GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
DAMAGE IN TRANSPORTATION
Please check all the packages carefully when delivering the device. If there
is any defect, please write the note “defect during transportation” on all the
copy of freight bill or delivery bill by GE’s agent or the principal of
hospital before deliver the goods. No matter the defect note is added or
not, let the deliveryman know the situation of the device. The freight
company is responsible of the goods within 14 days after receiving. If no
claim is made within this period, the freight company will not pay for
counterclaim.
GE Medical Systems (China) Co., Ltd.
Tel: +86-510-85225888
No.19, Changjiang Road, Wuxi National Hi-Tech
Development Zone,Jiangsu, P.R.C. 214028
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
Modification record
Vers-
Old
Chapter Page Summary Name Date
ion
Rev2 Rev1 1. DAMAGE IN
Numbers
TRANSPORT
ATION
2. CHAPTER 1
1. Page 1
2. Page 2
Update company
name, address, and
telephone number.
Hua Fang Oct. 25,
2006
1
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
6-4-1 Removal of the retainer ring of trackball................................................36
6-4-2 Cleaning the trackball .............................................................................36
6-4-3 Fixing the trackball and retainer ring.....................................................36
4
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
CHAPTER 1 Introduction
1-1 System Overview
LOGIQ A1 is an advanced and high quality ultrasound diagnostic. It is a linear and
convex scanning system with 3.5MHz C60 convex-array probe as standard configuration,
7.5MHz L700 linear array probe and 6.0MHz C14 micro-convex probe as optional ones.
This system can be used for diagnosis of abdomen, OB/GYN and small parts etc.
1-2 Purpose of Service Manual
This Service Manual provides service information for the LOGIQ A1 Ultrasound
Scanning System. It contains the following chapters:
Chapter 1 - Introduction: Contains a content summary and warnings.
Chapter 2 - Installation: Including electric request before installation and LOGIQ A1’s
installation process with whole bills of installation and examination.
Chapter 3 - System Outline: including the description of mechanicalelectric
environment property, peripheries and accessories.
Chapter 4 – Functional Check: including function examination for repairing and
regular maintenance after installation.
Chapter 5 – System Architecture: composing configuration of system.
Chapter 6 – Periodic Maintenance: offer the periodic maintenance process of LOGIQ
A1.
5
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 2 5176463-100
1-3 Safety Considerations
1-3-1 Warning
Read below notes carefully before using the system.
1. Read LOGIQ A1 User Manual before using the system and put it in an easy-to-reach
place.
2. Only the trained person can operate or repair LOGIQ A1. It is necessary to avoid the
unwanted ultrasound radicalization although emissive energy of LOGIQ A1
transducer is within standard limitation. ,
3. Only specified power cord is allowed to avoid shock. Power pin needs to be
connected to fixed socket with ground protection. Commutator is not allowed.
System is required to work within the rating voltage.
4. Probes need to be maintained carefully. It is dangerous to use damaged probe on
patient. Image quality is influenced by damaged probe too. The damaged probe
must be discarded, using repaired probe is forbidden. Do not touch the probe surface
with the speculate things or knock the probe.
5. Do not put any liquid over or on the top of mainframe. There might be current
leakage if electric parts are touched by liquid. Only trained service representative
can open the system mainframe to repair the system. Customer are not allowed to
open the system mainframe.
6. Do not use electric vibrating generator while using LOGIQ A1.
7. Use the machine in the prescriptive condition, otherwise it might disturb radiogram
and TV. In order to avoid this situation, please use the machine according to the
User Manual and Service Manual.
6
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
1-3-1 Warning
Fig1-1 describes the usage and position of safety labels and offers other important
information of this machine.
Label/Symbol DescriptionPosition
Nameplate and electrical
parameter
Type/grade label Indicate the degree of safety and
Equipment authentication
labels
Continued
Table 1-1 Warning labels
• Manufacture and address
• Manufacturing date
• Type and serial number
• Electrical parameter
protection
Indicate to have passed the trialproduce register of national
Back of mainframe
Back of mainframe
Back of mainframe
medicament government.
Type BF deviceSymbol is the
person in the frameindicate
the type B device equipped with
the type F applied parts(usually
to hang some probes)
Class equipment means the
equipment not only has the
basic insulation, but also has the
grounded to provide protection
against the electrical shock.
Power is on. Front panel (on the power
Warning, high voltage
Exterior
Pedestal (inside the
system)
CRT, transformer
switch)
Power is off. Front panel (on the power
switch)
7
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
1-3-2 Classification
Anti-electrical shock type: Class I equipment
Level of anti-shock: Type BF equipment
*1
*2
1. Class I equipment*
This means the equipment not only has the basic level insulation, but also has the
grounding to provide protection against the electrical shock. This additional protection
can prevent the exposed metal to be electrification when insulated malfunction happens.
2. Type BF equipment*
The type B equipment equipped with the Type F applied parts can provide the protection
against the electrical shock. When adding 1.1 times voltage between applied parts and
ground, it will not pass the permissible leakage current in the singularity malfunction
condition.
1-4 EMC (Electromagnetic Compatibility)
1-4-1 EMC performance
The electromagnetic interfere to other device through interspace transmission or the
linked cable depends on the specialities of electric device. EMC glossary shows that the
device has the capability not only to avoid the electromagnetism interfere from other
device, but also has the capability not to affect others because of its own
electromagnetism interfered.
This product is designed strictly according to EMC for medical electric device.
In order to guarantee EMC capability of device, please install the machine according to
service manual. Pay attention to the Note 1-4-2.
If there is something wrong related to EMC, please resolve it according to the section
1-4-2 Notice upon installation of product
1. Use the power cord supplied or appointer by GE. The power pin must connect to the
fixed power socket with ground.
8
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2. Connect three-pin to three-pin socket, do not use the commutator of three-pin to
two-pin.
3. Keep the machine far away from the other electronical-device.
4. Use the cable supplied or appointed by GE and connect it according to the
installation process. For example, separate the power cord from signal cord.
1-4-3 General Notice
1. Only the appointed external devices can connect with this system and cannot affect
EMC capability.
2. Note when doing the modification:
a) Do not change the collocation of this system. As such modification will worsen
EMC capability.
b) Modification of collocation including:
• Modification of cable (length, material, connection and so on.)
• Modification of installation/composition of system
• Modification of configuration/components.
• Modification in the way to fix system/components (turn on or off the cover
board, close the bolt of cover board)
3. Close all the cover boards when operating, if one cover board is on, make it off
before continuing the operation in spite of any reason. It will affect EMC capability
when the cover board is on.
1-4-4 Countermeasures against EMC related-issues
It is usually difficult to resolve EMC problems, it takes a lot of time and cost, and
make the electromagnetism interfere to other device.
1. Keep the system far away from other device, that can decrease electromagnetism
interfere.
2.
Change the relative position between this system to other devices, that can decrease
electromagnetism interfere.
3.
Change the position of other device’s power cord/signal cord, that can decrease
electromagnetism interfere.
4.
Change the power supply route of other device, that can decrease electromagnetism
interfere.
9
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
1-4-5 Notice on service
Close the bolts after repairing, flexible bolts will depress EMC capability.
Note:
The ultrasound system should be used far away from the electricity generator,
cellphone, amphibious transceiver, broadcasting station and electromotion control toy
etc to avoid the capability of LOGIQ A1 to pass the regulation. Turn off power supply
when such device beside LOGIQ A1.
Note:
It is responsible for the service person of LOGIQ A1 to guide the technician who
is related to system, patient and other persons to follow these regulations.
1-5 Contact Information
If this equipment does not work as indicated in this service manual or in the User
Manual, or if you required additional assistance, please contact the local distributor or
appropriate support resource, as listed below.
Prepare the following information before your call:
- System ID serial number.
- Software version.
NEEDS Local Service Contact Info (e.g. 800 number) HERE!!
Contact below system manufacturer for advanced technical support.
GE Medical Systems (China) Co., Ltd.
U/S Technical Support
No.19, Changjiang Road, Wuxi National Hi-Tech Development Zone,
Jiangsu, P.R.C. 214028
Tel: +86-51085225888
Fax: +86-510 – 85226688
10
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
CHAPTER 2 INSTALLATION
2-1 PRE-Installation
2-1-1 Overview
This section provides the information required to plan and prepare for the installation of
a LOGIQ A1. Included are descriptions of the facility and electrical needs to be met by
the purchaser of the unit.
2-1-2 Power line Requirements
Check the following wire’s parameters a week before installation. It is better to use
Drantez Model 605-3 power cord monitor with 101 options.
Parameter Rating
Power requirement AC 220V±22V
Power consumption No greater then 130VA
Power supply frequency 50Hz
Power supply twinkling change Less than 25% of regular peak value, less
than1ms of any instant-change status,
including frequencysynchronization
asynchronism or non-seasonal status’s
change.
Attenuation surge Less than 15% of peak value voltage, less
than 1ms.
2-1-3 Physical Specification
The weight of LOGIQ A1 is about 10Kg
Working Condition
11
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
Table 2-1 Working condition
Conditions
Operation Preservation Transportation
Parameter
Temperature
Humidity
Atmosphere
pressure
5
40-5
30%--80%no
condensation
86kPa106kPa 86kPa106kPa 50kPa106kPa
40-30
Less than80%no
condensation
55
Less than 95%no
condensation
To patient:
It is better for patient to keep the room temperature in the range of 20°C - 26°C(68°F 79°F), as the humidity 50% -70%.
EMI
The radio frequencymagnetic fieldspace twinkling status and electrical net is easy to
affect ultrasound device. The electrical equipment may cause the space
electromagnetism interfere or electrical net electromagnetism interfere. The resource of
the electromagnetism interfere including medical lasercautery electron gun
computer electric fanGel refiner microwave oven and cellphone, and
broadcasterbig truck also.
Read the following defend measurements carefully before installing the device:
1.
Connect all the devices’ power cord to the fixed power socket with ground.
2.
Make sure to connect the place of equipment with the ground of building.
3.
Keep the device far away from other electronic device.
If make sure or doubt that there is something wrong about EMI, deal with the jamming
devices as follows:
1.
Move the devices far away from jamming devices.
2.
Change the position of device in the room.
3.
Connect the power cord to another power socket.
4.
Move the device’s power cord and signal cord.
Make sure to screw down bolts after installing the device after repairing.
12
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2-2 Installation
2-2-1 Overview
This section describes the steps to install LOGIQ A1.
2-2-2 Average installation time
It takes about 1 hour to install GE LOGIQ A1.
2-2-3 Checking the components
Check the device carefully when getting a new one, if there is hard up of parts, contact
through the way in the Chapter 1.
Warning:
Connect LOGIQ A1 to the right grounded power socket to avoid electrical
shock, do not use three-pin to two-pin commutator that will destroy the grounded safety.
Warning:
There is no spare parts of an apparatus for operator. Do not open the outer
frame or panel in order to avoid electrical shock. Only the service person can do the
servicing when there is some wrong with the device except supply socket.
Warning:
Warning:
Make sure the back panel is well ventilated.
Do not hurt the rubber of probe surface, using of damaged probe is
prohibited.
Warning:
Although the emissive ultrasound energy is in the prescriptive range, it is
better to avoid unnecessary ultrasound radiation.
2-2-4 Unpacking LOGIQ A1
1.
Snip the package cingulum around the box.
2.
Remove the adhesive tape.
3.
Open the paper box.
4.
Take out the User manual, probe box, ultrasound gel and power cord.
5.
Take the device out of the package.
6.
Wipe off the plastic bag around device.
13
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2-2-5 Assembling LOGIQ A1
Assembling system
1.
Get the probe out of package and put in into the special probe holder.
2.
Make sure the mainframe off before connecting probe.
3.
Make sure any probe connects to the following socket:
a.
Turn the handle key to the position of 3 o’clock.
b.
Connect the linker of probe to the probe socket of main unit.
c.
Turn the handle key to the position of 12 o’clock.
4.
Connect the inner pin to AC socket in the back panel of main unit, connect another
pin to the hospital power socket with the regular press output, do not use three-pin
to two-pin commutator, that will destroy the grounded safety.
5.
Open the keyboard through the lock on the top of main unit.
Note:
2-2-6 Installation of optional accessories
1.
Do not move or institute mainframe when dropping the keyboard.
Connect VGP to suitable power socket, connect Video out port to Video In of VGP
with BNC-BNC cable in the back panel of LOGIQ A1.
Install the accessories
Illustration 2-1
14
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2-2-7 Adjustment of monitor contrast and brightness
Adjust contrast and brightness according to the brightness of room after turning on the
main unit.
Note:
Check the function in Chapter 4 before deliver the goods to customer.
2-3 Electrical safety tests
The following tests are performed at the factory and should be repeated at the sites as
part of periodic maintenance. These tests are: grounding continuity, classis leakage
current, probe leakage current. All measurements are made with an Electrical Safety
Analyzer Model Dale 600/600E built by Dale Technology Corporation or equipment
device.
Recommended Tool
Table-2-2
Recommended Tool
Part Number Name Description
46-285652G1GE
Electrical Safety Analyzer
For 120V Units
DALE 600
46-328406G2(GE) Electrical Safety Analyzer 600E For 220V Units
2113015(GE) Leakage Current Ultrasound Kit For 120V and 220V Units
2-3-1 Outlet test wiring Arrangement
Test all outlets in the area for proper wiring arrangement by plugging in the Dale
600/600E and noting the combination of lights that are illuminated.
15
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
Combination
of Lights
Condition
Note:
conductor and the Grounding (protective earth) conductor are reversed. If later tests
indicate high leakage currents, this should be suspected as a possible cause and the
No outlet tester can detect the condition where the Neutral (grounded supply)
Correct Wiring
Reverse Polarity
Outlet test
Illustration 2-2
Open Ground
Open Hot
Hot/Ground
Reversed
outlet wiring should be visually inspected.
2-3-2 Ground Continuity
Caution:
may cause electric shock.
Caution:
exposed metal parts of the case (See Illustration 2-3), this should be less than 0.2 ohms.
The patient must not be contacted to the equipment during the test because it
Measure the resistance from the third pin of the attachment plug to the
OHMMETER
Ground continuity test
Illustration 2-3
16
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
Meter Procedures
Follow these steps to test the ground wire resistance.
1. Turn the main unit LOGIQ A1.
2. Plug the unit into the meter, and the meter into the tested AC wall outlet. (Refer
Illustration 2-3)
3. Plug the black chassis cable into the meter’s “CHASSIS” connector and attach the
black chassis cable clamp to an exposed metal part of LOGIQ A1.
4. Set the meter’s “FUNCTION” switch to the RESISTANCE position.
5. Set the meter’s “POLARITY” switch to the OFF (Center) position.
6. Measure the ground wire resistance and keep a record of the results with other hard
copies of Preventive Maintenance (PM) data kept on site.
2-3-3 Chassis Leakage Current Test
Definition
Leakage current is the electrical current that could flow through the patient or
sonographer should a ground wire break. The unit, the probes and all external
peripherals must be tested.
The test verifies the isolation of the power line from the chassis. The testing meter is
connected from accessible metal parts of the case to ground. Measurements should be
made with the unit ON and OFF, with the power line polarity Normal and Reversed.
Record the highest reading of current.
SETUP FOR CHASSIS SOURCE LEAKAGE CURRENT
Illustration 2-4
17
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2-3-3 Chassis Leakage Current Test (Continued)
Meter Procedures
1. Follow these steps to test the unit for leakage current.
2. Turn the main unit of LOGIQ A1 off.
3. Plug the unit into meter, and the meter into the tested AC wall outlet.
GROUND AND CHASSIS LEAKAGE CURRENT TEST
Illustration 2-5
4. Plug the black chassis cable into the meter’s “CHASSIS” connector and attach the
black chassis cable clamp to an exposed metal part of LOGIQ A1. (Refer Illustration
2-4)
5. Set the tester’s “Function” switch to CHASSIS position.
6. Take readings for various locations as mentioned in Table 2-3.
Table 2-3
CHASSIS LEAKAGE CURRENT TEST CONDITIONS
Test Location
1 Mounting screw for probe receptacle
2 Real Panel
3 Mounting screw on the handle
4 Mounting screw for peripheral plugged into unit
18
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
7. Keep a record of the results with other hand copies of PM data kept on site.
2-3-3 Chassis Leakage Current Test (Continued)
Date sheet for Chassis Source Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than
the value shown in Table 2-4 below.
Table 2-3
MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE
CURRENT
Country Normal
Condition
Open Ground Reverse
Polarity
Open Neutral
USA N/A 0.3mA 0.3mA N/A
Others 0.1mA 0.5mA 0.5mA 0.5mA
Caution:
When the meter’s ground switch is OPEN, do not touch the unit to avoid
possible electric shock.
Table 2-4
DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT
Unit
Power
ON NORM OPEN
Tester
POLARIT
Y Switch
Tester
NEUTRA
L Switch
Test 1 Probe
Connector
Test 2 Real
Panel
Test 3 Mounting screw
on the handle
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED
19
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2-3-4 Probe Leakage Current Test
Description
This test measures the current that would flow to the ground from any of the probes
through a patient who is being scanned when he comes into contact with the ground
surfaces.
Measurements should be made with the ground open and closed, with power line
polarity normal and reversed and with the unit OFF and ON. For each combination, the
probe must be active to find the worst case condition.
Illustration 2-6 SET UP for Probe Leakage Current
Note:
Each probe will have some amount of leakage current, dependant on its design.
Small variations in probe leakage currents are normal from probe to probe. Other
variation will result from differences in line voltage and test lead placement.
Meter Procedure using Probe Adapter and Saline Bath
The dale 600/600E provides a method for testing probes independently from the system.
The meter utilizes a probe adapter to apply common test potential to all connector pins.
The probe’s imaging area is immersed in a saline solution along with a grounding probe
from the meter to complete the current path. Saline solution is a mixture of water and
salt. The salt adds a free ion to the water, making it conductive. Normal saline solution is
0.9% salt or 1/2 gram salt per 1 liter of water. If saline is not available, a mixture of 1
quart water with one or more grams of table sale, mixed thoroughly, will substitute.
CAUTION:
To avoid probe damaged and possible electric shock, do not immerse
probes into any liquid beyond the level indicated in the operator’s manual. Do not touch
20
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
the probe, conductive liquid or any part of the unit under test while the ISO TEST switch
is depressed.
Follow these steps to test each probe for leakage current.
1. Turn the main unit of LOGIQ A1 OFF.
2. Plug the probe to be tested to the meter’s appropriate adapter. (Refer Illustration 2-
6)
3. Connect the probe to be tested to the meter’s appropriate adapter. (Refer Illustration
2-6)
4. Plug the saline probe into the meter’s CHASSIS connector. (Refer Illustration 2-6)
TRANSDUCER SOURCE LEAKAGE CURRENT TEST
Illustration 2-7
5. Plug the probe adapter into the meter’s connector marked “EXTERNAL”.
6. Set the meter’s “FUNCTION” switch to EXTERNAL position.
7. Add the saline probe and the image area of the probe into the saline bath.
8. Have unit power ON for the first part; turn it OFF for the second half.
9. Depress the ISO TEST rocker switch and record the highest current reading.
10. Follow the test conditions described in TABLE 2-7 for every probe.
11. Keep a record of the results with other hand copies of PM data.
General Procedure
Follow these steps to test each probe for leaking current.
1. Turn the main unit of LOGIQ A1 OFF.
21
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
2. Plug the unit into the test meter, and the meter into the tested AC wall outlet.
3. Plug the external probe into the meter’s EXTERNAL connector.
4. Set the meter’ s FUNCTION switch to EXTERNAL position.
5. Connect the probe for test with the connector of the console.
6. Apply a liberal amount of gel to the probe’s lens.
7. Wrap the probe in aluminium foil, making sure there is good contact with its
acoustic aperture.
8. Either suspend the probe by its cable or place it on an insulated surface.
9. Have unit power ON for the first part; turn it OFF for the second half.
10. Hold the external probe against the aluminium foil on the probe when the unit is
ready to scan.
11. Depress the ISO TEST rocker switch and record when the unit is ready to scan.
12. Follow the test conditions described in the TABLE 2-7 for every probe.
13. Keep a record of the results with other hand copies of PM data.
Data sheet for Transducer Source Leakage Current
Follow the foregoing test procedure for every probe. The test passes when all readings
measure less than the values shown in the table below:
Table 2-6
MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT
Country Normal
Condition
Open
Ground
Reverse
Polarity
Open Neutral
USA 0.05mA 0.05mA 0.05mA 0.05mA
Others 0.1 mA 0.5 mA 0.5 mA 0.5 mA
Table 2-7
DATA SHEET FOR PROBE SOURCE LEAKAGE CURRENT
Probe tested
LOGIQ A1 Unit
Power
ON NORM ON
ON
ON
Test Power
Polarity
NORM CLOSED
REV ON
Tester Ground
22
Measurement
Switch
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
ON
REV CLOSED
CLOSED
CLOSED
CLOSED
CLOSED
NORM
NORM
REV
REV
ON
CLOSED
ON
CLOSED
2-3-5 When there’s too much Leakage Current…
Chassis Fails
Check the ground on the power cord and plug for continuity. Ensure the ground is not
broken, frayed, or intermittent. Replace any defective part.
Tighten all grounds.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet and verify if it is grounded. Notify the user or owner to correct any
deviations. Check the other outlets to see if they could be used instead.
Note:
No outlet tester can detect the condition where the white neutral wire and the
green grounding wire are reversed. If later tests indicate high leakage currents, this
should be suspected as a possible cause and the outlet wiring should be visually
inspected.
Probe Fails
Replace the probe and try again.
Peripherals Fail
Tighten all ground connection screws.
Inspect wiring for bad crimps, poor connections, loose screws or damage.
Still Fails
Begin isolation by removing the probes, external peripherals, then the on board ones, one
at a time while monitoring the leakage current measurement.
New Unit
If the leakage current measurement tests fail on a new unit and if situation cannot be
corrected, submit a Safety Failure Report to document the system problem. Remove the
unit from operation.
23
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
Ball
Contrast
Switch
REV 2 5176463-100
CHAPTER 3 SYSTEM OUTLINE
3-1 Overview
The outline and size of system is described in this chapter.
3
-2
Structure
3-2-1 Structure
Ultrasonic
gel
Monitor
Keyboard
Brightness
Knob
Probe
Track
3-2-2 Size
Knob
Power
Illustration 3- 1 Outline of main unit
450mm×280mm×300mm
24
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
3-2-3 Rear Panel
3-2-4 Control Panel
Illustration3-2 Rear Panel
1. Alphanumeric keys 2. Control key 3. Date and time key
4. Annotation key 5. Patient ID number key 6. Gynecological data
25
table selection key
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
7. Body marks key 8. Frequency steps selection key 9. Image processing selecting key
10. Measuring key 11. Storage control key 12. Set key
24. Total gain selecting key 25. CAPSLOCK key 26. SHIFT key
27. Far gain selecting key 28. Near gain selecting key 29. Clear key
30. Power indicator light
Illustration3-3 Control Panel
3-3 Electric performance
Power cord, tie-in, box, output socket, breaker and switches should be installed before
the mainframe of LOGIQ A1.
3-3-1 Power supply
All countries should install power cord according to their different requirements.
3-3-2 Electrical socket requirement
Separate electrical socket (220VAC) with 5A breaker is strongly recommended. This
kind of socket should adapt to the country’s criterion of power cord. It should be
approved by IEC or other related departments.
3-4 Storage and operating requirements
LOGIQ A1 is transported in separate package. It is about 10kg. Please find Table 3-2 to
see the temperature, atmospheric pressure and humidity range during transportation,
installation and operation.
26
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
Options could be connected to LOGIQ A1 through the linker on the back panel of
system. Linkers are: VIDEO OUT linker and power supply linker and VGA OUT linker.
30 ~ 95 30 ~ 95
27
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
CHAPTER 4 FUNCTIONAL CHECK
4-1 Overview
In this chapter you will learn how to check all the main functions of LOGIQ A1 rapidly.
Please read the user manual of LOGIQ A1 to see how to check the rest functions
4-2 Procedures of Functional Check
4-2-1 Monitor Display after turning on
The test is performed with the linear probe when installing. Check these function again
when repairing the system.
The display of monitor is as follows (Refer Fig 4-1):
1. B mode image
2. Year, Month, Day, Hour, Minute
3. Total Gain (0—99)
4. Near Gain (-30—0)
5. Far Gain (0—30)
6. Focus points
7. Displaying mode and zooming
8. Scanning position
9. Measuring and Body marks
10. Measuring value
11. Probe kind
12. Frequency
13. Linear average
14. Frame average
15. Post-process
16. Depth control
28
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
17. Freeze
18. Body marks
19. Probe position
20. Patient name
21. Gray scale
22. Image annotation
23. Hospital name
24. Level surveyor’s staff
Monitor Display after turning-on
Installation 4-1
4-2-2 Basic control
The basic control is described in the Table 4-1, it is useful to check the system when
installing.
29
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
Table 4-1
BASIC CONTROL
Examination Anticipant value
Connect probe to the linker at the right
of the system’s foundation
Turn supply power on Indicate B mode image.
Turn the knob of Gain Brightness of image changes with the
Adjust Near key Adjust the brightness of Near Gain.
Adjust Far key Adjust the brightness of Far Gain.
Adjust Depth key Adjust the depth of image.
Adjust Reverse key once
Press again
Press M key
Press B/M key
Press B/B key
adjustment.
Indicate the image from left to right
Image changes from right to left.
Indicate M mode image
Indicate B/M mode image( Move the scanning
cursor with trackball
Indicate B/B mode image
Move Trackball
Press Freeze key Freeze image.
Scanning cursor changes with trackball, and
the B/M mode image changes also.
Check the quality of image and capability by KS107BD tissue-imitated ultrasonic
phantom.
30
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
4-3 Diagnosis
4-3-1 Keyboard Test
Keyboard test
All the keys on the keyboard should be tested. In the test, you press the keys to see if
they are working well. (It is an iterative step.)
Trackball Test
When the cursor is shown on the screen, you touch the trackball to make it move on the
screen.
4-3-2 Image Memory Test
The image to be tested should be written on the memorizer and
unscrambled then. We try to see if the unscrambled image effective or not through the
test. If the image on the memorizer is not correctly displayed, it means the test is failed.
4-4 Maintenance and Adjustment
There is no extra spare part for maintenance and adjustment in the system of LOGIQ A1.
When the system has problem, it is recommended to exchange the trouble unit instead of
the spare part. For example, when the power supply has problem, you should exchange
the whole unit of power supply instead of adjusting the voltage-allocater or exchange the
fuse.
31
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
CHAPTER 5 SYSTEM ARCHITECTURE
5-1 Circuit Boards
The table below shows LOGIQ A1 system circuit boards.
Table 5-1 Circuit Boards
Circuit Boards of LOGIQ A1
DSC
TX
RV
Keyboard
Digital Scan Convert board
Transmission board
Receiving board
Keyboard PCB board
32
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
CHAPTER 6 PERIODIC MAINTENANCE
6-1 Overview
It has been determined by engineering that your LOGIQ A1 system does not have any
high wear components that fail with use, therefore no Periodic Maintenance Inspections
are mandatory. Some Customers Quality Assurance Programs may require additional
tasks and or inspections at a different frequency than listed in this manual.
6-2 Periodic Maintenance
Termly maintenance including eyeballingsystem capability examining and termly
cleaning relevant key parts. One time maintenance every six months to insure systemic
capability steady. Maintenance period is evaluated by operating equipment 40 hours
every week.
6-3 Procedure
6-3-1 Visual inspection
1.
Check if the cable is ruptured or fissured and the insulated layer of cable is abraded.
2.
Check if the cable jacket and cable airproof packet / link implement clamp well.
3.
Check if cables connect correctly.
4.
Check if all the bolts are there.
5.
Check if all cable plugs are connected to the corresponding sockets.
6.
Check if all parts are fixed to appropriate location and without damnification.
7.
Check if the keys and trackball on the keyboard work well and without mechanical
damnification.
8.
Check if the probe and probe cable is dilapidated or aging.
Check if the probe is dilapidated.
9.
33
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
10.
Check if has mechanical problems or keyboard problems.
11.
Check the equipment to see if any hardware has missed.
12.
Check the work of fan which is used to decrease temperature, and to ensure the air
circulation.
6-3-2 Cleaning
The surface of all parts can use snug humid soft cloth to clean. It is better to use
litmusless soap, but not alcohol generic liquid dissoluble substance or abrasive.
System
1. Cleaning CRT filtrate screen.
2. Use vacuum cleaner and scrubbing brush to clean fan and the mainframe of LOGIQ
A1.
3. Before cleaning system, turn off the power supply first, and then start to do the
cleaning.
4. Cleaning the place around mainframe and system shell.
5. Dab of the monitor with the soft humid cloth to dab, if it is too dirty, it is better to
use humid cloth with scour, and then soft clean cloth.
6. Dab keyboard with soft humid cloth.
7. Dab keyboard with dry soft cloth.
8. Dab the Video frequency printer with dry soft cloth.
Note:
Do not spill any liquid to get inside system when cleaning monitor, be carefully
do not abrade monitor.
Probe
Clean probe surfacecable and shellexamine probe whether damnification, in order
no liquid coming into probe. If find any dilapidation, do not use that probe before
recovered.
1. Wipe off the ultrasound gel with soft cloth or the running water.
2. Wipe off all the eyeable leftover with snug water soft sponge with litmusless soap
cotton yearn or cloth also. Do not use scour or abrasive to clean probe.
3. Wipe off all the soap leftover with distilled water.
4. Dry the probe with soft cloth or put the probe on the air.
34
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
5. Do not use liquid olefin, petroleum generic coupling liquid or other unapproved
materialsas these materials can damage the probe.
6. Do not put probe into any proportion immerge or dunk in the any material’s
cleanser. Do not use autoclaveultraviolet radiationgas or heat treatment
antisepsis technics to antisepsising probe.
7. When probe is immerged appropriate industry antiseptic, use for clinic medicine
applied registered EPA antiseptic. According to user operation’s note proceeding
chromacontiguity timestockpile and deal with operation of antisepsising
manufacture.
8. It is suggested to use 2% diacid aldehyde liquid without surface activation (for
example Cidex.
9. Do not use alcohol or any liquid like alcohol.
10. After use antiseptic to disinfect, use distilled water to wash all hangover in the
probe.
6-3-3 Testing
1. Make sure that the hospital’s power supply is regular.
2. Insure all power voltage is within the prescribe range. See chapter 4, function of
maintain adjust procedure examine.
3. Carry out electric safety testing, the contents are as follows:
• Hospital power socketTest of fault of power connection
• All commutatorsTest of commutator creep age
• Mainframe and external device (Test of creep age in shell and grounded anti-
resistance)
6-4 Trackball maintenance
Keep the trackball clean to make it work well. If there is dust in, it might work
abnormally or might be damaged.
Tools of trackball maintenance:
1. Trackball tooltools for opening
2.
Cotton stickeralcoholpropyl-alcohol or dry cloth.
35
GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 25176463-100
6-4-1 Removal of the retainer ring of trackball
1. Turn off power supplyand pull out power plug of system
2. Infix one end of convexity substance of trackball to the guiding hole of retainer ring
(2 holes).
3. Turn the trackball according to counter-clockwise slowly.
4. Take retainer ring out of keyboard.
5. Take out the trackball.
6-4-2 Cleaning the trackball
Wipe any oil or dust on the trackball with TFC dry cleaner or dry cloth.
Note:
Note:
the trackball parts with alcoholpropyl alcohol. Don’t use anyother impregnant to
avoid damaging rubber and mechanical parts of trackball.
6-4-3 Fixing the trackball and retainer ring
1. Put in the trackball.
2. Put in the retainer ring.
3. Aim trackball to the guiding hole, then fix retainer ring according to clockwise.
Note:
Note:
person or service person for help.
Keep retainer ring and trackball in certain place to avoid losing.
When clean trackball tee, do not let any liquid into it (system/keyboard). Clean
Don’t use huge strength to fix the retainer ring to avoid damage.
If there is need for maintenance tool or information of trackball, turn to sales’
1. It is suggested to use 2% diacid aldehyde liquid without surface activation (for
example Cidex.
2. Do not use alcohol or any liquid like alcohol.
3. Wipe off any leftover in the probe with distilled water after antisepsising with
antiseptic.
36
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.