GE INVENIA ABUS SET-UP AND BASIC SERVICE MANUAL_SM_4700-0021-00_6 CP 100 and CP 200TM 12-Lead Resting Electrocardiograph

Page 1
Technical Publication

4700-0021-00

Rev. 6
Invenia ABUS System Setup and Basic Service Manual
Operating Documentation
Copyright 2014, 2015 © by General Electric Company Inc. All Rights Reserved.
Page 2
Regulatory Requirement
InveniaTM Automated Breast Ultrasound System (ABUS) complies with regulatory requiem, concerning medical devices, and is certified by TUV Rheinland LGA Products GmbH (CE
This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices.
This manual is a reference for the Invenia ABUS. It applies to Version 1 Revision 0.0 software for the Invenia ABUS System Setup.
0197
).
Page 3
Change history
/Revision History
Date
Revision
Rev. 1 2013-11-4 Preliminary
Rev. 2 2014-01-16 Updated Version
Rev. 3 2014-02-08 Update Manual Name and Quality Assurance Personnel
Rev.4 2014-08-20 UDI Label Update
Rev. 5 2015/02/03 Workstation software only option update
Rev. 6 2015/07/09 CE Mark Labeling Update
(YYYY-MM-DD)
Reason for change
Invenia ABUS – System Setup and Basic Service Manual i-1 4700-0021-00 Rev. 6
Page 4
Important precautions
Translation policy
i-2 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 5
Invenia ABUS – System Setup and Basic Service Manual i-3 4700-0021-00 Rev. 6
Page 6
i-4 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 7
Invenia ABUS – System Setup and Basic Service Manual i-5 4700-0021-00 Rev. 6
Page 8
i-6 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 9
Invenia ABUS – System Setup and Basic Service Manual i-7 4700-0021-00 Rev. 6
Page 10
i-8 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 11
Invenia ABUS – System Setup and Basic Service Manual i-9 4700-0021-00 Rev. 6
Page 12
i-10 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 13
Damage in transportation
All packages should be closely examined at time of delivery. If damage is apparent, write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
Certified electrical contractor statement - For USA Only
All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
Invenia ABUS – System Setup and Basic Service Manual i-11 4700-0021-00 Rev. 6
Page 14
Omission and errors
If there are any omissions, errors or suggestions for improving this documentation, contact the Global Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details.
Mail the
information to: Service Documentation
447 Indio Way SUNNYVALE, CA 94085 USA Tel: (408) 328-7924 Fax: (408) 738-9156
GE employees should use the complaint handling system to report service documentation issues.
These issues will then be in the internal problem reporting tool and communicated to the writer.
i-12 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 15
Service Safety Considerations
DANGER
WARNING
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, refer to Chapter 1 in the Service Manual.
Invenia ABUS – System Setup and Basic Service Manual i-13 4700-0021-00 Rev. 6
Page 16
Legal notes
Trademarks
Copyrights
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission.
GE may revise this publication from time to time without written notice.
All products and their name brands are trademarks of their respective holders.
All Material Copyright 2014, 2015 © by General Electric Company Inc. All Rights Reserved.
i-14 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 17
Translation policy - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-2 Damage in transportation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-11 Certified electrical contractor statement - For USA Only - - - - - - - - - - - - i-11 Omission and errors - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-12 Service Safety Considerations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-13 Legal notes - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-14 Trademarks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-14 Copyrights - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-14
Table of Contents Chapter 1 — Introduction
Manual Overview
Contents in this manual - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-2 Typical users of the System Setup and Basic Service Manual- - - - - - - - 1-3 Invenia ABUS models covered by this manual- - - - - - - - - - - - - - - - - - - 1-3
Product Overview
Invenia ABUS Scan Station and System - - - - - - - - - - - - - - - - - - - - - - - 1-4 Product description in this Manual - - - - - - - - - - - - - - - - - - - - - - - - - - 1-11
Important conventions
Conventions used in book - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-12 Standard hazard icons - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-14 Standard Icons that indicate that a special procedure is to be used - - - 1-15 Product icons - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-15 Labels locations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-15
Safety considerations
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-16 Human Safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-16 Lockout/Tagout (LOTO) requirements- - - - - - - - - - - - - - - - - - - - - - - - 1-20 Dangerous procedure warnings - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-21 Electrical safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-22 Mechanical safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-23
Electromagnetic compatibility (EMC)
What is EMC? - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-26 Compliance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-26 Electrostatic discharge (ESD) prevention - - - - - - - - - - - - - - - - - - - - - 1-27
Customer assistance
Contact information - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-28 Phone numbers for Customer Assistance - - - - - - - - - - - - - - - - - - - - - 1-29 System manufacturer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-30

Table of Contents

Invenia ABUS – System Setup and Basic Service Manual i-15 4700-0021-00 Rev. 6
Page 18
Returning parts
Returning the probe and repair parts - - - - - - - - - - - - - - - - - - - - - - - - 1-31
Chapter 2 — Site Preparations
General Invenia ABUS requirements
Invenia ABUS Scan Station environmental requirements - - - - - - - - - - - 2-2 Electrical requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4 EMI limitations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-5
Facility needs
Purchaser responsibilities - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8 Time and manpower requirements - - - - - - - - - - - - - - - - - - - - - - - - - - 2-12 Facility needs - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-12
Networking setup requirements
Scanner connected to hospital’s network - - - - - - - - - - - - - - - - - - - - - 2-27 InSite Requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-27 Purpose of the DICOM network function - - - - - - - - - - - - - - - - - - - - - - 2-27 DICOM setup requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-28
Environmental Dangers
Patient Vicinity UL60601-1 (USA) - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-30
Chapter 3 — System Setup
Setup reminders
Average setup time - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2 Setup warnings- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
Receiving and unpacking the equipment
Warnings for receiving and unpacking - - - - - - - - - - - - - - - - - - - - - - - - 3-4 The Tilt & Shock indicators - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-5 Receiving the Invenia ABUS - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-6 Unpacking the Invenia ABUS - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-11
Packing materials - recycling information
Packing materials - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-15
Preparing for setup
Verify customer order - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-16 Physical inspection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-16 EMI protection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-16 Product Labels and Descriptions - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-17
Completing the setup
Electrical specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-22 Connections on the Rear Panel - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-23
Logging In System Customization
General Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-27 Scanning Protocols - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-28 Worklist Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-29 Storage Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-31 System Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-32 System Configuration (Administrator) - - - - - - - - - - - - - - - - - - - - - - - - 3-39 About Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-43 InSite Setup (Administrative User) - - - - - - - - - - - - - - - - - - - - - - - - - - 3-44
i-16 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 19
Connectivity overview
Physical connection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-48
Paperwork after setup
User’s Manual(s) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-49
Chapter 4 — Functional Checks
General procedures
General Precautions - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2 Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-3 Power ON/Boot Up - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-5 Moving and Transporting the Invenia ABUS - - - - - - - - - - - - - - - - - - - - 4-7 Where are the User Manuals and the Service Manual? - - - - - - - - - - - 4-11
Functional checks
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-12 Performance Functional Checks- - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-12 Mechanical Functions Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-14 Connectivity Functional Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-15
Chapter 5 — Components and Functions (Theory)
Invenia ABUS Components and Functions (Theory)
Invenia ABUS General Description - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-2 Air Filter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-3 Scanner Assembly - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-4
Chapter 6 — Service Adjustments
Service Adjustments
Chapter 7 — Diagnostics/Troubleshooting
Noise troubleshooting
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-2 General recommendations- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-3 EMI prevention/abatement - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-4 Different Power Outlet- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-5 Different system - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-5 Different location- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-5 Disconnect external cables - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-5 Overview of types of noise - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-6
InSite
InSite on the Invenia ABUS Scan Station - - - - - - - - - - - - - - - - - - - - - - 7-9 InSite on the Invenia ABUS Review Software - - - - - - - - - - - - - - - - - - 7-11 InSite Functionality - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-12
Troubleshooting and System Messages
Invenia ABUS Scan Station Troubleshooting and System Messages- - 7-15 Workstation Troubleshooting and System Messages - - - - - - - - - - - - - 7-19
Chapter 8 — Replacement Procedures
Warnings and important information
Warnings - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-2 Manpower - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-3 Returning/shipping probes and repair parts - - - - - - - - - - - - - - - - - - - - - 8-4
Invenia ABUS – System Setup and Basic Service Manual i-17 4700-0021-00 Rev. 6
Page 20
Parts Disposal
Used Media and Used Parts Disposal- - - - - - - - - - - - - - - - - - - - - - - - - 8-5
Cleaning the Filter / Replacement
Filter Replacement Functional Check - - - - - - - - - - - - - - - - - - - - - - - - - 8-7
Replacing the Transducer
Transducer Functional Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-10 B-Mode Functional Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-10 3D Volume Functional Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-10 Performing the Exam - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-10
Replacing the Power Cord Monitor (Touch Screen)
Replacing the Monitor - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-12
System Backup and Software Load
Notice against user modification- - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-16
Internal Component Replacement
Chapter 9 — Renewal Parts
Expected Service Life Reordering Supplies
Chapter 10 — Care & Maintenance
System Care and Maintenance
Tools and Maintenance Schedule- - - - - - - - - - - - - - - - - - - - - - - - - - - 10-2 Invenia ABUS Specific Information - - - - - - - - - - - - - - - - - - - - - - - - - - 10-4 Workstation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-39
Quality Assurance
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-40
Index
i-18 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 21
Chapter 1

Introduction

This chapter describes important issues related to safely servicing the Invenia ABUS Scan Station, Invenia ABUS Workstation, and Invenia ABUS Review Software. The service provider must read and understand all the information presented here before installing or servicing the Invenia ABUS Scan Station, Invenia ABUS Workstation, and Invenia ABUS Review Software.
Invenia ABUS – System Setup and Basic Service Manual 1-1 4700-0021-00 Rev. 6
Page 22
Introduction
Contents in this manual
This manual provides installation and service information for the Invenia ABUS Scan Station, Invenia ABUS Workstation, and Invenia ABUS Review Software.
The manual is divided into ten chapters.
In the beginning of the manual, before Chapter 1, you will find the Revision overview, the Important precautions including
Translation policy, Damage in transportation, Certified electrical contractor statement, Omission & errors, Service safety considerations and Legal notes, and the Table of Contents.
An Index has been included after Chapter 10.

Manual Overview

Table 1-1: Contents in this manual
Chapter Number
1. Introduction Contains a content summary and warnings.
2. Site preparations Contains pre-setup requirements for the Invenia ABUS.
3. System Setup Contains setup procedure with procedure checklist.
4. General Procedures and
5. Components and
6. Service Adjustments Contains instructions on how to make any available adjustments
7. Diagnostics/
8. Replacement procedures Provides disassembly procedures and reassembly procedures
9. Renewal Parts Contains a list of replacement parts for Invenia ABUS.
10. Care & Maintenance Provides maintenance procedures for Invenia ABUS.
ChapterTtitle Description
Contains functional checks that must be performed as part of the
Functional Checks
Functions (Theory)
Troubleshooting
installation, or as required during servicing and periodic maintenance.
Contains and functional explanations of the electronics.
to the Invenia ABUS.
Provides procedures for running diagnostic or related routines for the Invenia ABUS.
for all changeable FRU.
Index Index A quick way to the topic you’re looking for.
1-2 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 23
Manual Overview
Typical users of the System Setup and Basic Service Manual
GE Service Personnel (setup, maintenance, etc.)
Invenia ABUS User
Licensed Hospitals’ Service Providers
• Technical Trainers
Invenia ABUS models covered by this manual
Table 1-2: Invenia ABUS Models and Hardware/Software Compatibility
Model
Number Description System SW Appl. SW
H5014SS Invenia ABUS Scan Station Windows 7
Professional
H5014WS Invenia ABUS Workstation Windows 7
Professional
H5014SW Invenia ABUS Review Software Windows 7
Professional
The InveniaTM Automated Breast Ultrasound System (ABUS) comes with a review workstation available as software only (Invenia ABUS Review Software) or with dedicated hardware (Invenia ABUS Workstation).
NOTE: When not otherwise specified, the contents of this manual
applies to all Invenia ABUS models.
Version 1 Revision x.x
Version 2 Revision x.x
Version 2 Revision x.x
Invenia ABUS – System Setup and Basic Service Manual 1-3 4700-0021-00 Rev. 6
Page 24
Introduction

Product Overview

Invenia ABUS Scan Station and System
The Invenia ABUS is composed of the following components:
Invenia ABUS Scan Station
Invenia ABUS Workstation or Invenia ABUS Review Software
Figure 1-1. Invenia ABUS Scan Station, Invenia ABUS Workstation, or Invenia ABUS
Review Software
After performing the exam study on the Scan Station, the exam study is transferred to the Invenia ABUS over the DICOM network. After reviewing the patient’s study on the workstation, the study is sent to PACS for archiving, or to some other form of manual archive.
1-4 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 25
Invenia ABUS Scan Station and System (continued)
For more information on the Invenia ABUS Scan Station, please refer to the Invenia ABUS Scan Station Basic User Manual, Part Number 4700-0014-00.
For more information on the Workstation, please refer to the Invenia ABUS Workstation and Invenia ABUS Review Software Basic User Manual, Part Number 4700-0015-00.
Product Overview
Invenia ABUS – System Setup and Basic Service Manual 1-5 4700-0021-00 Rev. 6
Page 26
Introduction
Invenia ABUS Scan Station and System (continued)
The InveniaTM Automated Breast Ultrasound System (ABUS) comes with a review workstation available as software only (Invenia ABUS Review Software) or with dedicated hardware (Invenia ABUS Workstation).
There are three (3) distinctions to note with regard to the naming conventions used in this manual:
Workstation. A generic name for computer hardware, provided by GE or by a 3rd party, running the Invenia ABUS Review Software.
3rd Party Workstation. Computer hardware, provided by a 3rd party, running the Invenia ABUS Review Software.
Invenia ABUS Workstation. Computer hardware, provided by GE, running the Invenia ABUS Review Software.
Figure 1-2. Invenia ABUS Scan Station and Invenia ABUS Workstation
1-6 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 27
Invenia ABUS Scan Station
The Invenia ABUS is intended to be used in medical practices and in clinical departments and serves the purposes as a review station for the radiologist’s interpretation of images, electronic documentation of examinations in the form of text and images and generation of medical reports.
Ultrasound engine with articulating arm
Scanner Assembly with C15-6XW Reverse Curve transducer
Rotating Touch Screen monitor
Workstation (please refer to the Workstation User Manual)
Product Overview
TM
Figure 1-3. Invenia ABUS Overview
Table 1-3: Main System Components
1. Rotating Touch Screen Monitor
2. Ultrasound Coupling Lotion Holders
3. Power Cord, Mains Power Switch, and Network Connector (not shown, on back of system)
4. System Standby Button (On/ Off)
5. Articulating Arm
6. Scanner Assembly with
C15-6XW Reverse Curve transducer
TM
This section will walk you through how to use the Touch Screen monitor and scanner assembly to perform the ABUS scan. In addition, this section will discuss how to move the system and set the front and rear brakes.
NOTE: The Invenia ABUS Scan Station is ideally positioned with the
articulating arm adjacent to the top right or left corner of the bed.
Invenia ABUS – System Setup and Basic Service Manual 1-7 4700-0021-00 Rev. 6
Page 28
Introduction
Invenia ABUS Workstation and Review Software
The Invenia ABUS Review Software will typically be used in a mammography reading room environment, normally with low ambient light levels next to a mammography reading station and a DICOM workstation.
The Invenia ABUS Review Software is designed to accept, transfer, display, store and process medical images and data. Invenia ABUS Review Software enables the user to optimize, measure and annotate the images. The Invenia ABUS Review Software provides images to enable Physicians to differentiate normal and abnormal breast tissue and is intended for use by Health Care Professionals only.
The Invenia ABUS Review Software displays three-dimensional data sets for viewing in three orthogonal planes and standard Ultrasound images, permitting Interpreting Physicians to quickly rerview, locate, and mark regions of interest.
The Invenia ABUS Workstation, if provided, includes the following components:
1. Monitor and Computer, with proprietary GE software
2. Keyboard
3. Mouse
1-8 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 29
Product Overview
Direct Network Connection Between the Invenia ABUS Scan Station and Workstation to PACS (typical configuration)
Exam studies are sent from the Invenia ABUS Scan Station to the Invenia ABUS Review Software directly and reviewed by the Interpreting Physician before they are archived.
Figure 1-4. Review Prior to Archival
Invenia ABUS – System Setup and Basic Service Manual 1-9 4700-0021-00 Rev. 6
Page 30
Introduction
Connection Through a Storage system, such as PACS
Exams are archived first to Storage and then reviewed by an Interpreting Physician at any Invenia ABUS connected to the network.
Storage and Archive
Figure 1-5. Connection through a PACS
The Workstation provides temporary storage for approximately 200 patient studies. The workstation can be configured to maintain 10, 20, 30, 40, or 50 percent of the available disk space free to ensure good workstation performance. If free space reaches a percentage under the configured limit specified by the user, a warning appears when logging into the Invenia ABUS application. Old reviewed studies are then deleted (oldest first) overnight or when the workstation is powered up the next time. Transfers from the Invenia ABUS Scan Station are rejected if free disk space falls below 5%. Scans will not transfer to the Workstation if there is not enough disk space. Suspended or un-reviewed studies are never automatically deleted in this manner. Deleted studies can only be re-displayed if the whole study was stored to PACS or external storage and re-imported.
1-10 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 31
Product Overview
Product description in this Manual
Overview of the Invenia ABUS Scan Station and Workstation
This Invenia ABUS is a high performance digital ultrasound imaging system with total data management via the Workstation.
Signal flows from the Probe Connector Panel through the Invenia ABUS Scan Station and finally to the monitor.
System configuration is stored on the Scan Station and Workstation. Each have their configuration locally.
All necessary software is loaded from the hard drive on power up.
History - hardware/software versions
Please refer to Chapter 9 in this manual.
Field Replaceable Units (FRUs)
Please refer to Chapter 8 in this manual.
How to turn the Invenia ABUS Scan Station and Workstation ON and OFF
Please refer to Chapter 4 in this manual.
How to check for hardware/software version
To verify the hardware versions on the boards, please refer to Chapter 4 in this manual.
Purpose of the operator manual(s)
The operator manuals should be fully read and understood before operating the Invenia ABUS.
The online versions of the operator manuals are available via the Help function on Invenia ABUS’s operator panel.
Invenia ABUS – System Setup and Basic Service Manual 1-11 4700-0021-00 Rev. 6
Page 32
Introduction
Conventions used in book
Important conventions, used in this document, are described next.
Model designations
This manual covers the Invenia ABUS Scan Station and Workstation (Invenia ABUS Workstation and Invenia ABUS Review Software).
‘Invenia ABUS models covered by this manual’ on page 1-3.
Icons

Important conventions

Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels, and conventions used on the product and in the service information are described in this chapter.
1-12 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 33
Safety precaution messages
Various levels of safety precaution messages may be found on the equipment and in the service information. The different levels of concern are identified by a flag word that precedes the precautionary message. Known or potential hazards to personnel are labeled in one of three ways:
• DANGER
• WARNING
• CAUTION
Important conventions
DANGER
WARNING
CAUTION
NOTE: Notes are used to provide important information about an item
NOTE: Be sure to read the notes; the information contained in a note
Danger is used to indicate the presence of a hazard that will cause severe personal injury or death if the instructions are ignored.
Warning is used to indicate the presence of a hazard that can cause severe personal injury and property damage if instructions are ignored.
Caution is used to indicate the presence of a hazard that will or can cause minor personal injury and property damage if instructions are ignored. Equipment damage possible.
or a procedure.
can often save you time or effort.
Invenia ABUS – System Setup and Basic Service Manual 1-13 4700-0021-00 Rev. 6
Page 34
Introduction
Standard hazard icons
Important information will always be preceded by either the exclamation point (!) contained within a triangle, or the symbols for “Danger”, “Warning” or “Caution”, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could possibly cause harm. Even if a symbol isn’t used in this manual, it may be included for your reference.
Table 1-4: Standard hazard icons
ELECTRICAL
MECHANICAL
HEAT
PINCH
MAGNET PACEMAKER
1-14 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 35
Important conventions
Standard Icons that indicate that a special procedure is to be used
Some others icons make you aware of specific procedures that should be followed.
Table 1-5: Standard Icons that indicates that a special procedure is to be used
Avoid Static Electricity Tag and Lockout Wear Eye Protection
Hand Protection Foot Protection Wear Eye Protection
Be sure to read the notes; the information contained in a note can often save you time or effort.
Product icons
Always refer to the product’s User Manual for a full list of labels used on the Invenia ABUS Scan Station and Workstation.
Labels locations
It is important to refer to the current revision of the Invenia ABUS Scan Station and Workstation User Manuals for a full list of product labels prior to servicing the system.
Invenia ABUS – System Setup and Basic Service Manual 1-15 4700-0021-00 Rev. 6
Page 36
Introduction
Introduction
Human Safety

Safety considerations

The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
WARNING
This equipment contains dangerous voltages that are capable of serious injury or death.
If any defects are observed or malfunctions occur, stop operating the equipment and perform the proper action for the patient. Inform a qualified service person and contact a Service Representative for information.
There are no user serviceable components inside the console. Refer all servicing to qualified service personnel only.
Ensure that unauthorized personnel do not tamper with the unit.
The warranty is void if covers have been removed.
Only personnel who have participated in an Invenia ABUS Scan Station and Workstation Training Seminar are authorized to service the internal components of this equipment.
1-16 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 37
Human Safety (continued)
Safety considerations
DANGER
DANGER
WARNING
WARNING
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
WHERE APPLICABLE, THERE ARE SEVERAL PLACES ON THE IMAGING ENGINE CHASSIS, THE AC DISTRIBUTION, DC DISTRIBUTION, AND TIP BOARD THAT ARE DANGEROUS. BE SURE TO POWER DOWN THE SYSTEM, TURN OFF THE MAINS POWER SWITCH AND DISCONNECT THE MAINS POWER CABLE FROM THE WALL OUTLET BEFORE YOU REMOVE ANY PARTS. BE CAUTIOUS WHENEVER POWER IS STILL ON AND COVERS ARE REMOVED.
If the covers are removed from an operating Invenia ABUS Scan Station, some metal surfaces may be warm enough to pose a potential heat hazard if touched, even while in shutdown mode.
Because of the limited access to cabinets and equipment in the field, placing people in awkward positions, GE has limited the lifting weight for one person in the field to 16 KG (35 LBS). Anything over 16 KG (35 LBS) requires 2 people.
WARNING
Attempts to move the Invenia ABUS Scan Station considerable distances or on an incline by one person could result in injury or damage or both.
Invenia ABUS – System Setup and Basic Service Manual 1-17 4700-0021-00 Rev. 6
Page 38
Introduction
Human Safety (continued)
WARNING
WARNING
WARNING
WARNING
Explosion Warning
DO NOT operate the equipment in an explosive atmosphere. Operation of any electrical equipment in such an environment constitutes a definite safety hazard.
DO NOT substitute parts or modify equipment
Because of the danger of introducing additional hazards, ONLY install GE approved parts. DO NOT perform any unauthorized modification of the equipment.
Ensure that the Invenia ABUS Scan Station is turned off and unplugged
Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation.
Invenia ABUS Scan Station components may be energized. Always refer to LOTO warnings and cautions
Risk of electrical shock, Invenia ABUS Scan Station must be turned off and disconnected from power source. Cord must be controlled at all times.
Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation.
Invenia ABUS Scan Station components may be energized. Always refer to LOTO warnings and cautions
WARNING
1-18 Invenia ABUS – System Setup and Basic Service Manual
Tilting the console requires two people in order to avoid injury to service personnel and damage to the equipment.
4700-0021-00 Rev. 6
Page 39
Human Safety (continued)
Safety considerations
WARNING
WARNING
WARNING
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety glasses, and kneeling pad, to reduce the risk of injury.
Beware of possible sharp edges on all mechanical parts. If sharp edges are encountered, the appropriate PPE should be used to reduce the risk of injury.
Wear all PPE including gloves as indicated in the chemical MSDS.
Invenia ABUS – System Setup and Basic Service Manual 1-19 4700-0021-00 Rev. 6
Page 40
Introduction
Lockout/Tagout (LOTO) requirements
Lockout/Tagout
Follow Lockout/Tagout requirements by ensuring you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout (LOTO):
1. Plan and prepare for shutdown.
2. Shutdown the equipment.
3. Isolate the equipment.
4. Apply Lockout/Tagout Devices.
5. Control all stored and residual energy.
6. Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Energy Control and Power Lockout
WARNING
Energy Control and Power Lockout for Invenia ABUS.
When servicing parts of the Invenia ABUS Scan Station where there is exposure to voltage greater than 30 volts:
1. Follow LOCKOUT/TAGOUT procedures.
2. Turn off the Power Mains Switch.
3. Unplug the Invenia ABUS Scan Station.
4. Maintain control of the Invenia ABUS Scan Station power plug.
5. Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation.
Invenia ABUS Scan Station components may be energized.
1-20 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 41
Dangerous procedure warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
Safety considerations
DANGER
WARNING
WARNING
WARNING
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
If the covers are removed from an operating Invenia ABUS, some metal surfaces may be warm enough to pose a potential heat hazard if touched, even while in shutdown mode.
Explosion Warning
DO NOT operate the equipment in an explosive atmosphere. Operation of any electrical equipment in such an environment constitutes a definite safety hazard.
DO NOT substitute parts or modify equipment
Because of the danger of introducing additional hazards, ONLY install GE approved parts. DO NOT perform any unauthorized modification of the equipment.
Invenia ABUS – System Setup and Basic Service Manual 1-21 4700-0021-00 Rev. 6
Page 42
Introduction
Electrical safety
Safe practices
Follow these guidelines to minimize shock hazards whenever you are using the Invenia ABUS Scan Station:
To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
The Invenia ABUS Scan Station is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with safety ground.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector must meet international electrical standards
WARNING
Transducer
Connecting a Invenia ABUS to the wrong voltage level will most likely destroy it.
Follow these guidelines before connecting the transducer to the Invenia ABUS:
Inspect the probe prior to each use for damage or degradation to the:
• housing
• lens
• seal
• connector pins
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any liquid.
1-22 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 43
Mechanical safety
Safety considerations
WARNING
WARNING
WARNING
WARNING
While the software install procedure is designed to preserve data, ensure that all patient exams have transferred to the Workstation or PACS.
Ultrasound probes are highly sensitive medical instruments that can easily be damaged by improper handling. Use care when handling and protect from damage when not in use. do not use a damaged or defective probe. Failure to follow these precautions can result in serious injury and equipment damage.
Never use a probe that has fallen to the floor. Even if it looks OK, it may be damaged.
When the Invenia ABUS Scan Station is raised for a repair or moved along any incline, use extreme caution since it may become unstable and tip over.
CAUTION
Invenia ABUS – System Setup and Basic Service Manual 1-23 4700-0021-00 Rev. 6
Invenia ABUS weighs 123 kg (271 lbs.) when ready for use. Care must be used when moving it or replacing its parts.
Failure to follow the precautions listed below could result in injury, uncontrolled motion and costly damage.
ALWAYS:
be sure the pathway is clear
use slow, careful motions
use two people when moving on inclines or lifting more than 16 kg (35 lbs)
Page 44
Introduction
Mechanical safety (continued)
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
Before you move or transport the Invenia ABUS Scan Station, make sure to lock the Touch Screen monitor and Articulating Arm to prevent damage to the Invenia ABUS Scan Station.
To avoid injury when you move the Touch Screen monitor and the Articulating Arm, do not put your finger, hand, or object on the joint of the monitor or the monitor arm.
Keep the heat venting holes on the Touch Screen monitor unobstructed to avoid overheating of the monitor.
Do not transport Invenia ABUS in a vehicle without locking the casters (wheels) and securing it as described in the User Manual.
Use protective glasses during drilling, filing smooth surfaces, and during all other work where eyes need protection.
CAUTION
CAUTION
1-24 Invenia ABUS – System Setup and Basic Service Manual
Practice good ESD prevention. Wear an anti–static strap when handling electronic parts and even when disconnecting/ connecting cables.
Do not pull out or insert circuit boards while power is on.
4700-0021-00 Rev. 6
Page 45
Mechanical safety (continued)
Safety considerations
CAUTION
CAUTION
CAUTION
Do not operate this Invenia ABUS Scan Station unless all board covers and frame panels are securely in place. System performance and cooling require this.
Use safety shoes when doing work where there is any chance of foot injury.
Use protective gloves when working with sharp edges or when directed to wear PPE during a removal/replacement procedure.
CAUTION
NOTE: Special care should be taken when transporting the Invenia
Invenia ABUS – System Setup and Basic Service Manual 1-25 4700-0021-00 Rev. 6
Be careful not to pinch any of the cables.
ABUS in a vehicle:
• Before transporting, place the system in its special storage container.
Ensure that the system is firmly secured while inside the vehicle.
Secure system with straps or as directed otherwise to prevent motion during transport.
Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and erratic stops or starts.
Page 46
Introduction

Electromagnetic compatibility (EMC)

What is EMC?
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface. Inadequate compatibility results when a susceptible device fails to perform as intended due interference from its environment or when the device produces unacceptable levels of emission to its environment. This interference is often referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy, EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power supply.
Compliance
Invenia ABUS conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
For applicable standards, refer to the Safety Chapter of the Invenia ABUS Scan Station and Invenia ABUS Workstation and Invenia ABUS Review Software User’s Manuals.
NOTE: It is critical that all covers, screws, shielding, gaskets, mesh,
clamps, are in good condition, installed tightly without skew or stress. Proper installation following all comments noted in this service manual is required in order to achieve full EMC performance.
1-26 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 47
Electromagnetic compatibility (EMC)
Electrostatic discharge (ESD) prevention
WARNING
WARNING
DO NOT touch any boards with integrated circuits prior to taking the necessary ESD precautions.
Always connect yourself, via an arm-wrist strap, to the advised ESD connection point located on the rear of the Invenia ABUS Scan Station (near the power connector).
Follow general guidelines for handling of electrostatic sensitive equipment.
Risk of electrical shock, Invenia ABUS Scan Station must be turned off. Avoid all contact with electrical contacts, conductors and components. Always use non-conductive handles designed for the removal and replacement of ESD sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before making contact.
Invenia ABUS – System Setup and Basic Service Manual 1-27 4700-0021-00 Rev. 6
Page 48
Introduction
Contact information

Customer assistance

If this equipment does not work as indicated in this service manual or in the user manual, or if you require additional assistance, please contact the GE Service or appropriate support resource. Note the following:
1. System ID serial number.
2. Software version.
3. Date and time of occurrence.
4. Sequence of events leading to issue.
5. Is the issue repeatable?
Refer to Figure 3-1 to locate the System ID Serial Number on the Invenia ABUS Scan Station and Invenia ABUS Workstation. Specific label descriptions can be found in the user manuals.
1-28 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 49
Phone numbers for Customer Assistance
Table 1-6: Phone numbers for Customer Assistance
LOCATION PHONE NUMBER
Customer assistance
USA/Canada GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC 9900 Innovation Drive Wauwatosa, WI 53226
Latin America Service
EMEA GE Ultraschall Deutschland Gmbh & Co. KG Beethovenstrasse 239 Postfach 11 05 60, D-42655 Solingen Germany
EAGM Egypt Service Center UAE Service Center
Asia (Singapore) GE Ultrasound Asia Service Department - Ultrasound 298 Tiong Bahru Road #15-01/06 Central Placa Singapore 168730
Japan Support Center Phone:
Service: On-site 1-800-321-7937
Service Parts 1-800-558-2040
Application Support 1-800-682-5327 or
Application Support
Phone: +49 (0) 212-2802-652
Fax: +49 (0) 2122-8024-31
Service Tel
Tel: +65 6291-8528
Fax: +65-6291-7006
Fax:
1-262-524-5698
1-262-524-5300 1-262-524-5698
00202 2322 1252 00971 8003646
81-426-48-2940 81-426-48-2905
Australlia/New Zealand China India
South Korea
Online Services Ultrasound Asia
+(61) 1-800-647-855 +(86) 800-810-8188 +(91) 1800-425-8025 +(91) 1800-102-7750 +(91) 1800-425-7255 +(82) 2-1544--6119
Invenia ABUS – System Setup and Basic Service Manual 1-29 4700-0021-00 Rev. 6
Page 50
Introduction
System manufacturer
Table 1-7: System manufacturer
MANUFACTURER
U-Systems, Inc.
447 Indio Way
Sunnyvale, CA 94085
USA
1-30 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 51
Returning the probe and repair parts
Equipment being returned must be clean and free of blood and other infectious substances. Policy states that body fluids must be properly removed from any part or equipment prior to shipment. Employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment. Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the people who will receive or open this package.

Returning parts

Returning parts
NOTE: The US Department of Transportation (DOT) has ruled that
“items that were saturated and/or dripping with human blood that are now caked with dried blood; or which were used or intended for use in patient care” are “regulated medical waste” for transportation purposes and must be transported as a hazardous material.
NOTE: The USER/SERVICE staff should dispose of all the waste
properly, per federal, state, and local waste disposal regulations.
The Invenia ABUS is not meant to be used for long-term storage of patient data or images. The user is responsible for the data on the system.
If the system is sent for repair, please ensure that any patient information is backed up and erased from the system before shipping. It is always possible during system failure and repair to lose patient data. GE is not responsible for the loss of this data.
Invenia ABUS – System Setup and Basic Service Manual 1-31 4700-0021-00 Rev. 6
Page 52
Introduction
Returning the probe and repair parts (continued)
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes, GE will ascertain agreement from the customer. Patient information shall only be transferred by approved service processes, tools and devices restricting access, protecting or encrypting data where required, and providing traceability in the form of paper or electronic documents at each stage of the procedure while maintaining compliance with cross-border restrictions of patient information transfers.
1-32 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 53
Chapter 2

Site Preparations

This chapter provides the information required to plan and prepare for the setup of an Invenia ABUS Scan Station, Invenia ABUS Workstation, and Invenia ABUS Review Software. Included are descriptions of the facility and electrical needs to be met by the purchaser of the Invenia ABUS Scan Station, Invenia ABUS Workstation, and Invenia ABUS Review Software.
Invenia ABUS – System Setup and Basic Service Manual 2-1 4700-0021-00 Rev. 6
Page 54
Site Preparations

General Invenia ABUS requirements

Invenia ABUS Scan Station environmental requirements
If the Invenia ABUS Scan Station is very cold or hot
When unpacking the Invenia ABUS Scan Station, allow the temperature of the Invenia ABUS Scan Station to stabilize before powering up. The following table describes guidelines for reaching operational temperatures from storage or transport temperatures.
CAUTION
Cooling
If the Invenia ABUS Scan Station is very cold or hot, do not turn on its power until it has had a chance to acclimate to its operating environment.
Table 2-1: Invenia ABUS Scan Station acclimate time
The cooling requirement for a typical console Ultrasound system with monitor and on board peripherals, is up to 3800 BTU/h. This figure does not include cooling needed for lights, people, or other equipment in the room.
NOTE: Each person in the room places an additional 300 BTU/h
demand on the cooling system.
2-2 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 55
Lighting
General Invenia ABUS requirements
Bright light is needed for Invenia ABUS Scan Station setup, updates, and repairs. However, operator and patient comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source of Electromagnetic Interference (EMI) which could degrade image quality. These controls should be selected to minimize possible interference.
Invenia ABUS – System Setup and Basic Service Manual 2-3 4700-0021-00 Rev. 6
Page 56
Site Preparations
Electrical requirements
General requirements
The Invenia ABUS Scan Station requires a dedicated power and ground for the proper operation of its Ultrasound equipment. A separate power outlet with a minimum 10 amp circuit breaker for 100-240 VAC.
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power source to the Invenia ABUS Scan Station is only a conduit.
CAUTION
Power outage may occur. The Invenia ABUS requires a dedicated single branch circuit. To avoid circuit overload and possible loss of critical care equipment, make sure you do not have any other equipment operating on the same circuit.
Site power outlets
A dedicated AC power outlet must be within reach of the Invenia ABUS Scan Station without extension cords. Other outlets adequate for the external peripherals, medical and test equipment needed to support this Invenia ABUS Scan Station must also be present within 1 m (3.2 ft.) of the Invenia ABUS Scan Station. Electrical installation must meet all current local, state, and national electrical codes.
Unit power plug
Refer to Chapter 9 for Power Plug part numbers. If the Invenia ABUS Scan Station and Workstation arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required.
Power stability requirements
Voltage drop-out:
Max 10 ms.
Power transients (all applications):
Less than 25% of nominal peak voltage for less than 1 millisecond for any time of transient, including line frequency, synchronous, asynchronous, or aperiodic transients.
2-4 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 57
Power stability requirement
Table 2-2: Power Stability Requirement
General Invenia ABUS requirements
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
EMI limitations
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 s
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 s
Mains power quality should be that of a typical commercial or hospital environment.
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. They also generate EMI. The Invenia ABUS Scan Station and Invenia ABUS Workstation complies with limits as stated on the EMC label. However there is no guarantee that interference will not occur in a particular installation.
Possible EMI sources should be identified before the Invenia ABUS Scan Station and Workstation are set up.
Invenia ABUS – System Setup and Basic Service Manual 2-5 4700-0021-00 Rev. 6
Page 58
Site Preparations
EMI limitations (continued)
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Some of these sources include:
• medical lasers
• scanners
• cauterizing guns
• computers
• monitors
• fans
• gel warmers
• microwave ovens
• light dimmers
• mobile phones
in-house wireless phones (DECT phones)
wireless computer keyboard and mouse
air conditioning system
High Frequency (HF) surgery equipment
• general AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.
See: ‘EMI prevention/abatement’ on page 2-7 for EMI prevention tips.
2-6 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 59
General Invenia ABUS requirements
EMI prevention/abatement
Table 2-3: EMI prevention/abatement
EMI RULE DETAILS
Be aware of Radio Frequency sources
Ground the Ultrasound system
Replace all screws, Radio Frequency gaskets, covers, cores
Replace broken Radio Frequency gaskets
Do not place labels where Radio Frequency gaskets touch metal
Take care with cellular phones
• Keep the Invenia ABUS Scan Station at least 5 meters (15 feet) away from other EMI sources.
• Special shielding may be required to eliminate interference problems caused by high frequency, high powered radio or video broadcast signals.
Poor grounding is the most likely reason an Ultrasound system will have noisy images. Check grounding of the power cord and power outlet.
• After you finish repairing or updating the Ultrasound system, replace all covers and tighten all screws.
• Any cable with an external connection requires a magnet wrap at each end.
• Install all covers. Loose or missing covers or Radio Frequency gaskets allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an Radio Frequency gasket are broken, replace the gasket. Do not turn on the Ultrasound system until any loose metallic part is removed.
Where applicable, never place a label where Radio Frequency gaskets meet the Ultrasound system. Otherwise, the gap created will permit Radio Frequency leakage. Or, if a label has been found in such a position, move the label.
Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Invenia ABUS – System Setup and Basic Service Manual 2-7 4700-0021-00 Rev. 6
Page 60
Site Preparations
Purchaser responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre-installation work before delivery. Purchaser responsibility includes:
Procuring the materials required
Completing the preparations before delivery of the Ultrasound system (See Site Survey Planner below).
Paying the costs for any alterations and modifications not specifically provided in the sales contract
NOTE: All electrical installations that are preliminary to the positioning
of the equipment at the site prepared for the equipment must be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing must also be performed by qualified personnel. The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.

Facility needs

The desire to use a non–listed or customer provided product or to place an approved product further from the Ultrasound system than the interface kit allows, presents challenges to the installation team. To avoid delays during installation, such variances should be made known to the individuals or group performing the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the Ultrasound system. Carpet is not recommended because it collects dust and creates static. Potential sources of EMI (electromagnetic interference) should also be investigated before delivery. Dirt, static, and EMI can negatively impact Ultrasound system reliability.
2-8 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 61
Facility needs
Site Survey Planner
Use the Site Survey Planner on the next two pages to help with your planning activities.
You can also download this form (part number 9000-0014-01) from the Common
Documentation Library (http://www3.gehealthcare.com/en/Support/
Support_Documentation_Library).
Invenia ABUS – System Setup and Basic Service Manual 2-9 4700-0021-00 Rev. 6
Page 62
Site Preparations
Pre-Installation Site Survey Worksheet
SITE INFORMATION & CONTACT PERSONNEL
Facility Name:
Physical Address: _______________________________________________________________________________________________
Primary ABUS Contact:
Delivery/Install Contact:
IT/Network Admin Contact:
PACS or ABUS Image Data Storage Contact:
Secondary IT, Network Admin or PACS Contact:
To optimize the ecient implementation of your ABUS scanning prog ram, please consider purchase of a service contract.
________________________________________
____________________
____________________
____________________
____________________
____________________
Clinic Hours: _____ AM until _____ PM M-F Other______
Phone:
_____________________
Phone:
_____________________
Phone:
_____________________
Phone:
_____________________
Phone:
_____________________
E-Mail:
_________________________
E-Mail:
_________________________
E-Mail:
_________________________
E-Mail:
_________________________
E-Mail:
_________________________
INSTALLATION SCHEDULING INFORMATION
Scheduled Installation Date:
ABUS Delivery Date:
Comments: _______________________________________________________________________________________________________
_____-_____-_____
_____-_____-_____
PO Number:
Applications Date:
___________________
_____-_____-_____
SO Number: ___________________
ABUS “Go-Live” Date:
_____-_____-_____
SITE LAYOUT & MINIMUM REQUIREMENTS
Does the ABUS scan room meet the minimum size requirements of 96 Inches X 120 Inches?
ABUS Scan Station
Cat 5 LAN Connection: ABUS Workstation Table Size: (30”W X 24”D Recommended)
Comments regarding the size and layout of the ABUS exam room, ABUS reading room and/or ABUS Workstation Setup:
_________________________________________________________________________________________________________________
Direct (wall plate) Switch or Hub
_____
X
_____ In
Placement of CPU:
ABUS Workstation Cat 5 LAN Connection:
On Table with Monitor Under Table N/A (iMac) Other
Yes No, size: _____
Direct (wall plate) Switch or Hub
X
_____ Ft
NETWORKING INFORMATION
The ABUS System (Scan Station + Workstation) must be networked to operate outside of any domain:
If multiple Scan Stations and/or Workst ations will be networked, please supply additional information on the next page.
AE Title: _____________________
IP Address: _____._____._____._____ _____._____._____._____ Workstation DNS: _____._____._____._____
Subnet Mask: _____._____._____._____ _____._____._____._____ Workstation Alt DNS: _____._____._____._____
Gateway: _____._____._____._____ _____._____._____._____ Workstation Internet Access: Yes No
SCAN STATION 1
Will the facility’s mail server be accessible to the Scan Station for sending system lo gs for remote diagnostics? Yes No
If Yes, Host Name: _____________________ Port: _____ Auth Type: None SMTP POP3 DNS: _____._____._____._____
WORKSTATION 1
Network
OK
Speed:
_____________________ Workstation Port: _____________________
1G 100BaseT
Duplex Settings:
Half Full Auto
3 fo 2 egaP laitnedifnoC 3102/30/50 :etaD 30veR 10-4100-0009 :N/P
Figure 2-1. Site Survey
2-10 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 63
Facility needs
_
_
_
Pre-Installation Site Survey Worksheet
DICOM MODALITY WORKLIST
Does your facility intend to utilize a DICOM modality worklist to supply patient appointment data to ABUS? If yes, please provide the DICOM modality worklist information:
Comments: _______________________________________________________________________________________________________
AE Title: _____________________ Port: ______ IP Address: _____._____._____._____
Yes No
If Yes, where does your DICOM worklist reside?
HIS/RIS PACS Other
STORAGE of PATIENT IMAGE DATA
The 3D image data for a typical bilateral ABUS
exam (6 views) is approximately 600MB. How
will patient image data be managed long-term?
PACS STORAGE MODEL: __________________ OTHER STORAGE MODEL: _________________
AE Title: ______________________________ Port: ______ AE Title: ______________________________ Port: ______
IP Address: ____.____.____.____ Gateway: ____.____.____.____ IP Address: ____.____.____.____ Gateway: ____.____.____.____
Store Print Query/Retrieve Store Print Query/Retrieve
Please Note: The ABUS workstation does not accept compre ssed image data les. So, while ABUS 3D image data may be
compressed when it is pushed to PACS or an alternative NAS, it must be uncompressed before it is pulled back to the workstation . How many complete 3D ABUS exams do you plan to push to storage during a typical week?
If Yes, please describe: ______________________________________________________________________________________________
3D ABUS exams will be stored (600MB) Only select 2D images will be stored Storage on PACS Storage on facility’s network attached storage (NAS) No image data will be stored Other _________________________________
Do you plan to utilize a secondary PACS or other
______
long-term archival system for 3D ABUS exams?
Yes No
ABUS DELIVERY INSTRUCTIONS
Is a loading dock available at the facility? Yes No If Yes, what are the hours?
Will a delivery truck with a lift gate be required? Yes No How many drivers will be needed for unloading? One Two
What will happen to the ABUS crate after delivery? Return to U-Systems Disposal by Customer Storage by Customer
Site specied delivery time: _____ AM PM Special notes: _______________________________________________________
____ AM to _____ PM M-F Other____
ADDITIONAL SCAN STATIONS and/or WORKSTATIONS
Will multiple scan stations an d/or workstations be installed?
Scan Station 2 Room Number:
Scan Station 2 Location Relative to Scan Station 1:
Workstation 2 Room Number:
Workstation 2 Location Relative to Workstation 1:
AE Title: _____________________
IP Address: _____._____._____._____ _____._____._____._____ Workstation DNS: _____._____._____._____
Subnet Mask: _____._____._____._____ _____._____._____._____ Workstation Alt DNS: _____._____._____._____
Gateway: _____._____._____._____ _____._____._____._____ Workstation Internet Access: Yes No
SCAN STATION 2
SIGN OFF: Facility Representative Name: ______________________ Date: ___________ Signature: __________________________
______
______
Room Size:
Same Building Same Department Same Floor Same Room Off Site
Table Size:
Same Building Same Department Same Floor Same Room Off Site
_____________________ Workstation Port: _____________________
WORKSTATION 2
No or Yes and Now / Later
_____
_____
X
_____ Ft
X
_____ In
Room Usage:
Room Usage:
If Yes & Now, please supply the information below for each item
ABUS Only Other Procedures
Multiple Readers Single Reader
3 fo 3 egaP laitnedifnoC 3102/30/50 :etaD 30veR 10-4100-0009 :N/P
Figure 2-2. Site Survey
Invenia ABUS – System Setup and Basic Service Manual 2-11 4700-0021-00 Rev. 6
Page 64
Site Preparations
Time and manpower requirements
Site preparation takes time. Begin site preparation checks as soon as possible, if possible, six weeks before delivery, to allow enough time to make any changes.
Facility needs
The following are required:
Dedicated single branch power outlet of adequate amperage, meeting all local and national codes, which is located less than 2.5 m (8 ft.) from the Ultrasound system’s proposed location
Door opening is at least 76 cm (30 in) wide
Proposed location for Ultrasound system is at least 0.5 m (1.5 ft.) from the wall for cooling
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with peripheral within 1 m of the Ultrasound system to connect cables.
Power outlets for other medical equipment
Power outlets for test equipment within 1 m (3.2 ft.) of Ultrasound system
Material to safely clean probes (done with a plastic container, never metal)
The following are desired:
Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, for Single Use Stabilization Membrane
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Storage area for Single Use Stabilization Membranes and Ultrasound Coupling Lotion
2-12 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 65
Facility needs
Important Notices
General Electric, Affiliate, or Distributor Field Engineers and Application Specialists will setup the system.
NOTICE This medical equipment is approved, in terms of the prevention
of radio wave interference, to be used in hospitals, clinics and other institutions which are environmentally qualified. The use of this equipment in an inappropriate environment may cause some electronic interference to radios and televisions around the equipment.
Ensure that the following is provided for the new system:
A separate power outlet with a minimum of a minimum 10 amp circuit breaker for 100-240 VAC.
Take precautions to ensure that the console is protected from electromagnetic interference.
Precautions include:
Operate the console at least 5 meters (15 feet) away
from motors, typewriters, elevators, and other sources of strong electromagnetic radiation (non-medical grade UPS must be at least 2 meters (6 feet) away from console).
Operation in an enclosed area (wood, plaster or
concrete walls, floors and ceilings) helps prevent electromagnetic interference.
Special shielding may be required if the console is to be
operated in the vicinity of radio broadcast equipment.
Invenia ABUS – System Setup and Basic Service Manual 2-13 4700-0021-00 Rev. 6
Page 66
Site Preparations
Important Notices (continued)
WARNING
CAUTION
CAUTION
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
To avoid risk of fire, the system power must be supplied from a separate, properly rated outlet.
Under no circumstances should the AC power plug be altered, changed, or adapted to a configuration rated less than specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a “hospital grade” or “hospital only” grounded power outlet.
Use caution to ensure that the power cable does not disconnect during system use.
If the system is accidentally unplugged, data may be lost.
To avoid leakage current above safety limits as prescribed by IEC 60601-1 and to ensure continuity of protective earth. DO NOT connect Invenia ABUS Scan Station or Workstation and mains-operated accessories to a single or multiple socket extension cord or power strip.
The Invenia ABUS requires a dedicated power and ground for the proper operation of its Ultrasound equipment. A separate power outlet with a minimum 10 amp circuit breaker for 100-240 VAC.
2-14 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 67
Invenia ABUS Scan Station
Facility needs
Figure 2-3. Typical Room Layout for the Invenia ABUS Scan Station
Invenia ABUS – System Setup and Basic Service Manual 2-15 4700-0021-00 Rev. 6
Page 68
Site Preparations
Invenia ABUS Scan Station (continued)
Figure 2-4. Desired Footprint
1. Maximum arm reach is ~1 m, or 39”
2. ~0.76 m, or 30” clear space recommended
3. ~0.61 m, or 24” clear space recommended
2-16 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 69
Invenia ABUS Scan Station (continued)
Facility needs
Figure 2-5. Invenia ABUS Scan Station
Invenia ABUS – System Setup and Basic Service Manual 2-17 4700-0021-00 Rev. 6
Page 70
Site Preparations
Invenia ABUS Scan Station System Specifications
System Features
Multi-slice B-Mode image acquisition with frame-by-frame 3D position registration
User-customizable workflow protocols
Image Processing Algorithms
Tissue Equalization Algorithm (TEA)
• Speckle Reduction
Nipple Shadow Compensation (NSC)
• Breast Border Detection
Chest Wall Detection
• Acquisition Scout Images
Transverse Plane (Real-time)
Coronal Plane (Static, for Nipple Placement)
DICOM 3.0 Compliant
• Worklist (DMWL)
Modality Performed Procedure Step (MPPS)
• Store (SCU)
• 10/100/1000 Base-T Ethernet
• User-Replaceable Fan Filter
Remote Service Diagnostics
Embedded User Manual
2-18 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 71
Invenia ABUS Scan Station System Specifications (continued)
Scanner Assembly Features
• Compression Assist
• Increase/Decrease Compression
Start Scan (one-touch volume acquisition)
• Abort Scan
Ergonomic Arm with 3 Operator-selectable Levels of Compression between 5-20 lbs (2.25-9.7 kg)
• Integrated, extra-wide transducer:
• C15-6XW Reverse Curve
Frequency Range: 6-15 MHz
Aperture Length: 15.3 cm
Transducer Travel Distance: 16.9 cm
Number of Elements: 768
Element Pitch: 0.20 mm
• Out-of-plane Aperture: 3.5 mm
Transducer Bandwidth: 85%
Imaging Depth: Up to 5.0 cm
Facility needs
Single-Use Stabilization Membrane
Single Volume Acquisition in less than 60 Seconds
Multi-Row LED Task Lighting
Removeable Scanner View Window, for easier clean-up
Invenia ABUS – System Setup and Basic Service Manual 2-19 4700-0021-00 Rev. 6
Page 72
Site Preparations
Invenia ABUS Scan Station System Specifications (continued)
Specifications
Width: 23 inches (59 cm)
Length: 26 inches (66 cm)
• Height (Arm):
Lowest Position 65 inches (165 cm)
Highest Position 85 inches (216 cm)
Footprint: 4.2 sq. ft. (0.39 sq. m)
Weight: 271 lbs (123 kg)
Articulating Arm Reach (max): 39 inches (100 cm)
Main Voltage: 100-240 VAC
Maximum Current: 10 amps
Mains Frequency: 50/60 Hz
HDD Data Capacity: 200 GB (approximately 200 Studies), with Storage Status Indicator
Study Transfer: Auto and Manual
Display: 17” High Resolution LCD Touch Screen
Cart: Mobile platform with 4-wheel steering and braking
Operating System: MS Windows 7 Professional, 64 bit
2-20 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 73
Invenia ABUS Review Software
Facility needs
Figure 2-6. Invenia ABUS Review Software
Invenia ABUS – System Setup and Basic Service Manual 2-21 4700-0021-00 Rev. 6
Page 74
Site Preparations
Invenia ABUS Workstation System Specifications
NOTE: Workstation Specifications are only applicable for the Invenia
ABUS Workstation Option.
Image Display and Navigation
Individual user-defined Hanging and Viewing Protocols
Multi-Slice Viewing: 4-12 Images
Streamlined Review Protocols
• Standardized View Orientations: Thick-slice Coronal, Transverse, Sagittal, Radial, Anti-radial, and ROI Views
Variable Slice Thickness: 0.5 to 10.00 mm (increments of
0.5 mm)
Grayscale Windowing and Leveling
Pan and Zoom Feature
• 4X Magnifying Tool
CINE Review with Adjustable Playback Speed
360 Degree Any Point of Compass Display
Side-by-Side Display for Coronal View Comparison
Right-Click Graphical Navigational Tools
Networking/Storage/Archiving
• User Login Authentication
DICOM 3.0 Compliant:
• Query/Retrieval SCU
• Store SCP/SCU
• Print SCU
• 10/100/1000 Base-T Ethernet
Compatible Media Storage
USB Media Stick
DVD +/- RW
2-22 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 75
Invenia ABUS Workstation System Specifications (continued)
Annotation and Reporting
3-Dimensional Position Referencing By:
• Clock Position
Distance from Nipploe
Depth from Skin Surface
• Measurement Toolbox
Standardized Labeling for Scan Positions
Graphic Labels for Scan Position and Clock Position
Customizable auto Report Function with User-Selectable Images
• Patient Study Browser
Patient List Filtering Based on User’s Setting
Hardware Specifications
Tower: Workstation Class Hardware
Operating System: Windows 7 Professional, 64 bit
Display: 24-inch (diagonal) Display
Facility needs
Invenia ABUS – System Setup and Basic Service Manual 2-23 4700-0021-00 Rev. 6
Page 76
Site Preparations
Workstation Hardware Specifications for ABUS Review Software
Operating System (O.S.)
Windows 7 Professional 64 Bit
Monitor
Minimum Requirements:
Dedicated 24" Color Monitor
• Minimum resolution 1920x1200
Recommended Monitor Specs:
24” or greater diagonal size
1920 x 1200 resolution or higher
LED backlit, Color Panel
Brightness 250 cd/m or greater
Contrast Ratio 1000:1 or higher
Input Devices
• Standard Keyboard
Mouse with 3 Buttons and Scroll Wheel with individually distinct detents
Available Disk Space
Video Card
NOTE: On-board GPU (e.g Intel HD graphics) is not supported.
Hard Drive with 400MB reserved for application and separate Hard Drive/disk space for Exams (1GB per exam, recommend 300 exams cache)
Minimum requirements at installation:
300 GB of available space
Support for OpenGL 3.3 or Higher
2 GB GDDR5 Dedicated Video Memory or more
Memory Bandwidth 100GB/s or higher
Support Resolution of at least 1920 x 1200 at 60Hz
2-24 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 77
Facility needs
Workstation Hardware Specifications for ABUS Review Software (continued)
Processor, Memory
Xeon E5-1620 3.6GHz or better
8.0GB RAM or better
Minimum requirements at installation:
8.0 GB of RAM
Database
The Invenia ABUS software uses MySQL database. The database is installed and configure automatically during the installation process.
No other instances of MySQL database are supported on the same computer.
Networking/Storage/Archiving
• 10/100/1000 Base-T Ethernet
Compatible Media Storage
USB Media Stick
DVD +/- RW
I/O ports
One or more USB 2.0 port
Invenia ABUS – System Setup and Basic Service Manual 2-25 4700-0021-00 Rev. 6
Page 78
Site Preparations
Environmental Requirements
The system should be operated, stored, or transported within the parameters outlined below. Either its operational environment must be constantly maintained or the unit must be turned off.
NOTE: You may get an overheating message with regard to fan speed.
Ensure adequate system/room ventilation.
Table 2-4: System Environmental Requirements
Operational Storage Transport (<16hrs.)
Temperature 5° - 35°C
41° - 95°F
Humidity 30% to 75%
non-condensing
Altitude 0-3,000 m (9,800 feet) 0-5,000 m (16,400 feet) 0-5,000 m (16,400 feet)
-10° - 60°C 14° - 140°F
10% to 85% non-condensing
-10° - 60°C 14° - 140°F
10% to 85% non-condensing
Packaging Information
The shipping crate size is 55”L x 34”W x 73”H (1.4m L x 0.86m W x 1.85m H).
2-26 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 79

Networking setup requirements

Networking setup requirements
Scanner connected to hospital’s network
Supported networks:
10/100/1000 Base-T Ethernet
InSite Requirements
InSite ExC allows GE to remotely assist users, Biomedical Technicians, and Field Service personnel by using a secure remote connection across the network.
Internet access is needed to be able to connect to InSite ExC.
The following is needed to configure InSite:
• System IP Address
• Default Gateway
Proxy Server, if necessary, and Port
• Proxy Authentication, if necessary
User Name and Password
System ID (SID) Number
Purpose of the DICOM network function
DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
As an added benefit, transferring images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while scanning continues.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
Invenia ABUS – System Setup and Basic Service Manual 2-27 4700-0021-00 Rev. 6
Page 80
Site Preparations
DICOM setup requirements
To configure the Ultrasound system to work with other network connections, the site’s network administrator must provide information to complete the form “Worksheet for DICOM Network Information”. Ensure that there are no spaces in any field of the form.
Entries must include:
A host name, local port number, AE Title, IP address and Net Mask for the Ultrasound system.
The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
The host name, IP address, port and AE Title for each device the site wants connected to the Ultrasound system for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the revision of the device, is also included. This information may be useful for error solving.
Refer to the Site Planning worksheet on the Figure 2-1 on page 2-10.
2-28 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 81

Environmental Dangers

Environmental Dangers
Commercial devices such as laser cameras, printers, VCRs and external monitors, usually exceed allowable leakage current limits and, when plugged into separate AC outlets, are in violation of patient safety standards. Suitable electrical isolation of such external AC outlets, or providing the device with extra protective earth, will be required in order to meet UL60601-1 and IEC60601-1 / IEC60601-1-1 standards for electrical leakage.
Invenia ABUS – System Setup and Basic Service Manual 2-29 4700-0021-00 Rev. 6
Page 82
Site Preparations
Patient Vicinity UL60601-1 (USA)
NOTE: This section is only applicable if the Workstation is provided by
GE (Invenia ABUS Option).
2.12.20DV (UL60601-1:2003)
In area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to be contacted by the patient or attendant who can touch the patient. This encloses a space within the room 1.83 m (6 ft.) beyond the perimeter of the bed (examination table, dental chair, treatment booth, and the like) in its intended location, and extending vertically 2.29 m (7.5 ft.) above the floor.
1. Patient environment
2-30 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 83
Chapter 3

System Setup

This chapter contains information needed to install Invenia ABUS.
Included is a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to set up the system and how to check and test the unit, probes, and external peripherals for electrical safety are also included in this procedure.
Invenia ABUS – System Setup and Basic Service Manual 3-1 4700-0021-00 Rev. 6
Page 84
System Setup
Average setup time
Setup warnings

Setup reminders

Unpacking the Invenia ABUS: 0.5 hour
Set up Scan Station and Workstation: 3 hours or more, depending on the configuration
DICOM Network Configuration: 3 hours or more, depending on the configuration
Install InSite: 0.5 hour
DANGER
CAUTION
CAUTION
CAUTION
CAUTION
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE ULTRASOUND SYSTEM!
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a three to two prong adapter. This defeats safety grounding.
Do not wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present.
Do not operate this unit unless all board covers and frame panels are securely in place. System performance and cooling require this.
The system must be supplied by an adequately rated electrical circuit. The capacity of the supply circuit must be as specified.
3-2 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 85
Setup warnings (continued)
Setup reminders
CAUTION
CAUTION
If the Ultrasound system is very cold or hot, do not turn on its power until it has had a chance to acclimate to its operating environment.
The following table describes guidelines for reaching operational temperatures from storage or transport temperatures.
Operator Manual(s)
The User Manual(s) should be fully read and understood before operating the Invenia ABUS and kept near the Ultrasound system for quick reference.
Invenia ABUS – System Setup and Basic Service Manual 3-3 4700-0021-00 Rev. 6
Page 86
System Setup
Receiving and unpacking the
Warnings for receiving and unpacking
equipment
CAUTION
CAUTION
Attempts to move the Ultrasound system considerable distances or on an incline by one person could result in injury or damage or both.
Two people are required whenever a part weighing 16 KG (35 LBS) or more must be lifted.
Remember to use relevant personal protecting equipment (PPE) during packing and unpacking. Check with your local EHS representative.
3-4 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 87
The Tilt & Shock indicators
Overview
Improper handling during transportation may harm the equipment inside the package even if the package itself is undamaged.
To make it easier to detect if the handling during transportation has been improper, a set of Tilt & Shock indicators have been attached to the transportation box.
Table 3-1: Shock and Tilt Watch
Description Illustration
ShockWatch
Receiving and unpacking the equipment
Tilt Watch
Invenia ABUS – System Setup and Basic Service Manual 3-5 4700-0021-00 Rev. 6
Page 88
System Setup
Receiving the Invenia ABUS
Overview
Improper handling during transportation may harm the equipment inside the package even if the package itself is undamaged.
Examine all packages
Examine package closely at time of delivery, as described in the procedure below.
Table 3-2: Examine all packages
Step Task Illustrations
1. Is damage apparent?
• If YES; continue with the instructions in ‘Damage in transportation’ on page 3-10.
• If NO; continue with the next step.
2. Is the Shock Indicator red colored inside the middle of the indicator?
• If YES: The Shock Indicator has been
activated. Continue with the instructions in ‘Damage in transportation’ on page 3-10, then continue with the next step.
• If NO: continue with the next step.
1 - Red Color
3-6 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 89
Receiving and unpacking the equipment
Table 3-2: Examine all packages (Continued)
Step Task Illustrations
3. Is the Tilt Indicator red colored inside the middle of the indicator?
• If YES: The Tilt Indicator has been
activated. Continue with the instructions in ‘Damage in transportation’ on page 3-10 before you continue with the next step.
• If NO: continue with the next step.
4. Continue with the instructions in ‘Unpacking the Invenia ABUS’ on page 3-11.
1 - Red Color
Invenia ABUS – System Setup and Basic Service Manual 3-7 4700-0021-00 Rev. 6
Page 90
System Setup
Position of the Tilt & Shock indicators
The Tilt & Shock indicators have been attached to the transportation crate.
NOTE: Before cutting the straps, check Shock and Tilt Tags to make
sure they have not been triggered. If damaged, report it to the carrier. If not, then cut the straps around the crate.
If Shock Indicator has triggered or is missing
The purpose of a shock indicator label is NOT to tell if a product has been damaged during shipment. The purpose of these labels are to alert people handling a package that the product contained is very sensitive to shock damage and that it should be handled carefully. It is basically an active "Fragile" label that turns red if a predetermined shock does occur. Because the labels can receive false activation due to an impact shock directly on the label, an activated label must not be interpreted to mean product damage. It simply means that the receiver should note on the shipping papers at the time of receipt that the label was activated and the product should be inspected for possible concealed damage. Conversely, a high level, product-damaging shock could occur to the package in a way that does not activate the label, so a non activated label does not insure that the product is not damaged. Some degree of inspection is still required.
Table 3-3: Shock Indicator has triggered or is missing
Step Tas k
1. If the Shock Indicator is missing: Note on the shipping papers at the time of receipt that the Shock Indicator label is missing.
If the Shock Indicator has triggered:
Note on the shipping papers at the time of receipt that the Shock Indicator label was activated.
2. Inspect the product for possible concealed damage.
3-8 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 91
Receiving and unpacking the equipment
If Tilt Indicator has triggered or is missing
The purpose of the tilt indicator label is to alert people handling a product that it is sensitive to tipping and it must remain upright at all times. It is basically an active "Up Arrow" that changes color if the package is tipped 89 degrees or more from horizontal. These labels can be false activated if tipped less than 89 degrees, and shocked or vibrated at the same time. This event does occur, but is considered uncommon. If a package is received with an activated tilt indicator label, there is high degree of certainty it tipped 89 degrees or more from horizontal during shipment.
An activated tilt indicator label does not indicate if the package was simply “Tipped” (laid down with no impact shock) or “Tipped Over” (free fall, with an impact shock). Using both shock indicator labels and tilt indicator labels will help identify if a Tip Over impact shock occurred.
Table 3-4: Tilt Indicator has triggered or is missing
Step Tas k
1. If the Tilt Indicator is missing:
2. Inspect the product for possible concealed damage.
Note on the shipping papers at the time of receipt that the Tilt Indicator label is missing.
If the Tilt Indicator has triggered:
Note on the shipping papers at the time of receipt that the Tilt Indicator label was activated.
Invenia ABUS – System Setup and Basic Service Manual 3-9 4700-0021-00 Rev. 6
Page 92
System Setup
Damage in transportation
Follow this procedure if damage is apparent:
1. Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for“ by a GE representative or hospital receiving agent.
2. Report the damage to the carrier.
Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier.
• A transportation company will not pay a claim for
damage if an inspection is not requested within this 14 day period.
Invenia ABUS transportation box label
The Invenia ABUS transportation crate label is located at the front of the transportation crate.
Table 3-5: transportation box label
Symbols Definition/Comments
1. TOP,. UPRIGHT - Transportation and Storage
2. FRAGILE, Handle with Care
3. RECYCLING
4. KEEP DRY (protect from moisture)
1
3
2
4
3-10 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 93
Receiving and unpacking the equipment
Unpacking the Invenia ABUS
Instructions are provided in the materials attached to the Shipping Crate.
Table 3-6: Unpacking the Invenia ABUS Scan Station and Workstation
Steps Illustration
1. Carefully remove any shipping straps that secure the front door/ramp and back door.
2. Lift the hasp handles away from the crate.
Repeat for the step for the other three hasps.
3. Rotate the hasp handles counterclockwise to release.
Invenia ABUS – System Setup and Basic Service Manual 3-11 4700-0021-00 Rev. 6
Page 94
System Setup
Table 3-6: Unpacking the Invenia ABUS Scan Station and Workstation
Steps Illustration
4. Slowly lower the front door/ramp until it rests on the floor.
5. Remove any loose packaging, if present.
3-12 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 95
Table 3-6: Unpacking the Invenia ABUS Scan Station and Workstation
Steps Illustration
6. Lift the front blocking plate up and out of the crate.
7. If the wheel locks are locked, release the brakes.
Receiving and unpacking the equipment
Slowly roll/remove the Invenia ABUS Scan Station straight out of the crate.
8. Repeat the steps for opening the hasps, on the crate back door.
Invenia ABUS – System Setup and Basic Service Manual 3-13 4700-0021-00 Rev. 6
Page 96
System Setup
Table 3-6: Unpacking the Invenia ABUS Scan Station and Workstation
Steps Illustration
9. Slowly swing the door open and remove the Workstation (if ordered) and any other supplies.
10. Remove all of the packaging and place the removed packaging back into the crate to store for possible future use.
3-14 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 97
Packing materials

Packing materials - recycling information

Packing materials -
recycling information
The packing materials for Invenia ABUS are recyclable:
The Transportation Box is made of spruce or similar material. (“PHYTOSANITARY CERTIFICATE” included in all shipments to The People's Republic of China.)
Lever lockings (hinges) are made of zinc plated steel.
The inner reinforcements are made of Ethafoam (Polyethylene foam).
The plastic foil is made of LDPE (Low Density Polyethylene).
Invenia ABUS – System Setup and Basic Service Manual 3-15 4700-0021-00 Rev. 6
Page 98
System Setup
Verify customer order
Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered, or damaged.
Physical inspection
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to ‘Damage in transportation’ on page i-11 in the beginning of this manual.

Preparing for setup

EMI protection
The Invenia ABUS has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation.
See ‘EMI limitations’ on page 2-5 for more information about EMI protection.
The Invenia ABUS has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation.
3-16 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Page 99
Product Labels and Descriptions
Preparing for setup
Figure 3-1. Product Labels
Invenia ABUS – System Setup and Basic Service Manual 3-17 4700-0021-00 Rev. 6
Page 100
System Setup
Product Labels and Descriptions (continued)
Table 3-7: Rear Panel Label Explanations
1. “ATTENTION” - Consult accompanying documents” is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label.
2. TUV Rhineland Label: NRTL Listing and Certification Mark is used to designate conformance to nationally recognized product safety standards. The Mark bears the name and/or logo of the testing laboratory, product category, safety standard to which conformity is assessed, and a control number.
3. The CE Mark of Conformity indicates this equipment conforms with the Council Directive 93/42/EEC.
4. Prescription Device (For U.S.A. Only)
5. BF Applied Part
6. Identification and Rating Plate–USA/Asia 120V console
7. CISPR Emissions Label
8. System and Probe IP Code
9. Mercury Warning Label
10. WEEE Symbol
11. Possible shock hazard. Do not remove covers or panels. No user serviceable parts are inside. Refer servicing to qualified service personnel.
12. The equipment weighs approximately 123 kg (271 lbs). To avoid possible injury and equipment damage when transporting from one area of use to another:
• Be sure the pathway is clear.
• Limit movement to a slow careful walk.
• Use two or more persons to move the
equipment on inclines or long distance.
13. Fuse Label
14. Invenia ABUS Scan Station (A), Workstation Rating Plate (B) and Invenia ABUS Review Software CD Label (C)
• Manufacturer’s name and address
• Part Number
• Model/Serial Number/Date of Manufacturer
3-18 Invenia ABUS – System Setup and Basic Service Manual
4700-0021-00 Rev. 6
Loading...