State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 2 of 197
GE HEALTHCARE
IRECTION 5305880-100 , REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Your Notes:
Page i-ii -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 3 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
IMPORTANT PRECAUTIONS
LANGUAGE
WARNING
AVERTISSEMENT
•THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
•IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
•DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
•FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.
•CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
•SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
•NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
•LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
•DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
•FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
•VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
•WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER
DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER
SONSTIGE GEFAHREN KOMMEN.
- Page i-i
Approved Document - 5305880-100TPH_r3.pdf Page 4 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
AVISO
ATEN ÇÃO
•ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
•SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
•NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
•LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.
•ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL
EM INGLÊS.
•SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE
TRADUÇÃO.
•NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
•O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
AVVERTENZA
•IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
•SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
•SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
•NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
Page i-ii -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 5 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 6 of 197
- Page i-iii
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent wr ite “Damage
In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical installations that are preliminary to positioning of the equipment at the site prepared for
the equipment shall be performed by licensed electrical contractors. Other connections between
peices of electrical equipment, calibrations and testing shall be performed by qualified GE Medical
Systems personnel. In performing all electrical work on these products, GE will use its own specially
trained filed engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineer’s,
personnel of third party service companies with equivalent training, or licensed electricians ) to
perform electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact
the GE Medical Systems Global Documentation Group with specific information listing the system
type, manual title, part number, revision number, page number and suggestion details. Mail the
information to : Service Documentation, 9900 Innovation Dr ive (RP-2123), Wauwatosa, WI 532 26.
, USA.
GE Medical Systems employees should use the iTrack System to report all documentation errors
or omisions.
Page i-iv -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 7 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part,
without prior written permission of GE Medical Systems.
GE Medical Systems may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 8 of 197
- Page i-v
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Page i-vi -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 9 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Revision History
RevisionDateReason for change
1JAN 26 2008
2APRIL 24 2008
3August 4 2008
Preliminary Release.
Changes incorporated for correcting the SPR’s raised against DOC0421436.
Changes incorporated on contact details and added additional information based on service
teams requirements.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 10 of 197
- Page i-v
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Page i-vi -
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 11 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
3-5-2Installation of Optional Accessories ............................................................................... 3-7
3-5-3Installation of Two Probe Adapter & LOGIQ a200 Probe Adapter (Optional) 3-9
3-5-4Installation of LOGIQ™ 100 PRO with Global or Modular Trolley 3-11
3-5-5Adjustment of Monitor Brightness & Contrast 3-17
3-5-6Adjustment of System Clock/Date, Hospital Name & OB Version ............................... 3-17
Section 3-6 System Configuration............................................................... 3-17
10-6-6-2Generic Procedure for Leakage current .......................................... ... .... ... ... 10-19
10-6-6-3Meter Procedure Using Dale Meter to measure leakage current ................. 10-19
Meter Procedure using Probe Adapter to measure Probe isolation(Sink) Current 21
Section 10-7 When There's Too Much Leakage Current......................... 10-23
Page TOC-6
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 17 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Chapter 1 Introduction
Section 1-1Overview
1-1-1Overview
The LOGIQ™ 100 PRO is a compact portable ultrasound scanner which is designed for OB/GYN,
Abdomen, Urology and small part Scans using the convex, linear and micro-convex (Tr ansvaginal)
probes. High quality images can be obtained by the proper selection of scan control parameter s.
The diagnostic clarity is further enhanced by the different measurement and calculation packages
available in the system.
1-1-2Purpose of Chapter 1
This chapter describes important issues related to safely servicing this ultrasound machine. The
service provider must read and understand all the information presented here before installing or
servicing a unit.
This Service Manual provides installation and service information for the LOGIQ™ 100 PRO
Ultrasound Scanning System and contains the following chapters:
1.) Chapter 1, Introduction: Contains a content summary and warnings.
2.) Chapter 2, Pre-Installation: Contains any pre-installation requirements for the
LOGIQ™ 100 PRO.
3.) Chapter 3, Installation: Contains installation procedure with installation checklist.
4.) Chapter 4, Functional Checks: Contains functional checks that must be performed as
part of the installation, or as required during servicing and periodic maintenance.
5.) Chapter 5, Theory: Contains block diagrams and functional explanations of the
electronics.
6.) Chapter 6, Service Adjustments: Contains instructions on how to make any available
adjustments to the LOGIQ™ 100 PRO.
7.) Chapter 7, Diagnostics/Troubleshooting: Provides procedures for running and
diagnostic or related routines for the LOGIQ™ 100 PRO.
8.) Chapter 8, Replacement Procedures: Provides disassembly procedure s and
reassembly procedures for all changeable FRU.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 18 of 197
Chapter 1 - IntroductionPage 1-1
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
9.) Chapter 9, Replacement Parts: Contains a complete list of replacement parts for the
LOGIQ™ 100 PRO.
10.)Chapter 10, Quality Assurance: Provides periodic maintenance procedures for the
LOGIQ™ 100 PRO.
•Architects (Some parts of the Pre Installation Chapter)
1-1-5LOGIQ™ 100 PRO Models Covered by this Manual
Hcat NumberDescriptionComments
H41072LA,H41072LB,
H41072LC,H41072LD,
LOGIQ™ 100 PRO, 100V -
230V
H41072LE,H41072LF,
H41072LS
Table 1-2 LOGIQ™ 100 PRO Models Covered in this Manual
1-1-6Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before oper ating the LOGIQ™ 100
PRO and also kept near the unit for quick reference.
Page 1-2Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 19 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 1-2 Safety
1-2-1Warnings
WARNING CAREFULLY READ ALL THE WARNINGS BELOW
1.) The operator manual should be read and understood before op erating the LOGIQ™ 100
PRO and kept nearby for quick reference.
2.) Although the ultrasound energy transmitted from the LOGIQ™ 100 PRO transducer is
within AIUM/NEMA standards, unnecessary exposure should be avoided. Only trained
personnel should operate the LOGIQ™ 100 PRO.
3.) To prevent electric shock, the LOGIQ™ 100 PRO should be connected to a properly
grounded power receptacle.Do not use a three prong to two prong adapter. This defeats
safety grounding/
4.) Probes are fragile, Please handle with care.
5.) Concerning outside markings, refer to Figure 1-2 through 1-3.
6.) For the cleaning, disinfection, and sterilization, refer to Probe section in LOGIQ™ 100
PRO User Manual and Caution Sheet supplied with each probe.
NOTICE This m edical equipment is approved, in terms of the prevention of radi o wave interference,
to be used in hospitals, clinics and other institutions which are environmentally qualified. The use
of this equipment is an inappropriate environment may cause some electronic inte rference to radios
and televisions around the eqiupment. Proper handling of this equipment is required in order to
avoid such trouble according to the operator and service manuals. This eq iupments can be used in
residential areas only under the supervision of physicians and qualified technicians.
CAUTION Improper performance possibility. Do not use the following devices near this
equipment. Cellular phone, radio transciever, mobi le radio transmitter, radio-cont rolled toy,
etc. Use of these devices near this equipment c ould cause this equipment to perform outside
the published specifications. Keep power to these devices turned off when near this
equipment.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 20 of 197
Chapter 1 - IntroductionPage 1-3
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 1-3 Important Conventions
1-3-1Conventions Used in Book
Model Designations.
This manual covers the LOGIQ™ 100 PRO scanners.
(See “LOGIQ™ 100 PRO Models Covered by this Manual” on page 1-2.)
Icons.
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages.
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the
precautionary message. Known or potential hazards are labeled in one of three ways:
DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL
DANGER
WARNING
CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE
IGNORED.
WARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT CAN CAUSE
SEVERE PERSONAL INJURY OR PROPERTY DAMAGE IF INSTRUCTIONS ARE
IGNORED.
CAUTION
NOTICE
Caution is used to indicate the presence of a hazard that will or c an cause minor personal injury
and property damage if instructions are ignored.
Equipment Damage Possible
Notice is used when a hazard is present that can cause property damage but has absolute ly no personal
injury risk.
NOTE:Notes provide important information about an item or a procedure.
Information contained in a NOTE can often save you time or effort.
Page 1-4Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 21 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-2Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within a triangle,
as seen throughout this chapter. In addition to text, several different g raphical icons (symbols) may
be used to make you aware of specific types of hazards that could cause harm.
ELECTRICALMECHANICALRADIATION
LASERHEATPINCH
LASER
LIGHT
Table 1-3 Standard Hazard Icons
Other hazard icons make you aware of specific procedures that should be followed.
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
Table 1-4 Standard Icons Indicating a Special Procedure to be Used
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 22 of 197
Chapter 1 - IntroductionPage 1-5
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-3Product Icons
The following table describes the purpose and location of safety labels and other important
information provided on the equipment.
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
Identification and Rating PlateManufacturer's name and
address
Date of manufacture
Model and serial numbers
Electrical ratings
Type/Class LabelUsed to indicate the degree of
safety or protection.
IP Code (IPX8)Indicates the degree of
protection provided by the
enclosure per IEC 529. IPX8
indicates drip proof.
Equipment Type BF (man in
the box symbol) IEC 878-0203 indicates B Type
equipment having a floating
applied part.
Equipment Type CF (heart in
the box symbol) IEC 878-0205 indicates equipment
having a floating applied part
having a degree of protection
suitable for direct cardiac
contact.
Rear of console near power
inlet
Footswitch
Probe connectors and PCG
connector
ECG connector and surgical
probes
Device Listing/Certification
Labels
Laboratory logo or labels
denoting conformance with
industry safety standards
such as UL or IEC.
CAUTION - This unit
weighs...Special care must be
used to avoid..."
This precaution is intended to
prevent injury that may result
if one person attempt to move
the unit considerable
distances or on an incline due
to the weight of the unit.
"DANGER - Risk of explosion
used in..."
The system is not designed
for use with flammable
anesthetic gases.
Table 1-5 Product Icons
Page 1-6Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 23 of 197
Rear of console
On the console where easily
seen during transport?
Rear of console
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
"CAUTION" The equilateral
triangle is usually used in
combination with other
symbols to advise or warn the
user.
ATTENTION - Consult
accompanying documents " is
intended to alert the user to
refer to the operator manual or
other instructions when
complete information cannot
be provided on the label.
"CAUTION - Dangerous
voltage" (the lightning flash
with arrowhead in equilateral
triangle) is used to indicate
electric shock hazards.
"Mains OFF" Indicates the
power off position of the mains
power switch.
Various
Various
Various
Rear of system adjacent to
mains switch
Table 1-5 Product Icons
"OFF/Standby" Indicates the
power off/standby position of
the power switch.
CAUTION
This Power Switch DOES
NOT ISOLATE Mains
Supply
"Mains ON" Indicates the
power on position of the mains
power switch.
"ON" Indicates the power on
position of the power switch.
CAUTION
This Power Switch DOES
NOT ISOLATE Mains
Supply
"Protective Earth" Indicates
the protective earth
(grounding) terminal.
Adjacent to On-Off/Standby
Switch
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 24 of 197
Chapter 1 - IntroductionPage 1-7
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
"Equipotentiality" Indicates
the terminal to be used for
connecting equipotential
conductors when
interconnecting (grounding)
with other equipment.
"Non-Ionizing Radiation"
indicates that the system
applies RF energy.
This symbol indicates that
waste electrical and electronic
equipment must not be
disposed of as unsorted
municipal waste and must be
collected seperately. Please
contact an authorised
representative for
manufacturer for information
concerning the
decompositioning of your
requirement.
Rear of console
Rear of console near power
inlet
Rear Panel.
Indicates the presence of
hazardous substance(s)
above the maximum
concentration value.
Maximum concentration
values for electronic
information products, as set
by the People’s Republic of
China Electronic Industry
Standard SJ/T11364-2006,
include the hazardous
substances of lead, mercury,
hexavalent chromium,
cadmium, polybrominated
biphenyl (PBB), and
polybrominated diphenyl
ether (PBDE). "10" indicates
the number of years during
which the hazardous
substance(s) will not leak or
mutate so that the use of this
product will not result in any
severe environmental
pollution, bodily injury, or
damage to any assets.
Rear Panel.
Table 1-5 Product Icons
Page 1-8Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 25 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
Table 1-5 Product Icons
No hazardous substance,
above the maximum
concentration value, are
present. Maximum
concentration values for
electronic information
products, as set by the
People’s Republic of China
Electronic Industry Standard
SJ/T11364-2006, include the
hazardous substances of
lead, mercury, hexavalent
chromium, cadmium,
polybrominated biphenyl
(PBB), and polybrominated
diphenyl ether (PDBE).
Rear Panel.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 26 of 197
Chapter 1 - IntroductionPage 1-9
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 1-4Safety Considerations
1-4-1Introduction
The following safety precautions must be observed during all phases of operation, service and
repair of this equipment. Failure to comply with these precautions or with specific warnings
elsewhere in this manual, violates safety standards of design, manu facture and intended use o f the
equipment.
WARNING CARE FULLY READ ALL THE WARNINGS LISTED BELOW.
1.)Read the LOGIQ™ 100 PRO Operator Manual thoroughly before operating the
system and keep at hand for ready reference.
2.)Although the ultrasound energy transmitted from the LOGIQ™ 100 PRO
transducer is within AIUM/NEMA standards, unnecessary exposure should be
avoided. Only trained personnel should operate/service the LOGIQ™ 100
PRO.
3.)To avoid electrical shock, use only the supplied power cords and con nect them
to properly grounded power Sock et. Do not use a three pin to two p in adapter.
This defeats the purpose of safety groundin g. Syste m sh ou ld be oper at ed
within the voltage limits.
4.)Probes are fragile, please handle with care. A damaged probe may cause an electrically hazardous condition when coupled to the human body. A
damaged probe will not produce a desirable image. A damaged probe has to
be discarded. It cannot be repaired or reused. Do not allow the lens to come
into contact with a sharp object or to be knocked against an object.
5.)Do not place liquids on or above the console. If the liquid spills, it may come in
to contact with live parts and can cause an electric shock.This system
contains no operator serviceabl e com p on e nt s. To pr ev en t sho ck , do not re
move any covers or panels. Should problems or malfunctions occur, unplug
the power cord. Only qualified Service personnel should service the system.
Accidentally coming in contact with the electrical circuits inside the housing
could cause serious injury.
6.)Do not use Defibrillators when the LOGIQ™ 100 PRO is operated.
7.)Concerning outside markings, refer to ILLUSTRATION 1-1 .
8.)The LOGIQ™ 100 PRO system should not be placed on a soft surface, as it
prevents proper air circulation. The vents for air circulation are on the bottom
cover.
Note: This medical equipment is approved, in terms of the prevention of radio wave interferenc e, to be
used in hospitals, clinics and other institutions which are environmentally qualified. The use of this
equip ment in an inappropriate environment may cause some electronic interference to radios and
televi sions around the equipment. Proper handling of this equipment is required in order to avoid
such trouble according to the operator and service manuals.
1-4-2Human Safety
Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in a LOGIQ™ 100 PRO Training Seminar are au thorize d to
service the equipment.
Page 1-10Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 27 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-4-3Mechanical Safety
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
WARNING
WARNING
CAUTION
CAUTION
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT
MAY BE DAMAGED.
Always lock the Control Console in its parking (locked) position before moving the scanner
around.
The LOGIQ™ 100 PRO weights 10kg or more, depending on installed peripherals, (309 lbs, or
more) when ready for use. Care must be used when moving it or replacing its parts. Failure to
follow the precautions listed below could result in injury, uncontrolled motion and costly
damage.
NOTE:Special care should be taken when transporting the unit in a vehicle:
•Secure the unit in an upright position.
•Lock the wheels (brake)
•DO NOT use the Control Panel as an anchor point.
•Place the probes in their carrying case.
1-4-4Electrical Safety
To minimize shock hazard, the equipment chassis must be co nnected to an electrical groun d. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension cord is used with the system, make sure that the
total current rating of the system does not exceed the extension cord rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 28 of 197
Chapter 1 - IntroductionPage 1-11
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-4-5Labels Locations
GHOST Label
Figure 1-1 Warning Lables
Page 1-12Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 29 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-5 Labels Locations (cont’d)
Figure 1-2 Warning Lables
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 30 of 197
Chapter 1 - IntroductionPage 1-13
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-5 Labels Locations (cont’d)
Figure 1-3 Warning Labels
Page 1-14Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 31 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-5 Labels Locations (cont’d)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 32 of 197
Figure 1-4 Warning Labels
Chapter 1 - IntroductionPage 1-15
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-5 Labels Locations (cont’d)
CAUTION
!
Do not use the following devices
near this equipment.
Cellular phone, radio transceiver,
mobile radio transmitter,
radio-controller toy, etc.
Use of these devices could cause
this equipment to perform outside
the published specifications.
Keep power to these devices
turned off when near this
equipment.
AT TENTION
!
Ne pas utiliser les appareils suivants à
proximité de cet équipement : téléphones
mobiles, émetteurs-récepteurs, appareils
radiocommandés, etc., cela pouvant
entraîner des performances différentes des
spécifications annoncées. Maintenir les
appareils de ce type hors tension lorsqu'ils
se trouvent à proximité du système.
VORSICHT
AT TENZIONE
!
Non usare i seguenti apparecchi in prossimità di questa apparecchiatura:
cellulari, ricetrasmettitore radio, trasmettitore radio mobile, giocattoli radiocomandati ecc.
L'uso di questi apparecchi potrebbe modificare le specifiche dell'apparecchiatura.
Mantenere gli apparecchi interferenti
spenti, se in prossimità del sistema.
AVISO
VORSICHT
!
Verwenden Sie folgende Geräte nicht in
der Nähe der Anlage:
Funktelefon, Funkempfänger, mobiler
Funksender,funkgesteuertes Spielzeug
usw.
Bei Gebrauch dieser Geräte arbeitet die
Anlage evtl. nicht entsprechend der
veröffentlichten Angaben.
Belassen Sie diese Geräte in der Nähe
der Anlage ausgeschaltet.
CUIDADO
!
Não utilize os dispositivos seguintes perto deste equipamento:
Telefonecelular, radiotransceiver, radiotransmissor móvel, brinquedos de controle remoto, etc.
O uso destes dispositivos pode levar o
equipamento a ter um desempenho outro que aquele publicado.
Desligue estes dispositivos quando estiverem perto deste equipamento.
Teléfonoscelulares, receptores de radio,
juguetes controlados por transmisores
de radio móviles, etc.
El uso de estos dispositivos puede
hacer que el equipo funcione fuera de
las especificaciones publicadas.
Mantenga apagados dichos dispositivoscuando estén cerca del sistema.
Figure 1-5 Warning Labels
Page 1-16Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 33 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-3-5 Labels Locations (cont’d)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 34 of 197
Figure 1-6 Warning Lables
Chapter 1 - IntroductionPage 1-17
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-4-6Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
DANGER
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
EXPLOSION WARNING: DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE
WARNING
ATMOSPHERE. OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN
ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT: BECAUSE OF THE DANGER
WARNING
OF INTERDICTING ADDITIONAL HAZARDS, DO NOT INSTALL SUBSTITUTE PARTS OR
PERFORM ANY UNAUTHORIZED MODIFICATION OF THE EQUIPMENT.
1-4-7Lockout/Tagout Requirements (For USA Only)
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug.
1-4-8Classification
Type of protection against electric shock: Class I EQUIPMENT *1
Degree of protection against electric shock: Type BF EQUIPMENT**2
1.) *Class I EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC
INSULATION only, but which includes an additional safety precaution in that means
are provided for the connection of the EQUIPMENT to the protective earth conductor
in the fixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS
cannot become LIVE in the event of a failure of the BASIC INSULATION.
2.) **Type BF EQUIPMENT*
Type B EQUIPMENT with a F-TYPE isolated applied part providing a degree of
protection against electric shock to such a degree that the allowable PATIENT
LEAKAGE CURRENT under SINGLE FAULT CONDITIONS is not exceeded when
1.1 times the highest rated MAINS VOLTAGE is applied between the APPLIED PART
and earth.
Page 1-18Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 35 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 1-5 EMC, EMI, and ESD
1-5-1Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance of a device within its
electromagnetic environment. This environment consists of the device itself and its surroundings
including other equipment, power sources and persons with which the device must interface.
Inadequate compatibility results when a susceptible device fails to perform as intended due
interference from its environment or when the device produces unacceptable levels of emission to
its environment. This interference is often referred to as radio–freq uency or electromagnetic
interference (RFI/EMI) and can be radiated through space or condu cted over interconnecting power
of signal cables. In addition to electromagnetic energy, EMC also includes possible effects from
electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power
supply.
1-5-2CE Compliance
The LOGIQ™ 100 PRO unit conforms to all applicable conducted and radiated emission limits
and to immunity from electrostatic discharge, radiated and conducte d RF fields, magnetic fields and
power line transient requirements.
Applicable standards are: 47CFR Part 18, IEC 601–1–2, and 806–13.
For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps,
are in good condition, installed tightly without skew or stress. Proper installation f ollowing
all comments noted in this service manual is required in order to achieve full EMC
performance.
1-5-3EMC Performance
1-5-4Notice Upon Installation of Product
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
All types of electronic equipment may characteristically cause electromagnetic interference with
other equipment, ei ther through air or connecting cables. The term EMC (Electromagnetic
Compatibility) indicates capability of the equip ment, which curbs electromagnetic influence from
other equipment and at the same time does not affect other equip men t with similar electromagnetic
radiation from itself.
This product is designed to fully comply with the EN60601-1-2 (IEC601-1-2), in Medical electrical
equipment EMC regulations.
Proper installation following this service manual is required in order to achieve the full EMC
performance of the product.
The product must be installed as stipulated in 1-4-4, Notice upon Installation of Product.
In case of issues related to EMC, please follow proced ure s s tat ed in 1-4-6 , Co un te rm e asur es
against EMC-related Issues.
1.) Use either power supply cords provided by GEMS or ones approved by GEMS. Products
equipped with power source plug sho uld be plugged into th e fixed power socket which has the
protective grounding conductor
2.) Connect a three-pin plug to a three-pin socket without using a three-pin-to-two-pin converter.
3.) Keep the equipment as far as possible from other electronic equipment.
4.) Be sure to use the only cables provided by GEMS or ones approved by Wipro GE. Install the
unit/peripherals as per the installation procedure provided in Chapter 3 - INSTALLATION. The
layout of the LOGIQ™ 100 PRO & other peripherals should be as per installation procedures
described in Chapter 3 - INSTALLATION.
Chapter 1 - IntroductionPage 1-19
Approved Document - 5305880-100TPH_r3.pdf Page 36 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-5-5General Notice
1.) Designation of Peripheral Equipment Connectable to this Product. The peripheral which
conforms to EN60601-1-2 (IEC601-1-2), can be connected up to the LOGIQ™ 100 PRO
without compromising its EMC performance. Failure to comply with the above instruction may
result in poor EMC performance of the product.
2.) Notice against User Modification: Do not modify this product. Unilateral user modification
may cause degradation in EMC performance. Modification of the product includes:
a.) Changes in cables (length, material, wiring etc.)
b.) Changes in system installation/layout
c.) Changes in system configuration/components
d.) Changes in means of fixing system/parts (cover open/close, cover screwing)
3.) Operate the system with all covers closed. If you open any cover for some reason, be sure to
shut it before starting/ resuming operation. Operating the system with any cover open may
affect EMC performance.
1-5-6Countermeasures against EMC related Issues
Generally it is very difficult to handle with issues related to EMC. It is time consuming and
expensive. General counter measures for Electromagnetic interference with other equipment
1.) Electromagnetic interference may be reduced by positioning other equipment far away from
the system.
2.) Electromagnetic interference may be reduced by changing the relative location (installation
angle) of the system and other equipment.
3.) Electromagnetic interference may be reduced by changing wiring locations of power/signal
cables of other equip ment.
4.) Electromagnetic influence may be reduced by altering the path of power supply for oth er
equipment.
1-5-7Notice on Service
Ensure all screws are tight after servicing. Loose screws may cause degradation in EMC
performance.
CAUTION Do not use the following devices near the LOGIQ™ 100 PRO system. Devices which
intrinsically transmit radio waves such as cellular phone, radio transceiver, mobile radio
transmitter radio-controlled toy, etc. Use of the se devices could cause the LOGIQ™ 100 PRO
system to perform outside the published specif ications. Keep power to these devices turned
OFF when near the system. Medical staff in charge of the LOGIQ™ 100 PRO system is
required to instruct technicians, p atie nts and other people who may be around the system
to fully comply with the above regulation
Page 1-20Section 1-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 37 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
1-5-8Electrostatic Discharge (ESD) Prevention
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
WARNING
1.)Always connect yourself, via an arm-wrist strap, to the advised esd
connection point located on the rear of the scanner (to the right of the
power connector).
2.)Follow general guide lined for handling of electrostatic sensitive
equipment.
Section 1-6Customer Assistance
This system is not repairable by the customer. If this equipment does not work as indicated in the
Operator's Manual, please contact your service support center. If the service engineer needs
additional information to repair this equip ment, please contact any relevant addresses given below:
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 39 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Chapter 2 Pre Installation
Section 2-1Overview
2-1-1Purpose of Chapter 2
This chapter provides the information re qu ir ed to plan and prepare for the installation of a
LOGIQ™ 100 PRO. Included are descriptions of the facility and electrical needs to be met by the
purchaser of the unit. A checklist is also provided at the end of this section to help determine
whether the proper planning and preparation is accomplished before the actual equipment
installation is scheduled.
CONTENTS IN CHAPTER 2
SectionDescriptionPage Number
2-1Overview2-1
2-2General Console Requirements2-1
Table 2-1 Contents in Chapter 2
Section 2-2 General Console Requirements
2-2-1Console Environmental Requirements
Operating temperature:
Storage temperature:
Humidity: 30% to 75% rH non-condensing
Table 2-2 Environmental Specifications for LOGIQ™ 100 PRO Scanners
10 to 40 oC
-10 to 60 oC
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 40 of 197
Chapter 2 - Pre InstallationPage 2-1
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
2-2-1-1Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and in direct. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a
source of EMI which could degrade image quality. These controls should be selected to minimize
possible interference.
2-2-2 Electrical Requirements
2-2-2-1LOGIQ™ 100 PRO Power Requirements
Electrical Specifications for LOGIQ™ 100 PRO.
GE Healthcare
VoltageTolerancesCurrentFrequency
Hcat number
H41072LA,H41072LB,
H41072LC,H41072LD,
100-230
VAC
±10%0.9 -1.8 A
Max.
50-60 Hz
H41072LE,H41072LF,
H41072LS
Table 2-3 Electrical Specifications for LOGIQ™ 100 PRO
2-2-2-2Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
CAUTION POWER OUTAGE MAY OCCUR. The LOGIQ™ 100 PRO requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical care equipment, make sure you
DO NOT have any other equipment operating on the same circuit.
2-2-2-3Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension co rds. Other outlet s
adequate for the external periphe rals, medical and test equipment nee ded to support this unit m ust
also be present within 1 m (3.2 ft.) of the unit. Electric al installation must meet all current local, state,
and national electrical codes.
2-2-2-4Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or
the installation engineer must supply what is locally required.
2-2-2-5Power Stability Requirements
Voltage drop-out
Max 10 ms.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient,
including line frequency, synchronous, asynchronous, or aperiodic transients.
Page 2-2Section 2-2 - General Console Requirements
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 41 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
2-2-3EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The LOGIQ™ 100 PRO
complies with limits as stated on the EMC label. However there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect.
These sources include:
•medical lasers,
•scanners,
•cauterizing guns,
•computers,
•monitors,
•fans,
•gel warmers,
•microwave ovens,
•light dimmers
•portable phones.
The presence of a broadcast station or broadcast van may also cause interference.
See Table 2-4 for EMI Prevention tips.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 42 of 197
Chapter 2 - Pre InstallationPage 2-3
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
2-2-3 EMI Limitations (cont’d).
EMI RuleDetails
Be aware of RF
sources
Keep the unit at least 5 meters or 15 feet away from other EMI sources.
Special shielding may be required to eliminate interference problems
caused by high frequency, high powered radio or video broadcast
signals.
Ground the unitPoor grounding is the most likely reason a unit will have noisy images.
Check grounding of the power cord and power outl et.
Replace all screws,
RF gaskets, covers,
cores
After you finish repairing or updating the system, replace all covers and
tighten all screws. Any cable with an external connection requires a
magnet wrap at each end. Install the shield over the front of card cage.
Loose or missing covers or RF gaskets allow radio frequencies to
interfere with the ultrasound signals.
Replace broken RF
gaskets
If more than 20% or a pair of the fingers on an RF gasket are broken,
replace the gasket. Do not turn on the unit until any loose metallic part is
removed.
Do not place labels
where RF gaskets
touch metal
Use GE specified
harnesses and
peripherals
Never place a label where RF gaskets meet the unit. Otherwise, the ga p
created will permit RF leakage. Or, if a label has been found in such a
position, move the label.
The interconnect cables are grounded and require ferrite b eads and other
shielding. Also, cable length, material, and routing are all important; do
not change from what is specified.
Take care with
cellular phones
Properly dress
peripheral cables
Cellular phones may transmit a 5 V/m signal; that could cause image
artifacts.
Do not allow cables to lie across the top of the card cage or hang out of
the peripheral bays. Loop the excess length for peripheral cables inside
the peripheral bays. Attach the monitor cables to the frame.
Table 2-4 EMI Prevention/abatement
2-2-4Probes Environmental Requirements
ELECTRONICPAMPTE
Operation:
Storage:
Temperatures in degrees C, conversion to degrees F = oC*(9/5) + 32)
Table 2-5 Operation and Storage Temperatures for Probes.
NOTICE:SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES
OF -20 TO + 50 deg C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
10 to 40 oC 5 to 42.7 oC
-20 to 50 oC-20 to 50 oC
Page 2-4Section 2-2 - General Console Requirements
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 43 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
2-2-5Time and Manpower Requirements
Site preparation takes time. Begin Pre installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
2-2-6Checklist for Installation
2-2-6-1Checklist for Installation
ProceduresParagraphInitials
Observe WarningsSection 3-1-3
Unpack LOGIQ™ 100 PROSection 3-2
Assemble LOGIQ™ 100 PROSection 3-4
Perform Functional Checks Chapter 4-3
Perform Electrical Safety TestsChapter 10-6
Table 2-6 Checklist for Installat ion
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 44 of 197
Chapter 2 - Pre InstallationPage 2-5
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Your Notes:
Page 2-6Section 2-2 - General Console Requirements
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 45 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Chapter 3 Installation
Section 3-1Overview
3-1-1Purpose of Chapter 3
This chapter contains information needed to install the unit. Included are references to a procedure
that describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual installation, and how to check and test the unit,
probes, and external peripherals for electrical safety are included in this procedure. Also included
in this section are guidelines for transporting the unit to a new site.
CONTENTS IN CHAPTER 3
Table 3-1 Contents in Chapter 3
SectionDescriptionPage Number
3-1Overview
3-2Receiving and Unpacking the Equipment
3-3Preparing for Installation
3-5Completing the Installation
3-6System Configuration
3-7Paperwork
3-1
3-3
3-5
3-6
3-17
3-28
3-1-2Average Installation Time
DescriptionAverage
Installation
Time
Unpacking the scanner20 Minutes
Scanner wo/options40 MinutesDependant on the configuration that
Table 3-2 Average Installation Time
Comments
is required
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 46 of 197
Chapter 3 - InstallationPage 3-1
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-1-3Safety Reminders
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
DANGER
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
to its operating environment.
To prevent electrical shock, connect the unit t o a properly grounded power outlet. Do not use
a three to two prong adapter. This defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
Do not use a 20 Amp to 15 Amp adapter on the 120 Vac unit’s power cord. This unit requires
a dedicated 20 A circuit and can have a 15A plug if the on board peripher als do not cause the
unit to draw more than 14.0 amps.
Do not operate this unit unless all board covers and frame panels are securely in place.
System performance and cooling require this.
CAUTION
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQ™ 100
PRO and kept near the unit for quick reference.
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the LOGIQ™ 100 PRO probe is within
AIUM/NEMA standards, avoid unnecessary exposure. Ultrasound energy can produce heat
and mechanical damage.
Page 3-2Section 3-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 47 of 197
GE HEALTHCARE
p
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 3-2 Receiving and Unpacking the Equipment
Please read that procedure before packing/unpacking the LOGIQ™ 100 PRO.
We strongly advice you to store the LOGIQ™ 100 PRO packing material in undamaged condition
in case of future transportation.
1.) Cut the two straps wrapped around the box. Refer Figure 3-1.
2.) Remove the adhesive tape. Refer Figure 3-1
3.) Open the carton flaps. Refer Figure 3- 2.
4.) Remove the Operator Manual, Service Manual, Gel Bottle, Power Cord and other
accessories kept on either side of the equipment. Refer
5.) Lift the system (wrapped with stretch film) out of the box along with the packaging bu ffers.
Figure 3-2
Refer
6.) Slide the packaging buffers out to free the system.
7.) Remove the stretch film which is wrapped around the system. Refer Figure 3-2
Straps
Figure 3-3.
Adhesive Ta
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 48 of 197
Figure 3-1 Unpacking LOGIQ™ 100 PRO
Chapter 3 - InstallationPage 3-3
GE HEALTHCARE
fe
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Packaging Buf
Carton F laps
Stretch FilmPackaging Buffers
Figure 3-2 Unpacking LOGIQ™ 100 PRO
Note: The Service Manual, Operator Manual,Advanced Refernce Manual, Gel Bottle and
Power Cord are wrapped in bubble sheet and are kept along the sides of the
corrugated carton box. Refer Illustration below for details
Figure 3-3 Unpacking LOGIQ™ 100 PRO
Page 3-4Section 3-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 49 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 3-3 Preparing for Installation
3-3-1Physical Inspection
3-3-1-1Systems Voltage Settings
•Verify that the scanner is set to the correct voltage.
The Voltage settings for the LOGIQ™ 100 PRO Scanner is found on a label on the rear of the
system.
WARNING CONNECTING A LOGIQ™ 100 PRO SCANNER TO THE WRONG VOLTAGE LEVEL
WILL MOST LIKELY DESTROY THE SCANNER.
3-3-1-2Video Formats
Check that the video format is set to the locally used video standard, NTSC or PAL.
3-3-2EMI Protection
This unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many
of the covers, shields, and screws are provided primarily to protect the system from image artifacts
caused by this interference. For this reason, it is imperative that all covers and hardware are
installed and secured before the unit is put into operation.
Section 3-4 Assembling LOGIQ™ 100 PRO
CAUTION It is strongly recommended that the equipment should be installed on an even and hard
surface to provide proper cooling, as the vents for air circulat ion are on the bottom cover.
3-4-1Installation of the System
1.) Locate, remove and unpack the probes located in the separate container.
2.) Make sure that system power is OFF before connecting or disconnecting the probe .
3.) Secure the probe to the receptacle as follows. Refer to Fig 3-4
a.) Ensure probe `twist lock' lever points towards the top of the probe connector.
a.) Install probe connector on the receptacle guide pin until it touches the receptacle mating
surface.
a.) Twist the probe connector `twist lock' lever in the clockwise direction to lock it in place
(Twist the lever anticlockwise direction to disconnect the probe).
4.) Connect the LOGIQ™ 100 PRO Power cable's female part to the power con nector located on
the rear panel and the male p art to a hosp ital grade power socket of a proper voltage . Never
use a three-to-two pin adapter; this defeats the purpose of safety ground.
5.) Unlock the keyboard by pressing the lock release on top of the system.
CAUTION Do not move or lift the unit, holding the keyboard alone.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 50 of 197
Chapter 3 - InstallationPage 3-5
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Assembling LOGIQ™ 100 PRO (Contd.,)
Figure 3-4 Assembling LOGIQ™ 100 PRO
Section 3-5 Completing the Installation
3-5-1Transducer Connection
1.) Connect a transducer to transducer receptacle as follows,Refer to fig 3-4
A.) Ensure that the transducer twist lock lever points towards the top of the probe
connector.
B.) Insert the transducer connector on the receptacle guide pin until it touches the
receptacle mating surface.
C.) Twist the transducer twist lock lever in the clockwise direction to lock it in place. Twist
the lever in anti clockwise direction to disconnect the transducer.
Note: It is not necessary to turn OFF power to connect or disconnect a transducer.
2.) Connect the main power cable to a hospital grade power receptacle with the proper rated
voltage. Never use a three-to-two prong adapter; this defeats the safety ground.
Page 3-6Section 3-5 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 51 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3.5.1.1System Software Installation through USB Thumbdrive:
1.) Switch OFF the LOGIQ™ 100 PRO unit.Connect the USB thumb drive to the USB port
provided in the rear panel.
2.) Switch ON the unit and wait till the system boots up completely and the B mode image is
seen.
3.) Now Switch OFF the LOGIQ™ 100 PRO unit and switch ON again.
4.) The system should display the latest version of software that was loaded through the USB.
The USB thumb drive would have the latest software in it,which would be sent by the Manufactur er
before software upgrade.
Whenever Upgrade of LOGIQ™ 100 PRO is done at field, Ple ase ensure patient data back u p
is done before upgrading the machine.
Do not remove the USB thumb drive when the image is getting transfered.
3-5-2Installation of Optional Accessories
1.) Unpack the foot switch and connect it to the connector on the rear panel. (Refer Figure 3-5)
2.) Connect the VCR to a suitable power outlet. Using the BNC to BNC cable connect the Video
Out on the rear panel of LOGIQ™ 100 PRO to the Video IN of the VCR.
3.) Connect the Video Graphic Printer (VGP) to a suitable power outlet. Using the BNC to BNC
cable connect the Video Out on the rear panel of LOGIQ™ 100 PRO to the Video IN of the
VGP. Connect the shutter cable from the VGP printer to the rear panel of the Logiq 100 PRO.
4.) The images and report pages of the Logiq 100 PRO can be printed on HP Laser jet
printer(Model HP1020) through the USB port provided on the re ar panel.Images can be
transfered to a PC using USB Thumb drive.
Note: The LOGIQ™ 100 PRO has o nly one Video Out connection Therefore either a VGP
or a VCR only can be connected at a time
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 52 of 197
Chapter 3 - InstallationPage 3-7
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Figure 3-5 Installation of Accessories
Page 3-8Section 3-5 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 53 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-5-3Installation of Two Probe Adapter & LOGIQ α200 Probe Adapter (Optional)
1.) Remove the adhesive tape to open the flaps of two probe port or L OGIQ α 200 Probe Adapter
box
2.) Remove the PU Foams inside the two probe port or LOGIQ α 200 Probe Adapter box (Refer
Figure 3-6)
3.) Take out the two probe port or LOGIQ α 200 Probe Adapter (Refer Figure 3-6)
PU Foam
Note: The two probe port or LOGIQ α 200 Probe Adapter can be connected to or disconnected
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 54 of 197
Two Pr ob e po r t
or
LOGIQ a 200 Probe Adapter
Box
Figure 3-6 Unpacking Two Probe Adapter or LOGIQ α200 Probe Adapter
from the system at any time regardless of whether the system is powered ON or OFF.
Chapter 3 - InstallationPage 3-9
GE HEALTHCARE
(
)
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Installation of Two Probe Adapter & LOGIQ α200 Probe Adapter (Optional) (contd.,)
4.) Ensure that no probe is connected to the system. If connected remove it first.
5.) Remove the probe holder by lifting it vertically up
6.) Make sure that the two probe port or LOGIQ α 200 Probe Adapter lock points to the 12 o'clock
position.
7.) Align the Two Probe Port or LOGIQ α 200 Probe Adapter Connector with the System Probe
Port and carefully push it
8.) Align the slots on the upper flange to its counterparts on the top cover of the system.
9.) Turn the Two Probe Port or LOGIQ a 200 Lock to the 3 o'clock pos i tion to secur e the T wo
Probe Port or LOGIQ a 200 Probe Adapter to
10.) Carefully replace the probe holder to its place and press firmly till the projections at
the bottom fit into the slots in the upper flange and the system.
in.
the system.
Probe Hol der
Two Probe Port or
LOGIQ a 200 Probe Adapter lock
Two Probe Port Lock 12 o'clock
(Un Locked Position)
Hole on the upper flange
Two Probe port
LOGIQ a 200 Probe Adapter
Tw o Pr ob e Port Lock 3 o'clock
Locked Position
Figure 3-7 Installing Two Probe Port or LOGIQ 200 Probe Adapter
Page 3-10Section 3-5 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 55 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-5-4Installation of LOGIQ™ 100 PRO with Global or Modular Trolley
Note: LOGIQ™ 100 PRO has two types of Trolleys, the Global Trolley and the Modular Trolley.
The Global Trolley comes in the assembled form and the Modular T rolley comes in a modular
form which needs assembly at field.
1.) Unpack the global trolley packed in a wooden box by opening the top cover, refer Figure 3-8
(trolley which is inside the box is seen in the diagram).
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 56 of 197
Figure 3-8 Global Trolley Unpacking
Chapter 3 - InstallationPage 3-11
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,)
2.) Unpack the Modular Trolley packed in a box by opening the top cover, refer Figure 3-8
Figure 3-9 Modular Trolley Unpacking
Page 3-12Section 3-5 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 57 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 58 of 197
Figure 3-10 Modular Trolley Assembly
Chapter 3 - InstallationPage 3-13
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,)
3.) Remove two M4X10 screws from the base ofLOGIQ™ 100 PRO and fix lock plate. Refer
Figure 3-11
4.) Place the unit with lock plate fixed on Trolley, ensure that lock plate gets into th e projection on
the trolley. Refer
Figure 3-11
Lock Plate
Detail X
Console
Trolley
Loc k P late
Addi tional D etails
Figure 3-11 Installation of Trolley
Page 3-14Section 3-5 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 59 of 197
Detail X
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,)
5.) Take the belt (provided in the kit) over the unit and lock it. Refer Figure 3-12
6.) Release the Keyboard by pressing the lock release and make it rest on the handle. Refer
Figure 3-12
7.) Insert the VGP inside the Printer Tray and fix from the bottom using M3X12 screw. Refer Figure
3-12
Belt w i th Lock
Projection
VGP
Handle
VGP T ray (2179596)
Ensure that 2179596-2 is used
to take care of fitment problem of
VGP's .
Refer SV Note. SN LA100- 014
All the New Trolleys effective
September'1 2000 i s s uppl ied
wi th modified FRU (2179596- 2)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 60 of 197
Figure 3-12 Installation of LOGIQ™ 100 PRO with the Trolley
Chapter 3 - InstallationPage 3-15
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,)
Figure 3-13 Installation of LOGIQ™ 100 PRO with Global or Modular Trolley
Page 3-16Section 3-5 - Completing the Installation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 61 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-5-5Adjustment of Monitor Brightness & Contrast
Turn ON the Machine and then adjust the monitor contrast and brightness. Monitor contrast and
brightness should be adjusted according to the lighting in the room. For Monitor brightness and
contrast adjustment refer to the appropr iat e se ctio n in th e Use r Ma nu a l.
3-5-6Adjustment of System Clock/Date, Hospital Name & OB Version
Press CTRL-S-1 ENTER, an `Installation Setup window appears'. Set the system Clock/Date,
Hospital Name and OB Version. These settings can be modified by using the CONTROL-S-1 key if
required. Refer Section Control Keys of the User Manual for more details
CAUTION Functional checks in Chapter 4, to be carried out before handing over the system to the
customer
Section 3-6 System Configuration
3-6-1System Specifications
3-6-1-1Physical Dimensions
The physical dimensions of the LOGIQ™ 100 PRO unit are summarized in Table 3-2.
HeightWidthDepth
Table 3-3 Physical Dimensions of LOGIQ™ 100 PRO
28.930.2442.06
Unit
cm
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 62 of 197
Chapter 3 - InstallationPage 3-17
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-6-1-2Front View
Figure 3-14 Front View
Page 3-18Section 3-6 - System Configuration
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 63 of 197
GE HEALTHCARE
)
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-6-1-3Side View (Keyboard Closed)
Figure 3-15 Side View (Keyboard Closed)
3-6-1-4Side View (Keyboard Open)
420.6 (16.5)
289 (11.4)
Figure 3-16 Side View (Keyboard Open)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 64 of 197
596 (23.5
Chapter 3 - InstallationPage 3-19
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Table 3-5 Electrical Specifications for LOGIQ™ 10 0 PRO
Page 3-20Section 3-6 - System Configuration
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
VoltageTolerancesCurrentFrequency
100-230 VAC±10%0.9 -1.8 A Max.50-60 Hz
Approved Document - 5305880-100TPH_r3.pdf Page 65 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-6-2-1Power Supply
The Following Power Line requirement should be monitored one week before Installation. We
recommend that a Drantez model 605-3 Power line Monitor with options 101 to be used
Table 3-6
ParameterLimits
USA, JAPAN
Voltage Range
Power
Line FrequencyAll Applications
Power Transients
Decaying Oscillation
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of
transient, including line frequency, synchronous, asynchronous or periodic transients
Less than 15% of Peak Voltage for less than 1 millisecond
Europe
USA, JAPAN
Europe
3-6-2-2Facility Power Socket
A separate power outlet with a 5 amp circuit breaker for 120/220 VAC is recommended. The specific
power receptacle used depends on the country's power line standards.
This receptacle should have International Electro technical Commission (IEC) approval, or
equivalent
3-6-2-3Storage & Operation Requirements
The LOGIQ™ 100 PRO is shipped in a single container excluding probes. Shipping weight is
approximately 12 kgs.
Table 3-7 Table provides a summary of temperature, atmospher ic pressure,
and humidity tolerances for shipping, installation and operation:
100-115VAC ± 10% (90-127VAC)
200-240VAC ± 10% (198-264VAC)
MAX 145VA
MAX 175VA
50/60Hz ± 2Hz
Table 3-7 Storage & Operation Requirements
3-6-3Optional Peripherals
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
ParameterStorageOperation
Temperature (Deg.C)-10 to 6010-40
Atmospheric Pressure
(hPa)
Humidity (%)
Non- Condensing
700 to 1060700 to 1060
30 to 9035 to 75
LOGIQ™ 100 PRO peripherals and accessories can be properly connected using the connectors
on the rear panel of the LOGIQ™ 100 PRO system.
External Optional Peripherals
•VCR (Sony SVO-9500MD)
•VGP (Sony UP-895MD/MDW and UPD-897MDW)
•Printing through Laser Jet Printer with HP LaserJet 1020 Plus.
•Images can be transfered to memory stick from the image archive.
Chapter 3 - InstallationPage 3-21
Approved Document - 5305880-100TPH_r3.pdf Page 66 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Printing should not be done while live scanning patients. Also, the Printer cable should not be
attached to the LOGIQ™ 100 PRO during the scan.
For Europe Only: All devices connected to the LOGIQ™ 100 PRO must be CE Marked.
Located on the panel are video input & outp ut connectors, footswitch connector, shutter connector,
power connector & control connections for Video Graphic Printer and USB Port for the
LOGIQ™ 100 PRO.
Figure 3-17 Optional Peripherals
The Thumb Drive/Memory Stick (Model TS512MJF150), HP Laser Jet printer(HP1020), Sony VGP
Printer( Model UP-D897MD/ UP-897MD) can be connected through the USB slot provided in the
rear panel.
Page 3-22Section 3-6 - System Configuration
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 67 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-6-4Peripherals Connections and Installation
3.6.4.1FootSwitch Connection
The footswitch which is the remote FREEZE device is connected to the rear panel of the system.
this extra FREEZE switch is provided to enhance flexibility to freeze images when the system is not
within reach of the user.
2.) Connect the footswitch cable to the Footswitch slot provided in the rear panel.
3.) Press foot switch to check its functionality.
4.) By pressing the foot switch pedal the image in the screen should freeze.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 68 of 197
Chapter 3 - InstallationPage 3-23
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3.6.4.2Video Graphic Printer (VGP)
Connect the Video Graphic printer (with USB A to USB B cable) to the USB port located on the rear
panel of the system.
Model: (Sony UP-D897MD or UP-897MD)
Figure 3-20 Connection of USB cable to Printer
Installation of VGP:
1.) Place the SONY Digital Printer on the fixture.
2.) Fix one end of power cable and USB cable to the printer.
3.) Fix the other end of USB cable to the USB slot provided in the rear panel and power cable
to the Power socket.
4.) Switch ON the LOGIQ™ 100 PRO and VGP printer.
5.) Press the Recall button on the keyboard to view the stored images.
6.) Press Record button on the keyboard and check if the selected image matches with the
VGP print out.
3.6.4.3Pin Assignment of Mini Jack controlling VGP
Pin No.Signal
1Shutter
2Signal GND
Figure 3-21 Shutter Connector
3.6.4.4Video Casette Recorder
Connect the VCR Sony SVO-9500MD (H4120SR) for NTSC and Sony SVO 950 0MDP(E11801AA)
for PAL System to the Video IN socket in the rear panel. Press the “EXT. VIDEO” button on the
keyboard to enable an external video signal on the system monitor. SVHS tape should be used for
recording.
Page 3-24Section 3-6 - System Configuration
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 69 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Figure 3-22 VCR connection to Rear panel
Installation of VCR:
1.) Connect the SONY VCR
2.) Switch on the VCR and the LOGIQ™ 100 PRO unit.
3.) Press the Record key to record the image.
4.) Press the External video key on the keyboard to enable external video( i.e VCR Playback)to
be viewed on the LOGIQ™ 100 PRO system monitor.
3.6.4.5Laser jet printer
Connect the HP1020 Laser jet printer to the USB port provided in the rear panel of the system.This
enables you to take paper printout of the images and report pages.T he LOGIQ™ 100 PRO supports
printing in A4 size papers in the HP laser Jet 1020Series with standard USB cable.
Figure 3-23 Printer connection to Rear pane l
Installation of Laser Jet Printer:
1.) Switch ON on the LOGIQ™ 100 PRO unit.
2.) Connect the USB cable from the printer to the USB port provided in the rear pa nel of LOGIQ™
100 PRO unit.
3.) To Print images/Report pages, Press Shift+ Record.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 70 of 197
Chapter 3 - InstallationPage 3-25
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3.6.4.6USB MEMORY STICK
This module serves as an interface which enables users to transfer images from LOGIQ™ 100 PRO
to USB memory stick. LOGIQ™ 100 PRO in JPEG format.This can be further uploaded to any PC
having windows operating system and USB port support.
Figure 3-24 USB Memory Stick connection to Rear Panel
Installation of USB Memory Stick:
1.) Connect the USB Memory stick to the USB port provided in the rear panel of LOGIQ™ 100
PRO unit.
2.) Press the recall key. From the main menu Press 4 to enter the multiple selection menu.
3.) To transfer the images to a Memory stick move the trackball to the images that you want to
transfer and press 3.To transfer all images available in the archive press 2 to select images
and 3 to transfer images to Memory stick.
Please ensure that the USB Memory stick has to be connected befo re atte mpting
to transfer images to USB Memory stick. If USB is not connected to the LOGIQ™
100 PRO system, the system will prompt " USB not connected".
Do not remove the USB Memory stick when the image is getting transfered.
Note : USB hard disk drive may/may not be recognized by the system.
Page 3-26Section 3-6 - System Configuration
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 71 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3.6.4.7Available probes for LOGIQ™ 100 PRO
See in Specifications in the LOGIQ™ 100 PRO User manual for probes and intended use.
* This Probes are available only for LOGIQ™ 100 PRO with LOGIQ 200 Probe Adapter.
Use only approved probes, peripherals or accessories.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 72 of 197
Chapter 3 - InstallationPage 3-27
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-6-5Video Specification
Video specifications may be needed to be able to connect laser ca meras to the LOGIQ™ 100 PRO.
General
SLParameterNTSCPAL
1Total Number of Horizontal Lines
2Vertical Field Frequency
3Horizontal Scanning Frequency
4Vertical Scanning Method
5Sync Input
6Pixel Clock
7Display width and Height [pixel * line]
Horizontal Timing & Video Amplitude
SLParameterNTSCPAL
1Total H-Line Time
2H-Sync Pulse Width
3Back Porch
4Total Active H-line time
5Front Porch
6Video Amplitude (back porch to peak)
7Sync Amplitude (back porch to peak)
Note: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the
peripheral units must be kept as part of the original system documentation. This will ensure that all
relevant safety and user informations are available during the operation and service of the complete
system.
262.5[H]312.5[H]
1st Field 3[lines]1st Field 2.5[lines]
2nd Field 3[lines]2nd Field 2.5[lines]
1st Field 7.5[lines]1st Field 6.25[lines]
2nd Field 7.5[lines]2nd Field 6.25[lines]
1st Filed 30.5[lines]1st Filed 32[lines]
2nd Field 30.5[lines]2nd Field 32[lines]
Page 3-28Section 3-7 - Paperwork
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 73 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
3-7-1User Manual(s)
Check that the correct User Manual(s), per software (sw) revision, for the system is included.
3-7-2Product Locator Card
Fill out proper customer information on the Product Locator Installation Card. Refer to Illustration
below. Mail this Installation Card “Product Locator” to the address corresponding to your pole.
Note: The Product Locator Installation Card shown in Figure 3-25 may not be same as
the Product Locator card actually provided.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 74 of 197
Figure 3-25 Product Locator Card
Chapter 3 - InstallationPage 3-29
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Page 3-30Section 3-7 - Paperwork
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 75 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Chapter 4 Functional Checks
Section 4-1Overview
4-1-1Purpose of Chapter 4
This chapter provides procedures for quickly checking major functions of the LOGIQ™ 100 PRO
scanner, diagnostics by using the built-in service software, and power supply adjustments. For
remaining functional checks please refe r to the LOGIQ™ 100 PRO Operator's Manual.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 76 of 197
Chapter 4 - Functional ChecksPage 4-1
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 4-2 General Procedures
4-2-1Power On/Boot Up
Note: After turning off the system, wait at least ten seconds before turning it on again. The system may
not be able to boot if power is recycled too quickly.
CAUTION SYSTEM REQUIRES ALL COVERS
Operate this unit only when all board covers and frame panels are securely in place. The
covers are required for safe operation, good system performance and cooling purposes.
4-2-1-1System Boot-UP/Shutdown
Purpose: This is a description on how to Shut Down/Power-UP the system.
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug.
4-2-1-3Power Up
1.) Connect the Power Cable at the rear of the System.
2.) Connect the Power cable to an appropriate mains power outlet .
Mains Power
Cable Connector
Circuit Breaker
Figure 4-1 Mains Circuit Breaker and Mains Power Cable
Page 4-2Section 4-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 77 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
4-2-1-4Front Panel ON/OFF
Press the Front Panel ON/OFF key once. to Switch ON/OFF
Figure 4-2 ON/OFF key on Front Panel
When power is applied to the Scanner, Power is distributed to the Fan, Control panel, Monitor,
Internal and External I/O’s, Boards & Peripherals. Back End Processor starts and its software code
is distributed to initiate the scanner.
Power Up Display Comes Up. Refer figure 4-3.
4-2-1-5Power Shutdown
1.) Switch OFF the ON/OFF at the Front of the system.
2.) Disconnect the Mains Power Cable if needed.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 78 of 197
Chapter 4 - Functional ChecksPage 4-3
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 4-3 Functional Check Procedures
To perform these tests, you will need a convex or a linea r pro be. The se p rocedur es sho uld be pe r
formed during installation. These procedures are also used as basic checks to use when service of
the system is required.
Turn “ON” the LOGIQ™ 100 PRO system. The 2 LEDS, External Video and F reeze blink and go off.
The Monitor displays the following. Refer figure 4-3 for details.
1.) B-Mode Image appears
2.) Gray Scale Display
3.) Horizontal Scale Marker
4.) Vertical Scale Marker
5.) Hospital Name (If previously entered)
6.) Date
7.) Time
8.) Probe Type
9.) Focus Marker on vertical scale
10.) Depth (150mm)*
11.) Gain (G54) *
12.) Dynamic Range (D54) *
13.) Frequency **.
* Points 10-12 form a continuous code in the lower left hand corner of the monitor and they are
factory default values.
** Only for C36 Probe.
Page 4-4Section 4-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 79 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
4-3-1Power Up Monitor Display (Convex Probe)
Figure 4-3 Power Up Monitor Display (Convex Probe)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 80 of 197
Chapter 4 - Functional ChecksPage 4-5
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
4-3-2Basic Controls
Table 4-2 describes basic controls which helps in checking the LOGIQ™ 100 PRO during
installation.
Connect the sector or convex transducer to connector on the right side of the base.
TASKExpected Result
Turn `ON' Power SwitchB-Mode screen as in figure 4-3should appear
Press New Patient Key
A pop-up window appears. Enter Patient Name/ID. It
presets the system
Press ID/Name
A pop-up window appears. Enter Patient Name/ID. It
does not preset the system.
Press Dyn Range Up/DownImage grows softer and harder depending on position.
Rotate Gain KnobImage grows lighter and darker with rotation.
Press Reverse
Press Reverse a second time
Press SHIFT + Reverse
Press SHIFT+ Reverse a second
The image will be displayed Left/Right
The image will be displayed Right/Left.
The image will be displayed Top/Bottom
The image will be displayed Bottom/Top
time
Press M key
Roll Trackball
B/M Mode with M-line cursor appears. (Use trackball to
move the M-line cursor). Refer Figure
4-4
The M-Line cursor should follow the trackball movement
and the real-time image varies on the M-Mode dis play.
Press M key a second time
Only M-Mode image appears on the screen.
Press B Mode Key to exit M-Mode.
Press Freeze keyThe image will freeze.
Press ATO ATO, Automatic Tissue Optimization, optimizes the
Press Freeze and Zoom
Roll Trackball
Table 4-2 Basic Controls
Page 4-6Section 4-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 81 of 197
image based upon a specified Region Of Interest ( ROI)
or anatomy within the display. ATO Gray scale map is
available only in B-Mode.
Zoom is used to magnify an image. The magnification
factor for zoom is fixed at 2.0. The system adjusts all
imaging parameters accordingly.
To select the ROI for Zoom
Zoom feature is avialable in the freezed mode onl y.
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
TASKExpected Result
CINE
(Press Freeze and Rotate B/M Gain/
Cine Scroll )
The Cine Gauge, menu and the Cine frame number
appears at the bottom of the display.
Rotate the Cine Scroll dial to move through the images
in Cine Memory.
Cine Menu:
Cine frame number is displayed on the left side of the
screen.
Press 1(Start Frame)
Move the Cursor by rotating the Cine Scroll dial to the
frame you want and press 1 again to select the Start
frame.
Press 2(End Frame)
Move the Cursor by rotating the Cine Scroll dial to the
frame you want and press 2 again to select the End
frame.
Press 3(Cine Loop)
Enters into Cine loop within the selected Start & End
frames.
Press 4(Cine Gauge)
To toggle between the Cine Gauge display
Flash Memory (Image storage)
Comment column appears at the bottom of the display.
Freeze the image
Press Store
Enter the comment in that and press Store again to
Store the image.
Image Archive screen appears with option menu at the
Press Recall
bottom.
Select the image using the trackball (Use Dynamic
Range to page up/down in the Image Archive screen).
Press 1 to View the selected image.
Press 2 to Clear the selected image.
Press 3 to Clear All the stored images.
Press 4 to Sort the images by name or date.
To Exit Image Archive, press Clear.
Unfreeze to exit from the freezed image.
TGCThe TGC is used to adjust the Gain at a specific depth.
Slide the control to the right or left to increase or
decrease gain.
Table 4-2 Basic Controls
Refer User Manual for the Remaining Operations
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 82 of 197
Chapter 4 - Functional ChecksPage 4-7
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Basic Controls (Continued)
Figure 4-4 B/M Mode Display
Page 4-8Section 4-1 - Overview
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 83 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 4-4 Diagnostics
The LOGIQ™ 100 PRO System service diagnostics comprises of:
1.) Self test or power on diagnostics
2.) Service diagnostic tools
The self test or power on diagnostics are run every time the system is booted. The service
diagnostic tools include test procedures for testing the system at PCB level as well as block level.
The diagnostic tools provides a pop-up menu to enable selection of various tests to test various
blocks on the system.
4-4-1Power On Diagnostics
The power on diagnostics or self tests are run every time the system is booted. This self tests
include testing the validity of the system software through EPROM checksum test and testing the
system RAM. It checks whether the keyboard and the trackball are properly interfaced to the
system. It also initiates the self test of analog subsystem. The two LED's and the beeper are used
to indicate error conditions.
Note: Power on Diagnostics works only wh en the system is in application (imaging) mode
and not while on service diagnostics.
4-4-2Service Diagnostics
For More Details on Service Diagnostic refer Chapter 7.
PCBLED LocationFunction
LED - Location on FEB
FEB
LED - Location on FEB
FEB
LED - Location on FEB
FEB
LED - Location on FEB
FEB
LED - Location on Davinci
DAVINCI
LED - Location on Davinci
DAVINCI
D65
D66
D67
D68
CPU D12
CPU D11
This LED indicates the scan lines,the DSP is scanning.The
LED glows when the cursor is moved in the B/M mode
This LED indicates the probe connect signal is good.This
indicates the FEB is Good
This LED indicates the B mode signal is given to the
display.The intensity of this LED is more during B mode.
This LED indicates the M mode signal is given to the
display.This LED glows during M and B/M mode,t he inte nsit y
of this LED is less during B/M mode
This LED is ON during Acoustic Frame measurement
switching.This also indicates the communication between the
DSP and the Davinci CPU FPGA
This LED indicates the Video Sync is happening.This indicates
the video frame output from the Davinci CPU to the monitor is
ok.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 84 of 197
DAVINCI
DAVINCI
LED - Location on Davinci
CPU D10
LED - Location on Davinci
CPU D09
Chapter 4 - Functional ChecksPage 4-9
This LED indicates the KEY press on the keyboard.It glows
intermittantly when there is a key pressed.
This LED indicates the USB connection to the
Rearpanel.When a USB device or Pen drive is connected, the
LED glows to indicate it is to the Davinci CPU.(USB output of
the Davinci is Good)
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 4-5 Software Configuration Checks
StepCheckExpected Result(s)If Not Remedy
1.Check Date and Time settingDate and Time are correctAdjust the Date and Time setting
2.Check that Location (Hospital Name) is
correct
Location Name is correctRe-enter the correct Location Name
Table 4-3 Software Configuration Checks
Section 4-6 Peripheral Checks
Check that peripherals work as described below:.
StepCheck Expected Result(s)If Not Remedy
1.Press Freeze KeyTo stop image acquisition.
2.Press RECORD Key on the Control
panel
3.Press SHIFT + RECORD Key on the
Control Panel.
4.Press SHIFT + RECORD Key on the
Control Panel.
The image displayed on the screen is
printed on B&W Video Graphic Printer.
To image or report page displayed on
the screen is printed on the Laser Jet
Printer connected to the USB Port of
the System.
To print the displayed image on the
screen to the USB B&W Video Graphic
Printer.
5.Press EXTERNAL VIDEO Key on the
Control Panel
Table 4-4 Peripheral checks
Enables an External Video Input
Playback on the LOGIQ™ 100 PRO
monitor. Press once again to return to
the scan mode
Page 4-10Section 4-6 - Peripheral Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 85 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 4-7 Patient contact tools
4-7-1Probe/Connectors Check
Take the probes and check them as described below:.
StepCheckExpected Result(s)If Not Remedy
1.To test each delivered Probe, Connect it
to the System
2.Hold the probe connector horizontally
with the cable pointing Sideways.
Turn the connector locking handle to the
Vertical position.
Align the connector with the probe port
and carefully push into place.
Rotate the locking handle to the full
Horizontal position to lock in place.
Position the probe cable so that it is not
resting on the floor
CAUTION:
Do not allow the probe head to hang
freely. Impact to the probe head may
result in irreparable damage.
3.Rotate the lock handle counterclockwise to the Vertical position to
unlock the connector.
Remove the connector from the port
Ensure that the probe head is clean
before placing the probe in its storage
case.
4.To test the Two Probe Adapter,
Connect it to the System
It will display the probe name on the
screen
To connect a probe:
To disconnect probes:
The probes that are not connected
to the unit should be stored in their
storage case.
To connect to the probe port
5.Hold the Two Probe Port connector
6.Two Probes can be Switched by
7.Presetting Parameters to a Probe,
8.T o tes t the LO GIQ PROB E ADA PTER,
Table 4-5 Probe and connectors check
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
horizontally
Turn the connector locking handle to the
Vertical position.
Align the connector with the probe port
and carefully push into place.
Rotate the locking handle to the full
Horizontal position to lock in place.
pressing the Key on the Two Probe Port
Select the desired probe & adjust the
desired parameters
Press CTRL + W + Enter
connect it to the system
To connect a probe:
The LED toggles to indicate the probe
change & a click is also heard to
indicate change over
The system initalizes the new probe &
the image from the newly selected
probe is displayed in the B-Mode
The Parameters selected will be set as
Default Values
Connect the Adapter to the Port
Chapter 4 - Functional ChecksPage 4-11
Approved Document - 5305880-100TPH_r3.pdf Page 86 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3LOGIQ™ 100 PRO SERVICE MANUAL
D
StepCheckExpected Result(s)If Not Remedy
9.Hold the probe connector horizontally
with the cable pointing Sideways.
Turn the connector locking handle to the
Vertical position.
Align the connector with the probe port
and carefully push into place.
Rotate the locking handle to the full
Horizontal position to lock in place.
Position the probe cable so that it is not
resting on the floor
CAUTION:
Do not allow the probe head to hang
freely. Impact to the probe head may
result in irreparable damage.
It will display the Probe name on the
Screen
Table 4-5 Probe and connectors check
WARNING Probes can be connected at any time, whether the unit is on or off
WARNING Do NOT touch the patient and any of the connectors on the ult rasound unit simultaneously,
including ultrasound probe connectors.
Page 4-12Section 4-6 - Peripheral Checks
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 87 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 5-1Overview
5-1-1Purpose of Chapter 5
This chapter explains LOGIQ™ 100 PRO’s system concepts, component arrangement, and
subsystem function. It also describes the Power Distribution System (PDS) and probes.
Table 4-11 Power Distributor Input and Output Voltages of LOGIQ™ 100 PRO .
TEST point Locations on the FEB.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Location for Testing HV on
FEB
FEB Test Point 90.65 Vdc +/- 3Vdc.
FEB Test Point 91.80 Vdc +/- 3 vdc.
FEB Test Point 62, 65 and 73.12 Vdc +/- .3 vdc.
FEB Test Point 63, 67 and 70.5 V dc +/- .3 vdc.
FEB Test Point 110.-12V dc +/- .3 vdc.
FEB Test Point 60 and 61.3.3 V dc +/- .03 vdc.
Specification
Table 4-12 Voltage Test points on the FEB.
NOTE : The Test point information is provided on the Base of the console, which is visible after
opening the top cover.
- Page 5-9
Approved Document - 5305880-100TPH_r3.pdf Page 96 of 197
GE HEALTHCARE
Davinci b
d TGC
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
5-4-0-2Cable Assembly Diagrams
oar
connector
Figure 4-10 Cable Assembly Wiring Diagram
Page 5-10Section 5-3 - Block Diagram
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 97 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Cable Assembly Diagrams (Continued)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 98 of 197
Figure 4-10
* Ferrite Core Over which wires wound 2 Turns
- Page 5-11
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
Section 5-5 Functional Subsystems
5-5-1Front End Board
The Front End Bo ar d ge ner ates the strong tra nsmit bursts, t ransmitted by the probes as
ultrasound i nto the body. It also receives weak ultrasound echoes from blood cells and body
structure, amplifies the si gnals and converts them to an 8 bit digital signal. F EB support s 16
channels Beamformer. The beamf ormed data is passed through CH A COM ASIC and
subsequentl y passed thr ough Frame Averaging filter and Mean filter. The Output of FEB i s fed to
the Davinci CP U PWA through a boar d - to-board connect or. This connector also provides power
to the Davinci CPU PWA Board.
Page 5-12Section 5-3 - Block Diagram
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 99 of 197
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
5-5-2Davinci Board
The Davinci CPU board uses TMS320DM6446 c hip that contains an ARM Pr oc essor, DSP and
Video processors into a single chip.
The Major functionalities & corresponding implementation is described below.
1) System Memory - 256MB DDR2 Memory.
2) Program Storage – 16 Megaby tes of NOR Flash is used to store the Boot loader code, K ernel,
and Application software.
3) Image Storage – 64 Megabytes of NAND Flash is used to store the User Im ages. It is al so
used to store a part of Applic ation software and operating system code.
4) Keyboard interface (Ultrasound keyboard I /f , PS 2 trackball interfac e, G ain Controller
interfac e). Ultrasound Keyboard is interfaced to Davinci CPU PWA.The serial c ommunication
occurs at 9600-baud rate.
Optical Trackball TB_CLK and TB_DATA pins are connected t o Dav inci CP U PWA. Trackball
supports Ps2 standard.
GAIN Encoder - GAIN_CCW and GAIN_CW pins of the gain encoder are connected to the GPIO
pins of Davinci CPU.
5) On Screen Display (O S D), Vi deo Out and Monitor Display - The DAVINCI CPU generates the
OSD along with the image in NTS C or PAL vi deo output format (standard) t o the Video Out and
Monitor connector.
6) USB port interf ac e - The Davi nci CPU PWA prov ides one USB 2.0 Host connector. This port is
used to interface Thermal Printer, DeskJet pri nter as well as to transfer the im age to Mem ory
stick in JPEG format.
7) NTSC and PAL setting - A Dipswitc h is prov ided to select NTSC or PAL standard. This switch
is available on Dav inci CP U PWA.
8) Real Time Clock (RTC) - RTC chip is provided on Davinci CP U PWA that acts as a timekeeper
i.e. it maintains the system clock and date.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 100 of 197
- Page 5-13
GE HEALTHCARE
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
D
5-5-3Peripherals
VCR, Black & White Video Graphic Printer and HP Laser Jet Printer 1020 can be connected to the
scanner. For more details on recommended peripherals refer Chapter 3.
Section 5-6 Video Specifications
Refer to Section 3-6-5 for more details about the Video Specifications.
Section 5-7 Rear Panel Signal List
Refer to Section 3-6-4 for more details about the rear panel signal list.
Page 5-14Section 5-6 - Video Specifications
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 101 of 197
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.