Page 1

GEHealthcare
KISS™/KISS™Multilead
ElectrodeApplicationSystem
ServiceManual
22745010RevisionH
English
©1996-2017GeneralElectricCompany.
AllRightsReserved.
Page 2

PublicationInformation
TheinformationinthismanualappliesonlytoKISS/KISSMlutilead.Itdoesnotapplytoearlierproductversions.Duetocontinuingproduct
innovation,specicationsinthismanualaresubjecttochangewithoutnotice.
KISS,CASEandMACaretrademarksownedbyGEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompanygoingto
marketasGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheirrespectiveowners.
Thedocumentpartnumberandrevisionareatthebottomofeachpage.Therevisionidentiesthedocument’supdatelevel.Therevision
historyofthisdocumentissummarizedinthefollowingtable.
Revision
-
A
B
Date
September1996Initialrelease.
January2004ECO07578
April2006
Comments
ECO082331
C
D
E
F
GAugust2014
July2009
June2010
April2011changesonpages20and29,newsection3.5.4
March2014Updateofsection3.3.4andchapter6
ECO096124
TMsymbolforKISSadded,changesonpages3,4,21,22,28,
•Updateofchapters5and7duetonewRoHS-compliantACsuctionpump
•RemovalofKISS-3fromFSM
H
ToaccessotherGEHealthcareDiagnosticCardiologymanuals,gototheCommonDocumentationLibrary(CDL),locatedat
http://www.gehealthcare.com/usen/service/biomed_tech_selfservice/services_user_doc/products/support.html,andclickCardiology.
ToaccessOriginalEquipmentManufacturer(OEM)manuals,gotothedevicemanufacturer'sWebsite.
CAUTION:
Duringrepairs/serviceinterventions,observetheprotectivemeasuresagainstdamageduetoESD.
ManualPurpose
Thismanualsuppliestechnicalinformationforservicerepresentativeandtechnicalpersonnelsotheycanmaintaintheequipmenttothe
assemblylevel.Useitasaguideformaintenanceandelectricalrepairsconsideredeldrepairable.Wherenecessarythemanualidenties
additionalsourcesofrelevantinformationandortechnicalassistance.
Seetheoperatormanualfortheinstructionsnecessarytooperatetheequipmentsafelyinaccordancewithitsfunctionandintendeduse.
IntendedAudience
Thismanualisintendedforthepersonwhouses,maintains,ortroubleshootsthisequipment.
January2017
AddedinformationfortheO-Rings.
SafetyInformation
ResponsibilityoftheManufacturer
GEHealthcareisresponsiblefortheeffectsofsafety,reliability,andperformanceonlyif:
•Assemblyoperations,extensions,readjustments,modications,orrepairsarecarriedoutbypersonsauthorizedbyGEHealthcare.
•Theelectricalinstallationoftherelevantroomcomplieswiththerequirementsoftheappropriateregulations.
•Theequipmentisusedinaccordancewiththeinstructionsforuse.
•Thecountryofmanufactureappearsonthedevicelabel.
General
Thisdeviceisnotintendedforhomeuse.
ContactGEHealthcareforinformationbeforeconnectinganydevicestotheequipmentthatarenotrecommendedinthismanual.
PartsandaccessoriesusedmustmeettherequirementsoftheapplicableIEC601seriessafetystandards,and/orthesystemconguration
mustmeettherequirementsoftheIEC60601-1-1medicalelectricalsystemsstandard.
Periodically,andwhenevertheintegrityofthedeviceisindoubt,testallfunctions.
2
KISS™/KISS™Multilead
22745010H
January2017
Page 3

TheuseofACCESSORYequipmentnotcomplyingwiththeequivalentsafetyrequirementsofthisequipmentmayleadtoareducedlevelof
safetyoftheresultingsystem.Considerationrelatingtothechoiceshallinclude:
•useoftheaccessoryinthePA TIENTVICINITY;and
•evidencethatthesafetycerticationoftheACCESSORYhasbeenperformedinaccordancetotheappropriateIEC60601-1and/orIEC
60601-1-1harmonizednationalstandard.
TheauthorizedrepresentativeforGEMedicalSystemsInformationTechnologiesInc.inEuropeis:
GEMedicalSystemsInformationTechnologiesGmbH
MunzingerStrasse5
79111Freiburg
GERMANY
Phone+49-76145430
Fax+49-7614543233
Warnings,Cautions,andNotes
Thetermsdanger ,warning,andcautionareusedthroughoutthismanualtopointouthazardsandtodesignateadegreeorlevelor
seriousness.Familiarizeyourselfwiththeirdenitionsandsignicance.
Hazardisdenedasasourceofpotentialinjurytoaperson.
DANGER
Indicatesanimminenthazardwhich,ifnotavoided,willresultindeathorseriousinjury.
WARNING
Indicatesapotentialhazardorunsafepracticewhich,ifnotavoided,couldresultindeathorseriousinjury.
CAUTION
Indicatesapotentialhazardorunsafepracticewhich,ifnotavoided,couldresultinminorpersonalinjuryorproduct/propertydamage.
NOTE
Providesapplicationtipsorotherusefulinformationtoassurethatyougetthemostfromyourequipment.
ServiceRequirements
Followtheservicerequirementslistedbelow.
•ReferequipmentservicingtoGEHealthcare’sauthorizedservicepersonnelonly.
•Anyunauthorizedattempttorepairequipmentunderwarrantyvoidsthatwarranty.
•Itistheuser’sresponsibilitytoreporttheneedforservicetoGEHealthcareortooneoftheirauthorizedagents.
•Failureonthepartoftheresponsibleindividual,hospital,orinstitutionusingthisequipmenttoimplementasatisfactorymaintenance
schedulemaycauseundueequipmentfailureandpossiblehealthhazards.
•Regularmaintenance,irrespectiveofusage,isessentialtoensurethattheequipmentwillalwaysbefunctionalwhenrequired.
22745010H
KISS™/KISS™Multilead
3
Page 4

ServiceManualLanguageInformation
WARNING
(EN)
ПРЕДУПРЕЖДЕНИЕ
(BG)
警告
(ZH-CN)
警告
(ZH-TW)
UPOZORENJE
(HR)
VAROVÁNÍ
(CS)
ThisservicemanualisavailableinEnglishonly.
•Ifacustomer'sserviceproviderrequiresalanguageotherthanEnglish,itisthecustomer's
responsibilitytoprovidetranslationservices.
•Donotattempttoservicetheequipmentunlessthisservicemanualhasbeenconsulted
andisunderstood.
•Failuretoheedthiswarningmayresultininjurytotheserviceprovider ,operator,orpatient,
fromelectricshock,mechanicalorotherhazards.
Товаупътванезаработаеналичносамонаанглийскиезик.
•Акодоставчикътнауслугатанаклиентаизискадругезик,задължениенаклиентае
даосигурипревод.
•Неизползвайтеоборудването,предидастесеконсултиралииразбралиупътването
заработа.
•Неспазванетонатовапредупреждениеможедадоведедонараняваненадоставчика
науслугата,оператораилипациентврезултатнатоковударилимеханичнаили
другаопасност.
本维修手册仅提供英文版本。
•如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
•未详细阅读和完全理解本维修手册之前,不得进行维修。
•忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
本維修手冊只提供英文版。
•如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責提供翻
譯服務。
•除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維修。
•不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或其他因素
而受到傷害。
Oveuputezaservisiranjedostupnesusamonaengleskomjeziku.
•Ukolikokorisničkiserviszahtijevanekidrugijezik,korisnikovajeodgovornostosigurati
odgovarajućiprijevod.
•Nemojtepokušavatiservisiratiopremuukolikonistekonzultiraliirazumjelioveupute.
•Nepoštivanjeovogupozorenjamožerezultiratiozljedamaservisnogosoblja,korisnikaili
pacijentaprouzročenimelektričnimudaromtemehaničkimilinekimdrugimopasnostima.
Tentoprovoznínávodexistujepouzevanglickémjazyce.
•Vpřípadě,žeexterníslužbazákazníkůmpotřebujenávodvjinémjazyce,jezajištění
překladudoodpovídajícíhojazykaúkolemzákazníka.
•Nesnažteseoúdržbutohotozařízení,anižbystesipřečetlitentoprovoznínávoda
pochopilijehoobsah.
•Vpřípaděnedodržovánítétovarovánímůžedojítkporaněnípracovníkaprodejního
servisu,obslužnéhopersonálunebopacientůvlivemelektrickéhoproudu,respektive
vlivemmechanickýchčijinýchrizik.
4
KISS™/KISS™Multilead
22745010H
January2017
Page 5

ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(DA)
Denneservicemanualndeskunpåengelsk.
•Hvisenkundesteknikerharbrugforetandetsprogendengelsk,erdetkundensansvar
atsørgeforoversættelse.
•Forsøgikkeatservicereudstyretmedmindredenneservicemanualharværetkonsulteret
ogerforstået.
•Manglendeoverholdelseafdenneadvarselkanmedføreskadepågrundafelektrisk,
mekaniskellerandenfareforteknikeren,operatørenellerpatienten.
WAARSCHUWING
(NL)
DezeservicemanualisalleeninhetEngelsverkrijgbaar .
•Indienhetonderhoudspersoneeleenanderetaalnodigheeft,danisdeklant
verantwoordelijkvoordevertalingervan.
•Probeerdeapparatuurnietteonderhoudenvoordatdezeservicemanualgeraadpleegd
enbegrepenis.
•Indiendezewaarschuwingnietwordtopgevolgd,zouhetonderhoudspersoneel,de
gebruikerofeenpatiëntgewondkunnenrakenalsgevolgvaneenelektrischeschok,
mechanischeofanderegevaren.
HOIATUS
(ET)
Käesolevteenindusjuhendonsaadavalainultinglisekeeles.
•Kuiklienditeeninduseosutajanõuabjuhenditinglisekeelesterinevaskeeles,vastutab
klienttõlketeenuseosutamiseeest.
•Ärgeüritageseadmeidteenindadaenneeelnevaltkäesolevateenindusjuhendiga
tutvumistjasellestarusaamist.
•Käesolevahoiatuseeiraminevõibpõhjustadateenuseosutaja,operaatorivõipatsiendi
vigastamistelektrilöögi,mehaanilisevõimuuohutagajärjel.
VAROITUS
(FI)
Tämähuolto-ohjeonsaatavillavainenglanniksi.
•Josasiakkaanhuoltohenkilöstövaatiimuutakuinenglanninkielistämateriaalia,tarvittavan
käännöksenhankkiminenonasiakkaanvastuulla.
•Äläyritäkorjatalaitteistoaennenkuinoletvarmastilukenutjaymmärtänyttämän
huolto-ohjeen.
•Mikälitätävaroitustaeinoudateta,seurauksenavoiollahuoltohenkilöstön,laitteiston
käyttäjäntaipotilaanvahingoittuminensähköiskun,mekaanisenviantaimuun
vaaratilanteenvuoksi.
ATTENTION
(FR)
Cemanueltechniquen'estdisponiblequ'enanglais.
•Siunservicetechniqueclientsouhaiteobtenircemanueldansuneautrelangueque
l'anglais,ildevraprendreenchargelatraductionetlaresponsabilitéducontenu.
•Nepastenterd'intervenirsurleséquipementstantquelemanueltechniquen'apasété
consultéetcompris.
•Lenon-respectdecetavertissementpeutentraînerchezletechnicien,l'opérateuroule
patientdesblessuresduesàdesdangersélectriques,mécaniquesouautres.
22745010H
KISS™/KISS™Multilead
5
Page 6

ServiceManualLanguageInformation(cont'd.)
WARNUNG
(DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
AÐVÖRUN
(IS)
PERINGATAN
(ID)
DieseServiceanleitungistnurinenglischerSpracheverfügbar.
•FallsderKundendiensteineandereSprachebenötigt,musserfüreineentsprechende
Übersetzungsorgen.
•KeineWartungdurchführen,ohnedieseServiceanleitunggelesenundverstandenzu
haben.
•BeiZuwiderhandlungkanneszuVerletzungendesKundendiensttechnikers,des
AnwendersoderdesPatientendurchStromschläge,mechanischeodersonstigeGefahren
kommen.
Τοπαρόνεγχειρίδιοσέρβιςδιατίθεταιστααγγλικάμόνο.
•Εάντοάτομοπαροχήςσέρβιςενόςπελάτηαπαιτείτοπαρόνεγχειρίδιοσεγλώσσαεκτός
τωναγγλικών,αποτελείευθύνητουπελάτηναπαρέχειυπηρεσίεςμετάφρασης.
•Μηνεπιχειρήσετετηνεκτέλεσηεργασιώνσέρβιςστονεξοπλισμόεκτόςεάνέχετε
συμβουλευτείκαιέχετεκατανοήσειτοπαρόνεγχειρίδιοσέρβις.
•Εάνδενλάβετευπόψητηνπροειδοποίησηαυτή,ενδέχεταιναπροκληθείτραυματισμός
στοάτομοπαροχήςσέρβις,στοχειριστήήστονασθενήαπόηλεκτροπληξία,μηχανικούς
ήάλλουςκινδύνους.
Ezaszervizkézikönyvkizárólagangolnyelvenérhetőel.
•Haavevőszervizellátójaangoltóleltérőnyelvretartigényt,akkoravevőfelelőssége
afordításelkészíttetése.
•Nepróbáljaelkezdenihasználniaberendezést,amígaszervizkézikönyvbenleírtakat
nemértelmeztékésértettékmeg.
•Ezengyelmeztetésgyelmenkívülhagyásaaszervizellátó,aműködtetővagyapáciens
áramütés,mechanikaivagyegyébveszélyhelyzetmiattisérüléséteredményezheti.
Þessiþjónustuhandbókereingöngufáanlegáensku.
•Efaðþjónustuveitandiviðskiptamannsþarfnastannarstungumálsenensku,erþaðskylda
viðskiptamannsaðskaffatungumálaþjónustu.
•Reyniðekkiaðafgreiðatækiðnemaþessiþjónustuhandbókhefurveriðskoðuðogskilin.
•Brotáaðsinnaþessariaðvörungeturleitttilmeiðslaáþjónustuveitanda,stjórnandaeða
sjúklingifráraosti,vélrænumeðaöðrumáhættum.
ManualservisinihanyatersediadalambahasaInggris.
•JikapenyediajasaservispelangganmemerlukanbahasalainselaindariBahasaInggris,
merupakantanggungjawabdaripenyediajasaservistersebutuntukmenyediakan
terjemahannya.
•Janganmencobamelakukanservisterhadapperlengkapankecualitelahmembacadan
memahamimanualservisini.
•Mengabaikanperingataninibisamengakibatkancederapadapenyediaservis,operator,
ataupasien,karenaterkenakejutlistrik,bahayamekanisataubahayalainnya.
6
KISS™/KISS™Multilead
22745010H
January2017
Page 7

ServiceManualLanguageInformation(cont'd.)
AVVERTENZA
(IT)
警告
(JA)
경고
(KO)
ЕСКЕРТУ
(KK)
BRĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
IlpresentemanualedimanutenzioneèdisponibilesoltantoinInglese.
•Seunaddettoallamanutenzionerichiedeilmanualeinunalinguadiversa,ilclienteè
tenutoaprovvederedirettamenteallatraduzione.
•Siprocedaallamanutenzionedell'apparecchiaturasolodopoaverconsultatoilpresente
manualeedavernecompresoilcontenuto.
•Ilnonrispettodellapresenteavvertenzapotrebbefarcompiereoperazionidacuiderivino
lesioniall'addetto,allamanutenzione,all'utilizzatoreedalpazienteperfolgorazione
elettrica,perurtimeccaniciodaltririschi.
このサービスマニュアルは英語版しかありません。
•サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
•このサービスマニュアルを熟読し、十分に理解をした上で装置のサービスを
行ってください。
•この警告に従わない場合、サービスを担当される方、操作員あるいは患者が、感電
や機械的又はその他の危険により負傷する可能性があります。
본서비스지침서는영어로만이용하실수있습니다.
•고객의서비스제공자가영어이외의언어를요구할경우,번역서비스를제공하는것
은고객의책임입니다.
•본서비스지침서를참고했고이해하지않는한은해당장비를수리하려고시도하
지마십시오.
•이경고에유의하지않으면전기쇼크,기계상의혹은다른위험으로부터서비스제
공자,운영자혹은환자에게위해를가할수있습니다.
Бұлқызметкөрсетубойыншанұсқаулығытекағылшынтіліндеқолжетімді.
•Тұтынушыныңқызметпровайдеріағылшынтіліненбасқатілдегінұсқаныталапетсе,
аударубойыншақызметтеріменқамтамасызетутұтынушыжауапкершілігіндеболуы
тиіс.
•Бұлқызметкөрсетубойыншанұсқаулығынназарғаалып,түсінбегенше,жабдыққа
қызметкөрсетуденбастартыңыз.
•Бұлескертудіелемеуқызметпровайдері,операторнемесеемделушініңэлектр
шогынан,механикалықнемесебасқақауіптернәтижесіндежарақаталуынаәкелуі
мүмкін.
Šīapkalpotājurokasgrāmatairpieejamatikaiangļuvalodā.
•Jaapkalpošanassniedzējamnepieciešamainformācijacitā,nevisangļu,valodā,klienta
pienākumsirnodrošināttāstulkošanu.
•Neveicietaprīkojumaapkopi,neizlasotunnesaprototapkalpotājurokasgrāmatu.
•Šībrīdinājumaneievērošanavarradītelektriskāsstrāvastrieciena,mehāniskuvaiciturisku
izraisītutraumuapkopessniedzējam,operatoramvaipacientam.
Šiseksploatavimovadovasyraprieinamastikanglųkalba.
•Jeiklientopaslaugųtiekėjasreikalaujavadovokitakalba-neanglų,numatytivertimo
paslaugasyraklientoatsakomybė.
•Nemėginkiteatliktiįrangostechninėspriežiūros,nebentatsižvelgėteįšįeksploatavimo
vadovąirjįsupratote.
•Jeineatkreipsitedėmesioįšįperspėjimą,galimisužalojimaidėlelektrosšoko,mechaninių
arkitųpaslaugųtiekėjui,operatoriuiarpacientui.
22745010H
KISS™/KISS™Multilead
7
Page 8

ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(NO)
Denneservicehåndbokennnesbarepåengelsk.
•Hviskundensserviceleverandørtrengeretannetspråk,erdetkundensansvaråsørge
foroversettelse.
•Ikkeforsøkåreparereutstyretutenatdenneservicehåndbokenerlestogforstått.
•Manglendehensyntildenneadvarselenkanføretilatserviceleverandøren,operatøren
ellerpasientenskadespågrunnavelektriskstøt,mekaniskeellerandrefarer.
OSTRZEŻENIE
(PL)
Niniejszypodręcznikserwisowydostępnyjestjedyniewjęzykuangielskim.
•Jeślidostawcausługklientawymagajęzykainnegoniżangielski,zapewnienieusługi
tłumaczeniajestobowiązkiemklienta.
•Nienależyserwisowaćwyposażeniabezzapoznaniasięizrozumienianiniejszego
podręcznikaserwisowego.
•Niezastosowaniesiędotegoostrzeżeniamożespowodowaćurazydostawcyusług,
operatoralubpacjentawwynikuporażeniaelektrycznego,zagrożeniamechanicznego
bądźinnego.
AVISO
(PT-BR)
Estemanualdeassistênciatécnicasóseencontradisponíveleminglês.
•SeoserviçodeassistênciatécnicadoclientenãoforGE,eprecisardeoutroidioma,será
daresponsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanualde
assistênciatécnica.
•Onãocumprimentodesteavisopodeporemperigoasegurançadotécnico,operadorou
pacientedevidoachoqueselétricos,mecânicosououtros.
AVISO
(PT-PT)
Estemanualtécnicosóseencontradisponíveleminglês.
•Seaassistênciatécnicadoclientesolicitarestesmanuaisnoutroidioma,éda
responsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanual
técnico.
•Onãocumprimentodesteavisopodeprovocarlesõesaotécnico,aoutilizadorouao
pacientedevidoachoqueseléctricos,mecânicosououtros.
AVERTISMENT
(RO)
Acestmanualdeserviceestedisponibilnumaiînlimbaengleză.
•Dacăunfurnizordeserviciipentruclienţinecesităoaltălimbădecâtceaengleză,estede
datoriaclientuluisăfurnizezeotraducere.
•Nuîncercaţisăreparaţiechipamentuldecâtulteriorconsultăriişiînţelegeriiacestui
manualdeservice.
•Ignorareaacestuiavertismentarputeaducelarănireadepanatorului,operatoruluisau
pacientuluiînurmapericolelordeelectrocutare,mecanicesaudealtănatură.
ПРЕДУПРЕЖДЕНИЕ
(RU)
Настоящееруководствопообслуживаниюпредлагаетсятольконаанглийскомязыке.
•Еслисервисномуперсоналуклиентанеобходиморуководствоненаанглийском,ана
каком-тодругомязыке,клиентуследуетобеспечитьпереводсамостоятельно.
•Преждечемприступатькобслуживаниюоборудования,обязательнообратитеськ
настоящемуруководствуивнимательноизучитеизложенныевнемсведения.
•Несоблюдениетребованийданногопредупрежденияможетпривестиктому,что
специалистыпообслуживанию,операторыилипациентыполучатударэлектрическим
током,механическуютравмуилидругоеповреждение.
8
KISS™/KISS™Multilead
22745010H
January2017
Page 9

ServiceManualLanguageInformation(cont'd.)
UPOZORENJE
(SR)
VAROVANIE
(SK)
OPOZORILO
(SL)
ADVERTENCIA
(ES)
VARNING
(SV)
UYARI
(TR)
Ovoservisnouputstvojedostupnosamonaengleskomjeziku.
•Akoklijentovserviserzahtevanekidrugijezik,klijentjedužandaobezbediprevodilačke
usluge.
•Nepokušavajtedaopraviteuređajakonistepročitaliirazumeliovoservisnouputstvo.
•Zanemarivanjeovogupozorenjamožedovestidopovređivanjaservisera,rukovaocaili
pacijentausledstrujnogudara,ilimehaničkihidrugihopasnosti.
Tentonávodnaobsluhujekdispozíciilenvangličtine.
•Akzákazníkovposkytovateľslužiebvyžadujeinýjazykakoangličtinu,poskytnutie
prekladateľskýchslužiebjezodpovednosťouzákazníka.
•Nepokúšajtesaoobsluhuzariadeniaskôr,akosineprečítatenávodnaobsluhua
neporozumietemu.
•Zanedbanietohtovarovaniamôževyústiťdozraneniaposkytovateľaslužieb,obsluhujúcej
osobyalebopacientaelektrickýmprúdom,mechanickýmaleboinýmnebezpečenstvom.
Taservisnipriročnikjenavoljosamovangleškemjeziku.
•Čeponudnikstoritvestrankepotrebujepriročnikvdrugemjeziku,morastrankazagotoviti
prevod.
•Neposkušajteservisiratiopreme,četegapriročnikanistevcelotiprebraliinrazumeli.
•Četegaopozorilaneupoštevate,selahkozaradielektričnegaudara,mehanskihalidrugih
nevarnostipoškodujeponudnikstoritev,operateralibolnik.
Estemanualdeserviciosóloexisteeninglés.
•Sielencargadodemantenimientodeunclientenecesitaunidiomaquenoseaelinglés,el
clientedeberáencargarsedelatraduccióndelmanual.
•Nosedeberádarserviciotécnicoalequipo,sinhaberconsultadoycomprendidoeste
manualdeservicio.
•Lanoobservanciadelpresenteavisopuededarlugaraqueelproveedordeservicios,el
operadoroelpacientesufranlesionesprovocadasporcausaseléctricas,mecánicaso
deotranaturaleza.
Denhärservicehandbokennnsbaratillgängligpåengelska.
•Omenkundsserviceteknikerharbehovavettannatspråkänengelskaansvararkunden
föratttillhandahållaöversättningstjänster.
•Försökinteutföraservicepåutrustningenomduinteharlästochförstårdenhär
servicehandboken.
•Omduintetarhänsyntilldenhärvarningenkandetresulteraiskadorpåserviceteknikern,
operatörenellerpatiententillföljdavelektriskastötar,mekaniskafarorellerandrafaror.
Buserviskılavuzununsadeceİngilizcesimevcuttur.
•Eğermüşteriteknisyenibukılavuzuİngilizcedışındabirbaşkalisandantalepederse,bunu
tercümeettirmekmüşteriyedüşer.
•Serviskılavuzunuokuyupanlamadanekipmanlaramüdahaleetmeyiniz.
•Buuyarıyauyulmaması,elektrik,mekanikveyadiğertehlikelerdendolayıteknisyen,
operatörveyahastanınyaralanmasınayolaçabilir.
22745010H
KISS™/KISS™Multilead
9
Page 10

ServiceManualLanguageInformation(cont'd.)
ЗАСТЕРЕЖЕННЯ
(UK)
CẢNHBÁO
(VI)
Данекерівництвозсервісногообслуговуванняпостачаєтьсявиключноанглійськоюмовою.
•Якщосервіснийінженерпотребуєкерівництвоіншоюмовою,користувачзобов'язаний
забезпечитипослугиперекладача.
•Ненамагайтесяздійснюватитехнічнеобслуговуванняданогообладнання,якщо
винечитали,абонезрозумілиінформацію,наданувкерівництвізсервісного
обслуговування.
•Недотриманняцьогозастереженняможепризвестидотравмуваннясервісного
інженера,користувачаданогообладнанняабопацієнтавнаслідокелектричногошоку,
механічногоушкодженняабозіншихпричинневірногообслуговуванняобладнання.
TàiLiệuHướngDẫnSửaChữachỉcóbảntiếngAnh.
•Nếucácđơnvịcungcấpdịchvụchokháchhàngyêucầumộtngônngữnàokháctiếng
Anh,thìkháchhàngsẽcótráchnhiệmcungcấpcácdịchvụdịchthuật.
•KhôngđượcsửachữathiếtbịtrừkhiđãthamkhảovàhiểuTàiliệuHướngdẫnSửachữa.
•Khôngtuânthủnhữngcảnhbáonàycóthểdẫnđếncáctổnthươngchongườithực
hiệnsửachữa,ngườivậnhànhhaybệnhnhân,dosốcđiện,cácrủirovềcơkhíhay
cácrủirokhác.
10
KISS™/KISS™Multilead
22745010H
January2017
Page 11

Contents
1FunctionalDescription
GeneralInformation..................................................................................13
2OperatingControlsandSystemComponents
KISS™........................................................................................................15
KISS™Multilead.........................................................................................17
3TestinginstructionsforKISS™andKISS™Multilead
VisualCheck..............................................................................................19
SystemFunctionCheck.............................................................................19
EquipmentCheckforLeaksintheCompleteSystem,Pump,
Distributor,ElectrodeLeadsandSuctionCups.........................................20
TestingRequirements................................................................................20
CompleteSystemTest...............................................................................21
ElectrodeLeadValveLeak..........................................................................21
PumpLeakage...........................................................................................22
PumpLeakageTestwithPumpsRequiringPositivePressuretoStart......................22
PumpLeakageTestwithPumpsRequiringNegativePressuretoStart....................23
ElectrodeLeakage....................................................................................23
TestingElectricalFunction(KISS™)................................................................24
TestingElectricalFunction(KISS™Multilead)....................................................25
KISS™-Disassembly,Assembly................................................................25
DisassemblingandAssemblingtheControllerandtheEndPieceonthe
Distributor.............................................................................................25
DisassemblingtheTrunkCableandthePCB....................................................26
AssemblingthePCBandtheTrunkCable........................................................26
AssemblingtheController..........................................................................27
KISS™Multilead-Disassembling,Assembly.............................................27
DisassemblingandAssemblingtheController..................................................27
DisassemblingtheConnectorPCBandtheConnectorHousings...........................27
AirTubeandInputPCB..............................................................................28
ExchangeoftheKISSMultileadAirTube.........................................................29
KISS™/KISS™MultileadO-RingReplacement............................................34
4Maintenance,CleaningandTroubleshooting(KISS™/KISS™
Multilead)
Maintenance.............................................................................................37
22745010H
KISS™/KISS™Multilead
11
Page 12

CleaningandDisinfectingtheInstrumentSurfaces.................................37
CleaningandDisinfectingthePatientandLeadwires..............................37
CleaningElectrodesandElectrodeCare...................................................38
ReplacingLeadwires.................................................................................38
Troubleshooting........................................................................................39
5SparePartsList(KISS™/KISS™Multilead)
..................................................................................................................41
6TestingInstructionsforKISS™andKISS™Multilead
1.TestPlate(p/n30344490)......................................................................45
2.Y-Adapter(p/n30344489).....................................................................45
3.AdapterExtensionwithSyringe(p/n2081397-001)..............................45
4.DummyPlugforKISS™(p/n30344488).................................................46
7Specications
KISS™-10,KISS™-12..................................................................................47
Environment...........................................................................................48
KISS™Multilead.........................................................................................48
Environment...........................................................................................49
ExternalSuctionPumpwithPowerSupply................................................49
ASchematics
AssemblyDrawings...................................................................................51
12
KISS™/KISS™Multilead
22745010H
Page 13

1
FunctionalDescription
GeneralInformation
TheGEHealthcareElectrodeApplicationSystemKISS™isusedtoassistinthe
applicationofECGelectrodesfortheacquisitionofrestingandexerciseECGs.The
systemisoptimallysuitedfromanergonomicviewpoint.
AnumberofdifferentKISS™versionsareavailable:
•KISS™-10(Standard)andKISS™-12(Standard+Nehb).
•KISS™Multilead(KISS™forCAM-14AcquisitionModule):10-lead(default)and
optionalelectrodeextensionkitsforFRANK,NEHBandA1..A4
Theelectrodeleadsareconductedtoacommontrunkcableviaacentralelectrode
distributor.Integratedintheelectrodes-andtrunkcable-isthepneumatictubing
connector.
AnumberofinstrumentversionsareavailabletomatchtheECGrecorderand
acquisitionmoduleused.
AnoptionfortheelectrodeapplicationsystemKISS™comprisesapolewithan
extendibleswivelarmadjustableinlengthandheightforinstallationonaGE
Healthcareinstrumentcart.Withthebracketsprovidedintheaccessorykit,the
systemcanbemountedtoawalloronaninstrumenttable.
Anelectricallyoperatedsuctionpump,installedintheECGrecorder/stresssystemor
inasystempowersupplyunitgeneratesaconstantnegativepressure,thenegative
pressurecontrollerinthedistributorenablingcontrolledadjustment.
Thenegativepressurecanbeadjustedwithinarangeof80to220mbarinaccordance
withtherequiredapplication.
Thesuctionpumponlyoperateswhennegativepressuremustbecreatedafter
attachingtheelectrodesorwhenalossinpressureforelectrodesalreadyattached
hastobecompensatedfor.Thispermitstheuseofaquiet,lowWattratingpump.
Thepowersupplyunitsuppliesthesuctionpumpwithmainsvoltagesfrom110V
to240V.
Thesilver/silverchlorideelectrodesareattachedbysuctionbypressingbrieyonthe
suctioncupattheapplicationsite.Avalveopensintheelectrodelead,enablingthe
generationofnegativepressure.
22745010H
KISS™/KISS™Multilead
13
Page 14

FunctionalDescription
14
KISS™/KISS™Multilead
22745010H
Page 15

OperatingControlsandSystem
Components
KISS™
2
22745010H
KISS™/KISS™Multilead
15
Page 16

OperatingControlsandSystemComponents
ApplicationSystemKISS™andpole
Item
Description
1
Electrodedistributor
2
Negativepressurecontrol
3
Electrodeleadswithvalve
4
Silversilverchlorideelectrode
5
Filterdisk=>electrode
6
Suctioncup=>electrode
7
Electrocardiographandpump
connectingcable(trunkcable)
8
Connectorforelectrocardiograph
9
Luer-Lockconnectorforsuctionpump
10
Extensiontube
Item
Description
11
Powerconnectorforpumpmodule
12
Pumpmodule
13
Screwclampforxationofpoletotable
top
14
Wall-mountfacilityforpole
15
Poletopwithretainingscrew,
accommodatescontactspraybottle
16
Lockingcatchtoadjusttheheightof
theswivelarm
17
Lockingcatchtoadjustthelengthof
theswivelarm
18
Bracketforelectrodedistributor
19
Electrodeleadhanger
20
Powercordhanger
16
KISS™/KISS™Multilead
22745010H
Page 17

OperatingControlsandSystemComponents
KISS™Multilead
KISS™Multileadandpole
Item
Description
1
Bracketforelectrodedistributor
2
Lockingcatchtoadjustthelengthof
theswivelarm
3
Lockingcatchtoadjusttheheightof
theswivelarm
4
Poletopwithretainingscrew,
accommodatescontactspraybottle
5
Electrodeleadhanger
6
Powercordhanger
7
Electrodedistributor
8
Negativepressurecontrol
9
Electrodeleadswithvalve(A1-A4
optional)
10
Silverchlorideelectrode
Item
Description
11
Filterdisk=>electrode
12
Suctioncup=>electrode
13
CAM14acquisitionmodule
14
SignalleadtoCAM14acquisition
module
15
Suctionpumpconnector
16
CAMUSBA/TKISS™interface
17
ConnectstotheUSBportofthePC
18
ACpoweradapter
19
Screwclampforxationofpoletotable
top
20
Wall-mountfacilityforpole
22745010H
KISS™/KISS™Multilead
17
Page 18

OperatingControlsandSystemComponents
18
KISS™/KISS™Multilead
22745010H
Page 19

TestinginstructionsforKISS™and
KISS™Multilead
VisualCheck
Checktheapplicationsystemforexternalintactness:
•Thedistributorhousingisundamaged.
•Thepowersupplyunitwiththeintegratedsuctionpumpisundamaged.
•Thetrunkcableandtheelectrodeleadsareintact.Nocracks,kinksordeformations.
•Thesuctioncupshavenocracks.
•Surfacesofallsilver/silverchlorideelectrodesareundamaged,i.e.,unscratched,
unsoiled.
3
Removeelectrodeleadsfromtheelectrodes.Removeelectrodesfromthesuction
cups:
•Nosoilingbyelectrodegel,hair,etc.
•Moistelectrodelterdisksmustbereplaced.Electrodenozzlesintact.
Forinstrumentswithanextensionarm:
Checkinstrumentarmandelectrodedistributorhousingareundamagedandsliding
componentslocksecurely.
SystemFunctionCheck
1.SetnegativepressureregulatorontheelectrodedistributortoCENTER.
2.SwitchonECGrecorderorconnectuppowersupplyunit.
3.Presssuctioncupsgentlyontoasmoothevensurfaceandreleasethecup.
Thesuctionpumpstartsupandgeneratesthenegativepressure.
22745010H
KISS™/KISS™Multilead
19
Page 20

TestinginstructionsforKISS™andKISS™Multilead
Theelectrodesmustadheresecurely.
4.Changingthepressurewiththeregulatorresultsinavisibleeffect
onthesuctioncups.
Negativepressureregulatorinposition(left),theelectrodesshoulddropoff.
EquipmentCheckforLeaksintheComplete
System,Pump,Distributor,ElectrodeLeadsand
SuctionCups
TestingRequirements
•PneumaticTransducerTesterPTT,e.g.,X-Caliber(showninmmHg)
•Ohmmeter/Multimeter
•Testplateforconnectionofasuctionelectrode
see“1.TestPlate(p/n30344490)”onpage45
•Y-adapter(forconnectionnegativepressuretubing<->PTT)
see“2.Y-Adapter(p/n30344489)”onpage45
•Dummyplugfordistributor
see“4.DummyPlugforKISS™(p/n30344488)”onpage46
20
KISS™/KISS™Multilead
22745010H
Page 21

CompleteSystemTest
1.Startupthepumpbyattachinganelectrode.
TestinginstructionsforKISS™andKISS™Multilead
a.Thenremovetheelectrode.
b.Measurethenegativepressureafterabout10seconds:thereadingmust
2.Thenswitchoffthenegative-pressurepump.
Thedropinnegativepressuremustnotexceed2mmHg/s(2.66mbar/s).
3.Attachelectrodestoasmoothatsurfaceandrepeatabovetest.
Thedropinnegativepressuremustnotexceedmax.4mmHg/s(5.3mbar/s).
Ifthesetestvaluelimitsarenotobserved,i.e.,thedropinnegativepressureproceeds
tooquickly,thesystemshouldbetestedfor“leaks”.
ElectrodeLeadValveLeak
Allelectrodesmustbeconnectedtothedistributor.
Inordertotesttheelectrodesforleaks,proceedasfollows:
•Replaceanelectrodebyadummyplug(partno.30344488).
•Carryoutthetestasdescribedin“CompleteSystemTest”onpage21insteps
1and2.
Iftheleakisstillpresent,replacethenextelectrodewithadummyplugandtest
again.Repeattheabovetestuntilthedefectiveelectrodeisidentied.Ifallelectrodes
arereplacedwithdummyplugsandthesystemstillleaks,theremustbealeakinthe
distributororthepump.
bebetween225and265mmHg(300and350mbar).
22745010H
KISS™/KISS™Multilead
21
Page 22

TestinginstructionsforKISS™andKISS™Multilead
Forpumpleaks,proceedasdescribedin“PumpLeakage”onpage22.Ifthedistributor
leaks,theentiredistributorshouldbereplaced(seeChapter5“SparePartsList
(KISS™/KISS™Multilead)”onpage41).
PumpLeakage
PumpLeakageTestwithPumpsRequiringPositivePressure
toStart
Startthepumpbyblowingagainstthesuctionpipestillopenandthenclosingthepipe
withyourngerasshowninthetestsetup(within3s).Thenswitchoffthenegative
pressurepump(ECGrecorderorpowersupplyunit):Thedropinnegativepressure
mustnotexceedmax.1mmHg/s(1.33mbar/s).
Ifthisvalueisnotobserved,thepumpmustbereplaced(ECGrecorder/acquisition
module(acq.module)orthepowersupplyunit).
Ifthereadingiscorrect,theremustbealeakinthedistributororregulator.The
distributorincl.regulatorshouldbereplaced.See“KISS™-Disassembly,Assembly”on
page25.
22
KISS™/KISS™Multilead
22745010H
Page 23

TestinginstructionsforKISS™andKISS™Multilead
PumpLeakageTestwithPumpsRequiringNegative
PressuretoStart
Startthepumpbypullinggentlyonthesyringepiston.Thenswitchoffthenegative
pressurepump(ECGrecorderorpowersupplyunit):Thedropinnegativepressure
mustnotexceedmax.1mmHg/s(1.33mbar/s).
ElectrodeLeakage
Ifthisvalueisnotobserved,thepumpmustbereplaced(ECGrecorder/acquisition
module(acq.module)orthepowersupplyunit).
Ifthereadingiscorrect,theremustbealeakinthedistributororregulator.The
distributorincl.regulatorshouldbereplaced.See“KISS™-Disassembly,Assembly”on
page25.
Startthepumpbypressingtheelectrodetothetestplate(partno.30344490)
Negativepressureadjustedtomaximum:max.150...195mmHg(200...270mbar)
Negativepressureadjustedtominimum:min.37...60mmHg(50...80mbar)
Eachelectrodeplusleadistestedinsequence.
(Thiscanbealwaysbeconductedatthesamedistributorslot).
Ifthesevaluesarenotobserved,thereisleakageattheelectrodebody/suction
cupinterfaceorattheplug-inconnectiontotheelectrodelead.
22745010H
KISS™/KISS™Multilead
23
Page 24

TestinginstructionsforKISS™andKISS™Multilead
Remedy:Replaceelectrode;seeSeeChapter5.“SparePartsList(KISS™/KISS™
Multilead)”onpage41.
TestingElectricalFunction(KISS™ ™
ResistancebetweentheleadwireconnectorandtheKISS™electrodeconnector.The
measuredresistancemustbe10kohm±1kohm.
CAUTION:
Itisprudenttomakethismeasurementwithoutelectrodesasthesilver-silver
chlorideelectrodesmustnotcomeintocontactwithothermetals.
PinassigmentforECGunitswithIECplug:
R=9,L=10,F=11,N=14,C1=12,C2=1,C3=2,C4=3,C5=4,C6=5,shield=6
™
)
24
KISS™/KISS™Multilead
22745010H
Page 25

TestingElectricalFunction(KISS™Multilead)
ResistancebetweentheleadwireconnectorandtheKISS™electrodeconnector.The
measuredresistancemustbe10kohm±1kohm.
CAUTION:
Itisprudenttomakethismeasurementwithoutelectrodesasthesilver-silver
chlorideelectrodesmustnotcomeintocontactwithothermetals.
TestinginstructionsforKISS™andKISS™Multilead
Leadwireconnector(KISS™Multilead)
KISS™-Disassembly,Assembly
DisassemblingandAssemblingtheControllerandtheEnd
PieceontheDistributor
Removeallelectrodeleadsonthedistributor.Thereare4springcatchestobeopened
foreachlead.
Firstofallliftup2springcatchesononesidewithapairofpointedtweezers(Caution:
Donotoverstretch!)andliftcomponentslightly,thenliftupthenext2springcatches
ontheothersideandremovecomponent(controllerorendpiece).
22745010H
KISS™/KISS™Multilead
25
Page 26

TestinginstructionsforKISS™andKISS™Multilead
DisassemblingtheTrunkCableandthePCB
•Firstcarryout“DisassemblingandAssemblingtheControllerandtheEndPiece
ontheDistributor”onpage25.
•DisconnectplugfromPCB.
•Positionapairoftweezersundertheendofthetubingandeasethetubingoffthe
extensiontube.Donotpullonthetubingastheextensiontubecouldbreakoff!
•RemovePCB.
AssemblingthePCBandtheTrunkCable
1.PlugPCBintothedistributor.
2.Withnozzleretractedpushonthetrunkcableasfarasitwillgo.
3.Twistleadendstwoturnstotheleft.PlugleadplugsintoPCB(observecoding!)
4.Slideonnozzle.
5.Slideonendpieceuntilall4springcatcheslockin(Caution:Avoidpinchingwires!)
Final,checkbypullinggentlyontheendpiece.
26
KISS™/KISS™Multilead
22745010H
Page 27

TestinginstructionsforKISS™andKISS™Multilead
AssemblingtheController
Slideoncontrolleruntilallspringcatcheslockin.
Final,checkbypullinggentlyonthecontroller.
KISS™Multilead-Disassembling,Assembly
DisassemblingandAssemblingtheController
Removeallelectrodeleadsonthedistributor.Thereare4springcatchestobeopened
foreachlead.
Firstofallliftup2springcatchesononesidewithapairofpointedtweezers(Caution:
Donotoverstretch!)andliftcomponentslightly,thenliftupthenext2springcatches
ontheothersideandremovecomponent(controller).
DisassemblingtheConnectorPCBandtheConnector
Housings
•RemovetheCAM-14mountingunit:Todoso,loosenthefourscrewsandcarefully
liftofftheCAM-14mountingunit.
•LoosenthexationscrewoftheconnectorPCBandremoveconnectorPCB.
22745010H
KISS™/KISS™Multilead
27
Page 28

TestinginstructionsforKISS™andKISS™Multilead
•Youcanremovetheconnectorhousingsbypullingthemcarefullyoffthecontact
pins.
AirTubeandInputPCB
Thedistributorblockisavailableasasparepart(withoutregulator).Theregulator
mustbeorderedseparately,ifneeded.
28
KISS™/KISS™Multilead
22745010H
Page 29

Ifthiscomponentistobereplaced,adevicelabelmustbeorderedat
techsupport.gemsit@med.ge.comusingtheexistingCat.No.andS/N.
ExchangeoftheKISSMultileadAirTube
•Removeallelectrodeleadsfromthedistributorblockandloosenthescrewsas
shownongure1andgure2.
TestinginstructionsforKISS™andKISS™Multilead
•CarefullyliftofftheCAM-14mountingunit.
Figure3
22745010H
KISS™/KISS™Multilead
29
Page 30

TestinginstructionsforKISS™andKISS™Multilead
•Pullthelockingplates(A)onbothsidesofthedistributorblockbyusingasuitable
tool(seegure4).
Figure4
•Liftthetwospringcatchesonbothsides(seegure5).CAUTION:DoNOT
overstretch!
Figure5
30
KISS™/KISS™Multilead
22745010H
Page 31

TestinginstructionsforKISS™andKISS™Multilead
•Removethecoverfromthedistributorblock(seegure6).
Figure6
•Pullofftheairtubefromthedistributorblock(seegure7).
22745010H
KISS™/KISS™Multilead
31
Page 32

TestinginstructionsforKISS™andKISS™Multilead
•Slidetheairtubeoffthecoverasshowningure8.
•Insertthenewairtubeinthecoverasshowningure9.
Figure9
32
KISS™/KISS™Multilead
22745010H
Page 33

TestinginstructionsforKISS™andKISS™Multilead
•Widentheendoftheairtube(seeAongure10)byusingasuitabletoolsothatitts
overthenozzleofthetubeconnector.Pushtheairtubeoverthenozzle.Assemble
thecoverandthedistributorblock.Ensurethatthespringcatchesclickintoplace.
Figure10
•Slidethetwolockingplates(A)intothedistributorblock(seegure11).Carefully
connecttheCAM-14mountingunit(seegure3)tothedistributorblockand
tightenthefourscrews(seegure1andgure2).Connectallelectrodeleads
tothedistributor .
Figure11
22745010H
KISS™/KISS™Multilead
33
Page 34

TestinginstructionsforKISS™andKISS™Multilead
KISS™/KISS™MultileadO-RingReplacement
Removeallelectrodeleadsonthedistributor.Thereare4springcatchestobeopened.
•Liftuptwospringcatchesononesidewithapairofpointedtweezers(Caution:Do
nooverstretch!),insertpointedtweezers,carefullypressdownonthespringcatch
tounlockitandliftcomponentslightly.Thenrepeatprocedureonthetwospring
catchesontheotherside,andremovetheregulator(controller).
•RemovetheoldO-Ring,andmountthenewO-Ringstotheregulator(controller).
•Slideregulatorinparallelpositiontothedistributionblock,forthefourspring
catchestosnapintoposition.
34
KISS™/KISS™Multilead
22745010H
Page 35

•ConrmthattheO-Ringsarecorrectlypositioned.
•Performthefollowingtests:
•“VisualCheck”onpage19
TestinginstructionsforKISS™andKISS™Multilead
•“SystemFunctionCheck”onpage19
•“ElectrodeLeadValveLeak”onpage21
•“CleaningandDisinfectingtheInstrumentSurfaces”onpage37
22745010H
KISS™/KISS™Multilead
35
Page 36

TestinginstructionsforKISS™andKISS™Multilead
36
KISS™/KISS™Multilead
22745010H
Page 37

Maintenance,Cleaningand
Troubleshooting(KISS™/KISS™
Multilead)
Maintenance
Theapplicationsystemandnegativepressurepumpdonotrequireanymaintenance.
CleaningandDisinfectingtheInstrument
Surfaces
4
WARNING:
SHOCKHAZARD-DisconnecttheinstrumentfromtheECGrecorderorthemains
plugofthepowersupplyunitbeforecleaninganddisinfectingtheinstrument
surfaceand/ormaintenance!
UseamoistclothtowipethepoleandtheCAM-USBA/TKISS™interfaceclean.Do
notallowliquidtoentertheequipment.Whenthesurfaceisverysoiled,useamild
soapsolutionordetergent.Neveruseabrasivecleaners.
Alldisinfectantscontainingalcoholcanbeused.Followingaresomeproducts
foundinthelistofdisinfectantsissuedbyDGHM(GermanSocietyforHygieneand
Microbiology):FunktioseptA,Mucocit-BorMeliseptol.
CleaningandDisinfectingthePatientand
Leadwires
Disconnectthecablesfromthesystembeforecleaninganddisinfectingthem.
Whendisconnectingtheplug-inconnectorsalwayspullontheplugandnotthecables.
22745010H
KISS™/KISS™Multilead
37
Page 38

Maintenance,CleaningandTroubleshooting(KISS™/KISS™Multilead)
Cleanthecableswithaclothmoistenedwithsoapwater.Useadisinfectantfor
disinfectionoftheleadwiresandcables.
CAUTION:
EQUIPMENTDAMAGE-Undernocircumstancesshouldleadwiresbeimmersedin
thecleaningsolutionordisinfectant.Evenifthesystemwerenotundernegative
pressure,theliquidwouldenterthesystem,spoilingthelterdisksandclogging
thepneumaticsystem.
CleaningElectrodesandElectrodeCare
Forathoroughcleaningtheelectrodecable(3)canbedisconnectedfromthe
electrode(4).Whendoingthisitispracticaltodisconnectthecableplug.
Thesuctioncup(6)cannowberemovedfromtheelectrode.Thispermitsaccessto
thelterdisk(5),whichcanbereplaced.
•Thesurfaceofelectrodeswithminorcontaminationcanbecleanedwithacloth
moistenedwithwateranddetergent.
CAUTION:
EQUIPMENTDAMAGE-Donotscrapeelectrodeswithmetalobjects.The
silver-silverchloridestructurecouldbedamagedbythisandloseitsstabilizing
effectontheelectrodeinterface.
•Todisinfecttheelectrodes,wipethemdownwithaclothmoistenedwith80%
alcohol.Alldisinfectantscontainingalcoholcanbeusedfordisinfection.
CAUTION:
EQUIPMENTDAMAGE-Afterusedonotimmersetheelectrodesintothe
cleaningsolution.Theliquidenteringthenegativepressuresystemwouldsoak
andhencespoilthelterdisk.
•ObservetheinformationgivenintheKISS™MultileadOperatorManual.
Forasysteminfrequentusereplacethelterdisksonceaweek!
ReplacingLeadwires
Theleadwiresaresubjecttowearandtearand,therefore,theycanbereplaced.The
leadwiresareavailablewithouttheelectrodesasspareparts(forCatalogNos.see
Chapter5“SparePartsList(KISS™/KISS™Multilead)”onpage41).
Theleadwiresarepluggedintothedistributorheadforeasyreplacement.When
removingtheleadpullontheplug.Whenplugginginanewleaddonotuseforce;the
pneumaticandelectricalplug-inconnectorscouldbecomedamaged.
38
KISS™/KISS™Multilead
22745010H
Page 39

Maintenance,CleaningandTroubleshooting(KISS™/KISS™Multilead)
Troubleshooting
Possibleerror:Electrodesdonotbecomeattachedbysuction,insufcient
suction,orelectrodesfalloff
Causes:
Remedies:Removekinkinnegativepressuretubing,establishconnection
Noorinsufcientnegativepressuredueto:
•Kinkinnegativepressuretubing
•Negativepressuretubingconnectiontopumpinterrupted
•Filterdisk(5)soiled
•Powersupplyinterrupted
•Leakinnegativepressuresystem
•BrokenorfailedregulatorO-Ring
topump,replacelterdisk,replaceregulatorO-Rings,connect
powersupply.
Instrumentationtestasin“EquipmentCheckforLeaksinthe
CompleteSystem,Pump,Distributor,ElectrodeLeadsandSuction
Cups”onpage20toidentifyleak.
Possibleerror:ECGleadsyieldpoorrecording
Cause:
Remedy:Measureresistanceasin“TestingElectricalFunction(KISS™)”
Electrodeleads,trunkcableshielding,orplugsinterrupted
onpage24.
22745010H
KISS™/KISS™Multilead
39
Page 40

Maintenance,CleaningandTroubleshooting(KISS™/KISS™Multilead)
40
KISS™/KISS™Multilead
22745010H
Page 41

SparePartsList(KISS™/KISS™
Multilead)
SingleLeadwiresforKISS™andKISS™Multilead
5
30344365
30344366
30344367
30344368
30344369
30344370
30344371
30344372
30344373
30344374
30344375
30344376
2022862-001
2022862-003
2022862-004
2022862-005
2022862-006
2022862-007
2022862-008
2022862-009
LeadwireC1
LeadwireC2
LeadwireC3
LeadwireC4
LeadwireC5
LeadwireC6
LeadwireF
LeadwireL
LeadwireR
LeadwireN
LeadwireNst
LeadwireNax
LeadwireI
LeadwireE
LeadwireM
LeadwireH
LeadwireA1
LeadwireA2
LeadwireA3
LeadwireA4
LeadwireSetsforKISS™
38401590
38401591
LeadwireSetsforKISS™Multilead
2024037-001
2024038-001
22745010H
LeadwireSet(10-fold)
LeadwireSet(12-fold)
LeadwireSetStandard
UpgradeLeadwireSetFrank
KISS™/KISS™Multilead
41
Page 42

SparePartsList(KISS™/KISS™Multilead)
LeadwireSetsforKISS™Multilead(cont'd.)
2024039-001
2024040-001
21732801
RegulatorsforKISS™/KISS™Multilead
30344364
30344486
30344519
30344549
O-RingsforKISS™/KISS™Multilead
2100929-001
DistributorBlockforKISS™
30344453
30344461
30344460
30344462
UpgradeLeadwireSetNehb
UpgradeLeadwireSetA1-A4
SuctionelectrodeforKISS™/KISS™Multilead
International,withGElogo
OEMversionwithoutlogo
VersionforZimmer
VersionforHP
FRUO-RingKISS
Exchange10-folddistributorblockwithregulator,trunk
cableandGEHealthcareconnector
Exchange12-folddistributorblockwithregulator,trunk
cableandGEHealthcareconnector
Exchange10-folddistributorblockwithregulatorwithout
trunkcable
Exchange12-folddistributorblockwithregulatorwithout
trunkcable
TrunkCablesforKISS™
22341401
22341402
22341403
22341409
22341410
22341415
42
Trunkcablefor10-leadKISS™,standard(CardioSmartand
Corina)
Trunkcablefor12-leadKISS™,standard(CardioSmartand
Corina)
Trunkcablefor10-leadKISS™,universal(non-GEECG
devices)
Trunkcable,12-leadKISS™,Schwarzer
Trunkcable,10-leadKISS™(MicroSmart/MAC™1200/
MAC™1100)
Trunkcable,12-leadKISS™(MicroSmart/MAC™1200/
MAC™1100)
KISS™/KISS™Multilead
22745010H
Page 43

TrunkCablesforKISS™(cont'd.)
SparePartsList(KISS™/KISS™Multilead)
22341413
22341414
Trunkcable,10-leadKISS™(CASE™8000)
Trunkcable,10-leadKISS™(MAC™5000)
DistributorBlockPartsforKISS™Multilead
2028718-001
30344427
2056303-001FRUASSYTUBEAIRKISSMUL TILEAD
DistributorblockforKISS™Multilead
Blindplug
MiscellaneousAdaptersforKISS™withUniversalConnector
22336110
22336118
forMAC™1100/1200andMAC™500(MicroSmart)
forAM4andAM5
MiscellaneousItemsforCAM-14Holder
2022307-001
2022298-001PCBConnector
2022309-001
CAM-14holder
Connectorhousing
LuerLockConnector
92916655
92916656
PowerSupply(Pump)forKISS™
2080819-001
HolderforKISS™System
30344191
50465702
43252133
43252306
43252346
43252347
50465855
43252307
Nutcap
Nippleplug
SetExternalSuctionpump100-240V
Telescopearm
Stay(21)
Jointtelescopearm(18)
Brake
CaseShellright(15)
CaseShellleft(15)
Hook(20)
CableHolder(19)
22745010H
KISS™/KISS™Multilead
43
Page 44

SparePartsList(KISS™/KISS™Multilead)
44
KISS™/KISS™Multilead
22745010H
Page 45

TestingInstructionsforKISS™and
KISS™Multilead
1.TestPlate(p/n30344490)
2.Y-Adapter(p/n30344489)
6
3.AdapterExtensionwithSyringe(p/n
2081397-001)
22745010H
KISS™/KISS™Multilead
45
Page 46

TestingInstructionsforKISS™andKISS™Multilead
4.DummyPlugforKISS™(p/n30344488)
NOTE:
Model1wasreplacedwithmodel2.
46
KISS™/KISS™Multilead
22745010H
Page 47

Specications
KISS™-10,KISS™-12
7
ElectrodeapplicationsystemforacquiringECGsignalsfromapatient
Electrodedistributor
for10or12plug-inelectrodeleads
Electrodes
silver-silverchlorideelectrodes,withconnectors
Debrillationprotection
bymeansofintegratedcurrentandvoltagelimitation
Signalconnector
15-pinconnection,IECpinconguration
1=C2
2=C3
3=C4=Nap
4=C5
5=C6
6=shield
7=bridgeto8(onlywith12electrodes)
8=bridgeto7(onlywith12electrodes)
9=R
10=L
11=F
12=C1
13=Nst
14=N
15=Nax
22745010H
KISS™/KISS™Multilead
47
Page 48

Specications
Negativepressureadjustmentforpumpmodule
continuouslyadjustablebetween80and220mbar;
inpositionnegativepressure<30mbar
Mechanicaldata
swivelarmlength75to125cm(extendible)
verticalswivelrange
horizontalswivelrange
lengthofpatientcablefromelectrodedistributortoECGdeviceapprox.250cm
lengthofelectrodeleadslimbs140cm
lengthofnegativepressuretubeconnectedtoseparatepumporECGdevice:
250cm
45°
360°
chest120cm
Environment
Operation
•temperaturebetween+10to+40°C
•relativehumiditybetween25and95%,non-condensing
•atmosphericpressurebetween700and1060hPa
Storageandtransport
•KISSw/opump:
temperaturebetween-30°Cand+70°C
KISSwithpumpandKISSpump:
temperaturebetween-30°Cand+60°C
•relativehumiditybetween10and95%,non-condensing
•atmosphericpressurebetween500and1060hPa
Weight
electrodedistributorwith12electrodesapprox.800g
KISS™Multilead
ElectrodeapplicationsystemforCAM-14acquisitionmoduletoacquireECGsignals
fromapatient
Electrodedistributor
formax.14leads
Electrodes
silver-silverchlorideelectrodes,withconnectors
Debrillationprotection
bymeansofintegratedcurrentandvoltagelimitation
48
KISS™/KISS™Multilead
22745010H
Page 49

Specications
Signalconnector
ConnectortoaCAM-14AcquisitionModule
Negativepressureadjustmentforpumpmodule
continuouslyadjustablebetween80and220mbar;inpositionnegativepressure
<30mbar
Mechanicaldata
swivelarmlength75to125cm(extendible)
verticalswivelrange
horizontalswivelrange
lengthofelectrodeleadslimbs140cm
lengthofnegativepressuretubeconnectedtoseparatepumporECGDevice:
270cm
45°
360°
chest120cm
Environment
Operation
•temperaturebetween+10°Cand+40°C
•relativehumiditybetween25and95%,non-condensing
•atmosphericpressurebetween700and1060hPa
Storageandtransport
•temperaturebetween-30°Cand+70°C
•relativehumiditybetween10and95%,non-condensing
•atmosphericpressurebetween500and1060hPa
Weight
electrodedistributorwith14electrodesapprox.950g
ExternalSuctionPumpwithPowerSupply
Powersupply
fromthepowerline;designincompliancewithIEC60601-1/EN60601,protection
classII;forratedlinevoltagesfrom100to240VAC,50to60Hz(withoutswitchover)
inputoperatingvoltagerange
inputcurrent
outputvoltage
outputcurrent
22745010H
KISS™/KISS™Multilead
100to240VAC
700mA-350mA
12V
2.5A
49
Page 50

Specications
Dimensions(pumpmodule)
•length130mm
•width65mm
•height70mm
Weight
pumpmoduleapprox.410g,incl.powersupply:680g
50
KISS™/KISS™Multilead
22745010H
Page 51

A
Schematics
AssemblyDrawings
21612101/08
21612201
2022865-001
ElectrodeApplicationSystemKISS™10
ElectrodeApplicationSystemKISS™12
ElectrodeApplicationSystemKISS™Multilead
22745010H
KISS™/KISS™Multilead
51
Page 52

Schematics
52
KISS™/KISS™Multilead
22745010H
Page 53
Page 54

GEMedicalSystems
InformationTechnologies,Inc.
8200WestTowerAvenue
Milwaukee,WI53223USA
Tel:+14143555000
+18005587044(USOnly)
GEMedicalSystems
InformationTechnologiesGmbH
MunzingerStraße5
79111FreiburgGermany
Tel:+497614543-0
Fax:+497614543-233
AsiaHeadquarters
GEMedicalSystems
InformationTechnologies,Inc.
Asia;GE(China)Co.,Ltd.
1HuatuoRoad
ZhangjiangHi-techParkPudong
Shanghai,People'sRepublicofChina201203
Tel:+862138777888
Fax:+862138777402
GEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompany,goingtomarketas
GEHealthcare.
www.gehealthcare.com