavailabletraining,gototheGEHealthcaretrainingWebsite
(www.gehealthcare.com/training
).SelectEducation>ProductEducation-Technical>
DiagnosticCardiology .
Formoreself-pacedcourseofferings,tools,andreferenceguidesyou
maynduseful,pleasevisittheGEHealthcareEducationStoreat
www.gehealthcare.com/educationstore.
EquipmentIdentication
EveryGEHealthcareproducthasaproductlabelthatidentiestheproductname,
partnumber,manufacturinginformation,anduniqueserialnumber.Thisinformation
isrequiredwhencontactingGEHealthcareforsupport.
32 MAC™2000 2053535-003M
ProductLabel
Introduction
Theproductlabelislaidoutinthefollowingformat.Dependingontheproduct,the
labelmayvaryslightlyinformat,butitcontainsthesameinformation.
ProductLabelFormat
SerialNumberFormat
Item
1
2
3
4
5
Description
Productdescription
Productpartnumber
Deviceserialnumber(See“SerialNumberFormat”onpage33formoreinformation.)
Productbarcode
DateofmanufactureinYYYY-MMformat
Eachdevicehasaserialnumberthatuniquelyidentiesitandprovidesimportant
information.Youneedtheproductcodeandtheentireserialnumberbeforeservicing
orrequestingsupportforyourproduct.Theserialnumberformatisshowninthe
followingillustration:
SerialNumberFormat
Item Name
1
2
3
4
2053535-003M MAC™2000
ProductCode Three-lettercodethatuniquelyidentiestheproductline.
YearManufactured Two-digitcodeidentifyingtheyearthedevicewas
FiscalWeek
Manufactured
ProductSequence Four-digitnumberidentifyingtheorderinwhichthis
Description
See“ProductCodes”onpage34formoreinformation.
manufactured.Valuesrangefrom00to99
Forexample:00=2000,04=2004,05=2005(andsoon).
Two-digitcodeidentifyingtheweekthedevicewas
manufactured.
Valuesrangefrom01to52.GEHealthcare'sscalweeks
correspondtothecalendarweek.Forexample,01=rst
weekinJanuary.
devicewasmanufactured.Valuesrangefrom0001to
9999.
33
Introduction
SerialNumberFormat(cont'd.)
Item Name
5
6
ManufacturingSite One-lettercodeidentifyingthesitewherethedevicewas
Miscellaneous
Characteristic
Description
manufactured.Forexample,F=Milwaukee,N=Freiburg,
P=Bangalore
Forexample,P=deviceisaprototype,R=devicewas
refurbished,U=devicewasupgradedtomeetthe
specicationsofanotherproductcode,A=deviceisin
production.
DeviceAddressLabelandRatingPlate
TheDeviceAddresslabelandRatingPlateislaidoutinthefollowingformat.
Dependingontheproduct,thelabelmayvaryslightlyinformat.
Item
1
2
3
4
5
Description
Productdescription
Countryoforigin
Symbols
See“SymbolDescriptions”onpage25foradescriptionofthesymbolsusedon
thislabel.
Electricalratingofthedevice
Manufacturernameandaddress
ProductCodes
34
Theproductcodeidentiesspecicsystemplatforms.
Youcanidentifytheproductcodeusingtheserialnumberlistedontheproductlabel
locatedinoneofthefollowingplaces:
•Ontheproductlabelattachedtothedevice.
•OntheproductlabelprovidedwiththeapplicationCD.
Forsoftwareapplicationsystems,youcanviewtheserialnumberbylaunchingthe
systemapplicationandclickingHelp>About.
Forinformationonlaunchingtheapplication,refertotheserviceoroperator'smanual
forthissystem.
MAC™2000 2053535-003M
ServiceInformation
ServiceRequirements
WarrantyInformation
Introduction
Thissectionprovidesinformationpertainingtothemaintenanceandservicingofthe
system.FamiliarizeyourselfwiththisinformationbeforerequestingservicefromGE
Healthcareoritsauthorizedrepresentatives.
ForsystemswithhardwareprovidedbyGEHealthcare,failureonthepartofthe
responsibleindividual,hospital,orinstitutionusingthisequipmenttoimplement
asatisfactorymaintenanceschedulemayresultinequipmentfailureandpossible
safetyhazards.
Forsoftwareonlyproducts,maintenanceofthehardwareandoperatingsystemon
whichthesoftwareresidesistheresponsibilityofthecustomer.
Regularmaintenance,irrespectiveofusage,isessentialtoensurethatthecomponents
ofthissystemarealwaysfunctionalwhenrequired.
ThisdeviceisconsideredGEHealthcare-suppliedhardware.OnlyauthorizedGE
Healthcareservicepersonnelshouldservicethedevice.Anyunauthorizedattemptto
repairequipmentunderwarrantyvoidsthatwarranty.Itistheuser'sresponsibilityto
reporttheneedforservicetoGEHealthcareortooneoftheirauthorizedagents.
AdditionalAssistance
GEHealthcaremaintainsatrainedstaffofapplicationandtechnicalexpertstoanswer
questionsandrespondtoissuesandproblemsthatmayariseduringtheinstallation,
maintenance,anduseofthissystem.
ContactyourlocalGEHealthcarerepresentativetorequestadditionalassistance.
ManualInformation
Thissectionprovidesinformationforthecorrectuseofthismanual.
Keepthismanualwiththeequipmentatalltimesandperiodicallyreviewit.You
shouldrequesttrainingassistancefromGEHealthcare,ifneeded.
ManualPurpose
Thismanualprovidestechnicalinformationforservicerepresentativesandtechnical
personnelsotheycanmaintaintheequipmenttotheassemblylevel.Itservesasa
guideformaintenanceandelectricalrepairsthatareconsideredeld-repairable.
Wherenecessary,themanualidentiesadditionalsourcesofrelevantinformation
andtechnicalassistance.
Seetheoperator’smanualofthissystemfortheinstructionsnecessarytooperatethe
equipmentsafelyinaccordancewithitsfunctionandintendeduse.
DocumentConventions
Thismanualusesthefollowingconventions.
2053535-003M MAC™2000
35
Introduction
TypographicalConventions
Convention Description
BoldText Indicateskeysonthekeyboard,texttoenter,orhardwareitemssuchas
buttonsorswitchesontheequipment.
Italicized-Bold
Text
CTRL+ESC
<space>
Enter
>
Indicatessoftwaretermsthatidentifymenuitems,buttonsoroptionsin
variouswindows.
Indicatesakeyboardoperation.Aplus(+)signbetweenthenamesoftwo
keysindicatesthatwhileholdingtherstkey,youshouldpressandrelease
thesecondkey.Forexample,PressCTRL+ESCmeanstopressandholdthe
CTRL keyandthenpressandreleasetheESCkey.
Indicatesthatyoumustpressthespacebar.Wheninstructionsaregiven
fortypingaprecisetextstringwithoneormorespaces,thepointwhere
youmustpressthespacebarisindicatedas<space>.Thisensuresthat
thecorrectnumberofspacesisinsertedinthecorrectpositionswithin
theliteraltextstring.Thepurposeofthe<>bracketsistodistinguishthe
commandfromtheliteraltextwithinthestring.
IndicatesthatyoumustpresstheEnterorReturnkeyonthekeyboard.Do
nottypeEnter.
Thegreaterthansymbol,orrightanglebracket,isaconcisemethodto
indicateasequenceofmenuselections.
Forexample,thestatement“Fromthemainmenu,selectSystem>Setup>
Options toopentheOptionActivationwindow”replacesthefollowing:
1. Fromthemainmenu,selectSystemtoopentheSystemmenu.
2. FromtheSystemmenu,selectSetuptoopentheSetupmenu.
3. FromtheSetupmenu,selectOptionstoopentheOptionActivation
window.
Illustrations
Allillustrationsinthemanualareprovidedasexamplesonly.Dependingonsystem
conguration,screensinthemanualmaydifferfromthescreensonyoursystem.
Allpatientnamesanddataarectitious.Anysimilaritytoactualpersonsis
coincidental.
Notes
Notesprovideapplicationtipsoradditionalinformationthat,whileuseful,arenot
essentialtothecorrectoperationofthesystem.Theyarecalledoutfromthebody
textthroughaagwordandindentation,asfollows:
NOTE:
ThetiporadditionalinformationisindentedbelowtheNOTEagword.
36 MAC™2000 2053535-003M
ProductOverview
Thischapterprovidesadescriptionoftheproduct,itsfeatures,andtherequirements
necessarytooperatethissystem.
ProductDescription
Thissystemprovidestwobasicmodesofoperation:
•RestingECG
Thismodeisthestandardmodeforyoursystem.
•Arrhythmia
Thismodeisprovidedfortheconvenienceofautomaticallygenerating
documentation.
Youcanupgradethebasicsystemwithtwoothermodesofoperation:
2
•Exercise
Thismodeisforexercisestresstesting.
•RRAnalysis
ThismodeisforRRintervalsanalysis.
Thebasicsystemprints6or12leadsofECGandprovidesoptionaltransmissionand
receptionofECGdatatoandfromacentralECGcardiovascularinformationsystem.
Youcanalsoupgradeitwithsoftwareoptionssuchas12-leadECGmeasurement
andinterpretiveanalysis.
HardwareOverview
Thissectionidentiesthekeycomponentsofthesystemhardware.Familiarize
yourselfwiththesecomponents,theirlocation,andtheirusebeforeattempting
tousetheequipment.
2053535-003M MAC™2000
37
ProductOverview
FrontView
FrontViewofDevice
ItemName
1
Display Presentswaveformandtextdata.
2
FunctionKeys
3
Keypad Usetoselectmenuoptionsonthescreen.
4
Printerdoorpushbutton Openstheprinterdoor.
5
Printer/Printerdoor
Description
Selectsmenuoptionsonthescreen.
Printsreports.
RearView
38 MAC™2000 2053535-003M
RearViewofDevice
ProductOverview
ItemName
1
SDcardslot ConnectionforSecureDigital(SD)card.This
2
LANorWLANconnection
3
USBports(2)
4
COMMAport
5
COMMBport
6
Phonejack RJ11connectorfromtheinternalmodemtoan
7
ACPowerCordconnection StandardconnectorfortheACpowercable.
Description
systemsupportsSDcardsformattedfortheFATor
FAT16lesystems.
RJ45networkconnectorusedtoconnectEthernet
cableforwiredLANcommunicationorLAN
communicationusingWirelessbridge.
StandardUniversalSerialBus(USB)connector
forUSBdevices,suchastheoptionalbarcode
reader,optionalUSBWiFiDongle,anexternal
non-multimediaUSBkeyboardoroptionalUSB
poweredWirelessbridge.
Serialconnectorfordatacommunicationwith
CASE/CardioSoftorMUSEsystems.
Serialconnectorforstressdevices(bicycle,
ergometer,ortreadmill).
analogphoneline.
SideView
SideViewofDevice
ItemName
1
KISSconnection
2
ECGPatientCableconnection D-sub15–pinfemaleconnectorfortheacquisition
2053535-003M MAC™2000 39
Description
ConnectionportfortheoptionalKISSPump
system.
cable.
ProductOverview
InternalView
InternalViewofDevice
Item
1
2
3
4
5
6
7
8
9 COMMAport
10 COMMBport
11
12
Description
Printer
Batterycompartment
PrintedWiringAssembly(PWA)
KISSconnectorport
ECGpatientcableconnectorport
SDCardport
RJ45LANconnectionport
USBports(2)
RJ11PhoneporttoInternalModem
PowerSupply
40 MAC™2000 2053535-003M
Keypads
StandardKeypod
ProductOverview
Item Name
1
2
3 PowerLED
4
5
6
7
8
Poweron/off
BatteryLED
Leadskey Scrollsthroughtheleadsandallowsyoutoselect
ECGkey Acquiresandprintsa12–leadECG.
Rhythmkey
WriterStop
Trimpad/CursorControlkeys
Description
Turnsthesystemonoroff.
Indicatesvariousbatterystates:
•Steadyamberindicatesthebatteryischarging
•Flashingamberindicatesthebatteryislow
•Nolightindicatesthebatteryisneithercharging
norlow
ONindicatesthedeviceispluggedinandreceiving
power.
thedisplayformatsfortheleadsequence.
Printsreal-timecontinuousrhythm.
Stopstheprintingfunction.
Providesmovementthroughmenusandwindows.
Fordescriptiononusingthetrimpadandcursor
controlkeys,referto“UsingtheTrimpad”inthe
operatormanual.
2053535-003M MAC™2000 41
ProductOverview
Item Name
9
10
11
12
13
14
15
16
Description
Functionkeys Usetoselectmenuoptionsonthescreen.
NOTE:
Thereisnomarkingonthekeypadforthe
functionkeys.Uptosixmenuoptionsmaybe
availableatanygiventime,andeachoption
correspondstoafunctionkeydirectlybelow
thedisplay.
Backspacekey Deletescharacters.
Enter keyUsetoadvancethefocusinawindowortoselect
itemsfromthescreen.
Altkey
Spacebar Entersaspaceinthetext.
Optionkey Usetoenterspecialcharactersonnon-English
ESC(escape)
Shiftkey Usetoenteracapitalletter.
Switchesbetweendifferentinputmethodsfor
JapaneseandKoreankeyboardlanguages.
Asasecondaryfunction,itmovesthroughthe
menulists.
keyboards.
Closesawindowonascreen.
Forexample,pressShift+ptotypeacapitalP.
StressKeypad
Thestresskeypadhasthesamekeysasastandardkeypadwiththeadditionofspecic
stresskeys.Ifyoudonothavethestressoption,youdonothaveastresskeypad.
42 MAC™2000 2053535-003M
ProductOverview
StressKeys
Item Name
Item Name
1
StressKeys
Item Name
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
Description
Description
Stresskeys Controlsstressequipmentconnectedtothe
system.
Description
Preteststresskey Selectsthepretestphaseoradvancestothenext
stagewithinthephase.
Exercisestresskey
Recoverystresskey Selectstherecoveryphaseoradvancestothenext
TestEndstresskey Selectsthetestendphase.
HoldStagestresskey Remainsatthecurrentstage.
SpeedW+stresskey
(Speed/Loadup)
Gradeá á ástresskey
(Gradeup)
Tmilstresskey
Tmilstresskey
Gradeâ â âstresskey
(Gradedown)
SpeedW- stresskey
(Speed/Loaddown)
EnterBPstresskey Allowsyoutoenterbloodpressurevaluesorstart
Commentstresskey Allowsyoutoenteracommentduringthestress
Mediansstresskey
12ldstresskey Printsa12–leadreport
Recallstresskey Printstheprevious10secondsofECG
Selectstheexercisephaseoradvancestothenext
stagewithinthephase
stagewithinthephase
Manuallyincreasesthetreadmillspeedor
ergometerload.
Increasestheelevationofthetreadmill.
Startsthetreadmillduringthetest
Stopsthetreadmillduringthetest.
Decreasestheelevationofthetreadmill.
Manuallydecreasesthetreadmillspeedor
ergometerload.
abloodpressuremeasurement.
test.
Printsamedianreportduringthetest.
2053535-003M MAC™2000
43
ProductOverview
ArchitectureandTheoryofOperation
Thissectiondescribestheblockdiagram,hardware/rmwarearchitecture,andthe
productinterfacesforthissystem.
BlockDiagram
Thefollowingdiagramillustratesthebasiclayoutofthissystem.
44
MAC™2000 2053535-003M
ProductOverview
2053535-003M MAC™2000
45
ProductOverview
Hardware/FirmwareArchitecture
Thissystemincludesthefollowingsubsystems:
•PWAmainboard
ThemainboardhasamainCPUinwhichtheapplicationsoftwareruns.Allofthe
peripherals,suchasthedisplay,internalmodem,LAN,serialports,USBports,and
SDcardareinterfacedandcontrolledbythisCPU.ItalsohasaProgrammableSoC
(PSoC)thatcontrolsthepowersection,thermalprinterinterface,motor,andkeypad.
Theon-boardDSPcontrolstheintegratedACQmodule.
NOTE:
InitialshipmentsofMAC2000deviceshadRevEPWAwithTMSeriesPSoC
Chip.RevFPWAshaveanLPSeriesPSoCChipandstartedshippingwithSerial
NumberSJQ14220591PAonward.
•Display
Thedisplayisa7–inchTFTdisplaywith800x480resolutionandinterfacedtothe
mainboardthroughtheLVDSinterfacecable.
•Keypad
Thissystemhasamatrixkeypadwith10x9interfaces,66normalkeys,16stress
keys,andonededicatedpowerkey.ItalsohasLEDindicatorsforbatterystatus
andACmainsstatus.
ProductInterfaces
•Thermalprinter
TheprinterhasanA4/Letterthermalprinthead,motor,papersensors,anddoor
sensor.
•Powersupply
ThepowerinletconnectorisconnectedwiththeAD/DCmodulethroughanEMI
lter.TheAC/DCisa65WpowersupplyacceptingawideinputACvoltageand
delivering24DCtothemainboard.
•KISSpump
AninternalKISSpumpispoweredbythemainboard.Thesuctionconnectoris
terminatednexttothepatientconnectorintheassembly.
Thissystemoffersthefollowinginterfacesforconnectingtoexternaldevicesfordata
communication,softwareupdates,andthecontrolofworkloaddevices:
•RS232port(2)
•COMMPortAconnectstotheMUSEorCardioSoftsystem
•COMMPortBconnectstoatreadmillandotherstressdevices
•RJ45port(1)
Connectstonetworksusinga100baseTEthernetconnectorthroughanexternal
medicalisolator.
•USBconnector(2)
ConnectstoUSB-capabledevices,suchasanoptionalbarcodereader,external
USBkeyboard,WiFiDongleoranUSBpoweredexternalWirelessBridge.
•SecureDigital(SD)cardslot
46 MAC™2000 2053535-003M
InterfaceswithaSecureDigitalcard,whichisusedforthefollowing:
•StoringECGs
•Flashingthedevicewithsoftwareupdates
•ConnectingmemoryorfutureI/Oextensions
•RJ11port(1)
Connectsaninternalmedicalgradeanalogmodem(optional)toaphoneline.
ProductOverview
2053535-003M MAC™2000
47
ProductOverview
48 MAC™2000 2053535-003M
InstallingtheUSBPoweredSilex
WirelessBridge
Thischapteronlydescribesthehardwareinstallation.Fordetailedguidanceon
congurationandsetup,refertoMobileLinkWirelessCommunicationInstallation
Manual(2020299-021) .
ThenewSilexwirelesskit(PN2095170-001)comeswithSilexwirelessmodule,
EthernetCable,USBPowerCableandDualLocks(4pcs).
3
2053535-003M MAC™2000 49
InstallingtheUSBPoweredSilexWirelessBridge
InstallingtheSilexWirelessBridge
1. Pick2piecesofDualLocksfromtheSilexWirelessBridgeKitandengagethe
hookandloopsurfaceoftheDualLocksagainsteachothertoformapair.
Followthesameproceduretomakeonemorepair.
2. Fromeachpair,removetheadhesivelinerfromonesideoftheduallockand
stickeachpairtotheatbottomsurface(eithersideofthelabel)ofSilex
WirelessBridgemodule.
3. PeeltheadhesivelinerfromtheDualLockstuckontheSilexmodule(oneither
sideofthelabel)asshowninstep2.CarefullypositiontheSilexModuleonthe
50 MAC™2000 2053535-003M
InstallingtheUSBPoweredSilexWirelessBridge
MAC2000unitbyplacingitbehindtheDisplayinsuchawaythattheLANport
andPowerswitcharevisiblefromtherearoftheMAC2000device.
NOTE:
CareshouldbetakentopositiontheSilexwirelessmoduleapproximatelyon
thetangentofthecontouredgeofMAC2000topcover.
Ensurethat,thereisnopinchingoftheUSBPowercablebytheDisplay
housingwhileroutingthecable.
2053535-003M MAC™2000 51
InstallingtheUSBPoweredSilexWirelessBridge
4. RouteandconnecttheEthernetandUSBPowercablesfromSilexmodule
totheMAC2000unit.
5. ContinuewiththeappropriatefunctionalcheckoutprocedureforthisFRU.
See“FunctionalCheckout”onpage188formoreinformation.
6. Toverifytheinstallation,
a. TurnontheMAC2000ECGanalysissystem
b. PressFileManager.
c. SelecttheECGyouwouldliketotransmit.
NOTE:
IfnoECGisavailablefortheverication,insertanSDcardwithan
ECGstoredonitortakeaatlineECGwithnopatientIDnumberand
storeittotheSDcard.
d. VerifythattheECGtransmitssuccessfullyandcanbefoundintheMUSE
Edit list.
FormoreinformationonhowtotransmitanECGfromthecart,referto
theMAC2000ECGAnalysisSystemOperator’sManual(2053535–002).
52 MAC™2000 2053535-003M
InstallingtheEmbeddedWireless
Module
Thischapterprovidesinstructionsoninstallingthehardwarefortheembedded
wirelessmodule.See“CommunicationSetup”onpage252fordetailedguidanceon
congurationandsetup.
Theembeddedwirelessmoduleupgradekit(PN2072532-116)shipswiththefollowing:
4
Item
1
2
3 Antenna
4
5 Foam
2053535-003M MAC™2000
Description
MainBoard
Clips
EmbeddedWirelessModule
53
InstallingtheEmbeddedWirelessModule
InstallingtheEmbeddedWirelessModule
Performthefollowingstepstoinstalltheembeddedwirelessmodule.
1.“PreparingtheSystem”
2.“RemovingtheBattery”
3.“RemovingtheTopCoverAssembly”
4.“RemovingtheDisplayBottomCover”
5.“InstallingtheAntenna”
6.“RemovingtheMainBoard”
7.“ReplacingtheMainBoard”
8.“ReattachingtheTopCoverAssembly”
9.“InstallingtheBattery”
NOTE:
Takestrictprecautionsagainstelectrostaticdischargedamagewhilereplacing
eldreplaceableunits.
PreparingtheSystem
1. ArchivestoredECGs.
Ifthedevicehastheinternalstorageoption,transmitanyECGsremainingin
storagetoyourarchivalsystemand/orprintthemtoensureyouhaveaprinted
recordbeforeyoureplacethemainboard.RefertotheMAC2000operator's
manualforinstructionsontransmittingandprintingECGs.
2. Storeandprintyourdevicesettingssothatyoucanrestorethemormanually
recongurethesystemwhenyouaredone.
Forinformationonstoringyoursettings,see“SavingSystemConguration
Settings”onpage178.
Forinformationonprintingyoursettings,see“PrintingtheSystemSetupReport”
onpage178.
3. Powerdownthesystem.
4. DisconnectthedevicefromtheACwalloutlet.
5. Disconnectthepowercordfromtherearofthedevice.
6. Disconnectthepatientcablefromthedevice.
7. Disconnectallcablesfromthebackpanel.
8. Removeanypaperfromthepapercompartment.
RefertotheMAC2000operator'smanualformoreinformationonremovingthe
paperfromthepapercompartment.
9. Removethebatterybyperformingtheprocedurein“RemovingtheBattery”
onpage55.
54
MAC™2000 2053535-003M
RemovingtheBattery
1. Gentlyturnthedeviceoveranduseacross-headscrewdrivertoremovethe
screwholdingthebatterycover.
2. Pushthetabtoremovethebatterycompartmentcover .
InstallingtheEmbeddedWirelessModule
3. Gentlyliftthebatterycompartmentcover .
4. Removethebattery.
5. Removethetopcoverassemblybyperformingtheprocedurein“Removing
theTopCoverAssembly”onpage56.
2053535-003M MAC™2000
55
InstallingtheEmbeddedWirelessModule
RemovingtheTopCoverAssembly
1. Usingacross-headscrewdriver,removethefourscrewsonthebottomcover.
2. Gentlyturnthedevicebackoversothetopcoverisfacingup.
3. Opentheprinterdoorandlocatethelatchesatthetopleftandtoprightcorners
oftheprinter.
4. Usingaat-headscrewdriver ,gentlydisengagethetopcoverfromthelatches.
5. Gentlyliftthetopcoverpart-wayupanddetachthedisplaycable(1)and
keypadribbon(2).
56 MAC™2000 2053535-003M
InstallingtheEmbeddedWirelessModule
6. Liftthetopcoveroffcompletely.
Thetopcoverassembly,withthedisplayassemblymountedonit,isnow
completelyseparatedfromthedevice.
7. Removethedisplaybottomcoverbyperformingtheprocedurein“Removing
theDisplayBottomCover”onpage58.
2053535-003M MAC™2000
57
InstallingtheEmbeddedWirelessModule
RemovingtheDisplayBottomCover
1. Turnoverthetopcoverassemblyandremovethesixfastenersusinga
cross-headscrewdriver.
2. Turnoverthetopcoverassembly,slidethedisplayassemblytowardthefront
ofthetopcoverassembly,andliftthedisplayassembly,carefullyguidingthe
displaycableandkeypadribbonoutoftheslotsonthetopcover .
3. Turnthedisplayoversothebottomdisplaycoverisfacingyou.
58 MAC™2000 2053535-003M
InstallingtheEmbeddedWirelessModule
4. Usingacross-headscrewdriver,removethetenscrewsfromthedisplaybottom
cover.
5. Removethedisplaybottomcover,carefullythreadingthedisplaycableand
keypadribbonthroughtheopenings.
Thebottomcoverisnowremovedfromthedisplayassembly.
6. Installtheantennabyperformingtheprocedurein“InstallingtheAntenna”
onpage60.
2053535-003M MAC™2000 59
InstallingtheEmbeddedWirelessModule
InstallingtheAntenna
1. Takeouttheantennafromtheupgradekit,andremovetheadhesivelinerfrom
2. Turnoverthebottomcoverofthedisplayassemblyandadheretheantennato
theantenna.
thebottomcover.
3. Threadtheantennacablethroughtheopening.
4. Replacethedisplaybottomcover,carefullythreadingthekeypadribbonand
displaycablethroughtheopenings.
60 MAC™2000 2053535-003M
InstallingtheEmbeddedWirelessModule
5. Usingacross-headscrewdriver,securethedisplaybottomcoverwiththeten
screws.
6. Alignthetabsonthedisplaybottomcoverwiththeslotsonthedevicetopcover.
7. Lowerthedisplaybottomcovertothedevicetopcoverwhilecarefullythreading
thedisplaycable,keypadribbon,andantennathroughtheslotsonthetopcover.
2053535-003M MAC™2000 61
InstallingtheEmbeddedWirelessModule
8. Slidethedisplayassemblytowardthebackofthetopcovertoreattachitto
9. Turnoverthetopcoverassemblyanduseacross-headscrewdrivertosecure
thetopcover.
thedisplayassemblytothetopcoverwiththesixscrews.
10. Takeoutthefoamandclipsfromtheupgradekit,andremovetheadhesive
liners.
11. Adherethefoamandclipsonthetopcoverasshowninthefollowing
photograph.
12. Removethemainboardbyperformingtheprocedurein“RemovingtheMain
Board”onpage63.
62 MAC™2000 2053535-003M
RemovingtheMainBoard
1. Disconnectallribbonsandcablesfromthemainboard.
2. Removetherearconnectorpanel.
InstallingtheEmbeddedWirelessModule
3. Usingacross-headscrewdriver,removethesixscrewsholdingthemainboard
inplace.
2053535-003M MAC™2000 63
InstallingtheEmbeddedWirelessModule
4. Usingacross-headscrewdriver,removethetwoM3x4CSKscrewssecuringthe
5. Usingacross-headscrewdriver,removethepatientcableconnectorfromthe
patientcableconnectortothemainboard.
mainboard.
6. Removethemainboardfromthechassis.
7. Replacethemainboardbyperformingtheprocedurein“ReplacingtheMain
Board”onpage65.
64 MAC™2000 2053535-003M
ReplacingtheMainBoard
1. Insertthenewmainboardintoplaceonthechassis
2. Connectthepatientcableconnectortothemainboardanduseacross-head
screwdrivertosecureitwithtwoM3x4CSKscrews.
InstallingtheEmbeddedWirelessModule
3. Securethemainboardtothechassisbyinsertingthesixscrewsandtightening
themwithacross-headscrewdriver.
2053535-003M MAC™2000 65
InstallingtheEmbeddedWirelessModule
4. Snapthebackpanelintoplaceinthebottomcover .
5. Reconnectallribbonsandcablestothemainboard.
6. Reattachthetopcoverbyperformingtheprocedurein“ReattachingtheTop
CoverAssembly”onpage66.
ReattachingtheTopCoverAssembly
1. Alignthetopcovertothedevice.
2. Partiallylowerthetopcover.
66 MAC™2000 2053535-003M
InstallingtheEmbeddedWirelessModule
3. Connecttheantennalocatedontheinsideofthetopcover.
4. Threadtheantennathroughtheclips.
5. Connectthedisplaycable(1)andkeypadribbon(2)tothemainboard.
2053535-003M MAC™2000 67
InstallingtheEmbeddedWirelessModule
6. Lineupthetopcoverandgentlylowerituntilthelatchessnapintoplaceat
7. Turnthedeviceoversothebottomcoverisfacingupwardanduseacross-head
thetopleftandtoprightcornersoftheprintercompartment,justbelowthe
topcover.
screwdrivertotightenthevescrewssecuringthebottomcover.
8. Installthebatterybyperformingtheprocedurein“InstallingtheBattery”on
page69.
68 MAC™2000 2053535-003M
InstallingtheBattery
1. Placethebatteryinthecompartmentandpushuntilitclicksintoplace.
2. Replacethecoveronthebatterycompartment.
InstallingtheEmbeddedWirelessModule
3. Lowerthebatterycoveruntilitsnapsintoplace.
4. Usingacross-headscrewdriver,tightenthescrewsecuringthebatterycover.
2053535-003M MAC™2000 69
InstallingtheEmbeddedWirelessModule
70 MAC™2000 2053535-003M
ServiceSetup
Thediagnostictestsverifythatthesystemoperatesproperly.Runthediagnostictests
tochecktheoperationofthedisplayscreen,speaker,keypad,thermalwriter ,battery,
andcommunications.Thesediagnostictestsareusefultoolsfortroubleshooting
problems,andcanbeusefulaspartofsystemcheckoutprocedures.
AccessingServiceSetup
1. PoweronthesystembypressingthePowerbutton.
2. OntheMainMenu,pressSystemConguration.
3. PressMore>More>ServiceSetup.
AwindowopenspromptingyoutoentertheServicepassword.
5
4. Type prodandpressOK.
NOTE:
Ifthekeypaddoesnotincludethelettersprod,type7763andpressOK.
2053535-003M MAC™2000 71
ServiceSetup
TheServiceSetupmenuopens.
5. UsetheTrimpadtohighlightthebuttonyouwantandpressEnter.
AccessingDeviceSettings
1. OntheServiceSetupmenu,selectDeviceSettingsandpressEnter.
2. IntheSerialNumbereld,typetheserialnumberofthedevice.
3. SelecttheappropriatelanguagefromtheKeyboardLanguagelist.
4. PressSave.
EventLog
Eventloggingisawaytotracksystemerrors,warnings,andinformationalmessages.
Thefollowingwindowopens:
72 MAC™2000 2053535-003M
SettingUpEventLogging
ServiceSetup
1. OntheServiceSetupmenu,highlightEventLogandpressEnter.
Thefollowingwindowopens:
2. Enableordisableeventlogging.
•Toenableeventlogging,selecttheKeyEventLoggingcheckbox.
or
•Todisableeventlogging,deselecttheKeyEventLoggingcheckbox.
3. SelectalevelofseveritytologfromtheLogLevellist:
LogLevelSeverity
LogLevel WhatisloggedtotheEventLog
Error
Warning
InformationLogserrors,warningsandinformation
4. PressSavetosaveyoursettings.
Logsonlyerrors.
Logserrorsandwarnings.
messages.
2053535-003M MAC™2000
73
ServiceSetup
ExportingtheEventLog
Onceyouhavecapturedtheeventlog,youneedtoexportittoreviewthelogles.
1. InsertanSDcardintheSDcardslotinthebackpanel,asshowninthefollowing
illustration:
AccessingSystemDiagnostics
Thegoldcontactsareface-down.
2. OntheServiceSetupmenu,highlightEventLogandpressEnter.
3. PressExportLogFiles.
ThecurrentEventLogle,log_0.log,iscopiedtoalogdirectory
ontheSDcard.
NOTE:
Toaccessthelogle,inserttheSDcardintoanSDcardreaderthatis
connectedtoacomputerwithaWindowsoperatingsystemandatext
editorsuchasNotepadorWordPad.IfGEHealthcaretechnicalservice
requeststheEventLogfortroubleshootinganissue,sendtheleasan
emailattachment.
UsetheSystemDiagnosticsmenutoperformfunctionaldiagnostictests.Usethe
followingproceduretoaccesstheSystemDiagnosticsmenu.
UsetheTrimpadtohighlighttheSystemDiagnosticsbuttonandpressEnter.
74
MAC™2000 2053535-003M
TheDIAGNOSTICTESTSwindowopens.
ServiceSetup
TestingtheDisplay
Thefollowingsectionsdescribehowtoperformthespecicdiagnostictests.Proceed
totheappropriatesectionforthetestyouneedtoperform.
UsetheDisplayTesttodetermineifthedisplaypixelsareworkingproperly.
1. OpentheDIAGNOSTICTESTSwindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. SelectDisplayTest.
TheStartTestwindowopens.
2053535-003M MAC™2000
75
ServiceSetup
3. SelectStartTest.
Thefollowingwindowopens.
4. PresstherightarrowontheTrimpadrepeatedlytomovethecolorbars
horizontallyacrossthescreen.
5. Verifythatthecolorbandpattern(red,green,blue,white)scrollsacrossthe
screen.
Passthetestifthepatternisreplicatedwithoutdiscoloration.
6. Presstherstfunctionkeyonthelefttoswitchtohorizontalcolorbars.
7. PressthedownarrowontheTrimpadrepeatedly.
8. Verifythatthecolorbandpattern(red,green,blue,white)scrollsdownthe
screen.
Passthetestifthepatternisreplicatedwithoutdiscoloration.
9. Presstherstfunctionkeyagaintocyclethroughthesolidcolorpane(red,
green,blue,white).
10. Foreachpane,checkforblackpixels.
Passthetestifnomorethan4blackpixelsareobservedonanysinglecolor
pane.
NOTE:
Ablackpixelobservedononepanewillprobablybeobservedoneverypane.
11. PressEscorEnterwhenthetestisdone.
Thefollowingwindowopens.
76 MAC™2000 2053535-003M
12. PressEscorEnterwhenthetestisdone.
Thefollowingwindowopens.
13. Selectpassorfail:
•Ifthetestpassed,pressYes.
•Ifthetestfailed,pressNo.
a. Checktheconnection(cable)betweenthedisplayandPWAtoensureit
isconnectedproperlyandintact.
b. Runthedisplaytestagain.
c. Ifthedisplaytestfailsagain,replacethedisplayassemblyasdescribed
in“DisplayAssembly”onpage155.
ServiceSetup
TestingtheSpeaker
UsetheSpeakerTesttodetermineifthespeakerisworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. SelectSpeakerTest.
3. Listenforabriefaudibletonecomingfromthespeaker.
Thefollowingwindowopens.
4. Selectpassorfail:
•Ifyouheardanaudibletone,pressYes.
•Ifyoudidnothearanaudibletone,pressNo.
Ifthespeakertestfailed,replacethemainPWAassemblyasdescribedin
“PWA”onpage177.
2053535-003M MAC™2000
77
ServiceSetup
TestingtheKeypad
UsetheKeyboardTesttodetermineifthekeypadisworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. SelectKeyboardTest.
Thefollowingwindowopens.
3. Presseachkeyonthekeypadandverifythattwoasterisks(**)appearinthe
correspondingrepresentationofthatkeyonthescreen.
Akeypassesthetestifanasteriskappearsonthescreenwhenthe
correspondingkeyispressed.
4. Totestforstickykeys,continuetopresskeysthatalreadyhaveanasteriskon
thescreenandverifythatthescreenrepresentationofthekeyishighlighted
witheachsubsequentkeypress.
Akeypassesifthekeyonthescreenhighlightswitheachrepeatedkeypress.
5. Whenthetestisdone,pressShift+F6.
Thefollowingwindowopens.
6. Selectpassorfail:
•Ifeverykeypassesthetests,pressYes.
•Ifanykeyfailsthetests,pressNo.
Ifthekeyboardtestfailed:
a. Replacethekeypadassemblyasdescribedin“ReplacingtheKeyboard
Assembly”onpage166.
b. Runthetestagain.
•Ifthekeyboardteststillfails,replacethePWAasdescribedin“Replacing
thePWA”onpage181.
78 MAC™2000 2053535-003M
TestingtheAcquisitionModule
UsetheAcquisitionModuleTesttodetermineiftheacquisitionboardisworking
properly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. SelectAcquisitionModuleTest.
Awindowsimilartotheoneshowninthefollowingscreenopens.
NOTE:
Theversiondisplayeddependsonthesoftwareinstalled.
ServiceSetup
3. NotethetestresultandpressCancel.
4. IftheresultoftheAcquisitionModuleTestwasFailed,replacethemainPWA
asdescribedin“PWA”onpage177.
CheckingtheBatteryStatus
UsetheBatteryTesttodeterminethestatusoftheLithium-Ionbattery.Youmust
performthistestwhilerunningonbatterypower.
1. DisconnectthedevicefromACpowerandwaitatleastoneminute.
2. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
3. SelectBatteryTest.
Awindowsimilartotheoneshowninthefollowingscreenopens.
2053535-003M MAC™2000 79
ServiceSetup
Status
BatteryLifeRemainingTheestimatedtimeremainingbasedonthe
BatteryChargeRemainingAccordingtothebatteryspecicationsyou
Description
presentcurrentdraw.
Becausenoprintingisoccurringduringthis
statuscheck,thistimeiscalculatedbasedon
usingthesystemfordisplaypurposesonly.
Actualremainingbatterylifeislesswhen
usingthesystemforbothdisplayandprinting
purposes.
shouldsee3:00(threehours)orlessofBattery
LifeRemaining(displaytime)with100%Battery
ChargeRemaining.
Ifthenumbersdisplayedinthiswindoware
less,youmayrequiremorefrequentbattery
chargingoryoumaywishtoreplacethebattery.
See“ReplacingtheBattery”onpage110for
informationonhowtoreplacethebattery.
NOTE:
•Thissystemisinternallypoweredwithlithiumionbatterieswithdifferent
levelsofsafetyprotection.ItisnotrecommendedtochargeLithiumion
batteriesabove45°C(113°F).Ifthedeviceisoperatingatahighambient
temperature,andthebatterytemperatureexceedstherecommended
operatingconditions,chargingstopsautomatically.Undersuch
conditions,turnoffthedeviceandletitcoolbeforerestartingcharging.
4. NotethebatterystatusinformationandpressCanceltoclosetheBATTERY
STATUS window.
TestingtheWriter
UsetheWriterTesttodetermineifthewriterisworkingproperly.
NOTE:
BeforeperformingtheWriterTest,besurethatthermalpaperisloadedproperly.
Refertoyoursystem’sOperatorManualforinstructionsonloadingpaper.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. SelectWriterTest.
Thefollowingwindowopens.
•Youcanextendthebatterylifebyreducingthebrightnessonthedisplay.
•Todecreasethebrightness,pressShift+CursorDown.
•Toincreasethebrightness,pressShift+CursorUp.
80 MAC™2000 2053535-003M
3. Performthe50mm/sSpeedTest.
a. Select50mm/sSpeedTest.
Thewriterprintsthe50mm/sspeedtestreport.
ServiceSetup
b. Whenonepageofthereporthasprinted,pressStop.
Thefollowingwindowopens.
c. Examinetheprintedreport.
Usethefollowingcriteriatodetermineifthewriterpassedorfailedthe
50mm/sspeedtest.
•Ifonecycleofthesquarewavespans50mmonpaper,measuredfrom
cornertocornerofthewave,withanallowabletoleranceof1.0mm,
thetestpasses.
•Ifthiscriteriaisnotmet,thetestfails.
d. Ifthetestpassed,pressYes.
Ifthetestfailed,pressNo.
4. Repeatthepreviousstepfortheotherspeedtests.Thepass-failcriteriaforeach
oftheremainingtestsareasfollows:
•25mm/sSpeedTest
Ifonecycleofthesquarewavespans25mmonpaper ,measuredfrom
cornertocornerofthewave,withanallowabletoleranceof0.5mm(2%),
thetestpasses.
Ifthiscriteriaisnotmet,thetestfails.
•12.5mm/sSpeedTest
Measureveormorecyclesanddivideyourmeasurementbythenumber
ofcyclesmeasured.
2053535-003M MAC™2000 81
ServiceSetup
Ifonecycleofthesquarewavespans12.5mmonpaper,measuredfrom
cornertocornerofthewave,withanallowabletoleranceof0.5±1.25mm
(10%),thetestpasses.
Ifthiscriteriaisnotmet,thetestfails.
•5mm/sSpeedTest
Measureveormorecyclesanddivideyourmeasurementbythenumber
ofcyclesmeasured.Ifonecycleofthesquarewavespans5mmonpaper,
measuredfromcornertocornerofthewave,withanallowabletoleranceof
0.25±0.5mm(10%),thetestpasses.
Ifthiscriteriaisnotmet,thetestfails.
5. PerformthePrintHeadTest.
a. SelectPrintHeadTest.
Thewriterprintsa1–pageprintheadtestreport.
b. Verifythattherearenogapsinanyofthelinesprinted.
Upto5mmofblankpaperisallowableatthetopandbottomofthepage.
Whenthepageisdoneprinting,thefollowingwindowopens:
•Iftherearenogapsinthelinesontheprintedreport,pressYes.
•Iftherearegapsinthelinesontheprintedreport,pressNo.
6. Whenallwritertestsarecompleted,pressCanceltoclosethewindow.
7. Ifoneormoreofthewritertestsfail,usethefollowingtabletodeterminehowto
proceed.
StepsforWriterTestFailure
ObservedFailure Remedy
Improperpaperqueuingor
PaperJammessage
Missingdots(gapsinprinted
report)
Faultypapersensor
Replacethepapersensor.
See“PaperSensor”onpage142.
Faultyprinthead.
1. Cleantheprinthead.
See“CleaningthePrinthead”onpage144.
2. Replacetheprinthead.
Referto“ReplacingthePrintheadonPrinter
ModuleAssembly”onpage146.
82 MAC™2000 2053535-003M
StepsforWriterTestFailure(cont'd.)
ObservedFailure Remedy
ServiceSetup
Fadedprints
Incorrectprintspeedormotor
jerk
Writerpaperdoesnotmove
Faultyprinthead
Cleantheprinthead.
See“CleaningthePrinthead”onpage144.
Faultyroller
Replacetherollerintheprinterdoorassembly.
See“PrinterDoor”onpage147.
Printheadpressureproblem
Replacethespringsandprintheadholder.
See“Printhead”onpage143.
Faultysteppermotor .
Replacetherollerintheprinterdoorassembly.
See“PrinterDoor”onpage147.
FaultyPWA
ReplacethemainPWA.
Referto“PWA”onpage177.
Faultysteppermotor
Replacetherollerintheprinterdoorassembly.
Referto“PrinterDoor”onpage147.
Faultyprinterassembly
Replacetheprinterassembly.
Referto“PrinterModuleAssembly”onpage138.
FaultyPWA
ReplacethemainPWA.
Referto“PWA”onpage177.
Papermovesbutlinesdonot
printor
Printerinternalerrormessage
Faultycablesinprinterassembly.
1. Checkforpropercableassemblybetweenthe
printerandPWA.
2. Replacetheprinterassembly.
See“PrinterModuleAssembly”onpage138.
Faultyprinthead.
Replacetheprinthead.
Referto“Printhead”onpage143.
CommunicationsTests
Thereareseveraldifferentcommunicationsteststhatyoumayneedtoperform:
•RS232port
•LANoption(Network)
2053535-003M MAC™2000 83
ServiceSetup
•Modem(internal)
•USBport
Usetheproceduresinthefollowingsectionstoconductthenecessarytests.
TestingtheCOM(RS232)Ports
UsetheRS232TesttodetermineiftheCOMportsareworkingproperly.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. Useapapercliptoshortpins2and3onCOMA.
3. SelectRS232Test.
Thefollowingwindowopens:
4. PerformtheCOMPortLoopBackTestonCOMA.
a. SelectCOMAandpressEnter.
TheresultsoftheCOMPortLoopBackTestaredisplayed.
b. NotetheresultsoftheCOMPortLoopBackTest.
5. PerformtheCOMPortLoopBackTestonCOMB.
a. Usethepapercliptoshortpins2and3ontheCOMBconnector.
b. SelectCOMB,pressEnter.
TheresultsoftheCOMLoopBackTestforCOMBaredisplayed.
c. NotetheresultsoftheCOMLoopBackTestforCOMB.
6. Whenthetestisdone,pressEscorCanceltoclosetheresultswindow.
Ifeithertestfailed,replacethemainPWAmoduleasdescribedin“PWA”onpage177.
TestingtheLANOption
UsetheLANTesttotestwiredanddonglewirelessnetworkconnectivity.
1. Connectthedevicetoalocalareanetwork.
InthecaseofawiredLAN,ensurethatitisconnectedtoanactiveconnection.If
youconnecttoaninactivenetworktap,thetestresultmaybeafalsenegative.
84 MAC™2000 2053535-003M
ServiceSetup
InthecaseofdonglewirelessLAN,ensurethatthewirelessLANsettingsare
conguredasexplainedin“CommunicationSetup”onpage252.
2. UsetheTrimpadtohighlighttheSystemDiagnosticsbuttonandpressEnterto
opentheDiagnosticTestswindow
3. SelectLANTest.
Thefollowingwindowopens:
4. PressEntertostartTestnetworkconnectivity.
TestingtheWLANOption
Thefollowingmessageisdisplayed:Checkingconnectivity.Pleasewait.
•IfthemessageSystemConnectedtoNetworkisdisplayedinthewindow,
thetestpasses.
•IfthemessageNetworkUnavailableisdisplayedinthewindow,andyouare
surethesystemisconnectedtoanactivenetwork,thetestfails.
5. Whenthetestisdone,pressEscorCanceltoclosetheresultswindow.
UsetheWLANTesttotesttheembeddedwirelessnetworkconnectivity.
1. Connectthedevicetoalocalareanetwork.
EnsurethattheinternalwirelessLANsettingsareconguredasexplainedin
“CommunicationSetup”onpage252.
2. UsetheTrimpadtohighlighttheSystemDiagnosticsbuttonandpressEnterto
opentheDiagnosticTestswindow
3. SelectWLANTest.
Thefollowingwindowopens:
2053535-003M MAC™2000 85
ServiceSetup
4. ConguretheWLANDiagnosticwindowasfollows:
Field
LogLevelSettheloglevelfromlowtohightoincreasethelevelsofverbosity.
LogFilePathDenesthelocationwherethedriverlogwillbestored.
EnterIP
Address
Ping
Description
Optionsare:
•Off(Default)
•Low
•Medium
•High
MakesureyouhaveanSDcardinsertedintothedevice.
Thisisdisplay-onlyeld.Thedriverlogcanonlybestoredin
\SDMemory .
EntertheIPv4address.
NOTE:
GatewayIPaddressisrecommended.
ClicktopingtheIPaddress.
Theresultswillbedisplayedunderthetextboxandwillbeprinted
ontheservicereport.
Ifanerrormessagedisplays,refertothefollowingtable.
Message
TheIPtotestisinvalidTheIPaddressformatis
Nowirelessmodule
detected
86 MAC™2000 2053535-003M
Reason
incorrect.
Noembeddedwireless
moduleordonglewireless
moduleisavailableforthe
device
Solution
EnterthecorrectIP
address.
Embeddedwireless
moduleanddongle
wirelessmodulecan
beorderedseparately.
ContactyourGE
Healthcareservice
representativefor
information.
ServiceSetup
Message
Wirelessnetworkisnot
enabled
Wirelessnetworkis
disconnectedordeviceIP
isnotvalid
WiredLANisconnected,
pleasedisconnectwired
LANbeforetest
Pingtestfailed,no
receivedpackets
Reason
ThewirelessLANisnot
enabled.
Networkisnotconnected
orthereisnovalidIPfor
thedevice.
Thedeviceisconnectedto
thewiredLAN.
Youcanonlyconnect
eithertothewiredLANor
tothewirelessLAN.
IPiscorrectandvalid,but
theIPaddressmaynot
beinthesamesub-net
withthedevice,sothe
destinationIPisnot
reachable.
Solution
Enablethe Wireless
LANinWireless
NetworkingSettings
inCommunicationSetup.
See“Communication
Setup”onpage252for
detailedinformation.
ChecktheWireless
NetworkingSettings
inCommunicationSetup.
ConnectthedevicetoAP.
See“Communication
Setup”onpage252for
detailedinformation.
DisconnectthewiredLAN.
PingtheGatewayIP
address.
See“Communication
Setup”onpage252for
detailedinformation.
TestingtheInternalModem
SSIDisnotdetectable CannotdetecttheSSID.ConrmthecorrectSSID.
Conrmthewirelessuser
nameandpasswordifthe
SSIDisinvisible.
Incorrectcredentials Thewirelessusername
orpasswordortheEAP
certicateisincorrect.
Conrmthewirelessuser
nameandpassword.
Conrmthecerticateon
EAPCerticateSetting
inCommunicationSetup.
See“Communication
Setup”onpage252for
detailedinformation.
5. Whenthetestisdone,pressSaveorCanceltoclosetheresultswindow.
UsetheModemTesttotesttheinternalmodem.
1. Connectthedevicetoanactiveanalogphoneline.
Ensurethatthephonelineisactive.Ifyouconnecttoaninactivephoneline,the
testresultmaybeafalsenegative.
2. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2053535-003M MAC™2000 87
ServiceSetup
3. SelectModemTest.
Thefollowingwindowopens:
4. SelectInternalModemTest,pressEnter.
Thefollowingmessageisdisplayedinthewindow:TestinProgress.Please
wait.
Thentheresultsofthetestaredisplayed.
•Thetestpassesifthefollowingmessageisdisplayedinthewindow:Passed.
•Thetestfailsifthefollowingmessageisdisplayedinthewindow:Failed.
IftheInternalModemTestfails,dothefollowing:
a. Replacetheinternalmodemasdescribedin“ReplacingtheInternal
ModemonthePWA”onpage171.
5. Whenthetestisdone,pressCanceltoclosetheresultswindow.
TestingtheUSBPort
UsetheUSBTesttotesttheUSBport.
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. ConnectaUSBkeyboardtotheUSBportontherearpanelofthedevice.
NOTE:
3. SelectUSBTest.
Thefollowingwindowopens:
b. ReruntheInternalModemTest.
c. Iftheteststillfails,replacethePWAasdescribedin“PWA”onpage177
,andenablethemodemoption.
TheUSBkeyboardusedforthistestmustmatchthelanguagethatis
selectedinsetup.
88 MAC™2000 2053535-003M
4. PressanykeyontheUSBkeyboardandverifypassorfail:
•IfthecharacterthatappearsintheCharacterInputeldmatchesthekey
youpressed,thetestpassed.
•Ifthecharacterdoesnotmatchthekeyyoupressed,ornocharacterappears
intheCharacterInputeld,thetestfailed.
5. Whenthetestisdone,pressEscorCancel.
Thefollowingwindowopens:
6. Dooneofthefollowing,dependingontheresultsofthetest:
•Ifthetestpassed,pressYes
•Ifthetestfailed,pressNo.
ReplacethemainPWAasdescribedin“PWA”onpage177.
ServiceSetup
TestingthePatientLeadwires
1. OpentheDiagnosticTestswindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. Connectapatientcablewithleadwirestothedevice’spatientcableconnector.
3. Connectallleadstoapatientsimulatororshortingbar .
4. SelectPatientLeadWireCheck.
Thefollowingwindowopens:
2053535-003M MAC™2000 89
ServiceSetup
5. PressEntertoselectStartTest.
Thetestresultsaredisplayedforeachleadwire.
•Ifthefollowingmessageisdisplayed,theleadwirepassedthetest:
Connected .
•Ifthefollowingmessageisdisplayed,theleadwirefailedthetest:
Disconnected .
6. Whenthetestisdone,pressEscorCancel.
7. Replaceeveryleadwirethatfailedthetest.
8. Repeatthetest.
Iftheleadwirestillfailsthetest,replacethemainPWAasdescribedin“PWA”
onpage177.
PrintingaServiceReport
Usethefollowingproceduretoprintaservicereportaftercompletinganyofthe
diagnostictests.
1. OpentheDIAGNOSTICTESTSwindowasdescribedin“AccessingSystem
Diagnostics”onpage74.
2. SelectServiceReportandpressEnter.
UpdatingtheSoftware
SoftwareupdatesareprovidedonanSDcard.Performasoftwareupdateas
describedinthissection.
NOTE:
TheACpowermustbeconnectedwhileyouareupdatingthesoftwarethrough
theSDcard.
90 MAC™2000 2053535-003M
AccessingtheSoftwareUpdate
1. InserttheSDcardwiththesoftwareupdateintheSDcardslotintherearpanel
asshowninthefollowingphotograph.
Thegoldcontactsareface-down.
ServiceSetup
2. OntheMainMenu,pressSystemConguration.
3. PressMore>More>ServiceSetup.
AwindowopenspromptingyoutoentertheServicepassword.
4. TypeprodandpressOK.
NOTE:
Ifthekeypaddoesnotincludethelettersprod,type7763andpressOK.
2053535-003M MAC™2000 91
ServiceSetup
TheServiceSetupmenuopens.
5. UsethetrimpadtohighlightSoftwareUpdateandpressEnter.
Thefollowingwindowopens:
UpdatingtheOS
1. ClickOSUpdate.
•IfyoutrytoupdatetheOSwithouthavingACpowerconnected(youare
runningonbatterypower),thesystemdisplaysthefollowingmessage:
92 MAC™2000 2053535-003M
ConnecttheACpowercordbeforecontinuing.
•IfanSDcardisnotpresent,thesystemdisplaysthefollowingmessage:
a. InserttheSDcardandclickOK(onthetabatthebottomofthewindow).
b. AfteryouclickOK,oriftheSDcardisalreadypresent,thefollowing
messageisdisplayed:
ServiceSetup
2. ClickYestoproceedtoupdatingthesoftware.
Whilethesoftwareisupdating,thefollowingmessageisdisplayed:
2053535-003M MAC™2000 93
ServiceSetup
3. ClickOK(onthetabatthebottomofthescreen).
Thedateandtimefreezeandthefollowingmessageisashedonthescreen:
•Iftheupdateisnotsuccessful,correspondingmessagesaredisplayed.
CheckthelesontheSDcard.
•Iftheupdateissuccessful,thefollowingmessageisdisplayed:
4. Restartyoursystemfortheupdatetotakeeffect.
UsingForcedSoftwareUpdate
Thissystemsupportstheforcedsoftwareupdatefeature,whichyoucanusein
caseswherethesystemisnotbootingupafterasoftwareupdate.Usethefollowing
proceduretoinitiatetheforcedsoftwareupdate:
1. EnsuretheSDcardhasthecorrectOSsoftwareleversion.
2. CreateadummyleontheSDcardandrenameitupdate.com.
3. InserttheSDcardintheslotandthenpoweronthesystem.
Thesystemtakesafewminutestocomeup.
4. Whenitisdone,removetheSDcardanddeletetheupdate.comlefromthe
SDcard.
NOTE:
Aforcedsoftwareupdatetakesapproximately3to4minutes.
UpdatingtheAcquisitionFirmware
1. ClickAcquisitionFWUpdate.
94 MAC™2000 2053535-003M
Thefollowingwindowopensdisplayingthermwareversiononthedevice
andSDcard:
2. PressEntertoupdatethermware.
Whiletheupdateisinprogress,thefollowingmessageisdisplayed:
ServiceSetup
2053535-003M MAC™2000 95
ServiceSetup
3. PressEntertoselecttheOKbutton.
Thefollowingmessageisdisplayed:
•Iftheupdatefails,correspondingmessagesaredisplayed.Thefollowing
messageisanexample:
a. PressOKtoreturntotheupdatescreens.
b. CheckthelesintheSDcard.
•Ifthermwareupdateiscompleted,thefollowingmessageisdisplayed:
4. Restartyoursystemfortheupdatetotakeeffect.
96 MAC™2000 2053535-003M
UpdatingthePSOCFirmware
1. ClickonPSOCFWUpdate.
ServiceSetup
2053535-003M MAC™2000 97
ServiceSetup
Thefollowingwindowopensdisplayingthermwareversiononthedevice
andSDcard:
2. PressEntertoupdatethermware.
Whiletheupdateisinprogress,thefollowingmessageisdisplayed:
3. PressEntertoselecttheOKbutton.
Thefollowingmessageisdisplayed:
•Iftheupdatewasnotsuccessful,thesystemgivescorrespondingerror
messages.
a. PressOKontheerrormessagetoreturntotheupdatescreen.
b. CheckthelesintheSDcard.
•Iftheupdatewassuccessful,thesystemshutsdown.
FormatingtheFlashMemory
TheFormatFlashprocedureisusedtoresolvememoryordatacorruptionproblems.It
returnsthedevicetoitsfactorysettingsbyerasingallthedata,includingECGrecords,
systemcongurations,devicecongurations,optioncodes,logles,auditles,patient
lists,andexerciseprotocols.Beforeattemptingtoformatthedevice’sashmemory,
98 MAC™2000 2053535-003M
ServiceSetup
youshouldconrmthatanymemoryordatacorruptionissuesarenottheresultofa
changeinthedevice’scongurationsorsettings.
Usethefollowingproceduretoformatthedevice’sashmemory.
WARNING:
DATALOSS—Formattingthedevice’sinternalashdriveerasesallthedatain
memoryandreturnsthedevicetoitsfactorysettings.
Ifpossible,backuporrecordanydatathatyoudonotwanttolosebefore
performingthefollowingprocedure.
1. ConnectthedevicetoACpower.
2. GototheServiceSetupmenu,asdescribedin“AccessingServiceSetup”on
page71.
3. SelecttheFormatFlashbuttonandpressEnter.
IfthedeviceisnotconnectedtoACpower,awindowwillopen.Connectthe
devicetoACpowerandselectOK.
WhenthedeviceisconnectedtoACpower,awindowopenswiththefollowing
message:
4. SelectYes.
Thefollowingmessageisdisplayed:
5. ClickOK.
Thesystemstartstoformattheashmemory.
Whiletheashisbeingformatted,thedisplaywillseemtobelocked.
Whentheformatiscomplete,astatusmessageisdisplayed,andyouare
promptedtorestartthesystem.
2053535-003M MAC™2000 99
ServiceSetup
Afterthedevice’sashmemoryisformatted,youneedtorecongurethedevice,
RefertotheMAC™2000ECGAnalysisSystemOperator’sManualforinformationon
conguringthedevice.
Ifformattingthedevice’smemoryashdoesnotresolvethememoryordata
corruptionissues,youneedtoreplacethemainPWA.Forinstructionsonreplacingthe
PWA,see“ReplacingthePWA”onpage181.
OpeningaCommandPrompt
Usethefollowingprocedurestoopenacommandpromptfromwhichyoucanissue
variousnetworkdiagnosticcommands.
1. GototheServiceSetupmenu,asdescribedin“AccessingServiceSetup”on
page71.
2. SelecttheOpenCommandPromptbutton.
Thefollowingcommandpromptwindowopens:
3. Issueanyofthefollowingcommandsasnecessary.
Command
ping<ipaddress>
ipconfig
netstat
netstat-e
netstat-n
netstat-p
netstat—r
Description
ThesystemqueriesandreceivesrepliesfortheIPaddress
identiedinthecommand.
ReturnsinformationrelatedtotheTCP/IPcongurationof
thesystem.
Returnsnetworkstatisticsforthesystem.
Returnsethernetstatisticsforthenetwork.
Returnssystemnetworkaddressesandportnumbersin
numericalform.
ReturnsstatisticsspeciedfortheIP4protocol.
Returnstheroutingtable.
4. Whenyouaredone,typeexitandpressEntertoclosethecommandprompt
windowandreturntotheServiceMenu.
100 MAC™2000 2053535-003M