General Electric LOGIQ V3_V5_V5 Expert Basic Service Manual_SM_5726264-100_8 V5 Expert Basic Service Manual Direction Number: 5726264-100 English

Page 1
GE
Technical Publication

LOGIQ V3/V5/V5 Expert

Basic Service Manual Direction Number: 5726264-100 English
Rev.8
All Material Copyright © 2016-2017 by General Electric Company Inc. All Rights Reserved.
Page 2
Product Information
This Manual covers the software version of R2.x.x for LOGIQ V3, LOGIQ V5, LOGIQ V5 Expert ultrasound system.
Copyright © 2016-2017, General Electric Company. GE Medical Systems, a General Electric Company, going to market as GE
Manufacturer:
GE Medical Systems (China) Co., Ltd. No. 19, Changjiang Road WuXi National Hi-Tech Development Zone Jiangsu, P.R. China 214028 TEL: +86 510 85225888; FAX: +86 510 85226688 www.gehealthcare.com
Page 3
Revision history
Revision History
Date
Revision
Rev.1 2016/01/21 Initial Release
Rev.2 2016/06/12 Update control panel figure and book format
Rev.3 2016/08/10 Update warning label
Rev.4 2016/08/24 Update peripheral list
Rev.5 2016/10/12 Update unpacking process
Rev.6 2017/04/19 Update as software updates
Rev.7 2017/06/14 Add manufacturer information
Rev.8 2017/08/23 Update peripheral list
CHAPTER NUMBER Revision CHAPTER NUMBER Revision
(YYYY/MM/DD) Reason for change
List of Effected Pages (LOEP)
Front Rev.8 Chapter 6 Rev.8
Front matter Rev.8 Chapter 7 Rev.8
TOC Rev.8 Chapter 8 Rev.8
Chapter 1 Rev.8 Chapter 9 Rev.8
Chapter 2 Rev.8 Chapter 10 Rev.8
Chapter 3 Rev.8 Index Rev.8
Chapter 4 Rev.8 Rear Cover Rev.8
Chapter 5 Rev.8
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on MyWorkshop/ePDM (GE electronic Product Data Management). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-1 5726264-100 English Rev.8
Page 4
Important precautions

Translation policy

i-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 5
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-3 5726264-100 English Rev.8
Page 6
i-4 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 7
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-5 5726264-100 English Rev.8
Page 8
i-6 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 9
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-7 5726264-100 English Rev.8
Page 10
i-8 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 11
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-9 5726264-100 English Rev.8
Page 12
i-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 13
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-11 5726264-100 English Rev.8
Page 14

Damage in transportation

All packages should be closely examined at time of delivery. If damage is apparent, write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.

Certified electrical contractor statement - For USA Only

All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
i-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 15

Omission and errors

If there are any omissions, errors or suggestions for improving this documentation, contact the GE Global Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details.
Mail the
information to:
GE Medical Systems (China) Co., Ltd. No. 19 Changiang Road Wuxi National Hi-Tech Dev. Zone Jiangsu P.R.China 214028 TEL: +86 510 85225888; FAX: +86 510 85226688
GE employees should use TrackWise to report service documentation issues.
These issues will then be in the internal problem reporting tool and communicated to the writer.
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-13 5726264-100 English Rev.8
Page 16

Service Safety Considerations

WARNING
DANGER DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH,
ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, refer to Chapter 1 in the Service Manual.
i-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 17

Legal notes

Proprietary to GE

The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE.
GE may revise this publication from time to time without written notice.
Permission to use this Advanced Service Software and related documentation (herein called the material) by persons other than GE employees is provided only under an Advanced Service Package License relating specifically to this Proprietary Material. This is a different agreement from the one under which operating and basic service software is licensed. A license to use operating or basic service software does not extend to or cover this software or related documentation.
If you are a GE employee or a customer who has entered into such a license agreement with GE to use this proprietary software, you are authorized to use this Material according to the conditions stated in your license agreement.
However, you do not have the permission of GE to alter, decompose or reverse-assemble the software, and unless you are a GE employee, you may not copy the Material. The Material is protected by Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal prosecution.
If you are not party to such a license agreement or a GE Employee, you must exit this Material now.

Trademarks

All products and their name brands are trademarks of their respective holders.

Copyrights

All Material Copyright © 2016-2017 by General Electric Company Inc. All Rights Reserved.
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-15 5726264-100 English Rev.8
Page 18
i-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 19
Translation policy - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-2 Damage in transportation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-12 Certified electrical contractor statement - For USA Only - - - - - - - - - - - - i-12 Omission and errors - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-13 Service Safety Considerations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-14 Legal notes - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-15 Proprietary to GE - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-15 Trademarks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-15 Copyrights - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-15
Table of Contents Chapter 1 — Introduction
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-2
Manual Overview
Contents in this manual - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-3 Typical users of the Proprietary Service Manual - - - - - - - - - - - - - - - - - 1-4 Purpose of Operator Manual(s) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4 LOGIQ V3/V5/V5 Expert models covered by this manual - - - - - - - - - - - 1-5
Important conventions
Conventions used in book - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-6 Standard hazard icons - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-8
Product icons Labels locations
Rating Plate Location - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-11
Safety considerations
Contents in this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-16 Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-16 Human Safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-16 Mechanical safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-19 Electrical safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-22
Dangerous procedure warnings Lockout/Tagout (LOTO) requirements Returning probes and repair parts EMC, EMI and ESD
Electromagnetic Compatibility (EMC) - - - - - - - - - - - - - - - - - - - - - - - - 1-27 CE Compliance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-27 Electrostatic discharge (ESD) prevention - - - - - - - - - - - - - - - - - - - - - 1-28
Customer assistance
Contact information - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-29

Table of Contents

LOGIQ V3/V5/V5 Expert – Basic Service Manual i-17 5726264-100 English Rev.8
Page 20
Phone numbers for Customer Assistance - - - - - - - - - - - - - - - - - - - - - 1-30 System manufacturer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-30 Authorized Representative- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-31 Factory Site - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-31
Chapter 2 — Site Preparations
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2
General Ultrasound system requirements
Contents in this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-3 Ultrasound system environmental requirements- - - - - - - - - - - - - - - - - - 2-3 Electrical requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-6 EMI limitations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9 Probes environmental requirements - - - - - - - - - - - - - - - - - - - - - - - - - 2-11 Time and manpower requirements - - - - - - - - - - - - - - - - - - - - - - - - - - 2-11
Facility needs
Contents in this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-12 Purchaser responsibilities - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-13 Required facility needs - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-14 Desirable features- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-15 Minimal floor plan suggestion- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-16 Recommended floor plan suggestion - - - - - - - - - - - - - - - - - - - - - - - - 2-17 Networking setup requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-18
Chapter 3 — System Setup
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
Setup reminders
Average setup time - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-3 Setup warnings- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-3
Receiving and unpacking the equipment
Purpose of this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-6 Contents in this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-6 Warnings for receiving and unpacking - - - - - - - - - - - - - - - - - - - - - - - - 3-6 LOGIQ V3/V5/V5 Expert transportation box label- - - - - - - - - - - - - - - - - 3-7 Unpacking the LOGIQ V3/V5/V5 Expert - - - - - - - - - - - - - - - - - - - - - - - 3-8 Packing the Equipment - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-12
Preparing for setup
Verify customer order - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-13 Physical inspection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-13 EMI protection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-13
Completing the setup
Purpose of this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-14 Contents in this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-14 System specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-14 Electrical specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-16 Connecting Cables - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-17 Peripherals/Accessories Connector Panel- - - - - - - - - - - - - - - - - - - - - 3-18 Connections on the I/O Rear Panel - - - - - - - - - - - - - - - - - - - - - - - - - 3-20
i-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 21
Connecting probes - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-20 Powering the system - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-20
System Configuration
Purpose of this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-21 LOGIQ V3/V5/V5 Expert configuration - - - - - - - - - - - - - - - - - - - - - - - 3-21
Peripherals Installation
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-26 Furnished materials - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-27 Peripherals Installation Instructions - - - - - - - - - - - - - - - - - - - - - - - - - 3-29
Option Setup
Software Option Installation Procedure - - - - - - - - - - - - - - - - - - - - - - - 3-44
Paperwork after setup
Contents in this Section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-47 User’s Manual(s) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-47 Product Locator Installation Card - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-47
Chapter 4 — General Procedures and Functional Checks
Overview
Purpose of this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2 Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2 Special Equipment required - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2
General procedures
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-4 Power ON/Boot Up - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-5 Power off - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-8 Check System Date and Time - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-10 Manufacturer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-11 Removable media - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-11 Backup and Restore Database, Preset Configurations and Images - - - 4-12
Functional checks
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-21 Contents in this Section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-21 Preparation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-21 Basic Controls - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-22 Control Panel - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-22 LOGIQ V5/LOGIQ V3 SoftMenu Key Tour- - - - - - - - - - - - - - - - - - - - - 4-23 Control Panel - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-24 Performance Tests - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-25 2D Mode (B mode) Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-26 2D Mode Screen Example- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-26 Preparations- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-26 Adjust the 2D mode controls - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-26 Image Quality Check (IQC) preset for service - - - - - - - - - - - - - - - - - - 4-27 M Mode Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-29 M-Mode Screen Example - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-29 Preparations- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-29 Adjust the M Mode controls - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-29 Color Mode Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-30
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-19 5726264-100 English Rev.8
Page 22
Select Color 2D Mode - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-30 Adjust the Color Mode controls - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-31 Select Color M Mode- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-32 PW Doppler Mode Checks- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-34 Adjust the PW Doppler Mode controls- - - - - - - - - - - - - - - - - - - - - - - - 4-35 Tissue Velocity Imaging (TVI) Checks- - - - - - - - - - - - - - - - - - - - - - - - 4-36 Adjust the TVI Controls - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-37 Basic Measurements- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-38 Probe/Connectors Check- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-39 ECG Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-41 Cineloop Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-42 Preparation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-42 Adjust the Cineloop controls - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-42 Back End Processor checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-42 Operator Panel Test - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-43 Peripheral checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-43 Turn OFF Power to LOGIQ V3/V5/V5 Expert- - - - - - - - - - - - - - - - - - - 4-43 Mechanical Functions Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-43
Application Turnover Check List
Software Configuration Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-45
Site Log My Trainer
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-47
Chapter 5 — Components and Functions (Theory)
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-2
Block Diagram and Theory
Block Diagram - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-3
Common Service Platform
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-5
Chapter 6 — Service Adjustments
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-2
LCD Monitor Adjustments
Purpose of this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-3
Chapter 7 — Diagnostics/Troubleshooting
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-2
Gathering Trouble Data
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-3 Collect Vital System Information - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-3 Collect a Trouble Image with Logs - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-4
Screen Capture
Capturing a Screen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-6 Capturing a Screen with Service Dongle - - - - - - - - - - - - - - - - - - - - - - - 7-9
Global Service User Interface (GSUI)
Enter global service user interface - - - - - - - - - - - - - - - - - - - - - - - - - - 7-12
i-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 23
Control Frame - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-15 Common Diagnostics - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-16
Network and Insite II Configuration
Network Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-29
Chapter 8 — Replacement Procedures
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-2
Warnings and important information
Warnings - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-3 Returning/shipping probes and repair parts - - - - - - - - - - - - - - - - - - - - - 8-4
Loading the software
Purpose of this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-5 Customer provided prerequisite - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-5 Data Management - moving all images - - - - - - - - - - - - - - - - - - - - - - - - 8-5 Backing up the Patient Archive and System Configurations - - - - - - - - - 8-6 Restoring up the Patient Archive and System Configurations - - - - - - - - 8-6 Recording important settings and parameters - - - - - - - - - - - - - - - - - - - 8-7 Loading the System Software - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-8 Rewrite the Serial Number - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-22
Chapter 9 — Renewal Parts
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9-2
List of Abbreviations Renewal Parts Lists
Operator Console Assy - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9-4 Probe - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9-5 Peripheral - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9-5 Power Cord - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9-8 Manuals- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9-9
Chapter 10 — Care and Maintenance
Overview
Contents in this chapter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-2
Warnings Why do maintenance
Periodic maintenance inspections - - - - - - - - - - - - - - - - - - - - - - - - - - 10-4 Keeping records - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-4 Quality assurance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-5
Maintenance task schedule
How often should maintenance tasks be performed? - - - - - - - - - - - - - 10-6
Tools required
Standard GE tool kit - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-8 GE-2 tool kit - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-10 Special tools, supplies and equipment used for maintenance - - - - - - 10-11
System maintenance
Preliminary checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-12 Functional checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-13 Physical inspection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-15
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-21 5726264-100 English Rev.8
Page 24
Optional Diagnostic Checks- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-16 Probe maintenance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-17
Electrical safety tests
Safety test overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-19 Leakage current limits - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-21 Outlet test - wiring arrangement - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-23 Grounding continuity - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-24 Chassis leakage current test - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-25 Probe leakage current test- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-27
When there's too much leakage current …
AC/DC Fails - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-30 Chassis Fails - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-30 Probe Fails- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-31 Peripheral Fails - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-31 Still Fails - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-31 New Unit - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-31 ECG Fails - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-31
Inspection Paperwork
Ultrasound Inspection Forms - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-32
Electrical Safety Tests Log
Index
i-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 25

Chapter 1

Introduction

This chapter describes important issues related to safely servicing the Ultrasound system. The service provider must read and understand all the information presented here before installing or servicing the units.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-1 5726264-100 English Rev.8
Page 26
Introduction

1-1-1 Contents in this chapter

1-1 ‘Overview’ on page 1-2
1-2 ‘Manual Overview’ on page 1-3
1-3 ‘Important conventions’ on page 1-6
1-4 ‘Product icons’ on page 1-10
1-5 ‘Labels locations’ on page 1-11
1-6 ‘Safety considerations’ on page 1-16
1-7 ‘Dangerous procedure warnings’ on page 1-24
1-8 ‘Lockout/Tagout (LOTO) requirements’ on page 1-25
1-9 ‘Returning probes and repair parts’ on page 1-26
1-10 ‘EMC, EMI and ESD’ on page 1-27
1-11 ‘Customer assistance’ on page 1-29

1-1 Overview

1-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 27
This manual provides installation and service information for the LOGIQ V3/V5/V5 Expert Ultrasound system. It is divided in ten chapters as shown below.

1-2-1 Contents in this manual

The manual is divided into ten chapters.
In the beginning of the manual, before chapter 1, you will find the Revision overview, the Important precautions including
Translation policy, Damage in transportation, Certified electrical contractor statement, Omission & errors, Service safety considerations and Legal notes, and the Table of Contents (TOC).
Manual Overview

1-2 Manual Overview

Chapter number
1.
2.
3.
4.
5.
6.
An Index has been included after Chapter 10.
Table 1-1: Contents in this manual
Chapter title Description
Introduction Contains a content summary and warnings.
Site preparations Contains pre-setup requirements for the LOGIQ
V3/V5/V5 Expert.
System Setup Contains setup procedure with procedure
checklist for the system.
General Procedures and Functional Checks
Components and Functions (Theory)
Service Adjustments Contains instructions on how to make any
Contains functional checks that must be performed as part of the installation, or as required during servicing and periodic maintenance.
Contains block diagrams and functional explanations of the electronics.
available adjustments to the LOGIQ V3/V5/V5 Expert.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-3 5726264-100 English Rev.8
Page 28
Introduction
Table 1-1: Contents in this manual (Continued)
Chapter
Chapter title Description
number
7.
8.
9.
Diagnostics/ Troubleshooting
Replacement procedures
Provides procedures for running diagnostic or related routines for the LOGIQ V3/V5/V5 Expert.
Provides disassembly procedures and reassembly procedures for all changeable FRU.
Renewal Parts Contains a complete list of replacement parts for
LOGIQ V3/V5/V5 Expert.
10.
Care & Maintenance Provides periodic maintenance procedures for
LOGIQ V3/V5/V5 Expert.
N/A
Index A quick way to the topic you’re looking for.

1-2-2 Typical users of the Proprietary Service Manual

GEHC Service Personnel (setup, maintenance, etc.)
GEHC Online Center Personnel
Licensed Hospital’s Service Providers

1-2-3 Purpose of Operator Manual(s)

The Operator Manual(s) should be fully read and understood
before operating the LOGIQ V5/LOGIQ V3 and also kept near
the unit for quick reference.
NOTE: Probe information displayed on screen does not necessarily
reflect the probes available on your ultrasound system. Please refer to the probe list for available probes and features.
1-4 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 29
Manual Overview

1-2-4 LOGIQ V3/V5/V5 Expert models covered by this manual

Table 1-2: LOGIQ V3/V5/V5 Expert Model Designations
Model Number Description System SW
H48612AA LOGIQ V3 R2.x.x
H48612AB LOGIQ V5 R2.x.x
H48612AC LOGIQ V5 EXPERT R2.x.x
NOTE: When not otherwise specified, the contents of this manual
applies to all LOGIQ V3/V5/V5 Expert models.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-5 5726264-100 English Rev.8
Page 30
Introduction

1-3 Important conventions

1-3-1 Conventions used in book

Important conventions, used in this document, are described next.

1-3-1-1 Model designations

This manual covers the LOGIQ V3/V5/V5 Expert Ultrasound systems listed in:
1-2-4 ‘LOGIQ V3/V5/V5 Expert models covered by this manual’ on page 1-5.

1-3-1-2 Icons

Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels, and conventions used on the product and in the service information are described in this chapter.
1-6 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 31

1-3-1-3 Safety precaution messages

DANGER
WARNING
CAUTION
Various levels of safety precaution messages may be found on the equipment and in the service information. The different levels of concern are identified by a flag word that precedes the precautionary message. Known or potential hazards to personnel are labeled in one of three ways:
• DANGER
• WARNING
• CAUTION
Danger is used to indicate the presence of a hazard that will cause severe personal injury or death if the instructions are ignored.
Warning is used to indicate the presence of a hazard that can cause severe personal injury and property damage if instructions are ignored.
Important conventions
Caution is used to indicate the presence of a hazard that will or can cause minor personal injury and property damage if instructions are ignored. Equipment damage possible.
NOTE: Notes are used to provide important information about an item
or a procedure.
NOTE: Be sure to read the notes; the information contained in a note
can often save you time or effort.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-7 5726264-100 English Rev.8
Page 32
Introduction

1-3-2 Standard hazard icons

Important information will always be preceded by either the exclamation point (!) contained within a triangle, or the symbols for “Danger”, “Warning” or “Caution”, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could possibly cause harm.
Table 1-3: Standard hazard icons
ELECTRICAL
MECHANICAL
RADIATION
LASER
HEAT
PINCH
NOTE: Even if a symbol isn’t used on the product or in this manual, it
may be included for your reference.
1-8 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 33
Important conventions

1-3-2-1 Standard Icons that indicate that a special procedure is to be used

Some others icons make you aware of specific procedures that should be followed.
Table 1-4: Standard Icons that indicates that a special procedure is to be used
Avoid Static Electricity Tag and Lock Out Wear Eye Protection
Hand Protection Foot Protection Wear Eye Protection
Be sure to read the notes; the information contained in a note can often save you time or effort.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-9 5726264-100 English Rev.8
Page 34
Introduction

1-4 Product icons

It is important to refer to the current revision of the Ultrasound system’s Basic User Manual for a full list of product labels prior to servicing the system.
1-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 35
It is important to refer to the current revision of the Ultrasound system’s Basic User Manual for a full list of product labels prior to servicing the system.

1-5-1 Rating Plate Location

LOGIQ V3/V5/V5 Expert labels are provided in English.
The labels are at the bottom of the system. The label content may be different for different regions. Please refer to the labels on the system for the actual content.
Labels locations

1-5 Labels locations

LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-11 5726264-100 English Rev.8
Page 36
Introduction
1-5-1 Rating Plate Location(continued)
Figure 1-1. LOGIQ V5 Expert Label Location
1-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 37
1-5-1 Rating Plate Location(continued)
Labels locations
Figure 1-2. LOGIQ V5 Label Location
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-13 5726264-100 English Rev.8
Page 38
Introduction
1-5-1 Rating Plate Location(continued)
Figure 1-3. LOGIQ V3 Label Location
1-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 39
1-5-1 Rating Plate Location(continued)
1. Rating Plate
Table 1-5: Label Icons
Label/Icon Purpose/Meaning Location
Labels locations
Every system has a unique marking for identification, the Unique Device Identification (UDI) Label. The UDI label consists of a series of alpha-numeric characters and barcode which uniquely identify the LOGIQ V3/ V5/V5 Expert system as a medical device manufactured by General Electric. Scan or enter the UDI information into the patient health record as required by country-specific laws.
Serial Number.
Rating plate
Rating plate
NOTE: If the new label is needed during the service activities, please
click “Ask an Expert“ to submit the case in the support central: http://supportcentral.ge.com/products/sup_products.asp?prod_ id=44177.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-15 5726264-100 English Rev.8
Page 40
Introduction

1-6 Safety considerations

1-6-1 Contents in this section

1-6-2 ‘Introduction’ on page 1-16
1-6-3 ‘Human Safety’ on page 1-16
1-6-4 ‘Mechanical safety’ on page 1-19
1-6-5 ‘Electrical safety’ on page 1-22

1-6-2 Introduction

The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.

1-6-3 Human Safety

Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in a LOGIQ V3/V5/V5 Expert Training Seminar are authorized to service the equipment.
1-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 41
1-6-3 Human Safety(continued)
WARNING
WARNING
WARNING
Because of the limited access to cabinets and equipment in the field, placing people in awkward positions, GE has limited the lifting weight for one person in the field to 16 KG (35 LBS). Anything over 16 KG (35 LBS) requires 2 people.
Have two people available to deliver and unpack the LOGIQ V3/V5/V5 Expert.
Attempts to move the Ultrasound system considerable distances or on an incline by one person could result in injury or damage or both.
Ensure that the Ultrasound system is turned off and unplugged
Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation. The amber light on the OP panel ON/OFF button will turn off.
Safety considerations
Ultrasound system components may be energized. Always refer to the Ultrasound system's Proprietary Service Manual for LOTO warnings and cautions
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-17 5726264-100 English Rev.8
Page 42
Introduction
WARNING
WARNING
WARNING
WARNING
1-6-3 Human Safety(continued)
Risk of electrical shock, Ultrasound system must be turned off and disconnected from power source. Cord must be controlled at all times.
Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation. The amber light on the OP panel on/off button will turn off.
Ultrasound System components may be energized. Always refer to the Ultrasound system's Proprietary Service Manual for LOTO warnings and cautions
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety glasses, and kneeling pad, to reduce the risk of injury.
Beware of possible sharp edges on all mechanical parts. If sharp edges are encountered, the appropriate PPE should be used to reduce the risk of injury.
Wear all PPE including gloves as indicated in the chemical MSDS.
1-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 43

1-6-4 Mechanical safety

WARNING
WARNING
WARNING
WARNING
CAUTION
CAUTION
While the software install procedure is designed to preserve data, you should save any patient data, images, system setups to removable media or hardcopy before doing a software upgrade.
Ultrasound probes are highly sensitive medical instruments that can easily be damaged by improper handling. Use care when handling and protect from damage when not in use. Do NOT use a damaged or defective probe. Failure to follow these precautions can result in serious injury and equipment damage.
Never use a probe that has fallen to the floor. Even if it looks OK, it may be damaged.
Safety considerations
When the Ultrasound system is raised for a repair or moved along any incline, use extreme caution since it may become unstable and tip over.
Take extra care when moving the system.
The LOGIQ V3/V5/V5 Expert weighs approximately 65 kg (144 lbs) or more, depending on installed peripherals, when ready for use. To avoid possible injury and equipment damage when transporting from one area of use to another:
Be sure the pathway is clear.
Limit movement to a slow careful walk.
Use two or more persons to move the equipment on inclines or long distance.
Before you move or transport the Ultrasound system, make sure to lock the LCD monitor arm firmly and flip down the monitor to prevent damage to the Ultrasound system.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-19 5726264-100 English Rev.8
Page 44
Introduction
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
1-6-4 Mechanical safety(continued)
To avoid injury when you move the LCD monitor and the monitor arm, do not put your finger, hand, or object on the joint of the monitor or the monitor arm.
Ensure that nobody touches the console arm when moving the operator panel.
Do not move the Ultrasound system if the Operator Panel is in unlocked position.
Do not transport LOGIQ V3/V5/V5 Expert in a vehicle without locking the casters (wheels) and securing it as described in chapter 4.
Use protective glasses during drilling, filing smooth surfaces, and during all other work where eyes need protection.
Use safety shoes when doing work where there is any chance of foot injury.
1-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 45
1-6-4 Mechanical safety(continued)
CAUTION
CAUTION
Use protective gloves when working with sharp edges or when directed to wear PPE during a removal/replacement procedure.
Be careful not to pinch any of the cables.
NOTE: Special care should be taken when transporting the Ultrasound
system in a vehicle:
• Before transporting, place the system in its special storage case.
Lock the wheels (brake)
Ensure that the system is firmly secured while inside the vehicle.
Secure system with straps or as directed otherwise to prevent motion during transport.
Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and erratic stops or starts.
Safety considerations
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-21 5726264-100 English Rev.8
Page 46
Introduction
WARNING
WARNING

1-6-5 Electrical safety

1-6-5-1 Safe practices

Follow these guidelines to minimize shock hazards whenever you are using the Ultrasound system:
To minimize shock hazard, the equipment chassis must be
The Ultrasound system is equipped with a three-conductor
The power outlet used for this equipment should not be
Both the system power cable and the power connector must
Connecting a LOGIQ V3/V5/V5 Expert to the wrong voltage level will most likely destroy it.
connected to an electrical ground.
AC power cable. This must be plugged into an approved electrical outlet with safety ground.
shared with other types of equipment.
meet international electrical standards.
DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU UNIT LEVEL AT ANY CIRCUMSTANCES.
1-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 47

1-6-5-2 Probes

Safety considerations
Follow these guidelines before connecting a probe to the Ultrasound system:
Inspect the probe prior to each use for damage or degradation to the:
• housing
cable strain relief
• lens
• seal
• connector pins
• locking mechanism
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any liquid.
The system has more than one type of probe port. Use the appropriate probe port designed for the probe you are connecting.

1-6-5-3 Peripherals

Refer to the Patient Safety Environment section of the User’s Manual for peripheral isolation information.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-23 5726264-100 English Rev.8
Page 48
Introduction
DANGER
WARNING
WARNING
WARNING
WARNING
WARNING

1-7 Dangerous procedure warnings

Warnings, such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
If the covers are removed from an operating LOGIQ V3/V5/V5 Expert, some metal surfaces may be warm enough to pose a potential heat hazard if touched, even while in shutdown mode.
Explosion Warning
DO NOT operate the equipment in an explosive atmosphere. Operation of any electrical equipment in such an environment constitutes a definite safety hazard.
DO NOT substitute parts or modify equipment
Because of the danger of introducing additional hazards, ONLY install GE approved parts. DO NOT perform any unauthorized modification of the equipment.
SHUT DOWN FORCEDLY OR PLUG IN/OUT ACDC INVALID MAY CAUSE THE DAMAGE OF SYSTEM FILES.
AFTER UNPLUG POWER CORD, WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION.
1-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 49

Lockout/Tagout (LOTO) requirements

WARNING
1-8 Lockout/Tagout (LOTO)
requirements
Follow Lockout/Tagout requirements by ensuring you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout (LOTO):
1. Plan and prepare for shutdown.
2. Shutdown the equipment.
3. Isolate the equipment.
4. Apply Lockout/Tagout Devices.
5. Control all stored and residual energy.
6. Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Energy Control and Power Lockout for LOGIQ V3/V5/V5 Expert.
When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as there are no test points to verify isolation.
Ultrasound System components may be energized.
TEST POINTS TO VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-25 5726264-100 English Rev.8
Page 50
Introduction
1-9 Returning probes and repair
parts
Equipment being returned must be clean and free of blood and other infectious substances. GE policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment. Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that
“items that were saturated and/or dripping with human blood that are now caked with dried blood; or which were used or intended for use in patient care” are “regulated medical waste” for transportation purposes and must be transported as a hazardous material.
NOTE: The USER/SERVICE staff should dispose of all the waste
properly, per federal, state, and local waste disposal regulations.
The Ultrasound system is not meant to be used for long-term storage of patient data or images. The user is responsible for the data on the system and a regular backup is highly recommended.
If the system is sent for repair, please ensure that any patient information is backed up and erased from the system before shipping. It is always possible during system failure and repair to lose patient data. GE is not responsible for the loss of this data.
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes, GE will ascertain agreement from the customer. Patient information shall only be transferred by approved service processes, tools and devices restricting access, protecting or encrypting data where required, and providing traceability in the form of paper or electronic documents at each stage of the procedure while maintaining compliance with cross-border restrictions of patient information transfers.
1-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 51

1-10 EMC, EMI and ESD

1-10-1 Electromagnetic Compatibility (EMC)

Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface. Inadequate compatibility results when a susceptible device fails to perform as intended due interference from its environment or when the device produces unacceptable levels of emission to its environment. This interference is often referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy, EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power supply.
EMC, EMI and ESD

1-10-2 CE Compliance

LOGIQ V3/V5/V5 Expert conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
For applicable standards, refer to the Safety Chapter of the Ultrasound system User’s Manual.
NOTE: For CE Compliance, it is critical that all covers, screws,
shielding, gaskets, mesh, clamps, are in good condition, installed tightly without skew or stress. Proper installation following all comments noted in this service manual is required in order to achieve full EMC performance.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-27 5726264-100 English Rev.8
Page 52
Introduction
WARNING
WARNING

1-10-3 Electrostatic discharge (ESD) prevention

DO NOT touch any boards with integrated circuits prior to taking the necessary ESD precautions.
Always connect yourself, via an arm-wrist strap, to the advised ESD connection point located on the rear of the Ultrasound system (near the power connector).
Follow general guidelines for handling of electrostatic sensitive equipment.
Risk of electrical shock, Ultrasound system must be turned off. Avoid all contact with electrical contacts, conductors and components. Always use non-conductive handles designed for the removal and replacement of ESD sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before making contact.
1-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 53

1-11 Customer assistance

1-11-1 Contact information

If this equipment does not work as indicated in this service manual or in the basic user manual, or if you require additional assistance, please contact the local distributor or appropriate support resource, as listed below.
Before you call, identify the following information, and acquire log (Alt+D) to send to the Customer Care team:
1. System ID serial number.
2. Software version.
3. Date and time of occurrence.
4. Sequence of events leading to issue.
5. Is the issue repeatable?
6. Imaging mode, probe, preset/application.
7. Media brand, speed, capacity, type.
8. Save secondary image capture, cine loop.
Customer assistance
NOTE: Restart the application before resuming clinical scanning.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-29 5726264-100 English Rev.8
Page 54
Introduction

1-11-2 Phone numbers for Customer Assistance

Table 1-6: Phone numbers for Customer Assistance
LOCATION PHONE NUMBER
USA GE Healthcare - GE Medical Systems Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa, WI 53226
Canada Phone: 1-800-668-0732
Latin America Service
Europe (OLC-EMEA) GE Ultraschall Deutschland Gmbh & Co. KG Beethovenstraße 239 Postfach 11 05 60, D-42655 Solingen Germany
Online Services Ultrasound Asia Phone:
Service: On-site 1-800-437-1171
Service Parts 1-800-558-2040
Application Support 1-800-682-5327 or
1-262-524-5698
Application Support
OLC - EMEA Phone:
Fax: +49 (0) 2122-8024-31
• Australia
• China
• India
• Japan
• Korea
• Singapore
1-800-321-7937 1-262-524-5698
+49 (0) 212 2802 - 652 +33 1 3083 1300
+(61) 1-800-647-855 +(86) 800-810-8188 +(91) 1800-425-8025 +(81) 42-648-2940 +(82) 2620 13585 +(95) 6277-3444

1-11-3 System manufacturer

Table 1-7: System manufacturer
MANUFACTURER PHONE NUMBER FAX NUMBER
GE Medical Systems (China) Co., Ltd. No.19 Changjiang Road WuXi National Hi-Tech Development Zone Jiangsu P.R.China 214028
+86 510 85225888 +86 510 85226688
1-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 55

1-11-4 Authorized Representative

Table 1-8: Authorized Representative
AUTHORIZED REPRESENTATIVE
Customer assistance
TELEPHONE / FAX
NUMBER
The location of the CE marking is shown in the Safety chapter of the Basic User Manual.
Authorized EU Representative/European registered place of business: GE Medical Systems Information Technologies GmbH (GEMS IT GmbH) Munzinger Strasse 5, D-79111 Freiburg, GERMANY

1-11-5 Factory Site

Table 1-9: Factory Site
Factory Site
GE Medical Systems (China) Co., Ltd. No.19, Changjiang Road, Wuxi National Hi-Tech Development Zone, Jiangsu, P.R. China 214028
TEL: +86 510-85225888 FAX: +86 510-85226688
+49 761 45 43 -0 / +49 761 45 43 -233
TELEPHONE / FAX
NUMBER
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-31 5726264-100 English Rev.8
Page 56
Introduction
1-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 57

Chapter 2

Site Preparations

This chapter provides the information required to plan and prepare for the setup of an Ultrasound system. Included are descriptions of the facility and electrical needs to be met by the purchaser of the units.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-1 5726264-100 English Rev.8
Page 58
Site Preparations

2-1-1 Contents in this chapter

2-1 ‘Overview’ on page 2-2
2-2 ‘General Ultrasound system requirements’ on page 2-3
2-3 ‘Facility needs’ on page 2-12

2-1 Overview

2-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 59
2-2 General Ultrasound system
CAUTION

2-2-1 Contents in this section

2-2-2 ‘Ultrasound system environmental requirements’ on
page 2-3
2-2-3 ‘Electrical requirements’ on page 2-6
2-2-4 ‘EMI limitations’ on page 2-9
2-2-5 ‘Probes environmental requirements’ on page 2-11
2-2-6 ‘Time and manpower requirements’ on page 2-11

General Ultrasound system requirements

requirements

2-2-2 Ultrasound system environmental requirements

2-2-2-1 If the Ultrasound system is very cold or hot

When unpacking the Ultrasound system, allow the temperature of the Ultrasound system to stabilize before powering up. The following table describes guidelines for reaching operational temperatures from storage or transport temperatures.
If the Ultrasound system is very cold or hot, do not turn on its power until it has had a chance to acclimate to its operating environment.
Table 2-1: System Acclimation Time Chart
Degree C -4.5 -2 0.5 3 40 42.5 45 47.5 50 55 60
Degree F 23.9 28.4 32.9 37.4 104 108.5 113 117.5 122 131 140
hours
3
2
1
0 0 1 23 468
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-3 5726264-100 English Rev.8
Page 60
Site Preparations
CAUTION
CAUTION

2-2-2-2 Environmental specifications for Ultrasound system

The system should be operated, stored, or transported within the parameters outlined below. Either its operational environment must be constantly maintained or the unit must be turned off.
Table 2-2: System Environmental Requirements
Operational Storage Transport (<16hrs.)
Temperature 10° - 40°C
50° - 104°F
Humidity 30 - 80%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees
C * 9/5) + 32.
Ensure that the probe face temperature does not exceed the normal operation temperature range.
The LOGIQ V3/V5/V5 Expert system and probe connector is not waterproof. Do not expose the device to water or any kind of liquid.
-5° - 50°C
23° - 122°F
10 - 90% non-condensing
-5° - 50°C
23° - 122°F
10 - 90% non-condensing
2-4 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 61

2-2-2-3 Cooling

NOTE: Each person in the room places an additional 300 BTU/h

2-2-2-4 Lighting

General Ultrasound system requirements
The cooling requirement for a console Ultrasound system with monitor and on board peripherals, is up to 3800 BTU/h. This figure does not include cooling needed for lights, people, or other equipment in the room.
demand on the cooling system.
Bright light is needed for Ultrasound system installation, updates and repairs. However, operator and patient comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source of EMI which could degrade image quality. These controls should be selected to minimize possible interference.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-5 5726264-100 English Rev.8
Page 62
Site Preparations

2-2-3 Electrical requirements

2-2-3-1 General requirements

NOTE: GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the Ultrasound system.
Sites with a mains power system with defined Neutral and Live
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full size ground wire from the distribution panel to the Ultrasound outlet.
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power source to the Ultrasound system is only a conduit.

2-2-3-2 Electrical requirements for the Ultrasound system

In the table below, the electrical specifications for the Ultrasound system includes monitor and on board peripherals.
Table 2-3: Electrical Specifications for LOGIQ V3/V5/V5 Expert system
Vol tag e
100-240 VAC ±10% Max. 300VA 50/ 60HZ
To le r an c e Power Consumption Frequency
:
:

2-2-3-3 Site power outlets

A dedicated AC power outlet must be within reach of the Ultrasound system without extension cords. Other outlets adequate for the external peripherals, medical and test equipment needed to support this Ultrasound system must also be present within 1 m (3.2 ft.) of the Ultrasound system. Electrical installation must meet all current local, state, and national electrical codes.
2-6 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 63

2-2-3-4 Site circuit breaker

CAUTION
It is recommended that the branch circuit breaker for the Ultrasound system be readily accessible.
POWER OUTAGE MAY OCCURE.
The LOGIQ V3/V5/V5 Expert requires a dedicated single branch circuit. To avoid circuit overload and possible loss of critical care equipment, make sure you DO NOT have any other equipment operating on the same circuit.

2-2-3-5 Site power outlets

A dedicated AC power outlet must be within reach of the Ultrasound system without extension cords. Other outlets adequate for the external peripherals, medical and test equipment needed to support this Ultrasound system must also be present within 1 m (3.2 ft.) of the Ultrasound system. Electrical installation must meet all current local, state, and national electrical codes.
General Ultrasound system requirements

2-2-3-6 Unit power plug

If the Ultrasound system arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-7 5726264-100 English Rev.8
Page 64
Site Preparations

2-2-3-7 Power stability requirement

Table 2-4: Power stability requirement
IEC 61000-4-11
Voltage dips, short interruptions and voltage variations on mains supply
< 5%T (> 95% dip) for
0.5 cycle;
40%T (60% dip) for 5 cycles;
70%T (30 dip) for 25 cycles;
< 5%T (>95% dip) for 5 sec
< 5%T (> 95% dip) for
0.5 cycle;
40%T (60% dip) for 5 cycles;
70%T (30 dip) for 25 cycles;
< 5%T (>95% dip) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment.
2-8 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 65

2-2-4 EMI limitations

General Ultrasound system requirements
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. They also generate EMI. The Ultrasound system complies with limits as stated on the EMC label. However there is no guarantee that interference will not occur in a particular installation.
Possible EMI sources should be identified before the Ultrasound system is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Some of these sources include:
• medical lasers
• scanners
• cauterizing guns
• computers
• monitors
• fans
• gel warmers
• microwave ovens
• light dimmers
• mobile phones
in-house wireless phones (DECT phones)
wireless computer keyboard and mouse
air conditioning system
High Frequency (HF) surgery equipment
• general AC/DC adapters
• portable phones
The presence of a broadcast station or broadcast van may also cause interference.
See: 2-2-4-1 ‘EMI prevention/abatement’ on page 2-10 for EMI prevention tips.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-9 5726264-100 English Rev.8
Page 66
Site Preparations

2-2-4-1 EMI prevention/abatement

Table 2-5: EMI prevention/abatement
EMI RULE DETAILS
Be aware of Radio Frequency sources
Ground the Ultrasound system
Replace all screws, Radio Frequency gaskets, covers, cores
Replace broken Radio Frequency gaskets
Do not place labels where Radio Frequency gaskets touch metal
Use GE specified harnesses and peripherals
Take care with cellular phones
Properly route peripheral cables
• Keep the Ultrasound system at least 5 meters (15 feet) away from other EMI sources.
• Special shielding may be required to eliminate interference problems caused by high frequency, high powered radio or video broadcast signals.
Poor grounding is the most likely reason an Ultrasound system will have noisy images. Check grounding of the power cord and power outlet.
• After you finish repairing or updating the Ultrasound system, replace all covers and tighten all screws.
• Any cable with an external connection requires a magnet wrap at each end.
• Install all covers. Loose or missing covers or Radio Frequency gaskets allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an Radio Frequency gasket are broken, replace the gasket. Do not turn on the Ultrasound system until any loose metallic part is removed.
Where applicable, never place a label where Radio Frequency gaskets meet the Ultrasound system. Otherwise, the gap created will permit Radio Frequency leakage. Or, if a label has been found in such a position, move the label.
The interconnect cables are grounded and require ferrite beads and other shielding. Also, cable length, material, and routing are all important; do not change from what is specified.
Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Where applicable, do not allow cables to lie across the top of the Card Rack or hang out of the peripheral bays. Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor cables to the frame.
2-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 67
General Ultrasound system requirements
CAUTION
CAUTION
CAUTION

2-2-5 Probes environmental requirements

2-2-5-1 Operation, storage and transport temperatures for probes

Probes should be operated, stored, or transported within the parameters outlined below.
Ensure that the probe face temperature does not exceed the normal operation temperature range.
Table 2-6: Probe Environmental Requirements
Operational Storage Transport
Temperature 3° - 40° C
38° - 104° F
Humidity 30- 80%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
-5° - 50° C 23° - 122° F
10 - 90% non-condensing
-5° - 50° C 23° - 122° F
10 - 90% non-condensing
Check the room temperature before you use the probe.
Ensure that the probe face temperature does not exceed the normal operation temperature range.
NOTE: Refer to Table 2-1 on page 2-3 to determine the needed
settlement time.

2-2-6 Time and manpower requirements

Site preparation takes time. Begin site preparation checks as soon as possible, if possible, six weeks before delivery, to allow enough time to make any changes.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-11 5726264-100 English Rev.8
Page 68
Site Preparations

2-3-1 Contents in this section

2-3-2 ‘Purchaser responsibilities’ on page 2-13
2-3-3 ‘Required facility needs’ on page 2-14
2-3-4 ‘Desirable features’ on page 2-15
2-3-5 ‘Minimal floor plan suggestion’ on page 2-16
2-3-6 ‘Recommended floor plan suggestion’ on page 2-17
2-3-7 ‘Networking setup requirements’ on page 2-18

2-3 Facility needs

2-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 69

2-3-2 Purchaser responsibilities

The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre-installation work before delivery. Purchaser responsibility includes:
Procuring the materials required
Completing the preparations before delivery of the Ultrasound system
Paying the costs for any alterations and modifications not specifically provided in the sales contract
NOTE: All electrical installations that are preliminary to the positioning
of the equipment at the site prepared for the equipment must be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing must also be performed by qualified personnel. The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
Facility needs
The desire to use a non–listed or customer provided product or to place an approved product further from the Ultrasound system than the interface kit allows, presents challenges to the installation team. To avoid delays during installation, such variances should be made known to the individuals or group performing the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the Ultrasound system. Carpet is not recommended because it collects dust and creates static. Potential sources of EMI (electromagnetic interference) should also be investigated before delivery. Dirt, static, and EMI can negatively impact Ultrasound system reliability.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-13 5726264-100 English Rev.8
Page 70
Site Preparations

2-3-3 Required facility needs

NOTE: GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the Ultrasound system.
Sites with a mains power system with defined Neutral and Live
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full size ground wire from the distribution panel to the Ultrasound outlet.
:
:
2-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 71
2-3-3 Required facility needs(continued)
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power source to the Ultrasound unit is only a conduit.
Dedicated single branch power outlet of adequate amperage meeting all local and national codes which is located less than 2.5 m (8 ft.) from the unit’s proposed location
Door opening is at least 76 cm (30 in) wide
Proposed location for unit is at least 0.5m (1.5 ft.) from the wall for cooling
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with peripheral within 1 m of the unit to connect cables.
Power outlets for other medical equipment.
Power outlets for test equipment within 1 m (3.2 ft.) of Ultrasound system.
Clean and protected space to store probes (in their cases or on a rack)
Material to safely clean probes (done with a plastic container, never metal)
Facility needs
For the amperage requirements, see: 2-2-3 ‘Electrical
requirements’ on page 2-6.

2-3-4 Desirable features

Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe sheaths
• Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and proprietary manuals
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-15 5726264-100 English Rev.8
Page 72
Site Preparations
4
1
3
11
6
13
8
10
5
7
9
12
14
2

2-3-5 Minimal floor plan suggestion

CSI 8x10
Scale: Each square equals one square foot (app. 31 x 31 cm)
1. Door – at least 762 mm (30 inches)
2. Film Viewer
3. Counter Top, Sink with hot and cold water and Supplies Storage
4. Linen Supply
5. Probes/Supplies
6. Examination Table – 1930 x 610 mm (76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound system
10. External Peripherals
11. Dedicated Power Outlet ­Circuit Breaker protected and easily accessible
12. Network Interface
13. 457 mm (18 inches) distance of Ultrasound system from wall or objects
14. GE Cabinet for Software and Manuals
Figure 2-1. Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
2-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 73

2-3-6 Recommended floor plan suggestion

CSI 14x17
Facility needs
10
11
12
15
9
2
3
4
5
13 14
16
8
7
1
17
6
18
Scale: Each square equals one square foot (app. 31 x 31 cm)
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold water
6. Overhead Lights Dimmer ­Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet ­Circuit Breaker protected and easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of Ultrasound system from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
Figure 2-2. A 14 by 17 foot recommended floor plan
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-17 5726264-100 English Rev.8
Page 74
Site Preparations

2-3-7 Networking setup requirements

2-3-7-1 Stand alone Ultrasound system (without network connection)

None.

2-3-7-2 Scanner connected to hospital’s network

Supported networks:
100/1000 Mbit Ethernet/DICOM network (option)

2-3-7-3 InSite requirements

InSite requires an Ethernet connection via:
100/1000 Mbit Interface

2-3-7-4 Purpose of the DICOM network function

DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
As an added benefit, transferring images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while scanning continues.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 75

2-3-7-5 DICOM option setup requirements

To configure the Ultrasound system to work with other network connections, the site’s network administrator must provide information to complete the form “Worksheet for DICOM Network Information”. Ensure that there are no spaces in any field of the form.
See:
Entries must include:
A host name, local port number, AE Title, IP address and Net Mask for the Ultrasound system.
The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
The host name, IP address, port and AE Title for each device the site wants connected to the Ultrasound system for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the revision of the device, is also included. This information may be useful for error solving.
Facility needs
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-19 5726264-100 English Rev.8
Page 76
Site Preparations
2-3-7-5 DICOM option setup requirements(continued)
Figure 2-3. Worksheet for DICOM Network Information
2-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 77

Chapter 3

System Setup

This chapter contains information needed to install LOGIQ V3/V5/V5 Expert system.
Included is a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes, and external peripherals for electrical safety are also included in this procedure.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-1 5726264-100 English Rev.8
Page 78
System Setup

3-1-1 Contents in this chapter

3-1 ‘Overview’ on page 3-2
3-2 ‘Setup reminders’ on page 3-3
3-3 ‘Receiving and unpacking the equipment’ on page 3-6
3-4 ‘Preparing for setup’ on page 3-13
3-5 ‘Completing the setup’ on page 3-14
3-6 ‘System Configuration’ on page 3-21
3-7-3 ‘Peripherals Installation Instructions’ on page 3-29
3-9 ‘Paperwork after setup’ on page 3-47

3-1 Overview

3-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 79

3-2-1 Average setup time

DANGER
CAUTION
CAUTION
CAUTION
Unpacking the LOGIQ V3/V5/V5 Expert: 30 minutes
Set up LOGIQ V3/V5/V5 Expert options: 30 minutes
DICOM Network Configuration: 30 minutes or more, depending on the configuration
Install Insite: 0.5 hour
The LOGIQ V3/V5/V5 Expert installation and functional checkout will take approximately 1 hour. LOGIQ V3/V5/V5 Expert consoles with optional equipment may take slightly longer.
Setup reminders

3-2 Setup reminders

3-2-2 Setup warnings

WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE ULTRASOUND SYSTEM!
To prevent electrical shock, connect the unit to a properly grounded power outlet. DO NOT use a three to two prong adapter. This defeats safety grounding.
DO NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present.
DO NOT operate this unit unless all board covers and frame panels are securely in place. System performance and cooling require this.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-3 5726264-100 English Rev.8
Page 80
System Setup
DANGER
CAUTION
3-2-2 Setup warnings(continued)
1. There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servicing.
NOTE: For information regarding packing labels, refer to LABELS
ON PACKAGE.
2. After being transported, the unit may be very cold or hot. If this is the case, allow the unit to acclimate before you turn it on. It requires one hour for each 2.5°C increment it's temperature is below 3°C or above 40°C.
Equipment damage possibility. Turning the system on without acclimation after arriving at site may cause the system to be damaged.
If the Ultrasound system is very cold or hot, do not turn on its power until it has had a chance to acclimate to its operating environment.
Degree C
Degree F
hours
The following table describes guidelines for reaching operational temperatures from storage or transport temperatures.
Table 3-1: System Acclimation Time Chart
50 45 40 35 30 25 20 15 10 5 0 -5
122 113 104 95 86 77 68 59 50 41 23 23
420000000246
3-4 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 81
3-2-2 Setup warnings(continued)
CAUTION
CAUTION
Operator Manual(s)
The User Manual(s) should be fully read and understood before operating the LOGIQ V3/V5/V5 Expert and kept near the Ultrasound system for quick reference.
Acoustic Output Hazard
Although the ultrasound energy transmitted from the LOGIQ V3/V5/V5 Expert probe is within AIUM/NEMA standards, avoid unnecessary exposure. ultrasound energy can produce heat and mechanical damage.
Setup reminders
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-5 5726264-100 English Rev.8
Page 82
System Setup
CAUTION
CAUTION
3-3 Receiving and unpacking the

3-3-1 Purpose of this section

This section describes how to receive and unpack LOGIQ V3/ V5/V5 Expert.

3-3-2 Contents in this section

3-3-3 ‘Warnings for receiving and unpacking’ on page 3-6
3-3-5 ‘Unpacking the LOGIQ V3/V5/V5 Expert’ on page 3-8
equipment

3-3-3 Warnings for receiving and unpacking

Two people are needed to unpack the Ultrasound system because of its weight. Attempts to move the Ultrasound system considerable distances or on an incline by one person could result in injury or damage or both.
Two people are required whenever a part weighing 16 KG (35 LBS) or more must be lifted.
Remember to use relevant personal protecting equipment (PPE) during packing and unpacking. Check with your local EHS representative.
3-6 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 83
Receiving and unpacking the equipment

3-3-4 LOGIQ V3/V5/V5 Expert transportation box label

The LOGIQ V3/V5/V5 Expert transportation box label is located at the front of the transportation box.
Figure 3-1. LOGIQ V3/V5/V5 Expert transportation box label 1
Figure 3-2. LOGIQ V3/V5/V5 Expert transportation box label 2
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-7 5726264-100 English Rev.8
Page 84
System Setup
CAUTION
CAUTION
CAUTION

3-3-5 Unpacking the LOGIQ V3/V5/V5 Expert

When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
Please carefully unpack the system, and do not dispose the package of LOGIQ V3/V5/V5 Expert, so that it can be reused for service.
Do not lift the unit by the Keyboard. Equipment damage may result.
The crate with the LOGIQ V5/LOGIQ V3 weighs approximately 50kg. Be prepared for a sudden shift of weight as the unit is removed from its base (pallet).
NOTE: Please check the LOGIQ V3/V5/V5 Expert console is well
assembly after unpacking the system.
Table 3-2: Unpacking the LOGIQ V3/V5/V5 Expert
Step Description Corresponding Graphic
1 Tear off the "stop open" mark.
3-8 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 85
Receiving and unpacking the equipment
Table 3-2: Unpacking the LOGIQ V3/V5/V5 Expert
Step Description Corresponding Graphic
2 Cut off the two packing straps around the crate.
Note: To avoid injury, hold the strap clasp with one hand when cutting the strap.
3 Remove the top cover.
4 Remove the three plastic locks.
Note: Rotate the inside plastic lock
counterclockwise to remove it and then remove the outside lock.
5 Remove the outside shipping box.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-9 5726264-100 English Rev.8
Page 86
System Setup
Table 3-2: Unpacking the LOGIQ V3/V5/V5 Expert
Step Description Corresponding Graphic
6 Remove the dust bag from the unit.
Note: There is no dust bag if the system is transported by sea. Ignore this step if there is no dust bag.
7 Remove the clear plastic (wrapped around the
LOGIQ V3/V5/V5 Expert) from the unit.
8 Cut the packing straps around the four wheels.
3-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 87
Receiving and unpacking the equipment
Table 3-2: Unpacking the LOGIQ V3/V5/V5 Expert
Step Description Corresponding Graphic
9 Remove the 4 foams besides the wheels and
remove the aluminum foil bag from the unit.
Note: To avoid damaging the unit, please use a pair of scissors instead of the knife.
10 Remove the foams beside the LCD monitor.
11 Unlock the wheels, and then hold the control
panel at the front side to move the system until the front two wheels on the ground and then slowly move the whole system on the ground.

3-3-5-1 Moving into Position

Please refer to Basic User Manual on how to move the system.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-11 5726264-100 English Rev.8
Page 88
System Setup

3-3-6 Packing the Equipment

Please pack LOGIQ V3/V5/V5 Expert in the reverse order of unpacking.
3-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 89

3-4 Preparing for setup

3-4-1 Verify customer order

Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered, or damaged.

3-4-2 Physical inspection

Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to ‘Damage in
transportation’ on page i-12 in the beginning of this manual.
Preparing for setup

3-4-3 EMI protection

The LOGIQ V3/V5/V5 Expert has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation.
See 2-2-4 ‘EMI limitations’ on page 2-9 for more information about EMI protection.
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-13 5726264-100 English Rev.8
Page 90
System Setup

3-5 Completing the setup

3-5-1 Purpose of this section

This section describes how to complete the installation of LOGIQ V3/V5/V5 Expert.

3-5-2 Contents in this section

3-5-3 ‘System specifications’ on page 3-14
3-5-4 ‘Electrical specifications’ on page 3-16
3-5-7 ‘Connections on the I/O Rear Panel’ on page 3-20
3-5-8 ‘Connecting probes’ on page 3-20
3-5-9 ‘Powering the system’ on page 3-20

3-5-3 System specifications

3-5-3-1 System requirements verification

Verify that the site meets the requirements listed in Chapter 2.
(See: 2-3 ‘Facility needs’ on page 2-12.)
Verify that the specifications below don’t conflict with any on-site conditions.
3-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 91
Completing the setup

3-5-3-2 Physical dimensions

Table 3-3: Physical dimensions of LOGIQ V3/V5/V5 Expert
Height Width Depth Unit
1344 (system with 15" LCD Monitor) 1399 (system with 17" LCD Monitor) 1415 (system with 19" LCD Monitor)
52.91 (system with 15" LCD Monitor)
55.08 (system with 17" LCD Monitor)
55.71 (system with 19" LCD Monitor)
455 722
17.9 28.43

3-5-3-3 Console Weight

Weight (system with 15" LCD Monitor): approx. 43 kg (95
Weight (system with 17" LCD Monitor): approx. 45 kg (99
Weight (system with 19" LCD Monitor): approx. 48 kg (106
mm
Inches
lbs)
lbs)
lbs)
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-15 5726264-100 English Rev.8
Page 92
System Setup
WARNING

3-5-4 Electrical specifications

Connecting a LOGIQ V3/V5/V5 Expert to the wrong voltage level will most likely destroy it.

3-5-4-1 Verification of the system’s voltage setting

Verify that the mains voltage specified for the LOGIQ V3/V5/V5 Expert is available on-site.
Refer to the latest revision of the Basic User Manual for a full list of product labels prior to serving the system.

3-5-4-2 Electrical specifications for LOGIQ V3/V5/V5 Expert

In the table below, the electrical specifications for LOGIQ V3/V5/ V5 Expert includes monitor and on board peripherals.
Table 3-4: Electrical specifications for LOGIQ V3/V5/V5 Expert
Part
Number
5715379 LOGIQ V3
5718054 LOGIQ V5 EXPERT
Description Voltage Tol era nce s
Power
consumntion
100-240V ±10% Max.300VA 50/60 Hz5715380 LOGIQ V5
Frequenc
y
3-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 93
Completing the setup
WARNING

3-5-5 Connecting Cables

For software Option Installation, the operator must login as Administrator.
Table 3-5: List of Connecting Cables.
Name Part No. Figure NOTE
USB Cable Max.300VA
For USB Printer
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-17 5726264-100 English Rev.8
Page 94
System Setup

3-5-6 Peripherals/Accessories Connector Panel

LOGIQ V5/LOGIQ V3 peripherals and accessories can be properly connected using the side connector panel.
Figure 3-3. Rear Connector Panel
NOTE: The AC printer can be connected to the isolated Printer USB
port on the rear panel only.
1. Printer Dedicated external USB printer connection
2. Dual USB port USB for use
3. Ethernet 100MBase-T for net connectivity
4. HDMI port For video port
5. VGA VGA Video Out
6. Composite Out Composited Video Out
7. S-Video S-Video Out
8. Audio Out Audio Line Out
9. AC Inlet 100-240V
3-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 95

3-5-6-1 Pin assignment for each connector

Table 3-6: Pin Assignments of External VGA
Pin No. Signal Pin No. Signal
1RED9NC
2 GREEN 10 NC
3BLUE11NC
4NC12NC
5NC13HSY
6 GND 14 VSY
7 GND 15 NC
8GND
Table 3-7: Pin Assignments of USB
Pin No. Signal Pin No. Signal
1 +5 VDC 5 +5 VDC
Completing the setup
2DATA6DATA
3DATA7DATA
4GND8GND
Table 3-8: Pin Assignments of Audio
Pin No. Signal Pin No. Signal
1GND4NC
2L+5R+
3 Speaker L 6 Speaker R
Table 3-9: Pin Assignment of S-Video
Pin No. Output Signal Pin No. Output Signal
1GND 3Y
2GND 4C
Table 3-10: Pin Assignment of Composite Video Out
Output
Pin No. Output Signal Pin No.
Signal
1 Composite Out 2 GND
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-19 5726264-100 English Rev.8
Page 96
System Setup

3-5-7 Connections on the I/O Rear Panel

NOTE: Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical equipment). Furthermore, all complete configurations shall comply with the valid version of the system standard IEC60601-1. Everybody who connects additional equipment to the signal input part or signal output part of LOGIQ V3/V5/V5 Expert, configures a medical system, and is therefore responsible that the Ultrasound system complies with the requirements of the valid version of IEC60601-1. If in doubt, consult the technical service department or your local representative for GE.

3-5-7-1 Connect Ethernet

Connect the network cable to the Ethernet connector on the I/O Rear Panel.
The connector is located on the rear side of LOGIQ V3/V5/V5 Expert.

3-5-7-2 Connect USB Flash Drive

NOTE: USB Flash Drive approved for LOGIQ V3/V5/V5 Expert are
verified for EMC performance according to EN55011 class B. The use of any other USB Flash Drive will compromise this verification, and may cause interference on LOGIQ V3/V5/V5 Expert itself, or on other electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Drive in one of the USB ports on the LOGIQ V3/V5/V5 Expert.

3-5-8 Connecting probes

Please refer to Basic User Manual on how to connect/ disconnect a probe.

3-5-9 Powering the system

Please refer to Basic User Manual on how to power the system.
3-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 97

3-6 System Configuration

3-6-1 Purpose of this section

This section describes how to configure the LOGIQ V3/V5/V5 Expert.

3-6-2 LOGIQ V3/V5/V5 Expert configuration

For complete instructions, refer to the latest revision of the LOGIQ V3/V5/V5 Expert Basic User Manual, Chapter 16.
Information includes Entering Location, Adjusting Date and Time, Selecting User interface Language, Selecting Online Manual Language, Selecting Unites of Measure.
System Configuration

3-6-2-1 User Configurable Key installation

When the system is received, please check the software option installed on the system, you can refer to the label on the system’s rear neck for detailed information. The option with check mark enabled means it is installed.
Figure 3-4. Software Option Label
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-21 5726264-100 English Rev.8
Page 98
System Setup
3-6-2-1 User Configurable Key installation(continued)
TVI, LOGIQ View or 3D/4D, If any of them is installed, please replace the blank key cap according to the assigned software function.
Figure 3-5. Replace Key Cap
There is a key reserved for CWD function, replace the blank key cap for CWD option if CWD option installed.
Figure 3-6. CWD Key Cap
3-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Page 99
3-6-2-1 User Configurable Key installation(continued)
After the key cap is replaced, Enter Utility-> System-> User Configurable Key and assign the user defined key.
Figure 3-7. Assign User Configurable Key

3-6-2-2 Multi-language Keyboard Film installation

System Configuration
1. Take out the keyboard film.
Figure 3-8. Keyboard Film
LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-23 5726264-100 English Rev.8
Page 100
System Setup
3-6-2-2 Multi-language Keyboard Film installation(continued)
2. The paster is divided into four parts, peel off two parts of the paster in the middle of the keyboard film.
Figure 3-9. Peel Off the Paster
3. Make sure the keyboard is clean. Align the keyboard film on the keyboard and attach it.
Figure 3-10. Attach The Film
3-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
Loading...